SlideShare ist ein Scribd-Unternehmen logo
1 von 114
Dr Aftab Qadir
 Chest devices are encountered on daily basis by

radiologists
 Chest tubes, central lines, endotracheal tubes and NG
tubes are common
 New devices are constantly being introduced
 It is important to recognizer the presence of a device

and to have an understanding of its function as well as
complications associated with its use
 The proliferation of intensive care units and the

advances in the treatment of the very ill have greatly
increased the numbers of examinations performed at
the patient’s bedside.
 Obtaining a daily chest radiograph is standard practice
in most intensive care units
 Chest radiograph shows that the tip of the endotracheal tube (black arrow) is
slightly above the aortic arch and well above the carina, in good position. A right
chest tube (white arrow), ECG leads (E), a gown snap (G), and oxygen tubing (O)
are also visible.
Medical devices
 Extra thoracic devices
 Pleural devices
 Tracheal
 Esophageal

 Vascular
 Cardiac
 Circulatory assist devices
Extra thoracic devices
 Tubing, clamps, syringes
 Ventilator support tubing
 ECG electrodes
 Breast prostheses

 other apparatus often lie on or under the patient and

are imaged with the chest during the radiographic
examination
 Chest radiograph shows halo apparatus (with emergency

wrench close at hand) for cervical spine stabilization.
 ECG leads
Pleural Devices
 chest tubes are commonly used for evacuating fluid or

air from the pleural space
 Antero superiorly for pneumothorax
 Poster inferiorly for fluid collection
 Frontal (a) and lateral (b) views show a thoracostomy(chest) tube in

good position for treatment of a pneumothorax but not for an effusion.
 Chest tubes are commonly used for evacuating fluid or air from the pleural space.
 The normally positioned tube lies on the surface of the expanded lung, between the visceral
and parietal pleura.
 The tube is usually placed anterosuperiorly to evacuate a pneumothorax and
posteroinferiorly for fluid collections.
 the normally positioned tube lies on the surface of the

expanded lung, between the visceral and parietal
pleurae.
 Pigtail catheters may be used in place of standard
thoracostomy tubes, and they are popular for
empyema drainage and for installation of medication
for treatment of an empyema
 Frontal view of the chest shows a pigtail catheter that had been

inserted under fluoroscopic guidance into a loculated right
empyema for instillation of urokinase and fluid drainage.
 Assess the position of chest tube in both frontal and

lateral view and sometime may required CT scan

Complication
Malposition
 May enter the interlobar fissure
 Lung parenchyma
 Subcutaneous tissue
 Tube may be kinked
 58-year-old woman with extrapleural placement of chest

tube. Magnified anteroposterior chest radiograph shows
misplaced chest tube (arrows) within right chest wall.
 65-year-old woman with extrapleural placement of chest tube.
 A, Magnified anteroposterior chest radiograph shows left chest tube (arrow) in
apparently adequate position. CT scan was requested to further investigate
because of ineffective drainage of left pleural effusion.
 Magnified axial CT image shows misplacement of chest

tube within chest wall (arrow).
 69-year-old man with intrafissural placement of chest tube.
 A, Magnified anteroposterior chest radiograph shows

horizontal course of right chest tube (arrows).
 49-year-old man with intraparenchymal placement of

chest tube.
 A, Scout image shows chest tube (arrow) projecting
over right mid lung field.
 Magnified CT image shows chest tube (thick arrow)

coursing through right upper lobe. There is associated
small pneumothorax (asterisk) and subcutaneous
emphysema (thin arrow).
 48-year-old man with chest tube kinking. Magnified

posteroanterior chest radiograph performed after
chest tube placement shows kinking of chest tube
(arrow) precluding adequate pleural drainage.
 37-year-old man with mediastinal placement of chest tube.
 A, Anteroposterior chest radiograph shows left chest tube (arrows)

in inappropriate position, directed medially and projecting across
mediastinum. There is persistent left pleural effusion.
 37-year-old man with mediastinal placement of chest tube.

 B, CT image at level of pulmonary artery trunk confirms

that tip of chest tube (arrow) is in anterior mediastinum.
 30-year-old male victim of motor vehicle trauma with abdominal placement of

chest tube.
 A, Anteroposterior chest radiograph shows horizontally oriented chest tube
(arrow) in left lower hemithorax. There are several left rib
fractures, opacification of left hemithorax, and subcutaneous emphysema.
 30-year-old male victim of motor vehicle trauma with abdominal placement of chest tube.
 B, CT image shows traumatic left diaphragmatic rupture with migration of abdominal content
to left hemithorax. Chest tube (black arrow) is seen within mesenteric fat abutting small bowel
loops. Note splenic rupture (white arrow) related to trauma.
Tracheal Device
 Tracheal intubation is a life saving procedure but can

be life threatening if placed incorrectly
 The tip of the tube should be 5 cm above the carina
 The carina is just cauded to the aortic arch
 Chest radiograph shows that the tip of the endotracheal tube

(black arrow) is slightly above the aortic arch and well above the
carina, in good position. A right chest tube (white arrow), ECG
leads (E), a gown snap (G), and oxygen tubing (O) are also
visible.
Complications
 When advanced too far, the endotracheal tube usually

enters the right main bronchus, causing various
combinations of hyperinflation and atelectasis of the
two lungs, depending on the positions of the end and
side holes.
 endotracheal tubes can also be placed in the
esophagus
 the soft tissues of the neck.
 6-year-old woman with misplaced endotracheal tube. Magnified

anteroposterior chest radiograph shows that tip of endotracheal
tube (thick arrow) is too high, at level of thoracic inlet.
Endotracheal tube cuff (thin arrows) is overdistended. This
abnormal position may cause vocal cord injury.
 60-year-old woman with inadvertent right main bronchial

intubation. Anteroposterior chest radiograph shows endotracheal
tube tip (arrow) in right main bronchus, resulting in complete
collapse of left lung and leftward shift of mediastinum.
 Frontal view shows a double-lumen endotracheal tube

with selective intubation of the left main bronchus
(arrow).
 66-year-old man with left-sided double-lumen endotracheal tube. Magnified

anteroposterior chest radiograph shows double-lumen endotracheal tube with
its left tip (thick arrow) in left main bronchus. Right tip (thin arrow) is noted
within trachea for ventilation of right lung. Double-lumen endotracheal tube
allows control of distribution of ventilation to each lung. It is important to
differentiate between double-lumen endotracheal tubes and inadvertent
selective bronchial intubation with single-lumen catheters.
Esophageal devices
 NG tubes
 Esophageal balloons
 Esophageal stents
 PH probe
NG tubes
 Nasogastric tubes and feeding tubes are frequently

visualized passing through the mediastinum on their
way to the stomach and intestines.
 Esophageal balloons and esophageal stents used to
treat benign and malignant esophageal disease may
manifest themselves at chest radiography.
Vascular devices
Routinely used for
 Monitoring hemodynamic function
 Hemodialysis
 Administrating fluids
 Medication
 Nutrition
 Venous devices are usually inserted, either

percutaneously or surgically, via the
subclavian, internal jugular, or femoral veins. Arterial
devices usually are placed through the femoral artery
 central lines typically have one to three lumens
 A central venous catheter is ideally positioned in the
superior vena cava for the monitoring of pressure or
infusion of medication and nutrition.
 Swan-Ganz is a multilumen catheter used for

measuring hemodynamic pressures and cardiac
output.A better term to use is pulmonary artery
catheter.
 Accurate measurement of pulmonary arterial wedge
pressure
 chest radiograph shows a right subclavian single-lumen

central venous catheter and a left subcutaneous port
catheter, which enters via the left subclavian vein. Both
catheter tips are in the superior vena cava.
 The port is usually connected to a central venous

catheter or to an arterial catheter and can be used for
instillation of fluids, medications, chemotherapeutic
agents, parenteral nutritional solutions, and blood
products. It can also be used for withdrawal of blood
samples.
 Frontal view of the chest shows a left subclavian Groshong

catheter with its tip in the proximal most portion of the
superior vena cava. (b) Close up view of the catheter tip.
 Groshong catheter has closed, rounded tip Near the tip in the side of the
catheter is a three-position valve. The valve is designed to allow fluid to flow in
and out through the valve, but it remains closed when it is not in use. This
catheter does not require routine clamping or heparin solution to keep open. It
does require periodic flushing with 0.9% normal saline.
 A good rule of thumb is that the catheter tip

should be within the mediastinal shadow.
Placement more distally increases the chance of
pulmonary infarction or vessel rupture.
 Frontal chest radiograph shows a right jugular Swan-

Ganz catheter with its tip (arrow) in the right lower
pulmonary artery.
 Frontal chest radiograph shows a Swan-Ganz catheter

(white arrow) in the left pulmonary artery via the inferior
vena cava. Note also the bilateral chest tubes (black arrows)
and ECG leads (E).
 Complications of catheter insertion vary with the







catheters used and the sites employed
Pneumothorax is a common complication
Vessel lacerations and perforations can produce
hematomas, hemothorax, and infusion of fluid into
the mediastinum, thorax, or other inappropriate space
Nerve injury is usually a complication of improper
puncture technique
Looping of catheters may lead to knotting.
 Malpositioning of pulmonary artery catheter is

exceedingly common, found in approximately 25% of
catheters placed. This may lead to false readings and
an increased risk for complications.
 Complications of pulmonary artery catheter
placement include pneumothorax, pulmonary
infarction, cardiac arrhythmias, pulmonary artery
perforation, endocarditis, and sepsis
Cardiac Devices
 Cardiac pacemakers, valve prostheses, and artificial

hearts
 chest radiography is commonly employed in the
assessment of patients with heart disease
 recognition of cardiac devices and the problems
associated with them is important for all individuals
involved in the care of these patients.
Heart valves
 Mechanical
 Biologic
 Mechanical Heart valves most require life-long treatment

with anticoagulants.
 Biologic valves are less durable than mechanical
valves, with some deterioration developing, frequently 5–10
years after placement, but they do not usually require
anticoagulant treatment.
 It is not important and often impossible to know the
specific name of a particular prosthetic heart valve, but it is
important to recognize its presence
 The mitral and aortic valves are those most commonly
replaced
 Mechanical heart valve. Lateral view of the chest shows a Hemex

tilting bileaflet mechanical mitral valve prosthesis. Median
sternotomy wires and surgical clips are also evident.
 Mechanical heart valve. Lateral view of the chest shows a

Starr-Edwards caged ball mechanical mitral valve prosthesis.
 Biologic heart valve. Frontal view of the chest shows a

Hancock porcine mitral valve prosthesis (arrow). A singlelead pacemaker, ECG leads, and median sternotomy wires
are also seen.
 Biologic heart valve. Frontal view of the chest shows a

Hancock porcine valve prosthesis in a Rastelli conduit going
from the right ventricle to the pulmonary artery.
 Lateral view of the chest in an elderly patient shows a mitral annuloplasty

ring (black arrow) and a dual-lead cardiac pacemaker. Sternal
wires, surgical clips, and ECG leads are also present. The sternal wires are
used to close a sternal dehiscence. The patient has both horizontal sternal
wires and vertical intercostal wires (white arrows).
 Lateral view of the chest in a child shows an occlusion

basket (umbrella) for treatment of an atrial septal defect.
Cardiac Pacemakers
 common in older adults being treated for abnormal

cardiac rhythms caused by coronary artery disease
 Cardiac pacemakers improve cardiac function, reduce
the severity of clinical symptoms, and reduce mortality
and morbidity.
 A cardiac pacemaker is composed of two main
elements: (a) a pulse generator and (b) lead wires with
electrodes for contact with the endocardium or
myocardium
 Pacemakers range from simple temporary epicardial

electrodes to very complex pacemakers with multiple
atrial and ventricular leads
 Frontal (a) and lateral (b) views of the chest show a single electrode epicardial

“corkscrew” subxiphoid pacemaker (arrowhead in a, black arrow in b). There are
also coils (white arrow) occluding a previous right Blalock-Taussig shunt. In
addition, ECG leads and sternal wires are evident.
 Frontal (a) and lateral (b) views show an atrioventricular

sequential pacemaker with one electrode in the right atrial
appendage (RA) and the other at the right ventricular apex (RV).
Also shown are ECG leads (E) and the battery-control pack (B)
for the pacemaker.
 Biventricular devices (CRT)
 Because there can be such a wide variation in the

proper positioning of pacemaker leads, it is often
difficult for the to know if a pacemaker is properly
positioned.
 Pacemaker lead fracture is now rarely seen because of
improvements in the flexibility of the metal alloys
used in electrode construction.
Coronary Artery Surgery and Stents
 Revascularization techniques include CABG

surgery, coronary artery angioplasty, and coronary
artery stent placement.
 Median sternotomy is the usual surgical approach for
CABG surgery, and sternal wires the common method
of fixation of the two sternal segments
 At present, almost 90% of coronary interventions
include stent placement
 Complications associated with coronary artery stents
are stent thrombosis and restenosis.
 Lateral view of the chest shows sternal wires (arrowhead), vascular clips of

a saphenous vein bypass graft to the right coronary artery (curved
arrow), and those of the left internal mammary graft to the left anterior
descending coronary artery (straight arrow).
 PA chest radiograph demonstrates bilateral pulmonary artery stents in

a patient with bilateral pulmonary artery stenosis and aneurysmal right
ventricular outflow tract following tetralogy repair.
Circulatory Assist Devices
 The high mortality from cardiogenic shock continues

to spur efforts to develop mechanical support for the
circulatory system
 Most mechanical support for patients with heart
failure consists of devices that assist the heart without
replacing it.
Mechanical cardiac assist devices can be divided into
three groups:
 (a) temporary cardiac assist devices
 (b) permanent cardiac assist devices
 (c) heart replacement devices.
short-term cardiac assist devices are the intraaotic
balloon pump and newer left ventricular assist devices
(VADs)
 Frontal (a) and lateral (b) views of the chest show a

Thoratec left VAD (arrow).
 Abdominal image shows a HeartMate VAD.
 Abdominal image shows a Novacor VAD.
 chest show a CardioWest total artificial heart. the four prosthetic

valves and the two coil, reinforced polyurethane tubes carrying
pulses of compressed air to the two artificial ventricles.
 Cropped frontal view (a) and full lateral view (b) of the chest show a
CardioWest total artificial heart. Note the four prosthetic valves and the two
coil, reinforced polyurethane tubes carrying pulses of compressed air to the
two artificial ventricles.
 More x rays for interactive session
 30-year-old man with malpositioned feeding tube. Anteroposterior

chest radiograph shows that feeding tube has entered right main
bronchus, traversed right lower lobe bronchus (white arrows), and has
its tip overlying right upper quadrant of abdomen (black
arrow), raising concern for possible perforation of right
hemidiaphragm. Note associated right pneumothorax (asterisk).
 Frontal view of the chest shows “ping-pong ball plombage”

in the right apex, as well as a cardiac pacemaker.
 Frontal view of the chest shows a right apical oleothorax

(wax plombage). Extensive pleural calcification includes
the surface of the wax ball (arrows).
 Implanted bilateral brain stimulators
 A pacemaker is one of the common devices encountered on a chest x-ray. The
usual location for a pacemaker is the anterior left upper chest wall (black arrow).
Pacemakers may have either 1 or 2 leads. The wires connecting the pacemaker to
the intracardiac electrodes must be intact (yellow arrow). The typical position of

the cardiac electrodes is in the right ventricle (red arrow) for a single lead, and
also in the right atrium for a dual-lead pacemaker. It is important to compare the
electrode position to that in previous studies because an electrode may become
dislodged.
 a) Frontal view of the chest shows an esophageal stent (black

arrows) that was placed to ameliorate the effects of an esophageal
malignancy. There are also two chest tubes (), a peripherally
inserted central catheter (white arrow), ECG leads (E), a gown snap
(G), and a transjugular intrahepatic portosystemic shunt (T) in the
liver.
 Frontal view of the chest shows a left jugular Swan-Ganz catheter
(arrows), which passes through a persistent left superior vena cava into the
coronary sinus, through the right atrium and right ventricle, and into the right
pulmonary artery. Also seen are a subcutaneous port (P), an endotracheal tube
(ET), an ECG lead (E), and a nasogastric tube (not labeled).
 The electrode is placed in the epidural space adjacent to the spinal cord. The wires are
connected to a stimulating generator implanted subcutaneously. The electrode generates a
weak electrical current that interrupts the transmission of pain at a spinal cord level.
 Spinal cord stimulation (SCS) is recommended as a treatment option for adults with chronic
pain of neuropathic origin. Most frequently: Brachial plexopathy, Post-laminectomy
syndrome, Post Chemotherapy Neuropathy, Complex Regional Pain Syndrome (types I and II)
and HIV polyneuropathy.


The implantable loop recorder (ILR) is a subcutaneous electrocardiographic monitoring device that stores ECG data
automatically in response to specific rhythm anomalies or in response to patient activation.



It is mainly used for diagnosis in patients with recurrent unexplained episodes of syncope or palpitations, but is also
useful for long-term monitoring in patients with documented or suspected atrial fibrillation, for risk stratification in
patients who have sustained a myocardial infarction and those who have certain genetic disorders.
 48-year-old woman with extravascular placement of double-lumen

dialysis catheter.
 A, Anteroposterior chest radiograph shows catheter (arrow) inserted via
right subclavian vein with its tip projecting over right atrium.
 48-year-old woman with extravascular placement of double-lumen

dialysis catheter.
 B, Because there was clinical suspicion of malpositioning of
catheter, IV contrast medium was injected and was seen to
extravasate into pleural space (arrows).
 72-year-old woman with pulmonary infarction as complication of

pulmonary artery catheter placement. Magnified anteroposterior chest
radiograph shows that tip of catheter (black arrow) is too distal (i.e., >
2 cm lateral to hilum). There is wedged-shaped opacity (white arrows)
distal to catheter, consistent with pulmonary infarction.
 75-year-old woman with displacement of pacemaker

lead.
 A, Posteroanterior chest radiograph shows dual-lead
pacemaker. Tip of right ventricular lead (arrow) is
projected at edge cardiac silhouette.
 67-year-old man with pacemaker lead fracture. Magnified

anteroposterior chest radiograph shows fracture (arrow) in
pacemaker lead near battery-control pack. Lead fractures most
commonly occur at venous access site, near tip, or near
batterycontrol pack.
 7-year-old man with normal positioning of intraaortic counterpulsation balloon pump.
 B, Magnified anteroposterior chest radiograph obtained during diastole shows inflated
radiolucent balloon (thin arrows) as well as radiopaque tip (thick arrow) within upper
descending thoracic aorta. Catheter is inflated during diastole to increase myocardial perfusion
and is deflated during systole to decrease left ventricular afterload.
 misplaced central venous catheters.
 6-year-old girl. Posteroanterior chest radiograph shows two misplaced

catheters. Right internal jugular central venous catheter (black arrow)
has its tip in right atrium. Left subclavian central venous catheter
(white arrow) has its tip in right subclavian vein.
Conclusion
 Various devices are used to monitor and treat critically

ill patients.
 The radiographic evaluation of these devices is
important because the potentially serious
complications arising from their introduction and use
are often not clinically apparent.
 Familiarity with normal and abnormal radiographic
findings is critical for the detection of these
complications.
Recommendation
 The American College of Radiology recommends daily

chest radiography for critically ill patients who have acute
cardiopulmonary disease or are receiving mechanical
ventilation, as well as immediate imaging for all patients
who have undergone placement of endotracheal tubes
(ETTs), feeding tubes, vascular catheters, and chest tubes.
These recommendations are made because the
malpositioning of these devices and the serious
complications that may ensue are often not clinically
apparent. Radiographic evaluation of these devices is
important, albeit challenging, because of the technical
limitations of portable chest radiography and the inability
of patients to cooperate.
Thank You

Weitere ähnliche Inhalte

Was ist angesagt?

Medical devices on the chest x ray
Medical  devices on the chest x rayMedical  devices on the chest x ray
Medical devices on the chest x rayairwave12
 
Chest xray for evaluation of cardiovascular system
Chest xray for evaluation of cardiovascular systemChest xray for evaluation of cardiovascular system
Chest xray for evaluation of cardiovascular systemPRAVEEN GUPTA
 
Presentation1, radiological imaging of undescended testis.
Presentation1, radiological imaging of undescended testis.Presentation1, radiological imaging of undescended testis.
Presentation1, radiological imaging of undescended testis.Abdellah Nazeer
 
Mediastnum ppt
Mediastnum pptMediastnum ppt
Mediastnum pptjyomita
 
Peripheral angiography
Peripheral angiographyPeripheral angiography
Peripheral angiographyInosRagan
 
chest radiology in ICU
   chest radiology in ICU   chest radiology in ICU
chest radiology in ICUEman Mahmoud
 
CT Procedure of Thorax (CT Chest)
CT Procedure of Thorax (CT Chest)CT Procedure of Thorax (CT Chest)
CT Procedure of Thorax (CT Chest)Upakar Paudel
 
Pneumobilia vs portal vein gas
Pneumobilia vs portal vein gasPneumobilia vs portal vein gas
Pneumobilia vs portal vein gasairwave12
 
Fundamentals of chest radiology
Fundamentals of chest radiologyFundamentals of chest radiology
Fundamentals of chest radiologyDr. Sreedhar Rao
 
Chest x ray quality - how to interpret chest x-ray (1)
Chest x ray quality - how to interpret chest x-ray (1)Chest x ray quality - how to interpret chest x-ray (1)
Chest x ray quality - how to interpret chest x-ray (1)Yusuf Shieba Elhamd
 
Basic interpretation of cxr
Basic interpretation of cxrBasic interpretation of cxr
Basic interpretation of cxrKochi Chia
 
20 congenital heart disease Dr. Muhammmad Bin Zulfiqar
20 congenital heart disease Dr. Muhammmad Bin Zulfiqar20 congenital heart disease Dr. Muhammmad Bin Zulfiqar
20 congenital heart disease Dr. Muhammmad Bin ZulfiqarDr. Muhammad Bin Zulfiqar
 
076 advances in pulmonary imaging
076 advances in pulmonary imaging076 advances in pulmonary imaging
076 advances in pulmonary imagingmuhammed Yasar
 
Segmental approach to Congenital Heart Disease
Segmental approach to Congenital Heart DiseaseSegmental approach to Congenital Heart Disease
Segmental approach to Congenital Heart DiseaseTanat Tabtieang
 

Was ist angesagt? (20)

Normal chest xray
Normal chest xrayNormal chest xray
Normal chest xray
 
Medical devices on the chest x ray
Medical  devices on the chest x rayMedical  devices on the chest x ray
Medical devices on the chest x ray
 
Chest xray for evaluation of cardiovascular system
Chest xray for evaluation of cardiovascular systemChest xray for evaluation of cardiovascular system
Chest xray for evaluation of cardiovascular system
 
Presentation1, radiological imaging of undescended testis.
Presentation1, radiological imaging of undescended testis.Presentation1, radiological imaging of undescended testis.
Presentation1, radiological imaging of undescended testis.
 
Mediastnum ppt
Mediastnum pptMediastnum ppt
Mediastnum ppt
 
Chest xray
Chest xrayChest xray
Chest xray
 
Ultrasoud hernia
Ultrasoud herniaUltrasoud hernia
Ultrasoud hernia
 
Peripheral angiography
Peripheral angiographyPeripheral angiography
Peripheral angiography
 
chest radiology in ICU
   chest radiology in ICU   chest radiology in ICU
chest radiology in ICU
 
CT Pulmonary Angiography
CT Pulmonary AngiographyCT Pulmonary Angiography
CT Pulmonary Angiography
 
CT Procedure of Thorax (CT Chest)
CT Procedure of Thorax (CT Chest)CT Procedure of Thorax (CT Chest)
CT Procedure of Thorax (CT Chest)
 
Pneumobilia vs portal vein gas
Pneumobilia vs portal vein gasPneumobilia vs portal vein gas
Pneumobilia vs portal vein gas
 
Fundamentals of chest radiology
Fundamentals of chest radiologyFundamentals of chest radiology
Fundamentals of chest radiology
 
Chest x ray quality - how to interpret chest x-ray (1)
Chest x ray quality - how to interpret chest x-ray (1)Chest x ray quality - how to interpret chest x-ray (1)
Chest x ray quality - how to interpret chest x-ray (1)
 
Ct chest type
Ct chest typeCt chest type
Ct chest type
 
Chest imaging
Chest imagingChest imaging
Chest imaging
 
Basic interpretation of cxr
Basic interpretation of cxrBasic interpretation of cxr
Basic interpretation of cxr
 
20 congenital heart disease Dr. Muhammmad Bin Zulfiqar
20 congenital heart disease Dr. Muhammmad Bin Zulfiqar20 congenital heart disease Dr. Muhammmad Bin Zulfiqar
20 congenital heart disease Dr. Muhammmad Bin Zulfiqar
 
076 advances in pulmonary imaging
076 advances in pulmonary imaging076 advances in pulmonary imaging
076 advances in pulmonary imaging
 
Segmental approach to Congenital Heart Disease
Segmental approach to Congenital Heart DiseaseSegmental approach to Congenital Heart Disease
Segmental approach to Congenital Heart Disease
 

Ähnlich wie Medical devices on the chest x ray

Imaging in opaqe hemithorax [autosaved]
Imaging in opaqe hemithorax [autosaved]Imaging in opaqe hemithorax [autosaved]
Imaging in opaqe hemithorax [autosaved]Ajay Singh
 
RADIOLOGY OF LINES AND TUBES IN INTENSIVE CARE SEMINAR.pdf
RADIOLOGY OF LINES AND TUBES IN INTENSIVE CARE SEMINAR.pdfRADIOLOGY OF LINES AND TUBES IN INTENSIVE CARE SEMINAR.pdf
RADIOLOGY OF LINES AND TUBES IN INTENSIVE CARE SEMINAR.pdfTalabathulaMadhuri1
 
Radiology respiratory new.ppt
Radiology respiratory new.pptRadiology respiratory new.ppt
Radiology respiratory new.pptDarshuBoricha
 
role of radiology in pediatric.pptx
role of radiology in pediatric.pptxrole of radiology in pediatric.pptx
role of radiology in pediatric.pptxdypradio
 
abdxray-170716043820 (1).pptx
abdxray-170716043820 (1).pptxabdxray-170716043820 (1).pptx
abdxray-170716043820 (1).pptxRajVaghasia
 
Chest x ray - basics
Chest x ray - basicsChest x ray - basics
Chest x ray - basicsRikin Hasnani
 
Normal abd xray rjj.pptx
Normal abd xray rjj.pptxNormal abd xray rjj.pptx
Normal abd xray rjj.pptxrohanjohnjacob
 
Thoracic positioning
Thoracic positioningThoracic positioning
Thoracic positioningKushagra Garg
 
Middle mediastinal masses
Middle mediastinal massesMiddle mediastinal masses
Middle mediastinal massesairwave12
 
Congenital anomalies of respiratory system A Radiological approach
Congenital anomalies of respiratory system A Radiological approachCongenital anomalies of respiratory system A Radiological approach
Congenital anomalies of respiratory system A Radiological approachArif S
 
Tracheostomy and its care by Dr.Ashwin menon
Tracheostomy and its care by Dr.Ashwin menonTracheostomy and its care by Dr.Ashwin menon
Tracheostomy and its care by Dr.Ashwin menonDr.Ashwin Menon
 
common neonatal Radiology findings
common neonatal Radiology findingscommon neonatal Radiology findings
common neonatal Radiology findingsDr. Habibur Rahim
 
R NKUNA X_RAY INTERPRETATION 2020.pptx Physiotherapy
R NKUNA X_RAY INTERPRETATION 2020.pptx PhysiotherapyR NKUNA X_RAY INTERPRETATION 2020.pptx Physiotherapy
R NKUNA X_RAY INTERPRETATION 2020.pptx PhysiotherapySakhileKhoza2
 
Radiology in newborn collected by Dr. Saiful islam MD
Radiology in newborn collected by Dr. Saiful islam MDRadiology in newborn collected by Dr. Saiful islam MD
Radiology in newborn collected by Dr. Saiful islam MDDr. Habibur Rahim
 
abdxray-170716043820 (1).pdf
abdxray-170716043820 (1).pdfabdxray-170716043820 (1).pdf
abdxray-170716043820 (1).pdfRajVaghasia
 

Ähnlich wie Medical devices on the chest x ray (20)

Imaging in opaqe hemithorax [autosaved]
Imaging in opaqe hemithorax [autosaved]Imaging in opaqe hemithorax [autosaved]
Imaging in opaqe hemithorax [autosaved]
 
RADIOLOGY OF LINES AND TUBES IN INTENSIVE CARE SEMINAR.pdf
RADIOLOGY OF LINES AND TUBES IN INTENSIVE CARE SEMINAR.pdfRADIOLOGY OF LINES AND TUBES IN INTENSIVE CARE SEMINAR.pdf
RADIOLOGY OF LINES AND TUBES IN INTENSIVE CARE SEMINAR.pdf
 
Radiology respiratory new.ppt
Radiology respiratory new.pptRadiology respiratory new.ppt
Radiology respiratory new.ppt
 
role of radiology in pediatric.pptx
role of radiology in pediatric.pptxrole of radiology in pediatric.pptx
role of radiology in pediatric.pptx
 
USG chest
USG chestUSG chest
USG chest
 
Septal puncure ppt
Septal puncure pptSeptal puncure ppt
Septal puncure ppt
 
abdxray-170716043820 (1).pptx
abdxray-170716043820 (1).pptxabdxray-170716043820 (1).pptx
abdxray-170716043820 (1).pptx
 
Chest x ray - basics
Chest x ray - basicsChest x ray - basics
Chest x ray - basics
 
Normal abd xray rjj.pptx
Normal abd xray rjj.pptxNormal abd xray rjj.pptx
Normal abd xray rjj.pptx
 
tubes on chest xrays
tubes on chest xraystubes on chest xrays
tubes on chest xrays
 
Thoracic positioning
Thoracic positioningThoracic positioning
Thoracic positioning
 
Middle mediastinal masses
Middle mediastinal massesMiddle mediastinal masses
Middle mediastinal masses
 
Congenital anomalies of respiratory system A Radiological approach
Congenital anomalies of respiratory system A Radiological approachCongenital anomalies of respiratory system A Radiological approach
Congenital anomalies of respiratory system A Radiological approach
 
Tracheostomy and its care by Dr.Ashwin menon
Tracheostomy and its care by Dr.Ashwin menonTracheostomy and its care by Dr.Ashwin menon
Tracheostomy and its care by Dr.Ashwin menon
 
common neonatal Radiology findings
common neonatal Radiology findingscommon neonatal Radiology findings
common neonatal Radiology findings
 
R NKUNA X_RAY INTERPRETATION 2020.pptx Physiotherapy
R NKUNA X_RAY INTERPRETATION 2020.pptx PhysiotherapyR NKUNA X_RAY INTERPRETATION 2020.pptx Physiotherapy
R NKUNA X_RAY INTERPRETATION 2020.pptx Physiotherapy
 
Radiology in newborn collected by Dr. Saiful islam MD
Radiology in newborn collected by Dr. Saiful islam MDRadiology in newborn collected by Dr. Saiful islam MD
Radiology in newborn collected by Dr. Saiful islam MD
 
abdxray-170716043820 (1).pdf
abdxray-170716043820 (1).pdfabdxray-170716043820 (1).pdf
abdxray-170716043820 (1).pdf
 
Abd xray
Abd xrayAbd xray
Abd xray
 
USG chest
USG chestUSG chest
USG chest
 

Kürzlich hochgeladen

Independent Call Girls Andheri Nightlaila 9967584737
Independent Call Girls Andheri Nightlaila 9967584737Independent Call Girls Andheri Nightlaila 9967584737
Independent Call Girls Andheri Nightlaila 9967584737Riya Pathan
 
8447779800, Low rate Call girls in Shivaji Enclave Delhi NCR
8447779800, Low rate Call girls in Shivaji Enclave Delhi NCR8447779800, Low rate Call girls in Shivaji Enclave Delhi NCR
8447779800, Low rate Call girls in Shivaji Enclave Delhi NCRashishs7044
 
Future Of Sample Report 2024 | Redacted Version
Future Of Sample Report 2024 | Redacted VersionFuture Of Sample Report 2024 | Redacted Version
Future Of Sample Report 2024 | Redacted VersionMintel Group
 
Kenya’s Coconut Value Chain by Gatsby Africa
Kenya’s Coconut Value Chain by Gatsby AfricaKenya’s Coconut Value Chain by Gatsby Africa
Kenya’s Coconut Value Chain by Gatsby Africaictsugar
 
Case study on tata clothing brand zudio in detail
Case study on tata clothing brand zudio in detailCase study on tata clothing brand zudio in detail
Case study on tata clothing brand zudio in detailAriel592675
 
APRIL2024_UKRAINE_xml_0000000000000 .pdf
APRIL2024_UKRAINE_xml_0000000000000 .pdfAPRIL2024_UKRAINE_xml_0000000000000 .pdf
APRIL2024_UKRAINE_xml_0000000000000 .pdfRbc Rbcua
 
Call Girls In Radisson Blu Hotel New Delhi Paschim Vihar ❤️8860477959 Escorts...
Call Girls In Radisson Blu Hotel New Delhi Paschim Vihar ❤️8860477959 Escorts...Call Girls In Radisson Blu Hotel New Delhi Paschim Vihar ❤️8860477959 Escorts...
Call Girls In Radisson Blu Hotel New Delhi Paschim Vihar ❤️8860477959 Escorts...lizamodels9
 
8447779800, Low rate Call girls in New Ashok Nagar Delhi NCR
8447779800, Low rate Call girls in New Ashok Nagar Delhi NCR8447779800, Low rate Call girls in New Ashok Nagar Delhi NCR
8447779800, Low rate Call girls in New Ashok Nagar Delhi NCRashishs7044
 
Cybersecurity Awareness Training Presentation v2024.03
Cybersecurity Awareness Training Presentation v2024.03Cybersecurity Awareness Training Presentation v2024.03
Cybersecurity Awareness Training Presentation v2024.03DallasHaselhorst
 
Innovation Conference 5th March 2024.pdf
Innovation Conference 5th March 2024.pdfInnovation Conference 5th March 2024.pdf
Innovation Conference 5th March 2024.pdfrichard876048
 
Marketplace and Quality Assurance Presentation - Vincent Chirchir
Marketplace and Quality Assurance Presentation - Vincent ChirchirMarketplace and Quality Assurance Presentation - Vincent Chirchir
Marketplace and Quality Assurance Presentation - Vincent Chirchirictsugar
 
International Business Environments and Operations 16th Global Edition test b...
International Business Environments and Operations 16th Global Edition test b...International Business Environments and Operations 16th Global Edition test b...
International Business Environments and Operations 16th Global Edition test b...ssuserf63bd7
 
BEST Call Girls In Greater Noida ✨ 9773824855 ✨ Escorts Service In Delhi Ncr,
BEST Call Girls In Greater Noida ✨ 9773824855 ✨ Escorts Service In Delhi Ncr,BEST Call Girls In Greater Noida ✨ 9773824855 ✨ Escorts Service In Delhi Ncr,
BEST Call Girls In Greater Noida ✨ 9773824855 ✨ Escorts Service In Delhi Ncr,noida100girls
 
/:Call Girls In Indirapuram Ghaziabad ➥9990211544 Independent Best Escorts In...
/:Call Girls In Indirapuram Ghaziabad ➥9990211544 Independent Best Escorts In.../:Call Girls In Indirapuram Ghaziabad ➥9990211544 Independent Best Escorts In...
/:Call Girls In Indirapuram Ghaziabad ➥9990211544 Independent Best Escorts In...lizamodels9
 
Global Scenario On Sustainable and Resilient Coconut Industry by Dr. Jelfina...
Global Scenario On Sustainable  and Resilient Coconut Industry by Dr. Jelfina...Global Scenario On Sustainable  and Resilient Coconut Industry by Dr. Jelfina...
Global Scenario On Sustainable and Resilient Coconut Industry by Dr. Jelfina...ictsugar
 
The CMO Survey - Highlights and Insights Report - Spring 2024
The CMO Survey - Highlights and Insights Report - Spring 2024The CMO Survey - Highlights and Insights Report - Spring 2024
The CMO Survey - Highlights and Insights Report - Spring 2024christinemoorman
 
Market Sizes Sample Report - 2024 Edition
Market Sizes Sample Report - 2024 EditionMarket Sizes Sample Report - 2024 Edition
Market Sizes Sample Report - 2024 EditionMintel Group
 
Islamabad Escorts | Call 03070433345 | Escort Service in Islamabad
Islamabad Escorts | Call 03070433345 | Escort Service in IslamabadIslamabad Escorts | Call 03070433345 | Escort Service in Islamabad
Islamabad Escorts | Call 03070433345 | Escort Service in IslamabadAyesha Khan
 
8447779800, Low rate Call girls in Saket Delhi NCR
8447779800, Low rate Call girls in Saket Delhi NCR8447779800, Low rate Call girls in Saket Delhi NCR
8447779800, Low rate Call girls in Saket Delhi NCRashishs7044
 

Kürzlich hochgeladen (20)

Independent Call Girls Andheri Nightlaila 9967584737
Independent Call Girls Andheri Nightlaila 9967584737Independent Call Girls Andheri Nightlaila 9967584737
Independent Call Girls Andheri Nightlaila 9967584737
 
8447779800, Low rate Call girls in Shivaji Enclave Delhi NCR
8447779800, Low rate Call girls in Shivaji Enclave Delhi NCR8447779800, Low rate Call girls in Shivaji Enclave Delhi NCR
8447779800, Low rate Call girls in Shivaji Enclave Delhi NCR
 
Future Of Sample Report 2024 | Redacted Version
Future Of Sample Report 2024 | Redacted VersionFuture Of Sample Report 2024 | Redacted Version
Future Of Sample Report 2024 | Redacted Version
 
Kenya’s Coconut Value Chain by Gatsby Africa
Kenya’s Coconut Value Chain by Gatsby AfricaKenya’s Coconut Value Chain by Gatsby Africa
Kenya’s Coconut Value Chain by Gatsby Africa
 
Case study on tata clothing brand zudio in detail
Case study on tata clothing brand zudio in detailCase study on tata clothing brand zudio in detail
Case study on tata clothing brand zudio in detail
 
APRIL2024_UKRAINE_xml_0000000000000 .pdf
APRIL2024_UKRAINE_xml_0000000000000 .pdfAPRIL2024_UKRAINE_xml_0000000000000 .pdf
APRIL2024_UKRAINE_xml_0000000000000 .pdf
 
Call Girls In Radisson Blu Hotel New Delhi Paschim Vihar ❤️8860477959 Escorts...
Call Girls In Radisson Blu Hotel New Delhi Paschim Vihar ❤️8860477959 Escorts...Call Girls In Radisson Blu Hotel New Delhi Paschim Vihar ❤️8860477959 Escorts...
Call Girls In Radisson Blu Hotel New Delhi Paschim Vihar ❤️8860477959 Escorts...
 
8447779800, Low rate Call girls in New Ashok Nagar Delhi NCR
8447779800, Low rate Call girls in New Ashok Nagar Delhi NCR8447779800, Low rate Call girls in New Ashok Nagar Delhi NCR
8447779800, Low rate Call girls in New Ashok Nagar Delhi NCR
 
Cybersecurity Awareness Training Presentation v2024.03
Cybersecurity Awareness Training Presentation v2024.03Cybersecurity Awareness Training Presentation v2024.03
Cybersecurity Awareness Training Presentation v2024.03
 
Innovation Conference 5th March 2024.pdf
Innovation Conference 5th March 2024.pdfInnovation Conference 5th March 2024.pdf
Innovation Conference 5th March 2024.pdf
 
Marketplace and Quality Assurance Presentation - Vincent Chirchir
Marketplace and Quality Assurance Presentation - Vincent ChirchirMarketplace and Quality Assurance Presentation - Vincent Chirchir
Marketplace and Quality Assurance Presentation - Vincent Chirchir
 
International Business Environments and Operations 16th Global Edition test b...
International Business Environments and Operations 16th Global Edition test b...International Business Environments and Operations 16th Global Edition test b...
International Business Environments and Operations 16th Global Edition test b...
 
Japan IT Week 2024 Brochure by 47Billion (English)
Japan IT Week 2024 Brochure by 47Billion (English)Japan IT Week 2024 Brochure by 47Billion (English)
Japan IT Week 2024 Brochure by 47Billion (English)
 
BEST Call Girls In Greater Noida ✨ 9773824855 ✨ Escorts Service In Delhi Ncr,
BEST Call Girls In Greater Noida ✨ 9773824855 ✨ Escorts Service In Delhi Ncr,BEST Call Girls In Greater Noida ✨ 9773824855 ✨ Escorts Service In Delhi Ncr,
BEST Call Girls In Greater Noida ✨ 9773824855 ✨ Escorts Service In Delhi Ncr,
 
/:Call Girls In Indirapuram Ghaziabad ➥9990211544 Independent Best Escorts In...
/:Call Girls In Indirapuram Ghaziabad ➥9990211544 Independent Best Escorts In.../:Call Girls In Indirapuram Ghaziabad ➥9990211544 Independent Best Escorts In...
/:Call Girls In Indirapuram Ghaziabad ➥9990211544 Independent Best Escorts In...
 
Global Scenario On Sustainable and Resilient Coconut Industry by Dr. Jelfina...
Global Scenario On Sustainable  and Resilient Coconut Industry by Dr. Jelfina...Global Scenario On Sustainable  and Resilient Coconut Industry by Dr. Jelfina...
Global Scenario On Sustainable and Resilient Coconut Industry by Dr. Jelfina...
 
The CMO Survey - Highlights and Insights Report - Spring 2024
The CMO Survey - Highlights and Insights Report - Spring 2024The CMO Survey - Highlights and Insights Report - Spring 2024
The CMO Survey - Highlights and Insights Report - Spring 2024
 
Market Sizes Sample Report - 2024 Edition
Market Sizes Sample Report - 2024 EditionMarket Sizes Sample Report - 2024 Edition
Market Sizes Sample Report - 2024 Edition
 
Islamabad Escorts | Call 03070433345 | Escort Service in Islamabad
Islamabad Escorts | Call 03070433345 | Escort Service in IslamabadIslamabad Escorts | Call 03070433345 | Escort Service in Islamabad
Islamabad Escorts | Call 03070433345 | Escort Service in Islamabad
 
8447779800, Low rate Call girls in Saket Delhi NCR
8447779800, Low rate Call girls in Saket Delhi NCR8447779800, Low rate Call girls in Saket Delhi NCR
8447779800, Low rate Call girls in Saket Delhi NCR
 

Medical devices on the chest x ray

  • 2.  Chest devices are encountered on daily basis by radiologists  Chest tubes, central lines, endotracheal tubes and NG tubes are common  New devices are constantly being introduced
  • 3.  It is important to recognizer the presence of a device and to have an understanding of its function as well as complications associated with its use
  • 4.  The proliferation of intensive care units and the advances in the treatment of the very ill have greatly increased the numbers of examinations performed at the patient’s bedside.  Obtaining a daily chest radiograph is standard practice in most intensive care units
  • 5.  Chest radiograph shows that the tip of the endotracheal tube (black arrow) is slightly above the aortic arch and well above the carina, in good position. A right chest tube (white arrow), ECG leads (E), a gown snap (G), and oxygen tubing (O) are also visible.
  • 6. Medical devices  Extra thoracic devices  Pleural devices  Tracheal  Esophageal  Vascular  Cardiac  Circulatory assist devices
  • 7. Extra thoracic devices  Tubing, clamps, syringes  Ventilator support tubing  ECG electrodes  Breast prostheses  other apparatus often lie on or under the patient and are imaged with the chest during the radiographic examination
  • 8.  Chest radiograph shows halo apparatus (with emergency wrench close at hand) for cervical spine stabilization.
  • 9.
  • 11.
  • 12.
  • 13.
  • 14. Pleural Devices  chest tubes are commonly used for evacuating fluid or air from the pleural space  Antero superiorly for pneumothorax  Poster inferiorly for fluid collection
  • 15.  Frontal (a) and lateral (b) views show a thoracostomy(chest) tube in good position for treatment of a pneumothorax but not for an effusion.
  • 16.  Chest tubes are commonly used for evacuating fluid or air from the pleural space.  The normally positioned tube lies on the surface of the expanded lung, between the visceral and parietal pleura.  The tube is usually placed anterosuperiorly to evacuate a pneumothorax and posteroinferiorly for fluid collections.
  • 17.  the normally positioned tube lies on the surface of the expanded lung, between the visceral and parietal pleurae.  Pigtail catheters may be used in place of standard thoracostomy tubes, and they are popular for empyema drainage and for installation of medication for treatment of an empyema
  • 18.  Frontal view of the chest shows a pigtail catheter that had been inserted under fluoroscopic guidance into a loculated right empyema for instillation of urokinase and fluid drainage.
  • 19.  Assess the position of chest tube in both frontal and lateral view and sometime may required CT scan Complication Malposition  May enter the interlobar fissure  Lung parenchyma  Subcutaneous tissue  Tube may be kinked
  • 20.
  • 21.  58-year-old woman with extrapleural placement of chest tube. Magnified anteroposterior chest radiograph shows misplaced chest tube (arrows) within right chest wall.
  • 22.  65-year-old woman with extrapleural placement of chest tube.  A, Magnified anteroposterior chest radiograph shows left chest tube (arrow) in apparently adequate position. CT scan was requested to further investigate because of ineffective drainage of left pleural effusion.
  • 23.  Magnified axial CT image shows misplacement of chest tube within chest wall (arrow).
  • 24.  69-year-old man with intrafissural placement of chest tube.  A, Magnified anteroposterior chest radiograph shows horizontal course of right chest tube (arrows).
  • 25.  49-year-old man with intraparenchymal placement of chest tube.  A, Scout image shows chest tube (arrow) projecting over right mid lung field.
  • 26.  Magnified CT image shows chest tube (thick arrow) coursing through right upper lobe. There is associated small pneumothorax (asterisk) and subcutaneous emphysema (thin arrow).
  • 27.  48-year-old man with chest tube kinking. Magnified posteroanterior chest radiograph performed after chest tube placement shows kinking of chest tube (arrow) precluding adequate pleural drainage.
  • 28.  37-year-old man with mediastinal placement of chest tube.  A, Anteroposterior chest radiograph shows left chest tube (arrows) in inappropriate position, directed medially and projecting across mediastinum. There is persistent left pleural effusion.
  • 29.  37-year-old man with mediastinal placement of chest tube.  B, CT image at level of pulmonary artery trunk confirms that tip of chest tube (arrow) is in anterior mediastinum.
  • 30.  30-year-old male victim of motor vehicle trauma with abdominal placement of chest tube.  A, Anteroposterior chest radiograph shows horizontally oriented chest tube (arrow) in left lower hemithorax. There are several left rib fractures, opacification of left hemithorax, and subcutaneous emphysema.
  • 31.  30-year-old male victim of motor vehicle trauma with abdominal placement of chest tube.  B, CT image shows traumatic left diaphragmatic rupture with migration of abdominal content to left hemithorax. Chest tube (black arrow) is seen within mesenteric fat abutting small bowel loops. Note splenic rupture (white arrow) related to trauma.
  • 32. Tracheal Device  Tracheal intubation is a life saving procedure but can be life threatening if placed incorrectly  The tip of the tube should be 5 cm above the carina
  • 33.  The carina is just cauded to the aortic arch
  • 34.  Chest radiograph shows that the tip of the endotracheal tube (black arrow) is slightly above the aortic arch and well above the carina, in good position. A right chest tube (white arrow), ECG leads (E), a gown snap (G), and oxygen tubing (O) are also visible.
  • 35. Complications  When advanced too far, the endotracheal tube usually enters the right main bronchus, causing various combinations of hyperinflation and atelectasis of the two lungs, depending on the positions of the end and side holes.  endotracheal tubes can also be placed in the esophagus  the soft tissues of the neck.
  • 36.  6-year-old woman with misplaced endotracheal tube. Magnified anteroposterior chest radiograph shows that tip of endotracheal tube (thick arrow) is too high, at level of thoracic inlet. Endotracheal tube cuff (thin arrows) is overdistended. This abnormal position may cause vocal cord injury.
  • 37.  60-year-old woman with inadvertent right main bronchial intubation. Anteroposterior chest radiograph shows endotracheal tube tip (arrow) in right main bronchus, resulting in complete collapse of left lung and leftward shift of mediastinum.
  • 38.  Frontal view shows a double-lumen endotracheal tube with selective intubation of the left main bronchus (arrow).
  • 39.  66-year-old man with left-sided double-lumen endotracheal tube. Magnified anteroposterior chest radiograph shows double-lumen endotracheal tube with its left tip (thick arrow) in left main bronchus. Right tip (thin arrow) is noted within trachea for ventilation of right lung. Double-lumen endotracheal tube allows control of distribution of ventilation to each lung. It is important to differentiate between double-lumen endotracheal tubes and inadvertent selective bronchial intubation with single-lumen catheters.
  • 40. Esophageal devices  NG tubes  Esophageal balloons  Esophageal stents  PH probe
  • 41. NG tubes  Nasogastric tubes and feeding tubes are frequently visualized passing through the mediastinum on their way to the stomach and intestines.  Esophageal balloons and esophageal stents used to treat benign and malignant esophageal disease may manifest themselves at chest radiography.
  • 42.
  • 43.
  • 44.
  • 45.
  • 46.
  • 47.
  • 48. Vascular devices Routinely used for  Monitoring hemodynamic function  Hemodialysis  Administrating fluids  Medication  Nutrition
  • 49.  Venous devices are usually inserted, either percutaneously or surgically, via the subclavian, internal jugular, or femoral veins. Arterial devices usually are placed through the femoral artery  central lines typically have one to three lumens  A central venous catheter is ideally positioned in the superior vena cava for the monitoring of pressure or infusion of medication and nutrition.
  • 50.  Swan-Ganz is a multilumen catheter used for measuring hemodynamic pressures and cardiac output.A better term to use is pulmonary artery catheter.  Accurate measurement of pulmonary arterial wedge pressure
  • 51.
  • 52.
  • 53.  chest radiograph shows a right subclavian single-lumen central venous catheter and a left subcutaneous port catheter, which enters via the left subclavian vein. Both catheter tips are in the superior vena cava.
  • 54.  The port is usually connected to a central venous catheter or to an arterial catheter and can be used for instillation of fluids, medications, chemotherapeutic agents, parenteral nutritional solutions, and blood products. It can also be used for withdrawal of blood samples.
  • 55.  Frontal view of the chest shows a left subclavian Groshong catheter with its tip in the proximal most portion of the superior vena cava. (b) Close up view of the catheter tip.
  • 56.  Groshong catheter has closed, rounded tip Near the tip in the side of the catheter is a three-position valve. The valve is designed to allow fluid to flow in and out through the valve, but it remains closed when it is not in use. This catheter does not require routine clamping or heparin solution to keep open. It does require periodic flushing with 0.9% normal saline.
  • 57.  A good rule of thumb is that the catheter tip should be within the mediastinal shadow. Placement more distally increases the chance of pulmonary infarction or vessel rupture.
  • 58.  Frontal chest radiograph shows a right jugular Swan- Ganz catheter with its tip (arrow) in the right lower pulmonary artery.
  • 59.  Frontal chest radiograph shows a Swan-Ganz catheter (white arrow) in the left pulmonary artery via the inferior vena cava. Note also the bilateral chest tubes (black arrows) and ECG leads (E).
  • 60.  Complications of catheter insertion vary with the     catheters used and the sites employed Pneumothorax is a common complication Vessel lacerations and perforations can produce hematomas, hemothorax, and infusion of fluid into the mediastinum, thorax, or other inappropriate space Nerve injury is usually a complication of improper puncture technique Looping of catheters may lead to knotting.
  • 61.  Malpositioning of pulmonary artery catheter is exceedingly common, found in approximately 25% of catheters placed. This may lead to false readings and an increased risk for complications.  Complications of pulmonary artery catheter placement include pneumothorax, pulmonary infarction, cardiac arrhythmias, pulmonary artery perforation, endocarditis, and sepsis
  • 62.
  • 63. Cardiac Devices  Cardiac pacemakers, valve prostheses, and artificial hearts  chest radiography is commonly employed in the assessment of patients with heart disease  recognition of cardiac devices and the problems associated with them is important for all individuals involved in the care of these patients.
  • 64. Heart valves  Mechanical  Biologic  Mechanical Heart valves most require life-long treatment with anticoagulants.  Biologic valves are less durable than mechanical valves, with some deterioration developing, frequently 5–10 years after placement, but they do not usually require anticoagulant treatment.  It is not important and often impossible to know the specific name of a particular prosthetic heart valve, but it is important to recognize its presence  The mitral and aortic valves are those most commonly replaced
  • 65.  Mechanical heart valve. Lateral view of the chest shows a Hemex tilting bileaflet mechanical mitral valve prosthesis. Median sternotomy wires and surgical clips are also evident.
  • 66.  Mechanical heart valve. Lateral view of the chest shows a Starr-Edwards caged ball mechanical mitral valve prosthesis.
  • 67.  Biologic heart valve. Frontal view of the chest shows a Hancock porcine mitral valve prosthesis (arrow). A singlelead pacemaker, ECG leads, and median sternotomy wires are also seen.
  • 68.  Biologic heart valve. Frontal view of the chest shows a Hancock porcine valve prosthesis in a Rastelli conduit going from the right ventricle to the pulmonary artery.
  • 69.  Lateral view of the chest in an elderly patient shows a mitral annuloplasty ring (black arrow) and a dual-lead cardiac pacemaker. Sternal wires, surgical clips, and ECG leads are also present. The sternal wires are used to close a sternal dehiscence. The patient has both horizontal sternal wires and vertical intercostal wires (white arrows).
  • 70.  Lateral view of the chest in a child shows an occlusion basket (umbrella) for treatment of an atrial septal defect.
  • 71. Cardiac Pacemakers  common in older adults being treated for abnormal cardiac rhythms caused by coronary artery disease  Cardiac pacemakers improve cardiac function, reduce the severity of clinical symptoms, and reduce mortality and morbidity.  A cardiac pacemaker is composed of two main elements: (a) a pulse generator and (b) lead wires with electrodes for contact with the endocardium or myocardium
  • 72.  Pacemakers range from simple temporary epicardial electrodes to very complex pacemakers with multiple atrial and ventricular leads
  • 73.
  • 74.
  • 75.  Frontal (a) and lateral (b) views of the chest show a single electrode epicardial “corkscrew” subxiphoid pacemaker (arrowhead in a, black arrow in b). There are also coils (white arrow) occluding a previous right Blalock-Taussig shunt. In addition, ECG leads and sternal wires are evident.
  • 76.
  • 77.  Frontal (a) and lateral (b) views show an atrioventricular sequential pacemaker with one electrode in the right atrial appendage (RA) and the other at the right ventricular apex (RV). Also shown are ECG leads (E) and the battery-control pack (B) for the pacemaker.
  • 78.
  • 80.  Because there can be such a wide variation in the proper positioning of pacemaker leads, it is often difficult for the to know if a pacemaker is properly positioned.  Pacemaker lead fracture is now rarely seen because of improvements in the flexibility of the metal alloys used in electrode construction.
  • 81. Coronary Artery Surgery and Stents  Revascularization techniques include CABG surgery, coronary artery angioplasty, and coronary artery stent placement.  Median sternotomy is the usual surgical approach for CABG surgery, and sternal wires the common method of fixation of the two sternal segments  At present, almost 90% of coronary interventions include stent placement  Complications associated with coronary artery stents are stent thrombosis and restenosis.
  • 82.  Lateral view of the chest shows sternal wires (arrowhead), vascular clips of a saphenous vein bypass graft to the right coronary artery (curved arrow), and those of the left internal mammary graft to the left anterior descending coronary artery (straight arrow).
  • 83.
  • 84.
  • 85.  PA chest radiograph demonstrates bilateral pulmonary artery stents in a patient with bilateral pulmonary artery stenosis and aneurysmal right ventricular outflow tract following tetralogy repair.
  • 86. Circulatory Assist Devices  The high mortality from cardiogenic shock continues to spur efforts to develop mechanical support for the circulatory system  Most mechanical support for patients with heart failure consists of devices that assist the heart without replacing it.
  • 87. Mechanical cardiac assist devices can be divided into three groups:  (a) temporary cardiac assist devices  (b) permanent cardiac assist devices  (c) heart replacement devices. short-term cardiac assist devices are the intraaotic balloon pump and newer left ventricular assist devices (VADs)
  • 88.  Frontal (a) and lateral (b) views of the chest show a Thoratec left VAD (arrow).
  • 89.
  • 90.  Abdominal image shows a HeartMate VAD.
  • 91.  Abdominal image shows a Novacor VAD.
  • 92.  chest show a CardioWest total artificial heart. the four prosthetic valves and the two coil, reinforced polyurethane tubes carrying pulses of compressed air to the two artificial ventricles.
  • 93.  Cropped frontal view (a) and full lateral view (b) of the chest show a CardioWest total artificial heart. Note the four prosthetic valves and the two coil, reinforced polyurethane tubes carrying pulses of compressed air to the two artificial ventricles.
  • 94.  More x rays for interactive session
  • 95.  30-year-old man with malpositioned feeding tube. Anteroposterior chest radiograph shows that feeding tube has entered right main bronchus, traversed right lower lobe bronchus (white arrows), and has its tip overlying right upper quadrant of abdomen (black arrow), raising concern for possible perforation of right hemidiaphragm. Note associated right pneumothorax (asterisk).
  • 96.  Frontal view of the chest shows “ping-pong ball plombage” in the right apex, as well as a cardiac pacemaker.
  • 97.  Frontal view of the chest shows a right apical oleothorax (wax plombage). Extensive pleural calcification includes the surface of the wax ball (arrows).
  • 98.  Implanted bilateral brain stimulators
  • 99.  A pacemaker is one of the common devices encountered on a chest x-ray. The usual location for a pacemaker is the anterior left upper chest wall (black arrow). Pacemakers may have either 1 or 2 leads. The wires connecting the pacemaker to the intracardiac electrodes must be intact (yellow arrow). The typical position of the cardiac electrodes is in the right ventricle (red arrow) for a single lead, and also in the right atrium for a dual-lead pacemaker. It is important to compare the electrode position to that in previous studies because an electrode may become dislodged.
  • 100.
  • 101.  a) Frontal view of the chest shows an esophageal stent (black arrows) that was placed to ameliorate the effects of an esophageal malignancy. There are also two chest tubes (), a peripherally inserted central catheter (white arrow), ECG leads (E), a gown snap (G), and a transjugular intrahepatic portosystemic shunt (T) in the liver.
  • 102.  Frontal view of the chest shows a left jugular Swan-Ganz catheter (arrows), which passes through a persistent left superior vena cava into the coronary sinus, through the right atrium and right ventricle, and into the right pulmonary artery. Also seen are a subcutaneous port (P), an endotracheal tube (ET), an ECG lead (E), and a nasogastric tube (not labeled).
  • 103.  The electrode is placed in the epidural space adjacent to the spinal cord. The wires are connected to a stimulating generator implanted subcutaneously. The electrode generates a weak electrical current that interrupts the transmission of pain at a spinal cord level.  Spinal cord stimulation (SCS) is recommended as a treatment option for adults with chronic pain of neuropathic origin. Most frequently: Brachial plexopathy, Post-laminectomy syndrome, Post Chemotherapy Neuropathy, Complex Regional Pain Syndrome (types I and II) and HIV polyneuropathy.
  • 104.  The implantable loop recorder (ILR) is a subcutaneous electrocardiographic monitoring device that stores ECG data automatically in response to specific rhythm anomalies or in response to patient activation.  It is mainly used for diagnosis in patients with recurrent unexplained episodes of syncope or palpitations, but is also useful for long-term monitoring in patients with documented or suspected atrial fibrillation, for risk stratification in patients who have sustained a myocardial infarction and those who have certain genetic disorders.
  • 105.  48-year-old woman with extravascular placement of double-lumen dialysis catheter.  A, Anteroposterior chest radiograph shows catheter (arrow) inserted via right subclavian vein with its tip projecting over right atrium.
  • 106.  48-year-old woman with extravascular placement of double-lumen dialysis catheter.  B, Because there was clinical suspicion of malpositioning of catheter, IV contrast medium was injected and was seen to extravasate into pleural space (arrows).
  • 107.  72-year-old woman with pulmonary infarction as complication of pulmonary artery catheter placement. Magnified anteroposterior chest radiograph shows that tip of catheter (black arrow) is too distal (i.e., > 2 cm lateral to hilum). There is wedged-shaped opacity (white arrows) distal to catheter, consistent with pulmonary infarction.
  • 108.  75-year-old woman with displacement of pacemaker lead.  A, Posteroanterior chest radiograph shows dual-lead pacemaker. Tip of right ventricular lead (arrow) is projected at edge cardiac silhouette.
  • 109.  67-year-old man with pacemaker lead fracture. Magnified anteroposterior chest radiograph shows fracture (arrow) in pacemaker lead near battery-control pack. Lead fractures most commonly occur at venous access site, near tip, or near batterycontrol pack.
  • 110.  7-year-old man with normal positioning of intraaortic counterpulsation balloon pump.  B, Magnified anteroposterior chest radiograph obtained during diastole shows inflated radiolucent balloon (thin arrows) as well as radiopaque tip (thick arrow) within upper descending thoracic aorta. Catheter is inflated during diastole to increase myocardial perfusion and is deflated during systole to decrease left ventricular afterload.
  • 111.  misplaced central venous catheters.  6-year-old girl. Posteroanterior chest radiograph shows two misplaced catheters. Right internal jugular central venous catheter (black arrow) has its tip in right atrium. Left subclavian central venous catheter (white arrow) has its tip in right subclavian vein.
  • 112. Conclusion  Various devices are used to monitor and treat critically ill patients.  The radiographic evaluation of these devices is important because the potentially serious complications arising from their introduction and use are often not clinically apparent.  Familiarity with normal and abnormal radiographic findings is critical for the detection of these complications.
  • 113. Recommendation  The American College of Radiology recommends daily chest radiography for critically ill patients who have acute cardiopulmonary disease or are receiving mechanical ventilation, as well as immediate imaging for all patients who have undergone placement of endotracheal tubes (ETTs), feeding tubes, vascular catheters, and chest tubes. These recommendations are made because the malpositioning of these devices and the serious complications that may ensue are often not clinically apparent. Radiographic evaluation of these devices is important, albeit challenging, because of the technical limitations of portable chest radiography and the inability of patients to cooperate.