Echo planar imaging (EPI) is the method of rapid magnetic resonance imaging (MRI), overcoming one of the significant disadvantage of MRI concerning with slow imaging time. However, EPI-MRI imaging comes with it's own unique imaging artifacts.
3. Introduction
Echo-planar imaging (EPI) is capable of significantly shortening magnetic resonance (MR) imaging
times.
Echo-planar imaging allows acquisition of images in 20–100 msec.
This time resolution virtually eliminates motion-related artifacts.
Therefore, imaging of rapidly changing physiologic processes becomes possible.
Echo-planar images with resolution and contrast similar to those of conventional MR images can be
obtained by using mult-ishot acquisitions in only a few seconds.
3
4. K space
Before discussing the basic principles of echo planar imaging or any other
MR imaging technique, it is important to present the concept of k space.
k-space is the matrix containing the raw data from all the measurement
before the Fourier Transformation.
Rectangular in shape and has two axes perpendicular to each other (kx and
ky).
Each number in k-space refers to a spatial frequency.
MR images are obtained by performing FT of k-space.
4
The wavenumber (k) is
the number of waves
or cycles per unit
distance.
6. K space
The individual points (kx,ky) in k-space do not correspond one-
to-one with individual pixels (x,y) in the image. Each k-space
point contains spatial frequency and phase information about
every pixel in the final image. Conversely, each pixel in the
image maps to every point in k-space
6
7. K space
Facts about k-space-
The center of k-space (i.e., low spatial frequencies) contains
information on large-scale structures (e.g., contrast between
large objects).
The periphery of k-space (i.e., high spatial frequencies)
contains information on the fine structures (e.g., edges and
small-scale details).
Positive polarity phase fills the top half of K space;
negative polarity fills the bottom half
Steep gradients, (both positive and negative), select the
most outer lines and shallow gradients select the center
lines
Provided no phase errors occur during data collection, k-
space possesses a peculiar mirrored property known
as conjugate (or Hermitian) symmetry.Data are
symmetrical in K space.
7
8. 8
As an overview, the
center of k-space
contains low spatial
frequency information,
determining overall image
contrast, brightness, and
general shapes. The
periphery of k-space
contains high spatial
frequency information
(edges, details, sharp
transitions).
9. 9
As an overview, the
center of k-space
contains low spatial
frequency information,
determining overall image
contrast, brightness, and
general shapes. The
periphery of k-space
contains high spatial
frequency information
(edges, details, sharp
transitions).
10. K space
10
Phase-conjugate symmetry.
About half of k-space is sampled
by reducing the number of
phase-encoding steps. The other
half of k-space is
synthesized/reconstructed.
11. K space
11
Although phase-conjugate symmetry reduces imaging time
while preserving spatial resolution, this is accomplished at
the expense of the signal-to-noise ratio (SNR). For half-
Fourier imaging, SNR is reduced by a factor of √½ or 30% less
than a comparable sequence using the full number of phase-
encoding steps.
In theory, phase-conjugate symmetry allows one to acquire
data using only half the normal number of phase-encoding
steps, thus potentially reducing imaging time by as much as
50%. In practice, the time savings is closer to 40%, but this is
still a huge benefit widely used in modern MRI protocols.
12. K space trajectories
The main k space trajectories are:
Standard rectilinear/Non-EPI trajectory
EPI trajectory
Radial trajectory
Spiral trajectory
12
Rectilinear
16. Types of EPI trajectory
EP trajectories sweep back and forth across
k-space very rapidly such that the entire k-
space matrix, or a large portion of it, is
filled in one execution of the pulse
sequence.
If the entire k-space matrix is filled in one
execution of the pulse sequence, then it is
termed a single-shot acquisition.
If instead more than one execution of the
pulse sequence is required to collect all
the desired k-space data, then the scan is
a multi-shot acquisition and scan time is
TR times the number of shots (executions).
18
17. Types of EPI trajectory
Single-shot:
Traverse all of k-space, covering a rectilinear grid, starting in one corner, sweeping across
the kx axis for a given ky value, then jumping (“blipping”) to the next ky value and sweeping
back across the kx axis.
Another blip to the next ky value follows and so on until all of k-space is traversed.
For applications requiring very fast acquisitions.
Especially used for functional MRI (fMRI) studies due to high speed and sensitivity to BOLD
effect. Distortions and T2* dependent signal loss typically prohibit the acquisition of a high
resolution image.
19
18. Types of EPI trajectory
Multishot-EPI:
Sparsely traverse all of k-space, starting in one corner, sweeping across the kx axis for a
given ky value, then jumping (“blipping”) to a different ky value and sweeping back across
the kx axis.
Another blip to another ky value follows and so on until a fraction of k-space equal to
1/(#shots) is filled.
For applications requiring very fast acquisitions but not as fast as single shot EPI.
May also be used to fill a larger k-space matrix and thus obtain better spatial resolution
than possible with single shot EPI. Since time spent acquiring data is shorter than with
single shot EPI, distortions and signal loss are less severe
20
19. Basic principle
In echo-planar imaging, multiple lines of imaging data are acquired after a single RF excitation.
Like a conventional SE sequence, an SE echo planar imaging sequence begins with 90° and 180° RF
pulses.
However, after the 180° RF pulse, the frequency-encoding gradient oscillates rapidly from a
positive to a negative amplitude, forming a train of gradient echoes .
Each echo is phase encoded differently by phase-encoding blips on the phase-encoding axis.
Each oscillation of the frequency-encoding gradient corresponds to one line of imaging data in k
space, and each blip corresponds to a transition from one line to the next in k space.
This technique is called blipped echo-planar imaging
21
20. 22
Conventional SE imaging. Within each TR period, the pulse sequence
is executed and one line of imaging data or one phase-encoding step
is collected
22. EPI data acquisition
As with conventional MR- imaging, we put the acquired data for the frequency and phase encoding into
the 2D grid called k-space.
Also, the 2D Fourier transform is used to create the image.
In EPI, the data is filled into k-space in a rectangular “zig-zag”-like pattern.
23. Hardware
The requirements for gradient strength, rise time, and duty cycle are markedly increased because
all of k space should be traversed in a single RF excitation by using a rapidly oscillating frequency-
encoding gradient
Echo-planar imaging is accomplished by using gradient coils capable of a maximum amplitude of 20
mT/m, a minimum rise time of 0.1 msec, a slew rate of 200 T/m per second, and a duty cycle of
50%–60%
Higher bandwidth ADC in the order of MHz, upto 2 MHz ( in conventional sequence- kHz)
25. SNR
The SNR in EPI is inherently low due to wide bandwidth.
For same parameter ,SNR is lower in EPI than in conventional MRI.
Can be increased by Use of surface coils ,multi-shot imaging
27
26. Contrast in EPI
Different contrasts are achieved by either beginning the sequence by variable RF excitation pulse in GE–
EPI or with 90° and 180° RF pulses in SE– EPI with any type of RF pulse.
EPI– FLAIR (180° / 90° / 180° followed by EPI readout) nulls the CSF but the sequence is significantly
faster than in conventional FLAIR sequencing.
Proton density or T2 weighting is achieved by selecting either a short or long effective TE.
T1 weighting is possible by applying an inverting pulse before the excitation pulse to produce saturation.
Contrast in SE EPI is similar to that of CSE.
GRE EPI provides T2* weighting.
28
27. Resolution
Resolution depends on the maximum gradient amplitude-time product in the raw data.
To increase resolution requires:
increased gradient amplitude which may ultimately result in unacceptable safety problems)
increase in gradient duration (lowers the effective image bandwidth and increases the sensitivity
of the images to shape distortion and other artifacts) or both.
Higher bandwidth ADC in the order of MHz, upto 2 MHz ( in conventional sequence- kHz)
29
28. Resolution
EPI resolution is limited by the SNR of the images.
Since typical EPI data are collected over 40 to 50 msec , as compared to the nearly 1 second of
acquisition time spent on a conventional scan, the SNR is down by a factor of more than 4-fold at
comparable resolution on this basis alone.
With EPI at 3.0 T , spatial resolution can be improved tremendously up to 0.75 X 0.75 mm for a 2.5-
mm slice thickness; thus susceptibility artifacts at the bone-air interface do not constitute a problem.
At the present time, it seems that EPI spatial resolution is largely gradient limited
30
29. Types of epi
Types of EPI:
Single shot EPI
Multishot EPI
Preparation of EPI
Spin echo EPI
Gradient echo EPI
Hybrid EPI
Inversion recovery EPI
DW EPI
3D EPI
31
30. Single shot EPI
Collects all the data required to fill lines of k-space from a single echo train.
Multiple echoes are generated, and each is phase encoded by a different slope of gradient to fill all
the required lines of k-space in a single-shot.
Rapid change in gradient polarity rephases the FID produced after the RF excitation pulse to generate
gradient-echoes.
As the frequency-encoding gradient switches its polarity so rapidly, it is said to oscillate.
The phase-encoding gradient also switches on and off rapidly, but its polarity does not need to change
in this type of k-space traversal.
No TR in single-shot imaging .TR is therefore infinity. 32
31. Multishot EPI
The SS-EPI’s main advantage is reduced scan-time, it suffers from several limitations including low
signal-to-noise ratio (SNR), vulnerability to image distortions (or artifacts), and limited attainable
spatial-resolution due to the fast signal decay .
In order to overcome these limitations, multi-shot EPI (ms-EPI) can be used.
Multi-shot EPI follows the same principles as ss-EPI except it uses multiple excitations to acquire k-
space data. One way to acquire the data is in an interleaved manner, where the acquired lines of each
shot fill in the gaps from the other shots.
As an example, assuming 64 lines of k-space were desired, in two-shot EPI only 32 lines would be
acquired for each excitation. In an interleaved acquisition the first shot would acquire lines 1,3,5,7,
etc. while the second shot would acquire lines 2,4,6,8, etc. 33
32. MS- EPI
Each shot traverses the y-direction of k-space much faster than in ss-EPI, and this results in reduced
distortions that occur in ss-EPI due to its slow ky sampling rate.
One of ms-EPI’s main advantages is its ability to acquire a much greater number of k-space lines. In ss-
EPI the total number of attainable lines is limited because of the T2∗ signal decay of a single excitation.
In ms-EPI the number of lines in each shot is also limited by T2∗ decay, but by combining data from each
shot the total amount of lines is increased, and is limited only by the number of excitations used.
This results in a larger k-space coverage and therefore improved spatial resolution
In ms-EPI the number of k-space lines acquired in each shot is not stretched to the maximum as in ss-
EPI, the acquired signal from each shot does not decay as much as much as in ss-EPI, resulting in better
SNR.
The number of shots commonly used in practice are currently two, four, and eight, with two/four-shot
EPI being the most typical.
34
36. DWI -EPI
39
Diffusion weighted imaging. Two
diffusion gradient are place on either
side of 180 refocusing pulse. These
gradients have certain b value which
depend the strength the duration of
the gradient and the time between
two gradient lobes .the echo
generated are readout in epi
37. VOLUME-EPI
Idea of Echo Volume Imaging (EVI) is a direct extension of EPI to 3D imaging using a single RF
excitation.
Incorporates the oscillatory EP readout along the frequency-encoding and phase-encoding axes.
For the slice-select phase encoding a constant gradient is applied during the complete readout
(or small blips) so that the complete 3D data is encoded.
40
39. EPI ARTIFACTS
There are four major artifacts in EPI:
T2*weighting signal loss
Ghosting
Geometric distortion & Susceptibility artifacts
Fat water chemical shift
42
40. Geometric distortion
Geometric distortions manifest themselves as pixel movements predominantly in the phase
encoding direction in EPI images.
Appearance of stretching or bunching of signal near in the vicinity of field inhomogeneity.
The spatial distortions are greatly significant in the phase-encoding direction.
The pixels are therefore shifted in the direction in the phase encoding gradient
43
43. How to reduce distortion?
Attempts to reduce the total acquisition time of EPI so the result is less affected by magnetic disturbance
and generate images that have less artifact:
Partial Fourier Reconstruction
Parallel imaging
Segmented EPI
Multi-shot
46
44. Partial fourier
Partial echo:
Is performed when only part of echo is read during application of frequency encoding gradient.
The peak of echo is usually centered in the middle of sampling window.
When a very short TE is required, the echo rephases sooner than with a long TE.
This technique switches on the frequency encoding gradient as soon as it is possible to do so, but also
moves the peak of echo so that it is no longer at the centred in the middle of the sampling window
and occurs sooner.
This means that only the peak and dephasing part of echo are sampled and therefore initially only
half of the frequency area of k space is filled (right side of k space)
47
45. Partial fourier
Partial echo:
However due to right –to-left symmetry of k space ,the system can extrapolate the data in the right
side and place it in the left side.
Therefore although initially only the right side of the k space is filled with data, after extrapolation
both sides contain data and overall no data are lost.
Partial echo imaging is routinely used when a very short TE is selected in the scan protocol.
48
46. Partial fourier
Partial averaging:
The negative and positive lobes of k-space on each side of the phase axis are symmetrical and a mirror
image of ach other.
As long as over the half of the lines are filled during the acquisition ,the system has enough data to
create an image .
Scan time decreases as a result distortion also decreases.
Partial averaging is used where a reduction in scan time is necessary and where the resultant signal loss
is not of paramount importance.
49
49. Parallel imaging
Distortion in EPI is driven by the temporal separation between subsequent odd and even echoes
within the EPI readout train.
One way to reduce distortion in EPI is, therefore, through the use of parallel imaging with either
generalized auto calibrating partial parallel acquisition (GRAPPA) or sensitivity-encoding (SENSE) .
GRAPPA and SENSE speed up the traversal of k-space
52
51. How to reduce distortion?
however, image distortion in parallel imaging–enhanced EPI DWI remains problematic particularly at
high field strengths and for high-resolution imaging
Another way to reduce distortion in DWI is to speed k-space traversal through the use of alternate
trajectories.
These methods include but are not limited to:
Multishot EPI sequences
interleaved spiral sequences
and fast spin-echo (FSE)–based sequences such as periodically rotated overlapping parallel lines with
enhanced reconstruction (PROPELLER)
short-axis PROPELLER EPI
readout-segmented EPI
54
54. Propeller EPI
Traverses k-space radially in groups parallel and centered about the k(0,0) point.
The groups, also known as “blades” are rotated about the k(0,0)point of k-space like unto a
rotating propeller.
Advantages include a more rapid coverage of the blade in k-space for a scan duration similar to
EPI
speeds and a high coverage of the outer portions of k-space.
The technique maintains the robustness of radial sampling to bulk motion of the object but
without the associated blurring due to reduced higher k-space acquisition.
57
57. T2* signal loss
The transverse magnetization is T2* dependent.
The effect of field inhomogeneities is dependent upon the size of the phase dispersion caused
relative to the size of the pixel.
Intra-voxel dephasing is where field inhomogeneities occur within a voxel, resulting in signal loss
due to phase dispersion.
More prevalent in GE-EPI, as the 180 degree pulse in SE-EPI acts to reverse phase dispersions and
reduce dependence on T2*.
62
59. Chemical shift
Arises from the inherent 3.5 ppm frequency difference between fat and water protons under the
influence of an external magnetic field.
Chemical shift artifact causes the misplacement of signal from fat in the image which is a
misregistration of fat and water protons from a voxel that are mapped to different pixels.
It can also create both signal voids and signal superimposition (high signal) in areas where fat and
water interface.
Because the readout BW is so high, in EPI, chemical shift is not manifested in frequency readout
direction but in phase encoding direction if present.
64
62. How to reduce chemical shift?
Apply saturation band/Fat saturation
Decrease echo spacing
Under sample
69
63. N/2 GHOSTING
Images appear displaced by half the field-of-view (FOV) in the phase encoding direction, with the
bottom half of the (real) image generating a ghost image that appears at the top of the FOV and
the top half of the (real) image generating a ghost image that appears at the bottom of the FOV.
The readout gradients for odd and even echoes are of different polarity (one positive, the other,
negative).
While in MRI, any magnetic disturbance (such as inhomogeneity, eddy current, susceptibility,
gradient imperfection) will cause the center of the echo to shift and results in the mismatch of
centers of odd and even echoes.
It is due to this mismatch that the N/2 ghost appears.
70
64. 71
Simplified representation of signal collected in: a)
conventional sequences, b) EPI. One can notice that
center of echoes from odd and even echoes appear to be
misaligned in EPI. A typical N/2 ghost manifestation
69. Brain imaging
Brain Imaging
EPI–based diffusion imaging is routinely used for evaluation of early cerebral ischemia and
stroke
EPI–based perfusion imaging and DWI EPI is performed to evaluate cerebral ischemia and
differentiate different types of tumor from radiation necrosis
Evaluation of cortical activation with echo-planar imaging or functional MR imaging is an
active area of research in neuroscience
77
70. 78
Large right middle cerebral artery infarct in a 61-year-old woman. The infarct
was seen on a cerebral angiogram; MR imaging was performed approximately 5
hours after onset of the stroke. Axial T1- weighted (a) and T2-weighted (b)
images show only mild sulcal effacement and gyral edema on the right side of
the brain (arrow). Axial diffusion-weighted (c) and perfusion (d) images show
the extent of the infarct clearly
71. Abdominal imaging
The primary contribution of echo-planar imaging to abdominal imaging has been the increased
gradient capabilities of echo-planar imaging
the result has been shorter TEs and improved GRE sequences for breath-hold imaging and three-
dimensional MR angiography.
There are several potential uses for echo-planar imaging in abdominal imaging.
Multishot echo-planar imaging can be used to acquire breath-hold T2- weighted images, potentially
replacing SE and fast SE T2-weighted sequences.
In the uncooperative patient, single-shot echo-planar imaging of the upper abdomen can be
performed in about 2 seconds.
79
72. Abdominal imaging
Single-shot techniques, such as half-Fourier acquisition single-shot fast SE (HASTE) imaging or
single-shot fast SE (SSFSE) imaging, can be used for breath-hold imaging but are limited by image
blurring and suboptimal contrast resolution.
the echo-planar imaging pulse sequence can potentially combine the tissue contrast advantages of
conventional SE sequences and the speed of single-shot fast SE sequences
80
73. 81
Colon cancer metastatic to the liver. (a) Axial fast SE image shows one faint lesion
(arrow). (b) Axial single-shot fast SE image also shows only one lesion (arrow). (c)
Axial echo-planar image shows many lesions (arrows).
74. Cine Imaging
Echo-planar imaging facilitates rapid evaluation of cardiac function and anatomy.
Cine imaging of the heart is performed with GRE echo-planar imaging over multiple cardiac
cycles.
When singleshot echo-planar imaging is used, electrocardiographic gating is not required.
Therefore, this technique is particularly helpful in patients with arrhythmias.
82
75. Cine imaging
Cardiac perfusion imaging involves first-pass imaging after administration of a contrast
material bolus and is used for evaluation of myocardial ischemia.
Echo-planar imaging offers sufficient spatial and temporal resolution to distinguish
perfusion patterns
83
76. Three-dimensional breath-hold imaging of the coronary arteries with
segmented echo-planar imaging. Double oblique image (left image)
and axial images (all other images) show the right coronary artery
(RCA) and the left anterior descending artery (LAD)
Maximum-intensity projection
image
from breath-hold gadolinium-
enhanced three-dimensional MR
angiography. An ultrashort TR/TE
of 3.8/ 1.4 was used, with only 18
seconds of breath holding
78. Miscelleneous
Echo-planar imaging has potential value as an ultrafast method for MR angiography and
flow quantification .
Echo-planar imaging freezes fetal motion and has been used to identify fetal anomalies .
Echo-planar imaging can also be used to guide minimally invasive surgery performed with
focused ultrasound or a thermal laser .
Diffusion-weighted imaging allows real-time temperature monitoring during tissue
coagulation or freezing performed for tumor ablation.
Kinematic studies of joint motion can be performed with echo-planar imaging
possible to perform MR fluoroscopy (real-time MR imaging) and echo volume imaging,
which captures three-dimensional snapshot images in 80–100 msec.
86
80. Reference
Rad229 (2020) Lecture-10A: Echo Planar Imaging [Internet]. Youtube.com. 2021 [cited 25 October 2021]. Available from:
https://www.youtube.com/watch?v=lc5NoHS59Hk&list=PLscv4NA0bNSOWYDILwQPuXURMVkSaGuEU&index=29
Rad229 (2020) Lecture-10B: Echo Planar Imaging Artifacts [Internet]. Youtube.com. 2021 [cited 25 October 2021]. Available from:
https://www.youtube.com/watch?v=x17JmML0R40&list=PLscv4NA0bNSOWYDILwQPuXURMVkSaGuEU&index=30
Rad229 (2020) Lecture-10C: Advanced Echo Planar Imaging [Internet]. Youtube.com. 2021 [cited 25 October 2021]. Available from:
https://www.youtube.com/watch?v=mEEspC-V1l4&list=PLscv4NA0bNSOWYDILwQPuXURMVkSaGuEU&index=31
Echo-planar imaging (EPI) [Internet]. Questions and Answers in MRI. 2021 [cited 25 October 2021]. Available from:
http://mriquestions.com/echo-planar-imaging.html
Poustchi-Amin M, Mirowitz S, Brown J, McKinstry R, Li T. Principles and Applications of Echo-planar Imaging: A Review for the General
Radiologist. RadioGraphics. 2001;21(3):767-779.
Cohen M. Echo-planar imaging (EPI) and functional MRI [Internet]. Academia.edu. 2021 [cited 25 October 2021]. Available from:
http://www.academia.edu/2660043/Echo-planar_imaging_EPI_and_functional_MRI
Schmitt F, Stehling M, Turner R. Echo-Planar Imaging.
Paschal C, Morris H. K-space in the clinic. Journal of Magnetic Resonance Imaging. 2004;19(2):145-159.
MRI sequences - Echo planar (EPI) [Internet]. IMAIOS. 2021 [cited 25 October 2021]. Available from: https://www.imaios.com/en/e-
Courses/e-MRI/MRI-Sequences/echo-planar-imaging
(99+) Physics in Clinical MRI Series 7.5: Echo Planar Imaging (EPI) | LinkedIn 88
c) K space is filled in a linear fashion from top to bottom or bottom to top
c) K space is filled in a linear fashion from top to bottom or bottom to top
c) K space is filled in a linear fashion from top to bottom or bottom to top
(non-steady state imaging
Data from more than one excitation, each with a unique k-space trajectory are interleaved and then Fourier transformed to make one image.
Minimum gradient requirements for EPI sequences-
: maximum amplitude of 20 mT/m,
: minimum rise time of 0.1 msec,
: slew rate of 200 T/m per second,
: duty cycle of 50%–60%
Fig showing the spin echo with eppi readout.rf exciation 90 ulse followed by 180 refocusing pulse followed by rain of echo .hee echos are read out by epi
Shows gradient echo with epi readot.the ssexcitation pulse of certain flip angle is followed by gradient refosing pulseswith epi readout.
Inversion recovery with epi readut.180 inversion pulse followed by 90 excitation pules followed by 180 refocusing pulse .te echo is readout in epi
The blipped slice-select phase-encoding direction is possible, but the inaccuracy in applying very short blips of high amplitude can result in ringing and crosstalk between the reconstructed slices.
however, image distortion in parallel imaging–enhanced EPI DWI remains problematic particularly at high field strengths and for high-resolution imaging