 Basic concepts of hemodynamics and
it’s clinical application in arterial disease
Primary physiology of arterial
disease
 obstruction of the lumen
atherosclerosis, emboli, thrombi, fibromuscular
dysplasia,trauma or external compression
 disruption of the vessel wall
aneurysms or trauma
BASIC PRINCIPLES OF ARTERIAL
HEMODYNAMICS
 Fluid Pressure:
The pressure in a fluid system is defined as force per
unit area (in dynes per square centimeter).
(1) the dynamic pressure produced by contraction of
the heart
(2) the hydrostatic pressure
(3) the static filling pressure
Fluid Energy
 Potential energy can be expressed as follows:
predominant component of potential energy is
the pressure produced by cardiac contraction.
 Kinetic energy :
predominant component of kinetic energy is velocity.
Bernoulli’s Principle
 When fluid flows steadily(no acceleration or
deceleration)from one point in a system to another its
total energy content along any given streamline
remains constant,provided that there are no frictional
losses’
 This is in accordance with the law of conservation
of energy .
total fluid energy
remains constant
 increse in potential energy, (higher pressure)
(lower velocity )
proportional loss of kinetic energy .
 This situation is not observed in human arteries
because the ideal flow conditions specified in the
Bernoulli relationship are not present.
Fluid Energy Losses
 Energy losses in flowing blood
1. viscous losses:
viscosity= hematocrit of blood
2. inertial losses:
changes in the velocity or direction of flow.
viscosity
 The viscosity (η) is defined as the
ratio of shear stress (τ) to shear rate (D):
η= τ/D
 Shear stress :
proportional to the energy loss owing
to friction between adjacent
fluid layers,
 shear rate:
the relative velocity of adjacent fluid layers.
Rheologic Agents and viscus loss
.
1. Low-molecular-weight dextran.
2. pentoxifylline
are most often used in the immediate postoperative
period to increase flow through a reconstructed arterial
segment for lowering blood viscosity
inertial loss andPoiseuille’s law
 This law states that the pressure gradient along a tube
(P1 – P2, ) is directly proportional to the mean
flow velocity (V,)or volumeflow (Q), the tube
length(L), and the fluid viscosity (η) and it is inversely
proportional to either the second or fourth power of
the radius (r) .
 Therefore, radius has a profound
influence on energy losses
Stenosis Length and Multiple
stenosis
 Doubling the length of a stenosis results in a doubling of the
associated energy losses.
 But reducing the radius by half increases energy losses by a
factor of 16 .
 Thus separate short stenosis tend to be more significant than a
single longer stenosis.
 multiple subcritical stenosis may have the same effect as a
single critical stenosis
Inertial Energy Losses
 variations in lumen diameter
 at points of curvature and branching.
 the acceleration and deceleration of pulsatile flow,
 inertial energy losses result
 Energy losses related to inertia (ΔE) are proportional to a
and the square of blood velocity .

Application
 The relationship
between radius and
pressure drop for various
flow rates along a 10-cm
vessel segment .
Poiseuille’s law and inertial loss
 the pressure drop is negligible until the radius is
reduced to 0.3 cm or 3mm.
 if the radius less than 0.2 cm or 2mm, the pressure
drop increases rapidly.
 These observations may explain the preference of
autogenous vein bypass grafts of not less than 3-4 mm
in diameter .
Vascular Resistance
 Hemodynamic resistance (R) can be defined as the ratio of the
energy drop between two points along an artery (E1 – E2) to the
mean blood flow (Q):
 The standard physical units of hemodynamic resistance
peripheral resistance unit (PRU)
1PRU= 8 × 10^4 dyne-s/cm.^5 .
total vascular resistance
 arteries and capillaries=90%,
 arterioles and capillaries=> 60%
 large- and medium-sized arteries=15%
 venous flow=10% .
 atherosclerotic occlusive disease occur normally in vessels with
low resistance .
 The large and medium arteries are most commonly affected
Effects Resistance in Series and
Parallel
Flow in Parallel Graft and
Stenotic Artery
 Surgeons occasionally express
concern over the possibility that
continued patency of a stenotic
artery might lead to thrombosis
of a parallel graft.
 To allay this fear; they either
avoid end-to-side anastomoses or
ligate the stenotic artery.
 Theoretical considerations strongly suggest that
such concerns are not valid, provided that the arterial
segment is sufficiently diseased to merit bypass
grafting
vein Grafts with Double Lumens
 saphenous veins bifurcate into two
separate and parallel channels that
rejoin after a variable
distance to reconstitute a single
lumen.
 When this situation is
encountered, the surgeon must
decide whether or
not to include both channels in the
graft
 the combined resistance of the
two parallel channels exceeds that of the undivided
vein .
Sequential Grafts
theoretical advantages and
disadvantages .
 Flow rates in femorotibial grafts should theoretically
be lower than those in the proximal segment of
sequential grafts but higher than those in the distal
segment .
 Femorotibialgraft may be more likely to fail than the
sequential graft.
Blood Flow Patterns
Laminar Flow
 In steady-state conditions the flow pattern is laminar.
All motion is parallel to the walls of the tube and the
fluid is arranged in a series of concentric layers or
laminae.
 velocity profile that is parabolic in shape .
PARABOLA- Laminar Flow
 with velocity being highest in the center of the stream
and becoming progressively lower towards the vessel
wall.
 The layer of fluid or blood in contact with the wall is
stationary .
TURBULENCE
 In contrast to the linear streamlines of laminar flow,
turbulence is an irregular flow state in which velocity
varies rapidly with respect to space and time.
 These random velocity changes result in inertial
energy .and dissipation of fluid energy as heat.
To measure --- Reynolds number, Re
Reynolds number,
 The point at which flow changes from laminar to
turbulent is best defined in terms of a dimensionless
quantity known as the Reynolds number, Re.
 Re exceeds 2000= turbulence
 laminar if Re s <2000.
 Because Re values are well below 2000 in most
peripheral arteries, turbulence is unlikely to occur
under normal circumstances.
 However, turbulence does appear to develop in the
ascending aorta during the peak systolic ejection
phase .
Normal Pressure and Flow
 As the arterial pressure pulse moves distally, the
systolic pressure rises, the diastolic pressure falls, and
the pulse pressure becomes wider.
 The decrease in mean arterial pressure between the
heart and the ankle is normally less than 10 mm Hg .
 normal individuals, (ankle-brachial index) has a
mean value of 1.11 ± 0.10
in the resting state.
Boundary Layer Separation
 In fluid flowing through a tube, the portion of fluid
adjacent to the tube wall is referred to as the boundary
layer.
 This layer is subject to both frictional interactions with
the tube wall and viscous forces generated by the more
rapidly moving fluid toward the center of the tube.
.
 When the tube geometry changes suddenly, such as at
points of curvature, branching, or variations in lumen
diameter,small pressure gradients are created that
cause the boundary layer to stop or reverse direction.
 This change results in a complex, localized flow
pattern known as an area of boundary layer separation
or flow separation.
Flow Separation and Shear
 The clinical importance
of boundary layer
separation
is that these localized
flow disturbances may
contribute
to the formation of
atherosclerotic plaques
Anastomoses
Anastomotic Configuration
 To reduce energy losses due to flow disturbances, the
transition from graft to host vessel should be as
smooth as possible
 End-to-end anastomoses, therefore, most
closely approximate the ideal. End-to-side or side-to-
end anastomoses always result in alterations in flow
direction
 Tailoring the anastomosis to enter the recipient artery
or leave the donor artery at an acute angle will
minimize but can never eliminate flow disturbances.
Bifurcations and Branches
 The branches of the arterial system produce sudden
changes in the flow pattern that are potential sources
of energy loss.
 Flow patterns in a bifurcation are determined
mainly by the area ratio and the branch angle.
Area Ratio
 The area ratio is defined as the combined area of the
secondary branches divided by the area of the primary
artery.
 For efficient transmission of pulsatile energy across a
bifurcation, the vascular impedance of the primary
artery should equal that of the branches,
 ideal area ratio of 1.15 for larger arteries and 1.35 for
smaller arteries.
 infrarenal aorta
an area ratio of 0.8, approximately 22% of the incident
pulsatile energy is reflected .
 Hence atherosclerosis and aneurysms are common in
this arterial segment
 Despite their theoretical
disadvantages,
commercially
available bifurcation
grafts have functioned
extremely well in
a variety of clinical
applications.
The Branch Angle.
 Flow disturbances are minimized when the angle is
narrow and are exaggerated when the limbs are widely
spread
 The average angle between the human iliac
arteries is 54 degrees;
 however, with diseased or tortuous
iliac arteries, this angle can approach 180 degrees.
Angle Between The Limbs Of A
Bifurcation Graft
Bruits
 Stenosis or irregularities of the vessel lumen produce
turbulent flow patterns that set up vibrations in the
arterial wall.
 These vibrations generate displacement waves that
radiate through the surrounding tissues and can be
detected as audible sounds.
 Generally, a soft, midsystolic bruit is associated with a
relatively minor lesion that does not significantly
reduce flow or pressure.
 A bruit with a loud diastolic component
suggests a stenosis severe enough to reduce flow and
produce a pressure drop.
 A bruit may be absent when an artery is
nearly occluded or when the flow rate is extremely low
Poststenotic Dilatation
 The most likely explanation for this phenomenon
is that arterial wall vibrations result in structural
fatigue of elastin fibers.
 In a series of animal model studies, poststenotic
dilatations did not develop unless a bruit was
present distal to the stenosis.
CRITICAL STENOSIS
 The degree of narrowing at which pressure and flow
begin to be affected has been called the “critical
stenosis.
 appreciable changes in pressure and flow do not occur
until the cross-sectional area of a vessel has been
reduced by more than 75% or 50% reduction in
diameter.
VELOCITY OF BLOOD FLOW
 Critical stenosis-varies with resistance of run off bed.
 system with a high flow velocity (low
resistance) shows a reduction in pressure with less
narrowing than a system with low flow velocity.
Palpable Pulses
 Motion of the arterial wall is responsible for the
palpable pulses that are so important in the physical
examination of a patient in whom arterial disease is
suspected.
 However, a 7-mm femoral artery in a young subject
would expand only 0.2 mm under the influence of 50
mm Hg pulse pressure, and the stiffer arteries in an
older individual would expand even less.
 circular cross-section into an ellipse.
 It takes much less energy to bend the wall of an
elliptically shaped vessel than it does to stretch the
wall of a circular vessel.
 Therefore, when the artery is partially compressed, its
expansion in the direction of the compression is
greatly augmented
 The clinical value of pulse palpation is based on the
assumption that the strength of the pulse is directly
related to the pulse pressure, which should be
decreased distal to an obstruction.
 However, stiff, calcified vessels may display little or no
palpable pulse even though there is no decrease in
pulse pressure.
Arterial Obstruction
Collateral Circulation
 A circuit includes
diseased main artery a
parallel system of
collateral arteries, and
the peripheral runoff
bed.
Resistance
 fixed , collateral system resistance/segmental resistance.
 variable of a peripheral runoff bed resistance.
 Normally, the resting segmental resistance is low and
the peripheral resistance is relatively high;
Peripheral Runoff Bed
 terminal arterioles and precapillary sphincters
 Because of their small diameter and heavily muscled
walls, these vessels are ideally suited for regulatory
function.
 Their resistance is subject to control by
(1) the autonomic nervous system,
(2) circulating catecholamines,
(3) local metabolic products and
(4) myogenic influences
AUTOREGULATION
 Autoregulation can compensate for a drop in perfusion
pressure only until it falls below a critical level (e.g.,
about 20 to 30 mm Hg for skeletal muscle ).
 With pressures below this level, normal blood flow is
no longer maintained and flow responds passively to
changes in perfusion pressure
Exercise Therapy
 exercise therapy is best suited for patients with mild,
stable claudication who are not candidates for direct
intervention .
 In severe grade patients, the
peripheral resistance has already been lowered to
compensate for the increased segmental resistance,
attempts to further reduce the peripheral resistance
are seldom beneficial.
Vasodilators
 The rationale for the use of vasodilators is that they
lower peripheral vascular resistance and improve limb
blood flow.
 Although this may occur in normal limbs, it is
unlikely to be beneficial in limbs in which peripheral
resistance is already decreased as a result of arterial
disease.
Sympathectomy
 purpose of sympathectomy is to reduce
peripheral resistance by release of vasomotor tone.
 has little or no effect on collateral resistance.
 Clinical improvement can occur only if the ischemic tissues
are capable of further vasodilatation, as demonstrated by
reactive hyperemia testing.
VASCULAR STEAL
 Vascular “steal” may arise when two runoff beds with
different resistances must be supplied by a limited
source of arterial inflow.
 When an extraanatomic bypass is performed, a single
donor artery must supply several vascular beds.
 In the case of a femoral-femoral crossover graft, one
iliac artery is the donor artery, the leg ipsilateral to the
donor artery is the donor limb, and the contralateral
leg is the recipient limb.
 Studies of crossover grafts in animal models have
shown that the immediate effect of the graft is to
double flow in the donor artery.
 When an arteriovenous fistula is created in the
recipient limb, graft flow may increase by a factor of 10
without any evidence of steal from the donor limb
 The most important factor contributing to vascular
steal with a femoral-femoral graft is stenosis of the
donor iliac artery.
 Occlusive disease in the arteries of the donor limb
distal to the origin of the graft does not result in steal,
provided that the donor iliac artery is normal.
 These principles also apply to other types of
extraanatomic bypass grafts, including axillary-
axillary, carotid-subclavian, and axillofemoral grafts.
GRAVITY
 Many patients with severe lower extremity ischemia
discover that hanging their feet over the edge of the
bed or walking a few steps often provides some relief
from ischemic rest pain.
 The improvement in peripheral perfusion that
accompanies dependency can be documented
objectively with measurement of transcutaneous
oxygen tension, which may be increased severalfold
over levels measured when the patient is supine
 pressure in the dependent arteries, veins, and capillaries is
increased by gravity commensurate with the vertical
distance from the foot to the heart.
 Although there is no increase in the arteriovenous pressure
gradient,
 the increased hydrostatic pressure dilates capillaries and
microvascular vessels, thereby reducing their resistance, in
turn augmenting blood flow.
 Augmentation of blood flow does not occur in
nonischemic limbs because the venoarterial reflex that
serves to constrict arterioles in the dependent position
remains functional
Arterial Grafts
 Because radius is so important in determining both
viscous and inertial energy losses, the graft selected
should be large enough to carry all the flow required at
rest without causing a drop in pressure
 it should also be large enough to accommodate any
increased flow likely to be required during exercise
without an appreciable drop in pressure.
 Clearly, a graft with an inside diameter of less than 3
mm would be of marginal value at flow rates normally
observed at rest (60 to 150 mL/min) and would be
completely unsatisfactory during exercise (300 to 500
mL/min).
 Therefore, under most physiologic conditions, an
aortofemoral graft with 7-mm limbs should suffice,
with flow restricted only during strenuous exercise.
 However, 6-mm grafts might begin to show some
restriction of flow even with mild to moderate
exercise.
PSEUDOINTIMA
 prosthetic grafts develop a thin layer (0.5 to 1.0 mm) of
pseudointima.
 Therefore, after implantation, a 6-mm prosthetic graft
might have an internal diameter of only 4 to 5 mm, and a 7-
mm graft might have a luminal diameter of 5 to 6 mm.
 For this reason, when one is performing a femoropopliteal
bypass with a prosthetic graft, it is appropriate to select a
graft with an original diameter of at least 6 mm.
 Similarly, the original diameter of an aortofemoral graft
limb should be at least 7 mm.
 the diameter of a graft must not be too much larger
than that of the recipient arteries
 thrombus accumulates on the inner walls of grafts with
excessive diameter (much as they do in aneurysms) as
the flow stream tries to approximate the diameter of
the recipient vessels and achieve optimal flow
conditions
 High flow velocity (high shear) is conducive to the
formation of a thin, tightly adherent pseudointima,
 so the diameter of a prosthetic graft should be small
enough to ensure a rapid velocity of flow but large
enough to avoid restriction of arterial inflow
SHEAR RATES
 Long-term patency of autogenous vein grafts is
compromised by intimal hyperplasia, the development
of which has also been associated with low shear rates.
 Low shear rates cause smooth muscle cells to become
secretory and enhance platelet adherence;
 high shear rates foster continued patency and lessen
the tendency of the intima to become hyperplastic.
Aneurysm rupture and diameter
 Tangenital stress and
tensile stress,
 Laplace’s law, which
defines tangential
tension (T) as the
product of pressure and
radius
 Rupture occurs when the tangential stress within the
arterial wall becomes greater than the tensile strength.
 The tangential stress (τ) within the wall of a fluid-filled
cylindrical tube can be expressed as:
τ= P r/δ ,
 P is the pressure exerted by the fluid,
r is the internal radius,
δ is the thickness of the tube wall.
 The tendency of larger aneurysms to
rupture is explained by the effect of increased radius
on tangential stress .
 The relationship between tangential stress and
blood pressure accounts for the contribution of
hypertension.
Posterolateral rupture
 55% of ruptured abdominal aortic aneurysms, the site
of rupture is in the posterolateral aspect of the
aneurysm wall.

The posterior wall of the aorta is relatively fixed
against the spine, and repeated flexion in that area
could result in structural fatigue and a localized area
of weakness that might predispose to rupture,.
.
Arterial physiology

Weitere ähnliche Inhalte

Was ist angesagt?

No reflow and slow flow phenomenon during pci
No reflow and slow flow phenomenon during pciNo reflow and slow flow phenomenon during pci
No reflow and slow flow phenomenon during pci
rahul arora
 
Shunt Detection And Quantification
Shunt Detection And QuantificationShunt Detection And Quantification
Shunt Detection And Quantification
Dang Thanh Tuan
 

Was ist angesagt? (20)

FFR(fractional flow reserve)
FFR(fractional flow reserve)FFR(fractional flow reserve)
FFR(fractional flow reserve)
 
Pressure wave forms
Pressure wave formsPressure wave forms
Pressure wave forms
 
Echocardiography of Aortic stenosis
Echocardiography of Aortic stenosis Echocardiography of Aortic stenosis
Echocardiography of Aortic stenosis
 
Stent Thrombosis
Stent ThrombosisStent Thrombosis
Stent Thrombosis
 
No reflow and slow flow phenomenon during pci
No reflow and slow flow phenomenon during pciNo reflow and slow flow phenomenon during pci
No reflow and slow flow phenomenon during pci
 
Basic haemodynamic assessment with echo (iHeartScan)
Basic haemodynamic assessment with echo (iHeartScan)Basic haemodynamic assessment with echo (iHeartScan)
Basic haemodynamic assessment with echo (iHeartScan)
 
Coronary blood flow
Coronary blood flowCoronary blood flow
Coronary blood flow
 
Coronary anomalies
Coronary anomalies Coronary anomalies
Coronary anomalies
 
Aortic root anatomy
Aortic root anatomyAortic root anatomy
Aortic root anatomy
 
J. frederick ctsa summit tavr
J. frederick   ctsa summit tavrJ. frederick   ctsa summit tavr
J. frederick ctsa summit tavr
 
Hemodyanmic features of Constrictive pericarditis and Restrictive cardiomyopathy
Hemodyanmic features of Constrictive pericarditis and Restrictive cardiomyopathyHemodyanmic features of Constrictive pericarditis and Restrictive cardiomyopathy
Hemodyanmic features of Constrictive pericarditis and Restrictive cardiomyopathy
 
Shunt Detection And Quantification
Shunt Detection And QuantificationShunt Detection And Quantification
Shunt Detection And Quantification
 
PA Banding Rigel.pptx
PA Banding Rigel.pptxPA Banding Rigel.pptx
PA Banding Rigel.pptx
 
FRACTIONAL FLOW RESERVE
FRACTIONAL FLOW RESERVEFRACTIONAL FLOW RESERVE
FRACTIONAL FLOW RESERVE
 
Management of Small Aortic Root
Management of Small Aortic RootManagement of Small Aortic Root
Management of Small Aortic Root
 
Principles of CPB
Principles of CPBPrinciples of CPB
Principles of CPB
 
Elephant Trunk after Borst
Elephant Trunk after BorstElephant Trunk after Borst
Elephant Trunk after Borst
 
NO REFLOW
NO REFLOWNO REFLOW
NO REFLOW
 
Aortic valve repair
Aortic valve repairAortic valve repair
Aortic valve repair
 
Mitral valve anatomy
Mitral valve anatomyMitral valve anatomy
Mitral valve anatomy
 

Ähnlich wie Arterial physiology

C:\Documents And Settings\User\Desktop\Lec54
C:\Documents And Settings\User\Desktop\Lec54C:\Documents And Settings\User\Desktop\Lec54
C:\Documents And Settings\User\Desktop\Lec54
MBBS IMS MSU
 
31. cardiovascular 3-08-09
31. cardiovascular 3-08-0931. cardiovascular 3-08-09
31. cardiovascular 3-08-09
Nasir Koko
 
On the Analysis of the Laminar to Turbulent Flow Patterns in the Treatment of...
On the Analysis of the Laminar to Turbulent Flow Patterns in the Treatment of...On the Analysis of the Laminar to Turbulent Flow Patterns in the Treatment of...
On the Analysis of the Laminar to Turbulent Flow Patterns in the Treatment of...
Waqas Tariq
 
Vascular Sys Student Notes
Vascular Sys   Student NotesVascular Sys   Student Notes
Vascular Sys Student Notes
Adam
 

Ähnlich wie Arterial physiology (20)

General principals of circulation
General principals of circulationGeneral principals of circulation
General principals of circulation
 
C:\Documents And Settings\User\Desktop\Lec54
C:\Documents And Settings\User\Desktop\Lec54C:\Documents And Settings\User\Desktop\Lec54
C:\Documents And Settings\User\Desktop\Lec54
 
HEMODYNAMICS 202.pptx
HEMODYNAMICS 202.pptxHEMODYNAMICS 202.pptx
HEMODYNAMICS 202.pptx
 
Blood flowregulatingfactors
Blood flowregulatingfactorsBlood flowregulatingfactors
Blood flowregulatingfactors
 
Hemodynamics 202
Hemodynamics 202Hemodynamics 202
Hemodynamics 202
 
Cut Flow.Doc
Cut Flow.DocCut Flow.Doc
Cut Flow.Doc
 
Cvs ppt
Cvs pptCvs ppt
Cvs ppt
 
CORONARY BLOOD FLOW
CORONARY BLOOD FLOWCORONARY BLOOD FLOW
CORONARY BLOOD FLOW
 
Lecture: 57 pressure, blood flow, compliance and resistance
Lecture: 57 pressure, blood flow, compliance and resistanceLecture: 57 pressure, blood flow, compliance and resistance
Lecture: 57 pressure, blood flow, compliance and resistance
 
Hemodynamics by Dr Ambareesha.pptx
Hemodynamics by Dr Ambareesha.pptxHemodynamics by Dr Ambareesha.pptx
Hemodynamics by Dr Ambareesha.pptx
 
Hemodynamics.pdf
Hemodynamics.pdfHemodynamics.pdf
Hemodynamics.pdf
 
31. cardiovascular 3-08-09
31. cardiovascular 3-08-0931. cardiovascular 3-08-09
31. cardiovascular 3-08-09
 
Blood pressure kamla choudhary
Blood pressure kamla choudharyBlood pressure kamla choudhary
Blood pressure kamla choudhary
 
On the Analysis of the Laminar to Turbulent Flow Patterns in the Treatment of...
On the Analysis of the Laminar to Turbulent Flow Patterns in the Treatment of...On the Analysis of the Laminar to Turbulent Flow Patterns in the Treatment of...
On the Analysis of the Laminar to Turbulent Flow Patterns in the Treatment of...
 
Vascular Sys Student Notes
Vascular Sys   Student NotesVascular Sys   Student Notes
Vascular Sys Student Notes
 
FFR RAMDHAN.pptx
FFR RAMDHAN.pptxFFR RAMDHAN.pptx
FFR RAMDHAN.pptx
 
Physical Properties of Blood Flow
Physical Properties of Blood FlowPhysical Properties of Blood Flow
Physical Properties of Blood Flow
 
Echo assessment of aortic stenosis
Echo assessment of aortic stenosisEcho assessment of aortic stenosis
Echo assessment of aortic stenosis
 
Uploaded circulation1
Uploaded circulation1Uploaded circulation1
Uploaded circulation1
 
circulation1
circulation1circulation1
circulation1
 

Mehr von Tapish Sahu

Aneurysms of upper and lower extremities + aneurysms
Aneurysms of upper and lower extremities + aneurysmsAneurysms of upper and lower extremities + aneurysms
Aneurysms of upper and lower extremities + aneurysms
Tapish Sahu
 
Aneurysms of splanchnic and visceral arteries
Aneurysms of splanchnic and visceral arteriesAneurysms of splanchnic and visceral arteries
Aneurysms of splanchnic and visceral arteries
Tapish Sahu
 
Penetrating aortic ulcer
Penetrating aortic ulcerPenetrating aortic ulcer
Penetrating aortic ulcer
Tapish Sahu
 
Vascular Laboratory: Arterial Physiologic Assessment & Arterial Duplex Scanning
Vascular Laboratory: Arterial Physiologic Assessment & Arterial Duplex ScanningVascular Laboratory: Arterial Physiologic Assessment & Arterial Duplex Scanning
Vascular Laboratory: Arterial Physiologic Assessment & Arterial Duplex Scanning
Tapish Sahu
 

Mehr von Tapish Sahu (20)

Aortoiliac aneurysms evaluation
Aortoiliac aneurysms evaluationAortoiliac aneurysms evaluation
Aortoiliac aneurysms evaluation
 
Aneurysms of upper and lower extremities + aneurysms
Aneurysms of upper and lower extremities + aneurysmsAneurysms of upper and lower extremities + aneurysms
Aneurysms of upper and lower extremities + aneurysms
 
Aneurysms of splanchnic and visceral arteries
Aneurysms of splanchnic and visceral arteriesAneurysms of splanchnic and visceral arteries
Aneurysms of splanchnic and visceral arteries
 
Acute limb ischemia
Acute limb ischemiaAcute limb ischemia
Acute limb ischemia
 
Penetrating aortic ulcer
Penetrating aortic ulcerPenetrating aortic ulcer
Penetrating aortic ulcer
 
Arterial aneurysms
Arterial aneurysmsArterial aneurysms
Arterial aneurysms
 
Acute mesenteric arterial disease
Acute mesenteric arterial diseaseAcute mesenteric arterial disease
Acute mesenteric arterial disease
 
Upper extremity arterial disease
Upper extremity arterial diseaseUpper extremity arterial disease
Upper extremity arterial disease
 
Vascular grafts
Vascular graftsVascular grafts
Vascular grafts
 
Intraoperative management
Intraoperative managementIntraoperative management
Intraoperative management
 
Graft thrombosis
Graft thrombosisGraft thrombosis
Graft thrombosis
 
Anti coagulants
Anti coagulantsAnti coagulants
Anti coagulants
 
Preoperative evaluation and management
Preoperative evaluation and managementPreoperative evaluation and management
Preoperative evaluation and management
 
Normal coagulation
Normal coagulationNormal coagulation
Normal coagulation
 
Hypercoagulable states
Hypercoagulable statesHypercoagulable states
Hypercoagulable states
 
Post operative management
Post operative management Post operative management
Post operative management
 
CT, MRI in vascular surgery
CT, MRI in vascular surgery CT, MRI in vascular surgery
CT, MRI in vascular surgery
 
PET,SPECT, IVUS
PET,SPECT, IVUSPET,SPECT, IVUS
PET,SPECT, IVUS
 
Radiation safety
Radiation safetyRadiation safety
Radiation safety
 
Vascular Laboratory: Arterial Physiologic Assessment & Arterial Duplex Scanning
Vascular Laboratory: Arterial Physiologic Assessment & Arterial Duplex ScanningVascular Laboratory: Arterial Physiologic Assessment & Arterial Duplex Scanning
Vascular Laboratory: Arterial Physiologic Assessment & Arterial Duplex Scanning
 

Kürzlich hochgeladen

Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
chetankumar9855
 

Kürzlich hochgeladen (20)

Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
 
Call Girls Vadodara Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Vadodara Just Call 8617370543 Top Class Call Girl Service AvailableCall Girls Vadodara Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Vadodara Just Call 8617370543 Top Class Call Girl Service Available
 
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
 
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
 
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
 
Top Rated Bangalore Call Girls Majestic ⟟ 9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Majestic ⟟  9332606886 ⟟ Call Me For Genuine S...Top Rated Bangalore Call Girls Majestic ⟟  9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Majestic ⟟ 9332606886 ⟟ Call Me For Genuine S...
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
 
Top Rated Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...
Top Rated  Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...Top Rated  Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...
Top Rated Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...
 
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
 
Call Girls Kurnool Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kurnool Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Kurnool Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kurnool Just Call 8250077686 Top Class Call Girl Service Available
 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
 
Call Girls Visakhapatnam Just Call 8250077686 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 8250077686 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 8250077686 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 8250077686 Top Class Call Girl Service Ava...
 
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
 
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
 
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
 
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
 
Call Girls Kakinada Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kakinada Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Kakinada Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kakinada Just Call 9907093804 Top Class Call Girl Service Available
 

Arterial physiology

  • 2. Plan of presentation  Basic concepts of hemodynamics and it’s clinical application in arterial disease
  • 3. Primary physiology of arterial disease  obstruction of the lumen atherosclerosis, emboli, thrombi, fibromuscular dysplasia,trauma or external compression  disruption of the vessel wall aneurysms or trauma
  • 4. BASIC PRINCIPLES OF ARTERIAL HEMODYNAMICS  Fluid Pressure: The pressure in a fluid system is defined as force per unit area (in dynes per square centimeter). (1) the dynamic pressure produced by contraction of the heart (2) the hydrostatic pressure (3) the static filling pressure
  • 5. Fluid Energy  Potential energy can be expressed as follows: predominant component of potential energy is the pressure produced by cardiac contraction.  Kinetic energy : predominant component of kinetic energy is velocity.
  • 6. Bernoulli’s Principle  When fluid flows steadily(no acceleration or deceleration)from one point in a system to another its total energy content along any given streamline remains constant,provided that there are no frictional losses’  This is in accordance with the law of conservation of energy .
  • 7. total fluid energy remains constant  increse in potential energy, (higher pressure) (lower velocity ) proportional loss of kinetic energy .  This situation is not observed in human arteries because the ideal flow conditions specified in the Bernoulli relationship are not present.
  • 8. Fluid Energy Losses  Energy losses in flowing blood 1. viscous losses: viscosity= hematocrit of blood 2. inertial losses: changes in the velocity or direction of flow.
  • 9. viscosity  The viscosity (η) is defined as the ratio of shear stress (τ) to shear rate (D): η= τ/D  Shear stress : proportional to the energy loss owing to friction between adjacent fluid layers,  shear rate: the relative velocity of adjacent fluid layers.
  • 10. Rheologic Agents and viscus loss . 1. Low-molecular-weight dextran. 2. pentoxifylline are most often used in the immediate postoperative period to increase flow through a reconstructed arterial segment for lowering blood viscosity
  • 11. inertial loss andPoiseuille’s law  This law states that the pressure gradient along a tube (P1 – P2, ) is directly proportional to the mean flow velocity (V,)or volumeflow (Q), the tube length(L), and the fluid viscosity (η) and it is inversely proportional to either the second or fourth power of the radius (r) .  Therefore, radius has a profound influence on energy losses
  • 12. Stenosis Length and Multiple stenosis  Doubling the length of a stenosis results in a doubling of the associated energy losses.  But reducing the radius by half increases energy losses by a factor of 16 .  Thus separate short stenosis tend to be more significant than a single longer stenosis.  multiple subcritical stenosis may have the same effect as a single critical stenosis
  • 13. Inertial Energy Losses  variations in lumen diameter  at points of curvature and branching.  the acceleration and deceleration of pulsatile flow,  inertial energy losses result  Energy losses related to inertia (ΔE) are proportional to a and the square of blood velocity . 
  • 14. Application  The relationship between radius and pressure drop for various flow rates along a 10-cm vessel segment .
  • 15. Poiseuille’s law and inertial loss  the pressure drop is negligible until the radius is reduced to 0.3 cm or 3mm.  if the radius less than 0.2 cm or 2mm, the pressure drop increases rapidly.  These observations may explain the preference of autogenous vein bypass grafts of not less than 3-4 mm in diameter .
  • 16. Vascular Resistance  Hemodynamic resistance (R) can be defined as the ratio of the energy drop between two points along an artery (E1 – E2) to the mean blood flow (Q):  The standard physical units of hemodynamic resistance peripheral resistance unit (PRU) 1PRU= 8 × 10^4 dyne-s/cm.^5 .
  • 17. total vascular resistance  arteries and capillaries=90%,  arterioles and capillaries=> 60%  large- and medium-sized arteries=15%  venous flow=10% .  atherosclerotic occlusive disease occur normally in vessels with low resistance .  The large and medium arteries are most commonly affected
  • 18. Effects Resistance in Series and Parallel
  • 19. Flow in Parallel Graft and Stenotic Artery  Surgeons occasionally express concern over the possibility that continued patency of a stenotic artery might lead to thrombosis of a parallel graft.  To allay this fear; they either avoid end-to-side anastomoses or ligate the stenotic artery.
  • 20.  Theoretical considerations strongly suggest that such concerns are not valid, provided that the arterial segment is sufficiently diseased to merit bypass grafting
  • 21. vein Grafts with Double Lumens  saphenous veins bifurcate into two separate and parallel channels that rejoin after a variable distance to reconstitute a single lumen.  When this situation is encountered, the surgeon must decide whether or not to include both channels in the graft
  • 22.  the combined resistance of the two parallel channels exceeds that of the undivided vein .
  • 24. theoretical advantages and disadvantages .  Flow rates in femorotibial grafts should theoretically be lower than those in the proximal segment of sequential grafts but higher than those in the distal segment .  Femorotibialgraft may be more likely to fail than the sequential graft.
  • 25. Blood Flow Patterns Laminar Flow  In steady-state conditions the flow pattern is laminar. All motion is parallel to the walls of the tube and the fluid is arranged in a series of concentric layers or laminae.  velocity profile that is parabolic in shape .
  • 26. PARABOLA- Laminar Flow  with velocity being highest in the center of the stream and becoming progressively lower towards the vessel wall.  The layer of fluid or blood in contact with the wall is stationary .
  • 27. TURBULENCE  In contrast to the linear streamlines of laminar flow, turbulence is an irregular flow state in which velocity varies rapidly with respect to space and time.  These random velocity changes result in inertial energy .and dissipation of fluid energy as heat. To measure --- Reynolds number, Re
  • 28. Reynolds number,  The point at which flow changes from laminar to turbulent is best defined in terms of a dimensionless quantity known as the Reynolds number, Re.  Re exceeds 2000= turbulence  laminar if Re s <2000.
  • 29.  Because Re values are well below 2000 in most peripheral arteries, turbulence is unlikely to occur under normal circumstances.  However, turbulence does appear to develop in the ascending aorta during the peak systolic ejection phase .
  • 30. Normal Pressure and Flow  As the arterial pressure pulse moves distally, the systolic pressure rises, the diastolic pressure falls, and the pulse pressure becomes wider.  The decrease in mean arterial pressure between the heart and the ankle is normally less than 10 mm Hg .  normal individuals, (ankle-brachial index) has a mean value of 1.11 ± 0.10 in the resting state.
  • 31. Boundary Layer Separation  In fluid flowing through a tube, the portion of fluid adjacent to the tube wall is referred to as the boundary layer.  This layer is subject to both frictional interactions with the tube wall and viscous forces generated by the more rapidly moving fluid toward the center of the tube. .
  • 32.  When the tube geometry changes suddenly, such as at points of curvature, branching, or variations in lumen diameter,small pressure gradients are created that cause the boundary layer to stop or reverse direction.  This change results in a complex, localized flow pattern known as an area of boundary layer separation or flow separation.
  • 33. Flow Separation and Shear  The clinical importance of boundary layer separation is that these localized flow disturbances may contribute to the formation of atherosclerotic plaques
  • 35. Anastomotic Configuration  To reduce energy losses due to flow disturbances, the transition from graft to host vessel should be as smooth as possible  End-to-end anastomoses, therefore, most closely approximate the ideal. End-to-side or side-to- end anastomoses always result in alterations in flow direction
  • 36.  Tailoring the anastomosis to enter the recipient artery or leave the donor artery at an acute angle will minimize but can never eliminate flow disturbances.
  • 37. Bifurcations and Branches  The branches of the arterial system produce sudden changes in the flow pattern that are potential sources of energy loss.  Flow patterns in a bifurcation are determined mainly by the area ratio and the branch angle.
  • 38. Area Ratio  The area ratio is defined as the combined area of the secondary branches divided by the area of the primary artery.  For efficient transmission of pulsatile energy across a bifurcation, the vascular impedance of the primary artery should equal that of the branches,
  • 39.  ideal area ratio of 1.15 for larger arteries and 1.35 for smaller arteries.  infrarenal aorta an area ratio of 0.8, approximately 22% of the incident pulsatile energy is reflected .  Hence atherosclerosis and aneurysms are common in this arterial segment
  • 40.  Despite their theoretical disadvantages, commercially available bifurcation grafts have functioned extremely well in a variety of clinical applications.
  • 41. The Branch Angle.  Flow disturbances are minimized when the angle is narrow and are exaggerated when the limbs are widely spread  The average angle between the human iliac arteries is 54 degrees;  however, with diseased or tortuous iliac arteries, this angle can approach 180 degrees.
  • 42. Angle Between The Limbs Of A Bifurcation Graft
  • 43. Bruits  Stenosis or irregularities of the vessel lumen produce turbulent flow patterns that set up vibrations in the arterial wall.  These vibrations generate displacement waves that radiate through the surrounding tissues and can be detected as audible sounds.
  • 44.  Generally, a soft, midsystolic bruit is associated with a relatively minor lesion that does not significantly reduce flow or pressure.  A bruit with a loud diastolic component suggests a stenosis severe enough to reduce flow and produce a pressure drop.  A bruit may be absent when an artery is nearly occluded or when the flow rate is extremely low
  • 45. Poststenotic Dilatation  The most likely explanation for this phenomenon is that arterial wall vibrations result in structural fatigue of elastin fibers.  In a series of animal model studies, poststenotic dilatations did not develop unless a bruit was present distal to the stenosis.
  • 46. CRITICAL STENOSIS  The degree of narrowing at which pressure and flow begin to be affected has been called the “critical stenosis.  appreciable changes in pressure and flow do not occur until the cross-sectional area of a vessel has been reduced by more than 75% or 50% reduction in diameter.
  • 47. VELOCITY OF BLOOD FLOW  Critical stenosis-varies with resistance of run off bed.  system with a high flow velocity (low resistance) shows a reduction in pressure with less narrowing than a system with low flow velocity.
  • 48. Palpable Pulses  Motion of the arterial wall is responsible for the palpable pulses that are so important in the physical examination of a patient in whom arterial disease is suspected.  However, a 7-mm femoral artery in a young subject would expand only 0.2 mm under the influence of 50 mm Hg pulse pressure, and the stiffer arteries in an older individual would expand even less.
  • 49.  circular cross-section into an ellipse.  It takes much less energy to bend the wall of an elliptically shaped vessel than it does to stretch the wall of a circular vessel.  Therefore, when the artery is partially compressed, its expansion in the direction of the compression is greatly augmented
  • 50.  The clinical value of pulse palpation is based on the assumption that the strength of the pulse is directly related to the pulse pressure, which should be decreased distal to an obstruction.  However, stiff, calcified vessels may display little or no palpable pulse even though there is no decrease in pulse pressure.
  • 51. Arterial Obstruction Collateral Circulation  A circuit includes diseased main artery a parallel system of collateral arteries, and the peripheral runoff bed.
  • 52. Resistance  fixed , collateral system resistance/segmental resistance.  variable of a peripheral runoff bed resistance.  Normally, the resting segmental resistance is low and the peripheral resistance is relatively high;
  • 53. Peripheral Runoff Bed  terminal arterioles and precapillary sphincters  Because of their small diameter and heavily muscled walls, these vessels are ideally suited for regulatory function.  Their resistance is subject to control by (1) the autonomic nervous system, (2) circulating catecholamines, (3) local metabolic products and (4) myogenic influences
  • 54. AUTOREGULATION  Autoregulation can compensate for a drop in perfusion pressure only until it falls below a critical level (e.g., about 20 to 30 mm Hg for skeletal muscle ).  With pressures below this level, normal blood flow is no longer maintained and flow responds passively to changes in perfusion pressure
  • 55.
  • 56. Exercise Therapy  exercise therapy is best suited for patients with mild, stable claudication who are not candidates for direct intervention .  In severe grade patients, the peripheral resistance has already been lowered to compensate for the increased segmental resistance, attempts to further reduce the peripheral resistance are seldom beneficial.
  • 57. Vasodilators  The rationale for the use of vasodilators is that they lower peripheral vascular resistance and improve limb blood flow.  Although this may occur in normal limbs, it is unlikely to be beneficial in limbs in which peripheral resistance is already decreased as a result of arterial disease.
  • 58. Sympathectomy  purpose of sympathectomy is to reduce peripheral resistance by release of vasomotor tone.  has little or no effect on collateral resistance.  Clinical improvement can occur only if the ischemic tissues are capable of further vasodilatation, as demonstrated by reactive hyperemia testing.
  • 59. VASCULAR STEAL  Vascular “steal” may arise when two runoff beds with different resistances must be supplied by a limited source of arterial inflow.  When an extraanatomic bypass is performed, a single donor artery must supply several vascular beds.
  • 60.  In the case of a femoral-femoral crossover graft, one iliac artery is the donor artery, the leg ipsilateral to the donor artery is the donor limb, and the contralateral leg is the recipient limb.  Studies of crossover grafts in animal models have shown that the immediate effect of the graft is to double flow in the donor artery.  When an arteriovenous fistula is created in the recipient limb, graft flow may increase by a factor of 10 without any evidence of steal from the donor limb
  • 61.  The most important factor contributing to vascular steal with a femoral-femoral graft is stenosis of the donor iliac artery.  Occlusive disease in the arteries of the donor limb distal to the origin of the graft does not result in steal, provided that the donor iliac artery is normal.  These principles also apply to other types of extraanatomic bypass grafts, including axillary- axillary, carotid-subclavian, and axillofemoral grafts.
  • 62. GRAVITY  Many patients with severe lower extremity ischemia discover that hanging their feet over the edge of the bed or walking a few steps often provides some relief from ischemic rest pain.  The improvement in peripheral perfusion that accompanies dependency can be documented objectively with measurement of transcutaneous oxygen tension, which may be increased severalfold over levels measured when the patient is supine
  • 63.  pressure in the dependent arteries, veins, and capillaries is increased by gravity commensurate with the vertical distance from the foot to the heart.  Although there is no increase in the arteriovenous pressure gradient,  the increased hydrostatic pressure dilates capillaries and microvascular vessels, thereby reducing their resistance, in turn augmenting blood flow.  Augmentation of blood flow does not occur in nonischemic limbs because the venoarterial reflex that serves to constrict arterioles in the dependent position remains functional
  • 64. Arterial Grafts  Because radius is so important in determining both viscous and inertial energy losses, the graft selected should be large enough to carry all the flow required at rest without causing a drop in pressure  it should also be large enough to accommodate any increased flow likely to be required during exercise without an appreciable drop in pressure.
  • 65.
  • 66.  Clearly, a graft with an inside diameter of less than 3 mm would be of marginal value at flow rates normally observed at rest (60 to 150 mL/min) and would be completely unsatisfactory during exercise (300 to 500 mL/min).  Therefore, under most physiologic conditions, an aortofemoral graft with 7-mm limbs should suffice, with flow restricted only during strenuous exercise.  However, 6-mm grafts might begin to show some restriction of flow even with mild to moderate exercise.
  • 67. PSEUDOINTIMA  prosthetic grafts develop a thin layer (0.5 to 1.0 mm) of pseudointima.  Therefore, after implantation, a 6-mm prosthetic graft might have an internal diameter of only 4 to 5 mm, and a 7- mm graft might have a luminal diameter of 5 to 6 mm.  For this reason, when one is performing a femoropopliteal bypass with a prosthetic graft, it is appropriate to select a graft with an original diameter of at least 6 mm.  Similarly, the original diameter of an aortofemoral graft limb should be at least 7 mm.
  • 68.  the diameter of a graft must not be too much larger than that of the recipient arteries  thrombus accumulates on the inner walls of grafts with excessive diameter (much as they do in aneurysms) as the flow stream tries to approximate the diameter of the recipient vessels and achieve optimal flow conditions
  • 69.  High flow velocity (high shear) is conducive to the formation of a thin, tightly adherent pseudointima,  so the diameter of a prosthetic graft should be small enough to ensure a rapid velocity of flow but large enough to avoid restriction of arterial inflow
  • 70. SHEAR RATES  Long-term patency of autogenous vein grafts is compromised by intimal hyperplasia, the development of which has also been associated with low shear rates.  Low shear rates cause smooth muscle cells to become secretory and enhance platelet adherence;  high shear rates foster continued patency and lessen the tendency of the intima to become hyperplastic.
  • 71. Aneurysm rupture and diameter  Tangenital stress and tensile stress,  Laplace’s law, which defines tangential tension (T) as the product of pressure and radius
  • 72.  Rupture occurs when the tangential stress within the arterial wall becomes greater than the tensile strength.  The tangential stress (τ) within the wall of a fluid-filled cylindrical tube can be expressed as: τ= P r/δ ,  P is the pressure exerted by the fluid, r is the internal radius, δ is the thickness of the tube wall.
  • 73.  The tendency of larger aneurysms to rupture is explained by the effect of increased radius on tangential stress .  The relationship between tangential stress and blood pressure accounts for the contribution of hypertension.
  • 74. Posterolateral rupture  55% of ruptured abdominal aortic aneurysms, the site of rupture is in the posterolateral aspect of the aneurysm wall.  The posterior wall of the aorta is relatively fixed against the spine, and repeated flexion in that area could result in structural fatigue and a localized area of weakness that might predispose to rupture,. .