SlideShare ist ein Scribd-Unternehmen logo
1 von 43
By: Dr. Ankur Gupta
Resident, DM Cardiolgy
Introduction
Majority of the congenital lesions are asso. with
intracardiac shunts.
Detection, localisation and quantification of shunts –
integral part of hemodynamic evaluation of these
patients.
Quantification of shunts by:
Invasive
Noninvasive methods.
Non invasive methods
Clinical evaluation
CXR and ECG
Echo and Doppler
MRI
CT
Nuclear scanning
Invasive
Oximetry
Indicator dilution
2 –D Echocardiography:
direct visualization of any communication and their
shunting
excellent for localization of shunt
limited in the ability to quantify shunt
Indirect estimations
ASD with large shunting
 RA, RV dilated
 diastolic septal flattening
VSD with large shunting
 LA, LV dilatation
 Systolic and diastolic septal flattening
Contrast Echo
Agitated saline
 Detection of small R  L shunts
Large left to right shunts may cause a negative
contrast jet in the Rt. atrium
Doppler echocardiography
Most of the studies were done in ASD & VSD.
Based on the measurement of flow across the valves
or the defect itself.
Based on the formula that flow can be calculated
from CSA & VTI.
Volumetric flow measurement requires
- calculation of temporal mean flow velocity and
cross sectional area
- velocity profile is uniform across the vessel
- single sampling of the flow velocity in the center
of the vessel is recorded
To calculate 
computerised calculation of the area under the
doppler curve/ flow period
For systemic flow
Aortic velocity is recorded from the apical view.
Vessel diameter is measured from the parasternal long
axis view.
Pulmonary blood flow:
Pulm. mean velocity is recorded from the
parasternal short axis view
In case of large shunt and a shunt located close
to PA, the doppler tracing shows spectral
broadening or signs of disturbed flow, in such
cases uniform velocity cannot be assumed
Doppler tracing cannot be used if any PS is
present.
Measurement of the cross sectional area
- at the valve annulus (as it is the flow limiting
point of the vessel)
- measurement is made from the inner edge to
the inner edge.
- diameter is measured at the early systole –
time when vessel is at peak systolic dimension.
- flow velocity should be measured at same
location
Methods:
1.Qp/Qs calculation from flow velocity (velocity
time integral) and cross sectional area
Qp/Qs = Area of Pa x VTI
Area of Aorta x VTI
Sanders et al used pulmonary flow with good
correlation , r=0.85.
Barron et al used Mitral flow with better
correlation , r= 0.9 vs 0.69.
2. Simplified method - the square of the pulmonary
to aortic diameter ratio was substituted for the
area ratio and the flow peak velocity ratio for the
velocity time integral.
 Qp/Qs = Vel. Pk PA x r2
Vel. Pk Ao x r2
r= radius / diameter
square of the ratio of the pulmonary to aortic
luminal diameter (or radius)is used instead of
vessel areas.
3. Qp/Qs = (pulm + mitral flow)/2
(tricuspid + aortic flow)/2
 Assumption that tricuspid and aortic flow
represent systemic flow and pulmonary &
mitral flows represent pulmonary flow.
 Maximal diameter of mitral and tricuspid
valve annulus is measured during diastole in
an apical 4 –chamber view & area calculated
assuming the annulus to be circular.
 Laborious & time taking.
 Combines measurements from all the 4 valves
but was not shown to be better than the
previous 2.
4. Qp/Qs = Shunt flow + aortic flow
aortic flow
 Shunt (VSD) flow  as the product of the VTI
and color-derived cross-sectional area of VSD jet
measured at its narrowest point.
 This method does not use the pulmonary flow &
outflow diameter which are more variable.
Difficulties : It is assumed that
1. The defect is circular.
2. The size does not vary during systole.
3. No angle correction is needed.
4 . Flow velocity is uniform across the orifice.
Most studies show moderate correlation of
doppler techniques with oximetry data.
Sabry et al, studied above 4 methods with
cardiac cath calculated shunts.
First 2 methods showed moderate correlation
( r=0.54 & 0.56 ). Correlation improved
when large VSDs were excluded (r=0.62 &
0.66 ).
3rd
method showed no better correlation & it
was more time taking (r =0.57).
4th
method showed best correlation (r =0.82).
When large VSDs were excluded it was r = 0.9.
Sources of error
Mainly from the assumptions made
1) assumption that velocity profile is uniform
through out the cross sectional area
2) alignment of the doppler beam with the
direction of the flow ( angle < 200
)
3) turbulent flow in the vessels mainly in the
PA.
4) inaccuracy in measurement of cross
sectional area of vessels
- PA dimensions vary significantly during
cardiac cycle
5) errors in calculation because of presence of
additional defects
e.g PDA present downstream from the PA
6) semilunar valve regurgitation results in
overestimation of flow because of failure to measure
the regurgitant volume
7) All these methods overestimate the shunt in large
VSDs.
8) Quantification of PDA is not accurate due to the
turbulent flow across the defect & difficulty in
measuring the flow distal to the shunt.
Quantification of ASD
TTE estimation of ASD diameters & shunt
quantification by cath showed only fair correlation
(r=0.56).
• The size of the defect by transesophageal Doppler
color flow mapping correlated fairly well with the
size estimated at surgery (r = 0.73 ).
Other measurements by TTE for ASD are –
- RV/LV diameter (r=0.64)
- Area of PA (r=0.62)
- RV volume (r=o.71)
- PA/Ao (r=0.89)
Doppler colour flow mapping by
TEE
• Area of the ASD- calculated by assuming it to be
circular and taking the maximal Doppler color flow jet
width at the defect site as its diameter.
• The pulsed Doppler sample volume is to be placed
parallel to the shunt flow direction at the defect site to
obtain the mean velocity and flow duration.
• From these values, the shunt volume can
calculated as a product of the defect area, mean
velocity, flow duration and heart rate.
• The calculated shunt flow volume obtained by
transesophageal study showed a good correlation
with shunt flow volume (r = 0.91) and pulmonary to
systemic blood flow ratio (r = 0.84) obtained at
cardiac catheterization.
R  L shunt quantification
Aortic flow + VSD flow = Input to RV
Pulmonary flow + R  L shunt = output from RV.
Hence R  L shunt = Aortic flow + VSD flow –
Pulmonary flow.
Only fair correlation found with Cath data (r = 0.61 –
0.77 )
Automated cardiac flow
measurement method
ACM method – using spatial and temporal
integration of colour Doppler profiles .
Method  velocity profile in a region of interest
set on the flow tract is detected in each frame ,
imaged & recorded on the image memory
Flow volume rate is calculated by integration of
the velocity profile
Stroke volume is measured by temporal
integration, throughout the systolic period.
ACM method:
Advantages:
Quick and requires only two manual procedures.-
selection of systolic period in the stored image
memory, positioning at the area of interest
Calculations are done without tracing the
Doppler wave form to measure VTI
No need to measure area of outflow tract, because
the edge of the color profile is detected as the width
of the flow tract.
ACM method
ACM uses velocity profile across the flow tract
diameter, while in conventional pulsed Doppler
method the spectral velocity from a centrally
placed sample point is measured
Flow rate at each frame is temporally integrated
for the stroke volume calculation
Thus this method requires fewer assumptions
than pulsed Doppler method.
Martin et al found strong correlations between
ACM and invasive QP/QS ratio and the agreement
with invasive data was better using ACM than
using conventional echocardiographic method ( r
= 0.91 )
The only restriction remains to select manually
the systolic period and to carefully choose the
region of interest.
Gain must be optimised, allowing to see only one
colour throughout the outflow tract
Avoids the potential error of conventional PWD
method for the calculation of pulmonary output,
linked to the difficulty of measuring the exact
pulmonary diameter.
This method was applied to ASD & VSD with better
correlation .
Principle: oxygenated blood shunted from left
side of the heart to the rt cause an abnormal
increase (step – up) in the oxygen content or
saturation of blood in the chamber into which
shunting occurs
Dexter et al, max. increase in oxygen content
from,
RV to PA - 0.5 ml/dl
RA to RV - 1.0 ml/dl
SVC to RA - 2.0 ml/dl
Barratt -Boyes and Wood suggested that multiple
blood samples to be obtained from each Rt. heart
chamber.
Data to be averaged before applying Dexter
Criteria
Advantages
- easy to perform, results available immediately
- can ascertain the site of shunt
- magnitude of the shunt can be calculated
Detection of left to right shunt
Level of
shunt
Mean O2 sat % in
distal chambers
(Highest value in
proximal
chambers)
Mean O2
vol. % in
proximal
chambers
(Highest
value in
distal
chambers)
Min Qp/Qs
reqd. for
detection at
3L/min/m2
Possible causes
of step up
Atrial
(SVC/IVC
to RA)
≥7 (≥11) ≥1.3 (≥2.0) 1.5 – 1.9 ASD, RSOV,
VSD with TR,
coronary fistula
to RA,
anomalous PV
drainage
Ventricular
(RA to RV)
≥5 (≥10) ≥1.0 (≥1.7) 1.3 – 1.5 VSD, PDA with
PR, Primum
ASD, coronary
fistula to RV
Detection of left to right shunt
Level of
shunt
Mean O2 sat % in
distal chambers
(Highest value in
proximal
chambers)
Mean O2
vol. % in
proximal
chambers
(Highest
value in
distal
chambers)
Min Qp/Qs
reqd. for
detection at
3L/min/m2
Possible causes
of step up
Great
vessel
(RV to PA)
≥5 (≥5) ≥1.0 (≥1.0) ≥1.3 PDA, Aorta-
pulmonic
window,
Aberrant
coronary artery
origin
Any level
(SVC to
PA)
≥7 (≥8) ≥1.3 (≥1.5) ≥1.5 All of the above
Detection of left to right shunt
Chambers
sampled
Single
Sample
Multiple
samples
SVC – RA 7% 5%
RA – RV 5% 3%
RV – PA 4% 3%
PV/LA –
LV/SA
-3% -2%
Oximetry – limitations:
It lacks sensitivity – i.e. does not allow detection of
small left to right shunts
Requirement of steady state during collection of blood
samples
Calculation of Mixed venous saturation: Normal
variability of blood oxygen saturation in the Rt heart
chambers is strongly influenced by the magnitude of
systemic blood flow.
Higher levels of systemic blood flow  higher mixed
venous saturation  blunting of inter-chamber
variability.
Step up in oxygen content of the receiving chamber
depends also on the oxygen carrying capacity of blood
i.e. Hb conc.
Depends on the associated lesions also – like TR, PR.
Precautions
Samples at multiple sites to be obtained rapidly ,
should take < 7 min.
O2 saturation data rather than O2 content are
preferable,
Comparision of mean values is preferable to
highest values.
When the systemic blood flow at rest is low,
exercise shoud be used.
At higher FiO2 the dissolved O2 should also be
considered.
Shunt calculation
Qp = VO2 / (PVO2- PAO2) (Hb) (1.36) (10)
Qs = VO2 / (AoO2 – SVCO2) (Hb) (1.36) (10)
Qp eff. = VO2 / (PVO2 – SVC O2) (Hb) (1.36) (10)
Qp / Qs = (AoO2 – SVC O2) / (PVO2 – PAO2)
Q L  R shunt = (Qp – Qp eff)
 Q R  L = (Qs – Qs eff)
Carter formula
Determines the left to right shunting.
Best used when there is a smooth downstroke
They can detect a minimum shunt of 25 – 30%.
Simplified formula of Victoria & Gessner is used when
there is no smooth transition.
Calculation of Rt to Lt shunt
BT1
T2
%Qs = BT1 x P1x 100
BT1xP1 + 0.44xT2xP2
%Qs = percentage of systemic blood flow due to Rt to
Lt shunt.
PC1 = height of first peak.
BT1 = bulid up time from appearance to first peak.
T2 = time from injection to P2.
Shunt quantification

Weitere ähnliche Inhalte

Was ist angesagt?

Ventricular Septal defects Echocardiography
Ventricular Septal defects EchocardiographyVentricular Septal defects Echocardiography
Ventricular Septal defects EchocardiographySruthi Meenaxshi
 
Coronary anatomy and angiographic views
Coronary anatomy and angiographic viewsCoronary anatomy and angiographic views
Coronary anatomy and angiographic viewsthanigai arasu
 
BMV,PTMC,BALLOON MITRAL VALVOTOMY, BAL, VIRBHAN BALAI, DR VIRBHAN
BMV,PTMC,BALLOON MITRAL VALVOTOMY, BAL, VIRBHAN BALAI, DR VIRBHANBMV,PTMC,BALLOON MITRAL VALVOTOMY, BAL, VIRBHAN BALAI, DR VIRBHAN
BMV,PTMC,BALLOON MITRAL VALVOTOMY, BAL, VIRBHAN BALAI, DR VIRBHANDr Virbhan Balai
 
Percutaneous Pulmonary Valve Interventions
Percutaneous Pulmonary Valve InterventionsPercutaneous Pulmonary Valve Interventions
Percutaneous Pulmonary Valve InterventionsPraveen Nagula
 
Echo assessment of Aortic Regurgitation
Echo assessment of Aortic RegurgitationEcho assessment of Aortic Regurgitation
Echo assessment of Aortic RegurgitationMashiul Alam
 
ECHOCARDIOGRAPHIC EVALUATION OF AORTIC REGURGITATION
ECHOCARDIOGRAPHIC EVALUATION OF AORTIC REGURGITATIONECHOCARDIOGRAPHIC EVALUATION OF AORTIC REGURGITATION
ECHOCARDIOGRAPHIC EVALUATION OF AORTIC REGURGITATIONPraveen Nagula
 
FFR(fractional flow reserve)
FFR(fractional flow reserve)FFR(fractional flow reserve)
FFR(fractional flow reserve)DIPAK PATADE
 
Assessment of mitral valve by TEE
Assessment of mitral valve by TEEAssessment of mitral valve by TEE
Assessment of mitral valve by TEEjeetshitole
 
CORONARY BALLOONS PRACTICAL ASPECTS.pptx
CORONARY BALLOONS PRACTICAL ASPECTS.pptxCORONARY BALLOONS PRACTICAL ASPECTS.pptx
CORONARY BALLOONS PRACTICAL ASPECTS.pptxRohitWalse2
 
Basic haemodynamic assessment with echo (iHeartScan)
Basic haemodynamic assessment with echo (iHeartScan)Basic haemodynamic assessment with echo (iHeartScan)
Basic haemodynamic assessment with echo (iHeartScan)SCGH ED CME
 
HCM – Presentation, Hemodynamics and Intervention
HCM – Presentation, Hemodynamics and InterventionHCM – Presentation, Hemodynamics and Intervention
HCM – Presentation, Hemodynamics and InterventionAnkur Gupta
 
Echocardiography of Mitral regurgitation
Echocardiography of Mitral regurgitationEchocardiography of Mitral regurgitation
Echocardiography of Mitral regurgitationDr. Muhammad AzAm Shah
 
Diagnostic catheters for coronary angiography
Diagnostic catheters for coronary angiography Diagnostic catheters for coronary angiography
Diagnostic catheters for coronary angiography Aswin Rm
 
Aortic stenosis - Echocardiography
Aortic stenosis - EchocardiographyAortic stenosis - Echocardiography
Aortic stenosis - EchocardiographyAnkur Gupta
 
Assessment of prosthetic valve function
Assessment of prosthetic valve functionAssessment of prosthetic valve function
Assessment of prosthetic valve functionSwapnil Garde
 

Was ist angesagt? (20)

Ventricular Septal defects Echocardiography
Ventricular Septal defects EchocardiographyVentricular Septal defects Echocardiography
Ventricular Septal defects Echocardiography
 
Coronary anatomy and angiographic views
Coronary anatomy and angiographic viewsCoronary anatomy and angiographic views
Coronary anatomy and angiographic views
 
Echo assessment of mitral regurgitation
Echo assessment of mitral regurgitationEcho assessment of mitral regurgitation
Echo assessment of mitral regurgitation
 
BMV,PTMC,BALLOON MITRAL VALVOTOMY, BAL, VIRBHAN BALAI, DR VIRBHAN
BMV,PTMC,BALLOON MITRAL VALVOTOMY, BAL, VIRBHAN BALAI, DR VIRBHANBMV,PTMC,BALLOON MITRAL VALVOTOMY, BAL, VIRBHAN BALAI, DR VIRBHAN
BMV,PTMC,BALLOON MITRAL VALVOTOMY, BAL, VIRBHAN BALAI, DR VIRBHAN
 
Percutaneous Pulmonary Valve Interventions
Percutaneous Pulmonary Valve InterventionsPercutaneous Pulmonary Valve Interventions
Percutaneous Pulmonary Valve Interventions
 
Echo assessment of Aortic Regurgitation
Echo assessment of Aortic RegurgitationEcho assessment of Aortic Regurgitation
Echo assessment of Aortic Regurgitation
 
ECHOCARDIOGRAPHIC EVALUATION OF AORTIC REGURGITATION
ECHOCARDIOGRAPHIC EVALUATION OF AORTIC REGURGITATIONECHOCARDIOGRAPHIC EVALUATION OF AORTIC REGURGITATION
ECHOCARDIOGRAPHIC EVALUATION OF AORTIC REGURGITATION
 
FFR(fractional flow reserve)
FFR(fractional flow reserve)FFR(fractional flow reserve)
FFR(fractional flow reserve)
 
Assessment of mitral valve by TEE
Assessment of mitral valve by TEEAssessment of mitral valve by TEE
Assessment of mitral valve by TEE
 
Cath hemodynamics vir
Cath hemodynamics virCath hemodynamics vir
Cath hemodynamics vir
 
CORONARY BALLOONS PRACTICAL ASPECTS.pptx
CORONARY BALLOONS PRACTICAL ASPECTS.pptxCORONARY BALLOONS PRACTICAL ASPECTS.pptx
CORONARY BALLOONS PRACTICAL ASPECTS.pptx
 
Basic haemodynamic assessment with echo (iHeartScan)
Basic haemodynamic assessment with echo (iHeartScan)Basic haemodynamic assessment with echo (iHeartScan)
Basic haemodynamic assessment with echo (iHeartScan)
 
Trans septal puncture
Trans septal punctureTrans septal puncture
Trans septal puncture
 
Right Ventricle Echocardiography
Right Ventricle EchocardiographyRight Ventricle Echocardiography
Right Ventricle Echocardiography
 
HCM – Presentation, Hemodynamics and Intervention
HCM – Presentation, Hemodynamics and InterventionHCM – Presentation, Hemodynamics and Intervention
HCM – Presentation, Hemodynamics and Intervention
 
Echocardiography of Mitral regurgitation
Echocardiography of Mitral regurgitationEchocardiography of Mitral regurgitation
Echocardiography of Mitral regurgitation
 
Diagnostic catheters for coronary angiography
Diagnostic catheters for coronary angiography Diagnostic catheters for coronary angiography
Diagnostic catheters for coronary angiography
 
Echocardiography in mitral stenosis
Echocardiography in mitral stenosisEchocardiography in mitral stenosis
Echocardiography in mitral stenosis
 
Aortic stenosis - Echocardiography
Aortic stenosis - EchocardiographyAortic stenosis - Echocardiography
Aortic stenosis - Echocardiography
 
Assessment of prosthetic valve function
Assessment of prosthetic valve functionAssessment of prosthetic valve function
Assessment of prosthetic valve function
 

Andere mochten auch

Shunt quantification
Shunt quantificationShunt quantification
Shunt quantificationGirish Mishra
 
Intracardiac shunts
Intracardiac shuntsIntracardiac shunts
Intracardiac shuntsFuad Farooq
 
Pulmonary hypertension with cardiac shunt determination
Pulmonary hypertension with cardiac shunt determinationPulmonary hypertension with cardiac shunt determination
Pulmonary hypertension with cardiac shunt determinationDr. Rajesh Das
 
Calculation of bidirectional shunts
Calculation of bidirectional shuntsCalculation of bidirectional shunts
Calculation of bidirectional shuntsRamachandra Barik
 
Shunt quantification and reversibility
Shunt quantification and reversibilityShunt quantification and reversibility
Shunt quantification and reversibilityGOPAL GHOSH
 
Coronary Calcium and other CVD Risk Biomarkers: From Epidemiology to Comparat...
Coronary Calcium and other CVD Risk Biomarkers: From Epidemiology to Comparat...Coronary Calcium and other CVD Risk Biomarkers: From Epidemiology to Comparat...
Coronary Calcium and other CVD Risk Biomarkers: From Epidemiology to Comparat...CTSI at UCSF
 
Framingham
FraminghamFramingham
Framinghamkeisy PS
 
All About Framingham Heart Study
All About Framingham Heart StudyAll About Framingham Heart Study
All About Framingham Heart StudyALEXANDRU ANDRITOIU
 
Takayasu's arteritis
Takayasu's arteritisTakayasu's arteritis
Takayasu's arteritisAnkur Gupta
 
Natural history of right to left shunts
Natural history of right to left shuntsNatural history of right to left shunts
Natural history of right to left shuntsdinanathkumar
 
Portal hypertension surgical management
Portal hypertension surgical management Portal hypertension surgical management
Portal hypertension surgical management nikhilameerchetty
 
Constrictive pericarditis
Constrictive pericarditisConstrictive pericarditis
Constrictive pericarditisAnkur Gupta
 
Euroecho2010 intracardiac-shunts
Euroecho2010 intracardiac-shuntsEuroecho2010 intracardiac-shunts
Euroecho2010 intracardiac-shuntsoussama El-h
 
Echo Differentiation of Restrictive Cardiomyopathy and Constrictive Pericarditis
Echo Differentiation of Restrictive Cardiomyopathy and Constrictive PericarditisEcho Differentiation of Restrictive Cardiomyopathy and Constrictive Pericarditis
Echo Differentiation of Restrictive Cardiomyopathy and Constrictive PericarditisJunhao Koh
 
Intracoronary optical coherence tomography
Intracoronary optical coherence tomographyIntracoronary optical coherence tomography
Intracoronary optical coherence tomographyBHAWANI SHANKAR
 

Andere mochten auch (20)

Shunt quantification
Shunt quantificationShunt quantification
Shunt quantification
 
Intracardiac shunts
Intracardiac shuntsIntracardiac shunts
Intracardiac shunts
 
Pulmonary hypertension with cardiac shunt determination
Pulmonary hypertension with cardiac shunt determinationPulmonary hypertension with cardiac shunt determination
Pulmonary hypertension with cardiac shunt determination
 
Surgical shunts
Surgical shuntsSurgical shunts
Surgical shunts
 
Calculation of bidirectional shunts
Calculation of bidirectional shuntsCalculation of bidirectional shunts
Calculation of bidirectional shunts
 
Shunt quantification and reversibility
Shunt quantification and reversibilityShunt quantification and reversibility
Shunt quantification and reversibility
 
Portal hypertension
Portal hypertension Portal hypertension
Portal hypertension
 
Coronary Calcium and other CVD Risk Biomarkers: From Epidemiology to Comparat...
Coronary Calcium and other CVD Risk Biomarkers: From Epidemiology to Comparat...Coronary Calcium and other CVD Risk Biomarkers: From Epidemiology to Comparat...
Coronary Calcium and other CVD Risk Biomarkers: From Epidemiology to Comparat...
 
Framingham
FraminghamFramingham
Framingham
 
All About Framingham Heart Study
All About Framingham Heart StudyAll About Framingham Heart Study
All About Framingham Heart Study
 
Jupiter Trial
Jupiter TrialJupiter Trial
Jupiter Trial
 
Coronary physiology
Coronary physiologyCoronary physiology
Coronary physiology
 
Takayasu's arteritis
Takayasu's arteritisTakayasu's arteritis
Takayasu's arteritis
 
Natural history of right to left shunts
Natural history of right to left shuntsNatural history of right to left shunts
Natural history of right to left shunts
 
Portal hypertension surgical management
Portal hypertension surgical management Portal hypertension surgical management
Portal hypertension surgical management
 
Treatment of portal hypertension
Treatment of portal hypertensionTreatment of portal hypertension
Treatment of portal hypertension
 
Constrictive pericarditis
Constrictive pericarditisConstrictive pericarditis
Constrictive pericarditis
 
Euroecho2010 intracardiac-shunts
Euroecho2010 intracardiac-shuntsEuroecho2010 intracardiac-shunts
Euroecho2010 intracardiac-shunts
 
Echo Differentiation of Restrictive Cardiomyopathy and Constrictive Pericarditis
Echo Differentiation of Restrictive Cardiomyopathy and Constrictive PericarditisEcho Differentiation of Restrictive Cardiomyopathy and Constrictive Pericarditis
Echo Differentiation of Restrictive Cardiomyopathy and Constrictive Pericarditis
 
Intracoronary optical coherence tomography
Intracoronary optical coherence tomographyIntracoronary optical coherence tomography
Intracoronary optical coherence tomography
 

Ähnlich wie Shunt quantification

Flowvolumefinalversion
FlowvolumefinalversionFlowvolumefinalversion
FlowvolumefinalversionDuoc Vang
 
Basics in echocardiography - an initiative in evaluation of valvular heart di...
Basics in echocardiography - an initiative in evaluation of valvular heart di...Basics in echocardiography - an initiative in evaluation of valvular heart di...
Basics in echocardiography - an initiative in evaluation of valvular heart di...Praveen Nagula
 
Doppler echocardiography
Doppler echocardiographyDoppler echocardiography
Doppler echocardiographydian lestari
 
Dr masjedi hemodynamic monitoring in ICU
Dr masjedi hemodynamic monitoring in ICUDr masjedi hemodynamic monitoring in ICU
Dr masjedi hemodynamic monitoring in ICUmansoor masjedi
 
Echo assesmentof rv function
Echo assesmentof rv functionEcho assesmentof rv function
Echo assesmentof rv functionDeepak Agrawal
 
Aortic valve assessment
Aortic valve assessmentAortic valve assessment
Aortic valve assessmentPraveen Neema
 
Echo assessment of aortic stenosis
Echo assessment of aortic stenosisEcho assessment of aortic stenosis
Echo assessment of aortic stenosisNizam Uddin
 
Coronary CT- angiogram: physics
Coronary CT- angiogram:  physics Coronary CT- angiogram:  physics
Coronary CT- angiogram: physics REMIX MAHARJAN
 
Echocardiography assessment of Aortic Regurgitation severity
Echocardiography assessment of Aortic Regurgitation severityEchocardiography assessment of Aortic Regurgitation severity
Echocardiography assessment of Aortic Regurgitation severityPRAVEEN GUPTA
 
Hemodynamics in echo lab by Dr. Ranjeet S.Palkar
Hemodynamics  in echo lab by Dr. Ranjeet S.PalkarHemodynamics  in echo lab by Dr. Ranjeet S.Palkar
Hemodynamics in echo lab by Dr. Ranjeet S.PalkarRanjeet Palkar
 
ASSESSMENT OF RV FUNCTION.pptx
ASSESSMENT OF RV FUNCTION.pptxASSESSMENT OF RV FUNCTION.pptx
ASSESSMENT OF RV FUNCTION.pptxAlAmin837379
 
echoassessmentofrvfunction-190104172505.pptx
echoassessmentofrvfunction-190104172505.pptxechoassessmentofrvfunction-190104172505.pptx
echoassessmentofrvfunction-190104172505.pptxOmarMedina18478
 
Renal doppler ultrasound
Renal doppler ultrasoundRenal doppler ultrasound
Renal doppler ultrasoundAhmed Bahnassy
 
Echo assessment of Aortic Stenosis
Echo assessment of Aortic StenosisEcho assessment of Aortic Stenosis
Echo assessment of Aortic Stenosisdrranjithmp
 
Singular value decomposition filtering in high-frame-rate cardiac vector flow...
Singular value decomposition filtering in high-frame-rate cardiac vector flow...Singular value decomposition filtering in high-frame-rate cardiac vector flow...
Singular value decomposition filtering in high-frame-rate cardiac vector flow...journalBEEI
 
How to echo series....Aortic stenosis 2017 guidelines
How to echo series....Aortic stenosis 2017 guidelinesHow to echo series....Aortic stenosis 2017 guidelines
How to echo series....Aortic stenosis 2017 guidelinesVinayak Vadgaonkar
 
Non invasive estimation of pulmonary vascular resistance in patients of pulmo...
Non invasive estimation of pulmonary vascular resistance in patients of pulmo...Non invasive estimation of pulmonary vascular resistance in patients of pulmo...
Non invasive estimation of pulmonary vascular resistance in patients of pulmo...Arindam Pande
 
Ct angio in cardiology
Ct angio in cardiologyCt angio in cardiology
Ct angio in cardiologyAmit Verma
 

Ähnlich wie Shunt quantification (20)

MR Tawfeeq.pptx
MR Tawfeeq.pptxMR Tawfeeq.pptx
MR Tawfeeq.pptx
 
Flowvolumefinalversion
FlowvolumefinalversionFlowvolumefinalversion
Flowvolumefinalversion
 
Basics in echocardiography - an initiative in evaluation of valvular heart di...
Basics in echocardiography - an initiative in evaluation of valvular heart di...Basics in echocardiography - an initiative in evaluation of valvular heart di...
Basics in echocardiography - an initiative in evaluation of valvular heart di...
 
Doppler echocardiography
Doppler echocardiographyDoppler echocardiography
Doppler echocardiography
 
Dr masjedi hemodynamic monitoring in ICU
Dr masjedi hemodynamic monitoring in ICUDr masjedi hemodynamic monitoring in ICU
Dr masjedi hemodynamic monitoring in ICU
 
Echo assesmentof rv function
Echo assesmentof rv functionEcho assesmentof rv function
Echo assesmentof rv function
 
Aortic valve assessment
Aortic valve assessmentAortic valve assessment
Aortic valve assessment
 
Echo assessment of aortic stenosis
Echo assessment of aortic stenosisEcho assessment of aortic stenosis
Echo assessment of aortic stenosis
 
Coronary CT- angiogram: physics
Coronary CT- angiogram:  physics Coronary CT- angiogram:  physics
Coronary CT- angiogram: physics
 
Echocardiography assessment of Aortic Regurgitation severity
Echocardiography assessment of Aortic Regurgitation severityEchocardiography assessment of Aortic Regurgitation severity
Echocardiography assessment of Aortic Regurgitation severity
 
Hemodynamics in echo lab by Dr. Ranjeet S.Palkar
Hemodynamics  in echo lab by Dr. Ranjeet S.PalkarHemodynamics  in echo lab by Dr. Ranjeet S.Palkar
Hemodynamics in echo lab by Dr. Ranjeet S.Palkar
 
ASSESSMENT OF RV FUNCTION.pptx
ASSESSMENT OF RV FUNCTION.pptxASSESSMENT OF RV FUNCTION.pptx
ASSESSMENT OF RV FUNCTION.pptx
 
echoassessmentofrvfunction-190104172505.pptx
echoassessmentofrvfunction-190104172505.pptxechoassessmentofrvfunction-190104172505.pptx
echoassessmentofrvfunction-190104172505.pptx
 
Renal Doppler Ultrasound
Renal Doppler UltrasoundRenal Doppler Ultrasound
Renal Doppler Ultrasound
 
Renal doppler ultrasound
Renal doppler ultrasoundRenal doppler ultrasound
Renal doppler ultrasound
 
Echo assessment of Aortic Stenosis
Echo assessment of Aortic StenosisEcho assessment of Aortic Stenosis
Echo assessment of Aortic Stenosis
 
Singular value decomposition filtering in high-frame-rate cardiac vector flow...
Singular value decomposition filtering in high-frame-rate cardiac vector flow...Singular value decomposition filtering in high-frame-rate cardiac vector flow...
Singular value decomposition filtering in high-frame-rate cardiac vector flow...
 
How to echo series....Aortic stenosis 2017 guidelines
How to echo series....Aortic stenosis 2017 guidelinesHow to echo series....Aortic stenosis 2017 guidelines
How to echo series....Aortic stenosis 2017 guidelines
 
Non invasive estimation of pulmonary vascular resistance in patients of pulmo...
Non invasive estimation of pulmonary vascular resistance in patients of pulmo...Non invasive estimation of pulmonary vascular resistance in patients of pulmo...
Non invasive estimation of pulmonary vascular resistance in patients of pulmo...
 
Ct angio in cardiology
Ct angio in cardiologyCt angio in cardiology
Ct angio in cardiology
 

Kürzlich hochgeladen

VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋TANUJA PANDEY
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...narwatsonia7
 
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟ 9332606886 ⟟ Call Me For G...
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟  9332606886 ⟟ Call Me For G...Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟  9332606886 ⟟ Call Me For G...
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟ 9332606886 ⟟ Call Me For G...narwatsonia7
 
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Dipal Arora
 
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...tanya dube
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...chandars293
 
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...Taniya Sharma
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Servicevidya singh
 
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...perfect solution
 
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...indiancallgirl4rent
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...Taniya Sharma
 
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...parulsinha
 
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeCall Girls Delhi
 
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service AvailableCall Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...vidya singh
 
Bangalore Call Girls Nelamangala Number 9332606886 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 9332606886  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 9332606886  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 9332606886 Meetin With Bangalore Esc...narwatsonia7
 
Call Girls Bareilly Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Bareilly Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 

Kürzlich hochgeladen (20)

VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
 
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟ 9332606886 ⟟ Call Me For G...
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟  9332606886 ⟟ Call Me For G...Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟  9332606886 ⟟ Call Me For G...
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟ 9332606886 ⟟ Call Me For G...
 
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
 
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
 
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
 
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service Available
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
 
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
 
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
 
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
 
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service AvailableCall Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
 
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
 
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
 
Bangalore Call Girls Nelamangala Number 9332606886 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 9332606886  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 9332606886  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 9332606886 Meetin With Bangalore Esc...
 
Call Girls Bareilly Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Bareilly Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 8250077686 Top Class Call Girl Service Available
 

Shunt quantification

  • 1. By: Dr. Ankur Gupta Resident, DM Cardiolgy
  • 2. Introduction Majority of the congenital lesions are asso. with intracardiac shunts. Detection, localisation and quantification of shunts – integral part of hemodynamic evaluation of these patients. Quantification of shunts by: Invasive Noninvasive methods.
  • 3. Non invasive methods Clinical evaluation CXR and ECG Echo and Doppler MRI CT Nuclear scanning
  • 5.
  • 6. 2 –D Echocardiography: direct visualization of any communication and their shunting excellent for localization of shunt limited in the ability to quantify shunt Indirect estimations ASD with large shunting  RA, RV dilated  diastolic septal flattening VSD with large shunting  LA, LV dilatation  Systolic and diastolic septal flattening
  • 7. Contrast Echo Agitated saline  Detection of small R  L shunts Large left to right shunts may cause a negative contrast jet in the Rt. atrium
  • 8. Doppler echocardiography Most of the studies were done in ASD & VSD. Based on the measurement of flow across the valves or the defect itself. Based on the formula that flow can be calculated from CSA & VTI.
  • 9. Volumetric flow measurement requires - calculation of temporal mean flow velocity and cross sectional area - velocity profile is uniform across the vessel - single sampling of the flow velocity in the center of the vessel is recorded To calculate  computerised calculation of the area under the doppler curve/ flow period
  • 10. For systemic flow Aortic velocity is recorded from the apical view. Vessel diameter is measured from the parasternal long axis view.
  • 11. Pulmonary blood flow: Pulm. mean velocity is recorded from the parasternal short axis view In case of large shunt and a shunt located close to PA, the doppler tracing shows spectral broadening or signs of disturbed flow, in such cases uniform velocity cannot be assumed Doppler tracing cannot be used if any PS is present.
  • 12. Measurement of the cross sectional area - at the valve annulus (as it is the flow limiting point of the vessel) - measurement is made from the inner edge to the inner edge. - diameter is measured at the early systole – time when vessel is at peak systolic dimension. - flow velocity should be measured at same location
  • 13. Methods: 1.Qp/Qs calculation from flow velocity (velocity time integral) and cross sectional area Qp/Qs = Area of Pa x VTI Area of Aorta x VTI Sanders et al used pulmonary flow with good correlation , r=0.85. Barron et al used Mitral flow with better correlation , r= 0.9 vs 0.69.
  • 14. 2. Simplified method - the square of the pulmonary to aortic diameter ratio was substituted for the area ratio and the flow peak velocity ratio for the velocity time integral.  Qp/Qs = Vel. Pk PA x r2 Vel. Pk Ao x r2 r= radius / diameter square of the ratio of the pulmonary to aortic luminal diameter (or radius)is used instead of vessel areas.
  • 15. 3. Qp/Qs = (pulm + mitral flow)/2 (tricuspid + aortic flow)/2  Assumption that tricuspid and aortic flow represent systemic flow and pulmonary & mitral flows represent pulmonary flow.  Maximal diameter of mitral and tricuspid valve annulus is measured during diastole in an apical 4 –chamber view & area calculated assuming the annulus to be circular.  Laborious & time taking.  Combines measurements from all the 4 valves but was not shown to be better than the previous 2.
  • 16. 4. Qp/Qs = Shunt flow + aortic flow aortic flow  Shunt (VSD) flow  as the product of the VTI and color-derived cross-sectional area of VSD jet measured at its narrowest point.  This method does not use the pulmonary flow & outflow diameter which are more variable.
  • 17. Difficulties : It is assumed that 1. The defect is circular. 2. The size does not vary during systole. 3. No angle correction is needed. 4 . Flow velocity is uniform across the orifice.
  • 18. Most studies show moderate correlation of doppler techniques with oximetry data. Sabry et al, studied above 4 methods with cardiac cath calculated shunts. First 2 methods showed moderate correlation ( r=0.54 & 0.56 ). Correlation improved when large VSDs were excluded (r=0.62 & 0.66 ). 3rd method showed no better correlation & it was more time taking (r =0.57). 4th method showed best correlation (r =0.82). When large VSDs were excluded it was r = 0.9.
  • 19. Sources of error Mainly from the assumptions made 1) assumption that velocity profile is uniform through out the cross sectional area 2) alignment of the doppler beam with the direction of the flow ( angle < 200 ) 3) turbulent flow in the vessels mainly in the PA. 4) inaccuracy in measurement of cross sectional area of vessels - PA dimensions vary significantly during cardiac cycle
  • 20. 5) errors in calculation because of presence of additional defects e.g PDA present downstream from the PA 6) semilunar valve regurgitation results in overestimation of flow because of failure to measure the regurgitant volume 7) All these methods overestimate the shunt in large VSDs. 8) Quantification of PDA is not accurate due to the turbulent flow across the defect & difficulty in measuring the flow distal to the shunt.
  • 21. Quantification of ASD TTE estimation of ASD diameters & shunt quantification by cath showed only fair correlation (r=0.56). • The size of the defect by transesophageal Doppler color flow mapping correlated fairly well with the size estimated at surgery (r = 0.73 ). Other measurements by TTE for ASD are – - RV/LV diameter (r=0.64) - Area of PA (r=0.62) - RV volume (r=o.71) - PA/Ao (r=0.89)
  • 22. Doppler colour flow mapping by TEE • Area of the ASD- calculated by assuming it to be circular and taking the maximal Doppler color flow jet width at the defect site as its diameter. • The pulsed Doppler sample volume is to be placed parallel to the shunt flow direction at the defect site to obtain the mean velocity and flow duration.
  • 23. • From these values, the shunt volume can calculated as a product of the defect area, mean velocity, flow duration and heart rate. • The calculated shunt flow volume obtained by transesophageal study showed a good correlation with shunt flow volume (r = 0.91) and pulmonary to systemic blood flow ratio (r = 0.84) obtained at cardiac catheterization.
  • 24. R  L shunt quantification Aortic flow + VSD flow = Input to RV Pulmonary flow + R  L shunt = output from RV. Hence R  L shunt = Aortic flow + VSD flow – Pulmonary flow. Only fair correlation found with Cath data (r = 0.61 – 0.77 )
  • 25. Automated cardiac flow measurement method ACM method – using spatial and temporal integration of colour Doppler profiles . Method  velocity profile in a region of interest set on the flow tract is detected in each frame , imaged & recorded on the image memory Flow volume rate is calculated by integration of the velocity profile Stroke volume is measured by temporal integration, throughout the systolic period.
  • 26. ACM method: Advantages: Quick and requires only two manual procedures.- selection of systolic period in the stored image memory, positioning at the area of interest Calculations are done without tracing the Doppler wave form to measure VTI No need to measure area of outflow tract, because the edge of the color profile is detected as the width of the flow tract.
  • 27. ACM method ACM uses velocity profile across the flow tract diameter, while in conventional pulsed Doppler method the spectral velocity from a centrally placed sample point is measured Flow rate at each frame is temporally integrated for the stroke volume calculation Thus this method requires fewer assumptions than pulsed Doppler method.
  • 28. Martin et al found strong correlations between ACM and invasive QP/QS ratio and the agreement with invasive data was better using ACM than using conventional echocardiographic method ( r = 0.91 ) The only restriction remains to select manually the systolic period and to carefully choose the region of interest. Gain must be optimised, allowing to see only one colour throughout the outflow tract
  • 29. Avoids the potential error of conventional PWD method for the calculation of pulmonary output, linked to the difficulty of measuring the exact pulmonary diameter. This method was applied to ASD & VSD with better correlation .
  • 30.
  • 31. Principle: oxygenated blood shunted from left side of the heart to the rt cause an abnormal increase (step – up) in the oxygen content or saturation of blood in the chamber into which shunting occurs Dexter et al, max. increase in oxygen content from, RV to PA - 0.5 ml/dl RA to RV - 1.0 ml/dl SVC to RA - 2.0 ml/dl
  • 32. Barratt -Boyes and Wood suggested that multiple blood samples to be obtained from each Rt. heart chamber. Data to be averaged before applying Dexter Criteria Advantages - easy to perform, results available immediately - can ascertain the site of shunt - magnitude of the shunt can be calculated
  • 33. Detection of left to right shunt Level of shunt Mean O2 sat % in distal chambers (Highest value in proximal chambers) Mean O2 vol. % in proximal chambers (Highest value in distal chambers) Min Qp/Qs reqd. for detection at 3L/min/m2 Possible causes of step up Atrial (SVC/IVC to RA) ≥7 (≥11) ≥1.3 (≥2.0) 1.5 – 1.9 ASD, RSOV, VSD with TR, coronary fistula to RA, anomalous PV drainage Ventricular (RA to RV) ≥5 (≥10) ≥1.0 (≥1.7) 1.3 – 1.5 VSD, PDA with PR, Primum ASD, coronary fistula to RV
  • 34. Detection of left to right shunt Level of shunt Mean O2 sat % in distal chambers (Highest value in proximal chambers) Mean O2 vol. % in proximal chambers (Highest value in distal chambers) Min Qp/Qs reqd. for detection at 3L/min/m2 Possible causes of step up Great vessel (RV to PA) ≥5 (≥5) ≥1.0 (≥1.0) ≥1.3 PDA, Aorta- pulmonic window, Aberrant coronary artery origin Any level (SVC to PA) ≥7 (≥8) ≥1.3 (≥1.5) ≥1.5 All of the above
  • 35. Detection of left to right shunt Chambers sampled Single Sample Multiple samples SVC – RA 7% 5% RA – RV 5% 3% RV – PA 4% 3% PV/LA – LV/SA -3% -2%
  • 36. Oximetry – limitations: It lacks sensitivity – i.e. does not allow detection of small left to right shunts Requirement of steady state during collection of blood samples Calculation of Mixed venous saturation: Normal variability of blood oxygen saturation in the Rt heart chambers is strongly influenced by the magnitude of systemic blood flow. Higher levels of systemic blood flow  higher mixed venous saturation  blunting of inter-chamber variability.
  • 37. Step up in oxygen content of the receiving chamber depends also on the oxygen carrying capacity of blood i.e. Hb conc. Depends on the associated lesions also – like TR, PR.
  • 38. Precautions Samples at multiple sites to be obtained rapidly , should take < 7 min. O2 saturation data rather than O2 content are preferable, Comparision of mean values is preferable to highest values. When the systemic blood flow at rest is low, exercise shoud be used. At higher FiO2 the dissolved O2 should also be considered.
  • 39. Shunt calculation Qp = VO2 / (PVO2- PAO2) (Hb) (1.36) (10) Qs = VO2 / (AoO2 – SVCO2) (Hb) (1.36) (10) Qp eff. = VO2 / (PVO2 – SVC O2) (Hb) (1.36) (10) Qp / Qs = (AoO2 – SVC O2) / (PVO2 – PAO2) Q L  R shunt = (Qp – Qp eff)  Q R  L = (Qs – Qs eff)
  • 40. Carter formula Determines the left to right shunting. Best used when there is a smooth downstroke They can detect a minimum shunt of 25 – 30%. Simplified formula of Victoria & Gessner is used when there is no smooth transition.
  • 41. Calculation of Rt to Lt shunt BT1 T2
  • 42. %Qs = BT1 x P1x 100 BT1xP1 + 0.44xT2xP2 %Qs = percentage of systemic blood flow due to Rt to Lt shunt. PC1 = height of first peak. BT1 = bulid up time from appearance to first peak. T2 = time from injection to P2.