a p p l i c at i o n N o t e

Atomic Absorption
Author
Praveen Sarojam, Ph.D.
PerkinElmer, Inc.
Shelton, CT 06484 USA

Analysis of Pb, Cd and
As in Spice Mixtures
using Graphite Furnace
Atomic Absorption
Spectrophotometry

Introduction
The toxicity and effect of trace heavy metals on
human health and the environment has attracted
considerable attention and concern in recent years.
With an inherent toxicity, a tendency to accumulate
in the food chain and a particularly low removal rate
through excretion,1 lead (Pb), cadmium (Cd) and
arsenic (As) cause harm to humans even at low
concentrations. Exposure to trace and heavy metals
above the permissible level affects human health and
may result in teratogenicity (reproductive effects).
Individuals may also experience high blood pressure,
fatigue, as well as kidney and neurological disorders.

Spices, the dried parts of plants, grow widely in various regions of the world, are produced
either on small farmlands or naturally grown, and have been used for several purposes since
ancient times. Most are fragrant and flavorful and are used for culinary purposes to improve
the quality of food.2 Natural food spices, such as pepper, have been reported to contain
significant quantities of some heavy metals, including Pb, Cd and As. Contamination with
heavy metals may be accidental (e.g. contamination of the environment during plant
cultivation) or deliberate – in some cultures, according to traditional belief, specially treated
heavy metals are associated with health benefits and are thus an intentional ingredient of
traditional remedies. Spices and herbal plants may contain heavy metal ions over a wide
range of concentrations.3,4 There is often little information available about the safety of
those plants and their products in respect to heavy metal contamination. Due to the
significant amount of spices consumed, it is important to know the toxic metal
concentrations in them.5
The major challenge in performing spice analyses is the
extremely low analyte concentrations combined with high
matrix levels. Graphite furnace atomic absorption spectrophotometry (GFAAS) has been one of the more reliable
techniques used for many years and is therefore the preferred analytical method. The use of longitudinal Zeeman
background correction and the use of matrix modifiers
provides accurate results and low detection limits in samples
with high matrices. This makes GFAAS an indispensible tool
in carrying out such analyses.
The objective of this work is two-fold: (1) to use GFAAS to
accurately analyze the levels of Pb, Cd, and As present in
some major spices commonly available on the market; and
(2) to cross reference these measured levels to the recommended limits specified by the U.S. FDA.

Experimental Conditions
Instrumentation
The measurements were performed using a PerkinElmer®
PinAAcle™ 900T atomic absorption spectrophotometer
(Shelton, CT, USA) equipped with the intuitive WinLab32™
for AA software running under Microsoft® Windows™ 7,
which features all the tools to analyze samples, report and
archive data and ensure regulatory compliance (Figure 1).
The high-efficiency optical system and solid-state detector
used in the PinAAcle 900T spectrometer provide outstanding
signal-to-noise ratios. The longitudinal Zeeman-effect background correction for graphite furnace analysis provides
accurate background correction without the loss of light
that usually occurs in transverse Zeeman systems. The use
of a transversely heated graphite atomizer (THGA) provides
uniform temperature distribution across the entire length of
the graphite tube, eliminating memory effects and potential
interferences that may occur with high-matrix sample analysis.
Pyrolytically coated THGA tubes with end caps (Part No.
B3000655) were used for all measurements. The optimized
instrumental conditions for furnace experiments are given
in Table 1 (Page 3), and the graphite furnace temperature
programs are listed in Appendix I (Page 5). A heated injection
was used for lead; it can also be used for cadmium and arsenic.

Figure 1. PerkinElmer PinAAcle 900T atomic absorption spectrophotometer.

2

A microwave sample preparation system was used for the
microwave-assisted digestion. This is an industrial-type
microwave oven, which is equipped with various accessories
to optimize the sample digestion. The samples were digested
using ten 100 mL high-pressure vessels made of PTFE. The
microwave digestion program is listed in Table 2 (Page 3).
Standards, Chemicals and Certified Reference Materials
PerkinElmer Pure single-element calibration standards for
Pb, Cd, and As were used as the stock standards for preparing
the working standards (Part Nos. Pb: N9300128; Cd: N9300107;
As: N9300102). Working standards were prepared by serial
volume/volume (v/v) dilution in polypropylene vials (Part
Nos. B0193233 15 mL Conical; B0193234 50 mL Conical
Freestanding) ASTM® Type I deionized water (Millipore®
Corporation, Billerica, Massachusetts, U.S.) acidified with
0.2% nitric acid (HNO3) (Tamapure®, TAMA Chemicals,
Japan) was used as the calibration blank and for all dilutions.
30% hydrogen peroxide (H2O2) (Kanto Chemicals, Tokyo,
Japan) was used for digestion along with nitric acid.
Matrix modifiers were prepared from 10% NH4H2PO4 (Part
No. N9303445), 1% Mg as Mg(NO3)2 (Part No. B0190634)
and 1% Pd (Part No. B0190635) stock solutions, by diluting
with the 0.2% HNO3 made above. Matrix modifiers were
added automatically to each standard, blank and sample by
the AS 900 autosampler, an integral part of the PinAAcle
900T spectrometer.
NIST® 1568a Certified Reference Material (CRM) for Trace
Metals in Rice Flour was used to validate the method.
Quality control (QC) check standards were prepared at the
calibration curve midpoint concentration for each individual
element.
Sample and Certified Reference Material Preparation
Plastic bottles were cleaned by soaking with 10% v/v HNO3
for at least 24 hours and rinsed abundantly in deionized
water before use. The polypropylene autosampler cups
(Part No. B3001566) were soaked in 20% nitric acid overnight to minimize sample contamination, and thoroughly
rinsed with 0.5% HNO3 acid before use. Five-point calibration
curves (four standards and one blank) were constructed for
each analyte and the calibration curve correlation coefficient
was examined to ensure a value better than 0.998 before
the start of the sample analysis (Appendix II – Page 6).
Five branded powdered spice samples available in local supermarkets in India (coriander
powder, ginger powder, asafetida, black pepper powder and red chili powder) were analyzed.
Approximately 0.5 g of each sample or CRM, accurately weighed in duplicate, was transferred to the vessel of the microwave digestion system and the sample digestion method
(Table 2) was performed in accordance with U.S. Environmental Protection Agency (EPA)
Method 3052. The digested samples were diluted with 0.2% HNO3 and brought up to
25 mL in polypropylene vials.
Table 1. Optimized experimental conditions of PinAAcle 900T.
Analyte	Pb	

Cd	

As

Wavelength (nm)	

283.3	

228.8	

193.7

Lamp Type	

EDL 	

HCL	

EDL

	

(Part No. N3050657)	

(Part No. N3050115)	

(Part No. N3050605)

Standards (µg/L)	

5, 10, 15, 20	

0.5, 0.75, 1.0, 2.0	

10, 20, 30, 40

Correlation Coefficient	 0.9991	

0.9996	

0.9989

Read Time (sec)	

3	

5	

3

Injection Temp (˚C)	

90	

20	

20

0.05 mg NH4H2PO4	
0.005 mg Pd
Matrix Modifier (mg)	 0.05 mg NH4H2PO4 	
	
and 0.003 mg	
and 0.003 mg	
and 0.003 mg
	Mg(NO3)2	Mg(NO3)2	Mg(NO3)2
Modifier Volume (µL)	

5

Slit Width (nm)	

0.7

Sample Volume (µL)	

20

Measurement Type	

Peak Area

Calibration	

Linear through zero

Table 2. Program used for the digestion of spices and herbs.
Sequence	
Power (watts)	

1	2
1000	

0

Ramp Time (min)	

10	

0

Hold Time (min)	

10	

20

Weight Taken (mg)	

~500

H2O2 (mL)	

1.0

HNO3 (mL)	

7.0

Temperature (˚C)	

180

Results and Discussions
In the analysis of complex samples with the graphite furnace, obtaining reproducible results
can be a challenging task as one has to deal with analytes present at the low µg/L ppb level
and the potential for matrix interferences. With Stabilized Temperature Platform Furnace (STPF)
conditions (including: pyrolytically coated graphite tubes, platform atomization, rapid
heating, internal gas stop during atomization, fast signal processing and Zeeman background correction) the analysis of low level analytes in spice mixtures is an easy task.
The graphite furnace sample introduction system is of paramount importance in optimizing
the short-term stability of signals. The PinAAcle 900T spectrometer uses a unique built-in
camera to monitor sampler tip alignment and sample introduction into the graphite tube.
With the TubeView™ furnace camera, it is simple to position the AS 900 autosampler sample
capillary tip at the correct depth inside the tube so as to achieve highly reproducible pipetting.

3
Table 3. Analysis of certified reference material by GFAA.
	NIST® 1568a Rice Flour
	
Certified Value	
Analyte	(µg/g)	

Measured Value
(µg/g)

Pb	<0.010	

0.0093

Cd	

0.022 ±0.002	

0.020 ±0.004

As	

0.29 ±0.03	

0.24 ±0.02

Table 4. Estimated method detection limits (MDLs).
Analyte	

MDL (µg/L)

Pb	9.5
Cd	2.35
As	9.5

Table 5. Spike recovery studies.
		
		
		
Analyte	Sample	

Post-		 Pre-	
digestion	Spike	 digestion	Spike
Recovery	 Level	 Recovery 	 Level
(%)	 (µg/L)	 (%)	 (µg/L)

Pb	
	

Coriander powder	
Ginger powder	

91	
5.00	
90		

111	
*

5.00

Cd	
	

Coriander powder	
Ginger powder	

104	
0.625	
96		

119	
*

1.00

As	
	

Coriander powder	
Ginger powder	

108	
8.33	
106		

100	
*

5.00

*Pre-digestion spike recovery not performed

The developed method has been validated by incorporating
various QC checks and by the analysis of a CRM. The agreement between the certified and measured values was good
(Table 3), which demonstrates the accuracy of the direct
calibration being used as well as the overall accuracy of the
developed method. Method detection limits (MDLs) were calculated (Table 4) based on the standard deviation of seven
replicates of the reagent blank (t-value = 3.14) and took
into account the dilution factor for the samples. The low
method detection limits obtained show the capability of the
PinAAcle 900T spectrometer in analyzing difficult matrices at
the measured concentrations.
Incomplete mineralization of samples during the microwavedigestion process, due to high organic content, may cause
difficulty in transferring analytes into solution. These impurities
may also affect the analytical measurements.6 Application
of concentrated HNO3 along with H2O2 for mineralization of
spices and herbs leads to the complete digestion of samples,
which is proven by the determination of the values of the
analytes in the CRM (Table 3). Both a pre-digestion and an
automatic post-digestion recovery study were performed.
The recoveries obtained for the pre-digestion spike easily
meet the U.S. EPA recovery guidelines of ±30% and indicate
that neither analyte loss nor contamination was of major
concern throughout the analytycal procedure (Table 5).
The recoveries obtained for the post-digestion spike indicate
that no matrix interference towards the analyte signal was
seen. The results of the sample analysis are summarized in
Table 6. The level of arsenic, cadmium and lead in all the
samples analyzed were well within the permissible limits of
10, 0.3 and 10 mg/kg respectively, as specified by the U.S. FDA.

Table 6. Analysis of spices and herbs samples using GFAAS (µg/kg).
	
Analyte	

U.S. FDA 						
Limit	
Replicate	 Coriander Powder	 Ginger Powder	 Asafetida	 Black Pepper Powder	

Red Chili Powder

Pb 	

10,000	

Sample	

455	

551	

124	

9.8	

286

Cd 	

300	

Sample	

48	

89	

8.2	

25	

57

		Duplicate	

47	

92	

8.2	

28	

59

As 	

20	

49	

41	

49	

<MDL

24	

42	

38	

32	

<MDL

10,000	

Sample	

		Duplicate	

4
Conclusions
A complete method for the accurate determination of Pb,
Cd and As in various spice mixtures using the PinAAcle 900T
atomic absorption spectrophotometer in the GFAAS mode
was developed. The results confirmed that the determination
of arsenic, cadmium and lead in spice mixtures, after acid
solubilization by microwave digestion, can be performed
by GFAAS without any interference. The PinAAcle 900Z
(Longitudinal Zeeman Furnace only) spectrometer can also
be used for this application.

References
1.	O. Sadeghi, N. Tavassoli, M.M. Amini, H. Ebrahimzadeh,
N. Daei, Food Chemistry 127 (2011) 364–368.

2.	H. Mubeen, I. Naeem, A. Taskeen and Z. Saddiqe, New
York Science Journal, 2 (5) (2009) 1554-0200.
3.	K. K Gupta, S. Bhattacharjee, S. Kar, S. Chakrabarty, P.
Thakar, G. Bhattacharyya and S.C. Srivastava, Comm. Soil
Plant Anal., 34 (2003) 681-693.
4.	T. M. Ansari, N. Ikram, M. Najam-ul-Haq, O. Fayyaz, I.
Ghafoor and N. Khalid, J. Biol. Sci., 4 (2004) 95-99.
5.	R.P. Choudhury and A.N. Garg, Food Chemistry, 104
(2007) 1454-1463.
6.	I. Baranowska, K. Srogi, A. Włochowicz, K. Szczepanik,
Polish Journal of Environmental Studies, 11( 5) (2002)
467-471.

Appendix I – Graphite Furnace Temperature Program
Table 7. Furnace temperature program for Pb.
Analyte	
Step	
Temp ˚C	
Ramp Time (sec)	
Hold Time (sec)	
Internal Gas Flow (mL/min)	
Pb	 1	110	
1	
30	
250	
	
2	130	
15	
30	
250	
	
3	850	
10	
20	
250	
	
4	
1600	
0	
5	
0	
	
5	2450	
1	
3	
250	

Gas Type
Argon
Argon
Argon
–
Argon

Table 8. Furnace temperature program for Cd.
Analyte	
Step	
Temp (˚C)	
Ramp Time (sec)	
Cd	 1	110	
10	
	
2	130	
15	
	
3	500	
15	
	
4	
1500	
0	
	
5	2450	
1	

Hold Time (sec)	
Internal Gas Flow (mL/min)	
30	
250	
30	
250	
35	
250	
3	
0	
3	
250	

Gas Type
Argon
Argon
Argon
–
Argon

Table 9. Furnace temperature program for As.
Analyte	
Step	
Temp (˚C)	
Ramp Time (sec)	
Hold Time (sec)	
Internal Gas Flow (mL/min)	
As	 1	110	
5	
30	
250	
	
2	130	
20	
30	
250	
	
3	800	
15	
40	
250	
	
4	1200	
15	
30	
250	
	
5	
2200	
0	
5	
0	
	
6	2450	
1	
3	
250	

Gas Type
Argon
Argon
Argon
Argon
–
Argon

5
Appendix II – Calibration Graphs and Linearity

Figure 2. Calibration curve for the determination of Pb using GFAAS.

Figure 3. Calibration curve for the determination of Cd using GFAAS.

Figure 4. Calibration curve for the determination of As using GFAAS.

PerkinElmer, Inc.
940 Winter Street
Waltham, MA 02451 USA	
P: (800) 762-4000 or
(+1) 203-925-4602
www.perkinelmer.com

For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs
Copyright ©2011, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners.
009754A_01

Weitere ähnliche Inhalte

Was ist angesagt?

BioInnovation Leader Summit Transforming Process Efficiency
 BioInnovation Leader Summit Transforming Process Efficiency BioInnovation Leader Summit Transforming Process Efficiency
BioInnovation Leader Summit Transforming Process EfficiencyGBX Summits
 
Analysis of Fish and Seafoods with AAnalyst 800 Atomic Absorption Spectrophot...
Analysis of Fish and Seafoods with AAnalyst 800 Atomic Absorption Spectrophot...Analysis of Fish and Seafoods with AAnalyst 800 Atomic Absorption Spectrophot...
Analysis of Fish and Seafoods with AAnalyst 800 Atomic Absorption Spectrophot...PerkinElmer, Inc.
 
Toxicity immobilization of refinery sludge containing heavy metals via vitrif...
Toxicity immobilization of refinery sludge containing heavy metals via vitrif...Toxicity immobilization of refinery sludge containing heavy metals via vitrif...
Toxicity immobilization of refinery sludge containing heavy metals via vitrif...eSAT Journals
 
PerkinElmer Application Note: Monitoring volatile organic compounds in beer p...
PerkinElmer Application Note: Monitoring volatile organic compounds in beer p...PerkinElmer Application Note: Monitoring volatile organic compounds in beer p...
PerkinElmer Application Note: Monitoring volatile organic compounds in beer p...PerkinElmer, Inc.
 
Direct determination of bismuth, indium and lead in sea water by zeeman etaas...
Direct determination of bismuth, indium and lead in sea water by zeeman etaas...Direct determination of bismuth, indium and lead in sea water by zeeman etaas...
Direct determination of bismuth, indium and lead in sea water by zeeman etaas...Jorge Alexander Rocha Alvarez
 
Distillery Wastewater Decontamination by the Fenton Advanced Oxidation Method
Distillery Wastewater Decontamination by the Fenton Advanced Oxidation MethodDistillery Wastewater Decontamination by the Fenton Advanced Oxidation Method
Distillery Wastewater Decontamination by the Fenton Advanced Oxidation MethodIJRES Journal
 
Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinEl...
Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinEl...Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinEl...
Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinEl...PerkinElmer, Inc.
 
New Ultra-High Capacity TOYOPEARL® Protein A Resin
New Ultra-High Capacity TOYOPEARL® Protein A Resin New Ultra-High Capacity TOYOPEARL® Protein A Resin
New Ultra-High Capacity TOYOPEARL® Protein A Resin Tosoh Bioscience GmbH
 
BIODEGRADATION OF GASOIL BY FUNGAL ISOLATES FROM PETROLEUM CONTAMINATED SOILS
BIODEGRADATION OF GASOIL BY FUNGAL ISOLATES FROM PETROLEUM CONTAMINATED SOILS BIODEGRADATION OF GASOIL BY FUNGAL ISOLATES FROM PETROLEUM CONTAMINATED SOILS
BIODEGRADATION OF GASOIL BY FUNGAL ISOLATES FROM PETROLEUM CONTAMINATED SOILS AL-kitab university -IRAQ
 
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OESAnalysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OESPerkinElmer, Inc.
 
Alkaline Extraction of Cobia (Rachycentroncanadum) Proteins: Physicochemical ...
Alkaline Extraction of Cobia (Rachycentroncanadum) Proteins: Physicochemical ...Alkaline Extraction of Cobia (Rachycentroncanadum) Proteins: Physicochemical ...
Alkaline Extraction of Cobia (Rachycentroncanadum) Proteins: Physicochemical ...IJERA Editor
 

Was ist angesagt? (17)

BioInnovation Leader Summit Transforming Process Efficiency
 BioInnovation Leader Summit Transforming Process Efficiency BioInnovation Leader Summit Transforming Process Efficiency
BioInnovation Leader Summit Transforming Process Efficiency
 
P368995
P368995P368995
P368995
 
Chromatography: Feedstock Characterization and Fermentation Monitoring of Bio...
Chromatography: Feedstock Characterization and Fermentation Monitoring of Bio...Chromatography: Feedstock Characterization and Fermentation Monitoring of Bio...
Chromatography: Feedstock Characterization and Fermentation Monitoring of Bio...
 
Analysis of Fish and Seafoods with AAnalyst 800 Atomic Absorption Spectrophot...
Analysis of Fish and Seafoods with AAnalyst 800 Atomic Absorption Spectrophot...Analysis of Fish and Seafoods with AAnalyst 800 Atomic Absorption Spectrophot...
Analysis of Fish and Seafoods with AAnalyst 800 Atomic Absorption Spectrophot...
 
Toxicity immobilization of refinery sludge containing heavy metals via vitrif...
Toxicity immobilization of refinery sludge containing heavy metals via vitrif...Toxicity immobilization of refinery sludge containing heavy metals via vitrif...
Toxicity immobilization of refinery sludge containing heavy metals via vitrif...
 
PerkinElmer Application Note: Monitoring volatile organic compounds in beer p...
PerkinElmer Application Note: Monitoring volatile organic compounds in beer p...PerkinElmer Application Note: Monitoring volatile organic compounds in beer p...
PerkinElmer Application Note: Monitoring volatile organic compounds in beer p...
 
Determination of Carbohydrates in Various Matrices by Capillary High-Performa...
Determination of Carbohydrates in Various Matrices by Capillary High-Performa...Determination of Carbohydrates in Various Matrices by Capillary High-Performa...
Determination of Carbohydrates in Various Matrices by Capillary High-Performa...
 
Direct determination of bismuth, indium and lead in sea water by zeeman etaas...
Direct determination of bismuth, indium and lead in sea water by zeeman etaas...Direct determination of bismuth, indium and lead in sea water by zeeman etaas...
Direct determination of bismuth, indium and lead in sea water by zeeman etaas...
 
Distillery Wastewater Decontamination by the Fenton Advanced Oxidation Method
Distillery Wastewater Decontamination by the Fenton Advanced Oxidation MethodDistillery Wastewater Decontamination by the Fenton Advanced Oxidation Method
Distillery Wastewater Decontamination by the Fenton Advanced Oxidation Method
 
Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinEl...
Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinEl...Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinEl...
Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinEl...
 
Jurnal ultra1.pdf
Jurnal ultra1.pdfJurnal ultra1.pdf
Jurnal ultra1.pdf
 
New Ultra-High Capacity TOYOPEARL® Protein A Resin
New Ultra-High Capacity TOYOPEARL® Protein A Resin New Ultra-High Capacity TOYOPEARL® Protein A Resin
New Ultra-High Capacity TOYOPEARL® Protein A Resin
 
Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maxi...
Chromatography: Pesticide Residue Analysis Webinar Series:  Part 3 of 4: Maxi...Chromatography: Pesticide Residue Analysis Webinar Series:  Part 3 of 4: Maxi...
Chromatography: Pesticide Residue Analysis Webinar Series: Part 3 of 4: Maxi...
 
Service Analytics A4
Service Analytics A4Service Analytics A4
Service Analytics A4
 
BIODEGRADATION OF GASOIL BY FUNGAL ISOLATES FROM PETROLEUM CONTAMINATED SOILS
BIODEGRADATION OF GASOIL BY FUNGAL ISOLATES FROM PETROLEUM CONTAMINATED SOILS BIODEGRADATION OF GASOIL BY FUNGAL ISOLATES FROM PETROLEUM CONTAMINATED SOILS
BIODEGRADATION OF GASOIL BY FUNGAL ISOLATES FROM PETROLEUM CONTAMINATED SOILS
 
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OESAnalysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
 
Alkaline Extraction of Cobia (Rachycentroncanadum) Proteins: Physicochemical ...
Alkaline Extraction of Cobia (Rachycentroncanadum) Proteins: Physicochemical ...Alkaline Extraction of Cobia (Rachycentroncanadum) Proteins: Physicochemical ...
Alkaline Extraction of Cobia (Rachycentroncanadum) Proteins: Physicochemical ...
 

Ähnlich wie Analysis of Pb, Cd and As in Spice Mixtures using Graphite Furnace Atomic Absorption Spectrophotometry

Toxic Trace Metals in Edible Oils by Graphite Furnace Atomic Absorption Spect...
Toxic Trace Metals in Edible Oils by Graphite Furnace Atomic Absorption Spect...Toxic Trace Metals in Edible Oils by Graphite Furnace Atomic Absorption Spect...
Toxic Trace Metals in Edible Oils by Graphite Furnace Atomic Absorption Spect...PerkinElmer, Inc.
 
Nitrogen_Protein Determination of Spirulina Algae.pdf
Nitrogen_Protein Determination of Spirulina Algae.pdfNitrogen_Protein Determination of Spirulina Algae.pdf
Nitrogen_Protein Determination of Spirulina Algae.pdfNguyenMinhChau22
 
Analysis of heavy_metal_content_in_conventional_an
Analysis of heavy_metal_content_in_conventional_anAnalysis of heavy_metal_content_in_conventional_an
Analysis of heavy_metal_content_in_conventional_anZubariaKhan7
 
The Determination of Minerals and Metals in Multi-Mineral/Multi- Vitamin Tabl...
The Determination of Minerals and Metals in Multi-Mineral/Multi- Vitamin Tabl...The Determination of Minerals and Metals in Multi-Mineral/Multi- Vitamin Tabl...
The Determination of Minerals and Metals in Multi-Mineral/Multi- Vitamin Tabl...PerkinElmer, Inc.
 
Multi-Element Determination of Cu, Mn, and Se using Electrothermal Atomic Abs...
Multi-Element Determination of Cu, Mn, and Se using Electrothermal Atomic Abs...Multi-Element Determination of Cu, Mn, and Se using Electrothermal Atomic Abs...
Multi-Element Determination of Cu, Mn, and Se using Electrothermal Atomic Abs...IOSR Journals
 
Preparation of γ-Al2O3 and Prioritization of Affecting Factors on the Crystal...
Preparation of γ-Al2O3 and Prioritization of Affecting Factors on the Crystal...Preparation of γ-Al2O3 and Prioritization of Affecting Factors on the Crystal...
Preparation of γ-Al2O3 and Prioritization of Affecting Factors on the Crystal...A Behzadmehr
 
Spotlight on Analytical Applications e-Zine – Volume 14
Spotlight on Analytical Applications e-Zine – Volume 14Spotlight on Analytical Applications e-Zine – Volume 14
Spotlight on Analytical Applications e-Zine – Volume 14PerkinElmer, Inc.
 
Multi-residue pesticide analysis of food samples using acetonitrile extractio...
Multi-residue pesticide analysis of food samples using acetonitrile extractio...Multi-residue pesticide analysis of food samples using acetonitrile extractio...
Multi-residue pesticide analysis of food samples using acetonitrile extractio...Kate?ina Svobodov
 
Optimization of lutein production with mixotrophic cultivation of an indigeno...
Optimization of lutein production with mixotrophic cultivation of an indigeno...Optimization of lutein production with mixotrophic cultivation of an indigeno...
Optimization of lutein production with mixotrophic cultivation of an indigeno...Chen-chun Liu
 
Determination of Propionates and Propionic Acid in Bread Samples Using High P...
Determination of Propionates and Propionic Acid in Bread Samples Using High P...Determination of Propionates and Propionic Acid in Bread Samples Using High P...
Determination of Propionates and Propionic Acid in Bread Samples Using High P...theijes
 
APPLICATIONS OF GAS CHROMATOGRAPHY [APPLICATIONS OF GC] BY Prof. Dr. P.RAVISA...
APPLICATIONS OF GAS CHROMATOGRAPHY [APPLICATIONS OF GC] BY Prof. Dr. P.RAVISA...APPLICATIONS OF GAS CHROMATOGRAPHY [APPLICATIONS OF GC] BY Prof. Dr. P.RAVISA...
APPLICATIONS OF GAS CHROMATOGRAPHY [APPLICATIONS OF GC] BY Prof. Dr. P.RAVISA...Dr. Ravi Sankar
 
Determination of Arsenic in Baby Foods and Fruit Juices by GFAAS
Determination of Arsenic in Baby Foods and Fruit Juices by GFAASDetermination of Arsenic in Baby Foods and Fruit Juices by GFAAS
Determination of Arsenic in Baby Foods and Fruit Juices by GFAASPerkinElmer, Inc.
 
Enzyme Based Analytical Chemistry - Nitrate and the U.S. EPA
Enzyme Based Analytical Chemistry - Nitrate and the U.S. EPAEnzyme Based Analytical Chemistry - Nitrate and the U.S. EPA
Enzyme Based Analytical Chemistry - Nitrate and the U.S. EPAAnna-Marie Davidson
 
2018 sensors and actuators
2018 sensors and actuators2018 sensors and actuators
2018 sensors and actuatorsFilipa Correia
 
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...PerkinElmer, Inc.
 

Ähnlich wie Analysis of Pb, Cd and As in Spice Mixtures using Graphite Furnace Atomic Absorption Spectrophotometry (20)

Toxic Trace Metals in Edible Oils by Graphite Furnace Atomic Absorption Spect...
Toxic Trace Metals in Edible Oils by Graphite Furnace Atomic Absorption Spect...Toxic Trace Metals in Edible Oils by Graphite Furnace Atomic Absorption Spect...
Toxic Trace Metals in Edible Oils by Graphite Furnace Atomic Absorption Spect...
 
Nitrogen_Protein Determination of Spirulina Algae.pdf
Nitrogen_Protein Determination of Spirulina Algae.pdfNitrogen_Protein Determination of Spirulina Algae.pdf
Nitrogen_Protein Determination of Spirulina Algae.pdf
 
Analysis of heavy_metal_content_in_conventional_an
Analysis of heavy_metal_content_in_conventional_anAnalysis of heavy_metal_content_in_conventional_an
Analysis of heavy_metal_content_in_conventional_an
 
The Determination of Minerals and Metals in Multi-Mineral/Multi- Vitamin Tabl...
The Determination of Minerals and Metals in Multi-Mineral/Multi- Vitamin Tabl...The Determination of Minerals and Metals in Multi-Mineral/Multi- Vitamin Tabl...
The Determination of Minerals and Metals in Multi-Mineral/Multi- Vitamin Tabl...
 
Multi-Element Determination of Cu, Mn, and Se using Electrothermal Atomic Abs...
Multi-Element Determination of Cu, Mn, and Se using Electrothermal Atomic Abs...Multi-Element Determination of Cu, Mn, and Se using Electrothermal Atomic Abs...
Multi-Element Determination of Cu, Mn, and Se using Electrothermal Atomic Abs...
 
Determination of Heavy Metals in Baby Foods by ICP-MS
Determination of Heavy Metals in Baby Foods by ICP-MSDetermination of Heavy Metals in Baby Foods by ICP-MS
Determination of Heavy Metals in Baby Foods by ICP-MS
 
Preparation of γ-Al2O3 and Prioritization of Affecting Factors on the Crystal...
Preparation of γ-Al2O3 and Prioritization of Affecting Factors on the Crystal...Preparation of γ-Al2O3 and Prioritization of Affecting Factors on the Crystal...
Preparation of γ-Al2O3 and Prioritization of Affecting Factors on the Crystal...
 
Spotlight on Analytical Applications e-Zine – Volume 14
Spotlight on Analytical Applications e-Zine – Volume 14Spotlight on Analytical Applications e-Zine – Volume 14
Spotlight on Analytical Applications e-Zine – Volume 14
 
20320140501007
2032014050100720320140501007
20320140501007
 
Campanella2017
Campanella2017Campanella2017
Campanella2017
 
Multi-residue pesticide analysis of food samples using acetonitrile extractio...
Multi-residue pesticide analysis of food samples using acetonitrile extractio...Multi-residue pesticide analysis of food samples using acetonitrile extractio...
Multi-residue pesticide analysis of food samples using acetonitrile extractio...
 
Optimization of lutein production with mixotrophic cultivation of an indigeno...
Optimization of lutein production with mixotrophic cultivation of an indigeno...Optimization of lutein production with mixotrophic cultivation of an indigeno...
Optimization of lutein production with mixotrophic cultivation of an indigeno...
 
Determination of Propionates and Propionic Acid in Bread Samples Using High P...
Determination of Propionates and Propionic Acid in Bread Samples Using High P...Determination of Propionates and Propionic Acid in Bread Samples Using High P...
Determination of Propionates and Propionic Acid in Bread Samples Using High P...
 
MELA Nov2015 Presentation Deck
MELA Nov2015 Presentation DeckMELA Nov2015 Presentation Deck
MELA Nov2015 Presentation Deck
 
APPLICATIONS OF GAS CHROMATOGRAPHY [APPLICATIONS OF GC] BY Prof. Dr. P.RAVISA...
APPLICATIONS OF GAS CHROMATOGRAPHY [APPLICATIONS OF GC] BY Prof. Dr. P.RAVISA...APPLICATIONS OF GAS CHROMATOGRAPHY [APPLICATIONS OF GC] BY Prof. Dr. P.RAVISA...
APPLICATIONS OF GAS CHROMATOGRAPHY [APPLICATIONS OF GC] BY Prof. Dr. P.RAVISA...
 
Determination of Arsenic in Baby Foods and Fruit Juices by GFAAS
Determination of Arsenic in Baby Foods and Fruit Juices by GFAASDetermination of Arsenic in Baby Foods and Fruit Juices by GFAAS
Determination of Arsenic in Baby Foods and Fruit Juices by GFAAS
 
Enzyme Based Analytical Chemistry - Nitrate and the U.S. EPA
Enzyme Based Analytical Chemistry - Nitrate and the U.S. EPAEnzyme Based Analytical Chemistry - Nitrate and the U.S. EPA
Enzyme Based Analytical Chemistry - Nitrate and the U.S. EPA
 
2018 sensors and actuators
2018 sensors and actuators2018 sensors and actuators
2018 sensors and actuators
 
Magne gas
Magne gasMagne gas
Magne gas
 
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
 

Mehr von PerkinElmer, Inc.

Detectando a Pré Eclampsia no 1º Trimestre
Detectando a Pré Eclampsia no 1º TrimestreDetectando a Pré Eclampsia no 1º Trimestre
Detectando a Pré Eclampsia no 1º TrimestrePerkinElmer, Inc.
 
As Tecnologias que Definem o Desempenho
As Tecnologias que Definem o DesempenhoAs Tecnologias que Definem o Desempenho
As Tecnologias que Definem o DesempenhoPerkinElmer, Inc.
 
GCMS Portátil: Análises em 5 minutos!
GCMS Portátil: Análises em 5 minutos!GCMS Portátil: Análises em 5 minutos!
GCMS Portátil: Análises em 5 minutos!PerkinElmer, Inc.
 
Nós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidadeNós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidadePerkinElmer, Inc.
 
Janus G3: Rapidez e Automação que você precisa.
Janus G3: Rapidez e Automação que você precisa.Janus G3: Rapidez e Automação que você precisa.
Janus G3: Rapidez e Automação que você precisa.PerkinElmer, Inc.
 
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?PerkinElmer, Inc.
 
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?PerkinElmer, Inc.
 
O Mercado da Carne no Brasil
O Mercado da Carne no BrasilO Mercado da Carne no Brasil
O Mercado da Carne no BrasilPerkinElmer, Inc.
 
KITS de Preparo para Biblioteca NGS
KITS de Preparo para Biblioteca NGSKITS de Preparo para Biblioteca NGS
KITS de Preparo para Biblioteca NGSPerkinElmer, Inc.
 
Escolha um parceiro e não um fornecedor
Escolha um parceiro e não um fornecedorEscolha um parceiro e não um fornecedor
Escolha um parceiro e não um fornecedorPerkinElmer, Inc.
 
Eletroforese Capilar Automatizada
Eletroforese Capilar Automatizada Eletroforese Capilar Automatizada
Eletroforese Capilar Automatizada PerkinElmer, Inc.
 
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...PerkinElmer, Inc.
 
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...PerkinElmer, Inc.
 
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid SequencingHigh Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid SequencingPerkinElmer, Inc.
 

Mehr von PerkinElmer, Inc. (20)

Detectando a Pré Eclampsia no 1º Trimestre
Detectando a Pré Eclampsia no 1º TrimestreDetectando a Pré Eclampsia no 1º Trimestre
Detectando a Pré Eclampsia no 1º Trimestre
 
É A VEZ DO PRIME
É A VEZ DO PRIMEÉ A VEZ DO PRIME
É A VEZ DO PRIME
 
TREC PARA A VIDA
TREC PARA A VIDATREC PARA A VIDA
TREC PARA A VIDA
 
As Tecnologias que Definem o Desempenho
As Tecnologias que Definem o DesempenhoAs Tecnologias que Definem o Desempenho
As Tecnologias que Definem o Desempenho
 
GCMS Portátil: Análises em 5 minutos!
GCMS Portátil: Análises em 5 minutos!GCMS Portátil: Análises em 5 minutos!
GCMS Portátil: Análises em 5 minutos!
 
Nós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidadeNós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidade
 
Janus G3: Rapidez e Automação que você precisa.
Janus G3: Rapidez e Automação que você precisa.Janus G3: Rapidez e Automação que você precisa.
Janus G3: Rapidez e Automação que você precisa.
 
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?
 
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?
 
Aspirina Vs Pré Eclampsia
Aspirina Vs Pré Eclampsia Aspirina Vs Pré Eclampsia
Aspirina Vs Pré Eclampsia
 
O Mercado da Carne no Brasil
O Mercado da Carne no BrasilO Mercado da Carne no Brasil
O Mercado da Carne no Brasil
 
KITS de Preparo para Biblioteca NGS
KITS de Preparo para Biblioteca NGSKITS de Preparo para Biblioteca NGS
KITS de Preparo para Biblioteca NGS
 
O que é X Frágil?
O que é X Frágil?O que é X Frágil?
O que é X Frágil?
 
Escolha um parceiro e não um fornecedor
Escolha um parceiro e não um fornecedorEscolha um parceiro e não um fornecedor
Escolha um parceiro e não um fornecedor
 
Eletroforese Capilar Automatizada
Eletroforese Capilar Automatizada Eletroforese Capilar Automatizada
Eletroforese Capilar Automatizada
 
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
 
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
 
CG Clarus® All In One
CG Clarus® All In OneCG Clarus® All In One
CG Clarus® All In One
 
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid SequencingHigh Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing
 
XXI Encontro de Genética
XXI Encontro de GenéticaXXI Encontro de Genética
XXI Encontro de Genética
 

Kürzlich hochgeladen

Breaking the Kubernetes Kill Chain: Host Path Mount
Breaking the Kubernetes Kill Chain: Host Path MountBreaking the Kubernetes Kill Chain: Host Path Mount
Breaking the Kubernetes Kill Chain: Host Path MountPuma Security, LLC
 
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
04-2024-HHUG-Sales-and-Marketing-Alignment.pptxHampshireHUG
 
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...Integration and Automation in Practice: CI/CD in Mule Integration and Automat...
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...Patryk Bandurski
 
Pigging Solutions in Pet Food Manufacturing
Pigging Solutions in Pet Food ManufacturingPigging Solutions in Pet Food Manufacturing
Pigging Solutions in Pet Food ManufacturingPigging Solutions
 
Swan(sea) Song – personal research during my six years at Swansea ... and bey...
Swan(sea) Song – personal research during my six years at Swansea ... and bey...Swan(sea) Song – personal research during my six years at Swansea ... and bey...
Swan(sea) Song – personal research during my six years at Swansea ... and bey...Alan Dix
 
Presentation on how to chat with PDF using ChatGPT code interpreter
Presentation on how to chat with PDF using ChatGPT code interpreterPresentation on how to chat with PDF using ChatGPT code interpreter
Presentation on how to chat with PDF using ChatGPT code interpreternaman860154
 
Understanding the Laravel MVC Architecture
Understanding the Laravel MVC ArchitectureUnderstanding the Laravel MVC Architecture
Understanding the Laravel MVC ArchitecturePixlogix Infotech
 
Pigging Solutions Piggable Sweeping Elbows
Pigging Solutions Piggable Sweeping ElbowsPigging Solutions Piggable Sweeping Elbows
Pigging Solutions Piggable Sweeping ElbowsPigging Solutions
 
Benefits Of Flutter Compared To Other Frameworks
Benefits Of Flutter Compared To Other FrameworksBenefits Of Flutter Compared To Other Frameworks
Benefits Of Flutter Compared To Other FrameworksSoftradix Technologies
 
Neo4j - How KGs are shaping the future of Generative AI at AWS Summit London ...
Neo4j - How KGs are shaping the future of Generative AI at AWS Summit London ...Neo4j - How KGs are shaping the future of Generative AI at AWS Summit London ...
Neo4j - How KGs are shaping the future of Generative AI at AWS Summit London ...Neo4j
 
Beyond Boundaries: Leveraging No-Code Solutions for Industry Innovation
Beyond Boundaries: Leveraging No-Code Solutions for Industry InnovationBeyond Boundaries: Leveraging No-Code Solutions for Industry Innovation
Beyond Boundaries: Leveraging No-Code Solutions for Industry InnovationSafe Software
 
AI as an Interface for Commercial Buildings
AI as an Interface for Commercial BuildingsAI as an Interface for Commercial Buildings
AI as an Interface for Commercial BuildingsMemoori
 
The 7 Things I Know About Cyber Security After 25 Years | April 2024
The 7 Things I Know About Cyber Security After 25 Years | April 2024The 7 Things I Know About Cyber Security After 25 Years | April 2024
The 7 Things I Know About Cyber Security After 25 Years | April 2024Rafal Los
 
Install Stable Diffusion in windows machine
Install Stable Diffusion in windows machineInstall Stable Diffusion in windows machine
Install Stable Diffusion in windows machinePadma Pradeep
 
From Event to Action: Accelerate Your Decision Making with Real-Time Automation
From Event to Action: Accelerate Your Decision Making with Real-Time AutomationFrom Event to Action: Accelerate Your Decision Making with Real-Time Automation
From Event to Action: Accelerate Your Decision Making with Real-Time AutomationSafe Software
 
Enhancing Worker Digital Experience: A Hands-on Workshop for Partners
Enhancing Worker Digital Experience: A Hands-on Workshop for PartnersEnhancing Worker Digital Experience: A Hands-on Workshop for Partners
Enhancing Worker Digital Experience: A Hands-on Workshop for PartnersThousandEyes
 
Human Factors of XR: Using Human Factors to Design XR Systems
Human Factors of XR: Using Human Factors to Design XR SystemsHuman Factors of XR: Using Human Factors to Design XR Systems
Human Factors of XR: Using Human Factors to Design XR SystemsMark Billinghurst
 
How to Remove Document Management Hurdles with X-Docs?
How to Remove Document Management Hurdles with X-Docs?How to Remove Document Management Hurdles with X-Docs?
How to Remove Document Management Hurdles with X-Docs?XfilesPro
 
Handwritten Text Recognition for manuscripts and early printed texts
Handwritten Text Recognition for manuscripts and early printed textsHandwritten Text Recognition for manuscripts and early printed texts
Handwritten Text Recognition for manuscripts and early printed textsMaria Levchenko
 
Unblocking The Main Thread Solving ANRs and Frozen Frames
Unblocking The Main Thread Solving ANRs and Frozen FramesUnblocking The Main Thread Solving ANRs and Frozen Frames
Unblocking The Main Thread Solving ANRs and Frozen FramesSinan KOZAK
 

Kürzlich hochgeladen (20)

Breaking the Kubernetes Kill Chain: Host Path Mount
Breaking the Kubernetes Kill Chain: Host Path MountBreaking the Kubernetes Kill Chain: Host Path Mount
Breaking the Kubernetes Kill Chain: Host Path Mount
 
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
 
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...Integration and Automation in Practice: CI/CD in Mule Integration and Automat...
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...
 
Pigging Solutions in Pet Food Manufacturing
Pigging Solutions in Pet Food ManufacturingPigging Solutions in Pet Food Manufacturing
Pigging Solutions in Pet Food Manufacturing
 
Swan(sea) Song – personal research during my six years at Swansea ... and bey...
Swan(sea) Song – personal research during my six years at Swansea ... and bey...Swan(sea) Song – personal research during my six years at Swansea ... and bey...
Swan(sea) Song – personal research during my six years at Swansea ... and bey...
 
Presentation on how to chat with PDF using ChatGPT code interpreter
Presentation on how to chat with PDF using ChatGPT code interpreterPresentation on how to chat with PDF using ChatGPT code interpreter
Presentation on how to chat with PDF using ChatGPT code interpreter
 
Understanding the Laravel MVC Architecture
Understanding the Laravel MVC ArchitectureUnderstanding the Laravel MVC Architecture
Understanding the Laravel MVC Architecture
 
Pigging Solutions Piggable Sweeping Elbows
Pigging Solutions Piggable Sweeping ElbowsPigging Solutions Piggable Sweeping Elbows
Pigging Solutions Piggable Sweeping Elbows
 
Benefits Of Flutter Compared To Other Frameworks
Benefits Of Flutter Compared To Other FrameworksBenefits Of Flutter Compared To Other Frameworks
Benefits Of Flutter Compared To Other Frameworks
 
Neo4j - How KGs are shaping the future of Generative AI at AWS Summit London ...
Neo4j - How KGs are shaping the future of Generative AI at AWS Summit London ...Neo4j - How KGs are shaping the future of Generative AI at AWS Summit London ...
Neo4j - How KGs are shaping the future of Generative AI at AWS Summit London ...
 
Beyond Boundaries: Leveraging No-Code Solutions for Industry Innovation
Beyond Boundaries: Leveraging No-Code Solutions for Industry InnovationBeyond Boundaries: Leveraging No-Code Solutions for Industry Innovation
Beyond Boundaries: Leveraging No-Code Solutions for Industry Innovation
 
AI as an Interface for Commercial Buildings
AI as an Interface for Commercial BuildingsAI as an Interface for Commercial Buildings
AI as an Interface for Commercial Buildings
 
The 7 Things I Know About Cyber Security After 25 Years | April 2024
The 7 Things I Know About Cyber Security After 25 Years | April 2024The 7 Things I Know About Cyber Security After 25 Years | April 2024
The 7 Things I Know About Cyber Security After 25 Years | April 2024
 
Install Stable Diffusion in windows machine
Install Stable Diffusion in windows machineInstall Stable Diffusion in windows machine
Install Stable Diffusion in windows machine
 
From Event to Action: Accelerate Your Decision Making with Real-Time Automation
From Event to Action: Accelerate Your Decision Making with Real-Time AutomationFrom Event to Action: Accelerate Your Decision Making with Real-Time Automation
From Event to Action: Accelerate Your Decision Making with Real-Time Automation
 
Enhancing Worker Digital Experience: A Hands-on Workshop for Partners
Enhancing Worker Digital Experience: A Hands-on Workshop for PartnersEnhancing Worker Digital Experience: A Hands-on Workshop for Partners
Enhancing Worker Digital Experience: A Hands-on Workshop for Partners
 
Human Factors of XR: Using Human Factors to Design XR Systems
Human Factors of XR: Using Human Factors to Design XR SystemsHuman Factors of XR: Using Human Factors to Design XR Systems
Human Factors of XR: Using Human Factors to Design XR Systems
 
How to Remove Document Management Hurdles with X-Docs?
How to Remove Document Management Hurdles with X-Docs?How to Remove Document Management Hurdles with X-Docs?
How to Remove Document Management Hurdles with X-Docs?
 
Handwritten Text Recognition for manuscripts and early printed texts
Handwritten Text Recognition for manuscripts and early printed textsHandwritten Text Recognition for manuscripts and early printed texts
Handwritten Text Recognition for manuscripts and early printed texts
 
Unblocking The Main Thread Solving ANRs and Frozen Frames
Unblocking The Main Thread Solving ANRs and Frozen FramesUnblocking The Main Thread Solving ANRs and Frozen Frames
Unblocking The Main Thread Solving ANRs and Frozen Frames
 

Analysis of Pb, Cd and As in Spice Mixtures using Graphite Furnace Atomic Absorption Spectrophotometry

  • 1. a p p l i c at i o n N o t e Atomic Absorption Author Praveen Sarojam, Ph.D. PerkinElmer, Inc. Shelton, CT 06484 USA Analysis of Pb, Cd and As in Spice Mixtures using Graphite Furnace Atomic Absorption Spectrophotometry Introduction The toxicity and effect of trace heavy metals on human health and the environment has attracted considerable attention and concern in recent years. With an inherent toxicity, a tendency to accumulate in the food chain and a particularly low removal rate through excretion,1 lead (Pb), cadmium (Cd) and arsenic (As) cause harm to humans even at low concentrations. Exposure to trace and heavy metals above the permissible level affects human health and may result in teratogenicity (reproductive effects). Individuals may also experience high blood pressure, fatigue, as well as kidney and neurological disorders. Spices, the dried parts of plants, grow widely in various regions of the world, are produced either on small farmlands or naturally grown, and have been used for several purposes since ancient times. Most are fragrant and flavorful and are used for culinary purposes to improve the quality of food.2 Natural food spices, such as pepper, have been reported to contain significant quantities of some heavy metals, including Pb, Cd and As. Contamination with heavy metals may be accidental (e.g. contamination of the environment during plant cultivation) or deliberate – in some cultures, according to traditional belief, specially treated heavy metals are associated with health benefits and are thus an intentional ingredient of traditional remedies. Spices and herbal plants may contain heavy metal ions over a wide range of concentrations.3,4 There is often little information available about the safety of those plants and their products in respect to heavy metal contamination. Due to the significant amount of spices consumed, it is important to know the toxic metal concentrations in them.5
  • 2. The major challenge in performing spice analyses is the extremely low analyte concentrations combined with high matrix levels. Graphite furnace atomic absorption spectrophotometry (GFAAS) has been one of the more reliable techniques used for many years and is therefore the preferred analytical method. The use of longitudinal Zeeman background correction and the use of matrix modifiers provides accurate results and low detection limits in samples with high matrices. This makes GFAAS an indispensible tool in carrying out such analyses. The objective of this work is two-fold: (1) to use GFAAS to accurately analyze the levels of Pb, Cd, and As present in some major spices commonly available on the market; and (2) to cross reference these measured levels to the recommended limits specified by the U.S. FDA. Experimental Conditions Instrumentation The measurements were performed using a PerkinElmer® PinAAcle™ 900T atomic absorption spectrophotometer (Shelton, CT, USA) equipped with the intuitive WinLab32™ for AA software running under Microsoft® Windows™ 7, which features all the tools to analyze samples, report and archive data and ensure regulatory compliance (Figure 1). The high-efficiency optical system and solid-state detector used in the PinAAcle 900T spectrometer provide outstanding signal-to-noise ratios. The longitudinal Zeeman-effect background correction for graphite furnace analysis provides accurate background correction without the loss of light that usually occurs in transverse Zeeman systems. The use of a transversely heated graphite atomizer (THGA) provides uniform temperature distribution across the entire length of the graphite tube, eliminating memory effects and potential interferences that may occur with high-matrix sample analysis. Pyrolytically coated THGA tubes with end caps (Part No. B3000655) were used for all measurements. The optimized instrumental conditions for furnace experiments are given in Table 1 (Page 3), and the graphite furnace temperature programs are listed in Appendix I (Page 5). A heated injection was used for lead; it can also be used for cadmium and arsenic. Figure 1. PerkinElmer PinAAcle 900T atomic absorption spectrophotometer. 2 A microwave sample preparation system was used for the microwave-assisted digestion. This is an industrial-type microwave oven, which is equipped with various accessories to optimize the sample digestion. The samples were digested using ten 100 mL high-pressure vessels made of PTFE. The microwave digestion program is listed in Table 2 (Page 3). Standards, Chemicals and Certified Reference Materials PerkinElmer Pure single-element calibration standards for Pb, Cd, and As were used as the stock standards for preparing the working standards (Part Nos. Pb: N9300128; Cd: N9300107; As: N9300102). Working standards were prepared by serial volume/volume (v/v) dilution in polypropylene vials (Part Nos. B0193233 15 mL Conical; B0193234 50 mL Conical Freestanding) ASTM® Type I deionized water (Millipore® Corporation, Billerica, Massachusetts, U.S.) acidified with 0.2% nitric acid (HNO3) (Tamapure®, TAMA Chemicals, Japan) was used as the calibration blank and for all dilutions. 30% hydrogen peroxide (H2O2) (Kanto Chemicals, Tokyo, Japan) was used for digestion along with nitric acid. Matrix modifiers were prepared from 10% NH4H2PO4 (Part No. N9303445), 1% Mg as Mg(NO3)2 (Part No. B0190634) and 1% Pd (Part No. B0190635) stock solutions, by diluting with the 0.2% HNO3 made above. Matrix modifiers were added automatically to each standard, blank and sample by the AS 900 autosampler, an integral part of the PinAAcle 900T spectrometer. NIST® 1568a Certified Reference Material (CRM) for Trace Metals in Rice Flour was used to validate the method. Quality control (QC) check standards were prepared at the calibration curve midpoint concentration for each individual element. Sample and Certified Reference Material Preparation Plastic bottles were cleaned by soaking with 10% v/v HNO3 for at least 24 hours and rinsed abundantly in deionized water before use. The polypropylene autosampler cups (Part No. B3001566) were soaked in 20% nitric acid overnight to minimize sample contamination, and thoroughly rinsed with 0.5% HNO3 acid before use. Five-point calibration curves (four standards and one blank) were constructed for each analyte and the calibration curve correlation coefficient was examined to ensure a value better than 0.998 before the start of the sample analysis (Appendix II – Page 6).
  • 3. Five branded powdered spice samples available in local supermarkets in India (coriander powder, ginger powder, asafetida, black pepper powder and red chili powder) were analyzed. Approximately 0.5 g of each sample or CRM, accurately weighed in duplicate, was transferred to the vessel of the microwave digestion system and the sample digestion method (Table 2) was performed in accordance with U.S. Environmental Protection Agency (EPA) Method 3052. The digested samples were diluted with 0.2% HNO3 and brought up to 25 mL in polypropylene vials. Table 1. Optimized experimental conditions of PinAAcle 900T. Analyte Pb Cd As Wavelength (nm) 283.3 228.8 193.7 Lamp Type EDL HCL EDL (Part No. N3050657) (Part No. N3050115) (Part No. N3050605) Standards (µg/L) 5, 10, 15, 20 0.5, 0.75, 1.0, 2.0 10, 20, 30, 40 Correlation Coefficient 0.9991 0.9996 0.9989 Read Time (sec) 3 5 3 Injection Temp (˚C) 90 20 20 0.05 mg NH4H2PO4 0.005 mg Pd Matrix Modifier (mg) 0.05 mg NH4H2PO4 and 0.003 mg and 0.003 mg and 0.003 mg Mg(NO3)2 Mg(NO3)2 Mg(NO3)2 Modifier Volume (µL) 5 Slit Width (nm) 0.7 Sample Volume (µL) 20 Measurement Type Peak Area Calibration Linear through zero Table 2. Program used for the digestion of spices and herbs. Sequence Power (watts) 1 2 1000 0 Ramp Time (min) 10 0 Hold Time (min) 10 20 Weight Taken (mg) ~500 H2O2 (mL) 1.0 HNO3 (mL) 7.0 Temperature (˚C) 180 Results and Discussions In the analysis of complex samples with the graphite furnace, obtaining reproducible results can be a challenging task as one has to deal with analytes present at the low µg/L ppb level and the potential for matrix interferences. With Stabilized Temperature Platform Furnace (STPF) conditions (including: pyrolytically coated graphite tubes, platform atomization, rapid heating, internal gas stop during atomization, fast signal processing and Zeeman background correction) the analysis of low level analytes in spice mixtures is an easy task. The graphite furnace sample introduction system is of paramount importance in optimizing the short-term stability of signals. The PinAAcle 900T spectrometer uses a unique built-in camera to monitor sampler tip alignment and sample introduction into the graphite tube. With the TubeView™ furnace camera, it is simple to position the AS 900 autosampler sample capillary tip at the correct depth inside the tube so as to achieve highly reproducible pipetting. 3
  • 4. Table 3. Analysis of certified reference material by GFAA. NIST® 1568a Rice Flour Certified Value Analyte (µg/g) Measured Value (µg/g) Pb <0.010 0.0093 Cd 0.022 ±0.002 0.020 ±0.004 As 0.29 ±0.03 0.24 ±0.02 Table 4. Estimated method detection limits (MDLs). Analyte MDL (µg/L) Pb 9.5 Cd 2.35 As 9.5 Table 5. Spike recovery studies. Analyte Sample Post- Pre- digestion Spike digestion Spike Recovery Level Recovery Level (%) (µg/L) (%) (µg/L) Pb Coriander powder Ginger powder 91 5.00 90 111 * 5.00 Cd Coriander powder Ginger powder 104 0.625 96 119 * 1.00 As Coriander powder Ginger powder 108 8.33 106 100 * 5.00 *Pre-digestion spike recovery not performed The developed method has been validated by incorporating various QC checks and by the analysis of a CRM. The agreement between the certified and measured values was good (Table 3), which demonstrates the accuracy of the direct calibration being used as well as the overall accuracy of the developed method. Method detection limits (MDLs) were calculated (Table 4) based on the standard deviation of seven replicates of the reagent blank (t-value = 3.14) and took into account the dilution factor for the samples. The low method detection limits obtained show the capability of the PinAAcle 900T spectrometer in analyzing difficult matrices at the measured concentrations. Incomplete mineralization of samples during the microwavedigestion process, due to high organic content, may cause difficulty in transferring analytes into solution. These impurities may also affect the analytical measurements.6 Application of concentrated HNO3 along with H2O2 for mineralization of spices and herbs leads to the complete digestion of samples, which is proven by the determination of the values of the analytes in the CRM (Table 3). Both a pre-digestion and an automatic post-digestion recovery study were performed. The recoveries obtained for the pre-digestion spike easily meet the U.S. EPA recovery guidelines of ±30% and indicate that neither analyte loss nor contamination was of major concern throughout the analytycal procedure (Table 5). The recoveries obtained for the post-digestion spike indicate that no matrix interference towards the analyte signal was seen. The results of the sample analysis are summarized in Table 6. The level of arsenic, cadmium and lead in all the samples analyzed were well within the permissible limits of 10, 0.3 and 10 mg/kg respectively, as specified by the U.S. FDA. Table 6. Analysis of spices and herbs samples using GFAAS (µg/kg). Analyte U.S. FDA Limit Replicate Coriander Powder Ginger Powder Asafetida Black Pepper Powder Red Chili Powder Pb 10,000 Sample 455 551 124 9.8 286 Cd 300 Sample 48 89 8.2 25 57 Duplicate 47 92 8.2 28 59 As 20 49 41 49 <MDL 24 42 38 32 <MDL 10,000 Sample Duplicate 4
  • 5. Conclusions A complete method for the accurate determination of Pb, Cd and As in various spice mixtures using the PinAAcle 900T atomic absorption spectrophotometer in the GFAAS mode was developed. The results confirmed that the determination of arsenic, cadmium and lead in spice mixtures, after acid solubilization by microwave digestion, can be performed by GFAAS without any interference. The PinAAcle 900Z (Longitudinal Zeeman Furnace only) spectrometer can also be used for this application. References 1. O. Sadeghi, N. Tavassoli, M.M. Amini, H. Ebrahimzadeh, N. Daei, Food Chemistry 127 (2011) 364–368. 2. H. Mubeen, I. Naeem, A. Taskeen and Z. Saddiqe, New York Science Journal, 2 (5) (2009) 1554-0200. 3. K. K Gupta, S. Bhattacharjee, S. Kar, S. Chakrabarty, P. Thakar, G. Bhattacharyya and S.C. Srivastava, Comm. Soil Plant Anal., 34 (2003) 681-693. 4. T. M. Ansari, N. Ikram, M. Najam-ul-Haq, O. Fayyaz, I. Ghafoor and N. Khalid, J. Biol. Sci., 4 (2004) 95-99. 5. R.P. Choudhury and A.N. Garg, Food Chemistry, 104 (2007) 1454-1463. 6. I. Baranowska, K. Srogi, A. Włochowicz, K. Szczepanik, Polish Journal of Environmental Studies, 11( 5) (2002) 467-471. Appendix I – Graphite Furnace Temperature Program Table 7. Furnace temperature program for Pb. Analyte Step Temp ˚C Ramp Time (sec) Hold Time (sec) Internal Gas Flow (mL/min) Pb 1 110 1 30 250 2 130 15 30 250 3 850 10 20 250 4 1600 0 5 0 5 2450 1 3 250 Gas Type Argon Argon Argon – Argon Table 8. Furnace temperature program for Cd. Analyte Step Temp (˚C) Ramp Time (sec) Cd 1 110 10 2 130 15 3 500 15 4 1500 0 5 2450 1 Hold Time (sec) Internal Gas Flow (mL/min) 30 250 30 250 35 250 3 0 3 250 Gas Type Argon Argon Argon – Argon Table 9. Furnace temperature program for As. Analyte Step Temp (˚C) Ramp Time (sec) Hold Time (sec) Internal Gas Flow (mL/min) As 1 110 5 30 250 2 130 20 30 250 3 800 15 40 250 4 1200 15 30 250 5 2200 0 5 0 6 2450 1 3 250 Gas Type Argon Argon Argon Argon – Argon 5
  • 6. Appendix II – Calibration Graphs and Linearity Figure 2. Calibration curve for the determination of Pb using GFAAS. Figure 3. Calibration curve for the determination of Cd using GFAAS. Figure 4. Calibration curve for the determination of As using GFAAS. PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs Copyright ©2011, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 009754A_01