1. The right way to measure.
Important tips on the operation and maintenance of the testo 270
Preparation
<
cm
.1
>
in
m
! ! °C
Plastic compo- Do not touch Remove fried Wait until Observe minimum distance Measure in hot oil
nents may not the hot probe. goods before bubbles no of 1 cm to metal objects. (min. 40 °C, max. 190 °C).
come into Danger of measuring. longer rise to
contact with the burning! the surface
cooking oil. (approx. 5 min.).
Measurement
MAX
MIN
Switch on the instrument Immerse sensor in hot oil. 1. Move sensor in oil (adjustment time approx. 20 sec.) Note down
[ /Hold] Watch out for the min/max Auto Hold activated: the recorded
marks! • Auto Hold in the display indicates the end of the measu- measurement
rement. If set, an audible signal additionally sounds. values.
• Final value is fixed.
Auto Hold deactivated:
• When the temperature on the display no longer changes,
a stable TPM value has been reached.
• To fix measurement values: briefly press [ /Hold] < 1 sec.
2. To return to the measurement mode: briefly press
[ /Hold] < 1 sec.
Care and maintenance Cleaning Accuracy test
Your specific starting value:
___%TPM
Clean instrument with mild detergent. Rinse The TopSafe and For a fast and easy instrument test, For an accurate instrument test
sensor under running water and dry carefully hand strap can we recommend you to carry out and/or correction, or if measure-
with a soft paper towel. be conveniently an accuracy test according to the ment values are implausible,
cleaned in the instruction manual. we recommend calibration and
dishwasher. If the oil type or the supplier is adjustment in Testo reference oil.
changed, a new reference value
must be determined.
You can download this training card from: www.testo.com/trainingcard-oil-tester
For more information on the testo 270, visit: www.testo.com/oil-tester