2. H
P
igh
ressure
L iquid
C hromatography
Since high pressure is used when compared to classical chromatography
Separation of the components from a mixture is achieved by pumping
mobile phase at High pressure using appropriate displacement pumps
or gas pressure. (Due to the small particle size (3-5 um).
5. • Michael Tswett (1906) -separation of plant pigments
• Martin and Synge (1941) liquid-liquid partition chromatography
• 1952 Nobel Prize
• Other chromatography Techniques
• Thin-layer chromatography (TLC)
• Paper chromatography
• Preparative column chromatography
• Medium pressure chromatography
• Gas chromatography
• Ion-exchange chromatography
• Size-exclusion chromatography (Gel Permeation Chromatography)
• HPLC
• VLC
• Flash Chromatography
• Affinity Chromatography
• Chiral Chromatography
• Super Critical Fluid Chromatography
5
6. What means chromatography ?
AABC
AC
ABBC
C C
B
C
B
A
C
B
A
B
C
CC
C C
C
C
BB
B
B
B
B
A
A
A
A
A
A
Sample separated components
Chromatography is the chemical / physical separation of components for qualitative and
quantitative analysis
PrinciplePrinciple
6
7. Different chromatography methods
7
Name Mechanism Stat. Phase Mobile phase
Paper- partition liquid liquid
Thin layer- adsorption solid liquid
Gas- adsorption /
partition
solid or liquid gas
Column- adsorption /
partition
Solid/liquid liquid
8. Liquid Chromatography (LC)
• Two different phases are used to separate
components of the mixture: stationary and
mobile phase
• The solute separates on the column via
interactions based on physical mechanisms like
retention
8
9. Partitioning
• Separation is based on the analyte’s
relative solubility between two liquid
phases
9
Stationary PhaseMobile Phase
Solvent Bonded Phase
11. What is HPLC?
• The most widely used analytical separations technique
• Utilizes a liquid mobile phase & packed column to
separate components of mixture
• uses high pressure to push solvent through the column
• Popularity:
– sensitivity
– ready adaptability to accurate quantitative determination
– suitability for separating nonvolatile species or thermally
fragile ones
11
12. • Popularity:
– widespread applicability to substances that are of prime
interest to industry, to many fields of science, and to the
public
• Ideally suited for separation and identification of amino
acids, proteins, nucleic acids, hydrocarbons,
carbohydrates, pharmaceuticals, pesticides, pigments,
antibiotics, steroids, and a variety of other inorganic
substances
12
13. History
• Early LC carried out in glass columns
– diameters: 1-5 cm
– lengths: 50-500 cm
• Size of solid stationary phase
– diameters: 150-200 µm
• Flow rates still low ! Separation times long!
• Decrease particle size of packing causes increase in column
efficiency!
– diameters 3-10 µm
• This technology required sophisticated instruments
– new method called HPLC
13
14. Advantages of HPLC
• Higher resolution and speed of analysis
• Greater reproducibility due to close control of
the parameters affecting the efficiency of
separation
• Easy automation of instrument operation and
data analysis
• Adaptability to large-scale, preparative
procedures
14
15. Advantages to HPLC
• Advantages of HPLC are result of 2 major advances:
– stationary supports with very small particle sizes and large
surface areas
– appliance of high pressure to solvent flow
15
17. Parameters used in HPLC
• Retention parameters
• Column efficiency parameters
• Retention : When a component in a sample interacts
with the stationary phase in the column and a delay
in elution occurs
• Column efficiency : Goodness of a column
17
18. Column dead time, retention time
18
time
signal
t
t t
t
0
R 1 R 2
R 3
0
tR 1' tR 2' tR 3'
t´R = tR - t0
t0 = column dead time = time an unretarded compound needs to pass the
column
tR = retention time
20. Capacity Factor
• Capacity factor
20
k´ =
tR - t0
t
0
t´R
t
0
=
•If the substance is not retained by the stationary phase,
the capacity factor is k' = 0.
•Small k' (k < 1) values show that the components are
only retained slightly by the separation column. Their
peaks are located close to the non-retained peak (k' = 0).
•the optimum separation range to be k' values between 1
and 15. Values for k' > 5 mean long retention times with
associated band broadening.
21. Resolution
21
If the peaks are separated almost down to the baseline, R » 1.5.
Higher resolutions than R = 1.5 are not desirable because they
significantly extend the analysis time but do not result in additional
information.
Generally the values of R » 1.0 are sufficient to achieve qualitative or
quantitative results.Even values of R » 0.5 are sufficient to determine the
number of components present. For quantitative analysis, however, the
peak areas overlap too much
23. Plate NumberPlate Number
23
The theoretical plate number or N is a quantitative measure
for the column efficiency
In formulas (1) or (2) the peak base width WB or the half width
WH are compared with the retention time tR
24. Plate Number and Plate Height
Since an efficient separation column delivers sharp peaks with narrow base
widths, a better column has relatively high value of N
The concept of the theoretical plates is a useful tool to describe the efficiency of a
separation column
The number of theoretical plates is proportional to the column length L. The
longer a column, the more theoretical plates it has; however, the column back
pressure increases
To be able to compare separation columns of various lengths, the theoretical plate
height, H is used
24
H = L/N
26. Five modes in HPLC
LC mode Packing materials Mobile phase Interaction
Normal phase chromatography Silica gel n-Hexane/IPE Adsorption
Reversed phase chromatography Silica-C18(ODS) MeOH/Water Hydrophobic
Size exclusion chromatography Porous polymer THF Gel permeation
Ion exchange chromatography Ion exchange gel Buffer sol. Ion exchange
Affinity chromatography Packings with ligand Buffer sol. Affinity
26
29. Composition of HPLC System
• Solvent
• Solvent Delivery System (Pump)
• Injector
• Sample
• Column
• Detectors
• Waste Collector
• Recorder (Data Collection)
29
30. Mobile Phase Degassing
• Dissolved gases in the mobile phase can come out of
solution and form bubbles as the pressure changes
from the column entrance to the exit
– May block flow through the system
• Sparging is used to remove any dissolved gas from
the mobile phase
– An inert and virtually insoluble gas, such as helium, is
forced into the mobile phase solution and drives out any
dissolved gas.
• Degassing may also be achieved by filtering the
mobile phase under a vacuum
30
31. Are used to store Mobile-Phase. The solvent reservoir must be
made of inert material such as glass and must be smooth so as to
avoid growth of microorganisms on its walls. It can be transparent
or can be amber colored. A graduated bottle gives a rough
estimate of mobile-phase volume in the bottle. Solvent reservoirs
are placed above HPLC system (at higher level) in a tray. They
should never be kept directly above the system as any spillage of
solvent on the system may damage electronic parts of HPLC.
31
Solvent Reservoir
33. Mobile Phase Mixing
• Solvent proportioning valve(3.) can be
programmed to mix specific
amounts of solvent
from the various
reservoirs to
produce the
desired mobile
phase composition
33
3.
34. • Isocratic elution:
Use of a constant-composition mobile phase in liquid chromatography
• Gradient elution:
– Vary the mobile phase composition with time
– If there is a wide polarity range of components to be eluted.
– Allows for faster runs.
– Ex: mobile phase composition can be programmed to vary from
75% water: 25% acetonitrile at time zero to 25% water: 75%
acetonitrile at the end of the run.
• More polar components will tend to elute first.
• More non-polar components will elute later in the gradient
34
35. Common Reverse Phase Solvents
• Methanol
35
CH3OH
• Acetonitrile CH3CN
• Tetrahydrofuran
• Water H2O
37. In order to reduce separation time and allow the use of smaller particle size
packings (10 microns and below), we must force the liquid mobile phase
through the column under pressure. This is the function of the pump (also
called the "solvent delivery system") to maintain a constant flow of mobile
phase through the HPLC regardless of the pressure (back pressure) caused by
the flow resistance of the packed column.
There are several types of pumps are available,
Reciprocating Piston Pumps
Syringe Type Pumps
Constant Pressure Pumps
.
37
38. 38
Reciprocating Piston Pumps
Consist of a small motor driven piston which moves rapidly
back and forth in a hydraulic chamber that may vary from 35-
400 µL in volume.
On the back stroke, the separation column valve is closed, and
the piston pulls in solvent from the mobile phase reservoir. On
the forward stroke, the pump pushes solvent out to the column
from the reservoir.
A wide range of flow rates can be attained by altering the
piston stroke volume during each cycle, or by altering the stroke
frequency. Dual and triple head pumps consist of identical
piston-chamber units which operate at 180 or 120 degrees out
of phase
40. Syringe Type Pumps
The syringe pump is a large, electrically operated simulation of a
hypodermic syringe.
Are most suitable for small bore columns because this pump delivers
only a finite volume of mobile phase before it has to be refilled.
These pumps have a volume between 250 to 500 mL. The pump
operates by a motorized lead screw that delivers mobile phase to the
column at a constant rate. The rate of solvent delivery is controlled
by changing the voltage on the motor.
40
41. Constant Pressure Pumps
The mobile phase is driven through the column
with the use of pressure from a gas cylinder
A low-pressure gas source is needed to
generate high liquid pressures.
The valving arrangement allows the rapid refill
of the solvent chamber whose capacity is about
70 mL. This provides continuous mobile phase
flow rates.
41
42. Tubing
• Very small inner
diameter
• Consistent i.d.
• Very strong
• Easy to cut
• Fittings available
42
43. Auto samplers
Are fully automatic injection systems enabling greater productivity
and the highest level of precision.
The HPLC Autosampler incorporates an elegant swivel head
that allows a manual injection and a special Rheodyne injection
valve to give accurate full or partial loop fill injections.
43
44. Injection Port
The sample introduction device such as injector to introduce the sample in a
flow of mobile phase at high pressure.
It is not possible to use direct syringe injection on column like GC as the
inlet pressure in LC is too high.
The valve injection through fixed or variable loop is a common way of
introducing the sample.
The Rheodyne valve is the mostly used devise. The loop can be partially or
fully filled. There are both the types of injectors available.
The advantage of partial filling is the possibility of using small amount of
sample, when there is scarcity of sample.
The precision of the injection is 1% RSD
44
45. Samples are injected into the HPLC via an injection port. The injection port of an HPLC commonly
consists of an injection valve and the sample loop. The sample is typically dissolved in the mobile
phase before injection into the sample loop. The sample is then drawn into a syringe and injected into
the loop via the injection valve. A rotation of the valve rotor closes the valve and opens the loop in
order to inject the sample into the stream of the mobile phase. Loop volumes can range between 10
µl to over 500 µl. In modern HPLC systems, the sample injection is typically automated
45
47. Columns
• Solid Support - Backbone for bonded phases.
– Usually 10µ, 5µ or 3µ silica or polymeric particles.
• Bonded Phases - Functional groups firmly linked
(chemically bound) to the solid support.
– Extremely stable
– Reproducible
• Guard column - Protects the analytical column:
– Prolongs the life of the analytical column
47
• Analytical column - Performs the separation.
48. HPLC Columns
Particle size Column ID Sample Load
Analytical 3 – 5 µ 0.3 − 4.6 mm ng – µg
Semi-prep 10 µ 8 – 10 mm 1 – 100 mg
Preparative 10 – 30 µ 5 – 200 mm gram
scale
48
•An HPLC column consists of a stainless steel tube which is sealed with fittings
on both ends. Steel frits in the end fittings keep the packing material in the
column.
•Analytical columns have inner diameters of 1 - 10 mm and lengths of 25 - 250
mm. They are operated at flow rates of 60 µl - 5.0 ml/min.
•To protect the actual separation column from chemical contamination, a guard
column with the same packing material as the separation column is installed.
49. Column Packing
– Usually spherical silica particles of uniform
diameter (2-10µm)
• The smaller particles yield higher separation
efficiencies.
– The silica particles are very porous
• Allows for greater surface area for interactions
between the stationary phase and the analytes.
– Other packing materials may also be used:
• Zirconia (ZrO2) 49
http://hplc.chem.shu.edu/NEW/HP
LC_Book/Adsorbents/ads_part.html
50. Particle Diameter
• Has a greater effect on resolution than
column length
• Short columns with small particles ideal
• 5µm is standard size
• 3µm better, but restricted range of packings
available
• Downside is high back pressure and issues
with retention of small particles inside the
column, blockages
50
52. Refers to the solid support contained within the column over which the
mobile phase continuously flows.
The sample solution is injected into the mobile phase through the injector
port. As the sample solution flows with the mobile phase through the
stationary phase, the components of that solution will migrate according to
the non-covalent interactions of the compounds with the stationary phase.
The chemical interactions of the stationary phase and the sample with the
mobile phase, determines the degree of migration and separation of the
components contained in the sample.
For example, those samples which have stronger interactions with the
stationary phase than with the mobile phase will elute from the column less
quickly, and thus have a longer retention time, while the reverse is also true.
Columns containing various types of stationary phases are commercially
available.
Stationary Phases
52
53. Monofunctional surface modification of SiO2
OH
Si
O
Si
O
Si
O
SiO
OH
OH
OH
OH
X Si
CH3
CH3
CH2
CH2
CH2
CH2
CH2
CH2
CH2
CH2
CH2
CH2
CH2
CH2
C6H13
OH
Si
O
Si
O
Si
O
SiO
O
OH
O
OH
Si
CH3
C18H37
CH3
Si
CH3
C18H37
CH3
Si
CH3
CH3
CH3
OH
Si
O
Si
O
Si
O
SiO
O
O
O
O
Si
CH3
CH3
Si
CH3
CH3
Si
CH3
CH3
CH3
Endcapping
monofunctional
modification
Endcapping
Stationary Phases
53
54. Reversed Phase Chromatography
• Bonded phases made by covalently
bonding a molecule onto a solid
stationary phase like silica
• Typical stationary phases are
nonpolar hydrocarbons, waxy liquids
or bonded hydrocarbons (such as
C18, C8, C4, etc.)
• pH range 2.5 to 7.5
• 60-90% of all analytical LC
separations are done on bonded
phases in the reverse phase mode.
54
58. Reversed Phase ChromatographyReversed Phase Chromatography
modification: RP-8, RP-select B, RP-18, -CN, -Diol, -NH2
base material: LiChrosorb®
, LiChrospher®
, Superspher®
, Purospher®
,
Purospher®
STAR, Chromolith®
, Aluspher®
, Polyspher®
Normal Phase ChromatographyNormal Phase Chromatography
modification: Si, -Diol, -CN, -NH2
base material: LiChrosorb®
, LiChrospher®
, Superspher®
, Aluspher®
,Purospher®
STAR ,
Chromolith®
Ion Exchange ChromatographyIon Exchange Chromatography
modification: -NR+
3, -N+
HR2, NH2 for cations
-SO3
-
, -COO-
for anions
Base material: Polyspher®
, LiChrosil®
Size Exclusion ChromatographySize Exclusion Chromatography
Sorbents: LiChrogel®
PS for SEC in organic solvents
Fractogel®
for aqueous SEC
58
59. Detector
59
•Detect various compounds as they elute out from column. The
detector gives response in terms of a milivolt signal that is then
processed by the computer (integrator) to give a
chromatogram.
•Basically detector consists of a flow-cell through which the
mobile phase and resolved sample moves optics shine through
the detector cell and variation in optical properties are
detected.
•A Ultra violet or UV detector detects absorbance of UV light by
chromophores in the analyte compound.
•A refractive index detector will sense variation in refractive
index of mobile phase stream passing through flow-cell
• Similarly Fluorescence Detectors checks for Florescence.
60. Various Detetectors are Listed below
• Ultraviolet (UV)
Fixed wavelength detector
Variable wavelength detector
Diode Array
• Fluorescence
• Electrical Conductivity
• Refractive Index
• Electrochemical
• Light scattering
• IR Absorbance
• Mass-Spectrometric
60
68. • 2004: Further advances in column technology
and chromatography instrumentation
– Utilized even smaller packing particle sizes (1.7µm)
– Higher pressures (15000psi)
– Allowed for significant increases in LC speed,
reproducibility, and sensitivity.
• New research utilizing particle sizes as small as
1µm and pressures up to 100,000psi!
68
What is Ultra Performance
Liquid Chromatography?
70. Why UPLC more efficient
• Peak capacity (P) is the number of peaks that
can be resolved in a specific amount of time
• P is proportional to the inverse of the square
root of the Number of theoretical plates (N):
N = L/H
• Plate heights are correlated through the Van
Deemter equation
70
71. Contrasting HPLC and UPLC
• UPLC gives faster results with better
resolution
• UPLC uses less of valuable solvents like
acetonitrile which lowers cost
• The reduction of solvent use is more
environmentally friendly
• Increased productivity can increase you
revenue in an industrial setting
71
76. OVERVIEW
Detector Analyte /attributes sensitivity Nature
UV-Vis
Works with all molecules containing
chromophores absorbing UV-Vis
ng
Specific
PDA Works for all wavelengths in UV-Vis ng
Fluorescence
Compound with native fluorescence or
fluorescence tag
fg-pg
Conductivity
Anions, cations, organic acids,
surfactants
ng
Radio activity Radioactive labeled compounds ng-pg
76
77. Detector Analyte /attributes sensitivity Nature
Refractive Index
Temperature sensitive; polymers,
sugars, triglycerides, incompatible
with gradients
0.1-10 µg
Universal
Evaporative light
Scattering(ELSD)
Uniform response; nonvolatile to semi
volatile compounds; compatible to
gradients
ng
Electrochemical Redox reactions pg
Corona CAD
Can detect non UV absorbing
chromophores
ng
Mass Spectrometer Definite analyte identification fg-pg
Both universal and
specific
IR Works with all molecules mg
77
78. Uv- visible detector
Senstive
Wide linear range
Unaffected by changes in temp.
And mobile phase composition
Principle :
• Lambert-Beer’s law:
A= ε b c
At constant cell thickness and constant wavelength
a linear connection between the Absorbance A and
the Concentration C is achieved.78
79. Types of uv-visible detectors
A. Fixed wavelength detectors
B. Variable wavelength detectors
C. Photodiode array detectors
79
80. Fixed wavelength detectors
Operates only at 1 wavelength (254 nm)
Best overall precision is noted for peak
area measurements as compared to
variable wavelength detectors.
Uses a discrete source:
– Low pressure mercury lamp (253.6 nm)
– Silicone photodiode light detector
It can also be used at other wavelengths by
filtering the emission source to give other lines or
even using phosphor screens to give lines not
available from mercury.
80
81. Variable wavelength detectors
Uses continuous source of light
• Deuterium or xenon lamp or tugston lamp
• Desired wavelength is isolated by monochromator
Improved sensitivity from operating at λmax of solute of
interest
Drawback: limited lamp lifetime as compared to
mercury lamp (Deuterum lamp: 1000-2000 hrs)
Dual wavelength detection feature:
– Useful for simultaneous monitoring of two
substances
– Baseline noise is higher because this feature is
achieved by toggling the monochromator between 2
wavelengths 81
83. Photodiode array detectors
Uses charge coupled DA with 512 to 1024 diodes capable
of spectral resolution of 1 nm
Earlier, they suffered from sensitivity problems, which has
been solved by advanced flow cell design using fiber
optics technology to extend path length without
increasing noise or chromatographic band dispersion
Performs a simultaneous measurement of absorption as a
function of analysis time and over a chosen wavelength
range, so we get UV spectrum for each eluted peak
Used for method development where λmax of impurities in
drug are unknown
Every compound can be quantitated at its λmax , so useful
for trace analysis
Compounds not resolved chromatographically can
sometimes be resolved spectrally
Molecular absorption spectrum can be used for peak
identification and peak tracking
83
85. Indirect photometric detection
Used when solute of interest does not possess
chromophore
Here, mobile phase possess a chromophore and
absorbs light
When analyte without a significant
chromophore passes through the detector cell,
the absorption of mobile phase is decreased and
is recorded as negative peak
85
88. Features of modern uv-vis
detectors
Dual or multiple wavelength detection and stop
flow scanning features
Front panel access to self alligned sources and flow
cells for easy maintanance
Self validation features such as:
– Power up diagnostics
– Leak sensors
– Time logs for lamps
– Built in holmium oxide filters for wavelength calibration
– Filter wheels for linearity verification
88
89. Refractive index detectors
Measurement response is a function of the refraction
index difference between pure mobile phase and the
mobile phase with the dissolved separated
components
Due to the strong temperature dependence of the
refraction index, good temperature control of the
measurement cell must be ensured
The detection limit is in the range of 10-6
- 10-8
g/ml
To achieve a high sensitivity, in practice solvents are selected that have
a very high or very low refraction index
This ensures that the difference between the refractive indices of
89
91. One half of the measurement cell is purged with the flowing mobile phase, the second
half of the measurement cell (reference cell) is filled with mobile phase. The refraction
index n is at first identical in both cells.
If a sample component is added to the eluent, the refraction index n in the
measurement cells changes. The light beam experiences a deflection on the path
Operating principle
91
93. Fluorescence detectors
During fluorescence detection,
the sample is irradiated with
UV light of suitable wavelength
and excited for emission of long
wave light.
The Xenon lamp continuously sends light of 325 -
410 nm.
Provides LOD values 100 times lower than
absorption detectors
Highly selective
93
98. Electrochemical detectors
Electron transfer processes offer highly selective and
sensitive method
Easily adaptable for use with microcolumns
As background noise is dependent on mobile phase
conditions, it is difficult to utilize these detectors
with gradient elution separations
2 types:
1. Amperometric detection: fixed potential is applied to
the electrode (glassy carbon) and a solute which will
oxidize at that potential yields an output current
2. Coulometric detection: 100% of the solute species is
converted, which offers advantages of no mobile phase
flow dependence on signal and absolute quantitation
through Faraday’s law
98
100. Conductivity detectors
Measurement of electrical conductance is a subset of
ECD, although it is generally considered separately
since it is non-Faradaic electrochemistry; i.e. no
electron transfer reaction takes place
The electrical conductivity of the mobile phase is used
as characteristic for detection.
Universal technique for ionic solutes and are used
mainly for the separation of ions in water or polar
eluents
Since the conductivity can change approx. 2 % per °C,
conductivity detector are often equipped with
automatic temperature compensation
100
101. With the removal of electrolytes contained in the eluent system
when using ion exchangers, a significant increase in the sensitivity
can be achieved (otherwise it creates difficulty in measuring low
concentration of an ionic solute in the presence of highly
conducting mobile phase
The resulting extremely low basic conductivity makes it possible, to
generate good measurement signals from the smallest sample
volumes.
During conductivity detection the specific conductivity is measured
continuously.
The individual eluted components of the material sample are each
displayed through a change in conductivity.
101
Conductivity detectors
102. Principle of 4 electrode method
The two outer electrodes are used to transmit an
electronically stable alternating current through the cell
The electro-motor force (E), which is measured
deenergized on the two inner electrodes, is a measure for
the conductivity of the cell content according to the
equation.
χ = (K/E)*I
χ = electric conductivity
K = cell constant between the 2nd and 3rd electrode
E = electro-motor force
I = current strength
The deenergized measurement of the electro-motor force
at the two inner electrodes caused no passivity, hence the
instrument operates stably for a long period of time.
102
105. ELSD can outperform traditional detectors when analysing non-
chromophoric samples by HPLC
Traditional HPLC detectors such as UV and RI have limited capabilities
UV and RI are not compatible with a wide range of solvents
RI detection is not gradient compatible
Different analytes produce different UV responses depending on their
extinction co-efficient
ELSDs can detect anything that is less volatile than the mobile phase
ELSD is universal and compatible with a wide range of solvents
Evaporative Light Scattering Detector
105
106. Unique Method of Detection
Three steps:
• Nebulization
• Mobile Phase Evaporation
• Detection
106
107. Step 1: Nebulization
Column effluent passes through nebulizer needle
Mixes with nitrogen gas
Forms dispersion of droplet
107
108. Step 2: Mobile Phase Evaporation
Droplets pass through a heated zone
Mobile phase evaporates from the sample particle
Dried sample particles remain
108
109. Step 3: Detection
Sample particles pass through an optical cell
Sample particles interrupt laser beam and
scatter light
Photodiode detects the scattered light
109
110. ELSD – A Powerful Detector for HPLC
Universal
Sensitive
Gradient Compatible
AN ELSD IS AN EFFECTIVE REPLACEMENT OR A PERFECT COMPLEMENT TO EXISTING
LC DETECTORS
RI
UV
MS
Fluorescence
110
111. Four Reasons to Replace an RI with an
ELSD:
1. Better sensitivity
2. Gradient compatible
3. Stable baselines
4. No solvent front peaks
111
112. The ELSD Improves Baseline Stability and Detection Sensitivity Compared
to RI
112
113. Reasons to operate an ELS detector in series with a
UV detector:
1. Obtain a more accurate representation of
sample mass than UV
2. See what may be missing from your UV
chromatogram
113
114. The ELSD responds to all components in this antihistamine
formulation
1. Diphenhydramine
2. PEG, Gelatin
3. Sorbitan
4. Glycerol
5. Sorbitol
114
115. Detect Difficult Samples like Triglycerides without Derivatizing
1. LLO
2. LLP
3. OOL
4. POL
5. PPL
6. OOO
7. OOP
8. PPO
9. OOS
115
116. Although the Mass Spectrometer is a Universal
Detector, ELSDs offer many Benefits over MS:
1. Lower investment and operating costs
2. Less complicated operation
3. Less maintenance
Since chromatographic requirements for ELSD
and MS are similar, methods developed for
ELSD are usually transferable to MS without
Use ELSD in Parallel with MS to Obtain Maximum Structural and Concentration
Information
Use ELSD in Parallel with MS to Obtain Maximum Structural and Concentration
Information
116
118. Summary:
Benefits of ELSD as a Replacement or
Complement to Existing Detectors
• See what you might be missing
• More sensitive and stable than RI
• Get a more accurate representation of sample
mass than UV
• Simplify methods by eliminating derivatization
118
127. 127
APPLICATIONS OF HPLC :
Assay of cephalosporins
Assay of theophylline
Separation of barbiturates, phenothiazines, benzodiazepine
derivatives, rauwolfia alkaloids etc. by C-18 reversed phases
Quantitative analysis of several analgesics like aspirin, caffeine,
paracetamol, phenacetin, etc.
Analysis of urine and serum samples
Separation of antipyrine and benzocaine in ear drops
It is common practice to make up these organic solvents as mixtures with water or, in a lot of cases, have each pure solvent mixed under instrument control and changed at a certain rate with time (gradient). Gradients can be simple or complex. Simple - as a linear gradient (ramp); Complex as steps (start and hold) and ramps together. You can have a solvent or several solvents being controlled at the same time with a changing modifier such as a pH buffer. Methanol - Most common solvent. Close to water in structure. Miscible in all proportions with H 2 O so that for less polar organics you can have the power of 90% Methanol with 10% H 2 O. Acetonitrile - Highly polar, very low UV absorbance. Also completely miscible with H 2 O but lacking in hydrogen bonding capability thus affording a different partitioning effect. Tetrahydrofuran - Molecule has high dipole moment. More soluble with non-polar compounds. Water - Also a very common solvent. Used to make up solvent modifiers to adjust pH (buffers) as well as ion-pairing reagents. Emphasize degassing (“bends”- bubble in detector) and for particle-free (dust could be up to 10X size of particles of solid support). 555- 25mM H 3 PO 4 :MeCN(90:10)->10:90 605 - 0.1M pH4.7 OAc: MeCN (1:1) isocratic. 610 - MeCN:H 2 O -> MeCN 8316 - 100% H 2 O 8330 - MeOH:H 2 O isocratic 8331 - MeOH:H 2 O(HOAc-C 10 SO 3 H) 8332- MeOH:H 2 O (3:2-CN), (1:1-C-18) 8325-A=MeCN:0.01M OAc (75:25), B= MeCN -> 60%MeCN total. λ= 190 for MeCN, λ= 205 for MeOH, λ= 190 for H 2 O, λ~ 290 for THF, λ~ 255 for 1% HOAc and λ~ 260 NH 4 OAc (1M)
Here you can see the structure of the different packing materials of Merck. Starting from the irregular particles (e.g. LiChrosorb) there came the invention of spherical particles, used for the LiChrospher, Superspher, Purospher and Purospher STAR sorbents) and especially the latest innovation by Merck, the monolithic column technology in terms of the Chromolith column. On the different pictures you can compare the macropores (through-pores) and the mesopores of the different variations of these packing materials.