Scanning Electron Microscopy
M.Sc Material Science & Engineering
JIMMA UNIVERSITY
MScE6409 – Electron Spectroscopy
and Microscopy
What are Electron Microscopes?
Electron Microscopes are scientific instruments that use a beam of
energetic electrons to examine objects on a very fine scale.
Electron microscopes work by using an electron beam instead of visible
light and an electron detector instead of our eyes.
An electron beam allows us to see at very small scales because electrons
can also behave as light.
It has the properties of a wave with a wavelength that is much smaller
than visible light (a few trillionths of a meter).
With this wavelength we can distinguish features down to a fraction of a
nanometer.
Electron Microscopy
10,000x plus magnification, not possible using optical microscopes.
Uses electromagnetic lenses, electrons and fluorescent screen to produce
image
Resolution increased 1000 fold over bright field microscope to about 0.3
nm (1 X 10-9)
Magnification increased to 100,000x
Two types of electron microscopes
1. Scanning Electron Microscope (SEM)
2. Transmission Electron Microscope (TEM)
Principles of Electron Microscopy
What is SEM?
Working principles of SEM
Major Components and their functions
Electron beam – specimen interactions
Detection of the sample surface
Energy Dispersive X-ray Spectroscopy (EDS)
Characterization
Applications
Scanning Electron Microscopy
SEM uses a very fine probe of electrons
focused at the surface of the specimen
and scanned across it in a raster or
pattern of parallel lines.
It is a type of electron microscope
capable of producing high resolution
images of a sample surface (as low as
10nm or 100Å) in vacuum.
First constructed by Knoll and von
Ardenne in Germany, 1930s.
Commercially available in 1963 with
improvements made by McMullan and
Oatley at Cambridge University.
What is SEM?
The basic principle is that a beam of electrons is generated by a
suitable source, typically a tungsten filament (thermionic gun) or a
field emission gun (FEG).
The electron beam is accelerated through a high voltage (e.g. 20 kV)
and pass through a system of apertures and electromagnetic lenses to
produce a thin beam of electrons.
Working Principles of SEM
Working Principles of SEM
A beam of electrons scans the surface of a sample (Figure). The
electrons interact with the material in a way that triggers the emission
of secondary electrons.
These secondary electrons are captured by a detector, which forms an
image of the surface of the sample.
The direction of the emission of the secondary electrons depends on
the orientation of the features of the surface.
There, the image formed will reflect the characteristic feature of the
region of the surface that was exposed to the electron beam.
Major Components and their Functions
Major Components and their Functions
A tungsten (W) thermionic
filament is used as electron
gun.
Other examples: Lanthanum
or Cerium Hexaboride
When high voltage is applied
to tungsten filament which
produces the steady stream
of electrons.
Electron gun located at the
very top of a SEM fire a beam
of electrons at the object
under examination
Electron Gun
Thermionic Gun
makes use of a filament that is heated
up to make the electrons overcome the
work function
Field Emission Gun
makes use of two anodes: (1) to extract
electrons from the filament, (2) to
accelerate the electrons to the operating
voltage
Types of Electron Gun
1. Thermionic electron gun
Thermionic filaments are made of tungsten (W) in the form of a v-shaped wire.
Thermionic emission is the liberation of electrons from an electrode by virtue of its
temperature (releasing of energy supplied by heat).
This occurs because the thermal energy given to the carrier overcomes the work
function of the material.
The charge carriers can be electrons or ions, and in literature are sometimes
referred to as thermions.
The classical example of thermionic emission is that of electrons from a hot
cathode into a vacuum (also known as thermal electron emission or the Edison
effect) in a vacuum tube.
The hot cathode can be a metal filament e.g. Tungsten. Vacuum emission from
metals tends to become significant only for temperatures over 1,000 K (730 °C).
Types of Electron Gun
2. Field Emission Gun (FEG)
For a field emission source, a fine, sharp, single crystal tungsten tip is employed.
An FEG emitter gives a more coherent beam and its brightness is much higher than
the tungsten filament.
Electrons are emitted from a smaller area of the FEG source, giving a source size of
a few nanometers, compared to around 50 μm for the tungsten filament. This
leads to greatly improved image quality with the FEG source.
Comparison of Thermionic (tungsten) and Field Emission Gun
Types of Electron Gun
There are two types of FEG sources: Cold and Schottky FEGs
1. Cold emission source, heating of the filament is not required as it operates at
room temperature. However, this type of filament is prone to contamination and
requires more stringent vacuum conditions (10-8 Pa, 10-10 torr).
Regular and rapid heating (‘flashing’) is required in order to remove contamination.
The spread of electron energies is very small for a cold field emitter (0.3 eV) and the
source size is around 5 nm.
2. Thermal and Schottky sources, operate with lower field strengths. The surface of
the tungsten (W) tip is covered with a thin layer of zirconium oxide (ZrO) to further
decrease the potential barrier.
The Schottky source is also heated and dispenses zirconium dioxide onto the
tungsten tip to further lower its work function.
The Schottky source is slightly larger, 20–30 nm, with a small energy spread (about
1 eV).
Types of FEG Sources
SEMs use lenses to
produce clear and
detailed images.
They are not made of
glass but rather magnets
which are capable of
bending the path of
electrons.
The lenses focus and
control the electron
beam, ensuring that the
electrons end up
precisely on the desired
surface.
Condenser – Lens system
It accommodates the
specimen holder and
mechanisms for
manipulating it and
detectors for the various
emissions.
It must be very sturdy and
insulated from vibration
since the specimen must
be kept extremely still for
the microscope to produce
clear images.
SEM operates in a vacuum
to avoid interference of the
suspended particles with
the movement of incident
and reflected electron
beams.
Sample Chamber
Electron Collection System
It contains devices
detect the various ways
that the electron beam
interacts with the
sample object
The detector common to
all SEM systems is based
on the design of Everhart
and Thornley, which
detects secondary and
backscattered electrons
Electron beam – Specimen interactions
The result of the primary beam hitting the specimen is the formation of a
teardrop shaped reaction vessel (Fig.).
The reaction vessel by definition is where all the scattering events are
taking place. An increase in the topography will increase the surface area
of the reaction vessel resulting in more signal (Fig.).
Figure. Formation of teardrop shaped
reaction vessel
Figure. Surface area of teardrop
shaped reaction vessel
Electron beam – Specimen interactions
Electron beam – Specimen interactions
Electron beam – Specimen interactions
Six or more different events occur in the reaction vessel. These events
include:
1. Backscattered electrons. A primary beam electron may be scattered in
such a way that it escapes back from the specimen but does not go
through the specimen.
Backscattered electrons are the original beam electrons and thus, have a
high energy level, near that of the gun voltage.
Operating in the backscattered imaging mode is useful when relative
atomic density information in conjunction with topographical information
is to be displayed.
Electron beam – Specimen interactions
2. Secondary electrons are generated when a primary electron dislodges a
specimen electron from the specimen surface.
Secondary electrons have a low energy level of only a few electron volts,
thus, they can only be detected when they are dislodged near the surface of
the reaction vessel. Therefore, secondary electrons cannot escape from
deep within the reaction vessel.
Two of the foremost reasons for operating in the secondary electron
imaging mode are to obtain topographical information and high resolution.
3. X-rays. When electrons are dislodged from specific orbits of an atom in
the specimen, X-rays are omitted.
Elemental information can be obtained in the X-ray mode, because the X-ray
generated has a wavelength and energy characteristic of the elemental
atom from which it originated.
Electron beam – Specimen interactions
4. Cathode Luminescence. Some specimen molecule's florescence when
exposed to an electron beam. In the SEM, this reaction is called cathode
luminescence.
The florescence produces light photons that can be detected. A compound or
structure labeled with a luminescent molecule can be detected by using
cathode luminescence techniques.
5. Specimen Current. When the primary electron undergoes enough scattering
such that the energy of the electron is decrease to a point where the electron is
absorbed by the sample, this is known as specimen current.
6. Transmitted electrons. If the specimen is thin enough, primary electrons
may pass through the specimen. These electrons are known as transmitted
electrons and they provide some atomic density information.
The atomic density information is displayed as a shadow. The higher the atomic
number the darker the shadow until no electrons pass through the specimen.
Detectors
Detection of Sample Surface
Secondary electrons emitted from the specimen are detected using a
scintillator-photomultiplier, "Everhart-Thornley," detector.
Low energy secondary electrons are emitted from the sample in all directions.
They are gathered by a charged collector grid (or cage), which can be biased
from -50 to +300 V. This draws the secondary electrons towards the scintillator.
Detection of Sample Surface
The scintillator is a thin plastic disk coated with a short-persistence
phosphor that is highly efficient at converting the energy contained in the
electrons into ultraviolet light photons (4000 Å).
The response time of the phosphor is fast and permits high resolution
scanning.
The outer layer of the scintillator is coated with a thin layer [10-50 nm] of
aluminum, positively biased at approximately 10 KeV, which accelerates the
electrons to the scintillator surface.
The charged collector grid, in addition to collecting secondary electrons
from the sample, helps to alleviate some of the negative effects of the
scintillator aluminum layer bias, which can actually distort the incident
beam.
The topographical aspects of a secondary electron image depend on how
many of electrons actually reach the detector.
When the incident electron beam intersects the edges of topographically
high portions of a sample at lower angles, it puts more energy into the
volume of secondary electron production.
Thus, high points produce more secondary electrons, generating a larger
signal.
Faces oriented towards the detector also generate more secondary
electrons.
Secondary electrons that are prevented from reaching the detector do not
contribute to the final image and these areas will appear as shadows or
darker in contrast than those regions that have a clear electron path to the
detector.
Detection of Sample Surface
Signals are collected by appropriate detector resulting signal is amplified and
displayed.
The topographic information contained in the image is straightforward.
Detection of Sample Surface
1. Reduce the specimen into a size that fits the stub
2. Coat non-conducting specimens using either a vacuum evaporator or
sputter coater. This step increases primary and secondary emissions,
which enhances image quality.
For conventional imaging in the SEM, specimens must be electrically
conductive, at least at the surface, and electrically grounded to prevent
the accumulation of electrostatic charge (i.e. using metal coating).
Sample Preparation
Non-conducting materials are usually coated with an ultra-thin coating of
electrically conducting material, including gold, gold/palladium alloy,
platinum, platinum/palladium, iridium, tungsten, and chromium.
The recommended metals and sputter coating thickness are given below
for tungsten and FEG sources:
Metals:
Au, Au/Pd (Tungsten source)
Pt, Pd/Pt, Ir, W (FEG source)
Thickness:
5-10 nm for low magnification
2-3 nm for high resolution, the thinner the better
Sample Preparation
Type of Specimen Cutting Coating
Metallic  
Ceramic, conducting  
ceramic, non-conducting  
Semiconductor  
Polymer & polymer-matrix
composites  
Particles and fibers  
Biological and botanical
specimens  
Sample Preparation
Energy Dispersive X-ray spectroscopy
Energy-dispersive X-ray spectroscopy (EDS) is an analytical technique used for the
elemental analysis or chemical characterization of a sample. It relies on an
interaction of some source of X-ray excitation and a sample.
Its characterization is in large part to the fundamental principle that each element
has a unique atomic structure allowing a unique set of peaks on its electromagnetic
emission spectrum.
A sample is excited under high energy of electron
beam, the inner shell of electrons is ejected to
vacuum creating a vacancy in that shell.
Electrons from the outer shell jump into the
vacant site for filling the inner shell.
During this process, the sample fluoresces X-ray of
energy same as the energy difference between
the initial state and final state.
Energy Dispersive X-ray spectroscopy
Since each atom has its unique and discretized energy levels, the X-Ray
fluorescence is also characteristic of that atom.
Energy dispersive X ray spectroscopy is a technique that detects the X-ray
fluorescence to characterize the elements present in a material.
The most common detector used is Si(Li) detector cooled to cryogenic
temperatures with liquid nitrogen. Now, newer systems are often equipped
with silicon drift detectors (SDD) with Peltier cooling systems.
Figure. EDS spectrum of the mineral crust of the vent shrimp. Most of these
peaks are X-rays given off as electrons return to the K electron shell.(K-
alpha and K-beta lines) One peak is from the L shell of iron.
EDS Spectrum
Advantages of SEM
• Specimen size (small)
• Samples must be conductive, otherwise non-conductive material
should be metal coated
• Only surface morphology can be obtained
Limitations of SEM
1. SEM has a large depth of field, which allows more of a specimen to be in
focus at one time
2. SEM also has much higher resolution, so closely spaced specimens can
be magnified at much higher levels
3. Because the SEM uses electromagnets rather than lenses, the researcher
has much more control in the degree of magnification
All of these advantages, as well as the actual strikingly clear images, make
the SEM one of the most useful instruments in research today
SEM has a variety of applications in a number of scientific and industry-
related fields, especially where characterizations of solid materials is
beneficial.
In addition to topographical, morphological and compositional
information, an SEM can detect and analyze surface fractures, provide
information in microstructures, examine surface contaminations, reveal
spatial variations in chemical compositions, provide qualitative chemical
analyses and identify crystalline structures.
In addition, SEMs have practical industrial and technological applications
such as semiconductor inspection, production line of miniscule products and
assembly of microchips for computers.
SEMs can be as essential research tool in fields such as life science, biology,
gemology, medical and forensic science, metallurgy.
Applications
1. Image morphological of samples
2. Image compositional and some bonding differences
3. Examine wet and dry samples while viewing them (only in an
ESEM)
4. View frozen material (in a SEM with a cryostage)
5. Generate X-rays from samples for microanalysis (EDS)
6. View/map grain orientation/crystallographic orientation and
study related information like heterogeneity
Applications
Fractography
Biology
Crystal Structure
Surface Morphology
Applications
Sem md ppt

Weitere ähnliche Inhalte

Was ist angesagt?

Was ist angesagt? (20)

Electron Microscopy (SEM & TEM)
Electron Microscopy (SEM & TEM)Electron Microscopy (SEM & TEM)
Electron Microscopy (SEM & TEM)
 
Scanning Electron Microscope (SEM)
Scanning Electron Microscope (SEM)Scanning Electron Microscope (SEM)
Scanning Electron Microscope (SEM)
 
Atomic Force Microscope
Atomic Force MicroscopeAtomic Force Microscope
Atomic Force Microscope
 
Scanning electron microscopy
Scanning electron microscopyScanning electron microscopy
Scanning electron microscopy
 
Scanning and Transmission Electron Microscope
Scanning and Transmission Electron MicroscopeScanning and Transmission Electron Microscope
Scanning and Transmission Electron Microscope
 
Transmission Electron Microscope
Transmission Electron MicroscopeTransmission Electron Microscope
Transmission Electron Microscope
 
Transmission electron microscopy (1) (1)
Transmission electron microscopy (1) (1)Transmission electron microscopy (1) (1)
Transmission electron microscopy (1) (1)
 
Scanning Electron Microscopy - SEM
Scanning Electron Microscopy - SEM Scanning Electron Microscopy - SEM
Scanning Electron Microscopy - SEM
 
Scanning electron microscopy (sem)
Scanning electron microscopy (sem)Scanning electron microscopy (sem)
Scanning electron microscopy (sem)
 
Scanning Electron microscopy
Scanning Electron microscopy Scanning Electron microscopy
Scanning Electron microscopy
 
Scanning probe microscope
Scanning probe microscopeScanning probe microscope
Scanning probe microscope
 
2018 HM-scanning electron microscope
2018 HM-scanning electron microscope2018 HM-scanning electron microscope
2018 HM-scanning electron microscope
 
Scanning electron microscope (sem)
Scanning electron microscope (sem)Scanning electron microscope (sem)
Scanning electron microscope (sem)
 
Optical microscope
Optical microscopeOptical microscope
Optical microscope
 
Scanning Electron Microscopy (SEM) lecture
Scanning Electron Microscopy (SEM) lectureScanning Electron Microscopy (SEM) lecture
Scanning Electron Microscopy (SEM) lecture
 
transmission electron microscopy
transmission electron microscopytransmission electron microscopy
transmission electron microscopy
 
AFM (Atomic Force Microscopy)
AFM (Atomic Force Microscopy)AFM (Atomic Force Microscopy)
AFM (Atomic Force Microscopy)
 
Scanning and transmission electron microscopy
Scanning and transmission electron microscopyScanning and transmission electron microscopy
Scanning and transmission electron microscopy
 
Sem and tem
Sem and temSem and tem
Sem and tem
 
SCANNING ELECTRON MICROSCOPY (SEM)
SCANNING ELECTRON MICROSCOPY (SEM)SCANNING ELECTRON MICROSCOPY (SEM)
SCANNING ELECTRON MICROSCOPY (SEM)
 

Ähnlich wie Sem md ppt

The SEM can use either backscattered or secondary electrons to form .pdf
The SEM can use either backscattered or secondary electrons to form .pdfThe SEM can use either backscattered or secondary electrons to form .pdf
The SEM can use either backscattered or secondary electrons to form .pdf
rushabhshah600
 
Scanning electron microscope
Scanning electron microscopeScanning electron microscope
Scanning electron microscope
SuganyaPaulraj
 

Ähnlich wie Sem md ppt (20)

E mintro
E mintroE mintro
E mintro
 
Electron microscope (SEM and TEM)
Electron microscope (SEM and TEM)Electron microscope (SEM and TEM)
Electron microscope (SEM and TEM)
 
Electron Microscopy
Electron MicroscopyElectron Microscopy
Electron Microscopy
 
Chapter-3. Electron Microscopy.pptx
Chapter-3. Electron Microscopy.pptxChapter-3. Electron Microscopy.pptx
Chapter-3. Electron Microscopy.pptx
 
Scanning electron microscopy
Scanning electron microscopyScanning electron microscopy
Scanning electron microscopy
 
Sem
SemSem
Sem
 
lecture4-konsep dasar SEM.ppt
lecture4-konsep dasar SEM.pptlecture4-konsep dasar SEM.ppt
lecture4-konsep dasar SEM.ppt
 
Sem Poster
Sem PosterSem Poster
Sem Poster
 
SEM and TEM
SEM and  TEMSEM and  TEM
SEM and TEM
 
SEM,TEM & AFM
SEM,TEM & AFMSEM,TEM & AFM
SEM,TEM & AFM
 
electron microscopy
electron microscopyelectron microscopy
electron microscopy
 
ELECTRON MICROSCOPY-Scanning Electron MicroscopySEM.pptx
ELECTRON MICROSCOPY-Scanning Electron MicroscopySEM.pptxELECTRON MICROSCOPY-Scanning Electron MicroscopySEM.pptx
ELECTRON MICROSCOPY-Scanning Electron MicroscopySEM.pptx
 
overview on scanning electron microscope
overview on scanning electron microscopeoverview on scanning electron microscope
overview on scanning electron microscope
 
SCANNING ELECTRON MICROSCOPE MITHILESH CHOUDHARY
SCANNING ELECTRON MICROSCOPE MITHILESH CHOUDHARYSCANNING ELECTRON MICROSCOPE MITHILESH CHOUDHARY
SCANNING ELECTRON MICROSCOPE MITHILESH CHOUDHARY
 
SEM BRIEF.docx
SEM BRIEF.docxSEM BRIEF.docx
SEM BRIEF.docx
 
Scanning Electron Microscopy
Scanning Electron MicroscopyScanning Electron Microscopy
Scanning Electron Microscopy
 
The SEM can use either backscattered or secondary electrons to form .pdf
The SEM can use either backscattered or secondary electrons to form .pdfThe SEM can use either backscattered or secondary electrons to form .pdf
The SEM can use either backscattered or secondary electrons to form .pdf
 
Electron microscope
Electron microscopeElectron microscope
Electron microscope
 
Characterization of nanomaterials
Characterization of nanomaterialsCharacterization of nanomaterials
Characterization of nanomaterials
 
Scanning electron microscope
Scanning electron microscopeScanning electron microscope
Scanning electron microscope
 

Mehr von Mulugeta Abera (6)

Afm
AfmAfm
Afm
 
Chapter two new
Chapter two newChapter two new
Chapter two new
 
Chapter 1 (introduction)
Chapter 1 (introduction)Chapter 1 (introduction)
Chapter 1 (introduction)
 
Chapter 6 design of irrigation channel1
Chapter 6 design of irrigation channel1Chapter 6 design of irrigation channel1
Chapter 6 design of irrigation channel1
 
Chapter 3 soil water and irrigation practice1
Chapter 3  soil water and irrigation practice1Chapter 3  soil water and irrigation practice1
Chapter 3 soil water and irrigation practice1
 
Chapter 6 design of irrigation channel1
Chapter 6 design of irrigation channel1Chapter 6 design of irrigation channel1
Chapter 6 design of irrigation channel1
 

Kürzlich hochgeladen

Cara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak Hamil
Cara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak HamilCara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak Hamil
Cara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak Hamil
Cara Menggugurkan Kandungan 087776558899
 
DeepFakes presentation : brief idea of DeepFakes
DeepFakes presentation : brief idea of DeepFakesDeepFakes presentation : brief idea of DeepFakes
DeepFakes presentation : brief idea of DeepFakes
MayuraD1
 
Verification of thevenin's theorem for BEEE Lab (1).pptx
Verification of thevenin's theorem for BEEE Lab (1).pptxVerification of thevenin's theorem for BEEE Lab (1).pptx
Verification of thevenin's theorem for BEEE Lab (1).pptx
chumtiyababu
 

Kürzlich hochgeladen (20)

Navigating Complexity: The Role of Trusted Partners and VIAS3D in Dassault Sy...
Navigating Complexity: The Role of Trusted Partners and VIAS3D in Dassault Sy...Navigating Complexity: The Role of Trusted Partners and VIAS3D in Dassault Sy...
Navigating Complexity: The Role of Trusted Partners and VIAS3D in Dassault Sy...
 
Thermal Engineering Unit - I & II . ppt
Thermal Engineering  Unit - I & II . pptThermal Engineering  Unit - I & II . ppt
Thermal Engineering Unit - I & II . ppt
 
Cara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak Hamil
Cara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak HamilCara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak Hamil
Cara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak Hamil
 
DeepFakes presentation : brief idea of DeepFakes
DeepFakes presentation : brief idea of DeepFakesDeepFakes presentation : brief idea of DeepFakes
DeepFakes presentation : brief idea of DeepFakes
 
COST-EFFETIVE and Energy Efficient BUILDINGS ptx
COST-EFFETIVE  and Energy Efficient BUILDINGS ptxCOST-EFFETIVE  and Energy Efficient BUILDINGS ptx
COST-EFFETIVE and Energy Efficient BUILDINGS ptx
 
Computer Lecture 01.pptxIntroduction to Computers
Computer Lecture 01.pptxIntroduction to ComputersComputer Lecture 01.pptxIntroduction to Computers
Computer Lecture 01.pptxIntroduction to Computers
 
Design For Accessibility: Getting it right from the start
Design For Accessibility: Getting it right from the startDesign For Accessibility: Getting it right from the start
Design For Accessibility: Getting it right from the start
 
PE 459 LECTURE 2- natural gas basic concepts and properties
PE 459 LECTURE 2- natural gas basic concepts and propertiesPE 459 LECTURE 2- natural gas basic concepts and properties
PE 459 LECTURE 2- natural gas basic concepts and properties
 
S1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptx
S1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptxS1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptx
S1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptx
 
School management system project Report.pdf
School management system project Report.pdfSchool management system project Report.pdf
School management system project Report.pdf
 
Online food ordering system project report.pdf
Online food ordering system project report.pdfOnline food ordering system project report.pdf
Online food ordering system project report.pdf
 
Employee leave management system project.
Employee leave management system project.Employee leave management system project.
Employee leave management system project.
 
HAND TOOLS USED AT ELECTRONICS WORK PRESENTED BY KOUSTAV SARKAR
HAND TOOLS USED AT ELECTRONICS WORK PRESENTED BY KOUSTAV SARKARHAND TOOLS USED AT ELECTRONICS WORK PRESENTED BY KOUSTAV SARKAR
HAND TOOLS USED AT ELECTRONICS WORK PRESENTED BY KOUSTAV SARKAR
 
Tamil Call Girls Bhayandar WhatsApp +91-9930687706, Best Service
Tamil Call Girls Bhayandar WhatsApp +91-9930687706, Best ServiceTamil Call Girls Bhayandar WhatsApp +91-9930687706, Best Service
Tamil Call Girls Bhayandar WhatsApp +91-9930687706, Best Service
 
Double Revolving field theory-how the rotor develops torque
Double Revolving field theory-how the rotor develops torqueDouble Revolving field theory-how the rotor develops torque
Double Revolving field theory-how the rotor develops torque
 
Introduction to Serverless with AWS Lambda
Introduction to Serverless with AWS LambdaIntroduction to Serverless with AWS Lambda
Introduction to Serverless with AWS Lambda
 
Wadi Rum luxhotel lodge Analysis case study.pptx
Wadi Rum luxhotel lodge Analysis case study.pptxWadi Rum luxhotel lodge Analysis case study.pptx
Wadi Rum luxhotel lodge Analysis case study.pptx
 
Verification of thevenin's theorem for BEEE Lab (1).pptx
Verification of thevenin's theorem for BEEE Lab (1).pptxVerification of thevenin's theorem for BEEE Lab (1).pptx
Verification of thevenin's theorem for BEEE Lab (1).pptx
 
A CASE STUDY ON CERAMIC INDUSTRY OF BANGLADESH.pptx
A CASE STUDY ON CERAMIC INDUSTRY OF BANGLADESH.pptxA CASE STUDY ON CERAMIC INDUSTRY OF BANGLADESH.pptx
A CASE STUDY ON CERAMIC INDUSTRY OF BANGLADESH.pptx
 
FEA Based Level 3 Assessment of Deformed Tanks with Fluid Induced Loads
FEA Based Level 3 Assessment of Deformed Tanks with Fluid Induced LoadsFEA Based Level 3 Assessment of Deformed Tanks with Fluid Induced Loads
FEA Based Level 3 Assessment of Deformed Tanks with Fluid Induced Loads
 

Sem md ppt

  • 1. Scanning Electron Microscopy M.Sc Material Science & Engineering JIMMA UNIVERSITY MScE6409 – Electron Spectroscopy and Microscopy
  • 2. What are Electron Microscopes? Electron Microscopes are scientific instruments that use a beam of energetic electrons to examine objects on a very fine scale. Electron microscopes work by using an electron beam instead of visible light and an electron detector instead of our eyes. An electron beam allows us to see at very small scales because electrons can also behave as light. It has the properties of a wave with a wavelength that is much smaller than visible light (a few trillionths of a meter). With this wavelength we can distinguish features down to a fraction of a nanometer. Electron Microscopy
  • 3. 10,000x plus magnification, not possible using optical microscopes. Uses electromagnetic lenses, electrons and fluorescent screen to produce image Resolution increased 1000 fold over bright field microscope to about 0.3 nm (1 X 10-9) Magnification increased to 100,000x Two types of electron microscopes 1. Scanning Electron Microscope (SEM) 2. Transmission Electron Microscope (TEM) Principles of Electron Microscopy
  • 4. What is SEM? Working principles of SEM Major Components and their functions Electron beam – specimen interactions Detection of the sample surface Energy Dispersive X-ray Spectroscopy (EDS) Characterization Applications Scanning Electron Microscopy
  • 5. SEM uses a very fine probe of electrons focused at the surface of the specimen and scanned across it in a raster or pattern of parallel lines. It is a type of electron microscope capable of producing high resolution images of a sample surface (as low as 10nm or 100Å) in vacuum. First constructed by Knoll and von Ardenne in Germany, 1930s. Commercially available in 1963 with improvements made by McMullan and Oatley at Cambridge University. What is SEM?
  • 6. The basic principle is that a beam of electrons is generated by a suitable source, typically a tungsten filament (thermionic gun) or a field emission gun (FEG). The electron beam is accelerated through a high voltage (e.g. 20 kV) and pass through a system of apertures and electromagnetic lenses to produce a thin beam of electrons. Working Principles of SEM
  • 7. Working Principles of SEM A beam of electrons scans the surface of a sample (Figure). The electrons interact with the material in a way that triggers the emission of secondary electrons. These secondary electrons are captured by a detector, which forms an image of the surface of the sample. The direction of the emission of the secondary electrons depends on the orientation of the features of the surface. There, the image formed will reflect the characteristic feature of the region of the surface that was exposed to the electron beam.
  • 8. Major Components and their Functions
  • 9. Major Components and their Functions
  • 10. A tungsten (W) thermionic filament is used as electron gun. Other examples: Lanthanum or Cerium Hexaboride When high voltage is applied to tungsten filament which produces the steady stream of electrons. Electron gun located at the very top of a SEM fire a beam of electrons at the object under examination Electron Gun
  • 11. Thermionic Gun makes use of a filament that is heated up to make the electrons overcome the work function Field Emission Gun makes use of two anodes: (1) to extract electrons from the filament, (2) to accelerate the electrons to the operating voltage Types of Electron Gun
  • 12. 1. Thermionic electron gun Thermionic filaments are made of tungsten (W) in the form of a v-shaped wire. Thermionic emission is the liberation of electrons from an electrode by virtue of its temperature (releasing of energy supplied by heat). This occurs because the thermal energy given to the carrier overcomes the work function of the material. The charge carriers can be electrons or ions, and in literature are sometimes referred to as thermions. The classical example of thermionic emission is that of electrons from a hot cathode into a vacuum (also known as thermal electron emission or the Edison effect) in a vacuum tube. The hot cathode can be a metal filament e.g. Tungsten. Vacuum emission from metals tends to become significant only for temperatures over 1,000 K (730 °C). Types of Electron Gun
  • 13. 2. Field Emission Gun (FEG) For a field emission source, a fine, sharp, single crystal tungsten tip is employed. An FEG emitter gives a more coherent beam and its brightness is much higher than the tungsten filament. Electrons are emitted from a smaller area of the FEG source, giving a source size of a few nanometers, compared to around 50 μm for the tungsten filament. This leads to greatly improved image quality with the FEG source. Comparison of Thermionic (tungsten) and Field Emission Gun Types of Electron Gun
  • 14. There are two types of FEG sources: Cold and Schottky FEGs 1. Cold emission source, heating of the filament is not required as it operates at room temperature. However, this type of filament is prone to contamination and requires more stringent vacuum conditions (10-8 Pa, 10-10 torr). Regular and rapid heating (‘flashing’) is required in order to remove contamination. The spread of electron energies is very small for a cold field emitter (0.3 eV) and the source size is around 5 nm. 2. Thermal and Schottky sources, operate with lower field strengths. The surface of the tungsten (W) tip is covered with a thin layer of zirconium oxide (ZrO) to further decrease the potential barrier. The Schottky source is also heated and dispenses zirconium dioxide onto the tungsten tip to further lower its work function. The Schottky source is slightly larger, 20–30 nm, with a small energy spread (about 1 eV). Types of FEG Sources
  • 15. SEMs use lenses to produce clear and detailed images. They are not made of glass but rather magnets which are capable of bending the path of electrons. The lenses focus and control the electron beam, ensuring that the electrons end up precisely on the desired surface. Condenser – Lens system
  • 16. It accommodates the specimen holder and mechanisms for manipulating it and detectors for the various emissions. It must be very sturdy and insulated from vibration since the specimen must be kept extremely still for the microscope to produce clear images. SEM operates in a vacuum to avoid interference of the suspended particles with the movement of incident and reflected electron beams. Sample Chamber
  • 17. Electron Collection System It contains devices detect the various ways that the electron beam interacts with the sample object The detector common to all SEM systems is based on the design of Everhart and Thornley, which detects secondary and backscattered electrons
  • 18. Electron beam – Specimen interactions The result of the primary beam hitting the specimen is the formation of a teardrop shaped reaction vessel (Fig.). The reaction vessel by definition is where all the scattering events are taking place. An increase in the topography will increase the surface area of the reaction vessel resulting in more signal (Fig.). Figure. Formation of teardrop shaped reaction vessel Figure. Surface area of teardrop shaped reaction vessel
  • 19. Electron beam – Specimen interactions
  • 20. Electron beam – Specimen interactions
  • 21. Electron beam – Specimen interactions Six or more different events occur in the reaction vessel. These events include: 1. Backscattered electrons. A primary beam electron may be scattered in such a way that it escapes back from the specimen but does not go through the specimen. Backscattered electrons are the original beam electrons and thus, have a high energy level, near that of the gun voltage. Operating in the backscattered imaging mode is useful when relative atomic density information in conjunction with topographical information is to be displayed.
  • 22. Electron beam – Specimen interactions 2. Secondary electrons are generated when a primary electron dislodges a specimen electron from the specimen surface. Secondary electrons have a low energy level of only a few electron volts, thus, they can only be detected when they are dislodged near the surface of the reaction vessel. Therefore, secondary electrons cannot escape from deep within the reaction vessel. Two of the foremost reasons for operating in the secondary electron imaging mode are to obtain topographical information and high resolution. 3. X-rays. When electrons are dislodged from specific orbits of an atom in the specimen, X-rays are omitted. Elemental information can be obtained in the X-ray mode, because the X-ray generated has a wavelength and energy characteristic of the elemental atom from which it originated.
  • 23. Electron beam – Specimen interactions 4. Cathode Luminescence. Some specimen molecule's florescence when exposed to an electron beam. In the SEM, this reaction is called cathode luminescence. The florescence produces light photons that can be detected. A compound or structure labeled with a luminescent molecule can be detected by using cathode luminescence techniques. 5. Specimen Current. When the primary electron undergoes enough scattering such that the energy of the electron is decrease to a point where the electron is absorbed by the sample, this is known as specimen current. 6. Transmitted electrons. If the specimen is thin enough, primary electrons may pass through the specimen. These electrons are known as transmitted electrons and they provide some atomic density information. The atomic density information is displayed as a shadow. The higher the atomic number the darker the shadow until no electrons pass through the specimen.
  • 25. Detection of Sample Surface Secondary electrons emitted from the specimen are detected using a scintillator-photomultiplier, "Everhart-Thornley," detector. Low energy secondary electrons are emitted from the sample in all directions. They are gathered by a charged collector grid (or cage), which can be biased from -50 to +300 V. This draws the secondary electrons towards the scintillator.
  • 26. Detection of Sample Surface The scintillator is a thin plastic disk coated with a short-persistence phosphor that is highly efficient at converting the energy contained in the electrons into ultraviolet light photons (4000 Å). The response time of the phosphor is fast and permits high resolution scanning. The outer layer of the scintillator is coated with a thin layer [10-50 nm] of aluminum, positively biased at approximately 10 KeV, which accelerates the electrons to the scintillator surface. The charged collector grid, in addition to collecting secondary electrons from the sample, helps to alleviate some of the negative effects of the scintillator aluminum layer bias, which can actually distort the incident beam.
  • 27. The topographical aspects of a secondary electron image depend on how many of electrons actually reach the detector. When the incident electron beam intersects the edges of topographically high portions of a sample at lower angles, it puts more energy into the volume of secondary electron production. Thus, high points produce more secondary electrons, generating a larger signal. Faces oriented towards the detector also generate more secondary electrons. Secondary electrons that are prevented from reaching the detector do not contribute to the final image and these areas will appear as shadows or darker in contrast than those regions that have a clear electron path to the detector. Detection of Sample Surface
  • 28. Signals are collected by appropriate detector resulting signal is amplified and displayed. The topographic information contained in the image is straightforward. Detection of Sample Surface
  • 29. 1. Reduce the specimen into a size that fits the stub 2. Coat non-conducting specimens using either a vacuum evaporator or sputter coater. This step increases primary and secondary emissions, which enhances image quality. For conventional imaging in the SEM, specimens must be electrically conductive, at least at the surface, and electrically grounded to prevent the accumulation of electrostatic charge (i.e. using metal coating). Sample Preparation
  • 30. Non-conducting materials are usually coated with an ultra-thin coating of electrically conducting material, including gold, gold/palladium alloy, platinum, platinum/palladium, iridium, tungsten, and chromium. The recommended metals and sputter coating thickness are given below for tungsten and FEG sources: Metals: Au, Au/Pd (Tungsten source) Pt, Pd/Pt, Ir, W (FEG source) Thickness: 5-10 nm for low magnification 2-3 nm for high resolution, the thinner the better Sample Preparation
  • 31. Type of Specimen Cutting Coating Metallic   Ceramic, conducting   ceramic, non-conducting   Semiconductor   Polymer & polymer-matrix composites   Particles and fibers   Biological and botanical specimens   Sample Preparation
  • 32. Energy Dispersive X-ray spectroscopy Energy-dispersive X-ray spectroscopy (EDS) is an analytical technique used for the elemental analysis or chemical characterization of a sample. It relies on an interaction of some source of X-ray excitation and a sample. Its characterization is in large part to the fundamental principle that each element has a unique atomic structure allowing a unique set of peaks on its electromagnetic emission spectrum.
  • 33. A sample is excited under high energy of electron beam, the inner shell of electrons is ejected to vacuum creating a vacancy in that shell. Electrons from the outer shell jump into the vacant site for filling the inner shell. During this process, the sample fluoresces X-ray of energy same as the energy difference between the initial state and final state. Energy Dispersive X-ray spectroscopy Since each atom has its unique and discretized energy levels, the X-Ray fluorescence is also characteristic of that atom. Energy dispersive X ray spectroscopy is a technique that detects the X-ray fluorescence to characterize the elements present in a material. The most common detector used is Si(Li) detector cooled to cryogenic temperatures with liquid nitrogen. Now, newer systems are often equipped with silicon drift detectors (SDD) with Peltier cooling systems.
  • 34. Figure. EDS spectrum of the mineral crust of the vent shrimp. Most of these peaks are X-rays given off as electrons return to the K electron shell.(K- alpha and K-beta lines) One peak is from the L shell of iron. EDS Spectrum
  • 35. Advantages of SEM • Specimen size (small) • Samples must be conductive, otherwise non-conductive material should be metal coated • Only surface morphology can be obtained Limitations of SEM 1. SEM has a large depth of field, which allows more of a specimen to be in focus at one time 2. SEM also has much higher resolution, so closely spaced specimens can be magnified at much higher levels 3. Because the SEM uses electromagnets rather than lenses, the researcher has much more control in the degree of magnification All of these advantages, as well as the actual strikingly clear images, make the SEM one of the most useful instruments in research today
  • 36. SEM has a variety of applications in a number of scientific and industry- related fields, especially where characterizations of solid materials is beneficial. In addition to topographical, morphological and compositional information, an SEM can detect and analyze surface fractures, provide information in microstructures, examine surface contaminations, reveal spatial variations in chemical compositions, provide qualitative chemical analyses and identify crystalline structures. In addition, SEMs have practical industrial and technological applications such as semiconductor inspection, production line of miniscule products and assembly of microchips for computers. SEMs can be as essential research tool in fields such as life science, biology, gemology, medical and forensic science, metallurgy. Applications
  • 37. 1. Image morphological of samples 2. Image compositional and some bonding differences 3. Examine wet and dry samples while viewing them (only in an ESEM) 4. View frozen material (in a SEM with a cryostage) 5. Generate X-rays from samples for microanalysis (EDS) 6. View/map grain orientation/crystallographic orientation and study related information like heterogeneity Applications