DSA  ITS APPLICATION AND ADVANCEMENT Moderator: MR. S.R.CHOWDHURY TUTOR, DEPT.OF RADIO-DIAGNOSIS & IMAGING, PGIMER CHANDIGARH Presented by: DEEPAK GUPTA BSc  final year Student , DEPT.OF RADIO-DIAGNOSIS & IMAGING PGIMER CHANDIGARH
2/24/2010 2 ANGIO? ANGIO  means blood vessel And angiography is the radiological study of blood vessel in the body  after the  introduction of iodinated contrast media. SUBTRACTION?   It is simply a technique by which bone structures images are subtracted or canceled out from a film of bones plus opacified vessels, leaving an unobscured image of the vessels.
2/24/2010 3 WHAT DO YOU MEAN BY DSA DSA-The acquisition of digital fluoroscopic images combined with injection of contrast material and real-time subtraction of pre- and post contrast images to perform angiography is referred to as digital subtraction angiography
2/24/2010 4 HISTORY The Portuguese neurologist Egas Moniz,( Nobel Prize winner 1949), in 1927developed the technique of contrast x-ray cerebral angiography to diagnose diseases, such as tumors and arteriovenous malformations. The idea of subtraction images was first proposed by the Dutch radiologist Ziedses des Plantes in the 1935, when he was able to produce subtracted images using plain films. With the introduction of the Seldinger technique in 1953, the procedure became safer as no sharp devices need to remain inside the vascular lumen.
2/24/2010 5 HISTORICAL DEVELOPMENT CONVENTIONAL SUBTRACTION TECHNIQUE Photographic method used to eliminate unwanted images. No addition of information;only purpose to make diagnostically important information to see. First described by a Dutch radiologist, Zeides des Plantes, 1935. 3 conditions:SCOUT FILM                     ANGIOGRAM FILM-CONTRAST                                    NO MOTION OF HEAD
2/24/2010 6 CONTD. The principles of subtraction are based on the following: The scout film shows the structural details of the skull and the adjacent soft tissue. Angiogram film shows exactly the same anatomic details, if the patient does not move, plus the opacified blood vessels. If all the information in the scout film could be subtracted from the angiogram film, only the opacified vessel pattern would remain visible.
2/24/2010 7 EQUIPMENT THE  AOT CHANGER: Can change films –6/s or 1/5s   Two of these changers can be coupled electronically and mechanically and will operate out of phase or synchronously with each other. Radiographs in 2 planes simultaneously-one mounted vertically and other horizontal.
2/24/2010 8 AOT CONTD. LOADING MAGZINE Large steel container. Inside the container are arranged a no. of strong wire separators. 30 sheets of film. Precautions :each film should slide into position by it’s own wt. RECIEVING / LOADING MAGZINE
2/24/2010 9 EXPOSURE AREA : Rectangular, slightly recessed area at the top of the changer. Defined by heavy metal frame. Within this is a secondary grid. Supported on 4 small helical springs placed cornerwise and to consist of  a stout Al plate, to the lower side of which is attached an intensifying screen. 30 SEPERATING WIRES  RECEIVING CASSETTE: Shallow,polished metal container. Lid is a sliding section in a side which can be pushed down  by firm finger pressure and is retained by central spring loaded catch.
2/24/2010 10 THE ROLL FILM CHANGER: Mechanically easier to wind roll films. 12/s.
2/24/2010 11 DIGITAL  SUBTRACTION  ANGIOGRAPHY HISTORY: Developed in 1970s. University of Wisconsin, University of Arizona,University of Kiel. Commercial systems introduced in 1980.
2/24/2010 12 Angiography theory ages  Single shot angiography Manual changer AOT PUCK DSA
2/24/2010 13 EQUIPMENT AND   APPARATUS Both film and digital capabilities are usually present and the fluoroscopy equipment operates in the conventional way. Brief  review : Light detected amplified Input surface Visible light Output phosphor Electrical signal Light intensity
2/24/2010 14 DIGITAL IMAGE PROCESSOR DISPLAY  SECTION DIGITAL ARITHEMATIC  UNIT DIGITIZER DIGITAL IMAGE  MEMORY X-RAY CONTROL SIGNALS MONITOR COMPUTER MULTIFORMAT  CAMERA DIGITAL DISK STORAGE
2/24/2010 15 X-Ray Unit Should be capable of rapid serial exp.  @2-8 frames/sec for up to 20 sec. Near monochromatic beam consistent output. Isocentre motorized table ,tilt 90-0-90 , up-down,     sliding ,floating capable   Ceiling mounted C-arm with versatile movements II system with multi field ( 6”,9”,12”,16”) High resolution TV /CCD camera Computer with high definition monitor system
2/24/2010 16 Contd. High output preferably high frequency generator High heat loading capacity ,fast heat dissipation, high speed anode Target material tungsten,rhenium,graphite. Oil-water close loop cooling
2/24/2010 17 In our dept.Siemens polydoros SX 65-80 Polystar
2/24/2010 18 NECESSARY EQUIPMENT USED WITH ANGIO(DSA) GAS , AIR & VACUUM  PORTS Boyle apparatus
2/24/2010 19 BED SIDE  MONITOR INJECTOR
2/24/2010 20
2/24/2010 21 IMAGE INTENSIFIER-CONSTRUCTION
2/24/2010 22 WORKING
2/24/2010 23    Modern DSA systems are based on digital fluoroscopy/fluorography systems, which are equipped with special software and display facilities. Digital subtraction angiography (DSA) was developed to improve vessel contrast. This is a technique that uses a computer to subtract two images, obtained before and after contrast media is injected into the vessels of interest. The anatomical structures that are the same in the two images can be removed and the resulting image shows the vessels only.
2/24/2010 24  The image before the contrast agent is administered is called the mask image. Once the contrast is administered, a sequence of images are taken by a television camera in analog form, which is then digitised by computer.  The DSA processor has two separate image memories, one for the mask and the other for the images with contrast medium. These two image memories are subtracted from one another arithmetically, and the result goes to an image processing and display unit.
2/24/2010 25 IMAGE  PROCESSING  Adjust contrast & brightness. Remasking – correcting misregistered images. Pixel shifting. Edge enhancement – edges of the vessels can be enhanced so that small details can be made more obvious. Image zoom. land marking – a small amount of original image is added into the subtracted image. Noise smoothing – operates by reducing the statistical fluctuations in each pixel by averaging the pixel with it's closest neighbors. The visual prominence of noise has been suppressed by averaging, but resolution is decreased.
2/24/2010 26 Road Mapping  Road mapping is useful for the placement of catheters and wires in complex and small vasculature. A DSA sequence is performed, and the frame with maximum vessel opacification is identified;this frame becomes the road map mask. The road map mask is subtracted from subsequent live fluoroscopic images to produce real-time subtracted fluoroscopic images overlaid on a static image of the vasculature
2/24/2010 27     When a digital subtraction technique is used, patient motion that occurs between acquisition of the precontrast images and acquisition of the postcontrast images will result in artifacts due to misregistration of the two images. If these arti-facts are observed, it is possible to reregister the pre- and postcontrast images by shifting the subtraction mask (precontrast image) with respect to the postcontrast image and resubtracting the two images
2/24/2010 28 VIDICON CAMERA TUBE  ,[object Object]
Length 6 inch
Main parts -
Target section
Electron gun
Electromagnetic focusing coil
Two pair of electrostatic deflecting coil
A scanning section,[object Object]
2/24/2010 30 SCANNING SECTION/ SYSTEMS  ,[object Object]
The electrons are focused on the target of tube and
The beam is moved over the target area in an orderly scanning scan
Two types scanning system
Sequential scanning pattern and
Interlaced scanning pattern ,[object Object]
Requires a wide frequency band during transmission  ,[object Object]
In this pattern the transmission is easier and cheaper because the actual picture frequency is lower  ,[object Object]
Contrast
Brightness
Image lag
Distortion  ,[object Object]
The resolution of imaging system refers to the amount of detail which is observable
It may be formulated as the number of pairs of black and white lines which an image forming or image recording device can demonstrate in a length of 1 mm
Vertical resolution is determined by no. of vertical scan lines (e.g. 625).In recent years the manufactures of radio diagnostic imaging equipment have developed his definition TV  system which employ 1249 or even more lines.
2/24/2010 35 Contrast The contrast of the image on a TV screen clearly depends on the contrast of the original image focused onto the signal plate of the pick up tube. Both camera and monitor affect the contrast of a TV image.  A vidicon camera reduces contrast by a factor of approx. 0.8 and the monitor enhances by a factor of 2.
2/24/2010 36 Brightness The term luminance and brightness are often used when discussing the aspect of tv image Luminance can be defined as light emitted per unit area from a surface. It is measured in candela per square meter (cd/m2) Automatic brightness control (ABC) ,[object Object]
It is critical  to the patient dose and image quality,[object Object]
2/24/2010 38 Image lag Lag is the term used to describe inability of an imaging system to follow rapid changes in its input image  Distortion  Distortion of the image occur in the image intensifier or in the optical coupling system which links its output phosphor to the signal plate of the TV pick up tube  38
2/24/2010 39  Artifacts Lag Vignetting, Veiling glare  Pincushion distortion S distortion.  39
2/24/2010 40 Lag Lag is the persistence of luminescence after x-ray stimulation has been terminated.  	  Lag degrades the temporal resolution of the dynamic image.  	  Older imageintensifier tubes had phosphors with lag times on the order of 30–40 msec. Current imageintensifier tubes have lag times of approximately 1 msec. Vignetting A fall-off in brightness at the periphery of an image is called vignetting. Vignetting is caused by the unequal collection of light at the center of the imageintensifier compared with the light at its periphery. As a result, the center of an imageintensifier has better resolution, increased brightness, and less distortion.  40
2/24/2010 41 Veiling Glare Scattering of light and the defocusing of photoelectrons within the imageintensifier are called veiling glare. Veiling glare degrades object contrast at the output phosphor of the imageintensifier. As mentioned, the contrast ratio is a good measure of determining the veiling glare of an imageintensifier. X-ray, electron, and light scatter all contribute to veiling glare.  Pincushion Distortion ,[object Object]
The magnification difference at the peripheryof the image results from the projection of the x-ray beam ontoa curved input surface.
The distortion is easily visualizedby imaging a rectangular grid with the fluoroscope. 41
2/24/2010 42 S DISTORTION 	Electrons within the imageintensifier move in paths along designated lines of flux. External electromagnetic sources affect electron paths at the perimeter of the imageintensifier more so than those nearer the center. This characteristic causes the image in a fluoroscopic system to distort with an S shape . 	Manufacturers include a highly conductive mu-metal shield that lines the canister in which the vacuum bottle is positioned to reduce the effect of S distortion.  42
2/24/2010 43 INDICATIONS Non traumatic Subarachnoid Hemorrhage (SAH). Arterial dissection or laceration Aneurysm Pseudoaneurysm Thrombosis. Arterio-venous malformation (AVM).  Arterio-venous fistula (AVF). Tumor vascularity. Gastrointestinal bleed/Uterine artery bleed Therapeutic.  43
2/24/2010 44 CONTRAINDICATIONS ,[object Object],Poor renal reserve. Deranged coagulogram. Allergic to contrast media. Recent barium examination –abd.angio 44
2/24/2010 45    Contrast Media ,[object Object]
To increase image contrast, contrast agents, which are dense fluids with elements of high atomic numbers, such as iodine, are injected into a blood vessel during angiography. Because of its higher density and high atomic number, iodine absorbs photons more than blood and tissue.
This creates detailed images of the blood vessels in real time.
The first contrast media used for intravascular injection were called high-osmolar contrast media (HOCM). (osmolality is the measure of the particle concentration in a solution.)45
2/24/2010 46  HOCM had osmolarity seven to eight times higher than plasma. This high osmolarity caused adverse effects such as pain, endothelial damage, thrombosis, and increased pressure in the pulmonary circulation.  Low-osmolar contrast media (LOCM) were first developed in the 1970's and these helped to reduce these side effects.  One of the major risks of modern iodine contrast media is an allergic reaction to iodine. 46
2/24/2010 47 ,[object Object]
30-40% dilution with normal saline.
50 ml of diluted contrast media is enough to do a standard cerebral angiogram with total 8 projections.
Approx. 5-8 ml diluted contrast / injection.47
2/24/2010 48 Materials used ,[object Object]
Disposable syringes
Local anesthesia
Heparin
Surgical gloves
Elastoplast
Catheters
Arterial sheath
Medicut
Guidewires
Contrast
Connector/100 cm.tubing
Surgical blade48
2/24/2010 49 Cathetars      Picard Pigtail 49 Sim-1
2/24/2010 50         VASCULAR  SHEATH  GUIDE WIRES MEDICUT 50 ELASTOPLAST  STRIPS
2/24/2010 51 ANTICOAGULANT ANESTHESIA CONTRAST PVA-PARTICLES GEL FOAM
2/24/2010 52 PREPARATION Should be well hydrated. Should void before procedure. Peripheral pulses marked. I.V line in place. INFORMED CONSENT  MUST ,[object Object]
Nil orally 4-6 hrs.
On trolley
In hospital gown
Groin shave

Weitere ähnliche Inhalte

Was ist angesagt?

CT ITS BASIC PHYSICS
CT ITS BASIC PHYSICSCT ITS BASIC PHYSICS
CT ITS BASIC PHYSICSDEEPAK
 
Arthrography of the Shoulder, Ankle and Wrist.pptx
Arthrography of the Shoulder, Ankle and Wrist.pptxArthrography of the Shoulder, Ankle and Wrist.pptx
Arthrography of the Shoulder, Ankle and Wrist.pptxDeneicer Guy
 
Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionparthajyotidas11
 
Physics of Multidetector CT Scan
Physics of Multidetector CT ScanPhysics of Multidetector CT Scan
Physics of Multidetector CT ScanDr Varun Bansal
 
Flat Panel Digital Fluoroscopy by Akash Das
Flat Panel Digital Fluoroscopy by Akash Das Flat Panel Digital Fluoroscopy by Akash Das
Flat Panel Digital Fluoroscopy by Akash Das AkashDas169
 
CT image quality and image resolution.pptx
CT image quality and image resolution.pptxCT image quality and image resolution.pptx
CT image quality and image resolution.pptxRohit Bansal
 
DIGITAL SUBTRACTION ANGIOGRAPHY
DIGITAL SUBTRACTION ANGIOGRAPHYDIGITAL SUBTRACTION ANGIOGRAPHY
DIGITAL SUBTRACTION ANGIOGRAPHYJoshua Mathew
 
Digital fluoro
Digital fluoroDigital fluoro
Digital fluoroRad Tech
 
Image reconstrsuction in ct pdf
Image reconstrsuction in ct pdfImage reconstrsuction in ct pdf
Image reconstrsuction in ct pdfmitians
 
Portable ct mobile ct
Portable ct mobile ctPortable ct mobile ct
Portable ct mobile ctanilayyakutty
 
Mammography physics and technique
Mammography  physics and techniqueMammography  physics and technique
Mammography physics and techniqueArchana Koshy
 
Ct tube and detectors
Ct tube and detectorsCt tube and detectors
Ct tube and detectorssandip suman
 
magnetic resonance in angiography
magnetic resonance in  angiography magnetic resonance in  angiography
magnetic resonance in angiography BISHAL KHANAL
 
Chapter 6 image quality in ct
Chapter 6 image quality in ct Chapter 6 image quality in ct
Chapter 6 image quality in ct Muntaser S.Ahmad
 

Was ist angesagt? (20)

CT ITS BASIC PHYSICS
CT ITS BASIC PHYSICSCT ITS BASIC PHYSICS
CT ITS BASIC PHYSICS
 
Basics in ct
Basics in ctBasics in ct
Basics in ct
 
Arthrography of the Shoulder, Ankle and Wrist.pptx
Arthrography of the Shoulder, Ankle and Wrist.pptxArthrography of the Shoulder, Ankle and Wrist.pptx
Arthrography of the Shoulder, Ankle and Wrist.pptx
 
Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortion
 
Physics of Multidetector CT Scan
Physics of Multidetector CT ScanPhysics of Multidetector CT Scan
Physics of Multidetector CT Scan
 
Flat Panel Digital Fluoroscopy by Akash Das
Flat Panel Digital Fluoroscopy by Akash Das Flat Panel Digital Fluoroscopy by Akash Das
Flat Panel Digital Fluoroscopy by Akash Das
 
CT image quality and image resolution.pptx
CT image quality and image resolution.pptxCT image quality and image resolution.pptx
CT image quality and image resolution.pptx
 
MDCT (2)
MDCT (2)MDCT (2)
MDCT (2)
 
DIGITAL SUBTRACTION ANGIOGRAPHY
DIGITAL SUBTRACTION ANGIOGRAPHYDIGITAL SUBTRACTION ANGIOGRAPHY
DIGITAL SUBTRACTION ANGIOGRAPHY
 
CT History
CT HistoryCT History
CT History
 
Digital fluoro
Digital fluoroDigital fluoro
Digital fluoro
 
Image reconstrsuction in ct pdf
Image reconstrsuction in ct pdfImage reconstrsuction in ct pdf
Image reconstrsuction in ct pdf
 
Fluroscopy
Fluroscopy Fluroscopy
Fluroscopy
 
Portable ct mobile ct
Portable ct mobile ctPortable ct mobile ct
Portable ct mobile ct
 
Mammography physics and technique
Mammography  physics and techniqueMammography  physics and technique
Mammography physics and technique
 
Hypotonic duodenography
Hypotonic duodenographyHypotonic duodenography
Hypotonic duodenography
 
Ct tube and detectors
Ct tube and detectorsCt tube and detectors
Ct tube and detectors
 
Fluoroscopy
FluoroscopyFluoroscopy
Fluoroscopy
 
magnetic resonance in angiography
magnetic resonance in  angiography magnetic resonance in  angiography
magnetic resonance in angiography
 
Chapter 6 image quality in ct
Chapter 6 image quality in ct Chapter 6 image quality in ct
Chapter 6 image quality in ct
 

Andere mochten auch

Andere mochten auch (6)

Clase 5 vascularizacion, meninges y ventriculos
Clase 5   vascularizacion, meninges y ventriculosClase 5   vascularizacion, meninges y ventriculos
Clase 5 vascularizacion, meninges y ventriculos
 
Angiografia y TAC
Angiografia y TACAngiografia y TAC
Angiografia y TAC
 
Presentation1.pptx, radiological imaging of lower limb ischemia.
Presentation1.pptx, radiological imaging of lower limb ischemia.Presentation1.pptx, radiological imaging of lower limb ischemia.
Presentation1.pptx, radiological imaging of lower limb ischemia.
 
Digital subtraction angiography
Digital subtraction angiographyDigital subtraction angiography
Digital subtraction angiography
 
Digital subtraction angiography
Digital subtraction angiographyDigital subtraction angiography
Digital subtraction angiography
 
Vascularización Cerebral
Vascularización CerebralVascularización Cerebral
Vascularización Cerebral
 

Ähnlich wie DSA DEEPAK GUPTA

3 d imaging super resolution
3 d imaging super resolution3 d imaging super resolution
3 d imaging super resolutionredsea3h
 
CBCT in orthodontics
CBCT in orthodontics CBCT in orthodontics
CBCT in orthodontics DrAkshiSharma
 
EPID AND CBCT ON RADIATION THERAPY
EPID AND CBCT ON RADIATION THERAPYEPID AND CBCT ON RADIATION THERAPY
EPID AND CBCT ON RADIATION THERAPYanju k.v.
 
Treatment verification systems in radiation therapy
Treatment verification systems in radiation therapyTreatment verification systems in radiation therapy
Treatment verification systems in radiation therapyanju k.v.
 
Mobile c arm equip and dsa
Mobile c arm equip and dsaMobile c arm equip and dsa
Mobile c arm equip and dsaRad Tech
 
Cone beam computed tomography (2)
Cone beam computed tomography (2)Cone beam computed tomography (2)
Cone beam computed tomography (2)Mohit Kapoor
 
CBCT in dental practice
CBCT in dental practiceCBCT in dental practice
CBCT in dental practiceZana Hussein
 
2009- Smal Size Virtual Slides and 4M microscopes
2009- Smal Size Virtual Slides and 4M microscopes2009- Smal Size Virtual Slides and 4M microscopes
2009- Smal Size Virtual Slides and 4M microscopesUnesco Telemedicine
 
CBCT IN IMPLANTOLOGY.pptx
CBCT IN IMPLANTOLOGY.pptxCBCT IN IMPLANTOLOGY.pptx
CBCT IN IMPLANTOLOGY.pptxDr K. Abhilasha
 
The Role of Computers in Medical Physics
The Role of Computers in Medical PhysicsThe Role of Computers in Medical Physics
The Role of Computers in Medical PhysicsVictor Ekpo
 
Digital imaging of head and neck of the animals
Digital imaging of head and neck of the animalsDigital imaging of head and neck of the animals
Digital imaging of head and neck of the animalssozanmuhamad1
 
Digital imaging of the all body organ of
Digital imaging  of the all body organ ofDigital imaging  of the all body organ of
Digital imaging of the all body organ ofsozanmuhamad1
 
Go 2 d3d ceph_cataloge
Go 2 d3d ceph_catalogeGo 2 d3d ceph_cataloge
Go 2 d3d ceph_catalogeYenDoan19
 

Ähnlich wie DSA DEEPAK GUPTA (20)

3 d imaging super resolution
3 d imaging super resolution3 d imaging super resolution
3 d imaging super resolution
 
CBCT in orthodontics
CBCT in orthodontics CBCT in orthodontics
CBCT in orthodontics
 
EPID AND CBCT ON RADIATION THERAPY
EPID AND CBCT ON RADIATION THERAPYEPID AND CBCT ON RADIATION THERAPY
EPID AND CBCT ON RADIATION THERAPY
 
Treatment verification systems in radiation therapy
Treatment verification systems in radiation therapyTreatment verification systems in radiation therapy
Treatment verification systems in radiation therapy
 
Mobile c arm equip and dsa
Mobile c arm equip and dsaMobile c arm equip and dsa
Mobile c arm equip and dsa
 
Ccdcams
CcdcamsCcdcams
Ccdcams
 
Cone beam computed tomography (2)
Cone beam computed tomography (2)Cone beam computed tomography (2)
Cone beam computed tomography (2)
 
CBCT in dental practice
CBCT in dental practiceCBCT in dental practice
CBCT in dental practice
 
CBCT IN ORTHODONTICS
CBCT IN ORTHODONTICSCBCT IN ORTHODONTICS
CBCT IN ORTHODONTICS
 
2009- Smal Size Virtual Slides and 4M microscopes
2009- Smal Size Virtual Slides and 4M microscopes2009- Smal Size Virtual Slides and 4M microscopes
2009- Smal Size Virtual Slides and 4M microscopes
 
Digital cephalometry
Digital cephalometryDigital cephalometry
Digital cephalometry
 
Cone beam ct
Cone beam ctCone beam ct
Cone beam ct
 
CBCT IN IMPLANTOLOGY.pptx
CBCT IN IMPLANTOLOGY.pptxCBCT IN IMPLANTOLOGY.pptx
CBCT IN IMPLANTOLOGY.pptx
 
The Role of Computers in Medical Physics
The Role of Computers in Medical PhysicsThe Role of Computers in Medical Physics
The Role of Computers in Medical Physics
 
spect .pptx
spect .pptxspect .pptx
spect .pptx
 
Digital imaging of head and neck of the animals
Digital imaging of head and neck of the animalsDigital imaging of head and neck of the animals
Digital imaging of head and neck of the animals
 
Digital imaging of the all body organ of
Digital imaging  of the all body organ ofDigital imaging  of the all body organ of
Digital imaging of the all body organ of
 
DIGITAL IMAGING
DIGITAL IMAGINGDIGITAL IMAGING
DIGITAL IMAGING
 
Go 2 d3d ceph_cataloge
Go 2 d3d ceph_catalogeGo 2 d3d ceph_cataloge
Go 2 d3d ceph_cataloge
 
Digital imaging in orthodontics
Digital imaging in orthodontics Digital imaging in orthodontics
Digital imaging in orthodontics
 

Kürzlich hochgeladen

Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...Krashi Coaching
 
1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdfQucHHunhnh
 
Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3JemimahLaneBuaron
 
social pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajansocial pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajanpragatimahajan3
 
Key note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfKey note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfAdmir Softic
 
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdfBASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdfSoniaTolstoy
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdfQucHHunhnh
 
Paris 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activityParis 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activityGeoBlogs
 
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in DelhiRussian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhikauryashika82
 
microwave assisted reaction. General introduction
microwave assisted reaction. General introductionmicrowave assisted reaction. General introduction
microwave assisted reaction. General introductionMaksud Ahmed
 
Sanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfSanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfsanyamsingh5019
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104misteraugie
 
Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfciinovamais
 
Holdier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdfHoldier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdfagholdier
 
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...EduSkills OECD
 
Measures of Dispersion and Variability: Range, QD, AD and SD
Measures of Dispersion and Variability: Range, QD, AD and SDMeasures of Dispersion and Variability: Range, QD, AD and SD
Measures of Dispersion and Variability: Range, QD, AD and SDThiyagu K
 
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...Sapna Thakur
 
Accessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impactAccessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impactdawncurless
 

Kürzlich hochgeladen (20)

Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
 
Mattingly "AI & Prompt Design: The Basics of Prompt Design"
Mattingly "AI & Prompt Design: The Basics of Prompt Design"Mattingly "AI & Prompt Design: The Basics of Prompt Design"
Mattingly "AI & Prompt Design: The Basics of Prompt Design"
 
1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdf
 
Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3
 
social pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajansocial pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajan
 
Key note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfKey note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdf
 
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdfBASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdf
 
Paris 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activityParis 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activity
 
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
 
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in DelhiRussian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
 
microwave assisted reaction. General introduction
microwave assisted reaction. General introductionmicrowave assisted reaction. General introduction
microwave assisted reaction. General introduction
 
Sanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfSanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdf
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104
 
Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdf
 
Holdier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdfHoldier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdf
 
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
 
Measures of Dispersion and Variability: Range, QD, AD and SD
Measures of Dispersion and Variability: Range, QD, AD and SDMeasures of Dispersion and Variability: Range, QD, AD and SD
Measures of Dispersion and Variability: Range, QD, AD and SD
 
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
 
Accessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impactAccessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impact
 

DSA DEEPAK GUPTA

  • 1. DSA ITS APPLICATION AND ADVANCEMENT Moderator: MR. S.R.CHOWDHURY TUTOR, DEPT.OF RADIO-DIAGNOSIS & IMAGING, PGIMER CHANDIGARH Presented by: DEEPAK GUPTA BSc final year Student , DEPT.OF RADIO-DIAGNOSIS & IMAGING PGIMER CHANDIGARH
  • 2. 2/24/2010 2 ANGIO? ANGIO means blood vessel And angiography is the radiological study of blood vessel in the body after the introduction of iodinated contrast media. SUBTRACTION? It is simply a technique by which bone structures images are subtracted or canceled out from a film of bones plus opacified vessels, leaving an unobscured image of the vessels.
  • 3. 2/24/2010 3 WHAT DO YOU MEAN BY DSA DSA-The acquisition of digital fluoroscopic images combined with injection of contrast material and real-time subtraction of pre- and post contrast images to perform angiography is referred to as digital subtraction angiography
  • 4. 2/24/2010 4 HISTORY The Portuguese neurologist Egas Moniz,( Nobel Prize winner 1949), in 1927developed the technique of contrast x-ray cerebral angiography to diagnose diseases, such as tumors and arteriovenous malformations. The idea of subtraction images was first proposed by the Dutch radiologist Ziedses des Plantes in the 1935, when he was able to produce subtracted images using plain films. With the introduction of the Seldinger technique in 1953, the procedure became safer as no sharp devices need to remain inside the vascular lumen.
  • 5. 2/24/2010 5 HISTORICAL DEVELOPMENT CONVENTIONAL SUBTRACTION TECHNIQUE Photographic method used to eliminate unwanted images. No addition of information;only purpose to make diagnostically important information to see. First described by a Dutch radiologist, Zeides des Plantes, 1935. 3 conditions:SCOUT FILM ANGIOGRAM FILM-CONTRAST NO MOTION OF HEAD
  • 6. 2/24/2010 6 CONTD. The principles of subtraction are based on the following: The scout film shows the structural details of the skull and the adjacent soft tissue. Angiogram film shows exactly the same anatomic details, if the patient does not move, plus the opacified blood vessels. If all the information in the scout film could be subtracted from the angiogram film, only the opacified vessel pattern would remain visible.
  • 7. 2/24/2010 7 EQUIPMENT THE AOT CHANGER: Can change films –6/s or 1/5s Two of these changers can be coupled electronically and mechanically and will operate out of phase or synchronously with each other. Radiographs in 2 planes simultaneously-one mounted vertically and other horizontal.
  • 8. 2/24/2010 8 AOT CONTD. LOADING MAGZINE Large steel container. Inside the container are arranged a no. of strong wire separators. 30 sheets of film. Precautions :each film should slide into position by it’s own wt. RECIEVING / LOADING MAGZINE
  • 9. 2/24/2010 9 EXPOSURE AREA : Rectangular, slightly recessed area at the top of the changer. Defined by heavy metal frame. Within this is a secondary grid. Supported on 4 small helical springs placed cornerwise and to consist of a stout Al plate, to the lower side of which is attached an intensifying screen. 30 SEPERATING WIRES RECEIVING CASSETTE: Shallow,polished metal container. Lid is a sliding section in a side which can be pushed down by firm finger pressure and is retained by central spring loaded catch.
  • 10. 2/24/2010 10 THE ROLL FILM CHANGER: Mechanically easier to wind roll films. 12/s.
  • 11. 2/24/2010 11 DIGITAL SUBTRACTION ANGIOGRAPHY HISTORY: Developed in 1970s. University of Wisconsin, University of Arizona,University of Kiel. Commercial systems introduced in 1980.
  • 12. 2/24/2010 12 Angiography theory ages Single shot angiography Manual changer AOT PUCK DSA
  • 13. 2/24/2010 13 EQUIPMENT AND APPARATUS Both film and digital capabilities are usually present and the fluoroscopy equipment operates in the conventional way. Brief review : Light detected amplified Input surface Visible light Output phosphor Electrical signal Light intensity
  • 14. 2/24/2010 14 DIGITAL IMAGE PROCESSOR DISPLAY SECTION DIGITAL ARITHEMATIC UNIT DIGITIZER DIGITAL IMAGE MEMORY X-RAY CONTROL SIGNALS MONITOR COMPUTER MULTIFORMAT CAMERA DIGITAL DISK STORAGE
  • 15. 2/24/2010 15 X-Ray Unit Should be capable of rapid serial exp. @2-8 frames/sec for up to 20 sec. Near monochromatic beam consistent output. Isocentre motorized table ,tilt 90-0-90 , up-down, sliding ,floating capable Ceiling mounted C-arm with versatile movements II system with multi field ( 6”,9”,12”,16”) High resolution TV /CCD camera Computer with high definition monitor system
  • 16. 2/24/2010 16 Contd. High output preferably high frequency generator High heat loading capacity ,fast heat dissipation, high speed anode Target material tungsten,rhenium,graphite. Oil-water close loop cooling
  • 17. 2/24/2010 17 In our dept.Siemens polydoros SX 65-80 Polystar
  • 18. 2/24/2010 18 NECESSARY EQUIPMENT USED WITH ANGIO(DSA) GAS , AIR & VACUUM PORTS Boyle apparatus
  • 19. 2/24/2010 19 BED SIDE MONITOR INJECTOR
  • 21. 2/24/2010 21 IMAGE INTENSIFIER-CONSTRUCTION
  • 23. 2/24/2010 23 Modern DSA systems are based on digital fluoroscopy/fluorography systems, which are equipped with special software and display facilities. Digital subtraction angiography (DSA) was developed to improve vessel contrast. This is a technique that uses a computer to subtract two images, obtained before and after contrast media is injected into the vessels of interest. The anatomical structures that are the same in the two images can be removed and the resulting image shows the vessels only.
  • 24. 2/24/2010 24  The image before the contrast agent is administered is called the mask image. Once the contrast is administered, a sequence of images are taken by a television camera in analog form, which is then digitised by computer.  The DSA processor has two separate image memories, one for the mask and the other for the images with contrast medium. These two image memories are subtracted from one another arithmetically, and the result goes to an image processing and display unit.
  • 25. 2/24/2010 25 IMAGE PROCESSING Adjust contrast & brightness. Remasking – correcting misregistered images. Pixel shifting. Edge enhancement – edges of the vessels can be enhanced so that small details can be made more obvious. Image zoom. land marking – a small amount of original image is added into the subtracted image. Noise smoothing – operates by reducing the statistical fluctuations in each pixel by averaging the pixel with it's closest neighbors. The visual prominence of noise has been suppressed by averaging, but resolution is decreased.
  • 26. 2/24/2010 26 Road Mapping Road mapping is useful for the placement of catheters and wires in complex and small vasculature. A DSA sequence is performed, and the frame with maximum vessel opacification is identified;this frame becomes the road map mask. The road map mask is subtracted from subsequent live fluoroscopic images to produce real-time subtracted fluoroscopic images overlaid on a static image of the vasculature
  • 27. 2/24/2010 27 When a digital subtraction technique is used, patient motion that occurs between acquisition of the precontrast images and acquisition of the postcontrast images will result in artifacts due to misregistration of the two images. If these arti-facts are observed, it is possible to reregister the pre- and postcontrast images by shifting the subtraction mask (precontrast image) with respect to the postcontrast image and resubtracting the two images
  • 28.
  • 34. Two pair of electrostatic deflecting coil
  • 35.
  • 36.
  • 37. The electrons are focused on the target of tube and
  • 38. The beam is moved over the target area in an orderly scanning scan
  • 41.
  • 42.
  • 43.
  • 47.
  • 48. The resolution of imaging system refers to the amount of detail which is observable
  • 49. It may be formulated as the number of pairs of black and white lines which an image forming or image recording device can demonstrate in a length of 1 mm
  • 50. Vertical resolution is determined by no. of vertical scan lines (e.g. 625).In recent years the manufactures of radio diagnostic imaging equipment have developed his definition TV system which employ 1249 or even more lines.
  • 51. 2/24/2010 35 Contrast The contrast of the image on a TV screen clearly depends on the contrast of the original image focused onto the signal plate of the pick up tube. Both camera and monitor affect the contrast of a TV image. A vidicon camera reduces contrast by a factor of approx. 0.8 and the monitor enhances by a factor of 2.
  • 52.
  • 53.
  • 54. 2/24/2010 38 Image lag Lag is the term used to describe inability of an imaging system to follow rapid changes in its input image Distortion Distortion of the image occur in the image intensifier or in the optical coupling system which links its output phosphor to the signal plate of the TV pick up tube 38
  • 55. 2/24/2010 39 Artifacts Lag Vignetting, Veiling glare Pincushion distortion S distortion. 39
  • 56. 2/24/2010 40 Lag Lag is the persistence of luminescence after x-ray stimulation has been terminated. Lag degrades the temporal resolution of the dynamic image. Older imageintensifier tubes had phosphors with lag times on the order of 30–40 msec. Current imageintensifier tubes have lag times of approximately 1 msec. Vignetting A fall-off in brightness at the periphery of an image is called vignetting. Vignetting is caused by the unequal collection of light at the center of the imageintensifier compared with the light at its periphery. As a result, the center of an imageintensifier has better resolution, increased brightness, and less distortion. 40
  • 57.
  • 58. The magnification difference at the peripheryof the image results from the projection of the x-ray beam ontoa curved input surface.
  • 59. The distortion is easily visualizedby imaging a rectangular grid with the fluoroscope. 41
  • 60. 2/24/2010 42 S DISTORTION Electrons within the imageintensifier move in paths along designated lines of flux. External electromagnetic sources affect electron paths at the perimeter of the imageintensifier more so than those nearer the center. This characteristic causes the image in a fluoroscopic system to distort with an S shape . Manufacturers include a highly conductive mu-metal shield that lines the canister in which the vacuum bottle is positioned to reduce the effect of S distortion. 42
  • 61. 2/24/2010 43 INDICATIONS Non traumatic Subarachnoid Hemorrhage (SAH). Arterial dissection or laceration Aneurysm Pseudoaneurysm Thrombosis. Arterio-venous malformation (AVM). Arterio-venous fistula (AVF). Tumor vascularity. Gastrointestinal bleed/Uterine artery bleed Therapeutic. 43
  • 62.
  • 63.
  • 64. To increase image contrast, contrast agents, which are dense fluids with elements of high atomic numbers, such as iodine, are injected into a blood vessel during angiography. Because of its higher density and high atomic number, iodine absorbs photons more than blood and tissue.
  • 65. This creates detailed images of the blood vessels in real time.
  • 66. The first contrast media used for intravascular injection were called high-osmolar contrast media (HOCM). (osmolality is the measure of the particle concentration in a solution.)45
  • 67. 2/24/2010 46  HOCM had osmolarity seven to eight times higher than plasma. This high osmolarity caused adverse effects such as pain, endothelial damage, thrombosis, and increased pressure in the pulmonary circulation.  Low-osmolar contrast media (LOCM) were first developed in the 1970's and these helped to reduce these side effects.  One of the major risks of modern iodine contrast media is an allergic reaction to iodine. 46
  • 68.
  • 69. 30-40% dilution with normal saline.
  • 70. 50 ml of diluted contrast media is enough to do a standard cerebral angiogram with total 8 projections.
  • 71. Approx. 5-8 ml diluted contrast / injection.47
  • 72.
  • 85. 2/24/2010 49 Cathetars Picard Pigtail 49 Sim-1
  • 86. 2/24/2010 50 VASCULAR SHEATH GUIDE WIRES MEDICUT 50 ELASTOPLAST STRIPS
  • 87. 2/24/2010 51 ANTICOAGULANT ANESTHESIA CONTRAST PVA-PARTICLES GEL FOAM
  • 88.
  • 94. PTI52
  • 95.
  • 97. Renal
  • 100. Abdominal - coelic & branches
  • 101. -sup. mesentric A.
  • 102. - inf. mesentric A.
  • 104.
  • 105. - coil
  • 106. - gel
  • 107. - particle
  • 108. - ballon
  • 110. TACE
  • 112. 2/24/2010 54 PROCEDURE Gaining arterial access. Selective arterial catheterization. Image acquisition. Closure of arterial access. Post processing Hard copy 54
  • 114.
  • 115. Their heart rate and rhythm, breathing, and oxygen saturation are monitored throughout the procedure.
  • 117. A local anesthetic is usually used in the area where the catheter is to be inserted, most commonly the femoral artery.
  • 118. First, a small incision given, medicut is inserted into the artery. fluoroscopy is used to guide the needle to the proper position .
  • 119. The needle is then removed after placing guide wire in the artery and vascular sheath is inserted over the guide wire . The catheter is then inserted along the guide wire through the sheath.56
  • 120.
  • 121. Images are acquired during contrast injection. Injections can be made directly into the artery of interest (selective arteriography)
  • 122. Complications from an arteriogram are very rare, but there is some risk. Most problems that occur can be detected at the time of the procedure or immediately after the procedure. The artery may be injured at the puncture site or along the artery where the catheter is passed.57
  • 123. 2/24/2010 58 COMPLICATIONS 0.16% major complication rate. Local complications: hematoma, vessel laceration, dissection, peudoaneurysm, AV fistula. Systemic complications: contrast reactions, fever, sepsis, dehydration, death. CNS complication: aggravation of preexisting complaints, neurological deficit. 58
  • 124. 2/24/2010 59 POSTPROCEDURAL CARE After the catheter is removed compression is applied to the puncture site. The patient is asked for bed rest for a minimum of 4 hours During rest patient is monitored and vital sign like peripheral pulse like distal to Puncture are regularly The extremity is also checked for warmth, color, numbness to ensure circulation has not been disrupted.
  • 125. 2/24/2010 60 Oral fluid is given and analgesics are given if required. Special care should be given in case of children and geriatric patient since additional Patience is required. They have given warming blanket to keep the body temp. Normal. Geriatric patient also frequently feel nervous and afraid of falling off the Table ,reassurance and additional care from the technologist will enable the Patient to feel secure and comfortable.
  • 126.
  • 127.
  • 128. 2/24/2010 63 Dual energy subtraction 2 images are taken within a very short period (no pt. movement ). One at low KVp & other at high KVp. HYBRID K EDGE
  • 129.
  • 130. One method for achieving high iodine contrast is removal of all x-rays from the beam except those that lie below the k-shell B.E of ‘I’.
  • 131. An x-ray filter made of rare earth named cerium can do this well, since B.E = 40 KeV.
  • 132. Thick filter will remove most of the x-rays above 40 KeV.
  • 133. The image to be subtracted should contain few x-rays.
  • 134.
  • 135. No pt. movement – soft tissues cancel properly. but bone edges cause severe artifact.
  • 136. Consider the subtracted images to consist of only 2 atomic no. materials (iodine & bone).
  • 137. The hybrid subtraction produces 2 sets of subtracted images.
  • 138. The same bone & iodine structures are present on both sets.
  • 139. D.E.S can now be used to eliminate bone leaving only iodine.
  • 140.
  • 141. TID – a new mask for each subtraction.
  • 142.
  • 143.
  • 145. 2/24/2010 70 the Allura Xper FD20/10 balances superb image quality with excellent dose safety.  The flat detector's unique distortion-free imaging chain redefines image clarity and captures information at a resolution four times greater than conventional X-ray systems.
  • 146. 2/24/2010 71 Xper ct With XperCT clinicians can access CT-like imaging right on the angio system so that they can assess soft tissue, bone structure and other body structures before, during or after an interventional procedure. The XperCT reconstruction is created from a rotational acquisition performed on the Allura Xper system. This reconstruction can be overlaid with the 3D vascular image. IT Bring 3D soft tissue imaging to your angio suite 3D soft tissue imaging supports diagnosis, planning, interventions and treatment follow-up, XperCT can be combined with Allura 3D-RA images to visualize soft tissue and vascular anatomy on one image. Using CT-like imaging for a post-procedural check can help reduce delays and improve patient comfort.
  • 147. 2/24/2010 72 XperCT overlay on both the internal carotid and vertebral arteries
  • 148. 2/24/2010 73 3-D RECONSTRUCTION 73 73
  • 149. 2/24/2010 74 Spectra Beam It is basically a selectable copper beam filtration The combination of SpectraBeam with the MRC- tube allows increased X-ray output with better filtration of soft radiation. SpectraBeam offers filters of 200, 500 and 1000 microns thickness Cu equivalent to provide a high level of radiation protection regardless of the projection or patient absorption. This reduces patient X-ray dose for cardiac and vascular applications, while maintaining the same image quality.
  • 150. 2/24/2010 75 When the acquisition sequence is initiated, the C-arm rotates from left to right for 220o at 40o per second but data acquisition is accomplished at the range Of 200o, acquiring at 8.8 frames per second for 5 seconds resulting in 44mask images The gantry then rotates back to the starting position. When the C-arm returns to the starting position, the Power injector is initiated according to the injector delay protocol, which is Mask-to-Image delay of 1.1 seconds and Injector-to-Image delay of 1.1 seconds. The Power injector injects 18 milliliter of Visipaque 280 at the rate of 3.6 ml/second for Internal Carotid Artery (ICA). The gantry repeats the mask acquisition path and acquires another 44 images with contrast. Real time subtraction angiographs are displayed on post processing screen. 75 75
  • 151. 2/24/2010 76 76 System components Generators X-Ray Tubes II/Flat detector Imaging Chain Patient Table
  • 152. 2/24/2010 77 Generator characteristics X-ray generators High powered x-ray generators to provide almost constant potential power supply to the tube. Two Velara CVFD microprocessor-controlled 100 kW high-frequency converter generators. Minimum exposure time of 1 ms. Voltage range: 40 kV to 125 kV. Max current: - MRC-GS 04-07: 802 mA at 80 kV. - MRC-GS 05-08: 1062.5 mA at 80 kV. Automatic kV and mA control for optimal image quality prior to run to reduce dose. 77 77
  • 153. 2/24/2010 78 78 Generator characteristics cont. Max continuous power for fluoroscopy: 2.4 kW for 0.5 hour and 2 kW for 8 hours. Nominal power (highest electrical power): 100 kW (1000 mA at 100 kV) Reference loading conditions: 110 kV, 18 mA continuous. Pulsed X-ray of 3.75, 7.5, 15 and 30 frames/sec. in monoplane and biplane mode for pulsed fluoroscopy. Pulsed X-ray of 0.5 to 6 frames/sec. for digital subtracted acquisition in monoplane and biplane mode. Pulsed X-ray up to 6 frames/sec. for digital acquisition in monoplane and biplane mode. 15 and 30 frames/sec. Noise < 55 dB(A). 78
  • 154. 2/24/2010 79 The high performance metal ceramic tubes . Most relevant feature is support of the rotor anode in a spiral groove bearing The spiral groove bearing glides on a thin film of liquid metal to maximize heat dissipation allowing virtually unlimited X-ray sessions without forced cool down delays.• The MRC tube runs silently offering a more comfortable patient environment.• Another secret of the MRC tube is the enormous heat storage capacity of its 200 mm diameter anode, eliminating waiting times.• And, the MRC X-ray tube can image the heaviest patients even at steep angles, while maintaining image quality. 79 X-Ray Tubes 79
  • 155. 2/24/2010 80 Tube Specifications 80 80
  • 156. 2/24/2010 81 Tube Specifications … 81 81
  • 157.
  • 158. The flat detector converts the light generated by the absorbed X-ray energy in the scintillation layer directly into a digital electrical signal.82 Detectors 82
  • 159.
  • 161. Image matrix 2480x1920 pixel at 14 bit height
  • 162. Detector zoom fields 30x30cm,22x22cm, 16x16cm,13.5x13.5cm, 11x11cm formats Pixel size 154x154 μm Detective Quantum Efficiency (DQE): >73% at low spatial frequencies 83 Flat detector frontal 83
  • 163.
  • 165. Image matrix 1024x1024 pixel at 14 bit height
  • 166. Detector zoom fields 14.5x14.5cm, 11x11cm formatsPixel size 184x184 μm Detective Quantum Efficiency (DQE): 75% 84 Flat detector lateral 84
  • 167. 2/24/2010 85 Patient support AngioDIAGNOST 5 Length: 303 cm (119.3 inch). Metal-free overhang: 125 cm (49.2 inch). Width: 50 cm (19.7 inch). Longitudinal float: 100 cm (39.4 inch). Transverse float: 36 cm (14.2 inch). Height adjustment (without pivot option): 76 to 104 cm (29.9 to 40.9 inch). Maximum levelling speed (height adjustment): - 2 cm/s (0.8 inch/s). Table tilt: L-arc position Angulation (degrees) Rotation (degrees) Head-end of table 45 Caudal to 45 cranial 27 RAO to 115 RAO 85 85
  • 168. 2/24/2010 86 Light intensity: 30,000 Lux. Colour temperature: 4300° K. Focusable light field size: 14 - 25 cm (5.5 - 9.8 inch). Working distance: 70 -140 cm (27.6 - 55.1 inch). Light intensity at 30,000 Lux: 114 W/m2. Lamp type: halogen 22.8 / 24 V 50 W. Mains power: 220/240 V. 86 Examination light 86
  • 169. 2/24/2010 87 Quality Assurance Weekly 3D calibration is carried out. The calibration data is valid for only 14 days (7 days pre- and post- acquisition). The calibration is to rectify S-distortion as well as pin cushion distortion. S-distortion is caused by geo- magnetic field s on moving electrons within the Image Intensifier and pin cushion distortion is caused by the curvature of the input phosphor of the Image Intensifier.
  • 170. 2/24/2010 88 DoseWise is active throughout at every level of system - from spectra beam, pulsed fluoroscopy to syncratouch. It ensures every possible opportunity to reduce dose, while maintaining image quality. DoseWise includes three highly effective strategies for dose management: Spectra Beam technology to remove unwanted ‘soft’ radiation and minimize scatter radiation, automatically. Spectra Beam filters during fluoro and exposures remove unwanted soft radiation, i.e. those X-rays that hit the patient but do not have enough energy to reach the image detector. In this way, filtering significantly reduces patient dose and scattered radiation for staff while maintaining a high image quality.   88 Radiation Protection-Dose reduction 88
  • 171. 2/24/2010 89 89 Advantages Dose reduction 3-D road mapping Xper CT Bolus Chase Quantitative analysis packages 89
  • 172. 2/24/2010 90 DIS-ADVANTAGES OF DSA Mis-registration . Poor resolution compared to conventional angiography. User dependent success rate. Risk of emboli may reaching to healthy tissue. Not suitable for everyone.
  • 173.
  • 174. Three-dimensional (3D) reconstruction of the dataset acquired during rotational DSA represents the latest development in the neurovascular imaging armamentarium.
  • 175. This technique combines the anatomic resolution of DSA with the 3D visualization abilities previously offered by only CT or MR angiography and provides more detailed information than does DSA alone in the evaluation of neurovascular lesions, such as cerebral aneurysms .
  • 176. 3D-DSA has taken a prominent role in treatment planning by enabling better appreciation of the morphology of complex vascular lesions before endovascular or surgical management.
  • 177. It is also superior in the performance of sophisticated tasks such as aneurysm volume measurement
  • 178. On the other hand, the inability of 3D-DSA to simultaneously image osseous and vascular structures is noted as a weakness of this technique compared with CT angiography . 91 91