SlideShare ist ein Scribd-Unternehmen logo
1 von 30
Downloaden Sie, um offline zu lesen
GRR Made Simple
1
When looking at GRR and
MSA all the information is
“EXACTLY CORRECT”
There is however a LOT of detailed information to
understand
2
Is there a way to make it
simpler to understand
and implement?
Especially for Test
Engineers?
3
YES!!
4
There are 2 boundary conditions
that a test engineer can
understand that are only
“Sometimes” shown
5
This is the ïŹrst!!
6
This is the second!!
7
Using freely available
spreadsheet off of the Internet
for Excel analysis of GRR a
model was developed for
analysis using these 2
boundary conditions
8
Test Engineers
have been told to
follow 2 boundary
conditions
1. The error in the
measurement has to be 10
times less than the
measured value.
2. The error between the
USL and LSL has to be 10%
of the difference between
USL and LSL
X	Instrument	Error	
0
2X	Instrument	Error	
LSL
USL
ΔCV
ΔCV	-	Boundary Condition # 2
LCV - Boundary Condition # 1
Test Engineering Boundary Conditions
LCV
Where	
USL	=	Upper	SpeciïŹca>on	Limit								LSL	=	Lower	SpeciïŹca>on	Limit
LCV	=	Lowest	Capable	Value 							ΔCV	=	USL	-	LSL	=	DiïŹ€eren>al	Capability	
The variable, the RED “X”, needs to be understood
9
GRR Tools
Critical Evaluation with underlying Test
Engineering Boundary Conditions
The model created is using an OE (offset error of
±4mV with GE (gain error = 0%). The initial
evaluation used 10 samples, 3 tests, 3 Testers. One
tester had 0 offset, another -4mV, and the third
+4mV. Repeatability was 0.0001 to insure that just
equipment reproducibility is what is being examined.
Manage Instrument Error to achieve a passing
qualification on devices for GRR: ANOVA or XBAR&R
The experiment objective is to understand instrument
specification and how it affects GRR ALONE
10
GRR Tools
Manage Instrument Error to achieve a passing
qualification on devices for GRR: ANOVA or XBAR&R
GRR Calculation
# Samples
# Testers
Methods
# of Measurements
USL & LSL
Objective: Understand GRR from Instrument
Error Perspective
2-10 samples
Sample location: determines GRR
Random Selection
Intelligent selection
Plus Offset - one tester
Zero Offset - one tester
Minus Offset - one tester
ANOVA
Xbar&R
One value - Ideal
Second value: + 0.0001
Third value: - 0.0001
Note: sample location is
NOT usually looked at;
and it turns out to be
very important
11
How do these affect GRR
Results
‱ There are essentially 2 methods for doing GRR
‱ ANOVA
‱ Xbar&R
‱ Spreadsheets for each can be found on the internet free of
charge. Some with just one of the methods or the other, and
at least one with both methods in the spreadsheet.
‱ Each method was confirmed to give the same results with
identical data when comparing ANOVA to ANOVA and
Xbar&R to Xbar&R
12
The question is:
Are the 2 boundary conditions listed on slides 6 and 7 necesary
and sufficient to insure that the goal of GRR 10% passes?
1. Random samples within USL and LSL
2. 10 perfect evenly distributed samples within USL and LSL
3. 10 select samples to cause worst case within center ±50% of perfect samples placement
4. 10 select samples to cause worst case within center ±100% of perfect samples placement
5. 10 select samples to cause worst case within center ±150% of perfect samples placement
6. 10 select samples to cause worst case within center ±200% of perfect samples placement
6 ways to look at:
# Instrument errors range from 10X to shown on graphs
13
Necessary Equations
The diagram on Slide 2 shows 11X for instrument errors
to take into account the need for repeatability. This
works out to a GRR of approximately 9% for just
instrument errors, which is the objective of the exercise
USL =
LSL 1+10GE( )+ 20OE
1−10GE
USL − LSL = ΔCV
ΔCV = 10 GE i(USL + LSL)+ 2 iOE( )
ΔCV = 20 iOE +10 GE i(USL + LSL)( )
LSL = +LCV =
OE
0.1− GE
+LCV =
OE
0.1− GE
The model created us using an OE (offset error of ±4mV with GE
(gain error = 0%). The initial evaluation used 10 samples, 3 tests, 3
Testers. One tester had 0 offset, another -4mV, and the third +4mV.
Repeatability was 0.0001 to insure that just equipment reproducibility
is what is being examined.
14
0
LSL
USL
±100% of perfect
samples
placement
±50% of perfect
samples
placement
0
LSL
USL
Random perfect
samples
placement
0
LSL
USL
Perfect Placement
±100 % Placement - Worst Case
0
LSL
USL
0
LSL
USL
Perfect Placement
±50 % Placement - Worst Case
15
# Instrument errors range from 10X to what is shown on graph, 10 measurement points perfectly evenly distributed
0
LSL
USL
10X 11X
0
LSL
USL
12X
0
LSL
USL
13X
0
LSL
USL
14X
0
LSL
USL
15X
0
LSL
USL
16X
0
LSL
USL
24X
0
LSL
USL
23X
0
LSL
USL
22X
0
LSL
USL
21X
0
LSL
USL
20X
0
LSL
USL
19X
0
LSL
USL
18X
0
LSL
USL
17X
0
LSL
USL
Red lines represent perfect evenly spaced samples
16
Random selection of measurement for 10 DUTs with zero offset, +
offset and - offset.
1.5200 10 device randomly selected measurement values
using Monte Carlo simulations for Xbar&R from Statistical
Solutions, Tolerance and GRR results.
2.1010 10 device randomly selected measurement values
using Monte Carlo simulations using the ANOVA method
from www.dmaictools.com, Tolerance and GRR results.
Lower Specification limits started at 40mV (10X), Upper Specification limit start at 120mV (10X), ranging to
40X instrument errors. This means that the difference between USL-LSL, ranged from 10X to 40X.
Simulations were done for a process distribution width of 5.15 (encloses the central 99% of the process
distribution).
0
LSL
USL
Please note that the number of instrument
errors is the inverse of the instrument error
17
409 12 14 16 18 20 22 24 26 28 30 32 34 36 38
1
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
(# X) Number of Instrument Errors
Percentage(%)over10%
GRR%	Xbar&R
Xbar&R	Monte	Carlo	Simula3on	
in	Excel.	5200	simula3ons	of	10
devices.	
Random	devices	within	limits	
Using	Gage	R&R	
(XBar&R	Motorola	Version)	
from		Sta;s;cal		Solu;ons
Please	note	that	11X	and	
greater	is	acceptable.	However	
neither	the	Xbar&R	Method	
nor	ANOVA	correctly	calculate	
GRR	at	the	corners	of
instrument	limits
GRR%	ANOVA
ANOVA	Monte	Carlo	Simula3on	
in	Excel.	1010	simula3ons	of	10
devices.	
Random	devices	within	limits	
± 4mV offset Error
LSL starts at +40mV (10X)
USL starts at +120mV
USL-LSL starts at 80mV (10X)
Anova - www.dmaictools.com/measure/grr
30.0
Percentage of simulations (Monte Carlo) that showed the instances greater than GRR
10% for each Instrument X.
To read the graph, 20 on the X axis is the number of instrument errors. For GRR ANOVA,
approximately 0.5% of the simulation results showed greater than a GRR of 10%. For
GRR Xbar&R 1% of the simulation results showed greater than a GRR of 10%.
0
LSL
USL
Random
samples
placement
18
4. 10 select perfect samples to cause worst case
within center ±50% of perfect samples placement
NOTE: solid line on graph of next slide
1. ANOVA 5.15 Standard deviations
2.Xbar&R 5.15 Standard deviations
Lower Specification limits started at 40mV (10X), Upper Specification limit
start at 120mV (10X), ranging to 40X instrument errors. This means that
the difference between USL-LSL, ranged from 10X and following.
Calculations were done for a process distribution width of 5.15 (encloses
the central 99% of the process distribution) at worst case values..
0
LSL
USL
±50% of perfect
samples
placement
Please note that the number of instrument
errors is the inverse of the instrument error
19
4816 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47
0.1
0.035
0.04
0.045
0.05
0.055
0.06
0.065
0.07
0.075
0.08
0.085
0.09
0.095
# Instrument Errors
GRR
ANOVA		±50%	of	perfect	sample	placement
ANOVA		±100%	of	perfect	sample	placement
10	Perfectly	Measured	Device	Samples	for	QualiïŹca:on,	1	tester	0	oïŹ€set,	1	tester	+oïŹ€set,	1	tester	-	oïŹ€set
ANOVA		±150%	of	perfect	sample	placement
ANOVA		±200%	of	perfect	sample	placement
Anova	Perfect			16.5	(6.06%)
Anova	±50	 			17.8	(5.62%)
Anova	±100						19.3	(5.18%)
Anova	±150						21.0	(4.76%)
Anova	±200						22.9	(4.37%)
Xbar	Perfect					37.1	(2.70%)
Xbar	±50 				39.3	(2.54%)
Xbar	±100 				41.7	(2.40%)
Xbar	±150 				44.4	(2.25%)
Xbar	±200 				47.8	(2.09%)
©	Van	Brollini	2016
© Van Brollini 2016© Van Brollini 2016© Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016
© Van Brollini 2016© Van Brollini 2016
© Van Brollini 2016
© Van Brollini 2016
ANOVA		perfect	sample	placement
Anova	Perfect						33.1	(3.02%)
Anova	±50	 								35.7	(2.80%)
Anova	±100 								38.8	(2.58%)
Anova	±150 								42.2	(2.37%)
Anova	±200 								46.1	(2.17%)
Xbar	Perfect				18.4	(5.43%)
Xbar	±50											19.5	(5.13%)
Xbar	±100 		20.7	(4.83%)
Xbar	±150 		22.0	(4.55%)
Xbar	±200 		23.7	(4.22%)
Note:	%	numbers	shown	are	range	percent	error
EïŹ€ect	of	Only	Device	SpeciïŹca:on	
and	Sample	Placement	on	GRR
All	Device	Measurements	Perfect
±50% perfect sample placement
20
Next slide contains ANOVA
1. 10 perfect samples
2. 10 perfect samples ±50%
3. 10 perfect samples ±100%
4. 10 perfect samples ±150%
5. 10 perfect samples ±200%
21
4816 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47
0.1
0.035
0.04
0.045
0.05
0.055
0.06
0.065
0.07
0.075
0.08
0.085
0.09
0.095
# Instrument Errors
GRR
ANOVA		±50%	of	perfect	sample	placement
ANOVA		±100%	of	perfect	sample	placement
10	Perfectly	Measured	Device	Samples	for	QualiïŹca:on,	1	tester	0	oïŹ€set,	1	tester	+oïŹ€set,	1	tester	-	oïŹ€set
ANOVA		±150%	of	perfect	sample	placement
ANOVA		±200%	of	perfect	sample	placement
Anova	Perfect			16.5	(6.06%)
Anova	±50	 			17.8	(5.62%)
Anova	±100						19.3	(5.18%)
Anova	±150						21.0	(4.76%)
Anova	±200						22.9	(4.37%)
Xbar	Perfect					37.1	(2.70%)
Xbar	±50 				39.3	(2.54%)
Xbar	±100 				41.7	(2.40%)
Xbar	±150 				44.4	(2.25%)
Xbar	±200 				47.8	(2.09%)
©	Van	Brollini	2016
© Van Brollini 2016© Van Brollini 2016© Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016
© Van Brollini 2016© Van Brollini 2016
© Van Brollini 2016
© Van Brollini 2016
ANOVA		perfect	sample	placement
Anova	Perfect						33.1	(3.02%)
Anova	±50	 								35.7	(2.80%)
Anova	±100 								38.8	(2.58%)
Anova	±150 								42.2	(2.37%)
Anova	±200 								46.1	(2.17%)
Xbar	Perfect				18.4	(5.43%)
Xbar	±50											19.5	(5.13%)
Xbar	±100 		20.7	(4.83%)
Xbar	±150 		22.0	(4.55%)
Xbar	±200 		23.7	(4.22%)
Note:	%	numbers	shown	are	range	percent	error
EïŹ€ect	of	Only	Device	SpeciïŹca:on	
and	Sample	Placement	on	GRR
All	Device	Measurements	Perfect
22
Next 4 slides using
ANOVA show how to use
23
4816 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47
0.1
0.035
0.04
0.045
0.05
0.055
0.06
0.065
0.07
0.075
0.08
0.085
0.09
0.095
# Instrument Errors
GRR
ANOVA		±50%	of	perfect	sample	placement
ANOVA		±100%	of	perfect	sample	placement
10	Perfectly	Measured	Device	Samples	for	QualiïŹca:on,	1	tester	0	oïŹ€set,	1	tester	+oïŹ€set,	1	tester	-	oïŹ€set
ANOVA		±150%	of	perfect	sample	placement
ANOVA		±200%	of	perfect	sample	placement
Anova	Perfect			16.5	(6.06%)
Anova	±50	 			17.8	(5.62%)
Anova	±100						19.3	(5.18%)
Anova	±150						21.0	(4.76%)
Anova	±200						22.9	(4.37%)
Xbar	Perfect					37.1	(2.70%)
Xbar	±50 				39.3	(2.54%)
Xbar	±100 				41.7	(2.40%)
Xbar	±150 				44.4	(2.25%)
Xbar	±200 				47.8	(2.09%)
©	Van	Brollini	2016
© Van Brollini 2016© Van Brollini 2016© Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016
© Van Brollini 2016© Van Brollini 2016
© Van Brollini 2016
© Van Brollini 2016
ANOVA		perfect	sample	placement
Anova	Perfect						33.1	(3.02%)
Anova	±50	 								35.7	(2.80%)
Anova	±100 								38.8	(2.58%)
Anova	±150 								42.2	(2.37%)
Anova	±200 								46.1	(2.17%)
Xbar	Perfect				18.4	(5.43%)
Xbar	±50											19.5	(5.13%)
Xbar	±100 		20.7	(4.83%)
Xbar	±150 		22.0	(4.55%)
Xbar	±200 		23.7	(4.22%)
Note:	%	numbers	shown	are	range	percent	error
EïŹ€ect	of	Only	Device	SpeciïŹca:on	
and	Sample	Placement	on	GRR
All	Device	Measurements	Perfect
Have to choose something LESS than
10%. Need room for repeatability
24
4816 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47
0.1
0.035
0.04
0.045
0.05
0.055
0.06
0.065
0.07
0.075
0.08
0.085
0.09
0.095
# Instrument Errors
GRR
ANOVA		±50%	of	perfect	sample	placement
ANOVA		±100%	of	perfect	sample	placement
10	Perfectly	Measured	Device	Samples	for	QualiïŹca:on,	1	tester	0	oïŹ€set,	1	tester	+oïŹ€set,	1	tester	-	oïŹ€set
ANOVA		±150%	of	perfect	sample	placement
ANOVA		±200%	of	perfect	sample	placement
Anova	Perfect			16.5	(6.06%)
Anova	±50	 			17.8	(5.62%)
Anova	±100						19.3	(5.18%)
Anova	±150						21.0	(4.76%)
Anova	±200						22.9	(4.37%)
Xbar	Perfect					37.1	(2.70%)
Xbar	±50 				39.3	(2.54%)
Xbar	±100 				41.7	(2.40%)
Xbar	±150 				44.4	(2.25%)
Xbar	±200 				47.8	(2.09%)
©	Van	Brollini	2016
© Van Brollini 2016© Van Brollini 2016© Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016
© Van Brollini 2016© Van Brollini 2016
© Van Brollini 2016
© Van Brollini 2016
ANOVA		perfect	sample	placement
Anova	Perfect						33.1	(3.02%)
Anova	±50	 								35.7	(2.80%)
Anova	±100 								38.8	(2.58%)
Anova	±150 								42.2	(2.37%)
Anova	±200 								46.1	(2.17%)
Xbar	Perfect				18.4	(5.43%)
Xbar	±50											19.5	(5.13%)
Xbar	±100 		20.7	(4.83%)
Xbar	±150 		22.0	(4.55%)
Xbar	±200 		23.7	(4.22%)
Note:	%	numbers	shown	are	range	percent	error
EïŹ€ect	of	Only	Device	SpeciïŹca:on	
and	Sample	Placement	on	GRR
All	Device	Measurements	Perfect
Choose 24 Instrument Errors (4.17%)
25
4816 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47
0.1
0.035
0.04
0.045
0.05
0.055
0.06
0.065
0.07
0.075
0.08
0.085
0.09
0.095
# Instrument Errors
GRR
ANOVA		±50%	of	perfect	sample	placement
ANOVA		±100%	of	perfect	sample	placement
10	Perfectly	Measured	Device	Samples	for	QualiïŹca:on,	1	tester	0	oïŹ€set,	1	tester	+oïŹ€set,	1	tester	-	oïŹ€set
ANOVA		±150%	of	perfect	sample	placement
ANOVA		±200%	of	perfect	sample	placement
Anova	Perfect			16.5	(6.06%)
Anova	±50	 			17.8	(5.62%)
Anova	±100						19.3	(5.18%)
Anova	±150						21.0	(4.76%)
Anova	±200						22.9	(4.37%)
Xbar	Perfect					37.1	(2.70%)
Xbar	±50 				39.3	(2.54%)
Xbar	±100 				41.7	(2.40%)
Xbar	±150 				44.4	(2.25%)
Xbar	±200 				47.8	(2.09%)
©	Van	Brollini	2016
© Van Brollini 2016© Van Brollini 2016© Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016
© Van Brollini 2016© Van Brollini 2016
© Van Brollini 2016
© Van Brollini 2016
ANOVA		perfect	sample	placement
Anova	Perfect						33.1	(3.02%)
Anova	±50	 								35.7	(2.80%)
Anova	±100 								38.8	(2.58%)
Anova	±150 								42.2	(2.37%)
Anova	±200 								46.1	(2.17%)
Xbar	Perfect				18.4	(5.43%)
Xbar	±50											19.5	(5.13%)
Xbar	±100 		20.7	(4.83%)
Xbar	±150 		22.0	(4.55%)
Xbar	±200 		23.7	(4.22%)
Note:	%	numbers	shown	are	range	percent	error
EïŹ€ect	of	Only	Device	SpeciïŹca:on	
and	Sample	Placement	on	GRR
All	Device	Measurements	Perfect
If sample selection process is more or
less evenly distributed then error caused
by instrument specification will result in
≈ 6.6 - 7.9% GRR leaving plenty of room
for repeatability to achieve final goal of
10% GRR
26
Next slide contains Xbar&R
Xbar&R is more stringent than ANOVA
1. 10 perfect samples
2. 10 perfect samples ±50%
3. 10 perfect samples ±100%
4. 10 perfect samples ±150%
5. 10 perfect samples ±200%
27
4816 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47
0.1
0.035
0.04
0.045
0.05
0.055
0.06
0.065
0.07
0.075
0.08
0.085
0.09
0.095
# Instrument Errors
GRR
Xbar&R		±50%	of	perfect	sample	placement
Xbar&R		±100%	of	perfect	sample	placement
10	Perfectly	Measured	Device	Samples	for	QualiïŹca:on,	1	tester	0	oïŹ€set,	1	tester	+oïŹ€set,	1	tester	-	oïŹ€set
Xbar&R		±150%	of	perfect	sample	placement
Xbar&R		±200%	of	perfect	sample	placement
Xbar	Perfect					18.4	(5.43%)
Xbar	±50											19.5	(5.13%)
Xbar	±100 		20.7	(4.83%)
Xbar	±150 		22.0	(4.55%)
Xbar	±200 		23.7	(4.22%)
Xbar	Perfect					37.1	(2.70%)
Xbar	±50 				39.3	(2.54%)
Xbar	±100 				41.7	(2.40%)
Xbar	±150 				44.4	(2.25%)
Xbar	±200 				47.8	(2.09%)
©	Van	Brollini	2016
© Van Brollini 2016© Van Brollini 2016© Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016
© Van Brollini 2016© Van Brollini 2016
© Van Brollini 2016
© Van Brollini 2016
Xbar&R	perfect	sample	placement
Anova	Perfect			16.5	(6.06%)
Anova	±50	 				17.8	(5.62%)
Anova	±100							19.3	(5.18%)
Anova	±150							21.0	(4.76%)
Anova	±200							22.9	(4.37%)
Anova	Perfect						33.1	(3.02%)
Anova	±50	 								35.7	(2.80%)
Anova	±100 								38.8	(2.58%)
Anova	±150 								42.2	(2.37%)
Anova	±200 								46.1	(2.17%)
Note:	%	numbers	shown	are	range	percent	error
EïŹ€ect	of	Only	Device	SpeciïŹca:on	
and	Sample	Placement	on	GRR
All	Device	Measurements	Perfect
Note: Xbar&R is more stringent to pass.
Choose 25 instrument errors
28
4816 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47
0.1
0.035
0.04
0.045
0.05
0.055
0.06
0.065
0.07
0.075
0.08
0.085
0.09
0.095
# Instrument Errors
GRR
Xbar&R		±50%	of	perfect	sample	placement
Xbar&R		±100%	of	perfect	sample	placement
10	Perfectly	Measured	Device	Samples	for	QualiïŹca:on,	1	tester	0	oïŹ€set,	1	tester	+oïŹ€set,	1	tester	-	oïŹ€set
Xbar&R		±150%	of	perfect	sample	placement
Xbar&R		±200%	of	perfect	sample	placement
Xbar	Perfect					18.4	(5.43%)
Xbar	±50											19.5	(5.13%)
Xbar	±100 		20.7	(4.83%)
Xbar	±150 		22.0	(4.55%)
Xbar	±200 		23.7	(4.22%)
Xbar	Perfect					37.1	(2.70%)
Xbar	±50 				39.3	(2.54%)
Xbar	±100 				41.7	(2.40%)
Xbar	±150 				44.4	(2.25%)
Xbar	±200 				47.8	(2.09%)
©	Van	Brollini	2016
© Van Brollini 2016© Van Brollini 2016© Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016
© Van Brollini 2016© Van Brollini 2016
© Van Brollini 2016
© Van Brollini 2016
Xbar&R	perfect	sample	placement
Anova	Perfect			16.5	(6.06%)
Anova	±50	 				17.8	(5.62%)
Anova	±100							19.3	(5.18%)
Anova	±150							21.0	(4.76%)
Anova	±200							22.9	(4.37%)
Anova	Perfect						33.1	(3.02%)
Anova	±50	 								35.7	(2.80%)
Anova	±100 								38.8	(2.58%)
Anova	±150 								42.2	(2.37%)
Anova	±200 								46.1	(2.17%)
Note:	%	numbers	shown	are	range	percent	error
EïŹ€ect	of	Only	Device	SpeciïŹca:on	
and	Sample	Placement	on	GRR
All	Device	Measurements	Perfect
29
For 9% GRR the correct choices for allowable
instrument error to achieve the desired 10%
GRR with the 1% selection for repeatability are:
‱ Primary conclusion: Initial boundary condition is NOT SUFFICIENT to pass GRR at all!!
‱ Perfect parts and ONLY instrument error use 5.43% to 6.06% of the GRR 10% Goal
‱ Perfect parts and ONLY instrument error use 2.7% to 3.02% of the GRR 5% Goal
‱ Perfect parts, ONLY instrument and Sample Placement variation error use 4.22% to 5.62% of the GRR 10% Goal
‱ Perfect parts, ONLY instrument and Sample Placement variation error use 2.09% to 2.80% of the GRR 5% Goal
‱ Sample placement accounts for up to 1.3% of GRR 10% Goal for ANOVA
‱ Sample placement accounts for up to 0.63% of GRR 10% Goal for XBAR&R
Anova	Perfect			16.5	(6.06%)	
Anova	±50		 					17.8	(5.62%)	
Anova	±100						19.3	(5.18%)	
Anova	±150						21.0	(4.76%)	
Anova	±200						22.9	(4.37%)
Xbar	Perfect				18.4	(5.43%)	
Xbar	±50	 			19.5	(5.13%)	
Xbar	±100	 		20.7	(4.83%)	
Xbar	±150	 		22.0	(4.55%)	
Xbar	±200	 		23.7	(4.22%)
Xbar	Perfect					37.1	(2.70%)	
Xbar	±50	 				39.3	(2.54%)	
Xbar	±100	 				41.7	(2.40%)	
Xbar	±150	 				44.4	(2.25%)	
Xbar	±200	 				47.8	(2.09%)
Anova	Perfect						33.1	(3.02%)	
Anova	±50		 								35.7	(2.80%)	
Anova	±100	 								38.8	(2.58%)	
Anova	±150	 								42.2	(2.37%)	
Anova	±200	 								46.1	(2.17%)
9% GRR Target 4.5% GRR Target
30

Weitere Àhnliche Inhalte

Ähnlich wie GRR Made Easy

Identification of Outliersin Time Series Data via Simulation Study
Identification of Outliersin Time Series Data via Simulation StudyIdentification of Outliersin Time Series Data via Simulation Study
Identification of Outliersin Time Series Data via Simulation Studyiosrjce
 
Chap 9 A Process Capability & Spc Hk
Chap 9 A Process Capability & Spc HkChap 9 A Process Capability & Spc Hk
Chap 9 A Process Capability & Spc Hkajithsrc
 
A software fault localization technique based on program mutations
A software fault localization technique based on program mutationsA software fault localization technique based on program mutations
A software fault localization technique based on program mutationsTao He
 
Quality Control Chart
 Quality Control Chart Quality Control Chart
Quality Control ChartAshish Gupta
 
Multiple Sensors Soft-Failure Diagnosis Based on Kalman Filter
Multiple Sensors Soft-Failure Diagnosis Based on Kalman FilterMultiple Sensors Soft-Failure Diagnosis Based on Kalman Filter
Multiple Sensors Soft-Failure Diagnosis Based on Kalman Filtersipij
 
control Chart u np c p
control Chart u np c pcontrol Chart u np c p
control Chart u np c pblactigger401
 
DETECTION OF RELIABLE SOFTWARE USING SPRT ON TIME DOMAIN DATA
DETECTION OF RELIABLE SOFTWARE USING SPRT ON TIME DOMAIN DATADETECTION OF RELIABLE SOFTWARE USING SPRT ON TIME DOMAIN DATA
DETECTION OF RELIABLE SOFTWARE USING SPRT ON TIME DOMAIN DATAIJCSEA Journal
 
Summer 2015 Internship
Summer 2015 InternshipSummer 2015 Internship
Summer 2015 InternshipTaylor Martell
 
Error bounds for wireless localization in NLOS environments
Error bounds for wireless localization in NLOS environmentsError bounds for wireless localization in NLOS environments
Error bounds for wireless localization in NLOS environmentsIJECEIAES
 
Error bounds for wireless localization in NLOS environments
Error bounds for wireless localization in NLOS environmentsError bounds for wireless localization in NLOS environments
Error bounds for wireless localization in NLOS environmentsIJECEIAES
 
680report final
680report final680report final
680report finalRajesh M
 
Lecture Notes: EEEC6430312 Measurements And Instrumentation - Instrument Typ...
Lecture Notes:  EEEC6430312 Measurements And Instrumentation - Instrument Typ...Lecture Notes:  EEEC6430312 Measurements And Instrumentation - Instrument Typ...
Lecture Notes: EEEC6430312 Measurements And Instrumentation - Instrument Typ...AIMST University
 
Measurement systems analysis and a study of anova method
Measurement systems analysis and a study of anova methodMeasurement systems analysis and a study of anova method
Measurement systems analysis and a study of anova methodeSAT Journals
 
Ch04
Ch04Ch04
Ch04ajithsrc
 
Optimization and simulation of a New Low Density Parity-Check Decoder using t...
Optimization and simulation of a New Low Density Parity-Check Decoder using t...Optimization and simulation of a New Low Density Parity-Check Decoder using t...
Optimization and simulation of a New Low Density Parity-Check Decoder using t...IJERA Editor
 
Errors2
Errors2Errors2
Errors2sjsuchaya
 
2014 IIAG Imputation Assessments
2014 IIAG Imputation Assessments2014 IIAG Imputation Assessments
2014 IIAG Imputation AssessmentsDr Lendy Spires
 
Concept of Limit of Detection (LOD)
Concept of Limit of Detection (LOD)Concept of Limit of Detection (LOD)
Concept of Limit of Detection (LOD)GH Yeoh
 

Ähnlich wie GRR Made Easy (20)

Identification of Outliersin Time Series Data via Simulation Study
Identification of Outliersin Time Series Data via Simulation StudyIdentification of Outliersin Time Series Data via Simulation Study
Identification of Outliersin Time Series Data via Simulation Study
 
Chap 9 A Process Capability & Spc Hk
Chap 9 A Process Capability & Spc HkChap 9 A Process Capability & Spc Hk
Chap 9 A Process Capability & Spc Hk
 
TESCO Evaluation of Non-Normal Meter Data
TESCO Evaluation of Non-Normal Meter DataTESCO Evaluation of Non-Normal Meter Data
TESCO Evaluation of Non-Normal Meter Data
 
A software fault localization technique based on program mutations
A software fault localization technique based on program mutationsA software fault localization technique based on program mutations
A software fault localization technique based on program mutations
 
Quality Control Chart
 Quality Control Chart Quality Control Chart
Quality Control Chart
 
Multiple Sensors Soft-Failure Diagnosis Based on Kalman Filter
Multiple Sensors Soft-Failure Diagnosis Based on Kalman FilterMultiple Sensors Soft-Failure Diagnosis Based on Kalman Filter
Multiple Sensors Soft-Failure Diagnosis Based on Kalman Filter
 
control Chart u np c p
control Chart u np c pcontrol Chart u np c p
control Chart u np c p
 
DETECTION OF RELIABLE SOFTWARE USING SPRT ON TIME DOMAIN DATA
DETECTION OF RELIABLE SOFTWARE USING SPRT ON TIME DOMAIN DATADETECTION OF RELIABLE SOFTWARE USING SPRT ON TIME DOMAIN DATA
DETECTION OF RELIABLE SOFTWARE USING SPRT ON TIME DOMAIN DATA
 
Summer 2015 Internship
Summer 2015 InternshipSummer 2015 Internship
Summer 2015 Internship
 
Error bounds for wireless localization in NLOS environments
Error bounds for wireless localization in NLOS environmentsError bounds for wireless localization in NLOS environments
Error bounds for wireless localization in NLOS environments
 
Error bounds for wireless localization in NLOS environments
Error bounds for wireless localization in NLOS environmentsError bounds for wireless localization in NLOS environments
Error bounds for wireless localization in NLOS environments
 
680report final
680report final680report final
680report final
 
Skripta vezbe gradivo za 2 test
Skripta vezbe   gradivo za 2 testSkripta vezbe   gradivo za 2 test
Skripta vezbe gradivo za 2 test
 
Lecture Notes: EEEC6430312 Measurements And Instrumentation - Instrument Typ...
Lecture Notes:  EEEC6430312 Measurements And Instrumentation - Instrument Typ...Lecture Notes:  EEEC6430312 Measurements And Instrumentation - Instrument Typ...
Lecture Notes: EEEC6430312 Measurements And Instrumentation - Instrument Typ...
 
Measurement systems analysis and a study of anova method
Measurement systems analysis and a study of anova methodMeasurement systems analysis and a study of anova method
Measurement systems analysis and a study of anova method
 
Ch04
Ch04Ch04
Ch04
 
Optimization and simulation of a New Low Density Parity-Check Decoder using t...
Optimization and simulation of a New Low Density Parity-Check Decoder using t...Optimization and simulation of a New Low Density Parity-Check Decoder using t...
Optimization and simulation of a New Low Density Parity-Check Decoder using t...
 
Errors2
Errors2Errors2
Errors2
 
2014 IIAG Imputation Assessments
2014 IIAG Imputation Assessments2014 IIAG Imputation Assessments
2014 IIAG Imputation Assessments
 
Concept of Limit of Detection (LOD)
Concept of Limit of Detection (LOD)Concept of Limit of Detection (LOD)
Concept of Limit of Detection (LOD)
 

KĂŒrzlich hochgeladen

🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...
🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...
🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...9953056974 Low Rate Call Girls In Saket, Delhi NCR
 
Work Experience-Dalton Park.pptxfvvvvvvv
Work Experience-Dalton Park.pptxfvvvvvvvWork Experience-Dalton Park.pptxfvvvvvvv
Work Experience-Dalton Park.pptxfvvvvvvvLewisJB
 
Solving The Right Triangles PowerPoint 2.ppt
Solving The Right Triangles PowerPoint 2.pptSolving The Right Triangles PowerPoint 2.ppt
Solving The Right Triangles PowerPoint 2.pptJasonTagapanGulla
 
Application of Residue Theorem to evaluate real integrations.pptx
Application of Residue Theorem to evaluate real integrations.pptxApplication of Residue Theorem to evaluate real integrations.pptx
Application of Residue Theorem to evaluate real integrations.pptx959SahilShah
 
UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)
UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)
UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)Dr SOUNDIRARAJ N
 
Correctly Loading Incremental Data at Scale
Correctly Loading Incremental Data at ScaleCorrectly Loading Incremental Data at Scale
Correctly Loading Incremental Data at ScaleAlluxio, Inc.
 
computer application and construction management
computer application and construction managementcomputer application and construction management
computer application and construction managementMariconPadriquez1
 
Piping Basic stress analysis by engineering
Piping Basic stress analysis by engineeringPiping Basic stress analysis by engineering
Piping Basic stress analysis by engineeringJuanCarlosMorales19600
 
Class 1 | NFPA 72 | Overview Fire Alarm System
Class 1 | NFPA 72 | Overview Fire Alarm SystemClass 1 | NFPA 72 | Overview Fire Alarm System
Class 1 | NFPA 72 | Overview Fire Alarm Systemirfanmechengr
 
Introduction to Machine Learning Unit-3 for II MECH
Introduction to Machine Learning Unit-3 for II MECHIntroduction to Machine Learning Unit-3 for II MECH
Introduction to Machine Learning Unit-3 for II MECHC Sai Kiran
 
Arduino_CSE ece ppt for working and principal of arduino.ppt
Arduino_CSE ece ppt for working and principal of arduino.pptArduino_CSE ece ppt for working and principal of arduino.ppt
Arduino_CSE ece ppt for working and principal of arduino.pptSAURABHKUMAR892774
 
welding defects observed during the welding
welding defects observed during the weldingwelding defects observed during the welding
welding defects observed during the weldingMuhammadUzairLiaqat
 
Electronically Controlled suspensions system .pdf
Electronically Controlled suspensions system .pdfElectronically Controlled suspensions system .pdf
Electronically Controlled suspensions system .pdfme23b1001
 
8251 universal synchronous asynchronous receiver transmitter
8251 universal synchronous asynchronous receiver transmitter8251 universal synchronous asynchronous receiver transmitter
8251 universal synchronous asynchronous receiver transmitterShivangiSharma879191
 
Transport layer issues and challenges - Guide
Transport layer issues and challenges - GuideTransport layer issues and challenges - Guide
Transport layer issues and challenges - GuideGOPINATHS437943
 
Instrumentation, measurement and control of bio process parameters ( Temperat...
Instrumentation, measurement and control of bio process parameters ( Temperat...Instrumentation, measurement and control of bio process parameters ( Temperat...
Instrumentation, measurement and control of bio process parameters ( Temperat...121011101441
 
Call Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile serviceCall Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile servicerehmti665
 
Software and Systems Engineering Standards: Verification and Validation of Sy...
Software and Systems Engineering Standards: Verification and Validation of Sy...Software and Systems Engineering Standards: Verification and Validation of Sy...
Software and Systems Engineering Standards: Verification and Validation of Sy...VICTOR MAESTRE RAMIREZ
 
Call Girls Narol 7397865700 Independent Call Girls
Call Girls Narol 7397865700 Independent Call GirlsCall Girls Narol 7397865700 Independent Call Girls
Call Girls Narol 7397865700 Independent Call Girlsssuser7cb4ff
 

KĂŒrzlich hochgeladen (20)

Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptxExploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
 
🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...
🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...
🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...
 
Work Experience-Dalton Park.pptxfvvvvvvv
Work Experience-Dalton Park.pptxfvvvvvvvWork Experience-Dalton Park.pptxfvvvvvvv
Work Experience-Dalton Park.pptxfvvvvvvv
 
Solving The Right Triangles PowerPoint 2.ppt
Solving The Right Triangles PowerPoint 2.pptSolving The Right Triangles PowerPoint 2.ppt
Solving The Right Triangles PowerPoint 2.ppt
 
Application of Residue Theorem to evaluate real integrations.pptx
Application of Residue Theorem to evaluate real integrations.pptxApplication of Residue Theorem to evaluate real integrations.pptx
Application of Residue Theorem to evaluate real integrations.pptx
 
UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)
UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)
UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)
 
Correctly Loading Incremental Data at Scale
Correctly Loading Incremental Data at ScaleCorrectly Loading Incremental Data at Scale
Correctly Loading Incremental Data at Scale
 
computer application and construction management
computer application and construction managementcomputer application and construction management
computer application and construction management
 
Piping Basic stress analysis by engineering
Piping Basic stress analysis by engineeringPiping Basic stress analysis by engineering
Piping Basic stress analysis by engineering
 
Class 1 | NFPA 72 | Overview Fire Alarm System
Class 1 | NFPA 72 | Overview Fire Alarm SystemClass 1 | NFPA 72 | Overview Fire Alarm System
Class 1 | NFPA 72 | Overview Fire Alarm System
 
Introduction to Machine Learning Unit-3 for II MECH
Introduction to Machine Learning Unit-3 for II MECHIntroduction to Machine Learning Unit-3 for II MECH
Introduction to Machine Learning Unit-3 for II MECH
 
Arduino_CSE ece ppt for working and principal of arduino.ppt
Arduino_CSE ece ppt for working and principal of arduino.pptArduino_CSE ece ppt for working and principal of arduino.ppt
Arduino_CSE ece ppt for working and principal of arduino.ppt
 
welding defects observed during the welding
welding defects observed during the weldingwelding defects observed during the welding
welding defects observed during the welding
 
Electronically Controlled suspensions system .pdf
Electronically Controlled suspensions system .pdfElectronically Controlled suspensions system .pdf
Electronically Controlled suspensions system .pdf
 
8251 universal synchronous asynchronous receiver transmitter
8251 universal synchronous asynchronous receiver transmitter8251 universal synchronous asynchronous receiver transmitter
8251 universal synchronous asynchronous receiver transmitter
 
Transport layer issues and challenges - Guide
Transport layer issues and challenges - GuideTransport layer issues and challenges - Guide
Transport layer issues and challenges - Guide
 
Instrumentation, measurement and control of bio process parameters ( Temperat...
Instrumentation, measurement and control of bio process parameters ( Temperat...Instrumentation, measurement and control of bio process parameters ( Temperat...
Instrumentation, measurement and control of bio process parameters ( Temperat...
 
Call Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile serviceCall Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile service
 
Software and Systems Engineering Standards: Verification and Validation of Sy...
Software and Systems Engineering Standards: Verification and Validation of Sy...Software and Systems Engineering Standards: Verification and Validation of Sy...
Software and Systems Engineering Standards: Verification and Validation of Sy...
 
Call Girls Narol 7397865700 Independent Call Girls
Call Girls Narol 7397865700 Independent Call GirlsCall Girls Narol 7397865700 Independent Call Girls
Call Girls Narol 7397865700 Independent Call Girls
 

GRR Made Easy

  • 2. When looking at GRR and MSA all the information is “EXACTLY CORRECT” There is however a LOT of detailed information to understand 2
  • 3. Is there a way to make it simpler to understand and implement? Especially for Test Engineers? 3
  • 5. There are 2 boundary conditions that a test engineer can understand that are only “Sometimes” shown 5
  • 6. This is the ïŹrst!! 6
  • 7. This is the second!! 7
  • 8. Using freely available spreadsheet off of the Internet for Excel analysis of GRR a model was developed for analysis using these 2 boundary conditions 8
  • 9. Test Engineers have been told to follow 2 boundary conditions 1. The error in the measurement has to be 10 times less than the measured value. 2. The error between the USL and LSL has to be 10% of the difference between USL and LSL X Instrument Error 0 2X Instrument Error LSL USL ΔCV ΔCV - Boundary Condition # 2 LCV - Boundary Condition # 1 Test Engineering Boundary Conditions LCV Where USL = Upper SpeciïŹca>on Limit LSL = Lower SpeciïŹca>on Limit LCV = Lowest Capable Value ΔCV = USL - LSL = DiïŹ€eren>al Capability The variable, the RED “X”, needs to be understood 9
  • 10. GRR Tools Critical Evaluation with underlying Test Engineering Boundary Conditions The model created is using an OE (offset error of ±4mV with GE (gain error = 0%). The initial evaluation used 10 samples, 3 tests, 3 Testers. One tester had 0 offset, another -4mV, and the third +4mV. Repeatability was 0.0001 to insure that just equipment reproducibility is what is being examined. Manage Instrument Error to achieve a passing qualification on devices for GRR: ANOVA or XBAR&R The experiment objective is to understand instrument specification and how it affects GRR ALONE 10
  • 11. GRR Tools Manage Instrument Error to achieve a passing qualification on devices for GRR: ANOVA or XBAR&R GRR Calculation # Samples # Testers Methods # of Measurements USL & LSL Objective: Understand GRR from Instrument Error Perspective 2-10 samples Sample location: determines GRR Random Selection Intelligent selection Plus Offset - one tester Zero Offset - one tester Minus Offset - one tester ANOVA Xbar&R One value - Ideal Second value: + 0.0001 Third value: - 0.0001 Note: sample location is NOT usually looked at; and it turns out to be very important 11
  • 12. How do these affect GRR Results ‱ There are essentially 2 methods for doing GRR ‱ ANOVA ‱ Xbar&R ‱ Spreadsheets for each can be found on the internet free of charge. Some with just one of the methods or the other, and at least one with both methods in the spreadsheet. ‱ Each method was confirmed to give the same results with identical data when comparing ANOVA to ANOVA and Xbar&R to Xbar&R 12
  • 13. The question is: Are the 2 boundary conditions listed on slides 6 and 7 necesary and sufficient to insure that the goal of GRR 10% passes? 1. Random samples within USL and LSL 2. 10 perfect evenly distributed samples within USL and LSL 3. 10 select samples to cause worst case within center ±50% of perfect samples placement 4. 10 select samples to cause worst case within center ±100% of perfect samples placement 5. 10 select samples to cause worst case within center ±150% of perfect samples placement 6. 10 select samples to cause worst case within center ±200% of perfect samples placement 6 ways to look at: # Instrument errors range from 10X to shown on graphs 13
  • 14. Necessary Equations The diagram on Slide 2 shows 11X for instrument errors to take into account the need for repeatability. This works out to a GRR of approximately 9% for just instrument errors, which is the objective of the exercise USL = LSL 1+10GE( )+ 20OE 1−10GE USL − LSL = ΔCV ΔCV = 10 GE i(USL + LSL)+ 2 iOE( ) ΔCV = 20 iOE +10 GE i(USL + LSL)( ) LSL = +LCV = OE 0.1− GE +LCV = OE 0.1− GE The model created us using an OE (offset error of ±4mV with GE (gain error = 0%). The initial evaluation used 10 samples, 3 tests, 3 Testers. One tester had 0 offset, another -4mV, and the third +4mV. Repeatability was 0.0001 to insure that just equipment reproducibility is what is being examined. 14
  • 15. 0 LSL USL ±100% of perfect samples placement ±50% of perfect samples placement 0 LSL USL Random perfect samples placement 0 LSL USL Perfect Placement ±100 % Placement - Worst Case 0 LSL USL 0 LSL USL Perfect Placement ±50 % Placement - Worst Case 15
  • 16. # Instrument errors range from 10X to what is shown on graph, 10 measurement points perfectly evenly distributed 0 LSL USL 10X 11X 0 LSL USL 12X 0 LSL USL 13X 0 LSL USL 14X 0 LSL USL 15X 0 LSL USL 16X 0 LSL USL 24X 0 LSL USL 23X 0 LSL USL 22X 0 LSL USL 21X 0 LSL USL 20X 0 LSL USL 19X 0 LSL USL 18X 0 LSL USL 17X 0 LSL USL Red lines represent perfect evenly spaced samples 16
  • 17. Random selection of measurement for 10 DUTs with zero offset, + offset and - offset. 1.5200 10 device randomly selected measurement values using Monte Carlo simulations for Xbar&R from Statistical Solutions, Tolerance and GRR results. 2.1010 10 device randomly selected measurement values using Monte Carlo simulations using the ANOVA method from www.dmaictools.com, Tolerance and GRR results. Lower Specification limits started at 40mV (10X), Upper Specification limit start at 120mV (10X), ranging to 40X instrument errors. This means that the difference between USL-LSL, ranged from 10X to 40X. Simulations were done for a process distribution width of 5.15 (encloses the central 99% of the process distribution). 0 LSL USL Please note that the number of instrument errors is the inverse of the instrument error 17
  • 18. 409 12 14 16 18 20 22 24 26 28 30 32 34 36 38 1 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 (# X) Number of Instrument Errors Percentage(%)over10% GRR% Xbar&R Xbar&R Monte Carlo Simula3on in Excel. 5200 simula3ons of 10 devices. Random devices within limits Using Gage R&R (XBar&R Motorola Version) from Sta;s;cal Solu;ons Please note that 11X and greater is acceptable. However neither the Xbar&R Method nor ANOVA correctly calculate GRR at the corners of instrument limits GRR% ANOVA ANOVA Monte Carlo Simula3on in Excel. 1010 simula3ons of 10 devices. Random devices within limits ± 4mV offset Error LSL starts at +40mV (10X) USL starts at +120mV USL-LSL starts at 80mV (10X) Anova - www.dmaictools.com/measure/grr 30.0 Percentage of simulations (Monte Carlo) that showed the instances greater than GRR 10% for each Instrument X. To read the graph, 20 on the X axis is the number of instrument errors. For GRR ANOVA, approximately 0.5% of the simulation results showed greater than a GRR of 10%. For GRR Xbar&R 1% of the simulation results showed greater than a GRR of 10%. 0 LSL USL Random samples placement 18
  • 19. 4. 10 select perfect samples to cause worst case within center ±50% of perfect samples placement NOTE: solid line on graph of next slide 1. ANOVA 5.15 Standard deviations 2.Xbar&R 5.15 Standard deviations Lower Specification limits started at 40mV (10X), Upper Specification limit start at 120mV (10X), ranging to 40X instrument errors. This means that the difference between USL-LSL, ranged from 10X and following. Calculations were done for a process distribution width of 5.15 (encloses the central 99% of the process distribution) at worst case values.. 0 LSL USL ±50% of perfect samples placement Please note that the number of instrument errors is the inverse of the instrument error 19
  • 20. 4816 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 0.1 0.035 0.04 0.045 0.05 0.055 0.06 0.065 0.07 0.075 0.08 0.085 0.09 0.095 # Instrument Errors GRR ANOVA ±50% of perfect sample placement ANOVA ±100% of perfect sample placement 10 Perfectly Measured Device Samples for QualiïŹca:on, 1 tester 0 oïŹ€set, 1 tester +oïŹ€set, 1 tester - oïŹ€set ANOVA ±150% of perfect sample placement ANOVA ±200% of perfect sample placement Anova Perfect 16.5 (6.06%) Anova ±50 17.8 (5.62%) Anova ±100 19.3 (5.18%) Anova ±150 21.0 (4.76%) Anova ±200 22.9 (4.37%) Xbar Perfect 37.1 (2.70%) Xbar ±50 39.3 (2.54%) Xbar ±100 41.7 (2.40%) Xbar ±150 44.4 (2.25%) Xbar ±200 47.8 (2.09%) © Van Brollini 2016 © Van Brollini 2016© Van Brollini 2016© Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016© Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016 ANOVA perfect sample placement Anova Perfect 33.1 (3.02%) Anova ±50 35.7 (2.80%) Anova ±100 38.8 (2.58%) Anova ±150 42.2 (2.37%) Anova ±200 46.1 (2.17%) Xbar Perfect 18.4 (5.43%) Xbar ±50 19.5 (5.13%) Xbar ±100 20.7 (4.83%) Xbar ±150 22.0 (4.55%) Xbar ±200 23.7 (4.22%) Note: % numbers shown are range percent error EïŹ€ect of Only Device SpeciïŹca:on and Sample Placement on GRR All Device Measurements Perfect ±50% perfect sample placement 20
  • 21. Next slide contains ANOVA 1. 10 perfect samples 2. 10 perfect samples ±50% 3. 10 perfect samples ±100% 4. 10 perfect samples ±150% 5. 10 perfect samples ±200% 21
  • 22. 4816 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 0.1 0.035 0.04 0.045 0.05 0.055 0.06 0.065 0.07 0.075 0.08 0.085 0.09 0.095 # Instrument Errors GRR ANOVA ±50% of perfect sample placement ANOVA ±100% of perfect sample placement 10 Perfectly Measured Device Samples for QualiïŹca:on, 1 tester 0 oïŹ€set, 1 tester +oïŹ€set, 1 tester - oïŹ€set ANOVA ±150% of perfect sample placement ANOVA ±200% of perfect sample placement Anova Perfect 16.5 (6.06%) Anova ±50 17.8 (5.62%) Anova ±100 19.3 (5.18%) Anova ±150 21.0 (4.76%) Anova ±200 22.9 (4.37%) Xbar Perfect 37.1 (2.70%) Xbar ±50 39.3 (2.54%) Xbar ±100 41.7 (2.40%) Xbar ±150 44.4 (2.25%) Xbar ±200 47.8 (2.09%) © Van Brollini 2016 © Van Brollini 2016© Van Brollini 2016© Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016© Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016 ANOVA perfect sample placement Anova Perfect 33.1 (3.02%) Anova ±50 35.7 (2.80%) Anova ±100 38.8 (2.58%) Anova ±150 42.2 (2.37%) Anova ±200 46.1 (2.17%) Xbar Perfect 18.4 (5.43%) Xbar ±50 19.5 (5.13%) Xbar ±100 20.7 (4.83%) Xbar ±150 22.0 (4.55%) Xbar ±200 23.7 (4.22%) Note: % numbers shown are range percent error EïŹ€ect of Only Device SpeciïŹca:on and Sample Placement on GRR All Device Measurements Perfect 22
  • 23. Next 4 slides using ANOVA show how to use 23
  • 24. 4816 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 0.1 0.035 0.04 0.045 0.05 0.055 0.06 0.065 0.07 0.075 0.08 0.085 0.09 0.095 # Instrument Errors GRR ANOVA ±50% of perfect sample placement ANOVA ±100% of perfect sample placement 10 Perfectly Measured Device Samples for QualiïŹca:on, 1 tester 0 oïŹ€set, 1 tester +oïŹ€set, 1 tester - oïŹ€set ANOVA ±150% of perfect sample placement ANOVA ±200% of perfect sample placement Anova Perfect 16.5 (6.06%) Anova ±50 17.8 (5.62%) Anova ±100 19.3 (5.18%) Anova ±150 21.0 (4.76%) Anova ±200 22.9 (4.37%) Xbar Perfect 37.1 (2.70%) Xbar ±50 39.3 (2.54%) Xbar ±100 41.7 (2.40%) Xbar ±150 44.4 (2.25%) Xbar ±200 47.8 (2.09%) © Van Brollini 2016 © Van Brollini 2016© Van Brollini 2016© Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016© Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016 ANOVA perfect sample placement Anova Perfect 33.1 (3.02%) Anova ±50 35.7 (2.80%) Anova ±100 38.8 (2.58%) Anova ±150 42.2 (2.37%) Anova ±200 46.1 (2.17%) Xbar Perfect 18.4 (5.43%) Xbar ±50 19.5 (5.13%) Xbar ±100 20.7 (4.83%) Xbar ±150 22.0 (4.55%) Xbar ±200 23.7 (4.22%) Note: % numbers shown are range percent error EïŹ€ect of Only Device SpeciïŹca:on and Sample Placement on GRR All Device Measurements Perfect Have to choose something LESS than 10%. Need room for repeatability 24
  • 25. 4816 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 0.1 0.035 0.04 0.045 0.05 0.055 0.06 0.065 0.07 0.075 0.08 0.085 0.09 0.095 # Instrument Errors GRR ANOVA ±50% of perfect sample placement ANOVA ±100% of perfect sample placement 10 Perfectly Measured Device Samples for QualiïŹca:on, 1 tester 0 oïŹ€set, 1 tester +oïŹ€set, 1 tester - oïŹ€set ANOVA ±150% of perfect sample placement ANOVA ±200% of perfect sample placement Anova Perfect 16.5 (6.06%) Anova ±50 17.8 (5.62%) Anova ±100 19.3 (5.18%) Anova ±150 21.0 (4.76%) Anova ±200 22.9 (4.37%) Xbar Perfect 37.1 (2.70%) Xbar ±50 39.3 (2.54%) Xbar ±100 41.7 (2.40%) Xbar ±150 44.4 (2.25%) Xbar ±200 47.8 (2.09%) © Van Brollini 2016 © Van Brollini 2016© Van Brollini 2016© Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016© Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016 ANOVA perfect sample placement Anova Perfect 33.1 (3.02%) Anova ±50 35.7 (2.80%) Anova ±100 38.8 (2.58%) Anova ±150 42.2 (2.37%) Anova ±200 46.1 (2.17%) Xbar Perfect 18.4 (5.43%) Xbar ±50 19.5 (5.13%) Xbar ±100 20.7 (4.83%) Xbar ±150 22.0 (4.55%) Xbar ±200 23.7 (4.22%) Note: % numbers shown are range percent error EïŹ€ect of Only Device SpeciïŹca:on and Sample Placement on GRR All Device Measurements Perfect Choose 24 Instrument Errors (4.17%) 25
  • 26. 4816 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 0.1 0.035 0.04 0.045 0.05 0.055 0.06 0.065 0.07 0.075 0.08 0.085 0.09 0.095 # Instrument Errors GRR ANOVA ±50% of perfect sample placement ANOVA ±100% of perfect sample placement 10 Perfectly Measured Device Samples for QualiïŹca:on, 1 tester 0 oïŹ€set, 1 tester +oïŹ€set, 1 tester - oïŹ€set ANOVA ±150% of perfect sample placement ANOVA ±200% of perfect sample placement Anova Perfect 16.5 (6.06%) Anova ±50 17.8 (5.62%) Anova ±100 19.3 (5.18%) Anova ±150 21.0 (4.76%) Anova ±200 22.9 (4.37%) Xbar Perfect 37.1 (2.70%) Xbar ±50 39.3 (2.54%) Xbar ±100 41.7 (2.40%) Xbar ±150 44.4 (2.25%) Xbar ±200 47.8 (2.09%) © Van Brollini 2016 © Van Brollini 2016© Van Brollini 2016© Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016© Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016 ANOVA perfect sample placement Anova Perfect 33.1 (3.02%) Anova ±50 35.7 (2.80%) Anova ±100 38.8 (2.58%) Anova ±150 42.2 (2.37%) Anova ±200 46.1 (2.17%) Xbar Perfect 18.4 (5.43%) Xbar ±50 19.5 (5.13%) Xbar ±100 20.7 (4.83%) Xbar ±150 22.0 (4.55%) Xbar ±200 23.7 (4.22%) Note: % numbers shown are range percent error EïŹ€ect of Only Device SpeciïŹca:on and Sample Placement on GRR All Device Measurements Perfect If sample selection process is more or less evenly distributed then error caused by instrument specification will result in ≈ 6.6 - 7.9% GRR leaving plenty of room for repeatability to achieve final goal of 10% GRR 26
  • 27. Next slide contains Xbar&R Xbar&R is more stringent than ANOVA 1. 10 perfect samples 2. 10 perfect samples ±50% 3. 10 perfect samples ±100% 4. 10 perfect samples ±150% 5. 10 perfect samples ±200% 27
  • 28. 4816 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 0.1 0.035 0.04 0.045 0.05 0.055 0.06 0.065 0.07 0.075 0.08 0.085 0.09 0.095 # Instrument Errors GRR Xbar&R ±50% of perfect sample placement Xbar&R ±100% of perfect sample placement 10 Perfectly Measured Device Samples for QualiïŹca:on, 1 tester 0 oïŹ€set, 1 tester +oïŹ€set, 1 tester - oïŹ€set Xbar&R ±150% of perfect sample placement Xbar&R ±200% of perfect sample placement Xbar Perfect 18.4 (5.43%) Xbar ±50 19.5 (5.13%) Xbar ±100 20.7 (4.83%) Xbar ±150 22.0 (4.55%) Xbar ±200 23.7 (4.22%) Xbar Perfect 37.1 (2.70%) Xbar ±50 39.3 (2.54%) Xbar ±100 41.7 (2.40%) Xbar ±150 44.4 (2.25%) Xbar ±200 47.8 (2.09%) © Van Brollini 2016 © Van Brollini 2016© Van Brollini 2016© Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016© Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016 Xbar&R perfect sample placement Anova Perfect 16.5 (6.06%) Anova ±50 17.8 (5.62%) Anova ±100 19.3 (5.18%) Anova ±150 21.0 (4.76%) Anova ±200 22.9 (4.37%) Anova Perfect 33.1 (3.02%) Anova ±50 35.7 (2.80%) Anova ±100 38.8 (2.58%) Anova ±150 42.2 (2.37%) Anova ±200 46.1 (2.17%) Note: % numbers shown are range percent error EïŹ€ect of Only Device SpeciïŹca:on and Sample Placement on GRR All Device Measurements Perfect Note: Xbar&R is more stringent to pass. Choose 25 instrument errors 28
  • 29. 4816 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 0.1 0.035 0.04 0.045 0.05 0.055 0.06 0.065 0.07 0.075 0.08 0.085 0.09 0.095 # Instrument Errors GRR Xbar&R ±50% of perfect sample placement Xbar&R ±100% of perfect sample placement 10 Perfectly Measured Device Samples for QualiïŹca:on, 1 tester 0 oïŹ€set, 1 tester +oïŹ€set, 1 tester - oïŹ€set Xbar&R ±150% of perfect sample placement Xbar&R ±200% of perfect sample placement Xbar Perfect 18.4 (5.43%) Xbar ±50 19.5 (5.13%) Xbar ±100 20.7 (4.83%) Xbar ±150 22.0 (4.55%) Xbar ±200 23.7 (4.22%) Xbar Perfect 37.1 (2.70%) Xbar ±50 39.3 (2.54%) Xbar ±100 41.7 (2.40%) Xbar ±150 44.4 (2.25%) Xbar ±200 47.8 (2.09%) © Van Brollini 2016 © Van Brollini 2016© Van Brollini 2016© Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016© Van Brollini 2016 © Van Brollini 2016 © Van Brollini 2016 Xbar&R perfect sample placement Anova Perfect 16.5 (6.06%) Anova ±50 17.8 (5.62%) Anova ±100 19.3 (5.18%) Anova ±150 21.0 (4.76%) Anova ±200 22.9 (4.37%) Anova Perfect 33.1 (3.02%) Anova ±50 35.7 (2.80%) Anova ±100 38.8 (2.58%) Anova ±150 42.2 (2.37%) Anova ±200 46.1 (2.17%) Note: % numbers shown are range percent error EïŹ€ect of Only Device SpeciïŹca:on and Sample Placement on GRR All Device Measurements Perfect 29
  • 30. For 9% GRR the correct choices for allowable instrument error to achieve the desired 10% GRR with the 1% selection for repeatability are: ‱ Primary conclusion: Initial boundary condition is NOT SUFFICIENT to pass GRR at all!! ‱ Perfect parts and ONLY instrument error use 5.43% to 6.06% of the GRR 10% Goal ‱ Perfect parts and ONLY instrument error use 2.7% to 3.02% of the GRR 5% Goal ‱ Perfect parts, ONLY instrument and Sample Placement variation error use 4.22% to 5.62% of the GRR 10% Goal ‱ Perfect parts, ONLY instrument and Sample Placement variation error use 2.09% to 2.80% of the GRR 5% Goal ‱ Sample placement accounts for up to 1.3% of GRR 10% Goal for ANOVA ‱ Sample placement accounts for up to 0.63% of GRR 10% Goal for XBAR&R Anova Perfect 16.5 (6.06%) Anova ±50 17.8 (5.62%) Anova ±100 19.3 (5.18%) Anova ±150 21.0 (4.76%) Anova ±200 22.9 (4.37%) Xbar Perfect 18.4 (5.43%) Xbar ±50 19.5 (5.13%) Xbar ±100 20.7 (4.83%) Xbar ±150 22.0 (4.55%) Xbar ±200 23.7 (4.22%) Xbar Perfect 37.1 (2.70%) Xbar ±50 39.3 (2.54%) Xbar ±100 41.7 (2.40%) Xbar ±150 44.4 (2.25%) Xbar ±200 47.8 (2.09%) Anova Perfect 33.1 (3.02%) Anova ±50 35.7 (2.80%) Anova ±100 38.8 (2.58%) Anova ±150 42.2 (2.37%) Anova ±200 46.1 (2.17%) 9% GRR Target 4.5% GRR Target 30