Internal quality control:
what and how to change
Ferruccio Ceriotti
Servizio di Medicina di Laboratorio,
Ospedale San Raffaele, Milano
The work presented derives mainly from
the activity of the Working Group on
Analytical Quality of SIBioC and I have to
thank in particular:
– Duilio Brugnoni (current Chair of the WG)
– Sonia Mattioli
– Pasquale Iandolo
– Federica Braga
Ceriotti F. Milano, 18-11-2016 2
ISO 15189:2012
• 5.6 Ensuring quality of the examination results
• 5.6.2 Quality Control
• 5.6.2.1 General
– The laboratory shall design quality control procedures
that verify the attainment of the intended quality of
results
• 5.6.2.2 Quality control materials
– Quality control materials shall be periodically examined
with a frequency that is based on the stability of the
procedure and the risk of harm to the patient from an
erroneous result
Note 2. Use of independent third party control materials
should be considered, either instead of, or in addition to,
any control materials supplied by the …. manufacturer.
Ceriotti F. Milano, 18-11-2016 3
ISO 15189:2012
• 5.6.2.3 Quality control data
– The laboratory shall have a procedure to
prevent the release of patient results in the
event of quality control failure.
– When the quality control rules are violated
and indicated that the examination results are
likely to contain clinically significant errors, the
results shall be rejected ……
– Quality control results shall be reviewed at
regular intervals to detect trends ….
Ceriotti F. Milano, 18-11-2016 4
Braga F, Panteghini M. Clin Chim Acta 432 (2014) 55–61
Ceriotti F. Milano, 18-11-2016 5
Steps for IQC implementation
according CLSI C24-A4
1. Define the quality requirements
2. Select control materials
3. Determine target values and standard deviations
for quality control materials that represent stable
analytical performance
4. Set goals for quality control performance
5. Select a quality control strategy based on
performance goals
6. Design a quality control strategy for multiple
instruments
Ceriotti F. Milano, 18-11-2016 6
Ceriotti F. Milano, 18-11-2016 7
Implementing a Six Sigma quality system
Ceriotti F. Milano, 18-11-2016 8
Implementing a Six Sigma quality
system
• “Method validation and acceptable method
performance are prerequisite to the design and
implementation of SQC procedures. If
performance is NOT acceptable under stable
operating conditions, no amount of QC can
change or improve that performance. QC can
only monitor performance, and when properly
designed, alert analysts to the presence of
additional errors that occur because of unstable
performance.”
Westard JO, Westgard S. Ann Clin Biochem 2016;53:35-50
Ceriotti F. Milano, 18-11-2016 9
‘Westgard Sigma Rules’ for two
levels of controls
Westard JO, Westgard S. Ann Clin Biochem 2016;53:35-50
Ceriotti F. Milano, 18-11-2016 10
Conclusions: Our data from IQC do not
consistently demonstrate that the results from
clinical laboratories meet evidence-based
quality specifications. Therefore, we propose
that a graded scale of quality specifications
may be needed at this stage.
Ceriotti F. Milano, 18-11-2016 11
• “There are two possible reasons for the lack
of agreement between the proposed limits
and the routinely achieved analytical variation
by laboratories. First, currently used
technology is inherently insufficiently
robust to allow the achievement of a narrow
analytical variation regardless of the effort to
control the analytical process (i.e., creatinine
Jaffe method). Second, there is sub-optimal
control over the IQC process and a lack of
defined limits.”
Jassam et al. CCLM 2013;51:1579-84
Ceriotti F. Milano, 18-11-2016 12
Problems of the ICQ process
• Need to balance the metrological
complication and the practical simplicity
needed for adoption by medical laboratories
• Need to establish a direct link between the
performance characteristics of the method
and the QC rules
• Improve control on the bias component
• Need to demonstrate the “fitness for purpose”
Ceriotti F. Milano, 18-11-2016 13
Two components IQC
• Two independent components: one devoted
mainly to checking the alignment of the
analytical system and verification of the
consistency of declared traceability during
routine operations performed in accordance
with the manufacturer’s instructions (IQC
component I) and the other structured
particularly for estimating the measurement
uncertainty due to random effects (IQC
component II).
Panteghini M. Clin Chem Lab Med 2010;48(1):7–10
Ceriotti F. Milano, 18-11-2016 15
Internal QC, component I
• Aim: testing system alignment
• Materials: control materials supplied by
the system’s manufacturer with system
specific assigned values
• Use: acceptance/rejection of the
analytical run
• Rules: results within a stated acceptability
range, uncertainty based rules.
Ceriotti F. Milano, 18-11-2016 16
Internal QC, component II
• Aims: checking system stability (lot to lot
variations, possible drifts); providing data
for measurement uncertainty calculation
• Materials: third party control materials,
commutable, concentrations at clinical
decision limits
• Use: support the acceptance / rejection,
provide data for retrospective evaluation
• Rules: Westgard rules, uncertainty based
rules
Ceriotti F. Milano, 18-11-2016 17
Apply the uncertainty theory to IQC
• Estimated measurement uncertainty shall be taken
into account to prove the conformity or
nonconformity with the given specification.
• The complete measurement result, y’, is
represented by the measurement result Âą
measurement uncertainty (U).
• Conformity with a specification is proved when the
complete measurement result, y’, falls within the
zone defined by a Lower Specification Limit (LSL)
and an Upper Specification Limit (USL) (maximum
permissible error).
LSL ≤ y - U and y + U ≤ USL
Ceriotti F. Milano, 18-11-2016 18
The same conformity can be proved similarly when the measurement
result, y, falls within the zone of maximum permissible error reduced on
either side by the expanded measurement uncertainty.
LSL + U ≤ y ≤ USL - U
y
y’
U U
LSL USL
Ceriotti F. Milano, 18-11-2016 19
Quality control chart based with guard bands
based on measurement uncertainty
Ceriotti F. Milano, 18-11-2016 20
How to implement uncertainty
based IQC
• Define the acceptability range:
• Manufacturer
• Maximum acceptable TE based on biological variation
• Minimal performance specifications based on outcome
data or proposed by scientific societies
• Calculate the expanded uncertainty of your
procedure
• For this particular application only the random
component of uncertainty is relevant, so two times the
CV
Ceriotti F. Milano, 18-11-2016 21
How to implement uncertainty
based IQC
• Prepare the control chart reducing the
acceptability range on either side by
expanded measurement uncertainty
• Immediate understanding of the “process
capability”: in a “2 sigma” situation (acceptable
range Âą6%, CV=3%) the guard bands will cover all
the chart indicating that the performance of your
method is insufficient and there is no way to
guarantee the control of that level of quality
(change method or accept a lower quality
specification)
Ceriotti F. Milano, 18-11-2016 22
Problems with the presented
approach
• The Westgard’s criticisms is that “more [than
with the uncertainty based approach] specific guidance
on the selection of control rules and the
number of control measurements is provided
using the Sigma-related SQC planning
tools”.[1]
• The simulations Brugnoni et al.[2] indicate an
elevate number of “false alarms” in situation
of low sigma (around 3)
but:
Ceriotti F. Milano, 18-11-2016 23
[1] Ann Clin Biochem 2016 ;53:35-50
[2] Biochim Clin 2016;40SS:S83
• Violating a guard band does
not automatically mean to be
in error. The probability of
conformity, assuming a
symmetrical distribution, falls
to 50% when the
measurement result equals
the tolerance limit.
• Violating a guard band does not automatically mean
the presence of a systematic error, a second control
can be used to confirm or not this event
• The width of the “acceptance zone” automatically
prompts for a higher or lower level of control
Ceriotti F. Milano, 18-11-2016 24
LSL USL
Ceriotti F. Milano, 18-11-2016 25
Monte Carlo simulation
Comparison with the Westgard’s
strategy
• IQC results obtained during a 12 months period for glucose,
creatinine and total cholesterol were collected from 4 clinical
laboratories using 8 analyzers (4 Siemens Dimension Vista
1500, 2 Ortho Vitros 5600, 1 Siemens Advia 2400, 1
Beckman Coulter AU5800). For each measurand biological
variation derived TEa were used; Westgard rules were
selected according to the σ-metric calculated from a 6
months planning period and U was calculated from the
imprecision of the IQC results during the same period.
Based on these settings, the number of “out of control”
alarms occurred in the following 6 months with the
implemented Westgard’s rules were compared with the
number of QC results falling within the “alarm zones”.
Relationship between the alarm frequency and the DPMO
of the method derived from the σ value estimated during
the testing period.
6σ 5σ 4σ 3σ
Conclusions
• IQC plays a key role in the organization of a
clinical laboratory that wants to guarantee the
necessary analytical quality minimizing the risk of
delivering erroneous results
• To implement it correctly and effectively is not
easy, especially when thousands of results have
to be released every hour for dozens of different
tests like in a typical “core laboratory”
• The proposed approach simplifies the IQC
implementation without loosing any of the relevant
features that can be obtained with a correct
implementation of the classical approach
Ceriotti F. Milano, 18-11-2016 28
Ceriotti F. Milano, 18-11-2016 29

Weitere ähnliche Inhalte

Ähnlich wie Cerioti Internal quality control.pdf

ISO 15189 Accreditation Guide - Improving Laboratory Performance Through Qual...
ISO 15189 Accreditation Guide - Improving Laboratory Performance Through Qual...ISO 15189 Accreditation Guide - Improving Laboratory Performance Through Qual...
ISO 15189 Accreditation Guide - Improving Laboratory Performance Through Qual...Randox
 
App. Of Stat. Tools
App. Of Stat. ToolsApp. Of Stat. Tools
App. Of Stat. ToolsDenny Thayil
 
Harmonization of Laboratory Indicators, 09 03-2017
Harmonization of Laboratory Indicators, 09 03-2017Harmonization of Laboratory Indicators, 09 03-2017
Harmonization of Laboratory Indicators, 09 03-2017Ola Elgaddar
 
FDA Update: Inspections, Observations and Metrics - OMTEC 2017
FDA Update: Inspections, Observations and Metrics - OMTEC 2017FDA Update: Inspections, Observations and Metrics - OMTEC 2017
FDA Update: Inspections, Observations and Metrics - OMTEC 2017April Bright
 
Six system management.pptx in pharmaceutical industry
Six system management.pptx in pharmaceutical industrySix system management.pptx in pharmaceutical industry
Six system management.pptx in pharmaceutical industryvinitnai
 
The Importance of Third Party Controls
The Importance of Third Party ControlsThe Importance of Third Party Controls
The Importance of Third Party ControlsRandox
 
Pharmaceutical validation.pptx
Pharmaceutical validation.pptxPharmaceutical validation.pptx
Pharmaceutical validation.pptxArpithaNandakumar1
 
PET - Quality Assurance of PET Radiopharmaceutials
PET - Quality Assurance of PET RadiopharmaceutialsPET - Quality Assurance of PET Radiopharmaceutials
PET - Quality Assurance of PET Radiopharmaceutials@Saudi_nmc
 
boccardi_analitytical-procedures.pdf
boccardi_analitytical-procedures.pdfboccardi_analitytical-procedures.pdf
boccardi_analitytical-procedures.pdfssuser796efb
 
Six system inspection model pharmaceutical D-Astar.pptx
Six system inspection model pharmaceutical D-Astar.pptxSix system inspection model pharmaceutical D-Astar.pptx
Six system inspection model pharmaceutical D-Astar.pptxDevaPundkar
 
QMS Seminar.pptx
QMS Seminar.pptxQMS Seminar.pptx
QMS Seminar.pptxRohitKoli29
 
Presentation Process Validation A Lifecycle Approach .pdf
Presentation Process Validation A Lifecycle Approach .pdfPresentation Process Validation A Lifecycle Approach .pdf
Presentation Process Validation A Lifecycle Approach .pdfLalitaShukla4
 
Risk management tools
Risk management toolsRisk management tools
Risk management toolsDhruvi50
 
inspection , test and measurement
inspection , test and measurementinspection , test and measurement
inspection , test and measurementtanvikumbhar
 
Metrology & The Consequences of Bad Measurement Decisions
Metrology & The Consequences of Bad Measurement DecisionsMetrology & The Consequences of Bad Measurement Decisions
Metrology & The Consequences of Bad Measurement DecisionsRick Hogan
 
Validation (1).pptx
Validation (1).pptxValidation (1).pptx
Validation (1).pptxYunesalsayadi
 
Quality control in a virology laboratory.pdf
Quality control in a virology laboratory.pdfQuality control in a virology laboratory.pdf
Quality control in a virology laboratory.pdfsamwel18
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validationTausif Momin
 
PHARMACEUTICAL VALIADATION
PHARMACEUTICAL VALIADATIONPHARMACEUTICAL VALIADATION
PHARMACEUTICAL VALIADATIONYenda Manishankar
 

Ähnlich wie Cerioti Internal quality control.pdf (20)

Uncertainty (ASCP)
Uncertainty (ASCP)Uncertainty (ASCP)
Uncertainty (ASCP)
 
ISO 15189 Accreditation Guide - Improving Laboratory Performance Through Qual...
ISO 15189 Accreditation Guide - Improving Laboratory Performance Through Qual...ISO 15189 Accreditation Guide - Improving Laboratory Performance Through Qual...
ISO 15189 Accreditation Guide - Improving Laboratory Performance Through Qual...
 
App. Of Stat. Tools
App. Of Stat. ToolsApp. Of Stat. Tools
App. Of Stat. Tools
 
Harmonization of Laboratory Indicators, 09 03-2017
Harmonization of Laboratory Indicators, 09 03-2017Harmonization of Laboratory Indicators, 09 03-2017
Harmonization of Laboratory Indicators, 09 03-2017
 
FDA Update: Inspections, Observations and Metrics - OMTEC 2017
FDA Update: Inspections, Observations and Metrics - OMTEC 2017FDA Update: Inspections, Observations and Metrics - OMTEC 2017
FDA Update: Inspections, Observations and Metrics - OMTEC 2017
 
Six system management.pptx in pharmaceutical industry
Six system management.pptx in pharmaceutical industrySix system management.pptx in pharmaceutical industry
Six system management.pptx in pharmaceutical industry
 
The Importance of Third Party Controls
The Importance of Third Party ControlsThe Importance of Third Party Controls
The Importance of Third Party Controls
 
Pharmaceutical validation.pptx
Pharmaceutical validation.pptxPharmaceutical validation.pptx
Pharmaceutical validation.pptx
 
PET - Quality Assurance of PET Radiopharmaceutials
PET - Quality Assurance of PET RadiopharmaceutialsPET - Quality Assurance of PET Radiopharmaceutials
PET - Quality Assurance of PET Radiopharmaceutials
 
boccardi_analitytical-procedures.pdf
boccardi_analitytical-procedures.pdfboccardi_analitytical-procedures.pdf
boccardi_analitytical-procedures.pdf
 
Six system inspection model pharmaceutical D-Astar.pptx
Six system inspection model pharmaceutical D-Astar.pptxSix system inspection model pharmaceutical D-Astar.pptx
Six system inspection model pharmaceutical D-Astar.pptx
 
QMS Seminar.pptx
QMS Seminar.pptxQMS Seminar.pptx
QMS Seminar.pptx
 
Presentation Process Validation A Lifecycle Approach .pdf
Presentation Process Validation A Lifecycle Approach .pdfPresentation Process Validation A Lifecycle Approach .pdf
Presentation Process Validation A Lifecycle Approach .pdf
 
Risk management tools
Risk management toolsRisk management tools
Risk management tools
 
inspection , test and measurement
inspection , test and measurementinspection , test and measurement
inspection , test and measurement
 
Metrology & The Consequences of Bad Measurement Decisions
Metrology & The Consequences of Bad Measurement DecisionsMetrology & The Consequences of Bad Measurement Decisions
Metrology & The Consequences of Bad Measurement Decisions
 
Validation (1).pptx
Validation (1).pptxValidation (1).pptx
Validation (1).pptx
 
Quality control in a virology laboratory.pdf
Quality control in a virology laboratory.pdfQuality control in a virology laboratory.pdf
Quality control in a virology laboratory.pdf
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
PHARMACEUTICAL VALIADATION
PHARMACEUTICAL VALIADATIONPHARMACEUTICAL VALIADATION
PHARMACEUTICAL VALIADATION
 

KĂźrzlich hochgeladen

Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...narwatsonia7
 
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...narwatsonia7
 
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...saminamagar
 
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...narwatsonia7
 
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaCall Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaPooja Gupta
 
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceNehru place Escorts
 
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking ModelsMumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Modelssonalikaur4
 
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfHemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfMedicoseAcademics
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girlsnehamumbai
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...rajnisinghkjn
 
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbers
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbersBook Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbers
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbersnarwatsonia7
 
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safenarwatsonia7
 
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000aliya bhat
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Gabriel Guevara MD
 
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment BookingCall Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 

KĂźrzlich hochgeladen (20)

Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
 
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
 
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
 
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
 
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaCall Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
 
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
 
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking ModelsMumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
 
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfHemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
 
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
 
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
 
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbers
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbersBook Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbers
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbers
 
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
 
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Servicesauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
 
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024
 
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment BookingCall Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
 
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
 

Cerioti Internal quality control.pdf

  • 1. Internal quality control: what and how to change Ferruccio Ceriotti Servizio di Medicina di Laboratorio, Ospedale San Raffaele, Milano
  • 2. The work presented derives mainly from the activity of the Working Group on Analytical Quality of SIBioC and I have to thank in particular: – Duilio Brugnoni (current Chair of the WG) – Sonia Mattioli – Pasquale Iandolo – Federica Braga Ceriotti F. Milano, 18-11-2016 2
  • 3. ISO 15189:2012 • 5.6 Ensuring quality of the examination results • 5.6.2 Quality Control • 5.6.2.1 General – The laboratory shall design quality control procedures that verify the attainment of the intended quality of results • 5.6.2.2 Quality control materials – Quality control materials shall be periodically examined with a frequency that is based on the stability of the procedure and the risk of harm to the patient from an erroneous result Note 2. Use of independent third party control materials should be considered, either instead of, or in addition to, any control materials supplied by the …. manufacturer. Ceriotti F. Milano, 18-11-2016 3
  • 4. ISO 15189:2012 • 5.6.2.3 Quality control data – The laboratory shall have a procedure to prevent the release of patient results in the event of quality control failure. – When the quality control rules are violated and indicated that the examination results are likely to contain clinically significant errors, the results shall be rejected …… – Quality control results shall be reviewed at regular intervals to detect trends …. Ceriotti F. Milano, 18-11-2016 4
  • 5. Braga F, Panteghini M. Clin Chim Acta 432 (2014) 55–61 Ceriotti F. Milano, 18-11-2016 5
  • 6. Steps for IQC implementation according CLSI C24-A4 1. Define the quality requirements 2. Select control materials 3. Determine target values and standard deviations for quality control materials that represent stable analytical performance 4. Set goals for quality control performance 5. Select a quality control strategy based on performance goals 6. Design a quality control strategy for multiple instruments Ceriotti F. Milano, 18-11-2016 6
  • 7. Ceriotti F. Milano, 18-11-2016 7
  • 8. Implementing a Six Sigma quality system Ceriotti F. Milano, 18-11-2016 8
  • 9. Implementing a Six Sigma quality system • “Method validation and acceptable method performance are prerequisite to the design and implementation of SQC procedures. If performance is NOT acceptable under stable operating conditions, no amount of QC can change or improve that performance. QC can only monitor performance, and when properly designed, alert analysts to the presence of additional errors that occur because of unstable performance.” Westard JO, Westgard S. Ann Clin Biochem 2016;53:35-50 Ceriotti F. Milano, 18-11-2016 9
  • 10. ‘Westgard Sigma Rules’ for two levels of controls Westard JO, Westgard S. Ann Clin Biochem 2016;53:35-50 Ceriotti F. Milano, 18-11-2016 10
  • 11. Conclusions: Our data from IQC do not consistently demonstrate that the results from clinical laboratories meet evidence-based quality specifications. Therefore, we propose that a graded scale of quality specifications may be needed at this stage. Ceriotti F. Milano, 18-11-2016 11
  • 12. • “There are two possible reasons for the lack of agreement between the proposed limits and the routinely achieved analytical variation by laboratories. First, currently used technology is inherently insufficiently robust to allow the achievement of a narrow analytical variation regardless of the effort to control the analytical process (i.e., creatinine Jaffe method). Second, there is sub-optimal control over the IQC process and a lack of defined limits.” Jassam et al. CCLM 2013;51:1579-84 Ceriotti F. Milano, 18-11-2016 12
  • 13. Problems of the ICQ process • Need to balance the metrological complication and the practical simplicity needed for adoption by medical laboratories • Need to establish a direct link between the performance characteristics of the method and the QC rules • Improve control on the bias component • Need to demonstrate the “fitness for purpose” Ceriotti F. Milano, 18-11-2016 13
  • 14.
  • 15. Two components IQC • Two independent components: one devoted mainly to checking the alignment of the analytical system and verification of the consistency of declared traceability during routine operations performed in accordance with the manufacturer’s instructions (IQC component I) and the other structured particularly for estimating the measurement uncertainty due to random effects (IQC component II). Panteghini M. Clin Chem Lab Med 2010;48(1):7–10 Ceriotti F. Milano, 18-11-2016 15
  • 16. Internal QC, component I • Aim: testing system alignment • Materials: control materials supplied by the system’s manufacturer with system specific assigned values • Use: acceptance/rejection of the analytical run • Rules: results within a stated acceptability range, uncertainty based rules. Ceriotti F. Milano, 18-11-2016 16
  • 17. Internal QC, component II • Aims: checking system stability (lot to lot variations, possible drifts); providing data for measurement uncertainty calculation • Materials: third party control materials, commutable, concentrations at clinical decision limits • Use: support the acceptance / rejection, provide data for retrospective evaluation • Rules: Westgard rules, uncertainty based rules Ceriotti F. Milano, 18-11-2016 17
  • 18. Apply the uncertainty theory to IQC • Estimated measurement uncertainty shall be taken into account to prove the conformity or nonconformity with the given specification. • The complete measurement result, y’, is represented by the measurement result Âą measurement uncertainty (U). • Conformity with a specification is proved when the complete measurement result, y’, falls within the zone defined by a Lower Specification Limit (LSL) and an Upper Specification Limit (USL) (maximum permissible error). LSL ≤ y - U and y + U ≤ USL Ceriotti F. Milano, 18-11-2016 18
  • 19. The same conformity can be proved similarly when the measurement result, y, falls within the zone of maximum permissible error reduced on either side by the expanded measurement uncertainty. LSL + U ≤ y ≤ USL - U y y’ U U LSL USL Ceriotti F. Milano, 18-11-2016 19
  • 20. Quality control chart based with guard bands based on measurement uncertainty Ceriotti F. Milano, 18-11-2016 20
  • 21. How to implement uncertainty based IQC • Define the acceptability range: • Manufacturer • Maximum acceptable TE based on biological variation • Minimal performance specifications based on outcome data or proposed by scientific societies • Calculate the expanded uncertainty of your procedure • For this particular application only the random component of uncertainty is relevant, so two times the CV Ceriotti F. Milano, 18-11-2016 21
  • 22. How to implement uncertainty based IQC • Prepare the control chart reducing the acceptability range on either side by expanded measurement uncertainty • Immediate understanding of the “process capability”: in a “2 sigma” situation (acceptable range Âą6%, CV=3%) the guard bands will cover all the chart indicating that the performance of your method is insufficient and there is no way to guarantee the control of that level of quality (change method or accept a lower quality specification) Ceriotti F. Milano, 18-11-2016 22
  • 23. Problems with the presented approach • The Westgard’s criticisms is that “more [than with the uncertainty based approach] specific guidance on the selection of control rules and the number of control measurements is provided using the Sigma-related SQC planning tools”.[1] • The simulations Brugnoni et al.[2] indicate an elevate number of “false alarms” in situation of low sigma (around 3) but: Ceriotti F. Milano, 18-11-2016 23 [1] Ann Clin Biochem 2016 ;53:35-50 [2] Biochim Clin 2016;40SS:S83
  • 24. • Violating a guard band does not automatically mean to be in error. The probability of conformity, assuming a symmetrical distribution, falls to 50% when the measurement result equals the tolerance limit. • Violating a guard band does not automatically mean the presence of a systematic error, a second control can be used to confirm or not this event • The width of the “acceptance zone” automatically prompts for a higher or lower level of control Ceriotti F. Milano, 18-11-2016 24 LSL USL
  • 25. Ceriotti F. Milano, 18-11-2016 25 Monte Carlo simulation
  • 26. Comparison with the Westgard’s strategy • IQC results obtained during a 12 months period for glucose, creatinine and total cholesterol were collected from 4 clinical laboratories using 8 analyzers (4 Siemens Dimension Vista 1500, 2 Ortho Vitros 5600, 1 Siemens Advia 2400, 1 Beckman Coulter AU5800). For each measurand biological variation derived TEa were used; Westgard rules were selected according to the σ-metric calculated from a 6 months planning period and U was calculated from the imprecision of the IQC results during the same period. Based on these settings, the number of “out of control” alarms occurred in the following 6 months with the implemented Westgard’s rules were compared with the number of QC results falling within the “alarm zones”.
  • 27. Relationship between the alarm frequency and the DPMO of the method derived from the σ value estimated during the testing period. 6σ 5σ 4σ 3σ
  • 28. Conclusions • IQC plays a key role in the organization of a clinical laboratory that wants to guarantee the necessary analytical quality minimizing the risk of delivering erroneous results • To implement it correctly and effectively is not easy, especially when thousands of results have to be released every hour for dozens of different tests like in a typical “core laboratory” • The proposed approach simplifies the IQC implementation without loosing any of the relevant features that can be obtained with a correct implementation of the classical approach Ceriotti F. Milano, 18-11-2016 28
  • 29. Ceriotti F. Milano, 18-11-2016 29