2. Forma Light
Forma Light is an
innovative IPL Unit
developed to treat a range of
medical-aesthetic
applications
3. Forma Light Highlight
Full control over electro physical parameters such as
pulse duration, pulse wide and flunce energy .
Pre setting features based on the patient skin type
Treatments for all skin type
Innovative design including touch screen
Shortest pulse duration
4. Forma Light Highlight
System weight just 8.9kg
Light & ergonomic Applicator
Large tip for large area treatments
Multilingual support
The highest quality in competitive price
4 colors to choose from
6. Applicators
IPL F- HR 590nm
IPL F- HR 650nm
IPL F- SR 530nm
IPL F-AC 440 nm
IPL B- HR 590nm
IPL B - HR* 650nm
IPL HR 590nm
IPL HR* 650nm
IPL SR 530nm
Application
up to
pulses
up to
pulses
up to
pulses
up to
pulses
up to
pulses
up to
pulses
60,000 up to
120,000 pulses
60,000 up to
120,000 pulses
60,000 up to
120,000 pulses
1.6cm2
Hair Reduction
1.6cm2
Hair Reduction
1.6cm2
1.6cm2
Skin rejuvenation,
Pigmented/Vascular
Lesions
Acne
3 cm2
Hair Reduction
3 cm2
Hair Reduction
6.5cm2
Hair Reduction
6.5cm2
Hair Reduction
6.5cm2
Skin rejuvenation,
Pigmented lesions
8. Intense Pulse Light and
Thermal Relaxation Time - TRT
Represents the time taken for heated tissue to lose
50% of its heat through diffusion
If the light energy persists longer than TRT of the
target, heat is conduced to the surrounding tissue
leading to unwanted thermal injury. (Katsambas &
Lotti, 2003).
9. TRT in
Pigmented and Vascular lesions
Data & histological assessment of vessel injury strongly suggest
that pulse durations for ideal light treatment are in the
1 – 10ms region and depend on vessel diameter (Diedrickx, 1995).
Correct wavelength combine with correct TRT will achieve quick
coagulation of targeted chromophore (melanin/hemoglobin).
If TRT is exceeded, same clinical consistence is not achieved and
more treatments sessions are necessary.
Pulse wide choice is as important as that of the wavelength.
(X.Muller, Lasers for Ischemic Heart Disease,2000)
10. Forma Light – enjoy from all options
Innovation in Technology
Shortest pulse duration below 10 ms to in respect of Thermal
Relaxation Time parameters, and Long pulse for dark skin HR
11. Forma Light
Innovation in Technology
Fluence
=
[Power]
Area of Spot x Time (Pulse Duration)
(Energy Density)
Because of the short pulse reaches the highest intensity
13. B/F HR – Hair removal
Body and face hair removal.
2 modes pulse of work: single, double pulse.
Technical information:
Energy: up to 20 j/cm2.
Wavelength: 590/650 nm
Spot size: body 3cm²/6.5cm², face 1.6cm²
Pulse duration:
Single: from 3 to 10 milliseconds (for 3cm² and 1.6cm²)
Double: from 2 to 5 milliseconds (for 3cm² and 1.6cm²)
For the 6.5cm² 20 milliseconds and up
17. B/F SR –Skin Rejuvenation
Body and face Skin Rejuvenation and Fungus Treatment
2 modes of work: single pulse, double puls (single pulse
every 0.33 seconds).
Technical information:
Energy: up to 22 j/cm.
Wavelength: 530 nm
Spot size: body 3 cm²/6.5², face 1.6 cm²
Pulse duration:
Single: from 3 to 10 milliseconds (for 3cm² and 1.6cm²)
Double: from 2 to 5 milliseconds (for 3cm² and 1.6cm²)
For the 6.5cm² 20 milliseconds and up