Wuhan Feiyue Biotechnology CO.,LTD
1
FOR RESEARCH USE ONLY
NOT FOR IN VITRO CLINICAL DIAGNOSIS
Goat Interleukin 1 (IL-1) ELISA Kit
Catalog No: FY-EG4791 Size: 96T
This manual must be read attentively and completely before using this product.
If you have any problems, please contact us.
Phone: 86-027-87002654 86-180 6275 0658
Email: info@feiyuebio.com
Website: www.feiyuebio.com
Intended Use
The ELISA kit is a Sandwich enzyme immunoassay technique for the quantitative determination of Goat IL-1
concentrations in serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids.
Specification
 Sensitivity: 10pg/mL.
 Detection Range: 10-400pg/mL.
 No significant cross-reactivity or interference between Goat IL-1 and analogues was observed.
 Repeatability: Coefficient of variation is < 12%.
Principle of the Procedure
This ELISA kit uses the Sandwich-ELISA principle. The micro plate provided in this kit has been pre-coated with
an antibody specific to Goat IL-1. Standards or samples and Horseradish Peroxidase (HRP) labeled detection
antibody specific for Goat IL-1 are added to the plate wells together and incubated. After washing off unbound
material, TMB substrate solution is added to all wells and incubated. An enzyme-catalyzed reaction generates a
blue color in the solution, thereafter, stop solution is added to stop the substrate reaction and the color turns yellow.
The yellow solution is read at a wavelength of 450nm. The concentration of Goat IL-1 in the samples is then
calculated from the OD value by establishing a standard curve.
Limitations of the Procedure
1、This kit is for laboratory scientific research only, we will not be responsible for any consequences if this kit is
used for clinical diagnosis or any other procedures.
2、 Due to the uncertainty of its validity, this kit may not be suitable for testing some special experimental samples,
such as gene knockout experiments.
3、This kit should be used before its expiration date, and please strictly follow the instructions for storage.
4、Different manufacturers' kits or testing the same analyte by other methods may produce inconsistent results
because we do not compare our products with those of other manufacturers.
Wuhan Feiyue Biotechnology CO.,LTD
2
5、Since the antibodies used in the kit are usually prepared from recombinant proteins as immunogen, and
recombinant proteins can be limited by different fragmentation, expression and purification systems, it is not
recommended to use this kit to detect recombinant proteins
6、In order to get the best experimental results, please use only the reagents provided by the manufacturer, and do
not mix reagents from different batches and other manufacturers.
7、Due to the existing conditions and the limitations of science and technology, we cannot fully identify and
analyze the raw materials provided by the supplier comprehensively. Therefore, the kit may have some quality and
technical risks.
8、The possibility of interference cannot be excluded before all factors are tested in the ELISA immunoassay.
9、In order to obtain reproducible results, each step in the experiment should be controlled and variations in
sample collection, handling and storage may also lead to differences in sample measurements.
10、Although each kit passes rigorous quality testing, differences in measured values between batches of kits can
still be caused by factors such as shipping conditions and different laboratory equipment.
Reagents&Materials Provided
Materials & Equipment Required But Not Provided
Microplate reader with 450nm wavelength filter
Incubator capable of maintaining 37℃
Single or multi-channel pipettes with high precision
disposable pipette tips
EP tubes
Container for Wash Solution
Squirt bottle, manifold dispenser, or automated microplate washer
Deionized or distilled water
Absorbent paper for blotting the microplate
Sample Collection
Serum: Allow blood samples to clot for 2 hours at room temperature or overnight at 2-8℃ before centrifugation
for 15 min at 1000×g at 2~8℃. Supernatants should be taken for assay testing immediately or stored at -20℃ or -
80℃ for later use.
Plasma: Collect plasma using EDTA or heparin as an anticoagulant. Centrifuge samples for 15 min at 1000×g at
2~8℃ within 30 min of collection. Supernatants should be taken for assay testing immediately or stored at -20℃ or
-80℃ for later use.
Components Specifications Storage
Pre-coated ELISA Plate 12wells ×8 strips 2-8℃
Standard 6×0.3mL:400,200,100,50,25,0pg/mL 2-8℃
HRP-Conjugate reagent 1×10mL 2-8℃
Sample Diluent 1×6mL 2-8℃
Substrate Reagent A 1×6mL 2-8℃
Substrate Reagent B 1×6mL 2-8℃
Wash Buffer (20×) 1×25mL 2-8℃
Stop Solution 1×6mL 2-8℃
Plate Sealer 2 pieces
Manual 1 copy
Self-sealing bags 1pc
Wuhan Feiyue Biotechnology CO.,LTD
3
Tissue homogenates: Tissues should be rinsed in ice-cold PBS to remove excess blood thoroughly, weighed,
minced into small pieces. then Tissue pieces omogenized in PBS (tissue weight (g): PBS (mL) volume=1:9) with a
glass homogenizer on ice. The resulting suspension is sonicated with an ultrasonic cell disrupter till the solution is
clarified. The homogenates are then centrifuged for 5 min at 10000×g. Supernatants should be taken for assay
testing immediately or stored at -20℃ or -80℃ for later use.
Cell lysates: For adherent cells, gently wash the cells with moderate amount of cool PBS and detach the cells with
trypsin. Collect cells by centrifugation for 5 min at 1000×g(suspension cells can be collected by centrifugation
directly). Discard the supernate and wash cells 3 times with cool PBS. Resuspend cells in cool PBS with
concentration of 5×106
cells/mL. Repeat the freeze-thaw process several times until the cells are fully lysed.
Centrifuge for 10min at 1500×g at 2-8℃. Supernatants should be taken for assay testing immediately or stored at -
20℃ or -80℃ for later use.
Cell culture supernatant and other biological fluids: Centrifuge samples for 20 minutes at 1000xg at 2-8℃.
Supernatants should be taken for assay testing immediately or stored at -20℃ or -80℃ for later use.
Note for sample
1. Samples should be used within 6 days when stored at 2-8℃, otherwise samples must be stored at -20℃
(≤1month) or -80℃ (≤2months). Avoid repeated freeze-thaw cycles.
2. Please predict the concentration before assaying. If the sample concentration is not within the range of the
standard curve, users must determine the optimal sample dilutions for their particular experiments.
3. If the samples are not indicated in the manual, a preliminary experiment to determine the validity of the kit is
necessary.
4. If a lysis buffer is used to prepare tissue homogenates or cell culture supernatant, there is a possibility to cause a
deviation due to the introduced chemical substance.
5. Some recombinant protein may not be detected due to a mismatching with the coated antibody or detection
antibody.
6. Please do not use hemolyzed samples for ELISA as it will affect the test results.
7. Fresh samples without long time storage is recommended for the test. Otherwise, protein degradation and
denaturalization may occur in those samples and finally lead to wrong result.
Reagent Preparation
1. Bring all components to room temperature (18~25℃) before use. Follow the Microplate reader manual for set-
up and preheat it for 15 min before OD measurement.
2. Sample preparation: Add 40μL of Sample dilution to EP tube, then add 10μL of sample (sample final dilution
is 5-fold), mix thoroughly.
3. Wash Buffer: Dilute 25mL of Concentrated Wash Buffer with 475 mL of deionized or distilled water to prepare
500 mL of Wash Buffer.
Assay Protocol
1. Determine wells for diluted standard, sample and blank separately (blank wells don’t add sample and HRP-
Conjugate reagent, other each step operation is same). Prepare 5 wells for standard points, 1 well for blank, add
50μL of standard solution to standard well, add sample to wells, respectively.
2. Add 100μL of HRP-Conjugate reagent to each well, except blank well, closing plate with plate Sealer, incubate
for 60 min at 37℃.
3. Uncover plate Sealer, discard Liquid, add washing buffer to all wells, still for 30s then drain, repeat 5 times, dry
by pat.
4. Add 100μL of the equal volume mixture of Substrate Reagent A and Substrate Reagent B to each well, evade the
light preservation for 15min at 37℃.
5. Add 50μL of Stop Solution to each well, Stop the reaction.
6. Read absorbance at 450nm within 15min.
Wuhan Feiyue Biotechnology CO.,LTD
4
Calculation of results
Average the duplicate readings for each standard and samples, then subtract the average zero standard optical
density. Create a standard curve with standard concentration on the x-axis and OD values on the y-axis. Draw a
best fit curve through the points and it can be determined by regression analysis. If samples have been diluted, the
concentration read from the standard curve must be multiplied by the dilution factor.
Example data
The OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g.
operator, pipetting technique, washing technique or temperature effects), so plotting log of the data to establish a
standard curve for each test is strongly recommended. Typical standard curve is provided below for reference only.
Sensitivity
The minimum detectable dose of Goat IL-1 is typically less than 10pg/mL.
The sensitivity of this assay, or Limit of Detection (LOD) was defined as the lowest protein concentration that
could be differentiated from zero. It was determined by adding three standard deviations to the mean OD value of
twenty zero standard replicates and calculating the corresponding concentration.
Specificity
This assay has high sensitivity and excellent specificity for detection of Goat IL-1.
No significant cross-reactivity or interference between Goat IL-1 and analogues was observed.
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Goat IL-1 were tested
on one plate, 8 replicates in the plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Goat IL-1 were tested
on 3 different plates, 8 replicates in each plate.
CV(%) = SD/mean × 100%
Intra-Assay: CV< 10%
Inter-Assay: CV< 12%
Recovery
Three matrices listed below were spiked with certain level of Goat IL-1, The recovery rates of Goat IL-1 were
calculated by comparing the measured value to the expected amount of Goat IL-1 in samples.
Matrix type Recovery Range (%) Average (%)
Serum (n=5) 91-101 97
EDTA plasma (n=5) 95-115 98
Cell culture media (n=5) 95-106 98
Wuhan Feiyue Biotechnology CO.,LTD
5
Linearity
Three types of sample were spiked with appropriate concentrations of Goat IL-1 and diluted into a series of
concentration gradients, then the linearity of the assay was demonstrated by the percentage of comparing calculated
concentrations and expected values.
Dilution
times
Recovery Range (%)
Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 94-102 93-110 92-111
1:4 83-103 93-100 83-91
1:8 98-106 95-111 98-101
1:16 87-105 96-103 101-109
Summary
1. Add 50μL standard or sample to each well.
3. Aspirate and wash 5 times
2. Add 100μL HRP-Conjugate reagent. Incubate 60 min at 37℃
4. Add 100μL Substrate Reagent(A+B). Incubate 15 minutes at 37℃
5. Add 50μL Stop Solution
7. Calculation of results
6. Read at 450nm within 15min
Wuhan Feiyue Biotechnology CO.,LTD
6
Troubleshooting
Problem Possible Causes Subsequent Actions
Poor standard curve
Improper standard dilution
Ensure that standards are dissolved and
diluted in the recommended manner
Inaccurate pipetting
Periodically calibrate pipettes and check
the pipette tips
Evaporate the reaction solution Seal the enzyme plate with plate sealer
Incomplete plate washing
Adequate washing times and the amount
of washing solution added
Foreign matter in the bottom of wells
Clean the bottom of the plate before
reading
Weak or no color
development
Insufficient reaction of reagents
Ensure incubation time and incubate at
the recommended temperature
Inadequate reagent volumes
Check the pipette and Follow the steps
strictly to operate
Improper dilution Check the reagent dilution process
Inactivation of enzyme conjugate
Mix conjugate and substrate, check by
color development
Low OD value
Incorrect setting of microplate reader Check the wavelength of reader
No stop solution added Add appropriate amount of stop solution
Waiting too long time to read Read the plate in time
High background
Contaminated chromogenic solution Replace chromogenic solution
Coloring time is too long Control the coloring time
Wrong dilution of reaction reagent Use the recommended dilution
Inadequate washing of the plate
Adequate washing times and the amount
of washing solution added

Weitere ähnliche Inhalte

Mehr von DaisyFeiyue

Feiyuebio research-Monkeypox.pdf
Feiyuebio research-Monkeypox.pdfFeiyuebio research-Monkeypox.pdf
Feiyuebio research-Monkeypox.pdfDaisyFeiyue
 
S100 protein you need to know from Feiyuebio.pdf
S100 protein you need to know from Feiyuebio.pdfS100 protein you need to know from Feiyuebio.pdf
S100 protein you need to know from Feiyuebio.pdfDaisyFeiyue
 
Feiyuebio S100 protein.pdf
Feiyuebio S100 protein.pdfFeiyuebio S100 protein.pdf
Feiyuebio S100 protein.pdfDaisyFeiyue
 
Feiyuebio research-Infectious Diseases.pdf
Feiyuebio research-Infectious Diseases.pdfFeiyuebio research-Infectious Diseases.pdf
Feiyuebio research-Infectious Diseases.pdfDaisyFeiyue
 
Feiyuebio detailed introduction with QC tech service.pdf
Feiyuebio detailed introduction with QC tech service.pdfFeiyuebio detailed introduction with QC tech service.pdf
Feiyuebio detailed introduction with QC tech service.pdfDaisyFeiyue
 
Feiyuebio detailed introduction.pdf
Feiyuebio detailed introduction.pdfFeiyuebio detailed introduction.pdf
Feiyuebio detailed introduction.pdfDaisyFeiyue
 
Feiyuebio introduction.pdf
Feiyuebio introduction.pdfFeiyuebio introduction.pdf
Feiyuebio introduction.pdfDaisyFeiyue
 
Feiyuebio hot sale promotion discount.pdf
Feiyuebio hot sale promotion discount.pdfFeiyuebio hot sale promotion discount.pdf
Feiyuebio hot sale promotion discount.pdfDaisyFeiyue
 
Feiyuebio ELISA May promotion.pdf
Feiyuebio ELISA May promotion.pdfFeiyuebio ELISA May promotion.pdf
Feiyuebio ELISA May promotion.pdfDaisyFeiyue
 
Protein Daisy@feiyuebio.com.pdf
Protein Daisy@feiyuebio.com.pdfProtein Daisy@feiyuebio.com.pdf
Protein Daisy@feiyuebio.com.pdfDaisyFeiyue
 
Feiyuebio elisa kit
Feiyuebio elisa kitFeiyuebio elisa kit
Feiyuebio elisa kitDaisyFeiyue
 
Feiyuebio human tnf α elisa kit
Feiyuebio human tnf α elisa kitFeiyuebio human tnf α elisa kit
Feiyuebio human tnf α elisa kitDaisyFeiyue
 

Mehr von DaisyFeiyue (12)

Feiyuebio research-Monkeypox.pdf
Feiyuebio research-Monkeypox.pdfFeiyuebio research-Monkeypox.pdf
Feiyuebio research-Monkeypox.pdf
 
S100 protein you need to know from Feiyuebio.pdf
S100 protein you need to know from Feiyuebio.pdfS100 protein you need to know from Feiyuebio.pdf
S100 protein you need to know from Feiyuebio.pdf
 
Feiyuebio S100 protein.pdf
Feiyuebio S100 protein.pdfFeiyuebio S100 protein.pdf
Feiyuebio S100 protein.pdf
 
Feiyuebio research-Infectious Diseases.pdf
Feiyuebio research-Infectious Diseases.pdfFeiyuebio research-Infectious Diseases.pdf
Feiyuebio research-Infectious Diseases.pdf
 
Feiyuebio detailed introduction with QC tech service.pdf
Feiyuebio detailed introduction with QC tech service.pdfFeiyuebio detailed introduction with QC tech service.pdf
Feiyuebio detailed introduction with QC tech service.pdf
 
Feiyuebio detailed introduction.pdf
Feiyuebio detailed introduction.pdfFeiyuebio detailed introduction.pdf
Feiyuebio detailed introduction.pdf
 
Feiyuebio introduction.pdf
Feiyuebio introduction.pdfFeiyuebio introduction.pdf
Feiyuebio introduction.pdf
 
Feiyuebio hot sale promotion discount.pdf
Feiyuebio hot sale promotion discount.pdfFeiyuebio hot sale promotion discount.pdf
Feiyuebio hot sale promotion discount.pdf
 
Feiyuebio ELISA May promotion.pdf
Feiyuebio ELISA May promotion.pdfFeiyuebio ELISA May promotion.pdf
Feiyuebio ELISA May promotion.pdf
 
Protein Daisy@feiyuebio.com.pdf
Protein Daisy@feiyuebio.com.pdfProtein Daisy@feiyuebio.com.pdf
Protein Daisy@feiyuebio.com.pdf
 
Feiyuebio elisa kit
Feiyuebio elisa kitFeiyuebio elisa kit
Feiyuebio elisa kit
 
Feiyuebio human tnf α elisa kit
Feiyuebio human tnf α elisa kitFeiyuebio human tnf α elisa kit
Feiyuebio human tnf α elisa kit
 

Kürzlich hochgeladen

Best VIP Call Girls Noida Sector 40 Call Me: 8448380779
Best VIP Call Girls Noida Sector 40 Call Me: 8448380779Best VIP Call Girls Noida Sector 40 Call Me: 8448380779
Best VIP Call Girls Noida Sector 40 Call Me: 8448380779Delhi Call girls
 
HONOR Veterans Event Keynote by Michael Hawkins
HONOR Veterans Event Keynote by Michael HawkinsHONOR Veterans Event Keynote by Michael Hawkins
HONOR Veterans Event Keynote by Michael HawkinsMichael W. Hawkins
 
The Path to Product Excellence: Avoiding Common Pitfalls and Enhancing Commun...
The Path to Product Excellence: Avoiding Common Pitfalls and Enhancing Commun...The Path to Product Excellence: Avoiding Common Pitfalls and Enhancing Commun...
The Path to Product Excellence: Avoiding Common Pitfalls and Enhancing Commun...Aggregage
 
Call Girls Pune Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Pune Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Pune Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Pune Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
FULL ENJOY Call Girls In Majnu Ka Tilla, Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Majnu Ka Tilla, Delhi Contact Us 8377877756FULL ENJOY Call Girls In Majnu Ka Tilla, Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Majnu Ka Tilla, Delhi Contact Us 8377877756dollysharma2066
 
The Coffee Bean & Tea Leaf(CBTL), Business strategy case study
The Coffee Bean & Tea Leaf(CBTL), Business strategy case studyThe Coffee Bean & Tea Leaf(CBTL), Business strategy case study
The Coffee Bean & Tea Leaf(CBTL), Business strategy case studyEthan lee
 
Call Girls Jp Nagar Just Call 👗 7737669865 👗 Top Class Call Girl Service Bang...
Call Girls Jp Nagar Just Call 👗 7737669865 👗 Top Class Call Girl Service Bang...Call Girls Jp Nagar Just Call 👗 7737669865 👗 Top Class Call Girl Service Bang...
Call Girls Jp Nagar Just Call 👗 7737669865 👗 Top Class Call Girl Service Bang...amitlee9823
 
Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...
Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...
Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...anilsa9823
 
Famous Olympic Siblings from the 21st Century
Famous Olympic Siblings from the 21st CenturyFamous Olympic Siblings from the 21st Century
Famous Olympic Siblings from the 21st Centuryrwgiffor
 
Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...Dipal Arora
 
A DAY IN THE LIFE OF A SALESMAN / WOMAN
A DAY IN THE LIFE OF A  SALESMAN / WOMANA DAY IN THE LIFE OF A  SALESMAN / WOMAN
A DAY IN THE LIFE OF A SALESMAN / WOMANIlamathiKannappan
 
KYC-Verified Accounts: Helping Companies Handle Challenging Regulatory Enviro...
KYC-Verified Accounts: Helping Companies Handle Challenging Regulatory Enviro...KYC-Verified Accounts: Helping Companies Handle Challenging Regulatory Enviro...
KYC-Verified Accounts: Helping Companies Handle Challenging Regulatory Enviro...Any kyc Account
 
Call Girls Electronic City Just Call 👗 7737669865 👗 Top Class Call Girl Servi...
Call Girls Electronic City Just Call 👗 7737669865 👗 Top Class Call Girl Servi...Call Girls Electronic City Just Call 👗 7737669865 👗 Top Class Call Girl Servi...
Call Girls Electronic City Just Call 👗 7737669865 👗 Top Class Call Girl Servi...amitlee9823
 
Call Girls In Panjim North Goa 9971646499 Genuine Service
Call Girls In Panjim North Goa 9971646499 Genuine ServiceCall Girls In Panjim North Goa 9971646499 Genuine Service
Call Girls In Panjim North Goa 9971646499 Genuine Serviceritikaroy0888
 
Mysore Call Girls 8617370543 WhatsApp Number 24x7 Best Services
Mysore Call Girls 8617370543 WhatsApp Number 24x7 Best ServicesMysore Call Girls 8617370543 WhatsApp Number 24x7 Best Services
Mysore Call Girls 8617370543 WhatsApp Number 24x7 Best ServicesDipal Arora
 
MONA 98765-12871 CALL GIRLS IN LUDHIANA LUDHIANA CALL GIRL
MONA 98765-12871 CALL GIRLS IN LUDHIANA LUDHIANA CALL GIRLMONA 98765-12871 CALL GIRLS IN LUDHIANA LUDHIANA CALL GIRL
MONA 98765-12871 CALL GIRLS IN LUDHIANA LUDHIANA CALL GIRLSeo
 
Yaroslav Rozhankivskyy: Три складові і три передумови максимальної продуктивн...
Yaroslav Rozhankivskyy: Три складові і три передумови максимальної продуктивн...Yaroslav Rozhankivskyy: Три складові і три передумови максимальної продуктивн...
Yaroslav Rozhankivskyy: Три складові і три передумови максимальної продуктивн...Lviv Startup Club
 
Enhancing and Restoring Safety & Quality Cultures - Dave Litwiller - May 2024...
Enhancing and Restoring Safety & Quality Cultures - Dave Litwiller - May 2024...Enhancing and Restoring Safety & Quality Cultures - Dave Litwiller - May 2024...
Enhancing and Restoring Safety & Quality Cultures - Dave Litwiller - May 2024...Dave Litwiller
 
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756dollysharma2066
 

Kürzlich hochgeladen (20)

Best VIP Call Girls Noida Sector 40 Call Me: 8448380779
Best VIP Call Girls Noida Sector 40 Call Me: 8448380779Best VIP Call Girls Noida Sector 40 Call Me: 8448380779
Best VIP Call Girls Noida Sector 40 Call Me: 8448380779
 
HONOR Veterans Event Keynote by Michael Hawkins
HONOR Veterans Event Keynote by Michael HawkinsHONOR Veterans Event Keynote by Michael Hawkins
HONOR Veterans Event Keynote by Michael Hawkins
 
The Path to Product Excellence: Avoiding Common Pitfalls and Enhancing Commun...
The Path to Product Excellence: Avoiding Common Pitfalls and Enhancing Commun...The Path to Product Excellence: Avoiding Common Pitfalls and Enhancing Commun...
The Path to Product Excellence: Avoiding Common Pitfalls and Enhancing Commun...
 
Call Girls Pune Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Pune Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Pune Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Pune Just Call 9907093804 Top Class Call Girl Service Available
 
FULL ENJOY Call Girls In Majnu Ka Tilla, Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Majnu Ka Tilla, Delhi Contact Us 8377877756FULL ENJOY Call Girls In Majnu Ka Tilla, Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Majnu Ka Tilla, Delhi Contact Us 8377877756
 
The Coffee Bean & Tea Leaf(CBTL), Business strategy case study
The Coffee Bean & Tea Leaf(CBTL), Business strategy case studyThe Coffee Bean & Tea Leaf(CBTL), Business strategy case study
The Coffee Bean & Tea Leaf(CBTL), Business strategy case study
 
Call Girls Jp Nagar Just Call 👗 7737669865 👗 Top Class Call Girl Service Bang...
Call Girls Jp Nagar Just Call 👗 7737669865 👗 Top Class Call Girl Service Bang...Call Girls Jp Nagar Just Call 👗 7737669865 👗 Top Class Call Girl Service Bang...
Call Girls Jp Nagar Just Call 👗 7737669865 👗 Top Class Call Girl Service Bang...
 
Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...
Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...
Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...
 
Famous Olympic Siblings from the 21st Century
Famous Olympic Siblings from the 21st CenturyFamous Olympic Siblings from the 21st Century
Famous Olympic Siblings from the 21st Century
 
Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...
 
A DAY IN THE LIFE OF A SALESMAN / WOMAN
A DAY IN THE LIFE OF A  SALESMAN / WOMANA DAY IN THE LIFE OF A  SALESMAN / WOMAN
A DAY IN THE LIFE OF A SALESMAN / WOMAN
 
KYC-Verified Accounts: Helping Companies Handle Challenging Regulatory Enviro...
KYC-Verified Accounts: Helping Companies Handle Challenging Regulatory Enviro...KYC-Verified Accounts: Helping Companies Handle Challenging Regulatory Enviro...
KYC-Verified Accounts: Helping Companies Handle Challenging Regulatory Enviro...
 
Call Girls Electronic City Just Call 👗 7737669865 👗 Top Class Call Girl Servi...
Call Girls Electronic City Just Call 👗 7737669865 👗 Top Class Call Girl Servi...Call Girls Electronic City Just Call 👗 7737669865 👗 Top Class Call Girl Servi...
Call Girls Electronic City Just Call 👗 7737669865 👗 Top Class Call Girl Servi...
 
Call Girls In Panjim North Goa 9971646499 Genuine Service
Call Girls In Panjim North Goa 9971646499 Genuine ServiceCall Girls In Panjim North Goa 9971646499 Genuine Service
Call Girls In Panjim North Goa 9971646499 Genuine Service
 
Mysore Call Girls 8617370543 WhatsApp Number 24x7 Best Services
Mysore Call Girls 8617370543 WhatsApp Number 24x7 Best ServicesMysore Call Girls 8617370543 WhatsApp Number 24x7 Best Services
Mysore Call Girls 8617370543 WhatsApp Number 24x7 Best Services
 
MONA 98765-12871 CALL GIRLS IN LUDHIANA LUDHIANA CALL GIRL
MONA 98765-12871 CALL GIRLS IN LUDHIANA LUDHIANA CALL GIRLMONA 98765-12871 CALL GIRLS IN LUDHIANA LUDHIANA CALL GIRL
MONA 98765-12871 CALL GIRLS IN LUDHIANA LUDHIANA CALL GIRL
 
Yaroslav Rozhankivskyy: Три складові і три передумови максимальної продуктивн...
Yaroslav Rozhankivskyy: Три складові і три передумови максимальної продуктивн...Yaroslav Rozhankivskyy: Три складові і три передумови максимальної продуктивн...
Yaroslav Rozhankivskyy: Три складові і три передумови максимальної продуктивн...
 
Forklift Operations: Safety through Cartoons
Forklift Operations: Safety through CartoonsForklift Operations: Safety through Cartoons
Forklift Operations: Safety through Cartoons
 
Enhancing and Restoring Safety & Quality Cultures - Dave Litwiller - May 2024...
Enhancing and Restoring Safety & Quality Cultures - Dave Litwiller - May 2024...Enhancing and Restoring Safety & Quality Cultures - Dave Litwiller - May 2024...
Enhancing and Restoring Safety & Quality Cultures - Dave Litwiller - May 2024...
 
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
 

Goat IL-1 ELISA Kit

  • 1. Wuhan Feiyue Biotechnology CO.,LTD 1 FOR RESEARCH USE ONLY NOT FOR IN VITRO CLINICAL DIAGNOSIS Goat Interleukin 1 (IL-1) ELISA Kit Catalog No: FY-EG4791 Size: 96T This manual must be read attentively and completely before using this product. If you have any problems, please contact us. Phone: 86-027-87002654 86-180 6275 0658 Email: info@feiyuebio.com Website: www.feiyuebio.com Intended Use The ELISA kit is a Sandwich enzyme immunoassay technique for the quantitative determination of Goat IL-1 concentrations in serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids. Specification  Sensitivity: 10pg/mL.  Detection Range: 10-400pg/mL.  No significant cross-reactivity or interference between Goat IL-1 and analogues was observed.  Repeatability: Coefficient of variation is < 12%. Principle of the Procedure This ELISA kit uses the Sandwich-ELISA principle. The micro plate provided in this kit has been pre-coated with an antibody specific to Goat IL-1. Standards or samples and Horseradish Peroxidase (HRP) labeled detection antibody specific for Goat IL-1 are added to the plate wells together and incubated. After washing off unbound material, TMB substrate solution is added to all wells and incubated. An enzyme-catalyzed reaction generates a blue color in the solution, thereafter, stop solution is added to stop the substrate reaction and the color turns yellow. The yellow solution is read at a wavelength of 450nm. The concentration of Goat IL-1 in the samples is then calculated from the OD value by establishing a standard curve. Limitations of the Procedure 1、This kit is for laboratory scientific research only, we will not be responsible for any consequences if this kit is used for clinical diagnosis or any other procedures. 2、 Due to the uncertainty of its validity, this kit may not be suitable for testing some special experimental samples, such as gene knockout experiments. 3、This kit should be used before its expiration date, and please strictly follow the instructions for storage. 4、Different manufacturers' kits or testing the same analyte by other methods may produce inconsistent results because we do not compare our products with those of other manufacturers.
  • 2. Wuhan Feiyue Biotechnology CO.,LTD 2 5、Since the antibodies used in the kit are usually prepared from recombinant proteins as immunogen, and recombinant proteins can be limited by different fragmentation, expression and purification systems, it is not recommended to use this kit to detect recombinant proteins 6、In order to get the best experimental results, please use only the reagents provided by the manufacturer, and do not mix reagents from different batches and other manufacturers. 7、Due to the existing conditions and the limitations of science and technology, we cannot fully identify and analyze the raw materials provided by the supplier comprehensively. Therefore, the kit may have some quality and technical risks. 8、The possibility of interference cannot be excluded before all factors are tested in the ELISA immunoassay. 9、In order to obtain reproducible results, each step in the experiment should be controlled and variations in sample collection, handling and storage may also lead to differences in sample measurements. 10、Although each kit passes rigorous quality testing, differences in measured values between batches of kits can still be caused by factors such as shipping conditions and different laboratory equipment. Reagents&Materials Provided Materials & Equipment Required But Not Provided Microplate reader with 450nm wavelength filter Incubator capable of maintaining 37℃ Single or multi-channel pipettes with high precision disposable pipette tips EP tubes Container for Wash Solution Squirt bottle, manifold dispenser, or automated microplate washer Deionized or distilled water Absorbent paper for blotting the microplate Sample Collection Serum: Allow blood samples to clot for 2 hours at room temperature or overnight at 2-8℃ before centrifugation for 15 min at 1000×g at 2~8℃. Supernatants should be taken for assay testing immediately or stored at -20℃ or - 80℃ for later use. Plasma: Collect plasma using EDTA or heparin as an anticoagulant. Centrifuge samples for 15 min at 1000×g at 2~8℃ within 30 min of collection. Supernatants should be taken for assay testing immediately or stored at -20℃ or -80℃ for later use. Components Specifications Storage Pre-coated ELISA Plate 12wells ×8 strips 2-8℃ Standard 6×0.3mL:400,200,100,50,25,0pg/mL 2-8℃ HRP-Conjugate reagent 1×10mL 2-8℃ Sample Diluent 1×6mL 2-8℃ Substrate Reagent A 1×6mL 2-8℃ Substrate Reagent B 1×6mL 2-8℃ Wash Buffer (20×) 1×25mL 2-8℃ Stop Solution 1×6mL 2-8℃ Plate Sealer 2 pieces Manual 1 copy Self-sealing bags 1pc
  • 3. Wuhan Feiyue Biotechnology CO.,LTD 3 Tissue homogenates: Tissues should be rinsed in ice-cold PBS to remove excess blood thoroughly, weighed, minced into small pieces. then Tissue pieces omogenized in PBS (tissue weight (g): PBS (mL) volume=1:9) with a glass homogenizer on ice. The resulting suspension is sonicated with an ultrasonic cell disrupter till the solution is clarified. The homogenates are then centrifuged for 5 min at 10000×g. Supernatants should be taken for assay testing immediately or stored at -20℃ or -80℃ for later use. Cell lysates: For adherent cells, gently wash the cells with moderate amount of cool PBS and detach the cells with trypsin. Collect cells by centrifugation for 5 min at 1000×g(suspension cells can be collected by centrifugation directly). Discard the supernate and wash cells 3 times with cool PBS. Resuspend cells in cool PBS with concentration of 5×106 cells/mL. Repeat the freeze-thaw process several times until the cells are fully lysed. Centrifuge for 10min at 1500×g at 2-8℃. Supernatants should be taken for assay testing immediately or stored at - 20℃ or -80℃ for later use. Cell culture supernatant and other biological fluids: Centrifuge samples for 20 minutes at 1000xg at 2-8℃. Supernatants should be taken for assay testing immediately or stored at -20℃ or -80℃ for later use. Note for sample 1. Samples should be used within 6 days when stored at 2-8℃, otherwise samples must be stored at -20℃ (≤1month) or -80℃ (≤2months). Avoid repeated freeze-thaw cycles. 2. Please predict the concentration before assaying. If the sample concentration is not within the range of the standard curve, users must determine the optimal sample dilutions for their particular experiments. 3. If the samples are not indicated in the manual, a preliminary experiment to determine the validity of the kit is necessary. 4. If a lysis buffer is used to prepare tissue homogenates or cell culture supernatant, there is a possibility to cause a deviation due to the introduced chemical substance. 5. Some recombinant protein may not be detected due to a mismatching with the coated antibody or detection antibody. 6. Please do not use hemolyzed samples for ELISA as it will affect the test results. 7. Fresh samples without long time storage is recommended for the test. Otherwise, protein degradation and denaturalization may occur in those samples and finally lead to wrong result. Reagent Preparation 1. Bring all components to room temperature (18~25℃) before use. Follow the Microplate reader manual for set- up and preheat it for 15 min before OD measurement. 2. Sample preparation: Add 40μL of Sample dilution to EP tube, then add 10μL of sample (sample final dilution is 5-fold), mix thoroughly. 3. Wash Buffer: Dilute 25mL of Concentrated Wash Buffer with 475 mL of deionized or distilled water to prepare 500 mL of Wash Buffer. Assay Protocol 1. Determine wells for diluted standard, sample and blank separately (blank wells don’t add sample and HRP- Conjugate reagent, other each step operation is same). Prepare 5 wells for standard points, 1 well for blank, add 50μL of standard solution to standard well, add sample to wells, respectively. 2. Add 100μL of HRP-Conjugate reagent to each well, except blank well, closing plate with plate Sealer, incubate for 60 min at 37℃. 3. Uncover plate Sealer, discard Liquid, add washing buffer to all wells, still for 30s then drain, repeat 5 times, dry by pat. 4. Add 100μL of the equal volume mixture of Substrate Reagent A and Substrate Reagent B to each well, evade the light preservation for 15min at 37℃. 5. Add 50μL of Stop Solution to each well, Stop the reaction. 6. Read absorbance at 450nm within 15min.
  • 4. Wuhan Feiyue Biotechnology CO.,LTD 4 Calculation of results Average the duplicate readings for each standard and samples, then subtract the average zero standard optical density. Create a standard curve with standard concentration on the x-axis and OD values on the y-axis. Draw a best fit curve through the points and it can be determined by regression analysis. If samples have been diluted, the concentration read from the standard curve must be multiplied by the dilution factor. Example data The OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), so plotting log of the data to establish a standard curve for each test is strongly recommended. Typical standard curve is provided below for reference only. Sensitivity The minimum detectable dose of Goat IL-1 is typically less than 10pg/mL. The sensitivity of this assay, or Limit of Detection (LOD) was defined as the lowest protein concentration that could be differentiated from zero. It was determined by adding three standard deviations to the mean OD value of twenty zero standard replicates and calculating the corresponding concentration. Specificity This assay has high sensitivity and excellent specificity for detection of Goat IL-1. No significant cross-reactivity or interference between Goat IL-1 and analogues was observed. Precision Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Goat IL-1 were tested on one plate, 8 replicates in the plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Goat IL-1 were tested on 3 different plates, 8 replicates in each plate. CV(%) = SD/mean × 100% Intra-Assay: CV< 10% Inter-Assay: CV< 12% Recovery Three matrices listed below were spiked with certain level of Goat IL-1, The recovery rates of Goat IL-1 were calculated by comparing the measured value to the expected amount of Goat IL-1 in samples. Matrix type Recovery Range (%) Average (%) Serum (n=5) 91-101 97 EDTA plasma (n=5) 95-115 98 Cell culture media (n=5) 95-106 98
  • 5. Wuhan Feiyue Biotechnology CO.,LTD 5 Linearity Three types of sample were spiked with appropriate concentrations of Goat IL-1 and diluted into a series of concentration gradients, then the linearity of the assay was demonstrated by the percentage of comparing calculated concentrations and expected values. Dilution times Recovery Range (%) Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5) 1:2 94-102 93-110 92-111 1:4 83-103 93-100 83-91 1:8 98-106 95-111 98-101 1:16 87-105 96-103 101-109 Summary 1. Add 50μL standard or sample to each well. 3. Aspirate and wash 5 times 2. Add 100μL HRP-Conjugate reagent. Incubate 60 min at 37℃ 4. Add 100μL Substrate Reagent(A+B). Incubate 15 minutes at 37℃ 5. Add 50μL Stop Solution 7. Calculation of results 6. Read at 450nm within 15min
  • 6. Wuhan Feiyue Biotechnology CO.,LTD 6 Troubleshooting Problem Possible Causes Subsequent Actions Poor standard curve Improper standard dilution Ensure that standards are dissolved and diluted in the recommended manner Inaccurate pipetting Periodically calibrate pipettes and check the pipette tips Evaporate the reaction solution Seal the enzyme plate with plate sealer Incomplete plate washing Adequate washing times and the amount of washing solution added Foreign matter in the bottom of wells Clean the bottom of the plate before reading Weak or no color development Insufficient reaction of reagents Ensure incubation time and incubate at the recommended temperature Inadequate reagent volumes Check the pipette and Follow the steps strictly to operate Improper dilution Check the reagent dilution process Inactivation of enzyme conjugate Mix conjugate and substrate, check by color development Low OD value Incorrect setting of microplate reader Check the wavelength of reader No stop solution added Add appropriate amount of stop solution Waiting too long time to read Read the plate in time High background Contaminated chromogenic solution Replace chromogenic solution Coloring time is too long Control the coloring time Wrong dilution of reaction reagent Use the recommended dilution Inadequate washing of the plate Adequate washing times and the amount of washing solution added