Design And Analysis of Different Artificial Human Hip Stems Based on
Fenestrations Using PEEK Material
Mr. Rohith. N1, Dr. H. C. Chittappa2
1M. Tech Student, Department of Mechanical Engineering,
University Visvesvaraya College of Engineering, Bangalore, Karnataka, India.
2Professor, Department of Mechanical Engineering,
University Visvesvaraya College of Engineering, Bangalore, Karnataka, India.
---------------------------------------------------------------------***---------------------------------------------------------------------
Abstract - The Hip joint is a primary joint which provides
stability to the human body. These joints will be replaced by
the implants during hip replacement procedures due to
wear and strain brought on by ageing and other factors. Hip
joint prostheses are structural parts that still struggle with
complex issues including the interaction of the hip stem's
physical and biological characteristics with the human
femur bone. On the commercial market, there are numerous
varieties of artificial hip joints. The needs of the patients can
be accommodated by choosing from a variety of materials
and designs. One of them is the hip stem design with
fenestrations. Titanium and stainless steel are the materials
that are frequently utilised in hip joint replacement. The
design and analysis of complete joint replacements and
other orthopaedic devices have benefited from the use of the
finite element approach in orthopaedic biomechanics.
In this work, three different types of hip stems were
considered viz., hip stem with out fenestration, hip stem with
bigloop fenestration and hip stem with slot fenestration
were modeled using SOLIDWORKS 2018 by Dassault
Systems and these hip stems were analyzed for the stress–
strain distribution and deformation over the hip stem
prosthesis under different loading conditions like, standing,
walking, jumping and running for this we used two different
materials Polyether-ether-ketone (PEEK) and Ti-6Al-4V
using renowned tool ANSYS Workbench 2022 R2 version to
produce the hip stem with the best design and material
available.
Key Words: Finite Element Method, Biomaterials, Hip
implant, Artificial hip joint, PEEK.
1. INTRODUCTION
A biomaterial is a material that have been produced to
interact with living systems for therapaeutic (treating,
improving, fixing, or changing a tissue function of the
body) either diagnostic purpose. Bio-materials has been a
field of study for nearly 50 years. Bio-implants, which
include porous bone implants, dental implants, prostheses,
wearable biosensors, and drug delivery devices, are all
implants used in medical or clinical therapy. Typically,
they are placed within the body of a person for more than
30 days. They integrate with the human body, the qualities
of materials, and the intactness of bio-implants to fix,
sustain, recreate, either to improve the function of the
human tissues.
1.1 Total Hip Replacement
Hip joint replacement using artificial replacement surgery
needs to be performed for patients with permanent hip
joint damage due to, aging, necrosis, osteoarthritis,
osteoporosis or accidents. Many patients with hip joint
disease have difficulties carrying out for daily activities;
thus, hip replacement surgery has become a standard
method. THR is called as “Operation of Century” because it
has revolutionised the treatment of a patient with
advanced hip disorders. THR involves exchange of the
morbid hip with a hip implant referred to as prosthesis,
that is employed to transferring the load from socket to
the leg bone by implanting a hip stem into the latter.
Ideally, the stem area of a hip implant thought to be fully
contact with the surrounding femure cortical bone to
attain stable fixation during a THR. The hip implant is
additionally responsible for getting the physiological load
from the hip stem and transfers some amount of it to the
surrounding tissues. These are the major surgical
procedures, that are solely undertaken once the disease
has proceeded to the stage where no different treatment is
possible the aim is to alleviate pain and improve mobility
[1].
Fig. 1 Steps of Total Hip Replacement surgery
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 10 Issue: 08 | Aug 2023 www.irjet.net p-ISSN: 2395-0072
© 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal | Page 277
2. LITERATURE REVIEW
Solehuddin bin Shuib et al. Hip implants are prosthetic
joints made of metals like titanium alloy or stainless steel.
They are carried in place by lengthy stems that protrude
deep into the femur canal. When a human hip joint
withstands a biological injury, it is typically used. Analysis
also reveals a yearly rise in the number of people who get
this type of surgery, especially those who are older and fall
into the 65+ age group. The objective of this study is to
research hip implant advancements and to suggest a
design that fully satisfies the requirements for designing a
hip implant. Hip implants must take into consideration a
variety of factors, including stiffness, implant
characteristics, and implant size. Materials are among the
most essential design considerations for a hip implant
since they have a direct impact on the implant's stiffness.
Titanium and stainless steel are the commonly used
materials [2].
Kelechukwu Onuoha et al. The process of THR is
eminent. By reducing pain and functional impairment,
THR enhances the quality of life for persons with
moderate to serious hip arthritis. In this analysis, only hip
osteoarthritis patients were included. It was simple to
identify osteoarthritis as an indication for the surgery
because research also shown that arthritis was a common
reason for THR. 71.4 % of the patients in this work were
found to be older than 50 years [3].
Manuel Guzman et al. The femoral head and the hip stem
are the two primary parts of the prosthesis utilized in
THR. SS 316L, Ti alloys (Ti-6Al-4V and TMZF), and Co-Cr
alloys are the most commonly utilized metal alloys for
these components, despite the fact that they are prone to
loose particles. These materials after implanted, plays a
crucial role in the distribution of stress and strain because
their high elasticity modulus which is greater than that of
human bone. The requirements for a hip implant’s design
are complicated because there are so many factors to be
taken into an account, including the patient’s age, the
geometry of the implant, the bone structure, and the
biocompatibility of the material. All of these factors have
an impact on how much stress and strain the implant will
experience. One way to avoid future complications,
including aseptic loosening or implant fracture, is to
modify the shape of an existing prosthesis or use an ideal
biomaterial [4].
Anthony L. Sabatini et al. This article reevaluated a
number of hip stems. For all the stems, FE Analysis was
carried out using different cross-sections. In addition to
recording the displacement, von-Misses stress
comparisons were performed for each of the stems at the
predetermined places. The analysis was conducted using
three materials: Ti-6Al-4V, SS316L, and Cr-Co-Mo. Ti-6Al-
4V showed less stress than the other two materials. The
circular and eliptical cross-sections shaped even
distribution of stresses throughout the length of the hip
stem than the other cross-sections viz.; trapezoid and oval.
This analysis could clarify the scenarios that should be
avoided for hip implants that need stress shielding. For
total hip replacements, additional anatomical designs are
required [5].
M. Paglia et al. PEEK is a polyaromatic semi-crystalline
thermoplastic polymer which was first developed for
engineering applications by Victrex PLC in the early 1980s.
It was initially made available for industrial use in the
production of bearings, compressor plate valves, cable
insulation, turbine blades, piston parts, and aircraft. PEEK
is a polymer that is organic and colourless that is a
member of the PAEK (polyaryl-ether-ketone) family. A
single monomer makes it a homopolymer. This
thermoplastic semi-crystalline polymer retains its
outstanding chemical and mechanical characteristics even
at elevated temperature. PEEK is stable at high
temperatures due to its chemical compositions and
structure, which facilitates simple manufacturing of PEEK
implant components. When subjected to heat sterilising
processes, PEEK is resistant to degradation. Due to its
excellent thermal characteristics, superior wear
resistance, great processability, inertness, corrosion
resistance, high strength, and elastic modulus, PEEK is
widely employed in both medical and engineering fields.
In fact, it has an elastic modulus that is identical to human
bone, indicating uniform stress distribution to the tissues
around it. PEEK is a non-toxic, non-mutagenic substance
that is extremely compatible with the tissues in its
surrounding environment. As a result, it may be the
perfect replacement for those who are allergic to titanium
and other metals. PEEK has a more aesthetically pleasing
appearance and doesn't have a metallic tone it seems
beige with a hint of grey [6].
Jiaqi Zhang et al. Skull deformities are treated through
cranioplasty, which involves raising the scalp and
reshaping the skull with the original piece, titanium mesh,
or solid biomaterial. Medical practitioners are increasingly
using additive manufacturing technology, often known as
3D printing, to create specialized, replicas of tissues,
organs, and bones, providing a viable solution with ideal
anatomic fitting in the individual and skeletal
reconstruction. Here, we analyze a patient from 15 years
ago that had titanium mesh cranioplasty. The left eyebrow
arch was weakened by the titanium mesh's unsightly
appearance, which also caused a sinus tract to develop. An
additively produced polyether ether ketone (PEEK) skull
implant was used during cranioplasty. PEEK skull implants
have been inserted successfully without any issues. To our
information, this is the 1st reported case of direct use of
fused filament fabrication (FFF) fabricated PEEK implant
for cranial repair. The FFF-printed PEEK customized skull
implant can possess simultaneously with adjustable
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 10 Issue: 08 | Aug 2023 www.irjet.net p-ISSN: 2395-0072
© 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal | Page 278
material thickness and more complex structure, tunable
mechanical properties, and low processing cost compared
to traditional manufacturing processes [7].
3. PROBLEM IDENTIFICATION
At present in medical field Titanium alloys, Stainless steel
and Co-Cr alloy materials are been used for the artificial
human hip stems. These materials have good mechanical
and physical properties, they are very strong in nature.
But the stiffness / elastic modulus of these materials are
too much high when compared with the bone properties,
due to this when the load applied on the hip joint the
maximum load will be on the implant. When hip implant
carries more load than the femur bone then there will be a
wear between bone and the implant this effect is called as
shield stressing, by this it gets loosen very quickly before
its life span. Also, weight of the implant is more. Increases
in bone porosity. Decrease in bone mass. There will be a
magnetism effect and anode-cathode reactions.
4. OBJECTIVES
 To design for development of an artificial human
hip stems based on fenestration using CAD
modelling software i.e., SOLIDWORKS 2018.
 To perform FEA for all different types of
fenestrated artificial human hip stems for
Polyether–ether–ketone (PEEK) and Ti-6Al-4V
materials under different conditions such as
standing, walking, jumping and running using
ANSYS Workbench 2022 R2.
 Results of the three different hip stems with PEEK
and Ti-6Al-4V materials will be compared and one
of the best will be suggested.
 To reduce the weight of human hip implant by
using PEEK material instead of Titanium alloys,
Stainless steel and Co-Cr alloy.
 To improve the strength, corrosion resistance and
wear resistance of the hip stem by this stress
shielding can be prevented.
5. METHODOLOGY
Fig. 2 Methodology flow chart
6. DESIGN AND ANALYSIS OF HIP STEMS
The CAD models have to be created in the computer
environment. Here we designed a Hip Replacement of each
different parts, the dimensions of each part is collected
from previous record. SOLIDWORKS 2018 have been used
in the modeling process.
Fig. 3 Isometric view of hip implant
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 10 Issue: 08 | Aug 2023 www.irjet.net p-ISSN: 2395-0072
© 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal | Page 279
6.1 Different types of fenestration designs
Fig. 4 Types of fenestration
For hip replacement, artificial hip joints come in a variety
of designs and shapes. Three different hip joint designs
were used in this study, as shown in Fig. 4: a hip stem
without fenestration, a hip stem with a big loop
fenestration, and a hip stem with a slot fenestration [8].
The only difference in dimensions was fenestration. PEEK
and Ti-6Al-4V were both employed as the materials.
Materials were chosen for their high biocompatibility,
durability, resistance to wear and corrosion, and strength.
The material was thought to be uniform, isotropic, and
linearly elastic.
7. RESULTS AND DISCUSSION
The final model was assembled in SOLIDWORKS and
imported to ANSYS Workbench software in .IGS format
and the analysis is carried out in Static structural. So in
this project we are going to consider three different types
of hip stem designs namely, hip stem without fenestration,
hip stem with big loop fenestration and hip stem with slot
fenestration. Further these hip stems should be analyzed
for different loading conditions like, standing, walking,
jumping and running. Then comparing the results of PEEK
material with commonly used Titanium based alloy Ti-6Al-
4V for all these hip stems. The mechanical property of
these materials is given below Table 1.
Table 1 Mechanical properties of the materials used for
hip stems [9]
We are going to use some different activities like standing,
walking, etc. So the stress analysis on a hip prosthesis
during various patient activities, such as standing,
walking, jumping as well as in running. These loads are
considered according to Aftab S et al. [10] on a patient
who weighs 75 Kg, with an age group of around 35 years.
Due to the symmetry of the human body, when a person is
standing, their weight will be distributed evenly between
their two legs. For this reason, the weight was divided in
half and boundary conditions were applied to the finite
element model. The analysis was performed under the
assumption that the subject was standing normally, with a
load of 370 N applied to the head of the femur at the hip
joint and the hip stem fixed when in contact with the
femoral bone. Similar to this, the hip stem was examined
under a variety of situations derived from various human
everyday activities, such as jumping, where a force of 750
N acting on one leg was applied, for walking an additional
800 N was applied, and running, where a load of 1463 N
was applied. Also shown in the Table 2.
Table 2 Loading conditions of hip joint for different daily
activities [10]
Activity Maximum Load (N)
Standing 370
Walking 800
Jumping 750
Running 1463
The following Fig. 5 shows the meshing process which
chosen to be 1mm element size.
Fig. 5 Meshing process
Table 3 Number of Elements used in analysis
Element size: 1mm
Types of hip stem Nodes Elements
Without
fenestration
77915 32325
With big loop
fenestration
86111 36782
With slot
fenestration
89874 39068
Sl.N
o
Material
s
Young’s
Modulu
s (GPa)
Density
(kg/m3
)
Poisson’
s ratio
Yield
strengt
h (MPa)
1 PEEK 3.95 1320 0.3931 170
2 Ti-6Al-4V 120 4430 0.34 900
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 10 Issue: 08 | Aug 2023 www.irjet.net p-ISSN: 2395-0072
© 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal | Page 280
The below Fig. 6 indicates the fixed/support condition to
the hip implant.
Fig. 6 Support condition
The Fig. 7 shows the applying of loading conditions which
are mentioned in the Table 2. The force of 370 N
magnitude is applied along the -y direction as shown in
the following figure.
Fig. 7 Applying the load
7.1 Computational Results
7.1.1 Masses of the different types of fenestrated hip
stems
These below mentioned masses are taken from the
SOLIDWORKS software by applying the material
properties of the PEEK and Ti-6Al-4V for the different hip
stem designs.
Fig. 8 Mass of the hip stem with big loop fenestration for
PEEK material
Fig. 9 Comparison of masses for different types of hip
stem
7.1.2 Standing condition (370 N)
The Fig. 10(a) refers to Von-misess stress produced in the
hip stem without fenestration under standing condition
when PEEK material is applied to the hip stem which is
34.106 MPa, which is similar for the remaining
fenestrations.
The Fig. 10(b) refers to Von-misess stress produced in the
hip stem without fenestration under standing condition
when Ti-6Al-4V material is applied to the hip stem which
is 35.569 MPa, which is similar for the remaining
fenestrations.
74.77 69.61 72.17
250.94
233.6 242.21
0
50
100
150
200
250
300
Hip stem without
fenestration
Hip stem with big
loop fenestration
Hip stem with slot
fenestration
MASS
IN
GRAMS
TYPES OF HIP STEMS
Masses of the hip stems in grams for PEEK
Masses of the hip stems in grams for Ti-6Al-4V
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 10 Issue: 08 | Aug 2023 www.irjet.net p-ISSN: 2395-0072
© 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal | Page 281
(a) (b)
Fig. 10 Hip stem without fenestration under standing
condition for PEEK and Ti-6Al-4V material
Fig. 11 Comparison of Von-Misses stress under Standing
condition
From Fig. 11 the von-misses stress of hip stem with slot
fenestration for PEEK material is least when compared to
others in standing condition.
7.1.3 Walking condition (800 N)
The Fig. 12(a) refers to Von-misess stress produced in the
with bigloop fenestration under walking condition when
PEEK material is applied to the hip stem which is 78.393
MPa which is similar for the remaining fenestrations.
The Fig. 12(b) refers to Von-misess stress produced in the
hip stem with bigloop fenestration under walking
condition when Ti-6Al-4V material is applied to the hip
stem which is 84.475 MPa which is similar for the
remaining fenestrations.
(a) (b)
Fig. 12 Hip stem with big loop fenestration under walking
condition for PEEK and Ti-6Al-4V material
Fig. 13 Comparison of Von-Misses stress under Walking
condition
From Fig. 13 the von-misses stress of hip stem with slot
fenestration for PEEK material is least, when compared to
others in walking condition.
7.1.4 Jumping condition (750 N)
During jumping with the both legs 750 N load will be
applied on one leg. The Fig 14(a) refers to Von-misess
stress produced in the hip stem with slot fenestratation
when PEEK material is applied to the hip stem which is
63.687 MPa, which is similar for the remaining
fenestrations.
The Fig. 14(b) refers to Von-misess stress produced in the
hip stem with slot fenestratation when Ti-6Al-4V material
is applied to the hip stem which is 65.48 MPa, which is
similar for the remaining fenestrations.
34.106
36.257
31.419
35.569
39.07
32.303
0
5
10
15
20
25
30
35
40
45
Without fenestration With Big Loop
fenestration
With Slot fenestration
Von-Misses
Stress
in
MPa
Types Of Hip stem
PEEK Ti-6Al-4V
73.743
78.393
67.933
76.906
84.475
69.845
0
10
20
30
40
50
60
70
80
90
Without fenestration With Big Loop
fenestration
With Slot fenestration
Von-Misses
Stress
in
MPa
Types Of Hip stem
PEEK Ti-6Al-4V
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 10 Issue: 08 | Aug 2023 www.irjet.net p-ISSN: 2395-0072
© 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal | Page 282
(a) (b)
Fig. 14 Hip stem with slot fenestration under jumping
condition for PEEK and Ti-6Al-4V material
Fig. 15 Comparison of Von-Misses stress under Jumping
condition
From Fig. 15 the von-misses stress of hip stem with slot
fenestration for PEEK material is least, when compared to
others jumping condition.
7.1.5 Running condition (1463 N)
The Fig. 16(a) refers to Von-misess stress produced in the
hip stem without fenestration under running condition
when PEEK material is applied to the hip stem which is
134.86 MPa, which is similar for the remaining
fenestrations.
The Fig. 16(b) refers to Von-misess stress produced in the
hip stem without fenestration under jumping condition
when Ti-6Al-4V material is applied to the hip stem which
is 140.64 MPa, which is similar for the remaining
fenestrations.
(a) (b)
Fig. 16 Hip stem without fenestration under running
condition for PEEK and Ti-6Al-4V material
Fig. 17 Comparison of Von-Misses stress under Running
condition
From Fig. 17 the von-misses stress of hip stem with slot
fenestration for PEEK material is least, when compared to
others in running condition.
8. CONCLUSION
In this work, three hip stem designs, namely without
fenestration, with bigloop fenestration and with slot
fenestration designs were modeled. For all designs, two
materials PEEK and Ti-6Al-4V was used. PEEK is used
because it as the similar stiffness as human femur bone
and it as good wear properties, low moist absorption, low
co-efficient of friction etc. All of these designs, which use
materials with various physical properties, will be put
through the rigours of daily activities like standing,
walking, jumping, and running. Static structural analysis
was carried out using ANSYS Workbench with the
application of the loading boundary condition as per the
work by Aftab S. et al. All three designs were found to
suggest lower stress levels than the yield strength. Also, in
these three designs for PEEK and Ti-6Al-4V materials
under different loading conditions, the artificial human hip
stem with slot fenestration for PEEK material has lesser
Von-Misses stress under all different loading conditions as
69.134
73.494
63.687
72.1
79.196
65.48
0
10
20
30
40
50
60
70
80
90
Without fenestration With Big Loop
fenestration
With Slot fenestration
Von-Misses
Stress
in
MPa
Types Of Hip stem
PEEK Ti-6Al-4V
134.86
143.36
124.23
140.64
154.48
127.73
0
20
40
60
80
100
120
140
160
180
Without fenestration With Big Loop
fenestration
With Slot fenestration
Von-Misses
Stress
in
MPa
Types Of Hip stem
PEEK Ti-6Al-4V
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 10 Issue: 08 | Aug 2023 www.irjet.net p-ISSN: 2395-0072
© 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal | Page 283
mentioned above. Ti-6Al-4V has less deformation when
compared to PEEK material for all the designs which were
modeled in the current work. It is observed that for all the
designs under different loading conditions the von-misses
stress is almost similar for PEEK and Ti-6Al-4V material.
Also, weight of the PEEK material is lesser than the Ti-6Al-
4V. So, PEEK material with slot fenestration can be
suggested.
REFERENCES
[1] L. Guo et al., “On the design evolution of hip
implants: A review,” Materials and Design, vol. 216.
Elsevier Ltd, Apr. 01, 2022. doi:
10.1016/j.matdes.2022.110552.
[2] S. Shuib, B. Sahari, A. Ahmed Shokri, and C. Soon
Chai, “The Design Improvement of Hip Implant for
Total Hip Replacement (THR),” Jurnal
Kejuruteraan, vol. 20, no. 1, pp. 107–113, Dec.
2008, doi: 10.17576/jkukm-2008-20-10.
[3] “Clinical Outcomes Following Total Hip
Replacement,” EAS Journal of Medicine and Surgery,
vol. 4, no. 6, pp. 134–139, Jun. 2022, doi:
10.36349/easjms.2022.v04i06.003.
[4] M. Guzmán et al., “Finite Element Assessment of a
Hybrid Proposal for Hip Stem, from a Standardized
Base and Different Activities,” Applied Sciences
(Switzerland), vol. 12, no. 16, Aug. 2022, doi:
10.3390/app12167963.
[5] A. L. Sabatini and T. Goswami, “Hip implants VII:
Finite element analysis and optimization of cross-
sections,” Mater Des, vol. 29, no. 7, pp. 1438–1446,
2008, doi: 10.1016/j.matdes.2007.09.002.
[6] M. Paglia, M. Beretta, V. Quinzi, and S. Colombo,
“PEEK polymer in orthodontics: a scoping review,”
Eur J Paediatr Dent, vol. 23, no. 2, pp. 137–139,
2022, doi: 10.23804/ejpd.2022.23.02.10.
[7] J. Zhang et al., “Additively manufactured polyether
ether ketone (PEEK) skull implant as an alternative
to titanium mesh in cranioplasty,” Int J Bioprint,
vol. 9, no. 1, Nov. 2022, doi: 10.18063/ijb.v9i1.634.
[8] Ikhsan, A. R. Prabowo, J. M. Sohn, and J. Triyono,
“Finite element analysis of different artificial hip
stem designs based on fenestration under static
loading,” in Procedia Structural Integrity, 2020, vol.
27, pp. 101–108. doi:
10.1016/j.prostr.2020.07.014.
[9] Y. Jun and K. Choi, “Design of patient-specific hip
implants based on the 3D geometry of the human
femur,” Advances in Engineering Software, vol. 41,
no. 4, pp. 537–547, 2010, doi:
10.1016/j.advengsoft.2009.10.016.
[10] S. G. Aftab, A. Faisal, H. Hussain, B. Sreedhara, N. R.
Babu, and B. A. Praveen, “Structural analysis of
human femur bone to select an alternative
composite material,” Mater Today Proc, vol. 54, pp.
141–146, Jan. 2022, doi:
10.1016/j.matpr.2021.08.197.

Recomendados

FINITE ELEMENT ANALYSIS OF FEMORAL INTRAMEDULLARY NAILING von
FINITE ELEMENT ANALYSIS OF FEMORAL INTRAMEDULLARY NAILINGFINITE ELEMENT ANALYSIS OF FEMORAL INTRAMEDULLARY NAILING
FINITE ELEMENT ANALYSIS OF FEMORAL INTRAMEDULLARY NAILINGJournal For Research
369 views8 Folien
Modelling and static analysis of femur bone by using different implant materials von
Modelling and static analysis of femur bone by using different implant materialsModelling and static analysis of femur bone by using different implant materials
Modelling and static analysis of femur bone by using different implant materialsIOSR Journals
790 views10 Folien
H1302034856 von
H1302034856H1302034856
H1302034856IOSR Journals
233 views9 Folien
Total Hip Replacement Implant Designing and its Computational Analysis using ... von
Total Hip Replacement Implant Designing and its Computational Analysis using ...Total Hip Replacement Implant Designing and its Computational Analysis using ...
Total Hip Replacement Implant Designing and its Computational Analysis using ...IRJET Journal
6 views5 Folien
Finite element stress analysis of artificial femur head on hip joint von
Finite element stress analysis of artificial femur head on hip jointFinite element stress analysis of artificial femur head on hip joint
Finite element stress analysis of artificial femur head on hip jointSAURABH SINGH
343 views18 Folien
Biochmical1 pdf von
Biochmical1 pdfBiochmical1 pdf
Biochmical1 pdfKiran Kumar
20 views74 Folien

Más contenido relacionado

Similar a Design And Analysis of Different Artificial Human Hip Stems Based on Fenestrations Using PEEK Material

Evolution of Total Hip Replacement von
Evolution of Total Hip ReplacementEvolution of Total Hip Replacement
Evolution of Total Hip ReplacementTejasvi Agarwal
7.4K views50 Folien
bone and_cartilage_tissue_engineering by Sumit von
 bone and_cartilage_tissue_engineering by Sumit bone and_cartilage_tissue_engineering by Sumit
bone and_cartilage_tissue_engineering by SumitDcrust
511 views22 Folien
Extremity_Wear_Testing_-_Total_Ankle_and_Shoulder_Replacement_Testing_ von
Extremity_Wear_Testing_-_Total_Ankle_and_Shoulder_Replacement_Testing_Extremity_Wear_Testing_-_Total_Ankle_and_Shoulder_Replacement_Testing_
Extremity_Wear_Testing_-_Total_Ankle_and_Shoulder_Replacement_Testing_Evan Guilfoyle
259 views7 Folien
Review on alternative material for prostheesis.pdf von
Review on alternative material for prostheesis.pdfReview on alternative material for prostheesis.pdf
Review on alternative material for prostheesis.pdfEndalkachew Gashawtena
7 views6 Folien
Review on-alternate-materials-for-producing-low-cost-lower-limb-prosthetic-so... von
Review on-alternate-materials-for-producing-low-cost-lower-limb-prosthetic-so...Review on-alternate-materials-for-producing-low-cost-lower-limb-prosthetic-so...
Review on-alternate-materials-for-producing-low-cost-lower-limb-prosthetic-so...Endalkachew Gashawtena
74 views6 Folien
Review on-alternate-materials-for-producing-low-cost-lower-limb-prosthetic-so... von
Review on-alternate-materials-for-producing-low-cost-lower-limb-prosthetic-so...Review on-alternate-materials-for-producing-low-cost-lower-limb-prosthetic-so...
Review on-alternate-materials-for-producing-low-cost-lower-limb-prosthetic-so...Endalkachew Gashawtena
81 views6 Folien

Similar a Design And Analysis of Different Artificial Human Hip Stems Based on Fenestrations Using PEEK Material(20)

Evolution of Total Hip Replacement von Tejasvi Agarwal
Evolution of Total Hip ReplacementEvolution of Total Hip Replacement
Evolution of Total Hip Replacement
Tejasvi Agarwal7.4K views
bone and_cartilage_tissue_engineering by Sumit von Dcrust
 bone and_cartilage_tissue_engineering by Sumit bone and_cartilage_tissue_engineering by Sumit
bone and_cartilage_tissue_engineering by Sumit
Dcrust511 views
Extremity_Wear_Testing_-_Total_Ankle_and_Shoulder_Replacement_Testing_ von Evan Guilfoyle
Extremity_Wear_Testing_-_Total_Ankle_and_Shoulder_Replacement_Testing_Extremity_Wear_Testing_-_Total_Ankle_and_Shoulder_Replacement_Testing_
Extremity_Wear_Testing_-_Total_Ankle_and_Shoulder_Replacement_Testing_
Evan Guilfoyle259 views
Review on-alternate-materials-for-producing-low-cost-lower-limb-prosthetic-so... von Endalkachew Gashawtena
Review on-alternate-materials-for-producing-low-cost-lower-limb-prosthetic-so...Review on-alternate-materials-for-producing-low-cost-lower-limb-prosthetic-so...
Review on-alternate-materials-for-producing-low-cost-lower-limb-prosthetic-so...
Review on-alternate-materials-for-producing-low-cost-lower-limb-prosthetic-so... von Endalkachew Gashawtena
Review on-alternate-materials-for-producing-low-cost-lower-limb-prosthetic-so...Review on-alternate-materials-for-producing-low-cost-lower-limb-prosthetic-so...
Review on-alternate-materials-for-producing-low-cost-lower-limb-prosthetic-so...
Finite Element Investigation of Hybrid and Conventional Knee Implants von CSCJournals
Finite Element Investigation of Hybrid and Conventional Knee ImplantsFinite Element Investigation of Hybrid and Conventional Knee Implants
Finite Element Investigation of Hybrid and Conventional Knee Implants
CSCJournals389 views
Bone Grafts /certified fixed orthodontic courses by Indian dental academy von Indian dental academy
Bone Grafts   /certified fixed orthodontic courses by Indian dental academy Bone Grafts   /certified fixed orthodontic courses by Indian dental academy
Bone Grafts /certified fixed orthodontic courses by Indian dental academy
Fatigue Analysis of a Bone Implant Construct von Mert G
Fatigue Analysis of a Bone Implant ConstructFatigue Analysis of a Bone Implant Construct
Fatigue Analysis of a Bone Implant Construct
Mert G882 views
Bone grafts and Bone Substitutes/ dental implant courses von Indian dental academy
Bone grafts and Bone Substitutes/ dental implant coursesBone grafts and Bone Substitutes/ dental implant courses
Bone grafts and Bone Substitutes/ dental implant courses
Strontium-Coated Clay Nanoparticles in Calcium Phosphate Cement for Biomedica... von AnushaElumalai
Strontium-Coated Clay Nanoparticles in Calcium Phosphate Cement for Biomedica...Strontium-Coated Clay Nanoparticles in Calcium Phosphate Cement for Biomedica...
Strontium-Coated Clay Nanoparticles in Calcium Phosphate Cement for Biomedica...
AnushaElumalai84 views
Additive Manufacturing and Testing of a Prosthetic Foot Ankle Joint von ijtsrd
Additive Manufacturing and Testing of a Prosthetic Foot Ankle JointAdditive Manufacturing and Testing of a Prosthetic Foot Ankle Joint
Additive Manufacturing and Testing of a Prosthetic Foot Ankle Joint
ijtsrd337 views

Más de IRJET Journal

SOIL STABILIZATION USING WASTE FIBER MATERIAL von
SOIL STABILIZATION USING WASTE FIBER MATERIALSOIL STABILIZATION USING WASTE FIBER MATERIAL
SOIL STABILIZATION USING WASTE FIBER MATERIALIRJET Journal
22 views7 Folien
Sol-gel auto-combustion produced gamma irradiated Ni1-xCdxFe2O4 nanoparticles... von
Sol-gel auto-combustion produced gamma irradiated Ni1-xCdxFe2O4 nanoparticles...Sol-gel auto-combustion produced gamma irradiated Ni1-xCdxFe2O4 nanoparticles...
Sol-gel auto-combustion produced gamma irradiated Ni1-xCdxFe2O4 nanoparticles...IRJET Journal
8 views7 Folien
Identification, Discrimination and Classification of Cotton Crop by Using Mul... von
Identification, Discrimination and Classification of Cotton Crop by Using Mul...Identification, Discrimination and Classification of Cotton Crop by Using Mul...
Identification, Discrimination and Classification of Cotton Crop by Using Mul...IRJET Journal
8 views5 Folien
“Analysis of GDP, Unemployment and Inflation rates using mathematical formula... von
“Analysis of GDP, Unemployment and Inflation rates using mathematical formula...“Analysis of GDP, Unemployment and Inflation rates using mathematical formula...
“Analysis of GDP, Unemployment and Inflation rates using mathematical formula...IRJET Journal
13 views11 Folien
MAXIMUM POWER POINT TRACKING BASED PHOTO VOLTAIC SYSTEM FOR SMART GRID INTEGR... von
MAXIMUM POWER POINT TRACKING BASED PHOTO VOLTAIC SYSTEM FOR SMART GRID INTEGR...MAXIMUM POWER POINT TRACKING BASED PHOTO VOLTAIC SYSTEM FOR SMART GRID INTEGR...
MAXIMUM POWER POINT TRACKING BASED PHOTO VOLTAIC SYSTEM FOR SMART GRID INTEGR...IRJET Journal
14 views6 Folien
Performance Analysis of Aerodynamic Design for Wind Turbine Blade von
Performance Analysis of Aerodynamic Design for Wind Turbine BladePerformance Analysis of Aerodynamic Design for Wind Turbine Blade
Performance Analysis of Aerodynamic Design for Wind Turbine BladeIRJET Journal
7 views5 Folien

Más de IRJET Journal(20)

SOIL STABILIZATION USING WASTE FIBER MATERIAL von IRJET Journal
SOIL STABILIZATION USING WASTE FIBER MATERIALSOIL STABILIZATION USING WASTE FIBER MATERIAL
SOIL STABILIZATION USING WASTE FIBER MATERIAL
IRJET Journal22 views
Sol-gel auto-combustion produced gamma irradiated Ni1-xCdxFe2O4 nanoparticles... von IRJET Journal
Sol-gel auto-combustion produced gamma irradiated Ni1-xCdxFe2O4 nanoparticles...Sol-gel auto-combustion produced gamma irradiated Ni1-xCdxFe2O4 nanoparticles...
Sol-gel auto-combustion produced gamma irradiated Ni1-xCdxFe2O4 nanoparticles...
IRJET Journal8 views
Identification, Discrimination and Classification of Cotton Crop by Using Mul... von IRJET Journal
Identification, Discrimination and Classification of Cotton Crop by Using Mul...Identification, Discrimination and Classification of Cotton Crop by Using Mul...
Identification, Discrimination and Classification of Cotton Crop by Using Mul...
IRJET Journal8 views
“Analysis of GDP, Unemployment and Inflation rates using mathematical formula... von IRJET Journal
“Analysis of GDP, Unemployment and Inflation rates using mathematical formula...“Analysis of GDP, Unemployment and Inflation rates using mathematical formula...
“Analysis of GDP, Unemployment and Inflation rates using mathematical formula...
IRJET Journal13 views
MAXIMUM POWER POINT TRACKING BASED PHOTO VOLTAIC SYSTEM FOR SMART GRID INTEGR... von IRJET Journal
MAXIMUM POWER POINT TRACKING BASED PHOTO VOLTAIC SYSTEM FOR SMART GRID INTEGR...MAXIMUM POWER POINT TRACKING BASED PHOTO VOLTAIC SYSTEM FOR SMART GRID INTEGR...
MAXIMUM POWER POINT TRACKING BASED PHOTO VOLTAIC SYSTEM FOR SMART GRID INTEGR...
IRJET Journal14 views
Performance Analysis of Aerodynamic Design for Wind Turbine Blade von IRJET Journal
Performance Analysis of Aerodynamic Design for Wind Turbine BladePerformance Analysis of Aerodynamic Design for Wind Turbine Blade
Performance Analysis of Aerodynamic Design for Wind Turbine Blade
IRJET Journal7 views
Heart Failure Prediction using Different Machine Learning Techniques von IRJET Journal
Heart Failure Prediction using Different Machine Learning TechniquesHeart Failure Prediction using Different Machine Learning Techniques
Heart Failure Prediction using Different Machine Learning Techniques
IRJET Journal7 views
Experimental Investigation of Solar Hot Case Based on Photovoltaic Panel von IRJET Journal
Experimental Investigation of Solar Hot Case Based on Photovoltaic PanelExperimental Investigation of Solar Hot Case Based on Photovoltaic Panel
Experimental Investigation of Solar Hot Case Based on Photovoltaic Panel
IRJET Journal3 views
Metro Development and Pedestrian Concerns von IRJET Journal
Metro Development and Pedestrian ConcernsMetro Development and Pedestrian Concerns
Metro Development and Pedestrian Concerns
IRJET Journal2 views
Mapping the Crashworthiness Domains: Investigations Based on Scientometric An... von IRJET Journal
Mapping the Crashworthiness Domains: Investigations Based on Scientometric An...Mapping the Crashworthiness Domains: Investigations Based on Scientometric An...
Mapping the Crashworthiness Domains: Investigations Based on Scientometric An...
IRJET Journal3 views
Data Analytics and Artificial Intelligence in Healthcare Industry von IRJET Journal
Data Analytics and Artificial Intelligence in Healthcare IndustryData Analytics and Artificial Intelligence in Healthcare Industry
Data Analytics and Artificial Intelligence in Healthcare Industry
IRJET Journal3 views
DESIGN AND SIMULATION OF SOLAR BASED FAST CHARGING STATION FOR ELECTRIC VEHIC... von IRJET Journal
DESIGN AND SIMULATION OF SOLAR BASED FAST CHARGING STATION FOR ELECTRIC VEHIC...DESIGN AND SIMULATION OF SOLAR BASED FAST CHARGING STATION FOR ELECTRIC VEHIC...
DESIGN AND SIMULATION OF SOLAR BASED FAST CHARGING STATION FOR ELECTRIC VEHIC...
IRJET Journal52 views
Efficient Design for Multi-story Building Using Pre-Fabricated Steel Structur... von IRJET Journal
Efficient Design for Multi-story Building Using Pre-Fabricated Steel Structur...Efficient Design for Multi-story Building Using Pre-Fabricated Steel Structur...
Efficient Design for Multi-story Building Using Pre-Fabricated Steel Structur...
IRJET Journal11 views
Development of Effective Tomato Package for Post-Harvest Preservation von IRJET Journal
Development of Effective Tomato Package for Post-Harvest PreservationDevelopment of Effective Tomato Package for Post-Harvest Preservation
Development of Effective Tomato Package for Post-Harvest Preservation
IRJET Journal4 views
“DYNAMIC ANALYSIS OF GRAVITY RETAINING WALL WITH SOIL STRUCTURE INTERACTION” von IRJET Journal
“DYNAMIC ANALYSIS OF GRAVITY RETAINING WALL WITH SOIL STRUCTURE INTERACTION”“DYNAMIC ANALYSIS OF GRAVITY RETAINING WALL WITH SOIL STRUCTURE INTERACTION”
“DYNAMIC ANALYSIS OF GRAVITY RETAINING WALL WITH SOIL STRUCTURE INTERACTION”
IRJET Journal5 views
Understanding the Nature of Consciousness with AI von IRJET Journal
Understanding the Nature of Consciousness with AIUnderstanding the Nature of Consciousness with AI
Understanding the Nature of Consciousness with AI
IRJET Journal12 views
Augmented Reality App for Location based Exploration at JNTUK Kakinada von IRJET Journal
Augmented Reality App for Location based Exploration at JNTUK KakinadaAugmented Reality App for Location based Exploration at JNTUK Kakinada
Augmented Reality App for Location based Exploration at JNTUK Kakinada
IRJET Journal6 views
Smart Traffic Congestion Control System: Leveraging Machine Learning for Urba... von IRJET Journal
Smart Traffic Congestion Control System: Leveraging Machine Learning for Urba...Smart Traffic Congestion Control System: Leveraging Machine Learning for Urba...
Smart Traffic Congestion Control System: Leveraging Machine Learning for Urba...
IRJET Journal18 views
Enhancing Real Time Communication and Efficiency With Websocket von IRJET Journal
Enhancing Real Time Communication and Efficiency With WebsocketEnhancing Real Time Communication and Efficiency With Websocket
Enhancing Real Time Communication and Efficiency With Websocket
IRJET Journal5 views
Textile Industrial Wastewater Treatability Studies by Soil Aquifer Treatment ... von IRJET Journal
Textile Industrial Wastewater Treatability Studies by Soil Aquifer Treatment ...Textile Industrial Wastewater Treatability Studies by Soil Aquifer Treatment ...
Textile Industrial Wastewater Treatability Studies by Soil Aquifer Treatment ...
IRJET Journal4 views

Último

Generative AI Models & Their Applications von
Generative AI Models & Their ApplicationsGenerative AI Models & Their Applications
Generative AI Models & Their ApplicationsSN
8 views1 Folie
Proposal Presentation.pptx von
Proposal Presentation.pptxProposal Presentation.pptx
Proposal Presentation.pptxkeytonallamon
42 views36 Folien
Investor Presentation von
Investor PresentationInvestor Presentation
Investor Presentationeser sevinç
25 views26 Folien
Investigation of Physicochemical Changes of Soft Clay around Deep Geopolymer ... von
Investigation of Physicochemical Changes of Soft Clay around Deep Geopolymer ...Investigation of Physicochemical Changes of Soft Clay around Deep Geopolymer ...
Investigation of Physicochemical Changes of Soft Clay around Deep Geopolymer ...AltinKaradagli
12 views16 Folien
sam_software_eng_cv.pdf von
sam_software_eng_cv.pdfsam_software_eng_cv.pdf
sam_software_eng_cv.pdfsammyigbinovia
5 views5 Folien
SPICE PARK DEC2023 (6,625 SPICE Models) von
SPICE PARK DEC2023 (6,625 SPICE Models) SPICE PARK DEC2023 (6,625 SPICE Models)
SPICE PARK DEC2023 (6,625 SPICE Models) Tsuyoshi Horigome
28 views218 Folien

Último(20)

Generative AI Models & Their Applications von SN
Generative AI Models & Their ApplicationsGenerative AI Models & Their Applications
Generative AI Models & Their Applications
SN8 views
Investigation of Physicochemical Changes of Soft Clay around Deep Geopolymer ... von AltinKaradagli
Investigation of Physicochemical Changes of Soft Clay around Deep Geopolymer ...Investigation of Physicochemical Changes of Soft Clay around Deep Geopolymer ...
Investigation of Physicochemical Changes of Soft Clay around Deep Geopolymer ...
AltinKaradagli12 views
fakenews_DBDA_Mar23.pptx von deepmitra8
fakenews_DBDA_Mar23.pptxfakenews_DBDA_Mar23.pptx
fakenews_DBDA_Mar23.pptx
deepmitra815 views
MSA Website Slideshow (16).pdf von msaucla
MSA Website Slideshow (16).pdfMSA Website Slideshow (16).pdf
MSA Website Slideshow (16).pdf
msaucla76 views
Design of machine elements-UNIT 3.pptx von gopinathcreddy
Design of machine elements-UNIT 3.pptxDesign of machine elements-UNIT 3.pptx
Design of machine elements-UNIT 3.pptx
gopinathcreddy32 views
Control Systems Feedback.pdf von LGGaming5
Control Systems Feedback.pdfControl Systems Feedback.pdf
Control Systems Feedback.pdf
LGGaming56 views
GDSC Mikroskil Members Onboarding 2023.pdf von gdscmikroskil
GDSC Mikroskil Members Onboarding 2023.pdfGDSC Mikroskil Members Onboarding 2023.pdf
GDSC Mikroskil Members Onboarding 2023.pdf
gdscmikroskil53 views
Effect of deep chemical mixing columns on properties of surrounding soft clay... von AltinKaradagli
Effect of deep chemical mixing columns on properties of surrounding soft clay...Effect of deep chemical mixing columns on properties of surrounding soft clay...
Effect of deep chemical mixing columns on properties of surrounding soft clay...
AltinKaradagli9 views
Introduction to CAD-CAM.pptx von suyogpatil49
Introduction to CAD-CAM.pptxIntroduction to CAD-CAM.pptx
Introduction to CAD-CAM.pptx
suyogpatil495 views
Design of Structures and Foundations for Vibrating Machines, Arya-ONeill-Pinc... von csegroupvn
Design of Structures and Foundations for Vibrating Machines, Arya-ONeill-Pinc...Design of Structures and Foundations for Vibrating Machines, Arya-ONeill-Pinc...
Design of Structures and Foundations for Vibrating Machines, Arya-ONeill-Pinc...
csegroupvn5 views
Design_Discover_Develop_Campaign.pptx von ShivanshSeth6
Design_Discover_Develop_Campaign.pptxDesign_Discover_Develop_Campaign.pptx
Design_Discover_Develop_Campaign.pptx
ShivanshSeth632 views

Design And Analysis of Different Artificial Human Hip Stems Based on Fenestrations Using PEEK Material

  • 1. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 10 Issue: 08 | Aug 2023 www.irjet.net p-ISSN: 2395-0072 © 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal | Page 276 Design And Analysis of Different Artificial Human Hip Stems Based on Fenestrations Using PEEK Material Mr. Rohith. N1, Dr. H. C. Chittappa2 1M. Tech Student, Department of Mechanical Engineering, University Visvesvaraya College of Engineering, Bangalore, Karnataka, India. 2Professor, Department of Mechanical Engineering, University Visvesvaraya College of Engineering, Bangalore, Karnataka, India. ---------------------------------------------------------------------***--------------------------------------------------------------------- Abstract - The Hip joint is a primary joint which provides stability to the human body. These joints will be replaced by the implants during hip replacement procedures due to wear and strain brought on by ageing and other factors. Hip joint prostheses are structural parts that still struggle with complex issues including the interaction of the hip stem's physical and biological characteristics with the human femur bone. On the commercial market, there are numerous varieties of artificial hip joints. The needs of the patients can be accommodated by choosing from a variety of materials and designs. One of them is the hip stem design with fenestrations. Titanium and stainless steel are the materials that are frequently utilised in hip joint replacement. The design and analysis of complete joint replacements and other orthopaedic devices have benefited from the use of the finite element approach in orthopaedic biomechanics. In this work, three different types of hip stems were considered viz., hip stem with out fenestration, hip stem with bigloop fenestration and hip stem with slot fenestration were modeled using SOLIDWORKS 2018 by Dassault Systems and these hip stems were analyzed for the stress– strain distribution and deformation over the hip stem prosthesis under different loading conditions like, standing, walking, jumping and running for this we used two different materials Polyether-ether-ketone (PEEK) and Ti-6Al-4V using renowned tool ANSYS Workbench 2022 R2 version to produce the hip stem with the best design and material available. Key Words: Finite Element Method, Biomaterials, Hip implant, Artificial hip joint, PEEK. 1. INTRODUCTION A biomaterial is a material that have been produced to interact with living systems for therapaeutic (treating, improving, fixing, or changing a tissue function of the body) either diagnostic purpose. Bio-materials has been a field of study for nearly 50 years. Bio-implants, which include porous bone implants, dental implants, prostheses, wearable biosensors, and drug delivery devices, are all implants used in medical or clinical therapy. Typically, they are placed within the body of a person for more than 30 days. They integrate with the human body, the qualities of materials, and the intactness of bio-implants to fix, sustain, recreate, either to improve the function of the human tissues. 1.1 Total Hip Replacement Hip joint replacement using artificial replacement surgery needs to be performed for patients with permanent hip joint damage due to, aging, necrosis, osteoarthritis, osteoporosis or accidents. Many patients with hip joint disease have difficulties carrying out for daily activities; thus, hip replacement surgery has become a standard method. THR is called as “Operation of Century” because it has revolutionised the treatment of a patient with advanced hip disorders. THR involves exchange of the morbid hip with a hip implant referred to as prosthesis, that is employed to transferring the load from socket to the leg bone by implanting a hip stem into the latter. Ideally, the stem area of a hip implant thought to be fully contact with the surrounding femure cortical bone to attain stable fixation during a THR. The hip implant is additionally responsible for getting the physiological load from the hip stem and transfers some amount of it to the surrounding tissues. These are the major surgical procedures, that are solely undertaken once the disease has proceeded to the stage where no different treatment is possible the aim is to alleviate pain and improve mobility [1]. Fig. 1 Steps of Total Hip Replacement surgery
  • 2. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 10 Issue: 08 | Aug 2023 www.irjet.net p-ISSN: 2395-0072 © 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal | Page 277 2. LITERATURE REVIEW Solehuddin bin Shuib et al. Hip implants are prosthetic joints made of metals like titanium alloy or stainless steel. They are carried in place by lengthy stems that protrude deep into the femur canal. When a human hip joint withstands a biological injury, it is typically used. Analysis also reveals a yearly rise in the number of people who get this type of surgery, especially those who are older and fall into the 65+ age group. The objective of this study is to research hip implant advancements and to suggest a design that fully satisfies the requirements for designing a hip implant. Hip implants must take into consideration a variety of factors, including stiffness, implant characteristics, and implant size. Materials are among the most essential design considerations for a hip implant since they have a direct impact on the implant's stiffness. Titanium and stainless steel are the commonly used materials [2]. Kelechukwu Onuoha et al. The process of THR is eminent. By reducing pain and functional impairment, THR enhances the quality of life for persons with moderate to serious hip arthritis. In this analysis, only hip osteoarthritis patients were included. It was simple to identify osteoarthritis as an indication for the surgery because research also shown that arthritis was a common reason for THR. 71.4 % of the patients in this work were found to be older than 50 years [3]. Manuel Guzman et al. The femoral head and the hip stem are the two primary parts of the prosthesis utilized in THR. SS 316L, Ti alloys (Ti-6Al-4V and TMZF), and Co-Cr alloys are the most commonly utilized metal alloys for these components, despite the fact that they are prone to loose particles. These materials after implanted, plays a crucial role in the distribution of stress and strain because their high elasticity modulus which is greater than that of human bone. The requirements for a hip implant’s design are complicated because there are so many factors to be taken into an account, including the patient’s age, the geometry of the implant, the bone structure, and the biocompatibility of the material. All of these factors have an impact on how much stress and strain the implant will experience. One way to avoid future complications, including aseptic loosening or implant fracture, is to modify the shape of an existing prosthesis or use an ideal biomaterial [4]. Anthony L. Sabatini et al. This article reevaluated a number of hip stems. For all the stems, FE Analysis was carried out using different cross-sections. In addition to recording the displacement, von-Misses stress comparisons were performed for each of the stems at the predetermined places. The analysis was conducted using three materials: Ti-6Al-4V, SS316L, and Cr-Co-Mo. Ti-6Al- 4V showed less stress than the other two materials. The circular and eliptical cross-sections shaped even distribution of stresses throughout the length of the hip stem than the other cross-sections viz.; trapezoid and oval. This analysis could clarify the scenarios that should be avoided for hip implants that need stress shielding. For total hip replacements, additional anatomical designs are required [5]. M. Paglia et al. PEEK is a polyaromatic semi-crystalline thermoplastic polymer which was first developed for engineering applications by Victrex PLC in the early 1980s. It was initially made available for industrial use in the production of bearings, compressor plate valves, cable insulation, turbine blades, piston parts, and aircraft. PEEK is a polymer that is organic and colourless that is a member of the PAEK (polyaryl-ether-ketone) family. A single monomer makes it a homopolymer. This thermoplastic semi-crystalline polymer retains its outstanding chemical and mechanical characteristics even at elevated temperature. PEEK is stable at high temperatures due to its chemical compositions and structure, which facilitates simple manufacturing of PEEK implant components. When subjected to heat sterilising processes, PEEK is resistant to degradation. Due to its excellent thermal characteristics, superior wear resistance, great processability, inertness, corrosion resistance, high strength, and elastic modulus, PEEK is widely employed in both medical and engineering fields. In fact, it has an elastic modulus that is identical to human bone, indicating uniform stress distribution to the tissues around it. PEEK is a non-toxic, non-mutagenic substance that is extremely compatible with the tissues in its surrounding environment. As a result, it may be the perfect replacement for those who are allergic to titanium and other metals. PEEK has a more aesthetically pleasing appearance and doesn't have a metallic tone it seems beige with a hint of grey [6]. Jiaqi Zhang et al. Skull deformities are treated through cranioplasty, which involves raising the scalp and reshaping the skull with the original piece, titanium mesh, or solid biomaterial. Medical practitioners are increasingly using additive manufacturing technology, often known as 3D printing, to create specialized, replicas of tissues, organs, and bones, providing a viable solution with ideal anatomic fitting in the individual and skeletal reconstruction. Here, we analyze a patient from 15 years ago that had titanium mesh cranioplasty. The left eyebrow arch was weakened by the titanium mesh's unsightly appearance, which also caused a sinus tract to develop. An additively produced polyether ether ketone (PEEK) skull implant was used during cranioplasty. PEEK skull implants have been inserted successfully without any issues. To our information, this is the 1st reported case of direct use of fused filament fabrication (FFF) fabricated PEEK implant for cranial repair. The FFF-printed PEEK customized skull implant can possess simultaneously with adjustable
  • 3. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 10 Issue: 08 | Aug 2023 www.irjet.net p-ISSN: 2395-0072 © 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal | Page 278 material thickness and more complex structure, tunable mechanical properties, and low processing cost compared to traditional manufacturing processes [7]. 3. PROBLEM IDENTIFICATION At present in medical field Titanium alloys, Stainless steel and Co-Cr alloy materials are been used for the artificial human hip stems. These materials have good mechanical and physical properties, they are very strong in nature. But the stiffness / elastic modulus of these materials are too much high when compared with the bone properties, due to this when the load applied on the hip joint the maximum load will be on the implant. When hip implant carries more load than the femur bone then there will be a wear between bone and the implant this effect is called as shield stressing, by this it gets loosen very quickly before its life span. Also, weight of the implant is more. Increases in bone porosity. Decrease in bone mass. There will be a magnetism effect and anode-cathode reactions. 4. OBJECTIVES  To design for development of an artificial human hip stems based on fenestration using CAD modelling software i.e., SOLIDWORKS 2018.  To perform FEA for all different types of fenestrated artificial human hip stems for Polyether–ether–ketone (PEEK) and Ti-6Al-4V materials under different conditions such as standing, walking, jumping and running using ANSYS Workbench 2022 R2.  Results of the three different hip stems with PEEK and Ti-6Al-4V materials will be compared and one of the best will be suggested.  To reduce the weight of human hip implant by using PEEK material instead of Titanium alloys, Stainless steel and Co-Cr alloy.  To improve the strength, corrosion resistance and wear resistance of the hip stem by this stress shielding can be prevented. 5. METHODOLOGY Fig. 2 Methodology flow chart 6. DESIGN AND ANALYSIS OF HIP STEMS The CAD models have to be created in the computer environment. Here we designed a Hip Replacement of each different parts, the dimensions of each part is collected from previous record. SOLIDWORKS 2018 have been used in the modeling process. Fig. 3 Isometric view of hip implant
  • 4. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 10 Issue: 08 | Aug 2023 www.irjet.net p-ISSN: 2395-0072 © 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal | Page 279 6.1 Different types of fenestration designs Fig. 4 Types of fenestration For hip replacement, artificial hip joints come in a variety of designs and shapes. Three different hip joint designs were used in this study, as shown in Fig. 4: a hip stem without fenestration, a hip stem with a big loop fenestration, and a hip stem with a slot fenestration [8]. The only difference in dimensions was fenestration. PEEK and Ti-6Al-4V were both employed as the materials. Materials were chosen for their high biocompatibility, durability, resistance to wear and corrosion, and strength. The material was thought to be uniform, isotropic, and linearly elastic. 7. RESULTS AND DISCUSSION The final model was assembled in SOLIDWORKS and imported to ANSYS Workbench software in .IGS format and the analysis is carried out in Static structural. So in this project we are going to consider three different types of hip stem designs namely, hip stem without fenestration, hip stem with big loop fenestration and hip stem with slot fenestration. Further these hip stems should be analyzed for different loading conditions like, standing, walking, jumping and running. Then comparing the results of PEEK material with commonly used Titanium based alloy Ti-6Al- 4V for all these hip stems. The mechanical property of these materials is given below Table 1. Table 1 Mechanical properties of the materials used for hip stems [9] We are going to use some different activities like standing, walking, etc. So the stress analysis on a hip prosthesis during various patient activities, such as standing, walking, jumping as well as in running. These loads are considered according to Aftab S et al. [10] on a patient who weighs 75 Kg, with an age group of around 35 years. Due to the symmetry of the human body, when a person is standing, their weight will be distributed evenly between their two legs. For this reason, the weight was divided in half and boundary conditions were applied to the finite element model. The analysis was performed under the assumption that the subject was standing normally, with a load of 370 N applied to the head of the femur at the hip joint and the hip stem fixed when in contact with the femoral bone. Similar to this, the hip stem was examined under a variety of situations derived from various human everyday activities, such as jumping, where a force of 750 N acting on one leg was applied, for walking an additional 800 N was applied, and running, where a load of 1463 N was applied. Also shown in the Table 2. Table 2 Loading conditions of hip joint for different daily activities [10] Activity Maximum Load (N) Standing 370 Walking 800 Jumping 750 Running 1463 The following Fig. 5 shows the meshing process which chosen to be 1mm element size. Fig. 5 Meshing process Table 3 Number of Elements used in analysis Element size: 1mm Types of hip stem Nodes Elements Without fenestration 77915 32325 With big loop fenestration 86111 36782 With slot fenestration 89874 39068 Sl.N o Material s Young’s Modulu s (GPa) Density (kg/m3 ) Poisson’ s ratio Yield strengt h (MPa) 1 PEEK 3.95 1320 0.3931 170 2 Ti-6Al-4V 120 4430 0.34 900
  • 5. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 10 Issue: 08 | Aug 2023 www.irjet.net p-ISSN: 2395-0072 © 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal | Page 280 The below Fig. 6 indicates the fixed/support condition to the hip implant. Fig. 6 Support condition The Fig. 7 shows the applying of loading conditions which are mentioned in the Table 2. The force of 370 N magnitude is applied along the -y direction as shown in the following figure. Fig. 7 Applying the load 7.1 Computational Results 7.1.1 Masses of the different types of fenestrated hip stems These below mentioned masses are taken from the SOLIDWORKS software by applying the material properties of the PEEK and Ti-6Al-4V for the different hip stem designs. Fig. 8 Mass of the hip stem with big loop fenestration for PEEK material Fig. 9 Comparison of masses for different types of hip stem 7.1.2 Standing condition (370 N) The Fig. 10(a) refers to Von-misess stress produced in the hip stem without fenestration under standing condition when PEEK material is applied to the hip stem which is 34.106 MPa, which is similar for the remaining fenestrations. The Fig. 10(b) refers to Von-misess stress produced in the hip stem without fenestration under standing condition when Ti-6Al-4V material is applied to the hip stem which is 35.569 MPa, which is similar for the remaining fenestrations. 74.77 69.61 72.17 250.94 233.6 242.21 0 50 100 150 200 250 300 Hip stem without fenestration Hip stem with big loop fenestration Hip stem with slot fenestration MASS IN GRAMS TYPES OF HIP STEMS Masses of the hip stems in grams for PEEK Masses of the hip stems in grams for Ti-6Al-4V
  • 6. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 10 Issue: 08 | Aug 2023 www.irjet.net p-ISSN: 2395-0072 © 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal | Page 281 (a) (b) Fig. 10 Hip stem without fenestration under standing condition for PEEK and Ti-6Al-4V material Fig. 11 Comparison of Von-Misses stress under Standing condition From Fig. 11 the von-misses stress of hip stem with slot fenestration for PEEK material is least when compared to others in standing condition. 7.1.3 Walking condition (800 N) The Fig. 12(a) refers to Von-misess stress produced in the with bigloop fenestration under walking condition when PEEK material is applied to the hip stem which is 78.393 MPa which is similar for the remaining fenestrations. The Fig. 12(b) refers to Von-misess stress produced in the hip stem with bigloop fenestration under walking condition when Ti-6Al-4V material is applied to the hip stem which is 84.475 MPa which is similar for the remaining fenestrations. (a) (b) Fig. 12 Hip stem with big loop fenestration under walking condition for PEEK and Ti-6Al-4V material Fig. 13 Comparison of Von-Misses stress under Walking condition From Fig. 13 the von-misses stress of hip stem with slot fenestration for PEEK material is least, when compared to others in walking condition. 7.1.4 Jumping condition (750 N) During jumping with the both legs 750 N load will be applied on one leg. The Fig 14(a) refers to Von-misess stress produced in the hip stem with slot fenestratation when PEEK material is applied to the hip stem which is 63.687 MPa, which is similar for the remaining fenestrations. The Fig. 14(b) refers to Von-misess stress produced in the hip stem with slot fenestratation when Ti-6Al-4V material is applied to the hip stem which is 65.48 MPa, which is similar for the remaining fenestrations. 34.106 36.257 31.419 35.569 39.07 32.303 0 5 10 15 20 25 30 35 40 45 Without fenestration With Big Loop fenestration With Slot fenestration Von-Misses Stress in MPa Types Of Hip stem PEEK Ti-6Al-4V 73.743 78.393 67.933 76.906 84.475 69.845 0 10 20 30 40 50 60 70 80 90 Without fenestration With Big Loop fenestration With Slot fenestration Von-Misses Stress in MPa Types Of Hip stem PEEK Ti-6Al-4V
  • 7. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 10 Issue: 08 | Aug 2023 www.irjet.net p-ISSN: 2395-0072 © 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal | Page 282 (a) (b) Fig. 14 Hip stem with slot fenestration under jumping condition for PEEK and Ti-6Al-4V material Fig. 15 Comparison of Von-Misses stress under Jumping condition From Fig. 15 the von-misses stress of hip stem with slot fenestration for PEEK material is least, when compared to others jumping condition. 7.1.5 Running condition (1463 N) The Fig. 16(a) refers to Von-misess stress produced in the hip stem without fenestration under running condition when PEEK material is applied to the hip stem which is 134.86 MPa, which is similar for the remaining fenestrations. The Fig. 16(b) refers to Von-misess stress produced in the hip stem without fenestration under jumping condition when Ti-6Al-4V material is applied to the hip stem which is 140.64 MPa, which is similar for the remaining fenestrations. (a) (b) Fig. 16 Hip stem without fenestration under running condition for PEEK and Ti-6Al-4V material Fig. 17 Comparison of Von-Misses stress under Running condition From Fig. 17 the von-misses stress of hip stem with slot fenestration for PEEK material is least, when compared to others in running condition. 8. CONCLUSION In this work, three hip stem designs, namely without fenestration, with bigloop fenestration and with slot fenestration designs were modeled. For all designs, two materials PEEK and Ti-6Al-4V was used. PEEK is used because it as the similar stiffness as human femur bone and it as good wear properties, low moist absorption, low co-efficient of friction etc. All of these designs, which use materials with various physical properties, will be put through the rigours of daily activities like standing, walking, jumping, and running. Static structural analysis was carried out using ANSYS Workbench with the application of the loading boundary condition as per the work by Aftab S. et al. All three designs were found to suggest lower stress levels than the yield strength. Also, in these three designs for PEEK and Ti-6Al-4V materials under different loading conditions, the artificial human hip stem with slot fenestration for PEEK material has lesser Von-Misses stress under all different loading conditions as 69.134 73.494 63.687 72.1 79.196 65.48 0 10 20 30 40 50 60 70 80 90 Without fenestration With Big Loop fenestration With Slot fenestration Von-Misses Stress in MPa Types Of Hip stem PEEK Ti-6Al-4V 134.86 143.36 124.23 140.64 154.48 127.73 0 20 40 60 80 100 120 140 160 180 Without fenestration With Big Loop fenestration With Slot fenestration Von-Misses Stress in MPa Types Of Hip stem PEEK Ti-6Al-4V
  • 8. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 10 Issue: 08 | Aug 2023 www.irjet.net p-ISSN: 2395-0072 © 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal | Page 283 mentioned above. Ti-6Al-4V has less deformation when compared to PEEK material for all the designs which were modeled in the current work. It is observed that for all the designs under different loading conditions the von-misses stress is almost similar for PEEK and Ti-6Al-4V material. Also, weight of the PEEK material is lesser than the Ti-6Al- 4V. So, PEEK material with slot fenestration can be suggested. REFERENCES [1] L. Guo et al., “On the design evolution of hip implants: A review,” Materials and Design, vol. 216. Elsevier Ltd, Apr. 01, 2022. doi: 10.1016/j.matdes.2022.110552. [2] S. Shuib, B. Sahari, A. Ahmed Shokri, and C. Soon Chai, “The Design Improvement of Hip Implant for Total Hip Replacement (THR),” Jurnal Kejuruteraan, vol. 20, no. 1, pp. 107–113, Dec. 2008, doi: 10.17576/jkukm-2008-20-10. [3] “Clinical Outcomes Following Total Hip Replacement,” EAS Journal of Medicine and Surgery, vol. 4, no. 6, pp. 134–139, Jun. 2022, doi: 10.36349/easjms.2022.v04i06.003. [4] M. Guzmán et al., “Finite Element Assessment of a Hybrid Proposal for Hip Stem, from a Standardized Base and Different Activities,” Applied Sciences (Switzerland), vol. 12, no. 16, Aug. 2022, doi: 10.3390/app12167963. [5] A. L. Sabatini and T. Goswami, “Hip implants VII: Finite element analysis and optimization of cross- sections,” Mater Des, vol. 29, no. 7, pp. 1438–1446, 2008, doi: 10.1016/j.matdes.2007.09.002. [6] M. Paglia, M. Beretta, V. Quinzi, and S. Colombo, “PEEK polymer in orthodontics: a scoping review,” Eur J Paediatr Dent, vol. 23, no. 2, pp. 137–139, 2022, doi: 10.23804/ejpd.2022.23.02.10. [7] J. Zhang et al., “Additively manufactured polyether ether ketone (PEEK) skull implant as an alternative to titanium mesh in cranioplasty,” Int J Bioprint, vol. 9, no. 1, Nov. 2022, doi: 10.18063/ijb.v9i1.634. [8] Ikhsan, A. R. Prabowo, J. M. Sohn, and J. Triyono, “Finite element analysis of different artificial hip stem designs based on fenestration under static loading,” in Procedia Structural Integrity, 2020, vol. 27, pp. 101–108. doi: 10.1016/j.prostr.2020.07.014. [9] Y. Jun and K. Choi, “Design of patient-specific hip implants based on the 3D geometry of the human femur,” Advances in Engineering Software, vol. 41, no. 4, pp. 537–547, 2010, doi: 10.1016/j.advengsoft.2009.10.016. [10] S. G. Aftab, A. Faisal, H. Hussain, B. Sreedhara, N. R. Babu, and B. A. Praveen, “Structural analysis of human femur bone to select an alternative composite material,” Mater Today Proc, vol. 54, pp. 141–146, Jan. 2022, doi: 10.1016/j.matpr.2021.08.197.