SlideShare ist ein Scribd-Unternehmen logo
1 von 23
Differential Scanning
Calorimetry
Presented to:
Dr. Mohsin Ali Raza

Presenter:
Umair manzoor
Roll# E11-312
6th Semester

Department of Metallurgy & Materials
Engineering
INTRODUCTION











This technique is used to study what happens to polymers/samples upon heating
It is used to study thermal transitions of a polymer/sample (the changes that take
place on heating)
Need to know about the calorimeter , it is one who measures the heat in or out
of the sample.
and differential calorimeter is one who measures the heat of the sample relative
to the reference.
and the differential scanning calorimeter does all of the above functions and
heats the sample with the linear temperature.
Both the sample and reference are maintained at nearly the same temperature

through out the experiment in DSC.


The technique was developed by E.S. Watson and M.J. O'Neill in 1962



Most common than TGA technique
PRINCIPLE:
It is a technique in which the energy necessary to establish a zero temperature
difference between the sample & reference material is measured as a function
of temperature.
 Here, sample & reference material are heated by separate heaters in such a
way that their temp are kept equal while these temp. are increased or
decreased linearly.
 During heating two types of reactions can be take place one is the
endothermic and the other is the exothermic.
Endothermic reaction:








If sample absorbs some amount of heat during phase transition then reaction
is said to be endothermic
In endothermic reaction more energy needed to maintain zero temp
difference between sample & reference.
E.g. Melting, boiling, sublimation, vaporization
•Exothermic reaction:
if sample released some amount of heat during phase transition, then
reaction is said to be exothermic.
In exothermic reaction, less energy needed to maintain zero temp difference
between sample & reference.
E.g crystallization, degradation, polymerization
• what happens to a polymer when heated?
The polymer is heated in a device that looks
something like this:
•There are two pans, In sample pan, polymer is added, while the other,
reference pan is left empty
•Each pan sits on top of heaters which are controlled by a computer
•The computer turns on heaters, and let them heat the two pans at a
specific rate.
• The computer makes absolutely sure that the heating rate stays exactly the
same throughout the experiment
Why heaters don’t heat at the same rate?
•The simple reason is that the two pans are different. One has polymer in it,
and one doesn't. The polymer sample means there is extra material in the
sample pan. Having extra material means that it will take more heat to keep
the temperature of the sample pan increasing at the same rate as the
reference pan
• So the heater underneath the sample pan has to work harder than the
heater underneath the reference pan. It has to put out more heat.
DSC Curve:

•The result of a DSC experiment is a curve of heat flux versus
temperature or versus time. There are two different conventions:
exothermic reactions in the sample shown with a positive or negative
peak, depending on the kind of technology used in the experiment.
•This curve can be used to calculate enthalpies of transitions, which is
done by integrating the peak corresponding to a given transition. The
enthalpy of transition can be expressed using equation: ΔH = KA

• Where ΔH is the enthalpy of transition,
K is the calorimetric constant,
A is the area under the peak.
•The calorimetric constant varies from instrument to instrument, and can
be determined by analyzing a well-characterized material of known
enthalpies of transition.
•Area under the peak is directly proportional to heat absorbed or
evolved by the reaction,
•Height of the peak is directly proportional to rate of the reaction
Factors affecting DSC curve:
Two types of factors effect the DSC curve,
1-Instrumental
factors

2-Sample
characteristics

1-Instrumental factors
a- Furnace heating rate
b- Recording or chart speed
c- Furnace atmosphere
d- Geometry of sample holder/location of sensors
e- Sensitivity of the recoding system
f-Composition of sample containers.
2-Sample characteristics
a- Amount of sample
b- Nature of sample
c- Sample packing
d- Solubility of evolved gases in the sample
e- Particle size
f- Heat of reaction
g- Thermal conductivity
TYPES OF DSC INSTRUMENTS:
DSC HAS TWO TYPES OF INSTRUMENTS

1 .Heat –


1 Heat – flux
DSC

flux DSC

Sample holder:

PLATINUM,ALUMINIUM,STAINLESS STEEL.





SENSORS:

Temperature sensors

2. Power Compensated DSC






Sensors:



Pt resistance thermocouple



Separate sensors and heaters for the

one block for both sample and reference

sample and reference sample .

Furnace:





cells .



Temperature:



thermal power, which is supplied to the heaters
to maintain the temperature of the sample and
reference at the program value

Furnace:

separate blocks for sample and
reference cells.

The temperature difference between the
sample and reference is converted to differential

Sample holder:

Aluminum, platinum, stainless steel pans

usually thermocouples, which are same for
the sample and reference

2 Power
compensation
DSC




. Temperature:
differential thermal power is supplied to the
heaters to maintain the temperature of the
sample and reference at the program value


Heat flux DSC



Power compensated DCS

HEAT FLUX DSC AND POWER COMPENSATED DCS
Sample shape for DSC:
• Keep sample thin.
• The shape of the sample like so that it Cover as much as the
bottom of pan as possible.
• To make sample thin, cut the sample but don’t crush.
• If sample in pellet shape, cut cross section.
• If powder, spread evenly over the bottom of the pan.
• different shapes for DSC are shown in fig.
Sample preparation:
•

Accurately-weigh samples (~3-20 mg)

• Small sample pans (0.1 mL) of inert or treated metals (Al, Pt, Ni, etc.)
are used for sample preparation.
• Several pan configurations, e.g., open, pinhole, or hermetically-sealed
(airtight) pans

• The same material and configuration should be used for the sample
and the reference
• Material should completely cover the bottom of the pan to ensure
good thermal contact,
• Avoid overfilling the pan to minimize thermal lag from the bulk of the
material to the sensor
What can DSC measure?
•

Glass transitions

• Melting and boiling points
•

Crystallisation time and temperature

•

Percent crystallinity

•

Heats of fusion and reactions

•

Specific heat capacity

•

Oxidative/thermal stability

•

Reaction kinetics

•

Purity
Determination of Heat Capacity:
• By the DSC we can find out the heat capacity of the polymers and then
after futher heating we can find out the glass transition temperature and
after this we can find out the crystallinity and then the melting
temperature and then putting all above factors we get the DSC curve.
• In DSC curve we get crystallinity peaks, while melting dip.

exothermic

endothermic

• The crystallization peak and the melting dip will only show up for
polymers that can form crystals and melts.
• Completely amorphous polymers won't show any crystallization, or
any melting either. But polymers with both crystalline and
amorphous domains, will show all the features
: Determination

of Heat
Capacity

•DSC plot can be used to determine Heat Capacity. Suppose a polymer is
being heated.
Heating two
pans
Difference in
Heat output of
two heaters

temperature
against

• When we start heating two pans, the computer will plot the difference in

heat output of the two heaters against temperature that is plot of heat
absorbed by the polymer against temperature. The plot will look like this at
first.
The heat flow is heat (q) supplied per unit time (t), whereas
The heating rate is temperature increase (ΔT) per unit time (t)

By dividing heat flow (q/t) by the heating rate (ΔT/t). It ends up with heat
supplied divided by the temperature increase, which is called heat
capacity
The Glass Transition
Temperature by DSC

the glass transition
is a second order
transition.

On further heating the polymer to a certain temperature,
plot will shift downward suddenly, like this:
No peak and
dig

This means there is more heat flow. There is an increase in the heat
capacity of the polymer. This happens because the polymer has just
gone through the glass transition. Because of this change in heat
capacity that occurs at the glass transition, we can use DSC to measure a
polymer's glass transition temperature.
Crystallization by DSC
After glass transition, the polymers have a lot of mobility.
when they reach the right temperature, they will give off enough
energy to move into very ordered arrangements, which is called
crystals.
After
crystallization

Given off
heat
exothermic

heater to sample pan has
not to put a lot of heat into
the polymer in order to
keep the temperature rising
at the same rate as that of
the reference pan

this drop in the heat flow
as a big peak in the plot of
heat flow versus
temperature

If you analyze a 100% amorphous
polymer, like polystrene, you
wouldn't get this peak, because
such materials don't crystallize
melting temperature
by DSC
If we heat our polymer past its Tc, eventually we'll reach another thermal
transition, called melting. When we reach the polymer's melting
temperature, Tm, the polymer crystals begin to fall apart, that is they melt
When the polymer crystals melt, they
must absorb heat in order to do so.
After melting

Heat is
absorbed

heater to sample pan has to put a
lot of heat into the polymer in
order to both melt the crystals
and keep the temperature rising
at the same rate as that of the
reference pan

endothermic

this increase in the heat
flow as a big dig in the
plot of heat flow versus
temperature
Putting It All
Together
we saw a step in the plot when the polymer was heated past its glass
transition temperature. Then we saw a big peak when the polymer
reached its crystallization temperature. Then finally we saw a big dip
when the polymer reached its melting temperature. To put them all
together, a whole plot will often look something like this:
Peak
at Tc

Step at
Tg

Dig at
Tm
Best way to perform DSC
thermal analysis technique:
•proper instrument calibration

•use purge gas (N2 or He) to remove corrosive off-gases
•small sample size
•good thermal contact between the sample and the temperaturesensing device
•proper sample encapsulation
•start temperature well below expected transition temperature
•slow scanning speeds
•avoid decomposition in the DSC
Advantages of
using DSC:
•instruments can be used at very high temperatures
•instruments are highly sensitive
•flexibility in sample volume/form
•characteristic transition or reaction temperatures can be
determined
•High resolution obtained
•High sensitivity
•Stability of the material
limitations:
•DSC generally unsuitable for two-phase mixtures

• Difficulties in test cell preparation in avoiding evaporation of
volatile Solvents
• DSC is generally only used for thermal screening of
isolated intermediates and products
• Does not detect gas generation
• uncertainty of heats of fusion and transition temperatures
Thank you

!

Weitere ähnliche Inhalte

Was ist angesagt?

Differential thermal analysis & Differential Scanning Calorimetry
Differential thermal analysis & Differential Scanning CalorimetryDifferential thermal analysis & Differential Scanning Calorimetry
Differential thermal analysis & Differential Scanning Calorimetrysamira mohammadpour
 
Presentation2dsc
Presentation2dscPresentation2dsc
Presentation2dscSrota Dawn
 
Differential thermal analysis(dta)
Differential thermal analysis(dta)Differential thermal analysis(dta)
Differential thermal analysis(dta)Amruta Balekundri
 
Differential scanning calorimetry
Differential scanning calorimetryDifferential scanning calorimetry
Differential scanning calorimetryushaSanmugaraj
 
DIFFERENTIAL THERMAL ANALYSIS (DTA), ppt
DIFFERENTIAL THERMAL ANALYSIS (DTA),  pptDIFFERENTIAL THERMAL ANALYSIS (DTA),  ppt
DIFFERENTIAL THERMAL ANALYSIS (DTA), pptshaisejacob
 
Differential Scanning Colorimetry
Differential Scanning ColorimetryDifferential Scanning Colorimetry
Differential Scanning ColorimetryGautam Jha
 
Thermogravimetric analysis ppt
Thermogravimetric analysis pptThermogravimetric analysis ppt
Thermogravimetric analysis pptyudhishthir singh
 
Differential scanning calorimetry
Differential scanning calorimetryDifferential scanning calorimetry
Differential scanning calorimetryKhalid Hussain
 
Thermogravimetric Analysis
Thermogravimetric AnalysisThermogravimetric Analysis
Thermogravimetric AnalysisMedhaJoshi13
 
Thermogravimetric analysis
Thermogravimetric analysisThermogravimetric analysis
Thermogravimetric analysisSuresh Selvaraj
 
THERMOGRAVIMETRY ANALYSIS [TGA] AS PER PCI[M.PHARM]
THERMOGRAVIMETRY ANALYSIS [TGA] AS PER PCI[M.PHARM]THERMOGRAVIMETRY ANALYSIS [TGA] AS PER PCI[M.PHARM]
THERMOGRAVIMETRY ANALYSIS [TGA] AS PER PCI[M.PHARM]Shikha Popali
 
Thermal method of analysis
Thermal method of analysis Thermal method of analysis
Thermal method of analysis GANESH NIGADE
 
Dsc: interpretaion and application
Dsc: interpretaion and applicationDsc: interpretaion and application
Dsc: interpretaion and applicationRahul Kumar Maurya
 
DIFFERENTIAL THERMAL ANALYSIS & DIFFERENTIAL SCANNING CALORIMETRY
DIFFERENTIAL THERMAL ANALYSIS & DIFFERENTIAL SCANNING CALORIMETRYDIFFERENTIAL THERMAL ANALYSIS & DIFFERENTIAL SCANNING CALORIMETRY
DIFFERENTIAL THERMAL ANALYSIS & DIFFERENTIAL SCANNING CALORIMETRYPrachi Pathak
 
Differential thermal analysis - instrumental methods of analysis
Differential thermal analysis - instrumental methods of analysis Differential thermal analysis - instrumental methods of analysis
Differential thermal analysis - instrumental methods of analysis SIVASWAROOP YARASI
 
Presentation on DSC (differential scanning calorimetry )
Presentation on DSC (differential scanning calorimetry )Presentation on DSC (differential scanning calorimetry )
Presentation on DSC (differential scanning calorimetry )Hamza Suharwardi
 

Was ist angesagt? (20)

Differential thermal analysis & Differential Scanning Calorimetry
Differential thermal analysis & Differential Scanning CalorimetryDifferential thermal analysis & Differential Scanning Calorimetry
Differential thermal analysis & Differential Scanning Calorimetry
 
DSC and DTA
DSC and DTADSC and DTA
DSC and DTA
 
Presentation2dsc
Presentation2dscPresentation2dsc
Presentation2dsc
 
Differential thermal analysis(dta)
Differential thermal analysis(dta)Differential thermal analysis(dta)
Differential thermal analysis(dta)
 
Differential scanning calorimetry
Differential scanning calorimetryDifferential scanning calorimetry
Differential scanning calorimetry
 
DIFFERENTIAL THERMAL ANALYSIS (DTA), ppt
DIFFERENTIAL THERMAL ANALYSIS (DTA),  pptDIFFERENTIAL THERMAL ANALYSIS (DTA),  ppt
DIFFERENTIAL THERMAL ANALYSIS (DTA), ppt
 
TGA and DSC ppt
TGA and DSC ppt   TGA and DSC ppt
TGA and DSC ppt
 
Differential Scanning Colorimetry
Differential Scanning ColorimetryDifferential Scanning Colorimetry
Differential Scanning Colorimetry
 
Dta presentation
Dta presentationDta presentation
Dta presentation
 
Thermogravimetric analysis ppt
Thermogravimetric analysis pptThermogravimetric analysis ppt
Thermogravimetric analysis ppt
 
Differential scanning calorimetry
Differential scanning calorimetryDifferential scanning calorimetry
Differential scanning calorimetry
 
Thermogravimetric Analysis
Thermogravimetric AnalysisThermogravimetric Analysis
Thermogravimetric Analysis
 
Differential Thermal Analysis (DTA)
Differential Thermal Analysis (DTA)Differential Thermal Analysis (DTA)
Differential Thermal Analysis (DTA)
 
Thermogravimetric analysis
Thermogravimetric analysisThermogravimetric analysis
Thermogravimetric analysis
 
THERMOGRAVIMETRY ANALYSIS [TGA] AS PER PCI[M.PHARM]
THERMOGRAVIMETRY ANALYSIS [TGA] AS PER PCI[M.PHARM]THERMOGRAVIMETRY ANALYSIS [TGA] AS PER PCI[M.PHARM]
THERMOGRAVIMETRY ANALYSIS [TGA] AS PER PCI[M.PHARM]
 
Thermal method of analysis
Thermal method of analysis Thermal method of analysis
Thermal method of analysis
 
Dsc: interpretaion and application
Dsc: interpretaion and applicationDsc: interpretaion and application
Dsc: interpretaion and application
 
DIFFERENTIAL THERMAL ANALYSIS & DIFFERENTIAL SCANNING CALORIMETRY
DIFFERENTIAL THERMAL ANALYSIS & DIFFERENTIAL SCANNING CALORIMETRYDIFFERENTIAL THERMAL ANALYSIS & DIFFERENTIAL SCANNING CALORIMETRY
DIFFERENTIAL THERMAL ANALYSIS & DIFFERENTIAL SCANNING CALORIMETRY
 
Differential thermal analysis - instrumental methods of analysis
Differential thermal analysis - instrumental methods of analysis Differential thermal analysis - instrumental methods of analysis
Differential thermal analysis - instrumental methods of analysis
 
Presentation on DSC (differential scanning calorimetry )
Presentation on DSC (differential scanning calorimetry )Presentation on DSC (differential scanning calorimetry )
Presentation on DSC (differential scanning calorimetry )
 

Andere mochten auch

Andere mochten auch (18)

Electrochemistry
ElectrochemistryElectrochemistry
Electrochemistry
 
3rd Year Undergraduate Cyclic Voltammetry Practical
3rd Year Undergraduate Cyclic Voltammetry Practical3rd Year Undergraduate Cyclic Voltammetry Practical
3rd Year Undergraduate Cyclic Voltammetry Practical
 
DSC & TGA
DSC & TGA DSC & TGA
DSC & TGA
 
Dsc instrument 1
Dsc instrument 1Dsc instrument 1
Dsc instrument 1
 
Cyclic Voltammetry
Cyclic VoltammetryCyclic Voltammetry
Cyclic Voltammetry
 
15ab1s0308
15ab1s030815ab1s0308
15ab1s0308
 
Thermoanalytical techniques
Thermoanalytical techniquesThermoanalytical techniques
Thermoanalytical techniques
 
Ppp Dsc 1 Thermal Analysis Fundamentals Of Analysis
Ppp  Dsc 1 Thermal Analysis  Fundamentals Of AnalysisPpp  Dsc 1 Thermal Analysis  Fundamentals Of Analysis
Ppp Dsc 1 Thermal Analysis Fundamentals Of Analysis
 
Thermal analysis
Thermal analysisThermal analysis
Thermal analysis
 
shiva ram
shiva ramshiva ram
shiva ram
 
Dta
DtaDta
Dta
 
Cyclic voltammetry
Cyclic voltammetryCyclic voltammetry
Cyclic voltammetry
 
Thermometiric titration
Thermometiric titrationThermometiric titration
Thermometiric titration
 
Thermometric titration
Thermometric titrationThermometric titration
Thermometric titration
 
Voltammetry
VoltammetryVoltammetry
Voltammetry
 
Seminar on polymorphism
Seminar on polymorphismSeminar on polymorphism
Seminar on polymorphism
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 

Ähnlich wie Dsc presentation

Differential_Scanning_Calorimetry.ppt
Differential_Scanning_Calorimetry.pptDifferential_Scanning_Calorimetry.ppt
Differential_Scanning_Calorimetry.pptZORAIZ HAIDER
 
Differential Scanning Calorimetry.pptx
Differential Scanning Calorimetry.pptxDifferential Scanning Calorimetry.pptx
Differential Scanning Calorimetry.pptxMustafaIdais1
 
differential scanning calorimetry.pptx
differential scanning calorimetry.pptxdifferential scanning calorimetry.pptx
differential scanning calorimetry.pptxShashidarMShashi
 
Differential scanning calorimetry DSC
Differential scanning calorimetry DSCDifferential scanning calorimetry DSC
Differential scanning calorimetry DSCHaneenAbukatab
 
Thermal techniques- DSC- Durgashree Diwakar
Thermal techniques- DSC- Durgashree DiwakarThermal techniques- DSC- Durgashree Diwakar
Thermal techniques- DSC- Durgashree DiwakarDurgashree Diwakar
 
DSC | DIFFERENTIAL SCANNING CALORIMETRY
DSC | DIFFERENTIAL SCANNING CALORIMETRYDSC | DIFFERENTIAL SCANNING CALORIMETRY
DSC | DIFFERENTIAL SCANNING CALORIMETRYKarthic27
 
thermaltechniquesdsc-190602083135.pdf
thermaltechniquesdsc-190602083135.pdfthermaltechniquesdsc-190602083135.pdf
thermaltechniquesdsc-190602083135.pdfKajalMIshra63
 
Thermal charactrization of polymer
Thermal charactrization of polymerThermal charactrization of polymer
Thermal charactrization of polymerSharad Ghodake
 
Differential scanning Colorimetry (DSC).
Differential scanning Colorimetry (DSC).Differential scanning Colorimetry (DSC).
Differential scanning Colorimetry (DSC).Poonam Aher Patil
 
thermogravimetric analysis
thermogravimetric analysisthermogravimetric analysis
thermogravimetric analysisDalpat Singh
 
Differential Scanning Calorimetry (DSC).pdf
Differential Scanning Calorimetry (DSC).pdfDifferential Scanning Calorimetry (DSC).pdf
Differential Scanning Calorimetry (DSC).pdfSanDeepSharma926061
 
''Differential scanning calorimetry by Idrish Ansari''.pdf
''Differential scanning calorimetry by Idrish Ansari''.pdf''Differential scanning calorimetry by Idrish Ansari''.pdf
''Differential scanning calorimetry by Idrish Ansari''.pdfIdrishAnsari4
 

Ähnlich wie Dsc presentation (20)

DSC.pdf
DSC.pdfDSC.pdf
DSC.pdf
 
Differential_Scanning_Calorimetry.ppt
Differential_Scanning_Calorimetry.pptDifferential_Scanning_Calorimetry.ppt
Differential_Scanning_Calorimetry.ppt
 
Dsc
DscDsc
Dsc
 
Differential Scanning Calorimetry.pptx
Differential Scanning Calorimetry.pptxDifferential Scanning Calorimetry.pptx
Differential Scanning Calorimetry.pptx
 
Mahi
MahiMahi
Mahi
 
Thermal method of analysis(D.S.C)
Thermal method of analysis(D.S.C)Thermal method of analysis(D.S.C)
Thermal method of analysis(D.S.C)
 
differential scanning calorimetry.pptx
differential scanning calorimetry.pptxdifferential scanning calorimetry.pptx
differential scanning calorimetry.pptx
 
Differential scanning calorimetry DSC
Differential scanning calorimetry DSCDifferential scanning calorimetry DSC
Differential scanning calorimetry DSC
 
Thermal techniques- DSC- Durgashree Diwakar
Thermal techniques- DSC- Durgashree DiwakarThermal techniques- DSC- Durgashree Diwakar
Thermal techniques- DSC- Durgashree Diwakar
 
DSC | DIFFERENTIAL SCANNING CALORIMETRY
DSC | DIFFERENTIAL SCANNING CALORIMETRYDSC | DIFFERENTIAL SCANNING CALORIMETRY
DSC | DIFFERENTIAL SCANNING CALORIMETRY
 
thermaltechniquesdsc-190602083135.pdf
thermaltechniquesdsc-190602083135.pdfthermaltechniquesdsc-190602083135.pdf
thermaltechniquesdsc-190602083135.pdf
 
Thermal charactrization of polymer
Thermal charactrization of polymerThermal charactrization of polymer
Thermal charactrization of polymer
 
Differential scanning Colorimetry (DSC).
Differential scanning Colorimetry (DSC).Differential scanning Colorimetry (DSC).
Differential scanning Colorimetry (DSC).
 
thermogravimetric analysis
thermogravimetric analysisthermogravimetric analysis
thermogravimetric analysis
 
Differential Scanning Calorimetry (DSC).pdf
Differential Scanning Calorimetry (DSC).pdfDifferential Scanning Calorimetry (DSC).pdf
Differential Scanning Calorimetry (DSC).pdf
 
DTA
DTADTA
DTA
 
Tga
TgaTga
Tga
 
DSC.pdf
DSC.pdfDSC.pdf
DSC.pdf
 
Differential Thermal Analysis & Differential Scanning Calorimetry
Differential Thermal Analysis & Differential Scanning CalorimetryDifferential Thermal Analysis & Differential Scanning Calorimetry
Differential Thermal Analysis & Differential Scanning Calorimetry
 
''Differential scanning calorimetry by Idrish Ansari''.pdf
''Differential scanning calorimetry by Idrish Ansari''.pdf''Differential scanning calorimetry by Idrish Ansari''.pdf
''Differential scanning calorimetry by Idrish Ansari''.pdf
 

Dsc presentation

  • 1. Differential Scanning Calorimetry Presented to: Dr. Mohsin Ali Raza Presenter: Umair manzoor Roll# E11-312 6th Semester Department of Metallurgy & Materials Engineering
  • 2. INTRODUCTION       This technique is used to study what happens to polymers/samples upon heating It is used to study thermal transitions of a polymer/sample (the changes that take place on heating) Need to know about the calorimeter , it is one who measures the heat in or out of the sample. and differential calorimeter is one who measures the heat of the sample relative to the reference. and the differential scanning calorimeter does all of the above functions and heats the sample with the linear temperature. Both the sample and reference are maintained at nearly the same temperature through out the experiment in DSC.  The technique was developed by E.S. Watson and M.J. O'Neill in 1962  Most common than TGA technique
  • 3. PRINCIPLE: It is a technique in which the energy necessary to establish a zero temperature difference between the sample & reference material is measured as a function of temperature.  Here, sample & reference material are heated by separate heaters in such a way that their temp are kept equal while these temp. are increased or decreased linearly.  During heating two types of reactions can be take place one is the endothermic and the other is the exothermic. Endothermic reaction:     If sample absorbs some amount of heat during phase transition then reaction is said to be endothermic In endothermic reaction more energy needed to maintain zero temp difference between sample & reference. E.g. Melting, boiling, sublimation, vaporization
  • 4. •Exothermic reaction: if sample released some amount of heat during phase transition, then reaction is said to be exothermic. In exothermic reaction, less energy needed to maintain zero temp difference between sample & reference. E.g crystallization, degradation, polymerization • what happens to a polymer when heated? The polymer is heated in a device that looks something like this:
  • 5. •There are two pans, In sample pan, polymer is added, while the other, reference pan is left empty •Each pan sits on top of heaters which are controlled by a computer •The computer turns on heaters, and let them heat the two pans at a specific rate. • The computer makes absolutely sure that the heating rate stays exactly the same throughout the experiment Why heaters don’t heat at the same rate? •The simple reason is that the two pans are different. One has polymer in it, and one doesn't. The polymer sample means there is extra material in the sample pan. Having extra material means that it will take more heat to keep the temperature of the sample pan increasing at the same rate as the reference pan • So the heater underneath the sample pan has to work harder than the heater underneath the reference pan. It has to put out more heat.
  • 6. DSC Curve: •The result of a DSC experiment is a curve of heat flux versus temperature or versus time. There are two different conventions: exothermic reactions in the sample shown with a positive or negative peak, depending on the kind of technology used in the experiment. •This curve can be used to calculate enthalpies of transitions, which is done by integrating the peak corresponding to a given transition. The enthalpy of transition can be expressed using equation: ΔH = KA • Where ΔH is the enthalpy of transition, K is the calorimetric constant, A is the area under the peak. •The calorimetric constant varies from instrument to instrument, and can be determined by analyzing a well-characterized material of known enthalpies of transition. •Area under the peak is directly proportional to heat absorbed or evolved by the reaction, •Height of the peak is directly proportional to rate of the reaction
  • 7. Factors affecting DSC curve: Two types of factors effect the DSC curve, 1-Instrumental factors 2-Sample characteristics 1-Instrumental factors a- Furnace heating rate b- Recording or chart speed c- Furnace atmosphere d- Geometry of sample holder/location of sensors e- Sensitivity of the recoding system f-Composition of sample containers. 2-Sample characteristics a- Amount of sample b- Nature of sample c- Sample packing d- Solubility of evolved gases in the sample e- Particle size f- Heat of reaction g- Thermal conductivity
  • 8. TYPES OF DSC INSTRUMENTS: DSC HAS TWO TYPES OF INSTRUMENTS 1 .Heat –  1 Heat – flux DSC flux DSC Sample holder: PLATINUM,ALUMINIUM,STAINLESS STEEL.    SENSORS: Temperature sensors 2. Power Compensated DSC    Sensors:  Pt resistance thermocouple  Separate sensors and heaters for the one block for both sample and reference sample and reference sample . Furnace:   cells .   Temperature:  thermal power, which is supplied to the heaters to maintain the temperature of the sample and reference at the program value Furnace: separate blocks for sample and reference cells. The temperature difference between the sample and reference is converted to differential Sample holder: Aluminum, platinum, stainless steel pans usually thermocouples, which are same for the sample and reference 2 Power compensation DSC   . Temperature: differential thermal power is supplied to the heaters to maintain the temperature of the sample and reference at the program value
  • 9.  Heat flux DSC  Power compensated DCS HEAT FLUX DSC AND POWER COMPENSATED DCS
  • 10. Sample shape for DSC: • Keep sample thin. • The shape of the sample like so that it Cover as much as the bottom of pan as possible. • To make sample thin, cut the sample but don’t crush. • If sample in pellet shape, cut cross section. • If powder, spread evenly over the bottom of the pan. • different shapes for DSC are shown in fig.
  • 11. Sample preparation: • Accurately-weigh samples (~3-20 mg) • Small sample pans (0.1 mL) of inert or treated metals (Al, Pt, Ni, etc.) are used for sample preparation. • Several pan configurations, e.g., open, pinhole, or hermetically-sealed (airtight) pans • The same material and configuration should be used for the sample and the reference • Material should completely cover the bottom of the pan to ensure good thermal contact, • Avoid overfilling the pan to minimize thermal lag from the bulk of the material to the sensor
  • 12. What can DSC measure? • Glass transitions • Melting and boiling points • Crystallisation time and temperature • Percent crystallinity • Heats of fusion and reactions • Specific heat capacity • Oxidative/thermal stability • Reaction kinetics • Purity
  • 13. Determination of Heat Capacity: • By the DSC we can find out the heat capacity of the polymers and then after futher heating we can find out the glass transition temperature and after this we can find out the crystallinity and then the melting temperature and then putting all above factors we get the DSC curve. • In DSC curve we get crystallinity peaks, while melting dip. exothermic endothermic • The crystallization peak and the melting dip will only show up for polymers that can form crystals and melts. • Completely amorphous polymers won't show any crystallization, or any melting either. But polymers with both crystalline and amorphous domains, will show all the features
  • 14. : Determination of Heat Capacity •DSC plot can be used to determine Heat Capacity. Suppose a polymer is being heated. Heating two pans Difference in Heat output of two heaters temperature against • When we start heating two pans, the computer will plot the difference in heat output of the two heaters against temperature that is plot of heat absorbed by the polymer against temperature. The plot will look like this at first.
  • 15. The heat flow is heat (q) supplied per unit time (t), whereas The heating rate is temperature increase (ΔT) per unit time (t) By dividing heat flow (q/t) by the heating rate (ΔT/t). It ends up with heat supplied divided by the temperature increase, which is called heat capacity
  • 16. The Glass Transition Temperature by DSC the glass transition is a second order transition. On further heating the polymer to a certain temperature, plot will shift downward suddenly, like this: No peak and dig This means there is more heat flow. There is an increase in the heat capacity of the polymer. This happens because the polymer has just gone through the glass transition. Because of this change in heat capacity that occurs at the glass transition, we can use DSC to measure a polymer's glass transition temperature.
  • 17. Crystallization by DSC After glass transition, the polymers have a lot of mobility. when they reach the right temperature, they will give off enough energy to move into very ordered arrangements, which is called crystals. After crystallization Given off heat exothermic heater to sample pan has not to put a lot of heat into the polymer in order to keep the temperature rising at the same rate as that of the reference pan this drop in the heat flow as a big peak in the plot of heat flow versus temperature If you analyze a 100% amorphous polymer, like polystrene, you wouldn't get this peak, because such materials don't crystallize
  • 18. melting temperature by DSC If we heat our polymer past its Tc, eventually we'll reach another thermal transition, called melting. When we reach the polymer's melting temperature, Tm, the polymer crystals begin to fall apart, that is they melt When the polymer crystals melt, they must absorb heat in order to do so. After melting Heat is absorbed heater to sample pan has to put a lot of heat into the polymer in order to both melt the crystals and keep the temperature rising at the same rate as that of the reference pan endothermic this increase in the heat flow as a big dig in the plot of heat flow versus temperature
  • 19. Putting It All Together we saw a step in the plot when the polymer was heated past its glass transition temperature. Then we saw a big peak when the polymer reached its crystallization temperature. Then finally we saw a big dip when the polymer reached its melting temperature. To put them all together, a whole plot will often look something like this: Peak at Tc Step at Tg Dig at Tm
  • 20. Best way to perform DSC thermal analysis technique: •proper instrument calibration •use purge gas (N2 or He) to remove corrosive off-gases •small sample size •good thermal contact between the sample and the temperaturesensing device •proper sample encapsulation •start temperature well below expected transition temperature •slow scanning speeds •avoid decomposition in the DSC
  • 21. Advantages of using DSC: •instruments can be used at very high temperatures •instruments are highly sensitive •flexibility in sample volume/form •characteristic transition or reaction temperatures can be determined •High resolution obtained •High sensitivity •Stability of the material
  • 22. limitations: •DSC generally unsuitable for two-phase mixtures • Difficulties in test cell preparation in avoiding evaporation of volatile Solvents • DSC is generally only used for thermal screening of isolated intermediates and products • Does not detect gas generation • uncertainty of heats of fusion and transition temperatures