HPLC Problems, Probable Causes, and
Remedies
Problem No. 1: No Peaks/Very Small Peaks
• Probable Cause
1) Detector lamp off.
2) Loose/broken wire between detector and integrator or recorder.
3) No mobile phase flow.
4) No sample/deteriorated sample/ wrong sample.
5) Settings too high on detector or recorder.
• Remedy/Comments
1) Turn lamp on.
2) Check electrical connections and cables.
3) Pump off/ valve (knob) open
4) Be sure automatic sampler vials have sufficient liquid and no air bubbles in the
sample. Evaluate system performance with fresh standard to confirm sample as
source of problem.
5) Check attenuation or gain settings. Check lamp status. Auto-zero if necessary.
Problem No. 2: No Flow
• Probable Cause
1) Pump off.
2) Flow interrupted/obstructed.
3) Leak.
4) Air trapped in pump head. (Revealed by pressure fluctuations.)
• Remedy/Comments
1) Start pump.
2) Check mobile phase level in reservoir(s). Check flow throughout system. Examine sample loop for
obstruction or air lock. Make sure mobile phase components are miscible and mobile phase is properly
degassed.
3) Check system for loose fittings. Check pump for leaks, salt buildup, unusual noises. Change pump seals if
necessary.
4) Disconnect tubing at guard column (if present) or analytical column inlet. Check for flow. Purge pump at high
flow rate (e.g., 5-10 mL/min.), prime system if necessary. (Prime each pump head separately.) If system has
check valve, loosen valve to allow air to escape. If problem persists, flush system with 100% methanol or
isopropanol. If problem still persists, contact system manufacturer.
Problem No. 3: No Pressure/Pressure Lower
• Probable Cause
1) Leak.
2) Mobile phase flow interrupted/obstructed.
3) Air trapped in pump head. (Revealed by pressure fluctuations.)
4) Leak at column inlet end fitting.
5) Air trapped elsewhere in system.
6) Worn pump seal causing leaks around pump head.
7) Faulty check valve.
8) Faulty pump seals.
• Remedy/Comments
1) Check system for loose fittings. Check pump for leaks, salt buildup, unusual noises. Change pump seals if necessary.
2) Check mobile phase level in reservoir(s). Check flow throughout system. Examine sample loop for obstruction or air lock.
Make sure mobile phase components are miscible and mobile phase is properly degassed.
3) Disconnect tubing at guard column (if present) or analytical column inlet. Check for flow. Purge pump at high flow rate
(e.g., 10 mL/min.), prime system if necessary. (Prime each pump head separately.) If system has check valve, loosen valve
to allow air to escape.
4) Reconnect column and pump solvent at double the flow rate. If pressure is still low, check for leaks at inlet fitting or
column end fitting.
• Disconnect guard and analytical column and purge system. Reconnect column(s). If problem persists, flush system with
100% methanol or isopropanol.
• Replace seal. If problem persists, replace piston and seal.
• Rebuild or replace valve.
• Replace seals.
Problem No. 4: Pressure Higher
• Probable Cause
• Problem in pump, injector, in-line filter, or tubing.
• Obstructed guard column or analytical column.
• Remedy/Comments
• Remove guard column and analytical column from system. Replace with unions and 0.010'' I.D. or larger tubing to reconnect
injector to detector. Run pump at 2-5 mL/min. If pressure is minimal, see Cause 2. If not, isolate cause by systematically
eliminating system components, starting with detector, then in-line filter, and working back to pump. Replace filter in pump
if present.
• Remove guard column (if present) and check pressure. Replace guard column if necessary. If analytical column is
obstructed, reverse and flush the column, while disconnected from the detector. If problem persists, column may be
clogged with strongly retained contaminants. Use appropriate restoration procedure (Table 1). If problem still
persists, change inlet frit or replace column.
Problem No. 5: Variable Retention Times
• Probable Cause
1) Leak.
2) Change in mobile phase composition. (Small changes can lead to large changes in retention times.)
3) Air trapped in pump. (Retention times increase and decrease at random times.)
4) Column temperature fluctuations (especially evident in ion exchange systems).
5) Column overloading. (Retention times usually decrease as mass of solute injected on column exceeds column capacity.)
6) Sample solvent incompatible with mobile phase.
7) Column problem. (Not a common cause of erratic retention. As a column ages, retention times gradually decrease.)
• Remedy/Comments
1) Check system for loose fittings. Check pump for leaks, salt buildup, unusual noises. Change pump seals if necessary.
2) Check make-up of mobile phase. If mobile phase is machine mixed using proportioning values, hand mix and supply from
one reservoir.
3) Purge air from pump head or check valves. Change pump seals if necessary. Be sure mobile phase is degassed.
4) Use reliable column oven. (Note: higher column temperatures increase column efficiency. For optimum results, heat
eluant before introducing it onto column.)
5) Inject smaller volume (e.g., 10 μL vs. 100 μL) or inject the same volume after 1:10 or 1:100 dilutions of sample.
6) Adjust solvent. Whenever possible, inject samples in mobile phase.
7) Substitute new column of same type to confirm column as cause. Discard old column if restoration procedures fail.
Problem No. 6: Loss of Resolution
• Probable Cause
1) Mobile phase contaminated/deteriorated (causing retention times and/or selectivity to change).
2) Obstructed guard or analytical column.
• Remedy/Comments
1) Prepare fresh mobile phase.
2) Remove guard column (if present) and attempt analysis. Replace guard column if necessary. If analytical column is
obstructed, reverse and flush. If problem persists, column may be clogged with strongly retained contaminants. Use
appropriate restoration procedure (Table 1). If problem still persists, change inlet frit or replace column.
Problem No. 7: Split Peaks
• Probable Cause
1) Contamination on guard or analytical column inlet.
2) Partially blocked frit.
3) Small (uneven) void at column inlet.
4) Sample solvent incompatible with mobile phase.
• Remedy/Comments
1) Remove guard column (if present) and attempt analysis. Replace guard column if necessary. If analytical column is
obstructed, reverse and flush. If problem persists, column may be clogged with strongly retained contaminants. Use
appropriate restoration procedure (Table1 ). If problem still persists, inlet frit is probably (partially) plugged. Change
frit or replace column.
2) Replace frit (see above).
3) Repack top of column with pellicular particles of same bonded phase functionality. Continue using the column in reverse
flow direction.
4) Adjust solvent. Whenever possible, inject samples in mobile phase.
Problem No. 8: Tailing Peaks
• Probable Cause
1) Guard or analytical column contaminated/worn out.
2) Mobile phase contaminated/deteriorated.
3) Interfering components in sample.
• Remedy/Comments
1) Remove guard column (if present) and attempt analysis. Replace guard column if necessary. If analytical column is source
of problem, use appropriate restoration procedure (Table 1). If problem persists, replace column.
2) Check make-up of mobile phase.
3) Check column performance with standards.
Problem No. 9: Fronting Peaks
• Probable Cause
1) Column overloaded.
2) Sample solvent incompatible with mobile phase.
3) Shoulder or gradual baseline rise before a main peak may be another sample component.
• Remedy/Comments
1) Inject smaller volume (e.g., 10 μL vs. 100 μL). Dilute the sample 1:10 or 1:100 fold in case of mass overload.
2) Adjust solvent. Whenever possible, inject samples in mobile phase. Flush polar bonded phase column with 50 column
volumes HPLC grade ethyl acetate at 2-3 times the standard flow rate, then with intermediate polarity solvent prior to
analysis.
3) Increase efficiency or change selectivity of system to improve resolution. Try another column type if necessary (e.g.,
switch from nonpolar C18 to polar cyano phase).
Problem No. 10: Baseline Noise (regular)
• Probable Cause
1) Air in mobile phase, detector cell, or pump.
2) Pump pulsations.
3) Incomplete mobile phase mixing.
4) Temperature effect (column at high temperature, detector unheated).
5) Other electronic equipment on same line.
6) Leak.
• Remedy/Comments
1) Degas mobile phase. Flush system to remove air from detector cell or pump.
2) Incorporate pulse damper into system.
3) Mix mobile phase by hand or use less viscous solvent.
4) Reduce differential or add heat exchanger.
5) Isolate LC, detector, recorder to determine if source of problem is external. Correct as necessary.
6) Check system for loose fittings. Check pump for leaks, salt buildup, unusual noises. Change pump seals if necessary.
Problem No. 11: Negative Peak
• Probable Cause
1) Recorder leads reversed.
2) Refractive index of solute less than that of mobile phase (RI detector).
3) Sample solvent and mobile phase differ greatly in composition (vacancy peaks).
4) Mobile phase more absorptive than sample components to UV wavelength.
• Remedy/Comments
1) Check polarity.
2) Use mobile phase with lower refractive index, or reverse recorder leads.
3) Adjust or change sample solvent. Dilute sample in mobile phase whenever possible.
4) a. Change polarity when using indirect UV detection, or
b. Change UV wavelength or use mobile phase that does not adsorb chosen wavelength.
Problem No. 12: Ghost Peak
• Probable Cause
1) Contamination in injector or column.
2) Late eluting peak (usually broad) present in sample.
• Remedy/Comments
1) Flush injector between analyses (a good routine practice). If necessary, run strong solvent through column
to remove late eluters. Include final wash step in gradient analyses, to remove strongly retained compounds.
2) a. Check sample preparation.
b. Include (step) gradient to quickly elute component.
Table-1
• Table-1. Column Restoration Procedures
• Silica-Based Reversed Phase Column
(alkyl (C8, C18, etc.), phenyl, or diphenyl column, SUPELCOSIL LC-PAH column)
•A. Water Soluble Samples Flush with the following:
• Flush with warm (60 °C) distilled water
• 50 mL methanol
• 50 mL acetonitrile
• 25 mL methanol
• 25 mL mobile phase
• Evaluate column performance.
•B. Samples Not Soluble in Water Flush with the
following:
• 50 mL 2-propanol
• 50 mL methylene chloride
• 50 mL hexane
• 25 mL isopropanol
• 25 mL mobile phase
• Evaluate column performance.
THANK YOU

Weitere ähnliche Inhalte

Was ist angesagt?

Gas chromatography 2
Gas chromatography 2Gas chromatography 2
Gas chromatography 2
Afia Riaz
 
Sample injection system in hplc
Sample injection system in hplcSample injection system in hplc
Sample injection system in hplc
Ram Mohan S R
 

Was ist angesagt? (20)

High Performance Thin Layer Chromatography.pptx
High Performance Thin Layer Chromatography.pptxHigh Performance Thin Layer Chromatography.pptx
High Performance Thin Layer Chromatography.pptx
 
Stationary and mobile_phase_selection_m_ab_ph_gradient_analysis_33974
Stationary and mobile_phase_selection_m_ab_ph_gradient_analysis_33974Stationary and mobile_phase_selection_m_ab_ph_gradient_analysis_33974
Stationary and mobile_phase_selection_m_ab_ph_gradient_analysis_33974
 
Chromatographic Column chemistry
Chromatographic Column chemistryChromatographic Column chemistry
Chromatographic Column chemistry
 
Advantages of Liquid Liquid Extraction System
Advantages of Liquid Liquid Extraction SystemAdvantages of Liquid Liquid Extraction System
Advantages of Liquid Liquid Extraction System
 
Hplc
HplcHplc
Hplc
 
Gas chromatography 2
Gas chromatography 2Gas chromatography 2
Gas chromatography 2
 
HPTLC- Principle, Instrumentation and Software (Abhishek Gupta)
HPTLC- Principle, Instrumentation and Software (Abhishek Gupta)HPTLC- Principle, Instrumentation and Software (Abhishek Gupta)
HPTLC- Principle, Instrumentation and Software (Abhishek Gupta)
 
DERIVATIZATION IN GAS CHROMATOGRAPHY (GC), HIGHPERFORMANCE LIQUID CHROMATOGRA...
DERIVATIZATION IN GAS CHROMATOGRAPHY (GC), HIGHPERFORMANCE LIQUID CHROMATOGRA...DERIVATIZATION IN GAS CHROMATOGRAPHY (GC), HIGHPERFORMANCE LIQUID CHROMATOGRA...
DERIVATIZATION IN GAS CHROMATOGRAPHY (GC), HIGHPERFORMANCE LIQUID CHROMATOGRA...
 
Sample injection system in hplc
Sample injection system in hplcSample injection system in hplc
Sample injection system in hplc
 
HPLC BY MANOJ KUMAR.M
HPLC BY MANOJ KUMAR.MHPLC BY MANOJ KUMAR.M
HPLC BY MANOJ KUMAR.M
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
Thin Layer Chromatography
Thin Layer ChromatographyThin Layer Chromatography
Thin Layer Chromatography
 
Gas Chromatography
Gas ChromatographyGas Chromatography
Gas Chromatography
 
Gas Chromatography by Dr. A. Amsavel
Gas Chromatography  by Dr. A. AmsavelGas Chromatography  by Dr. A. Amsavel
Gas Chromatography by Dr. A. Amsavel
 
Column in gas chromatography
Column in gas chromatographyColumn in gas chromatography
Column in gas chromatography
 
Chromarogaphy system suitability ppt
Chromarogaphy system suitability pptChromarogaphy system suitability ppt
Chromarogaphy system suitability ppt
 
Hptlc vs tlc
Hptlc vs tlcHptlc vs tlc
Hptlc vs tlc
 
Gas chromatography by Dr. Anurag Yadav
Gas chromatography by Dr. Anurag YadavGas chromatography by Dr. Anurag Yadav
Gas chromatography by Dr. Anurag Yadav
 
High performance liquid chromatography
High performance liquid chromatographyHigh performance liquid chromatography
High performance liquid chromatography
 
Pumps
PumpsPumps
Pumps
 

Ähnlich wie HPLC Problems, Probable Causes, and Remedies2020.pptx

HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...
Vandana Devesh Sharma
 
ICS 5000+ Trouble-shooting guide
ICS 5000+ Trouble-shooting guideICS 5000+ Trouble-shooting guide
ICS 5000+ Trouble-shooting guide
Elliot Morgan
 

Ähnlich wie HPLC Problems, Probable Causes, and Remedies2020.pptx (20)

HPLC_basics.ppt
HPLC_basics.pptHPLC_basics.ppt
HPLC_basics.ppt
 
HPLC
HPLCHPLC
HPLC
 
Hplc
HplcHplc
Hplc
 
hplc-120629111345-phpapp01.ppt
hplc-120629111345-phpapp01.ppthplc-120629111345-phpapp01.ppt
hplc-120629111345-phpapp01.ppt
 
Tips and tricks_hplc_troubleshooting
Tips and tricks_hplc_troubleshootingTips and tricks_hplc_troubleshooting
Tips and tricks_hplc_troubleshooting
 
HPLC Instrumentation & Techniques_basics.ppt
HPLC Instrumentation & Techniques_basics.pptHPLC Instrumentation & Techniques_basics.ppt
HPLC Instrumentation & Techniques_basics.ppt
 
Informative Lecture about Atomic Absorption Spectrometer.pptx
Informative Lecture about Atomic Absorption Spectrometer.pptxInformative Lecture about Atomic Absorption Spectrometer.pptx
Informative Lecture about Atomic Absorption Spectrometer.pptx
 
High performance liquid chromatography
High performance liquid chromatographyHigh performance liquid chromatography
High performance liquid chromatography
 
HPLC.pptx
HPLC.pptxHPLC.pptx
HPLC.pptx
 
New microsoft office power point presentation
New microsoft office power point presentationNew microsoft office power point presentation
New microsoft office power point presentation
 
HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...
 
HPLC
HPLC HPLC
HPLC
 
Hplc.pptx
Hplc.pptxHplc.pptx
Hplc.pptx
 
ICS 5000+ Trouble-shooting guide
ICS 5000+ Trouble-shooting guideICS 5000+ Trouble-shooting guide
ICS 5000+ Trouble-shooting guide
 
Hplc by ikrm khan and groupe
Hplc by ikrm khan and groupeHplc by ikrm khan and groupe
Hplc by ikrm khan and groupe
 
introduccion a los Inyectores para cromatografía de gases
introduccion a los Inyectores para cromatografía de gasesintroduccion a los Inyectores para cromatografía de gases
introduccion a los Inyectores para cromatografía de gases
 
Care and use manual waters x bridge columns
Care and use manual waters x bridge columnsCare and use manual waters x bridge columns
Care and use manual waters x bridge columns
 
Gagandutt sharma(HPLC) [Pharmaceutics].pdf
Gagandutt sharma(HPLC) [Pharmaceutics].pdfGagandutt sharma(HPLC) [Pharmaceutics].pdf
Gagandutt sharma(HPLC) [Pharmaceutics].pdf
 
Gagandutt sharma (HPLC) ..pptx
Gagandutt sharma (HPLC) ..pptxGagandutt sharma (HPLC) ..pptx
Gagandutt sharma (HPLC) ..pptx
 
HPLC by Dr Santosh Gandhi
HPLC by Dr Santosh GandhiHPLC by Dr Santosh Gandhi
HPLC by Dr Santosh Gandhi
 

Kürzlich hochgeladen

➥🔝 7737669865 🔝▻ Muzaffarpur Call-girls in Women Seeking Men 🔝Muzaffarpur🔝 ...
➥🔝 7737669865 🔝▻ Muzaffarpur Call-girls in Women Seeking Men  🔝Muzaffarpur🔝  ...➥🔝 7737669865 🔝▻ Muzaffarpur Call-girls in Women Seeking Men  🔝Muzaffarpur🔝  ...
➥🔝 7737669865 🔝▻ Muzaffarpur Call-girls in Women Seeking Men 🔝Muzaffarpur🔝 ...
amitlee9823
 
Call Girls Chickpet ☎ 7737669865☎ Book Your One night Stand (Bangalore)
Call Girls Chickpet ☎ 7737669865☎ Book Your One night Stand (Bangalore)Call Girls Chickpet ☎ 7737669865☎ Book Your One night Stand (Bangalore)
Call Girls Chickpet ☎ 7737669865☎ Book Your One night Stand (Bangalore)
amitlee9823
 
Escorts Service Arekere ☎ 7737669865☎ Book Your One night Stand (Bangalore)
Escorts Service Arekere ☎ 7737669865☎ Book Your One night Stand (Bangalore)Escorts Service Arekere ☎ 7737669865☎ Book Your One night Stand (Bangalore)
Escorts Service Arekere ☎ 7737669865☎ Book Your One night Stand (Bangalore)
amitlee9823
 
Just Call Vip call girls Shillong Escorts ☎️9352988975 Two shot with one girl...
Just Call Vip call girls Shillong Escorts ☎️9352988975 Two shot with one girl...Just Call Vip call girls Shillong Escorts ☎️9352988975 Two shot with one girl...
Just Call Vip call girls Shillong Escorts ☎️9352988975 Two shot with one girl...
gajnagarg
 
Abortion Pill for sale in Riyadh ((+918761049707) Get Cytotec in Dammam
Abortion Pill for sale in Riyadh ((+918761049707) Get Cytotec in DammamAbortion Pill for sale in Riyadh ((+918761049707) Get Cytotec in Dammam
Abortion Pill for sale in Riyadh ((+918761049707) Get Cytotec in Dammam
ahmedjiabur940
 
Just Call Vip call girls Berhampur Escorts ☎️9352988975 Two shot with one gir...
Just Call Vip call girls Berhampur Escorts ☎️9352988975 Two shot with one gir...Just Call Vip call girls Berhampur Escorts ☎️9352988975 Two shot with one gir...
Just Call Vip call girls Berhampur Escorts ☎️9352988975 Two shot with one gir...
gajnagarg
 
Just Call Vip call girls Begusarai Escorts ☎️9352988975 Two shot with one gir...
Just Call Vip call girls Begusarai Escorts ☎️9352988975 Two shot with one gir...Just Call Vip call girls Begusarai Escorts ☎️9352988975 Two shot with one gir...
Just Call Vip call girls Begusarai Escorts ☎️9352988975 Two shot with one gir...
gajnagarg
 
Hosa Road Call Girls Service: ☎ 7737669865 ☎ High Profile Model Escorts | Ban...
Hosa Road Call Girls Service: ☎ 7737669865 ☎ High Profile Model Escorts | Ban...Hosa Road Call Girls Service: ☎ 7737669865 ☎ High Profile Model Escorts | Ban...
Hosa Road Call Girls Service: ☎ 7737669865 ☎ High Profile Model Escorts | Ban...
amitlee9823
 
怎样办理圣芭芭拉分校毕业证(UCSB毕业证书)成绩单留信认证
怎样办理圣芭芭拉分校毕业证(UCSB毕业证书)成绩单留信认证怎样办理圣芭芭拉分校毕业证(UCSB毕业证书)成绩单留信认证
怎样办理圣芭芭拉分校毕业证(UCSB毕业证书)成绩单留信认证
ehyxf
 
Just Call Vip call girls godhra Escorts ☎️9352988975 Two shot with one girl (...
Just Call Vip call girls godhra Escorts ☎️9352988975 Two shot with one girl (...Just Call Vip call girls godhra Escorts ☎️9352988975 Two shot with one girl (...
Just Call Vip call girls godhra Escorts ☎️9352988975 Two shot with one girl (...
gajnagarg
 
Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...
Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...
Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...
amitlee9823
 
Just Call Vip call girls daman Escorts ☎️9352988975 Two shot with one girl (d...
Just Call Vip call girls daman Escorts ☎️9352988975 Two shot with one girl (d...Just Call Vip call girls daman Escorts ☎️9352988975 Two shot with one girl (d...
Just Call Vip call girls daman Escorts ☎️9352988975 Two shot with one girl (d...
gajnagarg
 
➥🔝 7737669865 🔝▻ Deoghar Call-girls in Women Seeking Men 🔝Deoghar🔝 Escorts...
➥🔝 7737669865 🔝▻ Deoghar Call-girls in Women Seeking Men  🔝Deoghar🔝   Escorts...➥🔝 7737669865 🔝▻ Deoghar Call-girls in Women Seeking Men  🔝Deoghar🔝   Escorts...
➥🔝 7737669865 🔝▻ Deoghar Call-girls in Women Seeking Men 🔝Deoghar🔝 Escorts...
amitlee9823
 
Bommasandra Call Girls: 🍓 7737669865 🍓 High Profile Model Escorts | Bangalore...
Bommasandra Call Girls: 🍓 7737669865 🍓 High Profile Model Escorts | Bangalore...Bommasandra Call Girls: 🍓 7737669865 🍓 High Profile Model Escorts | Bangalore...
Bommasandra Call Girls: 🍓 7737669865 🍓 High Profile Model Escorts | Bangalore...
amitlee9823
 
Call Girls In RT Nagar ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In RT Nagar ☎ 7737669865 🥵 Book Your One night StandCall Girls In RT Nagar ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In RT Nagar ☎ 7737669865 🥵 Book Your One night Stand
amitlee9823
 
(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...
(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...
(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...
motiram463
 

Kürzlich hochgeladen (20)

➥🔝 7737669865 🔝▻ Muzaffarpur Call-girls in Women Seeking Men 🔝Muzaffarpur🔝 ...
➥🔝 7737669865 🔝▻ Muzaffarpur Call-girls in Women Seeking Men  🔝Muzaffarpur🔝  ...➥🔝 7737669865 🔝▻ Muzaffarpur Call-girls in Women Seeking Men  🔝Muzaffarpur🔝  ...
➥🔝 7737669865 🔝▻ Muzaffarpur Call-girls in Women Seeking Men 🔝Muzaffarpur🔝 ...
 
Call Girls Chickpet ☎ 7737669865☎ Book Your One night Stand (Bangalore)
Call Girls Chickpet ☎ 7737669865☎ Book Your One night Stand (Bangalore)Call Girls Chickpet ☎ 7737669865☎ Book Your One night Stand (Bangalore)
Call Girls Chickpet ☎ 7737669865☎ Book Your One night Stand (Bangalore)
 
Escorts Service Arekere ☎ 7737669865☎ Book Your One night Stand (Bangalore)
Escorts Service Arekere ☎ 7737669865☎ Book Your One night Stand (Bangalore)Escorts Service Arekere ☎ 7737669865☎ Book Your One night Stand (Bangalore)
Escorts Service Arekere ☎ 7737669865☎ Book Your One night Stand (Bangalore)
 
Just Call Vip call girls Shillong Escorts ☎️9352988975 Two shot with one girl...
Just Call Vip call girls Shillong Escorts ☎️9352988975 Two shot with one girl...Just Call Vip call girls Shillong Escorts ☎️9352988975 Two shot with one girl...
Just Call Vip call girls Shillong Escorts ☎️9352988975 Two shot with one girl...
 
Abortion Pill for sale in Riyadh ((+918761049707) Get Cytotec in Dammam
Abortion Pill for sale in Riyadh ((+918761049707) Get Cytotec in DammamAbortion Pill for sale in Riyadh ((+918761049707) Get Cytotec in Dammam
Abortion Pill for sale in Riyadh ((+918761049707) Get Cytotec in Dammam
 
Just Call Vip call girls Berhampur Escorts ☎️9352988975 Two shot with one gir...
Just Call Vip call girls Berhampur Escorts ☎️9352988975 Two shot with one gir...Just Call Vip call girls Berhampur Escorts ☎️9352988975 Two shot with one gir...
Just Call Vip call girls Berhampur Escorts ☎️9352988975 Two shot with one gir...
 
Just Call Vip call girls Begusarai Escorts ☎️9352988975 Two shot with one gir...
Just Call Vip call girls Begusarai Escorts ☎️9352988975 Two shot with one gir...Just Call Vip call girls Begusarai Escorts ☎️9352988975 Two shot with one gir...
Just Call Vip call girls Begusarai Escorts ☎️9352988975 Two shot with one gir...
 
Call Girls Chikhali Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Chikhali Call Me 7737669865 Budget Friendly No Advance BookingCall Girls Chikhali Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Chikhali Call Me 7737669865 Budget Friendly No Advance Booking
 
Hosa Road Call Girls Service: ☎ 7737669865 ☎ High Profile Model Escorts | Ban...
Hosa Road Call Girls Service: ☎ 7737669865 ☎ High Profile Model Escorts | Ban...Hosa Road Call Girls Service: ☎ 7737669865 ☎ High Profile Model Escorts | Ban...
Hosa Road Call Girls Service: ☎ 7737669865 ☎ High Profile Model Escorts | Ban...
 
怎样办理圣芭芭拉分校毕业证(UCSB毕业证书)成绩单留信认证
怎样办理圣芭芭拉分校毕业证(UCSB毕业证书)成绩单留信认证怎样办理圣芭芭拉分校毕业证(UCSB毕业证书)成绩单留信认证
怎样办理圣芭芭拉分校毕业证(UCSB毕业证书)成绩单留信认证
 
Just Call Vip call girls godhra Escorts ☎️9352988975 Two shot with one girl (...
Just Call Vip call girls godhra Escorts ☎️9352988975 Two shot with one girl (...Just Call Vip call girls godhra Escorts ☎️9352988975 Two shot with one girl (...
Just Call Vip call girls godhra Escorts ☎️9352988975 Two shot with one girl (...
 
Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...
Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...
Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...
 
Get Premium Pimple Saudagar Call Girls (8005736733) 24x7 Rate 15999 with A/c ...
Get Premium Pimple Saudagar Call Girls (8005736733) 24x7 Rate 15999 with A/c ...Get Premium Pimple Saudagar Call Girls (8005736733) 24x7 Rate 15999 with A/c ...
Get Premium Pimple Saudagar Call Girls (8005736733) 24x7 Rate 15999 with A/c ...
 
Just Call Vip call girls daman Escorts ☎️9352988975 Two shot with one girl (d...
Just Call Vip call girls daman Escorts ☎️9352988975 Two shot with one girl (d...Just Call Vip call girls daman Escorts ☎️9352988975 Two shot with one girl (d...
Just Call Vip call girls daman Escorts ☎️9352988975 Two shot with one girl (d...
 
Pooja 9892124323, Call girls Services and Mumbai Escort Service Near Hotel Th...
Pooja 9892124323, Call girls Services and Mumbai Escort Service Near Hotel Th...Pooja 9892124323, Call girls Services and Mumbai Escort Service Near Hotel Th...
Pooja 9892124323, Call girls Services and Mumbai Escort Service Near Hotel Th...
 
SM-N975F esquematico completo - reparación.pdf
SM-N975F esquematico completo - reparación.pdfSM-N975F esquematico completo - reparación.pdf
SM-N975F esquematico completo - reparación.pdf
 
➥🔝 7737669865 🔝▻ Deoghar Call-girls in Women Seeking Men 🔝Deoghar🔝 Escorts...
➥🔝 7737669865 🔝▻ Deoghar Call-girls in Women Seeking Men  🔝Deoghar🔝   Escorts...➥🔝 7737669865 🔝▻ Deoghar Call-girls in Women Seeking Men  🔝Deoghar🔝   Escorts...
➥🔝 7737669865 🔝▻ Deoghar Call-girls in Women Seeking Men 🔝Deoghar🔝 Escorts...
 
Bommasandra Call Girls: 🍓 7737669865 🍓 High Profile Model Escorts | Bangalore...
Bommasandra Call Girls: 🍓 7737669865 🍓 High Profile Model Escorts | Bangalore...Bommasandra Call Girls: 🍓 7737669865 🍓 High Profile Model Escorts | Bangalore...
Bommasandra Call Girls: 🍓 7737669865 🍓 High Profile Model Escorts | Bangalore...
 
Call Girls In RT Nagar ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In RT Nagar ☎ 7737669865 🥵 Book Your One night StandCall Girls In RT Nagar ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In RT Nagar ☎ 7737669865 🥵 Book Your One night Stand
 
(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...
(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...
(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...
 

HPLC Problems, Probable Causes, and Remedies2020.pptx

  • 1. HPLC Problems, Probable Causes, and Remedies
  • 2. Problem No. 1: No Peaks/Very Small Peaks
  • 3. • Probable Cause 1) Detector lamp off. 2) Loose/broken wire between detector and integrator or recorder. 3) No mobile phase flow. 4) No sample/deteriorated sample/ wrong sample. 5) Settings too high on detector or recorder. • Remedy/Comments 1) Turn lamp on. 2) Check electrical connections and cables. 3) Pump off/ valve (knob) open 4) Be sure automatic sampler vials have sufficient liquid and no air bubbles in the sample. Evaluate system performance with fresh standard to confirm sample as source of problem. 5) Check attenuation or gain settings. Check lamp status. Auto-zero if necessary.
  • 4. Problem No. 2: No Flow
  • 5. • Probable Cause 1) Pump off. 2) Flow interrupted/obstructed. 3) Leak. 4) Air trapped in pump head. (Revealed by pressure fluctuations.) • Remedy/Comments 1) Start pump. 2) Check mobile phase level in reservoir(s). Check flow throughout system. Examine sample loop for obstruction or air lock. Make sure mobile phase components are miscible and mobile phase is properly degassed. 3) Check system for loose fittings. Check pump for leaks, salt buildup, unusual noises. Change pump seals if necessary. 4) Disconnect tubing at guard column (if present) or analytical column inlet. Check for flow. Purge pump at high flow rate (e.g., 5-10 mL/min.), prime system if necessary. (Prime each pump head separately.) If system has check valve, loosen valve to allow air to escape. If problem persists, flush system with 100% methanol or isopropanol. If problem still persists, contact system manufacturer.
  • 6. Problem No. 3: No Pressure/Pressure Lower
  • 7. • Probable Cause 1) Leak. 2) Mobile phase flow interrupted/obstructed. 3) Air trapped in pump head. (Revealed by pressure fluctuations.) 4) Leak at column inlet end fitting. 5) Air trapped elsewhere in system. 6) Worn pump seal causing leaks around pump head. 7) Faulty check valve. 8) Faulty pump seals. • Remedy/Comments 1) Check system for loose fittings. Check pump for leaks, salt buildup, unusual noises. Change pump seals if necessary. 2) Check mobile phase level in reservoir(s). Check flow throughout system. Examine sample loop for obstruction or air lock. Make sure mobile phase components are miscible and mobile phase is properly degassed. 3) Disconnect tubing at guard column (if present) or analytical column inlet. Check for flow. Purge pump at high flow rate (e.g., 10 mL/min.), prime system if necessary. (Prime each pump head separately.) If system has check valve, loosen valve to allow air to escape. 4) Reconnect column and pump solvent at double the flow rate. If pressure is still low, check for leaks at inlet fitting or column end fitting.
  • 8. • Disconnect guard and analytical column and purge system. Reconnect column(s). If problem persists, flush system with 100% methanol or isopropanol. • Replace seal. If problem persists, replace piston and seal. • Rebuild or replace valve. • Replace seals.
  • 9. Problem No. 4: Pressure Higher
  • 10. • Probable Cause • Problem in pump, injector, in-line filter, or tubing. • Obstructed guard column or analytical column. • Remedy/Comments • Remove guard column and analytical column from system. Replace with unions and 0.010'' I.D. or larger tubing to reconnect injector to detector. Run pump at 2-5 mL/min. If pressure is minimal, see Cause 2. If not, isolate cause by systematically eliminating system components, starting with detector, then in-line filter, and working back to pump. Replace filter in pump if present. • Remove guard column (if present) and check pressure. Replace guard column if necessary. If analytical column is obstructed, reverse and flush the column, while disconnected from the detector. If problem persists, column may be clogged with strongly retained contaminants. Use appropriate restoration procedure (Table 1). If problem still persists, change inlet frit or replace column.
  • 11. Problem No. 5: Variable Retention Times
  • 12. • Probable Cause 1) Leak. 2) Change in mobile phase composition. (Small changes can lead to large changes in retention times.) 3) Air trapped in pump. (Retention times increase and decrease at random times.) 4) Column temperature fluctuations (especially evident in ion exchange systems). 5) Column overloading. (Retention times usually decrease as mass of solute injected on column exceeds column capacity.) 6) Sample solvent incompatible with mobile phase. 7) Column problem. (Not a common cause of erratic retention. As a column ages, retention times gradually decrease.) • Remedy/Comments 1) Check system for loose fittings. Check pump for leaks, salt buildup, unusual noises. Change pump seals if necessary. 2) Check make-up of mobile phase. If mobile phase is machine mixed using proportioning values, hand mix and supply from one reservoir. 3) Purge air from pump head or check valves. Change pump seals if necessary. Be sure mobile phase is degassed. 4) Use reliable column oven. (Note: higher column temperatures increase column efficiency. For optimum results, heat eluant before introducing it onto column.) 5) Inject smaller volume (e.g., 10 μL vs. 100 μL) or inject the same volume after 1:10 or 1:100 dilutions of sample. 6) Adjust solvent. Whenever possible, inject samples in mobile phase. 7) Substitute new column of same type to confirm column as cause. Discard old column if restoration procedures fail.
  • 13. Problem No. 6: Loss of Resolution
  • 14. • Probable Cause 1) Mobile phase contaminated/deteriorated (causing retention times and/or selectivity to change). 2) Obstructed guard or analytical column. • Remedy/Comments 1) Prepare fresh mobile phase. 2) Remove guard column (if present) and attempt analysis. Replace guard column if necessary. If analytical column is obstructed, reverse and flush. If problem persists, column may be clogged with strongly retained contaminants. Use appropriate restoration procedure (Table 1). If problem still persists, change inlet frit or replace column.
  • 15. Problem No. 7: Split Peaks
  • 16. • Probable Cause 1) Contamination on guard or analytical column inlet. 2) Partially blocked frit. 3) Small (uneven) void at column inlet. 4) Sample solvent incompatible with mobile phase. • Remedy/Comments 1) Remove guard column (if present) and attempt analysis. Replace guard column if necessary. If analytical column is obstructed, reverse and flush. If problem persists, column may be clogged with strongly retained contaminants. Use appropriate restoration procedure (Table1 ). If problem still persists, inlet frit is probably (partially) plugged. Change frit or replace column. 2) Replace frit (see above). 3) Repack top of column with pellicular particles of same bonded phase functionality. Continue using the column in reverse flow direction. 4) Adjust solvent. Whenever possible, inject samples in mobile phase.
  • 17. Problem No. 8: Tailing Peaks
  • 18. • Probable Cause 1) Guard or analytical column contaminated/worn out. 2) Mobile phase contaminated/deteriorated. 3) Interfering components in sample. • Remedy/Comments 1) Remove guard column (if present) and attempt analysis. Replace guard column if necessary. If analytical column is source of problem, use appropriate restoration procedure (Table 1). If problem persists, replace column. 2) Check make-up of mobile phase. 3) Check column performance with standards.
  • 19. Problem No. 9: Fronting Peaks
  • 20. • Probable Cause 1) Column overloaded. 2) Sample solvent incompatible with mobile phase. 3) Shoulder or gradual baseline rise before a main peak may be another sample component. • Remedy/Comments 1) Inject smaller volume (e.g., 10 μL vs. 100 μL). Dilute the sample 1:10 or 1:100 fold in case of mass overload. 2) Adjust solvent. Whenever possible, inject samples in mobile phase. Flush polar bonded phase column with 50 column volumes HPLC grade ethyl acetate at 2-3 times the standard flow rate, then with intermediate polarity solvent prior to analysis. 3) Increase efficiency or change selectivity of system to improve resolution. Try another column type if necessary (e.g., switch from nonpolar C18 to polar cyano phase).
  • 21. Problem No. 10: Baseline Noise (regular)
  • 22. • Probable Cause 1) Air in mobile phase, detector cell, or pump. 2) Pump pulsations. 3) Incomplete mobile phase mixing. 4) Temperature effect (column at high temperature, detector unheated). 5) Other electronic equipment on same line. 6) Leak. • Remedy/Comments 1) Degas mobile phase. Flush system to remove air from detector cell or pump. 2) Incorporate pulse damper into system. 3) Mix mobile phase by hand or use less viscous solvent. 4) Reduce differential or add heat exchanger. 5) Isolate LC, detector, recorder to determine if source of problem is external. Correct as necessary. 6) Check system for loose fittings. Check pump for leaks, salt buildup, unusual noises. Change pump seals if necessary.
  • 23. Problem No. 11: Negative Peak
  • 24. • Probable Cause 1) Recorder leads reversed. 2) Refractive index of solute less than that of mobile phase (RI detector). 3) Sample solvent and mobile phase differ greatly in composition (vacancy peaks). 4) Mobile phase more absorptive than sample components to UV wavelength. • Remedy/Comments 1) Check polarity. 2) Use mobile phase with lower refractive index, or reverse recorder leads. 3) Adjust or change sample solvent. Dilute sample in mobile phase whenever possible. 4) a. Change polarity when using indirect UV detection, or b. Change UV wavelength or use mobile phase that does not adsorb chosen wavelength.
  • 25. Problem No. 12: Ghost Peak
  • 26. • Probable Cause 1) Contamination in injector or column. 2) Late eluting peak (usually broad) present in sample. • Remedy/Comments 1) Flush injector between analyses (a good routine practice). If necessary, run strong solvent through column to remove late eluters. Include final wash step in gradient analyses, to remove strongly retained compounds. 2) a. Check sample preparation. b. Include (step) gradient to quickly elute component.
  • 27. Table-1 • Table-1. Column Restoration Procedures • Silica-Based Reversed Phase Column (alkyl (C8, C18, etc.), phenyl, or diphenyl column, SUPELCOSIL LC-PAH column) •A. Water Soluble Samples Flush with the following: • Flush with warm (60 °C) distilled water • 50 mL methanol • 50 mL acetonitrile • 25 mL methanol • 25 mL mobile phase • Evaluate column performance.
  • 28. •B. Samples Not Soluble in Water Flush with the following: • 50 mL 2-propanol • 50 mL methylene chloride • 50 mL hexane • 25 mL isopropanol • 25 mL mobile phase • Evaluate column performance.
  • 29.