SlideShare ist ein Scribd-Unternehmen logo
1 von 4
Downloaden Sie, um offline zu lesen
Silver/silver chloride electrode
This electrode is more rarely used but its robustness may make it worth
using. Its simplicity also makes it useful for miniaturized assemblies. It is
most often used in combination with a glass electrode or with ion
selective electrodes.
This electrode makes it possible to work at temperatures ranging from
−30 to +135°C. At 25°C, when filled with saturated potassium chloride,
15.2.4 Electrometric Checking and Calibration
The pH meter has to be calibrated before measurements are made. The
apparatus chosen must be able to correct slope and temperature. In most
cases, a precision of ± 0.1 pH unit is sufficient as other causes of errors
are greater (e.g. heterogeneity of the sample). For measurements on
suspensions, it is preferable to choose a device with a separate reference
electrode as the risk of errors caused by clogging of the porous junction is
reduced and it is easier to access when problems occur.
The apparatus should be switched on some time before beginning
the measurements (there is often a waiting position which maintains the
circuits in equilibrium), and then the electrodes are immersed in the
appropriate buffer. It is advisable to begin with buffer T4AC (cf.
Appendix 3, at the end of this chapter), which is particularly stable with
respect to temperature. Adjust the pH meter to the appropriate value, then
move on to the second buffer, which should be very close to the values to
be measured. The apparatus should then display the value for the new
buffer (corrected for temperature, see Appendix 3). Compensate for any
slight variation while correcting slope. If correction is not possible, the
coupling of the electrodes may not be satisfactory, particularly due to the
glass electrode which does not display a strictly linear response between
the two pH values selected. If this is the case, choose a narrower range of
pH: for example, instead of pH 4 and 9, choose 7 and 9 or 6 and 8. The
closer the value of the standard buffer to the value to be measured, the
better the measurement. Agitation is not advised during measurement. In
the case of problems of linearity, carefully clean the porous junction of
the electrode by brushing it with a hard brush. It can also be cleaned by
suction using a filter pump.
Inorganic Analysis564
its potential is +200 mV compared to a standard hydrogen electrode (see
appendix 1). Its potential is −45 mV compared to the calomel electrode
°at 25 C. It is advisable to monitor the stability of this electrode, as
measurement currents can cause the transformation of silver by micro-
electrolysis; it can be improved by a thin coating of Teflon film.
Measurements of soil pH can be classified in four main types:
1. On aqueous suspensions of soils
2. On saturated pastes
3. On extracts of saturation
4. “In situ” measurements
15.2.5 Measurement on Aqueous Soil Suspensions
Procedure
Fig. 15.2. Recommended
position of the electrodes
– Weigh 10 g of soil dried at room temperature and sieved to 2 mm
(a test specimen of at least 5 mL being the ISO standard), add 25 mL
(or five times the volume of the test specimen, ISO standard) of boiled
distilled water (CO2 removed); shake for 1 h on an oscillating table
(energetic agitation for 5 min, ISO standard).
– let decant for 30 min (at least 2 h but not more than 24 h, ISO standard),
immerse the electrodes taking care that the porous part of the reference
electrode is submerged in the clear part of the suspension (Fig. 15.2).
Read the pH value after stabilization of the measurement. Note the
temperature, and check correction for temperature on the pH meter.
Remarks
As the value of the pH drops with a rise in temperature, it is usual
to bring back all the values to 25°C. For precise measurements, a
thermostatic bath should be used; as soil suspensions often have a buffering
effect that it is impossible to correct by calculation.
pH Measurement 565
(1998) recommends a 1:2.5 soil:water ratio. The International standard NF
ISO 10390 (1994) recommends a ratio of 1:5.
Specifications vary with the soil:solution ratio due to the diversity of soil
materials. In France, the experimental AFNOR standard NF X-31-103
Measurements should be made without agitation. For certain soils, the
indications provided by the pH meter may not be stable, giving a
permanent drift. In such cases, it is advisable to read the pH after a
specific time interval, for example, 3 min, and to use the same time
interval for the other measurements, not forgetting to mention the fact
when noting the results.
Measurements on soil suspensions and more importantly on saturated
pastes induce a phenomenon called “paste effect” or “suspension effect”
which can modify the results by ± 1 pH unit (Fig. 15.3). This effect is
especially significant when the electrodes – and particularly the reference
electrode – are in contact with the sediment. This effect could be due to
the difference in mobility of the K+
and Cl−
ions of the diffusion solution
of the reference electrode when colloids (charged particles, strong cation
exchange capacity) are present. Grewling and Peech (1960) reported that
this phenomenon could be minimized by taking measurements with a
0.01 molar solution of CaCl2 (pHCa).
Although the soil:solution ratio influences the results of measurement
of pH in water, it has very little effect in saline solutions such as CaCl2
0.01 mol L−1
(Conyers and Davey 1988). Nilsson et al. (1995)
recommended water: soil ratios higher than 10 (v:w or 2 in v:v) to obtain
reliable measurements of pHwater on organic soils.
The time of contact between water and soil also influences the
pH. According to Conyers and Davey (1988) agitation times of more than a
few hours are not advisable because they generate variations and make it
impossible to obtain a stable value. Qiu and Zhu (1986) reported
stabilization of their measurements after 30 min (and up to 1 h) of agitation.
Fig. 15.3. Possible errors due to the effect of suspension for a soil in distilled
water (Bates 1973)
Inorganic Analysis566
15.2.6 Determination of the pH-K and pH-Ca
When measured in water, the pH value does not take total acidity into
account, particularly the protons and the aluminium forms fixed on the
exchange complex which represent potential acidity. In addition to the
first measurement in water, it is consequently necessary to carry out a
measurement in 1 mol (KCl) L−1
aqueous solution with the same soil:
solution ratio (1:2.5 for the procedure described in “Procedure” under
Sect. 15.2.5) and using the same technique. This gives pHK.
pHK measurements usually give lower pH values2
than pHH2O
measurements. The difference (ΔpH) can be as much as one pH unit. A
value of ΔpH > 0 indicates that the cation exchange capacity is higher than
the anion exchange capacity. There is a significant correlation between a
positive ΔpH and exchange acidity. However, although this is true in the
case of one family of soils, it is not true of a general comparison.
Measurements in 0.01 mol (CaCl2) L−1
aqueous solution are also used
because Ca ions cause flocculation of the solution and minimize the paste
and dilution effect. pH expression is thus less random and comparison is
easier between different soils and particularly saline soils. pHCa
measurements induce lower ΔpH than pHK measurement The study of
Conyers and Davey (1988) showed that pHwater, pHK, and pHCa are closely
correlated on soils with a broad range of pH. For non-saline soils with a
negative net charge, they found the relation:
pHCa = 1.05 pHeau − 0.9
The procedure described in “Procedure” under Sect. 15.2.5 is valid in
all cases, whether in water or saline solution.
15.2.7 Measurement on “Saturated Pastes”
pH measurement on saturated paste aims to reproduce the conditions of
the natural environment as closely as possible. This technique is very
delicate to implement. It is necessary to start by preparing a “saturated
paste”:
2
In some Andosols rich in allophane, pHKCl can be higher than pHH2O (amphoteric
medium).
pH Measurement 567
This measurement was the subject of an experimental an AFNOR
standard NF-X31-104 (1988) and became part of the international NF
ISO 10390 standard (1994).

Weitere ähnliche Inhalte

Was ist angesagt?

Report on pH and conductivity of different samples from delhi
Report on pH and conductivity of different samples from delhiReport on pH and conductivity of different samples from delhi
Report on pH and conductivity of different samples from delhimegha bagri
 
Soil analysis
Soil analysisSoil analysis
Soil analysisAkma Ija
 
Item 6: Uog soil labs functions aspac brisbane_oct 2019
Item 6: Uog soil labs functions aspac brisbane_oct 2019Item 6: Uog soil labs functions aspac brisbane_oct 2019
Item 6: Uog soil labs functions aspac brisbane_oct 2019Soils FAO-GSP
 
Liquid-Liquid Equilibria of Nitrobenzene-Inorganic Acid Systems at 298.15 K
Liquid-Liquid Equilibria of Nitrobenzene-Inorganic Acid Systems at 298.15 KLiquid-Liquid Equilibria of Nitrobenzene-Inorganic Acid Systems at 298.15 K
Liquid-Liquid Equilibria of Nitrobenzene-Inorganic Acid Systems at 298.15 KReddysuresh Kolavali
 
Dissolved Oxygen Measurement Techniques and Aplications for the Water Quality...
Dissolved Oxygen Measurement Techniques and Aplications for the Water Quality...Dissolved Oxygen Measurement Techniques and Aplications for the Water Quality...
Dissolved Oxygen Measurement Techniques and Aplications for the Water Quality...Valentina Giraldo
 
Paper id 312201524
Paper id 312201524Paper id 312201524
Paper id 312201524IJRAT
 
A water quality field study 2010
A water quality field study 2010A water quality field study 2010
A water quality field study 2010pauleyler
 
Absolute isotopic scale for deuterium analysis of natural waters
Absolute isotopic scale for deuterium analysis of natural watersAbsolute isotopic scale for deuterium analysis of natural waters
Absolute isotopic scale for deuterium analysis of natural watersMahbubul Hassan
 
A STUDY ON OCEAN ACIDIFICATION DUE TO CARBON DIOXIDE ALONG THE COAST OF VISAK...
A STUDY ON OCEAN ACIDIFICATION DUE TO CARBON DIOXIDE ALONG THE COAST OF VISAK...A STUDY ON OCEAN ACIDIFICATION DUE TO CARBON DIOXIDE ALONG THE COAST OF VISAK...
A STUDY ON OCEAN ACIDIFICATION DUE TO CARBON DIOXIDE ALONG THE COAST OF VISAK...Soma Sekhar Sriadibhatla
 
Tensammetric Analysis of Nonionic Surfactant Mixtures by Artificial Neural Ne...
Tensammetric Analysis of Nonionic Surfactant Mixtures by Artificial Neural Ne...Tensammetric Analysis of Nonionic Surfactant Mixtures by Artificial Neural Ne...
Tensammetric Analysis of Nonionic Surfactant Mixtures by Artificial Neural Ne...Fatemeh Sedaghatpour
 

Was ist angesagt? (20)

Exp 5
Exp 5Exp 5
Exp 5
 
EPC - Lab Report One
EPC - Lab Report OneEPC - Lab Report One
EPC - Lab Report One
 
Exp 4
Exp  4Exp  4
Exp 4
 
03 002
03 00203 002
03 002
 
Report on pH and conductivity of different samples from delhi
Report on pH and conductivity of different samples from delhiReport on pH and conductivity of different samples from delhi
Report on pH and conductivity of different samples from delhi
 
Exp 1 (B)
Exp 1 (B)Exp 1 (B)
Exp 1 (B)
 
Soil analysis
Soil analysisSoil analysis
Soil analysis
 
Item 6: Uog soil labs functions aspac brisbane_oct 2019
Item 6: Uog soil labs functions aspac brisbane_oct 2019Item 6: Uog soil labs functions aspac brisbane_oct 2019
Item 6: Uog soil labs functions aspac brisbane_oct 2019
 
Liquid-Liquid Equilibria of Nitrobenzene-Inorganic Acid Systems at 298.15 K
Liquid-Liquid Equilibria of Nitrobenzene-Inorganic Acid Systems at 298.15 KLiquid-Liquid Equilibria of Nitrobenzene-Inorganic Acid Systems at 298.15 K
Liquid-Liquid Equilibria of Nitrobenzene-Inorganic Acid Systems at 298.15 K
 
Environmental Engineering Lab Manual
Environmental Engineering Lab ManualEnvironmental Engineering Lab Manual
Environmental Engineering Lab Manual
 
Dissolved Oxygen Measurement Techniques and Aplications for the Water Quality...
Dissolved Oxygen Measurement Techniques and Aplications for the Water Quality...Dissolved Oxygen Measurement Techniques and Aplications for the Water Quality...
Dissolved Oxygen Measurement Techniques and Aplications for the Water Quality...
 
Soil science damini
Soil science daminiSoil science damini
Soil science damini
 
Paper id 312201524
Paper id 312201524Paper id 312201524
Paper id 312201524
 
Exp 1 (A)
Exp 1 (A)Exp 1 (A)
Exp 1 (A)
 
A water quality field study 2010
A water quality field study 2010A water quality field study 2010
A water quality field study 2010
 
Absolute isotopic scale for deuterium analysis of natural waters
Absolute isotopic scale for deuterium analysis of natural watersAbsolute isotopic scale for deuterium analysis of natural waters
Absolute isotopic scale for deuterium analysis of natural waters
 
pH indicator
pH indicatorpH indicator
pH indicator
 
A STUDY ON OCEAN ACIDIFICATION DUE TO CARBON DIOXIDE ALONG THE COAST OF VISAK...
A STUDY ON OCEAN ACIDIFICATION DUE TO CARBON DIOXIDE ALONG THE COAST OF VISAK...A STUDY ON OCEAN ACIDIFICATION DUE TO CARBON DIOXIDE ALONG THE COAST OF VISAK...
A STUDY ON OCEAN ACIDIFICATION DUE TO CARBON DIOXIDE ALONG THE COAST OF VISAK...
 
Tensammetric Analysis of Nonionic Surfactant Mixtures by Artificial Neural Ne...
Tensammetric Analysis of Nonionic Surfactant Mixtures by Artificial Neural Ne...Tensammetric Analysis of Nonionic Surfactant Mixtures by Artificial Neural Ne...
Tensammetric Analysis of Nonionic Surfactant Mixtures by Artificial Neural Ne...
 
Lab Report 1
Lab Report 1Lab Report 1
Lab Report 1
 

Andere mochten auch

XRF Analysis of Soil Samples from How Caverns
XRF Analysis of Soil Samples from How CavernsXRF Analysis of Soil Samples from How Caverns
XRF Analysis of Soil Samples from How CavernsJoseph Stabb, ABD
 
Characteristics of different types of soils
Characteristics of different types of soilsCharacteristics of different types of soils
Characteristics of different types of soilsParth Joshi
 
Different types of soil
Different types of soilDifferent types of soil
Different types of soilYash Kishore
 
Soil ppt[1][1]
Soil ppt[1][1]Soil ppt[1][1]
Soil ppt[1][1]trn2allah
 
Introduction to Soil Science
Introduction to Soil ScienceIntroduction to Soil Science
Introduction to Soil ScienceAndy Kleinschmidt
 
Properties of soils (teach)
Properties of soils (teach) Properties of soils (teach)
Properties of soils (teach) Moira Whitehouse
 

Andere mochten auch (11)

Ideafest Poster_Draft
Ideafest Poster_DraftIdeafest Poster_Draft
Ideafest Poster_Draft
 
FS205E
FS205EFS205E
FS205E
 
XRF Analysis of Soil Samples from How Caverns
XRF Analysis of Soil Samples from How CavernsXRF Analysis of Soil Samples from How Caverns
XRF Analysis of Soil Samples from How Caverns
 
Soil properties
Soil propertiesSoil properties
Soil properties
 
Characteristics of different types of soils
Characteristics of different types of soilsCharacteristics of different types of soils
Characteristics of different types of soils
 
Different types of soil
Different types of soilDifferent types of soil
Different types of soil
 
Soil ppt[1][1]
Soil ppt[1][1]Soil ppt[1][1]
Soil ppt[1][1]
 
Soil properties and
Soil properties andSoil properties and
Soil properties and
 
Introduction to Soil Science
Introduction to Soil ScienceIntroduction to Soil Science
Introduction to Soil Science
 
Properties of soils (teach)
Properties of soils (teach) Properties of soils (teach)
Properties of soils (teach)
 
Soil ppt
Soil pptSoil ppt
Soil ppt
 

Ähnlich wie pH of soil

Ähnlich wie pH of soil (20)

pH meter
pH meterpH meter
pH meter
 
DONDUCTOMETRIC TITRATION.
DONDUCTOMETRIC TITRATION.DONDUCTOMETRIC TITRATION.
DONDUCTOMETRIC TITRATION.
 
Physical Charecteristics of water
Physical Charecteristics of waterPhysical Charecteristics of water
Physical Charecteristics of water
 
pH electrodes
pH electrodespH electrodes
pH electrodes
 
CE8512- WATER & WASTE WATER ANALYSIS LAB MANUVAL
CE8512- WATER & WASTE WATER ANALYSIS LAB MANUVALCE8512- WATER & WASTE WATER ANALYSIS LAB MANUVAL
CE8512- WATER & WASTE WATER ANALYSIS LAB MANUVAL
 
Ph electrodes
Ph electrodesPh electrodes
Ph electrodes
 
DIFFERENT pH MEASUREMENT TECHNIQUE.pdf
DIFFERENT pH MEASUREMENT TECHNIQUE.pdfDIFFERENT pH MEASUREMENT TECHNIQUE.pdf
DIFFERENT pH MEASUREMENT TECHNIQUE.pdf
 
Experiment5 chem210
Experiment5 chem210Experiment5 chem210
Experiment5 chem210
 
pH - MEASUREMENT .ppt
pH - MEASUREMENT .pptpH - MEASUREMENT .ppt
pH - MEASUREMENT .ppt
 
is.3025.11.1983.pdf
is.3025.11.1983.pdfis.3025.11.1983.pdf
is.3025.11.1983.pdf
 
PA-I Condutometry.(HRB)
PA-I Condutometry.(HRB)PA-I Condutometry.(HRB)
PA-I Condutometry.(HRB)
 
unit 2.pptx
unit 2.pptxunit 2.pptx
unit 2.pptx
 
Ph meter.pptx
Ph meter.pptxPh meter.pptx
Ph meter.pptx
 
pH metry.pptx
pH metry.pptxpH metry.pptx
pH metry.pptx
 
Analisis de agua merck (1)
Analisis de agua merck (1)Analisis de agua merck (1)
Analisis de agua merck (1)
 
Reduce pH Measurement Errors
Reduce pH Measurement ErrorsReduce pH Measurement Errors
Reduce pH Measurement Errors
 
Conductometry
ConductometryConductometry
Conductometry
 
Conductometry (Pharmaceutical analysis)
Conductometry (Pharmaceutical analysis)Conductometry (Pharmaceutical analysis)
Conductometry (Pharmaceutical analysis)
 
JAndrew/PH meter.pptx
JAndrew/PH meter.pptxJAndrew/PH meter.pptx
JAndrew/PH meter.pptx
 
Amperometry.pptx
Amperometry.pptxAmperometry.pptx
Amperometry.pptx
 

pH of soil

  • 1. Silver/silver chloride electrode This electrode is more rarely used but its robustness may make it worth using. Its simplicity also makes it useful for miniaturized assemblies. It is most often used in combination with a glass electrode or with ion selective electrodes. This electrode makes it possible to work at temperatures ranging from −30 to +135°C. At 25°C, when filled with saturated potassium chloride, 15.2.4 Electrometric Checking and Calibration The pH meter has to be calibrated before measurements are made. The apparatus chosen must be able to correct slope and temperature. In most cases, a precision of ± 0.1 pH unit is sufficient as other causes of errors are greater (e.g. heterogeneity of the sample). For measurements on suspensions, it is preferable to choose a device with a separate reference electrode as the risk of errors caused by clogging of the porous junction is reduced and it is easier to access when problems occur. The apparatus should be switched on some time before beginning the measurements (there is often a waiting position which maintains the circuits in equilibrium), and then the electrodes are immersed in the appropriate buffer. It is advisable to begin with buffer T4AC (cf. Appendix 3, at the end of this chapter), which is particularly stable with respect to temperature. Adjust the pH meter to the appropriate value, then move on to the second buffer, which should be very close to the values to be measured. The apparatus should then display the value for the new buffer (corrected for temperature, see Appendix 3). Compensate for any slight variation while correcting slope. If correction is not possible, the coupling of the electrodes may not be satisfactory, particularly due to the glass electrode which does not display a strictly linear response between the two pH values selected. If this is the case, choose a narrower range of pH: for example, instead of pH 4 and 9, choose 7 and 9 or 6 and 8. The closer the value of the standard buffer to the value to be measured, the better the measurement. Agitation is not advised during measurement. In the case of problems of linearity, carefully clean the porous junction of the electrode by brushing it with a hard brush. It can also be cleaned by suction using a filter pump. Inorganic Analysis564 its potential is +200 mV compared to a standard hydrogen electrode (see appendix 1). Its potential is −45 mV compared to the calomel electrode °at 25 C. It is advisable to monitor the stability of this electrode, as measurement currents can cause the transformation of silver by micro- electrolysis; it can be improved by a thin coating of Teflon film.
  • 2. Measurements of soil pH can be classified in four main types: 1. On aqueous suspensions of soils 2. On saturated pastes 3. On extracts of saturation 4. “In situ” measurements 15.2.5 Measurement on Aqueous Soil Suspensions Procedure Fig. 15.2. Recommended position of the electrodes – Weigh 10 g of soil dried at room temperature and sieved to 2 mm (a test specimen of at least 5 mL being the ISO standard), add 25 mL (or five times the volume of the test specimen, ISO standard) of boiled distilled water (CO2 removed); shake for 1 h on an oscillating table (energetic agitation for 5 min, ISO standard). – let decant for 30 min (at least 2 h but not more than 24 h, ISO standard), immerse the electrodes taking care that the porous part of the reference electrode is submerged in the clear part of the suspension (Fig. 15.2). Read the pH value after stabilization of the measurement. Note the temperature, and check correction for temperature on the pH meter. Remarks As the value of the pH drops with a rise in temperature, it is usual to bring back all the values to 25°C. For precise measurements, a thermostatic bath should be used; as soil suspensions often have a buffering effect that it is impossible to correct by calculation. pH Measurement 565 (1998) recommends a 1:2.5 soil:water ratio. The International standard NF ISO 10390 (1994) recommends a ratio of 1:5. Specifications vary with the soil:solution ratio due to the diversity of soil materials. In France, the experimental AFNOR standard NF X-31-103
  • 3. Measurements should be made without agitation. For certain soils, the indications provided by the pH meter may not be stable, giving a permanent drift. In such cases, it is advisable to read the pH after a specific time interval, for example, 3 min, and to use the same time interval for the other measurements, not forgetting to mention the fact when noting the results. Measurements on soil suspensions and more importantly on saturated pastes induce a phenomenon called “paste effect” or “suspension effect” which can modify the results by ± 1 pH unit (Fig. 15.3). This effect is especially significant when the electrodes – and particularly the reference electrode – are in contact with the sediment. This effect could be due to the difference in mobility of the K+ and Cl− ions of the diffusion solution of the reference electrode when colloids (charged particles, strong cation exchange capacity) are present. Grewling and Peech (1960) reported that this phenomenon could be minimized by taking measurements with a 0.01 molar solution of CaCl2 (pHCa). Although the soil:solution ratio influences the results of measurement of pH in water, it has very little effect in saline solutions such as CaCl2 0.01 mol L−1 (Conyers and Davey 1988). Nilsson et al. (1995) recommended water: soil ratios higher than 10 (v:w or 2 in v:v) to obtain reliable measurements of pHwater on organic soils. The time of contact between water and soil also influences the pH. According to Conyers and Davey (1988) agitation times of more than a few hours are not advisable because they generate variations and make it impossible to obtain a stable value. Qiu and Zhu (1986) reported stabilization of their measurements after 30 min (and up to 1 h) of agitation. Fig. 15.3. Possible errors due to the effect of suspension for a soil in distilled water (Bates 1973) Inorganic Analysis566
  • 4. 15.2.6 Determination of the pH-K and pH-Ca When measured in water, the pH value does not take total acidity into account, particularly the protons and the aluminium forms fixed on the exchange complex which represent potential acidity. In addition to the first measurement in water, it is consequently necessary to carry out a measurement in 1 mol (KCl) L−1 aqueous solution with the same soil: solution ratio (1:2.5 for the procedure described in “Procedure” under Sect. 15.2.5) and using the same technique. This gives pHK. pHK measurements usually give lower pH values2 than pHH2O measurements. The difference (ΔpH) can be as much as one pH unit. A value of ΔpH > 0 indicates that the cation exchange capacity is higher than the anion exchange capacity. There is a significant correlation between a positive ΔpH and exchange acidity. However, although this is true in the case of one family of soils, it is not true of a general comparison. Measurements in 0.01 mol (CaCl2) L−1 aqueous solution are also used because Ca ions cause flocculation of the solution and minimize the paste and dilution effect. pH expression is thus less random and comparison is easier between different soils and particularly saline soils. pHCa measurements induce lower ΔpH than pHK measurement The study of Conyers and Davey (1988) showed that pHwater, pHK, and pHCa are closely correlated on soils with a broad range of pH. For non-saline soils with a negative net charge, they found the relation: pHCa = 1.05 pHeau − 0.9 The procedure described in “Procedure” under Sect. 15.2.5 is valid in all cases, whether in water or saline solution. 15.2.7 Measurement on “Saturated Pastes” pH measurement on saturated paste aims to reproduce the conditions of the natural environment as closely as possible. This technique is very delicate to implement. It is necessary to start by preparing a “saturated paste”: 2 In some Andosols rich in allophane, pHKCl can be higher than pHH2O (amphoteric medium). pH Measurement 567 This measurement was the subject of an experimental an AFNOR standard NF-X31-104 (1988) and became part of the international NF ISO 10390 standard (1994).