kV Meter
and
mA Meter
The x-ray machine takes energy from an
electrical source (usually a 220V outlet) and
converts it to two separate voltage streams:
• One stream is a low voltage source which
can be varied so that different amounts of
current can flow through the high resistance
wire that composes the filament in the
cathode. The current that flows through the
filament is measured in milliamperes (mA).
• The second stream is very high voltage,
measured in kilovolts (thousands of volts at
peak voltage -- kVp). This voltage is applied
across the gap between the anode and the
cathode.
When the kVp is
increased, the x-ray
photons that are
produced have a
shorter wavelength,
and thus have higher
energy.
The number of x-ray photons produced
(i.e. the intensity of the beam) depends
on the number of electrons that boil
off the filament. This depends on the
temperature of the heating element
which in turns depends on the current
flowing through the filament. The
amount of current flowing through the
filament is controlled by the mA
selector.
kV Meter
Kilovolt peak determines the quality
of the x-ray beam.
• When we say the quality of the x-ray
beam, we mean its energy and
penetrability.
–Penetrability refers to the ability of x-
rays to penetrate deeper in tissue.
• High energy x-rays are able to penetrate
tissue more deeply than low energy x-rays
A high kVp is available
at a relatively low mA
and vice versa.
↑ kVp = ↓ mA
↓ kVp = ↑ mA
A higher kVp will give your image
lower contrast [many shades of
gray (wide latitude)]. Lower kVp
will give your image a higher
contrast image [fewer shades of
gray (narrow latiude)].
↑ kVp = ↓ contrast = ↑ latitude
↓ kVp = ↑ contrast = ↓ latitude
X-rays are attenuated
exponentially.
– Attenuation is the reduction
in x-ray intensity that results
from absorption and
scattering.
Distance and mAs
DO NOT affect
radiation quality,
but they do affect
radiation quantity.
In radiography, the quality of x-rays is
measured by the Half Value Layer
(HVL).
The HVL of an x-ray beam is the
thickness of absorbing material
necessary to reduce the x-ray
intensity to half of its original value.
X-ray beam quality can be identified by
kVp or filtration, but HVL is most
appropriate as it is the best method
for specifying x-ray quality.
mA Meter
The number of electrons emitted by the
filament is determined by the temperature
of the filament.
– The filament temperature is in turn controlled
by the filament current, which is measured in
amperes (A).
– As the filament current increases, the
filament becomes hotter, and more
electrons are released by thermionic
emission.
↑ F. Current = Hotter F. = ↑ # of electrons
• Filaments normally operate at currents of
3 to 6 A.
• mA measures the x-ray
beam’s quantity, the number
of x-rays in the useful beam.
• X-ray quantity is also known
as :
–intensity of the x-ray beam
–radiation exposure.
• X-ray quantity is directly proportional to
mAs.
–When mAs is doubled, the number of
electrons striking the tube target is
doubled, therefore the number of x-rays
emitted is doubled.
↑ mAs = ↑ # of electrons = ↑ # of x-rays
Compensating for a change in SID (Source-
to-Image-Distance) by changing mAs by
the factor of SID2
is known as the
square law.
mAs1 = (SID1)2
_______ _________
mAs2 = (SID2)2
X-ray Quantity and kVp
I1 = (kVp1)2
__ _____
I2 = (kVp2)2
X-ray Quantity and Distance
aka INVERSE SQUARE LAW
I1 = (d1)2
__ ___
I2 = (d2)2
PREPARED BY GROUP TWO
JULABAR, MARONA YSABEL G.
ALIASUT, GAILRICH
PANGILINAN, RUFFA MAE
DATU, EUNICE LORENZ
MALIWAT, KEANA JAMILLA
RIVERA, JONIE
GAMBOA, DESIREE DAWN
OCOP, TRINA MILISE
PARAS, VANESSA YVRESSE
BALUYUT, PRINCESS PAULINE
PADILLON, GRACYN

Weitere ähnliche Inhalte

Was ist angesagt?

Radiographic film
Radiographic filmRadiographic film
Radiographic film
Rad Tech
 
Flouroscopic imging
Flouroscopic imgingFlouroscopic imging
Flouroscopic imging
Sahith Reddy
 
Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortion
parthajyotidas11
 
Radiographic exposure and image quality
Radiographic exposure and image qualityRadiographic exposure and image quality
Radiographic exposure and image quality
Rad Tech
 
Portable n mobile unit
Portable n mobile unitPortable n mobile unit
Portable n mobile unit
Sudil Paudyal
 

Was ist angesagt? (20)

Production of x rays & generators
Production of x rays & generatorsProduction of x rays & generators
Production of x rays & generators
 
Radiographic film
Radiographic filmRadiographic film
Radiographic film
 
Anode heel effect, line focus principle,
Anode heel effect, line focus principle,Anode heel effect, line focus principle,
Anode heel effect, line focus principle,
 
Factors affecting Quality and Quantity of X-ray beam
Factors affecting Quality and Quantity of X-ray beamFactors affecting Quality and Quantity of X-ray beam
Factors affecting Quality and Quantity of X-ray beam
 
Flouroscopic imging
Flouroscopic imgingFlouroscopic imging
Flouroscopic imging
 
Image characteristics of x ray film
Image characteristics of x ray filmImage characteristics of x ray film
Image characteristics of x ray film
 
Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortion
 
Beam restricted device and filter used in x ray
Beam restricted device and filter used in x rayBeam restricted device and filter used in x ray
Beam restricted device and filter used in x ray
 
Mobile & portable radiography
Mobile & portable radiographyMobile & portable radiography
Mobile & portable radiography
 
Radiographic exposure and image quality
Radiographic exposure and image qualityRadiographic exposure and image quality
Radiographic exposure and image quality
 
X ray generators
X ray generatorsX ray generators
X ray generators
 
Portable n mobile unit
Portable n mobile unitPortable n mobile unit
Portable n mobile unit
 
Fluroscopy
Fluroscopy Fluroscopy
Fluroscopy
 
X ray film
X  ray filmX  ray film
X ray film
 
MACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptxMACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptx
 
X-ray Tube (Cathode)
X-ray Tube (Cathode)X-ray Tube (Cathode)
X-ray Tube (Cathode)
 
Mammographic equipment
Mammographic equipmentMammographic equipment
Mammographic equipment
 
CHAPTER-7 - QUALITY OF X-RAY BEAM
 CHAPTER-7 - QUALITY OF X-RAY BEAM CHAPTER-7 - QUALITY OF X-RAY BEAM
CHAPTER-7 - QUALITY OF X-RAY BEAM
 
Fluoroscopy presentation
Fluoroscopy presentationFluoroscopy presentation
Fluoroscopy presentation
 
Fluroscopy
FluroscopyFluroscopy
Fluroscopy
 

Ähnlich wie kV and mA Meters

factorsaffecting-161206191607 (1).pdf
factorsaffecting-161206191607 (1).pdffactorsaffecting-161206191607 (1).pdf
factorsaffecting-161206191607 (1).pdf
CesarinaBlanca
 
X ray production & emission
X ray production & emissionX ray production & emission
X ray production & emission
airwave12
 

Ähnlich wie kV and mA Meters (20)

Factors affecting xray tube pptx
Factors affecting xray tube  pptxFactors affecting xray tube  pptx
Factors affecting xray tube pptx
 
factorsaffecting-161206191607 (1).pdf
factorsaffecting-161206191607 (1).pdffactorsaffecting-161206191607 (1).pdf
factorsaffecting-161206191607 (1).pdf
 
X ray production & emission
X ray production & emissionX ray production & emission
X ray production & emission
 
x-ray_safety_presentation__2_.ppt
x-ray_safety_presentation__2_.pptx-ray_safety_presentation__2_.ppt
x-ray_safety_presentation__2_.ppt
 
CLINICAL RADIATION GENERATORS
CLINICAL RADIATION GENERATORS CLINICAL RADIATION GENERATORS
CLINICAL RADIATION GENERATORS
 
X ray production (Emission and Filtration)
X ray production (Emission and Filtration)X ray production (Emission and Filtration)
X ray production (Emission and Filtration)
 
X-ray production & emission
X-ray production & emissionX-ray production & emission
X-ray production & emission
 
1-Radiation physics had.pptx
1-Radiation physics had.pptx1-Radiation physics had.pptx
1-Radiation physics had.pptx
 
Radiation physics
Radiation physicsRadiation physics
Radiation physics
 
Control of kV and mA in x-ray tube.pptx
Control of kV and mA in x-ray tube.pptxControl of kV and mA in x-ray tube.pptx
Control of kV and mA in x-ray tube.pptx
 
Production of x rays
Production of x raysProduction of x rays
Production of x rays
 
Linear Accelerators.pptx
Linear Accelerators.pptxLinear Accelerators.pptx
Linear Accelerators.pptx
 
xray production.pptx
xray production.pptxxray production.pptx
xray production.pptx
 
Linear accelerator
Linear acceleratorLinear accelerator
Linear accelerator
 
Chapter5 treatment machines for external beam
Chapter5 treatment machines for external beamChapter5 treatment machines for external beam
Chapter5 treatment machines for external beam
 
Chapter 3 Radiation Characteristics
Chapter 3 Radiation CharacteristicsChapter 3 Radiation Characteristics
Chapter 3 Radiation Characteristics
 
Production of x rays
Production of x rays Production of x rays
Production of x rays
 
Linear accelerator.pptx
Linear accelerator.pptxLinear accelerator.pptx
Linear accelerator.pptx
 
X-Ray Production.pptx
X-Ray Production.pptxX-Ray Production.pptx
X-Ray Production.pptx
 
X-Ray Tube complete-1.pptx
X-Ray Tube complete-1.pptxX-Ray Tube complete-1.pptx
X-Ray Tube complete-1.pptx
 

Mehr von Marona Ysabel Julabar

Mehr von Marona Ysabel Julabar (8)

Neuroanatomy: Cerebellum
Neuroanatomy: CerebellumNeuroanatomy: Cerebellum
Neuroanatomy: Cerebellum
 
Physiology of the Parathyroid Glands
Physiology of the Parathyroid GlandsPhysiology of the Parathyroid Glands
Physiology of the Parathyroid Glands
 
Female Endocrine Physiology
Female Endocrine PhysiologyFemale Endocrine Physiology
Female Endocrine Physiology
 
SKULL: Radiologic Positioning (Non-contrast)
SKULL: Radiologic Positioning (Non-contrast)SKULL: Radiologic Positioning (Non-contrast)
SKULL: Radiologic Positioning (Non-contrast)
 
Republic Act No. 9439
Republic Act No. 9439Republic Act No. 9439
Republic Act No. 9439
 
The Human Heart
The Human HeartThe Human Heart
The Human Heart
 
Reflection and Refraction
Reflection and RefractionReflection and Refraction
Reflection and Refraction
 
Social Doctrine of the Church and Encyclicals
Social Doctrine of the Church and EncyclicalsSocial Doctrine of the Church and Encyclicals
Social Doctrine of the Church and Encyclicals
 

Kürzlich hochgeladen

The Mariana Trench remarkable geological features on Earth.pptx
The Mariana Trench remarkable geological features on Earth.pptxThe Mariana Trench remarkable geological features on Earth.pptx
The Mariana Trench remarkable geological features on Earth.pptx
seri bangash
 
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 bAsymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
Sérgio Sacani
 
Digital Dentistry.Digital Dentistryvv.pptx
Digital Dentistry.Digital Dentistryvv.pptxDigital Dentistry.Digital Dentistryvv.pptx
Digital Dentistry.Digital Dentistryvv.pptx
MohamedFarag457087
 

Kürzlich hochgeladen (20)

❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.
❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.
❤Jammu Kashmir Call Girls 8617697112 Personal Whatsapp Number 💦✅.
 
Connaught Place, Delhi Call girls :8448380779 Model Escorts | 100% verified
Connaught Place, Delhi Call girls :8448380779 Model Escorts | 100% verifiedConnaught Place, Delhi Call girls :8448380779 Model Escorts | 100% verified
Connaught Place, Delhi Call girls :8448380779 Model Escorts | 100% verified
 
The Mariana Trench remarkable geological features on Earth.pptx
The Mariana Trench remarkable geological features on Earth.pptxThe Mariana Trench remarkable geological features on Earth.pptx
The Mariana Trench remarkable geological features on Earth.pptx
 
Factory Acceptance Test( FAT).pptx .
Factory Acceptance Test( FAT).pptx       .Factory Acceptance Test( FAT).pptx       .
Factory Acceptance Test( FAT).pptx .
 
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 bAsymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
 
PSYCHOSOCIAL NEEDS. in nursing II sem pptx
PSYCHOSOCIAL NEEDS. in nursing II sem pptxPSYCHOSOCIAL NEEDS. in nursing II sem pptx
PSYCHOSOCIAL NEEDS. in nursing II sem pptx
 
Site Acceptance Test .
Site Acceptance Test                    .Site Acceptance Test                    .
Site Acceptance Test .
 
Proteomics: types, protein profiling steps etc.
Proteomics: types, protein profiling steps etc.Proteomics: types, protein profiling steps etc.
Proteomics: types, protein profiling steps etc.
 
Pulmonary drug delivery system M.pharm -2nd sem P'ceutics
Pulmonary drug delivery system M.pharm -2nd sem P'ceuticsPulmonary drug delivery system M.pharm -2nd sem P'ceutics
Pulmonary drug delivery system M.pharm -2nd sem P'ceutics
 
Sector 62, Noida Call girls :8448380779 Model Escorts | 100% verified
Sector 62, Noida Call girls :8448380779 Model Escorts | 100% verifiedSector 62, Noida Call girls :8448380779 Model Escorts | 100% verified
Sector 62, Noida Call girls :8448380779 Model Escorts | 100% verified
 
Clean In Place(CIP).pptx .
Clean In Place(CIP).pptx                 .Clean In Place(CIP).pptx                 .
Clean In Place(CIP).pptx .
 
Grade 7 - Lesson 1 - Microscope and Its Functions
Grade 7 - Lesson 1 - Microscope and Its FunctionsGrade 7 - Lesson 1 - Microscope and Its Functions
Grade 7 - Lesson 1 - Microscope and Its Functions
 
COST ESTIMATION FOR A RESEARCH PROJECT.pptx
COST ESTIMATION FOR A RESEARCH PROJECT.pptxCOST ESTIMATION FOR A RESEARCH PROJECT.pptx
COST ESTIMATION FOR A RESEARCH PROJECT.pptx
 
9999266834 Call Girls In Noida Sector 22 (Delhi) Call Girl Service
9999266834 Call Girls In Noida Sector 22 (Delhi) Call Girl Service9999266834 Call Girls In Noida Sector 22 (Delhi) Call Girl Service
9999266834 Call Girls In Noida Sector 22 (Delhi) Call Girl Service
 
Human & Veterinary Respiratory Physilogy_DR.E.Muralinath_Associate Professor....
Human & Veterinary Respiratory Physilogy_DR.E.Muralinath_Associate Professor....Human & Veterinary Respiratory Physilogy_DR.E.Muralinath_Associate Professor....
Human & Veterinary Respiratory Physilogy_DR.E.Muralinath_Associate Professor....
 
Vip profile Call Girls In Lonavala 9748763073 For Genuine Sex Service At Just...
Vip profile Call Girls In Lonavala 9748763073 For Genuine Sex Service At Just...Vip profile Call Girls In Lonavala 9748763073 For Genuine Sex Service At Just...
Vip profile Call Girls In Lonavala 9748763073 For Genuine Sex Service At Just...
 
Digital Dentistry.Digital Dentistryvv.pptx
Digital Dentistry.Digital Dentistryvv.pptxDigital Dentistry.Digital Dentistryvv.pptx
Digital Dentistry.Digital Dentistryvv.pptx
 
High Profile 🔝 8250077686 📞 Call Girls Service in GTB Nagar🍑
High Profile 🔝 8250077686 📞 Call Girls Service in GTB Nagar🍑High Profile 🔝 8250077686 📞 Call Girls Service in GTB Nagar🍑
High Profile 🔝 8250077686 📞 Call Girls Service in GTB Nagar🍑
 
Locating and isolating a gene, FISH, GISH, Chromosome walking and jumping, te...
Locating and isolating a gene, FISH, GISH, Chromosome walking and jumping, te...Locating and isolating a gene, FISH, GISH, Chromosome walking and jumping, te...
Locating and isolating a gene, FISH, GISH, Chromosome walking and jumping, te...
 
Introduction to Viruses
Introduction to VirusesIntroduction to Viruses
Introduction to Viruses
 

kV and mA Meters

  • 2. The x-ray machine takes energy from an electrical source (usually a 220V outlet) and converts it to two separate voltage streams: • One stream is a low voltage source which can be varied so that different amounts of current can flow through the high resistance wire that composes the filament in the cathode. The current that flows through the filament is measured in milliamperes (mA). • The second stream is very high voltage, measured in kilovolts (thousands of volts at peak voltage -- kVp). This voltage is applied across the gap between the anode and the cathode.
  • 3. When the kVp is increased, the x-ray photons that are produced have a shorter wavelength, and thus have higher energy.
  • 4. The number of x-ray photons produced (i.e. the intensity of the beam) depends on the number of electrons that boil off the filament. This depends on the temperature of the heating element which in turns depends on the current flowing through the filament. The amount of current flowing through the filament is controlled by the mA selector.
  • 5. kV Meter Kilovolt peak determines the quality of the x-ray beam. • When we say the quality of the x-ray beam, we mean its energy and penetrability. –Penetrability refers to the ability of x- rays to penetrate deeper in tissue. • High energy x-rays are able to penetrate tissue more deeply than low energy x-rays
  • 6. A high kVp is available at a relatively low mA and vice versa. ↑ kVp = ↓ mA ↓ kVp = ↑ mA
  • 7. A higher kVp will give your image lower contrast [many shades of gray (wide latitude)]. Lower kVp will give your image a higher contrast image [fewer shades of gray (narrow latiude)]. ↑ kVp = ↓ contrast = ↑ latitude ↓ kVp = ↑ contrast = ↓ latitude
  • 8. X-rays are attenuated exponentially. – Attenuation is the reduction in x-ray intensity that results from absorption and scattering.
  • 9. Distance and mAs DO NOT affect radiation quality, but they do affect radiation quantity.
  • 10. In radiography, the quality of x-rays is measured by the Half Value Layer (HVL). The HVL of an x-ray beam is the thickness of absorbing material necessary to reduce the x-ray intensity to half of its original value. X-ray beam quality can be identified by kVp or filtration, but HVL is most appropriate as it is the best method for specifying x-ray quality.
  • 11. mA Meter The number of electrons emitted by the filament is determined by the temperature of the filament. – The filament temperature is in turn controlled by the filament current, which is measured in amperes (A). – As the filament current increases, the filament becomes hotter, and more electrons are released by thermionic emission. ↑ F. Current = Hotter F. = ↑ # of electrons • Filaments normally operate at currents of 3 to 6 A.
  • 12. • mA measures the x-ray beam’s quantity, the number of x-rays in the useful beam. • X-ray quantity is also known as : –intensity of the x-ray beam –radiation exposure.
  • 13. • X-ray quantity is directly proportional to mAs. –When mAs is doubled, the number of electrons striking the tube target is doubled, therefore the number of x-rays emitted is doubled. ↑ mAs = ↑ # of electrons = ↑ # of x-rays
  • 14. Compensating for a change in SID (Source- to-Image-Distance) by changing mAs by the factor of SID2 is known as the square law. mAs1 = (SID1)2 _______ _________ mAs2 = (SID2)2
  • 15. X-ray Quantity and kVp I1 = (kVp1)2 __ _____ I2 = (kVp2)2
  • 16. X-ray Quantity and Distance aka INVERSE SQUARE LAW I1 = (d1)2 __ ___ I2 = (d2)2
  • 17. PREPARED BY GROUP TWO JULABAR, MARONA YSABEL G. ALIASUT, GAILRICH PANGILINAN, RUFFA MAE DATU, EUNICE LORENZ MALIWAT, KEANA JAMILLA RIVERA, JONIE GAMBOA, DESIREE DAWN OCOP, TRINA MILISE PARAS, VANESSA YVRESSE BALUYUT, PRINCESS PAULINE PADILLON, GRACYN