SlideShare ist ein Scribd-Unternehmen logo
1 von 106
RADNOTI 130003
ISOLATED PERFUSED
   LIVER SYSTEM
Key system Components
Key system components:


•Primary reservoir
Key system components:


•Primary reservoir
•Secondary reservoir
Key system components:


•Primary reservoir
•Secondary reservoir
•Membrane Oxygenator
Key system components:


•Primary reservoir
•Secondary reservoir
•Membrane Oxygenator
•Bubble Trap Compliance Chamber
Key system components:


•Primary reservoir
•Secondary reservoir
•Membrane Oxygenator
•Bubble Trap Compliance Chamber
•Inflow Manifold
Key system components:


•Primary reservoir
•Secondary reservoir
•Membrane Oxygenator
•Bubble Trap Compliance Chamber
•Inflow Manifold
•Out Flow Manifold
Key system components:


•Out Flow Bubble Trap
Key system components:


•Out Flow Bubble Trap
•Peristaltic Pump
Key system components:


•Out Flow Bubble Trap
•Peristaltic Pump
•Liver Chamber
Key system components:


•Out Flow Bubble Trap
•Peristaltic Pump
•Liver Chamber
•3-Way Out Flow Manifold
Key system components:


•Out Flow Bubble Trap
•Peristaltic Pump
•Liver Chamber
•3-Way Out Flow Manifold
•3-Way Overflow Manifold
Flow Path Overview
Flow Path:
Perfusate is selected
either from the primary
reservoir
Flow Path:
Perfusate is selected
either from the primary
reservoir or the
secondary reservoir
Flow Path:
Perfusate is selected
either from the primary
reservoir or the
secondary reservoir
using the three way
stopcock located at the
outlet of the secondary
reservoir.
Flow Path:
Perfusate is selected
either from the primary
reservoir or the
secondary reservoir
using the three way
stopcock located at the
outlet of the secondary
reservoir.
Flow Path:
Perfusate is selected
either from the primary
reservoir or the
secondary reservoir
using the three way
stopcock located at the
outlet of the secondary
reservoir. The perfusate
then travels to the
peristaltic pump
Flow Path:
and is pumped through
the membrane
oxygenating chamber
Flow Path:
and up to the bubble trap
compliance chamber
Flow Path:
and up to the bubble trap
compliance chamber          Tip: on initial priming or
                            filling of the system you will
                            most likely need to close
                            the compliance port
                            stopcock
Flow Path:
and up to the bubble trap
compliance chamber          Tip: on initial priming or
                            filling of the system you will
                            most likely need to close
                            the compliance port
                            stopcock
                            and open the bubble trap
                            vent stopcock
Flow Path:
and up to the bubble trap
compliance chamber          Tip: on initial priming or
                            filling of the system you will
                            most likely need to close
                            the compliance port
                            stopcock
                            and open the bubble trap
                            vent stopcock
                            so that the trap will have the
                            opportunity to fill
Flow Path:
and up to the bubble trap
compliance chamber          When running in constant
                            pressure mode, perfusate
                            flow from the pump that is
                            greater than the flow rate of
                            the organ , will exit the
                            bubble trap via the
                            compliance port.
Flow Path:
and up to the bubble trap
compliance chamber          Generally you can leave this
                            port open after the system
                            has been primed.
Flow Path:
and up to the bubble trap
compliance chamber          Generally you can leave this
                            port open after the system
                            has been primed.
                            Over flow exiting the
                            compliance port is directed
                            to the overflow manifold.
Flow Path:
and up to the bubble trap
compliance chamber




                            The overflow manifold
                            will allow you to select
                            where the overflowing
                            perfusate is to be
                            directed.
Flow Path:
and up to the bubble trap
compliance chamber




                            Valve 1 directs flow
                            back to the primary
                            reservoir.
Flow Path:
and up to the bubble trap
compliance chamber




                            Valve 1 directs flow
                            back to the primary
                            reservoir.
                            Valve 2 directs flow
                            back to the secondary
                            reservoir.
Flow Path:
and up to the bubble trap
compliance chamber




                            Valve 1 directs flow
                            back to the primary
                            reservoir.
                            Valve 2 directs flow
                            back to the secondary
                            reservoir.
                            Valve 3 directs flow Out
                            to Waste.
Flow Path:
and up to the bubble trap
compliance chamber
the out flow of the bubble
trap then flows to the
inflow manifold.
Flow Path:
and up to the bubble trap
compliance chamber
the out flow of the bubble
trap then flows to the
inflow manifold.




                             Tip: When
                             initially priming
                             or flushing the
                             system, the
                             inflow and
                             outflow cannulae
                             will have to be
                             coupled. This
                             can be done by
                             using a small
                             section of tygon
                             tubing and
                             pushing the
                             cannulae tips in
                             either side.
Flow Path:
The perfusate then
passes through the
cannulated organ (or
tygon coupler when
priming or flushing) and
into the outflow manifold.
Flow Path:
Perfusate then flows out
to through the flow meter
( if so equipped) to the
outflow bubble trap.
Flow Path:
Perfusate then flows out
to through the flow meter
( if so equipped) to the
outflow bubble trap.




                            NOTE:
                            The vent
                            stopcock and
                            the out flow
                            stopcock of the
                            bubble trap
                            should be in the
                            closed position.
Flow Path:
Perfusate then flows out
to through the flow meter
( if so equipped) to the
outflow bubble trap.


The perfusate then is
drawn be the second
peristaltic pump head
and pushed to the three
way outflow manifold.
Flow Path:
At this point the
perfusate can be directed
to :
1. The primary reservoir
Flow Path:
At this point the
perfusate can be directed
to :
1. The primary reservoir
2. The secondary
reservoir
Flow Path:
At this point the
perfusate can be directed
to :
1. The primary reservoir
2. The secondary
reservoir
3. Waste




                            Sink or
                            collection flask
Flow Path:
Example:
If you are flushing the
system close valve 1
Flow Path:
Example:
If you are flushing the
system close valve 1 and
2
Flow Path:
Example:
If you are flushing the
system close valve 1 and
2
Leaving the valve open to
waste will direct the flow
out to waste.
Flow Path:

ALERT!




             SPECIAL NOTE:
             when flushing the
             organ you will
             want to protect
             components that
             may be sensitive to
             the initial effluent
             by diverting flow
             around the
             component.
Flow Path:




             SPECIAL NOTE:
             when flushing the
             organ you will
             want to protect
             components that
             may be sensitive to
             the initial effluent
             by diverting flow
             around the
             component.
Flow Path:




             SPECIAL NOTE:
             In this diagram the
             flow meter would
             need to be
             protected.
Flow Path:




             SPECIAL NOTE:
             In this diagram the
             flow meter would
             need to be
             protected. This is
             done by using the
             three way
             stopcock on the
             outflow manifold to
             bypass the flow
             meter.
Flow Path:




             SPECIAL NOTE:
             In this diagram the
             flow meter would
             need to be
             protected. This is
             done by using the
             three way
             stopcock on the
             outflow manifold to
             bypass the flow
             meter.
Flow Path:
Flow Path:
Example:
If the organ is flushed
and you wish to have a
recirculating path...
Flow Path:
Example:
If the organ is flushed
and you wish to have a
recirculating path...
In this case back to the
primary reservoir...
Flow Path:
Example:
If the organ is flushed
and you wish to have a
recirculating path...
In this case back to the
primary reservoir…
Close valve 3 (waste)
Flow Path:
Example:
If the organ is flushed
and you wish to have a
recirculating path...
In this case back to the
primary reservoir…
Close valve 3 (waste)
Close valve 2
Flow Path:
Example:
If the organ is flushed
and you wish to have a
recirculating path...
In this case back to the
primary reservoir…
Close valve 3 (waste)
Close valve 2.
Open valve 1
Flow Path:
Example:
If the organ is flushed
and you wish to have a
recirculating path...
Or
To recirculate back to the
secondary reservoir…
Flow Path:
Example:
If the organ is flushed
and you wish to have a
recirculating path...
Or
to recirculate back to the
secondary reservoir…
Close valve 1
Flow Path:
Example:
If the organ is flushed
and you wish to have a
recirculating path...
Or
to recirculate back to the
secondary reservoir…
Close valve 1
Open valve 2
Flow Path:
Example:
If the organ is flushed
and you wish to have a
recirculating path...
Or
to recirculate back to the
secondary reservoir…
Close valve 1
Open valve 2
Close valve 3 (waste)
INITIAL START UP
INITIAL START UP



♦Fill the primary
reservoir with buffer
INITIAL START UP

                        Make sure reservoir
♦Fill the primary       selection stopcock is
                        in the off position
reservoir with buffer
                        prior to filling.
INITIAL START UP

                          WARNING: ALWAYS MAKE SURE
                             THERE IS AN OPENING TO
♦Fill the primary         ATMOSPHERE ON THE PRIMARY
reservoir with buffer    RESERVOIR. IF FOR ANY REASON
                          GAS PREASURE IS ALLOWED TO
♦Turn on and adjust      BUILD IN THE RESERVOIR RISK OF
gas from tank           SERIOUS INJURY OR FATALITY MAY
                                     OCCUR.




            From tank
INITIAL START UP



♦Fill the primary       Note:
reservoir with buffer   Adjust gas so that
♦Turn on and adjust     a gentle stream of
gas from tank           bubbles flows from
                        the gas dispersion
                        frit in the primary
                        reservoir. This will
                        oxygenate the
                        buffer and serve as
                        a visual
                        representation that
                        gas is flowing
            From tank   through the
                        Membrane
                        Oxygenator.
INITIAL START UP



♦Fill the primary
reservoir with buffer
♦Turn on and adjust
gas from tank
♦Open compliance
port stopcock.
INITIAL START UP



♦Fill the primary
reservoir with buffer
♦Turn on and adjust
gas from tank
♦Open compliance
port stopcock.
♦Close valve 1 and 2
on three way out
flow manifold.
INITIAL START UP



♦Fill the primary
reservoir with buffer
♦Turn on and adjust
gas from tank
♦Open compliance
port stopcock.
♦Close valve 1 and 2
on three way out
flow manifold.
♦Open valve 3
(waste) on three way
out flow manifold
INITIAL START UP



♦Fill the primary
reservoir with buffer
♦Turn on and adjust
gas from tank
♦Open compliance
port stopcock.
♦Close valve 1 and 2
on three way out
flow manifold.
♦Open valve 3
(waste) on three way
out flow manifold
♦Close valve 2 and 3
on three-way
overflow manifold
INITIAL START UP



♦Fill the primary
reservoir with buffer
♦Turn on and adjust
gas from tank
♦Open compliance
port stopcock.
♦Close valve 1 and 2
on three way out
flow manifold.
♦Open valve 3
(waste) on three way
out flow manifold
♦Close valve 2 and 3
on three-way
overflow manifold
♦Open valve 1
(primary reservoir
return) on three-way
overflow manifold.
INITIAL START UP



♦Close the
stopcocks for the
inflow and outflow
pressure
transducers.
INITIAL START UP



♦Close the
stopcocks for the
inflow and outflow
pressure
transducers.
♦Direct the outflow
manifold three way
stopcock to pass
through the
manifold.
INITIAL START UP



♦Close the
stopcocks for the
inflow and outflow
pressure
transducers.
♦Direct the outflow
manifold three way
stopcock to pass
through the
manifold.
♦Close the drain and
relief ports
stopcocks on the out
flow bubble trap.
INITIAL START UP



♦Close the
stopcocks for the
inflow and outflow
pressure
transducers.
♦Direct the outflow
manifold three way
stopcock to pass
through the
manifold.
♦Close the drain and
relief ports
stopcocks on the out
flow bubble trap.
♦Couple the Inflow
and Outflow
cannulae.
INITIAL START UP



♦Close the
stopcocks for the
inflow and outflow
pressure
transducers.
♦Direct the outflow
manifold three way
stopcock to pass
through the
manifold.
♦Close the drain and
relief ports
stopcocks on the out
flow bubble trap.
♦Couple the Inflow
and Outflow
cannulae.
♦Turn on peristaltic
pump and turn
reservoir selection
three way stopcock
to feed from primary
                       ON
reservoir.
INITIAL START UP



The system will now
circulate buffer
driven by the
peristaltic pump.
It will take several
moments to purge
air from the lines.
 You will most likely
have to open and
close the vent and
overflow ports in the
bubble trap
compliance chamber
temporarily to build
up some perfusate.
Once it
approximately two
thirds full, return the
valves to their
previous position.



                          ON
INITIAL START UP



 Once the system is
primed, Turn off the
peristaltic pump.
Close stopcocks at
the inflow and
outflow manifold.


This will trap the
buffer in the lines
and keep the system
primed.
Signal Generation
Signal Generation



 Now is a good time
to calibrate the
pressure
transducers and ion
selective electrodes
(if so equipped.)
Signal Generation



 For the pressure
                          Pressure
                        Transducers
transducers typically
they will have to be
filled with fluid and
purged of bubbles.
Signal Generation Pressure Transducers



 For the pressure
transducers typically
they will have to be
filled with fluid and
purged of bubbles.


 Set the three-way
stopcock at the
outflow pressure
transducer so that
flow is accepted
from the outflow
manifold and the
purge port.
Signal Generation Pressure Transducers



 For the pressure
transducers typically
they will have to be
filled with fluid and
purged of bubbles.


 Set the three-way
stopcock at the
outflow pressure
transducer so that
flow is accepted
from the outflow
                                         Vent port
manifold and the
                                         stopcock
purge port.
                                         (not shown)
 Fill a disposable
syringe with buffer.
Open the transducer
purge port (one way
stopcock not
shown.) and gently
fill the pressure
transducer dome
causing air to be
purged.
Signal Generation Pressure Transducers



 Repeat the
procedure with the
inflow pressure
transducer.
Signal Generation Pressure Transducers



 The pressure
transducers can be
calibrated to your
data acquisition at
this time.
 Set the three-way
stopcock controlling
flow to the
transducer to the
closed position.
Signal Generation Pressure Transducers



 The pressure
transducers can be
calibrated to your
data acquisition at
this time.
 Set the three-way
stopcock controlling
flow to the
transducer to the
closed position.
 Open the purge
stopcock (not
shown) on the
transducer.
Signal Generation Pressure Transducers



 The pressure
transducers can be
calibrated to your
data acquisition at
this time.
 Set the three-way
stopcock controlling
flow to the
transducer to the
closed position.
 Open the purge
stopcock (not
shown) on the
transducer.
This will be your
zero pressure
calibration point.
Signal Generation Pressure Transducer



 Return the
stopcocks to their
previous position
(accepting flow from
the outflow
manifold) and
setting the purge
port stopcock (not
shown) to the closed
position.
This will be you're
high pressure
calibration point.
Signal Generation Pressure Transducer

                                        Note:

 Return the                             The pressure head is
stopcocks to their                      determined by the
previous position                       elevation of the
(accepting flow from                    bubble trap
the outflow                             compliance
manifold) and                           chamber. The
setting the purge                       distance from the
port stopcock (not                      chamber to the
shown) to the closed                    pressure transducer
position.                               can be calculated to
                                        a known pressure.
This will be you're
high pressure                           Distance in mm
calibration point.                      divided by 13.6 = mm
                                        of mercury
                                        perfusion pressure
                                        should be 10-15 mm
                                        Hg (15-25cm above
                                        the liver.)
Signal Generation Pressure Transducer

                                        Note:

 Repeat the                             The pressure head is
procedure for the                       determined by the
inflow manifold                         elevation of the
pressure transducer.                    bubble trap
                                        compliance
                                        chamber. The
                                        distance from the
                                        chamber to the
                                        pressure transducer
                                        can be calculated to
                                        a known pressure.


                                        Distance in mm
                                        divided by 13.6 = mm
                                        of mercury.
Signal Generation




                    pH ELECTRODES
Signal Generation pH Electrodes
The pH electrodes are
plugged directly from
the mili volt adapter to
the data acquisition
interface.
Ideally a pH electrode
will output a voltage of
0mV in a pH 7 buffer.
This can very by +/-
50mV based on the
individual pH
electrode.
The Nernst equation
tells us that a pH
buffer 4 should be
160mV greater (more
positive) than the
reading in a pH 7.
The reading in a pH10
should be -160mV less
(more negative than
that of a reading in a
pH7 buffer.
The 160 comes from
being 3(7-4)*59(Nernst
value at 20
degrees)=168mV for
100% slope and 160 is
slightly less than 100%
theoretical.
Signal Generation




                    O2 ELECTRODES
Signal Generation O2 Electrodes

Prior to calibration
of the oxygen
electrode, the
electrode should be
examined to insure
that the electrode
membrane is intact
and the interior
chamber is full of
buffer. If the
electrode requires
maintenance
please refer to the
manufacturers
instructions.
Signal Generation O2 electrodes

To obtain zero
oxygen reading, the
physiological
buffer, placed in a
vented calibration
container should be
gassed for at least
10 to 30 minutes
with pure nitrogen
at a rate of 3-6
bubbles per/sec to
maintain a constant
temperature and
gas saturation.
Signal Generation O2 Electrodes
To obtain zero
oxygen reading, the
physiological
buffer, placed in a
vented calibration
container should be
gassed for at least
10 to 30 minutes
with pure nitrogen
at a rate of 3-6
bubbles per/sec to
maintain a constant
temperature and
gas saturation.
The electrode is
then inserted into
the calibration
chamber and
monitored until the
reading is stabile.
Once the reading
has stabilized the
reading can bet set
to zero using the
amplifier gain and
offset adjustments.
The acquisition
software can use
this as the low
point calibration.
Signal Generation O2 Electrodes

The electrode is then
removed and inserted
into aerated container
of buffer. In this case
either the inflow or out
flow manifold with
perfusate being
pumped through. After
stabilizing the reading
would be adjusted
using the amplifier
gain and offset based
on the gas mixture
(room air 21% oxygen,
gas cylinder 100%
oxygen, gas cylinder
95% oxygen etc…)
This can be used as
your high point
calibration in the
datacquisiton
software.
Signal Generation O2 Electrodes

It is recommended
that the procedure
is repeated three
times in order to
insure the readings
are stable and
reproducible.
Signal Generation




                    Temperature
Signal Generation Temperature

Temperature probes
are preset and will not
require calibration. It
is recommended that
they be verified
periodically measuring
a known temperature
such as your heater
circulator bath and the
result compared to the
read out of the bath.
Signal Generation




                    Flow Meter
Signal Generation Flow Meter

The Flow Meter is
preset and will not
require calibration. It
is recommended that
it be verified
periodically measuring
a known flow and the
read out verified.
Liver Preparation
1. The animal is anesthetized and placed on its back; the anesthetic used may be a general
anesthetic such as isoflurane or phenobarbital, depending upon the protocol requirements. Test
for the depth of anesthesia via toe pinch, eye reflex, etc.

2. For best positioning, the limbs are retracted and secured with tape or string.


3. The abdomen is wiped with 70% alcohol; the abdomen can be shaved, although this is not
necessary.

4. A midline incision is made by lifting the skin with forceps and cutting the tissue. The abdomen
is cut with the blunt end of blunt/sharp scissors from above the bladder to just below the
diaphragm (rib cage). Care must be taken not to cut the abdomen or internal organs.

5. The incision is extended into horizontally flaps on both left and right to expose the liver and
intestines. The internal organ should be handled gingerly, especially the liver which is soft and
easily damaged.
Liver Preparation


6. The intestines are carefully moved to the left side of the animal, exposing the liver and surrounding
vasculature.

7. The vena cava, portal vein, mesenteric veins and arteries and bile duct are located.

8. Using a curved needle, non-cutting preferred and 00 or smaller silk suture, a suture is passed beneath
the portal vein, near the liver and past any branches. A second suture is passed beneath the vens cava
distal to this first suture(~5-10 mm). Note that the cannula will be inserted between the two sutures,
moved forward towards the liver and its tip secured by the suture closest to the liver. The distal suture
will be used to occlude portal vein blood flow.

9 A suture is passed beneath the vena cava, above the right renal vein. A suture is placed beneath the
mesenteric vein.

10. A suture is passed beneath the bile duct, the bile cannula is inserted and the cannula secured via
suture. (bile duct cannula is PE 10 tubing with cuff).

11. The appropriate sized portal vein cannula is selected by comparing the cannula tip to vein diameter.
The cannula is then placed on the end of the perfusion line and the line flushed by opening the stopcock
to clear air bubbles. The stopcock to the line is then closed. The pump should be cycling to permit
perfusion to occur as soon as the stopcock is opened.
Liver Preparation



12. Heparin (1000 units) is injected into the tail vein or into the vena cava, below the renal vein.

13. Note that at this point the experimenter must work efficiently so that the liver is not ischemic; total
time without blood flow to the liver should be less than 2 minutes from steps 14-17.

14. The distal portal vein suture is tied to occlude blow flow, the vessel then nicked to permit insertion of
the cannula, the cannula tip slid into the vein past the first suture and the cannula secured using the first
suture. Blood will back flush into the cannula.

15. The stopcock is opened to allow a small amount of fluid (<5ml) into the liver. The liver should not be
over expanded.

16. The vena cava is immediately cut below the suture loop to permit blood to drain out.

17. The stopcock is then re-opened and the liver perfused from the reservoir. A successful perfusion will
have the liver eventually blanching to an even beige color, without spotches or mottling.

18. The mesenteric veins are tied off with the previously placed suture.

19. The chest cavity is opened midline using scissors and the heart exposed. The atrial- thoracic vena
cava cannula is inserted through a cut placed in the right atria into the thoracic vena cava and secured
with suture. The cannula should fully dilated the vena cava to reduce backpressure.
Liver Preparation

20. The vena cava suture below the liver is secured and the stopcock to the atrial-thoracic vena cava
cannula is opened. This then forces fluid to exit from the liver into the atrial-thoracic vena cava cannula.

21. The lines to the bile duct, portal vein and vena cava are held in position
The liver is removed from the donor by supporting the diaphragm at the midline with forceps and
carefully cutting the diaphragm around the rib cage, first on side and then the other. Any attachment of
the liver to surrounding tissue are carefully removed.

22. The three lines are supported and the liver is then transferred to the organ chamber.

23. As indicators of a successful perfusion, liver flow should be at least 4 ml/minute/gram liver (for
perfusates without blood or other oxygen carriers) and bile flow 1-2 micro liters minute.
Post Experimental Clean Up

Glassware Maintenance & Post Experimental Cleanup
Post Experimental Cleanup
After the experiment has been completed, the experimenter should take care to scrupulously clean the
equipment. It is important to remember that the solutions that can sustain the heart and muscle will also provide
excellent media for bacteria. The cleaning procedures will be dependent upon the types of chemicals and
biological materials that are being used, the types of measurements that are being made and what substances
can interfere with those measurements and the frequency of the use of the equipment and number of operators
involved. Non-phosphate soaps are preferred, since insoluble phosphates can form from calcium and
magnesium in physiological salt solutions. Note that bactericidal soaps may contain iodine or other materials
which can affect isolated tissues and cells. Cleaning supplies and equipment, such as brushes, should be used
only for cleaning this glassware and not used for other lab cleaning procedures. Questions and procedures
noted here should be adjusted in accordance with your licensed procedures and the recommendations of your
safety personnel.
Post Experimental Cleanup


Shared equipment is the most difficult to maintain properly. In order to maintain equipment properly, it is generally
best (1) to assign the maintenance or the oversight of the equipment to one individual, who will monitor
equipment and maintain cleaning supplies (2) to have written protocols posted with the equipment (3) to have a
logbook where cleaning dates, as well as notification of problems, suggestions, etc., can be recorded.
Often overlooked as a source of contamination is the water circulator supply. This should be kept clean and the
bath rinsed and solution changed to reduce precipitate build up. Covering equipment to reduce air borne
contamination from microbes and spores is useful. Note that when baths are used intermittently, the lack of
frequent cleaning and the lack of solutions rinsing out bacteria that are deposited in the tubing may result in a
contamination problem when the system is finally used. A convenient rule of thumb for testing for contamination
in preparations that you have found reliable is that two consecutive experimental failures that are not explained
by an obviously damaged sample, poor surgical or dissection techniques or solution problems may be caused by
bath contamination.
Glassware
Much of the Radnoti apparatus is borosilicate glass, which can be cleaned with a wide range of soaps, ethyl
alcohol, dilute HCl or HNO3 (0.1 M) or other solvents. Extensive flushing with distilled, deionized water to remove
all traces of the cleaning agents and salts is recommended. Large glassware, such as reservoirs or assemblies
can be flushed in place, but care must be taken to thoroughly clean aerators, stopcocks and associated parts.
Aerators should be blown dry using gas or air at the final water rinse. If acid is used, the runoff water should not
be more acidic than the normal water pH. As with the use of any chemicals, proper protective gear and training
are essential to reduce personnel hazards and experimental and environmental contamination. Heated acid or
chromic acid is generally not recommended due to personnel hazards and possible heavy metal contamination of
the system.
If very lipohilic substances (prostaglandins, ionophores, certain dyes, etc.) are used, rinses with ethyl alcohol or
the most appropriate organic solvent can be used first, but this will necessitate thorough cleaning afterward to
remove any traces of the organic solvent.
Post Experimental Cleanup


Use of toxins, biohazardous materials and radiochemicals can present considerable complications to a
generalized cleaning procedure. Having an apparatus and a contained area dedicated to these procedures
reduces problems. Diluted bleach can be used on glassware, but must be rinsed extensively. The use of
disposable tubing and stopcocks will assist in cleanup, as will scheduling a run of these procedures, rather
than intermittent experiments, if non-dedicated equipment must be used. Glassware can be sterilized but all
fixtures, such as aerators, stopcocks caps, etc., should be removed prior to sterilization.
The glass aerators can be cleaned with water, or dilute acid if clogged. The use of water or gas under high
pressure can result in damage to the glassware and personnel and therefore is not recommended. After a
general soap and water rinse to remove soluble materials, cleaning with 0.1M HCl or 0.1 M HNO3 for several
hours or overnight, followed by an extensive water rinse, will usually remove most contaminants. If this does
not work, 1 M acid can be tried. Because the glass frit filaments are thin, high concentrations of acids, or
especially alkalis, can destroy them and are not recommended.
Non-glass items
Initial cleaning of non-glass items should be with aqueous soap solutions. Depending upon the chemical
resistance of the materials, the use of other solvents, cleaning procedures or sterilization may be possible.
Areas and items to be especially well cleaned are the aerator, tubing, syringe ports, cannulae, pressure
transducer fittings, septa, balloon and other catheters and electrodes (oxygen, pacing, ion selective, etc.).
Tubing should be inspected at the pump head for wear. Note that the interior of tubing can gradually be
roughened during use and the abraded areas will form sites for bacterial growth. Tubing should be a high
grade with low plasticizer leaching. Note that silicone tubing is very permeant to gases, so it should not be
generally used to transport gassed solutions.

Weitere ähnliche Inhalte

Was ist angesagt?

Was ist angesagt? (20)

Pneumatic valves
Pneumatic valvesPneumatic valves
Pneumatic valves
 
Hydraulic Valves (Valves) (Pneumatics also)
Hydraulic Valves (Valves) (Pneumatics also)Hydraulic Valves (Valves) (Pneumatics also)
Hydraulic Valves (Valves) (Pneumatics also)
 
Well control manual_bharat_petro_resource (1)
Well control manual_bharat_petro_resource (1)Well control manual_bharat_petro_resource (1)
Well control manual_bharat_petro_resource (1)
 
Presentation on solenoid valve
Presentation on solenoid valvePresentation on solenoid valve
Presentation on solenoid valve
 
Hydraulic valves
Hydraulic valvesHydraulic valves
Hydraulic valves
 
Control valve
Control valveControl valve
Control valve
 
Well control system
Well control systemWell control system
Well control system
 
Dc valve
Dc  valve Dc  valve
Dc valve
 
Control valve presentation
Control valve presentationControl valve presentation
Control valve presentation
 
Valve characteristics
Valve characteristicsValve characteristics
Valve characteristics
 
Marine engineering valve and cock
Marine engineering valve and cockMarine engineering valve and cock
Marine engineering valve and cock
 
Basic Control Valve Sizing and Selection
Basic Control Valve Sizing and SelectionBasic Control Valve Sizing and Selection
Basic Control Valve Sizing and Selection
 
Making a Hydraulic Ram Pump from Simple Plumbing Parts
Making a Hydraulic Ram Pump from Simple Plumbing PartsMaking a Hydraulic Ram Pump from Simple Plumbing Parts
Making a Hydraulic Ram Pump from Simple Plumbing Parts
 
Well control equipment
Well control equipmentWell control equipment
Well control equipment
 
Pressure control valves
Pressure control valvesPressure control valves
Pressure control valves
 
Pneumatics: Shuttle, Twin pressure, Quick Exhaust, Time Delay, FRL
Pneumatics: Shuttle, Twin pressure, Quick Exhaust, Time Delay, FRLPneumatics: Shuttle, Twin pressure, Quick Exhaust, Time Delay, FRL
Pneumatics: Shuttle, Twin pressure, Quick Exhaust, Time Delay, FRL
 
Control valve
Control valve Control valve
Control valve
 
Control Valves
Control ValvesControl Valves
Control Valves
 
Dk well control
Dk well controlDk well control
Dk well control
 
Control Valves And Actuators
Control Valves And ActuatorsControl Valves And Actuators
Control Valves And Actuators
 

Ähnlich wie Liver presentation

HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...
Vandana Devesh Sharma
 
Minimum flow system
Minimum flow systemMinimum flow system
Minimum flow system
taylorpo
 

Ähnlich wie Liver presentation (20)

Phacodynamics
PhacodynamicsPhacodynamics
Phacodynamics
 
Phacoemulsification part 2- Phacodynamics
Phacoemulsification part 2- PhacodynamicsPhacoemulsification part 2- Phacodynamics
Phacoemulsification part 2- Phacodynamics
 
Operation of Centrifugal Pumps
Operation of Centrifugal PumpsOperation of Centrifugal Pumps
Operation of Centrifugal Pumps
 
Phacodynamics
PhacodynamicsPhacodynamics
Phacodynamics
 
PUMPS+-
PUMPS+-PUMPS+-
PUMPS+-
 
3 valve shafts pneumatics and hydraulics
3 valve shafts pneumatics and hydraulics3 valve shafts pneumatics and hydraulics
3 valve shafts pneumatics and hydraulics
 
Phacodynamics in Phaco.pptx
Phacodynamics in Phaco.pptxPhacodynamics in Phaco.pptx
Phacodynamics in Phaco.pptx
 
780129
780129780129
780129
 
780129
780129780129
780129
 
Training materials for students final r4-uf-ro-1
Training materials for students final r4-uf-ro-1Training materials for students final r4-uf-ro-1
Training materials for students final r4-uf-ro-1
 
عرض تقديمي1.pptx
عرض تقديمي1.pptxعرض تقديمي1.pptx
عرض تقديمي1.pptx
 
Pumps
PumpsPumps
Pumps
 
Valve type-modi-11
Valve type-modi-11Valve type-modi-11
Valve type-modi-11
 
steam_blowing.pdf
steam_blowing.pdfsteam_blowing.pdf
steam_blowing.pdf
 
BREATHING_CIRCUITS POTC 2022 march.pptx
BREATHING_CIRCUITS POTC 2022 march.pptxBREATHING_CIRCUITS POTC 2022 march.pptx
BREATHING_CIRCUITS POTC 2022 march.pptx
 
hplc.pptx
hplc.pptxhplc.pptx
hplc.pptx
 
HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...
 
Minimum flow system
Minimum flow systemMinimum flow system
Minimum flow system
 
Phaco 3
Phaco 3Phaco 3
Phaco 3
 
HPLC_basics.ppt
HPLC_basics.pptHPLC_basics.ppt
HPLC_basics.ppt
 

Mehr von Radnoti

Working heart system set up 120101
Working heart system set up   120101Working heart system set up   120101
Working heart system set up 120101
Radnoti
 
Working Heart System Set Up - 120101
Working Heart System Set Up - 120101 Working Heart System Set Up - 120101
Working Heart System Set Up - 120101
Radnoti
 
Instrumentation of Radnoti Working Heart - Langendorff System
Instrumentation of Radnoti Working Heart - Langendorff SystemInstrumentation of Radnoti Working Heart - Langendorff System
Instrumentation of Radnoti Working Heart - Langendorff System
Radnoti
 
Radnoti Working Heart - Langendorff Application Notes
Radnoti Working Heart - Langendorff Application NotesRadnoti Working Heart - Langendorff Application Notes
Radnoti Working Heart - Langendorff Application Notes
Radnoti
 

Mehr von Radnoti (6)

Working heart system set up 120101
Working heart system set up   120101Working heart system set up   120101
Working heart system set up 120101
 
Working Heart System Set Up - 120101
Working Heart System Set Up - 120101 Working Heart System Set Up - 120101
Working Heart System Set Up - 120101
 
Instrumentation of Radnoti Working Heart - Langendorff System
Instrumentation of Radnoti Working Heart - Langendorff SystemInstrumentation of Radnoti Working Heart - Langendorff System
Instrumentation of Radnoti Working Heart - Langendorff System
 
Radnoti Working Heart - Langendorff Application Notes
Radnoti Working Heart - Langendorff Application NotesRadnoti Working Heart - Langendorff Application Notes
Radnoti Working Heart - Langendorff Application Notes
 
Langendorff System - 120102
Langendorff  System - 120102 Langendorff  System - 120102
Langendorff System - 120102
 
Working Heart System - 120101
Working Heart System - 120101Working Heart System - 120101
Working Heart System - 120101
 

Kürzlich hochgeladen

Challenges and Opportunities: A Qualitative Study on Tax Compliance in Pakistan
Challenges and Opportunities: A Qualitative Study on Tax Compliance in PakistanChallenges and Opportunities: A Qualitative Study on Tax Compliance in Pakistan
Challenges and Opportunities: A Qualitative Study on Tax Compliance in Pakistan
vineshkumarsajnani12
 
Al Mizhar Dubai Escorts +971561403006 Escorts Service In Al Mizhar
Al Mizhar Dubai Escorts +971561403006 Escorts Service In Al MizharAl Mizhar Dubai Escorts +971561403006 Escorts Service In Al Mizhar
Al Mizhar Dubai Escorts +971561403006 Escorts Service In Al Mizhar
allensay1
 

Kürzlich hochgeladen (20)

Falcon Invoice Discounting: Empowering Your Business Growth
Falcon Invoice Discounting: Empowering Your Business GrowthFalcon Invoice Discounting: Empowering Your Business Growth
Falcon Invoice Discounting: Empowering Your Business Growth
 
Putting the SPARK into Virtual Training.pptx
Putting the SPARK into Virtual Training.pptxPutting the SPARK into Virtual Training.pptx
Putting the SPARK into Virtual Training.pptx
 
Paradip CALL GIRL❤7091819311❤CALL GIRLS IN ESCORT SERVICE WE ARE PROVIDING
Paradip CALL GIRL❤7091819311❤CALL GIRLS IN ESCORT SERVICE WE ARE PROVIDINGParadip CALL GIRL❤7091819311❤CALL GIRLS IN ESCORT SERVICE WE ARE PROVIDING
Paradip CALL GIRL❤7091819311❤CALL GIRLS IN ESCORT SERVICE WE ARE PROVIDING
 
Berhampur CALL GIRL❤7091819311❤CALL GIRLS IN ESCORT SERVICE WE ARE PROVIDING
Berhampur CALL GIRL❤7091819311❤CALL GIRLS IN ESCORT SERVICE WE ARE PROVIDINGBerhampur CALL GIRL❤7091819311❤CALL GIRLS IN ESCORT SERVICE WE ARE PROVIDING
Berhampur CALL GIRL❤7091819311❤CALL GIRLS IN ESCORT SERVICE WE ARE PROVIDING
 
Pre Engineered Building Manufacturers Hyderabad.pptx
Pre Engineered  Building Manufacturers Hyderabad.pptxPre Engineered  Building Manufacturers Hyderabad.pptx
Pre Engineered Building Manufacturers Hyderabad.pptx
 
Lundin Gold - Q1 2024 Conference Call Presentation (Revised)
Lundin Gold - Q1 2024 Conference Call Presentation (Revised)Lundin Gold - Q1 2024 Conference Call Presentation (Revised)
Lundin Gold - Q1 2024 Conference Call Presentation (Revised)
 
Berhampur 70918*19311 CALL GIRLS IN ESCORT SERVICE WE ARE PROVIDING
Berhampur 70918*19311 CALL GIRLS IN ESCORT SERVICE WE ARE PROVIDINGBerhampur 70918*19311 CALL GIRLS IN ESCORT SERVICE WE ARE PROVIDING
Berhampur 70918*19311 CALL GIRLS IN ESCORT SERVICE WE ARE PROVIDING
 
Call 7737669865 Vadodara Call Girls Service at your Door Step Available All Time
Call 7737669865 Vadodara Call Girls Service at your Door Step Available All TimeCall 7737669865 Vadodara Call Girls Service at your Door Step Available All Time
Call 7737669865 Vadodara Call Girls Service at your Door Step Available All Time
 
UAE Bur Dubai Call Girls ☏ 0564401582 Call Girl in Bur Dubai
UAE Bur Dubai Call Girls ☏ 0564401582 Call Girl in Bur DubaiUAE Bur Dubai Call Girls ☏ 0564401582 Call Girl in Bur Dubai
UAE Bur Dubai Call Girls ☏ 0564401582 Call Girl in Bur Dubai
 
Durg CALL GIRL ❤ 82729*64427❤ CALL GIRLS IN durg ESCORTS
Durg CALL GIRL ❤ 82729*64427❤ CALL GIRLS IN durg ESCORTSDurg CALL GIRL ❤ 82729*64427❤ CALL GIRLS IN durg ESCORTS
Durg CALL GIRL ❤ 82729*64427❤ CALL GIRLS IN durg ESCORTS
 
Unveiling Falcon Invoice Discounting: Leading the Way as India's Premier Bill...
Unveiling Falcon Invoice Discounting: Leading the Way as India's Premier Bill...Unveiling Falcon Invoice Discounting: Leading the Way as India's Premier Bill...
Unveiling Falcon Invoice Discounting: Leading the Way as India's Premier Bill...
 
PARK STREET 💋 Call Girl 9827461493 Call Girls in Escort service book now
PARK STREET 💋 Call Girl 9827461493 Call Girls in  Escort service book nowPARK STREET 💋 Call Girl 9827461493 Call Girls in  Escort service book now
PARK STREET 💋 Call Girl 9827461493 Call Girls in Escort service book now
 
Arti Languages Pre Seed Teaser Deck 2024.pdf
Arti Languages Pre Seed Teaser Deck 2024.pdfArti Languages Pre Seed Teaser Deck 2024.pdf
Arti Languages Pre Seed Teaser Deck 2024.pdf
 
Falcon Invoice Discounting: Unlock Your Business Potential
Falcon Invoice Discounting: Unlock Your Business PotentialFalcon Invoice Discounting: Unlock Your Business Potential
Falcon Invoice Discounting: Unlock Your Business Potential
 
Marel Q1 2024 Investor Presentation from May 8, 2024
Marel Q1 2024 Investor Presentation from May 8, 2024Marel Q1 2024 Investor Presentation from May 8, 2024
Marel Q1 2024 Investor Presentation from May 8, 2024
 
Berhampur Call Girl Just Call 8084732287 Top Class Call Girl Service Available
Berhampur Call Girl Just Call 8084732287 Top Class Call Girl Service AvailableBerhampur Call Girl Just Call 8084732287 Top Class Call Girl Service Available
Berhampur Call Girl Just Call 8084732287 Top Class Call Girl Service Available
 
Berhampur 70918*19311 CALL GIRLS IN ESCORT SERVICE WE ARE PROVIDING
Berhampur 70918*19311 CALL GIRLS IN ESCORT SERVICE WE ARE PROVIDINGBerhampur 70918*19311 CALL GIRLS IN ESCORT SERVICE WE ARE PROVIDING
Berhampur 70918*19311 CALL GIRLS IN ESCORT SERVICE WE ARE PROVIDING
 
Katrina Personal Brand Project and portfolio 1
Katrina Personal Brand Project and portfolio 1Katrina Personal Brand Project and portfolio 1
Katrina Personal Brand Project and portfolio 1
 
Challenges and Opportunities: A Qualitative Study on Tax Compliance in Pakistan
Challenges and Opportunities: A Qualitative Study on Tax Compliance in PakistanChallenges and Opportunities: A Qualitative Study on Tax Compliance in Pakistan
Challenges and Opportunities: A Qualitative Study on Tax Compliance in Pakistan
 
Al Mizhar Dubai Escorts +971561403006 Escorts Service In Al Mizhar
Al Mizhar Dubai Escorts +971561403006 Escorts Service In Al MizharAl Mizhar Dubai Escorts +971561403006 Escorts Service In Al Mizhar
Al Mizhar Dubai Escorts +971561403006 Escorts Service In Al Mizhar
 

Liver presentation

  • 4. Key system components: •Primary reservoir •Secondary reservoir
  • 5. Key system components: •Primary reservoir •Secondary reservoir •Membrane Oxygenator
  • 6. Key system components: •Primary reservoir •Secondary reservoir •Membrane Oxygenator •Bubble Trap Compliance Chamber
  • 7. Key system components: •Primary reservoir •Secondary reservoir •Membrane Oxygenator •Bubble Trap Compliance Chamber •Inflow Manifold
  • 8. Key system components: •Primary reservoir •Secondary reservoir •Membrane Oxygenator •Bubble Trap Compliance Chamber •Inflow Manifold •Out Flow Manifold
  • 9. Key system components: •Out Flow Bubble Trap
  • 10. Key system components: •Out Flow Bubble Trap •Peristaltic Pump
  • 11. Key system components: •Out Flow Bubble Trap •Peristaltic Pump •Liver Chamber
  • 12. Key system components: •Out Flow Bubble Trap •Peristaltic Pump •Liver Chamber •3-Way Out Flow Manifold
  • 13. Key system components: •Out Flow Bubble Trap •Peristaltic Pump •Liver Chamber •3-Way Out Flow Manifold •3-Way Overflow Manifold
  • 15. Flow Path: Perfusate is selected either from the primary reservoir
  • 16. Flow Path: Perfusate is selected either from the primary reservoir or the secondary reservoir
  • 17. Flow Path: Perfusate is selected either from the primary reservoir or the secondary reservoir using the three way stopcock located at the outlet of the secondary reservoir.
  • 18. Flow Path: Perfusate is selected either from the primary reservoir or the secondary reservoir using the three way stopcock located at the outlet of the secondary reservoir.
  • 19. Flow Path: Perfusate is selected either from the primary reservoir or the secondary reservoir using the three way stopcock located at the outlet of the secondary reservoir. The perfusate then travels to the peristaltic pump
  • 20. Flow Path: and is pumped through the membrane oxygenating chamber
  • 21. Flow Path: and up to the bubble trap compliance chamber
  • 22. Flow Path: and up to the bubble trap compliance chamber Tip: on initial priming or filling of the system you will most likely need to close the compliance port stopcock
  • 23. Flow Path: and up to the bubble trap compliance chamber Tip: on initial priming or filling of the system you will most likely need to close the compliance port stopcock and open the bubble trap vent stopcock
  • 24. Flow Path: and up to the bubble trap compliance chamber Tip: on initial priming or filling of the system you will most likely need to close the compliance port stopcock and open the bubble trap vent stopcock so that the trap will have the opportunity to fill
  • 25. Flow Path: and up to the bubble trap compliance chamber When running in constant pressure mode, perfusate flow from the pump that is greater than the flow rate of the organ , will exit the bubble trap via the compliance port.
  • 26. Flow Path: and up to the bubble trap compliance chamber Generally you can leave this port open after the system has been primed.
  • 27. Flow Path: and up to the bubble trap compliance chamber Generally you can leave this port open after the system has been primed. Over flow exiting the compliance port is directed to the overflow manifold.
  • 28. Flow Path: and up to the bubble trap compliance chamber The overflow manifold will allow you to select where the overflowing perfusate is to be directed.
  • 29. Flow Path: and up to the bubble trap compliance chamber Valve 1 directs flow back to the primary reservoir.
  • 30. Flow Path: and up to the bubble trap compliance chamber Valve 1 directs flow back to the primary reservoir. Valve 2 directs flow back to the secondary reservoir.
  • 31. Flow Path: and up to the bubble trap compliance chamber Valve 1 directs flow back to the primary reservoir. Valve 2 directs flow back to the secondary reservoir. Valve 3 directs flow Out to Waste.
  • 32. Flow Path: and up to the bubble trap compliance chamber the out flow of the bubble trap then flows to the inflow manifold.
  • 33. Flow Path: and up to the bubble trap compliance chamber the out flow of the bubble trap then flows to the inflow manifold. Tip: When initially priming or flushing the system, the inflow and outflow cannulae will have to be coupled. This can be done by using a small section of tygon tubing and pushing the cannulae tips in either side.
  • 34. Flow Path: The perfusate then passes through the cannulated organ (or tygon coupler when priming or flushing) and into the outflow manifold.
  • 35. Flow Path: Perfusate then flows out to through the flow meter ( if so equipped) to the outflow bubble trap.
  • 36. Flow Path: Perfusate then flows out to through the flow meter ( if so equipped) to the outflow bubble trap. NOTE: The vent stopcock and the out flow stopcock of the bubble trap should be in the closed position.
  • 37. Flow Path: Perfusate then flows out to through the flow meter ( if so equipped) to the outflow bubble trap. The perfusate then is drawn be the second peristaltic pump head and pushed to the three way outflow manifold.
  • 38. Flow Path: At this point the perfusate can be directed to : 1. The primary reservoir
  • 39. Flow Path: At this point the perfusate can be directed to : 1. The primary reservoir 2. The secondary reservoir
  • 40. Flow Path: At this point the perfusate can be directed to : 1. The primary reservoir 2. The secondary reservoir 3. Waste Sink or collection flask
  • 41. Flow Path: Example: If you are flushing the system close valve 1
  • 42. Flow Path: Example: If you are flushing the system close valve 1 and 2
  • 43. Flow Path: Example: If you are flushing the system close valve 1 and 2 Leaving the valve open to waste will direct the flow out to waste.
  • 44. Flow Path: ALERT! SPECIAL NOTE: when flushing the organ you will want to protect components that may be sensitive to the initial effluent by diverting flow around the component.
  • 45. Flow Path: SPECIAL NOTE: when flushing the organ you will want to protect components that may be sensitive to the initial effluent by diverting flow around the component.
  • 46. Flow Path: SPECIAL NOTE: In this diagram the flow meter would need to be protected.
  • 47. Flow Path: SPECIAL NOTE: In this diagram the flow meter would need to be protected. This is done by using the three way stopcock on the outflow manifold to bypass the flow meter.
  • 48. Flow Path: SPECIAL NOTE: In this diagram the flow meter would need to be protected. This is done by using the three way stopcock on the outflow manifold to bypass the flow meter.
  • 50. Flow Path: Example: If the organ is flushed and you wish to have a recirculating path...
  • 51. Flow Path: Example: If the organ is flushed and you wish to have a recirculating path... In this case back to the primary reservoir...
  • 52. Flow Path: Example: If the organ is flushed and you wish to have a recirculating path... In this case back to the primary reservoir… Close valve 3 (waste)
  • 53. Flow Path: Example: If the organ is flushed and you wish to have a recirculating path... In this case back to the primary reservoir… Close valve 3 (waste) Close valve 2
  • 54. Flow Path: Example: If the organ is flushed and you wish to have a recirculating path... In this case back to the primary reservoir… Close valve 3 (waste) Close valve 2. Open valve 1
  • 55. Flow Path: Example: If the organ is flushed and you wish to have a recirculating path... Or To recirculate back to the secondary reservoir…
  • 56. Flow Path: Example: If the organ is flushed and you wish to have a recirculating path... Or to recirculate back to the secondary reservoir… Close valve 1
  • 57. Flow Path: Example: If the organ is flushed and you wish to have a recirculating path... Or to recirculate back to the secondary reservoir… Close valve 1 Open valve 2
  • 58. Flow Path: Example: If the organ is flushed and you wish to have a recirculating path... Or to recirculate back to the secondary reservoir… Close valve 1 Open valve 2 Close valve 3 (waste)
  • 60. INITIAL START UP ♦Fill the primary reservoir with buffer
  • 61. INITIAL START UP Make sure reservoir ♦Fill the primary selection stopcock is in the off position reservoir with buffer prior to filling.
  • 62. INITIAL START UP WARNING: ALWAYS MAKE SURE THERE IS AN OPENING TO ♦Fill the primary ATMOSPHERE ON THE PRIMARY reservoir with buffer RESERVOIR. IF FOR ANY REASON GAS PREASURE IS ALLOWED TO ♦Turn on and adjust BUILD IN THE RESERVOIR RISK OF gas from tank SERIOUS INJURY OR FATALITY MAY OCCUR. From tank
  • 63. INITIAL START UP ♦Fill the primary Note: reservoir with buffer Adjust gas so that ♦Turn on and adjust a gentle stream of gas from tank bubbles flows from the gas dispersion frit in the primary reservoir. This will oxygenate the buffer and serve as a visual representation that gas is flowing From tank through the Membrane Oxygenator.
  • 64. INITIAL START UP ♦Fill the primary reservoir with buffer ♦Turn on and adjust gas from tank ♦Open compliance port stopcock.
  • 65. INITIAL START UP ♦Fill the primary reservoir with buffer ♦Turn on and adjust gas from tank ♦Open compliance port stopcock. ♦Close valve 1 and 2 on three way out flow manifold.
  • 66. INITIAL START UP ♦Fill the primary reservoir with buffer ♦Turn on and adjust gas from tank ♦Open compliance port stopcock. ♦Close valve 1 and 2 on three way out flow manifold. ♦Open valve 3 (waste) on three way out flow manifold
  • 67. INITIAL START UP ♦Fill the primary reservoir with buffer ♦Turn on and adjust gas from tank ♦Open compliance port stopcock. ♦Close valve 1 and 2 on three way out flow manifold. ♦Open valve 3 (waste) on three way out flow manifold ♦Close valve 2 and 3 on three-way overflow manifold
  • 68. INITIAL START UP ♦Fill the primary reservoir with buffer ♦Turn on and adjust gas from tank ♦Open compliance port stopcock. ♦Close valve 1 and 2 on three way out flow manifold. ♦Open valve 3 (waste) on three way out flow manifold ♦Close valve 2 and 3 on three-way overflow manifold ♦Open valve 1 (primary reservoir return) on three-way overflow manifold.
  • 69. INITIAL START UP ♦Close the stopcocks for the inflow and outflow pressure transducers.
  • 70. INITIAL START UP ♦Close the stopcocks for the inflow and outflow pressure transducers. ♦Direct the outflow manifold three way stopcock to pass through the manifold.
  • 71. INITIAL START UP ♦Close the stopcocks for the inflow and outflow pressure transducers. ♦Direct the outflow manifold three way stopcock to pass through the manifold. ♦Close the drain and relief ports stopcocks on the out flow bubble trap.
  • 72. INITIAL START UP ♦Close the stopcocks for the inflow and outflow pressure transducers. ♦Direct the outflow manifold three way stopcock to pass through the manifold. ♦Close the drain and relief ports stopcocks on the out flow bubble trap. ♦Couple the Inflow and Outflow cannulae.
  • 73. INITIAL START UP ♦Close the stopcocks for the inflow and outflow pressure transducers. ♦Direct the outflow manifold three way stopcock to pass through the manifold. ♦Close the drain and relief ports stopcocks on the out flow bubble trap. ♦Couple the Inflow and Outflow cannulae. ♦Turn on peristaltic pump and turn reservoir selection three way stopcock to feed from primary ON reservoir.
  • 74. INITIAL START UP The system will now circulate buffer driven by the peristaltic pump. It will take several moments to purge air from the lines. You will most likely have to open and close the vent and overflow ports in the bubble trap compliance chamber temporarily to build up some perfusate. Once it approximately two thirds full, return the valves to their previous position. ON
  • 75. INITIAL START UP Once the system is primed, Turn off the peristaltic pump. Close stopcocks at the inflow and outflow manifold. This will trap the buffer in the lines and keep the system primed.
  • 77. Signal Generation Now is a good time to calibrate the pressure transducers and ion selective electrodes (if so equipped.)
  • 78. Signal Generation For the pressure Pressure Transducers transducers typically they will have to be filled with fluid and purged of bubbles.
  • 79. Signal Generation Pressure Transducers For the pressure transducers typically they will have to be filled with fluid and purged of bubbles. Set the three-way stopcock at the outflow pressure transducer so that flow is accepted from the outflow manifold and the purge port.
  • 80. Signal Generation Pressure Transducers For the pressure transducers typically they will have to be filled with fluid and purged of bubbles. Set the three-way stopcock at the outflow pressure transducer so that flow is accepted from the outflow Vent port manifold and the stopcock purge port. (not shown) Fill a disposable syringe with buffer. Open the transducer purge port (one way stopcock not shown.) and gently fill the pressure transducer dome causing air to be purged.
  • 81. Signal Generation Pressure Transducers Repeat the procedure with the inflow pressure transducer.
  • 82. Signal Generation Pressure Transducers The pressure transducers can be calibrated to your data acquisition at this time. Set the three-way stopcock controlling flow to the transducer to the closed position.
  • 83. Signal Generation Pressure Transducers The pressure transducers can be calibrated to your data acquisition at this time. Set the three-way stopcock controlling flow to the transducer to the closed position. Open the purge stopcock (not shown) on the transducer.
  • 84. Signal Generation Pressure Transducers The pressure transducers can be calibrated to your data acquisition at this time. Set the three-way stopcock controlling flow to the transducer to the closed position. Open the purge stopcock (not shown) on the transducer. This will be your zero pressure calibration point.
  • 85. Signal Generation Pressure Transducer Return the stopcocks to their previous position (accepting flow from the outflow manifold) and setting the purge port stopcock (not shown) to the closed position. This will be you're high pressure calibration point.
  • 86. Signal Generation Pressure Transducer Note: Return the The pressure head is stopcocks to their determined by the previous position elevation of the (accepting flow from bubble trap the outflow compliance manifold) and chamber. The setting the purge distance from the port stopcock (not chamber to the shown) to the closed pressure transducer position. can be calculated to a known pressure. This will be you're high pressure Distance in mm calibration point. divided by 13.6 = mm of mercury perfusion pressure should be 10-15 mm Hg (15-25cm above the liver.)
  • 87. Signal Generation Pressure Transducer Note: Repeat the The pressure head is procedure for the determined by the inflow manifold elevation of the pressure transducer. bubble trap compliance chamber. The distance from the chamber to the pressure transducer can be calculated to a known pressure. Distance in mm divided by 13.6 = mm of mercury.
  • 88. Signal Generation pH ELECTRODES
  • 89. Signal Generation pH Electrodes The pH electrodes are plugged directly from the mili volt adapter to the data acquisition interface. Ideally a pH electrode will output a voltage of 0mV in a pH 7 buffer. This can very by +/- 50mV based on the individual pH electrode. The Nernst equation tells us that a pH buffer 4 should be 160mV greater (more positive) than the reading in a pH 7. The reading in a pH10 should be -160mV less (more negative than that of a reading in a pH7 buffer. The 160 comes from being 3(7-4)*59(Nernst value at 20 degrees)=168mV for 100% slope and 160 is slightly less than 100% theoretical.
  • 90. Signal Generation O2 ELECTRODES
  • 91. Signal Generation O2 Electrodes Prior to calibration of the oxygen electrode, the electrode should be examined to insure that the electrode membrane is intact and the interior chamber is full of buffer. If the electrode requires maintenance please refer to the manufacturers instructions.
  • 92. Signal Generation O2 electrodes To obtain zero oxygen reading, the physiological buffer, placed in a vented calibration container should be gassed for at least 10 to 30 minutes with pure nitrogen at a rate of 3-6 bubbles per/sec to maintain a constant temperature and gas saturation.
  • 93. Signal Generation O2 Electrodes To obtain zero oxygen reading, the physiological buffer, placed in a vented calibration container should be gassed for at least 10 to 30 minutes with pure nitrogen at a rate of 3-6 bubbles per/sec to maintain a constant temperature and gas saturation. The electrode is then inserted into the calibration chamber and monitored until the reading is stabile. Once the reading has stabilized the reading can bet set to zero using the amplifier gain and offset adjustments. The acquisition software can use this as the low point calibration.
  • 94. Signal Generation O2 Electrodes The electrode is then removed and inserted into aerated container of buffer. In this case either the inflow or out flow manifold with perfusate being pumped through. After stabilizing the reading would be adjusted using the amplifier gain and offset based on the gas mixture (room air 21% oxygen, gas cylinder 100% oxygen, gas cylinder 95% oxygen etc…) This can be used as your high point calibration in the datacquisiton software.
  • 95. Signal Generation O2 Electrodes It is recommended that the procedure is repeated three times in order to insure the readings are stable and reproducible.
  • 96. Signal Generation Temperature
  • 97. Signal Generation Temperature Temperature probes are preset and will not require calibration. It is recommended that they be verified periodically measuring a known temperature such as your heater circulator bath and the result compared to the read out of the bath.
  • 98. Signal Generation Flow Meter
  • 99. Signal Generation Flow Meter The Flow Meter is preset and will not require calibration. It is recommended that it be verified periodically measuring a known flow and the read out verified.
  • 100. Liver Preparation 1. The animal is anesthetized and placed on its back; the anesthetic used may be a general anesthetic such as isoflurane or phenobarbital, depending upon the protocol requirements. Test for the depth of anesthesia via toe pinch, eye reflex, etc. 2. For best positioning, the limbs are retracted and secured with tape or string. 3. The abdomen is wiped with 70% alcohol; the abdomen can be shaved, although this is not necessary. 4. A midline incision is made by lifting the skin with forceps and cutting the tissue. The abdomen is cut with the blunt end of blunt/sharp scissors from above the bladder to just below the diaphragm (rib cage). Care must be taken not to cut the abdomen or internal organs. 5. The incision is extended into horizontally flaps on both left and right to expose the liver and intestines. The internal organ should be handled gingerly, especially the liver which is soft and easily damaged.
  • 101. Liver Preparation 6. The intestines are carefully moved to the left side of the animal, exposing the liver and surrounding vasculature. 7. The vena cava, portal vein, mesenteric veins and arteries and bile duct are located. 8. Using a curved needle, non-cutting preferred and 00 or smaller silk suture, a suture is passed beneath the portal vein, near the liver and past any branches. A second suture is passed beneath the vens cava distal to this first suture(~5-10 mm). Note that the cannula will be inserted between the two sutures, moved forward towards the liver and its tip secured by the suture closest to the liver. The distal suture will be used to occlude portal vein blood flow. 9 A suture is passed beneath the vena cava, above the right renal vein. A suture is placed beneath the mesenteric vein. 10. A suture is passed beneath the bile duct, the bile cannula is inserted and the cannula secured via suture. (bile duct cannula is PE 10 tubing with cuff). 11. The appropriate sized portal vein cannula is selected by comparing the cannula tip to vein diameter. The cannula is then placed on the end of the perfusion line and the line flushed by opening the stopcock to clear air bubbles. The stopcock to the line is then closed. The pump should be cycling to permit perfusion to occur as soon as the stopcock is opened.
  • 102. Liver Preparation 12. Heparin (1000 units) is injected into the tail vein or into the vena cava, below the renal vein. 13. Note that at this point the experimenter must work efficiently so that the liver is not ischemic; total time without blood flow to the liver should be less than 2 minutes from steps 14-17. 14. The distal portal vein suture is tied to occlude blow flow, the vessel then nicked to permit insertion of the cannula, the cannula tip slid into the vein past the first suture and the cannula secured using the first suture. Blood will back flush into the cannula. 15. The stopcock is opened to allow a small amount of fluid (<5ml) into the liver. The liver should not be over expanded. 16. The vena cava is immediately cut below the suture loop to permit blood to drain out. 17. The stopcock is then re-opened and the liver perfused from the reservoir. A successful perfusion will have the liver eventually blanching to an even beige color, without spotches or mottling. 18. The mesenteric veins are tied off with the previously placed suture. 19. The chest cavity is opened midline using scissors and the heart exposed. The atrial- thoracic vena cava cannula is inserted through a cut placed in the right atria into the thoracic vena cava and secured with suture. The cannula should fully dilated the vena cava to reduce backpressure.
  • 103. Liver Preparation 20. The vena cava suture below the liver is secured and the stopcock to the atrial-thoracic vena cava cannula is opened. This then forces fluid to exit from the liver into the atrial-thoracic vena cava cannula. 21. The lines to the bile duct, portal vein and vena cava are held in position The liver is removed from the donor by supporting the diaphragm at the midline with forceps and carefully cutting the diaphragm around the rib cage, first on side and then the other. Any attachment of the liver to surrounding tissue are carefully removed. 22. The three lines are supported and the liver is then transferred to the organ chamber. 23. As indicators of a successful perfusion, liver flow should be at least 4 ml/minute/gram liver (for perfusates without blood or other oxygen carriers) and bile flow 1-2 micro liters minute.
  • 104. Post Experimental Clean Up Glassware Maintenance & Post Experimental Cleanup Post Experimental Cleanup After the experiment has been completed, the experimenter should take care to scrupulously clean the equipment. It is important to remember that the solutions that can sustain the heart and muscle will also provide excellent media for bacteria. The cleaning procedures will be dependent upon the types of chemicals and biological materials that are being used, the types of measurements that are being made and what substances can interfere with those measurements and the frequency of the use of the equipment and number of operators involved. Non-phosphate soaps are preferred, since insoluble phosphates can form from calcium and magnesium in physiological salt solutions. Note that bactericidal soaps may contain iodine or other materials which can affect isolated tissues and cells. Cleaning supplies and equipment, such as brushes, should be used only for cleaning this glassware and not used for other lab cleaning procedures. Questions and procedures noted here should be adjusted in accordance with your licensed procedures and the recommendations of your safety personnel.
  • 105. Post Experimental Cleanup Shared equipment is the most difficult to maintain properly. In order to maintain equipment properly, it is generally best (1) to assign the maintenance or the oversight of the equipment to one individual, who will monitor equipment and maintain cleaning supplies (2) to have written protocols posted with the equipment (3) to have a logbook where cleaning dates, as well as notification of problems, suggestions, etc., can be recorded. Often overlooked as a source of contamination is the water circulator supply. This should be kept clean and the bath rinsed and solution changed to reduce precipitate build up. Covering equipment to reduce air borne contamination from microbes and spores is useful. Note that when baths are used intermittently, the lack of frequent cleaning and the lack of solutions rinsing out bacteria that are deposited in the tubing may result in a contamination problem when the system is finally used. A convenient rule of thumb for testing for contamination in preparations that you have found reliable is that two consecutive experimental failures that are not explained by an obviously damaged sample, poor surgical or dissection techniques or solution problems may be caused by bath contamination. Glassware Much of the Radnoti apparatus is borosilicate glass, which can be cleaned with a wide range of soaps, ethyl alcohol, dilute HCl or HNO3 (0.1 M) or other solvents. Extensive flushing with distilled, deionized water to remove all traces of the cleaning agents and salts is recommended. Large glassware, such as reservoirs or assemblies can be flushed in place, but care must be taken to thoroughly clean aerators, stopcocks and associated parts. Aerators should be blown dry using gas or air at the final water rinse. If acid is used, the runoff water should not be more acidic than the normal water pH. As with the use of any chemicals, proper protective gear and training are essential to reduce personnel hazards and experimental and environmental contamination. Heated acid or chromic acid is generally not recommended due to personnel hazards and possible heavy metal contamination of the system. If very lipohilic substances (prostaglandins, ionophores, certain dyes, etc.) are used, rinses with ethyl alcohol or the most appropriate organic solvent can be used first, but this will necessitate thorough cleaning afterward to remove any traces of the organic solvent.
  • 106. Post Experimental Cleanup Use of toxins, biohazardous materials and radiochemicals can present considerable complications to a generalized cleaning procedure. Having an apparatus and a contained area dedicated to these procedures reduces problems. Diluted bleach can be used on glassware, but must be rinsed extensively. The use of disposable tubing and stopcocks will assist in cleanup, as will scheduling a run of these procedures, rather than intermittent experiments, if non-dedicated equipment must be used. Glassware can be sterilized but all fixtures, such as aerators, stopcocks caps, etc., should be removed prior to sterilization. The glass aerators can be cleaned with water, or dilute acid if clogged. The use of water or gas under high pressure can result in damage to the glassware and personnel and therefore is not recommended. After a general soap and water rinse to remove soluble materials, cleaning with 0.1M HCl or 0.1 M HNO3 for several hours or overnight, followed by an extensive water rinse, will usually remove most contaminants. If this does not work, 1 M acid can be tried. Because the glass frit filaments are thin, high concentrations of acids, or especially alkalis, can destroy them and are not recommended. Non-glass items Initial cleaning of non-glass items should be with aqueous soap solutions. Depending upon the chemical resistance of the materials, the use of other solvents, cleaning procedures or sterilization may be possible. Areas and items to be especially well cleaned are the aerator, tubing, syringe ports, cannulae, pressure transducer fittings, septa, balloon and other catheters and electrodes (oxygen, pacing, ion selective, etc.). Tubing should be inspected at the pump head for wear. Note that the interior of tubing can gradually be roughened during use and the abraded areas will form sites for bacterial growth. Tubing should be a high grade with low plasticizer leaching. Note that silicone tubing is very permeant to gases, so it should not be generally used to transport gassed solutions.