Streamlining Instrument Calibration with
Calibration Management Software and
Documenting Field Calibrators
Documenting tools and calibration management
software combine for a paperless calibration
solution
1
Host: Transcat
Presenters:
Jim Shields and Bob Crepps
Fluke Process Calibration Group
Agenda
• Instrument calibration, challenges and elements
– Reasons to calibrate
• Establishing a process calibration maintenance management
program
– Using Calibration Management Software (CMS) to:
• Manage process instruments and calibration records
• Interface with Field Documenting Process Calibrators
• Predict instrument performance, minimize maintenance costs
– Using Documenting Process Calibrators to :
• Manage procedures or tasks downloaded from CMS
• Perform field calibrations and adjustments
• Record field calibration results
• Upload field calibration results to CMS
• Paperless calibration management program benefits
Why Calibrate?
• Regulations, standards and trade
• Custody Transfer
• Internal/External Quality Programs (like ISO-9000)
• Plant Safety and Uptime
• EPA, FDA, NRC, OSHA and other mandated
regulations
Why Calibrate?
• All instruments change with
– Time, temperature, humidity, environmental
exposure, vibration, wear and tear
– No two instruments change exactly the same
• Scheduled calibrations reduce process
downtime risk
– Improves the performance and reliability of critical
assets
• Product Quality / Yield
– Calibration helps processes comparable to traceable
calibration standards
• Helps minimize process deviations
– Helps drive process optimization and cost savings
Industry changes
• Regulatory forces increasing
• The plant isn’t getting smaller but shrinking budgets are reducing
the maintenance staff
– Driving efficiencies and the need to “Do more with less”
• Experienced instrumentation professionals retiring
– Technical and Vocational Schools working to fill voids
• Plant modernizations lead to control system changes
• Competitive pressure offshore push the need for plant
optimization, streamlined operations and maintenance practices
Process Calibration
Elements:
• Calibration equipment, procedures, instrument documentation,
calibration management software (CMS), trained personnel
Benefits:
• Consistent process performance and efficiency, improved product
quality, reduced downtime, safety, regulatory compliance,
maximized use of skilled maintenance personnel
Calibration, definitions…
Calibration: Comparison of a measured value to a traceable
measurement standard
– Adjust to agree with the standard if necessary
• Accuracy: The closeness of agreement between an observed
value and that of a traceable measurement standard
• Adjustment: Manual or digital compensation of the output signal
of device with a known input value applied
– Necessary if a device is not performing as required
• Need to understand adjustment tolerance
• Need communication tool for digital adjustment
Adjustment is part of a calibration process, it is not calibration by
itself.
Fluke Petrochemical Corporation By Tag ID
RESULTS FROM FLUKE 702 CALIBRATIONS
2-Feb-96
Fluke Petrochemical Corp.
By:
Title:
Cert. No.: 1
Model: FLUKE 702 Serial:6150203 User ID: JSmith 010
Cal. Status: 14 Loop Power: 24V
As found 25-Jan-96 4:46
Measure: 4.000 to 20.000 mA
Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90
Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321
Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234
1,996.001.0025 Input Value: Units Output value: Units Error % Status MeasUEV
3.906 mA 40.0 °C -0.59 Failed Disabled
11.918 mA 55.0 °C -0.51 Failed Disabled
19.923 mA 70.0 °C -0.48 Failed Disabled
11.918 mA 55.0 °C -0.51 Failed Disabled
3.907 mA 40.0 °C -0.58 Failed Disabled
Model: FLUKE 702 Serial:6150203 User ID: JSmith 010
Cal. Status: 14 Loop Power: 24V
As left 25-Jan-96 4:53
Measure: 4.000 to 20.000 mA
Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90
Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321
Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234
1,996.001.0025 Input Value: Units Output value: Units Error % Status MeasUEV
3.998 mA 40.0 °C -0.01 Passed Disabled
12.011 mA 55.0 °C 0.07 Passed Disabled
20.018 mA 70.0 °C 0.12 Passed Disabled
12.016 mA 55.0 °C 0.10 Passed Disabled
4.003 mA 40.0 °C 0.02 Passed Disabled
Fluke Petrochemical Corporation By Tag ID
RESULTS FROM FLUKE 702 CALIBRATIONS
2-Feb-96
Fluke Petrochemical Corp.
By:
Title:
Cert. No.: 1
Model: FLUKE 702 Serial:6150203 User ID: JSmith 010
Cal. Status: 14 Loop Power: 24V
As found 25-Jan-96 4:46
Measure: 4.000 to 20.000 mA
Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90
Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321
Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234
1,996.001.0025 Input Value: Units Output value: Units Error % Status MeasUEV
3.906 mA 40.0 °C -0.59 Failed Disabled
11.918 mA 55.0 °C -0.51 Failed Disabled
19.923 mA 70.0 °C -0.48 Failed Disabled
11.918 mA 55.0 °C -0.51 Failed Disabled
3.907 mA 40.0 °C -0.58 Failed Disabled
Model: FLUKE 702 Serial:6150203 User ID: JSmith 010
Cal. Status: 14 Loop Power: 24V
As left 25-Jan-96 4:53
Measure: 4.000 to 20.000 mA
Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90
Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321
Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234
1,996.001.0025 Input Value: Units Output value: Units Error % Status MeasUEV
3.998 mA 40.0 °C -0.01 Passed Disabled
12.011 mA 55.0 °C 0.07 Passed Disabled
20.018 mA 70.0 °C 0.12 Passed Disabled
12.016 mA 55.0 °C 0.10 Passed Disabled
4.003 mA 40.0 °C 0.02 Passed Disabled
Fluke Petrochemical Corporation By Tag ID
RESULTS FROM FLUKE 702 CALIBRATIONS
2-Feb-96
Fluke Petrochemical Corp.
By:
Title:
Cert. No.: 1
Model: FLUKE 702 Serial:6150203 User ID: JSmith 010
Cal. Status: 14 Loop Power: 24V
As found 25-Jan-96 4:46
Measure: 4.000 to 20.000 mA
Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90
Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321
Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234
1,996.001.0025 Input Value: Units Output value: Units Error % Status MeasUEV
3.906 mA 40.0 °C -0.59 Failed Disabled
11.918 mA 55.0 °C -0.51 Failed Disabled
19.923 mA 70.0 °C -0.48 Failed Disabled
11.918 mA 55.0 °C -0.51 Failed Disabled
3.907 mA 40.0 °C -0.58 Failed Disabled
Model: FLUKE 702 Serial:6150203 User ID: JSmith 010
Cal. Status: 14 Loop Power: 24V
As left 25-Jan-96 4:53
Measure: 4.000 to 20.000 mA
Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90
Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321
Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234
1,996.001.0025 Input Value: Units Output value: Units Error % Status MeasUEV
3.998 mA 40.0 °C -0.01 Passed Disabled
12.011 mA 55.0 °C 0.07 Passed Disabled
20.018 mA 70.0 °C 0.12 Passed Disabled
12.016 mA 55.0 °C 0.10 Passed Disabled
4.003 mA 40.0 °C 0.02 Passed Disabled
Fluke Petrochemical Corporation By Tag ID
RESULTS FROM FLUKE 702 CALIBRATIONS
2-Feb-96
Fluke Petrochemical Corp.
By:
Title:
Cert. No.: 1
Model: FLUKE 702 Serial:6150203 User ID: JSmith 010
Cal. Status: 14 Loop Power: 24V
As found 25-Jan-96 4:46
Measure: 4.000 to 20.000 mA
Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90
Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321
Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234
1,996.001.0025Input Value: Units Output value: Units Error % Status MeasUEV
3.906 mA 40.0 °C -0.59 Failed Disabled
11.918 mA 55.0 °C -0.51 Failed Disabled
19.923 mA 70.0 °C -0.48 Failed Disabled
11.918 mA 55.0 °C -0.51 Failed Disabled
3.907 mA 40.0 °C -0.58 Failed Disabled
Model: FLUKE 702 Serial:6150203 User ID: JSmith 010
Cal. Status: 14 Loop Power: 24V
As left 25-Jan-96 4:53
Measure: 4.000 to 20.000 mA
Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90
Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321
Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234
1,996.001.0025Input Value: Units Output value: Units Error % Status MeasUEV
3.998 mA 40.0 °C -0.01 Passed Disabled
12.011 mA 55.0 °C 0.07 Passed Disabled
20.018 mA 70.0 °C 0.12 Passed Disabled
12.016 mA 55.0 °C 0.10 Passed Disabled
4.003 mA 40.0 °C 0.02 Passed Disabled
Fluke Petrochemical Corporation By Tag ID
RESULTS FROM FLUKE 702 CALIBRATIONS
2-Feb-96
Fluke Petrochemical Corp.
By:
Title:
Cert. No.: 1
Model: FLUKE 702 Serial:6150203 User ID: JSmith 010
Cal. Status: 14 Loop Power: 24V
As found 25-Jan-96 4:46
Measure: 4.000 to 20.000 mA
Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90
Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321
Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234
1,996.001.0025Input Value: Units Output value: Units Error % Status MeasUEV
3.906 mA 40.0 °C -0.59 Failed Disabled
11.918 mA 55.0 °C -0.51 Failed Disabled
19.923 mA 70.0 °C -0.48 Failed Disabled
11.918 mA 55.0 °C -0.51 Failed Disabled
3.907 mA 40.0 °C -0.58 Failed Disabled
Model: FLUKE 702 Serial:6150203 User ID: JSmith 010
Cal. Status: 14 Loop Power: 24V
As left 25-Jan-96 4:53
Measure: 4.000 to 20.000 mA
Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90
Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321
Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234
1,996.001.0025Input Value: Units Output value: Units Error % Status MeasUEV
3.998 mA 40.0 °C -0.01 Passed Disabled
12.011 mA 55.0 °C 0.07 Passed Disabled
20.018 mA 70.0 °C 0.12 Passed Disabled
12.016 mA 55.0 °C 0.10 Passed Disabled
4.003 mA 40.0 °C 0.02 Passed Disabled
Fluke Petrochemical Corporation By Tag ID
RESULTS FROM FLUKE 702 CALIBRATIONS
2-Feb-96
Fluke Petrochemical Corp.
By:
Title:
Cert. No.: 1
Model: FLUKE 702 Serial:6150203 User ID: JSmith 010
Cal. Status: 14 Loop Power: 24V
As found 25-Jan-96 4:46
Measure: 4.000 to 20.000 mA
Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90
Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321
Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234
1,996.001.0025Input Value: Units Output value: Units Error % Status MeasUEV
3.906 mA 40.0 °C -0.59 Failed Disabled
11.918 mA 55.0 °C -0.51 Failed Disabled
19.923 mA 70.0 °C -0.48 Failed Disabled
11.918 mA 55.0 °C -0.51 Failed Disabled
3.907 mA 40.0 °C -0.58 Failed Disabled
Model: FLUKE 702 Serial:6150203 User ID: JSmith 010
Cal. Status: 14 Loop Power: 24V
As left 25-Jan-96 4:53
Measure: 4.000 to 20.000 mA
Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90
Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321
Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234
1,996.001.0025Input Value: Units Output value: Units Error % Status MeasUEV
3.998 mA 40.0 °C -0.01 Passed Disabled
12.011 mA 55.0 °C 0.07 Passed Disabled
20.018 mA 70.0 °C 0.12 Passed Disabled
12.016 mA 55.0 °C 0.10 Passed Disabled
4.003 mA 40.0 °C 0.02 Passed Disabled
National
Standards
Traceability
• Unbroken chain of
calibration.
• Documented
proof.
Process Calibration requires:
Traceable calibration tools, source functions:
– Temperature, electronic simulation of thermocouples or RTDs
• Does not test the sensor installed in a process
– Temperature, dry block or temperature bath calibrator
• Tests the temperature sensor and transmitter electronics
– Pressure, hand pump, or other means to apply pressure to the unit under
test
– Pressure measurement reference to measure the pressure generated (by a
pressure source) and applied to the UUT
• Transfer standard approach
– Other Electronic signals
• Frequency for flow, speed and rotation
• Current for I/P and analog I/O
• Voltage for signal conditioners and analog I/O
Process Calibration requires:
Traceable calibration tools, measurement functions:
– Temperature, thermocouple or RTD thermometer
• Measure the process temperature with a reference probe directly
and compare with control system
– Pressure, measure the pressure output from an I/P or other
pneumatic control signals
– Other Electronic signals
• Frequency for flow, speed and rotation
• Current for all transmitter mA loop signal outputs
• Voltage for signal conditioners other voltage output devices
– Many transmitters have voltage output instead of mA output.
Why document calibration, how?
Calibration Records may be:
– required for regulations
– required for quality standards
– required by your customers
Calibration records can:
– assist in understanding the performance of a process
– be used to determine a devices suitability for a process
– help assess if a device needs repair
– be used to help adjust maintenance intervals
• Calibration records can:
– be manually recorded on a paper form
– be recorded on a palm device or portable computer
– be recorded using a calibrator with documenting capabilities
Calibration management software + documenting calibrators enables easy
documenting and analysis of field calibrations
Calibration process, setup
• Identify critical instruments
– Safety instrumented systems
– Environmental critical (water and air outflow)
– Quality critical: directly effect product quality and value
– Revenue critical instruments:
• Custody transfer, trade related measurements
• Define the interval to maintain desired accuracy
– Instrument should perform in tolerance when tested
• Specifications and test tolerances need careful consideration
– Manufacturers claimed accuracies can be misleading
– Test tolerances should be specified by process engineer
– The accuracy of both the instrument being tested and the calibrator need to
be considered
• Define the procedures for the identified tags
Calibration process; execution
• Identify the tags that need calibration
• Organize the work into logical routes
• Collect the necessary procedures, documentation and tools
• Isolate the device from the process
– Lock-out, Tag-out procedure if warranted
• Connect the calibrator, begin to calibrate
– Perform the As-Found test, document the results
– Adjust the transmitter if the measured errors are greater than the
adjustment threshold defined by the procedure.
– Perform the As-Left test, document the results
– Complete the documented record with Tag#, time, date, Serial numbers of
the TAG, the test equipment used and its’ calibration data
Documented Calibration Requires – As Found Data
• Testing the instrument to
determine “As-Found” performance
• Accurate knowledge of the source
value is required
– The sourced value and measured values
need to be traceable
– The total uncertainty of the source should
be 3 to 4 times better than the device
under test
• Pass-fail assessment (fail)
• Adjustment required!
Graph of Percent Error
-0.3
-0.2
-0.1
0
0.1
0.2
0.3
0.4
0
10
20
30
40
50
60
70
30
90
100
Input % of Span
%Error
+Tolerance
-Tolerance
Graph of Percent Error
-0.4
-0.3
-0.2
-0.1
0
0.1
0.2
0.3
0.4
0
10
20
30
40
50
60
70
30
90
100
Input % of Span
%Error
+Tolerance
-Tolerance
Documented Calibration Requires – As Found Data
• Re-test to verify adjustments.
• “As Left” test documents the
condition of the transmitter after
adjustment
• Time and date of the test need to be
documented
• Instrument tag, serial number and
technician data need to be
documented
Paperless Instrument Calibration management
Why use a documenting calibrator?
• Multifunction source and measure
• Automatically capture results, date/time
user and tag information
• Eliminates manual transcription and
legibility errors
• Eliminates “pencil whipped” results
• Ensures calibrations performed per
documented procedures
• Coupled with CMS complete a paperless
documenting field calibration system
• Assists to comply with regulations and
audits
• Many have HART communication
capability built in
Why use a documenting calibrator?
• Multifunction source and measure
• Automatically capture results, date/time
user and tag information
• Eliminates manual transcription and
legibility errors
• Eliminates “pencil whipped” results
• Ensures calibrations performed per
documented procedures
• Coupled with CMS complete a paperless
documenting field calibration system
• Assists to comply with regulations and
audits
• Many have HART communication
capability built in
Why use a documenting calibrator?
• Multifunction source and measure
• Automatically capture results, date/time
user and tag information
• Eliminates manual transcription and
legibility errors
• Eliminates “pencil whipped” results
• Ensures calibrations performed per
documented procedures
• Coupled with CMS complete a paperless
documenting field calibration system
• Assists to comply with regulations and
audits
• Many have HART communication
capability built in
• Manage calibration records for field process instrumentation
• Print reports or run queries for instruments due for calibration to
help assess, schedule and manage
• Connect to Documenting Process Calibrators (DPCs)
– Load instrument procedures and upload documented calibrations
• Documented results enable analysis for drift, reliability or
maintenance costs
• Many CMS applications have connectivity to CMMS packages like
SAP, Oracle and Maximo
Why use Calibration Management Software (CMS)?
Examples:
• Fluke DPCTrack2
• Prime Technologies
ProCalV5
• Emerson AMS
• Yokogawa PRM
• Meridium
• Intergraph SmartPlant
Why use Calibration Management Software (CMS)?
• CMS completes the documenting
solution when coupled with a
documenting calibrator
• Many different solutions from fully
integrated CMMS to stand alone CMS
• Vary in complexity and price, $4K to
$100K
CMS Software Example, DPCTrack2
Modern calibration management software packages are
easy to use and deliver good reasons to adopt.
• Example entry level
single user CMS
application.
• Upgradeable to
ProCalV5 Enterprise
CMS software
What should a CMS package do?
• Manage and document process instrumentation performance
• Create custom procedures for download to DPCs
• Upload results from DPCs and store for later retrieval or analysis
• Print a variety of reports that support ISO 9000 or other
documentation needs
• Search for particular tags and report their history
• Search for tags due for calibration in the past or future
• Keep track of all Tags “touched” by a particular piece of test
equipment (reverse traceability)
Instrument View Search Queries
• Search queries can be customized to find populations of tags
differentiated by:
>Cal Due Date
>Tag Id
>S/N
>Location
>Device type
>Location
>Other fields
Tag Instrument Data
• Input the data describing the tag
Transmitter Data
• Describe the input, output and expected tolerance of the
transmitter
Upload and download information to DPCs
• Tags identified in the date based search are loaded into the
calibrator
Live DPCTrack2 demo
• Creating tags using example templates
• Error% calculation explanation
• Sorting the database for calibrations using database query tools
• Loading tags for a calibration route into the Fluke-729
Task Selection, (DPC)
• Tags downloaded to the DPC
available for selection
• Supervisor can assign as a
route to a technician
• Usage count indicates whether
a task has been completed
using the downloaded task
• Tasks must be performed for
usage count to increment
Tag
Number
Task Usage
& completion
Count
Downloaded transmitter test parameters
Predefined procedures
automatically fill in test
parameters:
• Source and measure end
points,
• Test tolerance,
• Test strategy (number of tests)
• Leak test available on 729 to
verify test setup for leakage
Pressure documenting made easy
• On the test screen:
– Applied pressure (top)
– Measured mA value (4.000 mA would be
the ideal value for this device)
– Error % of Range/Span
• Green for in tolerance
• Red for out of tolerance
– AutoTest and the 729 will automatically
apply the test pressures, measure mA
signals calculate Error % and document
• Out of tolerance tests flagged in Red
• Tag number, serial number and technician
name recorded.
Transmitter Adjustment, HART
PV zero trim
– Zero out transmitter offset
Output Trim
– Fetch retrieves the current value being
measured
– Trim 4 mA and Trim 20 mA performs mA
adjustment
Trim to applied Values
– Apply Expected 0 and 100% range values
– Adjust 4 and 20 mA outputs to match applied
pressures.
Transmitter calibration As-Left measurement
The As-Left test will repeat the initial
tests to verify and document the “As
Left”, post adjustment condition of
the transmitter
• Press “As Left”
• This As-Left result is successful as
noted by Pass
• The documented test is now ready
for upload
If the test fails As Left, additional
adjustment and additional As-Left
test(s) may be necessary.
Live DPCTrack2 demo
• Uploading documented calibration results
• Review and authorize the uploaded results
• Viewing the calibration certificate
• View other reporting capabilities
Uploaded Test Results
As-Found, As-Left results uploaded and authorized to
add to the database for analysis and printing
Calibration record
• Documented
calibration stored
for recall and
printing as
needed.
• Calibration date
updated in
database and
next calibration
due date is set.
Takeaways
• Efficiency of route based calibrations can save up to 50% on
calibration maintenance labor making technicians more efficient
• Multifunction documenting calibrators minimize returning to the
shop to get additional tools
• The efficiency of performing pressure tests with the 729
Automatic Pressure Calibrator both saves time making the
technician efficient but also improves the quality of the
calibration result.
• Documented calibration results can assist in passing audits or
other requirements where calibration records are required
Summary
• Calibration is a critical practice to remove variability and maintain
safety in a process
• Documentation of calibration history enables advanced analysis
techniques
• Field documentation with a documenting calibrator eliminates
transcription errors and enables data to be uploaded to
Calibration Management Software (CMS)
• CMS can be used to sort the calibrations due into routes for
download, maximizes use of skilled personnel
• CMS assists in managing and interpreting documented calibration
records and day to day workload assignments
Appendix, Fluke overview
Fluke Corporation
Fluke Headquarters, Everett
Washington, USA
From John Fluke’s basement to an industry
leader
Fluke Corporation
• 3000 employees worldwide
• Presence in more than 100 countries
• Products include:
• Process Calibration Tools
• DMMs
• Calibration Standards
• High Performance Oscilloscopes
• Power Quality Analyzers
• Electrical Testers
• Thermal imaging
• Condition monitoring
• More!
Current Offer
• Now through June 30th
– Free Fluke tools with purchases $250+
– www.transcat.com
42
Questions or Comments?
Email Nicole VanWert-Quinzi
nicole.vanwert@Transcat.com
Transcat: 800-800-5001
www.Transcat.com
For related product information, go to:
www.Transcat.com/Fluke

More Related Content

What's hot

Industrial Temperature Calibration Selection Guide by Fluke Calibration
Industrial Temperature Calibration Selection Guide by Fluke CalibrationIndustrial Temperature Calibration Selection Guide by Fluke Calibration
Industrial Temperature Calibration Selection Guide by Fluke CalibrationFluke Calibration
 
How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015
How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015
How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015Transcat
 
Evaluating Out Of Tolerance Instruments Webinar
Evaluating Out Of Tolerance Instruments WebinarEvaluating Out Of Tolerance Instruments Webinar
Evaluating Out Of Tolerance Instruments WebinarTranscat
 
Overhead & Underground Electric Safety Webinar Presented by Amprobe
Overhead & Underground Electric Safety Webinar Presented by AmprobeOverhead & Underground Electric Safety Webinar Presented by Amprobe
Overhead & Underground Electric Safety Webinar Presented by AmprobeTranscat
 
Improving Safety in the Industrial Workplace with Infrared Technology
Improving Safety in the Industrial Workplace with Infrared TechnologyImproving Safety in the Industrial Workplace with Infrared Technology
Improving Safety in the Industrial Workplace with Infrared TechnologyTranscat
 
Fluke Data Acquisition System Selection Guide. Hubungi PT. Siwali Swantika 02...
Fluke Data Acquisition System Selection Guide. Hubungi PT. Siwali Swantika 02...Fluke Data Acquisition System Selection Guide. Hubungi PT. Siwali Swantika 02...
Fluke Data Acquisition System Selection Guide. Hubungi PT. Siwali Swantika 02...PT. Siwali Swantika
 
Transcat's "Calibration's Role in the Manufacturing Jigsaw Puzzle" Presentation
Transcat's "Calibration's Role in the Manufacturing Jigsaw Puzzle" PresentationTranscat's "Calibration's Role in the Manufacturing Jigsaw Puzzle" Presentation
Transcat's "Calibration's Role in the Manufacturing Jigsaw Puzzle" PresentationTranscat
 
Brosur Fluke 5320A. Hubungi PT. Siwali Swantika 021-45850618
Brosur Fluke 5320A. Hubungi PT. Siwali Swantika 021-45850618Brosur Fluke 5320A. Hubungi PT. Siwali Swantika 021-45850618
Brosur Fluke 5320A. Hubungi PT. Siwali Swantika 021-45850618PT. Siwali Swantika
 
Katalog fluke-ir-and-contact-thermometers-by-tridinamika
Katalog fluke-ir-and-contact-thermometers-by-tridinamikaKatalog fluke-ir-and-contact-thermometers-by-tridinamika
Katalog fluke-ir-and-contact-thermometers-by-tridinamikaPT. Tridinamika Jaya Instrument
 
Datasheet Fluke Explosive Area. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke Explosive Area. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke Explosive Area. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke Explosive Area. Hubungi PT. Siwali Swantika 021-45850618PT. Siwali Swantika
 
Flow Instruments for Test and Measurement Selection Guide by Badger Meter
Flow Instruments for Test and Measurement Selection Guide by Badger MeterFlow Instruments for Test and Measurement Selection Guide by Badger Meter
Flow Instruments for Test and Measurement Selection Guide by Badger MeterThompson Equipment Company
 
Temperature Mapping_Kevin Loomis_Panasonic Healthcare
Temperature Mapping_Kevin Loomis_Panasonic HealthcareTemperature Mapping_Kevin Loomis_Panasonic Healthcare
Temperature Mapping_Kevin Loomis_Panasonic HealthcareBioPharmEquip LLC
 
Distributor presentation mx
Distributor presentation   mxDistributor presentation   mx
Distributor presentation mxmacnaught
 
Elevated Body Temperature Measurement with IR Thermometers and TI Cameras Web...
Elevated Body Temperature Measurement with IR Thermometers and TI Cameras Web...Elevated Body Temperature Measurement with IR Thermometers and TI Cameras Web...
Elevated Body Temperature Measurement with IR Thermometers and TI Cameras Web...Transcat
 
Validation Dataloggers - Jason Daoust
Validation Dataloggers - Jason Daoust Validation Dataloggers - Jason Daoust
Validation Dataloggers - Jason Daoust ISPE_LA
 
Vapour pressure-model-1354 atac
Vapour pressure-model-1354 atacVapour pressure-model-1354 atac
Vapour pressure-model-1354 atacEuropean Tech Serv
 
Lyssy l80 5000 Water Vapour Permeability Tester
Lyssy l80 5000 Water Vapour Permeability TesterLyssy l80 5000 Water Vapour Permeability Tester
Lyssy l80 5000 Water Vapour Permeability TesterSystechIllinois
 

What's hot (20)

Industrial Temperature Calibration Selection Guide by Fluke Calibration
Industrial Temperature Calibration Selection Guide by Fluke CalibrationIndustrial Temperature Calibration Selection Guide by Fluke Calibration
Industrial Temperature Calibration Selection Guide by Fluke Calibration
 
How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015
How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015
How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015
 
Evaluating Out Of Tolerance Instruments Webinar
Evaluating Out Of Tolerance Instruments WebinarEvaluating Out Of Tolerance Instruments Webinar
Evaluating Out Of Tolerance Instruments Webinar
 
Overhead & Underground Electric Safety Webinar Presented by Amprobe
Overhead & Underground Electric Safety Webinar Presented by AmprobeOverhead & Underground Electric Safety Webinar Presented by Amprobe
Overhead & Underground Electric Safety Webinar Presented by Amprobe
 
Improving Safety in the Industrial Workplace with Infrared Technology
Improving Safety in the Industrial Workplace with Infrared TechnologyImproving Safety in the Industrial Workplace with Infrared Technology
Improving Safety in the Industrial Workplace with Infrared Technology
 
Fluke Data Acquisition System Selection Guide. Hubungi PT. Siwali Swantika 02...
Fluke Data Acquisition System Selection Guide. Hubungi PT. Siwali Swantika 02...Fluke Data Acquisition System Selection Guide. Hubungi PT. Siwali Swantika 02...
Fluke Data Acquisition System Selection Guide. Hubungi PT. Siwali Swantika 02...
 
Transcat's "Calibration's Role in the Manufacturing Jigsaw Puzzle" Presentation
Transcat's "Calibration's Role in the Manufacturing Jigsaw Puzzle" PresentationTranscat's "Calibration's Role in the Manufacturing Jigsaw Puzzle" Presentation
Transcat's "Calibration's Role in the Manufacturing Jigsaw Puzzle" Presentation
 
Brosur Fluke 5320A. Hubungi PT. Siwali Swantika 021-45850618
Brosur Fluke 5320A. Hubungi PT. Siwali Swantika 021-45850618Brosur Fluke 5320A. Hubungi PT. Siwali Swantika 021-45850618
Brosur Fluke 5320A. Hubungi PT. Siwali Swantika 021-45850618
 
Katalog fluke-ir-and-contact-thermometers-by-tridinamika
Katalog fluke-ir-and-contact-thermometers-by-tridinamikaKatalog fluke-ir-and-contact-thermometers-by-tridinamika
Katalog fluke-ir-and-contact-thermometers-by-tridinamika
 
Datasheet Fluke Explosive Area. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke Explosive Area. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke Explosive Area. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke Explosive Area. Hubungi PT. Siwali Swantika 021-45850618
 
Flow Instruments for Test and Measurement Selection Guide by Badger Meter
Flow Instruments for Test and Measurement Selection Guide by Badger MeterFlow Instruments for Test and Measurement Selection Guide by Badger Meter
Flow Instruments for Test and Measurement Selection Guide by Badger Meter
 
Temperature Mapping_Kevin Loomis_Panasonic Healthcare
Temperature Mapping_Kevin Loomis_Panasonic HealthcareTemperature Mapping_Kevin Loomis_Panasonic Healthcare
Temperature Mapping_Kevin Loomis_Panasonic Healthcare
 
Distributor presentation mx
Distributor presentation   mxDistributor presentation   mx
Distributor presentation mx
 
Level and Pressure Instrumentation for Pulp & Paper Production
Level and Pressure Instrumentation for Pulp & Paper ProductionLevel and Pressure Instrumentation for Pulp & Paper Production
Level and Pressure Instrumentation for Pulp & Paper Production
 
Elevated Body Temperature Measurement with IR Thermometers and TI Cameras Web...
Elevated Body Temperature Measurement with IR Thermometers and TI Cameras Web...Elevated Body Temperature Measurement with IR Thermometers and TI Cameras Web...
Elevated Body Temperature Measurement with IR Thermometers and TI Cameras Web...
 
Azbil AX Series Vortex Flow Meters
Azbil AX Series Vortex Flow MetersAzbil AX Series Vortex Flow Meters
Azbil AX Series Vortex Flow Meters
 
Combination of pH and ORP electrode
Combination of pH and ORP electrodeCombination of pH and ORP electrode
Combination of pH and ORP electrode
 
Validation Dataloggers - Jason Daoust
Validation Dataloggers - Jason Daoust Validation Dataloggers - Jason Daoust
Validation Dataloggers - Jason Daoust
 
Vapour pressure-model-1354 atac
Vapour pressure-model-1354 atacVapour pressure-model-1354 atac
Vapour pressure-model-1354 atac
 
Lyssy l80 5000 Water Vapour Permeability Tester
Lyssy l80 5000 Water Vapour Permeability TesterLyssy l80 5000 Water Vapour Permeability Tester
Lyssy l80 5000 Water Vapour Permeability Tester
 

Similar to Transcat and Fluke Webinar: Streamlining Instrument Calibration

Managing Instrument Calibration and Maintenance Presented by Fluke.pptx
Managing Instrument Calibration and Maintenance Presented by Fluke.pptxManaging Instrument Calibration and Maintenance Presented by Fluke.pptx
Managing Instrument Calibration and Maintenance Presented by Fluke.pptxTranscat
 
Datasheet Fluke 726. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 726. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 726. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 726. Hubungi PT. Siwali Swantika 021-45850618PT. Siwali Swantika
 
Datasheet Fluke 1586A Super-DAQ Extended Specification. Hubungi PT. Siwali Sw...
Datasheet Fluke 1586A Super-DAQ Extended Specification. Hubungi PT. Siwali Sw...Datasheet Fluke 1586A Super-DAQ Extended Specification. Hubungi PT. Siwali Sw...
Datasheet Fluke 1586A Super-DAQ Extended Specification. Hubungi PT. Siwali Sw...PT. Siwali Swantika
 
Runlite 38mm Filament LED LM-80 test report.PDF
Runlite  38mm Filament LED  LM-80 test report.PDFRunlite  38mm Filament LED  LM-80 test report.PDF
Runlite 38mm Filament LED LM-80 test report.PDFNancy Yu
 
Runlite 26mm Filament LED LM-80 test report.PDF
Runlite  26mm Filament LED  LM-80 test report.PDFRunlite  26mm Filament LED  LM-80 test report.PDF
Runlite 26mm Filament LED LM-80 test report.PDFNancy Yu
 
Scope of Accreditation
Scope of AccreditationScope of Accreditation
Scope of AccreditationGRAFTEL LLC
 
Applied Reliability Symposium 2009 M Turner
Applied Reliability Symposium 2009 M TurnerApplied Reliability Symposium 2009 M Turner
Applied Reliability Symposium 2009 M TurnerMark Turner CRP
 
Datasheet Fluke 9190A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 9190A. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 9190A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 9190A. Hubungi PT. Siwali Swantika 021-45850618PT. Siwali Swantika
 
Datasheet Fluke 5502A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5502A. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 5502A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5502A. Hubungi PT. Siwali Swantika 021-45850618PT. Siwali Swantika
 
Split core ct, rogowski coil (2016)
Split core ct, rogowski coil (2016)Split core ct, rogowski coil (2016)
Split core ct, rogowski coil (2016)Eric Bin
 
Datasheet Fluke 5522A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5522A. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 5522A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5522A. Hubungi PT. Siwali Swantika 021-45850618PT. Siwali Swantika
 
rae_mic2_sensors_and_signals for study purposes
rae_mic2_sensors_and_signals for study purposesrae_mic2_sensors_and_signals for study purposes
rae_mic2_sensors_and_signals for study purposes134483
 
Electronic Ballast Tester
Electronic Ballast TesterElectronic Ballast Tester
Electronic Ballast TesterLisun Group
 

Similar to Transcat and Fluke Webinar: Streamlining Instrument Calibration (20)

Managing Instrument Calibration and Maintenance Presented by Fluke.pptx
Managing Instrument Calibration and Maintenance Presented by Fluke.pptxManaging Instrument Calibration and Maintenance Presented by Fluke.pptx
Managing Instrument Calibration and Maintenance Presented by Fluke.pptx
 
M4100
M4100M4100
M4100
 
Datasheet Fluke 726. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 726. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 726. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 726. Hubungi PT. Siwali Swantika 021-45850618
 
Datasheet Fluke 1586A Super-DAQ Extended Specification. Hubungi PT. Siwali Sw...
Datasheet Fluke 1586A Super-DAQ Extended Specification. Hubungi PT. Siwali Sw...Datasheet Fluke 1586A Super-DAQ Extended Specification. Hubungi PT. Siwali Sw...
Datasheet Fluke 1586A Super-DAQ Extended Specification. Hubungi PT. Siwali Sw...
 
Runlite 38mm Filament LED LM-80 test report.PDF
Runlite  38mm Filament LED  LM-80 test report.PDFRunlite  38mm Filament LED  LM-80 test report.PDF
Runlite 38mm Filament LED LM-80 test report.PDF
 
Runlite 26mm Filament LED LM-80 test report.PDF
Runlite  26mm Filament LED  LM-80 test report.PDFRunlite  26mm Filament LED  LM-80 test report.PDF
Runlite 26mm Filament LED LM-80 test report.PDF
 
DTT31
DTT31DTT31
DTT31
 
Scope of Accreditation
Scope of AccreditationScope of Accreditation
Scope of Accreditation
 
2 n6509g
2 n6509g2 n6509g
2 n6509g
 
TTR35
TTR35TTR35
TTR35
 
Applied Reliability Symposium 2009 M Turner
Applied Reliability Symposium 2009 M TurnerApplied Reliability Symposium 2009 M Turner
Applied Reliability Symposium 2009 M Turner
 
Datasheet Fluke 9190A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 9190A. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 9190A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 9190A. Hubungi PT. Siwali Swantika 021-45850618
 
EIT_Presentation
EIT_PresentationEIT_Presentation
EIT_Presentation
 
Datasheet Fluke 5502A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5502A. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 5502A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5502A. Hubungi PT. Siwali Swantika 021-45850618
 
Split core ct, rogowski coil (2016)
Split core ct, rogowski coil (2016)Split core ct, rogowski coil (2016)
Split core ct, rogowski coil (2016)
 
Promag 10D
Promag 10DPromag 10D
Promag 10D
 
Datasheet Fluke 5522A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5522A. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 5522A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5522A. Hubungi PT. Siwali Swantika 021-45850618
 
DTT35
DTT35DTT35
DTT35
 
rae_mic2_sensors_and_signals for study purposes
rae_mic2_sensors_and_signals for study purposesrae_mic2_sensors_and_signals for study purposes
rae_mic2_sensors_and_signals for study purposes
 
Electronic Ballast Tester
Electronic Ballast TesterElectronic Ballast Tester
Electronic Ballast Tester
 

More from Transcat

Predictive Maintenance Best Practices Presented by Fluke.pptx
Predictive Maintenance Best Practices Presented by Fluke.pptxPredictive Maintenance Best Practices Presented by Fluke.pptx
Predictive Maintenance Best Practices Presented by Fluke.pptxTranscat
 
Tools and Techniques for Commissioning and Maintaining PV Systems
Tools and Techniques for Commissioning and Maintaining PV SystemsTools and Techniques for Commissioning and Maintaining PV Systems
Tools and Techniques for Commissioning and Maintaining PV SystemsTranscat
 
Use of a Druck Pressure Calibrator as a Trouble Shooting Tool Presented by Druck
Use of a Druck Pressure Calibrator as a Trouble Shooting Tool Presented by DruckUse of a Druck Pressure Calibrator as a Trouble Shooting Tool Presented by Druck
Use of a Druck Pressure Calibrator as a Trouble Shooting Tool Presented by DruckTranscat
 
Electrical Measurement Safety and Best Practices Webinar Presented by Fluke
Electrical Measurement Safety and Best Practices Webinar Presented by FlukeElectrical Measurement Safety and Best Practices Webinar Presented by Fluke
Electrical Measurement Safety and Best Practices Webinar Presented by FlukeTranscat
 
Solar Market Opportunities & Solutions - Presented by Fluke
Solar Market Opportunities & Solutions - Presented by FlukeSolar Market Opportunities & Solutions - Presented by Fluke
Solar Market Opportunities & Solutions - Presented by FlukeTranscat
 
Pinpoint the leading causes of downtime with LinkIQ Cable and Network Tester
Pinpoint the leading causes of downtime with LinkIQ Cable and Network TesterPinpoint the leading causes of downtime with LinkIQ Cable and Network Tester
Pinpoint the leading causes of downtime with LinkIQ Cable and Network TesterTranscat
 
Identifying Energy Waste in your Plant/Facility Webinar with Fluke
Identifying Energy Waste in your Plant/Facility Webinar with FlukeIdentifying Energy Waste in your Plant/Facility Webinar with Fluke
Identifying Energy Waste in your Plant/Facility Webinar with FlukeTranscat
 
Principals of Low Resistance Testing Webinar with Megger
Principals of Low Resistance Testing Webinar with MeggerPrincipals of Low Resistance Testing Webinar with Megger
Principals of Low Resistance Testing Webinar with MeggerTranscat
 
Leak Detection & Predictive/Preventative Maintenance Solutions Presented by FLIR
Leak Detection & Predictive/Preventative Maintenance Solutions Presented by FLIRLeak Detection & Predictive/Preventative Maintenance Solutions Presented by FLIR
Leak Detection & Predictive/Preventative Maintenance Solutions Presented by FLIRTranscat
 
Field Pressure Calibration Tips Webinar with Fluke Calibration
Field Pressure Calibration Tips Webinar with Fluke CalibrationField Pressure Calibration Tips Webinar with Fluke Calibration
Field Pressure Calibration Tips Webinar with Fluke CalibrationTranscat
 
Elevated Skin Temperature Screening Solutions – How to Get Back to Work Safel...
Elevated Skin Temperature Screening Solutions – How to Get Back to Work Safel...Elevated Skin Temperature Screening Solutions – How to Get Back to Work Safel...
Elevated Skin Temperature Screening Solutions – How to Get Back to Work Safel...Transcat
 
Elevated Skin Temperature Screening - How to Get Back to Work Safely with FLIR
Elevated Skin Temperature Screening - How to Get Back to Work Safely with FLIRElevated Skin Temperature Screening - How to Get Back to Work Safely with FLIR
Elevated Skin Temperature Screening - How to Get Back to Work Safely with FLIRTranscat
 
Tips for Calibrating Analog Pressure Gauges Webinar Presented by Fluke Calibr...
Tips for Calibrating Analog Pressure Gauges Webinar Presented by Fluke Calibr...Tips for Calibrating Analog Pressure Gauges Webinar Presented by Fluke Calibr...
Tips for Calibrating Analog Pressure Gauges Webinar Presented by Fluke Calibr...Transcat
 
Blind Spots of Low Voltage Testing Webinar Presented by Megger Baker
Blind Spots of Low Voltage Testing Webinar Presented by Megger BakerBlind Spots of Low Voltage Testing Webinar Presented by Megger Baker
Blind Spots of Low Voltage Testing Webinar Presented by Megger BakerTranscat
 
Introduction to Temperature Measurement and Calibration Presented by Fluke Ca...
Introduction to Temperature Measurement and Calibration Presented by Fluke Ca...Introduction to Temperature Measurement and Calibration Presented by Fluke Ca...
Introduction to Temperature Measurement and Calibration Presented by Fluke Ca...Transcat
 
How to Calibrate an IR Thermometer Webinar Presented by Fluke Calibration
How to Calibrate an IR Thermometer Webinar Presented by Fluke CalibrationHow to Calibrate an IR Thermometer Webinar Presented by Fluke Calibration
How to Calibrate an IR Thermometer Webinar Presented by Fluke CalibrationTranscat
 
Solutions for your Thermography Program Presented by FLIR
Solutions for your Thermography Program Presented by FLIRSolutions for your Thermography Program Presented by FLIR
Solutions for your Thermography Program Presented by FLIRTranscat
 
Choosing the Right Calibrator Webinar Presented by Druck BHGE and Transcat
Choosing the Right Calibrator Webinar Presented by Druck BHGE and TranscatChoosing the Right Calibrator Webinar Presented by Druck BHGE and Transcat
Choosing the Right Calibrator Webinar Presented by Druck BHGE and TranscatTranscat
 
Using Acoustic Imaging to Detect Compressed Gas Leaks Presented by Fluke and ...
Using Acoustic Imaging to Detect Compressed Gas Leaks Presented by Fluke and ...Using Acoustic Imaging to Detect Compressed Gas Leaks Presented by Fluke and ...
Using Acoustic Imaging to Detect Compressed Gas Leaks Presented by Fluke and ...Transcat
 
How to Use a Portable Bath to Improve Calibration of Sanitary RTD's and Trans...
How to Use a Portable Bath to Improve Calibration of Sanitary RTD's and Trans...How to Use a Portable Bath to Improve Calibration of Sanitary RTD's and Trans...
How to Use a Portable Bath to Improve Calibration of Sanitary RTD's and Trans...Transcat
 

More from Transcat (20)

Predictive Maintenance Best Practices Presented by Fluke.pptx
Predictive Maintenance Best Practices Presented by Fluke.pptxPredictive Maintenance Best Practices Presented by Fluke.pptx
Predictive Maintenance Best Practices Presented by Fluke.pptx
 
Tools and Techniques for Commissioning and Maintaining PV Systems
Tools and Techniques for Commissioning and Maintaining PV SystemsTools and Techniques for Commissioning and Maintaining PV Systems
Tools and Techniques for Commissioning and Maintaining PV Systems
 
Use of a Druck Pressure Calibrator as a Trouble Shooting Tool Presented by Druck
Use of a Druck Pressure Calibrator as a Trouble Shooting Tool Presented by DruckUse of a Druck Pressure Calibrator as a Trouble Shooting Tool Presented by Druck
Use of a Druck Pressure Calibrator as a Trouble Shooting Tool Presented by Druck
 
Electrical Measurement Safety and Best Practices Webinar Presented by Fluke
Electrical Measurement Safety and Best Practices Webinar Presented by FlukeElectrical Measurement Safety and Best Practices Webinar Presented by Fluke
Electrical Measurement Safety and Best Practices Webinar Presented by Fluke
 
Solar Market Opportunities & Solutions - Presented by Fluke
Solar Market Opportunities & Solutions - Presented by FlukeSolar Market Opportunities & Solutions - Presented by Fluke
Solar Market Opportunities & Solutions - Presented by Fluke
 
Pinpoint the leading causes of downtime with LinkIQ Cable and Network Tester
Pinpoint the leading causes of downtime with LinkIQ Cable and Network TesterPinpoint the leading causes of downtime with LinkIQ Cable and Network Tester
Pinpoint the leading causes of downtime with LinkIQ Cable and Network Tester
 
Identifying Energy Waste in your Plant/Facility Webinar with Fluke
Identifying Energy Waste in your Plant/Facility Webinar with FlukeIdentifying Energy Waste in your Plant/Facility Webinar with Fluke
Identifying Energy Waste in your Plant/Facility Webinar with Fluke
 
Principals of Low Resistance Testing Webinar with Megger
Principals of Low Resistance Testing Webinar with MeggerPrincipals of Low Resistance Testing Webinar with Megger
Principals of Low Resistance Testing Webinar with Megger
 
Leak Detection & Predictive/Preventative Maintenance Solutions Presented by FLIR
Leak Detection & Predictive/Preventative Maintenance Solutions Presented by FLIRLeak Detection & Predictive/Preventative Maintenance Solutions Presented by FLIR
Leak Detection & Predictive/Preventative Maintenance Solutions Presented by FLIR
 
Field Pressure Calibration Tips Webinar with Fluke Calibration
Field Pressure Calibration Tips Webinar with Fluke CalibrationField Pressure Calibration Tips Webinar with Fluke Calibration
Field Pressure Calibration Tips Webinar with Fluke Calibration
 
Elevated Skin Temperature Screening Solutions – How to Get Back to Work Safel...
Elevated Skin Temperature Screening Solutions – How to Get Back to Work Safel...Elevated Skin Temperature Screening Solutions – How to Get Back to Work Safel...
Elevated Skin Temperature Screening Solutions – How to Get Back to Work Safel...
 
Elevated Skin Temperature Screening - How to Get Back to Work Safely with FLIR
Elevated Skin Temperature Screening - How to Get Back to Work Safely with FLIRElevated Skin Temperature Screening - How to Get Back to Work Safely with FLIR
Elevated Skin Temperature Screening - How to Get Back to Work Safely with FLIR
 
Tips for Calibrating Analog Pressure Gauges Webinar Presented by Fluke Calibr...
Tips for Calibrating Analog Pressure Gauges Webinar Presented by Fluke Calibr...Tips for Calibrating Analog Pressure Gauges Webinar Presented by Fluke Calibr...
Tips for Calibrating Analog Pressure Gauges Webinar Presented by Fluke Calibr...
 
Blind Spots of Low Voltage Testing Webinar Presented by Megger Baker
Blind Spots of Low Voltage Testing Webinar Presented by Megger BakerBlind Spots of Low Voltage Testing Webinar Presented by Megger Baker
Blind Spots of Low Voltage Testing Webinar Presented by Megger Baker
 
Introduction to Temperature Measurement and Calibration Presented by Fluke Ca...
Introduction to Temperature Measurement and Calibration Presented by Fluke Ca...Introduction to Temperature Measurement and Calibration Presented by Fluke Ca...
Introduction to Temperature Measurement and Calibration Presented by Fluke Ca...
 
How to Calibrate an IR Thermometer Webinar Presented by Fluke Calibration
How to Calibrate an IR Thermometer Webinar Presented by Fluke CalibrationHow to Calibrate an IR Thermometer Webinar Presented by Fluke Calibration
How to Calibrate an IR Thermometer Webinar Presented by Fluke Calibration
 
Solutions for your Thermography Program Presented by FLIR
Solutions for your Thermography Program Presented by FLIRSolutions for your Thermography Program Presented by FLIR
Solutions for your Thermography Program Presented by FLIR
 
Choosing the Right Calibrator Webinar Presented by Druck BHGE and Transcat
Choosing the Right Calibrator Webinar Presented by Druck BHGE and TranscatChoosing the Right Calibrator Webinar Presented by Druck BHGE and Transcat
Choosing the Right Calibrator Webinar Presented by Druck BHGE and Transcat
 
Using Acoustic Imaging to Detect Compressed Gas Leaks Presented by Fluke and ...
Using Acoustic Imaging to Detect Compressed Gas Leaks Presented by Fluke and ...Using Acoustic Imaging to Detect Compressed Gas Leaks Presented by Fluke and ...
Using Acoustic Imaging to Detect Compressed Gas Leaks Presented by Fluke and ...
 
How to Use a Portable Bath to Improve Calibration of Sanitary RTD's and Trans...
How to Use a Portable Bath to Improve Calibration of Sanitary RTD's and Trans...How to Use a Portable Bath to Improve Calibration of Sanitary RTD's and Trans...
How to Use a Portable Bath to Improve Calibration of Sanitary RTD's and Trans...
 

Recently uploaded

Introduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptxIntroduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptxupamatechverse
 
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...Christo Ananth
 
UNIT-III FMM. DIMENSIONAL ANALYSIS
UNIT-III FMM.        DIMENSIONAL ANALYSISUNIT-III FMM.        DIMENSIONAL ANALYSIS
UNIT-III FMM. DIMENSIONAL ANALYSISrknatarajan
 
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
MANUFACTURING PROCESS-II UNIT-1 THEORY OF METAL CUTTING
MANUFACTURING PROCESS-II UNIT-1 THEORY OF METAL CUTTINGMANUFACTURING PROCESS-II UNIT-1 THEORY OF METAL CUTTING
MANUFACTURING PROCESS-II UNIT-1 THEORY OF METAL CUTTINGSIVASHANKAR N
 
Glass Ceramics: Processing and Properties
Glass Ceramics: Processing and PropertiesGlass Ceramics: Processing and Properties
Glass Ceramics: Processing and PropertiesPrabhanshu Chaturvedi
 
UNIT - IV - Air Compressors and its Performance
UNIT - IV - Air Compressors and its PerformanceUNIT - IV - Air Compressors and its Performance
UNIT - IV - Air Compressors and its Performancesivaprakash250
 
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
Java Programming :Event Handling(Types of Events)
Java Programming :Event Handling(Types of Events)Java Programming :Event Handling(Types of Events)
Java Programming :Event Handling(Types of Events)simmis5
 
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service NashikCall Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service NashikCall Girls in Nagpur High Profile
 
UNIT-V FMM.HYDRAULIC TURBINE - Construction and working
UNIT-V FMM.HYDRAULIC TURBINE - Construction and workingUNIT-V FMM.HYDRAULIC TURBINE - Construction and working
UNIT-V FMM.HYDRAULIC TURBINE - Construction and workingrknatarajan
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Dr.Costas Sachpazis
 
Booking open Available Pune Call Girls Pargaon 6297143586 Call Hot Indian Gi...
Booking open Available Pune Call Girls Pargaon  6297143586 Call Hot Indian Gi...Booking open Available Pune Call Girls Pargaon  6297143586 Call Hot Indian Gi...
Booking open Available Pune Call Girls Pargaon 6297143586 Call Hot Indian Gi...Call Girls in Nagpur High Profile
 
KubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghlyKubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghlysanyuktamishra911
 
Coefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxCoefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxAsutosh Ranjan
 
UNIT-II FMM-Flow Through Circular Conduits
UNIT-II FMM-Flow Through Circular ConduitsUNIT-II FMM-Flow Through Circular Conduits
UNIT-II FMM-Flow Through Circular Conduitsrknatarajan
 
Russian Call Girls in Nagpur Grishma Call 7001035870 Meet With Nagpur Escorts
Russian Call Girls in Nagpur Grishma Call 7001035870 Meet With Nagpur EscortsRussian Call Girls in Nagpur Grishma Call 7001035870 Meet With Nagpur Escorts
Russian Call Girls in Nagpur Grishma Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 

Recently uploaded (20)

Introduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptxIntroduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptx
 
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
 
UNIT-III FMM. DIMENSIONAL ANALYSIS
UNIT-III FMM.        DIMENSIONAL ANALYSISUNIT-III FMM.        DIMENSIONAL ANALYSIS
UNIT-III FMM. DIMENSIONAL ANALYSIS
 
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
 
(INDIRA) Call Girl Aurangabad Call Now 8617697112 Aurangabad Escorts 24x7
(INDIRA) Call Girl Aurangabad Call Now 8617697112 Aurangabad Escorts 24x7(INDIRA) Call Girl Aurangabad Call Now 8617697112 Aurangabad Escorts 24x7
(INDIRA) Call Girl Aurangabad Call Now 8617697112 Aurangabad Escorts 24x7
 
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
MANUFACTURING PROCESS-II UNIT-1 THEORY OF METAL CUTTING
MANUFACTURING PROCESS-II UNIT-1 THEORY OF METAL CUTTINGMANUFACTURING PROCESS-II UNIT-1 THEORY OF METAL CUTTING
MANUFACTURING PROCESS-II UNIT-1 THEORY OF METAL CUTTING
 
Glass Ceramics: Processing and Properties
Glass Ceramics: Processing and PropertiesGlass Ceramics: Processing and Properties
Glass Ceramics: Processing and Properties
 
UNIT - IV - Air Compressors and its Performance
UNIT - IV - Air Compressors and its PerformanceUNIT - IV - Air Compressors and its Performance
UNIT - IV - Air Compressors and its Performance
 
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
 
Java Programming :Event Handling(Types of Events)
Java Programming :Event Handling(Types of Events)Java Programming :Event Handling(Types of Events)
Java Programming :Event Handling(Types of Events)
 
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service NashikCall Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
 
UNIT-V FMM.HYDRAULIC TURBINE - Construction and working
UNIT-V FMM.HYDRAULIC TURBINE - Construction and workingUNIT-V FMM.HYDRAULIC TURBINE - Construction and working
UNIT-V FMM.HYDRAULIC TURBINE - Construction and working
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
 
Roadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and RoutesRoadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and Routes
 
Booking open Available Pune Call Girls Pargaon 6297143586 Call Hot Indian Gi...
Booking open Available Pune Call Girls Pargaon  6297143586 Call Hot Indian Gi...Booking open Available Pune Call Girls Pargaon  6297143586 Call Hot Indian Gi...
Booking open Available Pune Call Girls Pargaon 6297143586 Call Hot Indian Gi...
 
KubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghlyKubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghly
 
Coefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxCoefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptx
 
UNIT-II FMM-Flow Through Circular Conduits
UNIT-II FMM-Flow Through Circular ConduitsUNIT-II FMM-Flow Through Circular Conduits
UNIT-II FMM-Flow Through Circular Conduits
 
Russian Call Girls in Nagpur Grishma Call 7001035870 Meet With Nagpur Escorts
Russian Call Girls in Nagpur Grishma Call 7001035870 Meet With Nagpur EscortsRussian Call Girls in Nagpur Grishma Call 7001035870 Meet With Nagpur Escorts
Russian Call Girls in Nagpur Grishma Call 7001035870 Meet With Nagpur Escorts
 

Transcat and Fluke Webinar: Streamlining Instrument Calibration

  • 1. Streamlining Instrument Calibration with Calibration Management Software and Documenting Field Calibrators Documenting tools and calibration management software combine for a paperless calibration solution 1 Host: Transcat Presenters: Jim Shields and Bob Crepps Fluke Process Calibration Group
  • 2. Agenda • Instrument calibration, challenges and elements – Reasons to calibrate • Establishing a process calibration maintenance management program – Using Calibration Management Software (CMS) to: • Manage process instruments and calibration records • Interface with Field Documenting Process Calibrators • Predict instrument performance, minimize maintenance costs – Using Documenting Process Calibrators to : • Manage procedures or tasks downloaded from CMS • Perform field calibrations and adjustments • Record field calibration results • Upload field calibration results to CMS • Paperless calibration management program benefits
  • 3. Why Calibrate? • Regulations, standards and trade • Custody Transfer • Internal/External Quality Programs (like ISO-9000) • Plant Safety and Uptime • EPA, FDA, NRC, OSHA and other mandated regulations
  • 4. Why Calibrate? • All instruments change with – Time, temperature, humidity, environmental exposure, vibration, wear and tear – No two instruments change exactly the same • Scheduled calibrations reduce process downtime risk – Improves the performance and reliability of critical assets • Product Quality / Yield – Calibration helps processes comparable to traceable calibration standards • Helps minimize process deviations – Helps drive process optimization and cost savings
  • 5. Industry changes • Regulatory forces increasing • The plant isn’t getting smaller but shrinking budgets are reducing the maintenance staff – Driving efficiencies and the need to “Do more with less” • Experienced instrumentation professionals retiring – Technical and Vocational Schools working to fill voids • Plant modernizations lead to control system changes • Competitive pressure offshore push the need for plant optimization, streamlined operations and maintenance practices
  • 6. Process Calibration Elements: • Calibration equipment, procedures, instrument documentation, calibration management software (CMS), trained personnel Benefits: • Consistent process performance and efficiency, improved product quality, reduced downtime, safety, regulatory compliance, maximized use of skilled maintenance personnel
  • 7. Calibration, definitions… Calibration: Comparison of a measured value to a traceable measurement standard – Adjust to agree with the standard if necessary • Accuracy: The closeness of agreement between an observed value and that of a traceable measurement standard • Adjustment: Manual or digital compensation of the output signal of device with a known input value applied – Necessary if a device is not performing as required • Need to understand adjustment tolerance • Need communication tool for digital adjustment Adjustment is part of a calibration process, it is not calibration by itself.
  • 8. Fluke Petrochemical Corporation By Tag ID RESULTS FROM FLUKE 702 CALIBRATIONS 2-Feb-96 Fluke Petrochemical Corp. By: Title: Cert. No.: 1 Model: FLUKE 702 Serial:6150203 User ID: JSmith 010 Cal. Status: 14 Loop Power: 24V As found 25-Jan-96 4:46 Measure: 4.000 to 20.000 mA Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90 Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321 Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234 1,996.001.0025 Input Value: Units Output value: Units Error % Status MeasUEV 3.906 mA 40.0 °C -0.59 Failed Disabled 11.918 mA 55.0 °C -0.51 Failed Disabled 19.923 mA 70.0 °C -0.48 Failed Disabled 11.918 mA 55.0 °C -0.51 Failed Disabled 3.907 mA 40.0 °C -0.58 Failed Disabled Model: FLUKE 702 Serial:6150203 User ID: JSmith 010 Cal. Status: 14 Loop Power: 24V As left 25-Jan-96 4:53 Measure: 4.000 to 20.000 mA Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90 Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321 Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234 1,996.001.0025 Input Value: Units Output value: Units Error % Status MeasUEV 3.998 mA 40.0 °C -0.01 Passed Disabled 12.011 mA 55.0 °C 0.07 Passed Disabled 20.018 mA 70.0 °C 0.12 Passed Disabled 12.016 mA 55.0 °C 0.10 Passed Disabled 4.003 mA 40.0 °C 0.02 Passed Disabled Fluke Petrochemical Corporation By Tag ID RESULTS FROM FLUKE 702 CALIBRATIONS 2-Feb-96 Fluke Petrochemical Corp. By: Title: Cert. No.: 1 Model: FLUKE 702 Serial:6150203 User ID: JSmith 010 Cal. Status: 14 Loop Power: 24V As found 25-Jan-96 4:46 Measure: 4.000 to 20.000 mA Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90 Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321 Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234 1,996.001.0025 Input Value: Units Output value: Units Error % Status MeasUEV 3.906 mA 40.0 °C -0.59 Failed Disabled 11.918 mA 55.0 °C -0.51 Failed Disabled 19.923 mA 70.0 °C -0.48 Failed Disabled 11.918 mA 55.0 °C -0.51 Failed Disabled 3.907 mA 40.0 °C -0.58 Failed Disabled Model: FLUKE 702 Serial:6150203 User ID: JSmith 010 Cal. Status: 14 Loop Power: 24V As left 25-Jan-96 4:53 Measure: 4.000 to 20.000 mA Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90 Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321 Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234 1,996.001.0025 Input Value: Units Output value: Units Error % Status MeasUEV 3.998 mA 40.0 °C -0.01 Passed Disabled 12.011 mA 55.0 °C 0.07 Passed Disabled 20.018 mA 70.0 °C 0.12 Passed Disabled 12.016 mA 55.0 °C 0.10 Passed Disabled 4.003 mA 40.0 °C 0.02 Passed Disabled Fluke Petrochemical Corporation By Tag ID RESULTS FROM FLUKE 702 CALIBRATIONS 2-Feb-96 Fluke Petrochemical Corp. By: Title: Cert. No.: 1 Model: FLUKE 702 Serial:6150203 User ID: JSmith 010 Cal. Status: 14 Loop Power: 24V As found 25-Jan-96 4:46 Measure: 4.000 to 20.000 mA Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90 Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321 Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234 1,996.001.0025 Input Value: Units Output value: Units Error % Status MeasUEV 3.906 mA 40.0 °C -0.59 Failed Disabled 11.918 mA 55.0 °C -0.51 Failed Disabled 19.923 mA 70.0 °C -0.48 Failed Disabled 11.918 mA 55.0 °C -0.51 Failed Disabled 3.907 mA 40.0 °C -0.58 Failed Disabled Model: FLUKE 702 Serial:6150203 User ID: JSmith 010 Cal. Status: 14 Loop Power: 24V As left 25-Jan-96 4:53 Measure: 4.000 to 20.000 mA Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90 Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321 Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234 1,996.001.0025 Input Value: Units Output value: Units Error % Status MeasUEV 3.998 mA 40.0 °C -0.01 Passed Disabled 12.011 mA 55.0 °C 0.07 Passed Disabled 20.018 mA 70.0 °C 0.12 Passed Disabled 12.016 mA 55.0 °C 0.10 Passed Disabled 4.003 mA 40.0 °C 0.02 Passed Disabled Fluke Petrochemical Corporation By Tag ID RESULTS FROM FLUKE 702 CALIBRATIONS 2-Feb-96 Fluke Petrochemical Corp. By: Title: Cert. No.: 1 Model: FLUKE 702 Serial:6150203 User ID: JSmith 010 Cal. Status: 14 Loop Power: 24V As found 25-Jan-96 4:46 Measure: 4.000 to 20.000 mA Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90 Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321 Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234 1,996.001.0025Input Value: Units Output value: Units Error % Status MeasUEV 3.906 mA 40.0 °C -0.59 Failed Disabled 11.918 mA 55.0 °C -0.51 Failed Disabled 19.923 mA 70.0 °C -0.48 Failed Disabled 11.918 mA 55.0 °C -0.51 Failed Disabled 3.907 mA 40.0 °C -0.58 Failed Disabled Model: FLUKE 702 Serial:6150203 User ID: JSmith 010 Cal. Status: 14 Loop Power: 24V As left 25-Jan-96 4:53 Measure: 4.000 to 20.000 mA Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90 Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321 Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234 1,996.001.0025Input Value: Units Output value: Units Error % Status MeasUEV 3.998 mA 40.0 °C -0.01 Passed Disabled 12.011 mA 55.0 °C 0.07 Passed Disabled 20.018 mA 70.0 °C 0.12 Passed Disabled 12.016 mA 55.0 °C 0.10 Passed Disabled 4.003 mA 40.0 °C 0.02 Passed Disabled Fluke Petrochemical Corporation By Tag ID RESULTS FROM FLUKE 702 CALIBRATIONS 2-Feb-96 Fluke Petrochemical Corp. By: Title: Cert. No.: 1 Model: FLUKE 702 Serial:6150203 User ID: JSmith 010 Cal. Status: 14 Loop Power: 24V As found 25-Jan-96 4:46 Measure: 4.000 to 20.000 mA Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90 Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321 Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234 1,996.001.0025Input Value: Units Output value: Units Error % Status MeasUEV 3.906 mA 40.0 °C -0.59 Failed Disabled 11.918 mA 55.0 °C -0.51 Failed Disabled 19.923 mA 70.0 °C -0.48 Failed Disabled 11.918 mA 55.0 °C -0.51 Failed Disabled 3.907 mA 40.0 °C -0.58 Failed Disabled Model: FLUKE 702 Serial:6150203 User ID: JSmith 010 Cal. Status: 14 Loop Power: 24V As left 25-Jan-96 4:53 Measure: 4.000 to 20.000 mA Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90 Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321 Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234 1,996.001.0025Input Value: Units Output value: Units Error % Status MeasUEV 3.998 mA 40.0 °C -0.01 Passed Disabled 12.011 mA 55.0 °C 0.07 Passed Disabled 20.018 mA 70.0 °C 0.12 Passed Disabled 12.016 mA 55.0 °C 0.10 Passed Disabled 4.003 mA 40.0 °C 0.02 Passed Disabled Fluke Petrochemical Corporation By Tag ID RESULTS FROM FLUKE 702 CALIBRATIONS 2-Feb-96 Fluke Petrochemical Corp. By: Title: Cert. No.: 1 Model: FLUKE 702 Serial:6150203 User ID: JSmith 010 Cal. Status: 14 Loop Power: 24V As found 25-Jan-96 4:46 Measure: 4.000 to 20.000 mA Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90 Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321 Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234 1,996.001.0025Input Value: Units Output value: Units Error % Status MeasUEV 3.906 mA 40.0 °C -0.59 Failed Disabled 11.918 mA 55.0 °C -0.51 Failed Disabled 19.923 mA 70.0 °C -0.48 Failed Disabled 11.918 mA 55.0 °C -0.51 Failed Disabled 3.907 mA 40.0 °C -0.58 Failed Disabled Model: FLUKE 702 Serial:6150203 User ID: JSmith 010 Cal. Status: 14 Loop Power: 24V As left 25-Jan-96 4:53 Measure: 4.000 to 20.000 mA Source: 40.0 to 70.0 °C Type: Pt100(3916) 2W RTD ITS-90 Test Tolerance: 0.25% Tag ID:TT451 Serial: 987654321 Test Strategy: 3 ^v Task Name: Temp Transmitter RTD1234 1,996.001.0025Input Value: Units Output value: Units Error % Status MeasUEV 3.998 mA 40.0 °C -0.01 Passed Disabled 12.011 mA 55.0 °C 0.07 Passed Disabled 20.018 mA 70.0 °C 0.12 Passed Disabled 12.016 mA 55.0 °C 0.10 Passed Disabled 4.003 mA 40.0 °C 0.02 Passed Disabled National Standards Traceability • Unbroken chain of calibration. • Documented proof.
  • 9. Process Calibration requires: Traceable calibration tools, source functions: – Temperature, electronic simulation of thermocouples or RTDs • Does not test the sensor installed in a process – Temperature, dry block or temperature bath calibrator • Tests the temperature sensor and transmitter electronics – Pressure, hand pump, or other means to apply pressure to the unit under test – Pressure measurement reference to measure the pressure generated (by a pressure source) and applied to the UUT • Transfer standard approach – Other Electronic signals • Frequency for flow, speed and rotation • Current for I/P and analog I/O • Voltage for signal conditioners and analog I/O
  • 10. Process Calibration requires: Traceable calibration tools, measurement functions: – Temperature, thermocouple or RTD thermometer • Measure the process temperature with a reference probe directly and compare with control system – Pressure, measure the pressure output from an I/P or other pneumatic control signals – Other Electronic signals • Frequency for flow, speed and rotation • Current for all transmitter mA loop signal outputs • Voltage for signal conditioners other voltage output devices – Many transmitters have voltage output instead of mA output.
  • 11. Why document calibration, how? Calibration Records may be: – required for regulations – required for quality standards – required by your customers Calibration records can: – assist in understanding the performance of a process – be used to determine a devices suitability for a process – help assess if a device needs repair – be used to help adjust maintenance intervals • Calibration records can: – be manually recorded on a paper form – be recorded on a palm device or portable computer – be recorded using a calibrator with documenting capabilities Calibration management software + documenting calibrators enables easy documenting and analysis of field calibrations
  • 12. Calibration process, setup • Identify critical instruments – Safety instrumented systems – Environmental critical (water and air outflow) – Quality critical: directly effect product quality and value – Revenue critical instruments: • Custody transfer, trade related measurements • Define the interval to maintain desired accuracy – Instrument should perform in tolerance when tested • Specifications and test tolerances need careful consideration – Manufacturers claimed accuracies can be misleading – Test tolerances should be specified by process engineer – The accuracy of both the instrument being tested and the calibrator need to be considered • Define the procedures for the identified tags
  • 13. Calibration process; execution • Identify the tags that need calibration • Organize the work into logical routes • Collect the necessary procedures, documentation and tools • Isolate the device from the process – Lock-out, Tag-out procedure if warranted • Connect the calibrator, begin to calibrate – Perform the As-Found test, document the results – Adjust the transmitter if the measured errors are greater than the adjustment threshold defined by the procedure. – Perform the As-Left test, document the results – Complete the documented record with Tag#, time, date, Serial numbers of the TAG, the test equipment used and its’ calibration data
  • 14. Documented Calibration Requires – As Found Data • Testing the instrument to determine “As-Found” performance • Accurate knowledge of the source value is required – The sourced value and measured values need to be traceable – The total uncertainty of the source should be 3 to 4 times better than the device under test • Pass-fail assessment (fail) • Adjustment required! Graph of Percent Error -0.3 -0.2 -0.1 0 0.1 0.2 0.3 0.4 0 10 20 30 40 50 60 70 30 90 100 Input % of Span %Error +Tolerance -Tolerance
  • 15. Graph of Percent Error -0.4 -0.3 -0.2 -0.1 0 0.1 0.2 0.3 0.4 0 10 20 30 40 50 60 70 30 90 100 Input % of Span %Error +Tolerance -Tolerance Documented Calibration Requires – As Found Data • Re-test to verify adjustments. • “As Left” test documents the condition of the transmitter after adjustment • Time and date of the test need to be documented • Instrument tag, serial number and technician data need to be documented
  • 17. Why use a documenting calibrator? • Multifunction source and measure • Automatically capture results, date/time user and tag information • Eliminates manual transcription and legibility errors • Eliminates “pencil whipped” results • Ensures calibrations performed per documented procedures • Coupled with CMS complete a paperless documenting field calibration system • Assists to comply with regulations and audits • Many have HART communication capability built in
  • 18. Why use a documenting calibrator? • Multifunction source and measure • Automatically capture results, date/time user and tag information • Eliminates manual transcription and legibility errors • Eliminates “pencil whipped” results • Ensures calibrations performed per documented procedures • Coupled with CMS complete a paperless documenting field calibration system • Assists to comply with regulations and audits • Many have HART communication capability built in
  • 19. Why use a documenting calibrator? • Multifunction source and measure • Automatically capture results, date/time user and tag information • Eliminates manual transcription and legibility errors • Eliminates “pencil whipped” results • Ensures calibrations performed per documented procedures • Coupled with CMS complete a paperless documenting field calibration system • Assists to comply with regulations and audits • Many have HART communication capability built in
  • 20. • Manage calibration records for field process instrumentation • Print reports or run queries for instruments due for calibration to help assess, schedule and manage • Connect to Documenting Process Calibrators (DPCs) – Load instrument procedures and upload documented calibrations • Documented results enable analysis for drift, reliability or maintenance costs • Many CMS applications have connectivity to CMMS packages like SAP, Oracle and Maximo Why use Calibration Management Software (CMS)?
  • 21. Examples: • Fluke DPCTrack2 • Prime Technologies ProCalV5 • Emerson AMS • Yokogawa PRM • Meridium • Intergraph SmartPlant Why use Calibration Management Software (CMS)? • CMS completes the documenting solution when coupled with a documenting calibrator • Many different solutions from fully integrated CMMS to stand alone CMS • Vary in complexity and price, $4K to $100K
  • 22. CMS Software Example, DPCTrack2 Modern calibration management software packages are easy to use and deliver good reasons to adopt. • Example entry level single user CMS application. • Upgradeable to ProCalV5 Enterprise CMS software
  • 23. What should a CMS package do? • Manage and document process instrumentation performance • Create custom procedures for download to DPCs • Upload results from DPCs and store for later retrieval or analysis • Print a variety of reports that support ISO 9000 or other documentation needs • Search for particular tags and report their history • Search for tags due for calibration in the past or future • Keep track of all Tags “touched” by a particular piece of test equipment (reverse traceability)
  • 24. Instrument View Search Queries • Search queries can be customized to find populations of tags differentiated by: >Cal Due Date >Tag Id >S/N >Location >Device type >Location >Other fields
  • 25. Tag Instrument Data • Input the data describing the tag
  • 26. Transmitter Data • Describe the input, output and expected tolerance of the transmitter
  • 27. Upload and download information to DPCs • Tags identified in the date based search are loaded into the calibrator
  • 28. Live DPCTrack2 demo • Creating tags using example templates • Error% calculation explanation • Sorting the database for calibrations using database query tools • Loading tags for a calibration route into the Fluke-729
  • 29. Task Selection, (DPC) • Tags downloaded to the DPC available for selection • Supervisor can assign as a route to a technician • Usage count indicates whether a task has been completed using the downloaded task • Tasks must be performed for usage count to increment Tag Number Task Usage & completion Count
  • 30. Downloaded transmitter test parameters Predefined procedures automatically fill in test parameters: • Source and measure end points, • Test tolerance, • Test strategy (number of tests) • Leak test available on 729 to verify test setup for leakage
  • 31. Pressure documenting made easy • On the test screen: – Applied pressure (top) – Measured mA value (4.000 mA would be the ideal value for this device) – Error % of Range/Span • Green for in tolerance • Red for out of tolerance – AutoTest and the 729 will automatically apply the test pressures, measure mA signals calculate Error % and document • Out of tolerance tests flagged in Red • Tag number, serial number and technician name recorded.
  • 32. Transmitter Adjustment, HART PV zero trim – Zero out transmitter offset Output Trim – Fetch retrieves the current value being measured – Trim 4 mA and Trim 20 mA performs mA adjustment Trim to applied Values – Apply Expected 0 and 100% range values – Adjust 4 and 20 mA outputs to match applied pressures.
  • 33. Transmitter calibration As-Left measurement The As-Left test will repeat the initial tests to verify and document the “As Left”, post adjustment condition of the transmitter • Press “As Left” • This As-Left result is successful as noted by Pass • The documented test is now ready for upload If the test fails As Left, additional adjustment and additional As-Left test(s) may be necessary.
  • 34. Live DPCTrack2 demo • Uploading documented calibration results • Review and authorize the uploaded results • Viewing the calibration certificate • View other reporting capabilities
  • 35. Uploaded Test Results As-Found, As-Left results uploaded and authorized to add to the database for analysis and printing
  • 36. Calibration record • Documented calibration stored for recall and printing as needed. • Calibration date updated in database and next calibration due date is set.
  • 37. Takeaways • Efficiency of route based calibrations can save up to 50% on calibration maintenance labor making technicians more efficient • Multifunction documenting calibrators minimize returning to the shop to get additional tools • The efficiency of performing pressure tests with the 729 Automatic Pressure Calibrator both saves time making the technician efficient but also improves the quality of the calibration result. • Documented calibration results can assist in passing audits or other requirements where calibration records are required
  • 38. Summary • Calibration is a critical practice to remove variability and maintain safety in a process • Documentation of calibration history enables advanced analysis techniques • Field documentation with a documenting calibrator eliminates transcription errors and enables data to be uploaded to Calibration Management Software (CMS) • CMS can be used to sort the calibrations due into routes for download, maximizes use of skilled personnel • CMS assists in managing and interpreting documented calibration records and day to day workload assignments
  • 40. Fluke Corporation Fluke Headquarters, Everett Washington, USA From John Fluke’s basement to an industry leader
  • 41. Fluke Corporation • 3000 employees worldwide • Presence in more than 100 countries • Products include: • Process Calibration Tools • DMMs • Calibration Standards • High Performance Oscilloscopes • Power Quality Analyzers • Electrical Testers • Thermal imaging • Condition monitoring • More!
  • 42. Current Offer • Now through June 30th – Free Fluke tools with purchases $250+ – www.transcat.com 42
  • 43. Questions or Comments? Email Nicole VanWert-Quinzi nicole.vanwert@Transcat.com Transcat: 800-800-5001 www.Transcat.com For related product information, go to: www.Transcat.com/Fluke