SlideShare ist ein Scribd-Unternehmen logo
1 von 58
Dr.V.Ramana reddy




                    1
Fluoroscopy
Purpose

To visualize, in real time:
 organ motion

 ingested or injected contrast agents

 insert stents

 cathetarize small blood vessels


             REAL TIME IMAGING           2
X-rays were discovered because of
  there ability to cause fluorescence .

Thefirst x-ray image of human part was
 observed fluoroscopically-Dr. Glasser.



                                          3
THE FLUOROSCOPE
   First generation fluoroscopes consisted of an x-ray
    tube, an x-ray table and a fluoroscopic screen.
   The fluorescent material used in screen was copper
    activated zinc cadmium sulfide that emitted light in
    yellow-green spectrum.
   A sheet of lead glass covered the screen, so that
    radiologist could stare directly into the screen with out
    having the x-ray beam strike his eyes.
    Screen fluoroscence was very faint so, the
    examination was carried out in a dark room by the
    radiologist who had to adapt his eyes by wearing red
    goggles for 20-30 mins prior to the examination 
    technique is now obsolete & gone.                      4
FLUOROSCOPE




              5
Photograph
shows an early
      (1933)
  fluoroscopic
 system in use
    before the
  development
     of image
intensification.
    An actual
  fluoroscopic
  examination
     with this
 device would
 have occurred
 in a darkened
      room.    6
IMAGE INTENSIFIER DESIGN
   Image intensifier was discovered in 1950s-to
    produce an image bright enough to allow cone
    vision without giving the pt an excess radiation
    exposure.
   The components of an x-ray image intensifier
   The tube itself is an evacuated glass
    envelope ,a vacuum tube containing-
          1.input phosphor and photocathode .
          2.electrostatic focusing lens.
          3.accelerating anode.
          4.out put phosphor.                    7
   After an x-ray beam passes the pt it enters the image
    intensifier tube the input fluorescent screen absorbs
    x-ray photons and converts their energy into light
    photons.

   The light photons strike the photo cathode, causing it
    to emit photoelectrons  these electrons are
    immediately drawn away from the photocathode by
    the high potential difference betn it &the accelerating
    anode.

   As the electrons flow from the cathode towards the
    anode, they are focused by an electrostatic lens which
    guides them to the output fluorescent screen without
    distorting their geometric configuration.

                                                         8
   The electrons strike the output screen, which
    emits the light photons that carry the
    fluoroscopic images to the eye of the observer.

   In intensifier tube, the image is first carried by
    the x-ray photons, then by the light photons,
    next by the electrons &finally by the light
    photons.




                                                     9
IMAGE INTENSIFIER




                    10
11
INPUT PHOSPHOR & PHOTO CATHODE:

   The input fluorescent screen in image intensifiers is
    cesium iodide (CsI). (older intensifier- silver activated
    zinc cadmium sulfide).
   CsI is deposited on a thin aluminum substrate by a
    process called “vapor deposition”.  an interesting &
    useful characteristic of CsI is that during the
    deposition process the crystals of CsI grow in tiny
    needles perpendicular to the substrate.  There by
    reducing scattering.




                                                          12
INPUT PHOSPHOR & PHOTO CATHODE:
   Image quality is dramatically better with CSI input
    screen than it was with zinc cadmium sulfide screens.
   Three physical characteristics of CsI make it superior.
        1. vertical orientation of the crystals.
        2. A greater packing density &
        3. A more favorable effective atomic number.
Phosphor   thickness have been reduced comparably
from app. 0.3mm with Zn-Cd su to 0.1mm with CsI. The
principal advantage of thinner phosphor layer combined
with needle shaped crystals is improved resolution.


                                                        13
PHOTO CATHODE:
   The photo cathode is a photoemissive metal
    (commonly a combination of antimony & cesium
    compounds).
   When the light from the fluorescent screen strikes the
    photo cathode, photo electrons are emitted in numbers
    proportional to the brightness of the screen.
   The photo cathode is applied directly to the CsI input
    phosphor.
   The photoelectrons thus produced has to be moved to
    the other end of the image intensifier. This can be
    done using an electrostatic focusing lens and an
    accelerating anode.
                                                        14
ELECTROSTATIC FOCUSING CUP:
   The lens is made up of a series positively charged
    electrodes that are usually plated on to the inside surface
    of the glass envelope.
   These electrodes focus the electron beam as it flows from
    the photocathode toward the output phosphor.
   Electron focusing inverts & reverses the image which is
    called “point inversion” because all the electrons pass
    through a common focal point on their way to output
    phosphor.
   The input phosphor is curved to ensure that electrons
    emitted at the peripheral regions of the photocathode
    travel the same distance as those emitted from the
    central region.
   The image on the output phosphor is reduced in size
    ,which is one of the principle reasons why it is brighter.
                                                            15
ACCELERATING ANODE :
   The anode is located in the neck of the image tube.
   Its function is to accelerate electrons emitted from the
    photocathode towards the output screen. the anode has a
    +ve potential of 25 to 35 kv relative to the photocathode,
    so it accelerates electrons to a tremendous velocity.
OUTPUT PHOSPHOR:
   The output fluorescent screen of image intensifiers is
    silver activated zn-cd sulfide, the same used in    Ist
    generation input phosphor.
   Crystal size and layer thickness are reduced to maintain
    resolution in the minified image.
   A thin layer of aluminum is plated onto the fluorescent
    screen prevent light from moving retrograde through
    the tube & activating the photocathode.
                                                          16
   The glass tube of the image intensifier is abt 2 to 4mm
    thick &is enclosed in a lead lined metal container
    protects the operator from stray radiation.
   The output phosphor image is viewed either directly
    through a series of lenses and mirrors or indirectly
    through closed circuit TV.
BRIGHTNESS GAIN:
 Two methods are used to evaluate the brightness gain
  of image intensifiers. the first compares the luminance
  of an intensifier output screen to that of a Patterson
  type B2 fluoroscopy screen when both are exposed to
  same quantity of radiation.
 The brightness gain is the ratio of the two
  illuminations.
 Brightness gain=intensifier luminance/Patterson b-2
  lumin.
                                                         17
   Another way of evaluation of brightness gain is called as
    “conversion factor”
   Conversion factor =cd/m2 by mR/sec
   Output screen luminance is measured in candelas.
   Radiation quality & output luminance are explicitly
    defined, so the method is accurate & reproducible.
   The brightness gain tends to deteriorate as the image
    intensifier ages. ie the pt dose with an older image
    intensifier tends to be higher than with a new intensifier
    of the same type.
   The brightness gain of an imag inten comes from 2
    completely unrelated sources called minification gain
    &flux gain.
                                                           18
MINIFICATION GAIN:
   The brightness gain from minification is produced by a
    reduction in image size.
   The quantity of gain depends on the relative areas of
    input &output screens. coz the size of the intensifier is
    usually indicated by its diameter, so minif gain is
    expressed as MG=(d1/d0)2.d1=diameter of input
    screen,d0=diameter of output screen.
   Most img inten have an input screen from 5 to9 in.& an
    output screen of app 1in in diameter.
   Theoretically minification can be increased indefinitely
    as we can see from above formula, but as the
    minification is increased the image becomes smaller.-
    disadvantage.
   Hence image has to be magnified and viewed which will
    result in reduce brightness & precipitous drop in
    resolution.                                             19
FLUX GAIN:

   FLUX gain increases the brightness of the fluoroscopic
    image by a factor of app 50.
   The total brightness of an imag intes is product of
    minification & flux gain: ie
     Brightness gain =minification gain x flux gain.


                                                          20
MULTIPLE-FIELD IMAGE INTENSIFIERS

   Dual field or triple field imsg intes attempt to
    resolve the conflicts btn image size & quality.
   They can be operated in several modes, including
    the 4.5in, a 6in, or a 9in mode. the 9in mode is
    used to view large anatomic areas. When size is
    unimportant the 4.5in or 6in mode is used coz of
    better resultant image quality.
   Field size is changed by a simple electronic
    principle: the higher the voltage on the
    electrostatic focusing lens, the more the electron
    beam is focused.                              21
   This figure shows this principle
    applied to a dual field imag
    intes.
   In the 9in mode, the
    electrostatic focusing voltage is
    decreased. the electrons focus to
    a point or cross, close to the
    output phosphor & the final
    image is actually smaller than
    the phosphor.
   In 6in mode the electrostatic
    focusing voltage is increased &
    the electrons focus farther away
    from the output phosphor. after
    the electrons cross they diverge,
    so the image on the output
    phosphor is larger than in the      22
Optical coupling
   Optical system transmits the output of the image
    intensifier to the light sensitive area of the video
    camera
   The optical distributor include beam-splitting
    mirror, which directs a portion of the light from
    the image intensifier output window to an
    accessory device for image recording and passes
    the remainder to the video camera.
   Two lenses are mounted in tandem
   The II and the vidicon are placed at the focal
    planes of the two lenses
                                                      23
24
Closed-circuit Television System
   used to view the image intensifier output image
   Consists of
       1)Television camera
       2) Camera control unit
       3) Monitor
   The television system allows for real-time
    viewing of the fluoroscopic image by several
    people at once from one monitor or multiple
    monitors

                                                      25
Television Camera Tube

   Output phosphor is directly coupled to a TV
    camera tube

   Plumbicon
   Vidicon
   CCD

                                              26
 The basic video camera consists of
1) vacuum tube cylinder (approximately 2.5 cm in
  diameter) surrounded by electromagnetic focusing
  coils ,2 pairs of electrostatic deflecting coils
 2) photoconductive target
3) a scanning electron beam


Target assembly:
a) Glass plate assembly
b) signal plate
c) Target
                                                     27
PICK UP TUBE   28
Target
   Functionally most important i n tube

   Thin film of photoconductive material, antimony sulfide
    suspended as globules in mica matrix.

   The optical coupling lens focuses the image intensifier output
    image onto the target, forming a charge image within the
    photoconductive layer

    This latent image is read out by the electron beam, which
    scans across the target in a series of horizontal raster lines.

   As the scanning electron beam moves across the target, a
    current signal is produced that represents the two-dimensional
    image as a continuous series of raster lines with varying
    voltage levels.
                                                                      29
Video signal
   When globules absorbs light ,photoelectrons are
    emitted
   The globule becomes positively charged
   The electron beam scans the electrical image stored
    on the target & fills in the holes left by the emitted
    photoelectrons, discharging the tiny globular
    capacitors
   When the electrons in beam neutralize the positive
    charge in the globules , the electrons on the signal
    plate leave the plate via resistor
   These moving electrons form a current flowing
    through a resistor and voltage across the resistor
   This voltage ,when collected for each neutralized
    globule, constitutes the video signal
                                                             30
31
Television monitor
   The video signal produced by the video camera is converted into a visible
    image by the monitor

   Contains picture tube & controls for regulating brightness & contrast

   Picture tube contains-
        Electron gun
        control grid
        anode
        focusing coil, deflecting coil-control the electron beam
                              synchrony with the camera tube

   Control grid-receive video signal from Camera control unit ,uses this signal
    to regulate the no. of electrons in electron beam
   Anode –carries higher potential (10,000V) accelerates the electron beam to
    much higher velocity

   The electron Strike the fluorescent screen ,emit large number of light
    photons.
                                                                             32
33
Image Recording
Two modes of recording th flouroscopic
 image -
1) light image from output phosphor of II
 recorded on film with a photospot camera
 or cine camera
 2) electrical signal generated by TV camera-
 includes magnetic tape, magnetic discs &
 optical discs

                                            34
   DIRECT FILM RECORDING
    SPOT FILM DEVICES:
   Fluoroscopic systems designed for gastrointestinal imaging
    are generally equipped with a spot film device.

   The spot film device allows exposure of a conventional
    screen-film cassette in conjunction with fluoroscopic
    viewing. This rather familiar system, located in front of the
    image intensifier, accepts the screen-film cassette and
    "parks" it out of the way during fluoroscopy

   Cassettes may be loaded from the front or rear depending on
    the design of the system.


                                                                35
  Standard spot film
 imaging configuration
         typical of
      gastrointestinal
fluoroscopy equipment.
The screen-film cassette
 is parked out of the x-
 ray field until the spot
 film trigger is pressed,
     causing both the
      cassette and the
   formatting mask to
   move into position.
                       36
   The x-ray field size is also reduced automatically by the
    collimators at the time of exposure to minimize scattered
    radiation and patient radiation dose.
   The fluoroscopist can override this automatic collimation to
    further reduce the x-ray field.
   Spot film imaging uses essentially the same technology as
    conventional screen-film radiography.

    Differences and limitations of spot film imaging
    compared with general radiography.
       One major limitation is the range of film sizes available for
        spot film imaging. Although some older fluoroscopy
        equipment is limited to a single size, usually 24 x 24 cm,
        current equipment allows a range of film sizes to be used,
        from 20 x 25 cm to 24 x 35 cm.
       Spot film devices usually allow more than one image to be
        obtained on a single film.
       Formats typically include one, two, three, four, or six images
        on a film.
                                                                         37
   Moving the spot film device closer to the patient
    reduces the amount of magnification and decreases
    the patient radiation dose.
   A number of factors affect patient doses in spot film
    imaging.
       The source-to-skin distance is shorter in spot film imaging
        than in general radiography.
       Although the automatic exposure control system fixes the
        exposure to the screen, the shorter source-to-skin distance
        increases the inverse square reduction in radiation intensity
        as it passes through the patient.
       This increase tends to make the skin entrance exposure
        higher.
       The field size in spot film imaging is generally smaller than
        that used in general radiography.
       This smaller field size reduces scatter and therefore tends to
        reduce dose. For the same reason, grids used in fluoroscopy
        generally have a lower grid ratio and therefore a smaller
        Bucky factor, which also leads to lower dose.
       These effects tend to offset each other to a large extent.     38
   One of the major shortcomings of conventional spot
    film devices is the delay involved in moving the
    cassette into position for exposure.
    In gastrointestinal imaging, this delay can be
    overcome by using photofluorography.
   In vascular imaging, more rapid film movement is
    achieved with automatic film changers.




                                                         39
AUTOMATIC FILM CHANGERS
   The automatic film changers used in vascular imaging are
    also screen-film systems.
   They can be found in several varieties. Some are large,
    floor-mounted boxes, but systems more commonly used
    today mount on the image intensifier
   The system consists of a supply magazine for holding
    unexposed film, a receiving magazine, a pair of
    radiographic screens, and a mechanism for transferring the
    film.
   When an exposure is required, the screens are mechanically
    separated, the film is pulled into place between them, and
    they are closed.
   After the film is exposed, the screens separate again.
   The film is moved to the receiver, and another film is pulled
    into place for the next exposure.
   The number of films and filming rates must be
    preprogrammed for proper operation.
                                                               40
Photospot camera
   Records the the image output of an II on a film
   Film – role film/cut film(10 cm)
   Advantage –1)reduction in pt exposure
                  2)film does not have to be changed
                     b/w exposures

                 3)exposure times are shorter-motion is less
                     likely problem
                4)films can be taken more rapidly
                 5)possible to record & view image at same
                     time

                                                               41
Framing with spot film cameras:
   Framing –utilisation of available area on film
   The output phosphor of II tube is round,
    shape of film is square
   4 framing patterns
    1) Exact framing
    2) Equal area framing
    3) mean diameter framing
    4) total area framing
   Equal area framing or mean diameter framing
    is recommended for most clinical situations.

                                                     42
43
Cineflourography
   Process of recording fluoroscopic images on
    movie (cine) film
   Two film sizes- 16 mm, 35 mm
   Cine camera-components are lens, iris
    diaphragm, shutter, aperture, pressure plate, pull
    down arm & film transport mechanism



                                                     44
45
TV image    recorders      :

3) TAPE RECORDER : used for both recoding& playback

•  as a recorder receives video signal from camera
   control unit
• for playback transmits the signal to one or more
  several TV monitors

Components:1)magnetic tube
           2)writing head
            3) tape transport system


Writing head converts an electrical signal in to magnetic
  field for recoding & converts magnetic signal to electric
  signal for replay
                                                          46
47
48
DIGITAL FLUOROGRAPHY
     Digital charge-coupled device (CCD) TV cameras are
rapidly replacing conventional TV cameras in fluoroscopic
systems.

     An analog, high-resolution (1,023-line) TV camera has
a vertical resolution of about 358 line pairs.

     A high-resolution CCD camera with a 1,024 x 1,024
matrix will provide equivalent vertical resolution. However,
the digital camera will have the same vertical and
horizontal resolution, whereas the horizontal resolution of
the analog camera is defined by its electronic band-pass.




                                                             49
For a 15-cm-diameter intensifier input, the
limiting resolution of the CCD camera would be
358 lp/150 mm or 2.4 lp/mm. This result is about
half the resolution of a photospot film. This
resolution loss is made up for by the ability to
digitally increase display contrast, reduce noise,
and enhance the edges of digital images.

    There are several other advantages to digital
photospot images. Mechanical devices are not
needed for film transport. Film processing is not
required. Images can be viewed immediately. The
linear response of the digital system makes it very
forgiving of under- or overexposure.
                                                     50
Fluoroscopic Equipment Configurations

The basic configurations include radiography/fluoroscopy (R/F) tables with
either an under-table or over-table x-ray tube and fixed C-arm, mobile C-arm,
and mini C-arm

R/F Units with Under-Table X-ray Tube:

most common fluoroscopic equipment configuration

The x-ray tube and collimator are mounted below the tabletop with the image
intensifier tower mounted above the table on a carriage that can be panned
over the patient.
In addition to the standard fluoroscopic imaging chain, R/F systems include
an overhead x-ray tube that can be used for regular radiography with a Bucky
incorporated into the table.

Other common features include a tilting table and image recording devices.
                                                                       51
Under-table x-ray tube R/F system. Photograph shows an example of an
R/F table that includes a spot film device and side-mounted video camera.




                                                                       52
R/F Units with Over-Table X-ray Tube
    x-ray tube mounted over the table with the
    image intensifier below.
   this configuration results in increased patient
    access, which is helpful for interventional
    procedures.
    Radiography can be performed with the same x-
    ray tube and a Bucky incorporated into the table.
    The x-ray tube can be angled to acquire
    angulated projections or tomograms.

                                                   53
Over-table x-ray tube R/F system. Photograph shows a sample system
that can be controlled from within the procedure room with the pedestal
control panel (left) or from outside the room from the remote desk controls
(right).




                                                                         54
SUMMARY:
   Early fluroscopy was accomplished by
    radiologists looking directly at a fluoroscopic
    screen.
   The image on the screen was only 0.0001 as bright
    as the image of a routinely viewed radiograph, so
    dark adaptation of eyes was required.
   In 1950s the image intensifier alleviated this
    situation by producing an image bright enough to
    be viewed with cone vision.
   The input phosphor of modern image intensifier
    is CsI; the output phosphor is zinc cadmium  55
   Brightness gain is the product of minification gain and
    flux gain.
   Imaging characteristics important in the evaluation of
    image intensifier fluoroscopy include contrast, lag and
    distortion.
   Large field of view image intensifier tubes are available
    to fill special needs, such as digital and spot film
    angiography.
   Most image intensifiers allow dual field or triple field
    imaging.

                                                         56
   Output phospher image is processed by television
    camera tube (vidicon, plumbicon, CCD)
   The image is displayed on TV monitor
   Standard x-ray closed-circuit television uses
    525x525 format with 5-MHZ band pass
   Vertical resolution limited by scan line format,
    horizontal resolution is a function of band pass
   Light image is recorded by photo spot camera or
    cine camera


                                                 57
58

Weitere ähnliche Inhalte

Was ist angesagt?

Sensitometry or characteristic curve
Sensitometry or characteristic curve   Sensitometry or characteristic curve
Sensitometry or characteristic curve
Upakar Paudel
 
Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortion
parthajyotidas11
 
Radiographic exposure and image quality
Radiographic exposure and image qualityRadiographic exposure and image quality
Radiographic exposure and image quality
Rad Tech
 

Was ist angesagt? (20)

Sensitometry or characteristic curve
Sensitometry or characteristic curve   Sensitometry or characteristic curve
Sensitometry or characteristic curve
 
MACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptxMACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptx
 
Flat Panel Digital Fluoroscopy by Akash Das
Flat Panel Digital Fluoroscopy by Akash Das Flat Panel Digital Fluoroscopy by Akash Das
Flat Panel Digital Fluoroscopy by Akash Das
 
Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortion
 
Production of x rays & generators
Production of x rays & generatorsProduction of x rays & generators
Production of x rays & generators
 
Fluoroscopy-Rohit.pptx
Fluoroscopy-Rohit.pptxFluoroscopy-Rohit.pptx
Fluoroscopy-Rohit.pptx
 
Image quality
Image qualityImage quality
Image quality
 
The control of scattered radiation
The control of scattered radiationThe control of scattered radiation
The control of scattered radiation
 
Digital Fluoroscopy Imaging System
Digital Fluoroscopy Imaging SystemDigital Fluoroscopy Imaging System
Digital Fluoroscopy Imaging System
 
Radiographic exposure and image quality
Radiographic exposure and image qualityRadiographic exposure and image quality
Radiographic exposure and image quality
 
Qc in xray
Qc in xrayQc in xray
Qc in xray
 
Filters and beam restrictors
Filters  and beam restrictorsFilters  and beam restrictors
Filters and beam restrictors
 
Ct tube and detectors
Ct tube and detectorsCt tube and detectors
Ct tube and detectors
 
CT ARTIFACTS.pptx
CT ARTIFACTS.pptxCT ARTIFACTS.pptx
CT ARTIFACTS.pptx
 
Mobile & portable radiography
Mobile & portable radiographyMobile & portable radiography
Mobile & portable radiography
 
Rectifiers
RectifiersRectifiers
Rectifiers
 
Computed radiography
Computed radiographyComputed radiography
Computed radiography
 
Flouroscopy ppt
Flouroscopy pptFlouroscopy ppt
Flouroscopy ppt
 
CT Image reconstruction
CT Image reconstructionCT Image reconstruction
CT Image reconstruction
 
CT detectors
CT detectorsCT detectors
CT detectors
 

Ähnlich wie Flouroscopic imging

flouroscopicimging-120617043323-phpapp01.pdf
flouroscopicimging-120617043323-phpapp01.pdfflouroscopicimging-120617043323-phpapp01.pdf
flouroscopicimging-120617043323-phpapp01.pdf
VishnuDutt40
 
Floroscopynewmicrosoftofficepowerpoint97 2003presentation2-131218145222-phpapp01
Floroscopynewmicrosoftofficepowerpoint97 2003presentation2-131218145222-phpapp01Floroscopynewmicrosoftofficepowerpoint97 2003presentation2-131218145222-phpapp01
Floroscopynewmicrosoftofficepowerpoint97 2003presentation2-131218145222-phpapp01
Ruta Angel
 
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
mr_koky
 

Ähnlich wie Flouroscopic imging (20)

flouroscopicimging-120617043323-phpapp01.pdf
flouroscopicimging-120617043323-phpapp01.pdfflouroscopicimging-120617043323-phpapp01.pdf
flouroscopicimging-120617043323-phpapp01.pdf
 
Flouroscopic imaging
Flouroscopic imagingFlouroscopic imaging
Flouroscopic imaging
 
Mi ppt fi print this
Mi ppt fi print thisMi ppt fi print this
Mi ppt fi print this
 
fluoroscopy-3.pptx
fluoroscopy-3.pptxfluoroscopy-3.pptx
fluoroscopy-3.pptx
 
Fluroscopic imaging ppt
Fluroscopic imaging pptFluroscopic imaging ppt
Fluroscopic imaging ppt
 
Fluoroscopic Imaging.pptx
Fluoroscopic Imaging.pptxFluoroscopic Imaging.pptx
Fluoroscopic Imaging.pptx
 
Floroscopynewmicrosoftofficepowerpoint97 2003presentation2-131218145222-phpapp01
Floroscopynewmicrosoftofficepowerpoint97 2003presentation2-131218145222-phpapp01Floroscopynewmicrosoftofficepowerpoint97 2003presentation2-131218145222-phpapp01
Floroscopynewmicrosoftofficepowerpoint97 2003presentation2-131218145222-phpapp01
 
Fluroscopy.pptx
Fluroscopy.pptxFluroscopy.pptx
Fluroscopy.pptx
 
Fluroscopy
Fluroscopy Fluroscopy
Fluroscopy
 
flouroscopy.pptx
flouroscopy.pptxflouroscopy.pptx
flouroscopy.pptx
 
Flouroscopy JP.pptx
Flouroscopy JP.pptxFlouroscopy JP.pptx
Flouroscopy JP.pptx
 
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
 
Fluoroscopy presentation1
Fluoroscopy presentation1Fluoroscopy presentation1
Fluoroscopy presentation1
 
Different types of microscopes
Different types of microscopesDifferent types of microscopes
Different types of microscopes
 
Lectures Notes Types of microscopes.pptx
Lectures Notes Types of microscopes.pptxLectures Notes Types of microscopes.pptx
Lectures Notes Types of microscopes.pptx
 
RDT-112-PRELIM-LESSON-1.docx
RDT-112-PRELIM-LESSON-1.docxRDT-112-PRELIM-LESSON-1.docx
RDT-112-PRELIM-LESSON-1.docx
 
fluoroscopy imaging
 fluoroscopy imaging  fluoroscopy imaging
fluoroscopy imaging
 
Tv transmitters & receiver
Tv transmitters & receiverTv transmitters & receiver
Tv transmitters & receiver
 
Night vision glasses
Night vision glassesNight vision glasses
Night vision glasses
 
2. Microscopy (Biochemistry)
2. Microscopy (Biochemistry)2. Microscopy (Biochemistry)
2. Microscopy (Biochemistry)
 

Kürzlich hochgeladen

EIS-Webinar-Prompt-Knowledge-Eng-2024-04-08.pptx
EIS-Webinar-Prompt-Knowledge-Eng-2024-04-08.pptxEIS-Webinar-Prompt-Knowledge-Eng-2024-04-08.pptx
EIS-Webinar-Prompt-Knowledge-Eng-2024-04-08.pptx
Earley Information Science
 
Artificial Intelligence: Facts and Myths
Artificial Intelligence: Facts and MythsArtificial Intelligence: Facts and Myths
Artificial Intelligence: Facts and Myths
Joaquim Jorge
 
CNv6 Instructor Chapter 6 Quality of Service
CNv6 Instructor Chapter 6 Quality of ServiceCNv6 Instructor Chapter 6 Quality of Service
CNv6 Instructor Chapter 6 Quality of Service
giselly40
 

Kürzlich hochgeladen (20)

How to Troubleshoot Apps for the Modern Connected Worker
How to Troubleshoot Apps for the Modern Connected WorkerHow to Troubleshoot Apps for the Modern Connected Worker
How to Troubleshoot Apps for the Modern Connected Worker
 
EIS-Webinar-Prompt-Knowledge-Eng-2024-04-08.pptx
EIS-Webinar-Prompt-Knowledge-Eng-2024-04-08.pptxEIS-Webinar-Prompt-Knowledge-Eng-2024-04-08.pptx
EIS-Webinar-Prompt-Knowledge-Eng-2024-04-08.pptx
 
Presentation on how to chat with PDF using ChatGPT code interpreter
Presentation on how to chat with PDF using ChatGPT code interpreterPresentation on how to chat with PDF using ChatGPT code interpreter
Presentation on how to chat with PDF using ChatGPT code interpreter
 
What Are The Drone Anti-jamming Systems Technology?
What Are The Drone Anti-jamming Systems Technology?What Are The Drone Anti-jamming Systems Technology?
What Are The Drone Anti-jamming Systems Technology?
 
Artificial Intelligence: Facts and Myths
Artificial Intelligence: Facts and MythsArtificial Intelligence: Facts and Myths
Artificial Intelligence: Facts and Myths
 
GenAI Risks & Security Meetup 01052024.pdf
GenAI Risks & Security Meetup 01052024.pdfGenAI Risks & Security Meetup 01052024.pdf
GenAI Risks & Security Meetup 01052024.pdf
 
TrustArc Webinar - Stay Ahead of US State Data Privacy Law Developments
TrustArc Webinar - Stay Ahead of US State Data Privacy Law DevelopmentsTrustArc Webinar - Stay Ahead of US State Data Privacy Law Developments
TrustArc Webinar - Stay Ahead of US State Data Privacy Law Developments
 
[2024]Digital Global Overview Report 2024 Meltwater.pdf
[2024]Digital Global Overview Report 2024 Meltwater.pdf[2024]Digital Global Overview Report 2024 Meltwater.pdf
[2024]Digital Global Overview Report 2024 Meltwater.pdf
 
08448380779 Call Girls In Friends Colony Women Seeking Men
08448380779 Call Girls In Friends Colony Women Seeking Men08448380779 Call Girls In Friends Colony Women Seeking Men
08448380779 Call Girls In Friends Colony Women Seeking Men
 
CNv6 Instructor Chapter 6 Quality of Service
CNv6 Instructor Chapter 6 Quality of ServiceCNv6 Instructor Chapter 6 Quality of Service
CNv6 Instructor Chapter 6 Quality of Service
 
Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...
Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...
Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...
 
GenCyber Cyber Security Day Presentation
GenCyber Cyber Security Day PresentationGenCyber Cyber Security Day Presentation
GenCyber Cyber Security Day Presentation
 
Powerful Google developer tools for immediate impact! (2023-24 C)
Powerful Google developer tools for immediate impact! (2023-24 C)Powerful Google developer tools for immediate impact! (2023-24 C)
Powerful Google developer tools for immediate impact! (2023-24 C)
 
🐬 The future of MySQL is Postgres 🐘
🐬  The future of MySQL is Postgres   🐘🐬  The future of MySQL is Postgres   🐘
🐬 The future of MySQL is Postgres 🐘
 
Axa Assurance Maroc - Insurer Innovation Award 2024
Axa Assurance Maroc - Insurer Innovation Award 2024Axa Assurance Maroc - Insurer Innovation Award 2024
Axa Assurance Maroc - Insurer Innovation Award 2024
 
Data Cloud, More than a CDP by Matt Robison
Data Cloud, More than a CDP by Matt RobisonData Cloud, More than a CDP by Matt Robison
Data Cloud, More than a CDP by Matt Robison
 
2024: Domino Containers - The Next Step. News from the Domino Container commu...
2024: Domino Containers - The Next Step. News from the Domino Container commu...2024: Domino Containers - The Next Step. News from the Domino Container commu...
2024: Domino Containers - The Next Step. News from the Domino Container commu...
 
Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...
Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...
Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...
 
From Event to Action: Accelerate Your Decision Making with Real-Time Automation
From Event to Action: Accelerate Your Decision Making with Real-Time AutomationFrom Event to Action: Accelerate Your Decision Making with Real-Time Automation
From Event to Action: Accelerate Your Decision Making with Real-Time Automation
 
Raspberry Pi 5: Challenges and Solutions in Bringing up an OpenGL/Vulkan Driv...
Raspberry Pi 5: Challenges and Solutions in Bringing up an OpenGL/Vulkan Driv...Raspberry Pi 5: Challenges and Solutions in Bringing up an OpenGL/Vulkan Driv...
Raspberry Pi 5: Challenges and Solutions in Bringing up an OpenGL/Vulkan Driv...
 

Flouroscopic imging

  • 2. Fluoroscopy Purpose To visualize, in real time:  organ motion  ingested or injected contrast agents  insert stents  cathetarize small blood vessels REAL TIME IMAGING 2
  • 3. X-rays were discovered because of there ability to cause fluorescence . Thefirst x-ray image of human part was observed fluoroscopically-Dr. Glasser. 3
  • 4. THE FLUOROSCOPE  First generation fluoroscopes consisted of an x-ray tube, an x-ray table and a fluoroscopic screen.  The fluorescent material used in screen was copper activated zinc cadmium sulfide that emitted light in yellow-green spectrum.  A sheet of lead glass covered the screen, so that radiologist could stare directly into the screen with out having the x-ray beam strike his eyes.  Screen fluoroscence was very faint so, the examination was carried out in a dark room by the radiologist who had to adapt his eyes by wearing red goggles for 20-30 mins prior to the examination  technique is now obsolete & gone. 4
  • 6. Photograph shows an early (1933) fluoroscopic system in use before the development of image intensification. An actual fluoroscopic examination with this device would have occurred in a darkened room. 6
  • 7. IMAGE INTENSIFIER DESIGN  Image intensifier was discovered in 1950s-to produce an image bright enough to allow cone vision without giving the pt an excess radiation exposure.  The components of an x-ray image intensifier  The tube itself is an evacuated glass envelope ,a vacuum tube containing- 1.input phosphor and photocathode . 2.electrostatic focusing lens. 3.accelerating anode. 4.out put phosphor. 7
  • 8. After an x-ray beam passes the pt it enters the image intensifier tube the input fluorescent screen absorbs x-ray photons and converts their energy into light photons.  The light photons strike the photo cathode, causing it to emit photoelectrons  these electrons are immediately drawn away from the photocathode by the high potential difference betn it &the accelerating anode.  As the electrons flow from the cathode towards the anode, they are focused by an electrostatic lens which guides them to the output fluorescent screen without distorting their geometric configuration. 8
  • 9. The electrons strike the output screen, which emits the light photons that carry the fluoroscopic images to the eye of the observer.  In intensifier tube, the image is first carried by the x-ray photons, then by the light photons, next by the electrons &finally by the light photons. 9
  • 11. 11
  • 12. INPUT PHOSPHOR & PHOTO CATHODE:  The input fluorescent screen in image intensifiers is cesium iodide (CsI). (older intensifier- silver activated zinc cadmium sulfide).  CsI is deposited on a thin aluminum substrate by a process called “vapor deposition”.  an interesting & useful characteristic of CsI is that during the deposition process the crystals of CsI grow in tiny needles perpendicular to the substrate.  There by reducing scattering. 12
  • 13. INPUT PHOSPHOR & PHOTO CATHODE:  Image quality is dramatically better with CSI input screen than it was with zinc cadmium sulfide screens.  Three physical characteristics of CsI make it superior. 1. vertical orientation of the crystals. 2. A greater packing density & 3. A more favorable effective atomic number. Phosphor thickness have been reduced comparably from app. 0.3mm with Zn-Cd su to 0.1mm with CsI. The principal advantage of thinner phosphor layer combined with needle shaped crystals is improved resolution. 13
  • 14. PHOTO CATHODE:  The photo cathode is a photoemissive metal (commonly a combination of antimony & cesium compounds).  When the light from the fluorescent screen strikes the photo cathode, photo electrons are emitted in numbers proportional to the brightness of the screen.  The photo cathode is applied directly to the CsI input phosphor.  The photoelectrons thus produced has to be moved to the other end of the image intensifier. This can be done using an electrostatic focusing lens and an accelerating anode. 14
  • 15. ELECTROSTATIC FOCUSING CUP:  The lens is made up of a series positively charged electrodes that are usually plated on to the inside surface of the glass envelope.  These electrodes focus the electron beam as it flows from the photocathode toward the output phosphor.  Electron focusing inverts & reverses the image which is called “point inversion” because all the electrons pass through a common focal point on their way to output phosphor.  The input phosphor is curved to ensure that electrons emitted at the peripheral regions of the photocathode travel the same distance as those emitted from the central region.  The image on the output phosphor is reduced in size ,which is one of the principle reasons why it is brighter. 15
  • 16. ACCELERATING ANODE :  The anode is located in the neck of the image tube.  Its function is to accelerate electrons emitted from the photocathode towards the output screen. the anode has a +ve potential of 25 to 35 kv relative to the photocathode, so it accelerates electrons to a tremendous velocity. OUTPUT PHOSPHOR:  The output fluorescent screen of image intensifiers is silver activated zn-cd sulfide, the same used in Ist generation input phosphor.  Crystal size and layer thickness are reduced to maintain resolution in the minified image.  A thin layer of aluminum is plated onto the fluorescent screen prevent light from moving retrograde through the tube & activating the photocathode. 16
  • 17. The glass tube of the image intensifier is abt 2 to 4mm thick &is enclosed in a lead lined metal container protects the operator from stray radiation.  The output phosphor image is viewed either directly through a series of lenses and mirrors or indirectly through closed circuit TV. BRIGHTNESS GAIN:  Two methods are used to evaluate the brightness gain of image intensifiers. the first compares the luminance of an intensifier output screen to that of a Patterson type B2 fluoroscopy screen when both are exposed to same quantity of radiation.  The brightness gain is the ratio of the two illuminations.  Brightness gain=intensifier luminance/Patterson b-2 lumin. 17
  • 18. Another way of evaluation of brightness gain is called as “conversion factor”  Conversion factor =cd/m2 by mR/sec  Output screen luminance is measured in candelas.  Radiation quality & output luminance are explicitly defined, so the method is accurate & reproducible.  The brightness gain tends to deteriorate as the image intensifier ages. ie the pt dose with an older image intensifier tends to be higher than with a new intensifier of the same type.  The brightness gain of an imag inten comes from 2 completely unrelated sources called minification gain &flux gain. 18
  • 19. MINIFICATION GAIN:  The brightness gain from minification is produced by a reduction in image size.  The quantity of gain depends on the relative areas of input &output screens. coz the size of the intensifier is usually indicated by its diameter, so minif gain is expressed as MG=(d1/d0)2.d1=diameter of input screen,d0=diameter of output screen.  Most img inten have an input screen from 5 to9 in.& an output screen of app 1in in diameter.  Theoretically minification can be increased indefinitely as we can see from above formula, but as the minification is increased the image becomes smaller.- disadvantage.  Hence image has to be magnified and viewed which will result in reduce brightness & precipitous drop in resolution. 19
  • 20. FLUX GAIN:  FLUX gain increases the brightness of the fluoroscopic image by a factor of app 50.  The total brightness of an imag intes is product of minification & flux gain: ie Brightness gain =minification gain x flux gain. 20
  • 21. MULTIPLE-FIELD IMAGE INTENSIFIERS  Dual field or triple field imsg intes attempt to resolve the conflicts btn image size & quality.  They can be operated in several modes, including the 4.5in, a 6in, or a 9in mode. the 9in mode is used to view large anatomic areas. When size is unimportant the 4.5in or 6in mode is used coz of better resultant image quality.  Field size is changed by a simple electronic principle: the higher the voltage on the electrostatic focusing lens, the more the electron beam is focused. 21
  • 22. This figure shows this principle applied to a dual field imag intes.  In the 9in mode, the electrostatic focusing voltage is decreased. the electrons focus to a point or cross, close to the output phosphor & the final image is actually smaller than the phosphor.  In 6in mode the electrostatic focusing voltage is increased & the electrons focus farther away from the output phosphor. after the electrons cross they diverge, so the image on the output phosphor is larger than in the 22
  • 23. Optical coupling  Optical system transmits the output of the image intensifier to the light sensitive area of the video camera  The optical distributor include beam-splitting mirror, which directs a portion of the light from the image intensifier output window to an accessory device for image recording and passes the remainder to the video camera.  Two lenses are mounted in tandem  The II and the vidicon are placed at the focal planes of the two lenses 23
  • 24. 24
  • 25. Closed-circuit Television System  used to view the image intensifier output image  Consists of 1)Television camera 2) Camera control unit 3) Monitor  The television system allows for real-time viewing of the fluoroscopic image by several people at once from one monitor or multiple monitors 25
  • 26. Television Camera Tube  Output phosphor is directly coupled to a TV camera tube  Plumbicon  Vidicon  CCD 26
  • 27.  The basic video camera consists of 1) vacuum tube cylinder (approximately 2.5 cm in diameter) surrounded by electromagnetic focusing coils ,2 pairs of electrostatic deflecting coils 2) photoconductive target 3) a scanning electron beam Target assembly: a) Glass plate assembly b) signal plate c) Target 27
  • 29. Target  Functionally most important i n tube  Thin film of photoconductive material, antimony sulfide suspended as globules in mica matrix.  The optical coupling lens focuses the image intensifier output image onto the target, forming a charge image within the photoconductive layer  This latent image is read out by the electron beam, which scans across the target in a series of horizontal raster lines.  As the scanning electron beam moves across the target, a current signal is produced that represents the two-dimensional image as a continuous series of raster lines with varying voltage levels. 29
  • 30. Video signal  When globules absorbs light ,photoelectrons are emitted  The globule becomes positively charged  The electron beam scans the electrical image stored on the target & fills in the holes left by the emitted photoelectrons, discharging the tiny globular capacitors  When the electrons in beam neutralize the positive charge in the globules , the electrons on the signal plate leave the plate via resistor  These moving electrons form a current flowing through a resistor and voltage across the resistor  This voltage ,when collected for each neutralized globule, constitutes the video signal 30
  • 31. 31
  • 32. Television monitor  The video signal produced by the video camera is converted into a visible image by the monitor  Contains picture tube & controls for regulating brightness & contrast  Picture tube contains- Electron gun control grid anode focusing coil, deflecting coil-control the electron beam synchrony with the camera tube  Control grid-receive video signal from Camera control unit ,uses this signal to regulate the no. of electrons in electron beam  Anode –carries higher potential (10,000V) accelerates the electron beam to much higher velocity  The electron Strike the fluorescent screen ,emit large number of light photons. 32
  • 33. 33
  • 34. Image Recording Two modes of recording th flouroscopic image - 1) light image from output phosphor of II recorded on film with a photospot camera or cine camera 2) electrical signal generated by TV camera- includes magnetic tape, magnetic discs & optical discs 34
  • 35.    DIRECT FILM RECORDING SPOT FILM DEVICES:  Fluoroscopic systems designed for gastrointestinal imaging are generally equipped with a spot film device.  The spot film device allows exposure of a conventional screen-film cassette in conjunction with fluoroscopic viewing. This rather familiar system, located in front of the image intensifier, accepts the screen-film cassette and "parks" it out of the way during fluoroscopy  Cassettes may be loaded from the front or rear depending on the design of the system. 35
  • 36.   Standard spot film imaging configuration typical of gastrointestinal fluoroscopy equipment. The screen-film cassette is parked out of the x- ray field until the spot film trigger is pressed, causing both the cassette and the formatting mask to move into position. 36
  • 37. The x-ray field size is also reduced automatically by the collimators at the time of exposure to minimize scattered radiation and patient radiation dose.  The fluoroscopist can override this automatic collimation to further reduce the x-ray field.  Spot film imaging uses essentially the same technology as conventional screen-film radiography. Differences and limitations of spot film imaging compared with general radiography.  One major limitation is the range of film sizes available for spot film imaging. Although some older fluoroscopy equipment is limited to a single size, usually 24 x 24 cm, current equipment allows a range of film sizes to be used, from 20 x 25 cm to 24 x 35 cm.  Spot film devices usually allow more than one image to be obtained on a single film.  Formats typically include one, two, three, four, or six images on a film. 37
  • 38. Moving the spot film device closer to the patient reduces the amount of magnification and decreases the patient radiation dose.  A number of factors affect patient doses in spot film imaging.  The source-to-skin distance is shorter in spot film imaging than in general radiography.  Although the automatic exposure control system fixes the exposure to the screen, the shorter source-to-skin distance increases the inverse square reduction in radiation intensity as it passes through the patient.  This increase tends to make the skin entrance exposure higher.  The field size in spot film imaging is generally smaller than that used in general radiography.  This smaller field size reduces scatter and therefore tends to reduce dose. For the same reason, grids used in fluoroscopy generally have a lower grid ratio and therefore a smaller Bucky factor, which also leads to lower dose.  These effects tend to offset each other to a large extent. 38
  • 39. One of the major shortcomings of conventional spot film devices is the delay involved in moving the cassette into position for exposure.  In gastrointestinal imaging, this delay can be overcome by using photofluorography.  In vascular imaging, more rapid film movement is achieved with automatic film changers. 39
  • 40. AUTOMATIC FILM CHANGERS  The automatic film changers used in vascular imaging are also screen-film systems.  They can be found in several varieties. Some are large, floor-mounted boxes, but systems more commonly used today mount on the image intensifier  The system consists of a supply magazine for holding unexposed film, a receiving magazine, a pair of radiographic screens, and a mechanism for transferring the film.  When an exposure is required, the screens are mechanically separated, the film is pulled into place between them, and they are closed.  After the film is exposed, the screens separate again.  The film is moved to the receiver, and another film is pulled into place for the next exposure.  The number of films and filming rates must be preprogrammed for proper operation. 40
  • 41. Photospot camera  Records the the image output of an II on a film  Film – role film/cut film(10 cm)  Advantage –1)reduction in pt exposure 2)film does not have to be changed b/w exposures 3)exposure times are shorter-motion is less likely problem 4)films can be taken more rapidly 5)possible to record & view image at same time 41
  • 42. Framing with spot film cameras:  Framing –utilisation of available area on film  The output phosphor of II tube is round, shape of film is square  4 framing patterns 1) Exact framing 2) Equal area framing 3) mean diameter framing 4) total area framing  Equal area framing or mean diameter framing is recommended for most clinical situations. 42
  • 43. 43
  • 44. Cineflourography  Process of recording fluoroscopic images on movie (cine) film  Two film sizes- 16 mm, 35 mm  Cine camera-components are lens, iris diaphragm, shutter, aperture, pressure plate, pull down arm & film transport mechanism 44
  • 45. 45
  • 46. TV image recorders : 3) TAPE RECORDER : used for both recoding& playback • as a recorder receives video signal from camera control unit • for playback transmits the signal to one or more several TV monitors Components:1)magnetic tube 2)writing head 3) tape transport system Writing head converts an electrical signal in to magnetic field for recoding & converts magnetic signal to electric signal for replay 46
  • 47. 47
  • 48. 48
  • 49. DIGITAL FLUOROGRAPHY Digital charge-coupled device (CCD) TV cameras are rapidly replacing conventional TV cameras in fluoroscopic systems. An analog, high-resolution (1,023-line) TV camera has a vertical resolution of about 358 line pairs. A high-resolution CCD camera with a 1,024 x 1,024 matrix will provide equivalent vertical resolution. However, the digital camera will have the same vertical and horizontal resolution, whereas the horizontal resolution of the analog camera is defined by its electronic band-pass. 49
  • 50. For a 15-cm-diameter intensifier input, the limiting resolution of the CCD camera would be 358 lp/150 mm or 2.4 lp/mm. This result is about half the resolution of a photospot film. This resolution loss is made up for by the ability to digitally increase display contrast, reduce noise, and enhance the edges of digital images. There are several other advantages to digital photospot images. Mechanical devices are not needed for film transport. Film processing is not required. Images can be viewed immediately. The linear response of the digital system makes it very forgiving of under- or overexposure. 50
  • 51. Fluoroscopic Equipment Configurations The basic configurations include radiography/fluoroscopy (R/F) tables with either an under-table or over-table x-ray tube and fixed C-arm, mobile C-arm, and mini C-arm R/F Units with Under-Table X-ray Tube: most common fluoroscopic equipment configuration The x-ray tube and collimator are mounted below the tabletop with the image intensifier tower mounted above the table on a carriage that can be panned over the patient. In addition to the standard fluoroscopic imaging chain, R/F systems include an overhead x-ray tube that can be used for regular radiography with a Bucky incorporated into the table. Other common features include a tilting table and image recording devices. 51
  • 52. Under-table x-ray tube R/F system. Photograph shows an example of an R/F table that includes a spot film device and side-mounted video camera. 52
  • 53. R/F Units with Over-Table X-ray Tube  x-ray tube mounted over the table with the image intensifier below.  this configuration results in increased patient access, which is helpful for interventional procedures.  Radiography can be performed with the same x- ray tube and a Bucky incorporated into the table.  The x-ray tube can be angled to acquire angulated projections or tomograms. 53
  • 54. Over-table x-ray tube R/F system. Photograph shows a sample system that can be controlled from within the procedure room with the pedestal control panel (left) or from outside the room from the remote desk controls (right). 54
  • 55. SUMMARY:  Early fluroscopy was accomplished by radiologists looking directly at a fluoroscopic screen.  The image on the screen was only 0.0001 as bright as the image of a routinely viewed radiograph, so dark adaptation of eyes was required.  In 1950s the image intensifier alleviated this situation by producing an image bright enough to be viewed with cone vision.  The input phosphor of modern image intensifier is CsI; the output phosphor is zinc cadmium 55
  • 56. Brightness gain is the product of minification gain and flux gain.  Imaging characteristics important in the evaluation of image intensifier fluoroscopy include contrast, lag and distortion.  Large field of view image intensifier tubes are available to fill special needs, such as digital and spot film angiography.  Most image intensifiers allow dual field or triple field imaging. 56
  • 57. Output phospher image is processed by television camera tube (vidicon, plumbicon, CCD)  The image is displayed on TV monitor  Standard x-ray closed-circuit television uses 525x525 format with 5-MHZ band pass  Vertical resolution limited by scan line format, horizontal resolution is a function of band pass  Light image is recorded by photo spot camera or cine camera 57
  • 58. 58