Presented By-
Sudipta Nandi
M. Pharm 1st Sem
Department of Pharmaceutical Sciences
Dr. Harisingh Gour Vishwavidyalaya Sagar, M.P.
(A Central University)
Content
 INTRODUCTION
 HISTORICAL BACKGROUND OF GC-MS
 PRINCIPLE
 INSTRUMENTATION
• GAS CHROMATOGRAPHY
• INTERFACE
• MASS SPECTROMETER
 WORKING
 LIMITATIONS
 APPLICATION
 CONCLUSION
 REFERENCES
INTRODUCTION
 Gas chromatography-mass spectrometry (GC-MS) is the synergistic
combination of two analytical method to separate and identify different
substances within a test sample.
 Gas chromatography separates the components of a mixture in time.
 Mass spectrometer provides information that aids in the identification and
structural elucidation of each component.
HYPHENATED
TECHNIQUES
Gas Chromatography-Mass Spectrometry
(GC-MS)
GAS
CHROMATOGRAPHY
MASS
SPECTROMETRY
GAS
CHROMATOGRAPHY
-MASS
SPECTROMETRY
Separated mixture of
chemicals so each can
be identified
individually
Identifies
(detects)chemicals
based on their molecular
weight or mass
A Chemical Analysis
Technique combining two
instruments to provide for
powerful separation and
identification capabilities
• Roland Gohlke and Fred mcLafferty introduce use of MS as
detector of GC in 1950s.
• Miniaturized computers has helped in the simplification of
instrument.
• In 1968, the Finnigan Instrument Corporation delivered the first
quadrupole GC-MS.
• By the 2000s computerized GC-MS instruments using quadrupole
technology had become essential.
• In 2005 GC tandem MS-MS have been introduced.
 The sample solution is injected into the GC inlet where it is
vaporized and swept onto a chromatographic column by the
carrier gas.
 The sample flows through column and compounds are
separated by virtue of their relative interaction with the
coating of the column and the carrier gas.
 Latter part of column passes through a heated transfer line
and ends at entrance to ion source where compounds eluting
from the column are converted to ions and detected
according to their mass to charge m/z ratio.
INSTRUMENTATION LAYOUT
Gas
chromatography
• Carrier gas
• Pneumatic control
• Injector
• Column
• Oven
Interface
• Jet Interface
• Direct Capillary
Infusion Interface
• Watson-Biemann
effusion separator
Mass
spectrometer
• Ion source
• High-vacuum
system
• Mass analyzer
• Ion collector
Data
system
Fig : Principle of GC-MS
1. Carrier Gas :
 Served as mobile phase supplied in the steel tank under high pressure.
 At pressure of 40-80 psi this passes into flow controllers.
 Example- nitrogen, helium; hydrogen and argon can also be used.
 Requirements –
 Inert
 Column requirements
 Detectors
 Purity- better than 99.995%
 Cost effective & available
2. Pneumatic Control :
 Gas supply is regulated to the correct pressure and then fed to the required
part of instrument.
 Older instruments – manual pressure control via regulators.
 Modern GC instruments – Electronic Pneumatic pressure controller.
3. Oven :
 Temperature programmable, typically range from 5°C - 400°C but can go as
low as -25°C with cryogenic cooling.
4. Sample Injection Port :
 Sample is made to vaporized rapidly before entering to column.
 Various kind of injectors :
 Packed column injectors
 Split injection
 Splitless injection
 Programmed Split/Splitless injection
 Programmed On-Column injector
5. Column :
 Two kinds of column used :
 Packed column
 Capillary column
 Gas chromatograph GC-MS utilizes capillary column.
 Here, stationary phase has been chemically bonded to the fused silica, e.g.,
DB-5.
Packed Column :-
• Less commonly used, having
diameter of 2-3 mm and length of
0.5-10 m.
• Manufactured from steel or glass,
internal wall is treated to avoid
catalytic effect with the sample.
• They can withstand a carrier gas
flow rate within range 10-
40ml/min.
Capillary Column :-
• Consist of long capillary tubing
10-100 m in length.
• Made up from stainless steel &
coil.
• Shorter columns - for fast analysis
• Larger columns – for high
resolution separation
6. Detectors :
 Simple and reliable
 Sensitive to electronegative groups (halogens)
 Largely non-destructive
 Limited dynamic range (10²)
 Mass sensitive detectors
-Thermal Conductivity Detector (TCD)
-Flame Ionization Detector (FID)
-Electron Capture Detector (ECD)
INTERFACE
 Pressure incompatibility problem between GC and MS was solved by
inserting an interface.
 Mainly three types of interfaces are there commercially available :
 Jet/Orifice separator
 Direct capillary infusion interface
 Watson-Biemann effusion separator
Fig : Direct capillary infusion
Fig : Jet separator Fig : Watson-Biemann effusion
MASS SPECTROMETER
 “ Mass spectrometry is a technique used for measuring the molecular weight
and determining the molecular formula of an organic compound.”
 In general a mass spectrometer consists of
 An ion source
 High-vacuum system
 A mass-selective analyzer
 An ion collector
 Data system
Ionization Method Typical Analytes Sample
Introduction
Mass
Range
Method
Highlights
Chemical Ionization (CI) Relatively small,
volatile
GC or
liquid/solid
probe
Upto
1000
Daltons
Soft method,
molecular ion peak
[M+H] +
Electron Impact
Ionization (EI)
Relatively small,
volatile
GC or
liquid/solid
probe
Upto
1000
Daltons
Hard method,
versatile, provides
structure info
Electrospray Ionization
(ESI)
Peptides, proteins,
nonvolatile
Liquid
chromatography
Upto
200000
Daltons
Soft method, ions
often multiply
charged
Fast Atom Bombardment
(FAB)
Carbohydrates,
organometallics,
peptides,
nonvolatile
Sample mixed in
viscous matrix
Upto
6000
Daltons
Harder than ESI or
MALDI
Matrix Assisted Laser
Desorption Ionization
(MALDI)
Peptides, proteins,
nucleotides
Sample mixed in
solid matrix
Upto
500000
Daltons
Soft method, very
high mass
Mass Selective Analyzer
• They deflects ions down a curved tubes in a magnetic fields based on their
kinetic energy determined by the mass, charge and velocity.
• The magnetic field is scanned to measure different ions.
Mass Analyzers
Quadrupole Ion Trap Time of flight
Quadrupole Analyzer :
• Also known as ‘Hewlett-Packard’ or ‘Mass Selective Detector’.
• In quadrupole mass analyser a set of four rods are arranged parallel to the
direction.
• Only m/z is been determined and stable oscillation takes place.
• Ions travels in quadrupole axis with cork screw type of trajectory.
• It functions as a mass filter
Ion Trap Mass Analyzer :
 The ion trap mass analyzer operates by similar principles where it consists
of circular ring electrode and two end caps that form a chamber.
 AC or DC power along RF potential is applied between the cups and the
ring electrode.
Ions entering into
chamber
Trapped by
electromagnetic fields
Ions oscillate in concentric
trajectories
Resonant Ejection
Time-of-flight Analyzer :
 The time-of-flight (TOF) analyser uses an electric field to accelerate the
ions through the same potential.
 Then it measures the time take to reach the detector.
 If the particles all have the same charge, the kinetic energies will be
identical and their velocities will depend upon only on their masses.
 Mass spectrometers employ computer control of same functions and also
use a computerised display and output.
 It is used to identify and measure the concentration of one or more analytes
in a complex mixture
 Advantages –
 Amount of data generated is very large
 It stores every sec for upto 90 min.
Data Handling
MASS
CHROMATOGRAMS
SELECTED ION
MONITORING
GC TRACE OF A THREE COMPONENT MIXTURE.
THE MASS SPECTROMETER GIVES A SPECTRUM FOR
EACH COMPONENT
Consider the mass spectrum of CH4 below:
LIMITATIONS
• Only compounds with vapour pressure exceeding about 10-10 torr can be
analysed by gas chromatography-mass spectrometry (GC-MS).
• Determining positional substitution on aromatic ring is often difficult.
• Certain isomeric compounds cannot be distinguished by mass spectrometry.
• Non-volatile matrices require additional preparation (extraction,
outgassing, etc.)
• Atmospheric gases are challenging (CO2, N2, O2, Ar, CO, H2O).
Applications of GC-MS
Petrochemical and hydrocarbon analysis
Geochemical research
Forensic (arsons, explosives, drugs, unknowns)
Environmental analysis
Pesticide analysis, food safety and quality
Clinical toxicology
Food, beverage and perfume analysis
CONCLUSION
 As we can see, GC-MS is used both in qualitative and quantitative
measurement.
 Being hyphenated technique, the cost of GC-MS becomes very high.
 Nowadays, cost is also cheaper than before as quadrupole and ion trap
instruments are used, which does not only save space but also save
manufacturing cost.
 GC-MS is an analytical method of increased sensitivity and reliability even
in very small quantities (ng).
 Braun R., Introduction To Instrumental Analysis, Second
Edition, PharmaMed Press, Hyderabad, Page no. 251-270.
 Chatwal G.R., Anand S.K., Instrumental method of Chemical
Analysis, Himalaya Publishing House, Fifth Edition-2012,
New Delhi, Page no. 420-449.
 Gohlke R.S.(1959), Analytical Chemistry, 36, Page no. 759-
764.
THANK YOU

Weitere ähnliche Inhalte

Was ist angesagt?

Affinity chromatography ppt
Affinity chromatography pptAffinity chromatography ppt
Affinity chromatography ppt
poojakamble1609
 
Ion exchange chromatography
Ion exchange chromatography Ion exchange chromatography
Ion exchange chromatography
Vharsha Haran
 

Was ist angesagt? (20)

Gas chromatography and its instrumentation
Gas chromatography and its instrumentationGas chromatography and its instrumentation
Gas chromatography and its instrumentation
 
Ion exchange chromatography
Ion  exchange chromatographyIon  exchange chromatography
Ion exchange chromatography
 
Lc ms
Lc msLc ms
Lc ms
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
Affinity chromatography ppt
Affinity chromatography pptAffinity chromatography ppt
Affinity chromatography ppt
 
Atomic absorption spectroscopy
Atomic absorption spectroscopy Atomic absorption spectroscopy
Atomic absorption spectroscopy
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
Gel permeation chromatography (GPC)
Gel permeation chromatography (GPC)Gel permeation chromatography (GPC)
Gel permeation chromatography (GPC)
 
Affinity chromatography
Affinity chromatographyAffinity chromatography
Affinity chromatography
 
Uv visible spectrophotometer
Uv visible spectrophotometerUv visible spectrophotometer
Uv visible spectrophotometer
 
Atomic absorption Spectrophotometry
Atomic absorption Spectrophotometry Atomic absorption Spectrophotometry
Atomic absorption Spectrophotometry
 
Infrared spectroscopy
Infrared spectroscopyInfrared spectroscopy
Infrared spectroscopy
 
HPLC - High Performance Liquid Chromatography
HPLC - High Performance Liquid ChromatographyHPLC - High Performance Liquid Chromatography
HPLC - High Performance Liquid Chromatography
 
High performance liquid chromatography (HPLC)
High performance liquid chromatography (HPLC)High performance liquid chromatography (HPLC)
High performance liquid chromatography (HPLC)
 
Mass spectrometry final.pptx
Mass spectrometry final.pptxMass spectrometry final.pptx
Mass spectrometry final.pptx
 
IR Spectroscopy
IR Spectroscopy IR Spectroscopy
IR Spectroscopy
 
Ion exchange chromatography
Ion exchange chromatography Ion exchange chromatography
Ion exchange chromatography
 
GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.
GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.
GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.
 
Mass spectrometry basic principle & Instrumentation
Mass spectrometry basic principle & InstrumentationMass spectrometry basic principle & Instrumentation
Mass spectrometry basic principle & Instrumentation
 
ATOMIC ABSORPTION SPECTROPHOTOMETRY
ATOMIC ABSORPTION SPECTROPHOTOMETRYATOMIC ABSORPTION SPECTROPHOTOMETRY
ATOMIC ABSORPTION SPECTROPHOTOMETRY
 

Ähnlich wie Gas chromatography mass spectrometry (GC-MS)

GC MASS SPECTROMETERY
GC MASS SPECTROMETERYGC MASS SPECTROMETERY
GC MASS SPECTROMETERY
SNEHA AGRAWAL GUPTA
 
Gas Chromatograph & Analyzers
Gas Chromatograph & AnalyzersGas Chromatograph & Analyzers
Gas Chromatograph & Analyzers
ENG AL AMERY
 

Ähnlich wie Gas chromatography mass spectrometry (GC-MS) (20)

Gc ms (Revised)
Gc ms (Revised)Gc ms (Revised)
Gc ms (Revised)
 
Gc Ms
Gc MsGc Ms
Gc Ms
 
Gas chromatography
Gas chromatography Gas chromatography
Gas chromatography
 
GC-MS .pptx
GC-MS .pptxGC-MS .pptx
GC-MS .pptx
 
Chromatography(gc ms & lc ms)
Chromatography(gc ms & lc ms)Chromatography(gc ms & lc ms)
Chromatography(gc ms & lc ms)
 
Gas chromatography mass spectrometry
Gas chromatography mass spectrometryGas chromatography mass spectrometry
Gas chromatography mass spectrometry
 
Gas Chromatography-Mass Spectroscopy - By Vishal Dakhale
Gas Chromatography-Mass Spectroscopy - By Vishal DakhaleGas Chromatography-Mass Spectroscopy - By Vishal Dakhale
Gas Chromatography-Mass Spectroscopy - By Vishal Dakhale
 
GC MASS SPECTROMETERY
GC MASS SPECTROMETERYGC MASS SPECTROMETERY
GC MASS SPECTROMETERY
 
Gas chromatography-Mass spectrometry (GC-MS)
Gas chromatography-Mass spectrometry (GC-MS)Gas chromatography-Mass spectrometry (GC-MS)
Gas chromatography-Mass spectrometry (GC-MS)
 
Gas chromatography mass spectrometry ppt
Gas chromatography mass spectrometry pptGas chromatography mass spectrometry ppt
Gas chromatography mass spectrometry ppt
 
GC-MS and GC-AAS Hyphenated Techniques.pptx
GC-MS and GC-AAS Hyphenated Techniques.pptxGC-MS and GC-AAS Hyphenated Techniques.pptx
GC-MS and GC-AAS Hyphenated Techniques.pptx
 
Gc ms techniques SEM4 PAPER2 MITHIBAI COLLEGE
Gc ms techniques SEM4 PAPER2 MITHIBAI COLLEGEGc ms techniques SEM4 PAPER2 MITHIBAI COLLEGE
Gc ms techniques SEM4 PAPER2 MITHIBAI COLLEGE
 
Gas Chromatograph & Analyzers
Gas Chromatograph & AnalyzersGas Chromatograph & Analyzers
Gas Chromatograph & Analyzers
 
Chromatography
ChromatographyChromatography
Chromatography
 
Gas Chromatography and mass spectrometry
Gas Chromatography and mass spectrometry Gas Chromatography and mass spectrometry
Gas Chromatography and mass spectrometry
 
Gas Chromatography.pptx
Gas Chromatography.pptxGas Chromatography.pptx
Gas Chromatography.pptx
 
Gas Chromatography.pptx
Gas Chromatography.pptxGas Chromatography.pptx
Gas Chromatography.pptx
 
Gas Chromatography in Analytical Analysis.pptx
Gas Chromatography in Analytical Analysis.pptxGas Chromatography in Analytical Analysis.pptx
Gas Chromatography in Analytical Analysis.pptx
 
GC-MS.-principle, instrumentation,working,Application .
GC-MS.-principle, instrumentation,working,Application .GC-MS.-principle, instrumentation,working,Application .
GC-MS.-principle, instrumentation,working,Application .
 
Gass Chromatography Atomic Absorption spectroscopy(GC-AAS)
Gass Chromatography Atomic Absorption spectroscopy(GC-AAS)Gass Chromatography Atomic Absorption spectroscopy(GC-AAS)
Gass Chromatography Atomic Absorption spectroscopy(GC-AAS)
 

Mehr von D.R. Chandravanshi

SYNTHETIC PEPTIDE VACCINES AND RECOMBINANT ANTIGEN VACCINE
SYNTHETIC PEPTIDE  VACCINES  AND RECOMBINANT  ANTIGEN VACCINESYNTHETIC PEPTIDE  VACCINES  AND RECOMBINANT  ANTIGEN VACCINE
SYNTHETIC PEPTIDE VACCINES AND RECOMBINANT ANTIGEN VACCINE
D.R. Chandravanshi
 

Mehr von D.R. Chandravanshi (20)

Questions of Central Nervous System
Questions of Central Nervous SystemQuestions of Central Nervous System
Questions of Central Nervous System
 
Drug Store and Business Management (DSBM).pdf
Drug Store and Business Management (DSBM).pdfDrug Store and Business Management (DSBM).pdf
Drug Store and Business Management (DSBM).pdf
 
Channel of Distribution.pdf
Channel of Distribution.pdfChannel of Distribution.pdf
Channel of Distribution.pdf
 
Channels of Distribution.pdf
Channels of Distribution.pdfChannels of Distribution.pdf
Channels of Distribution.pdf
 
Monoclonal Antibody Process.pdf
Monoclonal Antibody Process.pdfMonoclonal Antibody Process.pdf
Monoclonal Antibody Process.pdf
 
Chapter-1 Introduction to Human Anatomy and Physiology
Chapter-1 Introduction to Human Anatomy and PhysiologyChapter-1 Introduction to Human Anatomy and Physiology
Chapter-1 Introduction to Human Anatomy and Physiology
 
National Health Programs, Objectives, Fucntions
National Health Programs, Objectives, FucntionsNational Health Programs, Objectives, Fucntions
National Health Programs, Objectives, Fucntions
 
EU(European Union) and ICH Guidelines
EU(European Union) and ICH GuidelinesEU(European Union) and ICH Guidelines
EU(European Union) and ICH Guidelines
 
Forms of business organization, DSBM D.Pharma 2nd year
Forms of business organization, DSBM D.Pharma 2nd yearForms of business organization, DSBM D.Pharma 2nd year
Forms of business organization, DSBM D.Pharma 2nd year
 
Psoriasis: Natural Product and Phytomedicine
Psoriasis: Natural Product and Phytomedicine Psoriasis: Natural Product and Phytomedicine
Psoriasis: Natural Product and Phytomedicine
 
Omega 3 fatty acids d rc
Omega 3 fatty acids d rcOmega 3 fatty acids d rc
Omega 3 fatty acids d rc
 
Pilot Plant Techniques for SOLID dosage forms
Pilot Plant Techniques for SOLID dosage formsPilot Plant Techniques for SOLID dosage forms
Pilot Plant Techniques for SOLID dosage forms
 
Optimum Performance Laminar Chromatography (OPLC)
Optimum Performance Laminar Chromatography (OPLC) Optimum Performance Laminar Chromatography (OPLC)
Optimum Performance Laminar Chromatography (OPLC)
 
SYNTHETIC PEPTIDE VACCINES AND RECOMBINANT ANTIGEN VACCINE
SYNTHETIC PEPTIDE  VACCINES  AND RECOMBINANT  ANTIGEN VACCINESYNTHETIC PEPTIDE  VACCINES  AND RECOMBINANT  ANTIGEN VACCINE
SYNTHETIC PEPTIDE VACCINES AND RECOMBINANT ANTIGEN VACCINE
 
New generation vaccines production
New generation vaccines productionNew generation vaccines production
New generation vaccines production
 
Monoclonal Antibody and Hybridoma technology
Monoclonal Antibody and Hybridoma technologyMonoclonal Antibody and Hybridoma technology
Monoclonal Antibody and Hybridoma technology
 
Scanning Electron Microscope (SEM)
Scanning Electron Microscope (SEM)Scanning Electron Microscope (SEM)
Scanning Electron Microscope (SEM)
 
Formulation of parenteral products
Formulation of parenteral productsFormulation of parenteral products
Formulation of parenteral products
 
Evaluation of parenterals products
Evaluation of parenterals productsEvaluation of parenterals products
Evaluation of parenterals products
 
Factors affecting Design of Experiment (DOE) and softwares of DOE
Factors affecting Design of Experiment (DOE) and softwares of DOEFactors affecting Design of Experiment (DOE) and softwares of DOE
Factors affecting Design of Experiment (DOE) and softwares of DOE
 

Kürzlich hochgeladen

1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdf
QucHHunhnh
 
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in DelhiRussian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
kauryashika82
 
Seal of Good Local Governance (SGLG) 2024Final.pptx
Seal of Good Local Governance (SGLG) 2024Final.pptxSeal of Good Local Governance (SGLG) 2024Final.pptx
Seal of Good Local Governance (SGLG) 2024Final.pptx
negromaestrong
 

Kürzlich hochgeladen (20)

Micro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdfMicro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdf
 
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
 
Grant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingGrant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy Consulting
 
1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdf
 
Application orientated numerical on hev.ppt
Application orientated numerical on hev.pptApplication orientated numerical on hev.ppt
Application orientated numerical on hev.ppt
 
Magic bus Group work1and 2 (Team 3).pptx
Magic bus Group work1and 2 (Team 3).pptxMagic bus Group work1and 2 (Team 3).pptx
Magic bus Group work1and 2 (Team 3).pptx
 
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in DelhiRussian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
 
Unit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptxUnit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptx
 
microwave assisted reaction. General introduction
microwave assisted reaction. General introductionmicrowave assisted reaction. General introduction
microwave assisted reaction. General introduction
 
psychiatric nursing HISTORY COLLECTION .docx
psychiatric  nursing HISTORY  COLLECTION  .docxpsychiatric  nursing HISTORY  COLLECTION  .docx
psychiatric nursing HISTORY COLLECTION .docx
 
Introduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsIntroduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The Basics
 
SKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptx
SKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptxSKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptx
SKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptx
 
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdfUGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
 
Seal of Good Local Governance (SGLG) 2024Final.pptx
Seal of Good Local Governance (SGLG) 2024Final.pptxSeal of Good Local Governance (SGLG) 2024Final.pptx
Seal of Good Local Governance (SGLG) 2024Final.pptx
 
How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17
 
How to Manage Global Discount in Odoo 17 POS
How to Manage Global Discount in Odoo 17 POSHow to Manage Global Discount in Odoo 17 POS
How to Manage Global Discount in Odoo 17 POS
 
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
 
PROCESS RECORDING FORMAT.docx
PROCESS      RECORDING        FORMAT.docxPROCESS      RECORDING        FORMAT.docx
PROCESS RECORDING FORMAT.docx
 
ICT role in 21st century education and it's challenges.
ICT role in 21st century education and it's challenges.ICT role in 21st century education and it's challenges.
ICT role in 21st century education and it's challenges.
 
Asian American Pacific Islander Month DDSD 2024.pptx
Asian American Pacific Islander Month DDSD 2024.pptxAsian American Pacific Islander Month DDSD 2024.pptx
Asian American Pacific Islander Month DDSD 2024.pptx
 

Gas chromatography mass spectrometry (GC-MS)

  • 1. Presented By- Sudipta Nandi M. Pharm 1st Sem Department of Pharmaceutical Sciences Dr. Harisingh Gour Vishwavidyalaya Sagar, M.P. (A Central University)
  • 2. Content  INTRODUCTION  HISTORICAL BACKGROUND OF GC-MS  PRINCIPLE  INSTRUMENTATION • GAS CHROMATOGRAPHY • INTERFACE • MASS SPECTROMETER  WORKING  LIMITATIONS  APPLICATION  CONCLUSION  REFERENCES
  • 3. INTRODUCTION  Gas chromatography-mass spectrometry (GC-MS) is the synergistic combination of two analytical method to separate and identify different substances within a test sample.  Gas chromatography separates the components of a mixture in time.  Mass spectrometer provides information that aids in the identification and structural elucidation of each component. HYPHENATED TECHNIQUES
  • 4. Gas Chromatography-Mass Spectrometry (GC-MS) GAS CHROMATOGRAPHY MASS SPECTROMETRY GAS CHROMATOGRAPHY -MASS SPECTROMETRY Separated mixture of chemicals so each can be identified individually Identifies (detects)chemicals based on their molecular weight or mass A Chemical Analysis Technique combining two instruments to provide for powerful separation and identification capabilities
  • 5. • Roland Gohlke and Fred mcLafferty introduce use of MS as detector of GC in 1950s. • Miniaturized computers has helped in the simplification of instrument. • In 1968, the Finnigan Instrument Corporation delivered the first quadrupole GC-MS. • By the 2000s computerized GC-MS instruments using quadrupole technology had become essential. • In 2005 GC tandem MS-MS have been introduced.
  • 6.  The sample solution is injected into the GC inlet where it is vaporized and swept onto a chromatographic column by the carrier gas.  The sample flows through column and compounds are separated by virtue of their relative interaction with the coating of the column and the carrier gas.  Latter part of column passes through a heated transfer line and ends at entrance to ion source where compounds eluting from the column are converted to ions and detected according to their mass to charge m/z ratio.
  • 7. INSTRUMENTATION LAYOUT Gas chromatography • Carrier gas • Pneumatic control • Injector • Column • Oven Interface • Jet Interface • Direct Capillary Infusion Interface • Watson-Biemann effusion separator Mass spectrometer • Ion source • High-vacuum system • Mass analyzer • Ion collector Data system Fig : Principle of GC-MS
  • 8.
  • 9. 1. Carrier Gas :  Served as mobile phase supplied in the steel tank under high pressure.  At pressure of 40-80 psi this passes into flow controllers.  Example- nitrogen, helium; hydrogen and argon can also be used.  Requirements –  Inert  Column requirements  Detectors  Purity- better than 99.995%  Cost effective & available
  • 10. 2. Pneumatic Control :  Gas supply is regulated to the correct pressure and then fed to the required part of instrument.  Older instruments – manual pressure control via regulators.  Modern GC instruments – Electronic Pneumatic pressure controller. 3. Oven :  Temperature programmable, typically range from 5°C - 400°C but can go as low as -25°C with cryogenic cooling.
  • 11. 4. Sample Injection Port :  Sample is made to vaporized rapidly before entering to column.  Various kind of injectors :  Packed column injectors  Split injection  Splitless injection  Programmed Split/Splitless injection  Programmed On-Column injector 5. Column :  Two kinds of column used :  Packed column  Capillary column  Gas chromatograph GC-MS utilizes capillary column.  Here, stationary phase has been chemically bonded to the fused silica, e.g., DB-5.
  • 12. Packed Column :- • Less commonly used, having diameter of 2-3 mm and length of 0.5-10 m. • Manufactured from steel or glass, internal wall is treated to avoid catalytic effect with the sample. • They can withstand a carrier gas flow rate within range 10- 40ml/min. Capillary Column :- • Consist of long capillary tubing 10-100 m in length. • Made up from stainless steel & coil. • Shorter columns - for fast analysis • Larger columns – for high resolution separation
  • 13. 6. Detectors :  Simple and reliable  Sensitive to electronegative groups (halogens)  Largely non-destructive  Limited dynamic range (10²)  Mass sensitive detectors -Thermal Conductivity Detector (TCD) -Flame Ionization Detector (FID) -Electron Capture Detector (ECD)
  • 14. INTERFACE  Pressure incompatibility problem between GC and MS was solved by inserting an interface.  Mainly three types of interfaces are there commercially available :  Jet/Orifice separator  Direct capillary infusion interface  Watson-Biemann effusion separator Fig : Direct capillary infusion Fig : Jet separator Fig : Watson-Biemann effusion
  • 15. MASS SPECTROMETER  “ Mass spectrometry is a technique used for measuring the molecular weight and determining the molecular formula of an organic compound.”  In general a mass spectrometer consists of  An ion source  High-vacuum system  A mass-selective analyzer  An ion collector  Data system
  • 16. Ionization Method Typical Analytes Sample Introduction Mass Range Method Highlights Chemical Ionization (CI) Relatively small, volatile GC or liquid/solid probe Upto 1000 Daltons Soft method, molecular ion peak [M+H] + Electron Impact Ionization (EI) Relatively small, volatile GC or liquid/solid probe Upto 1000 Daltons Hard method, versatile, provides structure info Electrospray Ionization (ESI) Peptides, proteins, nonvolatile Liquid chromatography Upto 200000 Daltons Soft method, ions often multiply charged Fast Atom Bombardment (FAB) Carbohydrates, organometallics, peptides, nonvolatile Sample mixed in viscous matrix Upto 6000 Daltons Harder than ESI or MALDI Matrix Assisted Laser Desorption Ionization (MALDI) Peptides, proteins, nucleotides Sample mixed in solid matrix Upto 500000 Daltons Soft method, very high mass
  • 17. Mass Selective Analyzer • They deflects ions down a curved tubes in a magnetic fields based on their kinetic energy determined by the mass, charge and velocity. • The magnetic field is scanned to measure different ions. Mass Analyzers Quadrupole Ion Trap Time of flight
  • 18. Quadrupole Analyzer : • Also known as ‘Hewlett-Packard’ or ‘Mass Selective Detector’. • In quadrupole mass analyser a set of four rods are arranged parallel to the direction. • Only m/z is been determined and stable oscillation takes place. • Ions travels in quadrupole axis with cork screw type of trajectory. • It functions as a mass filter
  • 19. Ion Trap Mass Analyzer :  The ion trap mass analyzer operates by similar principles where it consists of circular ring electrode and two end caps that form a chamber.  AC or DC power along RF potential is applied between the cups and the ring electrode. Ions entering into chamber Trapped by electromagnetic fields Ions oscillate in concentric trajectories Resonant Ejection
  • 20. Time-of-flight Analyzer :  The time-of-flight (TOF) analyser uses an electric field to accelerate the ions through the same potential.  Then it measures the time take to reach the detector.  If the particles all have the same charge, the kinetic energies will be identical and their velocities will depend upon only on their masses.
  • 21.  Mass spectrometers employ computer control of same functions and also use a computerised display and output.  It is used to identify and measure the concentration of one or more analytes in a complex mixture  Advantages –  Amount of data generated is very large  It stores every sec for upto 90 min. Data Handling MASS CHROMATOGRAMS SELECTED ION MONITORING
  • 22. GC TRACE OF A THREE COMPONENT MIXTURE. THE MASS SPECTROMETER GIVES A SPECTRUM FOR EACH COMPONENT
  • 23. Consider the mass spectrum of CH4 below:
  • 24. LIMITATIONS • Only compounds with vapour pressure exceeding about 10-10 torr can be analysed by gas chromatography-mass spectrometry (GC-MS). • Determining positional substitution on aromatic ring is often difficult. • Certain isomeric compounds cannot be distinguished by mass spectrometry. • Non-volatile matrices require additional preparation (extraction, outgassing, etc.) • Atmospheric gases are challenging (CO2, N2, O2, Ar, CO, H2O).
  • 25. Applications of GC-MS Petrochemical and hydrocarbon analysis Geochemical research Forensic (arsons, explosives, drugs, unknowns) Environmental analysis Pesticide analysis, food safety and quality Clinical toxicology Food, beverage and perfume analysis
  • 26. CONCLUSION  As we can see, GC-MS is used both in qualitative and quantitative measurement.  Being hyphenated technique, the cost of GC-MS becomes very high.  Nowadays, cost is also cheaper than before as quadrupole and ion trap instruments are used, which does not only save space but also save manufacturing cost.  GC-MS is an analytical method of increased sensitivity and reliability even in very small quantities (ng).
  • 27.  Braun R., Introduction To Instrumental Analysis, Second Edition, PharmaMed Press, Hyderabad, Page no. 251-270.  Chatwal G.R., Anand S.K., Instrumental method of Chemical Analysis, Himalaya Publishing House, Fifth Edition-2012, New Delhi, Page no. 420-449.  Gohlke R.S.(1959), Analytical Chemistry, 36, Page no. 759- 764.