Flame emission and Atomic
absorption Spectroscopy
Presented by: Guided by:
Himal Barakoti Arup Chakroborty
M. Pharm, 1st Sem Assistant Professor
Department of Pharmacy Department of Pharmacy
Assam Down Town University Assam Down Town University
Contents:
 Introduction
 Theory and principle of Flame emission Spectroscopy
 Instrumentation
 Interferences
 Application
 Atomic Absorption Spectroscopy
 Theory and Principle
2
Contents:
 Instrumentation
 Interferences
 Application
 Comparison between Flame emission and Atomic absorption
Spectroscopy
 References
3
Introduction
 Atomic spectroscopy is thought to be the oldest
instrumental method for the determination of
elements.
 These techniques are introduced in the mid of 19th
Century during which Gustav Kirchhoff and Robert
Bunsen showed that the radiation emitted from the
flames depends on the characteristic element present in
the flame.
 The developments in the instrumentation area led to the
widespread application of atomic spectroscopy.
 The absorption and emission of radiant energy by
atoms provide powerful analytical tools for both
quantitative and qualitative analysis of substance.
4
Flame emission Spectroscopy
 Flame emission spectroscopy (FES) is a method of chemical analysis that
uses intensity of light emitted from flame, arc or spark at particular
wavelength to determine quantity of element in sample.
 The basis of flame photometric working is that, the species of alkali metals
(Group 1) and alkaline earth metals (Group II) metals are dissociated due to
the thermal energy provided by the flame source.
 Due to this thermal excitation, some of the atoms are excited to a higher energy
level where they are not stable.
 The absorbance of light due to the electrons excitation can be measured by using
the direct absorption techniques.
5
 The subsequent loss of energy will result in the movement of excited atoms
to the low energy ground state with emission of some radiations, which can
be visualized in the visible region of the spectrum.
 The intensity of radiation emitted by these excited atoms returning to the
ground state provides the basic for analytical determination in FES.
 The wavelength of emitted light is specific for specific elements.
 In flame emission spectroscopy,
o Wavelength of spectral lines give identity of elements.
o Intensity of emitted light is directly proportional to the number of atoms
present.
6
Instrumentation
 Flame photometers are the simplest type of atomic spectrometers.
 Solution is introduced into a fine spray.
 Solvents evaporates leaving dehydrated salt.
 Certain fraction of atoms absorbs energy and are raised to excited state.
 These excited atoms on returning to ground state emits photons of certain wavelength.
 Flame emission passes through monochromator which filters all emitted light expect the
wavelength of our interest.
 Photoelectric Detector measures the intensity of filtered light.
7
1. Atomizer/Nebulizer
 Process of conversion of sample to a fine mist of finely divided droplets using a
jet of compressed gas.
o Pneumatic nebulizers
o Electro thermal vaporizer
o Ultrasound nebulizer
2. Burner
 They must have ability to evaporate the liquid droplets from sample solution and
capacity to excite atoms formed and cause them to emit radiant energy.
o Mecker burner
o Total consumption burner
o Laminar flow burner
o Shileded burner
o Lundergraph burner
8
3. Monochromator (Grating or prism monochromator)
They disperse radiation coming from the flame and falling on it. Dispersed
radiation goes to detector from exit slit.
Filters chosen has wavelength range transparent to emission from element
of interest.
4. Detectors (photoemissive cells or photomultiplier tubes)
They measure intensity of radiation falling on it
5. Amplifier & readout devices
It suitably amplify and display detector’s output.
Amplifier sensativity can be changed to be able to analyze sample of
varying concentration.
9
Types of flame used
The most common instrument uses air as oxidant. The temperature of the
flames produced is relatively low so the technique is only suitable for the
elements that are easily excited such as alkali and alkali earth elements.
10
Flame Temp/˚C
Gas/Air 1700-1900
Gas/O2 2700-2800
H2/Air 2000-2100
H2/O2 2550-2700
C2H2/Air 2100-2400
C2H2/O2 3050-3150
C2H2/N2O 2600-2800
 A higher temperature will tend to increase the number of atoms in the excited
state and hence the signal.
 Some detection limits:
11
Elements Spectral lines/nm Flame Detection limits/ppm
Al 396 C2H2/N2O 0.1
Ba 553 C2H2/N2O 0.01
K 766 C2H2/O2 0.001
Li 671 C2H2/N2O 0.0001
Interferences
 Number of factors beside the analyte affect the intensity of emitted radiation. The
analytical signals often include contributions from constituents other than
analyte termed as interferences and are found to interfere the outcome of
procedure.
 3 types:
1. Spectral Interferences: interference that affect spectral intensity or resolution.
a. 1st type of interference arises when two elements exhibit spectra, which
partially overlap, and both emit radiation at particular wavelength. The
detector cannot distinguish between the source of radiation and records the
total sign, thus resulting in incorrect answer. This type of interferences are
more common in high flame temperature as numerous spectral lines are
produced in high temperature.
E.g., Fe line at 324.73 nm overlaps with Cu line at 324.75 nm. This Can be
overcome by taking alternative wavelength with no overlap.
12
b. 2nd type of spectral interference deals with spectral lines of two or more
elements which are close but their spectra do not overlap. The filter may allow
spectral lines separated by 5.0-10.0 nm to pass through, thus resulting in error in
analysis. This can be reduced by increasing the resolution of spectral
isolation system.
c. 3rd type of spectral interference occurs due to presence of continuous
background which arises due to high concentration of salts in the sample. Some
organic solvents also produce continuous background. This can be corrected by
using suitable scanning technique.
2. Ionization Interference: In certain cases, high temperature flame may cause
ionization of the metal atoms. E.g., In case of sodium
Na Na+ + e-
The Na+ ion possesses an emission spectrum of its own frequencies, which
are different from those of Na atom. This reduces the radiant power of atomic
emission. This type of interference can be eliminated by addition of large
quantity of potassium salt to standard and sample solution. Potassium itself
undergoes ionization due to low ionization energy and suppresses the ionization
of sodium.
13
3. Chemical Interferences: This arises out of reaction between different
interferents and the analyte.
a. Cation-cation interference: The presence of certain anions, such as oxalate,
phosphate, sulphate and aluminate in a solution may affect the intensity of
radiation emitted by an element, resulting in serious analytical error. For e.g.,
calcium in presence of phosphate ion forms a stable substance as Ca3(PO4)2
which does not decompose easily, resulting in production of lesser atom. Thus
calcium signal is depressed. This can be removed either by extraction of anion
or by using calibration curve prepared from standard solutions containing
same concentrations of the anion as found in the sample.
b. Cation-cation interference: In many cases, mutual interferences of cations results
in reducing element signal intensity. These interference are neither spectral nor
ionic in nature and the actual mechanism has not been well understood. E.g.,
Al interferes with Ca and Mg. Na interferes with Mg.
c. Oxide formation: arises due to formation of stable metal oxide if oxygen is
present in flame, resulting in reduced signal intensity. E.g., Alkaline earth metal.
This can be eliminated by either using high flame temperature to dissociate
the oxides or by using oxygen-deficient environment to produce excited
atom.
14
Limitations
 As natural gas and air flame is employed for excitation, the temperature is not
high enough to excite transition metals, therefore the method is selective
towards detection of alkali and alkaline earth metals.
 Low temperature makes this method susceptible to certain disadvantages, most
of them related to flame stability and aspiration conditions. Fuel and oxidation
flow rates and purity, aspiration rates, solution viscosity affects these.
 FES is a means of determining the total metal concentration of a sample; it tells
us nothing about the molecular form of the metal in original sample.
 Only liquid samples can be used.
15
Applications
 Flame Photometers are widely used in quality control where a simple and quick
determination of alkali or alkali earth metal is required. They have the
advantage of being significantly lower priced than most other atomic
spectrometers.
 Biological/medical applications- notable applications are the determinations of
Na, K, Ca and Mg in body fluids and other biological samples.
 Food industry – Determination of calcium and iron in beer
16
Atomic Absorption Spectroscopy
 Guystav Kirchoff and Robert Bunsen first used atomic absorption
spectroscopy—along with atomic emission—in 1859 and 1860 as a means for
identify atoms in flames and hot gases.
 Modern atomic absorption spectroscopy has its beginnings in 1955 as a result of
the independent work of Alan. C. Walsh and C. T. J. Alkemade.
17
 Atomic absorption spectroscopy (AAS) is an absorption spectroscopic method
where radiation from a source is absorbed by non-excited atoms in vapour
state.
 AAS deals with the absorption of specific wavelength of radiation by neutral
atoms in the ground state.
 Reliable for detecting over 70 elements with metals and metalloids.
18
Theory and Principle
 The technique uses basically the principle that free atoms (gas) generated in an
atomizer can absorb radiations at specific frequency.
 AAS quantifies the absorption of ground state atoms in the gaseous form.
 Atoms absorb UV/visible light and make transition to higher electronic energy levels. The
analyte concentration is determined from the amount of absorption.
 Concentration measurement are usually determined from a working curve after
calibrating instrument with standard known concentration.
Liquid sample Formation of droplets Fine residue Formation of neutral ion
Neutral atom absorbs specific wavelength of radiation from hollow cathode lamp
Measurement of intensity of radiation absorbed by the neutral atom using the detector
19
 The extent to which radiation of a particular frequency is absorbed by an atomic
vapour is related to the length of the path transverse and to the concentration of
absorbing atoms in the vapour.
 This is analogues to the Beer-Lamberts law relating to samples in solution.
Thus, for a collimated monochromatic beam of radiation of incident Io passing
through an atomic vapour of thickness I
Iv = Io eл-kvI
Where,
Iv is the intensity of the transmitted radiation
Kv is absorption co efficient
20
Instrumentations21
1. Sharp line radiation source (hollow cathode lamp)
 Hollow cathode lamps are most commonly used radiation containing tungsten
anode and hollow cylindrical cathode made of element to be determined.
 These are sealed with gas tube filled with an inert gas.
 When a current flows between the anode and cathode in this lamps, metal atoms
are sputtered from the cathode cup, and collisions occur with the filler gas. A
number of metal atoms become excited and give off their characteristics radiation.
22
2. A solution Nebulizer and burner
 Nebulizer suck up liquid samples at controlled rate and create a fine aerosol spray
for introduction into flame.
 Aerosol, fuel and oxidant are thoroughly mixed for introduction into flame.
 Atomization is carried out by separation of particles into individual molecules and
breaking molecules into atoms. This is done by exposing analyte to high
temperature in flame or graphite furnace.
 In graphite furnace, samples are deposited in a small graphite coated tube which
can then be heated to vaporize and atomize the analyte.
 Graphite tubes are heated using a high current power supply.
23
3. Monochromator
 They are used to select specific wavelength of light which s absorbed by the
sample, and to exclude other wavelengths.
 Selections of specific light allows the determination of selected element in the
presence of others.
4. Detectors
 The light selected by monochromator is directed onto a detector that is typically a
photomultiplier tube which convert the light signal into electrical signal
proportional to the light intensity.
 The signal is amplified, displayed for readout of fed to data station for print out.
24
Interferences
 Spectral Interferences: Spectral interferences are caused by presence of
another atomic absorption line or a molecular absorbance band close to the
spectral line of element of interest.
 Chemical interference: If a sample contains a species which forms a thermally
stable compound with the analyte that is not completely decomposed by the
energy available in the flame then chemical interference exists. Refractory
elements such as Ti, W, Zr, Mo and Al may combine with oxygen to form thermally
stable oxides. Analysis of such elements can be carried out at higher flame
temperatures using nitrous oxide – acetylene flame instead of air-acetylene to
provide higher dissociation energy.
25
 Matrix interference: When a sample is more viscous or has different surface
tension than the standard it can result in differences in sample uptake rate due
to changes in nebulization efficiency. Such interferences are minimized by
matching as closely as possible the matrix composition of standard and
sample.
 Ionization interference: Excess energy of the flame can lead to excitation of
ground state atoms to ionic state by loss of electrons thereby resulting in
depletion of ground state atoms.
26
limitations
 Expensive
 Low precision
 Low sample throughput
 Requires high level of operator skill
 Sample must be in solution or at least volatile
 Individual source lamps required for each element
27
Applications
 Determination of metal at trace level in solution.
 Determination of purity:
o Presence of heavy metal in body fluids
o Pollution of water by metals
o Food stuffs
o Soft drinks and beer
o Analysis of Geochemical exploration for mineral
o Soils, crude oils, petroleum products, plastics.
28
Comparison between FES and AAS
Flame Emission spectroscopy Atomic Absorption spectroscopy
1. Amount of light emitted by excited atom is
measured.
1. Amount of light absorbed by ground state
atom is measured.
2. Absorption intensity and signal response
greatly influenced by temperature variation.
2. Absorption intensity and signal response
does not depend upon temperature.
3. Beer’s law is not obeyed. 3. Beers law is obeyed.
4. Intensity of emitted radiation is directly
proportional to the number of atoms in excited
state.
4. Intensity of absorbed radiation is directly
proportional to the number of atoms in ground
state.
5. Relation between emission intensity vs.
concentration is not much linear.
5. Absorption intensity vs concentration of
analyte is much linear.
6.Atomization and excitation flame used. 6. Atomization flame used.
7. Intensity vs concentration data is obtained. 7. Absorbance vs concentration data is
obtained
8. Limited to alkali and alkali earth metals. 8. Useful for more than 70 metals.
29
References
 G. Gauglitz and T. Vo-Dinh “Handbook of spectroscopy” pg 421-493
 J. Michael Hollas “Modern Spectroscopy” 4th Ed, pg 64-65
 https://socratic.org/questions/what-is-the-principle-of-flame-emission-spectroscopy
 http://vlab.amrita.edu/?sub=2&brch=294&sim=1351&cnt=1
 https://www.britannica.com/science/flame-emission-spectroscopy
 https://chem.libretexts.org/Textbook_Maps/Analytical_Chemistry_Textbook_Maps/
Map%3A_Analytical_Chemistry_2.0_(Harvey)/10_Spectroscopic_Methods/10.4%3
A_Atomic_Absorption_Spectroscopy
 https://web.nmsu.edu/~esevosti/report.htm
30
Thank you
31

Weitere ähnliche Inhalte

Was ist angesagt?

Was ist angesagt? (20)

PRINCIPLES of FT-NMR & 13C NMR
PRINCIPLES of FT-NMR & 13C NMRPRINCIPLES of FT-NMR & 13C NMR
PRINCIPLES of FT-NMR & 13C NMR
 
Nmr instrumentation
Nmr  instrumentationNmr  instrumentation
Nmr instrumentation
 
Thermogravimetric analysis ppt
Thermogravimetric analysis pptThermogravimetric analysis ppt
Thermogravimetric analysis ppt
 
C 13 NMR Spectroscopy
C 13 NMR SpectroscopyC 13 NMR Spectroscopy
C 13 NMR Spectroscopy
 
Differential Scanning Calorimeter Instrumentation.(DSC)
Differential Scanning Calorimeter Instrumentation.(DSC)Differential Scanning Calorimeter Instrumentation.(DSC)
Differential Scanning Calorimeter Instrumentation.(DSC)
 
Ft ir instrumentation
Ft ir instrumentationFt ir instrumentation
Ft ir instrumentation
 
Mass spectroscopy, Ionization techniques and types of mass analyzers
Mass spectroscopy, Ionization techniques and types of mass analyzers Mass spectroscopy, Ionization techniques and types of mass analyzers
Mass spectroscopy, Ionization techniques and types of mass analyzers
 
Solvents and solvent effect in UV - Vis Spectroscopy, By Dr. Umesh Kumar sh...
Solvents and  solvent effect in UV -  Vis Spectroscopy, By Dr. Umesh Kumar sh...Solvents and  solvent effect in UV -  Vis Spectroscopy, By Dr. Umesh Kumar sh...
Solvents and solvent effect in UV - Vis Spectroscopy, By Dr. Umesh Kumar sh...
 
Mass spectrometry
Mass spectrometryMass spectrometry
Mass spectrometry
 
Flame emission spectroscopy
Flame emission spectroscopyFlame emission spectroscopy
Flame emission spectroscopy
 
Fragmentation techniques
Fragmentation techniquesFragmentation techniques
Fragmentation techniques
 
Plate theory of Chromatography
Plate theory of ChromatographyPlate theory of Chromatography
Plate theory of Chromatography
 
NMR SPECTROSCOPY
NMR SPECTROSCOPYNMR SPECTROSCOPY
NMR SPECTROSCOPY
 
Uv spectroscopy
Uv  spectroscopyUv  spectroscopy
Uv spectroscopy
 
C-13 NMR Spectroscopy
C-13 NMR SpectroscopyC-13 NMR Spectroscopy
C-13 NMR Spectroscopy
 
NMR Spectroscopy
NMR SpectroscopyNMR Spectroscopy
NMR Spectroscopy
 
Fragmentation rules mass spectroscopy
Fragmentation rules mass spectroscopyFragmentation rules mass spectroscopy
Fragmentation rules mass spectroscopy
 
IR - Spectroscopy, theory, modes of vibration and sample handling. by Dr. Ume...
IR - Spectroscopy, theory, modes of vibration and sample handling. by Dr. Ume...IR - Spectroscopy, theory, modes of vibration and sample handling. by Dr. Ume...
IR - Spectroscopy, theory, modes of vibration and sample handling. by Dr. Ume...
 
Fragmentation techniques in mass spectroscopy
Fragmentation techniques in mass spectroscopyFragmentation techniques in mass spectroscopy
Fragmentation techniques in mass spectroscopy
 
Infrared spectroscopy
Infrared spectroscopyInfrared spectroscopy
Infrared spectroscopy
 

Ähnlich wie Flame emission & atomic absorption spectroscopy

Mridu- AAS.pptx
Mridu- AAS.pptxMridu- AAS.pptx
Mridu- AAS.pptx
SweetuCutie
 
Flame emission spectroscopy and atomic absorption spectroscopy ppt
Flame emission spectroscopy and atomic absorption spectroscopy pptFlame emission spectroscopy and atomic absorption spectroscopy ppt
Flame emission spectroscopy and atomic absorption spectroscopy ppt
Sachin G
 
Atomic absorption spectroscopy (AAS).pptx
Atomic absorption spectroscopy (AAS).pptxAtomic absorption spectroscopy (AAS).pptx
Atomic absorption spectroscopy (AAS).pptx
Priyankashah645680
 
Atomic absorption-spectrometry
Atomic absorption-spectrometryAtomic absorption-spectrometry
Atomic absorption-spectrometry
MUsama10
 

Ähnlich wie Flame emission & atomic absorption spectroscopy (20)

Aas by i k dhoot
Aas by i k  dhootAas by i k  dhoot
Aas by i k dhoot
 
Atomic absorption & flame emissionn.pptx
Atomic absorption & flame emissionn.pptxAtomic absorption & flame emissionn.pptx
Atomic absorption & flame emissionn.pptx
 
Absorption and emission spectroscopy
Absorption and emission spectroscopyAbsorption and emission spectroscopy
Absorption and emission spectroscopy
 
Mridu- AAS.pptx
Mridu- AAS.pptxMridu- AAS.pptx
Mridu- AAS.pptx
 
Lecture 3. Atomic-absorption spectrophotometry.pptx
Lecture 3. Atomic-absorption spectrophotometry.pptxLecture 3. Atomic-absorption spectrophotometry.pptx
Lecture 3. Atomic-absorption spectrophotometry.pptx
 
Flame emission Spectroscopy SlideShare mineeta mahra
Flame emission Spectroscopy SlideShare mineeta mahraFlame emission Spectroscopy SlideShare mineeta mahra
Flame emission Spectroscopy SlideShare mineeta mahra
 
Aas and fes
Aas and fesAas and fes
Aas and fes
 
AAS interference.pptx
AAS interference.pptxAAS interference.pptx
AAS interference.pptx
 
Anta sharma
Anta sharmaAnta sharma
Anta sharma
 
Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)
 
Flame emission spectroscopy and atomic absorption spectroscopy ppt
Flame emission spectroscopy and atomic absorption spectroscopy pptFlame emission spectroscopy and atomic absorption spectroscopy ppt
Flame emission spectroscopy and atomic absorption spectroscopy ppt
 
Atomic absorption & flame emission spectrophotometry by Dr. Anurag Yadav
Atomic absorption & flame emission spectrophotometry by Dr. Anurag YadavAtomic absorption & flame emission spectrophotometry by Dr. Anurag Yadav
Atomic absorption & flame emission spectrophotometry by Dr. Anurag Yadav
 
Instru booklet (1)
Instru booklet (1)Instru booklet (1)
Instru booklet (1)
 
APA seminar
APA seminarAPA seminar
APA seminar
 
Instru booklet
Instru bookletInstru booklet
Instru booklet
 
Atomic absorption spectroscopy (AAS).pptx
Atomic absorption spectroscopy (AAS).pptxAtomic absorption spectroscopy (AAS).pptx
Atomic absorption spectroscopy (AAS).pptx
 
Atomic absorption spectroscopy
Atomic absorption spectroscopy Atomic absorption spectroscopy
Atomic absorption spectroscopy
 
Atomic absorption-spectrometry
Atomic absorption-spectrometryAtomic absorption-spectrometry
Atomic absorption-spectrometry
 
Atomic absorption-spectrometry
Atomic absorption-spectrometryAtomic absorption-spectrometry
Atomic absorption-spectrometry
 
Atomic absorption spectroscopy lec
Atomic absorption spectroscopy lecAtomic absorption spectroscopy lec
Atomic absorption spectroscopy lec
 

Mehr von Himal Barakoti

Mehr von Himal Barakoti (10)

Monoclonal antibodies
Monoclonal antibodiesMonoclonal antibodies
Monoclonal antibodies
 
Challenges in formulating Herbal cosmetics
Challenges in formulating Herbal cosmeticsChallenges in formulating Herbal cosmetics
Challenges in formulating Herbal cosmetics
 
Computational modeling in drug disposition
Computational modeling in drug dispositionComputational modeling in drug disposition
Computational modeling in drug disposition
 
Oligonucleotide
OligonucleotideOligonucleotide
Oligonucleotide
 
Spin spin coupling and coupling constant
Spin spin coupling and coupling constantSpin spin coupling and coupling constant
Spin spin coupling and coupling constant
 
Self micro-emulsifying drug delivery system (SMEDDS)
Self micro-emulsifying drug delivery system (SMEDDS)Self micro-emulsifying drug delivery system (SMEDDS)
Self micro-emulsifying drug delivery system (SMEDDS)
 
Regulatory requirement of EU, MHRA and TGA
Regulatory requirement of EU, MHRA and TGARegulatory requirement of EU, MHRA and TGA
Regulatory requirement of EU, MHRA and TGA
 
Layout of pharmaceutical building and services
Layout of pharmaceutical building and servicesLayout of pharmaceutical building and services
Layout of pharmaceutical building and services
 
Investigation of medicinal product dossier (IMPD)
Investigation of medicinal product dossier (IMPD)Investigation of medicinal product dossier (IMPD)
Investigation of medicinal product dossier (IMPD)
 
Cmc, post approval and regulation
Cmc, post approval and regulationCmc, post approval and regulation
Cmc, post approval and regulation
 

Kürzlich hochgeladen

🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
Call Girls In Delhi Whatsup 9873940964 Enjoy Unlimited Pleasure
 
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
9953056974 Low Rate Call Girls In Saket, Delhi NCR
 
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
chetankumar9855
 

Kürzlich hochgeladen (20)

Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
 
Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...
Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...
Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
 
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
 
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
 
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
 
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
 
Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...
Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...
Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...
 
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
 
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
 
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
 
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
 
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
 
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
 
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
 
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
 
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
 
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
 

Flame emission & atomic absorption spectroscopy

  • 1. Flame emission and Atomic absorption Spectroscopy Presented by: Guided by: Himal Barakoti Arup Chakroborty M. Pharm, 1st Sem Assistant Professor Department of Pharmacy Department of Pharmacy Assam Down Town University Assam Down Town University
  • 2. Contents:  Introduction  Theory and principle of Flame emission Spectroscopy  Instrumentation  Interferences  Application  Atomic Absorption Spectroscopy  Theory and Principle 2
  • 3. Contents:  Instrumentation  Interferences  Application  Comparison between Flame emission and Atomic absorption Spectroscopy  References 3
  • 4. Introduction  Atomic spectroscopy is thought to be the oldest instrumental method for the determination of elements.  These techniques are introduced in the mid of 19th Century during which Gustav Kirchhoff and Robert Bunsen showed that the radiation emitted from the flames depends on the characteristic element present in the flame.  The developments in the instrumentation area led to the widespread application of atomic spectroscopy.  The absorption and emission of radiant energy by atoms provide powerful analytical tools for both quantitative and qualitative analysis of substance. 4
  • 5. Flame emission Spectroscopy  Flame emission spectroscopy (FES) is a method of chemical analysis that uses intensity of light emitted from flame, arc or spark at particular wavelength to determine quantity of element in sample.  The basis of flame photometric working is that, the species of alkali metals (Group 1) and alkaline earth metals (Group II) metals are dissociated due to the thermal energy provided by the flame source.  Due to this thermal excitation, some of the atoms are excited to a higher energy level where they are not stable.  The absorbance of light due to the electrons excitation can be measured by using the direct absorption techniques. 5
  • 6.  The subsequent loss of energy will result in the movement of excited atoms to the low energy ground state with emission of some radiations, which can be visualized in the visible region of the spectrum.  The intensity of radiation emitted by these excited atoms returning to the ground state provides the basic for analytical determination in FES.  The wavelength of emitted light is specific for specific elements.  In flame emission spectroscopy, o Wavelength of spectral lines give identity of elements. o Intensity of emitted light is directly proportional to the number of atoms present. 6
  • 7. Instrumentation  Flame photometers are the simplest type of atomic spectrometers.  Solution is introduced into a fine spray.  Solvents evaporates leaving dehydrated salt.  Certain fraction of atoms absorbs energy and are raised to excited state.  These excited atoms on returning to ground state emits photons of certain wavelength.  Flame emission passes through monochromator which filters all emitted light expect the wavelength of our interest.  Photoelectric Detector measures the intensity of filtered light. 7
  • 8. 1. Atomizer/Nebulizer  Process of conversion of sample to a fine mist of finely divided droplets using a jet of compressed gas. o Pneumatic nebulizers o Electro thermal vaporizer o Ultrasound nebulizer 2. Burner  They must have ability to evaporate the liquid droplets from sample solution and capacity to excite atoms formed and cause them to emit radiant energy. o Mecker burner o Total consumption burner o Laminar flow burner o Shileded burner o Lundergraph burner 8
  • 9. 3. Monochromator (Grating or prism monochromator) They disperse radiation coming from the flame and falling on it. Dispersed radiation goes to detector from exit slit. Filters chosen has wavelength range transparent to emission from element of interest. 4. Detectors (photoemissive cells or photomultiplier tubes) They measure intensity of radiation falling on it 5. Amplifier & readout devices It suitably amplify and display detector’s output. Amplifier sensativity can be changed to be able to analyze sample of varying concentration. 9
  • 10. Types of flame used The most common instrument uses air as oxidant. The temperature of the flames produced is relatively low so the technique is only suitable for the elements that are easily excited such as alkali and alkali earth elements. 10 Flame Temp/˚C Gas/Air 1700-1900 Gas/O2 2700-2800 H2/Air 2000-2100 H2/O2 2550-2700 C2H2/Air 2100-2400 C2H2/O2 3050-3150 C2H2/N2O 2600-2800
  • 11.  A higher temperature will tend to increase the number of atoms in the excited state and hence the signal.  Some detection limits: 11 Elements Spectral lines/nm Flame Detection limits/ppm Al 396 C2H2/N2O 0.1 Ba 553 C2H2/N2O 0.01 K 766 C2H2/O2 0.001 Li 671 C2H2/N2O 0.0001
  • 12. Interferences  Number of factors beside the analyte affect the intensity of emitted radiation. The analytical signals often include contributions from constituents other than analyte termed as interferences and are found to interfere the outcome of procedure.  3 types: 1. Spectral Interferences: interference that affect spectral intensity or resolution. a. 1st type of interference arises when two elements exhibit spectra, which partially overlap, and both emit radiation at particular wavelength. The detector cannot distinguish between the source of radiation and records the total sign, thus resulting in incorrect answer. This type of interferences are more common in high flame temperature as numerous spectral lines are produced in high temperature. E.g., Fe line at 324.73 nm overlaps with Cu line at 324.75 nm. This Can be overcome by taking alternative wavelength with no overlap. 12
  • 13. b. 2nd type of spectral interference deals with spectral lines of two or more elements which are close but their spectra do not overlap. The filter may allow spectral lines separated by 5.0-10.0 nm to pass through, thus resulting in error in analysis. This can be reduced by increasing the resolution of spectral isolation system. c. 3rd type of spectral interference occurs due to presence of continuous background which arises due to high concentration of salts in the sample. Some organic solvents also produce continuous background. This can be corrected by using suitable scanning technique. 2. Ionization Interference: In certain cases, high temperature flame may cause ionization of the metal atoms. E.g., In case of sodium Na Na+ + e- The Na+ ion possesses an emission spectrum of its own frequencies, which are different from those of Na atom. This reduces the radiant power of atomic emission. This type of interference can be eliminated by addition of large quantity of potassium salt to standard and sample solution. Potassium itself undergoes ionization due to low ionization energy and suppresses the ionization of sodium. 13
  • 14. 3. Chemical Interferences: This arises out of reaction between different interferents and the analyte. a. Cation-cation interference: The presence of certain anions, such as oxalate, phosphate, sulphate and aluminate in a solution may affect the intensity of radiation emitted by an element, resulting in serious analytical error. For e.g., calcium in presence of phosphate ion forms a stable substance as Ca3(PO4)2 which does not decompose easily, resulting in production of lesser atom. Thus calcium signal is depressed. This can be removed either by extraction of anion or by using calibration curve prepared from standard solutions containing same concentrations of the anion as found in the sample. b. Cation-cation interference: In many cases, mutual interferences of cations results in reducing element signal intensity. These interference are neither spectral nor ionic in nature and the actual mechanism has not been well understood. E.g., Al interferes with Ca and Mg. Na interferes with Mg. c. Oxide formation: arises due to formation of stable metal oxide if oxygen is present in flame, resulting in reduced signal intensity. E.g., Alkaline earth metal. This can be eliminated by either using high flame temperature to dissociate the oxides or by using oxygen-deficient environment to produce excited atom. 14
  • 15. Limitations  As natural gas and air flame is employed for excitation, the temperature is not high enough to excite transition metals, therefore the method is selective towards detection of alkali and alkaline earth metals.  Low temperature makes this method susceptible to certain disadvantages, most of them related to flame stability and aspiration conditions. Fuel and oxidation flow rates and purity, aspiration rates, solution viscosity affects these.  FES is a means of determining the total metal concentration of a sample; it tells us nothing about the molecular form of the metal in original sample.  Only liquid samples can be used. 15
  • 16. Applications  Flame Photometers are widely used in quality control where a simple and quick determination of alkali or alkali earth metal is required. They have the advantage of being significantly lower priced than most other atomic spectrometers.  Biological/medical applications- notable applications are the determinations of Na, K, Ca and Mg in body fluids and other biological samples.  Food industry – Determination of calcium and iron in beer 16
  • 17. Atomic Absorption Spectroscopy  Guystav Kirchoff and Robert Bunsen first used atomic absorption spectroscopy—along with atomic emission—in 1859 and 1860 as a means for identify atoms in flames and hot gases.  Modern atomic absorption spectroscopy has its beginnings in 1955 as a result of the independent work of Alan. C. Walsh and C. T. J. Alkemade. 17
  • 18.  Atomic absorption spectroscopy (AAS) is an absorption spectroscopic method where radiation from a source is absorbed by non-excited atoms in vapour state.  AAS deals with the absorption of specific wavelength of radiation by neutral atoms in the ground state.  Reliable for detecting over 70 elements with metals and metalloids. 18
  • 19. Theory and Principle  The technique uses basically the principle that free atoms (gas) generated in an atomizer can absorb radiations at specific frequency.  AAS quantifies the absorption of ground state atoms in the gaseous form.  Atoms absorb UV/visible light and make transition to higher electronic energy levels. The analyte concentration is determined from the amount of absorption.  Concentration measurement are usually determined from a working curve after calibrating instrument with standard known concentration. Liquid sample Formation of droplets Fine residue Formation of neutral ion Neutral atom absorbs specific wavelength of radiation from hollow cathode lamp Measurement of intensity of radiation absorbed by the neutral atom using the detector 19
  • 20.  The extent to which radiation of a particular frequency is absorbed by an atomic vapour is related to the length of the path transverse and to the concentration of absorbing atoms in the vapour.  This is analogues to the Beer-Lamberts law relating to samples in solution. Thus, for a collimated monochromatic beam of radiation of incident Io passing through an atomic vapour of thickness I Iv = Io eл-kvI Where, Iv is the intensity of the transmitted radiation Kv is absorption co efficient 20
  • 22. 1. Sharp line radiation source (hollow cathode lamp)  Hollow cathode lamps are most commonly used radiation containing tungsten anode and hollow cylindrical cathode made of element to be determined.  These are sealed with gas tube filled with an inert gas.  When a current flows between the anode and cathode in this lamps, metal atoms are sputtered from the cathode cup, and collisions occur with the filler gas. A number of metal atoms become excited and give off their characteristics radiation. 22
  • 23. 2. A solution Nebulizer and burner  Nebulizer suck up liquid samples at controlled rate and create a fine aerosol spray for introduction into flame.  Aerosol, fuel and oxidant are thoroughly mixed for introduction into flame.  Atomization is carried out by separation of particles into individual molecules and breaking molecules into atoms. This is done by exposing analyte to high temperature in flame or graphite furnace.  In graphite furnace, samples are deposited in a small graphite coated tube which can then be heated to vaporize and atomize the analyte.  Graphite tubes are heated using a high current power supply. 23
  • 24. 3. Monochromator  They are used to select specific wavelength of light which s absorbed by the sample, and to exclude other wavelengths.  Selections of specific light allows the determination of selected element in the presence of others. 4. Detectors  The light selected by monochromator is directed onto a detector that is typically a photomultiplier tube which convert the light signal into electrical signal proportional to the light intensity.  The signal is amplified, displayed for readout of fed to data station for print out. 24
  • 25. Interferences  Spectral Interferences: Spectral interferences are caused by presence of another atomic absorption line or a molecular absorbance band close to the spectral line of element of interest.  Chemical interference: If a sample contains a species which forms a thermally stable compound with the analyte that is not completely decomposed by the energy available in the flame then chemical interference exists. Refractory elements such as Ti, W, Zr, Mo and Al may combine with oxygen to form thermally stable oxides. Analysis of such elements can be carried out at higher flame temperatures using nitrous oxide – acetylene flame instead of air-acetylene to provide higher dissociation energy. 25
  • 26.  Matrix interference: When a sample is more viscous or has different surface tension than the standard it can result in differences in sample uptake rate due to changes in nebulization efficiency. Such interferences are minimized by matching as closely as possible the matrix composition of standard and sample.  Ionization interference: Excess energy of the flame can lead to excitation of ground state atoms to ionic state by loss of electrons thereby resulting in depletion of ground state atoms. 26
  • 27. limitations  Expensive  Low precision  Low sample throughput  Requires high level of operator skill  Sample must be in solution or at least volatile  Individual source lamps required for each element 27
  • 28. Applications  Determination of metal at trace level in solution.  Determination of purity: o Presence of heavy metal in body fluids o Pollution of water by metals o Food stuffs o Soft drinks and beer o Analysis of Geochemical exploration for mineral o Soils, crude oils, petroleum products, plastics. 28
  • 29. Comparison between FES and AAS Flame Emission spectroscopy Atomic Absorption spectroscopy 1. Amount of light emitted by excited atom is measured. 1. Amount of light absorbed by ground state atom is measured. 2. Absorption intensity and signal response greatly influenced by temperature variation. 2. Absorption intensity and signal response does not depend upon temperature. 3. Beer’s law is not obeyed. 3. Beers law is obeyed. 4. Intensity of emitted radiation is directly proportional to the number of atoms in excited state. 4. Intensity of absorbed radiation is directly proportional to the number of atoms in ground state. 5. Relation between emission intensity vs. concentration is not much linear. 5. Absorption intensity vs concentration of analyte is much linear. 6.Atomization and excitation flame used. 6. Atomization flame used. 7. Intensity vs concentration data is obtained. 7. Absorbance vs concentration data is obtained 8. Limited to alkali and alkali earth metals. 8. Useful for more than 70 metals. 29
  • 30. References  G. Gauglitz and T. Vo-Dinh “Handbook of spectroscopy” pg 421-493  J. Michael Hollas “Modern Spectroscopy” 4th Ed, pg 64-65  https://socratic.org/questions/what-is-the-principle-of-flame-emission-spectroscopy  http://vlab.amrita.edu/?sub=2&brch=294&sim=1351&cnt=1  https://www.britannica.com/science/flame-emission-spectroscopy  https://chem.libretexts.org/Textbook_Maps/Analytical_Chemistry_Textbook_Maps/ Map%3A_Analytical_Chemistry_2.0_(Harvey)/10_Spectroscopic_Methods/10.4%3 A_Atomic_Absorption_Spectroscopy  https://web.nmsu.edu/~esevosti/report.htm 30