2. 2
What makes a good test?
1. Test is acceptable or compliant
2. Test is appropriate for the staff technical
ability
3. Test is robust, reagents and equipment
are reliable & stable in summer heat
4. Test is not affected by interferences
5. Test is accurate when staff performs the
test correctly
3. eXact®
Photometers
• 0.01 ppm (mg/L) precision
• 4 ml built-in, built-in cell
• Automatic countdown timing
• 20 Test memory for each direct read test
• Incorporates eXact®
Strip technology
patented EZ-3 method (Fill, Dip and Read)
• No color matching
• Lightweight and waterproof (IP-67) meter
• CE certified
3
5. eXact®
Micro 10 Instructions
5
I n d e x :
P a r a m e t e r P a g e M E N U
T o t a l A lk a lin it y 4 A L 1
p H 5 P H 2
F r e e C h lo r in e ( D P D - 1 ) 6 C L 3
C o m b in e d C h lo r in e ( D P D - 3 ) 7 C L 3
B r o m in e 8 C L 3
T o t a l C h lo r in e ( D P D - 4 ) 8 C L 3
P h o s p h a t e 9 P O 4
C a lc iu m H a r d n e s s 1 0 C A 5
C o p p e r 1 0 C U 8
C h lo r id e ( S a lt ) 1 1 C H 6
C y a n u r ic A c id 1 2 C Y 7
H y d r o g e n P e r o x id e C o n v e r s io n 1 5 C Y 7
A b o u t Y o u r P h o t o m e t e r 3
T ip s f o r B e s t A c c u r a c y 1 3
B a t t e r y In s t a lla t io n 1 4
W a r r a n t y ( 2 y e a r ) 1 4
R e o r d e r In f o r m a t io n 1 5
U S E P A C o m p lia n c e 2 , 1 6
K it C o m p o n e n t s 1 6
V is it u s o n lin e a t s e n s a f e .c o m /m ic r o 1 0 f o r u p -t o -d a t e
p r o d u c t in f o r m a t io n & N E W t e s t s a v a ila b le .
I D E A L F O R T E S T I N G P O O L S A N D S P A S .
A C C E P T E D B Y M O S T H E A L T H D E P A R T M E N T S
F O R C O M M E R C I A L P O O L A N D S P A T E S T I N G .
T h e e X a c t ®
M i c r o 1 0 A d v a n c e d
P h o t o m e t e r S y s t e m h a s b e e n
d e s i g n e d f o r u s e w i t h t h e e X a c t ®
S t r i p
M i c r o r e a g e n t d e l i v e r y s y s t e m .
®
A d v a n c e d P h o to m e te r S y s te m
U S E P A , D IN , & IS O C o m p lia n t f o r F r e e , C o m b in e d , & T o t a l C h lo r in e T e s t in g
( 4 5 0 0 - C L G , D IN S t a n d a r d 3 8 4 0 8 G 4 , IS O 7 3 9 3 / 2 )
®
M a n u f a c t u r e d B y : I n d u s t r ia l T e s t S y s te m s , I n c .
P h o n e : - I N S I D E T H E U . S .
- O U T S I D E T H E U . S .
F a x :
I T S @ S E N S A F E .C O M
W W W . S E N S A F E .C O M
w w w .p o o lc h e c k o n lin e .c o m
M i c r o 1 0 i s
M a n u f a c t u r e d
a n d t e s t e d i n
a n I S O 9 0 0 1
F a c i l i t y
T o p r e v e n t w a r r a n t y is s u e s , y o u m u s t r e g is te r y o u r
o w n e r s h ip w ith in 3 0 d a y s o f p u r c h a s e a n d in c lu d e th e s e r i a l
n u m b e r ( o n b o tto m o f M ic r o 1 0 M e te r ) a n d v a lid e -m a il
a d d r e s s . V is it o u r w e b s ite s e n s a f e .c o m /m ic r o 1 0 t o r e g is te r .
B y r e g is te r in g , y o u r 2 -y e a r w a r r a n ty is a c tiv a te d a n d y o u
w ill b e u p d a te d w h e n n e e d e d r e g a r d i n g th e M ic r o 1 0 .
R e v i s io n 0 5 /1 9 /1 4
S t a n d a r d K it
S t a r t e r K it
6. 6
2
1
2
e X a c t ®
R 0 5 1 9 1 4
A L 1
P H 2
C L 3
C L 3
P O 4
C A 5
C H 6
C Y 7
C Y 7
C U 8
2
+ 2
1 2
1 2
1 2
1
1
1
1
1
1 E x p e c t e d M e t e r
3
e X a c t ®
S t r ip M ic r o 1 0 h a s b e e n d e s ig n e d to o f f e r t h e u s e r a m o r e “ G r e e n ” a n d c o s t - e f fe c tiv e a lte r n a t iv e to te s t in g .
In s te a d o f u s in g a 1 0 m l w a te r s a m p le , e X a c t®
S tr ip M ic r o 1 0 u s e s a 4 m l w a t e r s a m p le , w h ic h u s e s u p to 6 0 %
le s s c h e m ic a l p e r t e s t. T h e a c c u r a c y o f th e m e te r is m a in ta in e d b y d e s ig n in g th e p h o t o c e ll w ith a 2 0 m m p a th le n g t h .
D im e n s io n s : In s t r u m e n t : 5 ( W ) x 3 . 5 ( D ) x 1 6 .5 ( H ) c m ;
( 2 x 1 .4 x 6 . 3 7 5 in )
42. • Ampoule (solution) used to verify system (operator, reagent,
and photometer) is performing correctly
• Available for Total Alkalinity, pH, Calcium Hardness, Copper,
Cyanuric Acid, Phosphate, and Salt,
• No dilution needed, just snap ampoule open, fill cell with
sample, and run test
42
P a r a m e t e r T e s tM e n u
I t e m
D e s i r e d
V a l u e
A c c e p t a b le
R a n g e
B o r d e r li n e
R a n g e
A L 1 T o t a l A lk a lin it y 5 2 p p m 4 2 - 6 2 3 8 - 6 7
P H 2 p H 7 . 5 p H 7 . 3 - 7 . 7 7 . 2 - 7 . 8
P O 4 P h o s p h a t e 2 p p m 1 . 7 - 2 . 3 1 . 6 - 2 . 4
C A 5 C a lc iu m a s C a C O 3 2 2 2 p p m 1 9 2 - 2 5 2 1 8 0 - 2 6 4
C Y 7 C y a n u r ic A c id 4 0 p p m 3 0 - 5 0 2 6 - 5 4
C U 8 C o p p e r a s C u + 2
0 . 2 2 p p m 0 . 1 7 - 0 . 2 7 0 . 1 4 - 0 . 3 0
C U 8 T o t a l C o p p e r a s C u + 2
0 . 6 4 p p m 0 . 5 6 - 0 . 7 2 p p m 0 . 5 3 - 0 . 7 5 p p m
R051914
Assigned Value Chart for READY SNAP®
1P (lot 4813) Solution:
43. 43
P a r a m e t e r T e s tM e n u
I t e m
D e s i r e d
V a l u e
A c c e p t a b l e
R a n g e
B o r d e r l i n e
R a n g e
A L 1 T o t a l A lk a lin it y 5 2 p p m 4 2 - 6 2 3 8 - 6 7
P H 2 p H 7 .5 p H 7 . 3 - 7 . 7 7 .2 - 7 .8
P O 4 P h o s p h a t e 2 p p m 1 . 7 - 2 . 3 1 .6 - 2 .4
C A 5 C a lc iu m a s C a C O 3 2 2 2 p p m 1 9 2 - 2 5 2 1 8 0 - 2 6 4
C Y 7 C y a n u r ic A c id 4 0 p p m 3 0 - 5 0 2 6 - 5 4
C U 8 C o p p e r a s C u + 2
0 .2 2 p p m 0 .1 7 - 0 . 2 7 0 . 1 4 - 0 .3 0
C U 8 T o t a l C o p p e r a s C u + 2
0 .6 4 p p m 0 .5 6 - 0 .7 2 p p m 0 . 5 3 - 0 . 7 5 p p m
R051914
Assigned Value Chart for READY SNAP®
1P (lot 4813) Solution:
1. Turn on the eXact®
Micro
three times with clean
2.
test method procedure you
want verified (for example
3.
Snap®
Verification Standard by
twisting the top 180° (see
Image 1
4 . S q u e e z e p la s t ic a m p o u le
s e e Im a g e 2
5.
sample followed with a
quick shake of the meter to
remaining water drops.
6. S q u e e z e p la s tic a m p o u le
s e e Im a g e 2
7 .
y o u d o n o r m a lly u s in g t h e
c o r r e c t p r o c e d u r e ( s e e
I m a g e 4 a n d r e f e r t o
p h o t o m e t e r m a n u a l
8 .
V a lu e c h a r t p r o v id e d b e lo w
a n d o n b a c k ( s e e I m a g e 5 .
9 . I f y o u r v a lu e is w ith in t h e
a r e o p e r a tin g th e e X a c t®
f o r t h is M E N U t e s t m e th o d .
10. If your value is
Borderline, review the
Instruction Manual and
verify that you are following
the proper procedure.
Step 1 through Step 8.
3
74
5
N O T E :
CA5 test
result
was 220
46. ITS ManufacturingITS Manufacturing
• Manufacturing system based on Good Manufacturing
Practices (GMP)
• Temperature and humidity controlled environment
• Quality Control laboratory
• 3 Research and Development laboratories
• Latest manufacturing and packaging equipment, with
2011 investment of $500,000 in new equipment
• Currently adding 10,000 square feet of additional
manufacturing space to complete Dec 2014
46
47. ITS Quality ControlITS Quality Control
• Material is QC tested during filter paper impregnation
and after final assembly
• Testing is performed with water standards prepared
with ACS grade chemicals and Industry Accepted
procedures
• Standards are verified by Industry Accepted method
using colorimetric, titrimetric and gravimetric methods
• All reagent strip lots are confirmed to be match listed
specifications
47
48. Shelf LifeShelf Life
• Accelerated Shelf Life Study
–Product is placed in 55° C ovens for fixed times that simulate 6 month, 12
month, 18 month, 24 month, and 30 month product aging. The material is
then tested for performance to estimate shelf life.
• Real Time Shelf Life Study
–Product is stored at room temperature and evaluated at six month
intervals to confirm real shelf life. This study is repeated for every product
at documented frequencies to update shelf life dating given to products
48