The new frontier in
    Microwave thermoablation




Technical Presentation
Indice
Index

• Principi di termoablazione a microonde
• Principles of microwave thermoablation

• Microonde vs RF
• MWs vs RF currents

• Recenti sviluppi nei sistemi per termoablazione a microonde:
  AMICA
• Recent enhancements in MW ablation systems:
 AMICA
Termoablazione
    Thermoablation
    Necrosi coagulativa delle cellule indotta da un aumento locale della temperatura
    Coagulative necrosis of cells induced by a local temperature increase




    Morte cellulare pressochè istantanea per T >60°C.
    A temperature più basse (dell’ordine di 50°C) si ottiene lo stesso effetto per esposizioni prolungate al
    calore (dell’ordine di almeno 5/10 minuti).

4
    Almost instantaneous cell death for T >60°C.
    At lower temperatures (about 50°C) the same effect is obtained for longer exposures
    (about 5/10 minutes).
Requisiti fondamentali di un sistema termoablativo
Basic requirements for a thermoablative system
Controllo sulla figura di necrosi
       • Prevedibilità del risultato terapeutico (forma e dimensioni del volume ablato) in condizioni
       operative assegnate (tipo di tessuto-bersaglio, tipo e dose di energia somministrata, modalità
       e durata dell’erogazione di energia)
       • Prestazioni coagulative minimali e massimali (modellabilità della termolesione)
       • Confinamento (sfericità, preservazione di tessuti adiacenti al bersaglio, assenza di interferenze
       con eventuali dispositivi medici impiantati o protesi)
Velocità di riscaldamento
       • Breve durata del trattamento
       • Bassa sensibilità agli effetti di sottrazione del calore (ad es. heat sinking dovuto al circolo san-
       guigno o ai dotti biliari)
Sicurezza
       • Minimo grado di invasività, basso tasso di complicanze intra- e post-operatorie, compatibilità
       EMC e sicurezza elettrica, disponibilità di allarmi e protezioni hardware e software
Facilità d’uso
       • Curva d’apprendimento breve, compatibilità piena con le principali tecniche di imaging, accesso
       facile alla lesione, controllabilità in tempo reale dell’andamento del trattamento
Costi accessibili
                                                                                                               5


• Control over coagulation pattern
       • Predictability of treatment outcome (shape and size of coagulation pattern) for a given set of
       working parameters (type of target tissue, type and quantity of energy administered, modality and
       duration of energy deposition)
       • Capability of obtaining both big or small lesions (max/min lesion diameter/length)
       • Confinement (sphericity, sparing of healthy tissues in the neighbourhoods of target tissues, no in-
       terference with implanted medical devices)
• Heating velocity
       • Short treatment duration
       • Low sensitivity to heat sinking effects (blood vessels, biliary ducts)
• Safety of use
       • Low degree of invasiveness, low rate of intra- and post-operative complications, EMC compliance,
       availability of hardware and software alarms and protections
• Ease of use
       • Fast learning curve, full compatibility with available imaging technologies, easy access to lesion,
       real time control of treatment progress
• Cost effectiveness
Scelta dello strumento ablativo più appropriato
    Choice of most appropriate ablative tool



                                                                             ͖
                                        Correnti a radiofrequenza (RFs)
                                                      Microonde (MWs)
                                                                                 Azione EM
                                                                    Laser
                                           Elettroporazione irreversibile



                                                                         ͖
                           Ultrasuoni focalizzati ad alta intensità (HIFU)       Azione meccanica
                                                           Crioablazione ͖       Azione termica




                                                                             ͖
                                   Iniezione percutanea di etanolo (PEI)
                                  Chemoembolizzazione trans-arteriale
                                                                                 Azione chimica
                                                              (TACE)


6




                                                                             ͖
                                          Radiofrequency currents (RFs)
                                            Microwave radiations (MWs)
                                                                                 EM action
                                                                    Laser
                                              Irreversible Electroporation



                               High Intensity Focused Ultrasound (HIFU)   ͖      Mechanical action
                                                             Cryoablation
                                                                          ͖      Thermal action




                                                                             ͖
                                    Percutaneous ethanol injection (PEI)
                                Transarterial Chemoembolization (TACE)
                                                                                 Chemical action
Termoablazione EM
EM Thermoablation
Ad oggi, il riscaldamento elettromagnetico ancora appare il miglior compromesso tra esigenze molto
diverse:
• controllo sulla geometria della termolesione indotta
• sicurezza e semplicità d’uso
• accessibilità dei costi

                                              Tuttavia…

Le varie sorgenti di energia EM disponibili (RF, MW, Laser) differiscono enormemente l’una dall’altra
nel modo in cui interagiscono con i tessuti biologici e nella performance coagulativa complessiva.




                                                                                                            7



At present, electromagnetic heating still seems to offer the best trade off among very different require-
ments:
• control over the geometry of the induced necrosis
• safety and ease of use
• cost effectiveness

                                              However…

The various available EM energy sources (RFs, MWs, Laser) greatly differ one from the other in the way
they interact with biological tissues and in their final coagulative performance.
Efficienza del riscaldamento EM
    EM heating effectiveness
    Ai fini dell’efficacia del riscaldamento da fonti EM occorre individuare il miglior compromesso
    tra due parametri inversamente proporzionali:

                        Pv (potenza assorbita per unità di volume) α frequenza

                          Dp (profondità di penetrazione) α 1 / frequenza


    For maximum EM heating effectiveness the best trade off must be found between two inversely
    proportional parameters:

                           Pv (power absorbed per unit volume) α frequency

                              Dp (depth of penetration) α 1 / frequency



8
    Tasso d’assorbimento dell’energia EM da parte delle molecole
    d’acqua
    EM energy absorption rate of water
Quale fonte d’energia EM?
Which EM energy source?
                                     Radiazioni non ionizzanti
• Laser
• Correnti a radiofrequenza (RFs)
• Microonde (MWs)

                                             Tuttavia…

• Per il trattamento di lesioni di medie e grandi dimensioni le radiazioni laser appaiono inadeguate
 (elevata Pv, ma bassa Dp!)
• RF e MW trovano un’applicabilità clinica più estesa
• Pertanto, la vera sfida tecnologica è:

                                           RFs vs MWs




                                                                                                       9


                                       Non ionizing radiations
• Laser
• Radiofrequency currents (RFs)
• Microwaves (MWs)

                                             However…

• For the treatment of medium and big size lesions laser radiations are inadequate (high Pv but very
low Dp!)
• RFs and MWs have a considerably wider clinical scope
• Therefore, the true clinical and technological challenge is:

                                           RFs vs MWs
RF vs Microonde / Meccanismo di riscaldamento
     RF vs Microwaves / Heating mechanism
                            RFs                                                  MWs




     Le correnti RF usate nella pratica clinica (fre-     Le oscillazioni dei dipoli atomici, che seguono le
     quenza operativa: qualche centinaio di KHz) in-      alternanze del campo elettrico, generano calore
     ducono il riscaldamento nei tessuti per              per attrito. Tale meccanismo si amplifica nelle
     dissipazione ohmica (effetto Joule e agitazione       molecole POLARI, come l’acqua, di cui i tessuti
     ionica).                                             sono molto ricchi...


     The radiofrequency currents (RFs) used in clinical
10
     practice (operating frequency: a few hundreds        The oscillations of atomic dipoles, striving to catch
     KHz) induce tissue heating simply through ohmic      up with the continuously switching electric field,
     dissipation (Joule effect) and ionic agitation       generates frictional heat. This mechanism is am-
                                                          plified in POLAR molecules, such as water mole-
                                                          cules. Biological tissues are very rich in water…
RF vs Microonde / Velocità di riscaldamento
      RF vs Microwaves / Heating velocity
                       RFs                                                  MWs
Le correnti RF riscaldano per azione diretta sola-     Le microonde riscaldano per azione diretta e in
mente i tessuti nelle vicinanze della punta espo-      simultanea l’intero volume tessutale raggiunto,
sta dell’elettrodo: la propagazione del calore a       senza ritardi di propagazione né disomogeneità.
distanza avviene, più lentamente, per condu-
zione termica, che dipende dalle proprietà fisiche      Microwaves actively and simultaneously heat the
e biologiche dei tessuti.                              whole irradiated tissue, with no propagation delay
                                                       nor unhomogeneous distribution.
RF currents directly heat only tissues in touch with
or in close proximity of the current injecting elec-
trode: heat propagation to a greater distance is
mainly due to a slower thermal conduction
process, depending on the physical and biological
properties of the tissues.



                                                                                                            11




                             Le microonde scaldano più velocemente!
La velocità di riscaldamento delle microonde contrasta l’effetto di “heat sinking” del circolo sanguigno:
si può operare in prossimità dei grandi vasi senza significative perdite di performance coagulativa. La
nettezza dei gradienti di temperatura indotti dalle microonde porta a termolesioni dai margini altret-
tanto netti.


                                            MWs heat faster!
The microwave heating velocity effectively opposes the “heat sinking” due to blood circulation: one may
operate very close to large vessels without sensible loss of coagulative performance. Steep temperature
gradients also yield very clear-cut lesion margins.
RF vs Microonde / Uniformità di riscaldamento
           RF vs Microwaves / Heating uniformity
                             RFs                                                  MWs
     La circolazione delle correnti RF nel corpo del pa-     Le microonde si propagano anche attraverso il
     ziente richiede un percorso ininterrotto e a bassa      vuoto o regioni a bassa conducibilità, garan-
     impedenza tra l’elettrodo impiantato e le piastre       tendo un notevole grado di omogeneità e ripe-
     di riferimento applicate al paziente: qualora il        tibilità delle figure di necrosi indotte.
     tessuto-bersaglio sia scarsamente conduttivo o
     fortemente irregolare, il trattamento RF può ri-        Microwaves propagate even across vacuum or
     sultare inefficace, o incompleto, o difficilmente           poorly conductive regions, allowing sensibly
     controllabile e predicibile.                            higher homogeneity and repeatability of their
                                                             heating patterns
     RFs circulation within the patient’s body requires
     a continuous, low impedance path between the
     implanted current injecting electrode and the
     grounding pad applied to the patient: whenever
     the target tissues are poorly electrically conductive
     or markedly irregular, RF treatments may prove
     ineffective, or incomplete, or hardly controllable
12   and predictable.




                           Le microonde ottengono lesioni più uniformi e ripetibili!
     La propagazione di microonde è soggetta a meno vincoli legati alle proprietà fisiche dei tessuti e alla
     loro variabilità locale: le microonde garantiscono prestazioni più uniformi e ripetibili, su una più ampia
     gamma di tessuti.


                        MWs provide uniform and repeatable coagulative performances!
     Microwaves are less affected by the constraints due to the physical properties of tissues and their local
     variability. As a consequence microwaves produce more repeatable and uniform performances over a
     wider range of tissues.
RF vs Microonde / Confinamento
      RF vs Microwaves / Confinement
                        RFs                                                     MWs
Le correnti RF si disperdono attraverso il corpo         I radiatori a microonde hanno un raggio d’azione
del paziente lungo percorsi non predicibili; per-        intrinsicamente confinato: le onde EM decadono
tanto non dovrebbero subire trattamenti RF pa-           esponenzialmente con la distanza dal punto di
zienti con pacemakers, protesi metalliche o altri        emissione. Le antenne a microonde convenzio-
dispositivi impiantati. Inoltre, l’esito di un’abla-     nali esibiscono una figura di radiazione ellissoi-
zione RF può cambiare apprezzabilmente in fun-           dale attorno alla parte emettitrice.
zione della posizione delle piastre di dispersione.

RF currents disperse through the patient’s body          Microwave radiators exhibit an intrinsically con-
along unpredictable paths. For this reason, RF ab-       fined range of action: indeed, EM waves amplitude
lation should not be performed on patients bear-         falls exponentially with increasing distance from
ing pacemakers or other current-sensitive devices        the emitting point. Conventional microwave an-
or metal prosthesis. Also, the outcome of an RF          tennas feature an ellipsoidal radiation pattern
treatment may markedly change depending on               around their emitting tips.
the grounding pads placement.

                                                                                                                 13




            DELOCALIZED ACTION                                           CONFINED ACTION


                      Le microonde sono più sicure e meglio confinate!
  Le microonde agiscono solo sul tessuto-bersaglio senza risentire o interferire con altre parti del
                                paziente al di fuori della lesione.

                        MWs are inherently safer and more “on the target”!
MWs provide a truly target-focussed ablative tool, not affected by nor interfering with parts of the patient’s
body apart from the target lesion.
RF vs Microonde / Bipolarità
           RF vs Microwaves / Bipolarity
                           RFs                                                    MWs
     Nella maggior parte dei casi, i sistemi per abla-      I radiatori MW sono forzatamente bipolari: non
     zione RF sono monopolari (cioè, un solo elet-          sono richieste piastre di dispersione, eliminando
     trodo impiantato in funzione di iniettore di           così le criticità a queste associate (ustioni cuta-
     corrente e una o più piastre di dispersione ap-        nee, incertezze sul posizionamento, allunga-
     plicate al paziente.                                   mento dei tempi di preparazione del paziente
                                                            all’intervento).
     In the majority of cases, RF ablative systems are
     monopolar (that is, just one implanted electrode       Microwave radiators are bipolar in nature:
     serving as current injector and one or more dis-       grounding pads are not required!!!
     persing electrodes applied to the patient).            All problems arising from the pads (inappropriate
                                                            positioning, skin burns, longer procedure set-up
                                                            time) are automatically overcome.




14




                             Gli applicatori MW sono intrinsicamente bipolari!
     Gli applicatori a MW sono antenne, intrinsecamente bipolari, che non necessitano quindi di piastre
                di dispersione né risentono dei problemi legati al posizionamento delle stesse.

                                  MW applicators are intrinsically bipolar!
     MW applicators are EM antennas, inherently bipolar, thus needless of dispersing electrodes and unaffected
     by grounding pad-related issues.
RF vs Microonde / Versatilità d’impiego
      RF vs Microwaves / Versatility
                        RFs                                                    MWs
Le possibilità di utilizzo “fuori standard” della tec-   La tecnologia MW offre una varietà illimitata di
nologia RF sono fortemente limitate:                     setup operativi:
Non è possibile impiegare elettrodi RF alimentati         Si può utilizzare un numero arbitrario di appli-
simultaneamente, a causa delle reciproche in-            catori MW simultaneamente attivi, senza rischi
terferenze che ne scaturirebbero: si deve ricorrere      o vincoli, accrescendo le dimensioni della termo-
a sistemi ad alimentazione alternata (switching),        lesione ottenibile nel medesimo tempo d’inter-
con tempi d’intervento più lunghi                        vento.
Un applicatore RF è semplicemente un elettrodo           Teoricamente, utilizzando una schiera di antenne
che inietta corrente: si dispone di pochi parame-        in coerenza di fase (phased array) si può ablare
tri operativi e di design su cui agire per modifi-        una regione tessutale di forma e localizzazione
care forma ed estensione della figura di                  arbitrarie.
riscaldamento qualora se ne presenti la necessità         Virtualmente, esiste uno specifico design d’an-
in particolari applicazioni.                             tenna per ottenere qualsivoglia figura di riscal-
                                                         damento.
Out of standard operability is very limited with RF
technology:                                              MW technology provides an unlimited range of
Multiple RF electrodes may not be used simulta-          possible operative setups:                             15
neously, due to possible mutual interferences and        An arbitrary number of MW radiators may be used
loss of control over current paths: switching meth-      at once, without risks or limitations. Big or multi-
ods must be used, which require longer treatment         ple lesions in a short time!
times                                                     In principle, by operating a phased MW antennas
An RF applicator is simply an electrode injecting        array, one may selectively burn out a volume of
current through single or multiple active tips: very     tissue of arbitrary shape, size and location.
few design and operative parameters are available         Specific antenna designs may be used to attain
to modify the final heating pattern in order to          virtually any coagulative pattern
match special requirements                                For a given antenna design and a given energy
                                                         deposition rate, the lesion dimensions may be
                                                         made to vary within a sufficiently wide range of
                                                         values simply by changing the treatment duration




Gli applicatori MW possono essere progettati per operare individualmente o in schiera e otte-
                     nere termolesioni di forma e dimensioni arbitrarie!

 MW applicators may be designed to work individually or in group so as to provide thermal le-
                            sions of arbitrary shape and size!
Le ragioni del predominio RF
     Why RFs are still leading
     • Minor costo e complessità rispetto alla tecnologia MW. L’erogazione sicura di potenza MW per usi
       medicali richiede la conversione dai comuni magnetron (che operano ad elevate tensioni AC) alla
       più sofisticata tecnologia a stato solido (speciali transistor, alimentati a basse tensioni DC), decisa-
       mente più costosa, sebbene le cose ultimamente stiano cambiando…

     • Miglior controllo sulla figura di necrosi. Le MW risentono di fenomeni di retro-riscaldamento (EF-
       FETTO COMETA) dovuto sia alla presenza di radiazione non assorbita dai tessuti-bersaglio e che si
       propaga a ritroso lungo le pareti esterne dell’applicatore),sia alle rilevanti perdite di potenza in calore
       lungo il cavo coassiale che alimenta l’antenna (con il rischio di surriscaldamento dell’applicatore).

     • Lesser cost and complexity with respect to MW technology. Safe delivery of MW power in clinical treat-
       ments requires conversion from cheap magnetron sources (which are operated at high AC voltages) to
       solid state technology (transistors, fed at low DC voltages). The latter technology is definitely more ex-
       pensive, though things have started changing lately…


     • Better control over coagulation pattern. MWs are affected by back heating phenomena (COMET
16     EFFECT) both due to poor control over reflected power (that is, the portion of MW power unabsorbed
       by the target-tissue and propagating backwards along the antenna shaft) and due to intense power
       losses along the antenna feeding cable (turning into severe shaft overheating).
Applicatori MW convenzionali
Conventional MW probes




                                                                                                   17




I comuni rimedi al fenomeno di retro-riscaldamento (sia dovuto alle riflessioni che alle perdite
ohmiche lungo la lina coassiale) comportano un considerevole incremento del calibro degli ap-
plicatori MW.
In alternativa, si possono utilizzare applicatori MW sottili, ma a potenze limitate.

                                In pratica, fino a ieri si aveva:
                                 Calibro applicatori RF: 14-17G
                                Calibro applicatori MW: 11-13G


Common technical remedies to back heating effects (both due to reflected waves and due to ohmic
dissipation along the feeding coaxial line) bring a remarkable increase in MW applicators gauge.
Alternatively, slender MW probes may be used, but with limited power output.

                                  In practice, till yesterday:
                                 RF applicators gauge: 14-17G
                                MW applicators gauge : 11-13G
In sintesi…
     Summarizing…
     Se si potesse garantire per gli applicatori MW, contemporaneamente:
      • un piccolo calibro, ai fini della minima invasività
      • un efficace sistema di intrappolamento delle microonde
      • un adeguato sistema di raffreddamento dell’antenna
     allora le MW potrebbero realmente candidarsi a sostituire le RF nel ruolo di tecnologia termoablativa
     d’elezione, in quanto le MW assicurano:
     • Un riscaldamento più rapido
     • Accresciuta affidabilità e versatilità d’impiego
     • Performance coagulativa più omogenea e ripetibile
     Per queste ragioni la comunità medica mondiale ha grandi aspettative riguardo le MW. Negli
     ultimi anni, diversi gruppi industriali hanno raccolto la sfida tecnologica finalizzata allo sviluppo di
     nuovi e migliori sistemi per ablazione MW con tutti i necessari requisiti di sicurezza, efficienza e
     accessibilità dei costi.




18



     If one could warrant for microwave applicators, at once:
     • a small gauge, so as to attain minimum invasiveness
     • an effective system for microwaves trapping
     • a suitable antenna cooling system
     then microwaves would certainly challenge RFs supremacy in the field of thermoablation, since mi-
     crowaves attain:
     • Faster heating
     • Enhanced safety and versatility of use
     • Homogeneous and repeatable coagulative performance
     For these reasons, the medical community worldwide has great expectations concerning mi-
     crowaves. In the last few years, a technological challenge has started among several important in-
     dustrial groups in different corners of the world, in order to develop new microwave systems finally
     fulfilling all requirements for a safe, efficient and cost effective clinical use.
AMICA

                                          AMICA-PROBE
Applicatore interstiziale monouso ad introduzione diretta, dotato di dispositivo miniaturizzato per il
confinamento delle MW (BREVETTO MONDIALE!), circuito di raffreddamento, termocoppia integrata
                                        e memoria digitale.

                                          AMICA-GEN
            Generatore di microonde (2450 MHz/100W), programmabile, a stato solido.

                                          AMICA-PUMP
       Pompa peristaltica per il raffreddamento dell’applicatore, interamente automatizzata.




                                                                                                         19




                                             AMICA-PROBE
   Disposable interstitial applicator, internally cooled, featuring a patented miniaturized device for
 microwaves confinement (WORLDWIDE PATENT!), an integrated thermocouple for probe temperature
                                 monitoring and a digital memory chip.

                                           AMICA-GEN
                  Programmable, solid state microwave source (2450 MHz/100W).

                                          AMICA-PUMP
                         Fully automated peristaltic pump for probe cooling.
20
Struttura di AMICA-PROBE
AMICA-PROBE structure




• “Punta calda”: l’antenna viene raffreddata per tutta la sua lunghezza, ad eccezione della porzione
  distale (~ 2.5 cm)
• Rivestimento in teflon anti-aderente
• Punta attiva di lunghezza assegnata (~ 2 cm): le dimensioni delle termolesioni indotte cambiano
  modificando opportunamente potenza erogata e durata del trattamento. Il concetto di “punta
  esposta” (proprio degli elettrodi RF) non si applica più.
                                                                                                        21

• “Hot tip”: the antenna is cooled throughout its length, except for its distal end (~ 2.5 cm)
• Anti-sticking teflon coating
• Active tip of assigned length (~ 2 cm): lesion dimensions may be varied by changing delivered power
  and treatment duration. The “exposed tip” concept is no longer applicable.
AMICA-PROBE: performance coagulativa
     AMICA-PROBE: coagulative performance

                Dati ex vivo su fegato bovino a temperatura ambiente
                   Ex vivo data on bovine liver at room temperature




                                           Mean aspect ratio (D/L) = 0.8




22
AMICA-GEN




• Sorgente a stato solido (transistor alimentati a 12 VDC)
• Potenza disponibile: 100W CW
• Frequenza: 2450 MHz
• Interfaccia-utente essenziale e intuitiva
• Architettura digitale : massima versatilità!
• Interattività: collegabile a un PC e a un numero illimitato di altre periferiche
• Massima sicurezza: sofisticato sistema di allarmi e protezioni
                                                                                          23



•   Solid state source (transistors fed at 12 VDC)
•   Available power: 100W CW
•   Frequency: 2450 MHz
•   Essential and intuitive user interface
•   Digital architecture: maximum versatility!
•   Interactivity: connectable to PC and to an unlimited number of external peripherals
•   Maximum safety: sophisticated system of alarms and protections
Feedback in tempo reale dalla lesione
     Real time feedback from lesion




24

     S11 è il coefficiente di riflessione, cioè il rapporto tra potenza a MW riflessa e potenza
     erogata (in valore assoluto).

     S11 is the MW probe reflection coefficient, that is the ratio between reflected and direct MW power
     (in absolute value).
     AMICA-GEN consente il monitoraggio continuo del coefficiente di riflessione e della tempe-
     ratura dell’applicatore.
     La lesione “ci parla”…

     AMICA-GEN allows for continous monitoring of probe temperature and reflection coefficient.

     The lesion “speaks out”…
HS AMICA: riassumendo…
HS AMICA: summarizing…
Frequenza operativa                            2450 MHz
Operating frequency                            2450 MHz
Max potenza all’applicatore                    70 W
Max available power on probe                   70 W
Calibri degli applicatori                      11G, 14G , 16G
Applicator gauge                               11G, 14G , 16G
                                               Mini-choke (brevetto mondiale)
Controllo dell’effetto-cometa                   Patented mini-choke
Back heating control                           Raffreddamento interno
                                               Internal probe cooling
Controllo della temperatura dell’applicatore   Termocoppia integrata nell’applicatore
Probe temperature control                      Embedded Thermocouple
Max. lesione (ex vivo su fegato bovino,        Più di 4 cm di diametro in 10 min
con ago singolo in un solo passaggio)
Max. lesion diam. (ex vivo bovine liver,       More than 4 cm in 10 min
single probe, single insertion)
Accessori nel kit monouso                      Mini-bisturi per la pre-incisione della cute;
                                               introduttore coassiale; termocoppia                 25
                                               interstiziale; sacca di raccolta del refrigerante
                                                refluo; telo
Probe kit accessories                          Mini-scalpel for skin pre-incision; coaxial
                                               introducer; interstitial thermocouple;
                                               integrated refluent coolant collector;
                                               drape
Applicatori multipli                           Possibile per un numero arbitrario di applicatori
                                               in simultanea, a qualunque distanza reciproca
Multiple probes operation                      Possible, with any number of probes operated
                                               simultaneously, at any reciprocal distance
Generatore                                     Architettura digitale “aperta” : interfaccia-
                                               utente interattive e personalizzabile.
                                               Monitoraggio continuo di temperatura e
                                               riflessioni MW.
Generator                                      “Open” digital architecture: interactive
                                               and customizable user-interface.
                                               Continuous monitoring of probe
                                               temperature and MW reflections.
Certificazioni                                  CE, FDA
Approvals                                      CE, FDA
Caratteristiche esclusive di AMICA
     Unique features of AMICA

     • Applicatori sottili (finanche 16G)
     • Ampia varietà di applicatori (11G per ablazione ossea o per intra-operatorio; 14G e 16G per uso
       percutaneo; applicatori “lunghi” per laparoscopia, e “corti” per uso sotto guida TAC; applicatori flessibili
       presto disponibili!)
     • Elevata potenza disponibile (70W netti all’applicatore)
     • Grandi termolesioni (oltre 4cm di diametro, con applicatore singolo in 10 minuti)
     • Controllo ottimale di forma e dimensioni del coagulo. Niente effetto-cometa (mini-choke per la
       soppressione delle MWs riflesse e sistema di raffreddamento integrato per prevenire il surriscalda-
       mento dello stelo)
     • Monitoraggio costante delle riflessioni e della temperatura dell’applicatore, che fornisce un
       feedback in tempo reale sull’andamento del trattamento
     • Chip di memoria integrato per un’accresciuta sicurezza d’uso (auto-limitazione della massima po-
       tenza erogabile ad un dato applicatore, auto-diagnosi di integrità e scadenza dell’applicatore)
     • Generatore ad architettura digitale aperta : interfaccia-utente intuitiva e altamente interattiva,
       personalizzabile a piacere; salvataggio dei dati operativi e collegabilità a PC;
     • Pompa peristaltica completamente automatizzata
26


     • Small applicator gauge (down to 16G)
     • Wide variety of probes (11G for bone ablation or for intra-surgical treatments; 14G and 16G for in-
       terstitial ablations; oversized lengths for laparascopic treatments, undersized lengths for CT-guided
       procedures; flexible probes coming soon!)
     • High available power (70W net at applicator input section)
     • Big lesions (over 4cm in diameter, with single applicator in 10 minutes)
     • Optimal control on the coagulative pattern size and shape. No back-heating (mini-choke for reflected
       MWs trapping and integrated cooling system to avoid shaft overheating)
     • Constant monitoring of MW reflections and probe temperature, providing real time feedback on
       treatment progression
     • Built-in memory chip for enhanced safety of use (self-limitation of maximum delivered power de-
       pending on applicator characteristics, self-diagnosis of probe expiration and integrity)
     • Open-architecture, digital microwave generator: very intuitive and highly interactive user interface,
       infinitely customizable; ablation data storage and PC link available;
     • Fully automated peristaltic pump
2450 MHz vs 915 MHz
2450 MHz vs 915 MHz




Calcolo dell’intensità di campo elettrico (figura dx) e del tasso d’assorbimento specifico (figura sx) in
fantoccio muscolo-equivalente in funzione della distanza da dipoli elementari rispettivamente operati
a 915MHz (curve rosse) e a 2450 MHz (curve blu). A parità di potenza erogata, il campo a 2450 MHz è
nettamente superiore al campo a 915 MHz fino a 4cm di distanza dall’antenna: come a dire che in
tutto il volume di un nodulo fino a 8 cm di diametro l’azione di un applicatore a 2450 MHz è marcata-
mente più efficace rispetto ad un applicatore a 915 MHz.
(A cura del Dipartimento di Ingegneria Elettronica dell’Università degli Studi di Roma “La Sapienza”)             27



Calculation of the electric field intensity (on the right) and of the specific absorption rate (on the left) in
a muscle-equivalent phantom as functions of the distance from elementary dipoles, respectively operated
at 915MHz (red curves) and at 2450 MHz (blue curves). The 2450 MHz field turns out to be definitely
higher than the 915 MHz field up to 4cm away from the antenna. It means that in lesions up to 8 cm
of diameter the efficacy of a 2450 MHz applicator is higher than that of a 915 MHz applicator.
(Courtesy of the Electronic Engineering Department of the University of Rome “La Sapienza”)
AMICA: Clinical Results - Liver ablation (HCC)




28
29
Hepatic metastases from lungs




                                Single lesion: 54x48 mm
                Courtesy of Dr. M. F. Meloni, S. Gerardo Hospital, Monza, Italy




30
     Multiple metastases from myosarcoma
Hepatic metastases from lungs (US-guided laparoscopy)




           Courtesy of Dr.Santambrogio, S. Giuseppe Hospital, Milan, Italy




                                                                              31
HCC / Pre- and post-ablation CT images




              Courtesy of Dr. L. Solbiati, Hospital of Busto Arsizio, Italy
HCC / Pre- and post-ablation CT images




               Courtesy of Dr. L. Solbiati, Hospital of Busto Arsizio, Italy

32
     HCC / Pre- and post-ablation CT images




               Courtesy of Dr. L. Solbiati, Hospital of Busto Arsizio, Italy
Phase I clinical trial on AMICA (safety and tolerability):
(the study was published on the Journal of Endourology in July 2008)




 “TRANS PERINEAL MICROWAVE THERMOABLATION IN PATIENTS WITH OBSTRUCTIVE BENIGN
PROSTATIC HYPERPLASIA: A PHASE I CLINICAL STUDY WITH A NEW MINI-CHOKED MICROWAVE
                                    APPLICATOR”.

        Bartoletti R.1, Cai T. 1, Pinzi N. 2, Longo I. 3, Ricci A. 3, Massaro MP. 4, Tosoratti N. 4

                      1- Department of Urology, University of Florence, Italy
                      2- Urologic Unit, Campo di Marte Hospital - Lucca, Italy
                      3- Institute Chemo-Physical Processes, CNR - Pisa, Italy
                           4- HS Hospital Service SpA - Aprilia (LT), Italy



                                                                                                      33
                RESULTS (20 W / 5 min)
RESULTS (20 W / 5 min)

     Pathological analyses showed a quasi-spheroidal lesion (longitudinal axis: 18 mm; transversal
     axis: 16 mm), characterized by a central coagulative necrosis surrounded by an inflammatory
     infiltrate reaction (thickness: 2.1 mm), beyond which healthy tissue was found.




34
                                       CONCLUSIONS




           The AMICA-PROBE treatment is a safe, well tolerated and repeatable treatment
           for benign prostatic hypertrophy. Further clinical phase II-III studies are needed
           in order to evaluate its efficacy and morbidity in comparison with classical TUMT
           or trans urethral resection of the prostate.
HS HOSPITAL SERVICE S.P.A

Via Zosimo 13 - 00178 Roma, ITALY

Stabilimento di produzione/Factory:
Via Angela Vacchi 24/26
04011 Aprilia (Latina), ITALY

Tel.: +39 06 9201961
Fax: +39 06 92727871
email: hs@hshospitalservice.com
www.hshospitalservice.com

Weitere ähnliche Inhalte

Andere mochten auch

How To Make Sure Your Brand Is Dressed To The Nines
How To Make Sure Your Brand Is Dressed To The NinesHow To Make Sure Your Brand Is Dressed To The Nines
How To Make Sure Your Brand Is Dressed To The NinesVerticalResponse
 
LiveHelp Server for Joomla Quick start v3
LiveHelp Server for Joomla Quick start   v3LiveHelp Server for Joomla Quick start   v3
LiveHelp Server for Joomla Quick start v3ActiveHelper
 
Milestone Group Plc
Milestone Group PlcMilestone Group Plc
Milestone Group PlcDeborahWhite
 
2013 cillo laparoscopic ablation of hepatocellular carcinoma in cirrhotic pat...
2013 cillo laparoscopic ablation of hepatocellular carcinoma in cirrhotic pat...2013 cillo laparoscopic ablation of hepatocellular carcinoma in cirrhotic pat...
2013 cillo laparoscopic ablation of hepatocellular carcinoma in cirrhotic pat...Marco Zaccaria
 
Understanding Social Media
Understanding Social MediaUnderstanding Social Media
Understanding Social MediaVerticalResponse
 
Email and social trends for 2013
Email and social trends for 2013 Email and social trends for 2013
Email and social trends for 2013 VerticalResponse
 
Chelsiapowerpoint22
Chelsiapowerpoint22Chelsiapowerpoint22
Chelsiapowerpoint22cjbates89
 
How to Make Social Media and PR Work Together Like PB&J
How to Make Social Media and PR Work Together Like PB&JHow to Make Social Media and PR Work Together Like PB&J
How to Make Social Media and PR Work Together Like PB&JVerticalResponse
 
A 3-869e531bc383bc2db2d9e22b083ea95f5f0fa454-141125232253-conversion-gate02
A 3-869e531bc383bc2db2d9e22b083ea95f5f0fa454-141125232253-conversion-gate02A 3-869e531bc383bc2db2d9e22b083ea95f5f0fa454-141125232253-conversion-gate02
A 3-869e531bc383bc2db2d9e22b083ea95f5f0fa454-141125232253-conversion-gate02Chris Humphreys
 
Tips Menjadi Penulis Produktif
Tips Menjadi Penulis ProduktifTips Menjadi Penulis Produktif
Tips Menjadi Penulis ProduktifArry Rahmawan
 
Learnings from Building a Lean Startup Within 27 Years Old 150MEUR+ Business
Learnings from Building a Lean Startup Within 27 Years Old 150MEUR+ BusinessLearnings from Building a Lean Startup Within 27 Years Old 150MEUR+ Business
Learnings from Building a Lean Startup Within 27 Years Old 150MEUR+ BusinessJanne Pirttilahti
 
SXSW 2013: Show & Smell 2: Marketing Experiences Beyond the Visual
SXSW 2013: Show & Smell 2: Marketing Experiences Beyond the VisualSXSW 2013: Show & Smell 2: Marketing Experiences Beyond the Visual
SXSW 2013: Show & Smell 2: Marketing Experiences Beyond the VisualTaylor Host
 
Lotus notes-703-1232746032205457-2
Lotus notes-703-1232746032205457-2Lotus notes-703-1232746032205457-2
Lotus notes-703-1232746032205457-2IN023227
 
Au Psy492 E Portfolio Template For Slide Share Revised
Au Psy492 E Portfolio Template For Slide Share RevisedAu Psy492 E Portfolio Template For Slide Share Revised
Au Psy492 E Portfolio Template For Slide Share Revisedcharmaine03
 

Andere mochten auch (20)

How To Make Sure Your Brand Is Dressed To The Nines
How To Make Sure Your Brand Is Dressed To The NinesHow To Make Sure Your Brand Is Dressed To The Nines
How To Make Sure Your Brand Is Dressed To The Nines
 
LiveHelp Server for Joomla Quick start v3
LiveHelp Server for Joomla Quick start   v3LiveHelp Server for Joomla Quick start   v3
LiveHelp Server for Joomla Quick start v3
 
Milestone Group Plc
Milestone Group PlcMilestone Group Plc
Milestone Group Plc
 
2013 cillo laparoscopic ablation of hepatocellular carcinoma in cirrhotic pat...
2013 cillo laparoscopic ablation of hepatocellular carcinoma in cirrhotic pat...2013 cillo laparoscopic ablation of hepatocellular carcinoma in cirrhotic pat...
2013 cillo laparoscopic ablation of hepatocellular carcinoma in cirrhotic pat...
 
Understanding Social Media
Understanding Social MediaUnderstanding Social Media
Understanding Social Media
 
Email and social trends for 2013
Email and social trends for 2013 Email and social trends for 2013
Email and social trends for 2013
 
Centro de Atención al Migrante
Centro de Atención al MigranteCentro de Atención al Migrante
Centro de Atención al Migrante
 
30-Minute Email Recipes
30-Minute Email Recipes 30-Minute Email Recipes
30-Minute Email Recipes
 
Chelsiapowerpoint22
Chelsiapowerpoint22Chelsiapowerpoint22
Chelsiapowerpoint22
 
How to Make Social Media and PR Work Together Like PB&J
How to Make Social Media and PR Work Together Like PB&JHow to Make Social Media and PR Work Together Like PB&J
How to Make Social Media and PR Work Together Like PB&J
 
Subject Line Savvy
Subject Line SavvySubject Line Savvy
Subject Line Savvy
 
A 3-869e531bc383bc2db2d9e22b083ea95f5f0fa454-141125232253-conversion-gate02
A 3-869e531bc383bc2db2d9e22b083ea95f5f0fa454-141125232253-conversion-gate02A 3-869e531bc383bc2db2d9e22b083ea95f5f0fa454-141125232253-conversion-gate02
A 3-869e531bc383bc2db2d9e22b083ea95f5f0fa454-141125232253-conversion-gate02
 
Tips Menjadi Penulis Produktif
Tips Menjadi Penulis ProduktifTips Menjadi Penulis Produktif
Tips Menjadi Penulis Produktif
 
Learnings from Building a Lean Startup Within 27 Years Old 150MEUR+ Business
Learnings from Building a Lean Startup Within 27 Years Old 150MEUR+ BusinessLearnings from Building a Lean Startup Within 27 Years Old 150MEUR+ Business
Learnings from Building a Lean Startup Within 27 Years Old 150MEUR+ Business
 
Successful email2013
Successful email2013Successful email2013
Successful email2013
 
SXSW 2013: Show & Smell 2: Marketing Experiences Beyond the Visual
SXSW 2013: Show & Smell 2: Marketing Experiences Beyond the VisualSXSW 2013: Show & Smell 2: Marketing Experiences Beyond the Visual
SXSW 2013: Show & Smell 2: Marketing Experiences Beyond the Visual
 
What is coaching?
What is coaching?What is coaching?
What is coaching?
 
Lotus notes-703-1232746032205457-2
Lotus notes-703-1232746032205457-2Lotus notes-703-1232746032205457-2
Lotus notes-703-1232746032205457-2
 
Prova
ProvaProva
Prova
 
Au Psy492 E Portfolio Template For Slide Share Revised
Au Psy492 E Portfolio Template For Slide Share RevisedAu Psy492 E Portfolio Template For Slide Share Revised
Au Psy492 E Portfolio Template For Slide Share Revised
 

Ähnlich wie Hs amica apparatus for mi crowave ablation

Rischio campi elettromagnetici
Rischio campi elettromagneticiRischio campi elettromagnetici
Rischio campi elettromagneticiseagruppo
 
2.Effetti Fisici e Biologici RM_uninovara
2.Effetti Fisici e Biologici RM_uninovara2.Effetti Fisici e Biologici RM_uninovara
2.Effetti Fisici e Biologici RM_uninovaraLuca Gastaldi
 
2005 terni, università di medicina. corso di elettrofisiologia, la biofisica...
2005 terni, università di medicina. corso di elettrofisiologia, la biofisica...2005 terni, università di medicina. corso di elettrofisiologia, la biofisica...
2005 terni, università di medicina. corso di elettrofisiologia, la biofisica...Centro Diagnostico Nardi
 
La Valutazione dei Rischi per la Presenza dei Campi Elettromagnetici - Road S...
La Valutazione dei Rischi per la Presenza dei Campi Elettromagnetici - Road S...La Valutazione dei Rischi per la Presenza dei Campi Elettromagnetici - Road S...
La Valutazione dei Rischi per la Presenza dei Campi Elettromagnetici - Road S...ghirardo
 
2.effetti fisici e_biologici
2.effetti fisici e_biologici2.effetti fisici e_biologici
2.effetti fisici e_biologiciLuca Gastaldi
 
Compatibilità Elettromagnetica Sfide Attuali [Modalità Compatibilità]
Compatibilità Elettromagnetica Sfide Attuali [Modalità Compatibilità]Compatibilità Elettromagnetica Sfide Attuali [Modalità Compatibilità]
Compatibilità Elettromagnetica Sfide Attuali [Modalità Compatibilità]marco_crescenzi
 
Radiazioni non ionizzanti
Radiazioni non ionizzantiRadiazioni non ionizzanti
Radiazioni non ionizzantimaurocicognani
 
Implicazioni mediche e biologiche derivanti dall’utilizzo dei principali elet...
Implicazioni mediche e biologiche derivanti dall’utilizzo dei principali elet...Implicazioni mediche e biologiche derivanti dall’utilizzo dei principali elet...
Implicazioni mediche e biologiche derivanti dall’utilizzo dei principali elet...Roberto Bonin
 
Multicentric research on triple therapy
Multicentric research on triple therapyMulticentric research on triple therapy
Multicentric research on triple therapyLevel Medical
 
MICRODOSIMETRIA DEL CAMPO ELETTROMAGNETICO INDOTTO A LIVELLO CELLULARE: UNO S...
MICRODOSIMETRIA DEL CAMPO ELETTROMAGNETICO INDOTTO A LIVELLO CELLULARE: UNO S...MICRODOSIMETRIA DEL CAMPO ELETTROMAGNETICO INDOTTO A LIVELLO CELLULARE: UNO S...
MICRODOSIMETRIA DEL CAMPO ELETTROMAGNETICO INDOTTO A LIVELLO CELLULARE: UNO S...annachiara_sepiacci
 
Rotary Forum 19.05.2012
Rotary Forum 19.05.2012Rotary Forum 19.05.2012
Rotary Forum 19.05.2012Sergio_Barta
 
OPUSCOLO INFORMATIVO SU CAMPI ELETTROMAGNETICI.pdf
OPUSCOLO INFORMATIVO SU CAMPI ELETTROMAGNETICI.pdfOPUSCOLO INFORMATIVO SU CAMPI ELETTROMAGNETICI.pdf
OPUSCOLO INFORMATIVO SU CAMPI ELETTROMAGNETICI.pdfISEA ODV
 
Inquinamento elettromagnetico
Inquinamento elettromagneticoInquinamento elettromagnetico
Inquinamento elettromagneticoISEA ODV
 
SISTEMA DI RIVELAZIONE INNOVATIVO PER IL MONITORAGGIO REAL-TIME IMPULSO PER I...
SISTEMA DI RIVELAZIONE INNOVATIVO PER IL MONITORAGGIO REAL-TIME IMPULSO PER I...SISTEMA DI RIVELAZIONE INNOVATIVO PER IL MONITORAGGIO REAL-TIME IMPULSO PER I...
SISTEMA DI RIVELAZIONE INNOVATIVO PER IL MONITORAGGIO REAL-TIME IMPULSO PER I...convegnonazionaleaiic
 

Ähnlich wie Hs amica apparatus for mi crowave ablation (20)

Principi fisici e sicurezza RM - Gastaldi
Principi fisici e sicurezza RM - GastaldiPrincipi fisici e sicurezza RM - Gastaldi
Principi fisici e sicurezza RM - Gastaldi
 
Rischio campi elettromagnetici
Rischio campi elettromagneticiRischio campi elettromagnetici
Rischio campi elettromagnetici
 
2.Effetti Fisici e Biologici RM_uninovara
2.Effetti Fisici e Biologici RM_uninovara2.Effetti Fisici e Biologici RM_uninovara
2.Effetti Fisici e Biologici RM_uninovara
 
1 P Introduzione
1 P  Introduzione1 P  Introduzione
1 P Introduzione
 
P Introduzione
P  IntroduzioneP  Introduzione
P Introduzione
 
Poster Amica
Poster AmicaPoster Amica
Poster Amica
 
2005 terni, università di medicina. corso di elettrofisiologia, la biofisica...
2005 terni, università di medicina. corso di elettrofisiologia, la biofisica...2005 terni, università di medicina. corso di elettrofisiologia, la biofisica...
2005 terni, università di medicina. corso di elettrofisiologia, la biofisica...
 
La Valutazione dei Rischi per la Presenza dei Campi Elettromagnetici - Road S...
La Valutazione dei Rischi per la Presenza dei Campi Elettromagnetici - Road S...La Valutazione dei Rischi per la Presenza dei Campi Elettromagnetici - Road S...
La Valutazione dei Rischi per la Presenza dei Campi Elettromagnetici - Road S...
 
2.effetti fisici e_biologici
2.effetti fisici e_biologici2.effetti fisici e_biologici
2.effetti fisici e_biologici
 
Compatibilità Elettromagnetica Sfide Attuali [Modalità Compatibilità]
Compatibilità Elettromagnetica Sfide Attuali [Modalità Compatibilità]Compatibilità Elettromagnetica Sfide Attuali [Modalità Compatibilità]
Compatibilità Elettromagnetica Sfide Attuali [Modalità Compatibilità]
 
Tipi
TipiTipi
Tipi
 
Radiazioni non ionizzanti
Radiazioni non ionizzantiRadiazioni non ionizzanti
Radiazioni non ionizzanti
 
Implicazioni mediche e biologiche derivanti dall’utilizzo dei principali elet...
Implicazioni mediche e biologiche derivanti dall’utilizzo dei principali elet...Implicazioni mediche e biologiche derivanti dall’utilizzo dei principali elet...
Implicazioni mediche e biologiche derivanti dall’utilizzo dei principali elet...
 
Multicentric research on triple therapy
Multicentric research on triple therapyMulticentric research on triple therapy
Multicentric research on triple therapy
 
MICRODOSIMETRIA DEL CAMPO ELETTROMAGNETICO INDOTTO A LIVELLO CELLULARE: UNO S...
MICRODOSIMETRIA DEL CAMPO ELETTROMAGNETICO INDOTTO A LIVELLO CELLULARE: UNO S...MICRODOSIMETRIA DEL CAMPO ELETTROMAGNETICO INDOTTO A LIVELLO CELLULARE: UNO S...
MICRODOSIMETRIA DEL CAMPO ELETTROMAGNETICO INDOTTO A LIVELLO CELLULARE: UNO S...
 
6.3 P Applicazioni
6.3   P  Applicazioni6.3   P  Applicazioni
6.3 P Applicazioni
 
Rotary Forum 19.05.2012
Rotary Forum 19.05.2012Rotary Forum 19.05.2012
Rotary Forum 19.05.2012
 
OPUSCOLO INFORMATIVO SU CAMPI ELETTROMAGNETICI.pdf
OPUSCOLO INFORMATIVO SU CAMPI ELETTROMAGNETICI.pdfOPUSCOLO INFORMATIVO SU CAMPI ELETTROMAGNETICI.pdf
OPUSCOLO INFORMATIVO SU CAMPI ELETTROMAGNETICI.pdf
 
Inquinamento elettromagnetico
Inquinamento elettromagneticoInquinamento elettromagnetico
Inquinamento elettromagnetico
 
SISTEMA DI RIVELAZIONE INNOVATIVO PER IL MONITORAGGIO REAL-TIME IMPULSO PER I...
SISTEMA DI RIVELAZIONE INNOVATIVO PER IL MONITORAGGIO REAL-TIME IMPULSO PER I...SISTEMA DI RIVELAZIONE INNOVATIVO PER IL MONITORAGGIO REAL-TIME IMPULSO PER I...
SISTEMA DI RIVELAZIONE INNOVATIVO PER IL MONITORAGGIO REAL-TIME IMPULSO PER I...
 

Mehr von Marco Zaccaria

EZ Trocar Hakko Medical JP
EZ Trocar Hakko Medical JPEZ Trocar Hakko Medical JP
EZ Trocar Hakko Medical JPMarco Zaccaria
 
Menzione di merito progetto hs amica
Menzione di merito progetto hs amicaMenzione di merito progetto hs amica
Menzione di merito progetto hs amicaMarco Zaccaria
 
2013 poggi anticancer research mw ablation of hcc using a new percutaneous de...
2013 poggi anticancer research mw ablation of hcc using a new percutaneous de...2013 poggi anticancer research mw ablation of hcc using a new percutaneous de...
2013 poggi anticancer research mw ablation of hcc using a new percutaneous de...Marco Zaccaria
 
2013 pusceddu jvir treatment of_bonemetastaseswithmicrowavethermal ablation
2013 pusceddu jvir treatment of_bonemetastaseswithmicrowavethermal ablation2013 pusceddu jvir treatment of_bonemetastaseswithmicrowavethermal ablation
2013 pusceddu jvir treatment of_bonemetastaseswithmicrowavethermal ablationMarco Zaccaria
 
Na int pascale lesioni polmonari inoperabili trattate con termoablazione mw
Na int pascale lesioni polmonari inoperabili trattate con termoablazione mwNa int pascale lesioni polmonari inoperabili trattate con termoablazione mw
Na int pascale lesioni polmonari inoperabili trattate con termoablazione mwMarco Zaccaria
 
Microwave Thermal Ablation For Hepatocarcinoma Six Liver Transplantation Cases
Microwave Thermal Ablation For Hepatocarcinoma Six Liver Transplantation CasesMicrowave Thermal Ablation For Hepatocarcinoma Six Liver Transplantation Cases
Microwave Thermal Ablation For Hepatocarcinoma Six Liver Transplantation CasesMarco Zaccaria
 
Tesi di Laurea*Facoltà di Scienze Matematiche Fisiche e Naturali*Corso di Lau...
Tesi di Laurea*Facoltà di Scienze Matematiche Fisiche e Naturali*Corso di Lau...Tesi di Laurea*Facoltà di Scienze Matematiche Fisiche e Naturali*Corso di Lau...
Tesi di Laurea*Facoltà di Scienze Matematiche Fisiche e Naturali*Corso di Lau...Marco Zaccaria
 

Mehr von Marco Zaccaria (12)

Osteoid Osteomas
Osteoid OsteomasOsteoid Osteomas
Osteoid Osteomas
 
EZ Trocar Hakko Medical JP
EZ Trocar Hakko Medical JPEZ Trocar Hakko Medical JP
EZ Trocar Hakko Medical JP
 
Endo Relief
Endo ReliefEndo Relief
Endo Relief
 
Menzione di merito progetto hs amica
Menzione di merito progetto hs amicaMenzione di merito progetto hs amica
Menzione di merito progetto hs amica
 
2013 poggi anticancer research mw ablation of hcc using a new percutaneous de...
2013 poggi anticancer research mw ablation of hcc using a new percutaneous de...2013 poggi anticancer research mw ablation of hcc using a new percutaneous de...
2013 poggi anticancer research mw ablation of hcc using a new percutaneous de...
 
2013 pusceddu jvir treatment of_bonemetastaseswithmicrowavethermal ablation
2013 pusceddu jvir treatment of_bonemetastaseswithmicrowavethermal ablation2013 pusceddu jvir treatment of_bonemetastaseswithmicrowavethermal ablation
2013 pusceddu jvir treatment of_bonemetastaseswithmicrowavethermal ablation
 
Lauro.wcio2011 ny
Lauro.wcio2011 nyLauro.wcio2011 ny
Lauro.wcio2011 ny
 
Steve Jobs
Steve JobsSteve Jobs
Steve Jobs
 
Na int pascale lesioni polmonari inoperabili trattate con termoablazione mw
Na int pascale lesioni polmonari inoperabili trattate con termoablazione mwNa int pascale lesioni polmonari inoperabili trattate con termoablazione mw
Na int pascale lesioni polmonari inoperabili trattate con termoablazione mw
 
Credo
CredoCredo
Credo
 
Microwave Thermal Ablation For Hepatocarcinoma Six Liver Transplantation Cases
Microwave Thermal Ablation For Hepatocarcinoma Six Liver Transplantation CasesMicrowave Thermal Ablation For Hepatocarcinoma Six Liver Transplantation Cases
Microwave Thermal Ablation For Hepatocarcinoma Six Liver Transplantation Cases
 
Tesi di Laurea*Facoltà di Scienze Matematiche Fisiche e Naturali*Corso di Lau...
Tesi di Laurea*Facoltà di Scienze Matematiche Fisiche e Naturali*Corso di Lau...Tesi di Laurea*Facoltà di Scienze Matematiche Fisiche e Naturali*Corso di Lau...
Tesi di Laurea*Facoltà di Scienze Matematiche Fisiche e Naturali*Corso di Lau...
 

Hs amica apparatus for mi crowave ablation

  • 1. The new frontier in Microwave thermoablation Technical Presentation
  • 2.
  • 3. Indice Index • Principi di termoablazione a microonde • Principles of microwave thermoablation • Microonde vs RF • MWs vs RF currents • Recenti sviluppi nei sistemi per termoablazione a microonde: AMICA • Recent enhancements in MW ablation systems: AMICA
  • 4. Termoablazione Thermoablation Necrosi coagulativa delle cellule indotta da un aumento locale della temperatura Coagulative necrosis of cells induced by a local temperature increase Morte cellulare pressochè istantanea per T >60°C. A temperature più basse (dell’ordine di 50°C) si ottiene lo stesso effetto per esposizioni prolungate al calore (dell’ordine di almeno 5/10 minuti). 4 Almost instantaneous cell death for T >60°C. At lower temperatures (about 50°C) the same effect is obtained for longer exposures (about 5/10 minutes).
  • 5. Requisiti fondamentali di un sistema termoablativo Basic requirements for a thermoablative system Controllo sulla figura di necrosi • Prevedibilità del risultato terapeutico (forma e dimensioni del volume ablato) in condizioni operative assegnate (tipo di tessuto-bersaglio, tipo e dose di energia somministrata, modalità e durata dell’erogazione di energia) • Prestazioni coagulative minimali e massimali (modellabilità della termolesione) • Confinamento (sfericità, preservazione di tessuti adiacenti al bersaglio, assenza di interferenze con eventuali dispositivi medici impiantati o protesi) Velocità di riscaldamento • Breve durata del trattamento • Bassa sensibilità agli effetti di sottrazione del calore (ad es. heat sinking dovuto al circolo san- guigno o ai dotti biliari) Sicurezza • Minimo grado di invasività, basso tasso di complicanze intra- e post-operatorie, compatibilità EMC e sicurezza elettrica, disponibilità di allarmi e protezioni hardware e software Facilità d’uso • Curva d’apprendimento breve, compatibilità piena con le principali tecniche di imaging, accesso facile alla lesione, controllabilità in tempo reale dell’andamento del trattamento Costi accessibili 5 • Control over coagulation pattern • Predictability of treatment outcome (shape and size of coagulation pattern) for a given set of working parameters (type of target tissue, type and quantity of energy administered, modality and duration of energy deposition) • Capability of obtaining both big or small lesions (max/min lesion diameter/length) • Confinement (sphericity, sparing of healthy tissues in the neighbourhoods of target tissues, no in- terference with implanted medical devices) • Heating velocity • Short treatment duration • Low sensitivity to heat sinking effects (blood vessels, biliary ducts) • Safety of use • Low degree of invasiveness, low rate of intra- and post-operative complications, EMC compliance, availability of hardware and software alarms and protections • Ease of use • Fast learning curve, full compatibility with available imaging technologies, easy access to lesion, real time control of treatment progress • Cost effectiveness
  • 6. Scelta dello strumento ablativo più appropriato Choice of most appropriate ablative tool ͖ Correnti a radiofrequenza (RFs) Microonde (MWs) Azione EM Laser Elettroporazione irreversibile ͖ Ultrasuoni focalizzati ad alta intensità (HIFU) Azione meccanica Crioablazione ͖ Azione termica ͖ Iniezione percutanea di etanolo (PEI) Chemoembolizzazione trans-arteriale Azione chimica (TACE) 6 ͖ Radiofrequency currents (RFs) Microwave radiations (MWs) EM action Laser Irreversible Electroporation High Intensity Focused Ultrasound (HIFU) ͖ Mechanical action Cryoablation ͖ Thermal action ͖ Percutaneous ethanol injection (PEI) Transarterial Chemoembolization (TACE) Chemical action
  • 7. Termoablazione EM EM Thermoablation Ad oggi, il riscaldamento elettromagnetico ancora appare il miglior compromesso tra esigenze molto diverse: • controllo sulla geometria della termolesione indotta • sicurezza e semplicità d’uso • accessibilità dei costi Tuttavia… Le varie sorgenti di energia EM disponibili (RF, MW, Laser) differiscono enormemente l’una dall’altra nel modo in cui interagiscono con i tessuti biologici e nella performance coagulativa complessiva. 7 At present, electromagnetic heating still seems to offer the best trade off among very different require- ments: • control over the geometry of the induced necrosis • safety and ease of use • cost effectiveness However… The various available EM energy sources (RFs, MWs, Laser) greatly differ one from the other in the way they interact with biological tissues and in their final coagulative performance.
  • 8. Efficienza del riscaldamento EM EM heating effectiveness Ai fini dell’efficacia del riscaldamento da fonti EM occorre individuare il miglior compromesso tra due parametri inversamente proporzionali: Pv (potenza assorbita per unità di volume) α frequenza Dp (profondità di penetrazione) α 1 / frequenza For maximum EM heating effectiveness the best trade off must be found between two inversely proportional parameters: Pv (power absorbed per unit volume) α frequency Dp (depth of penetration) α 1 / frequency 8 Tasso d’assorbimento dell’energia EM da parte delle molecole d’acqua EM energy absorption rate of water
  • 9. Quale fonte d’energia EM? Which EM energy source? Radiazioni non ionizzanti • Laser • Correnti a radiofrequenza (RFs) • Microonde (MWs) Tuttavia… • Per il trattamento di lesioni di medie e grandi dimensioni le radiazioni laser appaiono inadeguate (elevata Pv, ma bassa Dp!) • RF e MW trovano un’applicabilità clinica più estesa • Pertanto, la vera sfida tecnologica è: RFs vs MWs 9 Non ionizing radiations • Laser • Radiofrequency currents (RFs) • Microwaves (MWs) However… • For the treatment of medium and big size lesions laser radiations are inadequate (high Pv but very low Dp!) • RFs and MWs have a considerably wider clinical scope • Therefore, the true clinical and technological challenge is: RFs vs MWs
  • 10. RF vs Microonde / Meccanismo di riscaldamento RF vs Microwaves / Heating mechanism RFs MWs Le correnti RF usate nella pratica clinica (fre- Le oscillazioni dei dipoli atomici, che seguono le quenza operativa: qualche centinaio di KHz) in- alternanze del campo elettrico, generano calore ducono il riscaldamento nei tessuti per per attrito. Tale meccanismo si amplifica nelle dissipazione ohmica (effetto Joule e agitazione molecole POLARI, come l’acqua, di cui i tessuti ionica). sono molto ricchi... The radiofrequency currents (RFs) used in clinical 10 practice (operating frequency: a few hundreds The oscillations of atomic dipoles, striving to catch KHz) induce tissue heating simply through ohmic up with the continuously switching electric field, dissipation (Joule effect) and ionic agitation generates frictional heat. This mechanism is am- plified in POLAR molecules, such as water mole- cules. Biological tissues are very rich in water…
  • 11. RF vs Microonde / Velocità di riscaldamento RF vs Microwaves / Heating velocity RFs MWs Le correnti RF riscaldano per azione diretta sola- Le microonde riscaldano per azione diretta e in mente i tessuti nelle vicinanze della punta espo- simultanea l’intero volume tessutale raggiunto, sta dell’elettrodo: la propagazione del calore a senza ritardi di propagazione né disomogeneità. distanza avviene, più lentamente, per condu- zione termica, che dipende dalle proprietà fisiche Microwaves actively and simultaneously heat the e biologiche dei tessuti. whole irradiated tissue, with no propagation delay nor unhomogeneous distribution. RF currents directly heat only tissues in touch with or in close proximity of the current injecting elec- trode: heat propagation to a greater distance is mainly due to a slower thermal conduction process, depending on the physical and biological properties of the tissues. 11 Le microonde scaldano più velocemente! La velocità di riscaldamento delle microonde contrasta l’effetto di “heat sinking” del circolo sanguigno: si può operare in prossimità dei grandi vasi senza significative perdite di performance coagulativa. La nettezza dei gradienti di temperatura indotti dalle microonde porta a termolesioni dai margini altret- tanto netti. MWs heat faster! The microwave heating velocity effectively opposes the “heat sinking” due to blood circulation: one may operate very close to large vessels without sensible loss of coagulative performance. Steep temperature gradients also yield very clear-cut lesion margins.
  • 12. RF vs Microonde / Uniformità di riscaldamento RF vs Microwaves / Heating uniformity RFs MWs La circolazione delle correnti RF nel corpo del pa- Le microonde si propagano anche attraverso il ziente richiede un percorso ininterrotto e a bassa vuoto o regioni a bassa conducibilità, garan- impedenza tra l’elettrodo impiantato e le piastre tendo un notevole grado di omogeneità e ripe- di riferimento applicate al paziente: qualora il tibilità delle figure di necrosi indotte. tessuto-bersaglio sia scarsamente conduttivo o fortemente irregolare, il trattamento RF può ri- Microwaves propagate even across vacuum or sultare inefficace, o incompleto, o difficilmente poorly conductive regions, allowing sensibly controllabile e predicibile. higher homogeneity and repeatability of their heating patterns RFs circulation within the patient’s body requires a continuous, low impedance path between the implanted current injecting electrode and the grounding pad applied to the patient: whenever the target tissues are poorly electrically conductive or markedly irregular, RF treatments may prove ineffective, or incomplete, or hardly controllable 12 and predictable. Le microonde ottengono lesioni più uniformi e ripetibili! La propagazione di microonde è soggetta a meno vincoli legati alle proprietà fisiche dei tessuti e alla loro variabilità locale: le microonde garantiscono prestazioni più uniformi e ripetibili, su una più ampia gamma di tessuti. MWs provide uniform and repeatable coagulative performances! Microwaves are less affected by the constraints due to the physical properties of tissues and their local variability. As a consequence microwaves produce more repeatable and uniform performances over a wider range of tissues.
  • 13. RF vs Microonde / Confinamento RF vs Microwaves / Confinement RFs MWs Le correnti RF si disperdono attraverso il corpo I radiatori a microonde hanno un raggio d’azione del paziente lungo percorsi non predicibili; per- intrinsicamente confinato: le onde EM decadono tanto non dovrebbero subire trattamenti RF pa- esponenzialmente con la distanza dal punto di zienti con pacemakers, protesi metalliche o altri emissione. Le antenne a microonde convenzio- dispositivi impiantati. Inoltre, l’esito di un’abla- nali esibiscono una figura di radiazione ellissoi- zione RF può cambiare apprezzabilmente in fun- dale attorno alla parte emettitrice. zione della posizione delle piastre di dispersione. RF currents disperse through the patient’s body Microwave radiators exhibit an intrinsically con- along unpredictable paths. For this reason, RF ab- fined range of action: indeed, EM waves amplitude lation should not be performed on patients bear- falls exponentially with increasing distance from ing pacemakers or other current-sensitive devices the emitting point. Conventional microwave an- or metal prosthesis. Also, the outcome of an RF tennas feature an ellipsoidal radiation pattern treatment may markedly change depending on around their emitting tips. the grounding pads placement. 13 DELOCALIZED ACTION CONFINED ACTION Le microonde sono più sicure e meglio confinate! Le microonde agiscono solo sul tessuto-bersaglio senza risentire o interferire con altre parti del paziente al di fuori della lesione. MWs are inherently safer and more “on the target”! MWs provide a truly target-focussed ablative tool, not affected by nor interfering with parts of the patient’s body apart from the target lesion.
  • 14. RF vs Microonde / Bipolarità RF vs Microwaves / Bipolarity RFs MWs Nella maggior parte dei casi, i sistemi per abla- I radiatori MW sono forzatamente bipolari: non zione RF sono monopolari (cioè, un solo elet- sono richieste piastre di dispersione, eliminando trodo impiantato in funzione di iniettore di così le criticità a queste associate (ustioni cuta- corrente e una o più piastre di dispersione ap- nee, incertezze sul posizionamento, allunga- plicate al paziente. mento dei tempi di preparazione del paziente all’intervento). In the majority of cases, RF ablative systems are monopolar (that is, just one implanted electrode Microwave radiators are bipolar in nature: serving as current injector and one or more dis- grounding pads are not required!!! persing electrodes applied to the patient). All problems arising from the pads (inappropriate positioning, skin burns, longer procedure set-up time) are automatically overcome. 14 Gli applicatori MW sono intrinsicamente bipolari! Gli applicatori a MW sono antenne, intrinsecamente bipolari, che non necessitano quindi di piastre di dispersione né risentono dei problemi legati al posizionamento delle stesse. MW applicators are intrinsically bipolar! MW applicators are EM antennas, inherently bipolar, thus needless of dispersing electrodes and unaffected by grounding pad-related issues.
  • 15. RF vs Microonde / Versatilità d’impiego RF vs Microwaves / Versatility RFs MWs Le possibilità di utilizzo “fuori standard” della tec- La tecnologia MW offre una varietà illimitata di nologia RF sono fortemente limitate: setup operativi: Non è possibile impiegare elettrodi RF alimentati Si può utilizzare un numero arbitrario di appli- simultaneamente, a causa delle reciproche in- catori MW simultaneamente attivi, senza rischi terferenze che ne scaturirebbero: si deve ricorrere o vincoli, accrescendo le dimensioni della termo- a sistemi ad alimentazione alternata (switching), lesione ottenibile nel medesimo tempo d’inter- con tempi d’intervento più lunghi vento. Un applicatore RF è semplicemente un elettrodo Teoricamente, utilizzando una schiera di antenne che inietta corrente: si dispone di pochi parame- in coerenza di fase (phased array) si può ablare tri operativi e di design su cui agire per modifi- una regione tessutale di forma e localizzazione care forma ed estensione della figura di arbitrarie. riscaldamento qualora se ne presenti la necessità Virtualmente, esiste uno specifico design d’an- in particolari applicazioni. tenna per ottenere qualsivoglia figura di riscal- damento. Out of standard operability is very limited with RF technology: MW technology provides an unlimited range of Multiple RF electrodes may not be used simulta- possible operative setups: 15 neously, due to possible mutual interferences and An arbitrary number of MW radiators may be used loss of control over current paths: switching meth- at once, without risks or limitations. Big or multi- ods must be used, which require longer treatment ple lesions in a short time! times In principle, by operating a phased MW antennas An RF applicator is simply an electrode injecting array, one may selectively burn out a volume of current through single or multiple active tips: very tissue of arbitrary shape, size and location. few design and operative parameters are available Specific antenna designs may be used to attain to modify the final heating pattern in order to virtually any coagulative pattern match special requirements For a given antenna design and a given energy deposition rate, the lesion dimensions may be made to vary within a sufficiently wide range of values simply by changing the treatment duration Gli applicatori MW possono essere progettati per operare individualmente o in schiera e otte- nere termolesioni di forma e dimensioni arbitrarie! MW applicators may be designed to work individually or in group so as to provide thermal le- sions of arbitrary shape and size!
  • 16. Le ragioni del predominio RF Why RFs are still leading • Minor costo e complessità rispetto alla tecnologia MW. L’erogazione sicura di potenza MW per usi medicali richiede la conversione dai comuni magnetron (che operano ad elevate tensioni AC) alla più sofisticata tecnologia a stato solido (speciali transistor, alimentati a basse tensioni DC), decisa- mente più costosa, sebbene le cose ultimamente stiano cambiando… • Miglior controllo sulla figura di necrosi. Le MW risentono di fenomeni di retro-riscaldamento (EF- FETTO COMETA) dovuto sia alla presenza di radiazione non assorbita dai tessuti-bersaglio e che si propaga a ritroso lungo le pareti esterne dell’applicatore),sia alle rilevanti perdite di potenza in calore lungo il cavo coassiale che alimenta l’antenna (con il rischio di surriscaldamento dell’applicatore). • Lesser cost and complexity with respect to MW technology. Safe delivery of MW power in clinical treat- ments requires conversion from cheap magnetron sources (which are operated at high AC voltages) to solid state technology (transistors, fed at low DC voltages). The latter technology is definitely more ex- pensive, though things have started changing lately… • Better control over coagulation pattern. MWs are affected by back heating phenomena (COMET 16 EFFECT) both due to poor control over reflected power (that is, the portion of MW power unabsorbed by the target-tissue and propagating backwards along the antenna shaft) and due to intense power losses along the antenna feeding cable (turning into severe shaft overheating).
  • 17. Applicatori MW convenzionali Conventional MW probes 17 I comuni rimedi al fenomeno di retro-riscaldamento (sia dovuto alle riflessioni che alle perdite ohmiche lungo la lina coassiale) comportano un considerevole incremento del calibro degli ap- plicatori MW. In alternativa, si possono utilizzare applicatori MW sottili, ma a potenze limitate. In pratica, fino a ieri si aveva: Calibro applicatori RF: 14-17G Calibro applicatori MW: 11-13G Common technical remedies to back heating effects (both due to reflected waves and due to ohmic dissipation along the feeding coaxial line) bring a remarkable increase in MW applicators gauge. Alternatively, slender MW probes may be used, but with limited power output. In practice, till yesterday: RF applicators gauge: 14-17G MW applicators gauge : 11-13G
  • 18. In sintesi… Summarizing… Se si potesse garantire per gli applicatori MW, contemporaneamente: • un piccolo calibro, ai fini della minima invasività • un efficace sistema di intrappolamento delle microonde • un adeguato sistema di raffreddamento dell’antenna allora le MW potrebbero realmente candidarsi a sostituire le RF nel ruolo di tecnologia termoablativa d’elezione, in quanto le MW assicurano: • Un riscaldamento più rapido • Accresciuta affidabilità e versatilità d’impiego • Performance coagulativa più omogenea e ripetibile Per queste ragioni la comunità medica mondiale ha grandi aspettative riguardo le MW. Negli ultimi anni, diversi gruppi industriali hanno raccolto la sfida tecnologica finalizzata allo sviluppo di nuovi e migliori sistemi per ablazione MW con tutti i necessari requisiti di sicurezza, efficienza e accessibilità dei costi. 18 If one could warrant for microwave applicators, at once: • a small gauge, so as to attain minimum invasiveness • an effective system for microwaves trapping • a suitable antenna cooling system then microwaves would certainly challenge RFs supremacy in the field of thermoablation, since mi- crowaves attain: • Faster heating • Enhanced safety and versatility of use • Homogeneous and repeatable coagulative performance For these reasons, the medical community worldwide has great expectations concerning mi- crowaves. In the last few years, a technological challenge has started among several important in- dustrial groups in different corners of the world, in order to develop new microwave systems finally fulfilling all requirements for a safe, efficient and cost effective clinical use.
  • 19. AMICA AMICA-PROBE Applicatore interstiziale monouso ad introduzione diretta, dotato di dispositivo miniaturizzato per il confinamento delle MW (BREVETTO MONDIALE!), circuito di raffreddamento, termocoppia integrata e memoria digitale. AMICA-GEN Generatore di microonde (2450 MHz/100W), programmabile, a stato solido. AMICA-PUMP Pompa peristaltica per il raffreddamento dell’applicatore, interamente automatizzata. 19 AMICA-PROBE Disposable interstitial applicator, internally cooled, featuring a patented miniaturized device for microwaves confinement (WORLDWIDE PATENT!), an integrated thermocouple for probe temperature monitoring and a digital memory chip. AMICA-GEN Programmable, solid state microwave source (2450 MHz/100W). AMICA-PUMP Fully automated peristaltic pump for probe cooling.
  • 20. 20
  • 21. Struttura di AMICA-PROBE AMICA-PROBE structure • “Punta calda”: l’antenna viene raffreddata per tutta la sua lunghezza, ad eccezione della porzione distale (~ 2.5 cm) • Rivestimento in teflon anti-aderente • Punta attiva di lunghezza assegnata (~ 2 cm): le dimensioni delle termolesioni indotte cambiano modificando opportunamente potenza erogata e durata del trattamento. Il concetto di “punta esposta” (proprio degli elettrodi RF) non si applica più. 21 • “Hot tip”: the antenna is cooled throughout its length, except for its distal end (~ 2.5 cm) • Anti-sticking teflon coating • Active tip of assigned length (~ 2 cm): lesion dimensions may be varied by changing delivered power and treatment duration. The “exposed tip” concept is no longer applicable.
  • 22. AMICA-PROBE: performance coagulativa AMICA-PROBE: coagulative performance Dati ex vivo su fegato bovino a temperatura ambiente Ex vivo data on bovine liver at room temperature Mean aspect ratio (D/L) = 0.8 22
  • 23. AMICA-GEN • Sorgente a stato solido (transistor alimentati a 12 VDC) • Potenza disponibile: 100W CW • Frequenza: 2450 MHz • Interfaccia-utente essenziale e intuitiva • Architettura digitale : massima versatilità! • Interattività: collegabile a un PC e a un numero illimitato di altre periferiche • Massima sicurezza: sofisticato sistema di allarmi e protezioni 23 • Solid state source (transistors fed at 12 VDC) • Available power: 100W CW • Frequency: 2450 MHz • Essential and intuitive user interface • Digital architecture: maximum versatility! • Interactivity: connectable to PC and to an unlimited number of external peripherals • Maximum safety: sophisticated system of alarms and protections
  • 24. Feedback in tempo reale dalla lesione Real time feedback from lesion 24 S11 è il coefficiente di riflessione, cioè il rapporto tra potenza a MW riflessa e potenza erogata (in valore assoluto). S11 is the MW probe reflection coefficient, that is the ratio between reflected and direct MW power (in absolute value). AMICA-GEN consente il monitoraggio continuo del coefficiente di riflessione e della tempe- ratura dell’applicatore. La lesione “ci parla”… AMICA-GEN allows for continous monitoring of probe temperature and reflection coefficient. The lesion “speaks out”…
  • 25. HS AMICA: riassumendo… HS AMICA: summarizing… Frequenza operativa 2450 MHz Operating frequency 2450 MHz Max potenza all’applicatore 70 W Max available power on probe 70 W Calibri degli applicatori 11G, 14G , 16G Applicator gauge 11G, 14G , 16G Mini-choke (brevetto mondiale) Controllo dell’effetto-cometa Patented mini-choke Back heating control Raffreddamento interno Internal probe cooling Controllo della temperatura dell’applicatore Termocoppia integrata nell’applicatore Probe temperature control Embedded Thermocouple Max. lesione (ex vivo su fegato bovino, Più di 4 cm di diametro in 10 min con ago singolo in un solo passaggio) Max. lesion diam. (ex vivo bovine liver, More than 4 cm in 10 min single probe, single insertion) Accessori nel kit monouso Mini-bisturi per la pre-incisione della cute; introduttore coassiale; termocoppia 25 interstiziale; sacca di raccolta del refrigerante refluo; telo Probe kit accessories Mini-scalpel for skin pre-incision; coaxial introducer; interstitial thermocouple; integrated refluent coolant collector; drape Applicatori multipli Possibile per un numero arbitrario di applicatori in simultanea, a qualunque distanza reciproca Multiple probes operation Possible, with any number of probes operated simultaneously, at any reciprocal distance Generatore Architettura digitale “aperta” : interfaccia- utente interattive e personalizzabile. Monitoraggio continuo di temperatura e riflessioni MW. Generator “Open” digital architecture: interactive and customizable user-interface. Continuous monitoring of probe temperature and MW reflections. Certificazioni CE, FDA Approvals CE, FDA
  • 26. Caratteristiche esclusive di AMICA Unique features of AMICA • Applicatori sottili (finanche 16G) • Ampia varietà di applicatori (11G per ablazione ossea o per intra-operatorio; 14G e 16G per uso percutaneo; applicatori “lunghi” per laparoscopia, e “corti” per uso sotto guida TAC; applicatori flessibili presto disponibili!) • Elevata potenza disponibile (70W netti all’applicatore) • Grandi termolesioni (oltre 4cm di diametro, con applicatore singolo in 10 minuti) • Controllo ottimale di forma e dimensioni del coagulo. Niente effetto-cometa (mini-choke per la soppressione delle MWs riflesse e sistema di raffreddamento integrato per prevenire il surriscalda- mento dello stelo) • Monitoraggio costante delle riflessioni e della temperatura dell’applicatore, che fornisce un feedback in tempo reale sull’andamento del trattamento • Chip di memoria integrato per un’accresciuta sicurezza d’uso (auto-limitazione della massima po- tenza erogabile ad un dato applicatore, auto-diagnosi di integrità e scadenza dell’applicatore) • Generatore ad architettura digitale aperta : interfaccia-utente intuitiva e altamente interattiva, personalizzabile a piacere; salvataggio dei dati operativi e collegabilità a PC; • Pompa peristaltica completamente automatizzata 26 • Small applicator gauge (down to 16G) • Wide variety of probes (11G for bone ablation or for intra-surgical treatments; 14G and 16G for in- terstitial ablations; oversized lengths for laparascopic treatments, undersized lengths for CT-guided procedures; flexible probes coming soon!) • High available power (70W net at applicator input section) • Big lesions (over 4cm in diameter, with single applicator in 10 minutes) • Optimal control on the coagulative pattern size and shape. No back-heating (mini-choke for reflected MWs trapping and integrated cooling system to avoid shaft overheating) • Constant monitoring of MW reflections and probe temperature, providing real time feedback on treatment progression • Built-in memory chip for enhanced safety of use (self-limitation of maximum delivered power de- pending on applicator characteristics, self-diagnosis of probe expiration and integrity) • Open-architecture, digital microwave generator: very intuitive and highly interactive user interface, infinitely customizable; ablation data storage and PC link available; • Fully automated peristaltic pump
  • 27. 2450 MHz vs 915 MHz 2450 MHz vs 915 MHz Calcolo dell’intensità di campo elettrico (figura dx) e del tasso d’assorbimento specifico (figura sx) in fantoccio muscolo-equivalente in funzione della distanza da dipoli elementari rispettivamente operati a 915MHz (curve rosse) e a 2450 MHz (curve blu). A parità di potenza erogata, il campo a 2450 MHz è nettamente superiore al campo a 915 MHz fino a 4cm di distanza dall’antenna: come a dire che in tutto il volume di un nodulo fino a 8 cm di diametro l’azione di un applicatore a 2450 MHz è marcata- mente più efficace rispetto ad un applicatore a 915 MHz. (A cura del Dipartimento di Ingegneria Elettronica dell’Università degli Studi di Roma “La Sapienza”) 27 Calculation of the electric field intensity (on the right) and of the specific absorption rate (on the left) in a muscle-equivalent phantom as functions of the distance from elementary dipoles, respectively operated at 915MHz (red curves) and at 2450 MHz (blue curves). The 2450 MHz field turns out to be definitely higher than the 915 MHz field up to 4cm away from the antenna. It means that in lesions up to 8 cm of diameter the efficacy of a 2450 MHz applicator is higher than that of a 915 MHz applicator. (Courtesy of the Electronic Engineering Department of the University of Rome “La Sapienza”)
  • 28. AMICA: Clinical Results - Liver ablation (HCC) 28
  • 29. 29
  • 30. Hepatic metastases from lungs Single lesion: 54x48 mm Courtesy of Dr. M. F. Meloni, S. Gerardo Hospital, Monza, Italy 30 Multiple metastases from myosarcoma
  • 31. Hepatic metastases from lungs (US-guided laparoscopy) Courtesy of Dr.Santambrogio, S. Giuseppe Hospital, Milan, Italy 31 HCC / Pre- and post-ablation CT images Courtesy of Dr. L. Solbiati, Hospital of Busto Arsizio, Italy
  • 32. HCC / Pre- and post-ablation CT images Courtesy of Dr. L. Solbiati, Hospital of Busto Arsizio, Italy 32 HCC / Pre- and post-ablation CT images Courtesy of Dr. L. Solbiati, Hospital of Busto Arsizio, Italy
  • 33. Phase I clinical trial on AMICA (safety and tolerability): (the study was published on the Journal of Endourology in July 2008) “TRANS PERINEAL MICROWAVE THERMOABLATION IN PATIENTS WITH OBSTRUCTIVE BENIGN PROSTATIC HYPERPLASIA: A PHASE I CLINICAL STUDY WITH A NEW MINI-CHOKED MICROWAVE APPLICATOR”. Bartoletti R.1, Cai T. 1, Pinzi N. 2, Longo I. 3, Ricci A. 3, Massaro MP. 4, Tosoratti N. 4 1- Department of Urology, University of Florence, Italy 2- Urologic Unit, Campo di Marte Hospital - Lucca, Italy 3- Institute Chemo-Physical Processes, CNR - Pisa, Italy 4- HS Hospital Service SpA - Aprilia (LT), Italy 33 RESULTS (20 W / 5 min)
  • 34. RESULTS (20 W / 5 min) Pathological analyses showed a quasi-spheroidal lesion (longitudinal axis: 18 mm; transversal axis: 16 mm), characterized by a central coagulative necrosis surrounded by an inflammatory infiltrate reaction (thickness: 2.1 mm), beyond which healthy tissue was found. 34 CONCLUSIONS The AMICA-PROBE treatment is a safe, well tolerated and repeatable treatment for benign prostatic hypertrophy. Further clinical phase II-III studies are needed in order to evaluate its efficacy and morbidity in comparison with classical TUMT or trans urethral resection of the prostate.
  • 35.
  • 36. HS HOSPITAL SERVICE S.P.A Via Zosimo 13 - 00178 Roma, ITALY Stabilimento di produzione/Factory: Via Angela Vacchi 24/26 04011 Aprilia (Latina), ITALY Tel.: +39 06 9201961 Fax: +39 06 92727871 email: hs@hshospitalservice.com www.hshospitalservice.com