NIGHT DREAN Genuine Call girls in Vasant Vihar Delhi | 83778 77756
RBFM Clamp-on Ultrsonic Flow Meter/ Heat Meter Catalog
1. www.rbflowmeters.com
ULTRASONIC FLOW METER/HEAT METER Catalog
1
Catalog
1. Ultrasonic Flow meters …………………………………………… 2
Fixed Type ……………………………………………….………………3
Handheld/Portable Type……………………….…………………….5
2. Ultrasonic Heat meters …………………………………………….9
3. Flanged Type Ultrasonic Flow meter …………………………10
4. Ultrasonic Thickness gauge ………………………….………. 14
Fixed Type Handheld/Portable
2. www.rbflowmeters.com
ULTRASONIC FLOW METER/HEAT METER Catalog
2
RBFM Series Ultrasonic Flow Meter
Principle of Measurement
The RBFM transit-time flow meter utilizes two transducers that function as both ultrasonic
transmitters and receivers. The transducers are clamped on the outside of a closed pipe at a specific
distance from each other. The transducers can be mounted in V-method where the sound
transverses the pipe twice, or W-method where the sound transverses the pipe four times, or in
Z-method where the transducers are mounted on opposite sides of the pipe and the sound crosses
the pipe once. This selection of the mounting method depends on pipe and liquid characteristics.
The flow meter operates by alternately transmitting and receiving a frequency modulated burst of
sound energy between the two transducers and measuring the transit time that it takes for sound to
travel between the two transducers. The difference in the transit time measured is directly and
exactly related to the velocity of the liquid in the pipe, show as follows:
Figure 1
Where:
M is the travel times of the ultrasonic beam
D is the pipe diameter
T up is the time for the beam from upstream transducer to the downstream one
T down is the time for the beam from downstream transducer to the upstream one
ΔT=Tup –Tdown
3. www.rbflowmeters.com
ULTRASONIC FLOW METER/HEAT METER Catalog
3
Fixed Type: the RBFM-F is a fixed-mount, low cost transit-time ultrasonic flow meter
with clamp-on transducers for non-invasive liquid flow rate measurement. Our
microprocessor-based, user friendly field programmable creates no interruption of the process flow
and has low installation costs.
This versatile series of flow meters is available in a
variety of configurations. Available transducer
types include clamp-on, in-line, and insertion.
Wall-mount, panel-mount, pipe-mounted and
compact din-rail mounted display types are also
offered. Additional available features include pulse,
0-20mA & 4-20mA outputs, RS-232
communication, user-selectable engineering units,
and a data logger function.
1. Feature
Fast response time - less than 1 second.
2 line LCD display with flow rate, tantalizer & signal condition.
Stores up 64 Day/months tantalizer daily value.
Batch control function.
Wide range velocities of ±0.1~ ± 12m/s (0.03 ~ ± 105 feet/sec.)
Transducers for pipe sizes ranging from13 to 6000 mm (½" to 240").
High accuracy of ± 1% of reading.
Clamp-on transducers feature integral magnet for installation on metal pipe without use of
a mounting belt
Data logger function includes date, totalized, & signal condition diagnostic
2. Specification
Parts Picture Range
Transmitter
S type Pipe size (15 mm ~ 100 mm) ½" ~ 4"
Temp. SN -30 ~ 90℃
SH -30~160℃
Size(mm) SN: 45×25×23
4. www.rbflowmeters.com
ULTRASONIC FLOW METER/HEAT METER Catalog
4
Transducer
Standard Type
High Temp. type
SH 45×25×28
M Type Pipe size 50 mm ~ 700 mm (2" ~ 28" )
Temp. MN -30 ~ 90℃
MH -30~160℃
Size(mm) MN: 64×35×32
MH 64×39×44
L Type Pipe size 300~ 6000(12" ~ 240" )
Temp. LN -30 ~ 90℃
LH -30~160℃
Size(mm) LN 98×49×45
LH 97×54×53
Insertion
Type
Pipe size 80~6000mm
Temp. RCN -40 ~ 160℃
RCL(Extended) -40 ~ 160℃
Temp. Sensor
Pt100 3 wires
Pt100 3 wires(only for small pipes)
Fluid Water, sea water, industrial sewage, acid & alkali liquid, alcohol, beer, all kinds
of oils which can transmit ultrasonic single uniform liquid
Flow Direction Bi-directional measuring, net flow/heat measuring
Accuracy ± 1% ~ ± 2% of reading (1.5 ~ 100 feet/sec.)
Repeatability ± 0.2% of reading
Flow rate range 0.01 ~ ± 12 m/s
Response Time Less than 1 second
Keypad 16 Key with tactile action
Pipe Material Steel, stainless steel, cast iron, cement pipe, copper, PVC, aluminum, FRP etc.
Liner is allowed
Liner Material Tar Epoxy, Rubber, Mortar, Polypropylene, Polyserial, Polystyrenes, Polyester,
Ebonite, Polyethylene, Teflon, etc.
Mounting wall mounting, local
Display 2 Line LCD with backlight; 5 digits with decimal point for flow rate, 8 digits for
Forward, Reverse & Net totalized values
Output Analog 4-20mA or 0-20mA current output. Impedance 0∼1k. Accuracy
0.1%.
Relay output Over 20 source signals (no signal, reverse flow etc.)
Pulse Output Frequency signal :1~9999 Hz
Input Three analogue input
5. www.rbflowmeters.com
ULTRASONIC FLOW METER/HEAT METER Catalog
5
Three-wire PT100 resistor input (optional)
Other functions
Automatically record the tantalizer data of the last 64 days / 64 months / 5
years;
The power-on time and corresponding flow rate of the last 64 power on and off
events. Allow manual or automatic flow loss compensation
The instrument working status of the last 64 days
Communication RS-232/RS485
Alarm High/Low with buzzer
Protection Transmitter IP 67 Sensor IP67/IP68 (optional)
Straight pipe
section
In the upstream it must be beyond 10D, in the downstream it must be beyond
5D, in the upstream the length must be beyond 30D from the access of the
pump. (D stands for pipe diameter)
Cable Length Twisted Pair Line, standard length of 10m, can be extended to 500m (not
recommended); Contact the manufacturer for longer cable requirement. RS-485
interface, transmission distance up to 1000m
Power Supply Remote Type 90 ~ 260VAC 50/60 Hz; 9-28VDC
Compact Type 9-28VDC
Power
Consumption
Less than 2W
Data Storage Operation parameters and tantalization data are stored in EEPROM for more
than 10 years
Engineer Unit M3, Liter, US gallon, Imperial gallon, Million gallons, Cubic feet, US Barrels,
Imperial barrels, Oil barrels
Time Units Second, Minute, Hour, Day
Protocols MODBUS, M-BUS, Fuji extended protocol and another factory protocol
Humidity Main Unit: 85% RH
Remarks: With a high strength carrying case suitable for normal and harsh environment
Handheld/ Portable Type
Feature
4-line LCD display with flow rate, totalizer & diagnostics
Small sized (8.25" x 3.50" x 1.48") and lightweight (≈1 lb.)
Wide range of fluid velocities - 0.03 ~ ±105 feet/sec. (0.01 ~ ± 32 m/s)
Long-life Ni-MH battery for up to 12 hours of continuous operation
Transducers for pipe sizes ranging from ½" ~ 240" (15 to 6000 mm)
High accuracy - ±0.5% of reading
Clamp-on transducers feature integral magnet for installation on metal pipe without use of a
mounting belt
Data logger function includes date, totalizer, & signal condition diagnostic
Fast response time - less than 1 second
6. www.rbflowmeters.com
ULTRASONIC FLOW METER/HEAT METER Catalog
6
Specification
Parts Picture Range
Transmitter:
Portable/
Handheld type:
RBFM-H
Transducer
Stand type
High Temp.
type
S type Pipe size (15 mm ~ 100 mm) ½" ~ 4"
Temp. SPN -30 ~ 90℃
SPH -30~160℃
Size(mm) SPN: 45×25×23
SPH 45×25×28
M Type Pipe size 50 mm ~ 1000 mm (2" ~ 38”)
Temp. MPN -30 ~ 90℃
MPH -30~160℃
Size(mm) MPN: 64×35×32
MPH 64×39×44
L Type Pipe size 300~ 6000(12" ~ 240”)
Temp. LPN -30 ~ 90℃
LPH -30~160℃
Size(mm) LPN 98×49×45
LPH 97×54×53
Transducer
Mounting Frame
S Type Pipe size (15 mm ~ 100 mm) ½" ~ 4"
Temp. SPMN -30 ~ 90℃
SPMH -30~160℃
M Type Pipe size 50 mm ~ 300 mm (2" ~ 38”)
Temp. MPMN -30 ~ 90℃
MPMH -30~160℃
Temp. Sensor
(only portable
type)
Pt100 3 wires (optional)
Pt100 3 wires (only for small pipes)
Fluid Water, sea water, industrial sewage, acid and alkali liquid, various oils etc.
liquid which can transmit sound wave.
Accuracy ± 1% ~ ± 2% of reading (1.5 ~ 100 feet/sec.)
Flow rate Range 0.01 ~ ± 12 m/s
Repeatability ±0.0001 m/s
7. www.rbflowmeters.com
ULTRASONIC FLOW METER/HEAT METER Catalog
7
Response Time Less than 1 second
Ambient
Temperature
-4 ~ 122 °F (-20 ~ 50 °C)
Keypad 16 Key with tactile action
Pipe Material Cast Iron, Stainless Steel, Ductile Iron Copper, PVC, Aluminum, Asbestos
Fiberglass, etc.
Liner Material Tar Epoxy, Rubber, Mortar, Polypropylene, Polyserial, Polystyrenes, Polyester,
Ebonite, Polyethylene, Teflon, etc.
Mounting wall mounting, local
Display RBFP
RBFH 4 Line LCD with backlight; 5 digits with decimal point for flow
rate, 8 digits for Forward, Reverse & Net totalized values
Output Analog 4-20mA or 0-20mA ((External, optional, if need it, when you
order, pls specify)
Pulse Output 1~9999 Hz (external, optional)
Communication RS485 or Data logger for optional
Alarm High/Low with buzzer
Built-in battery Ni-MH
Battery
RBFP > 20 Hours
RBFH > 12 Hours
Charger RBHF/RBPF 90 ~ 260VAC 50/60 Hz, 8-10 hr charging
Protection Transmitter IP 65 Sensor IP67/IP68 (optional)
Cable Length 150 m
Power
Consumption
Less than 2W
Data Storage RBHF 64 data points including flow rate, totalizer, time, & date
RBPF SD card, can be to 2G
Engineer Unit M3, Liter, US gallon, Imperial gallon, Million gallons, Cubic feet, US Barrels,
Imperial barrels, Oil barrels
Time Units Second, Minute, Hour, Day
Remarks With a high strength carrying case suitable for normal and harsh environment
Typical Applications
The RBPF flow meter can be virtually applied to a wide range of measurements. The measured pipe
ranges 15-6000mm. Because the instrument and transducers are non-contacting and have no
moving parts; the flow meter can not be affected by system pressure, fouling or wear. Standard
transducers are rated to 110 ºC. For further information, please consult the manufacturer for
assistance.
1. Water、industrial sewage、sea
water
10. Operation and maintenance of ship
2. Acid & Alkali Liquid 11. Monitoring and management of water
saving
3. Various oil 12. Papermaking and pulping
8. www.rbflowmeters.com
ULTRASONIC FLOW METER/HEAT METER Catalog
8
4. Feed water and drainage 13. Leak detection
5. Detection of water resources 14. Heating
6. Petrochemical 15. Manufacturing of pump、boiler、cooling
tower
7. Food and pharmaceutical industry 16. Management of flow and heat、control
network system
8. Power plant 17. Patrol, tracing and collection of flow
9. Metallurgical mines 18. Measurement and balance of heat
Ultrasonic Heat meter
Overview
RB series ultrasonic heat meters employ the proven transit-time ultrasonic flow
measurement technique combined with two PT 100 temperature sensors (or customer-supplied
RTD sensors) for calculating total and instantaneous energy consumption. Flow measurement may
be achieved via clamp-on, spool piece, or insertion-type sensors. Our microprocessor based, user
friendly field programmable flow measurement technique creates no interruption of the process
flow and allows for easy installation.
RB series ultrasonic heat meter is available with a variety of features that make it adaptable to
many different applications. Available transducer types include clamp-on, in-line, and insertion.
Wall-mount, panel-mount, pipe-mounted and compact din-rail mounted display types are also
offered. Additional available features include pulse, 0-20mA & 4-20mA outputs, RS-232
communication, user-selectable engineering units, and a data logger function. This variety of
features, combined with its low cost, make the RB series an ideal choice for economical BTU
monitoring in many applications involving chillers, radiators, and heat exchangers.
Feature
Measures energy consumption rate and total
Fixed and Portable version
Proven ultrasonic technology for flow measurement and PT 100 for temperature
Clamp-on, spool piece or insertion mounting for flow measurement
Wide range of fluid velocities - 0.03 ~ ± 105 feet/sec.
Transducers for pipe sizes ranging from ½" to 240" (15 to 6000 mm)
High accuracy - ±1% of reading
RS485 and Modbus communication protocols available
4-20mA and pulse outputs with available relays and alarms
Data logger function - includes time & date, totalizer, diagnostics
9. www.rbflowmeters.com
ULTRASONIC FLOW METER/HEAT METER Catalog
9
Fast response time - less than 1 second
NIST traceable calibration certificate
Specification: See Fixed and Portable Type
Measuring Principle Transit time difference
Mounting wall mounting, panel, local
Transducer type clamp-on, spool piece, or insertion
Display LCD with backlight, 2 x 20 characters 1-line, non-backlit 7 digit LCD
(ILD version) Displays energy rate, total consumption, temperature,
instantaneous flow rate, accumulated flow rate, velocity, time
Keypad 4 x 4 tactile-feedback membrane keypad
Pipe Size 25-6000mm
Pipe Material Carbon Steel, Stainless Steel, Ductile Iron Copper, PVC, Aluminum,
Asbestos Fiberglass, etc.
Output Signals 4-20mA , Impedance 0-1kΩ, pulse, relays, Isolated OCT for alarms
(on/off with buzzer)
Inputs Two RTD channels, and additional inputs
10. www.rbflowmeters.com
ULTRASONIC FLOW METER/HEAT METER Catalog
10
Flanged Type Ultrasonic Flow Meter
Features
Calibrated in manufacture's lab, field setup is unnecessary.
Measurement is independent of fluid nature with wider applications than magnetic meters.
Available including up to 2GB event data logger, can search the event of totalizer flow, flow
rate, velocity etc.
Parallel operation of positive, negative and net flow totalizer, while the output of totalizer pulse
and frequency output are transmitted via open collector.
Re-calibration or maintenance is easy, no processing interruption (just plug out the inserted
transducers from pipe line, when re-calibration or maintenance is finished), can be hot-tapped.
Transmitter
Type Π Type Flanged Pipe size DN32/DN40 Pipe size≥DN5
Transducer
Material 304 Stainless Steel Carbon Steel
Pipe Size DN15 ~DN40 DN32~DN40 ≥DN50
12. www.rbflowmeters.com
ULTRASONIC FLOW METER/HEAT METER Catalog
12
Model Selection
RBFM X X X DNxx X X X X X
Instrument Type
F Ultrasonic Flowmeter
H Ultrasonic Heat meter
A
D
B
F
H
F
B
SN
SH
SPMN
SPMH
MN M Type(pipe size: 50-700mm, temp.: -30 to 90℃))
MH M Type(pipe size: 50-700mm, temp.: -40 to 160℃))
MPMN
MPMH
RB-1
RBH
LN L type(pipe size: 300-6000mm, temp.: -30 to 90℃)
LH L type(pipe size: 300-6000mm, temp.: -40 to 160℃)
I
IE
F
Pt100
DNxx
C
SS
CI
F
P
C
X
X
Output
N
A
P
R
2
4
5
N
1 1*4-20mA
2 2*4-20mA
3 3*4-20mA
Cable length
X meter
S type with mounting frame only for Handheld (pipe size:
15-100mm, temp.: -40 to 90℃)
S type with mounting frame only for Handheld(pipe size:
15-100mm, temp.: -40 to 160℃)
M type with mounting frame only for Handheld (pipe size:
50-300mm, temp.: -40 to 90℃)
M type with mounting frame only for Handheld (pipe size:
50-300mm, temp.: -40 to 160℃)
Power supply
220VAC
24VDC
Battery
Handheld type
Flanged type
Battery type
S type(pipe size: 15-100mm, temp.: -30 to 90℃)
S type(pipe size: 15-100mm, temp.: -40 to 160℃)
Insertion Type
Extended type
4-20mA
Relay
RS232
RS485
Fibreglass
PVC
cement pipe
Others, clarify
RBFM Ultrasonic Flow/Heat meter Model Selection
NO
Datalogger
Input
NO
Pulse
Flow/Heat meter Type
Transducer Type
Temperature sensor
Flanged type
Fixed type
extended the mounting frame, for pipe:DN300-700mm,
temp.: -40 to 90℃
extended the mounting frame, for pipe:DN300-700mm,
temp.: -40 to 90℃
Pipe size
Pipe material
Pressure(MPa)
Carbon Steel
Stainless Steel
cast iron
13. www.rbflowmeters.com
ULTRASONIC FLOW METER/HEAT METER Catalog
13
Application
The wall-mounting flow meter can be applied to a wide range of pipe flow measurements. Applicable
liquids include pure liquids as well as liquid with small quantity of tiny particles.
Examples are:
Water (hot water, chilled water, city water, sea water, waste water, etc.);
Sewage with small particle content;
Oil (crude oil, lubricating oil, diesel oil, fuel oil, etc.);
Chemicals (alcohol, acids, etc.);
Plant effluent;
Beverage, liquid food;
Ultra-pure liquids;
Solvents and other liquids
14. www.rbflowmeters.com
ULTRASONIC FLOW METER/HEAT METER Catalog
14
Ultrasonic Thickness gauge
Summary
RBTG Ultrasonic Thickness gauge is used for measuring the
thickness of materials where access to only one side of the test
piece is available for many different materials including steel,
cast iron, aluminum, red copper, brass, zinc, quartz glass,
polyethylene PVC, gray cast iron, nodular cast iron and other
ultrasonic wave well-conductive materials. Dual probe
extended your measure range to every possible need in your
area. Transducer models are available for special application,
including for coarse grain material and high temperature
applications. Two-Point Calibration function; Two work
modes: Single point mode and Scan mode;Coupling status
indicator showing the coupling status;Auto sleep and auto
power off function to conserve battery life.
Features
Large LCD for easy reading
Suitable for all metallic and non-metallic materials
Easy to operate
Wide measuring ranges from 0.75mm to 300mm (Steel)
Sound Speed Range up to 9999m/s
Display resolution 0.1 mm
Handheld and robust, USB or through coating type optional;
Function
1. Applicable to measure the thickness of any hard materials, e.g. steel, cast iron, aluminum, red
copper, brass, zinc, quartz glass, polyethylene PVC, gray cast iron, nodular cast iron
2. Automatic Zero Adjustment;
3. Data Store Function for up to 20 t files (up to 99 values for each file) of stored value;
4. Auto Power Shut Off;
5. Linear compensation circuit for high accuracy (1%H+0.1);
6. Speed of sound testing (m/s) with 12 storage memory;
7. 2.5 MHz for rough surface measurement
17. www.rbflowmeters.com
ULTRASONIC FLOW METER/HEAT METER Catalog
17
Application
Used for measuring thickness and corrosion of pressure vessels, chemical equipment, boilers, oil storage tanks, etc.