A presentation on Regulatory Approval Process for Medical Devices in European Union that explains in brief about the various aspects including the EU Medical Device Directives, Classifications, CE Certification, Medical Device Registration & Timelines. This was presented as a part of curriculum by Akshay Anand in JSS College of Pharmacy, Mysuru during January 2015
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Â
Regulatory Approval Process for Medical Devices in EU - Presentation by Akshay Anand
1. AKSHAY ANAND M
I Year â Master of Pharmacy
Pharmaceutical Regulatory Affairs
JSS College Of Pharmacy
Mysuru 570015, Karnataka
REGULATORYAPPROVAL
PROCESSFOR
MEDICALDEVICESINEUROPEANUNION
Š2015
akshayanand
2. ASPECTS
Š2015
akshayanand
⢠Medical Devices âWhat are they?
⢠European Union â An Overview
⢠EU Directives that impact Medical Devices
⢠Medical Devices as defined by EU
⢠Classification of Medical Devices as per EU
⢠The CE Certification
⢠The Registration Process for MD
⢠The ProcessTimeline for Registration
OF THIS PRESENTATION
3. 01
⢠A medical device is an instrument, apparatus, implant, in
vitro reagent, or similar or related article that is used to
diagnose, prevent, or treat disease or other conditions
⢠It does not achieve its purposes through chemical action within
or on the body.
⢠Medical devices vary greatly in complexity and application.
⢠The design of medical devices constitutes a major segment of
the field of biomedical engineering.
⢠Examples range from simple devices such as tongue
depressors, medical thermometers, and disposable gloves to
advanced devices such as computers which assist in the
conduct of medical testing, implants, and prostheses.
⢠The global medical device market reached roughly $209 billion
in 2006.
Š2015
akshayanand
5. 03
⢠28 member states plus
European FreeTrade
Association members.
⢠24 languages.
⢠Population of more than
500 million people.
⢠GDP of $18.124 trillion.
⢠Euro, the currency
equalling 71.47 Indian
Rupee per unit.
Š2015
akshayanand
6. 04
⢠Medical Devices (MDD) 93/42/EEC
⢠Covers the bulk of medical devices marketed in the EU
⢠Examples: orthopaedic implants, heart valves, medical software
Š2015
akshayanand
⢠Active Implantable Medical Devices (AIMDD) 90/385/EEC
⢠Covers devices that require external power sources in order to
function properly
⢠Examples: pacemakers, implantable defibrillators
⢠InVitro Diagnostics (IVDD) 98/79/EC
⢠Covers devices used for the examination of specimens taken from the
human body
⢠Examples: pregnancy self-testing strips, blood glucose self-testing
strips
7. 05
Š2015
akshayanand
⢠Personal Protective Equipment 89/686/EEC
⢠LowVoltage 2006/95/EC
⢠Electromagnetic Compatibility 2004/108/EC
⢠Blood Product Directive 2002/98/EC
⢠AnimalTissue Use in Medical Devices 2003/32/EC
⢠HumanTissue Products Directives 2004 & 2005
⢠Breast Implants Reclassification Directive 2003/12/EC
⢠Hip, Knee, Shoulder Joint Replacement Reclassification
Directive 2005/50/EC
8. 06
This guidance document under Directive 93/42/EEC on Medical
Devices provides definition of Medical Devices to include the
following:
Medical Devices: Guidance Document
MEDDEV 2.1/1 APRIL 1994
a. Devices â Accessory
b. Medical Purpose
c. Customizing
d. Medical âToiletry Purpose
e. Aids for Handicapped Persons
f. Software
g. Multipurpose Products
Š2015
akshayanand
9. 07
⢠The classification of medical devices is a ârisk basedâ system based on the
vulnerability of the human body taking account of the potential risks
associated with the devices.
⢠This approach allows the use of a set of criteria that can be combined in
various ways in order to determine classification, e.g. duration of contact
with the body, degree of invasiveness and local vs. systemic effect. These
criteria can then be applied to a vast range of different medical devices
and technologies.
⢠These are referred to as the âclassification rulesâ and are set out in Annex
IX of Directive 93/42/EEC.
Medical Devices: Guidance Document
MEDDEV 2.4/1 Rev.8 JULY 2001
MEDDEV 2.4/1 Rev.9 JUNE 2010
Š2015
akshayanand
10. ⢠The CE mark, or formerly EC mark, is a mandatory conformity marking
for certain products sold within the European Economic Area (EEA) since
1985.
⢠The CE marking is also found on products sold outside the EEA that are
manufactured in, or designed to be sold in, the EEA.
⢠This makes the CE marking recognizable worldwide even to people who
are not familiar with the European Economic Area.
⢠It is in that sense similar to the FCC Declaration of Conformity used on
certain electronic devices sold in the United States.
⢠It consists of the CE logo and, if applicable, the four digit identification
number of the notified body involved in the conformity assessment
procedure.
⢠The CE marking is the manufacturer's declaration that the product meets
the requirements of the applicable EC directives.
08
Š2015
akshayanand
12. 10
Š2015
akshayanand
KEY PLAYERSIN THE
CE MARKING PROCESS(REGULATORY APPROVAL PROCESS FOR MEDICAL DEVICES)
⢠Manufacturers and
Subcontractors
⢠Competent Authorities
⢠Notified Bodies
⢠Authorized Representatives
⢠Distributors
13. 11
Š2015
akshayanand
⢠Competent Authorities are entities
enforcing Medical Device Directives at
the national level in each EU member
state.
⢠Member states and their Competent
Authorities each have their own
interpretations of how Directives
should be carried out.
⢠Manufacturers seeking to sell in more
than one EU member state should take
this into account.
14. 12
Š2015
akshayanand
⢠Notified Bodies are authorized to audit manufacturersâ quality
systems and test/review devices for compliance with applicable
EU Directives and standards.
⢠Their responsibilities include:
⢠Advising on device classification and ConformityAssessment routes
⢠Pre-assessment of devices
⢠Product and quality system evaluation
⢠Concluding and evaluating manufacturersâCorrective Actions
⢠Issuance of certifications
⢠Maintaining programmed surveillance of devices in the EU
15. 13
Š2015
akshayanand
⢠Required for any firm without a location in the EU.
⢠Their responsibilities include:
⢠Liaising between manufacturers and Competent Authorities
⢠Listing their names on clientsâ device labels and packaging
⢠Notifying Competent Authorities of manufacturersâ and devicesâ names
⢠KeepingTechnical Files/Design Dossiers available for review by
Competent Authorities
⢠Participating in vigilance/post-market surveillance procedures
EC REP
16. 14
Š2015
akshayanand
Identify Directives and Regulations appropriate for your device
Classify your device according to MDD Annex IX
Implement Quality System in compliance with ISO 13485.
PrepareTechnical File or Design Dossier demonstrating
compliance with MDD 93/42/EEC.
17. 15
Š2015
akshayanand
Appoint your Authorized Representative.
Have Notified Body audit your Quality System
andTechnical File or Design Dossier.
Register your device with Competent Authorities if necessary.
Prepare Declaration of Conformity and affix
CE Marking to your device.
18. 16
Š2015
akshayanand
⢠First, determine which Medical Device Directive (MDD,
AIMDD or IVDD) applies to your device.
⢠Then determine device classification using Annex IX of
MDD 93/42/EEC.
Class I Non-Sterile
Class I Non-Measuring
Class I Sterile
Class I Measuring
Low-risk, non-invasive devices such as non-sterile
wound dressings and stethoscopes
19. 17
Š2015
akshayanand
Class IIa
Medium-risk, short-term invasive devices such as
tracheal tubes and lancets
Class IIb
Higher-risk, often longer-term surgically invasive,
sometimes implantable devices including intra-ocular
lenses and surgical lasers
Class III
Highest-risk devices, including all active
implantable devices such as replacement heart
valves and vascular stents
20. 18
Š2015
akshayanand
⢠MDD 93/42/EEC Annex IX contains 18 rules for medical
device classification based on factors including the
following:
⢠Deviceâs intended use
⢠Active versus non-active functionality
⢠Deviceâs duration of contact with patient
⢠Degree of invasiveness
⢠Part of body contacted by device
⢠Special situations
21. 19
Š2015
akshayanand
⢠Duration of Contact Criteria
⢠Less than 60 minutes =Transient
⢠Less than 30 days = Short-term
⢠More than 30 days = Long-term
⢠Degree of InvasivenessCriteria
⢠Application to body surface versus an orifice
⢠Surgically invasive or non-invasive
⢠Implantable
Determining your deviceâs duration of contact and degree of
invasiveness will help determine how to classify the device.
23. 21
Š2015
akshayanand
Manufacturers of Class I
sterile/measuring, IIa, IIb and III devices
must implement quality management
systems (QMS) (in compliance with
Annex II orV of the MDD).
Most manufacturers meet this
requirement using ISO 13485.
24. 22
Š2015
akshayanand
Phase I: Quality system preparation and development
⢠Management buy-in
⢠Assign resources and staff
⢠Conduct gap analysis
⢠Develop quality manual and documentation controls
Phase II: Implementation
⢠Employee training
⢠Internal auditing
⢠Corrective and preventative actions
Phase III: Registration
⢠Registration audit performed by Notified Body
⢠Final ISO 13485 certification
25. 23
Š2015
akshayanand
⢠All Class I, IIa and IIb device manufacturers
must prepareTechnical Files containing data
to demonstrate MDD 93/42/EEC compliance.
⢠Class III device manufacturers must prepare
Design Dossiers to demonstrate compliance
to MDD.
⢠All manufacturers must have clinical data to
demonstrate compliance to MDD.
26. 24
Š2015
akshayanand
1
Description of product family and justification for why your device
falls into that family
2 Device intended use
3
Description of device components, specifications, packaging and
literature
4 Device manufacturing Process
5 List of accessories to your device
6 Location of design responsibility and manufacturing facilities
7 Classification along with rationale for classification
27. 25
Š2015
akshayanand
8 Chosen compliance route according to applicable Directive(s)
9
Declaration of Conformity stating manufacturerâs compliance
with applicable Directive(s)
10 Shelf life and environmental limitations of device
11
Retention of quality assurance, Competent Authority and Notified
Body records
12 Vigilance reporting and Medical Device Reporting procedures
13 How and when to contact Competent Authorities
14 Name of and contract with yourAuthorized Representative
28. 26
Š2015
akshayanand
15 Subcontractor names and addresses if applicable
16 Essential Requirements
17 Design input specifications
18 Application and references to Standards and Guidelines
19 Testing results and clinical evaluations
20 Risk analysis
21 Instructions for Use and Labeling
29. 27
Š2015
akshayanand
⢠EU regulators have increased their
requirements for clinical data even for Class I
medical devices in recent years.
⢠In many cases, manufacturers must submit
Clinical Evaluation Summary Reports that
verify the safety and performance of their
products along withTechnical Files or
Design Dossiers.
31. 29
Š2015
akshayanand
⢠Registering your device with Competent Authorities before commercialization
⢠Serving as primary point of contact with all EU Competent Authorities
⢠Keeping a current copy of yourTechnical File/Design Dossier available for
inspections by Competent Authorities
⢠Authorizing you to place EC REP name and address on your device labels,
packaging and instructions for use
⢠Handling Incident and Field Safety Corrective Action (FSCA) reporting to
Competent Authorities
⢠Representing you before the European Commission in the event that your
product is withdrawn in a member state
⢠Protecting the confidentiality of your product information
⢠Maintaining reports on clinical evaluation data for review by Competent
Authorities
⢠Communicating with Competent Authorities on serious device incidents or FSCAs
32. 30
Š2015
akshayanand
Following appointment of your EC REP, your quality system and
Technical File/Design Dossier must be audited by a Notified Body.
Upon Successful completion of your audit, the Notified Body will
issue a CE Certificate.
Manufacturers of Class I Non-sterile/Non-measuring devices self-declare MDD
conformity and do not require CE Marking certificates.
Class IIa Class IIb Class IIIClass I S/M
33. 31
Š2015
akshayanand
Class I device manufacturers must register with the Competent
Authority of the country where their EC REPs are based.
Class I S/MClass I NS/NM
Class IIa Class IIb Class III
Some EU member states do not require registration of Class IIa,
IIb or III devices with Competent Authorities. Other countries do
not require registration of devices.
34. 32
Š2015
akshayanand
⢠In order to complete the CE Marking process, all
manufacturers must prepare a Declaration of
Conformity stating their device is in compliance
with all applicable Directives.
⢠Upon completion of your Declaration of
Conformity, you may affix CE Marking to your
product.
36. 34
Š2015
akshayanand
⢠EU member states MAY require labeling
information be made available in their local
languages or another European Community
language to patients and users in
accordance with Annex I, point 13 of the
MDD.
⢠All languages subject to subsidiarity All EU
member states require device labeling in
their official languages Label information in
yourTechnical File should only appear in
ONE language
37. 35
Š2015
akshayanand
Source: EmergoGroup.com
Device
Classification in
Europe
Wait Time after
submission until
approval is
granted
Validity
Period for
CE Marking
Certificate
Advance Start
Time for
Registration
Renewal
Complexity of
Registration
Process
Overall Cost of
gaining
Regulatory
Approval
Class I NS/NM <1 Month No Expiry Not Applicable
Class I S/M 3-5 Months 3 Years 2 Months
Class IIa 3-5 Months 3 Years 2 Months
Class IIb 3-6 Months 3 Years 2 Months
Class III 6-9 Months 3 Years 2 Months
Simple Complex Low High
Simple Complex Low High
Simple Complex Low High
Simple Complex Low High
Simple Complex Low High
38. 36
Š2015
akshayanand
⢠PMS required for all medical devices sold in the EU
⢠Manufacturers are required to develop proactive plans for PMS
⢠PMS plans should include reviews of market and field data, complaints and
corrective actions
⢠Systematic records of PMS findings must be kept
⢠Purpose ofVigilance Requirements:
⢠Protecting public health and safety
⢠Evaluating incidents to prevent recurrence
⢠Determining effectiveness of corrective actions taken
⢠Monitoring and learning from experience
⢠Vigilance Requirements for Manufacturers:
⢠Establish and maintain systematic procedures to review experience
gained from devices in the post-production phase
⢠Implement appropriate means to apply any necessary corrective actions
39. 37
⢠https://en.wikipedia.org/wiki/Medical_device - Wikipedia article on Medical Devices
⢠http://eur-lex.europa.eu/LexUriServ/LexUriServ.do?uri=CONSLEG:1993L0042:20071011:en:PDF â Official EU
Document : Council Directive 93/42/EEC
⢠http://ec.europa.eu/health/medical-devices/documents/guidelines/index_en.htm - Guidance MEDDEVs Online
Directory
⢠http://ec.europa.eu/health/medical-devices/files/meddev/2_1-1___04-1994_en.pdf - Guidance Document
MEDDEV 2.1/1April 1994
⢠http://ec.europa.eu/health/medical-devices/files/meddev/2_2_4-1part1_07-2001_en.pdf - Guidance Document
MEDDEV 2.4/1 Rev.8 July 2001
⢠http://ec.europa.eu/health/medical-devices/files/meddev/2_4_1_rev_9_classification_en.pdf - Guidance
Document MEDDEV 2.4/1 Rev.9 June 2010
⢠https://en.wikipedia.org/wiki/CE_marking- Wikipedia article on CE Certification
⢠http://www.lne-gmed.com/pdf/en/annex9-directive-93-42-amended.pdf - Official EU Document : Annex IX of
Council Directive 93/42/EEC
⢠http://ec.europa.eu/health/medical-devices/faq/market_en.htm - Information about the Medical Device
Registration on the Official Website of European Commission
⢠http://medicaldevices.bsigroup.com/en-GB/our-services/ce-marking/ - Information about CE Marking Approval
for Medical Devices in Europe
⢠http://www.emergogroup.com/resources/europe-process-chart - Europe CE Approval Process for Medical
Devices â Charts &Timelines