We are from Adyar Cancer Institute Chennai 20. A magnificent
hospital providing state of the art cancer care for over 100,000
patients annually. Celebrating Diamond Jubilee and it has several
firsts.
1
BY
G. Yogananthem and C. Palanivelu
Cancer Institute, Adyar, Chennai
2
Medical imaging technique is well advanced today
and we are measuring signals from protons level in
MRI imaging. But we left behind a error in 2-D
measurement in digital X- ray in imaging. This we
found out during imaging of Scanogram. In this
presentation we are explaining that how we are
controlling this error by applying small trick for
given plane interest on the digital X-ray Scanogram
image.
2-D measurement displayed in the Digital X-ray monitor is incorrect
4
For experimental proof, we developed a wax phantom by keeping 5 lead balls in
plane such that 100 mm distance is kept between the first and fifth lead ball
and the plane is at unknown height from the bottom (the plane of interest).
5
To prove geometrical error we did the imaging in DX and CT on the
same phantom.
Getting ready CT AP Scout Getting ready DX AP view
6
CT AP SCOUT VIEW OF THE PHANTOM.
We did AP scout view
of this phantom and
the distance between
the first and fifth lead
ball is measured in
straight line and the
distance is 100 mm.
Due to slit beam
technology there will
not be any geometrical
error. The measured
distance is displayed at
bottom of the monitor.
7
Digital X-ray AP view
This image is obtained
by AP view of same wax
phantom exposed in
digital X-ray equipment
(Defineium TM 6000)
The measured distance
is 114.2 mm and not
100mm. Due to
geometrical error it is
14.2% more .
8
scanogram
Scanogram is a special
type of X-ray to
measure the bone
length. In orthopedic
field the Scanogram
are used to measure
real size or length of
the bone. An accurate
and reproducible
assessment of limb
length is required for
successful treatment.
Scanogram by using
conventional x-ray
equipment
Geometrical error exists
from conventional X-ray
period and all technical
people are managing it
by keeping lead rulers at
anatomy level and 6 feet
film focal distance. They
measuring the length of
the long bone with the
help of scale projected
over the film or image.
Digital X rays are widely used and images on the films are either magnified or
reduced. With the help of workstations and CPU the distance of the required
images are measured and these measurements are inaccurate and discomfort
of taking scanogram started here and we are not able increase FFD beyond 40
inches in table x ray. Because of these reasons many of our orthopedic doctors
are not ready to give up conventional x ray machines.
X-Ray Image
The digital medical
X-ray image is
superimposed
image of multiple
planes our body.
In digital X-ray equipment, 2D Measurements are inaccurate
since they are measured at a detector level and not at an
anatomical level.
13
The plane away from the detector will have higher magnification
and plane near to the detector will have least magnification. Tan
angle formula can be used to know correct 2-D measurement.
14
We have developed a technique , for long bones, where we can get images
without geometrical errors and images are imaged in the film as real size for
the given plane of interest. We can measure the actual length of the given
anatomy with the help of an ordinary measuring ruler on the film itself .
15
We can call this as real-size imaging, when the virtual ruler readings on the film
matches (coincide) with external ruler for the given plane of interest.
Virtual ruler in the DX image
External ruler readings matches
with the virtual readings
16
REAL-SIZE IMAGING IN FOUR STEPS BY USING DX
1. DESIGN A LEAD RULER .
2. SELECT THE PLANE OF INTEREST
3. POSITION THE ANATOMY
4. CONTROL ZOOMING ON THE MONITOR
17
STEP 1. Take a lead ruler and keep it over a processed film from laser
camera as shown below. Measure the width of exposed area of the film
and keep two pointers or markers such that distance between is equal to
the width of the exposing area of the film by omitting the border.
18
STEP 2. Select the plane of interest by using axial cut of CT image or
lateral CT Scout image.
19
STEP 3. Position the anatomy (Long bone) parallel to the detector by keeping
pads. Keep the designed lead ruler by side at the level of plane of interest as
shown below and expose. This will help us to get parallel or equal magnification
of long bone and lead ruler and thus we get some control or idea of
magnification of the plane of interest.
20
STEP 4. Expose the images in landscape format by using laser camera. The
arrows shows the pointer or marker location on the X-ray image. Now we have
to zoom the images in such a way so that both the pointers should touch or
reach the upper and lower limits of film format. Then give print in order to get
real size imaging film from the camera.
21
This is video demo of zooming of exposed images on
laser film format by using advanced work station.
22
This is one of the real size imaged film from the camera. We
brought same sample films for display
23
CT images do not have geometrical error. Matching image width and film
format width we can produce real size imaging on the film by integrating
laser camera and advance work station by feeding simple software. The
following film is an example real size images on the film did manually.
 This CT film shows cerebral
hemorrhage in multiple film
formats. All images are equally
zoomed in real size. The size of
the hematoma will be easily
measured with the help of
ordinary ruler on the film itself
for the treatment and to
comparing with the CT films
repeated later.
 The same method is adopted
to get the multiple real size
images on the same film for
MRI images also.
CT guided J needle biopsy is the procedure of taking tissue for study
in operation theater. For this, one lead marker is placed on the body
and one axial cut is taken. Radiologist marks the entry and distance
for the procedure. Only one measurement is on the film. In some
cases surgeon may select another entry at the last minute. In such
cases , real time image will help them to select another entry and
this real time film will help to select and measure on the film itself by
the ruler
The below picture shows comparison of different modality images on
single film with real size imaging. The breath of film is matched with
lengths of images. Each modality is best at giving its own details
26
We applied for patent right for this methodology and it is
under process.
WE PROVED 2-D MEASUREMENT IN DX AT DECTECTOR LEVEL
IS IN CORRECT AND KINDLY UPLOAD TAN ANGLE SOFT WARE
SO THAT CUSTERMERS CAN AWARE AT WHAT HEIGHT FROM
DETECTOR ARE MEASURING.
KINDLY INTAGRATE LAZER CAMERA AND ADVANCE WORK
STATITION SO THAT AT SINGLE TOUCH TECHNOLOGIST CAN
GET REAL SIZE IMAGING.
We proved, all Digital X ray equipments provide inaccurate
2D measurements, since they are measured at detector level
and not at an anatomy level.
By feeding Tan Angle software, we can compute accurate 2-D
measurements at desired anatomy level.
By matching film size and image size we can produce real size
CT and MRI images on the film.
By matching Screen size and image size we can produce real
size images in PACS monitor also.
we are developing soft wares for all above purpose.
30
DR.V.SHANTA EXECUTIVE CHAIRMAN
CANCER INSTITUTE, ADYAR, CHENNAI, INDIA
DR.A.V.LAKSHMANAN, ADVISOR
CANCER INSITITUTE
DR.KATHIRESAN, HOD, SURGICAL ONCOLOGY
CANCER INSTITUTE
DR.G .SELVALUXMY , ADDITIONAL DIRECTOR
CANCER INSTITUTE
ALL STAFF MEMBERS, RADIOLOGY
CANCER INSTITUTE
NCISRT AND ALL MEMBERS
AND AUDIENCE
32

Weitere ähnliche Inhalte

Was ist angesagt?

Upper extremity anatomy & positioning
Upper extremity anatomy & positioningUpper extremity anatomy & positioning
Upper extremity anatomy & positioningInfoUtilRT
 
Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionparthajyotidas11
 
Flat Panel Digital Fluoroscopy by Akash Das
Flat Panel Digital Fluoroscopy by Akash Das Flat Panel Digital Fluoroscopy by Akash Das
Flat Panel Digital Fluoroscopy by Akash Das AkashDas169
 
Radiograpic views for shoulder joint
Radiograpic views  for shoulder jointRadiograpic views  for shoulder joint
Radiograpic views for shoulder jointGanesan Yogananthem
 
Radiographic positioning of Upper limb (ELBOW & HUMERUS)
Radiographic positioning of Upper limb (ELBOW & HUMERUS)Radiographic positioning of Upper limb (ELBOW & HUMERUS)
Radiographic positioning of Upper limb (ELBOW & HUMERUS)Nasir Mohiudin
 
Mri coils
Mri coilsMri coils
Mri coilsSelf
 
Mammographic equipment
Mammographic equipmentMammographic equipment
Mammographic equipmentSharmaRajan4
 
Ct instrumentation and types of detector configuration
Ct instrumentation and types of detector configurationCt instrumentation and types of detector configuration
Ct instrumentation and types of detector configurationSUJAN KARKI
 
Image Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh ShresthaImage Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh ShresthaAvinesh Shrestha
 
Fluoroscopy presentation
Fluoroscopy presentationFluoroscopy presentation
Fluoroscopy presentationHuzaifa Oxford
 
Mammography positioning technique for Cranio Caudal (CC)
Mammography positioning technique for Cranio Caudal (CC) Mammography positioning technique for Cranio Caudal (CC)
Mammography positioning technique for Cranio Caudal (CC) Selin Prasad
 
Arthrography of the Shoulder, Ankle and Wrist.pptx
Arthrography of the Shoulder, Ankle and Wrist.pptxArthrography of the Shoulder, Ankle and Wrist.pptx
Arthrography of the Shoulder, Ankle and Wrist.pptxDeneicer Guy
 
Image reconstrsuction in ct pdf
Image reconstrsuction in ct pdfImage reconstrsuction in ct pdf
Image reconstrsuction in ct pdfmitians
 
Physics of Multidetector CT Scan
Physics of Multidetector CT ScanPhysics of Multidetector CT Scan
Physics of Multidetector CT ScanDr Varun Bansal
 
Computed Radiography and digital radiography
Computed Radiography and  digital radiographyComputed Radiography and  digital radiography
Computed Radiography and digital radiographyDurga Singh
 

Was ist angesagt? (20)

Upper extremity anatomy & positioning
Upper extremity anatomy & positioningUpper extremity anatomy & positioning
Upper extremity anatomy & positioning
 
Dexa scan
Dexa scanDexa scan
Dexa scan
 
Mammography
MammographyMammography
Mammography
 
Ct image artifacts
Ct image artifactsCt image artifacts
Ct image artifacts
 
Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortion
 
Flat Panel Digital Fluoroscopy by Akash Das
Flat Panel Digital Fluoroscopy by Akash Das Flat Panel Digital Fluoroscopy by Akash Das
Flat Panel Digital Fluoroscopy by Akash Das
 
Radiograpic views for shoulder joint
Radiograpic views  for shoulder jointRadiograpic views  for shoulder joint
Radiograpic views for shoulder joint
 
Radiographic positioning of Upper limb (ELBOW & HUMERUS)
Radiographic positioning of Upper limb (ELBOW & HUMERUS)Radiographic positioning of Upper limb (ELBOW & HUMERUS)
Radiographic positioning of Upper limb (ELBOW & HUMERUS)
 
Quality assurance
Quality  assuranceQuality  assurance
Quality assurance
 
Mri coils
Mri coilsMri coils
Mri coils
 
Mammographic equipment
Mammographic equipmentMammographic equipment
Mammographic equipment
 
Ct instrumentation and types of detector configuration
Ct instrumentation and types of detector configurationCt instrumentation and types of detector configuration
Ct instrumentation and types of detector configuration
 
Image Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh ShresthaImage Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
 
CT RADIATION DOSE REDUCTION
CT RADIATION DOSE REDUCTION CT RADIATION DOSE REDUCTION
CT RADIATION DOSE REDUCTION
 
Fluoroscopy presentation
Fluoroscopy presentationFluoroscopy presentation
Fluoroscopy presentation
 
Mammography positioning technique for Cranio Caudal (CC)
Mammography positioning technique for Cranio Caudal (CC) Mammography positioning technique for Cranio Caudal (CC)
Mammography positioning technique for Cranio Caudal (CC)
 
Arthrography of the Shoulder, Ankle and Wrist.pptx
Arthrography of the Shoulder, Ankle and Wrist.pptxArthrography of the Shoulder, Ankle and Wrist.pptx
Arthrography of the Shoulder, Ankle and Wrist.pptx
 
Image reconstrsuction in ct pdf
Image reconstrsuction in ct pdfImage reconstrsuction in ct pdf
Image reconstrsuction in ct pdf
 
Physics of Multidetector CT Scan
Physics of Multidetector CT ScanPhysics of Multidetector CT Scan
Physics of Multidetector CT Scan
 
Computed Radiography and digital radiography
Computed Radiography and  digital radiographyComputed Radiography and  digital radiography
Computed Radiography and digital radiography
 

Andere mochten auch

The bleeding frogs
The bleeding frogsThe bleeding frogs
The bleeding frogskruegs44
 
Alpha beta and_gamma
Alpha beta and_gammaAlpha beta and_gamma
Alpha beta and_gammabillbarker830
 
Factors affecting Quality and Quantity of X-ray beam
Factors affecting Quality and Quantity of X-ray beamFactors affecting Quality and Quantity of X-ray beam
Factors affecting Quality and Quantity of X-ray beamVinay Desai
 
Image quality
Image qualityImage quality
Image qualityairwave12
 
Specialised Techniques in Oral Radiology
Specialised Techniques in Oral RadiologySpecialised Techniques in Oral Radiology
Specialised Techniques in Oral RadiologyIAU Dent
 
Examination in paediatric medicine
Examination in paediatric medicineExamination in paediatric medicine
Examination in paediatric medicineVarsha Shah
 
Radioactivity
RadioactivityRadioactivity
RadioactivityE H Annex
 
Limb length discrepancy evaluation
Limb length discrepancy evaluationLimb length discrepancy evaluation
Limb length discrepancy evaluationAbdulla Kamal
 
Radiographic exposure and image quality
Radiographic exposure and image qualityRadiographic exposure and image quality
Radiographic exposure and image qualityRad Tech
 
Radioactivity
RadioactivityRadioactivity
Radioactivitywilsone
 
Components of CT Scan Machine
Components of CT Scan MachineComponents of CT Scan Machine
Components of CT Scan MachineSyed Jamal
 
How to Make Awesome SlideShares: Tips & Tricks
How to Make Awesome SlideShares: Tips & TricksHow to Make Awesome SlideShares: Tips & Tricks
How to Make Awesome SlideShares: Tips & TricksSlideShare
 
Getting Started With SlideShare
Getting Started With SlideShareGetting Started With SlideShare
Getting Started With SlideShareSlideShare
 

Andere mochten auch (16)

The bleeding frogs
The bleeding frogsThe bleeding frogs
The bleeding frogs
 
Long bone measurement
Long bone measurementLong bone measurement
Long bone measurement
 
Alpha beta and_gamma
Alpha beta and_gammaAlpha beta and_gamma
Alpha beta and_gamma
 
Factors affecting Quality and Quantity of X-ray beam
Factors affecting Quality and Quantity of X-ray beamFactors affecting Quality and Quantity of X-ray beam
Factors affecting Quality and Quantity of X-ray beam
 
Image quality
Image qualityImage quality
Image quality
 
Ideal radiography
Ideal radiographyIdeal radiography
Ideal radiography
 
Specialised Techniques in Oral Radiology
Specialised Techniques in Oral RadiologySpecialised Techniques in Oral Radiology
Specialised Techniques in Oral Radiology
 
Radioactivity.ppt
Radioactivity.pptRadioactivity.ppt
Radioactivity.ppt
 
Examination in paediatric medicine
Examination in paediatric medicineExamination in paediatric medicine
Examination in paediatric medicine
 
Radioactivity
RadioactivityRadioactivity
Radioactivity
 
Limb length discrepancy evaluation
Limb length discrepancy evaluationLimb length discrepancy evaluation
Limb length discrepancy evaluation
 
Radiographic exposure and image quality
Radiographic exposure and image qualityRadiographic exposure and image quality
Radiographic exposure and image quality
 
Radioactivity
RadioactivityRadioactivity
Radioactivity
 
Components of CT Scan Machine
Components of CT Scan MachineComponents of CT Scan Machine
Components of CT Scan Machine
 
How to Make Awesome SlideShares: Tips & Tricks
How to Make Awesome SlideShares: Tips & TricksHow to Make Awesome SlideShares: Tips & Tricks
How to Make Awesome SlideShares: Tips & Tricks
 
Getting Started With SlideShare
Getting Started With SlideShareGetting Started With SlideShare
Getting Started With SlideShare
 

Ähnlich wie Digital X-ray Problems..

Towards fine-precision automated immobilization in maskless radiosurgery
Towards fine-precision automated immobilization in maskless radiosurgeryTowards fine-precision automated immobilization in maskless radiosurgery
Towards fine-precision automated immobilization in maskless radiosurgeryOlalekan Ogunmolu
 
Dominion Vi 3D
Dominion Vi 3DDominion Vi 3D
Dominion Vi 3Dhjuhan
 
Welcome to International Journal of Engineering Research and Development (IJERD)
Welcome to International Journal of Engineering Research and Development (IJERD)Welcome to International Journal of Engineering Research and Development (IJERD)
Welcome to International Journal of Engineering Research and Development (IJERD)IJERD Editor
 
COMPUTER APPLICATIONS IN DENTISTRY.pptx
COMPUTER APPLICATIONS IN DENTISTRY.pptxCOMPUTER APPLICATIONS IN DENTISTRY.pptx
COMPUTER APPLICATIONS IN DENTISTRY.pptxParkhiBhatngar
 
Cinematic rendering (2).pptx
Cinematic rendering (2).pptxCinematic rendering (2).pptx
Cinematic rendering (2).pptxCamrynHarris1
 
Recent advances in diagnosis and treatment planning1 /certified fixed orthod...
Recent advances in diagnosis and treatment  planning1 /certified fixed orthod...Recent advances in diagnosis and treatment  planning1 /certified fixed orthod...
Recent advances in diagnosis and treatment planning1 /certified fixed orthod...Indian dental academy
 
LCU RDG 402 PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
LCU RDG 402  PRINCIPLES OF COMPUTED TOMOGRAPHY.pptxLCU RDG 402  PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
LCU RDG 402 PRINCIPLES OF COMPUTED TOMOGRAPHY.pptxEmmanuelOluseyi1
 
Recent advances in diagnosis and treatment planning1 /certified fixed orthod...
Recent advances in diagnosis and treatment  planning1 /certified fixed orthod...Recent advances in diagnosis and treatment  planning1 /certified fixed orthod...
Recent advances in diagnosis and treatment planning1 /certified fixed orthod...Indian dental academy
 
Cephalometry (2) /certified fixed orthodontic courses by Indian dental academy
Cephalometry (2) /certified fixed orthodontic courses by Indian dental academy Cephalometry (2) /certified fixed orthodontic courses by Indian dental academy
Cephalometry (2) /certified fixed orthodontic courses by Indian dental academy Indian dental academy
 
StereoCheck, StereoCheckRM, and CheckBRW - Stereotactic Coordinates Verification
StereoCheck, StereoCheckRM, and CheckBRW - Stereotactic Coordinates VerificationStereoCheck, StereoCheckRM, and CheckBRW - Stereotactic Coordinates Verification
StereoCheck, StereoCheckRM, and CheckBRW - Stereotactic Coordinates VerificationArmando Alaminos Bouza
 
EOS in MEDIC presented in Thailand, Dr NGUYEN VAN CONG, MEDIC MEDICAL CENTER
EOS in MEDIC presented in Thailand, Dr NGUYEN VAN CONG, MEDIC MEDICAL CENTEREOS in MEDIC presented in Thailand, Dr NGUYEN VAN CONG, MEDIC MEDICAL CENTER
EOS in MEDIC presented in Thailand, Dr NGUYEN VAN CONG, MEDIC MEDICAL CENTERhungnguyenthien
 
Cone Beam Computed Tomography.pptx
Cone Beam Computed Tomography.pptxCone Beam Computed Tomography.pptx
Cone Beam Computed Tomography.pptxDrAnand22
 
Vision based non-invasive tool for facial swelling assessment
Vision based non-invasive tool for facial swelling assessment Vision based non-invasive tool for facial swelling assessment
Vision based non-invasive tool for facial swelling assessment University of Moratuwa
 
Go 2 d3d ceph_cataloge
Go 2 d3d ceph_catalogeGo 2 d3d ceph_cataloge
Go 2 d3d ceph_catalogeYenDoan19
 

Ähnlich wie Digital X-ray Problems.. (20)

Towards fine-precision automated immobilization in maskless radiosurgery
Towards fine-precision automated immobilization in maskless radiosurgeryTowards fine-precision automated immobilization in maskless radiosurgery
Towards fine-precision automated immobilization in maskless radiosurgery
 
Dominion Vi 3D
Dominion Vi 3DDominion Vi 3D
Dominion Vi 3D
 
Welcome to International Journal of Engineering Research and Development (IJERD)
Welcome to International Journal of Engineering Research and Development (IJERD)Welcome to International Journal of Engineering Research and Development (IJERD)
Welcome to International Journal of Engineering Research and Development (IJERD)
 
COMPUTER APPLICATIONS IN DENTISTRY.pptx
COMPUTER APPLICATIONS IN DENTISTRY.pptxCOMPUTER APPLICATIONS IN DENTISTRY.pptx
COMPUTER APPLICATIONS IN DENTISTRY.pptx
 
Cinematic rendering (2).pptx
Cinematic rendering (2).pptxCinematic rendering (2).pptx
Cinematic rendering (2).pptx
 
Recent advances in diagnosis and treatment planning1 /certified fixed orthod...
Recent advances in diagnosis and treatment  planning1 /certified fixed orthod...Recent advances in diagnosis and treatment  planning1 /certified fixed orthod...
Recent advances in diagnosis and treatment planning1 /certified fixed orthod...
 
IMAGE GUIDED SURGERY
IMAGE GUIDED SURGERYIMAGE GUIDED SURGERY
IMAGE GUIDED SURGERY
 
LCU RDG 402 PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
LCU RDG 402  PRINCIPLES OF COMPUTED TOMOGRAPHY.pptxLCU RDG 402  PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
LCU RDG 402 PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
 
Recent advances in diagnosis and treatment planning1 /certified fixed orthod...
Recent advances in diagnosis and treatment  planning1 /certified fixed orthod...Recent advances in diagnosis and treatment  planning1 /certified fixed orthod...
Recent advances in diagnosis and treatment planning1 /certified fixed orthod...
 
Cephalometry (2) /certified fixed orthodontic courses by Indian dental academy
Cephalometry (2) /certified fixed orthodontic courses by Indian dental academy Cephalometry (2) /certified fixed orthodontic courses by Indian dental academy
Cephalometry (2) /certified fixed orthodontic courses by Indian dental academy
 
IRJET-V9I1137.pdf
IRJET-V9I1137.pdfIRJET-V9I1137.pdf
IRJET-V9I1137.pdf
 
StereoCheck, StereoCheckRM, and CheckBRW - Stereotactic Coordinates Verification
StereoCheck, StereoCheckRM, and CheckBRW - Stereotactic Coordinates VerificationStereoCheck, StereoCheckRM, and CheckBRW - Stereotactic Coordinates Verification
StereoCheck, StereoCheckRM, and CheckBRW - Stereotactic Coordinates Verification
 
EOS in MEDIC presented in Thailand, Dr NGUYEN VAN CONG, MEDIC MEDICAL CENTER
EOS in MEDIC presented in Thailand, Dr NGUYEN VAN CONG, MEDIC MEDICAL CENTEREOS in MEDIC presented in Thailand, Dr NGUYEN VAN CONG, MEDIC MEDICAL CENTER
EOS in MEDIC presented in Thailand, Dr NGUYEN VAN CONG, MEDIC MEDICAL CENTER
 
Imaging in periodontics
Imaging in periodonticsImaging in periodontics
Imaging in periodontics
 
7200 35328
7200 353287200 35328
7200 35328
 
Cone Beam Computed Tomography.pptx
Cone Beam Computed Tomography.pptxCone Beam Computed Tomography.pptx
Cone Beam Computed Tomography.pptx
 
Vision based non-invasive tool for facial swelling assessment
Vision based non-invasive tool for facial swelling assessment Vision based non-invasive tool for facial swelling assessment
Vision based non-invasive tool for facial swelling assessment
 
Go 2 d3d ceph_cataloge
Go 2 d3d ceph_catalogeGo 2 d3d ceph_cataloge
Go 2 d3d ceph_cataloge
 
CT Scanner Trends
CT Scanner TrendsCT Scanner Trends
CT Scanner Trends
 
2011BMES Abstract radDE5DE
2011BMES Abstract radDE5DE2011BMES Abstract radDE5DE
2011BMES Abstract radDE5DE
 

Mehr von Ganesan Yogananthem

nuclear medicine 10 marks questions and answers.docx
nuclear medicine 10 marks questions and answers.docxnuclear medicine 10 marks questions and answers.docx
nuclear medicine 10 marks questions and answers.docxGanesan Yogananthem
 
nuclear medicine 3 marks questions and answers.pdf
nuclear medicine 3 marks questions and answers.pdfnuclear medicine 3 marks questions and answers.pdf
nuclear medicine 3 marks questions and answers.pdfGanesan Yogananthem
 
MAGNETIC RESONANCE IMAGING QUESTION AND ANSWER 3 MARKS.pdf
MAGNETIC RESONANCE IMAGING QUESTION AND ANSWER 3 MARKS.pdfMAGNETIC RESONANCE IMAGING QUESTION AND ANSWER 3 MARKS.pdf
MAGNETIC RESONANCE IMAGING QUESTION AND ANSWER 3 MARKS.pdfGanesan Yogananthem
 
QUESTIONS & ANSWERS FOR QUALITY ASSURANCE, RADIATIONBIOLOGY& RADIATION HAZARD...
QUESTIONS & ANSWERS FOR QUALITY ASSURANCE, RADIATIONBIOLOGY& RADIATION HAZARD...QUESTIONS & ANSWERS FOR QUALITY ASSURANCE, RADIATIONBIOLOGY& RADIATION HAZARD...
QUESTIONS & ANSWERS FOR QUALITY ASSURANCE, RADIATIONBIOLOGY& RADIATION HAZARD...Ganesan Yogananthem
 
MRI 3-Marks QUESTIONS AND ANSWERS
MRI 3-Marks QUESTIONS AND ANSWERSMRI 3-Marks QUESTIONS AND ANSWERS
MRI 3-Marks QUESTIONS AND ANSWERSGanesan Yogananthem
 
CT SCAN - 3 Marks - QUESTION AND ANSWERS
CT SCAN - 3 Marks - QUESTION AND ANSWERSCT SCAN - 3 Marks - QUESTION AND ANSWERS
CT SCAN - 3 Marks - QUESTION AND ANSWERSGanesan Yogananthem
 
Ultrasound assignment 3 mark question and answer
Ultrasound assignment 3 mark question and answer Ultrasound assignment 3 mark question and answer
Ultrasound assignment 3 mark question and answer Ganesan Yogananthem
 
Radiology procedure questions and answer 1
Radiology procedure questions and answer 1Radiology procedure questions and answer 1
Radiology procedure questions and answer 1Ganesan Yogananthem
 
Radiological procedure questions and answers2 converted
Radiological procedure questions and answers2 convertedRadiological procedure questions and answers2 converted
Radiological procedure questions and answers2 convertedGanesan Yogananthem
 
B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING
B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING
B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING Ganesan Yogananthem
 
BIOLOGICAL EFFECTS OF IONIZING RADIATION
BIOLOGICAL EFFECTS OF IONIZING RADIATIONBIOLOGICAL EFFECTS OF IONIZING RADIATION
BIOLOGICAL EFFECTS OF IONIZING RADIATIONGanesan Yogananthem
 
CT numbers, window width and window level
CT numbers, window width and window levelCT numbers, window width and window level
CT numbers, window width and window levelGanesan Yogananthem
 
RADIOGRAPHIC VIEWS FOR HIP JOINT
RADIOGRAPHIC VIEWS FOR HIP JOINTRADIOGRAPHIC VIEWS FOR HIP JOINT
RADIOGRAPHIC VIEWS FOR HIP JOINTGanesan Yogananthem
 
RADIOGRAPHIC VIEWS OF ANKLE JOINT
RADIOGRAPHIC VIEWS OF ANKLE JOINTRADIOGRAPHIC VIEWS OF ANKLE JOINT
RADIOGRAPHIC VIEWS OF ANKLE JOINTGanesan Yogananthem
 

Mehr von Ganesan Yogananthem (18)

nuclear medicine 10 marks questions and answers.docx
nuclear medicine 10 marks questions and answers.docxnuclear medicine 10 marks questions and answers.docx
nuclear medicine 10 marks questions and answers.docx
 
nuclear medicine 3 marks questions and answers.pdf
nuclear medicine 3 marks questions and answers.pdfnuclear medicine 3 marks questions and answers.pdf
nuclear medicine 3 marks questions and answers.pdf
 
MAGNETIC RESONANCE IMAGING QUESTION AND ANSWER 3 MARKS.pdf
MAGNETIC RESONANCE IMAGING QUESTION AND ANSWER 3 MARKS.pdfMAGNETIC RESONANCE IMAGING QUESTION AND ANSWER 3 MARKS.pdf
MAGNETIC RESONANCE IMAGING QUESTION AND ANSWER 3 MARKS.pdf
 
QUESTIONS & ANSWERS FOR QUALITY ASSURANCE, RADIATIONBIOLOGY& RADIATION HAZARD...
QUESTIONS & ANSWERS FOR QUALITY ASSURANCE, RADIATIONBIOLOGY& RADIATION HAZARD...QUESTIONS & ANSWERS FOR QUALITY ASSURANCE, RADIATIONBIOLOGY& RADIATION HAZARD...
QUESTIONS & ANSWERS FOR QUALITY ASSURANCE, RADIATIONBIOLOGY& RADIATION HAZARD...
 
MRI 3-Marks QUESTIONS AND ANSWERS
MRI 3-Marks QUESTIONS AND ANSWERSMRI 3-Marks QUESTIONS AND ANSWERS
MRI 3-Marks QUESTIONS AND ANSWERS
 
CT SCAN - 3 Marks - QUESTION AND ANSWERS
CT SCAN - 3 Marks - QUESTION AND ANSWERSCT SCAN - 3 Marks - QUESTION AND ANSWERS
CT SCAN - 3 Marks - QUESTION AND ANSWERS
 
Ultrasound assignment 3 mark question and answer
Ultrasound assignment 3 mark question and answer Ultrasound assignment 3 mark question and answer
Ultrasound assignment 3 mark question and answer
 
Radiology procedure questions and answer 1
Radiology procedure questions and answer 1Radiology procedure questions and answer 1
Radiology procedure questions and answer 1
 
Radiological procedure questions and answers2 converted
Radiological procedure questions and answers2 convertedRadiological procedure questions and answers2 converted
Radiological procedure questions and answers2 converted
 
B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING
B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING
B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING
 
BIOLOGICAL EFFECTS OF IONIZING RADIATION
BIOLOGICAL EFFECTS OF IONIZING RADIATIONBIOLOGICAL EFFECTS OF IONIZING RADIATION
BIOLOGICAL EFFECTS OF IONIZING RADIATION
 
CT numbers, window width and window level
CT numbers, window width and window levelCT numbers, window width and window level
CT numbers, window width and window level
 
CT artifact
CT artifact CT artifact
CT artifact
 
RADIOGRAPHIC VIEWS FOR HIP JOINT
RADIOGRAPHIC VIEWS FOR HIP JOINTRADIOGRAPHIC VIEWS FOR HIP JOINT
RADIOGRAPHIC VIEWS FOR HIP JOINT
 
RADIOGRAPHIC VIEWS OF ANKLE JOINT
RADIOGRAPHIC VIEWS OF ANKLE JOINTRADIOGRAPHIC VIEWS OF ANKLE JOINT
RADIOGRAPHIC VIEWS OF ANKLE JOINT
 
BASIC MRI SEQUENCES
BASIC MRI SEQUENCESBASIC MRI SEQUENCES
BASIC MRI SEQUENCES
 
Portal venous doppler
Portal venous dopplerPortal venous doppler
Portal venous doppler
 
CSF flow study
CSF flow studyCSF flow study
CSF flow study
 

Kürzlich hochgeladen

Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetChandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meetpriyashah722354
 
dhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
dhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetdhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
dhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Chandigarh Escorts, 😋9988299661 😋50% off at Escort Service in Chandigarh
Chandigarh Escorts, 😋9988299661 😋50% off at Escort Service in ChandigarhChandigarh Escorts, 😋9988299661 😋50% off at Escort Service in Chandigarh
Chandigarh Escorts, 😋9988299661 😋50% off at Escort Service in ChandigarhSheetaleventcompany
 
Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★
Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★
Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★indiancallgirl4rent
 
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetSambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Call Girl Raipur 📲 9999965857 whatsapp live cam sex service available
Call Girl Raipur 📲 9999965857 whatsapp live cam sex service availableCall Girl Raipur 📲 9999965857 whatsapp live cam sex service available
Call Girl Raipur 📲 9999965857 whatsapp live cam sex service availablegragmanisha42
 
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591adityaroy0215
 
VIP Call Girl Sector 10 Noida Call Me: 9711199171
VIP Call Girl Sector 10 Noida Call Me: 9711199171VIP Call Girl Sector 10 Noida Call Me: 9711199171
VIP Call Girl Sector 10 Noida Call Me: 9711199171Call Girls Service Gurgaon
 
Call Girls Service Faridabad 📲 9999965857 ヅ10k NiGhT Call Girls In Faridabad
Call Girls Service Faridabad 📲 9999965857 ヅ10k NiGhT Call Girls In FaridabadCall Girls Service Faridabad 📲 9999965857 ヅ10k NiGhT Call Girls In Faridabad
Call Girls Service Faridabad 📲 9999965857 ヅ10k NiGhT Call Girls In Faridabadgragmanisha42
 
Call Girls Patiala Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Patiala Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Patiala Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Patiala Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅
Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅
Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅gragmanisha42
 
bhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
bhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetbhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
bhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near MeVIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Memriyagarg453
 
Call Girl Amritsar ❤️♀️@ 8725944379 Amritsar Call Girls Near Me ❤️♀️@ Sexy Ca...
Call Girl Amritsar ❤️♀️@ 8725944379 Amritsar Call Girls Near Me ❤️♀️@ Sexy Ca...Call Girl Amritsar ❤️♀️@ 8725944379 Amritsar Call Girls Near Me ❤️♀️@ Sexy Ca...
Call Girl Amritsar ❤️♀️@ 8725944379 Amritsar Call Girls Near Me ❤️♀️@ Sexy Ca...Sheetaleventcompany
 
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetErnakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Chandigarh
 
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF ...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF  ...❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF  ...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF ...Gfnyt.com
 
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetnagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetOzhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetNanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 

Kürzlich hochgeladen (20)

Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetChandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
 
dhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
dhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetdhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
dhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Chandigarh Escorts, 😋9988299661 😋50% off at Escort Service in Chandigarh
Chandigarh Escorts, 😋9988299661 😋50% off at Escort Service in ChandigarhChandigarh Escorts, 😋9988299661 😋50% off at Escort Service in Chandigarh
Chandigarh Escorts, 😋9988299661 😋50% off at Escort Service in Chandigarh
 
Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★
Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★
Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★
 
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetSambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Call Girl Raipur 📲 9999965857 whatsapp live cam sex service available
Call Girl Raipur 📲 9999965857 whatsapp live cam sex service availableCall Girl Raipur 📲 9999965857 whatsapp live cam sex service available
Call Girl Raipur 📲 9999965857 whatsapp live cam sex service available
 
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
 
VIP Call Girl Sector 10 Noida Call Me: 9711199171
VIP Call Girl Sector 10 Noida Call Me: 9711199171VIP Call Girl Sector 10 Noida Call Me: 9711199171
VIP Call Girl Sector 10 Noida Call Me: 9711199171
 
Call Girls Service Faridabad 📲 9999965857 ヅ10k NiGhT Call Girls In Faridabad
Call Girls Service Faridabad 📲 9999965857 ヅ10k NiGhT Call Girls In FaridabadCall Girls Service Faridabad 📲 9999965857 ヅ10k NiGhT Call Girls In Faridabad
Call Girls Service Faridabad 📲 9999965857 ヅ10k NiGhT Call Girls In Faridabad
 
Call Girls Patiala Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Patiala Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Patiala Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Patiala Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅
Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅
Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅
 
bhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
bhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetbhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
bhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near MeVIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
 
Call Girl Amritsar ❤️♀️@ 8725944379 Amritsar Call Girls Near Me ❤️♀️@ Sexy Ca...
Call Girl Amritsar ❤️♀️@ 8725944379 Amritsar Call Girls Near Me ❤️♀️@ Sexy Ca...Call Girl Amritsar ❤️♀️@ 8725944379 Amritsar Call Girls Near Me ❤️♀️@ Sexy Ca...
Call Girl Amritsar ❤️♀️@ 8725944379 Amritsar Call Girls Near Me ❤️♀️@ Sexy Ca...
 
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetErnakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
 
❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF ...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF  ...❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF  ...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF ...
 
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetnagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetOzhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetNanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 

Digital X-ray Problems..

  • 1. We are from Adyar Cancer Institute Chennai 20. A magnificent hospital providing state of the art cancer care for over 100,000 patients annually. Celebrating Diamond Jubilee and it has several firsts. 1
  • 2. BY G. Yogananthem and C. Palanivelu Cancer Institute, Adyar, Chennai 2
  • 3. Medical imaging technique is well advanced today and we are measuring signals from protons level in MRI imaging. But we left behind a error in 2-D measurement in digital X- ray in imaging. This we found out during imaging of Scanogram. In this presentation we are explaining that how we are controlling this error by applying small trick for given plane interest on the digital X-ray Scanogram image.
  • 4. 2-D measurement displayed in the Digital X-ray monitor is incorrect 4
  • 5. For experimental proof, we developed a wax phantom by keeping 5 lead balls in plane such that 100 mm distance is kept between the first and fifth lead ball and the plane is at unknown height from the bottom (the plane of interest). 5
  • 6. To prove geometrical error we did the imaging in DX and CT on the same phantom. Getting ready CT AP Scout Getting ready DX AP view 6
  • 7. CT AP SCOUT VIEW OF THE PHANTOM. We did AP scout view of this phantom and the distance between the first and fifth lead ball is measured in straight line and the distance is 100 mm. Due to slit beam technology there will not be any geometrical error. The measured distance is displayed at bottom of the monitor. 7
  • 8. Digital X-ray AP view This image is obtained by AP view of same wax phantom exposed in digital X-ray equipment (Defineium TM 6000) The measured distance is 114.2 mm and not 100mm. Due to geometrical error it is 14.2% more . 8
  • 9. scanogram Scanogram is a special type of X-ray to measure the bone length. In orthopedic field the Scanogram are used to measure real size or length of the bone. An accurate and reproducible assessment of limb length is required for successful treatment.
  • 10. Scanogram by using conventional x-ray equipment Geometrical error exists from conventional X-ray period and all technical people are managing it by keeping lead rulers at anatomy level and 6 feet film focal distance. They measuring the length of the long bone with the help of scale projected over the film or image.
  • 11. Digital X rays are widely used and images on the films are either magnified or reduced. With the help of workstations and CPU the distance of the required images are measured and these measurements are inaccurate and discomfort of taking scanogram started here and we are not able increase FFD beyond 40 inches in table x ray. Because of these reasons many of our orthopedic doctors are not ready to give up conventional x ray machines.
  • 12. X-Ray Image The digital medical X-ray image is superimposed image of multiple planes our body.
  • 13. In digital X-ray equipment, 2D Measurements are inaccurate since they are measured at a detector level and not at an anatomical level. 13
  • 14. The plane away from the detector will have higher magnification and plane near to the detector will have least magnification. Tan angle formula can be used to know correct 2-D measurement. 14
  • 15. We have developed a technique , for long bones, where we can get images without geometrical errors and images are imaged in the film as real size for the given plane of interest. We can measure the actual length of the given anatomy with the help of an ordinary measuring ruler on the film itself . 15
  • 16. We can call this as real-size imaging, when the virtual ruler readings on the film matches (coincide) with external ruler for the given plane of interest. Virtual ruler in the DX image External ruler readings matches with the virtual readings 16
  • 17. REAL-SIZE IMAGING IN FOUR STEPS BY USING DX 1. DESIGN A LEAD RULER . 2. SELECT THE PLANE OF INTEREST 3. POSITION THE ANATOMY 4. CONTROL ZOOMING ON THE MONITOR 17
  • 18. STEP 1. Take a lead ruler and keep it over a processed film from laser camera as shown below. Measure the width of exposed area of the film and keep two pointers or markers such that distance between is equal to the width of the exposing area of the film by omitting the border. 18
  • 19. STEP 2. Select the plane of interest by using axial cut of CT image or lateral CT Scout image. 19
  • 20. STEP 3. Position the anatomy (Long bone) parallel to the detector by keeping pads. Keep the designed lead ruler by side at the level of plane of interest as shown below and expose. This will help us to get parallel or equal magnification of long bone and lead ruler and thus we get some control or idea of magnification of the plane of interest. 20
  • 21. STEP 4. Expose the images in landscape format by using laser camera. The arrows shows the pointer or marker location on the X-ray image. Now we have to zoom the images in such a way so that both the pointers should touch or reach the upper and lower limits of film format. Then give print in order to get real size imaging film from the camera. 21
  • 22. This is video demo of zooming of exposed images on laser film format by using advanced work station. 22
  • 23. This is one of the real size imaged film from the camera. We brought same sample films for display 23
  • 24. CT images do not have geometrical error. Matching image width and film format width we can produce real size imaging on the film by integrating laser camera and advance work station by feeding simple software. The following film is an example real size images on the film did manually.  This CT film shows cerebral hemorrhage in multiple film formats. All images are equally zoomed in real size. The size of the hematoma will be easily measured with the help of ordinary ruler on the film itself for the treatment and to comparing with the CT films repeated later.  The same method is adopted to get the multiple real size images on the same film for MRI images also.
  • 25. CT guided J needle biopsy is the procedure of taking tissue for study in operation theater. For this, one lead marker is placed on the body and one axial cut is taken. Radiologist marks the entry and distance for the procedure. Only one measurement is on the film. In some cases surgeon may select another entry at the last minute. In such cases , real time image will help them to select another entry and this real time film will help to select and measure on the film itself by the ruler
  • 26. The below picture shows comparison of different modality images on single film with real size imaging. The breath of film is matched with lengths of images. Each modality is best at giving its own details 26
  • 27.
  • 28. We applied for patent right for this methodology and it is under process.
  • 29. WE PROVED 2-D MEASUREMENT IN DX AT DECTECTOR LEVEL IS IN CORRECT AND KINDLY UPLOAD TAN ANGLE SOFT WARE SO THAT CUSTERMERS CAN AWARE AT WHAT HEIGHT FROM DETECTOR ARE MEASURING. KINDLY INTAGRATE LAZER CAMERA AND ADVANCE WORK STATITION SO THAT AT SINGLE TOUCH TECHNOLOGIST CAN GET REAL SIZE IMAGING.
  • 30. We proved, all Digital X ray equipments provide inaccurate 2D measurements, since they are measured at detector level and not at an anatomy level. By feeding Tan Angle software, we can compute accurate 2-D measurements at desired anatomy level. By matching film size and image size we can produce real size CT and MRI images on the film. By matching Screen size and image size we can produce real size images in PACS monitor also. we are developing soft wares for all above purpose. 30
  • 31. DR.V.SHANTA EXECUTIVE CHAIRMAN CANCER INSTITUTE, ADYAR, CHENNAI, INDIA DR.A.V.LAKSHMANAN, ADVISOR CANCER INSITITUTE DR.KATHIRESAN, HOD, SURGICAL ONCOLOGY CANCER INSTITUTE DR.G .SELVALUXMY , ADDITIONAL DIRECTOR CANCER INSTITUTE ALL STAFF MEMBERS, RADIOLOGY CANCER INSTITUTE
  • 32. NCISRT AND ALL MEMBERS AND AUDIENCE 32