11. Recommended Sample Taps X Poor , may draw sediment X Poor , avoid ells, valves, T’s and other areas of turbulence Better Poor , may draw air Best X X X
71. Turbidity Measurement Particle Count Measurement Measurement of light scattered at an angle. For municipal water/wastewater applications light scattering measurements at 90º to the incident light path. Particle counting measurements can be light scattering or light blocking. Light scattering technology is appropriate for particle sizes <1µm. Light blocking technology is appropriate for particle sizes > 1µm. For municipal drinking water applications, light blocking > 1µm (typically >2µm) is appropriate. Not a specific measurement of anything, it is a qualitative measurement A quantitative measurement of particle size and particle number. Measurement is independent of volume Measurement is volume dependent Measurement is relatively independent of flow rate. Sample can be flowing or static Sample must be flowing and flowing at a constant rate. Unit of measurement is nephelometric turbidity units, NTU Unit of measurement is particle counting must state the number of particles, particle size or range of sizes and unit volume. For example 10 particles per ml > 5µm or 200 particles per ml 2-5µm. Peak wavelength response for lab, SS7 and 1720 series process is ~560nm, FT660 is 660 nm, for Accu4 ~ 850nm Wavelength is 790 nm Theoretical particle size sensitivity 10 -8 m (0.01µm) 2200 PCX sensitivity is > 2µm Turbidity Measurement Particle Count Measurement Size range from approximately 10 -8 m - 10 -3 m (large molecules to sand) For the 2200 PCX: 2-750 µm Color in water is a negative interference except for the Accu 4 Color does not interfere with particle count measurements Turbidity interferes. High turbidity is a negative interference. At high turbidity scattered light is blocked or absorbed by the large amount of turbidity and thus does not reach the detector. The turbidity will be false negative. This phenomenon is called ‘going blind.’ Turbidity interferes. High turbidity is a negative interference. Particle counters typically have a range of approximately 17,000 particles/ml > 2µm. The particle counter may be over range at turbidity between 1 and 10 NTU – typically approximately 5 NTU. The particle counts will be false negative. Light absorbing materials (i.e. activated carbon) are negative interferences. Light absorbing materials (i.e. carbon) block light well and thus are counted. They do not interfere Accuracy of measurement is influenced by particle size Accuracy of measurement is influenced by particle size Accuracy of measurement is influenced by particle shape Accuracy of measurement is influenced by particle shape Accuracy of measurement is influenced by a particle’s refractive index Accuracy of measurement is influenced by particle’s refractive index
113. Large Filters Filter Side A Filter Side B BW Trough/Launderer BW Trough/Launderer BW Trough/Launderer BW Trough/Launderer BW Trough/Launderer BW Trough/Launderer Flow Common Drain Gullet Mount Permanent Probe here to look into the stream Use second probe to move around the filter to assess uniformity of backwash
114.
115.
116.
117.
118.
119.
120.
121. Surface Wash On 4,000 gpm 6,000 gpm Surface Wash Off Surface Wash On 3,900 gpm Surface Wash Off 6,000gpm 5,300 gpm 2,000 gpm
157. Chlorination Curve Chlorine Added Chlorine Measured Chloramination I Breakpoint 9:1 Cl 2 :N Ratio II 5:1 Cl 2 :N Ratio
158.
159. Chlorination Curve Free Chlorination III Chlorine Added Chlorine Measured Chloramination I Breakpoint 9:1 Cl 2 :N Ratio II 5:1 Cl 2 :N Ratio
160.
161.
162.
163. Chloramination Species Free Chlorination III Total Chlorine Free Ammonia Chlorine Added Chlorine Measured Chloramination I II
164.
165. Chloramination Species Free Chlorination III Total Chlorine Free Ammonia Monochloramine Chlorine Added Chlorine Measured Chloramination I II
166.
167. Chloramination Species Free Chlorination III Total Chlorine Free Ammonia Monochloramine Free Chlorine Chlorine Added Chlorine Measured Chloramination I II
168.
169. Where Am I When Total Chlorine = 3mg/L? Free Chlorination III Chlorine Added Chlorine Measured Chloramination I II
170. I Am Here! Free Chlorination III NH 2 Cl = t-DPD f-NH 3 N > 0 NH 2 Cl < t-DPD f-NH 3 N = 0 t-DPD > 0 NH 2 Cl = 0 f-NH 3 N = 0 Chlorine Added Chlorine Measured Chloramination I II
182. How the Algorithm Works Baseline Estimator Gain Matrix A Distance Measure Unit Vector Formation using Y(t) Vector Search X(t) Five Parameter Signal Vector Baseline Deviation Y(t) + - Resultant Vector Report Best Match Vector Libraries ( Agent, Plant ) A two-step process is used: Trigger when deviations indicate agent Classify agent in response to Trigger Is Threshold exceeded? No Yes Trigger Signal Threshold Level Trigger
184. The plant uses caustic feed to control water pH and experienced an operational problem that resulted in the feed of excess caustic. That affected the pH and the conductivity of the water, causing the Event Monitor to alarm. The Event Monitor learned this Plant Event and can identify a recurrence of the event.
185. Road work near a distribution line dislodged biomass and other particulate matter from the lining of the pipe. There was a massive increase in turbidity, which not only showed up on the turbidimeter, but also showed up as an interference in the chlorine measurement ( optical ). As expected, the conductivity and pH also showed minor changes. The increase in biomass in the water was indicated by the TOC analyzer. This event illustrates the ability of the Event Monitor to detect and alarm on unanticipated events. This event also provides a signature for the materials adhering to the walls of the pipes in this location.
186. The Event Monitor is located in a building which experiences a daily variation in water pressure. The sample variation is associated with a turbidity increase that causes a Trigger. There is also a small pH decrease at that time, possible because of increased solubility of CO2 in the water, dropping the pH slightly. This pattern is recognized by the Event Monitor as a "Normal" event, rather than an alarm condition.
187. Effects of Variable Demand Daily events influencing turbidity, chlorine, pH and possibly conductivity are not completely understood but suspected to be caused by water demand fluctuations in the area. May indicate need for flushing and chlorine booster.
188.
189.
190. Possible Chlorine Feed Event On April 3, 2007 there was a turbidity and pH increase and a decrease in chlorine and conductivity. Operator believes that there might have been a problem with the chlorine feed at the plant. However, this cannot be confirmed. The plant was using free chlorine instead of chloramines at the time which rules out the ammonia feed problem.
191.
192. Main Break #1 The break occurred on July 30 th , 2006. The line was a 36” water main located 1.0 miles from the WDMP. Conductivity, turbidity, and chlorine spiked. There appears to be two water flow interruptions to the WDMP the night before but it’s unclear if they are related to the break on the 30 th .
193. Main Break #2 The break occurred at night on September 20 th , 2006. The exact time of the break is currently unknown, although there appears to be a flow interruption to the WDMP the previous morning. The line involved was a 12” water main which is 0.4 miles from the WDMP.
A consdieration in any process for water production and/or treatment is the best use of data accumulation whether it comes from laboratory analysis or from online continuous monitoring. Each type of analysis has its own advantages and disadvantages. The user must decide what combination works the best.
Online, continuouos monitoring provides more timely data for quicker reaction to problems.
Laboratory analysis always has a place even when on-line process analyzers are in use.
1) The Series 5000 silica performs 96 tests/day in the long cycle and almost twice as many in the short cycle. The CL17 performs an analysis every 2.5 minutes, for a total of 576 tests per day. The cycle-times on other instruments will vary. 2) If a lab test is only performed twice a shift, problems in a process can occur between tests without being noticed. With on-line instrumentation, problems will show up almost immediately. Delays can cause major problems in a process. 3) The CL17 and EC 1000 can be used to operate chemical feed pumps. 4) In most cases, all an operator must do is change reagents and perform routine maintenance. Time is not spent collecting and running samples. 5) Lab tests leave room for errors between technicians and within a technician. Sample collection errors , dilutions, contamination, dirty glassware, etc., can all occur when performing lab tests. 6) If many lab tests are being run, in some cases reagent costs can be reduced when using a process instrument.
1) One of the reasons a process instrument is used, is to reduce the time and effort spent monitoring a process. If an instrument requires a great deal of maintenance, it defeats the purpose of having the analyzer. Hach Company has been trying to reduce maintenance on our instruments. Compare maintenance for the Series 5000 to a PCA. 2) Get the specifications on an analyzer and make sure it meets the needs of the customer. Be sure to look at detection limits, range, accuracy, resolution and repeatability. 3) Is the instrument reliable for the application? If the samples are dirty, there will be more problems than with clean samples. Check with others in the industry who have similar applications. Check with Hach Regional Managers or Hach Dealers for information. 4) Is the instrument easy for the operator to use? Is it easy to read and understand the display? Is the manual easy to read and understand? 5) Take a look at the features of the analyzer. Is everything that the customer needs and wants there? Are there a lot of features that are not really needed, or is the customer paying extra for features he will never use? 6) Cost should be last on the list of considerations. Consider both the cost of operation and of the analyzer itself, but cost shouldn’t be the determining factor in buying an instrument. If it is necessary to pay a little more to get what is wanted and needed, then do it.
1) One of the reasons a process instrument is used, is to reduce the time and effort spent monitoring a process. If an instrument requires a great deal of maintenance, it defeats the purpose of having the analyzer. Hach Company has been trying to reduce maintenance on our instruments. Compare maintenance for the Series 5000 to a PCA. 2) Get the specifications on an analyzer and make sure it meets the needs of the customer. Be sure to look at detection limits, range, accuracy, resolution and repeatability. 3) Is the instrument reliable for the application? If the samples are dirty, there will be more problems than with clean samples. Check with others in the industry who have similar applications. Check with Hach Regional Managers or Hach Dealers for information. 4) Is the instrument easy for the operator to use? Is it easy to read and understand the display? Is the manual easy to read and understand? 5) Take a look at the features of the analyzer. Is everything that the customer needs and wants there? Are there a lot of features that are not really needed, or is the customer paying extra for features he will never use? 6) Cost should be last on the list of considerations. Consider both the cost of operation and of the analyzer itself, but cost shouldn’t be the determining factor in buying an instrument. If it is necessary to pay a little more to get what is wanted and needed, then do it.
1) One of the reasons a process instrument is used, is to reduce the time and effort spent monitoring a process. If an instrument requires a great deal of maintenance, it defeats the purpose of having the analyzer. Hach Company has been trying to reduce maintenance on our instruments. Compare maintenance for the Series 5000 to a PCA. 2) Get the specifications on an analyzer and make sure it meets the needs of the customer. Be sure to look at detection limits, range, accuracy, resolution and repeatability. 3) Is the instrument reliable for the application? If the samples are dirty, there will be more problems than with clean samples. Check with others in the industry who have similar applications. Check with Hach Regional Managers or Hach Dealers for information. 4) Is the instrument easy for the operator to use? Is it easy to read and understand the display? Is the manual easy to read and understand? 5) Take a look at the features of the analyzer. Is everything that the customer needs and wants there? Are there a lot of features that are not really needed, or is the customer paying extra for features he will never use? 6) Cost should be last on the list of considerations. Consider both the cost of operation and of the analyzer itself, but cost shouldn’t be the determining factor in buying an instrument. If it is necessary to pay a little more to get what is wanted and needed, then do it.
1) One of the reasons a process instrument is used, is to reduce the time and effort spent monitoring a process. If an instrument requires a great deal of maintenance, it defeats the purpose of having the analyzer. Hach Company has been trying to reduce maintenance on our instruments. Compare maintenance for the Series 5000 to a PCA. 2) Get the specifications on an analyzer and make sure it meets the needs of the customer. Be sure to look at detection limits, range, accuracy, resolution and repeatability. 3) Is the instrument reliable for the application? If the samples are dirty, there will be more problems than with clean samples. Check with others in the industry who have similar applications. Check with Hach Regional Managers or Hach Dealers for information. 4) Is the instrument easy for the operator to use? Is it easy to read and understand the display? Is the manual easy to read and understand? 5) Take a look at the features of the analyzer. Is everything that the customer needs and wants there? Are there a lot of features that are not really needed, or is the customer paying extra for features he will never use? 6) Cost should be last on the list of considerations. Consider both the cost of operation and of the analyzer itself, but cost shouldn’t be the determining factor in buying an instrument. If it is necessary to pay a little more to get what is wanted and needed, then do it.
This slide is the flow diagram for the production of potable (drinking) water, and the various parameters and sampling points for monitoring.