3. SAFETY DEVICES IN
ANAESTHESIA MACHINE
Pneumatic components
High Pressure Systems
Low Pressure Systems
Pin Index System
Colour Coding
Pressure Regulator
Intermediate Pressure System
5. • Mechanical Linkage
• Electronic linkage
• Unidirectional Check Valve
• Pressure Relief Device
• Common Gas Outlets
• Different Tubes for different Gas
8. PNEUMATIC COMPONENTS
• The anaesthesia machine has
pneumatic components that
ensures safety foe the patients as
follows:
• 1) High Pressure System
• 2) Intermediate Pressure System
• 3) Low Pressure System.
10. PIN INDEX SYSTEM
• Prevents wrong cylinder attachment. -
Presence & Integrity of pins should be
checked.
• Filter :- Prevents dust & foreign
particles from entering.
• Unidirectional valve :- - Prevents
backflow into the cylinder when
cylinder pressure falls below the
pressure inside the machine.
11. COLOUR CODING
• Colour coding of cylinders used for
same gas in machine. caps, hoses,
flow control knobs, Pressure gauges
ensures patient safety
12. PRESSURE REGULATOR
• There are separate regulator for
each gas, and cylinder output
reduced to 45 psig Pipeline output
reduced to 50-55 psig
13. INTERMEDIATE PRESSURE SYSTEM
• The intermediate pressure system
comprises of pipeline inlet
connections, Fitted with threaded
non-interchangeable DISS and
colour coded for gases.
14. PIPE LINE PRESSURE GAUGE
• Is Colour coded for each gas.
• Is Present on the pipeline side of the
unidirectional valve ( If present on the
machine side – will show cylinder
pressures)
• Thus wont read 0 if pipeline supply
fails.
• The piping inside the machine is able to
withstand 4 times the intended service
pressure.
15. OXYGEN FAILURE SAFETY DEVICE
• With respect to Oxygen Fail Safe Valve
& N2O Shut off Valve ASTM Stds require
that, whenever O2 supply pressure falls
below manufacturer specified min
pressure, the delivered O2 conc at the
CGO should not fall below 19%. • If O2
pressure falls – N2O pressure falls
proportionately. • If O2 supply fails –
N2O is shut off.
16. CHECKING O2 SAFE VALVE
• Switch on the flows of N2O & O2 in flow
meters.
• Disconnect O2 pipeline/ Close the cylinder
valve.
• N2O flow indicators falls to 0 even before the
O2 indicator falls
• O2 Failure Safety Alarm :- Alarm will set of
when O2 pressures fall below specified
minimum. (Usually 30 psig) Not possible to
silence unless the pressure is restored
17. GAS SELECTOR SWITCH
• Present in some machines prevents
Air & N2O from being used together.
18. STAGE PRESSURE REGULATOR
• Reduces pressures further to : O2 –
14 psi N2O – 26 psi
• Provides a stable pressure to flow
meters i.c.o. pipeline pressure
fluctuation.
19. OXYGEN FLUSH
• Delivers high flows (35-75 L/min) at
intermediate pressures.
•Can be activated even if the master switch is
off.
• Inappropriate & long activation –
Barotrauma & Awareness to dilution is
designed as “Single purpose, self closing
device & recessed”
• Usually recessed/ placed in a collar – to
prevent accidental activation
20. FLOW CONTROL KNOB
• There is only 1 flow control for each
gas.
• Is colour coded for each gas.
• Appears in different size,
consistency & position. eg :- O2 –
larger & fluted & on the right side/
downstream.
21. LOW PRESSURE SYSTEM
• Flow meter assembly : Stops @ full ON &
OFF pos. Stop at full on pos – at top ↓ a)
Prevents escape of indicator from tube to
common manifold.
• Prevents blockade of common manifold by
the stop.
• Prevents indicator from ascending to the top
where it cant be seen clearly. Stop collar to
flow meter knob – prevents damage to seat
by pin when tightened.
22. DIFFERENT TUBES FOR
DIFFERENT GAS
• Tubes vary in diameter & calibration.
• Tubes made leak proof by “O-rings”
(Neoprene washers) at both the ends.
• Tubes are coated with anti-static coating
on both surfaces ↓ prevents indicator
from sticking.
• Made of glass – transparent material ↓
indicator clearly visible throughout the
tube.
23. HYPOXIA PREVENTION DEVICE
• Mandatory Min O2 Flow (MMOF) A
MMOF – 50-250 ml/min – must as soon
as the master switch is turned on and O2
flow – can’t be reduced below that.
Some machines alarms when O2 flow
falls below MMOF.
• Min O2 Ratio : ASTM Stds require that;
A Min O2 ratio/ Min O2 conc of 21% in
the FGF at CGO. 2 types of linkages to
achieve this.
24. MECHANICAL LINKAGE
• Mechanical Linking of O2 & N2O. O2
– 14 tooth sprocket, & N2O – 29
tooth sprocket Connected by metal
chain.
• Both knobs turn together – to
maintain min 25% O2 conc.
25. ELECTRONIC LINKAGE
• An electronic proportionating valve ↓
maintains min 25% O2 conc in FGF.
• A computer continuously calculates the
N2O flow for the given O2 flow ↓ If
higher N2O flow & O2 conc <25% ↓
Electronic proportionating valve
automatically reduces N2O flow
• Alarms is set off when O2:N2O falls
below a preset value.
26. UNIDIRECTIONAL CHECK VALVE
• Location :- Downstream to vaporisers
Upstream to the point where O2 flush meets
FGF.
• Positive pressures generated within the ckt.s
(eg – during controlled ventilation, when O2
flush is used) ↓ if transferred back to the
machine ↓ dilution of volatile agents &
interfere with the outflow through flow
meters & vaporisers.
27. • Prevents +ve pressures within the
ckt.s to be transferred to the
machine.
• Disadvantage :- Leaks upstream to
the valve cant be detected.
28. PRESSURE RELIEF DEVICES
• Location :- Downstream to the
unidirectional valve Upstream to the
point where O2 flush joins FGF.
• To protect machine from high pressures.
Opens & vents gases to atmosphere
when preset pressure is exceeded.
• Disadvantage :- Limits ability of the
machine to provide adequate pressures
↓ Jet ventilation cant be provided
29. COMMON GAS OUTLETS
• ASTM stds require that;- It should be
difficult to accidently disconnect the
delivery hose from the CGO. ↓ Thus
fitted with std 15mm female slip joint
fitting with 22mm coaxial male
connector.
• Some machines have 2 CGOs ↓ ASTM
stds require that;- Only 1 CGO must be
functional at a time.