1. PRODUCT CONFORMITY CERTIFICATE
This is to certify that the
Graphite 52M FID Analyser
manufactured by:
Environnement SA
111 Boulevard Robespierre
78304 Poissy Cedex
France
has been assessed by Sira Certification Service
and for the conditions stated on this certificate complies with:
MCERTS Performance Standards for Continuous Emission
Monitoring Systems, Version 2, Revision 1 (April 2003)
Certification Range
TOC
Project No:
Certificate No:
Initial Certification:
This Certificate Issued
Renewal Date:
:
0 to 15 mgC/m3
674/0104
Sira MC 060082/05
22 September 2006
23 August 2011
21 September 2011
Technical Director
MCERTS is operated on behalf of the Environment Agency by
Sira Certification Service
12 Acorn Industrial Park, Crayford Road, Crayford
Dartford, Kent, UK, DA1 4AL
Tel: 01322 520500 Fax: 01322 520501
This certificate may only be reproduced in its entirety and without change
Page 1 of 4
2. Approved Site Application
On the basis of the assessment and the ranges required for compliance with EU Directives this
instrument is considered suitable for use on waste incineration applications.
Any potential user should ensure, in consultation with the manufacturer, that the emission
monitoring system is suitable for the process on which it will be installed. For general
guidance on stack emission monitoring techniques refer to Environment Agency Technical
Guidance Note M2: Monitoring of stack emissions to air. This is available on the Agency’s
website at www.mcerts.net.
Fuel type H2/He mix (40% / 60%) was used during the field test.
Basis of Certification
This certification is based on the following Test Report(s) and on Sira’s assessment and
ongoing surveillance of the product and the manufacturing process:
TÜV KÖln Report Number: 936/983008 dated October 1994 (R1)
TÜV KÖln Report Number: 936/21205431/A dated 29/03/06 (R2)
Scientifics Report Number: LAB 06450/Version 1 dated 03/05/06 (R3)
TÜV reports are accepted on the basis of the Environment Agency’s document ‘MCERTS –
Guidance on the acceptance of German type approval test reports for CEMS’ Version 2
(October 2003)
Product Certified
The Graphite 52M measuring system consists of the following parts:
Graphite 52M analyser
Heated sampling system which comprises a probe manufactured by Environnement SA
model CA-P (certified SEC MIR9000 system) and HOFI box including a heated filter
model F05-150-57-0.5
This certificate applies to all instruments fitted with software version V2 onwards (serial
number 405).
Certificate No:
This Certificate Issued:
Sira MC 060082/05
23 August 2011
This certificate may only be reproduced in its entirety and without change
Page 2 of 4
3. Certified Performance
The instrument was evaluated for use under the following conditions:
Ambient Temperature Range: +50C to +400C
Unless otherwise stated the evaluation was carried out on the certification range 0 to 15 mgC/m3.
Test
Results expressed as % of the
certification range
<1
<2
<4
<0.5
Other results
3
MCERTS
specification
Linearity
0.15 mgCm
<0.4mgCm
3
Cross-sensitivity
0.5 mgCm
3
<1.0mgCm
3
o
Temperature dependent zero shift
-0.05
<0.3%/ C
Temperature dependent upper
reference point shift
0.07
<0.3%/ C
o
Response time
3s
Detection Limit
0.08 mgCm
<0.4mgC/m3
Sample gas pressure
See note 2
To be reported
Sample gas temperature
See note 2
To be reported
-8.5%
<15%
<60s
3
Control gas mixture – low levels of TOC
Effect of oxygen synergism
-0.2mgC/m
3
<0.8mgC/m
3
Range of response factors:
1.00
1.02
0.89
Aliphatic hydrocarbons
Aromatic hydrocarbons
Dichloromethane
090-1.10
0.85-1.10
0.75-1.15
Vibration test
See note 1
(10 to 60Hz (0.3mm), 60Hz to 150Hz at
2
19.6m/s )
To be reported
Mains voltage
(10V increments, varying voltage from
230V to 190V and then 230V to 250V)
Certificate No:
This Certificate Issued:
-0.07
<2%
Sira MC 060082/05
23 August 2011
This certificate may only be reproduced in its entirety and without change
Page 3 of 4
4. Test
Results expressed as % of the
certification range
<1
<2
<4
<0.5
Availability
99.9%
Zero drift during field trial
Note 3
>95%
-0.61
Upper reference point drift during field
2%/week
-2.27
Note 3
Note 1:
Note 2:
Note 3:
>95%
Note 3
3
Maintenance Interval
MCERTS
specification
98.8%
Analysis function (field)
Other results
Note 3
4%/week
2 week
To be reported
Test not applicable as Graphite 52M is an extractive analyser.
Test not applicable as the product is an extractive analyser
The three months field test was performed on a clinical waste incineration site.
Description:
The Graphite is a totally heated TOC analyser which uses FID (Flames Ionisation Detector) to
measure organic gaseous components in an extractive system. The system is heated
throughout at a nominal temperature of 191oC, that is adjustable and the analyser is coupled
to a heated line.
The unit is used with a heated sampling system, with a heated sample probe, a heated filter
and a heated line. The unit can measure Total VOC expressed in methane, expressed in
carbon or can measure selectively total non methane compounds and methane.
General Notes
1.
This certificate is based upon the equipment tested. The Manufacturer is responsible for
ensuring that on-going production complies with the standard(s) and performance
criteria defined in this Certificate. The Manufacturer is required to maintain an approved
quality management system controlling the manufacture of the certified product. Both
the product and the quality management system shall be subject to regular surveillance
according to ‘Regulations Applicable to the Holders of Sira Certificates’. The design of
the product certified is defined in the Sira Design Schedule for certificate No. Sira MC
060082/04.
2.
If certified product is found not to comply, Sira Certification Service should be notified
immediately at the address shown on this certificate.
3.
The Certification Marks that can be applied to the product or used in publicity material
are defined in ‘Regulations Applicable to the Holders of Sira Certificates’.
4.
This document remains the property of Sira and shall be returned when requested by
the company.
Certificate No:
This Certificate Issued:
Sira MC 060082/05
23 August 2011
This certificate may only be reproduced in its entirety and without change
Page 4 of 4