3. • Defibrillators are devices that
restore a normal heartbeat by
sending an electric pulse or shock
to the heart
• Defibrillators can also restore the
heart's beating if the heart
suddenly stops
4. • Some times defibrillators are used
to prevent or correct an
arrhythmia (a heartbeat that is
uneven or that is too slow or too
fast. The parameters affected are
rate & rhythm of the heart beat )
8. INDICATION FOR DEFIBRILLATION
• Defibrillation is a treatment for life-
threatening cardiac dysrhythmias,
specifically ventricular fibrillation
and non-perfusing ventricular
tachycardia.
• A defibrillator delivers a dose of
electric current to the heart and
restores the heart’s pulsation
10. MECHANISM OF FUNCTIONING
• A defibrillator delivers a dose
of electric current (often called
a counter-shock) to the heart.
• This process depolarizes a large
amount of the heart muscle, ending
the dysrhythmia.
11. • Subsequently, the body's natural
pacemaker in the sinoatrial node of the
heart is able to re-establish normal sinus
rhythm.
• A heart which is in asystole (flat line)
cannot be restarted by a defibrillator,
but would be treated
by cardiopulmonary resuscitation (CPR).
12. TYPES OF DEFIBRILLATOR
• Defibrillators can be external,
transvenous,or implanted (impla
ntable cardioverter-defibrillator),
depending on the type of device
used or needed
13. AUTOMATED EXTERNAL
DEFIBRILLATORS
• Some external units, known
as automated external
defibrillators (AEDs), automate the
diagnosis of treatable rhythms,
meaning that lay responders or by
standers to use them successfully
with little or no training.
15. CONTRA INDICATION FOR
DEFIBRILLATION
• Defibrillation is also not indicated if
the patient is conscious or has a
pulse.
• Improperly given electrical shocks
can cause dangerous dysrhythmias,
such as ventricular fibrillation.
16. MANUAL EXTERNAL DEFIBRILLATOR
• Manual external defibrillators require
the expertise of a healthcare
professional
• They are used in conjunction with
an electrocardiogram, which can be
separate or built-in
17. • A healthcare provider first diagnose
the cardiac rhythm and then
manually determine the voltage and
timing for the electrical shock.
These units are primarily found
in hospitals and on
some ambulances
18. MANUAL INTERNAL DEFIBRILLATOR
Manual internal defibrillators deliver
the shock through paddles placed
directly on the heart.
• They are mostly used in the operating
room and, in rare circumstances, in the
emergency room during an open heart
procedure.
19. AUTOMATED EXTERNAL
DEFIBRILLATOR
• An AED installed outside is designed
for public use. Automated external
defibrillators are designed for use
by untrained personnel
20. • AEDs contain technology for analysis of
heart rhythms
• As a result, it does not require a trained
health provider to determine whether
or not a rhythm is shockable. AEDs have
improved outcomes for sudden out-of-
hospital cardiac arrests.
21. • AEDs have set voltages and does not
allow the operator to vary voltage
according to need.
• AEDs may also delay delivery of
effective CPR.
• AED require the stopping of chest
compressions and rescue breathing for
diagnosis of heart rhythm
22. • AEDs have been incorporated into the
algorithm for basic life support (BLS).
Many first responders, such as fire
fighters, policemen, and security guards,
are equipped with them.
• AEDs can be fully automatic or semi-
automatic.
23. SEMI AUTOMATIC AED
• A semi-automatic AED automatically
diagnoses heart rhythms and
determines if a shock is necessary.
• If a shock is advised, the user must then
push a button to administer the shock.
24. FULLY AUTOMATED AED
• A fully automated AED automatically
diagnoses the heart rhythm and advises
the user to stand back while the shock is
automatically given.
• Some types of AEDs come with
advanced features, such as a manual
override or an ECG display.
26. • AICD constantly monitor the
patient's heart rhythm, and
automatically administer shocks for
various life-threatening
arrhythmias, according to the
device's programming.
27. • Many modern AICD devices can
distinguish between ventricular
fibrillation, ventricular tachycardia,
and more benign arrhythmias
like supraventricular
tachycardia and atrial fibrillation.
28. • Some AICD devices may attempt
overdrive pacing prior to
synchronised cardioversion. When
the life-threatening arrhythmia is
ventricular fibrillation, the device is
programmed to proceed
immediately to an unsynchronized
shock.
30. • The unit monitors the patient 24 hours a
day and can automatically deliver a
biphasic shock if VF or VT is detected.
• This device is mainly indicated in
patients who are not immediate
candidates for ICDs
31. INTERNAL DEFIBRILLATOR
• Internal defibrillator is used to
defibrillate the heart during or after
cardiac surgery such as a heart bypass.
• The electrodes consist of round metal
plates that come in direct contact with
the myocardium.
32. THE PROCEDURE
• The connection between the
defibrillator and the patient consists of
a pair of electrodes, each provided
with electrically conductive gel in order
to ensure a good connection and to
minimize electrical resistance, also
called chest impedance (despite the DC
discharge) which would burn the
patient.
33. • Gel may be either wet (similar in
consistency to surgical lubricant) or
solid (similar to gummi candy).
• Solid-gel is more convenient, because
there is no need to clean the used gel
off the person's skin after
defibrillation.
34. • The use of solid-gel presents a
higher risk of burns during
defibrillation, since wet-gel
electrodes more evenly conduct
electricity into the body.
35. PADDLE ELECTRODES
An AED with electrodes attached.
• The most well-known type of electrode
(widely depicted in films and television)
is the traditional metal paddle with an
insulated (usually plastic) handle.
36. • This type must be held in place on the
patient's skin with approximately 25 lbs
of force while a shock or a series of
shocks is delivered. Paddles offer a few
advantages over self-adhesive pads.
• Many hospitals in the use of paddles,
with disposable gel pads attached due
to the inherent speed with which these
electrodes can be placed and used.
37. • This is critical during cardiac arrest,
as each second
of nonperfusion means tissue loss.
Modern paddles allow for
monitoring (electrocardiography)
38. • Paddles are reusable, being cleaned
after use and stored for the next
patient.
• Gel is therefore not pre applied, and
must be added before these paddles are
used on the patient. Paddles are
generally only found on manual external
units.
39. SELF-ADHESIVE ELECTRODES
Newer types of resuscitation electrodes
are designed as an adhesive pad, which
includes either solid or wet gel.
These are peeled off their backing and
applied to the patient's chest when
deemed necessary, much the same as
any other sticker.
40. • The electrodes are then connected
to a defibrillator.
• If defibrillation is required, the
machine is charged, and the shock is
delivered, without any need to
apply any additional gel or to
retrieve and place any paddles.
41. • Most adhesive electrodes are
designed to be used not only for
defibrillation, but also
for transcutaneous pacing and
synchronized electrical cardioversion
42. • These adhesive pads are found on
most automated and semi-
automated units and are replacing
paddles entirely in non-hospital
settings.
• In hospital, for cases where cardiac
arrest is likely to occur , self-
adhesive pads may be placed
prophylactically.
43. • Pads also offer an advantage to the
untrained user, and to medics
working in the sub-optimal
conditions of the field.
• Pads do not require extra leads to
be attached for monitoring, and
they do not require any force to be
applied as the shock is delivered.
44. • The adhesive electrodes minimize
the risk of the operator coming into
physical (and thus electrical) contact
with the patient as the shock is
delivered by allowing the operator
to be up to several feet away.
45. • The risk of electrical shock to others
remains unchanged, as does that of
shock due to operator misuse. Self-
adhesive electrodes are single-use only.
• They may be used for multiple shocks in
a single course of treatment, but are
replaced if (or in case) the patient
recovers then re enters cardiac arrest.
46. PLACEMENT OF PADDLES
PLACEMENT OF ELECTRODES FOR
DEFIBRILLATION
• Resuscitation electrodes are placed
according to one of two scheme as
follows.
47. ANTERIOR POSTERIOR SCHEME
• The anterior-posterior scheme is the
preferred scheme for long-term
electrode placement. Where one
electrode is placed over the left
precordium (the lower part of the
chest, in front of the heart).
48. The other electrode is placed on the
back, behind the heart in the region
between the scapula.
This placement is preferred because
it is best for non-invasive pacing.
49. ANTERIOR-APEX SCHEME
• The anterior-apex scheme can be
used when the anterior-posterior
scheme is inconvenient or
unnecessary.
50. • In this scheme, the anterior electrode is
placed on the right, below the clavicle.
The apex electrode is applied to the left
side of the patient, just below and to
the left of the pectoral muscle.
• This scheme works well for
defibrillation and cardioversion, as well
as for monitoring an ECG.
52. • Researchers have created a software
modeling system capable of mapping an
individual's chest and determining the best
position for an external or internal cardiac
defibrillator.
53. MECHANISM OF ACTION
• The exact mechanism of defibrillation is
not well understood.
• One theory is that successful
defibrillation affects most of the heart,
resulting in insufficient remaining heart
muscle to continue the arrhythmia.
54. REFERENCE
• Ong, ME; Lim, S; Venkataraman, A (2016). "Defibrillation and cardioversion". In Tintinalli
JE; et al. (eds.). Tintinalli's Emergency Medicine: A Comprehensive Study Guide,
8e. McGraw-Hill (New York, NY).
• ^ Jump up to:a b c d e f Kerber, RE (2011). "Chapter 46. Indications and Techniques of
Electrical Defibrillation and Cardioversion". In Fuster V; Walsh RA; Harrington RA
(eds.). Hurst's The Heart (13th ed.). New York, NY: McGraw-Hill – via AccessMedicine.
• ^ Werman, Howard A.; Karren, K; Mistovich, Joseph (2014). "Automated External
Defibrillation and Cardiopulmonary Resuscitation". In Werman A. Howard; Mistovich J;
Karren K (eds.). Prehospital Emergency Care, 10e. Pearson Education, Inc. p. 425.
• ^ Author:Bradley P Knight, MD, FACCSection Editor:Richard L Page, MDDeputy Editor:Brian
C Downey, MD, FACC. "Basic principles and technique of external electrical cardioversion
and defibrillation". UpToDate. Retrieved 2019-07-24.
• ^ Hoskins, MH; De Lurgio, DB (2012). "Chapter 129. Pacemakers, Defibrillators, and Cardiac
Resynchronization Devices in Hospital Medicine". In McKean SC; Ross JJ; Dressler DD;
Brotman DJ; Ginsberg JS (eds.). Principles and Practice of Hospital Medicine. New York, NY:
McGraw-Hill – via Access Medicine.
• ^ Jump up to:a b c Venegas-Borsellino, C; Bangar, MD (2016). "CPR and ACLS Updates". In
Orpello JM; et al. (eds.). Critical Care. McGraw-Hill