Mrs. BHUVANESHWARI.P
M.Sc. MEDICAL PHYSICS
LECTURER
GANGA COLLEGE OF NURSING
COIMBATORE
PERFORMANCE EVALUATION OF COMPUTED
TOMOGRAPHY (CT) SCANNERS
American Association of Physicists in Medicine initiated
pioneering R &D activities in the field.
These initial works were brought out through:
1.Phantoms for Performance Evaluation and Quality Assurance
of CT scanners (AAPM Report No.1)
2.Specification and Acceptance Testing of Computed
Tomography Scanners (AAPM Report No.39)
Though CT scanners underwent tremendous design
modifications with respect to source -detector configurations
and their relative movements , the basic principles adopted in
the performance evaluation of these scanners still are based on
the above reports.
CT Performance evaluation tests can be broadly
classified into:
1.Electromechanical Tests
2.X-ray Generator (Electrical ) Tests
3.Image Quality Tests
4.Radiation Dose Tests
5.General Tests Related to CT Number
ELECTRO MECHANICAL TESTS
1.Scan Localization laser light accuracy
Scan localization laser lights are provided in the gantry to
locate the scan plane and to position the isocentre of the gantry
at the centre of the slice being imaged.
Modern units are provided with external and internal
localization lights at fixed distance apart (60-70 cm).
External for positioning the patient and internal to define the
plane of slice being imaged.
Test Tool : Laser Light Test Tool
2. Vertical Alignment of Table to Imaging Plane
If alignment fails, isocentre of different slices would be at different
locations. This situation affects the accuracy of treatment plan on the
basis of virtual simulation.
Test Tool : Laser Light Test Tool
3. Table increment /backlash
Table increment or backlash is controlled from computer console.
Hence correspondence between the applied increment/backlash and
the effected increment or backlash is necessary as it determines the
relative location of image slices.
Test Tool : Film (Kodak X-Omat V / industrial films Agfa D2 or D4
pre packed in a light proof envelop so as to get a film density of 1 to
1.5 under exposure parameters 120-140 kVp and 50 –100 mAs)
4. Gantry Tilt
To determine the accuracy of tilt indicators and to ensure that
specified tilt can be accomplished under clinical condition.
Test Method : Test 3 and 4 can be combined using the same
film.
5. Slice Localization from Radiographic (scout) image
To determine the correspondence of the slice localization at
computer consol with the actual slice position and angle.
Test Tool : 450 cross wire tool as described in AAPM report No.
39.
6. Collimation Tests
Collimation determines the slice thickness.
Pre patient collimation/ post patient collimation or both can be
present depending upon the make and model of the scanner.
In pre patient collimation system, slice thickness is defined at
isocentre.
In post patient collimation system, slice thickness is determined
by detector collimators.
In multi channel detector array systems, slice thickness is
determined by the detector array being activated for the scan.
Test Methods:
Slice thickness defined at isocentre can be checked with film.
FWHM of the radiation profile width gives slice thickness.
If slice thickness is determined by detector collimation or
activated detector channels 
slice thickness has to be measured by the response of the scanner
to an attenuating impulse along Z axis at an angle to the the
imaging plane (Z sensitivity)
AAPM -39 suggests a RAMP phantom or WIRE HELIX
phantom for this test.
Above methods for evaluation of slice thickness, however are not
valid for spiral scanners because of orientation difficulties.
Line response phantom is ideal for evaluating spiral CT Scan
Sensitivity Profile.
Line Response phantom :
Comprises of a small thin circular foil (0.1 mm) of high CT number
material (e.g. PTFE) sandwiched between two discs of rigid foam
having CT number nearly equal to that of air.
ImPACT recommends the use of a phantom containing a tungsten
foil (6 mm diameter , 0.05 mm thick sandwiched ) within a perspex
rod.
Test Method : CT number along Z axis of the images of the phantom
rises from background to maximum at the disc level and again falls
of to background because of partial volume averaging. FWHM of CT
number profile along Z axis provides slice thickness.
X-RAY GENERATOR (ELECTRICAL ) TESTS
1. Accelerating Voltage
Non-Invasive Methods for routine evaluation.
Digital kVp meters commercially available are used.
These meters consists of a pair of solid state detectors shielded
with beam hardening filters of different thickness. The ratio of
the signals under different filtration varies with kVp.
Tolerance :  2 kV
2. Measurement of mA linearity
Linearity of mA is inferred by measuring radiation output at
isocentre in air using a dose meter.
mA Linearity:
Keep kV , time and slice thickness constant and measure the
radiation output for different mAs
Tabulate the results as follows:
Sr.
No
Radiation output (mGy) Rav/mAs
mGy/mAs
(X)
R1 R2 R3 R4 Rav
mAs1 X1
mAs2 X2
mAs3 X3
mAs4 X4
mAs5 X5
Coefficient of linearity (COL) is estimated as :
(Xmax-Xmin) / (Xmax + Xmin)
COL  0.01
3.Reproducibility of Radiation Output
Verified by analyzing exposure to exposure variation of radiation
output for applied exposure parameters
kV mAs Radiation output (mGy/mAs) (Xi – Xm) 2
X1 X2 X3 X4 X5 Xm
Coefficient of variation (COV ) =  /m = (1/ Xm ) [  (Xi-Xm)2 / (n-1) ] ½
Tolerance  0.05 i.e within  5 %
IMAGE QUALITY TEST
1.Low Contrast Resolution
Low contrast resolution refers to the capability of the
scanner to image small objects having attenuation properties not
very much different from that of surrounding tissues.
Test Tool : Consists of polystyrene sheet, on which holes of
different diameters are milled, is placed in a 20 cm diameter
perspex phantom filled with water (or 10% dextrose (C6H12O6) in
water).
CT number of polystyrene = -24
CT number of water = 0
Hence contrast variation between polystyrene and water is 2.4% ,
which is low contrast.
Depending upon the selected slice thickness, polystyrene sheet
and water portion of different thickness provides varying low
contrast ( 0.5 to 2.4 %) by partial volume effect.
Low contrast resolution is normally quoted by the minimum hole
size (lp/mm) that could be resolved at a particular low contrast
variation (in %).
Low contrast resolution test
tool developed by RPAD,
BARC
Low contrast is achieved
between polystyrene and
10% dextrose solution
Tolerance: 5 mm hole should
be observable for 1%
contrast change
High Contrast Resolution
Refers to the ability of scanner to distinguish small objects of
high subject contrast, situated close by.
A linear attenuation difference of 10 % is considered as high
contrast in CT.
Perspex of CT number 120 HU and water (CT number 0 HU)
provides contrast of this order (12%)
High Contrast Resolution Test Tool
Comprises of a Perspex disc ( 19 cm diameter , 2 cm thick) with
line pair /hole pair pattern milled on it and positioned in the
middle of a cylindrical water phantom so that the line pair/hole
pair patterns are filled with water.
High contrast is normally specified by the smallest line/hole size
(lp/mm) resolvable at 10- 12 % contrast variation.
Typically,
1.0 lp/mm (object size 0.5 mm) to 0.06 lp/mm (object size 8mm )
A resolution of 0.64 mm ( 0.78 lp/mm should be typically
distinguishable.
High Contrast Resolution Test Tool
developed by RPAD, BARC
Hole size varies from 2.5 mm
diameter (0.2 lp/mm) to 0.5 mm
diameter (1.0 lp/mm) in eight steps.
Tolerance : 1mm hole should be
resolved at 10% contrast change
Noise
The random variation of CT number around a mean value in
the image of a homogeneous body is known as noise.
Test Tool – Water Phantom
Noise is expressed as a percentage
Noise (%) = [  / (CT water – CT air) ] x 100
The dose at which noise is measured should be stated.
RADIATION DOSE TESTS
Dose Descriptors : Computed Tomography Dose Index (CTDI)
and Multiple Scan Average Dose (MSAD)
CTDI is equivalent of the dose value inside the irradiated slice
that would result if the absorbed radiation profile were entirely
concentrated to a rectangular profile of width equal to the nominal
slice thickness.
-  + D(z) dz = CTDI x h
CTDI = (1/h) x -  + D(z) dz
MSAD
In case of multiple scan, the dose in the central region of the
total dose profile increase as a result of superimposition of all
single dose profiles. This increased value is called the Multiple
Scan Average Dose (MSAD).
If the examination is done with overlapping slices, that is by
using a table feed smaller than the slice thickness the increase in
dose becomes even larger.
MSAD = CTDI , if table feed (TF) is equal to slice thickness (h)
i.e , if pitch (p) = 1
Pitch (P) = Table feed (TF) / Slice Thickness (h)
If p is not equal to 1,
MSAD = 1/p (CTDI)
Above relation implies that in order to obtain average dose for a
scan series, it is sufficient to obtain the CTDI from a single scan.
Different CTDI definitions:
1. CTDI FDA = (1/h ) -7h  +7h D(z) dz
Measured in Perspex phantom
Disadvantage : With decreasing slice thickness the length over which
the dose contributions are summed becomes shorter . However the
length of the tails of on the dose profile is not reduced to the same
extent. As a consequence, the dose from smaller slice thickness will
be underestimated.
To avoid this underestimation, the definition of CTDI is modified as
2. CTDI 100 =(1/h) -50mm  +50mm K air(z) dz
Instead of dose, air kerma (Kair ) is used and error of about 10% due
to Kair to dose in Perspex (D(z)) is avoided.
3. Normalized CTDI
n CTDI XYZ = CTDI XYZ / mAs
Normalized CTDI represents the capacity of scanner in terms
of its radiation output and does not convey anything about
patient dose.
Reference to patient dose can be made by multiplying n CTDI
with current-time product (mAs).
CTDI measurement to relate it to patient dose is done in Head
and Body phantoms
Head/Body phantoms : Perspex solid cylinder (Head:16 cm
dia. Body: 32 cm dia. and each with15 cm height) with
through holes at centre and at four points at vertically opposite
directions at periphery (Centres of holes at periphery at 1cm
from the edge of phantom)
Body and head phantoms for CTDI measurement
CTDI was earlier specified as the CTDI at the centre and the
average CTDI at the periphery separately for Head and body
phantoms.
In order to represent both axial and peripheral dose together,
weighted CTDI is introduced.
4. Weighted CTDI
CTDIW = 1/3 CTDI 100 c + 2/3 CTDI 100p
The advantage of CTDI that it enables the use of a single number
instead of two.
The dose display at the operators’ console (in terms of CTDIW)
is presently a mandatory requirement in European Community.
If the pitch factor is not equal to 1, CTDI value has to be divided
by the pitch factor to yield effective CTDI.
5. CTDI w.eff = (1/p ) x CTDI w
Dose is not appropriate to indicate the extent of body irradiation.
Dose Length Product (DLP) is introduced to mention the extent
of body irradiation.
6. DLP = CTDI x n x h (mGy.cm)
n  number of slices, h  slice thickness
DLP w = CTDI w.eff .p. n. h = CTDI w.eff . n . TF ( because p.h
is equal to the table feed)
Measured CTDI value should be preferably within  20 % of the
value quoted by the vendor and should not exceed  40%
GENERAL TESTS RELATED TO CT NUMBER
1.CT Number uniformity
The mean CT number over 100 pixels over various regions
of an image of a homogeneous medium should not be very much
different.
Test Method : In the image of a water phantom , note down CT
number values at centre and four locations at periphery using an
ROI of 2 cm diameter.
Variation should not be more than  5 HU among 100 pixels
Image uniformity checks can be helpful in identifying perturbations
such as beam hardening artifacts , ring artifacts etc.
CT number variations due to slice thickness, phantom size,
phantom position and reconstruction algorithm also can be checked
with the same experimental set up.
1.CT Number Linearity
CT number implies the relative variation in attenuation
properties of a medium in comparison with that of water.
Ideally, different linear attenuation coefficients and
corresponding CT number should have a linear relationship.
This linearity is important when CT scanner is used for
quantitative studies and virtual simulation.
Test Method
CT number linearity can be assessed by scanning samples of
material of known composition having different attenuation
coefficients.
The plot of CT numbers against  values should be a straight
line.
CT Number Linearity Phantom
developed by RPAD, BARC
Built in attenuating materials
are selected to cover the
attenuation coefficients of
lung to bone.
Material used include:
1. Bakelite
2. Perspex
3. Polystyrene
4. Nylon
5. Teflon
6. Bakelite
7. Polypropylene
Quantitative Tests
When CT scanner is used for Bone Mineral Analysis
(BMA), CT numbers are used directly for quantification.
Hence linear relationship between mineral concentration and
corresponding CT numbers is necessary for the effective
BMA.
Test Tool : Elliptical body phantom (Perspex) with hole for
mineral vial inserts at spinal position.
Bone mineral
equivalent
K2HPO4 of
different
concentrations (0,
50, 100, 200
mg/ml) in the vials
were scanned
PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERS

Weitere ähnliche Inhalte

Was ist angesagt?

CTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose IndexCTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose IndexVivek Elangovan
 
Quality assurance of treatment planning system by Rahim Gohar
Quality assurance of treatment planning system by Rahim GoharQuality assurance of treatment planning system by Rahim Gohar
Quality assurance of treatment planning system by Rahim GoharRahim Gohar
 
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEXTISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEXVictor Ekpo
 
Shielding calculation
Shielding calculationShielding calculation
Shielding calculationBBAdhikari
 
MDCT Principles and Applications- Avinesh Shrestha
MDCT Principles and Applications- Avinesh ShresthaMDCT Principles and Applications- Avinesh Shrestha
MDCT Principles and Applications- Avinesh ShresthaAvinesh Shrestha
 
Ct quality control
Ct quality controlCt quality control
Ct quality controlSAJIDEJAZ1
 
TRS 398 (Technical Report Series)
TRS 398 (Technical Report Series)TRS 398 (Technical Report Series)
TRS 398 (Technical Report Series)Vinay Desai
 
Computed Tomography Dose Index
Computed Tomography Dose IndexComputed Tomography Dose Index
Computed Tomography Dose IndexAnjan Dangal
 
Beam modification devices
Beam modification devicesBeam modification devices
Beam modification devicesVibhay Pareek
 
Beam Modification in Radiotherapy
Beam Modification in RadiotherapyBeam Modification in Radiotherapy
Beam Modification in RadiotherapySantam Chakraborty
 
Quality Assurance in Radiotherapy and Dosimetry
Quality Assurance in Radiotherapy and DosimetryQuality Assurance in Radiotherapy and Dosimetry
Quality Assurance in Radiotherapy and Dosimetrymahbubul hassan
 
Image guided radiation therapy (IGRT)
Image guided radiation therapy (IGRT)Image guided radiation therapy (IGRT)
Image guided radiation therapy (IGRT)Anees Muhammed
 

Was ist angesagt? (20)

CTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose IndexCTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose Index
 
Quality assurance of treatment planning system by Rahim Gohar
Quality assurance of treatment planning system by Rahim GoharQuality assurance of treatment planning system by Rahim Gohar
Quality assurance of treatment planning system by Rahim Gohar
 
Ultrasound contrast agents
Ultrasound contrast agentsUltrasound contrast agents
Ultrasound contrast agents
 
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEXTISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
 
Quality Assurance in Radiotherapy
Quality Assurance in RadiotherapyQuality Assurance in Radiotherapy
Quality Assurance in Radiotherapy
 
Qc in xray
Qc in xrayQc in xray
Qc in xray
 
Shielding calculation
Shielding calculationShielding calculation
Shielding calculation
 
Nuclear imaging
Nuclear imagingNuclear imaging
Nuclear imaging
 
MDCT Principles and Applications- Avinesh Shrestha
MDCT Principles and Applications- Avinesh ShresthaMDCT Principles and Applications- Avinesh Shrestha
MDCT Principles and Applications- Avinesh Shrestha
 
Quality assurance
Quality  assuranceQuality  assurance
Quality assurance
 
Ct quality control
Ct quality controlCt quality control
Ct quality control
 
TRS 398 (Technical Report Series)
TRS 398 (Technical Report Series)TRS 398 (Technical Report Series)
TRS 398 (Technical Report Series)
 
Computed Tomography Dose Index
Computed Tomography Dose IndexComputed Tomography Dose Index
Computed Tomography Dose Index
 
Beam modification devices
Beam modification devicesBeam modification devices
Beam modification devices
 
CT History
CT HistoryCT History
CT History
 
Cobalt & linac
Cobalt & linacCobalt & linac
Cobalt & linac
 
Fluoroscopy
FluoroscopyFluoroscopy
Fluoroscopy
 
Beam Modification in Radiotherapy
Beam Modification in RadiotherapyBeam Modification in Radiotherapy
Beam Modification in Radiotherapy
 
Quality Assurance in Radiotherapy and Dosimetry
Quality Assurance in Radiotherapy and DosimetryQuality Assurance in Radiotherapy and Dosimetry
Quality Assurance in Radiotherapy and Dosimetry
 
Image guided radiation therapy (IGRT)
Image guided radiation therapy (IGRT)Image guided radiation therapy (IGRT)
Image guided radiation therapy (IGRT)
 

Ähnlich wie PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERS

Image Quality And Artifacts in Computed Tomography.pptx
Image Quality And Artifacts in Computed Tomography.pptxImage Quality And Artifacts in Computed Tomography.pptx
Image Quality And Artifacts in Computed Tomography.pptxivanKeshari
 
Physical society presentation ashik.pptx
Physical society presentation ashik.pptxPhysical society presentation ashik.pptx
Physical society presentation ashik.pptxAshikurRahman204838
 
Computed Tomography
Computed TomographyComputed Tomography
Computed TomographySujan Poudel
 
L10 Patient Dose
L10 Patient DoseL10 Patient Dose
L10 Patient Doselidgor
 
Dose reduction technique in ct scan
Dose reduction technique in ct scanDose reduction technique in ct scan
Dose reduction technique in ct scanMohd Aiman Azmardi
 
UV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.pptUV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.pptJgdishrathi
 
Qa of ct and mamography
Qa of ct and mamographyQa of ct and mamography
Qa of ct and mamographySelf
 
Exposure factors2
Exposure factors2Exposure factors2
Exposure factors2mr_koky
 
A method for the systematic selection of technique factors in paediatric ct c...
A method for the systematic selection of technique factors in paediatric ct c...A method for the systematic selection of technique factors in paediatric ct c...
A method for the systematic selection of technique factors in paediatric ct c...Nuke_slidesshare
 
Radiographic exposure and image quality
Radiographic exposure and image qualityRadiographic exposure and image quality
Radiographic exposure and image qualityRad Tech
 
Image Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh ShresthaImage Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh ShresthaAvinesh Shrestha
 

Ähnlich wie PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERS (20)

Image Quality And Artifacts in Computed Tomography.pptx
Image Quality And Artifacts in Computed Tomography.pptxImage Quality And Artifacts in Computed Tomography.pptx
Image Quality And Artifacts in Computed Tomography.pptx
 
Physical society presentation ashik.pptx
Physical society presentation ashik.pptxPhysical society presentation ashik.pptx
Physical society presentation ashik.pptx
 
Ayaz_portfolio
Ayaz_portfolioAyaz_portfolio
Ayaz_portfolio
 
Computed Tomography
Computed TomographyComputed Tomography
Computed Tomography
 
computed Tomography
computed Tomographycomputed Tomography
computed Tomography
 
L10 Patient Dose
L10 Patient DoseL10 Patient Dose
L10 Patient Dose
 
Hrct i
Hrct iHrct i
Hrct i
 
Dose reduction technique in ct scan
Dose reduction technique in ct scanDose reduction technique in ct scan
Dose reduction technique in ct scan
 
UV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.pptUV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.ppt
 
Session 6 image quality ct and session 8 dose ct
Session 6  image quality ct and  session 8 dose ctSession 6  image quality ct and  session 8 dose ct
Session 6 image quality ct and session 8 dose ct
 
A_custom-made_linear_array_transducer
A_custom-made_linear_array_transducerA_custom-made_linear_array_transducer
A_custom-made_linear_array_transducer
 
Qa of ct and mamography
Qa of ct and mamographyQa of ct and mamography
Qa of ct and mamography
 
31075
3107531075
31075
 
Exposure factors2
Exposure factors2Exposure factors2
Exposure factors2
 
Basics in ct
Basics in ctBasics in ct
Basics in ct
 
A method for the systematic selection of technique factors in paediatric ct c...
A method for the systematic selection of technique factors in paediatric ct c...A method for the systematic selection of technique factors in paediatric ct c...
A method for the systematic selection of technique factors in paediatric ct c...
 
Radiographic exposure and image quality
Radiographic exposure and image qualityRadiographic exposure and image quality
Radiographic exposure and image quality
 
Tearhertz Sub-Nanometer Sub-Surface Imaging of 2D Materials
Tearhertz Sub-Nanometer Sub-Surface Imaging of 2D MaterialsTearhertz Sub-Nanometer Sub-Surface Imaging of 2D Materials
Tearhertz Sub-Nanometer Sub-Surface Imaging of 2D Materials
 
_kt-110222
_kt-110222_kt-110222
_kt-110222
 
Image Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh ShresthaImage Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
 

Kürzlich hochgeladen

Micro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdfMicro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdfPoh-Sun Goh
 
Web & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdfWeb & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdfJayanti Pande
 
Beyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactBeyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactPECB
 
Z Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphZ Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphThiyagu K
 
Introduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsIntroduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsTechSoup
 
ComPTIA Overview | Comptia Security+ Book SY0-701
ComPTIA Overview | Comptia Security+ Book SY0-701ComPTIA Overview | Comptia Security+ Book SY0-701
ComPTIA Overview | Comptia Security+ Book SY0-701bronxfugly43
 
How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17Celine George
 
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...christianmathematics
 
Food Chain and Food Web (Ecosystem) EVS, B. Pharmacy 1st Year, Sem-II
Food Chain and Food Web (Ecosystem) EVS, B. Pharmacy 1st Year, Sem-IIFood Chain and Food Web (Ecosystem) EVS, B. Pharmacy 1st Year, Sem-II
Food Chain and Food Web (Ecosystem) EVS, B. Pharmacy 1st Year, Sem-IIShubhangi Sonawane
 
Measures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeMeasures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeThiyagu K
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104misteraugie
 
General Principles of Intellectual Property: Concepts of Intellectual Proper...
General Principles of Intellectual Property: Concepts of Intellectual  Proper...General Principles of Intellectual Property: Concepts of Intellectual  Proper...
General Principles of Intellectual Property: Concepts of Intellectual Proper...Poonam Aher Patil
 
PROCESS RECORDING FORMAT.docx
PROCESS      RECORDING        FORMAT.docxPROCESS      RECORDING        FORMAT.docx
PROCESS RECORDING FORMAT.docxPoojaSen20
 
This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.christianmathematics
 
Application orientated numerical on hev.ppt
Application orientated numerical on hev.pptApplication orientated numerical on hev.ppt
Application orientated numerical on hev.pptRamjanShidvankar
 
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...Nguyen Thanh Tu Collection
 
Unit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptxUnit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptxVishalSingh1417
 
Making and Justifying Mathematical Decisions.pdf
Making and Justifying Mathematical Decisions.pdfMaking and Justifying Mathematical Decisions.pdf
Making and Justifying Mathematical Decisions.pdfChris Hunter
 

Kürzlich hochgeladen (20)

Micro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdfMicro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdf
 
Web & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdfWeb & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdf
 
Beyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactBeyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global Impact
 
Z Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphZ Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot Graph
 
Introduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsIntroduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The Basics
 
Asian American Pacific Islander Month DDSD 2024.pptx
Asian American Pacific Islander Month DDSD 2024.pptxAsian American Pacific Islander Month DDSD 2024.pptx
Asian American Pacific Islander Month DDSD 2024.pptx
 
ComPTIA Overview | Comptia Security+ Book SY0-701
ComPTIA Overview | Comptia Security+ Book SY0-701ComPTIA Overview | Comptia Security+ Book SY0-701
ComPTIA Overview | Comptia Security+ Book SY0-701
 
How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17
 
Mehran University Newsletter Vol-X, Issue-I, 2024
Mehran University Newsletter Vol-X, Issue-I, 2024Mehran University Newsletter Vol-X, Issue-I, 2024
Mehran University Newsletter Vol-X, Issue-I, 2024
 
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
 
Food Chain and Food Web (Ecosystem) EVS, B. Pharmacy 1st Year, Sem-II
Food Chain and Food Web (Ecosystem) EVS, B. Pharmacy 1st Year, Sem-IIFood Chain and Food Web (Ecosystem) EVS, B. Pharmacy 1st Year, Sem-II
Food Chain and Food Web (Ecosystem) EVS, B. Pharmacy 1st Year, Sem-II
 
Measures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeMeasures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and Mode
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104
 
General Principles of Intellectual Property: Concepts of Intellectual Proper...
General Principles of Intellectual Property: Concepts of Intellectual  Proper...General Principles of Intellectual Property: Concepts of Intellectual  Proper...
General Principles of Intellectual Property: Concepts of Intellectual Proper...
 
PROCESS RECORDING FORMAT.docx
PROCESS      RECORDING        FORMAT.docxPROCESS      RECORDING        FORMAT.docx
PROCESS RECORDING FORMAT.docx
 
This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.
 
Application orientated numerical on hev.ppt
Application orientated numerical on hev.pptApplication orientated numerical on hev.ppt
Application orientated numerical on hev.ppt
 
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
 
Unit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptxUnit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptx
 
Making and Justifying Mathematical Decisions.pdf
Making and Justifying Mathematical Decisions.pdfMaking and Justifying Mathematical Decisions.pdf
Making and Justifying Mathematical Decisions.pdf
 

PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERS

  • 1. Mrs. BHUVANESHWARI.P M.Sc. MEDICAL PHYSICS LECTURER GANGA COLLEGE OF NURSING COIMBATORE
  • 2. PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERS American Association of Physicists in Medicine initiated pioneering R &D activities in the field. These initial works were brought out through: 1.Phantoms for Performance Evaluation and Quality Assurance of CT scanners (AAPM Report No.1) 2.Specification and Acceptance Testing of Computed Tomography Scanners (AAPM Report No.39) Though CT scanners underwent tremendous design modifications with respect to source -detector configurations and their relative movements , the basic principles adopted in the performance evaluation of these scanners still are based on the above reports.
  • 3. CT Performance evaluation tests can be broadly classified into: 1.Electromechanical Tests 2.X-ray Generator (Electrical ) Tests 3.Image Quality Tests 4.Radiation Dose Tests 5.General Tests Related to CT Number
  • 4. ELECTRO MECHANICAL TESTS 1.Scan Localization laser light accuracy Scan localization laser lights are provided in the gantry to locate the scan plane and to position the isocentre of the gantry at the centre of the slice being imaged. Modern units are provided with external and internal localization lights at fixed distance apart (60-70 cm). External for positioning the patient and internal to define the plane of slice being imaged. Test Tool : Laser Light Test Tool
  • 5. 2. Vertical Alignment of Table to Imaging Plane If alignment fails, isocentre of different slices would be at different locations. This situation affects the accuracy of treatment plan on the basis of virtual simulation. Test Tool : Laser Light Test Tool 3. Table increment /backlash Table increment or backlash is controlled from computer console. Hence correspondence between the applied increment/backlash and the effected increment or backlash is necessary as it determines the relative location of image slices. Test Tool : Film (Kodak X-Omat V / industrial films Agfa D2 or D4 pre packed in a light proof envelop so as to get a film density of 1 to 1.5 under exposure parameters 120-140 kVp and 50 –100 mAs)
  • 6. 4. Gantry Tilt To determine the accuracy of tilt indicators and to ensure that specified tilt can be accomplished under clinical condition. Test Method : Test 3 and 4 can be combined using the same film. 5. Slice Localization from Radiographic (scout) image To determine the correspondence of the slice localization at computer consol with the actual slice position and angle. Test Tool : 450 cross wire tool as described in AAPM report No. 39.
  • 7. 6. Collimation Tests Collimation determines the slice thickness. Pre patient collimation/ post patient collimation or both can be present depending upon the make and model of the scanner. In pre patient collimation system, slice thickness is defined at isocentre. In post patient collimation system, slice thickness is determined by detector collimators. In multi channel detector array systems, slice thickness is determined by the detector array being activated for the scan.
  • 8. Test Methods: Slice thickness defined at isocentre can be checked with film. FWHM of the radiation profile width gives slice thickness. If slice thickness is determined by detector collimation or activated detector channels  slice thickness has to be measured by the response of the scanner to an attenuating impulse along Z axis at an angle to the the imaging plane (Z sensitivity) AAPM -39 suggests a RAMP phantom or WIRE HELIX phantom for this test.
  • 9.
  • 10. Above methods for evaluation of slice thickness, however are not valid for spiral scanners because of orientation difficulties. Line response phantom is ideal for evaluating spiral CT Scan Sensitivity Profile. Line Response phantom : Comprises of a small thin circular foil (0.1 mm) of high CT number material (e.g. PTFE) sandwiched between two discs of rigid foam having CT number nearly equal to that of air. ImPACT recommends the use of a phantom containing a tungsten foil (6 mm diameter , 0.05 mm thick sandwiched ) within a perspex rod. Test Method : CT number along Z axis of the images of the phantom rises from background to maximum at the disc level and again falls of to background because of partial volume averaging. FWHM of CT number profile along Z axis provides slice thickness.
  • 11. X-RAY GENERATOR (ELECTRICAL ) TESTS 1. Accelerating Voltage Non-Invasive Methods for routine evaluation. Digital kVp meters commercially available are used. These meters consists of a pair of solid state detectors shielded with beam hardening filters of different thickness. The ratio of the signals under different filtration varies with kVp. Tolerance :  2 kV 2. Measurement of mA linearity Linearity of mA is inferred by measuring radiation output at isocentre in air using a dose meter.
  • 12. mA Linearity: Keep kV , time and slice thickness constant and measure the radiation output for different mAs Tabulate the results as follows: Sr. No Radiation output (mGy) Rav/mAs mGy/mAs (X) R1 R2 R3 R4 Rav mAs1 X1 mAs2 X2 mAs3 X3 mAs4 X4 mAs5 X5 Coefficient of linearity (COL) is estimated as : (Xmax-Xmin) / (Xmax + Xmin) COL  0.01
  • 13. 3.Reproducibility of Radiation Output Verified by analyzing exposure to exposure variation of radiation output for applied exposure parameters kV mAs Radiation output (mGy/mAs) (Xi – Xm) 2 X1 X2 X3 X4 X5 Xm Coefficient of variation (COV ) =  /m = (1/ Xm ) [  (Xi-Xm)2 / (n-1) ] ½ Tolerance  0.05 i.e within  5 %
  • 14. IMAGE QUALITY TEST 1.Low Contrast Resolution Low contrast resolution refers to the capability of the scanner to image small objects having attenuation properties not very much different from that of surrounding tissues. Test Tool : Consists of polystyrene sheet, on which holes of different diameters are milled, is placed in a 20 cm diameter perspex phantom filled with water (or 10% dextrose (C6H12O6) in water). CT number of polystyrene = -24 CT number of water = 0 Hence contrast variation between polystyrene and water is 2.4% , which is low contrast.
  • 15. Depending upon the selected slice thickness, polystyrene sheet and water portion of different thickness provides varying low contrast ( 0.5 to 2.4 %) by partial volume effect. Low contrast resolution is normally quoted by the minimum hole size (lp/mm) that could be resolved at a particular low contrast variation (in %). Low contrast resolution test tool developed by RPAD, BARC Low contrast is achieved between polystyrene and 10% dextrose solution Tolerance: 5 mm hole should be observable for 1% contrast change
  • 16. High Contrast Resolution Refers to the ability of scanner to distinguish small objects of high subject contrast, situated close by. A linear attenuation difference of 10 % is considered as high contrast in CT. Perspex of CT number 120 HU and water (CT number 0 HU) provides contrast of this order (12%) High Contrast Resolution Test Tool Comprises of a Perspex disc ( 19 cm diameter , 2 cm thick) with line pair /hole pair pattern milled on it and positioned in the middle of a cylindrical water phantom so that the line pair/hole pair patterns are filled with water.
  • 17. High contrast is normally specified by the smallest line/hole size (lp/mm) resolvable at 10- 12 % contrast variation. Typically, 1.0 lp/mm (object size 0.5 mm) to 0.06 lp/mm (object size 8mm ) A resolution of 0.64 mm ( 0.78 lp/mm should be typically distinguishable. High Contrast Resolution Test Tool developed by RPAD, BARC Hole size varies from 2.5 mm diameter (0.2 lp/mm) to 0.5 mm diameter (1.0 lp/mm) in eight steps. Tolerance : 1mm hole should be resolved at 10% contrast change
  • 18. Noise The random variation of CT number around a mean value in the image of a homogeneous body is known as noise. Test Tool – Water Phantom Noise is expressed as a percentage Noise (%) = [  / (CT water – CT air) ] x 100 The dose at which noise is measured should be stated.
  • 19. RADIATION DOSE TESTS Dose Descriptors : Computed Tomography Dose Index (CTDI) and Multiple Scan Average Dose (MSAD) CTDI is equivalent of the dose value inside the irradiated slice that would result if the absorbed radiation profile were entirely concentrated to a rectangular profile of width equal to the nominal slice thickness. -  + D(z) dz = CTDI x h CTDI = (1/h) x -  + D(z) dz
  • 20. MSAD In case of multiple scan, the dose in the central region of the total dose profile increase as a result of superimposition of all single dose profiles. This increased value is called the Multiple Scan Average Dose (MSAD). If the examination is done with overlapping slices, that is by using a table feed smaller than the slice thickness the increase in dose becomes even larger.
  • 21. MSAD = CTDI , if table feed (TF) is equal to slice thickness (h) i.e , if pitch (p) = 1 Pitch (P) = Table feed (TF) / Slice Thickness (h) If p is not equal to 1, MSAD = 1/p (CTDI) Above relation implies that in order to obtain average dose for a scan series, it is sufficient to obtain the CTDI from a single scan.
  • 22. Different CTDI definitions: 1. CTDI FDA = (1/h ) -7h  +7h D(z) dz Measured in Perspex phantom Disadvantage : With decreasing slice thickness the length over which the dose contributions are summed becomes shorter . However the length of the tails of on the dose profile is not reduced to the same extent. As a consequence, the dose from smaller slice thickness will be underestimated. To avoid this underestimation, the definition of CTDI is modified as 2. CTDI 100 =(1/h) -50mm  +50mm K air(z) dz Instead of dose, air kerma (Kair ) is used and error of about 10% due to Kair to dose in Perspex (D(z)) is avoided.
  • 23. 3. Normalized CTDI n CTDI XYZ = CTDI XYZ / mAs Normalized CTDI represents the capacity of scanner in terms of its radiation output and does not convey anything about patient dose. Reference to patient dose can be made by multiplying n CTDI with current-time product (mAs). CTDI measurement to relate it to patient dose is done in Head and Body phantoms Head/Body phantoms : Perspex solid cylinder (Head:16 cm dia. Body: 32 cm dia. and each with15 cm height) with through holes at centre and at four points at vertically opposite directions at periphery (Centres of holes at periphery at 1cm from the edge of phantom)
  • 24. Body and head phantoms for CTDI measurement
  • 25. CTDI was earlier specified as the CTDI at the centre and the average CTDI at the periphery separately for Head and body phantoms. In order to represent both axial and peripheral dose together, weighted CTDI is introduced. 4. Weighted CTDI CTDIW = 1/3 CTDI 100 c + 2/3 CTDI 100p The advantage of CTDI that it enables the use of a single number instead of two. The dose display at the operators’ console (in terms of CTDIW) is presently a mandatory requirement in European Community.
  • 26. If the pitch factor is not equal to 1, CTDI value has to be divided by the pitch factor to yield effective CTDI. 5. CTDI w.eff = (1/p ) x CTDI w Dose is not appropriate to indicate the extent of body irradiation. Dose Length Product (DLP) is introduced to mention the extent of body irradiation. 6. DLP = CTDI x n x h (mGy.cm) n  number of slices, h  slice thickness DLP w = CTDI w.eff .p. n. h = CTDI w.eff . n . TF ( because p.h is equal to the table feed) Measured CTDI value should be preferably within  20 % of the value quoted by the vendor and should not exceed  40%
  • 27. GENERAL TESTS RELATED TO CT NUMBER 1.CT Number uniformity The mean CT number over 100 pixels over various regions of an image of a homogeneous medium should not be very much different. Test Method : In the image of a water phantom , note down CT number values at centre and four locations at periphery using an ROI of 2 cm diameter. Variation should not be more than  5 HU among 100 pixels Image uniformity checks can be helpful in identifying perturbations such as beam hardening artifacts , ring artifacts etc. CT number variations due to slice thickness, phantom size, phantom position and reconstruction algorithm also can be checked with the same experimental set up.
  • 28. 1.CT Number Linearity CT number implies the relative variation in attenuation properties of a medium in comparison with that of water. Ideally, different linear attenuation coefficients and corresponding CT number should have a linear relationship. This linearity is important when CT scanner is used for quantitative studies and virtual simulation. Test Method CT number linearity can be assessed by scanning samples of material of known composition having different attenuation coefficients. The plot of CT numbers against  values should be a straight line.
  • 29. CT Number Linearity Phantom developed by RPAD, BARC Built in attenuating materials are selected to cover the attenuation coefficients of lung to bone. Material used include: 1. Bakelite 2. Perspex 3. Polystyrene 4. Nylon 5. Teflon 6. Bakelite 7. Polypropylene
  • 30. Quantitative Tests When CT scanner is used for Bone Mineral Analysis (BMA), CT numbers are used directly for quantification. Hence linear relationship between mineral concentration and corresponding CT numbers is necessary for the effective BMA. Test Tool : Elliptical body phantom (Perspex) with hole for mineral vial inserts at spinal position. Bone mineral equivalent K2HPO4 of different concentrations (0, 50, 100, 200 mg/ml) in the vials were scanned