CLINICAL LAB QUALITY
CONTROL
Dr.S. Sethupathy, M.D,Ph.D.,
Professor of Biochemistry,
Rajah Muthiah Medical College,
Annamalai University.
PURPOSE OF LAB TESTING
1. Single patient testing for diagnosis
2. Repeated testing in a patient for
monitoring
3. Testing a population –
Epidemiological work
QUALITY LAB SERVICE
Importance
• To produce reliable reports and timely reports
for diagnostic, prognostic and screening
purpose to achieve best patient care
• Cost-effective service to reduce burden on the
patient.
QUALITY CONTROL
• Good laboratory practice (GLP) requires Internal
Quality Control (IQC)
For Lab accreditation , QC is a must. Why?
• Quality must be designed from the front end and
not on the back end.
TOTAL QUALITY MANAGEMENT
• It is a continuous approach to improve quality and performance
towards meeting out the demands and satisfaction of the
customers.
• It involves :
• Management
• Work force
• Suppliers
• Feed back from customers
• Plan, do, review and act- sequence of QC
GOAL FOR QC
• To provide cost-effective , quality lab service
for the best patient care
How to achieve?
1. By increasing the error detection, reliability
and quality is assured.
2. By reducing the false rejection of runs, it will
be cost effective.
GOALS OF MEASUREMENT
• Accuracy and Precision
• Accuracy is the closeness of measured value with the
true value.
• The method adopted for clinical lab has certain level of
inaccuracy.
• To minimize inaccuracy ,Quality assurance program is
run .
PRECISION
• It is the closeness among replicative
measurements under specified conditions.
• The wide and random dispersion of values of
the replicative measurements indicate
imprecision of the assay.
• Imprecision leads to loss of reliability and
affects serial measurements. eg- bl. glucose.
VARIABLE
• Measurement is a varying one.
Variation can be due to
• a. Analytical variation – under control .
• b. Biological variation – Physiological
• c. Pathological variation - our concern
VARIABILITY OF LAB MEASUREMENTS
Due to :
• Sample collection
• Sample transport and storage
• Preparation of reagents, performance
• Maintenance of instruments
• Type of method, etc.
• Variation inherent to the biological sample.
To obtain reliable results
• All the sources of variation should be
minimized to differentiate pathological
variation.
• To achieve this, Minimize laboratory error.
• Perfectly estimate the components of
biological variation ( The inherent variation of
the sample)
CHECK POINTS OF CONTROL
• Pre-analytical
• Analytical
• Post-analytical
IQC AND EQA
• IQC is done on daily basis .
• It is useful for accepting or rejecting the runs based on
the variation limits.
• It is done for monitoring the quality of analysis.
• EQA is used to compare the quality between
laboratories of that area and to confirm the IQC
results.
HOW TO ACHIEVE QUALITY RESULTS?
• Adopting approved methods
• Adopting proper calibration protocol
• Adhering to maintenance schedule
• Training the technicians
• Internal quality control
• Adopting quality assurance program
• Proper documentation and regular audits for improvement
INTERNAL QUALITY CONTROL
• Internal Quality Control (IQC). It includes all
SQC methods which are performed every day by
the laboratory personnel with the laboratory’s
materials and equipment.
• It checks primarily the precision (repeatability
or reproducibility) of the method.
EXTERNAL QUALITY ASSESSMENT (EQA)
• External Quality Control (EQC). It includes all SQC
methods which are performed periodically (i.e. every
month, every two months, twice a year) by the
laboratory personnel with the contribution of an external
center .
• It checks primarily the accuracy of the laboratory’s
analytical methods.
• Certain EQC schemes check both the accuracy and
the precision.
HOW MUCH VARIATION IS ALLOWED IN
MEASUREMENTS?
• Variation is unavoidable.
• It should be controlled for clinical utility.
• It is needed to reject or accept analytical runs for
reliable reports.
• Fixing up the control limits is the first step of IQC.
CALCULATION OF CONTROL LIMITS
• Control limits : They consist of a center value (CL) and
an upper and low control limit (UCL &LCL).
• They are created by repetitive measurements of control
samples.
• In internal SQC two or more control samples are
assayed every day and at least once per day before the
patients’ samples.
The steps for their calculation are the following:
1. Collects 20 – 30 successive measurements from any control
level.
2. Standard deviation (s) and mean value (μ) are calculated. The
range μ±3s is considered as “trial limits”
3. If any of the measurements exceed the range μ±3s, the outlier
is rejected. The standard deviation and mean value are calculated
once more.
4. Repeat the previous procedure until no measurement exceeds
the range μ±3s.
5. The final μ and s are the mean value and the standard
deviation of the control limits.
• Control limits correspond to a normal distribution.
• “μ” is considered a true value of the daily control
values .
• “s” is equal to the inherent error.
• The mean value of the daily control values is
symbolized as x bar and its standard deviation as
“SD”.
• SD = the inherent error and any other RE.
ERRORS AND MISTAKES
• Errors: Non-conforming results with “statistical
meaning” due to non- human action.
• Mistakes: Non-conforming results with “no statistical
meaning” due to the human errors e.g. mixing up
samples, the special samples, the control samples.
• Random error- It is any positive or negative
unexpected deviation from the calculated
mean.
• Systematic error- It is a change in the mean
of control values.
• It may be gradual (Trend ) or abrupt (Shift ).
They are reproducible inaccuracies and are
often due to a persistent problem .
•TE (total error, total
analytical error -TAE)
The sum of random error
(imprecision) and systematic
error (bias).
TOTAL ANALYTIC ERROR
LEVEY- JENNINGS CHART
• It is used to plot control values for monitoring
• Done for successive day-to-day or run-to-run .
• A chart is prepared for each test.
• For level of control, the decision limits are first fixed.
• The limits are ± 1s, ± 2s, ± 3s from the mean.
LEVEY-JENNINGS CHART
RANDOM AND SYSTEMATIC ERRORS
RANDOM ERRORS
SYSTEMATIC ERRORS
QC STATISTICS
Mean
• To find out mean, the sum of the values in the data set
is divided by number of values in the data set.
Formula : ∑ Xn/n
• ∑ - sum
• Xn – Each value in the data set
• n – number of values in the data set
• Two groups having same mean can have different
spread or range.
VARIANCE
• Variance is the average squared deviation from the
mean:
• s2 = Σ ( xi - ẋ )2 / ( n - 1 )
• where s2 is the sample variance, ẋ is the sample mean,
xi is the ith element from the sample, and n is the
number of elements in the sample.
• The standard deviation is the square root of the
variance.
STANDARD DEVIATION (S)
• It is a measure of precision.
• It indicates how close the QC values are to
each other.
• Wide s indicates the imprecision and loss of
reliability.
• It is also useful for monitoring daily
performance.
STANDARD DEVIATION (S)
Formula
• s = √ ∑ (xi - ẋ) 2 / n-1
• s- standard deviation
• ẋ - mean
• ∑ (xi - ẋ) 2 - The sum of the squares of the
differences between individual QC values and the
mean
• n- number of the values in the data set
IF ANALYTICAL PROCESS IN CONTROL
• 68% of QC values are within ± 1s
• 95.5% of QC values are within ± 2s
• 99.7% of QC values are within ± 3s.
• 0.3 % of QC values are outside ± 3s and are
associated with significant error . So patient
results should not be reported.
STANDARD DEVIATION
• In case of serious loss of precision,
the following things are to be looked in to.
1. Change of reagent
2. Regarding maintenance schedule
3. Recent Calibration
4. Reagent , sample pipettes status
5. Any change of operator
COEFFICIENT OF VARIATION
• It is the ratio of standard deviation to the mean of the data
set expressed as percentage.
• CV = ( s / ẋ ) 100
• It is used
1. To compare precision of two methods.
2. To compare two different methods with
different s .
3.To compare instrument performance.
Reagent hCG (low level)
CV
hCG (Normal)
CV
hCG( high)
CV
Reagent 1 6.0% 4.5% 12%
Reagent 2 5.7% 5.0% 11%
Reagent 3 15.2% 5.7% 10%
DECISION ABOUT OF THE CHOICE OF
THE KIT
• Since the clinical decision about pregnancy
is based on low level, only kit 1 or 2 is
useful and so the kit 3 is not selected.
• Imprecision and inaccuracy are most
important at the clinical decision levels.
WHAT IS AN ACCEPTABLE CV?
It is decided based on
• Precision information given in the product insert or
instrument manual
• Inter-laboratory comparison (QC) programs
• Proficiency surveys
• Evaluations of methods, instruments published in journals
• CLIA proficiency limits (USA)
COEFFICIENT OF VARIATION RATIO (CVR)
• It is useful for comparing the precision of the lab test with peer
lab group using the same reagent, same instrument.
• CVR = Within lab CV / peer group CV
• If ratio is less than 1, the precision is better than peer group lab.
• Ratio greater than 1.5 needs investigation for the cause of
imprecision.
• Ratio greater than 2 needs troubleshooting and corrective
action.
CONTROL CHART
• The control chart is a graph used to study how a process
changes over time. Data are plotted in time order.
• A control chart always has a central line for the average, an
upper line for the upper control limit and a lower line for the
lower control limit.
• Lines are determined from historical data. By comparing current
data to these lines, you can draw conclusions about whether the
process variation is consistent (in control) or is unpredictable
(out of control, affected by special causes of variation).
WESTGARD RULES
• Dr.James Westgard’s six rules are used
individually or in combination to evaluate the
quality of analytical runs.
• The rules are expressed as NL where N stands
for number of control observations and
subscript L stands for statistical limits for
evaluating control observations.
WESTGARD RULES
• Rule 12s - single control observation lies outside the ± 2s
limits. It is a warning sign and the run is not rejected.
• In the absence of error, 4.5% of quality results lie between ± 2s
and ± 3s limits.
• The relationship between current, previous analytical runs
should be examined .
• If no source of error is found, it is an acceptable random error
and the patient results can be reported.
APPLICATIONS OF RULE 22S
Within run : It affects the current analytical run. If both
level 1 and level 2 controls are greater than 2s limits
on the same side of the mean, this violates within run
for systematic error.
Across runs : If level one is -1s and level 2 is +2.5s,
then previous run is examined and if level 2 is more
than + 2s, then it violates for systematic error and the
current run is rejected.
• Rule 13s - Any QC value outside
3s limits violates this rule for
random error and the run is
rejected. It may be the beginning
of a large systematic error.
Rule R4s : It identifies random error within
run only .
• If there is more than 4s difference between the
controls in the same run, this rule is violated for
random error. Eg: If level 1 is +2.8s and level 2 is
– 1.4 s , the difference 4.2s . It violates the run for
random error and the run is rejected.
• For three level controls, two of three shows more
than 4s difference, the run violates for random
error and it may be the beginning of systematic
error. So it should be rectified.
• Violation of the following rules does not require rejection of
analytical runs. These indicate smaller systematic error or
analytical bias which may not be clinically relevant. These can be
eliminated by calibration and or instrument maintenance.
• Rule 31s : Three consecutive results greater than 1s on the
same side of the mean.
• Rule 41s : Four consecutive results greater than 1s on the
same side of the mean.
• Violation of these rules - applications
• If level 1 control is used, it is systematic bias over a narrow
concentration. If three level controls are used in combination , it
violates for systematic error over a broader concentration.
• Rule 10 ẋ : 7 ẋ, 8 ẋ, 9 ẋ, 10 ẋ, 12 ẋ
• These rules are violated when 7or 8 or 9 or 10 or
12 control results are on the same side of the
mean regardless of the s.
• Applications
• If level 1 control is used, it is systematic bias over
a narrow concentration. If three level controls are
used in combination , it indicates systematic bias
over a broader concentration.
QC LOG
It should contain :
• Name of the test
• Name of the instrument
• Units
• Date
• Initials of the person performed
• Results of each level of control assayed
• Actions taken in case of out of control QC values
• Name of the method
• Assay temperature in case of enzymes
• Supervisor’s notes
Thank you

Weitere ähnliche Inhalte

Was ist angesagt?

Understanding statistics in laboratory quality control
Understanding statistics in laboratory quality controlUnderstanding statistics in laboratory quality control
Understanding statistics in laboratory quality control
Randox
 
Quality control in the medical laboratory
Quality control in the medical laboratoryQuality control in the medical laboratory
Quality control in the medical laboratory
Adnan Jaran
 

Was ist angesagt? (20)

Laboratory Method Verification, March 2017
Laboratory Method Verification, March 2017Laboratory Method Verification, March 2017
Laboratory Method Verification, March 2017
 
Quality control lecture CPath master 2014 Ain Shams
Quality control lecture CPath master 2014 Ain ShamsQuality control lecture CPath master 2014 Ain Shams
Quality control lecture CPath master 2014 Ain Shams
 
Lecture , quality_control_in_clinical_biochemistry_laboratory
Lecture , quality_control_in_clinical_biochemistry_laboratoryLecture , quality_control_in_clinical_biochemistry_laboratory
Lecture , quality_control_in_clinical_biochemistry_laboratory
 
Understanding statistics in laboratory quality control
Understanding statistics in laboratory quality controlUnderstanding statistics in laboratory quality control
Understanding statistics in laboratory quality control
 
Quality control
Quality controlQuality control
Quality control
 
Quality Control in a Medical Testing Laboratory
Quality Control in a Medical Testing LaboratoryQuality Control in a Medical Testing Laboratory
Quality Control in a Medical Testing Laboratory
 
External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
 
Quality Control in Pathological Laboratory
Quality Control in Pathological LaboratoryQuality Control in Pathological Laboratory
Quality Control in Pathological Laboratory
 
Quality Control in Laboratory
Quality Control in LaboratoryQuality Control in Laboratory
Quality Control in Laboratory
 
Quality control in the medical laboratory
Quality control in the medical laboratoryQuality control in the medical laboratory
Quality control in the medical laboratory
 
Troubleshooting iqc eqas 19.07.2018
Troubleshooting iqc eqas 19.07.2018Troubleshooting iqc eqas 19.07.2018
Troubleshooting iqc eqas 19.07.2018
 
WESTGARD RULES
WESTGARD RULESWESTGARD RULES
WESTGARD RULES
 
Quality control in clinical laboratory
Quality control in clinical laboratoryQuality control in clinical laboratory
Quality control in clinical laboratory
 
Quality control
Quality controlQuality control
Quality control
 
Laboratory Internal Quality Control presentation master revision, 2014
Laboratory Internal Quality Control presentation master revision, 2014Laboratory Internal Quality Control presentation master revision, 2014
Laboratory Internal Quality Control presentation master revision, 2014
 
Troubleshooting Poor EQA/QC Performance in the Laboratory
Troubleshooting Poor EQA/QC Performance in the Laboratory Troubleshooting Poor EQA/QC Performance in the Laboratory
Troubleshooting Poor EQA/QC Performance in the Laboratory
 
Quality assurance in a medical laboratory
Quality assurance in a medical laboratoryQuality assurance in a medical laboratory
Quality assurance in a medical laboratory
 
Medical Laboratory Quality Assurance
Medical Laboratory Quality AssuranceMedical Laboratory Quality Assurance
Medical Laboratory Quality Assurance
 
Quality control
Quality controlQuality control
Quality control
 
Quality control in clinical biochemistry
Quality control in clinical biochemistryQuality control in clinical biochemistry
Quality control in clinical biochemistry
 

Ähnlich wie Clinical lab qc sethu

Ähnlich wie Clinical lab qc sethu (20)

Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)
Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)
Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)
 
Overview of Quality Control and its implementation in the laboratory.pptx
Overview of Quality Control and its implementation in the laboratory.pptxOverview of Quality Control and its implementation in the laboratory.pptx
Overview of Quality Control and its implementation in the laboratory.pptx
 
quality control in clinical laboratory
quality control in clinical laboratory quality control in clinical laboratory
quality control in clinical laboratory
 
Basics of laboratory internal quality control, Ola Elgaddar, 2012
Basics of laboratory internal quality control, Ola Elgaddar, 2012Basics of laboratory internal quality control, Ola Elgaddar, 2012
Basics of laboratory internal quality control, Ola Elgaddar, 2012
 
QUALITY control in hematology by Dr yogeeta.pptx
QUALITY  control in hematology by Dr yogeeta.pptxQUALITY  control in hematology by Dr yogeeta.pptx
QUALITY control in hematology by Dr yogeeta.pptx
 
Control of analytical quality using stable control materials postgrad
Control of analytical quality using stable control materials postgradControl of analytical quality using stable control materials postgrad
Control of analytical quality using stable control materials postgrad
 
2. Quality Control Notes
2. Quality Control Notes2. Quality Control Notes
2. Quality Control Notes
 
Laboratory Quality Control .ppt
Laboratory Quality Control .pptLaboratory Quality Control .ppt
Laboratory Quality Control .ppt
 
Quality Assurance and Quality Control
Quality Assurance and Quality ControlQuality Assurance and Quality Control
Quality Assurance and Quality Control
 
Quality assurance and quality control
Quality assurance and quality controlQuality assurance and quality control
Quality assurance and quality control
 
Chapter 12
Chapter 12Chapter 12
Chapter 12
 
Quality control.pptx
Quality control.pptxQuality control.pptx
Quality control.pptx
 
Qc and qa
Qc and qaQc and qa
Qc and qa
 
Biological variation as an uncertainty component
Biological variation as an uncertainty componentBiological variation as an uncertainty component
Biological variation as an uncertainty component
 
QA-QC-PSK.pdf
QA-QC-PSK.pdfQA-QC-PSK.pdf
QA-QC-PSK.pdf
 
qc final final.pptx
qc final final.pptxqc final final.pptx
qc final final.pptx
 
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
 
Quality control
Quality controlQuality control
Quality control
 
Quality Control
Quality ControlQuality Control
Quality Control
 
qualitycontrolinbiochemistry-190731013722.pdf
qualitycontrolinbiochemistry-190731013722.pdfqualitycontrolinbiochemistry-190731013722.pdf
qualitycontrolinbiochemistry-190731013722.pdf
 

Mehr von subramaniam sethupathy

Mehr von subramaniam sethupathy (20)

clinical cases aminoacid metabolism.pptx
clinical cases aminoacid metabolism.pptxclinical cases aminoacid metabolism.pptx
clinical cases aminoacid metabolism.pptx
 
carbo case reports.pptx
carbo case reports.pptxcarbo case reports.pptx
carbo case reports.pptx
 
clinical cases lipid metabolism.pptx
clinical cases lipid metabolism.pptxclinical cases lipid metabolism.pptx
clinical cases lipid metabolism.pptx
 
Extra Cellular Matrix (ECM)
Extra Cellular Matrix (ECM)Extra Cellular Matrix (ECM)
Extra Cellular Matrix (ECM)
 
Neurotransmitters
Neurotransmitters Neurotransmitters
Neurotransmitters
 
Renal tubular acidosis
Renal tubular acidosisRenal tubular acidosis
Renal tubular acidosis
 
Parathyroid hormone
Parathyroid hormoneParathyroid hormone
Parathyroid hormone
 
Microsatellite
MicrosatelliteMicrosatellite
Microsatellite
 
Porphyrias
Porphyrias Porphyrias
Porphyrias
 
Multiple myeloma
Multiple myelomaMultiple myeloma
Multiple myeloma
 
Renal function tests
Renal function testsRenal function tests
Renal function tests
 
Tests for pancreatic and intestinal functions
Tests for pancreatic and intestinal functionsTests for pancreatic and intestinal functions
Tests for pancreatic and intestinal functions
 
Liver function tests 2020
Liver function tests 2020Liver function tests 2020
Liver function tests 2020
 
Monoclonal antibodies
Monoclonal antibodiesMonoclonal antibodies
Monoclonal antibodies
 
Acid base disorders
Acid base disordersAcid base disorders
Acid base disorders
 
Lipid transport 2020
Lipid transport 2020Lipid transport 2020
Lipid transport 2020
 
Water and electrolytes balance ss
Water and electrolytes balance ss Water and electrolytes balance ss
Water and electrolytes balance ss
 
Urine examination
Urine examinationUrine examination
Urine examination
 
Overview of Inborn errors of metabolism
Overview of Inborn errors of metabolism Overview of Inborn errors of metabolism
Overview of Inborn errors of metabolism
 
Thyroid disorders- recent advances
Thyroid disorders- recent advancesThyroid disorders- recent advances
Thyroid disorders- recent advances
 

Kürzlich hochgeladen

💚 Punjabi Call Girls In Chandigarh 💯Lucky 🔝8868886958🔝Call Girl In Chandigarh
💚 Punjabi Call Girls In Chandigarh 💯Lucky 🔝8868886958🔝Call Girl In Chandigarh💚 Punjabi Call Girls In Chandigarh 💯Lucky 🔝8868886958🔝Call Girl In Chandigarh
💚 Punjabi Call Girls In Chandigarh 💯Lucky 🔝8868886958🔝Call Girl In Chandigarh
Sheetaleventcompany
 
Mathura Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mathura Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMathura Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mathura Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Bhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetBhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...
Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...
Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...
Sheetaleventcompany
 
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
kozhikode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
kozhikode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetkozhikode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
kozhikode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetBareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetraisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
dhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
dhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetdhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
dhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Top 20 Famous Indian Female Pornstars Name List 2024
Top 20 Famous Indian Female Pornstars Name List 2024Top 20 Famous Indian Female Pornstars Name List 2024
Top 20 Famous Indian Female Pornstars Name List 2024
Sheetaleventcompany
 
Jalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Jalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetJalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Jalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Thoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Thoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetThoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Thoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in Lahore
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in LahoreBest Lahore Escorts 😮‍💨03250114445 || VIP escorts in Lahore
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in Lahore
Deny Daniel
 
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetNanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 

Kürzlich hochgeladen (20)

💚 Punjabi Call Girls In Chandigarh 💯Lucky 🔝8868886958🔝Call Girl In Chandigarh
💚 Punjabi Call Girls In Chandigarh 💯Lucky 🔝8868886958🔝Call Girl In Chandigarh💚 Punjabi Call Girls In Chandigarh 💯Lucky 🔝8868886958🔝Call Girl In Chandigarh
💚 Punjabi Call Girls In Chandigarh 💯Lucky 🔝8868886958🔝Call Girl In Chandigarh
 
Mathura Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mathura Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMathura Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mathura Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Escorts Service Ahmedabad🌹6367187148 🌹 No Need For Advance Payments
Escorts Service Ahmedabad🌹6367187148 🌹 No Need For Advance PaymentsEscorts Service Ahmedabad🌹6367187148 🌹 No Need For Advance Payments
Escorts Service Ahmedabad🌹6367187148 🌹 No Need For Advance Payments
 
Jaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
Jaipur Call Girls 9257276172 Call Girl in Jaipur RajasthanJaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
Jaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
 
Bhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetBhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...
Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...
Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...
 
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
kozhikode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
kozhikode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetkozhikode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
kozhikode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Kolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girl
Kolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girlKolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girl
Kolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girl
 
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetBareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetraisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
dhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
dhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetdhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
dhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Top 20 Famous Indian Female Pornstars Name List 2024
Top 20 Famous Indian Female Pornstars Name List 2024Top 20 Famous Indian Female Pornstars Name List 2024
Top 20 Famous Indian Female Pornstars Name List 2024
 
Jalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Jalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetJalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Jalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Thoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Thoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetThoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Thoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in Lahore
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in LahoreBest Lahore Escorts 😮‍💨03250114445 || VIP escorts in Lahore
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in Lahore
 
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetNanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Kochi call girls Mallu escort girls available 7877702510
Kochi call girls Mallu escort girls available 7877702510Kochi call girls Mallu escort girls available 7877702510
Kochi call girls Mallu escort girls available 7877702510
 
Call Girls Service Mohali {7435815124} ❤️VVIP PALAK Call Girl in Mohali Punjab
Call Girls Service Mohali {7435815124} ❤️VVIP PALAK Call Girl in Mohali PunjabCall Girls Service Mohali {7435815124} ❤️VVIP PALAK Call Girl in Mohali Punjab
Call Girls Service Mohali {7435815124} ❤️VVIP PALAK Call Girl in Mohali Punjab
 

Clinical lab qc sethu

  • 1. CLINICAL LAB QUALITY CONTROL Dr.S. Sethupathy, M.D,Ph.D., Professor of Biochemistry, Rajah Muthiah Medical College, Annamalai University.
  • 2. PURPOSE OF LAB TESTING 1. Single patient testing for diagnosis 2. Repeated testing in a patient for monitoring 3. Testing a population – Epidemiological work
  • 3. QUALITY LAB SERVICE Importance • To produce reliable reports and timely reports for diagnostic, prognostic and screening purpose to achieve best patient care • Cost-effective service to reduce burden on the patient.
  • 4. QUALITY CONTROL • Good laboratory practice (GLP) requires Internal Quality Control (IQC) For Lab accreditation , QC is a must. Why? • Quality must be designed from the front end and not on the back end.
  • 5.
  • 6. TOTAL QUALITY MANAGEMENT • It is a continuous approach to improve quality and performance towards meeting out the demands and satisfaction of the customers. • It involves : • Management • Work force • Suppliers • Feed back from customers • Plan, do, review and act- sequence of QC
  • 7. GOAL FOR QC • To provide cost-effective , quality lab service for the best patient care How to achieve? 1. By increasing the error detection, reliability and quality is assured. 2. By reducing the false rejection of runs, it will be cost effective.
  • 8. GOALS OF MEASUREMENT • Accuracy and Precision • Accuracy is the closeness of measured value with the true value. • The method adopted for clinical lab has certain level of inaccuracy. • To minimize inaccuracy ,Quality assurance program is run .
  • 9. PRECISION • It is the closeness among replicative measurements under specified conditions. • The wide and random dispersion of values of the replicative measurements indicate imprecision of the assay. • Imprecision leads to loss of reliability and affects serial measurements. eg- bl. glucose.
  • 10.
  • 11. VARIABLE • Measurement is a varying one. Variation can be due to • a. Analytical variation – under control . • b. Biological variation – Physiological • c. Pathological variation - our concern
  • 12. VARIABILITY OF LAB MEASUREMENTS Due to : • Sample collection • Sample transport and storage • Preparation of reagents, performance • Maintenance of instruments • Type of method, etc. • Variation inherent to the biological sample.
  • 13. To obtain reliable results • All the sources of variation should be minimized to differentiate pathological variation. • To achieve this, Minimize laboratory error. • Perfectly estimate the components of biological variation ( The inherent variation of the sample)
  • 14. CHECK POINTS OF CONTROL • Pre-analytical • Analytical • Post-analytical
  • 15.
  • 16.
  • 17.
  • 18.
  • 19.
  • 20.
  • 21.
  • 22.
  • 23. IQC AND EQA • IQC is done on daily basis . • It is useful for accepting or rejecting the runs based on the variation limits. • It is done for monitoring the quality of analysis. • EQA is used to compare the quality between laboratories of that area and to confirm the IQC results.
  • 24.
  • 25. HOW TO ACHIEVE QUALITY RESULTS? • Adopting approved methods • Adopting proper calibration protocol • Adhering to maintenance schedule • Training the technicians • Internal quality control • Adopting quality assurance program • Proper documentation and regular audits for improvement
  • 26.
  • 27.
  • 28.
  • 29.
  • 30.
  • 31.
  • 32.
  • 33.
  • 34. INTERNAL QUALITY CONTROL • Internal Quality Control (IQC). It includes all SQC methods which are performed every day by the laboratory personnel with the laboratory’s materials and equipment. • It checks primarily the precision (repeatability or reproducibility) of the method.
  • 35. EXTERNAL QUALITY ASSESSMENT (EQA) • External Quality Control (EQC). It includes all SQC methods which are performed periodically (i.e. every month, every two months, twice a year) by the laboratory personnel with the contribution of an external center . • It checks primarily the accuracy of the laboratory’s analytical methods. • Certain EQC schemes check both the accuracy and the precision.
  • 36. HOW MUCH VARIATION IS ALLOWED IN MEASUREMENTS? • Variation is unavoidable. • It should be controlled for clinical utility. • It is needed to reject or accept analytical runs for reliable reports. • Fixing up the control limits is the first step of IQC.
  • 37. CALCULATION OF CONTROL LIMITS • Control limits : They consist of a center value (CL) and an upper and low control limit (UCL &LCL). • They are created by repetitive measurements of control samples. • In internal SQC two or more control samples are assayed every day and at least once per day before the patients’ samples.
  • 38. The steps for their calculation are the following: 1. Collects 20 – 30 successive measurements from any control level. 2. Standard deviation (s) and mean value (μ) are calculated. The range μ±3s is considered as “trial limits” 3. If any of the measurements exceed the range μ±3s, the outlier is rejected. The standard deviation and mean value are calculated once more. 4. Repeat the previous procedure until no measurement exceeds the range μ±3s. 5. The final μ and s are the mean value and the standard deviation of the control limits.
  • 39. • Control limits correspond to a normal distribution. • “μ” is considered a true value of the daily control values . • “s” is equal to the inherent error. • The mean value of the daily control values is symbolized as x bar and its standard deviation as “SD”. • SD = the inherent error and any other RE.
  • 40. ERRORS AND MISTAKES • Errors: Non-conforming results with “statistical meaning” due to non- human action. • Mistakes: Non-conforming results with “no statistical meaning” due to the human errors e.g. mixing up samples, the special samples, the control samples.
  • 41. • Random error- It is any positive or negative unexpected deviation from the calculated mean. • Systematic error- It is a change in the mean of control values. • It may be gradual (Trend ) or abrupt (Shift ). They are reproducible inaccuracies and are often due to a persistent problem .
  • 42. •TE (total error, total analytical error -TAE) The sum of random error (imprecision) and systematic error (bias).
  • 44. LEVEY- JENNINGS CHART • It is used to plot control values for monitoring • Done for successive day-to-day or run-to-run . • A chart is prepared for each test. • For level of control, the decision limits are first fixed. • The limits are ± 1s, ± 2s, ± 3s from the mean.
  • 49. QC STATISTICS Mean • To find out mean, the sum of the values in the data set is divided by number of values in the data set. Formula : ∑ Xn/n • ∑ - sum • Xn – Each value in the data set • n – number of values in the data set • Two groups having same mean can have different spread or range.
  • 50. VARIANCE • Variance is the average squared deviation from the mean: • s2 = Σ ( xi - ẋ )2 / ( n - 1 ) • where s2 is the sample variance, ẋ is the sample mean, xi is the ith element from the sample, and n is the number of elements in the sample. • The standard deviation is the square root of the variance.
  • 51. STANDARD DEVIATION (S) • It is a measure of precision. • It indicates how close the QC values are to each other. • Wide s indicates the imprecision and loss of reliability. • It is also useful for monitoring daily performance.
  • 52. STANDARD DEVIATION (S) Formula • s = √ ∑ (xi - ẋ) 2 / n-1 • s- standard deviation • ẋ - mean • ∑ (xi - ẋ) 2 - The sum of the squares of the differences between individual QC values and the mean • n- number of the values in the data set
  • 53. IF ANALYTICAL PROCESS IN CONTROL • 68% of QC values are within ± 1s • 95.5% of QC values are within ± 2s • 99.7% of QC values are within ± 3s. • 0.3 % of QC values are outside ± 3s and are associated with significant error . So patient results should not be reported.
  • 54. STANDARD DEVIATION • In case of serious loss of precision, the following things are to be looked in to. 1. Change of reagent 2. Regarding maintenance schedule 3. Recent Calibration 4. Reagent , sample pipettes status 5. Any change of operator
  • 55. COEFFICIENT OF VARIATION • It is the ratio of standard deviation to the mean of the data set expressed as percentage. • CV = ( s / ẋ ) 100 • It is used 1. To compare precision of two methods. 2. To compare two different methods with different s . 3.To compare instrument performance.
  • 56. Reagent hCG (low level) CV hCG (Normal) CV hCG( high) CV Reagent 1 6.0% 4.5% 12% Reagent 2 5.7% 5.0% 11% Reagent 3 15.2% 5.7% 10%
  • 57. DECISION ABOUT OF THE CHOICE OF THE KIT • Since the clinical decision about pregnancy is based on low level, only kit 1 or 2 is useful and so the kit 3 is not selected. • Imprecision and inaccuracy are most important at the clinical decision levels.
  • 58. WHAT IS AN ACCEPTABLE CV? It is decided based on • Precision information given in the product insert or instrument manual • Inter-laboratory comparison (QC) programs • Proficiency surveys • Evaluations of methods, instruments published in journals • CLIA proficiency limits (USA)
  • 59. COEFFICIENT OF VARIATION RATIO (CVR) • It is useful for comparing the precision of the lab test with peer lab group using the same reagent, same instrument. • CVR = Within lab CV / peer group CV • If ratio is less than 1, the precision is better than peer group lab. • Ratio greater than 1.5 needs investigation for the cause of imprecision. • Ratio greater than 2 needs troubleshooting and corrective action.
  • 60. CONTROL CHART • The control chart is a graph used to study how a process changes over time. Data are plotted in time order. • A control chart always has a central line for the average, an upper line for the upper control limit and a lower line for the lower control limit. • Lines are determined from historical data. By comparing current data to these lines, you can draw conclusions about whether the process variation is consistent (in control) or is unpredictable (out of control, affected by special causes of variation).
  • 61.
  • 62. WESTGARD RULES • Dr.James Westgard’s six rules are used individually or in combination to evaluate the quality of analytical runs. • The rules are expressed as NL where N stands for number of control observations and subscript L stands for statistical limits for evaluating control observations.
  • 63. WESTGARD RULES • Rule 12s - single control observation lies outside the ± 2s limits. It is a warning sign and the run is not rejected. • In the absence of error, 4.5% of quality results lie between ± 2s and ± 3s limits. • The relationship between current, previous analytical runs should be examined . • If no source of error is found, it is an acceptable random error and the patient results can be reported.
  • 64.
  • 65. APPLICATIONS OF RULE 22S Within run : It affects the current analytical run. If both level 1 and level 2 controls are greater than 2s limits on the same side of the mean, this violates within run for systematic error. Across runs : If level one is -1s and level 2 is +2.5s, then previous run is examined and if level 2 is more than + 2s, then it violates for systematic error and the current run is rejected.
  • 66.
  • 67. • Rule 13s - Any QC value outside 3s limits violates this rule for random error and the run is rejected. It may be the beginning of a large systematic error.
  • 68.
  • 69. Rule R4s : It identifies random error within run only . • If there is more than 4s difference between the controls in the same run, this rule is violated for random error. Eg: If level 1 is +2.8s and level 2 is – 1.4 s , the difference 4.2s . It violates the run for random error and the run is rejected. • For three level controls, two of three shows more than 4s difference, the run violates for random error and it may be the beginning of systematic error. So it should be rectified.
  • 70.
  • 71. • Violation of the following rules does not require rejection of analytical runs. These indicate smaller systematic error or analytical bias which may not be clinically relevant. These can be eliminated by calibration and or instrument maintenance. • Rule 31s : Three consecutive results greater than 1s on the same side of the mean. • Rule 41s : Four consecutive results greater than 1s on the same side of the mean. • Violation of these rules - applications • If level 1 control is used, it is systematic bias over a narrow concentration. If three level controls are used in combination , it violates for systematic error over a broader concentration.
  • 72.
  • 73. • Rule 10 ẋ : 7 ẋ, 8 ẋ, 9 ẋ, 10 ẋ, 12 ẋ • These rules are violated when 7or 8 or 9 or 10 or 12 control results are on the same side of the mean regardless of the s. • Applications • If level 1 control is used, it is systematic bias over a narrow concentration. If three level controls are used in combination , it indicates systematic bias over a broader concentration.
  • 74.
  • 75. QC LOG It should contain : • Name of the test • Name of the instrument • Units • Date • Initials of the person performed • Results of each level of control assayed • Actions taken in case of out of control QC values • Name of the method • Assay temperature in case of enzymes • Supervisor’s notes

Hinweis der Redaktion

  1. ∑∑