Ionizing Radiation -How is Gray different from Sievert -Deterministic & Stochastic Radiation Risks -Air Kerma-Time, Distance and Shielding Principles -Dosimetry
9. These types of radiation are called “ionizing” radiation, because they have
enough energy to knock an electron out of its orbit around an atom – creating an ion.
13. Complete penetration; X-ray passes completely through tissue and into the image recording device.
Total absorption; X-ray energy is completely absorbed by the tissue. No imaging information results.
Partial absorption with scatter; Scattering involves a partial transfer of energy to tissue, with the
resulting scattered X-ray having less energy and a different trajectory. Scattered radiation tends to
degrade image quality and is the primary source of radiation exposure to operator and staff
There are three outcomes when X-rays traverse tissue.
17. How is Gray different from Sievert?
Absorbed
radiation dose
Gray
Biological
radiation dose
Sievert
Allow to estimate
risk in a tissue or
organ
The Sievert is similar to Gray but takes into account the
potential ability of the radiation to cause a biological effect
18. What you need to know?
Quick ABC’s of X-ray
Occupational Risks In Cath Lab
Radiation Safety Basics
Optimizing Radiation Safety in Cath Lab
19. How high is the patient exposure in cardiac
interventions in comparison to chest radiograph?
20. Radiation exposure distribution in the interventional cardiologist.
Radiation
exposure on the
left is almost
double that on
the right side.
23. Your exposure today may not be felt for years to come.
Radiation-induced cancers have a biological
latency of more than 10 years
Radiation in cardiology: can’t live without it !
Working towards zero operator exposure
24.
25. Bad backs and aching necks:
Occupational hazards of the cath lab
26. “Probabilistic”
Deterministic effects, which only occur above a certain dose threshold
Stochastic effects, which have a chance of occurring at any range of dose
27. Biologic Effects of Radiation :
Radiation skin (deterministic) effects
A. Dry desquamation
(Poikiloderma) at one month in
a patient receiving 11 Gy
calculated peak skin dose.
B. Skin Necrosis at 6 months in
a patient who received 18 Gy
calculated peak skin dose.
28. The wound on the right back healed into a scar while the injury on
the arm ultimately required grafting.
The arm was too close to the x-ray source.
29. Radiation can harm biological systems by damaging the DNA of cells.
If this damage is not properly repaired, the cells may divide in an uncontrolled
manner and cause cancer.
Biologic Effects of Radiation :
Cancer-inducing (Stochastic)
31. Dose exposure is described in terms of the following
parameters
Fluoroscopic Time (min): This is the time during a procedure that fluoroscopy
is used but does not include cine acquisition imaging. Therefore,
considered alone, it tends to underestimate the total
radiation dose received.
Air Kerma (Gy): The cumulative air kerma is a measure of X-ray energy
delivered to air at the interventional reference point (15 cm from the
isocenter in the direction of the focal spot). Kerma (Kinetic Energy Released in MAtter)
This measurement has been closely associated with
deterministic skin effects
Dose-Area Product (Gy.cm2): This is the cumulative sum of the instantaneous
air kerma and the X-ray field area.
This monitors the patient dose burden and is a good
indicator of stochastic effects.
32. The axis of rotation of the C-arm is
depicted as a dashed line.
The isocenter lies on the rotational
axis, between the source and detector.
Air Kerma is a measure of X-ray
energy delivered to air at the
interventional reference point
(15 cm from the isocenter in the
direction of the focal spot).
33.
34. What you need to know?
Quick ABC’s of X-ray
Occupational Risks In Cath Lab
Radiation Safety Basics
Optimizing Radiation Safety in Cath Lab
35.
36.
37. Justification
Appropriate selection of patients for cardiac imaging is
the first step toward enhancing radiation safety.
Circulation
November 4,
2014
Procedure justification and assuring the
right test is done on the right patient for
the right reason
38. No Idea What I’m DoingAs Low As Reasonably Achievable
I Have ”No Idea What I'm Doing”
ALARA
NIWID
39.
40. Using appropriate shielding, keeping a distance as safely as possible and
reducing radiation time are essential principles for radiation reduction
TDS
41. What you need to know?
Quick ABC’s of X-ray
Occupational Risks In Cath Lab
Radiation Safety Basics
Optimizing Radiation Safety in Cath Lab
42.
43. What is Dosimetry?
Dosimetry is the measurement of radiation dose received.
How
much
dose
did I
get?
Energy in the form of
radiation
Units of Dosimetry
Absorbed Dose
Equivalent Dose
Effective Dose
Do You Know Your
Radiation Dose
During Your Cath?
44. Summary of dose quantities commonly used in medical
imaging dosimetry with definitions and units.
Equivalent Dose (H) is the
absorbed dose (D)
multiplied by a radiation
weighting factor (WR)
H = D x WR
Effective Dose (E) is the
equivalent dose (H)
multiplied by a tissue
weighting factor (WT)
E = H x WT
45. Recommended use of at least two dosimeters, one above and one underneath the lead
apron. They allow risk estimation for the deterministic effects (such as cataract) and the
stochastic effects (such as cancer risks), respectively.
Dosimetry
46. Real Time Monitoring of Staff Dose in the
Cath Lab
Real time radiation dose monitoring in the cath lab enables staff to
see their level of exposure any time and can alert them when their
levels are spiking.
47. The patient should be placed
away from the radiation
source and close to the image
intensifier
A lower table setting
without changing the
source-intensifier
distance results in higher
dose due to proximity of
the patient to the
radiation source
Elevation of the image
intensifier results in higher
dose owing to geometric
magnification by the
intensifier
Low Subject-Image Distance
48. The influence of patient size to patient radiation dose
Thin patient Thick patient
For a larger patient, operators commonly increase the source to image distance(SID )as
well as lower the patient table to facilitate the size of the patient.
Such changes, in addition to high patient attenuation, contributes
to an increase of patient dose.
Source
Image
SID
49. Whenever possible, angulations should be avoided.
In lateral (or craniocaudal) angulations, x-rays cross more tissues, which increases
attenuation and decreases image quality. To compensate, the system increases the
beam energy to maintain image quality.
50. ( a ) Posteroanterior projection where the dose rate is less than the oblique angulation ( b )
Effect of angulation on patient dose.
51. Magnification
increases the
dose
*Field of view, diameter 17 cm
Dose rate = 0.6 mGy/s
*Field or view, diameter 12 cm
Dose rate = 1.23 mGy/s.
Normal mode Mag mode
*Field of view, diameter 25 cm
Dose rate= 0.3 mGy/s
The FOV will decrease,
and the dose delivered
to the patient’s skin
will increase
52. Benefits from using collimation
Optimal collimation on the area of interest allows significant dose reduction
The collimator is an adjustable lead shutter attached to the beam exit port of the x-ray
source that can be closed down to limit the area of the body that is irradiated.
By collimating the beam to the diagnostically appropriate field of view, you will minimize
radiation to the patient, as well as to yourself.
53. Does moving the X ray beam to different areas of the patient’s body during a
procedure have an effect on the exposure to the patient?
The peak skin dose is the absorbed dose at the skin location that has received the
highest dose. This quantity is used to predict a skin injury.
Spread the Dose
54. Minimize frame rate of fluoroscopy
Lower pulse rates lead to greater dose reduction per unit time.
(pulses per second)
A reduction of the fluoroscopic pulse rate from 15 frames/sec to 7.5 frames/sec
with a fluoroscopic mode to low dose reduces the radiation exposure by 67%.
55. The diagrams depict scatter radiation for a C-arm fluoroscopy system with the x-ray tube
under the table (left) and in lateral projection on the same side as the operator (right).
Note the high dose to the operator when standing on the same side of the patient as the
tube.
If the operator stands upright, scattered radiation to the face is perhaps one-fourth as
great as when the operator is leaning down toward the patient.
Short operators receive more radiation to the face than do tall operators. They may wish
to stand on a platform.
56. The lateral projection is not
recommended when the lead
shield is not protecting the
operator.
The lateral projection is
recommended when the
lead shield is protecting the
operator.
57. Do not step on fluoroscopy pedal
when not looking at screen
Decrease Cine Use
59. Radiation shields
A – image intensifier;
B – Articulated, ceiling-
mounted radiation
protection screen;
C – patient position;
D – table-side shields.
60. Patient position and shielding in the cath lab.
A – digital flat panel
detector mounted
on C-arm;
B – ceiling-mounted
articulated
protection screen;
C – monitors;
D – patient;
E – C-arm and image
control panel;
F – tableside
protective shielding.
62. The “Interventionalist disc disease”:
How to protect against it?
The Zero-Gravity™ radiation protection system
63. The future Robots Moving Into The Cath Lab
Robotic-Assisted PCI
The CorPath 200 cath lab robotic system is designed for more precise
movements and less radiation exposure to physicians during PCI.
65. 1
•Precautions to Minimize
Exposure to Patient and Operator
2
•Precautions to Specifically
Minimize Exposure to Operator
3
•Precautions to Specifically
Minimize Exposure to Patient
66. Precautions to Minimize Exposure to Patient and Operator
Utilize radiation only when imaging is necessary .Avoid the”heavy foot“
Minimize use of cine.
Minimize use of steep angles of X-ray beam.( LAO Cr – AP Cr )
Minimize use of magnification modes.
Minimize frame rate of fluoroscopy and cine.(7.5 frames/sec fluoroscopy
setting)
Keep the image detector close to the patient (low subject-image
distance)
Utilize collimation to the fullest extent possible.
Monitor radiation dose in real time.
67. Precautions to Specifically Minimize Exposure to Operator
Use and maintain appropriate protective lead garments.
Maximize distance of operator from X-ray source and patient.
Keep above-table (hanging) and below-table shields in optimal
position at all times.
Keep all body parts out of the field of view at all times
A robotic PCI system may be considered
68. Precautions to Specifically Minimize Exposure to Patient
Keep table height as high as comfortably possible for the operator.
Every 30 minutes, vary the imaging beam angle to minimize exposure to
any specific skin area .
Keep the patient's extremities out of the beam.