ELECTRON BEAM THERAPY
DEFINITION
It is a type of particulate beam
radiation therapy used for the
treatment of Superficial tumors
BEAM
PHOTON PARTICULATE
X ray γ ray
α,β,proton,e,He,C
WHY ELECTRON BEAM
 No exit dose like X-ray
 Delivery of reasonably uniform dose from skin to a specific
depth
 abrupt dose fall off after 90% to 80% isodose curve to near
zero level
Photon BeamPit Falls we want
 Exit dose
 No 100% Skin sparing
 No 100% skin dose
 No 100% tumor control
 No normal tissue sparing
 Side scatter
 Low LET
 High OER
 Low RBE
 No exit dose
 100% Skin sparing
 100% skin dose
 100% tumor control
 Normal tissue sparing
 No Side scatter
 High LET
 Low OER
 High RBE
PHOTON vs. ELECTRON
 Exit dose
 Near 100% Skin sparing
 No 100% skin dose
 High penetration
 No DOSE uniformity
 No exit dose
 near 100% skin dose
 abrupt dose fall off after
80% isodose curve
 High scatter
 Dose uniformity
MACHINES
 Van De Graff generator.
 Betatron
 Linear accelerator – a device that uses
high frequency EM waves to accelerate
charged particles viz. electrons to a high
energy through a linear tube
Mechanism of production
X-ray Mode/ electron Mode
Parts of treatment head
1. Primary Collimator
2. Scattering foil
3. Flattening filter
4. Ionization chamber
5. Secondary collimator
6. Tertiary collimator ( trimmer , electron
cone, auxiliary collimator )
SCATTERING FOIL
1. The scattering foils are made up of high
dense material interposed in electron beam
2. Different scattering foils are used for
different e- energy
3. The electron beam edges can be sharply
defined only if the collimator extended
towards the skin of the pt by attachment of
trimmers/applicators
4. The electron beam trimmers are optimally
designed to give uniform fluency
ELECTRON BEAM FROM
MACHINE TO PATIENT
 The energetic electron beam emerging from
accelerators is pencil beam.it is not suitable for
treatment
 Hence the beam has to be spread to a larger area
for treatment.
 Spread can be done by- electromagnetic scattering
device or scattering foil
 To further spreading electron applicator is used
Mechanism of interaction of electron with matter
 Inelastic collision with atomic electron
 Inelastic collision with nuclei
 elastic collision with atomic electron
 elastic collision with nuclei
Fundamental questions before EBT
 Indication of EBT
 Goal of EBT
 Planned Treatment Volume
 Planned Treatment dose
 Planned Treatment Technique
Indications
 Treating skin cancers.
 Chest wall irradiation in Ca breast.
 Boosting to neck node after 45 Gy.
 Total skin irradiation in mycosis fungoides.
 Total limb irradiation
 Total scalp irradiation
 Craniospinal irradiation
 Intracavitary irradiation
Dosimetry
 Measured by pencil beam algorithm
 Accurate in water at standard and extended
SSD
 Correctly predicts changes in Penumbra
 Predict changes in dose in oblique incidence
or irregular surfaces
Isodose Curves
Isodose curve characteristic
 Rapid dose falloff below 80% isodose
 Bulging towards bottom
 Ballooning towards edge
Isodose curve in slopping
surface
Determination of absorbed dose-
 Calorimeter
 TLD
 Solid state diode
 Ionization chamber
Depth dose characteristic/isodose
characteristic
 There is a abrupt fall of doses beyond 90% to
80% of isodose curve
 80% or 90% isodose curve is taken as standard
dose for prescription
 Depth (cm)of 90% isodose=E(Mev)/4
 Depth (cm)of 80% isodose=E(Mev)/3
 Electron beam penetrates a finite depth with clear
cut range,the photon beam in other hand proceeds
infinite range.
Dose dependence on incident energy
Dose dependence on field size
Depth Dose dependence on SSD
dmax
 Maximum dose is not on skin
 It is somehow away from skin
 It is due to electronic equilibrium
 6Mev-10mm
 9Mev-15mm
 12Mev-19mm
Approximation
Range of depth
 Roughly E/2
 E=energy
 E.g –for 6Mev range is 6/2=3cm
Surface dose in 10x10 cm field
 For 5mev=74%
 10mev=82%
 16mev=93%
 25mev=96%
 It depends upon energy,field size and
thickness of scattering foil
Energy range
 6-20 Mev
6,9,12,15,18,20 mev
Selection of energy
Ep0(MeV) = 3.3 x R90 (cm) [R90 exceeds max
depth of PTV]
Ep0(MeV) = 2 x Rp [Rp is the practical range
of electrons ]
Ideal condition for EBT
 Electron beam incidence normal to flat
surface
 Underlying homogenous soft tissue
(provide uniform dose in penumbra from
surface to R90 after which there is rapid
dose fall off)
USE OF BOLUS
 To flatten out irregular surface
 Reduce the penetration of electrons in part of the field
 Increase the surface dose
 To act as a missing tissue compensator
 Thickness of bolus required increases as the surface dose
increases e.g
– 6-10mev=1cm
– 10-15mev=0.5cm
– >15mev=no bolus
BOLUS MATERIAL
 Paraffin wax,polysterene,lucite,superstoff,superflab
 Flexible bolus that confirms to surface is desirable
 Large air gap between the absorber and the surface
would result in scattering of e- outside the field and
reduction in dose that may not be easily predictable
unless specially measured for this condition
MLC vs. BLOCK
 No MLC
 Wooden MLC
 Wax MLC
 Customized block
BLOCKING MATERIAL
 Usually Cerroband
 Cerroband-Pb+Bi+Sn+Cu
 Melting point=
 Customized block
 Thickness of block required to use as block
=E/2 e.g for 6mev=6/2=3cm
Monitor unit
CGy/MU
Decelerators
 A plate of low atomic wt. material such as
lucite and polysterene is sometimes used to
reduce the energy of electron beam known
as decelerators
 It must be placed in close contact with
patient surface with bolus
Electron arc
therapy
Electron field-photon field
combination
Total skin irradiation-
Total scalp irradiation
Craniospinal irradiation
Total limb irradiation
ELECTRON BEAM RADIOTHERAPY

More Related Content

What's hot

Electron Beam Therapy
Electron Beam TherapyElectron Beam Therapy
Electron Beam Therapy
ajjitchandran
 
4D Scan and Respiratory Gating
4D Scan and Respiratory Gating4D Scan and Respiratory Gating
4D Scan and Respiratory Gating
Kothanda Raman
 
Beam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principlesBeam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principles
Santam Chakraborty
 

What's hot (20)

Treatment plannings i kiran
Treatment plannings i   kiranTreatment plannings i   kiran
Treatment plannings i kiran
 
Total body irradiation
Total body irradiationTotal body irradiation
Total body irradiation
 
TSET
TSETTSET
TSET
 
Icru 50,62,83 volume deliniation
Icru 50,62,83 volume deliniationIcru 50,62,83 volume deliniation
Icru 50,62,83 volume deliniation
 
Arc therapy [autosaved] [autosaved]
Arc therapy [autosaved] [autosaved]Arc therapy [autosaved] [autosaved]
Arc therapy [autosaved] [autosaved]
 
Electron Beam Therapy
Electron Beam TherapyElectron Beam Therapy
Electron Beam Therapy
 
Electron beam radiotherapy
Electron beam radiotherapyElectron beam radiotherapy
Electron beam radiotherapy
 
Concept of bed in radiobiology
Concept of bed in radiobiologyConcept of bed in radiobiology
Concept of bed in radiobiology
 
Srs and sbrt 2 dr.kiran
Srs and sbrt 2 dr.kiranSrs and sbrt 2 dr.kiran
Srs and sbrt 2 dr.kiran
 
Icru 50
Icru 50Icru 50
Icru 50
 
ICRU CONCEPT
ICRU CONCEPTICRU CONCEPT
ICRU CONCEPT
 
4D Scan and Respiratory Gating
4D Scan and Respiratory Gating4D Scan and Respiratory Gating
4D Scan and Respiratory Gating
 
Electron beam therapy
Electron beam therapyElectron beam therapy
Electron beam therapy
 
The vmat vs other recent radiotherapy techniques
The vmat vs other recent radiotherapy techniquesThe vmat vs other recent radiotherapy techniques
The vmat vs other recent radiotherapy techniques
 
Lung sbrt ppt
Lung  sbrt pptLung  sbrt ppt
Lung sbrt ppt
 
EPID AND CBCT ON RADIATION THERAPY
EPID AND CBCT ON RADIATION THERAPYEPID AND CBCT ON RADIATION THERAPY
EPID AND CBCT ON RADIATION THERAPY
 
Electron beam therapy
Electron beam therapyElectron beam therapy
Electron beam therapy
 
Gap correction
Gap correctionGap correction
Gap correction
 
Mind the Gap: Dealing with Interruptions in Radiotherapy Treatment
Mind the Gap: Dealing with Interruptions in Radiotherapy TreatmentMind the Gap: Dealing with Interruptions in Radiotherapy Treatment
Mind the Gap: Dealing with Interruptions in Radiotherapy Treatment
 
Beam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principlesBeam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principles
 

Similar to ELECTRON BEAM RADIOTHERAPY

Best therapy for small superficial treatment areas
Best therapy for small superficial treatment areasBest therapy for small superficial treatment areas
Best therapy for small superficial treatment areas
Rosannah Smith
 
Isodose curves RADIATION ONCOLOGY
Isodose curves RADIATION ONCOLOGYIsodose curves RADIATION ONCOLOGY
Isodose curves RADIATION ONCOLOGY
Paul George
 
Radiographic exposure and image quality
Radiographic exposure and image qualityRadiographic exposure and image quality
Radiographic exposure and image quality
Rad Tech
 
Exposure factors2
Exposure factors2Exposure factors2
Exposure factors2
mr_koky
 

Similar to ELECTRON BEAM RADIOTHERAPY (20)

Electron Beam Therapy
Electron Beam TherapyElectron Beam Therapy
Electron Beam Therapy
 
Electron beam radiotherapy
Electron beam radiotherapyElectron beam radiotherapy
Electron beam radiotherapy
 
ELECTRON BEAM THERAPY
ELECTRON BEAM THERAPYELECTRON BEAM THERAPY
ELECTRON BEAM THERAPY
 
RON BEAM THEELECT
RON BEAM THEELECTRON BEAM THEELECT
RON BEAM THEELECT
 
Gowtham's 5th electrons
Gowtham's 5th   electronsGowtham's 5th   electrons
Gowtham's 5th electrons
 
Icru 71 electrons
Icru 71 electrons Icru 71 electrons
Icru 71 electrons
 
Dosimetric Evaluation of High Energy Electron Beams Applied in Radiotherapy
Dosimetric Evaluation of High Energy Electron Beams Applied in RadiotherapyDosimetric Evaluation of High Energy Electron Beams Applied in Radiotherapy
Dosimetric Evaluation of High Energy Electron Beams Applied in Radiotherapy
 
Film Image Acquisition.pptx
Film Image Acquisition.pptxFilm Image Acquisition.pptx
Film Image Acquisition.pptx
 
gr 8.pdf
gr 8.pdfgr 8.pdf
gr 8.pdf
 
Gowtham's 4th phsysics photons
Gowtham's 4th phsysics   photonsGowtham's 4th phsysics   photons
Gowtham's 4th phsysics photons
 
Best therapy for small superficial treatment areas
Best therapy for small superficial treatment areasBest therapy for small superficial treatment areas
Best therapy for small superficial treatment areas
 
ICRU 71
ICRU 71ICRU 71
ICRU 71
 
Copy of Osric · SlidesCarnival.pdf
Copy of Osric · SlidesCarnival.pdfCopy of Osric · SlidesCarnival.pdf
Copy of Osric · SlidesCarnival.pdf
 
Isodose curves RADIATION ONCOLOGY
Isodose curves RADIATION ONCOLOGYIsodose curves RADIATION ONCOLOGY
Isodose curves RADIATION ONCOLOGY
 
Tseb
TsebTseb
Tseb
 
Radiographic exposure and image quality
Radiographic exposure and image qualityRadiographic exposure and image quality
Radiographic exposure and image quality
 
Exposure factors2
Exposure factors2Exposure factors2
Exposure factors2
 
Medical uses of ionising radiation
Medical uses of ionising radiationMedical uses of ionising radiation
Medical uses of ionising radiation
 
10. laser
10. laser10. laser
10. laser
 
LINAC final.pptx
LINAC final.pptxLINAC final.pptx
LINAC final.pptx
 

More from Kanhu Charan

More from Kanhu Charan (20)

MOTION MANAGEMANT IN LUNG SBRT BY DR KANHU CHARAN PATRO
MOTION MANAGEMANT IN LUNG SBRT BY DR KANHU CHARAN PATROMOTION MANAGEMANT IN LUNG SBRT BY DR KANHU CHARAN PATRO
MOTION MANAGEMANT IN LUNG SBRT BY DR KANHU CHARAN PATRO
 
MAY 2024 ONCOLOGY CARTOONS BY DR KANHU CHARAN PATRO
MAY 2024 ONCOLOGY CARTOONS BY DR KANHU CHARAN PATROMAY 2024 ONCOLOGY CARTOONS BY DR KANHU CHARAN PATRO
MAY 2024 ONCOLOGY CARTOONS BY DR KANHU CHARAN PATRO
 
April 2024 ONCOLOGY CARTOON by DR KANHU CHARAN PATRO
April 2024 ONCOLOGY CARTOON by  DR KANHU CHARAN PATROApril 2024 ONCOLOGY CARTOON by  DR KANHU CHARAN PATRO
April 2024 ONCOLOGY CARTOON by DR KANHU CHARAN PATRO
 
TARGET DELINEATION OF THORACIC NODAL. STATION
TARGET DELINEATION OF THORACIC NODAL. STATIONTARGET DELINEATION OF THORACIC NODAL. STATION
TARGET DELINEATION OF THORACIC NODAL. STATION
 
TARGET DELINEATION IN RECTUM CANCER BY DR KANHU
TARGET DELINEATION IN RECTUM  CANCER BY DR KANHUTARGET DELINEATION IN RECTUM  CANCER BY DR KANHU
TARGET DELINEATION IN RECTUM CANCER BY DR KANHU
 
TARGET DELINEATION IN ANAL CANAL CANCER BY DR KANHU
TARGET DELINEATION IN ANAL CANAL CANCER BY DR KANHUTARGET DELINEATION IN ANAL CANAL CANCER BY DR KANHU
TARGET DELINEATION IN ANAL CANAL CANCER BY DR KANHU
 
TARGET DELINEATION IN VULVAL CANCER BY DR KANHU
TARGET DELINEATION IN VULVAL CANCER BY DR KANHUTARGET DELINEATION IN VULVAL CANCER BY DR KANHU
TARGET DELINEATION IN VULVAL CANCER BY DR KANHU
 
TARGET DELINEATION IN CERVIX CANCER BY DR KANHU
TARGET DELINEATION IN CERVIX CANCER BY DR KANHUTARGET DELINEATION IN CERVIX CANCER BY DR KANHU
TARGET DELINEATION IN CERVIX CANCER BY DR KANHU
 
Oncology cartoons by Dr Kanhu Charan Patro
Oncology cartoons by Dr Kanhu Charan PatroOncology cartoons by Dr Kanhu Charan Patro
Oncology cartoons by Dr Kanhu Charan Patro
 
RADIATION THERAPY IN BILIARY TRACT CANCER
RADIATION THERAPY IN BILIARY TRACT CANCERRADIATION THERAPY IN BILIARY TRACT CANCER
RADIATION THERAPY IN BILIARY TRACT CANCER
 
FEBRUARY 2024 ONCOLOGY CARTOON /95TH VOLUME
FEBRUARY 2024 ONCOLOGY CARTOON /95TH VOLUMEFEBRUARY 2024 ONCOLOGY CARTOON /95TH VOLUME
FEBRUARY 2024 ONCOLOGY CARTOON /95TH VOLUME
 
Molecular Profile of Endometrial cancer.
Molecular Profile of Endometrial cancer.Molecular Profile of Endometrial cancer.
Molecular Profile of Endometrial cancer.
 
ONCOLOGY CARTOONS JANUARY 2024 BY DR KANHU CHARAN PATRO
ONCOLOGY CARTOONS JANUARY 2024 BY DR KANHU CHARAN PATROONCOLOGY CARTOONS JANUARY 2024 BY DR KANHU CHARAN PATRO
ONCOLOGY CARTOONS JANUARY 2024 BY DR KANHU CHARAN PATRO
 
TYPES OF STATISTICAL DATA BY DR KANHU CHARAN PATRO
TYPES OF STATISTICAL DATA  BY DR KANHU CHARAN PATROTYPES OF STATISTICAL DATA  BY DR KANHU CHARAN PATRO
TYPES OF STATISTICAL DATA BY DR KANHU CHARAN PATRO
 
WHY STEREOTATXY IN CRANIAL AVM / DR KANHU CHARAN PATRO
WHY STEREOTATXY IN CRANIAL AVM / DR KANHU CHARAN PATROWHY STEREOTATXY IN CRANIAL AVM / DR KANHU CHARAN PATRO
WHY STEREOTATXY IN CRANIAL AVM / DR KANHU CHARAN PATRO
 
PORTAL VEIN TUMOR THROMBOSIS SBRT/DR KANHU CHRAN PATRO
PORTAL VEIN TUMOR THROMBOSIS SBRT/DR KANHU CHRAN PATROPORTAL VEIN TUMOR THROMBOSIS SBRT/DR KANHU CHRAN PATRO
PORTAL VEIN TUMOR THROMBOSIS SBRT/DR KANHU CHRAN PATRO
 
DR KANHU CHARTAN PATRO/ FOR ENT SURGEONS
DR KANHU CHARTAN PATRO/ FOR ENT SURGEONSDR KANHU CHARTAN PATRO/ FOR ENT SURGEONS
DR KANHU CHARTAN PATRO/ FOR ENT SURGEONS
 
DECEMBER 2023 ONCOLOGY CARTOONS DRKANHU CHARAN PATRO
DECEMBER 2023 ONCOLOGY CARTOONS DRKANHU CHARAN PATRODECEMBER 2023 ONCOLOGY CARTOONS DRKANHU CHARAN PATRO
DECEMBER 2023 ONCOLOGY CARTOONS DRKANHU CHARAN PATRO
 
DEBATE IN CA BLADDER TMT VS CYSTECTOMY
DEBATE IN CA BLADDER TMT VS CYSTECTOMYDEBATE IN CA BLADDER TMT VS CYSTECTOMY
DEBATE IN CA BLADDER TMT VS CYSTECTOMY
 
ROSE CASE CARDIAC ARRHYTHMIA SBRT
ROSE CASE CARDIAC  ARRHYTHMIA SBRTROSE CASE CARDIAC  ARRHYTHMIA SBRT
ROSE CASE CARDIAC ARRHYTHMIA SBRT
 

Recently uploaded

🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
Call Girls In Delhi Whatsup 9873940964 Enjoy Unlimited Pleasure
 
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
adilkhan87451
 
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
Sheetaleventcompany
 
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
chetankumar9855
 

Recently uploaded (20)

🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
 
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service AvailableCall Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
 
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
 
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
 
Call Girls Varanasi Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Varanasi Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
 
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
 
Top Rated Call Girls Kerala ☎ 8250092165👄 Delivery in 20 Mins Near Me
Top Rated Call Girls Kerala ☎ 8250092165👄 Delivery in 20 Mins Near MeTop Rated Call Girls Kerala ☎ 8250092165👄 Delivery in 20 Mins Near Me
Top Rated Call Girls Kerala ☎ 8250092165👄 Delivery in 20 Mins Near Me
 
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
 
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
 
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
 
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
 
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
 
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
 
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
 
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
 
Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...
Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...
Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...
 
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
 
Call Girls Madurai Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Madurai Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Madurai Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Madurai Just Call 9630942363 Top Class Call Girl Service Available
 

ELECTRON BEAM RADIOTHERAPY

  • 2. DEFINITION It is a type of particulate beam radiation therapy used for the treatment of Superficial tumors
  • 3. BEAM PHOTON PARTICULATE X ray γ ray α,β,proton,e,He,C
  • 4. WHY ELECTRON BEAM  No exit dose like X-ray  Delivery of reasonably uniform dose from skin to a specific depth  abrupt dose fall off after 90% to 80% isodose curve to near zero level
  • 5. Photon BeamPit Falls we want  Exit dose  No 100% Skin sparing  No 100% skin dose  No 100% tumor control  No normal tissue sparing  Side scatter  Low LET  High OER  Low RBE  No exit dose  100% Skin sparing  100% skin dose  100% tumor control  Normal tissue sparing  No Side scatter  High LET  Low OER  High RBE
  • 6. PHOTON vs. ELECTRON  Exit dose  Near 100% Skin sparing  No 100% skin dose  High penetration  No DOSE uniformity  No exit dose  near 100% skin dose  abrupt dose fall off after 80% isodose curve  High scatter  Dose uniformity
  • 7. MACHINES  Van De Graff generator.  Betatron  Linear accelerator – a device that uses high frequency EM waves to accelerate charged particles viz. electrons to a high energy through a linear tube
  • 8.
  • 11. Parts of treatment head 1. Primary Collimator 2. Scattering foil 3. Flattening filter 4. Ionization chamber 5. Secondary collimator 6. Tertiary collimator ( trimmer , electron cone, auxiliary collimator )
  • 12. SCATTERING FOIL 1. The scattering foils are made up of high dense material interposed in electron beam 2. Different scattering foils are used for different e- energy 3. The electron beam edges can be sharply defined only if the collimator extended towards the skin of the pt by attachment of trimmers/applicators 4. The electron beam trimmers are optimally designed to give uniform fluency
  • 13. ELECTRON BEAM FROM MACHINE TO PATIENT  The energetic electron beam emerging from accelerators is pencil beam.it is not suitable for treatment  Hence the beam has to be spread to a larger area for treatment.  Spread can be done by- electromagnetic scattering device or scattering foil  To further spreading electron applicator is used
  • 14. Mechanism of interaction of electron with matter  Inelastic collision with atomic electron  Inelastic collision with nuclei  elastic collision with atomic electron  elastic collision with nuclei
  • 15. Fundamental questions before EBT  Indication of EBT  Goal of EBT  Planned Treatment Volume  Planned Treatment dose  Planned Treatment Technique
  • 16. Indications  Treating skin cancers.  Chest wall irradiation in Ca breast.  Boosting to neck node after 45 Gy.  Total skin irradiation in mycosis fungoides.  Total limb irradiation  Total scalp irradiation  Craniospinal irradiation  Intracavitary irradiation
  • 17. Dosimetry  Measured by pencil beam algorithm  Accurate in water at standard and extended SSD  Correctly predicts changes in Penumbra  Predict changes in dose in oblique incidence or irregular surfaces
  • 19. Isodose curve characteristic  Rapid dose falloff below 80% isodose  Bulging towards bottom  Ballooning towards edge
  • 20. Isodose curve in slopping surface
  • 21. Determination of absorbed dose-  Calorimeter  TLD  Solid state diode  Ionization chamber
  • 22. Depth dose characteristic/isodose characteristic  There is a abrupt fall of doses beyond 90% to 80% of isodose curve  80% or 90% isodose curve is taken as standard dose for prescription  Depth (cm)of 90% isodose=E(Mev)/4  Depth (cm)of 80% isodose=E(Mev)/3  Electron beam penetrates a finite depth with clear cut range,the photon beam in other hand proceeds infinite range.
  • 23. Dose dependence on incident energy
  • 24. Dose dependence on field size
  • 26. dmax  Maximum dose is not on skin  It is somehow away from skin  It is due to electronic equilibrium  6Mev-10mm  9Mev-15mm  12Mev-19mm Approximation
  • 27. Range of depth  Roughly E/2  E=energy  E.g –for 6Mev range is 6/2=3cm
  • 28. Surface dose in 10x10 cm field  For 5mev=74%  10mev=82%  16mev=93%  25mev=96%  It depends upon energy,field size and thickness of scattering foil
  • 29. Energy range  6-20 Mev 6,9,12,15,18,20 mev
  • 30. Selection of energy Ep0(MeV) = 3.3 x R90 (cm) [R90 exceeds max depth of PTV] Ep0(MeV) = 2 x Rp [Rp is the practical range of electrons ]
  • 31. Ideal condition for EBT  Electron beam incidence normal to flat surface  Underlying homogenous soft tissue (provide uniform dose in penumbra from surface to R90 after which there is rapid dose fall off)
  • 32. USE OF BOLUS  To flatten out irregular surface  Reduce the penetration of electrons in part of the field  Increase the surface dose  To act as a missing tissue compensator  Thickness of bolus required increases as the surface dose increases e.g – 6-10mev=1cm – 10-15mev=0.5cm – >15mev=no bolus
  • 33. BOLUS MATERIAL  Paraffin wax,polysterene,lucite,superstoff,superflab  Flexible bolus that confirms to surface is desirable  Large air gap between the absorber and the surface would result in scattering of e- outside the field and reduction in dose that may not be easily predictable unless specially measured for this condition
  • 34.
  • 35.
  • 36. MLC vs. BLOCK  No MLC  Wooden MLC  Wax MLC  Customized block
  • 37. BLOCKING MATERIAL  Usually Cerroband  Cerroband-Pb+Bi+Sn+Cu  Melting point=  Customized block  Thickness of block required to use as block =E/2 e.g for 6mev=6/2=3cm
  • 39. Decelerators  A plate of low atomic wt. material such as lucite and polysterene is sometimes used to reduce the energy of electron beam known as decelerators  It must be placed in close contact with patient surface with bolus