1. Some important points to remember! 1.) Typical response obtained by
chromatography (i.e., a chromatogram): Where: tR = retention time tM =
void time Wb = baseline width of the peak in time units Wh = half-height
width of the peak in time units
2. 2.) Solute Retention: A solute’s retention time or retention volume in
chromatography is directly related to the strength of the solute’s interaction
with the mobile and stationary phases. Retention on a given column pertain
to the particulars of that system: - size of the column - flow rate of the
mobile phase
smaller difference in interactions in order to separate two solutes
Efficiency is related theoretically to the various kinetic processes that are
involved in solute retention and transport in the column - determine the
width or standard deviation (σ) of peaks Estimate σ from peak widths,
assuming Gaussian shaped peak: Wh Wb = 4σ Wh = 2.354σ Dependent
on the amount of time that a solute spends in the column (k’ or tR) 3. 3.)
Efficiency: Efficiency is related experimentally to a solute’s peak width. - an
efficient system will produce narrow peaks - narrow peaks
4. Number of theoretical plates (N): compare efficiencies of a system for
solutes that have different retention times N = (tR/σ)2 or for a Gaussian
shaped peak N = 16 (tR/Wb)2 N = 5.54 (tR/Wh)2 The larger the value of N
is for a column, the better the column will be able to separate two
compounds. - the better the ability to resolve solutes that have small
differences in retention - N is independent of solute retention - N is
dependent on the length of the column
5. Plate height or height equivalent of a theoretical plate (H or HETP):
compare efficiencies ofcolumns with different lengths: H = L/N where: L =
column length N = number of theoretical plates for the columnNote: H
simply gives the length of the column that corresponds to one theoretical
plateH can be also used to relate various chromatographic parameters
(e.g., flow rate, particle size, etc.) to the kinetic processes that give rise to
peak broadening: Why Do Bands Spread? a. Eddy diffusion b. Mobile
phase mass transfer c. Stagnant mobile phase mass transfer d. Stationary
phase mass transfer e. Longitudinal diffusion
6. Van Deemter equation: relates flow-rate or linear velocity to H: H = A +
B/µ + Cµ where: µ = linear velocity (flow-rate x Vm/L) H = total plate height
of the column A = constant representing eddy diffusion & mobile phase
mass transfer B = constant representing longitudinal diffusion C = constant
representing stagnant mobile phase & stationary phase mass transfer One
use of plate height (H) is to relate these kinetic process to band broadening
to a parameter of the chromatographic system (e.g., flow-rate). This
relationship is used to predict what the resulting effect would be of varying
this parameter on the overall efficiency of the chromatographic system.
7. Plot of van Deemter equation shows how H changes with the linear
velocity (flow-rate) of the mobile phase µ optimum Optimum linear velocity
(µopt) - where H has a minimum value and the point of maximum column
efficiency. µopt is easy to achieve for gas chromatography, but is usually
too small for liquid chromatography requiring flow-rates higher than optimal
to separate compounds
8. 4.) Measures of Solute Separation: separation factor (α) – parameter
used to describe how well two solutes are separated by a chromatographic
system: α = k’2/k’1 k’ = (tR –tM)/tM where: k’1 = the capacity factor of the
first solute k’2 = the capacity factor of the second solute, with k’2 ~ k’ A
value of α ~ 1.1 is usually indicative of a good separation Does not
consider the effect of column efficiency or peak widths, only retention.
ideal case. Rs ~1.0 considered adequate for most separations. 9.
Resolution (RS) – resolution between two peaks is a second measure of
howwell two peaks are separated: tr2 – tr1 RS = (Wb2 + Wb1)/2 where:
tr1, Wb1 = retention time and baseline width for the first eluting peak tr2,
Wb2 = retention time and baseline width for the second eluting peak Rs is
preferred over α since both retention (tr) and column efficiency (Wb) are
considered in defining peak separation. Rs ~1.5 represents baseline
resolution, or complete separation of two neighboring solutes
10. Gas ChromatographyA.) Introduction: Gas Chromatography (GC) is a
chromatographic technique in which the mobile phase is a gas. GC is
currently one of the most popular methods for separating and analyzing
compounds. This is due to its high resolution, low limits of detection,
speed, accuracy and reproducibility. GC can be applied to the separation
of any compound that is either naturally volatile (i.e., readily goes into the
gas phase) or can be converted to a volatile derivative. This makes GC
useful in the separation of a number of small organic and inorganic
compounds.
11. Gas ChromatographyB.) Equipment: A simple GC system consists of:
1. Gas source (with pressure and flow regulators) 2. Injector or sample
application system 3. Chromatographic column (with oven for temperature
control) 4. Detector & computer or recorder
12. A typical GC system used is shown below (a gas chromatograph)
Carrier gas: He (common), N2, H2 Pinlet 10-50 psig Flow = 25-150 mL/min
packed column Flow = 1-25 mL/min open tubular column Column: 2-50 m
coiled stainless steel/glass/Teflon Oven: 0-400 °C ~ average boiling point
of sample Accurate to <1 °C Detectors: FID, TCD, ECD, (MS)
13. Schematic Diagram of Gas Chromatography
14. C.) Mobile Phase: GC separates solutes based on their different
interactions with the mobile and stationary phases. - solute’s retention is
determined mostly by its vapor pressure and volatility - solute’s retention is
controlled by its interaction with the stationary phase - gas mobile phase
has much lower density ‚decreased chance for interacting with solute
‚increased chance that solid or liquid stationary phase interacts with solute
15. Carrier gas – main purpose of the gas in GC is to move the solutes
along the column, mobile phase is often referred to as carrier gas.Common
carrier gas: include He, Ar, H2, N2
16. Carrier Gas or Mobile phase does not affect solute retention, butdoes
affect: 1.) faster, more efficient separations 2.) Stability of column and
solutes - H2 or O2 can react with functional groups on solutes and
stationary phase or with surfaces of theinjector, connections and detector
3.) Response of the detector - thermal conductor requires H2 or He - other
detectors require specific carrier gas larger diffusion coefficients - low
molecular weight gases Desired efficiency for the GC System - low
molecular weight gases (He, H2)
17. D.) Stationary Phases: Stationary phase in GC is the main factor
determining the selectivity and retention of solutes. There are three types
of stationary phases used in GC: Solid adsorbents Liquids coated on solid
supports Bonded-phase supports
18. 1.) Gas-solid chromatography (GSC) - same material is used as both
the stationary phase and support material - common adsorbents include: ‚
alumina ‚l molecular sieves (crystalline aluminosilicates [zeolites] and clay)
‚ silica ‚ active carbon Magnified Pores in activated carbon
19. Gas-solid chromatography (GSC):advantages: - long column lifetimes -
ability to retain and separate some compounds not easily resolved by other
GC methods ‚ geometrical isomers ‚ permanent gasesdisadvantage: - very
strong retention of low volatility or polar solutes - catalytic changes that can
occur on GSC supports - GSC supports have a range of chemical and
physical environments ‚ different strength retention sites ‚ non-symmetrical
peaks ‚ variable retention times
20. 2.) Gas-liquid chromatography (GLC) - stationary phase is some liquid
coated on a solid support - over 400 liquid stationary phases available for
GLC ‚ many stationary phases are very similar in terms of their retention
properties - material range from polymers (polysiloxanes, polyesters,
polyethylene glycols) to fluorocarbons, molten salts and liquid crystals
21. Based on polarity, of the 400 phases available only 6-12 are needed for
most separations. The routinely recommended phases are listed below:
Chemical nature of Max. McReynolds’ constantsName polysiloxane temp.
x’ y’ z’ µ’ s’SE-30 Dimethyl 350 14 53 44 64 41Dexsil300 Carborane-
dimethyl 450 43 64 111 151 101OV-17 50% Phenyl methyl 375 119 158
162 243 202 Higher the number theOV-210 50% Trifluoropropyl 270 146
238 358 468 310 higher the absorption.OV-225 25% Cyanopropyl- 250 238
369 338 492 386 25% phenylSilar-SCP 50% Cyanopropyl- 275 319 495
446 637 531 50% phenylSP-2340 75% Cyanopropyl 275 520 757 659 942
804OV-275 Dicyanoallyl 250 629 872 763 1106 849 McReynolds’
constants based on retention of 5 standard “probe” analytes – Benzene, n-
butanol, 2-pentanone, nitropropanone, pyridine
22. Preparing a stationary phase for GLC: - slurry of the desired liquid
phase and solvent is made with a solid support ‚i w solid support is usually
diatomaceous earth (fossilized shells of ancient aquatic algae (diatoms),
silica-based material) - solvent is evaporated off, coating the liquid
stationary phase on the support - the resulting material is then packed into
the column
23. disadvantage: - liquid may slowly bleed off with time ‚ especially if high
temperatures are used ‚ contribute to background ‚ change characteristics
of the column with time
24. 3.) Bonded-Phase Gas chromatography - covalently attach stationary
phase to the solid support material - avoids column bleeding in GLC -
bonded phases are prepared by reacting the desired phase with the
surface of a silica-based supportreactions form an Si-O-Si bond between
the stationary phase and support orreactions form an Si-C-C-Si bond
between the stationary phase and support
25. - many bonded phases exist, but most separations can be formed
withthe following commonly recommended bonded-phases: ‚
Dimethylpolysiloxane ‚ Methyl(phenyl)polysiloxane ‚ Polyethylene glycol
(Carbowax 20M) ‚ Trifluoropropylpolysiloxane ‚ Cyanopropylpolysiloxane
CH3 CH3 C6H5 H HO Si O Si O Si HO C C O H CH3 n CH3 C6H5 n n m
H H advantages: - more stable than coated liquid phases - can be placed
on support with thinner and more uniform thickness than liquid phases
26. E.) Support Material: There are two main types of supports used in GC:
Packed columns ‚ large sample capacity ‚ preparative work Capillary (open-
tubular) columns ‚ahigher efficiency ‚i smaller sample size ‚ analytical
applications
general elution problem Gradient elution - change column condition with
time which changes retention of solutes to overcome general elution
problem 27. F.) Elution Methods: A common problem to all
chromatographic techniques is that in any one sample there may be many
solutes present, each retained by the column to a different degree: Best
separation and limits of detection are usually obtained with solutes with k’
values of 2-10 Difficult to find one condition that elutes all solutes in this k’
range
28. Temperature Programming – changing the temperature on the column
with time tosimulate gradient elution in GC since a solute’s retention in GC
is related to itsvolatility. Comparison of a GC separation using isothermal
conditions and temperature programming is shown below ISOTHERMAL
Column temp. 120oC Programmed temp. (30oC to 180oC) (5o/min)
Temperature programming is usually done either in a stepwise change, a
linear change or a combination of several linear changes. A single linear
change or ramp is the most common
29. G.) GC Detectors: The following devices are common types of GC
detectors: 1. Thermal Conductivity Detector (TCD) 2. Flame Ionization
Detector (FID) 3. Nitrogen-phosphorus Detector 4. Electron Capture
Detector (ECD) 5. Mass Spectrometers (discussed later in the course) The
choice of detector will depend on the analyte and how the GC method is
being used (i.e., analytical or preparative scale)
30. 1.) Thermal Conductivity Detector (TCD) - katherometer or hot-wire
detector - first universal detector developed for GCProcess - measures a
bulk property of the mobile phase leaving the column. - measures ability to
conduct heat away from a hot-wire (i.e., thermal conductivity) - thermal
conductivity changes with presence of other components in the mobile
phase
change in voltage difference between points (+) and (-). change in
resistor’s temperature and resistance change in amount of heat removed
from resistor 31. Design- based on electronic circuit known as a
Wheatstone bridge.- circuit consists of an arrangement of four resistors
with a fixed current applied to them.- thermal conductivity changes with
presence of other components in the mobile phase.- the voltage between
points (+) and (-) will be zero as long as the resistances in the different
arms of the circuit are properly balanced -one resistor in contact with
mobile phase leaving column -another in contact with reference stream of
pure mobile phase as solute emerge from column: change in thermal
conductivity
32. Considerations - mobile phase must have very different thermal
conductivity then solutes being separated. - most compounds separated in
GC have thermal conductivity of about 1-4X10-5. - H2 and He are carrier
gases with significantly different thermal conductivity values. - H2 reacts
with metal oxides present on the resistors, so not used
33. advantages: - truly universal detector ‚ applicable to the detection of
any compound in GC - non-destructive ‚ useful for detecting compounds
from preparative-scale columns ‚ useful in combination with other types of
GC detectorsdisadvantage: - detect mobile phase impurities - sensitive to
changes in flow-rates - limit of detection ‚ ~ 10-7 M ‚ much higher then
other GC detectors
34. 2.) Flame Ionization Detector (FID) - most common type of GC detector
- “universal” detector capable of measuring the presence of almost any
organic and many inorganic compoundProcess - measures the production
of ions when a solute is burned in a flame. - ions are collected at an
electrode to create a current
35. advantages: - universal detector for organics ‚ doesn’t respond to
common inorganic compounds - mobile phase impurities not detected -
carrier gases not detected - limit of detection: FID is 1000x better than TCD
- linear and dynamic range better than TCDdisadvantage: - destructive
detector
36. 3.) Nitrogen-Phosphorus Detector (NPD) - used for detecting nitrogen-
or phosphorus containing compounds - also known as alkali flame
ionization detector or thermionic detectorProcess Alkali Bead - same basic
principal as FID - measures production of ions when a solute is burned in a
flame - ions are collected at an electrode to create a current - contains a
small amount of alkali metal vapor in the flame - enhances the formation of
ions from nitrogen- and phosphorus- containing compounds
37. 3.) Nitrogen-Phosphorus Detector (NPD)advantages: - useful for
environmental testing ‚ detection of organophosphate pesticides - useful for
drug analysis - ‚ determination of amine-containing or basic drugs - Like
FID, does not detect common mobile phase impurities or carrier gases -
limit of detection: NPD is 500x better than FID in detecting nitrogen and
phosphorus- containing compounds - NPD more sensitive to other
heterocompounds, such as sulfur-, halogen-, and arsenic- containing
moleculesdisadvantage: - destructive detector - NPD is less sensitive to
organic compounds compared to FID
38. 4.) Electron Capture Detector (ECD) - radiation-based detector -
selective for compounds containing electronegative atoms, such as
halogensProcess - based on the capture of electrons by electronegative
atoms in a molecule - electrons are produced by ionization of the carrier
gas with a radioactive source ‚ 3H or 63Ni - in absence of solute, steady
stream of these electrons is produced - electrons go to collector electrode
where they produce a current - compounds with electronegative atoms
capture electrons, reducing current
39. advantages: - useful for environmental testing ‚ detection of chlorinated
pesticides or herbicides ‚ detection of polynuclear aromatic carcinogens ‚
detection of organometallic compounds - selective for halogen- (I, Br, Cl,
F), nitro-, and sulfur- containing compounds - detects polynuclear aromatic
compounds, anhydrides and conjugated carbonyl compounds
40. Gas Chromatography Application
41. Gas Chromatography can only be used toanalyze volatile components
(i.e. compoundsthat can be volatilized at 250oC
withoutdecomposition).Non-volatile components have to bederivatized
before GC analysis.
42. Retention TimesResponse RT= 4.0 min on SE-30 Hexane GC
Retention Time on SE-30Response RT= 4 min on SE-30 Unknown
compound GC Retention Time on SE-30
43. TENTATIVE IDENTIFICATION OF UNKNOWN
COMPOUNDSResponse Mixture of known compounds Octane Decane 1.6
min = RT Hexane GC Retention Time on Carbowax-20 (min)Response
Unknown compound may be Hexane 1.6 min = RT Retention Time on
Carbowax-20 (min)
44. SEMI- QUANTITATIVE ANALYSIS OF FATTY ACIDSDetector
Response C18 Peak Area (cm2 ) C 16 10 8 C14 6 4 2 0.5 1.0 1.5 2.0 2.5
3.0 Retention Time Sample Concentration (mg/ml) C4 1 The content % of
C fatty acids = 14 1 ∗0 0 C + C11 + C 8 6 1 4 = the content % of C fatty
acids 14
45. GLC ADVANTAGES1. Very good separation2. Time (analysis is
short)3. Small sample is needed - µl4. Good detection system5. Can be
used for both qualitative and quantitative analysis
46. DISADVANTAGES OF GAS CHROMATOGRAPHYMaterial has to be
volatilized at 250oCwithout decomposition. If not the analyte hasto be
derivatized.e.g. Fatty acids → Methyl esters Phenols → Silanized Sugars
→ Silanized
47. Fatty Acids Methylester O OR C OH + CH3 OH + H2 SO4 R C O CH3
Reflux Volatile in Gas Chromatography O CH2 O C R O CH3 ONa O CH O
C R + CH 3 OH 3 R C O CH3 Volatile in Gas O Chromatography CH2 O C
R
48. Gas Chromatogram of Methyl Esters of Fatty Acids
49. Derivation of Glucose with Trimethylchlorosilane 6 CH2 OH CH O 3 5 4
1 + 5Cl Si CH OH 3 HO OH CH 3 2 3 OH Glucose Trimethylchlorosilane 6
CH2 O-Si(CH3)3 O 5 + 5HCl 4 O-Si(CH ) 13 3 (CH3)3-Si-O 3 2 O-
Si(CH3)3 O-Si(CH3)3
50. Effects of Derivation1. Time consumption2. Side reaction3. Loss of
sample

Weitere ähnliche Inhalte

Was ist angesagt?

Q913 re1 w1 lec 4
Q913 re1 w1 lec 4Q913 re1 w1 lec 4
Q913 re1 w1 lec 4
AFATous
 
Q922+rfp+l05 v1
Q922+rfp+l05 v1Q922+rfp+l05 v1
Q922+rfp+l05 v1
AFATous
 
Q913 rfp w2 lec 5
Q913 rfp w2 lec 5Q913 rfp w2 lec 5
Q913 rfp w2 lec 5
AFATous
 
Q913 rfp w1 lec 3
Q913 rfp w1 lec 3Q913 rfp w1 lec 3
Q913 rfp w1 lec 3
AFATous
 
Q921 re1 lec2 v1
Q921 re1 lec2 v1Q921 re1 lec2 v1
Q921 re1 lec2 v1
AFATous
 
Q922+rfp+l08 v1
Q922+rfp+l08 v1Q922+rfp+l08 v1
Q922+rfp+l08 v1
AFATous
 
Q921 rfp lec3
Q921 rfp lec3Q921 rfp lec3
Q921 rfp lec3
AFATous
 
Q913 rfp w2 lec 8
Q913 rfp w2 lec 8Q913 rfp w2 lec 8
Q913 rfp w2 lec 8
AFATous
 
Q921 rfp lec2 v1
Q921 rfp lec2 v1Q921 rfp lec2 v1
Q921 rfp lec2 v1
AFATous
 
Q913 rfp w1 lec 2
Q913 rfp w1 lec 2Q913 rfp w1 lec 2
Q913 rfp w1 lec 2
AFATous
 
Final Presentation_Ekramul_Haque_Ehite_04042016
Final Presentation_Ekramul_Haque_Ehite_04042016Final Presentation_Ekramul_Haque_Ehite_04042016
Final Presentation_Ekramul_Haque_Ehite_04042016
Ekramul Haque Ehite
 
Q922+rfp+l04 v1
Q922+rfp+l04 v1Q922+rfp+l04 v1
Q922+rfp+l04 v1
AFATous
 
Q922+rfp+l06 v1
Q922+rfp+l06 v1Q922+rfp+l06 v1
Q922+rfp+l06 v1
AFATous
 
Q913 rfp w2 lec 6
Q913 rfp w2 lec 6Q913 rfp w2 lec 6
Q913 rfp w2 lec 6
AFATous
 
Q913 rfp w3 lec 10
Q913 rfp w3 lec 10Q913 rfp w3 lec 10
Q913 rfp w3 lec 10
AFATous
 

Was ist angesagt? (20)

Q913 re1 w1 lec 4
Q913 re1 w1 lec 4Q913 re1 w1 lec 4
Q913 re1 w1 lec 4
 
Chromatography
ChromatographyChromatography
Chromatography
 
Q922+rfp+l05 v1
Q922+rfp+l05 v1Q922+rfp+l05 v1
Q922+rfp+l05 v1
 
Q913 rfp w2 lec 5
Q913 rfp w2 lec 5Q913 rfp w2 lec 5
Q913 rfp w2 lec 5
 
Q913 rfp w1 lec 3
Q913 rfp w1 lec 3Q913 rfp w1 lec 3
Q913 rfp w1 lec 3
 
Q921 re1 lec2 v1
Q921 re1 lec2 v1Q921 re1 lec2 v1
Q921 re1 lec2 v1
 
Q922+rfp+l08 v1
Q922+rfp+l08 v1Q922+rfp+l08 v1
Q922+rfp+l08 v1
 
Chapter7gaschromatograph 110507221015-phpapp01
Chapter7gaschromatograph 110507221015-phpapp01Chapter7gaschromatograph 110507221015-phpapp01
Chapter7gaschromatograph 110507221015-phpapp01
 
Q921 rfp lec3
Q921 rfp lec3Q921 rfp lec3
Q921 rfp lec3
 
Q913 rfp w2 lec 8
Q913 rfp w2 lec 8Q913 rfp w2 lec 8
Q913 rfp w2 lec 8
 
Absorption
AbsorptionAbsorption
Absorption
 
Q921 rfp lec2 v1
Q921 rfp lec2 v1Q921 rfp lec2 v1
Q921 rfp lec2 v1
 
Analisis de agua merck (1)
Analisis de agua merck (1)Analisis de agua merck (1)
Analisis de agua merck (1)
 
Gas lift cad-model-project report
Gas lift cad-model-project reportGas lift cad-model-project report
Gas lift cad-model-project report
 
Q913 rfp w1 lec 2
Q913 rfp w1 lec 2Q913 rfp w1 lec 2
Q913 rfp w1 lec 2
 
Final Presentation_Ekramul_Haque_Ehite_04042016
Final Presentation_Ekramul_Haque_Ehite_04042016Final Presentation_Ekramul_Haque_Ehite_04042016
Final Presentation_Ekramul_Haque_Ehite_04042016
 
Q922+rfp+l04 v1
Q922+rfp+l04 v1Q922+rfp+l04 v1
Q922+rfp+l04 v1
 
Q922+rfp+l06 v1
Q922+rfp+l06 v1Q922+rfp+l06 v1
Q922+rfp+l06 v1
 
Q913 rfp w2 lec 6
Q913 rfp w2 lec 6Q913 rfp w2 lec 6
Q913 rfp w2 lec 6
 
Q913 rfp w3 lec 10
Q913 rfp w3 lec 10Q913 rfp w3 lec 10
Q913 rfp w3 lec 10
 

Andere mochten auch (9)

Wind energy
Wind energyWind energy
Wind energy
 
Six stroke-engine-presenation
Six stroke-engine-presenationSix stroke-engine-presenation
Six stroke-engine-presenation
 
Hike 2630
Hike 2630Hike 2630
Hike 2630
 
991491214
991491214991491214
991491214
 
9221492614
92214926149221492614
9221492614
 
Hike10211025
Hike10211025Hike10211025
Hike10211025
 
Hike114118
Hike114118Hike114118
Hike114118
 
Hike 2630
Hike 2630Hike 2630
Hike 2630
 
120213120613
120213120613120213120613
120213120613
 

Ähnlich wie Gc chromatography

Presentation of chromatography by Shaiq Ali
Presentation of chromatography by Shaiq AliPresentation of chromatography by Shaiq Ali
Presentation of chromatography by Shaiq Ali
Shaiq Ali
 
Liquid chromatography still striving for high efficiency
Liquid chromatography still striving for high efficiencyLiquid chromatography still striving for high efficiency
Liquid chromatography still striving for high efficiency
guest63ff7d
 

Ähnlich wie Gc chromatography (20)

Chromatography
ChromatographyChromatography
Chromatography
 
Presentation of chromatography by Shaiq Ali
Presentation of chromatography by Shaiq AliPresentation of chromatography by Shaiq Ali
Presentation of chromatography by Shaiq Ali
 
Lec#4_Column Efficiency_GC.ppt
Lec#4_Column Efficiency_GC.pptLec#4_Column Efficiency_GC.ppt
Lec#4_Column Efficiency_GC.ppt
 
Gas Chromatography
Gas ChromatographyGas Chromatography
Gas Chromatography
 
Liquid chromatography still striving for high efficiency
Liquid chromatography still striving for high efficiencyLiquid chromatography still striving for high efficiency
Liquid chromatography still striving for high efficiency
 
GAS CHROMATOGRAPHY
GAS CHROMATOGRAPHYGAS CHROMATOGRAPHY
GAS CHROMATOGRAPHY
 
Chromarogaphy system suitability ppt
Chromarogaphy system suitability pptChromarogaphy system suitability ppt
Chromarogaphy system suitability ppt
 
Lec#3_Separation by Chromatography.ppt
Lec#3_Separation by Chromatography.pptLec#3_Separation by Chromatography.ppt
Lec#3_Separation by Chromatography.ppt
 
Chapter 4 chromatography_b_2010
Chapter 4 chromatography_b_2010Chapter 4 chromatography_b_2010
Chapter 4 chromatography_b_2010
 
Super Critical Fluid Separation Process
Super Critical Fluid Separation ProcessSuper Critical Fluid Separation Process
Super Critical Fluid Separation Process
 
Gas Chromatography PPT GCW.pptx
Gas Chromatography PPT GCW.pptxGas Chromatography PPT GCW.pptx
Gas Chromatography PPT GCW.pptx
 
Chromatography Basics
Chromatography BasicsChromatography Basics
Chromatography Basics
 
Chromatography
Chromatography Chromatography
Chromatography
 
Gas chromatograph analyzer
Gas chromatograph analyzerGas chromatograph analyzer
Gas chromatograph analyzer
 
Hplc by niper pharma analysis
Hplc by niper pharma analysisHplc by niper pharma analysis
Hplc by niper pharma analysis
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
Gas Chromatography
Gas ChromatographyGas Chromatography
Gas Chromatography
 
gas chromat.Final.pptx
gas chromat.Final.pptxgas chromat.Final.pptx
gas chromat.Final.pptx
 
HPLC
HPLCHPLC
HPLC
 
Gas Chromatography
Gas ChromatographyGas Chromatography
Gas Chromatography
 

Kürzlich hochgeladen

Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers:  A Deep Dive into Serverless Spatial Data and FMECloud Frontiers:  A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Safe Software
 
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers:  A Deep Dive into Serverless Spatial Data and FMECloud Frontiers:  A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Safe Software
 

Kürzlich hochgeladen (20)

EMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWER
EMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWEREMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWER
EMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWER
 
DBX First Quarter 2024 Investor Presentation
DBX First Quarter 2024 Investor PresentationDBX First Quarter 2024 Investor Presentation
DBX First Quarter 2024 Investor Presentation
 
presentation ICT roal in 21st century education
presentation ICT roal in 21st century educationpresentation ICT roal in 21st century education
presentation ICT roal in 21st century education
 
Manulife - Insurer Transformation Award 2024
Manulife - Insurer Transformation Award 2024Manulife - Insurer Transformation Award 2024
Manulife - Insurer Transformation Award 2024
 
FWD Group - Insurer Innovation Award 2024
FWD Group - Insurer Innovation Award 2024FWD Group - Insurer Innovation Award 2024
FWD Group - Insurer Innovation Award 2024
 
MS Copilot expands with MS Graph connectors
MS Copilot expands with MS Graph connectorsMS Copilot expands with MS Graph connectors
MS Copilot expands with MS Graph connectors
 
Spring Boot vs Quarkus the ultimate battle - DevoxxUK
Spring Boot vs Quarkus the ultimate battle - DevoxxUKSpring Boot vs Quarkus the ultimate battle - DevoxxUK
Spring Boot vs Quarkus the ultimate battle - DevoxxUK
 
Navigating the Deluge_ Dubai Floods and the Resilience of Dubai International...
Navigating the Deluge_ Dubai Floods and the Resilience of Dubai International...Navigating the Deluge_ Dubai Floods and the Resilience of Dubai International...
Navigating the Deluge_ Dubai Floods and the Resilience of Dubai International...
 
[BuildWithAI] Introduction to Gemini.pdf
[BuildWithAI] Introduction to Gemini.pdf[BuildWithAI] Introduction to Gemini.pdf
[BuildWithAI] Introduction to Gemini.pdf
 
ProductAnonymous-April2024-WinProductDiscovery-MelissaKlemke
ProductAnonymous-April2024-WinProductDiscovery-MelissaKlemkeProductAnonymous-April2024-WinProductDiscovery-MelissaKlemke
ProductAnonymous-April2024-WinProductDiscovery-MelissaKlemke
 
CNIC Information System with Pakdata Cf In Pakistan
CNIC Information System with Pakdata Cf In PakistanCNIC Information System with Pakdata Cf In Pakistan
CNIC Information System with Pakdata Cf In Pakistan
 
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers:  A Deep Dive into Serverless Spatial Data and FMECloud Frontiers:  A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
 
AWS Community Day CPH - Three problems of Terraform
AWS Community Day CPH - Three problems of TerraformAWS Community Day CPH - Three problems of Terraform
AWS Community Day CPH - Three problems of Terraform
 
Apidays New York 2024 - The value of a flexible API Management solution for O...
Apidays New York 2024 - The value of a flexible API Management solution for O...Apidays New York 2024 - The value of a flexible API Management solution for O...
Apidays New York 2024 - The value of a flexible API Management solution for O...
 
Strategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
Strategize a Smooth Tenant-to-tenant Migration and Copilot TakeoffStrategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
Strategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
 
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers:  A Deep Dive into Serverless Spatial Data and FMECloud Frontiers:  A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
 
How to Troubleshoot Apps for the Modern Connected Worker
How to Troubleshoot Apps for the Modern Connected WorkerHow to Troubleshoot Apps for the Modern Connected Worker
How to Troubleshoot Apps for the Modern Connected Worker
 
Ransomware_Q4_2023. The report. [EN].pdf
Ransomware_Q4_2023. The report. [EN].pdfRansomware_Q4_2023. The report. [EN].pdf
Ransomware_Q4_2023. The report. [EN].pdf
 
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
 
Repurposing LNG terminals for Hydrogen Ammonia: Feasibility and Cost Saving
Repurposing LNG terminals for Hydrogen Ammonia: Feasibility and Cost SavingRepurposing LNG terminals for Hydrogen Ammonia: Feasibility and Cost Saving
Repurposing LNG terminals for Hydrogen Ammonia: Feasibility and Cost Saving
 

Gc chromatography

  • 1. 1. Some important points to remember! 1.) Typical response obtained by chromatography (i.e., a chromatogram): Where: tR = retention time tM = void time Wb = baseline width of the peak in time units Wh = half-height width of the peak in time units 2. 2.) Solute Retention: A solute’s retention time or retention volume in chromatography is directly related to the strength of the solute’s interaction with the mobile and stationary phases. Retention on a given column pertain to the particulars of that system: - size of the column - flow rate of the mobile phase smaller difference in interactions in order to separate two solutes Efficiency is related theoretically to the various kinetic processes that are involved in solute retention and transport in the column - determine the width or standard deviation (σ) of peaks Estimate σ from peak widths, assuming Gaussian shaped peak: Wh Wb = 4σ Wh = 2.354σ Dependent on the amount of time that a solute spends in the column (k’ or tR) 3. 3.) Efficiency: Efficiency is related experimentally to a solute’s peak width. - an efficient system will produce narrow peaks - narrow peaks 4. Number of theoretical plates (N): compare efficiencies of a system for solutes that have different retention times N = (tR/σ)2 or for a Gaussian shaped peak N = 16 (tR/Wb)2 N = 5.54 (tR/Wh)2 The larger the value of N is for a column, the better the column will be able to separate two compounds. - the better the ability to resolve solutes that have small differences in retention - N is independent of solute retention - N is dependent on the length of the column 5. Plate height or height equivalent of a theoretical plate (H or HETP): compare efficiencies ofcolumns with different lengths: H = L/N where: L = column length N = number of theoretical plates for the columnNote: H simply gives the length of the column that corresponds to one theoretical plateH can be also used to relate various chromatographic parameters (e.g., flow rate, particle size, etc.) to the kinetic processes that give rise to
  • 2. peak broadening: Why Do Bands Spread? a. Eddy diffusion b. Mobile phase mass transfer c. Stagnant mobile phase mass transfer d. Stationary phase mass transfer e. Longitudinal diffusion 6. Van Deemter equation: relates flow-rate or linear velocity to H: H = A + B/µ + Cµ where: µ = linear velocity (flow-rate x Vm/L) H = total plate height of the column A = constant representing eddy diffusion & mobile phase mass transfer B = constant representing longitudinal diffusion C = constant representing stagnant mobile phase & stationary phase mass transfer One use of plate height (H) is to relate these kinetic process to band broadening to a parameter of the chromatographic system (e.g., flow-rate). This relationship is used to predict what the resulting effect would be of varying this parameter on the overall efficiency of the chromatographic system. 7. Plot of van Deemter equation shows how H changes with the linear velocity (flow-rate) of the mobile phase µ optimum Optimum linear velocity (µopt) - where H has a minimum value and the point of maximum column efficiency. µopt is easy to achieve for gas chromatography, but is usually too small for liquid chromatography requiring flow-rates higher than optimal to separate compounds 8. 4.) Measures of Solute Separation: separation factor (α) – parameter used to describe how well two solutes are separated by a chromatographic system: α = k’2/k’1 k’ = (tR –tM)/tM where: k’1 = the capacity factor of the first solute k’2 = the capacity factor of the second solute, with k’2 ~ k’ A value of α ~ 1.1 is usually indicative of a good separation Does not consider the effect of column efficiency or peak widths, only retention. ideal case. Rs ~1.0 considered adequate for most separations. 9. Resolution (RS) – resolution between two peaks is a second measure of howwell two peaks are separated: tr2 – tr1 RS = (Wb2 + Wb1)/2 where: tr1, Wb1 = retention time and baseline width for the first eluting peak tr2, Wb2 = retention time and baseline width for the second eluting peak Rs is preferred over α since both retention (tr) and column efficiency (Wb) are
  • 3. considered in defining peak separation. Rs ~1.5 represents baseline resolution, or complete separation of two neighboring solutes 10. Gas ChromatographyA.) Introduction: Gas Chromatography (GC) is a chromatographic technique in which the mobile phase is a gas. GC is currently one of the most popular methods for separating and analyzing compounds. This is due to its high resolution, low limits of detection, speed, accuracy and reproducibility. GC can be applied to the separation of any compound that is either naturally volatile (i.e., readily goes into the gas phase) or can be converted to a volatile derivative. This makes GC useful in the separation of a number of small organic and inorganic compounds. 11. Gas ChromatographyB.) Equipment: A simple GC system consists of: 1. Gas source (with pressure and flow regulators) 2. Injector or sample application system 3. Chromatographic column (with oven for temperature control) 4. Detector & computer or recorder 12. A typical GC system used is shown below (a gas chromatograph) Carrier gas: He (common), N2, H2 Pinlet 10-50 psig Flow = 25-150 mL/min packed column Flow = 1-25 mL/min open tubular column Column: 2-50 m coiled stainless steel/glass/Teflon Oven: 0-400 °C ~ average boiling point of sample Accurate to <1 °C Detectors: FID, TCD, ECD, (MS) 13. Schematic Diagram of Gas Chromatography 14. C.) Mobile Phase: GC separates solutes based on their different interactions with the mobile and stationary phases. - solute’s retention is determined mostly by its vapor pressure and volatility - solute’s retention is controlled by its interaction with the stationary phase - gas mobile phase has much lower density ‚decreased chance for interacting with solute ‚increased chance that solid or liquid stationary phase interacts with solute
  • 4. 15. Carrier gas – main purpose of the gas in GC is to move the solutes along the column, mobile phase is often referred to as carrier gas.Common carrier gas: include He, Ar, H2, N2 16. Carrier Gas or Mobile phase does not affect solute retention, butdoes affect: 1.) faster, more efficient separations 2.) Stability of column and solutes - H2 or O2 can react with functional groups on solutes and stationary phase or with surfaces of theinjector, connections and detector 3.) Response of the detector - thermal conductor requires H2 or He - other detectors require specific carrier gas larger diffusion coefficients - low molecular weight gases Desired efficiency for the GC System - low molecular weight gases (He, H2) 17. D.) Stationary Phases: Stationary phase in GC is the main factor determining the selectivity and retention of solutes. There are three types of stationary phases used in GC: Solid adsorbents Liquids coated on solid supports Bonded-phase supports 18. 1.) Gas-solid chromatography (GSC) - same material is used as both the stationary phase and support material - common adsorbents include: ‚ alumina ‚l molecular sieves (crystalline aluminosilicates [zeolites] and clay) ‚ silica ‚ active carbon Magnified Pores in activated carbon 19. Gas-solid chromatography (GSC):advantages: - long column lifetimes - ability to retain and separate some compounds not easily resolved by other GC methods ‚ geometrical isomers ‚ permanent gasesdisadvantage: - very strong retention of low volatility or polar solutes - catalytic changes that can occur on GSC supports - GSC supports have a range of chemical and physical environments ‚ different strength retention sites ‚ non-symmetrical peaks ‚ variable retention times 20. 2.) Gas-liquid chromatography (GLC) - stationary phase is some liquid coated on a solid support - over 400 liquid stationary phases available for GLC ‚ many stationary phases are very similar in terms of their retention
  • 5. properties - material range from polymers (polysiloxanes, polyesters, polyethylene glycols) to fluorocarbons, molten salts and liquid crystals 21. Based on polarity, of the 400 phases available only 6-12 are needed for most separations. The routinely recommended phases are listed below: Chemical nature of Max. McReynolds’ constantsName polysiloxane temp. x’ y’ z’ µ’ s’SE-30 Dimethyl 350 14 53 44 64 41Dexsil300 Carborane- dimethyl 450 43 64 111 151 101OV-17 50% Phenyl methyl 375 119 158 162 243 202 Higher the number theOV-210 50% Trifluoropropyl 270 146 238 358 468 310 higher the absorption.OV-225 25% Cyanopropyl- 250 238 369 338 492 386 25% phenylSilar-SCP 50% Cyanopropyl- 275 319 495 446 637 531 50% phenylSP-2340 75% Cyanopropyl 275 520 757 659 942 804OV-275 Dicyanoallyl 250 629 872 763 1106 849 McReynolds’ constants based on retention of 5 standard “probe” analytes – Benzene, n- butanol, 2-pentanone, nitropropanone, pyridine 22. Preparing a stationary phase for GLC: - slurry of the desired liquid phase and solvent is made with a solid support ‚i w solid support is usually diatomaceous earth (fossilized shells of ancient aquatic algae (diatoms), silica-based material) - solvent is evaporated off, coating the liquid stationary phase on the support - the resulting material is then packed into the column 23. disadvantage: - liquid may slowly bleed off with time ‚ especially if high temperatures are used ‚ contribute to background ‚ change characteristics of the column with time 24. 3.) Bonded-Phase Gas chromatography - covalently attach stationary phase to the solid support material - avoids column bleeding in GLC - bonded phases are prepared by reacting the desired phase with the surface of a silica-based supportreactions form an Si-O-Si bond between the stationary phase and support orreactions form an Si-C-C-Si bond between the stationary phase and support
  • 6. 25. - many bonded phases exist, but most separations can be formed withthe following commonly recommended bonded-phases: ‚ Dimethylpolysiloxane ‚ Methyl(phenyl)polysiloxane ‚ Polyethylene glycol (Carbowax 20M) ‚ Trifluoropropylpolysiloxane ‚ Cyanopropylpolysiloxane CH3 CH3 C6H5 H HO Si O Si O Si HO C C O H CH3 n CH3 C6H5 n n m H H advantages: - more stable than coated liquid phases - can be placed on support with thinner and more uniform thickness than liquid phases 26. E.) Support Material: There are two main types of supports used in GC: Packed columns ‚ large sample capacity ‚ preparative work Capillary (open- tubular) columns ‚ahigher efficiency ‚i smaller sample size ‚ analytical applications general elution problem Gradient elution - change column condition with time which changes retention of solutes to overcome general elution problem 27. F.) Elution Methods: A common problem to all chromatographic techniques is that in any one sample there may be many solutes present, each retained by the column to a different degree: Best separation and limits of detection are usually obtained with solutes with k’ values of 2-10 Difficult to find one condition that elutes all solutes in this k’ range 28. Temperature Programming – changing the temperature on the column with time tosimulate gradient elution in GC since a solute’s retention in GC is related to itsvolatility. Comparison of a GC separation using isothermal conditions and temperature programming is shown below ISOTHERMAL Column temp. 120oC Programmed temp. (30oC to 180oC) (5o/min) Temperature programming is usually done either in a stepwise change, a linear change or a combination of several linear changes. A single linear change or ramp is the most common 29. G.) GC Detectors: The following devices are common types of GC detectors: 1. Thermal Conductivity Detector (TCD) 2. Flame Ionization Detector (FID) 3. Nitrogen-phosphorus Detector 4. Electron Capture
  • 7. Detector (ECD) 5. Mass Spectrometers (discussed later in the course) The choice of detector will depend on the analyte and how the GC method is being used (i.e., analytical or preparative scale) 30. 1.) Thermal Conductivity Detector (TCD) - katherometer or hot-wire detector - first universal detector developed for GCProcess - measures a bulk property of the mobile phase leaving the column. - measures ability to conduct heat away from a hot-wire (i.e., thermal conductivity) - thermal conductivity changes with presence of other components in the mobile phase change in voltage difference between points (+) and (-). change in resistor’s temperature and resistance change in amount of heat removed from resistor 31. Design- based on electronic circuit known as a Wheatstone bridge.- circuit consists of an arrangement of four resistors with a fixed current applied to them.- thermal conductivity changes with presence of other components in the mobile phase.- the voltage between points (+) and (-) will be zero as long as the resistances in the different arms of the circuit are properly balanced -one resistor in contact with mobile phase leaving column -another in contact with reference stream of pure mobile phase as solute emerge from column: change in thermal conductivity 32. Considerations - mobile phase must have very different thermal conductivity then solutes being separated. - most compounds separated in GC have thermal conductivity of about 1-4X10-5. - H2 and He are carrier gases with significantly different thermal conductivity values. - H2 reacts with metal oxides present on the resistors, so not used 33. advantages: - truly universal detector ‚ applicable to the detection of any compound in GC - non-destructive ‚ useful for detecting compounds from preparative-scale columns ‚ useful in combination with other types of GC detectorsdisadvantage: - detect mobile phase impurities - sensitive to
  • 8. changes in flow-rates - limit of detection ‚ ~ 10-7 M ‚ much higher then other GC detectors 34. 2.) Flame Ionization Detector (FID) - most common type of GC detector - “universal” detector capable of measuring the presence of almost any organic and many inorganic compoundProcess - measures the production of ions when a solute is burned in a flame. - ions are collected at an electrode to create a current 35. advantages: - universal detector for organics ‚ doesn’t respond to common inorganic compounds - mobile phase impurities not detected - carrier gases not detected - limit of detection: FID is 1000x better than TCD - linear and dynamic range better than TCDdisadvantage: - destructive detector 36. 3.) Nitrogen-Phosphorus Detector (NPD) - used for detecting nitrogen- or phosphorus containing compounds - also known as alkali flame ionization detector or thermionic detectorProcess Alkali Bead - same basic principal as FID - measures production of ions when a solute is burned in a flame - ions are collected at an electrode to create a current - contains a small amount of alkali metal vapor in the flame - enhances the formation of ions from nitrogen- and phosphorus- containing compounds 37. 3.) Nitrogen-Phosphorus Detector (NPD)advantages: - useful for environmental testing ‚ detection of organophosphate pesticides - useful for drug analysis - ‚ determination of amine-containing or basic drugs - Like FID, does not detect common mobile phase impurities or carrier gases - limit of detection: NPD is 500x better than FID in detecting nitrogen and phosphorus- containing compounds - NPD more sensitive to other heterocompounds, such as sulfur-, halogen-, and arsenic- containing moleculesdisadvantage: - destructive detector - NPD is less sensitive to organic compounds compared to FID 38. 4.) Electron Capture Detector (ECD) - radiation-based detector - selective for compounds containing electronegative atoms, such as
  • 9. halogensProcess - based on the capture of electrons by electronegative atoms in a molecule - electrons are produced by ionization of the carrier gas with a radioactive source ‚ 3H or 63Ni - in absence of solute, steady stream of these electrons is produced - electrons go to collector electrode where they produce a current - compounds with electronegative atoms capture electrons, reducing current 39. advantages: - useful for environmental testing ‚ detection of chlorinated pesticides or herbicides ‚ detection of polynuclear aromatic carcinogens ‚ detection of organometallic compounds - selective for halogen- (I, Br, Cl, F), nitro-, and sulfur- containing compounds - detects polynuclear aromatic compounds, anhydrides and conjugated carbonyl compounds 40. Gas Chromatography Application 41. Gas Chromatography can only be used toanalyze volatile components (i.e. compoundsthat can be volatilized at 250oC withoutdecomposition).Non-volatile components have to bederivatized before GC analysis. 42. Retention TimesResponse RT= 4.0 min on SE-30 Hexane GC Retention Time on SE-30Response RT= 4 min on SE-30 Unknown compound GC Retention Time on SE-30 43. TENTATIVE IDENTIFICATION OF UNKNOWN COMPOUNDSResponse Mixture of known compounds Octane Decane 1.6 min = RT Hexane GC Retention Time on Carbowax-20 (min)Response Unknown compound may be Hexane 1.6 min = RT Retention Time on Carbowax-20 (min) 44. SEMI- QUANTITATIVE ANALYSIS OF FATTY ACIDSDetector Response C18 Peak Area (cm2 ) C 16 10 8 C14 6 4 2 0.5 1.0 1.5 2.0 2.5 3.0 Retention Time Sample Concentration (mg/ml) C4 1 The content % of C fatty acids = 14 1 ∗0 0 C + C11 + C 8 6 1 4 = the content % of C fatty acids 14
  • 10. 45. GLC ADVANTAGES1. Very good separation2. Time (analysis is short)3. Small sample is needed - µl4. Good detection system5. Can be used for both qualitative and quantitative analysis 46. DISADVANTAGES OF GAS CHROMATOGRAPHYMaterial has to be volatilized at 250oCwithout decomposition. If not the analyte hasto be derivatized.e.g. Fatty acids → Methyl esters Phenols → Silanized Sugars → Silanized 47. Fatty Acids Methylester O OR C OH + CH3 OH + H2 SO4 R C O CH3 Reflux Volatile in Gas Chromatography O CH2 O C R O CH3 ONa O CH O C R + CH 3 OH 3 R C O CH3 Volatile in Gas O Chromatography CH2 O C R 48. Gas Chromatogram of Methyl Esters of Fatty Acids 49. Derivation of Glucose with Trimethylchlorosilane 6 CH2 OH CH O 3 5 4 1 + 5Cl Si CH OH 3 HO OH CH 3 2 3 OH Glucose Trimethylchlorosilane 6 CH2 O-Si(CH3)3 O 5 + 5HCl 4 O-Si(CH ) 13 3 (CH3)3-Si-O 3 2 O- Si(CH3)3 O-Si(CH3)3 50. Effects of Derivation1. Time consumption2. Side reaction3. Loss of sample