Presented at the 10th International NDE Conference in Nice, France in 2013, this paper outlines new methods and technologies for obtaining representative samples without tendon can disassembly, and a more robust and reliable test slate to evaluate corrosion failure modes including MIC, water, corrosive ions, and the presence of oxidized ferrous debris. Methods developed in response to a containment tendon failure, and the resulting corrective action plan to evaluate an entire assembly of tendons.
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
Concrete Containment Tendon Grease Sampling and Analysis
1. Advanced methods for
sampling and analysis of
post-tensioned concrete
tendon greases
Rich Wurzbach
ASNT Level III, Infrared and Thermal Testing
Laboratory Lubricant Analyst (LLA-I)
MRG Labs, York, Pennsylvania, USA
2. Overview
Background of tendon grease testing
Failures of post-tension tendons
Comparison of as-found conditions,
failures, to routine grease testing
Root-Cause failure analysis and corrective
actions
Enhanced sampling methods
Expanded grease analysis techniques
3. Background
Some nuclear containment designs
include ungrouted post-tension concrete
tendons
Tendon sheaths and anchorhead areas
are filled with a corrosion protection
grease
ASME Section XI, IWL-2525 Table 1
contains guidance for corrosion protection
grease testing, requires 125g sample
minimum
5. Status of referenced tests
ASTM D992 was withdrawn in 1983
Current version of ASTM D4327 allows for
testing of all three ions, but IWL-2525-1
specifies this for nitrates only, requires two other
tests to be performed.
APHA tests are for environmental and
wastewater; often requires samples being split
between labs, or being run by inexperienced
personnel, may require refrigeration of prepared
sample
Total alkalinity test does not fall within the scope
of current ASTM standard.
6. Sampling Challenges
Quantity required for current IWL-2525
test slate requires disassembly of tendon
anchors can, disturbing anchor area
during sampling
Harvesting of grease may disturb position
of buttonheads, hiding broken strand
indications
Invasive nature of sampling requires costly
logistical support, equipment, time,
scaffold, etc.
8. Containment Tendon Grease
Tendon failure at US
nuclear power plant
prompts RCFA, CAR
IWL, Section IX
requires 125g grease
Test slate for 2 grams
grease developed; tool
for sampling remotely
9. Grease thief and T-handle
Provide method to sample grease near anchor without
disassembly
Design consistent with ASTM D7718 Standard
Practice for Obtaining In-Service Samples of
Lubricating Grease
T-handle and stinger probe allow sampling grease
from face of anchor
Results from Small Volume Test (SVT) correlated to
disassembly method (large volume sample)
Sampling saved ~US$500,000 in maintenance costs
10. Water – D95 (IWL-2525)
Uses distillation glassware
Per ASTM: “The amount of water as determined
by this test method…to the nearest 0.05 or 0.1
volume%”, which is 500-1000ppm.
Per Noria Training: D95 “is fairly cumbersome
and requires a comparatively large sample to
ensure accuracy, which is why it is rarely used in
production-style oil analysis labs today”
11. Water – Karl Fischer (SVT)
Per ASTM: KF allows “direct determination
of water in the range of 10 to 25,000” ppm
Can be performed on grease via “oven
method” to eliminate interferences and
improve accuracy
Only a small volume of grease required
(about ¼ gram)
13. Total Alkalinity Testing
Uses an “ASTM Modified” procedure.
Per ASTM: “determination of acidic or basic
constituents in petroleum products and
lubricants soluble or nearly soluble in mixtures
of toluene and isopropyl alcohol.”
Per ASTM: “This test method is not suitable
for measuring the basic constituents of many
basic additive-type lubricating oils. Test
Method D4739 can be used for this purpose.”
Visconorust contains basic detergents
14. Total Alkalinty Comparison
Method developed using 0.5 grams
grease and smaller solvent amount
Grease fully dissolves, acidification
process followed
Titration performed with both colorimetric
and potentiometric (D4793) methods,
correlating results
Titration performed per IWL-2525 and new
small volume method, correlating results
15. Ion testing
Two of three methods in IWL-2525 allow
for ion-specific electrodes
New method for small volume ion-specific
electrode method under development
Can create more repeatable, reliable, and
quicker test for ions, Chloride, Nitrate and
Sulfide
16. SVT Small Volume Test Slate
Ferromagnetic Iron (ppm) – tested with a Hall-effect
sensor
Metals spectroscopy – multiple elements, ppm (iron,
calcium, etc)
Moisture, ppm
Total Base Number (alkalinity) – using 0.5g sample
Red oxide count (corrosion particulate by microscope or
direct-imaging sensor
Die Extrusion Index – evaluating unexpected softening or
hardening of the grease
Microbial content – quantification of ATP by grease thin-
film extraction
17. Analysis Techniques
Sample is received. fdM+ is run Grease Thief Analyzer is performed and substrate is made
Two strips are used to make
a dilution to run RDE/ICP.
One Strip is used
for FT-IR.
One Strip is Dissolved in Green
RULER solution to run RULER.
19. Results
Data compiled for 500 tendon locations
Strongest correlation between related
parameters (Calcium and TBN, Red
Oxides and Ferrous Debris)
Some high Red Oxides were found in
higher Total Alkalinity locations
Multiple potential failure modes likely
contributor to lack of consistent correlation
24. Analysis of Data
Not a clear single failure mode
represented by correlated data
Several tendons identified for further
testing: high ATP (microbial), several high
Red Oxides, & high Red Oxide/Low TBN
Testing showed majority of tendons in
good condition, small percentage
candidates for disassembly and inspection
Advanced statistics package used for
ranking metrics
25. Analysis of Data
Dispersion analysis: failed tendon
exceeded 2σ outside the mean for the
failed tendon for three parameters
Data analysis by Dr. David Polk, Polk-Lepson Research, USA http://www.polk-lepsonresearchgroup.com/
26. Analysis of Data
Regression analysis performed using Red
Oxides as the dependent variable (actual
measure of corrosion) and assigning
weighting factors for scoring
Failed tendon and one other scored 5, and
twelve others scored as 4.
Data analysis by Dr. David Polk, Polk-Lepson Research, USA http://www.polk-lepsonresearchgroup.com/
27. Summary
Tendon failure modes require reliable and
cost-effective method for sampling and
analysis
Using new sampling tools per ASTM
D7718 without tendon can removal
Small volume testing methods shown to
be equivalent or superior to traditional
testing
Additional failure modes detectable with
small volume test slate
Hinweis der Redaktion
APHA American Public Health Association
The Grease ThiefTM is pictured here. The red arrow shows the location of the purge hole, which allows excessive grease to purge out without creating backpressue in the bearing. It is recognized that many electric motor designs do not regularly purge grease out the drain hole when properly greased. It is not suggested that greasers pump enough grease in to fill this fitting. However, it can be substituted for the existing drain plug in many motors, and standby to catch any grease that might come out, instead of it falling to the floor. Then the captured grease can be analyzed. The red piston provides a visual indication of when the grease has begun to slump, leak or pressure drain into the fitting, indication of overgreasing, grease property changes, overheating of the bearing, etc.
fdM+ ferrous debris analyzer
Determines the amount of ferrous debris in a sample
Grease Thief Analyzer
Measures the consistency of the grease
FT-IR
Tests for mixing, contamination, and oxidation
RULER
Tests for amount of useful antioxidant life remaining
RDE Spectroscopy
metals spectroscopy
Rheometer
Tests the consistency, pumpability, and likeliness of tunneling
We can prepare a ferrographic slide for inspection. If there is a signficant amount of non-ferrous particulate of interest, a micropatch can be prepared with the same solution for microscopic evaluation.