Diese Präsentation wurde erfolgreich gemeldet.
Die SlideShare-Präsentation wird heruntergeladen. ×
Anzeige
Anzeige
Anzeige
Anzeige
Anzeige
Anzeige
Anzeige
Anzeige
Anzeige
Anzeige
Anzeige
Anzeige
Nächste SlideShare
ECG
ECG
Wird geladen in …3
×

Hier ansehen

1 von 12 Anzeige

Weitere Verwandte Inhalte

Diashows für Sie (20)

Ähnlich wie Pulse detector (20)

Anzeige

Aktuellste (20)

Pulse detector

  1. 1. A Minor Project Report On “REAL TIME PULSE DETECTOR” by: SHATRUGHAN PATEL DEPARTMENT OF ELECTRICAL & ELECTRONICS ENGINEERING BRANCH: ELECTRONICS & COMMUNICATIONS
  2. 2. INTRODUCATION • The heart is the organ that responsible for pumping blood throughout the body. It is located in the middle of the thorax, slightly offset to the left and surrounded by the lungs basically; the human heart is composed of four chambers which are two atriums and two ventricles. The right atrium receives blood returning to the heart from the whole body. That blood passes through the right ventricle and is pumped to the lungs where it is oxygenated and goes back to the heart through the left atrium, and then the blood passes through the left ventricle and is pumped again to be distributed to the entire body through the arteries
  3. 3. MATERIAL AND TOOLS • The fingertip sensor - The heart rate monitor builds to test the value of users’ heart rate currently. The circuit will base on the non-invasive PPG sensor which detects the variation of the blood flow in finger when the mechanical contraction of heart. The sensor contains an infrared LED as an IR transmitter and photodiode as an IR receiver. • ATMEGA32 Microcontroller - The high-performance, low-power Atmel 8-bit AVR RISC-based microcontroller combines 16KB of programmable flash memory, 2KB SRAM, 1KB EEPROM, an 8channel 10-bit A/D converter, 2 capture/compare/PWM functions, 3 timers, a synchronous serial port that can be configured as either 3-wire SPI or 2-wire I2C bus, a USART, and a Parallel Slave Port. The device supports throughput of 16 MIPS at 16 MHz and operates between 4.5-5.5 volts
  4. 4. ATMEGA32 Microcontroller PIN DIAGRAM -
  5. 5. Fingertip sensor - Use of light to measure heart rate is a field of study where abundant research has been done in the past few decades. Fingertip sensor relies on measurement of a physiological signal called Photoplethysmogrpahy (PPG) [14.], which is an optical measurement of the change in blood volume in the arteries. Fingertip sensor acquires PPG signals by irradiating wavelength of light through the tissue, and compares the light absorption characteristics of blood under these wavelengths High level design - By high level design we mean to see the project from an overall view point, then to get down and consider the project as a system of main parts. Fingertip sensor is used in the device as transducer converts the physical PPG signal to into an electrical signal. This electrical signal is applied to the microcontroller which is the main brain of the heart rate monitor. The microcontroller, according to the
  6. 6. CIRCUIT DIAGRAM
  7. 7. • This project used LM358 two-class amplifier to adjust the voltage value to 5V for being received by ATMEGA32. Because the signal acquisition is extremely weak and vulnerable to interference by the external circuit, high-pass filter and low-pass filter are used for normal heart beat signals. • The amplifier (see figure 4.9) uses an LM358 dual op amp to provide two identical broadly-tuned low pass stages with the gain of each filter stage is set to 101 and a cut-off frequency of about 2.34 Hz. So the maximum HR that could be read is approximately equal 150. The equations for calculating gain and cut-off frequency of the active low pass filter are expressed as
  8. 8. LCD • In this project, a 16×2 LCD display will be used for displaying the value of heart rate. There are 16 pins in whole in LCD. The pin diagram of LCD is shown as below. From the datasheet, it is known that pin 1 & pin 2 and pin 15 & pin 16 both functions as power supply ports. However, just pin 1 & pin 2 needed, and the lightness of LCD should be controlled by a 10kΩ variable resistor.
  9. 9. RESULT gender Age HR on display HR on oscilloscope Error rate (%) Male 22 97 96 1.03 Male 22 83 81 2.41 Male 20 78 78 0 Male 22 90 87 3.33 Male 20 80 79 1.25 Female 22 77 77 0 Female 22 104 103 0.96 Female 19 75 75 0 Female 20 69 71 2.81 The accuracy of the device depends on the testing performed is about 1.414. In general, the HR of females is higher than males. But this is not true where difference in weight, daily life (i.e. kind of work, health, inheritance of heart disease, doing exercises… etc).
  10. 10. CONCLUSION Infrared sensor instead of pressure sensor was used in Heart Rate Monitor, it has higher sensitivity, and the output signal is more stability. The infrared sensor depends on the fact that the speed of blood is proportional to heart rate. That is to say, when people are in high active, the heart rate will change. As the same theory as previous research, heart rate changes when people are excited. The HR value displayed on LCD shows the information that when people took exercise, the heart rate will increase; sometimes it is out of healthy range. When people take a break, the normal heart rate will recover. In totally, heart rate is physiological indicators to show what is going on in body. People need to care about their healthy condition with the help of Heart Rate Monitor.
  11. 11. References [1] Sokwoo Rhee “DESIGN AND ANALYSIS ARTIFACTRESISTIVEFINGER PHOTOPLETHYSMOGRAPHIC SENSORS FOR VITAL SIGN MONITORING” Massachusetts,U.S, 2001, chapter1,P9. [2] World health organization, Mortality country fact sheet 2006, Sudan. . WEB URL- [3] Wikipedia – http://en.wikipedia.org/wiki/Main_Page.
  12. 12. THE END THANK YOU

×