Factors Affecting ECG signal
Introduction
Electronic advancements have produced instrumentation with the
potential of providing accurate information and reliable heart rate
alarms. Computerized monitors enhance patient assessment by
allowing uninterrupted electronic observation of heart rate and
ectopic activity with comprehensive automatic reporting
mechanisms. Despite these improvements in instrumentation,
monitoring systems continue to be plagued by problems related to
poor signals. Wandering baselines, small complexes, fuzzy tracings,
and frequent electrode replacement make patient assessment
difficult and induce false heart rate alarms. Improvement in electrode
preparation techniques and a better understanding of the sources of
artifact can enhance equipment performance, resulting in improved
patient assessment, more effective utilization of nursing time, and
reduced operating costs.
ECG MONITORING ARTIFACT
In ECG, artifact is defined as electrical interference or, “noise,” that
is recorded from sources other than the electronic signals of the
heart. Artifact is a distortion of the signal being recorded.
Sources of ECG artifacts
There are basically two types of ECG artifacts:-
1. Internal or physiologic sources
2. External or non-physiologic sources
1.Internal or physiologic sources - .Internal or physiologic
sources of artifact are: (a) signals from other muscles
(electromyographic signals) and (b) signals produced in the
epidermis.
a) Electromyographic (EMG) Signals - All muscle activity
produces electrical signals. Signals from muscles other
than the heart are called EMG signals and appear on the monitor as
narrow, rapid spikes associated with muscle
movement. These signals are sufficiently dissimilar to the ECG
signals that they can be electronically reduced or
"filtered" from the trace. Research shows that the typical value of
30mv skin potential exists in the inside and outside of interface of
skin. When the skin stretches the potential fell to about 25mv, the 5
mv skin potential variation is reflected in the ECG, which is noise
observed by muscle contractions.
Epidermal Signals -
The skin is a source of electrical signals which produce motion
artifact. Studies have revealed that a voltage of several millivolts can
be generated by stretching the epidermis, the outer layer of the
skin. This type of artifact is visible as large baseline shifts occurring
when the patient changes positions in bed, eats or ambulates.
Epidermal artifact is more troublesome than other types of artifact
because: (1) it is difficult to filter electronically and (2) its
amplitude is often larger than the ECG signal.
2.Non-Physiologic Sources of ECG Artifact - There are basically
two types of Non-physiologic sources of ECG artifact-
a) 60 Hz. Pickup - This type of artifact, also called 60 Hz. Interference,
produces a wide, fuzzy baseline. It is related to poor electrode contact
associated with poor skin preparation techniques, dried electrode gel, or
defective patient cables or lead wires. The source of 60 Hz. pickup is the
60 Hz. current which supplies power to the electrical wall outlets. The 60
Hz. energy "radiates" from the electrical wiring in the patient's room and
is received in the lead wires and by the patient. The source of radiation
cannot be eliminated, but modern monitors can reduce 60 Hz. pickup by
filtering and by an electronic technique called common mode rejection.
This technique requires good skin contact by all electrodes. One or more
electrodes with poor contact will result in the wide fuzzy baseline .
Alternating current (AC) describes the type of electricity that we get from
the wall. In the United States, the 60 hertz. Many places in Europe use
50 Hz AC electricity. When an ECG machine is poorly grounded or not
equipped to filter out this interference.
Fig 3 shows that when AC power to the electrical wall outlets.
b) Offset Potentials - An offset potential is a voltage that is stored by the
electrode. This stored voltage will add to the ECG signal and interfere with
it. The amount of offset potential and the length of time required for it to
dissipate are determined by the materials used for the electrode and the
gel. Certain combinations of metals and gels generate large voltages (up to
200 millivolts) with the ability to hold this voltage for long periods of time.
Electrode materials such as silver-silver chloride do not allow significant
buildup of offset potential, whereas stainless steel electrodes have poor
offset characteristics. Most electrodes used for ECG monitoring today are
made of silver-silver chloride. With the common use of the silver-silver
chloride electrode, offset potentials are no longer considered to be a
significant problem.
Fig 4. Shows that when offset potential voltage that is stored by the electrode
Signals produced by the interaction of body fluids and
electrode gel:
Electrode Gel -Until recently, movement of the electrode gel under
the electrode was thought to be the primary cause of motion artifact;
however, studies have revealed that this effect is minimal. The
electrode gel does, however, significantly affect the transmission of
signals from the skin to the electrode. The lack of sufficient electrode
gel, frequently due to evaporation caused by improper storage,
results in 60 Hz. pickup and extremely unstable traces. This type of
artifact is easily identified by very high electrode impedances.
Electrode Impedance- In order for ECG signals to pass from the
body to the electrode, an electrically conductive path between the
skin and electrode must be established. The conductive ability of this
path is referred to as electrode impedance or contact impedance.
Electrode impedance is measured in ohms. High impedance
decreases the conduction of the ECG signal. Low impedances
improve this conduction. The major factors affecting electrode
impedance are: (1) the quantity and quality of gel between the
electrode and the patient and (2) the degree to which the outer layer
of the epidermis (the stratum corneum) has been bridged by the
conductive gel.
Shows that when movement of electrode.
Lead wire and patient cable problems:
-Breaks in the wires and connections between the electrode and the
monitor will always be a source of monitoring problems. Poor contact
at any snap connection, loose pins at the cable end of the lead wire,
and breaks in the conductors of the lead wire or patient cable can
cause intermittent loss of the ECG tracing, 60 Hz. pickup, or trace
instability.
Reversed leads:
Electrode/lead placement is very important. If one were to
accidentally confuse the red and white lead cables i.e. place the
white one where the red one should go, vice versa, he might get an
ECG that looks like this:-
Shows that when elerode are wrong placed.
Reversed leads:
Electrode/lead placement is very important. If one were to
accidentally confuse the red and white lead cables i.e. place the
white one where the red one should go, vice versa, he might get an
ECG that looks like this:-
Shows that when elerode are wrong placed.

Weitere ähnliche Inhalte

Was ist angesagt?

Pre amplifier desiging for phsiological
Pre amplifier desiging for phsiologicalPre amplifier desiging for phsiological
Pre amplifier desiging for phsiological
Surendra Meena
 

Was ist angesagt? (19)

Electrosurgical unit
Electrosurgical unitElectrosurgical unit
Electrosurgical unit
 
Bioelectrodes for ecg- Raghav Mishra
Bioelectrodes for ecg- Raghav MishraBioelectrodes for ecg- Raghav Mishra
Bioelectrodes for ecg- Raghav Mishra
 
Bioelectric potentials
Bioelectric potentialsBioelectric potentials
Bioelectric potentials
 
Electromyography (EMG)
Electromyography (EMG)Electromyography (EMG)
Electromyography (EMG)
 
Electrosurgery and Energised Dissection
Electrosurgery and Energised Dissection Electrosurgery and Energised Dissection
Electrosurgery and Energised Dissection
 
Ec 2021 medical electronics
Ec 2021 medical electronicsEc 2021 medical electronics
Ec 2021 medical electronics
 
Management of Electrical Injuries
Management of Electrical InjuriesManagement of Electrical Injuries
Management of Electrical Injuries
 
Elecrosurgery in hysteroscopy
Elecrosurgery in hysteroscopyElecrosurgery in hysteroscopy
Elecrosurgery in hysteroscopy
 
Electrosurgery
ElectrosurgeryElectrosurgery
Electrosurgery
 
Electrosurgery
ElectrosurgeryElectrosurgery
Electrosurgery
 
mini 2016
mini 2016mini 2016
mini 2016
 
6004312 0000 eng_a_w
6004312 0000 eng_a_w6004312 0000 eng_a_w
6004312 0000 eng_a_w
 
Power sources
Power sourcesPower sources
Power sources
 
Pre amplifier desiging for phsiological
Pre amplifier desiging for phsiologicalPre amplifier desiging for phsiological
Pre amplifier desiging for phsiological
 
Biomedical Instrumentation
Biomedical InstrumentationBiomedical Instrumentation
Biomedical Instrumentation
 
Principles in Electrosurgery
Principles in ElectrosurgeryPrinciples in Electrosurgery
Principles in Electrosurgery
 
Energy sources in surgery
Energy sources in surgery Energy sources in surgery
Energy sources in surgery
 
CLASSIFICATION OF MEDICAL EQUIPMENT
CLASSIFICATION OF MEDICAL EQUIPMENTCLASSIFICATION OF MEDICAL EQUIPMENT
CLASSIFICATION OF MEDICAL EQUIPMENT
 
Energy devices in surgery
Energy devices in surgeryEnergy devices in surgery
Energy devices in surgery
 

Andere mochten auch

Project Customer Relationship Management (CRM)
Project Customer Relationship Management (CRM)Project Customer Relationship Management (CRM)
Project Customer Relationship Management (CRM)
SigridVanneste
 
CAREER OBJECTIVE milon
CAREER OBJECTIVE milonCAREER OBJECTIVE milon
CAREER OBJECTIVE milon
Milon Paul
 

Andere mochten auch (11)

My Resume
My ResumeMy Resume
My Resume
 
WU Best
WU BestWU Best
WU Best
 
Project Customer Relationship Management (CRM)
Project Customer Relationship Management (CRM)Project Customer Relationship Management (CRM)
Project Customer Relationship Management (CRM)
 
Announcements - March 7, 2016
Announcements - March 7, 2016Announcements - March 7, 2016
Announcements - March 7, 2016
 
High School Talent Battles
High School Talent BattlesHigh School Talent Battles
High School Talent Battles
 
WU-3
WU-3WU-3
WU-3
 
148012541 90397156-homemade-recipes-book-soaps-ointments-candles-shampoos-and...
148012541 90397156-homemade-recipes-book-soaps-ointments-candles-shampoos-and...148012541 90397156-homemade-recipes-book-soaps-ointments-candles-shampoos-and...
148012541 90397156-homemade-recipes-book-soaps-ointments-candles-shampoos-and...
 
CAREER OBJECTIVE milon
CAREER OBJECTIVE milonCAREER OBJECTIVE milon
CAREER OBJECTIVE milon
 
เอกสารประกอบการสอน: การพยาบาลผู้ป่วยผู้ใหญ่ที่มีปัญหาเกี่ยวกับระบบสืบพันธุ์แล...
เอกสารประกอบการสอน: การพยาบาลผู้ป่วยผู้ใหญ่ที่มีปัญหาเกี่ยวกับระบบสืบพันธุ์แล...เอกสารประกอบการสอน: การพยาบาลผู้ป่วยผู้ใหญ่ที่มีปัญหาเกี่ยวกับระบบสืบพันธุ์แล...
เอกสารประกอบการสอน: การพยาบาลผู้ป่วยผู้ใหญ่ที่มีปัญหาเกี่ยวกับระบบสืบพันธุ์แล...
 
FAST (Focused Assessment Sonography for Trauma)
FAST (Focused Assessment Sonography for Trauma)FAST (Focused Assessment Sonography for Trauma)
FAST (Focused Assessment Sonography for Trauma)
 
Wounds
WoundsWounds
Wounds
 

Ähnlich wie Factors

various type of artifacts in ECG signal & how it's remove
various type of artifacts in ECG signal & how it's removevarious type of artifacts in ECG signal & how it's remove
various type of artifacts in ECG signal & how it's remove
Manish Kumar
 
Biosignals andthermometry fin
Biosignals andthermometry finBiosignals andthermometry fin
Biosignals andthermometry fin
MUBOSScz
 
unit1 IPORTANT.pptx
unit1 IPORTANT.pptxunit1 IPORTANT.pptx
unit1 IPORTANT.pptx
ponni2
 
BIO ELECTRODES PPT.pptx
BIO ELECTRODES PPT.pptxBIO ELECTRODES PPT.pptx
BIO ELECTRODES PPT.pptx
ponni2
 

Ähnlich wie Factors (20)

various type of artifacts in ECG signal & how it's remove
various type of artifacts in ECG signal & how it's removevarious type of artifacts in ECG signal & how it's remove
various type of artifacts in ECG signal & how it's remove
 
B010521318
B010521318B010521318
B010521318
 
Ieeepro techno solutions ieee 2013 embedded project a review on design of ...
Ieeepro techno solutions    ieee 2013 embedded project a review on design of ...Ieeepro techno solutions    ieee 2013 embedded project a review on design of ...
Ieeepro techno solutions ieee 2013 embedded project a review on design of ...
 
Phonocardiography(PCG)
Phonocardiography(PCG)Phonocardiography(PCG)
Phonocardiography(PCG)
 
Project Report on ECG Transmitter using Agilent ADS (Advance System Design)
Project Report on ECG Transmitter using Agilent ADS (Advance System Design)Project Report on ECG Transmitter using Agilent ADS (Advance System Design)
Project Report on ECG Transmitter using Agilent ADS (Advance System Design)
 
Eg24842846
Eg24842846Eg24842846
Eg24842846
 
intermiate equipments(1).ppt
intermiate equipments(1).pptintermiate equipments(1).ppt
intermiate equipments(1).ppt
 
MEASUREMENT OF BIO POTENTIAL USING TWO ELECTRODES AND RECORDING PROBLEMS
MEASUREMENT OF BIO POTENTIAL  USING TWO ELECTRODES AND RECORDING PROBLEMSMEASUREMENT OF BIO POTENTIAL  USING TWO ELECTRODES AND RECORDING PROBLEMS
MEASUREMENT OF BIO POTENTIAL USING TWO ELECTRODES AND RECORDING PROBLEMS
 
Ix3515341540
Ix3515341540Ix3515341540
Ix3515341540
 
Electroencephalography (EEG)
Electroencephalography (EEG)Electroencephalography (EEG)
Electroencephalography (EEG)
 
Biomedical system
Biomedical systemBiomedical system
Biomedical system
 
Biosignals andthermometry fin
Biosignals andthermometry finBiosignals andthermometry fin
Biosignals andthermometry fin
 
Electro presentation
Electro presentationElectro presentation
Electro presentation
 
11
1111
11
 
Electromyography (emg) basics
Electromyography (emg)   basicsElectromyography (emg)   basics
Electromyography (emg) basics
 
Evaluating ECG Capturing Using Sound-Card of PC/Laptop
Evaluating ECG Capturing Using Sound-Card of PC/LaptopEvaluating ECG Capturing Using Sound-Card of PC/Laptop
Evaluating ECG Capturing Using Sound-Card of PC/Laptop
 
ELECTROSURGERY VSR.pptx
ELECTROSURGERY VSR.pptxELECTROSURGERY VSR.pptx
ELECTROSURGERY VSR.pptx
 
Ccna1 3.0 03 networking media
Ccna1 3.0 03 networking mediaCcna1 3.0 03 networking media
Ccna1 3.0 03 networking media
 
unit1 IPORTANT.pptx
unit1 IPORTANT.pptxunit1 IPORTANT.pptx
unit1 IPORTANT.pptx
 
BIO ELECTRODES PPT.pptx
BIO ELECTRODES PPT.pptxBIO ELECTRODES PPT.pptx
BIO ELECTRODES PPT.pptx
 

Factors

  • 1. Factors Affecting ECG signal Introduction Electronic advancements have produced instrumentation with the potential of providing accurate information and reliable heart rate alarms. Computerized monitors enhance patient assessment by allowing uninterrupted electronic observation of heart rate and ectopic activity with comprehensive automatic reporting mechanisms. Despite these improvements in instrumentation, monitoring systems continue to be plagued by problems related to poor signals. Wandering baselines, small complexes, fuzzy tracings, and frequent electrode replacement make patient assessment difficult and induce false heart rate alarms. Improvement in electrode preparation techniques and a better understanding of the sources of artifact can enhance equipment performance, resulting in improved patient assessment, more effective utilization of nursing time, and reduced operating costs. ECG MONITORING ARTIFACT In ECG, artifact is defined as electrical interference or, “noise,” that is recorded from sources other than the electronic signals of the heart. Artifact is a distortion of the signal being recorded. Sources of ECG artifacts There are basically two types of ECG artifacts:- 1. Internal or physiologic sources 2. External or non-physiologic sources 1.Internal or physiologic sources - .Internal or physiologic sources of artifact are: (a) signals from other muscles (electromyographic signals) and (b) signals produced in the epidermis. a) Electromyographic (EMG) Signals - All muscle activity produces electrical signals. Signals from muscles other than the heart are called EMG signals and appear on the monitor as narrow, rapid spikes associated with muscle movement. These signals are sufficiently dissimilar to the ECG signals that they can be electronically reduced or "filtered" from the trace. Research shows that the typical value of 30mv skin potential exists in the inside and outside of interface of
  • 2. skin. When the skin stretches the potential fell to about 25mv, the 5 mv skin potential variation is reflected in the ECG, which is noise observed by muscle contractions. Epidermal Signals - The skin is a source of electrical signals which produce motion artifact. Studies have revealed that a voltage of several millivolts can be generated by stretching the epidermis, the outer layer of the skin. This type of artifact is visible as large baseline shifts occurring when the patient changes positions in bed, eats or ambulates. Epidermal artifact is more troublesome than other types of artifact because: (1) it is difficult to filter electronically and (2) its amplitude is often larger than the ECG signal. 2.Non-Physiologic Sources of ECG Artifact - There are basically two types of Non-physiologic sources of ECG artifact- a) 60 Hz. Pickup - This type of artifact, also called 60 Hz. Interference, produces a wide, fuzzy baseline. It is related to poor electrode contact associated with poor skin preparation techniques, dried electrode gel, or defective patient cables or lead wires. The source of 60 Hz. pickup is the 60 Hz. current which supplies power to the electrical wall outlets. The 60 Hz. energy "radiates" from the electrical wiring in the patient's room and
  • 3. is received in the lead wires and by the patient. The source of radiation cannot be eliminated, but modern monitors can reduce 60 Hz. pickup by filtering and by an electronic technique called common mode rejection. This technique requires good skin contact by all electrodes. One or more electrodes with poor contact will result in the wide fuzzy baseline . Alternating current (AC) describes the type of electricity that we get from the wall. In the United States, the 60 hertz. Many places in Europe use 50 Hz AC electricity. When an ECG machine is poorly grounded or not equipped to filter out this interference. Fig 3 shows that when AC power to the electrical wall outlets. b) Offset Potentials - An offset potential is a voltage that is stored by the electrode. This stored voltage will add to the ECG signal and interfere with it. The amount of offset potential and the length of time required for it to dissipate are determined by the materials used for the electrode and the gel. Certain combinations of metals and gels generate large voltages (up to 200 millivolts) with the ability to hold this voltage for long periods of time. Electrode materials such as silver-silver chloride do not allow significant buildup of offset potential, whereas stainless steel electrodes have poor offset characteristics. Most electrodes used for ECG monitoring today are made of silver-silver chloride. With the common use of the silver-silver chloride electrode, offset potentials are no longer considered to be a significant problem. Fig 4. Shows that when offset potential voltage that is stored by the electrode
  • 4. Signals produced by the interaction of body fluids and electrode gel: Electrode Gel -Until recently, movement of the electrode gel under the electrode was thought to be the primary cause of motion artifact; however, studies have revealed that this effect is minimal. The electrode gel does, however, significantly affect the transmission of signals from the skin to the electrode. The lack of sufficient electrode gel, frequently due to evaporation caused by improper storage, results in 60 Hz. pickup and extremely unstable traces. This type of artifact is easily identified by very high electrode impedances. Electrode Impedance- In order for ECG signals to pass from the body to the electrode, an electrically conductive path between the skin and electrode must be established. The conductive ability of this path is referred to as electrode impedance or contact impedance. Electrode impedance is measured in ohms. High impedance decreases the conduction of the ECG signal. Low impedances improve this conduction. The major factors affecting electrode impedance are: (1) the quantity and quality of gel between the electrode and the patient and (2) the degree to which the outer layer of the epidermis (the stratum corneum) has been bridged by the conductive gel. Shows that when movement of electrode. Lead wire and patient cable problems: -Breaks in the wires and connections between the electrode and the monitor will always be a source of monitoring problems. Poor contact at any snap connection, loose pins at the cable end of the lead wire, and breaks in the conductors of the lead wire or patient cable can cause intermittent loss of the ECG tracing, 60 Hz. pickup, or trace instability.
  • 5. Reversed leads: Electrode/lead placement is very important. If one were to accidentally confuse the red and white lead cables i.e. place the white one where the red one should go, vice versa, he might get an ECG that looks like this:- Shows that when elerode are wrong placed.
  • 6. Reversed leads: Electrode/lead placement is very important. If one were to accidentally confuse the red and white lead cables i.e. place the white one where the red one should go, vice versa, he might get an ECG that looks like this:- Shows that when elerode are wrong placed.