CS 3600
Summary from “Trueness and precision of 5 intraoral scanners in
the impressions of single and multiple implants: a comparative in
vitro study”
Overview
Article Name
“Trueness and precision of 5 intraoral scanners in the
impressions of single and multiple implants: a comparative in
vitro study”
Authors
‱ Francesco Guido Mangano, Uli Hauschild, Giovanni Veronesi,
Mario Imburgia, Carlo Mangano, and Oleg Admakin
Date Published
6 June 2019
Background
‱ Very few peer reviewed studies have evaluated the clinical
efficacy of intraoral scanners in implantology.
‱ Even fewer studies compare the accuracy of the various
intraoral scanners.
‱ This study compares the trueness and precision of five
intraoral scanners, and determines which have both accurate
and consistent results.
2
What are Trueness and Precision?
In this study:
Trueness is quantified by distances between generated model and reference model
(smaller value means better trueness).
Precision is quantified by distances between repeatedly generated models (smaller
value means better precision).
Accuracy is a combination of trueness and precision. An intraoral scanner must
perform well in both trueness and precision tests to be considered “highly accurate.”
3
Study Material and Methods
10 scans were performed by a single digital dentistry expert per indication per intraoral
scanner device with a zig-zag technique in the same environmental conditions.
2 reference maxillary plaster models were used as a reference:
‱ A - Partially edentulous maxillary arch with:
‱ Implant analog on tooth #23 (implant single support crown simulation)
‱ Implant analog on teeth #14 & 16 (partial implant supported case simulation)
‱ B - Totally edentulous maxillary arch:
‱ Implant analog on teeth #11, 14, 16, 21, 24 & 26 (fixed full arch implant supported denture simulation)
Scanbodies used : MegagenÂź PEEK scanbodies to enhance scans (no light reflection).
Reference : 3 scans made per indication with Freedom UHD Âź desktop scanner and
superimposed in Geomagic Studio 2012Âź to validate the reference models.
4
How do these intraoral scanners compare in accuracy?
Implantology Intraoral Scanner Comparative Study
5
TRIOS 3
EMERALD
DWIO
CS 3600
OMNICAM
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
Implantology Intraoral Scanner Comparative Study
Scanner Producer Technology of
Acquisition
Powder Color System
CS 3600 Carestream
Dental,
Atlanta, Georgia, USA
Structured light-Active
Speed 3D Videoℱ
No Yes Proprietary files (CSZ), but
also open formats (PLY, STL)
immediately available
Trios 3 3-Shape,
Copenhagen,
Denmark
Structured light –Confocal
microscopy and Ultrafast
Optical Scanningℱ
No Yes Proprietary files (DCM)
available, but possibility to
export STL files via the
new Trios on Dental DesktopÂź
Omnicam Dentsply-Sirona,
York, Pennsylvania,
USA
Structured light -Optical
triangulation and confocal
microscopy
No Yes Proprietary files (CS3, SDT,
CDT, IDT) are available,
possible to export STL files via
the Cerec ConnectÂź
DWIO Dentalwings,
Montreal, Quebec,
Canada
Blue laser-Multiscan
Imagingℱ technology
No No Proprietary files (XORDER),
but also open formats (STL)
immediately available
Emerald Planmeca,
Helsinki, Finland
Red, green and blue
lasers-Projected Pattern
Triangulationℱ
No Yes Open formats (PLY, STL)
immediately available
6
Summary of 5 Intraoral Scanners Used in the Study
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
Single Crown, Best Result in Trueness (ÎŒM)
7
Deviation: 22 ÎŒm (20)
Triangles: 179,496
Deviation: 14 ÎŒm (13)
Triangles: 34,980
Deviation: 31 ÎŒm (26)
Triangles: 81,712
Deviation: 23 ÎŒm (21)
Triangles: 68,281
Deviation: 21 ÎŒm (28)
Triangles: 123,417
TRIOS 3 CS 3600 EMERALD DWIO OMNICAM
Best Result
[Carestream Dental Note: Triangle number is not a measure of acquisition resolution or
mesh quality. In fact, in having the best trueness with the smallest number of triangles, CS
3600 demonstrates the most efficient/optimal mesh generation (by placing more triangles in
areas with more details and vice versa), with minimized mesh file size while maintaining
accuracy.]
Single crown: best result in trueness (standard deviation) in ÎŒm, for the 5 examined
scanners, and the number of triangles composing each mesh.
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
Partial Prosthesis, Best Result in Trueness (ÎŒM)
8
Deviation: 28 ÎŒm (24)
Triangles: 207,159
Deviation: 21 ÎŒm (25)
Triangles: 55,968
Deviation: 42 ÎŒm (32)
Triangles: 113,162
Deviation: 39 ÎŒm (39)
Triangles: 135,028
Deviation: 29 ÎŒm (32)
Triangles: 192,967
8
TRIOS 3 CS 3600 EMERALD DWIO OMNICAM
Best Result
Partial prosthesis: best result in trueness (standard deviation) in ÎŒm, for the 5 examined
scanners, and the number of triangles composing each mesh.
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
[Carestream Dental Note: Triangle number is not a measure of acquisition resolution or
mesh quality. In fact, in having the best trueness with the smallest number of triangles, CS
3600 demonstrates the most efficient/optimal mesh generation (by placing more triangles in
areas with more details and vice versa), with minimized mesh file size while maintaining
accuracy.]
Full Arch, Best Result in Trueness (ÎŒM)
9
Deviation: 37 ÎŒm (44)
Triangles: 326,696
Deviation: 32 ÎŒm (49)
Triangles: 83,735
Deviation: 57 ÎŒm (58)
Triangles: 178,507
Deviation: 73 ÎŒm (64)
Triangles: 227, 491
Deviation: 47 ÎŒm (55)
Triangles: 322,929
9
TRIOS 3 CS 3600 EMERALD DWIO OMNICAM
Best Result
Full arch: best result in trueness (standard deviation) in ÎŒm, for the 5 examined
scanners, and the number of triangles composing each mesh.
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
[Carestream Dental Note: Triangle number is not a measure of acquisition resolution or
mesh quality. In fact, in having the best trueness with the smallest number of triangles, CS
3600 demonstrates the most efficient/optimal mesh generation (by placing more triangles in
areas with more details and vice versa), with minimized mesh file size while maintaining
accuracy.]
Mean Trueness/Standard Deviation
Scanner Single Crown
Mean ± SD
Partial Prosthesis
Mean ± SD
Full Arch
Mean ± SD
p-value1
Trios 3
22.3 ± 0.5† 28.5 ± 0.5†,‡,‱ 46.3 ± 4.9†,‡,‱ <.0001
CS 3600
15.2 ± 0.8‡,#,§ 23.0 ± 1.1^,§,# 44.9 ± 8.9^,§,# <.0001
Emerald
43.1 ± 11.5†,‡,‱,^ 49.3 ± 5.5†,^,° 66.3 ± 5.6†,^,° <.0001
DWIO
27.8±3.2#,‱ 49.8±5.0‡,§,* 92.1±24.1‡,§,°,* <.0001
Omnicam
28.4±4.5§,^ 38.1±8.8‱,#,°,* 70.4±11.9‱,#,* <.0001
10
The same symbol after SD indicates differences in trueness between scanner pairs (Tukey-adjustment for multiple comparison). Minimum
significant difference across scanners: 7.3ÎŒm, 6.6ÎŒm, 16.8ÎŒm for single crown (SC), partial prosthesis (PP) and full arch (FA), respectively. 1p-
value testing the interaction between scanner and context (SC vs. PP vs. FA) from non-parametric, Kruskall-Wallis test. A p-value >0.05
indicates no difference in scanner trueness according to the context.
Best Result
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
Colorimetric Trueness Map Comparison
11
TRIOS 3 CS 3600 EMERALD DWIO OMNICAM
+0.100 mm
+0.088 mm
+0.077 mm
+0.065 mm
+0.053 mm
+0.042 mm
+0.030 mm
-0.030 mm
-0.042 mm
-0.053 mm
-0.065 mm
-0.077 mm
-0.088 mm
-0.100 mm
Best Result
Trueness in the single crown (SC), partial prosthesis (PP) and full-arch (FA) with the 5 examined intraoral scanners (IOSs): colorimetric maps.
The color maps indicated inward (blue) or outward (red) displacement between overlaid structures, whereas a minimal change was indicated by
a green color. For all three models (SC, PP, FA): the color scale ranged from a maximum deviation of +100ÎŒm and−100ÎŒm, with the best result
given by the deviations comprised between +30ÎŒm and−30ÎŒm (green color).
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
Changes in Mean Trueness Comparison
Changes in mean trueness (standard deviation), in ÎŒm, for the 5 examined scanners, in
the different applications (single crown, SC vs. partial prosthesis, PP vs. full-arch, FA).
12
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
CS 3600
Mean Precision/Standard Deviation
Scanner Single Crown
Mean ± SD
Partial Prosthesis
Mean ± SD
Full Arch
Mean ± SD
p-value1
Trios 3
15.2 ± 0.8†,‡,‱ 21.0 ± 1.9‡,‱ 35.6 ± 3.4†,‡,‱ <.0001
CS 3600
11.3 ± 1.1^,§,# 17.0 ± 2.3^,§,# 35.7 ± 4.3^,§,# <.0001
Emerald 32.8 ± 10.7†,^ 29.9 ± 8.9^,° 61.5 ± 18.1†,^,°,* 0.0007
DWIO 27.1 ± 10.7‡,§ 34.8 ± 10.8‡,§ 111.0 ± 24.8‡,§,°,ç <.0001
Omnicam
30.6 ± 3.3‱,# 43.2 ± 9.4‱,#,° 89.3 ± 14.0‱,#,*,ç <.0001
13
The same symbol after SD indicates differences in precision between scanner pairs (Tukey-adjustment for multiple comparison). Minimum
significant difference across scanners: 8.8 Όm, 9.8 Όm, 19.4 Όm for single crown (SC), partial prosthesis (PP) and full arch (FA), respectively.
1p-value testing the interaction between scanner and context (SC vs. PP vs. FA) from non-parametric, Kruskall-Wallis test. A p-value > 0.05
indicates no difference in scanner precision according to the context.
Best Result Very closely follows
best result
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
Colorimetric Precision Map Comparison
14
Very Close to
Best Results
TRIOS 3 CS 3600 EMERALD DWIO OMNICAM
Best Result
Trueness in the single crown (SC), partial prosthesis (PP) and full-arch (FA) with the 5 examined intraoral scanners (IOSs): colorimetric maps.
The color maps indicated inward (blue) or outward (red) displacement between overlaid structures, whereas a minimal change was indicated by
a green color. For all three models (SC, PP, FA): the color scale ranged from a maximum deviation of +100ÎŒm and−100ÎŒm, with the best result
given by the deviations comprised between +30ÎŒm and−30ÎŒm (green color).
+0.100 mm
+0.088 mm
+0.077 mm
+0.065 mm
+0.053 mm
+0.042 mm
+0.030 mm
-0.030 mm
-0.042 mm
-0.053 mm
-0.065 mm
-0.077 mm
-0.088 mm
-0.100 mm
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
Changes in Mean Precision Comparison
Changes in mean precision (standard deviation), in ÎŒm, for the 5 examined scanners, in
the different applications (single crown vs. partial prosthesis vs. full-arch)
15
CS 3600
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
CS 3600
Results Summary
Single Crown:
CS 3600 had the best trueness (15.2 ± 0.8 Όm) and the best
precision (11.3 ± 1.1 Όm)
Partial Prosthesis
CS 3600 had the best trueness (23 ± 1.1 Όm) and the best
precision(21 ± 1.9 Όm)
Full Arch
CS 3600 had the best trueness (44.9 ± 8.9 Όm)
Trios3 had the best precision (35.6 ± 3.4 Όm), followed very
closely by CS 3600 (35.7 ± 4.3 Όm)
16
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7

Weitere Àhnliche Inhalte

Ähnlich wie CS 3600 True Precision Implantology Study

Micro CT settings for caries detection: how to optimize
Micro CT settings for caries detection: how to optimizeMicro CT settings for caries detection: how to optimize
Micro CT settings for caries detection: how to optimizeIJERA Editor
 
Diagnosis of Vertical Root Fracture Using Digital Radiography, Helical Comput...
Diagnosis of Vertical Root Fracture Using Digital Radiography, Helical Comput...Diagnosis of Vertical Root Fracture Using Digital Radiography, Helical Comput...
Diagnosis of Vertical Root Fracture Using Digital Radiography, Helical Comput...iosrjce
 
190330 AI & cloud based medical/dental SW (KIST êč€ì˜ì€€)
190330 AI & cloud based medical/dental SW (KIST êč€ì˜ì€€)190330 AI & cloud based medical/dental SW (KIST êč€ì˜ì€€)
190330 AI & cloud based medical/dental SW (KIST êč€ì˜ì€€)Youngjun Kim
 
Jc14-12/2016-subharti dental collegeand hospital,Meerut
Jc14-12/2016-subharti dental collegeand hospital,MeerutJc14-12/2016-subharti dental collegeand hospital,Meerut
Jc14-12/2016-subharti dental collegeand hospital,Meerutdeekshasaxena12
 
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...IJERD Editor
 
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...IJERD Editor
 
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...IJERD Editor
 
Implications of cbct in orthodontics
Implications of cbct in orthodonticsImplications of cbct in orthodontics
Implications of cbct in orthodonticsAli Waqar Hasan
 
3D tumor spheroid models for in vitro therapeutic screening: a systematic app...
3D tumor spheroid models for in vitro therapeutic screening: a systematic app...3D tumor spheroid models for in vitro therapeutic screening: a systematic app...
3D tumor spheroid models for in vitro therapeutic screening: a systematic app...Arun kumar
 
Jung2019
Jung2019Jung2019
Jung2019soranrs
 
Clinical accuracy outcome of open and closed trayimpressions
Clinical accuracy outcome of open and closed trayimpressionsClinical accuracy outcome of open and closed trayimpressions
Clinical accuracy outcome of open and closed trayimpressionsNavneet Randhawa
 
Journal club on Connective tissue graft associated or not with low laser ther...
Journal club on Connective tissue graft associated or not with low laser ther...Journal club on Connective tissue graft associated or not with low laser ther...
Journal club on Connective tissue graft associated or not with low laser ther...Shilpa Shiv
 
20221026 implant LL - Association of prosthetic features and peri-implantitis...
20221026 implant LL - Association of prosthetic features and peri-implantitis...20221026 implant LL - Association of prosthetic features and peri-implantitis...
20221026 implant LL - Association of prosthetic features and peri-implantitis...CurtisWeng1
 
International Journal of Engineering Research and Development
International Journal of Engineering Research and DevelopmentInternational Journal of Engineering Research and Development
International Journal of Engineering Research and DevelopmentIJERD Editor
 

Ähnlich wie CS 3600 True Precision Implantology Study (20)

Micro CT settings for caries detection: how to optimize
Micro CT settings for caries detection: how to optimizeMicro CT settings for caries detection: how to optimize
Micro CT settings for caries detection: how to optimize
 
Diagnosis of Vertical Root Fracture Using Digital Radiography, Helical Comput...
Diagnosis of Vertical Root Fracture Using Digital Radiography, Helical Comput...Diagnosis of Vertical Root Fracture Using Digital Radiography, Helical Comput...
Diagnosis of Vertical Root Fracture Using Digital Radiography, Helical Comput...
 
1111.pdf
1111.pdf1111.pdf
1111.pdf
 
1111_2.pdf
1111_2.pdf1111_2.pdf
1111_2.pdf
 
190330 AI & cloud based medical/dental SW (KIST êč€ì˜ì€€)
190330 AI & cloud based medical/dental SW (KIST êč€ì˜ì€€)190330 AI & cloud based medical/dental SW (KIST êč€ì˜ì€€)
190330 AI & cloud based medical/dental SW (KIST êč€ì˜ì€€)
 
Jc14-12/2016-subharti dental collegeand hospital,Meerut
Jc14-12/2016-subharti dental collegeand hospital,MeerutJc14-12/2016-subharti dental collegeand hospital,Meerut
Jc14-12/2016-subharti dental collegeand hospital,Meerut
 
Cone beam
Cone beamCone beam
Cone beam
 
Korte-PachymetryACT13
Korte-PachymetryACT13Korte-PachymetryACT13
Korte-PachymetryACT13
 
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
 
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
 
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
 
Implications of cbct in orthodontics
Implications of cbct in orthodonticsImplications of cbct in orthodontics
Implications of cbct in orthodontics
 
3D tumor spheroid models for in vitro therapeutic screening: a systematic app...
3D tumor spheroid models for in vitro therapeutic screening: a systematic app...3D tumor spheroid models for in vitro therapeutic screening: a systematic app...
3D tumor spheroid models for in vitro therapeutic screening: a systematic app...
 
BIOSTATS
BIOSTATSBIOSTATS
BIOSTATS
 
Jung2019
Jung2019Jung2019
Jung2019
 
Clinical accuracy outcome of open and closed trayimpressions
Clinical accuracy outcome of open and closed trayimpressionsClinical accuracy outcome of open and closed trayimpressions
Clinical accuracy outcome of open and closed trayimpressions
 
Journal club on Connective tissue graft associated or not with low laser ther...
Journal club on Connective tissue graft associated or not with low laser ther...Journal club on Connective tissue graft associated or not with low laser ther...
Journal club on Connective tissue graft associated or not with low laser ther...
 
Muller-JonesHPF_4185022
Muller-JonesHPF_4185022Muller-JonesHPF_4185022
Muller-JonesHPF_4185022
 
20221026 implant LL - Association of prosthetic features and peri-implantitis...
20221026 implant LL - Association of prosthetic features and peri-implantitis...20221026 implant LL - Association of prosthetic features and peri-implantitis...
20221026 implant LL - Association of prosthetic features and peri-implantitis...
 
International Journal of Engineering Research and Development
International Journal of Engineering Research and DevelopmentInternational Journal of Engineering Research and Development
International Journal of Engineering Research and Development
 

KĂŒrzlich hochgeladen

All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...Arohi Goyal
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...chandars293
 
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...GENUINE ESCORT AGENCY
 
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...Anamika Rawat
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀ night ...aartirawatdelhi
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 â‚č5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 â‚č5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 â‚č5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 â‚č5000 To 25K With AC Room 💚😋TANUJA PANDEY
 
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service AvailableCall Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service AvailableJanvi Singh
 
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X79953056974 Low Rate Call Girls In Saket, Delhi NCR
 
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...Namrata Singh
 
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...karishmasinghjnh
 
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service AvailableGENUINE ESCORT AGENCY
 
Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...
Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...
Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...GENUINE ESCORT AGENCY
 
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls ServiceGENUINE ESCORT AGENCY
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...chandars293
 
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappMost Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappInaaya Sharma
 
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...adilkhan87451
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Dipal Arora
 
Call Girls Service Jaipur {8445551418} ❀VVIP BHAWNA Call Girl in Jaipur Raja...
Call Girls Service Jaipur {8445551418} ❀VVIP BHAWNA Call Girl in Jaipur Raja...Call Girls Service Jaipur {8445551418} ❀VVIP BHAWNA Call Girl in Jaipur Raja...
Call Girls Service Jaipur {8445551418} ❀VVIP BHAWNA Call Girl in Jaipur Raja...parulsinha
 
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Availableperfect solution
 

KĂŒrzlich hochgeladen (20)

All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
 
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
 
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀ night ...
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 â‚č5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 â‚č5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 â‚č5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 â‚č5000 To 25K With AC Room 💚😋
 
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service AvailableCall Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
 
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
 
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
 
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
 
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
 
Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...
Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...
Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...
 
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
 
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappMost Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
 
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 
Call Girls Service Jaipur {8445551418} ❀VVIP BHAWNA Call Girl in Jaipur Raja...
Call Girls Service Jaipur {8445551418} ❀VVIP BHAWNA Call Girl in Jaipur Raja...Call Girls Service Jaipur {8445551418} ❀VVIP BHAWNA Call Girl in Jaipur Raja...
Call Girls Service Jaipur {8445551418} ❀VVIP BHAWNA Call Girl in Jaipur Raja...
 
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
 

CS 3600 True Precision Implantology Study

  • 1. CS 3600 Summary from “Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study”
  • 2. Overview Article Name “Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study” Authors ‱ Francesco Guido Mangano, Uli Hauschild, Giovanni Veronesi, Mario Imburgia, Carlo Mangano, and Oleg Admakin Date Published 6 June 2019 Background ‱ Very few peer reviewed studies have evaluated the clinical efficacy of intraoral scanners in implantology. ‱ Even fewer studies compare the accuracy of the various intraoral scanners. ‱ This study compares the trueness and precision of five intraoral scanners, and determines which have both accurate and consistent results. 2
  • 3. What are Trueness and Precision? In this study: Trueness is quantified by distances between generated model and reference model (smaller value means better trueness). Precision is quantified by distances between repeatedly generated models (smaller value means better precision). Accuracy is a combination of trueness and precision. An intraoral scanner must perform well in both trueness and precision tests to be considered “highly accurate.” 3
  • 4. Study Material and Methods 10 scans were performed by a single digital dentistry expert per indication per intraoral scanner device with a zig-zag technique in the same environmental conditions. 2 reference maxillary plaster models were used as a reference: ‱ A - Partially edentulous maxillary arch with: ‱ Implant analog on tooth #23 (implant single support crown simulation) ‱ Implant analog on teeth #14 & 16 (partial implant supported case simulation) ‱ B - Totally edentulous maxillary arch: ‱ Implant analog on teeth #11, 14, 16, 21, 24 & 26 (fixed full arch implant supported denture simulation) Scanbodies used : MegagenÂź PEEK scanbodies to enhance scans (no light reflection). Reference : 3 scans made per indication with Freedom UHD Âź desktop scanner and superimposed in Geomagic Studio 2012Âź to validate the reference models. 4
  • 5. How do these intraoral scanners compare in accuracy? Implantology Intraoral Scanner Comparative Study 5 TRIOS 3 EMERALD DWIO CS 3600 OMNICAM Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
  • 6. Implantology Intraoral Scanner Comparative Study Scanner Producer Technology of Acquisition Powder Color System CS 3600 Carestream Dental, Atlanta, Georgia, USA Structured light-Active Speed 3D Videoℱ No Yes Proprietary files (CSZ), but also open formats (PLY, STL) immediately available Trios 3 3-Shape, Copenhagen, Denmark Structured light –Confocal microscopy and Ultrafast Optical Scanningℱ No Yes Proprietary files (DCM) available, but possibility to export STL files via the new Trios on Dental DesktopÂź Omnicam Dentsply-Sirona, York, Pennsylvania, USA Structured light -Optical triangulation and confocal microscopy No Yes Proprietary files (CS3, SDT, CDT, IDT) are available, possible to export STL files via the Cerec ConnectÂź DWIO Dentalwings, Montreal, Quebec, Canada Blue laser-Multiscan Imagingℱ technology No No Proprietary files (XORDER), but also open formats (STL) immediately available Emerald Planmeca, Helsinki, Finland Red, green and blue lasers-Projected Pattern Triangulationℱ No Yes Open formats (PLY, STL) immediately available 6 Summary of 5 Intraoral Scanners Used in the Study Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
  • 7. Single Crown, Best Result in Trueness (ÎŒM) 7 Deviation: 22 ÎŒm (20) Triangles: 179,496 Deviation: 14 ÎŒm (13) Triangles: 34,980 Deviation: 31 ÎŒm (26) Triangles: 81,712 Deviation: 23 ÎŒm (21) Triangles: 68,281 Deviation: 21 ÎŒm (28) Triangles: 123,417 TRIOS 3 CS 3600 EMERALD DWIO OMNICAM Best Result [Carestream Dental Note: Triangle number is not a measure of acquisition resolution or mesh quality. In fact, in having the best trueness with the smallest number of triangles, CS 3600 demonstrates the most efficient/optimal mesh generation (by placing more triangles in areas with more details and vice versa), with minimized mesh file size while maintaining accuracy.] Single crown: best result in trueness (standard deviation) in ÎŒm, for the 5 examined scanners, and the number of triangles composing each mesh. Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
  • 8. Partial Prosthesis, Best Result in Trueness (ÎŒM) 8 Deviation: 28 ÎŒm (24) Triangles: 207,159 Deviation: 21 ÎŒm (25) Triangles: 55,968 Deviation: 42 ÎŒm (32) Triangles: 113,162 Deviation: 39 ÎŒm (39) Triangles: 135,028 Deviation: 29 ÎŒm (32) Triangles: 192,967 8 TRIOS 3 CS 3600 EMERALD DWIO OMNICAM Best Result Partial prosthesis: best result in trueness (standard deviation) in ÎŒm, for the 5 examined scanners, and the number of triangles composing each mesh. Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7 [Carestream Dental Note: Triangle number is not a measure of acquisition resolution or mesh quality. In fact, in having the best trueness with the smallest number of triangles, CS 3600 demonstrates the most efficient/optimal mesh generation (by placing more triangles in areas with more details and vice versa), with minimized mesh file size while maintaining accuracy.]
  • 9. Full Arch, Best Result in Trueness (ÎŒM) 9 Deviation: 37 ÎŒm (44) Triangles: 326,696 Deviation: 32 ÎŒm (49) Triangles: 83,735 Deviation: 57 ÎŒm (58) Triangles: 178,507 Deviation: 73 ÎŒm (64) Triangles: 227, 491 Deviation: 47 ÎŒm (55) Triangles: 322,929 9 TRIOS 3 CS 3600 EMERALD DWIO OMNICAM Best Result Full arch: best result in trueness (standard deviation) in ÎŒm, for the 5 examined scanners, and the number of triangles composing each mesh. Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7 [Carestream Dental Note: Triangle number is not a measure of acquisition resolution or mesh quality. In fact, in having the best trueness with the smallest number of triangles, CS 3600 demonstrates the most efficient/optimal mesh generation (by placing more triangles in areas with more details and vice versa), with minimized mesh file size while maintaining accuracy.]
  • 10. Mean Trueness/Standard Deviation Scanner Single Crown Mean ± SD Partial Prosthesis Mean ± SD Full Arch Mean ± SD p-value1 Trios 3 22.3 ± 0.5† 28.5 ± 0.5†,‡,‱ 46.3 ± 4.9†,‡,‱ <.0001 CS 3600 15.2 ± 0.8‡,#,§ 23.0 ± 1.1^,§,# 44.9 ± 8.9^,§,# <.0001 Emerald 43.1 ± 11.5†,‡,‱,^ 49.3 ± 5.5†,^,° 66.3 ± 5.6†,^,° <.0001 DWIO 27.8±3.2#,‱ 49.8±5.0‡,§,* 92.1±24.1‡,§,°,* <.0001 Omnicam 28.4±4.5§,^ 38.1±8.8‱,#,°,* 70.4±11.9‱,#,* <.0001 10 The same symbol after SD indicates differences in trueness between scanner pairs (Tukey-adjustment for multiple comparison). Minimum significant difference across scanners: 7.3ÎŒm, 6.6ÎŒm, 16.8ÎŒm for single crown (SC), partial prosthesis (PP) and full arch (FA), respectively. 1p- value testing the interaction between scanner and context (SC vs. PP vs. FA) from non-parametric, Kruskall-Wallis test. A p-value >0.05 indicates no difference in scanner trueness according to the context. Best Result Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
  • 11. Colorimetric Trueness Map Comparison 11 TRIOS 3 CS 3600 EMERALD DWIO OMNICAM +0.100 mm +0.088 mm +0.077 mm +0.065 mm +0.053 mm +0.042 mm +0.030 mm -0.030 mm -0.042 mm -0.053 mm -0.065 mm -0.077 mm -0.088 mm -0.100 mm Best Result Trueness in the single crown (SC), partial prosthesis (PP) and full-arch (FA) with the 5 examined intraoral scanners (IOSs): colorimetric maps. The color maps indicated inward (blue) or outward (red) displacement between overlaid structures, whereas a minimal change was indicated by a green color. For all three models (SC, PP, FA): the color scale ranged from a maximum deviation of +100ÎŒm and−100ÎŒm, with the best result given by the deviations comprised between +30ÎŒm and−30ÎŒm (green color). Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
  • 12. Changes in Mean Trueness Comparison Changes in mean trueness (standard deviation), in ÎŒm, for the 5 examined scanners, in the different applications (single crown, SC vs. partial prosthesis, PP vs. full-arch, FA). 12 Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7 CS 3600
  • 13. Mean Precision/Standard Deviation Scanner Single Crown Mean ± SD Partial Prosthesis Mean ± SD Full Arch Mean ± SD p-value1 Trios 3 15.2 ± 0.8†,‡,‱ 21.0 ± 1.9‡,‱ 35.6 ± 3.4†,‡,‱ <.0001 CS 3600 11.3 ± 1.1^,§,# 17.0 ± 2.3^,§,# 35.7 ± 4.3^,§,# <.0001 Emerald 32.8 ± 10.7†,^ 29.9 ± 8.9^,° 61.5 ± 18.1†,^,°,* 0.0007 DWIO 27.1 ± 10.7‡,§ 34.8 ± 10.8‡,§ 111.0 ± 24.8‡,§,°,ç <.0001 Omnicam 30.6 ± 3.3‱,# 43.2 ± 9.4‱,#,° 89.3 ± 14.0‱,#,*,ç <.0001 13 The same symbol after SD indicates differences in precision between scanner pairs (Tukey-adjustment for multiple comparison). Minimum significant difference across scanners: 8.8 Όm, 9.8 Όm, 19.4 Όm for single crown (SC), partial prosthesis (PP) and full arch (FA), respectively. 1p-value testing the interaction between scanner and context (SC vs. PP vs. FA) from non-parametric, Kruskall-Wallis test. A p-value > 0.05 indicates no difference in scanner precision according to the context. Best Result Very closely follows best result Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
  • 14. Colorimetric Precision Map Comparison 14 Very Close to Best Results TRIOS 3 CS 3600 EMERALD DWIO OMNICAM Best Result Trueness in the single crown (SC), partial prosthesis (PP) and full-arch (FA) with the 5 examined intraoral scanners (IOSs): colorimetric maps. The color maps indicated inward (blue) or outward (red) displacement between overlaid structures, whereas a minimal change was indicated by a green color. For all three models (SC, PP, FA): the color scale ranged from a maximum deviation of +100ÎŒm and−100ÎŒm, with the best result given by the deviations comprised between +30ÎŒm and−30ÎŒm (green color). +0.100 mm +0.088 mm +0.077 mm +0.065 mm +0.053 mm +0.042 mm +0.030 mm -0.030 mm -0.042 mm -0.053 mm -0.065 mm -0.077 mm -0.088 mm -0.100 mm Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
  • 15. Changes in Mean Precision Comparison Changes in mean precision (standard deviation), in ÎŒm, for the 5 examined scanners, in the different applications (single crown vs. partial prosthesis vs. full-arch) 15 CS 3600 Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7 CS 3600
  • 16. Results Summary Single Crown: CS 3600 had the best trueness (15.2 ± 0.8 ÎŒm) and the best precision (11.3 ± 1.1 ÎŒm) Partial Prosthesis CS 3600 had the best trueness (23 ± 1.1 ÎŒm) and the best precision(21 ± 1.9 ÎŒm) Full Arch CS 3600 had the best trueness (44.9 ± 8.9 ÎŒm) Trios3 had the best precision (35.6 ± 3.4 ÎŒm), followed very closely by CS 3600 (35.7 ± 4.3 ÎŒm) 16 Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7