I o n
                                                  Metrohm
                                                  a n a l y s i s




                  743 Rancimat




Determination of the oxidation stability of
• Oils and fats
• Biodiesel
• Food and cosmetics
The 743 Rancimat in brief

    The 743 Rancimat is a modern, PC-controlled analytical instrument for the comfortable determi-
    nation of the oxidation stability of oils and fats, also referred to as Oil Stability Index (OSI). The
    software of the third Rancimat generation comprises automatic data acquisition and evaluation
    as well as a database and permits the convenient management of large amounts of data. The
    unique temperature extrapolation allows a rough estimation of the storage stability of a product.

    Disposable reaction vessels reduce the amount of cleaning of the accessories to an absolute mini-
    mum. This saves both time and money. The accuracy and reproducibility of the measurements are
    also significantly improved. The GLP Set with certified temperature sensor allows to set the tempe-
    rature very accurately and reproducibly, which in turn improves the reproducibility of the results.

    The most important applications

    Oxidation stability of natural     Oxidation stability of biodiesel    Oxidation stability of food
    oils and fats                      (fatty acid methyl esters,          and cosmetics
                                       FAME)




2
The most important advantages
at a glance
• All instrument functions controlled from PC
• Unparalleled reliability and simple operation due to unique accessories
  - Favorably-priced disposable glass parts
  - Robust conductivity cell with integral electrical connections in cover
• Excellent data security by storing all measuring, method and
  instrument data
• Re-evaluation and recalculation of all measuring data
• Database functions such as filtering and sorting
• User administration with freely configurable access rights
• Convenient user interface
• GLP and validation functions
• Estimation of the storage stability of products
• Compact dimensions and low weight
• 2 heating blocks with 8 measuring positions per instrument. Up to
  4 instruments can be connected to one PC.
• Independent start of each channel




                                                                             3
Oxidation stability of natural oils
    and fats




    The determination of the oxidation stability of oils and fats is the classical application for the 743
    Rancimat. The resulting induction time characterizes the resistance of oils and fats to oxidation.
    In addition to the term «induction time», the expression Oil Stability Index (OSI) is in common use.
    The Rancimat method is also called the automated Swift test or accelerated oxidation test and is
    in general use today as an automated version of the previously used and extremely complicated
    Active Oxygen Method (AOM).

    During the measurement a stream of air is passed through the oil or fat sample contained in a
    sealed and heated reaction vessel. This treatment results in oxidation of the oil or fat molecules
    in the sample, with peroxides initially being formed as the primary oxidation products. After some
    time the fatty acids are completely destroyed; the secondary oxidation products formed include
    low-molecular organic acids in addition to other volatile organic compounds. These are transpor-
    ted in the stream of air to a second vessel containing distilled water. The conductivity in this vessel
    is recorded continuously. The organic acids can be detected by the increase in conductivity. The
    time that elapses until these secondary reaction products appear is known as the induction time,
    induction period or Oil Stability Index (OSI).

    The Rancimat method yields standard quality control parameters for the production of oils and fats
    in the food industry or for incoming goods checks in further processing. In addition to oils and fats
    from vegetable sources, the oxidation stability of animal fats such as lard, tallow and fish oil can
    also be checked with the 743 Rancimat as a matter of course.

    Foodstuffs may contain added antioxidants in order to delay the oxidative decomposition of their
    oils and fats. With the 743 Rancimat it is possible to characterize the effectiveness of antioxidants.
    This opens up a further field of application for the instrument.




4
Standards
The Rancimat method is included in various national and internationals standards, such as:
• AOCS Cd 12b-92 (AOCS – American Oil Chemists’ Society)
  Sampling and analysis of commercial fats and oils: Oil Stability Index
• ISO 6886
  Animal and vegetable fats and oils – Determination of oxidation stability (accelerated oxidation test)
• 2.4.28.2-93
  Fat stability test on Autoxidation. CDM, Japan
• Swiss Food Manual (Schweizerisches Lebensmittelbuch), section 7.5.4

Estimating the storage stability
An extension of the previous possibilities is provided by a new function – the so-called temperature
extrapolation. This is an aid for estimating the shelf life of oils and fats. The extrapolation makes use
of the relationship between the measured induction time and the temperature given by van’t Hoff’s
law. Several measurements are made at different temperatures and then extrapolated to the storage
temperature. The value obtained allows a rough estimation of the storage stability of the oil or fat.




                                                                                                            5
Oxidation stability of biodiesel
    (fatty acid methyl esters, FAME)




    In addition to other alternative fuels such as ethanol, methanol or biogas (methane), fatty acid
    methyl esters are increasingly being found on the market; these are also known as biodiesel or
    FAME (fatty acid methyl esters). Fatty acid methyl esters are usually obtained from oil seeds and
    are mainly used in their pure form or mixed with conventional diesel fuel in the transport sector.

    During manufacture the vegetable oil is transesterified with methanol. This produces the methyl
    esters of the fatty acids present in the oil together with glycerol as a byproduct. Fatty acid methyl
    esters are relatively unstable on storage, as like all natural oils and fats they are slowly oxidized
    by atmospheric oxygen. The substances produced in this way could cause motor damage. This is
    why the oxidation stability is an important quality criterion for biodiesel; it is regularly determined
    during the production process. With the 743 Rancimat this determination can be carried out
    quickly and simply. The oxidation process can be delayed by the addition of antioxidants. The
    743 Rancimat can also be used to determine the effectiveness of the antioxidants.

    The use of alternative fuels from renewable vegetable sources has become increasingly widespread
    in recent years, as the life cycle of biogenic fuels is neutral with respect to CO2 release. The carbon
    dioxide released has already been assimilated from the atmosphere during plant growth. In con-
    trast to fossil fuels, biogenic fuels neither contribute to the accumulation of CO2 in the atmos-
    phere nor to global warming. In addition, alternative fuels are biodegradable and thus less envi-
    ronmentally harmful.

    The oxidation stability of fatty acid methyl esters has already been included in various test methods
    as a standard parameter used to define the minimum quality requirements of biodiesel:

                          • EN 14214 «Automotive fuels – Fatty acid methyl esters (FAME) for diesel
                             engines – Requirements and test methods»
                             • EN 14112 «Fat and oil derivatives – Fatty acid methyl esters (FAME) –
                               Determination of oxidation stability (accelerated oxidation test)»




6
7
Oxidation stability of foods and
    cosmetics




    Just like the pure substances, the oils and fats contained in foods are also subject to oxidation, which
    contributes to their spoilage. In such cases the 743 Rancimat can be used to determine the oxidation
    stability of foods containing oils and fats.

    Meltable foods with a high fat content, such as butter, margarine, lard or tallow, can be analyzed
    directly without any further sample preparation. For liquid or semi-liquid foods, such as salad dress-
    ings or mayonnaise, it is better to split the emulsion and analyze the separated fat phase. For solid,
    non-meltable foods it is also necessary to separate off the fat phase. In this case the fat is normally
    cold-extracted with petroleum ether and the isolated fat is then analyzed.

    Examples of typical applications are:
    • Mayonnaise
    • Salad dressings
    • Biscuits
    • Waffles
    • Grain flakes
    • Nuts
    • Chocolate
    • Bacon
    • Sausages




    As well as for determining the oxidation stability of foods, the 743 Rancimat is also used for testing
    cosmetics or cosmetic additives. With meltable products such as lipsticks this can be done without
    any further sample preparation. With creams or salves the fat phase must normally be separated off
    before the analysis.




8
9
The 743 Rancimat in detail




     Two heating blocks with two temperatures
     The 743 Rancimat has two independent heating
     blocks that allow up to eight samples to be ana-
     lyzed at one or two temperatures. Up to four
     Rancimats can be connected to one PC, so that
     the maximum number of samples that can be
     analyzed in parallel can be increased to 32. Each
     measuring position can be started individually.
     As soon as the measurement has been completed
     the measuring position is immediately ready for
     a new sample, which means that the instrument
     can be used to its full capacity.




10
Simple reaction vial handling
Favorably-priced disposable reaction vessels are
used with the 743 Rancimat. Weighing out the
sample and assembling the reaction vessel are
extremely simple and safe. As the vessel does not
need to be expensively cleaned at the end of the
measurement this means that the personnel are
available for other tasks. This reduces the analysis
costs.


Measuring vessels
Easy-to-clean polycarbonate beakers are used for
the automatic conductivity measurement. Glass
beakers are available as an alternative.




Cover with built-in conductivity cell
The conductivity cell is incorporated in the measur-
ing vessel cover. When the cover is placed in posi-
tion the cell is immersed in the water. At the same
time electrical contact is made to the electronics in
the instrument. The use of fragile glass conductivi-
ty electrodes with lengthy connecting cables went
out of fashion a long time ago. The new conducti-
vity cell is also very easy to clean.


Connections
In order to make operation as simple as possible
there are no controls at all on the instrument. All
its functions are controlled from the PC. Apart
from the power switch, the only features you will
find on the instrument are the RS-232 socket for
connection to a PC and a socket for connecting
the Pt-100 temperature sensor.



Validation with the GLP Set
The optionally available GLP Set facilitates the
validation of your 743 Rancimat. It contains a
certified Pt-100 temperature sensor with accesso-
ries that can be used for testing the temperature
regulation of the heating block. A test plug for
checking the conductivity measurement inputs
is also supplied.



Air inlet filter and molecular sieve
The air used for the measurement is aspirated
through a filter that prevents particles from en-
tering the instrument. The molecular sieve remo-
ves water vapor from the aspirated air; as water
contributes to the hydrolytic decomposition of
the fat molecules, it could interfere with the
measurement.



Air supply line
The amount of air that passes through the sam-
ple is automatically controlled via the rotation
rate of the built-in pump according to the
method settings. A separate supply of compres-
sed air is not necessary.




                                                        11
Software functions




     All functions of the 743 Rancimat are controlled by the Rancimat software, which excels by its
     user-friendliness. All the functions are clearly arranged in just a few windows, the operation is
     intuitive.

     Rancimat control
     This is where the measuring parameters can be called up and edited. The instrument functions are
     controlled directly from here, the measurements are also started and shown in the live display
     field. The arrangement of this window corresponds to a view of the instrument from above. This
     means that the assignment of sample information and measuring position is perfectly clear. The
     timer function can be used to automatically switch on the heating blocks before the start of work,
     so that it is no longer necessary to wait while they warm up.

     The functions in a nutshell:
     • Individual start/stop for each position
     • Live display
     • Temperature display
     • Method definition
     • Instrument controls
     • Calculation formulas, automatic result
       transformation to other temperatures
     • Timer function




     Rancimat results
     At the end of each determination the measuring data is stored in a database and can be viewed
     by the user in the results window. Sample information and results are shown in tabular form and
     can be exported in various formats. The measured curves can be shown individually or in groups.
     It is also possible to edit the automatic evaluation and recalculate the results. The temperature
     extrapolation function for estimating the storage stability is available in this section of the soft-
     ware. You can, of course, sort or filter all the displayed data and adapt the display to meet your
     own requirements.

     • Overview table
     • Curve display: individual or multiple plots
     • Re-evaluation: induction time, stability time
       and manual tangent method
     • Report printout
     • Temperature extrapolation (estimation of
       storage stability)
     • Database functions: filtering, sorting
     • Data export




12
GLP functions
By using the built-in monitoring of the validation intervals, nothing can go wrong with your quality
management procedure. The software requests you to carry out the validation in good time. The
wizards for temperature and conductivity validation help you to do this.

• Monitoring the validation intervals
• Wizards for temperature and conductivity
• Electronic validation logbook




User administration
The software includes a flexible and comfortable user administration with login function; this can
be used to define access rights for groups of users and individuals in great detail.

• Definition of user access rights
• Login function




                                                                                                       13
Technical specifications


     Heating blocks
     2 aluminum heating blocks; electrically heated; can be set to different temperatures

     Number of samples
     8 sample (4 measuring positions per heating block)

     Temperature control and measurement
     Temperature range                     50...220 °C, adjustable in 1 °C steps
     Temperature correction                –9.9...+9.9 °C, adjustable in 0.1 °C steps
     Reproducibility of set temperature    Typically better than ±0.2 °C*
     Temperature variation                 Typically <0.1 °C*
     Temperature difference between
     different measuring positions         Typically <0.3 °C*
     Instrument heating-up time from 20 °C
     to 120 °C                             Approx. 45 min (to ±0.1 °C temperature stability)
     to 220 °C                             Approx. 60 min (to ±0.1 °C temperature stability)
     Outer temperature of instrument       <50 °C (at an operating temperature of 220 °C)
     Response temperature of thermal
     protection device                     260 °C
     * When operating temperature has been reached, with inserted reaction vessels with an identical filling and 20 L/h air throughput.


     Air throughput
     Pump                                              Diaphragm pump
     Output range                                      7...25 L/h

     Conductivity measurement
     Measuring range                                   0...400 μS/cm
     Electrodes                                        6.0913.130 conductivity cell with double steel-pin electrode
                                                       built into vessel cover

     Ambient temperature
     Nominal working range                             +5...+40 °C (at 20...80% relative humidity)
     Storage                                           –20...+70 °C
     Transport                                         -40...+70 °C

     Line power
     Voltage                                           2.743.0014 / 2.743.0114: 230 V (220...240 V ±10%)
                                                       2.743.0015 / 2.743.0115: 115 V (100...120 V ±10%)
     Frequency                                         50...60 Hz
     Power consumption                                 <450 VA (depending on heating power)

     Dimensions
     Width                                             405   mm
     Height                                            268   mm (without accessories)
                                                       353   mm (with accessories)
     Depth                                             466   mm

     Weight
     27.6 kg (with accessories)

     PC requirements
     Processor                                         Pentium III with 700 MHz or higher
     Operating system                                  WindowsTM NT, WindowsTM 2000 or WindowsTM XP
     Memory                                            20 MB for program files, 200 MB recommended for
                                                       measuring data storage
     RAM working memory                                128 MB, recommended 256 MB or higher
                                                       (particularly for WindowsTM XP)
     Graphics resolution                               min. 800 x 600, recommended 1024 x 768 or higher
     Interface                                         1 free RS-232C interface (COM)
     Printer                                           All printers supported by WindowsTM




14
Ordering information

743 Rancimat
Instrument for the automatic determination of the oxidation stability of natural oils and fats with
conductometric indication. Including Rancimat software.
2.743.0014       743 Rancimat (230 V, 50...60 Hz) for oils and fats
2.743.0015       743 Rancimat (115 V, 50...60 Hz) for oils and fats

Instrument for the automatic determination of the oxidation stability of biodiesel with conducto-
metric indication. Including Rancimat software.
2.743.0114       743 Rancimat (230 V, 50...60 Hz) for biodiesel
2.743.0115       743 Rancimat (115 V, 50...60 Hz) for biodiesel

Options
6.2059.000      Turntable for 743 Rancimat
6.1428.020      Glass conductivity measuring vessel for 743 Rancimat (8 pcs. are required)
6.2757.000      Exhaust collection tube for 743 Rancimat (with 8 stoppers)
6.1816.010      Silicone tubing, i.d. 6 mm, o.d. 9 mm; length 22 cm (8 pcs. are required)
6.5616.000      GLP Test Set for 743 Rancimat; consisting of
                6.1111.010 Pt-100 temperature sensor 16 cm
                6.1253.000 Measuring insert for heating block
                6.2109.030 Test plug 10 kOhm
6.2301.060      KCl conductivity standard, c(KCl) = 0.1 mol/L, 250 mL

Consumables
6.1429.040      Reaction vessels for 743 Rancimat (117 pcs.)
6.2418.100      Air inlet tube for oil measurements (117 pcs.)
6.1428.100      Polycarbonate conductivity measuring vessel for 743 Rancimat
6.1816.010      Silicone tubing for fat/oil measurements, i.d. 6 mm, o.d. 9 mm; length 22 cm
6.1839.000      Iso-Versinic tubing for biodiesel measurements, i.d. 6 mm, o.d. 9 mm;
                length 22 cm
6.2724.010      Dust filter for 743 Rancimat
6.2811.000      Molecular sieve 250 g




                                                                                                      15
I o n    a n a l y s i s

Metrohm Ltd.
CH-9101 Herisau/Switzerland
Phone +41 71 353 85 85
Fax +41 71 353 89 01
www.metrohm.com
info@metrohm.com




                              Layout by Ecknauer+Schoch ASW, printed in Switzerland by Metrohm Ltd., Herisau
                              Subject to modifications

                              8.743.6013 – 2006-03

Weitere ähnliche Inhalte

Was ist angesagt?

Was ist angesagt? (20)

Soap Manufacturing CPI
Soap Manufacturing CPISoap Manufacturing CPI
Soap Manufacturing CPI
 
Types of cheese by pauly
Types of cheese by paulyTypes of cheese by pauly
Types of cheese by pauly
 
Colloidal systems in food
Colloidal systems in foodColloidal systems in food
Colloidal systems in food
 
Butter Analysis
Butter AnalysisButter Analysis
Butter Analysis
 
Analysis of milk and butter
Analysis of milk and butterAnalysis of milk and butter
Analysis of milk and butter
 
Carbohydrates
CarbohydratesCarbohydrates
Carbohydrates
 
Ice cream ppt
Ice cream pptIce cream ppt
Ice cream ppt
 
Processing of cereals
Processing of cerealsProcessing of cereals
Processing of cereals
 
Flavors - An Overview
Flavors - An OverviewFlavors - An Overview
Flavors - An Overview
 
Ashing techniques
Ashing techniques Ashing techniques
Ashing techniques
 
Fat analysis
Fat analysisFat analysis
Fat analysis
 
CARBOHYDRATES IN FOOD ANALYSIS
CARBOHYDRATES IN FOOD ANALYSISCARBOHYDRATES IN FOOD ANALYSIS
CARBOHYDRATES IN FOOD ANALYSIS
 
Starch manufacturing from Corn and Industrial Application
Starch manufacturing from Corn and Industrial ApplicationStarch manufacturing from Corn and Industrial Application
Starch manufacturing from Corn and Industrial Application
 
Milk powder
Milk powderMilk powder
Milk powder
 
Analysis of fibre
Analysis of fibreAnalysis of fibre
Analysis of fibre
 
Milk
MilkMilk
Milk
 
Chapter 2 Food Evaluation
Chapter 2 Food EvaluationChapter 2 Food Evaluation
Chapter 2 Food Evaluation
 
Acceptance-preference test
Acceptance-preference testAcceptance-preference test
Acceptance-preference test
 
How to make ice cream
How to make ice creamHow to make ice cream
How to make ice cream
 
Lecture 1 : Grading, Testing & Quality of raw milk
Lecture 1 : Grading, Testing & Quality of raw milkLecture 1 : Grading, Testing & Quality of raw milk
Lecture 1 : Grading, Testing & Quality of raw milk
 

Ähnlich wie Rancimat by Metrohm

BR-70701-LC-GC-NIR-ICP-Edible-Oil-Workflows
BR-70701-LC-GC-NIR-ICP-Edible-Oil-WorkflowsBR-70701-LC-GC-NIR-ICP-Edible-Oil-Workflows
BR-70701-LC-GC-NIR-ICP-Edible-Oil-Workflows
dmend129
 
Laboratory role in petroleum industry
Laboratory role  in petroleum industryLaboratory role  in petroleum industry
Laboratory role in petroleum industry
Fouad Agamy
 

Ähnlich wie Rancimat by Metrohm (20)

Instrumental method of analysis Oil and Fat(Unit 7 b)
Instrumental method of analysis Oil and Fat(Unit 7 b)Instrumental method of analysis Oil and Fat(Unit 7 b)
Instrumental method of analysis Oil and Fat(Unit 7 b)
 
Engineers and Consultants Services By SPEC Engineers & Consultants Pvt. Ltd, ...
Engineers and Consultants Services By SPEC Engineers & Consultants Pvt. Ltd, ...Engineers and Consultants Services By SPEC Engineers & Consultants Pvt. Ltd, ...
Engineers and Consultants Services By SPEC Engineers & Consultants Pvt. Ltd, ...
 
Analysis of oil and Fat(Unit 7)
Analysis of oil and Fat(Unit 7)Analysis of oil and Fat(Unit 7)
Analysis of oil and Fat(Unit 7)
 
How to test good the fish oil any time, in factory
How to test good the fish oil any time, in factoryHow to test good the fish oil any time, in factory
How to test good the fish oil any time, in factory
 
Determination of the Thermal Oxidation Stability and the Kinetic Parameters o...
Determination of the Thermal Oxidation Stability and the Kinetic Parameters o...Determination of the Thermal Oxidation Stability and the Kinetic Parameters o...
Determination of the Thermal Oxidation Stability and the Kinetic Parameters o...
 
Biofuels
BiofuelsBiofuels
Biofuels
 
Boortsog
BoortsogBoortsog
Boortsog
 
11.[1 7]evaluation of temperature & solvent effect on peach kernel oil extrac...
11.[1 7]evaluation of temperature & solvent effect on peach kernel oil extrac...11.[1 7]evaluation of temperature & solvent effect on peach kernel oil extrac...
11.[1 7]evaluation of temperature & solvent effect on peach kernel oil extrac...
 
ajfst-3-4-1.pdf
ajfst-3-4-1.pdfajfst-3-4-1.pdf
ajfst-3-4-1.pdf
 
Detection of oxidative status of some edible oil
Detection of oxidative status of some edible oilDetection of oxidative status of some edible oil
Detection of oxidative status of some edible oil
 
BR-70701-LC-GC-NIR-ICP-Edible-Oil-Workflows
BR-70701-LC-GC-NIR-ICP-Edible-Oil-WorkflowsBR-70701-LC-GC-NIR-ICP-Edible-Oil-Workflows
BR-70701-LC-GC-NIR-ICP-Edible-Oil-Workflows
 
Presentación estefanyrosalesinglestécnico
Presentación estefanyrosalesinglestécnicoPresentación estefanyrosalesinglestécnico
Presentación estefanyrosalesinglestécnico
 
Biodiesel production
Biodiesel productionBiodiesel production
Biodiesel production
 
Biodiesel production
Biodiesel productionBiodiesel production
Biodiesel production
 
Equipment used for Fat Analysis.pptx....
Equipment used for Fat Analysis.pptx....Equipment used for Fat Analysis.pptx....
Equipment used for Fat Analysis.pptx....
 
Episode 46 : PRODUCTION OF OLEOCHEMICAL METHYL ESTER FROM RBD PALM KERNEL OIL
Episode 46 :  PRODUCTION OF OLEOCHEMICAL METHYL ESTER FROM  RBD PALM KERNEL OIL Episode 46 :  PRODUCTION OF OLEOCHEMICAL METHYL ESTER FROM  RBD PALM KERNEL OIL
Episode 46 : PRODUCTION OF OLEOCHEMICAL METHYL ESTER FROM RBD PALM KERNEL OIL
 
Biodiesel
Biodiesel Biodiesel
Biodiesel
 
Characterization of biodiesel produced by meth butanolysis of castor oil
Characterization of biodiesel produced by meth butanolysis of castor oilCharacterization of biodiesel produced by meth butanolysis of castor oil
Characterization of biodiesel produced by meth butanolysis of castor oil
 
Production of Biodiesel from Vernonia Galamensis Oil using Ethanol with Alkal...
Production of Biodiesel from Vernonia Galamensis Oil using Ethanol with Alkal...Production of Biodiesel from Vernonia Galamensis Oil using Ethanol with Alkal...
Production of Biodiesel from Vernonia Galamensis Oil using Ethanol with Alkal...
 
Laboratory role in petroleum industry
Laboratory role  in petroleum industryLaboratory role  in petroleum industry
Laboratory role in petroleum industry
 

Mehr von Metrohm India Limited

Metrohm – customized analysis for surface finishing, Process control in elect...
Metrohm – customized analysis for surface finishing, Process control in elect...Metrohm – customized analysis for surface finishing, Process control in elect...
Metrohm – customized analysis for surface finishing, Process control in elect...
Metrohm India Limited
 

Mehr von Metrohm India Limited (20)

Manual titration can be automated with ECO TITRATOR - save time and increase ...
Manual titration can be automated with ECO TITRATOR - save time and increase ...Manual titration can be automated with ECO TITRATOR - save time and increase ...
Manual titration can be automated with ECO TITRATOR - save time and increase ...
 
Global webinar on Advanced solutions for battery material characterization
Global webinar on Advanced solutions for battery material characterizationGlobal webinar on Advanced solutions for battery material characterization
Global webinar on Advanced solutions for battery material characterization
 
IIT Bombay Metrohm Young Chemist Award 2019
IIT Bombay Metrohm Young Chemist Award 2019IIT Bombay Metrohm Young Chemist Award 2019
IIT Bombay Metrohm Young Chemist Award 2019
 
Application note on purified water for mammalian and bacterial cell culture
Application note on purified water for mammalian and bacterial cell cultureApplication note on purified water for mammalian and bacterial cell culture
Application note on purified water for mammalian and bacterial cell culture
 
Brochure power plantanalysis
Brochure power plantanalysisBrochure power plantanalysis
Brochure power plantanalysis
 
875 kf gas analyzer
875 kf gas analyzer875 kf gas analyzer
875 kf gas analyzer
 
Petrochemical Analysis
Petrochemical AnalysisPetrochemical Analysis
Petrochemical Analysis
 
Environmental analysis
Environmental analysisEnvironmental analysis
Environmental analysis
 
Ti touch brochure
Ti touch brochureTi touch brochure
Ti touch brochure
 
Ion Meter from Metrohm
Ion Meter from MetrohmIon Meter from Metrohm
Ion Meter from Metrohm
 
Coulometer for low level water determination in sample matrix
Coulometer for low level water determination in sample matrixCoulometer for low level water determination in sample matrix
Coulometer for low level water determination in sample matrix
 
Metrohm – customized analysis for surface finishing, Process control in elect...
Metrohm – customized analysis for surface finishing, Process control in elect...Metrohm – customized analysis for surface finishing, Process control in elect...
Metrohm – customized analysis for surface finishing, Process control in elect...
 
Surface finishing
Surface finishingSurface finishing
Surface finishing
 
pH / Ion meter from Metrohm
pH / Ion meter from MetrohmpH / Ion meter from Metrohm
pH / Ion meter from Metrohm
 
Ion Chromatography in determining mineral acids in Etching baths
Ion Chromatography in determining mineral acids in Etching bathsIon Chromatography in determining mineral acids in Etching baths
Ion Chromatography in determining mineral acids in Etching baths
 
Battery research
Battery research Battery research
Battery research
 
Biofuels analysis from Metrohm
Biofuels analysis from MetrohmBiofuels analysis from Metrohm
Biofuels analysis from Metrohm
 
Accessories by Metrohm India
Accessories by Metrohm IndiaAccessories by Metrohm India
Accessories by Metrohm India
 
888 Titrando for Titration Applications from Metrohm
888 Titrando for Titration Applications from Metrohm888 Titrando for Titration Applications from Metrohm
888 Titrando for Titration Applications from Metrohm
 
Tiamo software from Metrohm
Tiamo software from MetrohmTiamo software from Metrohm
Tiamo software from Metrohm
 

Rancimat by Metrohm

  • 1. I o n Metrohm a n a l y s i s 743 Rancimat Determination of the oxidation stability of • Oils and fats • Biodiesel • Food and cosmetics
  • 2. The 743 Rancimat in brief The 743 Rancimat is a modern, PC-controlled analytical instrument for the comfortable determi- nation of the oxidation stability of oils and fats, also referred to as Oil Stability Index (OSI). The software of the third Rancimat generation comprises automatic data acquisition and evaluation as well as a database and permits the convenient management of large amounts of data. The unique temperature extrapolation allows a rough estimation of the storage stability of a product. Disposable reaction vessels reduce the amount of cleaning of the accessories to an absolute mini- mum. This saves both time and money. The accuracy and reproducibility of the measurements are also significantly improved. The GLP Set with certified temperature sensor allows to set the tempe- rature very accurately and reproducibly, which in turn improves the reproducibility of the results. The most important applications Oxidation stability of natural Oxidation stability of biodiesel Oxidation stability of food oils and fats (fatty acid methyl esters, and cosmetics FAME) 2
  • 3. The most important advantages at a glance • All instrument functions controlled from PC • Unparalleled reliability and simple operation due to unique accessories - Favorably-priced disposable glass parts - Robust conductivity cell with integral electrical connections in cover • Excellent data security by storing all measuring, method and instrument data • Re-evaluation and recalculation of all measuring data • Database functions such as filtering and sorting • User administration with freely configurable access rights • Convenient user interface • GLP and validation functions • Estimation of the storage stability of products • Compact dimensions and low weight • 2 heating blocks with 8 measuring positions per instrument. Up to 4 instruments can be connected to one PC. • Independent start of each channel 3
  • 4. Oxidation stability of natural oils and fats The determination of the oxidation stability of oils and fats is the classical application for the 743 Rancimat. The resulting induction time characterizes the resistance of oils and fats to oxidation. In addition to the term «induction time», the expression Oil Stability Index (OSI) is in common use. The Rancimat method is also called the automated Swift test or accelerated oxidation test and is in general use today as an automated version of the previously used and extremely complicated Active Oxygen Method (AOM). During the measurement a stream of air is passed through the oil or fat sample contained in a sealed and heated reaction vessel. This treatment results in oxidation of the oil or fat molecules in the sample, with peroxides initially being formed as the primary oxidation products. After some time the fatty acids are completely destroyed; the secondary oxidation products formed include low-molecular organic acids in addition to other volatile organic compounds. These are transpor- ted in the stream of air to a second vessel containing distilled water. The conductivity in this vessel is recorded continuously. The organic acids can be detected by the increase in conductivity. The time that elapses until these secondary reaction products appear is known as the induction time, induction period or Oil Stability Index (OSI). The Rancimat method yields standard quality control parameters for the production of oils and fats in the food industry or for incoming goods checks in further processing. In addition to oils and fats from vegetable sources, the oxidation stability of animal fats such as lard, tallow and fish oil can also be checked with the 743 Rancimat as a matter of course. Foodstuffs may contain added antioxidants in order to delay the oxidative decomposition of their oils and fats. With the 743 Rancimat it is possible to characterize the effectiveness of antioxidants. This opens up a further field of application for the instrument. 4
  • 5. Standards The Rancimat method is included in various national and internationals standards, such as: • AOCS Cd 12b-92 (AOCS – American Oil Chemists’ Society) Sampling and analysis of commercial fats and oils: Oil Stability Index • ISO 6886 Animal and vegetable fats and oils – Determination of oxidation stability (accelerated oxidation test) • 2.4.28.2-93 Fat stability test on Autoxidation. CDM, Japan • Swiss Food Manual (Schweizerisches Lebensmittelbuch), section 7.5.4 Estimating the storage stability An extension of the previous possibilities is provided by a new function – the so-called temperature extrapolation. This is an aid for estimating the shelf life of oils and fats. The extrapolation makes use of the relationship between the measured induction time and the temperature given by van’t Hoff’s law. Several measurements are made at different temperatures and then extrapolated to the storage temperature. The value obtained allows a rough estimation of the storage stability of the oil or fat. 5
  • 6. Oxidation stability of biodiesel (fatty acid methyl esters, FAME) In addition to other alternative fuels such as ethanol, methanol or biogas (methane), fatty acid methyl esters are increasingly being found on the market; these are also known as biodiesel or FAME (fatty acid methyl esters). Fatty acid methyl esters are usually obtained from oil seeds and are mainly used in their pure form or mixed with conventional diesel fuel in the transport sector. During manufacture the vegetable oil is transesterified with methanol. This produces the methyl esters of the fatty acids present in the oil together with glycerol as a byproduct. Fatty acid methyl esters are relatively unstable on storage, as like all natural oils and fats they are slowly oxidized by atmospheric oxygen. The substances produced in this way could cause motor damage. This is why the oxidation stability is an important quality criterion for biodiesel; it is regularly determined during the production process. With the 743 Rancimat this determination can be carried out quickly and simply. The oxidation process can be delayed by the addition of antioxidants. The 743 Rancimat can also be used to determine the effectiveness of the antioxidants. The use of alternative fuels from renewable vegetable sources has become increasingly widespread in recent years, as the life cycle of biogenic fuels is neutral with respect to CO2 release. The carbon dioxide released has already been assimilated from the atmosphere during plant growth. In con- trast to fossil fuels, biogenic fuels neither contribute to the accumulation of CO2 in the atmos- phere nor to global warming. In addition, alternative fuels are biodegradable and thus less envi- ronmentally harmful. The oxidation stability of fatty acid methyl esters has already been included in various test methods as a standard parameter used to define the minimum quality requirements of biodiesel: • EN 14214 «Automotive fuels – Fatty acid methyl esters (FAME) for diesel engines – Requirements and test methods» • EN 14112 «Fat and oil derivatives – Fatty acid methyl esters (FAME) – Determination of oxidation stability (accelerated oxidation test)» 6
  • 7. 7
  • 8. Oxidation stability of foods and cosmetics Just like the pure substances, the oils and fats contained in foods are also subject to oxidation, which contributes to their spoilage. In such cases the 743 Rancimat can be used to determine the oxidation stability of foods containing oils and fats. Meltable foods with a high fat content, such as butter, margarine, lard or tallow, can be analyzed directly without any further sample preparation. For liquid or semi-liquid foods, such as salad dress- ings or mayonnaise, it is better to split the emulsion and analyze the separated fat phase. For solid, non-meltable foods it is also necessary to separate off the fat phase. In this case the fat is normally cold-extracted with petroleum ether and the isolated fat is then analyzed. Examples of typical applications are: • Mayonnaise • Salad dressings • Biscuits • Waffles • Grain flakes • Nuts • Chocolate • Bacon • Sausages As well as for determining the oxidation stability of foods, the 743 Rancimat is also used for testing cosmetics or cosmetic additives. With meltable products such as lipsticks this can be done without any further sample preparation. With creams or salves the fat phase must normally be separated off before the analysis. 8
  • 9. 9
  • 10. The 743 Rancimat in detail Two heating blocks with two temperatures The 743 Rancimat has two independent heating blocks that allow up to eight samples to be ana- lyzed at one or two temperatures. Up to four Rancimats can be connected to one PC, so that the maximum number of samples that can be analyzed in parallel can be increased to 32. Each measuring position can be started individually. As soon as the measurement has been completed the measuring position is immediately ready for a new sample, which means that the instrument can be used to its full capacity. 10
  • 11. Simple reaction vial handling Favorably-priced disposable reaction vessels are used with the 743 Rancimat. Weighing out the sample and assembling the reaction vessel are extremely simple and safe. As the vessel does not need to be expensively cleaned at the end of the measurement this means that the personnel are available for other tasks. This reduces the analysis costs. Measuring vessels Easy-to-clean polycarbonate beakers are used for the automatic conductivity measurement. Glass beakers are available as an alternative. Cover with built-in conductivity cell The conductivity cell is incorporated in the measur- ing vessel cover. When the cover is placed in posi- tion the cell is immersed in the water. At the same time electrical contact is made to the electronics in the instrument. The use of fragile glass conductivi- ty electrodes with lengthy connecting cables went out of fashion a long time ago. The new conducti- vity cell is also very easy to clean. Connections In order to make operation as simple as possible there are no controls at all on the instrument. All its functions are controlled from the PC. Apart from the power switch, the only features you will find on the instrument are the RS-232 socket for connection to a PC and a socket for connecting the Pt-100 temperature sensor. Validation with the GLP Set The optionally available GLP Set facilitates the validation of your 743 Rancimat. It contains a certified Pt-100 temperature sensor with accesso- ries that can be used for testing the temperature regulation of the heating block. A test plug for checking the conductivity measurement inputs is also supplied. Air inlet filter and molecular sieve The air used for the measurement is aspirated through a filter that prevents particles from en- tering the instrument. The molecular sieve remo- ves water vapor from the aspirated air; as water contributes to the hydrolytic decomposition of the fat molecules, it could interfere with the measurement. Air supply line The amount of air that passes through the sam- ple is automatically controlled via the rotation rate of the built-in pump according to the method settings. A separate supply of compres- sed air is not necessary. 11
  • 12. Software functions All functions of the 743 Rancimat are controlled by the Rancimat software, which excels by its user-friendliness. All the functions are clearly arranged in just a few windows, the operation is intuitive. Rancimat control This is where the measuring parameters can be called up and edited. The instrument functions are controlled directly from here, the measurements are also started and shown in the live display field. The arrangement of this window corresponds to a view of the instrument from above. This means that the assignment of sample information and measuring position is perfectly clear. The timer function can be used to automatically switch on the heating blocks before the start of work, so that it is no longer necessary to wait while they warm up. The functions in a nutshell: • Individual start/stop for each position • Live display • Temperature display • Method definition • Instrument controls • Calculation formulas, automatic result transformation to other temperatures • Timer function Rancimat results At the end of each determination the measuring data is stored in a database and can be viewed by the user in the results window. Sample information and results are shown in tabular form and can be exported in various formats. The measured curves can be shown individually or in groups. It is also possible to edit the automatic evaluation and recalculate the results. The temperature extrapolation function for estimating the storage stability is available in this section of the soft- ware. You can, of course, sort or filter all the displayed data and adapt the display to meet your own requirements. • Overview table • Curve display: individual or multiple plots • Re-evaluation: induction time, stability time and manual tangent method • Report printout • Temperature extrapolation (estimation of storage stability) • Database functions: filtering, sorting • Data export 12
  • 13. GLP functions By using the built-in monitoring of the validation intervals, nothing can go wrong with your quality management procedure. The software requests you to carry out the validation in good time. The wizards for temperature and conductivity validation help you to do this. • Monitoring the validation intervals • Wizards for temperature and conductivity • Electronic validation logbook User administration The software includes a flexible and comfortable user administration with login function; this can be used to define access rights for groups of users and individuals in great detail. • Definition of user access rights • Login function 13
  • 14. Technical specifications Heating blocks 2 aluminum heating blocks; electrically heated; can be set to different temperatures Number of samples 8 sample (4 measuring positions per heating block) Temperature control and measurement Temperature range 50...220 °C, adjustable in 1 °C steps Temperature correction –9.9...+9.9 °C, adjustable in 0.1 °C steps Reproducibility of set temperature Typically better than ±0.2 °C* Temperature variation Typically <0.1 °C* Temperature difference between different measuring positions Typically <0.3 °C* Instrument heating-up time from 20 °C to 120 °C Approx. 45 min (to ±0.1 °C temperature stability) to 220 °C Approx. 60 min (to ±0.1 °C temperature stability) Outer temperature of instrument <50 °C (at an operating temperature of 220 °C) Response temperature of thermal protection device 260 °C * When operating temperature has been reached, with inserted reaction vessels with an identical filling and 20 L/h air throughput. Air throughput Pump Diaphragm pump Output range 7...25 L/h Conductivity measurement Measuring range 0...400 μS/cm Electrodes 6.0913.130 conductivity cell with double steel-pin electrode built into vessel cover Ambient temperature Nominal working range +5...+40 °C (at 20...80% relative humidity) Storage –20...+70 °C Transport -40...+70 °C Line power Voltage 2.743.0014 / 2.743.0114: 230 V (220...240 V ±10%) 2.743.0015 / 2.743.0115: 115 V (100...120 V ±10%) Frequency 50...60 Hz Power consumption <450 VA (depending on heating power) Dimensions Width 405 mm Height 268 mm (without accessories) 353 mm (with accessories) Depth 466 mm Weight 27.6 kg (with accessories) PC requirements Processor Pentium III with 700 MHz or higher Operating system WindowsTM NT, WindowsTM 2000 or WindowsTM XP Memory 20 MB for program files, 200 MB recommended for measuring data storage RAM working memory 128 MB, recommended 256 MB or higher (particularly for WindowsTM XP) Graphics resolution min. 800 x 600, recommended 1024 x 768 or higher Interface 1 free RS-232C interface (COM) Printer All printers supported by WindowsTM 14
  • 15. Ordering information 743 Rancimat Instrument for the automatic determination of the oxidation stability of natural oils and fats with conductometric indication. Including Rancimat software. 2.743.0014 743 Rancimat (230 V, 50...60 Hz) for oils and fats 2.743.0015 743 Rancimat (115 V, 50...60 Hz) for oils and fats Instrument for the automatic determination of the oxidation stability of biodiesel with conducto- metric indication. Including Rancimat software. 2.743.0114 743 Rancimat (230 V, 50...60 Hz) for biodiesel 2.743.0115 743 Rancimat (115 V, 50...60 Hz) for biodiesel Options 6.2059.000 Turntable for 743 Rancimat 6.1428.020 Glass conductivity measuring vessel for 743 Rancimat (8 pcs. are required) 6.2757.000 Exhaust collection tube for 743 Rancimat (with 8 stoppers) 6.1816.010 Silicone tubing, i.d. 6 mm, o.d. 9 mm; length 22 cm (8 pcs. are required) 6.5616.000 GLP Test Set for 743 Rancimat; consisting of 6.1111.010 Pt-100 temperature sensor 16 cm 6.1253.000 Measuring insert for heating block 6.2109.030 Test plug 10 kOhm 6.2301.060 KCl conductivity standard, c(KCl) = 0.1 mol/L, 250 mL Consumables 6.1429.040 Reaction vessels for 743 Rancimat (117 pcs.) 6.2418.100 Air inlet tube for oil measurements (117 pcs.) 6.1428.100 Polycarbonate conductivity measuring vessel for 743 Rancimat 6.1816.010 Silicone tubing for fat/oil measurements, i.d. 6 mm, o.d. 9 mm; length 22 cm 6.1839.000 Iso-Versinic tubing for biodiesel measurements, i.d. 6 mm, o.d. 9 mm; length 22 cm 6.2724.010 Dust filter for 743 Rancimat 6.2811.000 Molecular sieve 250 g 15
  • 16. I o n a n a l y s i s Metrohm Ltd. CH-9101 Herisau/Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 www.metrohm.com info@metrohm.com Layout by Ecknauer+Schoch ASW, printed in Switzerland by Metrohm Ltd., Herisau Subject to modifications 8.743.6013 – 2006-03