RTM ASSIGNMENT 2: THE WATER EQUIVALENCE OF LIF TLDS
i
TABLE OF CONTENTS
TABLE OF CONTENTS...................................................................................................................................................i
ABSTRACT................................................................................................................................................................... ii
INTRODUCTION..........................................................................................................................................................1
METHODOLOGY .........................................................................................................................................................2
Clinical measurements...........................................................................................................................................2
Equipment..............................................................................................................................................................2
Entrance dose calibration ......................................................................................................................................2
Midline dose determination ..................................................................................................................................3
RESULTS AND DISCUSSION.........................................................................................................................................4
CONCLUSION..............................................................................................................................................................8
References .................................................................................................................................................................9
ii
ABSTRACT
The dosimetry was carried out for radiotherapy patients, and measurements were performed using
LiF and thermoluminescent dosimeters (TLDs). Evaluations were done for water-equivalent
(effective) thicknesses and target dose with transmission data. Considerations were made for the
accuracy of the parameter for the ratio of measured to expected value for each quantity. The
entrance dose was estimated as 1.01 ± 0.07. The mean ratio of effective to contour depth was 1.00
± 0.13, showing a wide distribution reflecting the influence of contour inaccuracies. The mean
ratio of the measured contour dose prescription was 1.00 ± 0.07. The difference in depths that is
patient and effective depth is a reflection of target dose discrepancies. Graphical simulations were
done using Monte-Carlo Simulations and presented.
1
INTRODUCTION
Probability relationship approximates the control of local tumor and normal tissue injury (Leunens,
et al., 1990; Leunens, et al., 1990). Thus procedural therapy for weak point determination is key
to treatment. The most and effective method for this probabilistic determination weak point is in
vivo dosimetry. The commonly used method is thermoluminescent dosimeter (TLD) (Tung, et al.,
2004). This application si popular since it permits the use of small dependencies on dose rate,
temperature, energy and high spatial sensitivity, and resolution (Moscovitch, 1999). TLDs are
popular also since they are unattached to the bias voltage and electrometer though they are passive
and their answers are not prompt.
In vivo dosimetry are in various levels: intracavity dose; entrance dose and exit dose. The coherent
objective of dosimetry is the determination of value of doses. Therefore, the study is first-quality
assurance program for determination of target dose based on water-equivalent (effective) diameter.
This was done for comparison with the actual contour diameter for investigation of the effect of
the corresponding variation of the delivered dose to the target.
Photons on the other hand interacts with atoms in many ways (Corvan, et al., 2014): the multi-
MeV regime, where photoionization occurs, electron-positron pair production and compton
scattering. There are also a number of ways of detection of MeV gamma rays which in most cases
exploits quantum electromagnetic cascade initiated by photon solidification and compton
scattering.
2
METHODOLOGY
Clinical measurements
The study was forced on the Source Axis Distance (SAD) since the study is focused on the midline.
Measurements were taken for dosage at the entrance and exit doses. TLDs of 50 𝑚𝑚 build up was
used to cover material for placement of the surface perpendicular to the beam central axis at each
measuring point. Shadow effect was avoided by either the entrance detector or the exit detector by
shifting the beam slightly out of his axis. Accuracy of the treatment procedure is the ratio of the
measured to the expected values.
Equipment
Since TLD is water equivalent, the analysis of TLDs were done with the Harshaw-3500 manual
reader. With an absorbed dose of 1 Gy at the center of a 10 × 10 cm2
field size and 95 cm source-
skin distance (SSD), the residual signal was noted to be < 1%. Thus batch uniformity = 3.5%
which is equivalent to 1 SD.
Entrance dose calibration
TLDs were positioned on phantom surface for measurement of entrance dose (Den) measurement
and calibration were done for the determination of dose inside the phantom at a depth of the
maximum dose (𝑑 𝑚).
Sensitivity loss factor was estimated by:
𝑆𝑓 =
(
𝑅̅ 𝑟𝑒𝑓
𝐼𝐶 𝑎𝑏𝑠𝑜𝑟𝑏𝑒𝑑 𝑑𝑜𝑠𝑒)
1
(
𝑅̅ 𝑟𝑒𝑓
𝐼𝐶 𝑎𝑏𝑠𝑜𝑟𝑏𝑒𝑑 𝑑𝑜𝑠𝑒)
𝑓
3
Midline dose determination
Determination of Midline dose was based on measurements taken during transmission. From the
theory, 𝑇𝑒𝑥 =
𝐷 𝑒𝑥
𝐷 𝑒𝑛
, while 𝑇 𝑚𝑖𝑑 =
𝐷 𝑚𝑖𝑑
𝐷 𝑒𝑛
. However, 𝑇𝑒𝑥 and 𝑇 𝑚𝑖𝑑 could be calculated from
𝑇𝑒𝑥(𝐴′
, 𝑑 𝑧−𝑒𝑛) =
𝑇𝑀𝑅(𝐴′
, 𝑑 𝑧−𝑑 𝑒𝑛
) (
𝑆𝑆𝐷 + 𝑑 𝑚
𝑆𝑆𝐷 + 𝑍 − 𝑑 𝑚
)
2
𝐵𝑆𝐹(𝐴′)
𝐵𝑆𝐹(𝐴0)
𝑇 𝑚𝑖𝑑 (𝐴, 𝑑 𝑍
2
) =
𝑇𝑀𝑅 (𝐴′
, 𝑑 𝑍
2
) (
𝑆𝑆𝐷 + 𝑑 𝑚
𝑆𝑆𝐷 +
𝑍
𝑑 𝑚
)
2
𝐵𝑆𝐹(𝐴)
𝐵𝑆𝐹(𝐴0)
where 𝐴0, 𝐴, and 𝐴′
are the field sizes at the entrance, midline ((
𝑆𝑆𝐷+
𝑍
2
𝑆𝑆𝐷+𝑑 𝑚
) 𝐴0) and exit levels
((
𝑆𝑆𝐷 + 𝑧 − 𝑑 𝑚
𝑆𝑆𝐷 + 𝑑 𝑚
) 𝐴0)
and 𝑍 is the water-equivalent (effective) depth, 𝑑 𝑚 maximum dose depth.
4
RESULTS AND DISCUSSION
Correction factors
Correction factors have to be determined when irradiation geometry differs from the reference
geometry used in the calibration procedure. These variations in TLDs’ response were examined in
different SSDs and field sizes. Correction factor (CF) is defined as
𝐶𝐹 =
𝑅̅ 𝑟𝑒𝑓
𝐼𝐶 (𝑎𝑏𝑠𝑜𝑟𝑏𝑒𝑑 𝑑𝑜𝑠𝑒) 𝑟𝑒𝑓𝑒𝑟𝑒𝑛𝑐𝑒 𝑔𝑒𝑜𝑔𝑚𝑒𝑡𝑟𝑦
𝑅̅ 𝑟𝑒𝑓
𝐼𝐶 (𝑎𝑏𝑠𝑜𝑟𝑏𝑒𝑑 𝑑𝑜𝑠𝑒) 𝑔𝑒𝑜𝑔𝑚𝑒𝑡𝑟𝑦 𝑜𝑓 𝑖𝑛𝑡𝑒𝑟𝑒𝑠𝑡
The attention of the experiment was focused on the LiF, a material of the lowest value of Z. Based
on the database, the Compton scattering was noted as the overriding interaction between
mechanisms for a multi-MeV gamma-ray beam up to energies of 20 MeV as shown in the figure
below.
Figure 1: Energy loss for a multi-MeV gamma ray beam as it interacts with LiF.
5
The lines in the square corresponds to nuclear filed while the blue is for diamond for Compton
scattering. Compton scattering is dominant up to around 1.5 MeV, after which pair production
dominates.
To estimate the effect of LiF on the spectrum of the electrons, simulation was done using the
Monte-Carlo code FLUKA and the simulation present in the figure below.
Figure 2: Spectra of the electrons generated.
However, the Gaussian gave appropriate estimation of the relationship between the peak values of
the conversion factor and is show in the Figure 3 below. Furthermore, the peak of the conversion
factors in the electron spectra plotted as a function of the initial photon energy. The peaks rise in
magnitude up to 5 MeV where the peaks remain endless up to 10 MeV after which they start to
decrease again. This shown in Figure 4.
6
Figure 3: The peak of the conversion factors in the electron spectra plotted as a function of the
initial photon energy
7
Figure 4: Fluka simulations.
8
CONCLUSION
Lif and TLD was used to carry a compressive series evaluation of in vivo dosimetry measurements.
Determination of water-equivalent depth and midline dose showed a close association between the
midline dose imprecision and discrepancy of water-equivalent and contour depths.
Furthermore, the modification of compact gamma-ray spectrometer for in vivo dosimetry for
treatment presented showed that for energy windows around 20 MeV, using Monte-Carlo
simulations, the performance of the system can be checked. The results of this experiments were
in line with other conducted by Aad, et al., (2014).
9
References
Aad, G. a. A. T. a. A. B. a. A. J. a. K. S. A. a. A. A. et al., 2014. Measurement of the inclusive
isolated prompt photons cross section in p p collisions at s= 7 TeV with the ATLAS
detector using 4.6 fb- 1. hysical Review D, 89(5), p. 052004.
Corvan, D. J., Sarri, G. & Zepf, M., 2014. Design of a compact spectrometer for high-flux MeV
gamma-ray beams. Review of Scientific Instruments, 85(6), p. 065119.
Leunens, G., Van Dam, J., Dutreix, A. & Van der Schueren, E., 1990. Quality assurance in
radiotherapy by in vivo dosimetry. 1. Entrance dose measurements, a reliable procedure.
Radiotherapy and Oncology, 17(2), pp. 141-151.
Leunens, G., Van Dam, J., Dutreix, A. & Van der Schueren, E., 1990. Quality assurance in
radiotherapy by in vivo dosimetry. 2. Determination of the target absorbed dose.
Radiotherapy and Oncology, 19(1), pp. 73-87.
Moscovitch, M., 1999. Personnel dosimetry using LiF: Mg, Cu, P. Radiation protection
dosimetry, 85(1-4), pp. 49-56.
Tung, C. et al., 2004. In vivo dosimetry for external photon treatments of head and neck cancers
by diodes and TLDS. Radiation protection dosimetry, 111(1), pp. 45-50.

Weitere ähnliche Inhalte

Was ist angesagt?

DLS and Fluroscence spectroscopy
DLS and Fluroscence spectroscopyDLS and Fluroscence spectroscopy
DLS and Fluroscence spectroscopySMALLYSINHA
 
Dynamic light scattering
Dynamic light scatteringDynamic light scattering
Dynamic light scatteringMannuMaken92
 
Light Scattering: Fundamentals (Old Logo)
Light Scattering: Fundamentals (Old Logo)Light Scattering: Fundamentals (Old Logo)
Light Scattering: Fundamentals (Old Logo)LS Instruments
 
Dynamic Light Scattering Anjali Devi J S
Dynamic Light Scattering Anjali Devi J SDynamic Light Scattering Anjali Devi J S
Dynamic Light Scattering Anjali Devi J SAnjali Devi J S
 
Light Scattering Fundamentals
Light Scattering FundamentalsLight Scattering Fundamentals
Light Scattering FundamentalsLS Instruments
 
Dynamic Light Scattering
Dynamic Light Scattering Dynamic Light Scattering
Dynamic Light Scattering Pooja Mehta
 
Trojan Scaling BWTS Systems Employing Filtration and UV
Trojan Scaling BWTS Systems Employing Filtration and UVTrojan Scaling BWTS Systems Employing Filtration and UV
Trojan Scaling BWTS Systems Employing Filtration and UVJim Cosman
 
Dls presentation edited
Dls presentation editedDls presentation edited
Dls presentation editedPRIYA KUMARI
 
Nanotech and biotech
Nanotech and biotechNanotech and biotech
Nanotech and biotechTorrey DeLuca
 
Radiation physics 2
Radiation physics 2Radiation physics 2
Radiation physics 2Rad Tech
 
Final HPC Poster Compact
Final HPC Poster CompactFinal HPC Poster Compact
Final HPC Poster CompactAngelica Kiser
 
STUDY OF NANOSECOND ND: YAG 213 nm LASER ABLATION OF EXTRA HEAVY CRUDE OIL B...
STUDY OF NANOSECOND ND: YAG 213 nm LASER ABLATION OF EXTRA  HEAVY CRUDE OIL B...STUDY OF NANOSECOND ND: YAG 213 nm LASER ABLATION OF EXTRA  HEAVY CRUDE OIL B...
STUDY OF NANOSECOND ND: YAG 213 nm LASER ABLATION OF EXTRA HEAVY CRUDE OIL B...IJSIT Editor
 
Understanding climate model evaluation and validation
Understanding climate model evaluation and validationUnderstanding climate model evaluation and validation
Understanding climate model evaluation and validationPuneet Sharma
 
Engineering Fluid Mechanics 10th Edition Elger Solutions Manual
Engineering Fluid Mechanics 10th Edition Elger Solutions ManualEngineering Fluid Mechanics 10th Edition Elger Solutions Manual
Engineering Fluid Mechanics 10th Edition Elger Solutions ManualNortoner
 

Was ist angesagt? (20)

dynamic light scattering
dynamic light scatteringdynamic light scattering
dynamic light scattering
 
DLS and Fluroscence spectroscopy
DLS and Fluroscence spectroscopyDLS and Fluroscence spectroscopy
DLS and Fluroscence spectroscopy
 
Dynamic light scattering
Dynamic light scatteringDynamic light scattering
Dynamic light scattering
 
Infrared spectroscopy2
Infrared spectroscopy2Infrared spectroscopy2
Infrared spectroscopy2
 
Light Scattering: Fundamentals (Old Logo)
Light Scattering: Fundamentals (Old Logo)Light Scattering: Fundamentals (Old Logo)
Light Scattering: Fundamentals (Old Logo)
 
Dynamic Light Scattering Anjali Devi J S
Dynamic Light Scattering Anjali Devi J SDynamic Light Scattering Anjali Devi J S
Dynamic Light Scattering Anjali Devi J S
 
DLS
DLSDLS
DLS
 
Light Scattering Fundamentals
Light Scattering FundamentalsLight Scattering Fundamentals
Light Scattering Fundamentals
 
Dynamic Light Scattering
Dynamic Light Scattering Dynamic Light Scattering
Dynamic Light Scattering
 
Trojan Scaling BWTS Systems Employing Filtration and UV
Trojan Scaling BWTS Systems Employing Filtration and UVTrojan Scaling BWTS Systems Employing Filtration and UV
Trojan Scaling BWTS Systems Employing Filtration and UV
 
Dls presentation edited
Dls presentation editedDls presentation edited
Dls presentation edited
 
Nanotech and biotech
Nanotech and biotechNanotech and biotech
Nanotech and biotech
 
P367
P367P367
P367
 
Radiation physics 2
Radiation physics 2Radiation physics 2
Radiation physics 2
 
Final HPC Poster Compact
Final HPC Poster CompactFinal HPC Poster Compact
Final HPC Poster Compact
 
Photoacoustics
PhotoacousticsPhotoacoustics
Photoacoustics
 
03 sp2013
03 sp201303 sp2013
03 sp2013
 
STUDY OF NANOSECOND ND: YAG 213 nm LASER ABLATION OF EXTRA HEAVY CRUDE OIL B...
STUDY OF NANOSECOND ND: YAG 213 nm LASER ABLATION OF EXTRA  HEAVY CRUDE OIL B...STUDY OF NANOSECOND ND: YAG 213 nm LASER ABLATION OF EXTRA  HEAVY CRUDE OIL B...
STUDY OF NANOSECOND ND: YAG 213 nm LASER ABLATION OF EXTRA HEAVY CRUDE OIL B...
 
Understanding climate model evaluation and validation
Understanding climate model evaluation and validationUnderstanding climate model evaluation and validation
Understanding climate model evaluation and validation
 
Engineering Fluid Mechanics 10th Edition Elger Solutions Manual
Engineering Fluid Mechanics 10th Edition Elger Solutions ManualEngineering Fluid Mechanics 10th Edition Elger Solutions Manual
Engineering Fluid Mechanics 10th Edition Elger Solutions Manual
 

Ähnlich wie Rtm assignment 2

Future ground arrays for ultrahigh-energy cosmic rays: recent updates and per...
Future ground arrays for ultrahigh-energy cosmic rays: recent updates and per...Future ground arrays for ultrahigh-energy cosmic rays: recent updates and per...
Future ground arrays for ultrahigh-energy cosmic rays: recent updates and per...Toshihiro FUJII
 
Results from telescope_array_experiment
Results from telescope_array_experimentResults from telescope_array_experiment
Results from telescope_array_experimentSérgio Sacani
 
Variation of dose distribution with depth and incident energy using EGSnrc Mo...
Variation of dose distribution with depth and incident energy using EGSnrc Mo...Variation of dose distribution with depth and incident energy using EGSnrc Mo...
Variation of dose distribution with depth and incident energy using EGSnrc Mo...iosrjce
 
MODELING THE CHLOROPHYLL-A FROM SEA SURFACE REFLECTANCE IN WEST AFRICA BY DEE...
MODELING THE CHLOROPHYLL-A FROM SEA SURFACE REFLECTANCE IN WEST AFRICA BY DEE...MODELING THE CHLOROPHYLL-A FROM SEA SURFACE REFLECTANCE IN WEST AFRICA BY DEE...
MODELING THE CHLOROPHYLL-A FROM SEA SURFACE REFLECTANCE IN WEST AFRICA BY DEE...gerogepatton
 
MODELING THE CHLOROPHYLL-A FROM SEA SURFACE REFLECTANCE IN WEST AFRICA BY DEE...
MODELING THE CHLOROPHYLL-A FROM SEA SURFACE REFLECTANCE IN WEST AFRICA BY DEE...MODELING THE CHLOROPHYLL-A FROM SEA SURFACE REFLECTANCE IN WEST AFRICA BY DEE...
MODELING THE CHLOROPHYLL-A FROM SEA SURFACE REFLECTANCE IN WEST AFRICA BY DEE...ijaia
 
Physical society presentation ashik.pptx
Physical society presentation ashik.pptxPhysical society presentation ashik.pptx
Physical society presentation ashik.pptxAshikurRahman204838
 
Direct detection of a break in the teraelectronvolt cosmic-ray spectrum of el...
Direct detection of a break in the teraelectronvolt cosmic-ray spectrum of el...Direct detection of a break in the teraelectronvolt cosmic-ray spectrum of el...
Direct detection of a break in the teraelectronvolt cosmic-ray spectrum of el...Sérgio Sacani
 
Laser Physics Department -ALNeelain University 5 th
Laser Physics Department -ALNeelain University 5 thLaser Physics Department -ALNeelain University 5 th
Laser Physics Department -ALNeelain University 5 thGazy Khatmi
 
Modeling the Chlorophyll-a from Sea Surface Reflectance in West Africa by Dee...
Modeling the Chlorophyll-a from Sea Surface Reflectance in West Africa by Dee...Modeling the Chlorophyll-a from Sea Surface Reflectance in West Africa by Dee...
Modeling the Chlorophyll-a from Sea Surface Reflectance in West Africa by Dee...gerogepatton
 
Laser Pulsing in Linear Compton Scattering
Laser Pulsing in Linear Compton ScatteringLaser Pulsing in Linear Compton Scattering
Laser Pulsing in Linear Compton ScatteringTodd Hodges
 
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEXTISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEXVictor Ekpo
 
New folderelec425_2016_hw5.pdfMar 25, 2016 ELEC 425 S.docx
New folderelec425_2016_hw5.pdfMar 25, 2016 ELEC 425 S.docxNew folderelec425_2016_hw5.pdfMar 25, 2016 ELEC 425 S.docx
New folderelec425_2016_hw5.pdfMar 25, 2016 ELEC 425 S.docxcurwenmichaela
 
PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERS
PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERSPERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERS
PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERSBhuvi palaniswamy
 
Performance of cognitive radio networks with maximal ratio combining over cor...
Performance of cognitive radio networks with maximal ratio combining over cor...Performance of cognitive radio networks with maximal ratio combining over cor...
Performance of cognitive radio networks with maximal ratio combining over cor...Polytechnique Montreal
 
Some possible interpretations from data of the CODALEMA experiment
Some possible interpretations from data of the CODALEMA experimentSome possible interpretations from data of the CODALEMA experiment
Some possible interpretations from data of the CODALEMA experimentAhmed Ammar Rebai PhD
 
Characterisation of neutron meriaty
Characterisation of neutron  meriatyCharacterisation of neutron  meriaty
Characterisation of neutron meriatyLeishman Associates
 
Optical interferometery to detect sound waves as an analogue for gravitationa...
Optical interferometery to detect sound waves as an analogue for gravitationa...Optical interferometery to detect sound waves as an analogue for gravitationa...
Optical interferometery to detect sound waves as an analogue for gravitationa...Thomas Actn
 

Ähnlich wie Rtm assignment 2 (20)

Future ground arrays for ultrahigh-energy cosmic rays: recent updates and per...
Future ground arrays for ultrahigh-energy cosmic rays: recent updates and per...Future ground arrays for ultrahigh-energy cosmic rays: recent updates and per...
Future ground arrays for ultrahigh-energy cosmic rays: recent updates and per...
 
Results from telescope_array_experiment
Results from telescope_array_experimentResults from telescope_array_experiment
Results from telescope_array_experiment
 
Variation of dose distribution with depth and incident energy using EGSnrc Mo...
Variation of dose distribution with depth and incident energy using EGSnrc Mo...Variation of dose distribution with depth and incident energy using EGSnrc Mo...
Variation of dose distribution with depth and incident energy using EGSnrc Mo...
 
MODELING THE CHLOROPHYLL-A FROM SEA SURFACE REFLECTANCE IN WEST AFRICA BY DEE...
MODELING THE CHLOROPHYLL-A FROM SEA SURFACE REFLECTANCE IN WEST AFRICA BY DEE...MODELING THE CHLOROPHYLL-A FROM SEA SURFACE REFLECTANCE IN WEST AFRICA BY DEE...
MODELING THE CHLOROPHYLL-A FROM SEA SURFACE REFLECTANCE IN WEST AFRICA BY DEE...
 
MODELING THE CHLOROPHYLL-A FROM SEA SURFACE REFLECTANCE IN WEST AFRICA BY DEE...
MODELING THE CHLOROPHYLL-A FROM SEA SURFACE REFLECTANCE IN WEST AFRICA BY DEE...MODELING THE CHLOROPHYLL-A FROM SEA SURFACE REFLECTANCE IN WEST AFRICA BY DEE...
MODELING THE CHLOROPHYLL-A FROM SEA SURFACE REFLECTANCE IN WEST AFRICA BY DEE...
 
Physical society presentation ashik.pptx
Physical society presentation ashik.pptxPhysical society presentation ashik.pptx
Physical society presentation ashik.pptx
 
Direct detection of a break in the teraelectronvolt cosmic-ray spectrum of el...
Direct detection of a break in the teraelectronvolt cosmic-ray spectrum of el...Direct detection of a break in the teraelectronvolt cosmic-ray spectrum of el...
Direct detection of a break in the teraelectronvolt cosmic-ray spectrum of el...
 
E04401032038
E04401032038E04401032038
E04401032038
 
Laser Physics Department -ALNeelain University 5 th
Laser Physics Department -ALNeelain University 5 thLaser Physics Department -ALNeelain University 5 th
Laser Physics Department -ALNeelain University 5 th
 
Modeling the Chlorophyll-a from Sea Surface Reflectance in West Africa by Dee...
Modeling the Chlorophyll-a from Sea Surface Reflectance in West Africa by Dee...Modeling the Chlorophyll-a from Sea Surface Reflectance in West Africa by Dee...
Modeling the Chlorophyll-a from Sea Surface Reflectance in West Africa by Dee...
 
Laser Pulsing in Linear Compton Scattering
Laser Pulsing in Linear Compton ScatteringLaser Pulsing in Linear Compton Scattering
Laser Pulsing in Linear Compton Scattering
 
NON PARAMETRIC METHOD
NON PARAMETRIC METHODNON PARAMETRIC METHOD
NON PARAMETRIC METHOD
 
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEXTISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
 
New folderelec425_2016_hw5.pdfMar 25, 2016 ELEC 425 S.docx
New folderelec425_2016_hw5.pdfMar 25, 2016 ELEC 425 S.docxNew folderelec425_2016_hw5.pdfMar 25, 2016 ELEC 425 S.docx
New folderelec425_2016_hw5.pdfMar 25, 2016 ELEC 425 S.docx
 
PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERS
PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERSPERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERS
PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERS
 
Performance of cognitive radio networks with maximal ratio combining over cor...
Performance of cognitive radio networks with maximal ratio combining over cor...Performance of cognitive radio networks with maximal ratio combining over cor...
Performance of cognitive radio networks with maximal ratio combining over cor...
 
Some possible interpretations from data of the CODALEMA experiment
Some possible interpretations from data of the CODALEMA experimentSome possible interpretations from data of the CODALEMA experiment
Some possible interpretations from data of the CODALEMA experiment
 
Characterisation of neutron meriaty
Characterisation of neutron  meriatyCharacterisation of neutron  meriaty
Characterisation of neutron meriaty
 
I010415255
I010415255I010415255
I010415255
 
Optical interferometery to detect sound waves as an analogue for gravitationa...
Optical interferometery to detect sound waves as an analogue for gravitationa...Optical interferometery to detect sound waves as an analogue for gravitationa...
Optical interferometery to detect sound waves as an analogue for gravitationa...
 

Kürzlich hochgeladen

Call Girls Visakhapatnam Just Call 8250077686 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 8250077686 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 8250077686 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 8250077686 Top Class Call Girl Service Ava...Dipal Arora
 
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappMost Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappInaaya Sharma
 
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426jennyeacort
 
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟ 9332606886 ⟟ Call Me For G...
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟  9332606886 ⟟ Call Me For G...Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟  9332606886 ⟟ Call Me For G...
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟ 9332606886 ⟟ Call Me For G...narwatsonia7
 
Call Girls Shimla Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Shimla Just Call 8617370543 Top Class Call Girl Service AvailableCall Girls Shimla Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Shimla Just Call 8617370543 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service AvailableGENUINE ESCORT AGENCY
 
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...Ishani Gupta
 
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...jageshsingh5554
 
Trichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service Available
Trichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service AvailableTrichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service Available
Trichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service AvailableGENUINE ESCORT AGENCY
 
Call Girls Guntur Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Guntur  Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Guntur  Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Guntur Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...
Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...
Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...GENUINE ESCORT AGENCY
 
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In AhmedabadGENUINE ESCORT AGENCY
 
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...narwatsonia7
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...hotbabesbook
 
Call Girls Vadodara Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Vadodara Just Call 8617370543 Top Class Call Girl Service AvailableCall Girls Vadodara Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Vadodara Just Call 8617370543 Top Class Call Girl Service AvailableDipal Arora
 
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls * UPA...
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls  * UPA...Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls  * UPA...
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls * UPA...mahaiklolahd
 
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...Anamika Rawat
 
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Dipal Arora
 

Kürzlich hochgeladen (20)

Call Girls Visakhapatnam Just Call 8250077686 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 8250077686 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 8250077686 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 8250077686 Top Class Call Girl Service Ava...
 
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappMost Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
 
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
 
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟ 9332606886 ⟟ Call Me For G...
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟  9332606886 ⟟ Call Me For G...Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟  9332606886 ⟟ Call Me For G...
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟ 9332606886 ⟟ Call Me For G...
 
Call Girls Shimla Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Shimla Just Call 8617370543 Top Class Call Girl Service AvailableCall Girls Shimla Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Shimla Just Call 8617370543 Top Class Call Girl Service Available
 
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
 
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
 
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
 
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
 
Trichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service Available
Trichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service AvailableTrichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service Available
Trichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service Available
 
Call Girls Guntur Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Guntur  Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Guntur  Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Guntur Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...
Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...
Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...
 
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
 
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
 
Call Girls Vadodara Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Vadodara Just Call 8617370543 Top Class Call Girl Service AvailableCall Girls Vadodara Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Vadodara Just Call 8617370543 Top Class Call Girl Service Available
 
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls * UPA...
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls  * UPA...Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls  * UPA...
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls * UPA...
 
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
 
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
 

Rtm assignment 2

  • 1. RTM ASSIGNMENT 2: THE WATER EQUIVALENCE OF LIF TLDS
  • 2. i TABLE OF CONTENTS TABLE OF CONTENTS...................................................................................................................................................i ABSTRACT................................................................................................................................................................... ii INTRODUCTION..........................................................................................................................................................1 METHODOLOGY .........................................................................................................................................................2 Clinical measurements...........................................................................................................................................2 Equipment..............................................................................................................................................................2 Entrance dose calibration ......................................................................................................................................2 Midline dose determination ..................................................................................................................................3 RESULTS AND DISCUSSION.........................................................................................................................................4 CONCLUSION..............................................................................................................................................................8 References .................................................................................................................................................................9
  • 3. ii ABSTRACT The dosimetry was carried out for radiotherapy patients, and measurements were performed using LiF and thermoluminescent dosimeters (TLDs). Evaluations were done for water-equivalent (effective) thicknesses and target dose with transmission data. Considerations were made for the accuracy of the parameter for the ratio of measured to expected value for each quantity. The entrance dose was estimated as 1.01 ± 0.07. The mean ratio of effective to contour depth was 1.00 ± 0.13, showing a wide distribution reflecting the influence of contour inaccuracies. The mean ratio of the measured contour dose prescription was 1.00 ± 0.07. The difference in depths that is patient and effective depth is a reflection of target dose discrepancies. Graphical simulations were done using Monte-Carlo Simulations and presented.
  • 4. 1 INTRODUCTION Probability relationship approximates the control of local tumor and normal tissue injury (Leunens, et al., 1990; Leunens, et al., 1990). Thus procedural therapy for weak point determination is key to treatment. The most and effective method for this probabilistic determination weak point is in vivo dosimetry. The commonly used method is thermoluminescent dosimeter (TLD) (Tung, et al., 2004). This application si popular since it permits the use of small dependencies on dose rate, temperature, energy and high spatial sensitivity, and resolution (Moscovitch, 1999). TLDs are popular also since they are unattached to the bias voltage and electrometer though they are passive and their answers are not prompt. In vivo dosimetry are in various levels: intracavity dose; entrance dose and exit dose. The coherent objective of dosimetry is the determination of value of doses. Therefore, the study is first-quality assurance program for determination of target dose based on water-equivalent (effective) diameter. This was done for comparison with the actual contour diameter for investigation of the effect of the corresponding variation of the delivered dose to the target. Photons on the other hand interacts with atoms in many ways (Corvan, et al., 2014): the multi- MeV regime, where photoionization occurs, electron-positron pair production and compton scattering. There are also a number of ways of detection of MeV gamma rays which in most cases exploits quantum electromagnetic cascade initiated by photon solidification and compton scattering.
  • 5. 2 METHODOLOGY Clinical measurements The study was forced on the Source Axis Distance (SAD) since the study is focused on the midline. Measurements were taken for dosage at the entrance and exit doses. TLDs of 50 𝑚𝑚 build up was used to cover material for placement of the surface perpendicular to the beam central axis at each measuring point. Shadow effect was avoided by either the entrance detector or the exit detector by shifting the beam slightly out of his axis. Accuracy of the treatment procedure is the ratio of the measured to the expected values. Equipment Since TLD is water equivalent, the analysis of TLDs were done with the Harshaw-3500 manual reader. With an absorbed dose of 1 Gy at the center of a 10 × 10 cm2 field size and 95 cm source- skin distance (SSD), the residual signal was noted to be < 1%. Thus batch uniformity = 3.5% which is equivalent to 1 SD. Entrance dose calibration TLDs were positioned on phantom surface for measurement of entrance dose (Den) measurement and calibration were done for the determination of dose inside the phantom at a depth of the maximum dose (𝑑 𝑚). Sensitivity loss factor was estimated by: 𝑆𝑓 = ( 𝑅̅ 𝑟𝑒𝑓 𝐼𝐶 𝑎𝑏𝑠𝑜𝑟𝑏𝑒𝑑 𝑑𝑜𝑠𝑒) 1 ( 𝑅̅ 𝑟𝑒𝑓 𝐼𝐶 𝑎𝑏𝑠𝑜𝑟𝑏𝑒𝑑 𝑑𝑜𝑠𝑒) 𝑓
  • 6. 3 Midline dose determination Determination of Midline dose was based on measurements taken during transmission. From the theory, 𝑇𝑒𝑥 = 𝐷 𝑒𝑥 𝐷 𝑒𝑛 , while 𝑇 𝑚𝑖𝑑 = 𝐷 𝑚𝑖𝑑 𝐷 𝑒𝑛 . However, 𝑇𝑒𝑥 and 𝑇 𝑚𝑖𝑑 could be calculated from 𝑇𝑒𝑥(𝐴′ , 𝑑 𝑧−𝑒𝑛) = 𝑇𝑀𝑅(𝐴′ , 𝑑 𝑧−𝑑 𝑒𝑛 ) ( 𝑆𝑆𝐷 + 𝑑 𝑚 𝑆𝑆𝐷 + 𝑍 − 𝑑 𝑚 ) 2 𝐵𝑆𝐹(𝐴′) 𝐵𝑆𝐹(𝐴0) 𝑇 𝑚𝑖𝑑 (𝐴, 𝑑 𝑍 2 ) = 𝑇𝑀𝑅 (𝐴′ , 𝑑 𝑍 2 ) ( 𝑆𝑆𝐷 + 𝑑 𝑚 𝑆𝑆𝐷 + 𝑍 𝑑 𝑚 ) 2 𝐵𝑆𝐹(𝐴) 𝐵𝑆𝐹(𝐴0) where 𝐴0, 𝐴, and 𝐴′ are the field sizes at the entrance, midline (( 𝑆𝑆𝐷+ 𝑍 2 𝑆𝑆𝐷+𝑑 𝑚 ) 𝐴0) and exit levels (( 𝑆𝑆𝐷 + 𝑧 − 𝑑 𝑚 𝑆𝑆𝐷 + 𝑑 𝑚 ) 𝐴0) and 𝑍 is the water-equivalent (effective) depth, 𝑑 𝑚 maximum dose depth.
  • 7. 4 RESULTS AND DISCUSSION Correction factors Correction factors have to be determined when irradiation geometry differs from the reference geometry used in the calibration procedure. These variations in TLDs’ response were examined in different SSDs and field sizes. Correction factor (CF) is defined as 𝐶𝐹 = 𝑅̅ 𝑟𝑒𝑓 𝐼𝐶 (𝑎𝑏𝑠𝑜𝑟𝑏𝑒𝑑 𝑑𝑜𝑠𝑒) 𝑟𝑒𝑓𝑒𝑟𝑒𝑛𝑐𝑒 𝑔𝑒𝑜𝑔𝑚𝑒𝑡𝑟𝑦 𝑅̅ 𝑟𝑒𝑓 𝐼𝐶 (𝑎𝑏𝑠𝑜𝑟𝑏𝑒𝑑 𝑑𝑜𝑠𝑒) 𝑔𝑒𝑜𝑔𝑚𝑒𝑡𝑟𝑦 𝑜𝑓 𝑖𝑛𝑡𝑒𝑟𝑒𝑠𝑡 The attention of the experiment was focused on the LiF, a material of the lowest value of Z. Based on the database, the Compton scattering was noted as the overriding interaction between mechanisms for a multi-MeV gamma-ray beam up to energies of 20 MeV as shown in the figure below. Figure 1: Energy loss for a multi-MeV gamma ray beam as it interacts with LiF.
  • 8. 5 The lines in the square corresponds to nuclear filed while the blue is for diamond for Compton scattering. Compton scattering is dominant up to around 1.5 MeV, after which pair production dominates. To estimate the effect of LiF on the spectrum of the electrons, simulation was done using the Monte-Carlo code FLUKA and the simulation present in the figure below. Figure 2: Spectra of the electrons generated. However, the Gaussian gave appropriate estimation of the relationship between the peak values of the conversion factor and is show in the Figure 3 below. Furthermore, the peak of the conversion factors in the electron spectra plotted as a function of the initial photon energy. The peaks rise in magnitude up to 5 MeV where the peaks remain endless up to 10 MeV after which they start to decrease again. This shown in Figure 4.
  • 9. 6 Figure 3: The peak of the conversion factors in the electron spectra plotted as a function of the initial photon energy
  • 10. 7 Figure 4: Fluka simulations.
  • 11. 8 CONCLUSION Lif and TLD was used to carry a compressive series evaluation of in vivo dosimetry measurements. Determination of water-equivalent depth and midline dose showed a close association between the midline dose imprecision and discrepancy of water-equivalent and contour depths. Furthermore, the modification of compact gamma-ray spectrometer for in vivo dosimetry for treatment presented showed that for energy windows around 20 MeV, using Monte-Carlo simulations, the performance of the system can be checked. The results of this experiments were in line with other conducted by Aad, et al., (2014).
  • 12. 9 References Aad, G. a. A. T. a. A. B. a. A. J. a. K. S. A. a. A. A. et al., 2014. Measurement of the inclusive isolated prompt photons cross section in p p collisions at s= 7 TeV with the ATLAS detector using 4.6 fb- 1. hysical Review D, 89(5), p. 052004. Corvan, D. J., Sarri, G. & Zepf, M., 2014. Design of a compact spectrometer for high-flux MeV gamma-ray beams. Review of Scientific Instruments, 85(6), p. 065119. Leunens, G., Van Dam, J., Dutreix, A. & Van der Schueren, E., 1990. Quality assurance in radiotherapy by in vivo dosimetry. 1. Entrance dose measurements, a reliable procedure. Radiotherapy and Oncology, 17(2), pp. 141-151. Leunens, G., Van Dam, J., Dutreix, A. & Van der Schueren, E., 1990. Quality assurance in radiotherapy by in vivo dosimetry. 2. Determination of the target absorbed dose. Radiotherapy and Oncology, 19(1), pp. 73-87. Moscovitch, M., 1999. Personnel dosimetry using LiF: Mg, Cu, P. Radiation protection dosimetry, 85(1-4), pp. 49-56. Tung, C. et al., 2004. In vivo dosimetry for external photon treatments of head and neck cancers by diodes and TLDS. Radiation protection dosimetry, 111(1), pp. 45-50.