MRI ARTIFACTS
 What is an artifact??
 An image artifact is a structure not normally
present.
 But visible as a result of a limitation or
malfunction in the hardware or software in the
MRI device.
 Some affect the quality of the MRI exam while
others do not affect the diagnostic quality but
may be confused with pathology.
 The knowledge of MRI artifacts and noise
producing factors is important for continuing
maintenance of image quality.
 Sources of artifacts
Hardware issues
Software problems
Physiological phenomena
Physical limitations
 Types of artifacts
 Ghosting/Motion artifacts
 Aliasing
 Chemical shift
Gibbs or truncation
 Magnetic Susceptibility
 Zipper artifact/Straight Lines
 Shading
Cross Excitation and Cross Talk
 Parallel Imaging
 Black boundary
 Entry slice phenomenon
 Field inhomogenity
 Slice overlapping
Magic angle
Moire fringes
RF overflow
Eddy current artifacts
Center point artifact
Motion artifacts
 Voluntary/involuntary patient motion.
 Blurness in the image.
 Caused by
 Arterial pulsations
 CSF pulsations
 Swallowing
 Breathing
 Peristalsis
 Physical movement
 Solutions
 Patient education
 Sedation
 Motion correction
sequences e.g.
HASTE
 Navigator echo
 Respiratory gating
 Cardiac triggering
Motion artifacts
Aliasing artifact or
wraparound
 It occurs when the dimensions of the body part
being imaged exceed the FoV.
 The part beyond the FoV is projected to the
other side of the image.
 It is caused by under sampling in the phase
encoded direction.
 Choosing an FoV that is smaller than the area
imaged leads to wraparound or aliasing
artifacts.
 Appearance
 signals from outside the FoV seen in the
imaging volume.
 Solution
 Larger FoV and oversampling
 Use saturation bands outside the FoV.
Chemical shift artifact
 In the frequency direction, the MR scanner uses
the frequency of the signal to indicate spatial
resolution
 Since water in organs and muscles resonate at
different frequency than fat.
 The MR scanner mistakes the frequency as a
spatial difference.
 Commonly noticed in vertebral end plates,
abdomen, and orbits where fat and other tissues
form borders.
 Water protons frequency 3.5 ppm precess at
 Appearance
 Spine -one end plate to appear thicker than
the opposite one.
 Abdomen &orbits -Black border at fat water
interfaces and bright border at opposite
border.
 Solution
fat suppression technique
Use wide receiver bandwidth
Gibbs or truncation artifact
 Occurs at high contrast boundaries.
 Due to truncation(omission) of sampled signals.
 Commonly seen at the low signal intensity spinal
cord with high signal intensity CSF on T2WI of the
spine.
 As the signal is sampled, some data is necessarily
omitted in k-space, causing the signal intensity of
a given pixel to vary from its ideal signal intensity.
 Appearance
 Bright and dark lines.
 Solution
 Increase matrix size.
Magnetic suscepbility artifact
 Distortion in the MR image especially seen in
while imaging with metallic orthopedic hardware
or dental work.
 Magnetic field inhomogenetics introduced by the
metallic object in to homogeneous magnetic
field.
 Greater at high magnetic field strength.
 Worst with long TE and gradient echo
sequences.
 Appearance
Bright and dark areas
 Solution
Larger receiver bandwidth
Gradient echo and echo planar sequences
should be avoided.
The use of echo spin particularly fast spin
echo sequences should be considered.
Zipper artifact
 Most of them related to hardware or software
problems.
 RF from some radio transmitters will cause
zipper that are oriented perpendicular to
frequency direction.
 Occur in either frequency or phase encoding
directions.
 Solutions
MR scanner room is shut down.
Remove all electric devices from patient.
Field inhomogenity
 Contribute to signal non uniformity.
 May arise from RF coil construction or standing
wave effects.
 When the RF is shorter than the dimensions of
the anatomic structures examined standing
waves may result.
 Leads to inhomogeneous fat suppression.
 Solution
 Shimming –allow precise control of overall
homogeneity of RF field.
 Use STIR for Fat Sat.
 Coil- use volume coils, allow space between body
and coil.
 Dielectric –use phased array coils.
RF overflow/Shading artifact
 Non uniform, washed out appearance in an
image.
 Occurs when the signal received from the patient
is too intense to be digitized by analog to digital
converter.
 Auto prescanning usually adjusts the receiver
gain.
 Post processing methods also existing but may
be time consuming.
Central Point/Parallel Imaging
Artifact
 A focal dot of increased or decreased signal in
the center of the image.
 Caused by constant offset of the DC voltage in
the amplifiers.
 Solution
 Requires calibration.
 Maintain a constant temperature.
Black Boundary artifact
 Artificially created black line located at water fat
interfaces.
 This results in sharp delineation of the muscle
fat boundary that is sometimes visually
appealing but not an anatomical structure.
 Occurs at TE when the fat and water spins
located in the same pixel are out of phase,
cancelling each others signal.
 Particularly noticeable in GRE sequences.
 Both in frequency and phase directions.
 Remedy
Use phase TE’s
Fat suppression technique
Increase matrix size and bandwidth.
Entry slice phenomenon
 When unsaturated spins in blood first enter in to
a slice or slices.
 Characterized by bright signal in a blood vessel
at the first slice.
 Confused with thrombosis.
 Gradient echo flow techniques to differentiate.
 Solution
 Use spatial saturation bands before the first and
after the last.
Slice overlapping
 Loss of signal due to multi slice, multi angle
acquisition or imperfect slice profile.
 Mechanism : Spin saturation.
 If slices at different angles cross, then spins that
have previously excited, could be excited again.
 Mostly seen in spine imaging e.g. L4-L5 or L5-
S1.
 Appearance
 Band of signal loss crossing horizontally in the
image, usually posteriorly.
 Solution
 Continuous imaging
 Increase slice gap.
Magic angle artifact
 Seen most frequently in tendons and ligaments that
are oriented at a 55 degree to main magnetic field.
 Normally dipolar interaction between water
molecules in ligaments are strong.
 Which implies that T2 relaxation is very fast leading
to signal loss.
 Dipolar interaction go to zero.
 Solution
 Lengthen TE
 Use T1 weighted sequences since T1 relaxation is
unaffected by this.
Moire fringes
 An interference pattern most commonly when
doing gradient echo images.
 One cause is aliasing of one side of the body to
the other results in superimposition.
 Can also be caused by receiver picking up a
stimulated echo.
 Similar to the effect of looking through two
window screens.
 Solution
 Improve shimming
Eddy current artifact
 Varying magnetic field induces electric current
which distort gradient waveforms.
 When diffusion gradient applied, change in
magnetic field creates electric current .
 Such current creates smaller magnetic field that
Bo.
 Modern gradient coils equipped with active
shielding to avoid these effects of electric
conduction.
 Solution
 Shielding gradients.
 A distorted gradient
waveform is used
Spike artifact
 Caused by one bad data point k-space.
 Shows one data point in k-space, which is out of
ordinary.
 Diagonal lines in the image.
 Solution
Repeat the scan
Annefact artifact
 Due to anatomy within the active volume of the
coil, but outside the Fov.
 Signals from such region give rise to ghost
imaging in the phase encoding direction.
 Most commonly in sagittal spine imaging
especially in T and L spines.
 Always turn off the coils which is not in use.
MRI Artifacts

Weitere ähnliche Inhalte

Was ist angesagt?

magnetic resonance in angiography
magnetic resonance in  angiography magnetic resonance in  angiography
magnetic resonance in angiography BISHAL KHANAL
 
MRI Artifact Radiology
MRI Artifact RadiologyMRI Artifact Radiology
MRI Artifact RadiologyHenock Negasi
 
Generations and advancement of CT
Generations and advancement of CTGenerations and advancement of CT
Generations and advancement of CTsuman duwal
 
PARAMETERS AND ASSOCIATED Trade-Offs, MRI
PARAMETERS AND ASSOCIATED  Trade-Offs, MRIPARAMETERS AND ASSOCIATED  Trade-Offs, MRI
PARAMETERS AND ASSOCIATED Trade-Offs, MRIBeuniquewithNehaSing
 
CT numbers, window width and window level
CT numbers, window width and window levelCT numbers, window width and window level
CT numbers, window width and window levelGanesan Yogananthem
 
Mri image quality gamal mahdaly
Mri image quality gamal mahdalyMri image quality gamal mahdaly
Mri image quality gamal mahdalyGamal Mahdaly
 
The Physics of MRI
The Physics of MRIThe Physics of MRI
The Physics of MRIbiomedicz
 
Ultrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pkUltrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pkDr pradeep Kumar
 
CT Physics
CT PhysicsCT Physics
CT PhysicsRMLIMS
 
Gradient echo pulse sequence swastik
Gradient echo pulse sequence  swastikGradient echo pulse sequence  swastik
Gradient echo pulse sequence swastikBipul Poudel
 
Tissue harmonic imaging
Tissue harmonic imaging Tissue harmonic imaging
Tissue harmonic imaging MAMTA PANDA
 
CT Scan Image reconstruction
CT Scan Image reconstructionCT Scan Image reconstruction
CT Scan Image reconstructionGunjan Patel
 
K Space in MRI
K Space in MRIK Space in MRI
K Space in MRIKajal Jha
 
Side lobes and Harmonic Imaging
Side lobes and Harmonic ImagingSide lobes and Harmonic Imaging
Side lobes and Harmonic ImagingRamzee Small
 
Physics of Multidetector CT Scan
Physics of Multidetector CT ScanPhysics of Multidetector CT Scan
Physics of Multidetector CT ScanDr Varun Bansal
 
Ultrasound instrumentation
Ultrasound instrumentationUltrasound instrumentation
Ultrasound instrumentationVnAy Kris
 

Was ist angesagt? (20)

magnetic resonance in angiography
magnetic resonance in  angiography magnetic resonance in  angiography
magnetic resonance in angiography
 
MRI Artifact Radiology
MRI Artifact RadiologyMRI Artifact Radiology
MRI Artifact Radiology
 
MRI Artefacts
MRI ArtefactsMRI Artefacts
MRI Artefacts
 
Generations and advancement of CT
Generations and advancement of CTGenerations and advancement of CT
Generations and advancement of CT
 
PARAMETERS AND ASSOCIATED Trade-Offs, MRI
PARAMETERS AND ASSOCIATED  Trade-Offs, MRIPARAMETERS AND ASSOCIATED  Trade-Offs, MRI
PARAMETERS AND ASSOCIATED Trade-Offs, MRI
 
CT numbers, window width and window level
CT numbers, window width and window levelCT numbers, window width and window level
CT numbers, window width and window level
 
Mri image quality gamal mahdaly
Mri image quality gamal mahdalyMri image quality gamal mahdaly
Mri image quality gamal mahdaly
 
The Physics of MRI
The Physics of MRIThe Physics of MRI
The Physics of MRI
 
Ultrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pkUltrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pk
 
CT Physics
CT PhysicsCT Physics
CT Physics
 
Gradient echo pulse sequence swastik
Gradient echo pulse sequence  swastikGradient echo pulse sequence  swastik
Gradient echo pulse sequence swastik
 
CT Artifacts
CT ArtifactsCT Artifacts
CT Artifacts
 
Mri physics
Mri physicsMri physics
Mri physics
 
Tissue harmonic imaging
Tissue harmonic imaging Tissue harmonic imaging
Tissue harmonic imaging
 
CT Scan Image reconstruction
CT Scan Image reconstructionCT Scan Image reconstruction
CT Scan Image reconstruction
 
K Space in MRI
K Space in MRIK Space in MRI
K Space in MRI
 
Side lobes and Harmonic Imaging
Side lobes and Harmonic ImagingSide lobes and Harmonic Imaging
Side lobes and Harmonic Imaging
 
Physics of Multidetector CT Scan
Physics of Multidetector CT ScanPhysics of Multidetector CT Scan
Physics of Multidetector CT Scan
 
Artifacts
ArtifactsArtifacts
Artifacts
 
Ultrasound instrumentation
Ultrasound instrumentationUltrasound instrumentation
Ultrasound instrumentation
 

Ähnlich wie MRI Artifacts

Artifacts and Quality Control
Artifacts and Quality ControlArtifacts and Quality Control
Artifacts and Quality ControlChaun Dee
 
Navneet.mr hardware & artifacts related to it
Navneet.mr hardware & artifacts related to itNavneet.mr hardware & artifacts related to it
Navneet.mr hardware & artifacts related to itNavneet Ranjan
 
artifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxartifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxShubham661884
 
ppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxKeerthan39
 
ICU Echo, now in 3D - by Janin
ICU Echo, now in 3D - by JaninICU Echo, now in 3D - by Janin
ICU Echo, now in 3D - by JaninSMACC Conference
 
Laser and its use in veterinary practice
Laser and its use in veterinary practiceLaser and its use in veterinary practice
Laser and its use in veterinary practiceManzoor Bhat
 
Neuroimaging Lecture
Neuroimaging LectureNeuroimaging Lecture
Neuroimaging Lecturetest
 
1 diagnostic imaging
1 diagnostic imaging1 diagnostic imaging
1 diagnostic imagingSimba Syed
 
Distortion Artifacts in MRI and their correction
Distortion Artifacts in MRI and their correctionDistortion Artifacts in MRI and their correction
Distortion Artifacts in MRI and their correctionMiami Cancer Institute
 
MRI 3-Marks QUESTIONS AND ANSWERS
MRI 3-Marks QUESTIONS AND ANSWERSMRI 3-Marks QUESTIONS AND ANSWERS
MRI 3-Marks QUESTIONS AND ANSWERSGanesan Yogananthem
 
Mrisequences 130118064505-phpapp02
Mrisequences 130118064505-phpapp02Mrisequences 130118064505-phpapp02
Mrisequences 130118064505-phpapp02Hereward College
 
Biopotentials
BiopotentialsBiopotentials
Biopotentialsstooty s
 
Emmanuel w. fiagbedzi presentation on MRI ARTIFACTS(WATER FAT SHIFT)
Emmanuel w. fiagbedzi presentation on MRI ARTIFACTS(WATER FAT SHIFT)Emmanuel w. fiagbedzi presentation on MRI ARTIFACTS(WATER FAT SHIFT)
Emmanuel w. fiagbedzi presentation on MRI ARTIFACTS(WATER FAT SHIFT)efiagbedzi
 

Ähnlich wie MRI Artifacts (20)

Artifacts and Quality Control
Artifacts and Quality ControlArtifacts and Quality Control
Artifacts and Quality Control
 
Navneet.mr hardware & artifacts related to it
Navneet.mr hardware & artifacts related to itNavneet.mr hardware & artifacts related to it
Navneet.mr hardware & artifacts related to it
 
artifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxartifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptx
 
mri artefacts.pptx
mri artefacts.pptxmri artefacts.pptx
mri artefacts.pptx
 
ppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptx
 
ppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptx
 
MRIARTIFACTS.pptx
MRIARTIFACTS.pptxMRIARTIFACTS.pptx
MRIARTIFACTS.pptx
 
ICU Echo, now in 3D - by Janin
ICU Echo, now in 3D - by JaninICU Echo, now in 3D - by Janin
ICU Echo, now in 3D - by Janin
 
CT and MRI
CT and MRICT and MRI
CT and MRI
 
Laser and its use in veterinary practice
Laser and its use in veterinary practiceLaser and its use in veterinary practice
Laser and its use in veterinary practice
 
Mri artifacts
Mri artifactsMri artifacts
Mri artifacts
 
LASER SURGERY VSR.pptx
LASER SURGERY VSR.pptxLASER SURGERY VSR.pptx
LASER SURGERY VSR.pptx
 
Neuroimaging Lecture
Neuroimaging LectureNeuroimaging Lecture
Neuroimaging Lecture
 
1 diagnostic imaging
1 diagnostic imaging1 diagnostic imaging
1 diagnostic imaging
 
Distortion Artifacts in MRI and their correction
Distortion Artifacts in MRI and their correctionDistortion Artifacts in MRI and their correction
Distortion Artifacts in MRI and their correction
 
MRI 3-Marks QUESTIONS AND ANSWERS
MRI 3-Marks QUESTIONS AND ANSWERSMRI 3-Marks QUESTIONS AND ANSWERS
MRI 3-Marks QUESTIONS AND ANSWERS
 
Mrisequences 130118064505-phpapp02
Mrisequences 130118064505-phpapp02Mrisequences 130118064505-phpapp02
Mrisequences 130118064505-phpapp02
 
Biopotentials
BiopotentialsBiopotentials
Biopotentials
 
MRI Basics by Saleh Asad
MRI  Basics  by Saleh AsadMRI  Basics  by Saleh Asad
MRI Basics by Saleh Asad
 
Emmanuel w. fiagbedzi presentation on MRI ARTIFACTS(WATER FAT SHIFT)
Emmanuel w. fiagbedzi presentation on MRI ARTIFACTS(WATER FAT SHIFT)Emmanuel w. fiagbedzi presentation on MRI ARTIFACTS(WATER FAT SHIFT)
Emmanuel w. fiagbedzi presentation on MRI ARTIFACTS(WATER FAT SHIFT)
 

Mehr von Rahman Ud Din

Radiobiology-Human Cell
Radiobiology-Human CellRadiobiology-Human Cell
Radiobiology-Human CellRahman Ud Din
 
Steady State Free Precession MRA
Steady State Free Precession MRASteady State Free Precession MRA
Steady State Free Precession MRARahman Ud Din
 
Phase Contrast and ECG-Gated MRA
Phase Contrast and ECG-Gated MRA Phase Contrast and ECG-Gated MRA
Phase Contrast and ECG-Gated MRA Rahman Ud Din
 
Magnetic Resonance Cholangiopancreatography (MRCP)
Magnetic Resonance Cholangiopancreatography (MRCP)Magnetic Resonance Cholangiopancreatography (MRCP)
Magnetic Resonance Cholangiopancreatography (MRCP)Rahman Ud Din
 
Cardiac Magnetic Resonance Imaging
Cardiac Magnetic Resonance ImagingCardiac Magnetic Resonance Imaging
Cardiac Magnetic Resonance ImagingRahman Ud Din
 
Echo in Pericardial Diseases
Echo in Pericardial DiseasesEcho in Pericardial Diseases
Echo in Pericardial DiseasesRahman Ud Din
 
Magnetic resonance spectroscopy
Magnetic resonance spectroscopyMagnetic resonance spectroscopy
Magnetic resonance spectroscopyRahman Ud Din
 
Magnetic Resonance Perfusion
Magnetic Resonance PerfusionMagnetic Resonance Perfusion
Magnetic Resonance PerfusionRahman Ud Din
 
Magnetic Resonance Diffusion
Magnetic Resonance DiffusionMagnetic Resonance Diffusion
Magnetic Resonance DiffusionRahman Ud Din
 
Magnetic resonance angiography
Magnetic resonance angiographyMagnetic resonance angiography
Magnetic resonance angiographyRahman Ud Din
 

Mehr von Rahman Ud Din (12)

Human cell division
Human cell divisionHuman cell division
Human cell division
 
Radiobiology-Human Cell
Radiobiology-Human CellRadiobiology-Human Cell
Radiobiology-Human Cell
 
Steady State Free Precession MRA
Steady State Free Precession MRASteady State Free Precession MRA
Steady State Free Precession MRA
 
Phase Contrast and ECG-Gated MRA
Phase Contrast and ECG-Gated MRA Phase Contrast and ECG-Gated MRA
Phase Contrast and ECG-Gated MRA
 
Magnetic Resonance Cholangiopancreatography (MRCP)
Magnetic Resonance Cholangiopancreatography (MRCP)Magnetic Resonance Cholangiopancreatography (MRCP)
Magnetic Resonance Cholangiopancreatography (MRCP)
 
Cardiac Magnetic Resonance Imaging
Cardiac Magnetic Resonance ImagingCardiac Magnetic Resonance Imaging
Cardiac Magnetic Resonance Imaging
 
Echo in Pericardial Diseases
Echo in Pericardial DiseasesEcho in Pericardial Diseases
Echo in Pericardial Diseases
 
Magnetic resonance spectroscopy
Magnetic resonance spectroscopyMagnetic resonance spectroscopy
Magnetic resonance spectroscopy
 
PWI ASL
PWI ASLPWI ASL
PWI ASL
 
Magnetic Resonance Perfusion
Magnetic Resonance PerfusionMagnetic Resonance Perfusion
Magnetic Resonance Perfusion
 
Magnetic Resonance Diffusion
Magnetic Resonance DiffusionMagnetic Resonance Diffusion
Magnetic Resonance Diffusion
 
Magnetic resonance angiography
Magnetic resonance angiographyMagnetic resonance angiography
Magnetic resonance angiography
 

Kürzlich hochgeladen

Kolkata Call Girls Naktala 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Kolkata Call Girls Naktala  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Girl Se...Kolkata Call Girls Naktala  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Girl Se...
Kolkata Call Girls Naktala 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...Namrata Singh
 
❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...
❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...
❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...Sheetaleventcompany
 
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Availableperfect solution
 
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...Sheetaleventcompany
 
Ahmedabad Call Girls Book Now 9630942363 Top Class Ahmedabad Escort Service A...
Ahmedabad Call Girls Book Now 9630942363 Top Class Ahmedabad Escort Service A...Ahmedabad Call Girls Book Now 9630942363 Top Class Ahmedabad Escort Service A...
Ahmedabad Call Girls Book Now 9630942363 Top Class Ahmedabad Escort Service A...Sheetaleventcompany
 
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...Sheetaleventcompany
 
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...soniyagrag336
 
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana GuptaLifecare Centre
 
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...gragneelam30
 
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service AvailableCall Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service AvailableJanvi Singh
 
Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...
Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...
Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...Sheetaleventcompany
 
Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...
Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...
Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...Sheetaleventcompany
 
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...call girls hydrabad
 
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room DeliveryCall 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room DeliveryJyoti singh
 
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...Janvi Singh
 
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...Sheetaleventcompany
 
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...TanyaAhuja34
 
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book nowChennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book nowtanudubay92
 
Call Girl In Chandigarh 📞9809698092📞 Just📲 Call Inaaya Chandigarh Call Girls ...
Call Girl In Chandigarh 📞9809698092📞 Just📲 Call Inaaya Chandigarh Call Girls ...Call Girl In Chandigarh 📞9809698092📞 Just📲 Call Inaaya Chandigarh Call Girls ...
Call Girl In Chandigarh 📞9809698092📞 Just📲 Call Inaaya Chandigarh Call Girls ...Sheetaleventcompany
 

Kürzlich hochgeladen (20)

Kolkata Call Girls Naktala 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Kolkata Call Girls Naktala  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Girl Se...Kolkata Call Girls Naktala  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Girl Se...
Kolkata Call Girls Naktala 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
 
❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...
❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...
❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...
 
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
 
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
 
Ahmedabad Call Girls Book Now 9630942363 Top Class Ahmedabad Escort Service A...
Ahmedabad Call Girls Book Now 9630942363 Top Class Ahmedabad Escort Service A...Ahmedabad Call Girls Book Now 9630942363 Top Class Ahmedabad Escort Service A...
Ahmedabad Call Girls Book Now 9630942363 Top Class Ahmedabad Escort Service A...
 
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
 
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
 
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
 
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
 
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service AvailableCall Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
 
Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...
Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...
Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...
 
Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...
Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...
Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...
 
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
 
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room DeliveryCall 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
 
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
 
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
 
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
 
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
 
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book nowChennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
 
Call Girl In Chandigarh 📞9809698092📞 Just📲 Call Inaaya Chandigarh Call Girls ...
Call Girl In Chandigarh 📞9809698092📞 Just📲 Call Inaaya Chandigarh Call Girls ...Call Girl In Chandigarh 📞9809698092📞 Just📲 Call Inaaya Chandigarh Call Girls ...
Call Girl In Chandigarh 📞9809698092📞 Just📲 Call Inaaya Chandigarh Call Girls ...
 

MRI Artifacts

  • 2.  What is an artifact??  An image artifact is a structure not normally present.  But visible as a result of a limitation or malfunction in the hardware or software in the MRI device.  Some affect the quality of the MRI exam while others do not affect the diagnostic quality but may be confused with pathology.  The knowledge of MRI artifacts and noise producing factors is important for continuing maintenance of image quality.
  • 3.  Sources of artifacts Hardware issues Software problems Physiological phenomena Physical limitations
  • 4.  Types of artifacts  Ghosting/Motion artifacts  Aliasing  Chemical shift Gibbs or truncation  Magnetic Susceptibility  Zipper artifact/Straight Lines  Shading Cross Excitation and Cross Talk  Parallel Imaging
  • 5.  Black boundary  Entry slice phenomenon  Field inhomogenity  Slice overlapping Magic angle Moire fringes RF overflow Eddy current artifacts Center point artifact
  • 6. Motion artifacts  Voluntary/involuntary patient motion.  Blurness in the image.  Caused by  Arterial pulsations  CSF pulsations  Swallowing  Breathing  Peristalsis  Physical movement  Solutions  Patient education  Sedation  Motion correction sequences e.g. HASTE  Navigator echo  Respiratory gating  Cardiac triggering
  • 8. Aliasing artifact or wraparound  It occurs when the dimensions of the body part being imaged exceed the FoV.  The part beyond the FoV is projected to the other side of the image.  It is caused by under sampling in the phase encoded direction.  Choosing an FoV that is smaller than the area imaged leads to wraparound or aliasing artifacts.
  • 9.  Appearance  signals from outside the FoV seen in the imaging volume.  Solution  Larger FoV and oversampling  Use saturation bands outside the FoV.
  • 10.
  • 11.
  • 12. Chemical shift artifact  In the frequency direction, the MR scanner uses the frequency of the signal to indicate spatial resolution  Since water in organs and muscles resonate at different frequency than fat.  The MR scanner mistakes the frequency as a spatial difference.  Commonly noticed in vertebral end plates, abdomen, and orbits where fat and other tissues form borders.  Water protons frequency 3.5 ppm precess at
  • 13.  Appearance  Spine -one end plate to appear thicker than the opposite one.  Abdomen &orbits -Black border at fat water interfaces and bright border at opposite border.  Solution fat suppression technique Use wide receiver bandwidth
  • 14.
  • 15.
  • 16. Gibbs or truncation artifact  Occurs at high contrast boundaries.  Due to truncation(omission) of sampled signals.  Commonly seen at the low signal intensity spinal cord with high signal intensity CSF on T2WI of the spine.  As the signal is sampled, some data is necessarily omitted in k-space, causing the signal intensity of a given pixel to vary from its ideal signal intensity.  Appearance  Bright and dark lines.  Solution  Increase matrix size.
  • 17.
  • 18. Magnetic suscepbility artifact  Distortion in the MR image especially seen in while imaging with metallic orthopedic hardware or dental work.  Magnetic field inhomogenetics introduced by the metallic object in to homogeneous magnetic field.  Greater at high magnetic field strength.  Worst with long TE and gradient echo sequences.
  • 19.  Appearance Bright and dark areas  Solution Larger receiver bandwidth Gradient echo and echo planar sequences should be avoided. The use of echo spin particularly fast spin echo sequences should be considered.
  • 20.
  • 21. Zipper artifact  Most of them related to hardware or software problems.  RF from some radio transmitters will cause zipper that are oriented perpendicular to frequency direction.  Occur in either frequency or phase encoding directions.  Solutions MR scanner room is shut down. Remove all electric devices from patient.
  • 22.
  • 23. Field inhomogenity  Contribute to signal non uniformity.  May arise from RF coil construction or standing wave effects.  When the RF is shorter than the dimensions of the anatomic structures examined standing waves may result.  Leads to inhomogeneous fat suppression.
  • 24.  Solution  Shimming –allow precise control of overall homogeneity of RF field.  Use STIR for Fat Sat.  Coil- use volume coils, allow space between body and coil.  Dielectric –use phased array coils.
  • 25.
  • 26. RF overflow/Shading artifact  Non uniform, washed out appearance in an image.  Occurs when the signal received from the patient is too intense to be digitized by analog to digital converter.  Auto prescanning usually adjusts the receiver gain.  Post processing methods also existing but may be time consuming.
  • 27.
  • 28. Central Point/Parallel Imaging Artifact  A focal dot of increased or decreased signal in the center of the image.  Caused by constant offset of the DC voltage in the amplifiers.  Solution  Requires calibration.  Maintain a constant temperature.
  • 29.
  • 30. Black Boundary artifact  Artificially created black line located at water fat interfaces.  This results in sharp delineation of the muscle fat boundary that is sometimes visually appealing but not an anatomical structure.  Occurs at TE when the fat and water spins located in the same pixel are out of phase, cancelling each others signal.  Particularly noticeable in GRE sequences.  Both in frequency and phase directions.
  • 31.  Remedy Use phase TE’s Fat suppression technique Increase matrix size and bandwidth.
  • 32. Entry slice phenomenon  When unsaturated spins in blood first enter in to a slice or slices.  Characterized by bright signal in a blood vessel at the first slice.  Confused with thrombosis.  Gradient echo flow techniques to differentiate.  Solution  Use spatial saturation bands before the first and after the last.
  • 33.
  • 34. Slice overlapping  Loss of signal due to multi slice, multi angle acquisition or imperfect slice profile.  Mechanism : Spin saturation.  If slices at different angles cross, then spins that have previously excited, could be excited again.  Mostly seen in spine imaging e.g. L4-L5 or L5- S1.
  • 35.  Appearance  Band of signal loss crossing horizontally in the image, usually posteriorly.  Solution  Continuous imaging  Increase slice gap.
  • 36.
  • 37. Magic angle artifact  Seen most frequently in tendons and ligaments that are oriented at a 55 degree to main magnetic field.  Normally dipolar interaction between water molecules in ligaments are strong.  Which implies that T2 relaxation is very fast leading to signal loss.  Dipolar interaction go to zero.  Solution  Lengthen TE  Use T1 weighted sequences since T1 relaxation is unaffected by this.
  • 38.
  • 39. Moire fringes  An interference pattern most commonly when doing gradient echo images.  One cause is aliasing of one side of the body to the other results in superimposition.  Can also be caused by receiver picking up a stimulated echo.  Similar to the effect of looking through two window screens.
  • 41. Eddy current artifact  Varying magnetic field induces electric current which distort gradient waveforms.  When diffusion gradient applied, change in magnetic field creates electric current .  Such current creates smaller magnetic field that Bo.  Modern gradient coils equipped with active shielding to avoid these effects of electric conduction.
  • 42.  Solution  Shielding gradients.  A distorted gradient waveform is used
  • 43. Spike artifact  Caused by one bad data point k-space.  Shows one data point in k-space, which is out of ordinary.  Diagonal lines in the image.  Solution Repeat the scan
  • 44.
  • 45. Annefact artifact  Due to anatomy within the active volume of the coil, but outside the Fov.  Signals from such region give rise to ghost imaging in the phase encoding direction.  Most commonly in sagittal spine imaging especially in T and L spines.  Always turn off the coils which is not in use.