creating sound valueTM
New Approaches to Preserve Residual
Hearing and Improve Performance for
Cochlear Implant Recipients
Robert Cowan
Carrie Newbold, Silvana Mergen, Dimitra Stathopoulos, Scott Chambers
Chris Miller, Robert Briggs, BinBin Zhang, Gordon Wallace & Stephen O’Leary
The HEARing Cooperative Research Centre, Australia
Depts of Otolaryngology & Audiology, The University of Melbourne, Australia
Royal Victorian Eye and Ear Hospital, Melbourne, Australia
Intelligent Polymer Research Institute, University of Wollongong, Australia
Cochlear Limited
www.hearingcrc.orgcreating sound valueTM
www.hearingcrc.org
creating sound valueTM 2
Human “Medical’ Bionics ‘Interface’
‘seamless transition
between the hard
world of electronics
and the soft
world of biology’
Nanowerk, July 17 2007
creating sound valueTM
Cochlear Implant (CI)
“cochlear implants, which use multiple electrodes that are implanted
into the cochlea to stimulate auditory nerves, are one of the most
successful medical bionic prosthetics”
Modern Cochlear Implant Systems
creating sound valueTM
Med-EL
Neurelec
Advanced
Bionics
Cochlear
creating sound valueTMcreating sound valueTM 5
Development of Nucleus ® Technology
CI24M CI24R
(Contour™)
CI24RE
(Freedom™ )
1986 1997 2000 2005
CI512
2009
CI22M
WSP ESPrit 3GMSP SPrint™
1983 1989 1994 1997 1998 2002 2005
CP810
2009ESPrit™ Freedom™
Spectra
creating sound valueTM
Contour
Advance
Half-banded
Contour
Technology: Perimodiolar Positioning
creating sound valueTM 7
S C
Shepherd et al.,1993
 EABR recordings at differing
positions in the cochlea
Outer wall Mid-scala
Peri-cortiPerimodiolar
Improving the Electro-neural Interface - Positioning
creating sound valueTM
Psychophysics
Pre-Curled Electrode Studies
creating sound valueTM
1994
1995
1997
1998
1999
2001
2003
Perimodiolar Curling Array using hydrogels
concept studies initiated with IPRI
Histology on first perimodiolar designs
Curly with Stylet studies
First bioengineering & insertion studies
Curling and pre-curled arrays
Prototype pre-curled electrode
implanted 3 adult patients
First Contour implants
Contour Array
Phase II Trials
Advance &
AOS Insertion
Applying Polymer Technology to Improve the
Electro-neural Interface
creating sound valueTM
Technology: Perimodiolar Positioning
Standard Nucleus 24 electrode Nucleus 24 Contour electrode
Standard Contour
creating sound valueTM
Technology: Perimodiolar Positioning
 AOS surgical technique critical to atraumatic insertion
Xu et al, 2001
Microfocus
Radiography
creating sound valueTM
Standard
insertion
technique
AOS™
insertion
technique
• Flexible, pre-
curved
electrode
• SoftipTM
• Minimal
lateral wall
forces with
AOS1
1 Roland, J.T. Jnr. A model for Cochlear Implant Electrode Insertion and Force Evaluation: Results with a New Electrode Design and Insertion Technique (2005).
• Standard insertion technique
• AOS insertion technique
Electrode insertion: Contour & Contour Advance
creating sound valueTM 13
 no insertion trauma cochlear structures
ITB5, side section 10
10 Jun 2002
Contour Advance
Advance
Applying Polymer Technology to Improve the
Electro-neural Interface
creating sound valueTM
CI Patients: Pre-Operative Hearing Levels
creating sound valueTM
Nucleus® Electrode Portfolio
Hybrid L
RW Insertion
Hybrid S8
Research
N24 Straight
Contour Advance
creating sound valueTM
17-19mm perimodiolar
16mm lateral wall
Tip : 0.5 mm
Tip : 0.35 x 0.25mm
Base : 0.8mm
Base : 0.55 x 0.4mm
Nucleus® electrodes
Contour advance
Hybrid-L™ Electrode
Straight Banded array
24mm overall Base : 0.6mm Tip : 0.4 mm17mm lateral wall
creating sound valueTM
Nucleus® Hybrid-L CI System
Hybrid L
RW Insertion
Frequency (Hertz)
-10
0
10
20
30
40
50
60
70
80
90
100
110
120
HearingThresholdLevels(dBHL)
125 250 500 1000 2000 4000 8000
HA stimulates
low frequencies
CI stimulates
high frequencies
Shorter electrode to avoid potential for
trauma to apical end of cochlea
creating sound valueTM
Unaided Thresholds -Pre-Op & 1 Month Post-Op
1-month post-operative data selected to indicate hearing
preservation with Hybrid-L
creating sound valueTM
Unaided Thresholds – Pre & I month Post Op
creating sound valueTM
Clinical Trial – Benefits of Hybrid-L
COMBINED MODE BIMODAL MODE
HA HACI + HA CI
Electroacoustic Acoustic AcousticElectric+ +
Speech &
Noise Front
Localisation
creating sound valueTM
creating sound valueTM
Two subjects showed unstable hearing post-operatively; one at 3-months
(coincided with labyrinthitis) and one at 6-months (no clinical signs)
Hearing Stability (all subjects with 6m+ data)
creating sound valueTM
Preoperative versus Postoperative Benefit?
Benefit from 20%-34% Benefit from 37%-49%
Benefit greater when acoustic stimulation combined
with electric stimulation compared to electric alone
Electric + Acoustic
Word scores are better postoperatively (electric
alone) compared to preoperatively (HAs).
Electric alone
What is the benefit of adding acoustic to electric stimulation?
creating sound valueTM
How does Post-op Combined (EA+A)
use compare with Bilateral Acoustic
Inputs ( A+A)?
Paola Incerti
creating sound valueTM
Horizontal Localisation
• N = 14, aged 45 – 81years
• Nucleus CI system
- Hybrid L24 (6)
- CI422 (SRA) (6)
- MRA (1)
- Freedom CA (1)
• MAP parameters
- ACE sound coding
- MP1+2 stim mode
- 25 μsec pulse width
- 7200 Hz total stim rate
• Test condition: EA+A vs A+A
creating sound valueTM
Localisation
Effect of device: p=0.48
EA + A EA + A
creating sound valueTM
Speech Perception
Rader et al (2013) Ear Hear 34:324-32.
Fig. 3. Individual and mean
SRT in dB SNR are shown for
the bimodal (gray bars) and
best-aided EAS (black bars)
listening conditions.
Gifford et al (2013) Ear Hear 34:413-25.
EA + A EA + A
creating sound valueTM
Or:
Is hearing preservation in the implanted ear useful?
How does Post-op Combined (EA+A)
use compare with Bimodal (E +A)?
Kerrie Plant
creating sound valueTM
• n = 13, aged 55 – 83 years
• Nucleus CI system
-Hybrid L24 (5)
-CI422 (SRA) (7)
-MRA (1)
• Two test conditions:
– EA+A vs E+A
Measures:
• Subjective ratings (SSQ)
• Musical sound quality
ratings
• Speech perception in noise
• Localisation ability
Clinical Trial – Benefits of Hearing Preservation
creating sound valueTM
Subjective Rating
EA+A vs E+A
EA+AE+A
creating sound valueTM
BetterPerformance
Speech Perception and Localization
EA+AE+A EA+AE+A
creating sound valueTM
Skarzynski et al (2014) Ear Hear
Skarzynski et al (2014) Cochlear implantation with the Nucleus slim straight electrode
in subjects with residual low-frequency hearing. Ear Hear.
EA + A E + A
n = 30
p = 0.34 p < 0.05
creating sound valueTM
What about if all acoustic hearing
is lost in the implanted ear?
creating sound valueTM
Speech Perception
Best Aided; Pre-Operative HAs, Post-Operative Bimodal (n=19)
creating sound valueTM
CVC words at 65dBSPL
creating sound valueTM
Spatial Release from Masking
Best Aided; Pre-Operative HAs, Post-Operative Bimodal
creating sound valueTM
HACI
Speech (S0)
SRM: SRT S0N90 – S0N0
Positive Value = advantage
provided by spatial
separation
NOISE CI NOISE HA
creating sound valueTM
Localisation
Best Aided; Pre-Operative HAs, Post-Operative Bimodal (n=14)
creating sound valueTM
Consistent trend all subjects
50% subjects had poorer post-
operative localisation ability
25º
1
2
3
4 5
7
6
8
creating sound valueTM
Subjective Ratings of Benefit:
Best Aided; Pre-Operative HAs, Post-Operative Bimodal
creating sound valueTM
SSQ Questionnaire
creating sound valueTM
Nucleus® Electrode Portfolio
Hybrid L
RW Insertion
N24 Straight
Contour Advance
Slim Straight
(422)
creating sound valueTM
17-19mm perimodiolar
16mm lateral wall
Tip : 0.5 mm
Tip : 0.35 x 0.25mm
Base : 0.8mm
Base : 0.55 x 0.4mm
Nucleus® electrodes
Contour advance
Hybrid-L™ Electrode
CI422 (Slim Straight)
25mm overall
20mm lateral wall
Base : 0.6mm Tip : 0.3mm
Straight Banded array
24mm overall Base : 0.6mm Tip : 0.4 mm17mm lateral wall
creating sound valueTM
Unaided Thresholds (Pre- vs Post-Operative)
creating sound valueTM
creating sound valueTM
Unaided Thresholds (Pre- vs Post-Op): Straight Slim
creating sound valueTM
creating sound valueTM
Nucleus CI422 Slim Straight electrode
• Prospective study of outcomes for 90 adult
CI422 implant recipients at RVEEH
Photos courtesy of Cochlear
CI422
Recipients
Mean (SD)
Age at implant (years) 61.2 (16.2)
Duration of loss - CI ear
(years)
15.4 (11.8)
Duration of loss -
contralateral ear (years)
16.9 (12.9)
creating sound valueTM
Pre-op vs Post-op Hearing
Pre-Op
thresholds:
• 250 = 52.8
• 500 = 65.4
• 1K = 87.4
• 2K = 102.5
• 4K = 107.2
(n=55)
Post-Op (3mth)
thresholds:
• 250 = 76.6
• 500 = 88.0
• 1K = 103.4
• 2K = 110.7
• 4K = 115.5
Post Op (12mth)
thresholds:
• 250 = 88.5
• 500 = 99.0
• 1K = 112.3
• 2K = 117.4
• 4K = 118.9
(n=55) (n=21)
creating sound valueTM 43
Hearing Preservation – Slim Straight 422
30% had drop of greater than 30dB in acoustic hearing
creating sound valueTM
Preservation of Residual Hearing
• Mean change in hearing thresholds at 500 Hz by
3M post-op is -23.3 dB
• Examining the degree of change to 500 Hz
threshold from pre-op to 3M post-op:
• Is the hearing useful?
– 67.2 % (37) of CI422 recipients had hearing
thresholds of 80 dB or better pre-operatively
– At 3 months post-op, 40.5 % (15) of these 37
had hearing thresholds of 80 dB or better
Degree of Change % of CI422 recipients (n=55)
≤ 10 dB loss at 500 Hz 36.4 %
15 – 25 dB loss at 500 Hz 23.6 %
≥ 30 dB loss at 500 Hz 40.0 %
creating sound valueTM 45
Effect of Depth of Insertion
-70
-60
-50
-40
-30
-20
-10
0
300 350 400 450 500 550DegreeofChangeto500Hz
Threshold(dBHL)
Insertion Angle (degrees)
For insertion >360 degrees no apparent increase in hearing loss with
deeper insertion as might be predicted from increase in lateral wall force
creating sound valueTM 46
Comparing Hearing Preservation
0
20
40
60
80
100
120
250 500 1000 2000 4000
Hearingthresholdlevel(dBHL)
Frequency (Hz)
0
20
40
60
80
100
120
250 500 1000 2000 4000
Hearingthresholdlevel(dBHL)
Frequency (Hz)
0
20
40
60
80
100
120
250 500 1000 2000 4000
Hearingthresholdlevel(dBHL)
Frequency (Hz)
0
20
40
60
80
100
120
250 500 1000 2000 4000
Hearingthresholdlevel(dBHL)
Frequency (Hz)
Preoperative Postoperative
Hybrid L24Hybrid L24
CI422 CI422
creating sound valueTM 47
Comparing Speech Perception
PRE-OP
Monaural
Sentence in
Noise (%)
PRE-OP
Binaural
Sentence in
Noise (%)
3M
Monaural
Sentence in
Noise (%)
3M
Binaural
Sentence in
Noise (%)
Hybrid-L24 45.91 56.54 64.94 77.66
CI422 21.40 43.91 58.94 75.53
Pre-op, Hybrid-L24 patients had better pre-op speech perception
in the implanted ear (p=0.003)
No difference post-op between the groups on monaural or binaural
creating sound valueTM
Future Directions for Electrodes?
Where are we headed?, and how to get there?
creating sound valueTM
Nucleus® Electrode Portfolio
Hybrid L
RW Insertion
N24 Straight
Contour Advance
Slim Straight
(422)
Modiolar
Research Array
Note: MRA research device not approved for clinical use.
creating sound valueTM
CA
MRA
MRA - Concept
SOFT TUBE
0.64mm
0.8mm
Volume = 1.769mm3
Volume = 8.266mm3
Ø0.5 Ø0.3
Ø0.8
Ø0.5
SHEATH
LUMEN/STYLET
Perimodiolar Positioning
creating sound valueTM
Electrode
Handle
Sheath Handle
Split along lateral wall
Sheath opens when pulled
back against electrode handle
MRA Electrode - Concept
• Design goal: thinner electrode for atraumatic
insertion
• Insertion concept: precurved electrode held
straight for insertion by soft polymer sheath
Note: Research device, not approved for clinical use.
creating sound valueTM 52
creating sound valueTM
Electrode array
Auditory Nerve
Cells Bodies
Electro-neural Interface
OSL
+ +
+
+
-
-
-
e-
e-
e-
Lateral Wall
+ +
+
+
-
-
-
e-
e-
e-
Living biological
material
• current spread is through an ‘electrolytic’ medium
• Fibrous tissue sheath may significantly dissipate electric current
• mechanisms to control tissue reaction could be beneficial
creating sound valueTM 54
1
2
st
sv
1 =New bone 2 = Fibrous tissue
Low impedance High impedance
st: scala tympani sv: scala vestibuli
Electro-neural Interface & Tissue Reaction
creating sound valueTM
Eastwood, Chang, Kel, Sly, Richardson and O’Leary. Round window delivery of dexamethasone
ameliorates local and remote hearing loss produced by cochlear implantation into the second turn
of the guinea pig cochlea. Hearing Research 265 (2010) 25–29
 studies have confirmed that round window delivery of
dexamethasone can reduce inflammatory response
Effects of dex in Animal Model
Dex-eluting array reduced hearing loss from implantation
in the guinea pig model
creating sound valueTM
Passive Drug Delivery from Surface
drug
drug
Cochlear Implant-based Steroid Delivery
drug
drug
drug
drug
drug
Natural diffusion
drug
drug
drug
drug
drug
Pt electrode
Silicone carrier
DEX within silicone
DEX can be incorporated into the silicone
by physical hand-mixing or by solvent casting
creating sound valueTM
Faster Dex release from solvent casting films
0
10
20
30
40
50
0 20 40 60 80 100
Dexreleaseefficiency%
Time (day)
S5
M5
0
10
20
30
40
50
0 20 40 60 80 100
Time (day)
Dexreleaseefficiency%
S10
M10
0
10
20
30
40
50
0 20 40 60 80 100
Time (day)
Dexreleaseefficiency%
S15
M15
0
10
20
30
40
50
0 20 40 60 80 100
Time (day)
Dexreleaseefficiency%
S20
M20
Evaluating Drug-eluting Processing
S = solvent casting
M = physically mixed
creating sound valueTM
Electrode Design for Passive Delivery
An augmented Cochlear Nucleus Freedom implant with
Contour Advance electrode
40%w/w
dexamethasone
(DXB, Sanofi) in the
tip & spine regions
of the intra-cochlear
array
creating sound valueTM
Shelf Life Validation
No significant difference in purity (demonstrated by amount
of substance recovered), left, and in vitro pharmacokinetics,
right, was evident after normal and accelerated aging.
creating sound valueTM
Inner Ear Safety
Mean SGN densities in each turn of implanted
cochleae
Proportion of new Loose Fibrous Tissue, LFT, or
Dense Fibrous Tissue, DFT and bone, NB,
evident in the scala tympani
8 Dunkin-Hartley guinea pigs implanted with electrode
arrays loaded with 50% w/w DXB were compared to a
control group implanted with a drug-free variant
creating sound valueTM
Meningitis Risk
30 male Hooded-Wistar rats were inoculated with 103, 104, 105 CFU
of Streptococcus pneumoniae to determine the level of innoculum
required to induce meningitis (left).
10 rats were then implanted with a device composed of 50%w/w
DXB and inoculated with 104 CFU Streptococcus pneumoniae 5
days later (right). A control group of n =10 was also included,
implanted with the same device without DXB
creating sound valueTM
First Time in Human Study
Centre RVEEH RPAH
Contour Advance with 40%w/w DXB 8 2
Contour Advance (no DXB) 30 6
Objectives: 1. Ease of use and effectiveness
2. Safety in controlled adult population
Subjects:
Evaluation Schedule:
Pre-op Surgery +1wk +2wk +1, 3, 6, 12,
24M
4-point (local)
impedance
X X X X
Speech
perception X 3 & 12M
creating sound valueTM
Interim Findings
Passive delivery of DXB is safe for humans, as demonstrated
by in vitro and pre-clinical investigations.
A modified commercial device has been used in a FTIH study
and has demonstrated initial benefits in impedance
creating sound valueTM
Controlled Drug Delivery with Hollow Fibre Systems
• Controlled membrane porosity by redox induced conformational change
• Solvated ion flux activated by doping / dedoping process
Polymer doping / dedoping induced release of drug
drug
drug
drug
drug
drug
drug
drug
drug
drug
drug
drug
drug
drug
drug
drug
drug
drug
drug
drug
drug
drug
drug
drug
drug
drug
drug
drug
Cochlear implant based steroid delivery
drug
drug
drug
drug
drug
drug
0
0.01
0.02
0.03
0.04
0.05
0.06
0.07
0.08
0 10 20 30 40
Time / minutes
ABSat554nm
3minPPyPVDF
3minPPyPVDF pulsed potential
creating sound valueTMcreating sound valueTM
1. Electrosynthesis of polypyrrole incorporating Dexamethsasone 21-phosphate
disodium salt under different conditions and their relation with Dex release.
Current Density 0.5mA/cm2 1.0mA/cm2 1.5mA/cm2
A. ~81nm B. ~124nm C. ~198nm 5mM
Roughness
(RMS)
D. ~69.8nm E. ~121nm 10mM
F. ~68nm 20mM
Dex Concentration
Tab 1. Film growth condition vs. film roughness ([Py]=0.2M, Q=1C/cm2)
0
10
20
30
40
50
60
70
80
90
0 50 100 150 200 250 300
DexPReleased(%)
Time (min)
A (0.5mA/cm2, Dex 5mM)
B (1.0mA/cm2, Dex 5mM)
C (1.5mA/cm2, Dex 5mM)
D (1.0mA/cm2, Dex 10mM)
E (1.5mA/cm2, Dex 10mM)
F (1.5mA/cm2, Dex 20mM)
Fig 1. Dex release efficiency vs. different film growth conditions
(Potential: -500mV, Drug loading: 380ug/cm2)
• With higher current
density and lower Dex
concentration, higher
roughness obtained
• 80% of Dex released
out of the film with
highest roughness
Drug Elution with Electrical Stimulation
creating sound valueTM
CA
MRA
Focus in Electrode Design
Focus of electrode design on:
• preservation of internal cochlear structures
• positioning to reduce spread of current
• designs for electro-acoustic stimuli
MRA MRA
CA
creating sound valueTM
• preservation of acoustic residual hearing can be
accomplished with good electrode design and
attention to surgical procedure ~ but outcomes
remain unpredictable
•a drug eluting electrode is a feasible vehicle for
delivery of therapeutic agents into the cochlea,
(but must not compromise insertion characteristics)
• elution rates in passive designs are dependent on
drug loading and processes of formulation
•active systems may enable similar drug load delivery
through a smaller cross-sectional area electrode array
Summary
creating sound valueTM
SYDNEY
CLUSTER
MELBOURNE
CLUSTER
BRISBANE
CLUSTER
HEARing CRC Members
68
hear and
say centre
This research was financially supported by the HEARing CRC established and supported
under the Australian Government’s Cooperative Research Centres Program
www.hearingcrc.org

Weitere ähnliche Inhalte

Was ist angesagt?

Acoustic Immittance Measurements
Acoustic Immittance MeasurementsAcoustic Immittance Measurements
Acoustic Immittance Measurements
bethfernandezaud
 
MSD ASSESSMENT MASLP
MSD ASSESSMENT MASLPMSD ASSESSMENT MASLP
MSD ASSESSMENT MASLP
HimaniBansal15
 
Speech audiometry & masking
Speech audiometry & maskingSpeech audiometry & masking
Speech audiometry & masking
bethfernandezaud
 

Was ist angesagt? (20)

Cochlear implant
Cochlear implantCochlear implant
Cochlear implant
 
Introduction to CAPD - Signs & Symptoms
Introduction to CAPD - Signs & SymptomsIntroduction to CAPD - Signs & Symptoms
Introduction to CAPD - Signs & Symptoms
 
Electroglottography & Inverse Filtering Procedures
Electroglottography & Inverse Filtering ProceduresElectroglottography & Inverse Filtering Procedures
Electroglottography & Inverse Filtering Procedures
 
Specific features of hearing aids
Specific features of hearing aidsSpecific features of hearing aids
Specific features of hearing aids
 
Acoustic Immittance Measurements
Acoustic Immittance MeasurementsAcoustic Immittance Measurements
Acoustic Immittance Measurements
 
Lecture 5 description of electro acoustic characteristics of hearing instrume...
Lecture 5 description of electro acoustic characteristics of hearing instrume...Lecture 5 description of electro acoustic characteristics of hearing instrume...
Lecture 5 description of electro acoustic characteristics of hearing instrume...
 
MSD ASSESSMENT MASLP
MSD ASSESSMENT MASLPMSD ASSESSMENT MASLP
MSD ASSESSMENT MASLP
 
ecochG.pptx
ecochG.pptxecochG.pptx
ecochG.pptx
 
Speech audiometry & masking
Speech audiometry & maskingSpeech audiometry & masking
Speech audiometry & masking
 
Assr
AssrAssr
Assr
 
Voice therapy
Voice therapyVoice therapy
Voice therapy
 
Periodic and aperiodic sounds (2)
Periodic and aperiodic sounds (2)Periodic and aperiodic sounds (2)
Periodic and aperiodic sounds (2)
 
Assessment of voice in professional voice users
Assessment of voice in professional voice usersAssessment of voice in professional voice users
Assessment of voice in professional voice users
 
Voice disorders
Voice disordersVoice disorders
Voice disorders
 
The Nucleus Hybrid L24 Cochlear Implant System for sensorineural hearing loss
The Nucleus Hybrid L24 Cochlear Implant System for sensorineural hearing lossThe Nucleus Hybrid L24 Cochlear Implant System for sensorineural hearing loss
The Nucleus Hybrid L24 Cochlear Implant System for sensorineural hearing loss
 
Cochlear Implants
Cochlear ImplantsCochlear Implants
Cochlear Implants
 
Immittance audiometry
Immittance audiometryImmittance audiometry
Immittance audiometry
 
Otoacoustic emission
Otoacoustic emissionOtoacoustic emission
Otoacoustic emission
 
Introduction to Basic Audiology
Introduction to Basic AudiologyIntroduction to Basic Audiology
Introduction to Basic Audiology
 
Speech acoustics
Speech acousticsSpeech acoustics
Speech acoustics
 

Ähnlich wie New Approaches to Preserve Residual Hearing and Improve Performance for Cochlear Implant Recipients

Coordinated Adaptive Processing.pdf
Coordinated Adaptive Processing.pdfCoordinated Adaptive Processing.pdf
Coordinated Adaptive Processing.pdf
MaiGaber4
 
Acoustical Design of Digital Stethoscope for Improved Performance: ATOA CAE
Acoustical Design of Digital Stethoscope for Improved Performance: ATOA CAE Acoustical Design of Digital Stethoscope for Improved Performance: ATOA CAE
Acoustical Design of Digital Stethoscope for Improved Performance: ATOA CAE
ATOA Scientific Technologies
 

Ähnlich wie New Approaches to Preserve Residual Hearing and Improve Performance for Cochlear Implant Recipients (20)

Electro Acoustic Stimulation ( EAS )
Electro Acoustic Stimulation ( EAS ) Electro Acoustic Stimulation ( EAS )
Electro Acoustic Stimulation ( EAS )
 
Cochlear implant systems
Cochlear implant systemsCochlear implant systems
Cochlear implant systems
 
A NOVEL METHOD FOR OBTAINING A BETTER QUALITY SPEECH SIGNAL FOR COCHLEAR IMPL...
A NOVEL METHOD FOR OBTAINING A BETTER QUALITY SPEECH SIGNAL FOR COCHLEAR IMPL...A NOVEL METHOD FOR OBTAINING A BETTER QUALITY SPEECH SIGNAL FOR COCHLEAR IMPL...
A NOVEL METHOD FOR OBTAINING A BETTER QUALITY SPEECH SIGNAL FOR COCHLEAR IMPL...
 
Eas n hp zee
Eas n hp zeeEas n hp zee
Eas n hp zee
 
Advancement in acoustic reflex measures
Advancement in acoustic reflex measuresAdvancement in acoustic reflex measures
Advancement in acoustic reflex measures
 
Hearing Aid Design for Disabling Hearing Loss
Hearing Aid Design for Disabling Hearing LossHearing Aid Design for Disabling Hearing Loss
Hearing Aid Design for Disabling Hearing Loss
 
البحث الثامن.pdf
البحث الثامن.pdfالبحث الثامن.pdf
البحث الثامن.pdf
 
Biometry immersion optical helpful tips & key points
Biometry immersion optical  helpful tips & key pointsBiometry immersion optical  helpful tips & key points
Biometry immersion optical helpful tips & key points
 
Hearing After Hemispherectomy
Hearing After HemispherectomyHearing After Hemispherectomy
Hearing After Hemispherectomy
 
SubBottom Profiler training
SubBottom Profiler trainingSubBottom Profiler training
SubBottom Profiler training
 
Coordinated Adaptive Processing.pdf
Coordinated Adaptive Processing.pdfCoordinated Adaptive Processing.pdf
Coordinated Adaptive Processing.pdf
 
Bone Anchored Hearing Aid JC
Bone Anchored Hearing Aid JCBone Anchored Hearing Aid JC
Bone Anchored Hearing Aid JC
 
Audiometry class by Dr. Kavitha Ashok Kumar MSU Malaysia
Audiometry class by Dr. Kavitha Ashok Kumar MSU MalaysiaAudiometry class by Dr. Kavitha Ashok Kumar MSU Malaysia
Audiometry class by Dr. Kavitha Ashok Kumar MSU Malaysia
 
Rehear OTC HA.pdf
Rehear OTC HA.pdfRehear OTC HA.pdf
Rehear OTC HA.pdf
 
Recording Distortion Product Otoacoustic Emissions using the Adaptive Noise C...
Recording Distortion Product Otoacoustic Emissions using the Adaptive Noise C...Recording Distortion Product Otoacoustic Emissions using the Adaptive Noise C...
Recording Distortion Product Otoacoustic Emissions using the Adaptive Noise C...
 
Predicted impairment and handicap from exposure to steady state broad-band in...
Predicted impairment and handicap from exposure to steady state broad-band in...Predicted impairment and handicap from exposure to steady state broad-band in...
Predicted impairment and handicap from exposure to steady state broad-band in...
 
QoMEX2014 - Analysing the Quality of Experience of Multisensory Media from Me...
QoMEX2014 - Analysing the Quality of Experience of Multisensory Media from Me...QoMEX2014 - Analysing the Quality of Experience of Multisensory Media from Me...
QoMEX2014 - Analysing the Quality of Experience of Multisensory Media from Me...
 
Acoustical Design of Digital Stethoscope for Improved Performance: ATOA CAE
Acoustical Design of Digital Stethoscope for Improved Performance: ATOA CAE Acoustical Design of Digital Stethoscope for Improved Performance: ATOA CAE
Acoustical Design of Digital Stethoscope for Improved Performance: ATOA CAE
 
cochlear corporation.pptx
cochlear corporation.pptxcochlear corporation.pptx
cochlear corporation.pptx
 
Reduction of audio acoustic in Audio-visual transceiving with single port
Reduction of audio acoustic in Audio-visual transceiving with single portReduction of audio acoustic in Audio-visual transceiving with single port
Reduction of audio acoustic in Audio-visual transceiving with single port
 

Mehr von HEARnet _

Mehr von HEARnet _ (20)

Use of ConeBeam Computed Tomography to determine cochlear length and implant ...
Use of ConeBeam Computed Tomography to determine cochlear length and implant ...Use of ConeBeam Computed Tomography to determine cochlear length and implant ...
Use of ConeBeam Computed Tomography to determine cochlear length and implant ...
 
Training in adult cochlear iimpant users
Training in adult cochlear iimpant usersTraining in adult cochlear iimpant users
Training in adult cochlear iimpant users
 
How far down does the top down control of speech processing go? - HEARing CRC...
How far down does the top down control of speech processing go? - HEARing CRC...How far down does the top down control of speech processing go? - HEARing CRC...
How far down does the top down control of speech processing go? - HEARing CRC...
 
Should i get a hearing aid now - HEARing CRC PhD presentation
Should i get a hearing aid now - HEARing CRC PhD presentationShould i get a hearing aid now - HEARing CRC PhD presentation
Should i get a hearing aid now - HEARing CRC PhD presentation
 
Role of infant-directed speech in language development of hearing impaired in...
Role of infant-directed speech in language development of hearing impaired in...Role of infant-directed speech in language development of hearing impaired in...
Role of infant-directed speech in language development of hearing impaired in...
 
Patient and family centred care for young adults - HEARing CRC PhD presentation
Patient and family centred care for young adults - HEARing CRC PhD presentationPatient and family centred care for young adults - HEARing CRC PhD presentation
Patient and family centred care for young adults - HEARing CRC PhD presentation
 
Organisational outcomes of person centred hearing care - HEARing CRC PhD pres...
Organisational outcomes of person centred hearing care - HEARing CRC PhD pres...Organisational outcomes of person centred hearing care - HEARing CRC PhD pres...
Organisational outcomes of person centred hearing care - HEARing CRC PhD pres...
 
MEG system for cochlear implants ecipients and auditory entrainment - HEARing...
MEG system for cochlear implants ecipients and auditory entrainment - HEARing...MEG system for cochlear implants ecipients and auditory entrainment - HEARing...
MEG system for cochlear implants ecipients and auditory entrainment - HEARing...
 
Longterm effects of conductive hearing loss -HEARing CRC PhD presentation
Longterm effects of conductive hearing loss -HEARing CRC PhD presentationLongterm effects of conductive hearing loss -HEARing CRC PhD presentation
Longterm effects of conductive hearing loss -HEARing CRC PhD presentation
 
Hearing loss and the acquistion of english plural morphology - HEARing CRC Ph...
Hearing loss and the acquistion of english plural morphology - HEARing CRC Ph...Hearing loss and the acquistion of english plural morphology - HEARing CRC Ph...
Hearing loss and the acquistion of english plural morphology - HEARing CRC Ph...
 
Hearing, listening and reading: A complex interplay of factors that contribut...
Hearing, listening and reading: A complex interplay of factors that contribut...Hearing, listening and reading: A complex interplay of factors that contribut...
Hearing, listening and reading: A complex interplay of factors that contribut...
 
Hearing health of live music sound engineers - HEARing CRC PhD presentation
Hearing health of live music sound engineers - HEARing CRC PhD presentationHearing health of live music sound engineers - HEARing CRC PhD presentation
Hearing health of live music sound engineers - HEARing CRC PhD presentation
 
Evaluating audiological intervention options for people with dementia living ...
Evaluating audiological intervention options for people with dementia living ...Evaluating audiological intervention options for people with dementia living ...
Evaluating audiological intervention options for people with dementia living ...
 
Effects of age and working memory in understanding speech in conversational n...
Effects of age and working memory in understanding speech in conversational n...Effects of age and working memory in understanding speech in conversational n...
Effects of age and working memory in understanding speech in conversational n...
 
Development of tone sandhi in mandarin speaking children with Cochlear Implan...
Development of tone sandhi in mandarin speaking children with Cochlear Implan...Development of tone sandhi in mandarin speaking children with Cochlear Implan...
Development of tone sandhi in mandarin speaking children with Cochlear Implan...
 
Development of an on-line assessment of speech perception - HEARing CRC PhD p...
Development of an on-line assessment of speech perception - HEARing CRC PhD p...Development of an on-line assessment of speech perception - HEARing CRC PhD p...
Development of an on-line assessment of speech perception - HEARing CRC PhD p...
 
Benefits of music training for children with hearing loss - HEARing CRC PhD p...
Benefits of music training for children with hearing loss - HEARing CRC PhD p...Benefits of music training for children with hearing loss - HEARing CRC PhD p...
Benefits of music training for children with hearing loss - HEARing CRC PhD p...
 
Efficacy of auditory training in adults with hearing loss and auditory proces...
Efficacy of auditory training in adults with hearing loss and auditory proces...Efficacy of auditory training in adults with hearing loss and auditory proces...
Efficacy of auditory training in adults with hearing loss and auditory proces...
 
Audiology Students and Patient-Centred Communication - HEARing CRC PhD presen...
Audiology Students and Patient-Centred Communication - HEARing CRC PhD presen...Audiology Students and Patient-Centred Communication - HEARing CRC PhD presen...
Audiology Students and Patient-Centred Communication - HEARing CRC PhD presen...
 
Artificially enhancing better-ear glimpsing cues to improve understanding of ...
Artificially enhancing better-ear glimpsing cues to improve understanding of ...Artificially enhancing better-ear glimpsing cues to improve understanding of ...
Artificially enhancing better-ear glimpsing cues to improve understanding of ...
 

Kürzlich hochgeladen

College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
perfect solution
 
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service Available
Dipal Arora
 

Kürzlich hochgeladen (20)

Call Girls Guntur Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Guntur  Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Guntur  Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Guntur Just Call 8250077686 Top Class Call Girl Service Available
 
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
 
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
 
Call Girls Bangalore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 8250077686 Top Class Call Girl Service Available
 
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
 
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟ 9332606886 ⟟ Call Me For G...
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟  9332606886 ⟟ Call Me For G...Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟  9332606886 ⟟ Call Me For G...
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟ 9332606886 ⟟ Call Me For G...
 
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
 
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
 
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
 
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
 
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
 
Call Girls Bareilly Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Bareilly Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls Agra Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Agra Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Agra Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Agra Just Call 8250077686 Top Class Call Girl Service Available
 
O963O942363 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
O963O942363 Call Girls In Ahmedabad Escort Service Available 24×7 In AhmedabadO963O942363 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
O963O942363 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
 
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service Available
 

New Approaches to Preserve Residual Hearing and Improve Performance for Cochlear Implant Recipients

  • 1. creating sound valueTM New Approaches to Preserve Residual Hearing and Improve Performance for Cochlear Implant Recipients Robert Cowan Carrie Newbold, Silvana Mergen, Dimitra Stathopoulos, Scott Chambers Chris Miller, Robert Briggs, BinBin Zhang, Gordon Wallace & Stephen O’Leary The HEARing Cooperative Research Centre, Australia Depts of Otolaryngology & Audiology, The University of Melbourne, Australia Royal Victorian Eye and Ear Hospital, Melbourne, Australia Intelligent Polymer Research Institute, University of Wollongong, Australia Cochlear Limited www.hearingcrc.orgcreating sound valueTM www.hearingcrc.org
  • 2. creating sound valueTM 2 Human “Medical’ Bionics ‘Interface’ ‘seamless transition between the hard world of electronics and the soft world of biology’ Nanowerk, July 17 2007
  • 3. creating sound valueTM Cochlear Implant (CI) “cochlear implants, which use multiple electrodes that are implanted into the cochlea to stimulate auditory nerves, are one of the most successful medical bionic prosthetics”
  • 4. Modern Cochlear Implant Systems creating sound valueTM Med-EL Neurelec Advanced Bionics Cochlear
  • 5. creating sound valueTMcreating sound valueTM 5 Development of Nucleus ® Technology CI24M CI24R (Contour™) CI24RE (Freedom™ ) 1986 1997 2000 2005 CI512 2009 CI22M WSP ESPrit 3GMSP SPrint™ 1983 1989 1994 1997 1998 2002 2005 CP810 2009ESPrit™ Freedom™ Spectra
  • 7. creating sound valueTM 7 S C Shepherd et al.,1993  EABR recordings at differing positions in the cochlea Outer wall Mid-scala Peri-cortiPerimodiolar Improving the Electro-neural Interface - Positioning
  • 9. creating sound valueTM 1994 1995 1997 1998 1999 2001 2003 Perimodiolar Curling Array using hydrogels concept studies initiated with IPRI Histology on first perimodiolar designs Curly with Stylet studies First bioengineering & insertion studies Curling and pre-curled arrays Prototype pre-curled electrode implanted 3 adult patients First Contour implants Contour Array Phase II Trials Advance & AOS Insertion Applying Polymer Technology to Improve the Electro-neural Interface
  • 10. creating sound valueTM Technology: Perimodiolar Positioning Standard Nucleus 24 electrode Nucleus 24 Contour electrode Standard Contour
  • 11. creating sound valueTM Technology: Perimodiolar Positioning  AOS surgical technique critical to atraumatic insertion Xu et al, 2001 Microfocus Radiography
  • 12. creating sound valueTM Standard insertion technique AOS™ insertion technique • Flexible, pre- curved electrode • SoftipTM • Minimal lateral wall forces with AOS1 1 Roland, J.T. Jnr. A model for Cochlear Implant Electrode Insertion and Force Evaluation: Results with a New Electrode Design and Insertion Technique (2005). • Standard insertion technique • AOS insertion technique Electrode insertion: Contour & Contour Advance
  • 13. creating sound valueTM 13  no insertion trauma cochlear structures ITB5, side section 10 10 Jun 2002 Contour Advance Advance Applying Polymer Technology to Improve the Electro-neural Interface
  • 14. creating sound valueTM CI Patients: Pre-Operative Hearing Levels
  • 15. creating sound valueTM Nucleus® Electrode Portfolio Hybrid L RW Insertion Hybrid S8 Research N24 Straight Contour Advance
  • 16. creating sound valueTM 17-19mm perimodiolar 16mm lateral wall Tip : 0.5 mm Tip : 0.35 x 0.25mm Base : 0.8mm Base : 0.55 x 0.4mm Nucleus® electrodes Contour advance Hybrid-L™ Electrode Straight Banded array 24mm overall Base : 0.6mm Tip : 0.4 mm17mm lateral wall
  • 17. creating sound valueTM Nucleus® Hybrid-L CI System Hybrid L RW Insertion Frequency (Hertz) -10 0 10 20 30 40 50 60 70 80 90 100 110 120 HearingThresholdLevels(dBHL) 125 250 500 1000 2000 4000 8000 HA stimulates low frequencies CI stimulates high frequencies Shorter electrode to avoid potential for trauma to apical end of cochlea
  • 18. creating sound valueTM Unaided Thresholds -Pre-Op & 1 Month Post-Op 1-month post-operative data selected to indicate hearing preservation with Hybrid-L
  • 19. creating sound valueTM Unaided Thresholds – Pre & I month Post Op
  • 20. creating sound valueTM Clinical Trial – Benefits of Hybrid-L COMBINED MODE BIMODAL MODE HA HACI + HA CI Electroacoustic Acoustic AcousticElectric+ + Speech & Noise Front Localisation creating sound valueTM
  • 21. creating sound valueTM Two subjects showed unstable hearing post-operatively; one at 3-months (coincided with labyrinthitis) and one at 6-months (no clinical signs) Hearing Stability (all subjects with 6m+ data)
  • 22. creating sound valueTM Preoperative versus Postoperative Benefit? Benefit from 20%-34% Benefit from 37%-49% Benefit greater when acoustic stimulation combined with electric stimulation compared to electric alone Electric + Acoustic Word scores are better postoperatively (electric alone) compared to preoperatively (HAs). Electric alone What is the benefit of adding acoustic to electric stimulation?
  • 23. creating sound valueTM How does Post-op Combined (EA+A) use compare with Bilateral Acoustic Inputs ( A+A)? Paola Incerti
  • 24. creating sound valueTM Horizontal Localisation • N = 14, aged 45 – 81years • Nucleus CI system - Hybrid L24 (6) - CI422 (SRA) (6) - MRA (1) - Freedom CA (1) • MAP parameters - ACE sound coding - MP1+2 stim mode - 25 μsec pulse width - 7200 Hz total stim rate • Test condition: EA+A vs A+A
  • 25. creating sound valueTM Localisation Effect of device: p=0.48 EA + A EA + A
  • 26. creating sound valueTM Speech Perception Rader et al (2013) Ear Hear 34:324-32. Fig. 3. Individual and mean SRT in dB SNR are shown for the bimodal (gray bars) and best-aided EAS (black bars) listening conditions. Gifford et al (2013) Ear Hear 34:413-25. EA + A EA + A
  • 27. creating sound valueTM Or: Is hearing preservation in the implanted ear useful? How does Post-op Combined (EA+A) use compare with Bimodal (E +A)? Kerrie Plant
  • 28. creating sound valueTM • n = 13, aged 55 – 83 years • Nucleus CI system -Hybrid L24 (5) -CI422 (SRA) (7) -MRA (1) • Two test conditions: – EA+A vs E+A Measures: • Subjective ratings (SSQ) • Musical sound quality ratings • Speech perception in noise • Localisation ability Clinical Trial – Benefits of Hearing Preservation
  • 29. creating sound valueTM Subjective Rating EA+A vs E+A EA+AE+A
  • 30. creating sound valueTM BetterPerformance Speech Perception and Localization EA+AE+A EA+AE+A
  • 31. creating sound valueTM Skarzynski et al (2014) Ear Hear Skarzynski et al (2014) Cochlear implantation with the Nucleus slim straight electrode in subjects with residual low-frequency hearing. Ear Hear. EA + A E + A n = 30 p = 0.34 p < 0.05
  • 32. creating sound valueTM What about if all acoustic hearing is lost in the implanted ear?
  • 33. creating sound valueTM Speech Perception Best Aided; Pre-Operative HAs, Post-Operative Bimodal (n=19) creating sound valueTM CVC words at 65dBSPL
  • 34. creating sound valueTM Spatial Release from Masking Best Aided; Pre-Operative HAs, Post-Operative Bimodal creating sound valueTM HACI Speech (S0) SRM: SRT S0N90 – S0N0 Positive Value = advantage provided by spatial separation NOISE CI NOISE HA
  • 35. creating sound valueTM Localisation Best Aided; Pre-Operative HAs, Post-Operative Bimodal (n=14) creating sound valueTM Consistent trend all subjects 50% subjects had poorer post- operative localisation ability 25º 1 2 3 4 5 7 6 8
  • 36. creating sound valueTM Subjective Ratings of Benefit: Best Aided; Pre-Operative HAs, Post-Operative Bimodal creating sound valueTM SSQ Questionnaire
  • 37. creating sound valueTM Nucleus® Electrode Portfolio Hybrid L RW Insertion N24 Straight Contour Advance Slim Straight (422)
  • 38. creating sound valueTM 17-19mm perimodiolar 16mm lateral wall Tip : 0.5 mm Tip : 0.35 x 0.25mm Base : 0.8mm Base : 0.55 x 0.4mm Nucleus® electrodes Contour advance Hybrid-L™ Electrode CI422 (Slim Straight) 25mm overall 20mm lateral wall Base : 0.6mm Tip : 0.3mm Straight Banded array 24mm overall Base : 0.6mm Tip : 0.4 mm17mm lateral wall
  • 39. creating sound valueTM Unaided Thresholds (Pre- vs Post-Operative) creating sound valueTM
  • 40. creating sound valueTM Unaided Thresholds (Pre- vs Post-Op): Straight Slim creating sound valueTM
  • 41. creating sound valueTM Nucleus CI422 Slim Straight electrode • Prospective study of outcomes for 90 adult CI422 implant recipients at RVEEH Photos courtesy of Cochlear CI422 Recipients Mean (SD) Age at implant (years) 61.2 (16.2) Duration of loss - CI ear (years) 15.4 (11.8) Duration of loss - contralateral ear (years) 16.9 (12.9)
  • 42. creating sound valueTM Pre-op vs Post-op Hearing Pre-Op thresholds: • 250 = 52.8 • 500 = 65.4 • 1K = 87.4 • 2K = 102.5 • 4K = 107.2 (n=55) Post-Op (3mth) thresholds: • 250 = 76.6 • 500 = 88.0 • 1K = 103.4 • 2K = 110.7 • 4K = 115.5 Post Op (12mth) thresholds: • 250 = 88.5 • 500 = 99.0 • 1K = 112.3 • 2K = 117.4 • 4K = 118.9 (n=55) (n=21)
  • 43. creating sound valueTM 43 Hearing Preservation – Slim Straight 422 30% had drop of greater than 30dB in acoustic hearing
  • 44. creating sound valueTM Preservation of Residual Hearing • Mean change in hearing thresholds at 500 Hz by 3M post-op is -23.3 dB • Examining the degree of change to 500 Hz threshold from pre-op to 3M post-op: • Is the hearing useful? – 67.2 % (37) of CI422 recipients had hearing thresholds of 80 dB or better pre-operatively – At 3 months post-op, 40.5 % (15) of these 37 had hearing thresholds of 80 dB or better Degree of Change % of CI422 recipients (n=55) ≤ 10 dB loss at 500 Hz 36.4 % 15 – 25 dB loss at 500 Hz 23.6 % ≥ 30 dB loss at 500 Hz 40.0 %
  • 45. creating sound valueTM 45 Effect of Depth of Insertion -70 -60 -50 -40 -30 -20 -10 0 300 350 400 450 500 550DegreeofChangeto500Hz Threshold(dBHL) Insertion Angle (degrees) For insertion >360 degrees no apparent increase in hearing loss with deeper insertion as might be predicted from increase in lateral wall force
  • 46. creating sound valueTM 46 Comparing Hearing Preservation 0 20 40 60 80 100 120 250 500 1000 2000 4000 Hearingthresholdlevel(dBHL) Frequency (Hz) 0 20 40 60 80 100 120 250 500 1000 2000 4000 Hearingthresholdlevel(dBHL) Frequency (Hz) 0 20 40 60 80 100 120 250 500 1000 2000 4000 Hearingthresholdlevel(dBHL) Frequency (Hz) 0 20 40 60 80 100 120 250 500 1000 2000 4000 Hearingthresholdlevel(dBHL) Frequency (Hz) Preoperative Postoperative Hybrid L24Hybrid L24 CI422 CI422
  • 47. creating sound valueTM 47 Comparing Speech Perception PRE-OP Monaural Sentence in Noise (%) PRE-OP Binaural Sentence in Noise (%) 3M Monaural Sentence in Noise (%) 3M Binaural Sentence in Noise (%) Hybrid-L24 45.91 56.54 64.94 77.66 CI422 21.40 43.91 58.94 75.53 Pre-op, Hybrid-L24 patients had better pre-op speech perception in the implanted ear (p=0.003) No difference post-op between the groups on monaural or binaural
  • 48. creating sound valueTM Future Directions for Electrodes? Where are we headed?, and how to get there?
  • 49. creating sound valueTM Nucleus® Electrode Portfolio Hybrid L RW Insertion N24 Straight Contour Advance Slim Straight (422) Modiolar Research Array Note: MRA research device not approved for clinical use.
  • 50. creating sound valueTM CA MRA MRA - Concept SOFT TUBE 0.64mm 0.8mm Volume = 1.769mm3 Volume = 8.266mm3 Ø0.5 Ø0.3 Ø0.8 Ø0.5 SHEATH LUMEN/STYLET Perimodiolar Positioning
  • 51. creating sound valueTM Electrode Handle Sheath Handle Split along lateral wall Sheath opens when pulled back against electrode handle MRA Electrode - Concept • Design goal: thinner electrode for atraumatic insertion • Insertion concept: precurved electrode held straight for insertion by soft polymer sheath Note: Research device, not approved for clinical use.
  • 53. creating sound valueTM Electrode array Auditory Nerve Cells Bodies Electro-neural Interface OSL + + + + - - - e- e- e- Lateral Wall + + + + - - - e- e- e- Living biological material • current spread is through an ‘electrolytic’ medium • Fibrous tissue sheath may significantly dissipate electric current • mechanisms to control tissue reaction could be beneficial
  • 54. creating sound valueTM 54 1 2 st sv 1 =New bone 2 = Fibrous tissue Low impedance High impedance st: scala tympani sv: scala vestibuli Electro-neural Interface & Tissue Reaction
  • 55. creating sound valueTM Eastwood, Chang, Kel, Sly, Richardson and O’Leary. Round window delivery of dexamethasone ameliorates local and remote hearing loss produced by cochlear implantation into the second turn of the guinea pig cochlea. Hearing Research 265 (2010) 25–29  studies have confirmed that round window delivery of dexamethasone can reduce inflammatory response Effects of dex in Animal Model Dex-eluting array reduced hearing loss from implantation in the guinea pig model
  • 56. creating sound valueTM Passive Drug Delivery from Surface drug drug Cochlear Implant-based Steroid Delivery drug drug drug drug drug Natural diffusion drug drug drug drug drug Pt electrode Silicone carrier DEX within silicone DEX can be incorporated into the silicone by physical hand-mixing or by solvent casting
  • 57. creating sound valueTM Faster Dex release from solvent casting films 0 10 20 30 40 50 0 20 40 60 80 100 Dexreleaseefficiency% Time (day) S5 M5 0 10 20 30 40 50 0 20 40 60 80 100 Time (day) Dexreleaseefficiency% S10 M10 0 10 20 30 40 50 0 20 40 60 80 100 Time (day) Dexreleaseefficiency% S15 M15 0 10 20 30 40 50 0 20 40 60 80 100 Time (day) Dexreleaseefficiency% S20 M20 Evaluating Drug-eluting Processing S = solvent casting M = physically mixed
  • 58. creating sound valueTM Electrode Design for Passive Delivery An augmented Cochlear Nucleus Freedom implant with Contour Advance electrode 40%w/w dexamethasone (DXB, Sanofi) in the tip & spine regions of the intra-cochlear array
  • 59. creating sound valueTM Shelf Life Validation No significant difference in purity (demonstrated by amount of substance recovered), left, and in vitro pharmacokinetics, right, was evident after normal and accelerated aging.
  • 60. creating sound valueTM Inner Ear Safety Mean SGN densities in each turn of implanted cochleae Proportion of new Loose Fibrous Tissue, LFT, or Dense Fibrous Tissue, DFT and bone, NB, evident in the scala tympani 8 Dunkin-Hartley guinea pigs implanted with electrode arrays loaded with 50% w/w DXB were compared to a control group implanted with a drug-free variant
  • 61. creating sound valueTM Meningitis Risk 30 male Hooded-Wistar rats were inoculated with 103, 104, 105 CFU of Streptococcus pneumoniae to determine the level of innoculum required to induce meningitis (left). 10 rats were then implanted with a device composed of 50%w/w DXB and inoculated with 104 CFU Streptococcus pneumoniae 5 days later (right). A control group of n =10 was also included, implanted with the same device without DXB
  • 62. creating sound valueTM First Time in Human Study Centre RVEEH RPAH Contour Advance with 40%w/w DXB 8 2 Contour Advance (no DXB) 30 6 Objectives: 1. Ease of use and effectiveness 2. Safety in controlled adult population Subjects: Evaluation Schedule: Pre-op Surgery +1wk +2wk +1, 3, 6, 12, 24M 4-point (local) impedance X X X X Speech perception X 3 & 12M
  • 63. creating sound valueTM Interim Findings Passive delivery of DXB is safe for humans, as demonstrated by in vitro and pre-clinical investigations. A modified commercial device has been used in a FTIH study and has demonstrated initial benefits in impedance
  • 64. creating sound valueTM Controlled Drug Delivery with Hollow Fibre Systems • Controlled membrane porosity by redox induced conformational change • Solvated ion flux activated by doping / dedoping process Polymer doping / dedoping induced release of drug drug drug drug drug drug drug drug drug drug drug drug drug drug drug drug drug drug drug drug drug drug drug drug drug drug drug drug Cochlear implant based steroid delivery drug drug drug drug drug drug 0 0.01 0.02 0.03 0.04 0.05 0.06 0.07 0.08 0 10 20 30 40 Time / minutes ABSat554nm 3minPPyPVDF 3minPPyPVDF pulsed potential
  • 65. creating sound valueTMcreating sound valueTM 1. Electrosynthesis of polypyrrole incorporating Dexamethsasone 21-phosphate disodium salt under different conditions and their relation with Dex release. Current Density 0.5mA/cm2 1.0mA/cm2 1.5mA/cm2 A. ~81nm B. ~124nm C. ~198nm 5mM Roughness (RMS) D. ~69.8nm E. ~121nm 10mM F. ~68nm 20mM Dex Concentration Tab 1. Film growth condition vs. film roughness ([Py]=0.2M, Q=1C/cm2) 0 10 20 30 40 50 60 70 80 90 0 50 100 150 200 250 300 DexPReleased(%) Time (min) A (0.5mA/cm2, Dex 5mM) B (1.0mA/cm2, Dex 5mM) C (1.5mA/cm2, Dex 5mM) D (1.0mA/cm2, Dex 10mM) E (1.5mA/cm2, Dex 10mM) F (1.5mA/cm2, Dex 20mM) Fig 1. Dex release efficiency vs. different film growth conditions (Potential: -500mV, Drug loading: 380ug/cm2) • With higher current density and lower Dex concentration, higher roughness obtained • 80% of Dex released out of the film with highest roughness Drug Elution with Electrical Stimulation
  • 66. creating sound valueTM CA MRA Focus in Electrode Design Focus of electrode design on: • preservation of internal cochlear structures • positioning to reduce spread of current • designs for electro-acoustic stimuli MRA MRA CA
  • 67. creating sound valueTM • preservation of acoustic residual hearing can be accomplished with good electrode design and attention to surgical procedure ~ but outcomes remain unpredictable •a drug eluting electrode is a feasible vehicle for delivery of therapeutic agents into the cochlea, (but must not compromise insertion characteristics) • elution rates in passive designs are dependent on drug loading and processes of formulation •active systems may enable similar drug load delivery through a smaller cross-sectional area electrode array Summary
  • 68. creating sound valueTM SYDNEY CLUSTER MELBOURNE CLUSTER BRISBANE CLUSTER HEARing CRC Members 68 hear and say centre This research was financially supported by the HEARing CRC established and supported under the Australian Government’s Cooperative Research Centres Program www.hearingcrc.org