The Analytical Method Transfer Process
PDAAMD-AMV Workshop
Baltimore, MD
7-8 October 2013
Stephan O. Krause, Ph.D.
Principal Scientist, Regulatory Science, Development
MedImmune
2
The Analytical Method Transfer Process
Agenda:
 Introduction to Analytical Method Transfer (AMT)
 Mapping out the Overall Product Development and Tech Transfer
Process
 Method Types and Acceptance Criteria
 Analytical Platform Methods vs. New Methods
 AMT Example
Krause/PDA, 2012
3
The Analytical Method Life Cycle
Krause/PDA, 2012
CQA Development, CMC Changes, Specifications
4
5
The Five General AMV/AMT Classes
AMV Class Description
Typical Risk /
Uncertainty Level
(1=Low, 5=High)
Suggested Prospective
AMV Studies
AMV
Class
No.
Analytical Method
Product /
Process Sample
A New New 4-5 Full Validation
B New Old (Validated) 3-4(1) Full Validation Plus
AMR(2)
Studies
C
Analytical Platform
Technology (not
validated “as run”)
New 2-3 Partial Validation
D Old (Validated) New 1-2
Partial Validation or
Verification
E Compendial New 1-2
Verification per USP
<1226>
(1) If a new analytical method (forced method replacement) is needed due to supply reasons, the risk level can be generally considered higher
because no other option may exist. Unforced test method replacements can be considered to be a lower risk level as more time may be available
to optimize the method performance.
(2) AMR = Analytical Method Replacement. A study to confirm that a new analytical method can perform equally or better than the existing one.
From Krause, PDA/DHI 2007.
6
Risk-Based AMT Protocol Acceptance Criteria
Krause/PDA, 2012
7
General AMT Strategy
Krause/PDA, 2012
 Co-validation/Co-qualification – this may be used early in the life cycle of a test
method when appropriate.
 Comparative study – AMT study performed concurrently by sending and
receiving laboratories. Acceptance criteria determine the equivalence of the two
laboratories. Historical and validation data may be used when appropriate for
parts of the method transfer study. The sending laboratory typically has collected
a significant amount of historical data for test method performance results in
addition to test results for the samples to be tested at the receiving laboratory.
 Performance Verification - The receiving laboratory may already perform the
method for a similar product or for another type of sample for the same product.
In this case, a formal method transfer may not be required. Any reduced
prospective study considered should be properly justified.
8
Retrospective and Prospective Use of Data for AMV Studies from
other Processes Prior to AMV – New Method
9
Retrospective and Prospective Use of Data for AMV Studies from
other Processes Prior to AMV – Platform Method
10
“Fixed” vs. “Variable” AMT Execution Plan
 A fixed execution plan does not integrate test method result variation
and the minimum number of test replicates needed to obtain a
desirable statistical confidence in the transfer results. A fixed
execution matrix can be more advantageous when transferring
multiple products to/from multiple locations.
 A variable execution plan does consider test method result variation
and may require a larger data sets, especially for test methods with
relatively high result variation. For example, a variable execution
plan may be advantageous when transferring bioassays with an
expected high degree of test result variation.
Krause/PDA, 2012
11
Typical “Fixed” AMT Execution Plan
for Late-Stage/Commercial Products
Laboratory Day Analyst Instrument Replicates
Sending 1 1 1 3
Sending 1 2 2 3
Sending 2 1 1 3
Sending 2 2 2 3
Sending 3 1 1 3
Sending 3 2 2 3
Receiving 1 1 1 3
Receiving 1 2 2 3
Receiving 2 1 1 3
Receiving 2 2 2 3
Receiving 3 1 1 3
Receiving 3 2 2 3
From Krause, PDA/DHI 2007.
12
Analytical Method Transfer (AMT)
Example
 A validated analytical method for potency is to be transferred from the
original QC laboratory to another QC laboratory to release drug product
(DP). The analytical method generates potency (dose) results for
lyophilized DP.
 The vials are available in three nominal doses between 500 – 2000 IU/vial
using an identical formulation. Release testing is performed using three
replicate preparations from each of three vials.
 Before analysis the content of a vial is reconstituted with 5.0 mL of WFI
water and the potency is measured in IU/mL (100 – 400 IU/mL).
 The samples and a product-specific reference standard are prepared
similarly. The analytical method procedure and statistical evaluation are
performed with the parallel-line concept.
 The “variable” AMT execution plan is used.
13
AMT Study Design and Acceptance Criteria
Characteristics
Evaluated
Accuracy/Matching:
The relative difference between lab means should at 90% confidence not be less than -Θ= 10% and not
more than +Θ = 10%. The 10% difference limit was set with consideration of product specification.
Intermediate Precision:
RSD 6 % for all sample types, with appropriate homoscedasticity throughout the potency range (from
validation results).
This means that any RSD from a sample of n=8 should not exceed 9.43 %
Number of
Replicates
Nreplicates
= at least 23 independent replicates
The confidence interval for the lab-to-lab difference for N determinations to less than the [10%, +10%]. As
above the 10% difference limit was set with consideration of product specification.
Samples to test Nlevel
= 3
The range of potency/dosing results is covered by:
Lowest dose 500 IU/vial or 100 IU/mL
Medium dose 1000 IU/vial or 200 IU/mL
Highest dose 2000 IU/vial or 400 IU/mL
Testing design,
each sample
Number of operators, n = 2
Number of days, n = 2
Number of replicates per day per operator, n = 2
N = 8 in each lab for each of n= 3 potency levels. Results are converted to “% recoveries vs. expected” to
allow pooling
Total NTotal
= 24 individual observations will be recorded for each laboratory. N=24 individual
observations are needed as N=23 is the minimum number of replicates calculated.
14
AMT Results from Sending and Receiving Labs
Theoretical
Potency Level
in IU/mL
Operator Day Replicate
Sending lab Receiving lab
Experimental
Potency in IU/mL
%Recovery vs.
Theoretical
Potency
Experimental
Potency in IU/mL
%Recovery vs.
Theoretical
Potency
100 1 1 1 103 103.0 95 95.0
100 1 1 2 104 104.0 99 99.0
100 1 2 1 108 108.0 104 104.0
100 1 2 2 101 101.0 103 103.0
100 2 1 1 94 94.0 93 93.0
100 2 1 2 99 99.0 96 96.0
100 2 2 1 102 102.0 92 92.0
100 2 2 2 104 104.0 100 100.0
200 1 1 1 212 106.0 208 104.0
200 1 1 2 208 104.0 192 96.0
200 1 2 1 191 95.5 199 99.5
200 1 2 2 201 100.5 195 97.5
200 2 1 1 204 102.0 208 104.0
200 2 1 2 206 103.0 211 105.5
200 2 2 1 198 99.0 203 101.5
200 2 2 2 200 100.0 183 91.5
400 1 1 1 375 93.8 383 95.8
400 1 1 2 401 100.3 401 100.3
400 1 2 1 408 102.0 389 97.3
400 1 2 2 388 97.0 391 97.8
400 2 1 1 402 100.5 408 102.0
400 2 1 2 415 103.8 421 105.3
400 2 2 1 406 101.5 415 103.8
400 2 2 2 410 102.5 403 100.8
15
AMT Result Summary from Sending and
Receiving Labs
(1) Raw data was used unrounded. Upper and lower 90% CIs were calculated using equation below.
(2) N=24 data points are not independent.
Separate and Pooled Potency Levels
Evaluated
Sending lab Receiving lab
Statistical
Parameters
%Recovery vs.
Theoretical
Potency
Statistical
Parameters
%Recovery vs.
Theoretical
Potency
TOST with acceptance criteria
[-10%, +10%] (1)
N1 24(2)
N2 24
Mean1 101.1 Mean2 99.3
SD 3.5(2)
SD 4.2
RSD 3.4 RSD 4.2
Pooled SD(2)
3.9
Mean1-Mean2 1.8
t-value 1.679
Upper 90%
CI limit(1)
4 (3.6)
Lower 90%
CI limit(1)
0 (-0.1)
Transfer Acceptance Conclusion Pass
Θ+<





+××±−<Θ− −+−
21
2,2121
11
)( 21
nn
stxx pnnα
16
Graphical Representation of Potency Results Per
Potency Level Between Laboratories
 The boxes represent the 25th – 75th percentile distribution of the results for the two laboratories. Medians
(line in the box) and means (cross in the box) are approximately centered while the medians are equidistant
from the box hinges, providing a visual indication for a normal data distribution(s) among data points
within each laboratory set.
 One potential outlier (lower open circle outside of the whiskers) is observed in the sending lab, however,
this does not change the overall interpretation for the demonstration of lab-to-lab equivalence.
 The variation in the test results (wider 25th – 75th percentile boxes) appears to be higher in the receiving
laboratory which may be attributed to less test method execution experience.
80
90
100
110
120
Sending lab Receiving lab
%Recoveryversustheoreticalpotency(in%)
17
Graphical Representation of the Combined Percent Recoveries
Between Laboratories for all Concentration Levels
18
Summary
 AMTs may occur at different product development stages.
 AMT study execution matrix and the acceptance criteria should be
developed based on risk(s).
 Many thanks to:
• Rashmi Rawat (Product Quality Reviewer, CDER)
• Pat Cash (Sr. Director, Analytical Biochemistry, MedImmune)
• Mark Schenerman (VP, Analytical Biochemistry, MedImmune)
• Martin Van Trieste (SVP, Quality, Amgen)
• Rich Levy (SVP, PDA)
• Pierre Douette, Ph.D., Eurogentec S.A., Belgium
• Michael Warncke, Ph.D., Bayer HealthCare, USA
• Earl K. Zablackis, Ph.D., Sanofi Pasteur, USA
Krause/PDA, 2012

More Related Content

What's hot (20)

Analytical Method Validation
Analytical Method ValidationAnalytical Method Validation
Analytical Method Validation
 
Q6 guidelines
Q6 guidelinesQ6 guidelines
Q6 guidelines
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
 
BACPAC
BACPACBACPAC
BACPAC
 
CGMP guidelines
CGMP guidelinesCGMP guidelines
CGMP guidelines
 
Case study on Out of Specification (OOS).
Case study on Out of Specification (OOS).Case study on Out of Specification (OOS).
Case study on Out of Specification (OOS).
 
Handling of Out of Specification Results
Handling of Out of Specification ResultsHandling of Out of Specification Results
Handling of Out of Specification Results
 
Validation master plan
Validation master planValidation master plan
Validation master plan
 
Six system inspection model
Six system inspection modelSix system inspection model
Six system inspection model
 
Handling of OOS Dr.A. Amsavel
Handling of OOS  Dr.A. AmsavelHandling of OOS  Dr.A. Amsavel
Handling of OOS Dr.A. Amsavel
 
Validation master plan
Validation master planValidation master plan
Validation master plan
 
CMA-CPP-CQA for oral solid dosageform
CMA-CPP-CQA for oral solid dosageformCMA-CPP-CQA for oral solid dosageform
CMA-CPP-CQA for oral solid dosageform
 
Pharmaceutical process validation (PV)
Pharmaceutical process validation (PV)  Pharmaceutical process validation (PV)
Pharmaceutical process validation (PV)
 
Pharmaceutical validation
Pharmaceutical validationPharmaceutical validation
Pharmaceutical validation
 
Snda
SndaSnda
Snda
 
Qualification of laboratory equipments
Qualification of laboratory equipmentsQualification of laboratory equipments
Qualification of laboratory equipments
 
Analytical QbD
Analytical QbDAnalytical QbD
Analytical QbD
 
Batch Review And Batch Release.pptx
Batch Review And Batch Release.pptxBatch Review And Batch Release.pptx
Batch Review And Batch Release.pptx
 
Computer system validations
Computer system validationsComputer system validations
Computer system validations
 
Validation master plan
Validation master planValidation master plan
Validation master plan
 

Viewers also liked

FDA (invited) Presentation - Specifications and Analytical Method Lifecycle f...
FDA (invited) Presentation - Specifications and Analytical Method Lifecycle f...FDA (invited) Presentation - Specifications and Analytical Method Lifecycle f...
FDA (invited) Presentation - Specifications and Analytical Method Lifecycle f...Stephan O. Krause, PhD
 
Bioprocess Summit - 03Aug15 - final SK27Jul15
Bioprocess Summit - 03Aug15  - final SK27Jul15Bioprocess Summit - 03Aug15  - final SK27Jul15
Bioprocess Summit - 03Aug15 - final SK27Jul15Stephan O. Krause, PhD
 
CPV Acceptance Criteria and Conditions SK09Aug16
CPV Acceptance Criteria and Conditions SK09Aug16CPV Acceptance Criteria and Conditions SK09Aug16
CPV Acceptance Criteria and Conditions SK09Aug16Stephan O. Krause, PhD
 
Analytical Control Strategy for Biologics SK17Jan13
Analytical Control Strategy for Biologics SK17Jan13Analytical Control Strategy for Biologics SK17Jan13
Analytical Control Strategy for Biologics SK17Jan13Stephan O. Krause, PhD
 
Analytical Method &amp; Technology Transfer Ispe Guide
Analytical Method &amp; Technology Transfer Ispe GuideAnalytical Method &amp; Technology Transfer Ispe Guide
Analytical Method &amp; Technology Transfer Ispe GuideCrown Cork & Seal
 

Viewers also liked (6)

FDA (invited) Presentation - Specifications and Analytical Method Lifecycle f...
FDA (invited) Presentation - Specifications and Analytical Method Lifecycle f...FDA (invited) Presentation - Specifications and Analytical Method Lifecycle f...
FDA (invited) Presentation - Specifications and Analytical Method Lifecycle f...
 
Bioprocess Summit - 03Aug15 - final SK27Jul15
Bioprocess Summit - 03Aug15  - final SK27Jul15Bioprocess Summit - 03Aug15  - final SK27Jul15
Bioprocess Summit - 03Aug15 - final SK27Jul15
 
CPV Acceptance Criteria and Conditions SK09Aug16
CPV Acceptance Criteria and Conditions SK09Aug16CPV Acceptance Criteria and Conditions SK09Aug16
CPV Acceptance Criteria and Conditions SK09Aug16
 
Analytical Control Strategy for Biologics SK17Jan13
Analytical Control Strategy for Biologics SK17Jan13Analytical Control Strategy for Biologics SK17Jan13
Analytical Control Strategy for Biologics SK17Jan13
 
Sun pharma ppt
Sun pharma pptSun pharma ppt
Sun pharma ppt
 
Analytical Method &amp; Technology Transfer Ispe Guide
Analytical Method &amp; Technology Transfer Ispe GuideAnalytical Method &amp; Technology Transfer Ispe Guide
Analytical Method &amp; Technology Transfer Ispe Guide
 

Similar to The Analytical Method Transfer Process SK-Sep 2013

Analytical mehod validation explained sadasiva
Analytical mehod validation explained sadasivaAnalytical mehod validation explained sadasiva
Analytical mehod validation explained sadasivaSada Siva Rao Maddiguntla
 
Analytical mehod validation explained sadasiva
Analytical mehod validation explained sadasivaAnalytical mehod validation explained sadasiva
Analytical mehod validation explained sadasivaSada Siva Rao Maddiguntla
 
The late-stage analytical method lifecycle - Risk-based validation and extens...
The late-stage analytical method lifecycle - Risk-based validation and extens...The late-stage analytical method lifecycle - Risk-based validation and extens...
The late-stage analytical method lifecycle - Risk-based validation and extens...Stephan O. Krause, PhD
 
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...Randox
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validationSai Praveen Reddy
 
Risk-based Analytical Method Validation and Maintenance Strategies SK-Sep13
Risk-based Analytical Method Validation and Maintenance Strategies SK-Sep13Risk-based Analytical Method Validation and Maintenance Strategies SK-Sep13
Risk-based Analytical Method Validation and Maintenance Strategies SK-Sep13Stephan O. Krause, PhD
 
Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods Mostafa Mahmoud
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validationroshankhetade2
 
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdfANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdfBALASUNDARESAN M
 
validation and calibration of HPLC
validation and calibration of HPLCvalidation and calibration of HPLC
validation and calibration of HPLCSakshi Garg
 
Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)
Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)
Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)Aamir Ijaz Brig
 
Analytical Method Validation.pptx
Analytical Method Validation.pptxAnalytical Method Validation.pptx
Analytical Method Validation.pptxBholakant raut
 
Validation of Analytical Procedures.pdf
Validation of Analytical Procedures.pdfValidation of Analytical Procedures.pdf
Validation of Analytical Procedures.pdfpharma344
 
Analytical Method Development
Analytical Method DevelopmentAnalytical Method Development
Analytical Method DevelopmentBijesh Verma
 
Points to Consider in QC Method Validation and Transfer for Biological Products
Points to Consider in QC Method Validation and Transfer for Biological ProductsPoints to Consider in QC Method Validation and Transfer for Biological Products
Points to Consider in QC Method Validation and Transfer for Biological ProductsWeijun Li
 
analytical method validation and validation of hplc
analytical method validation and validation of hplcanalytical method validation and validation of hplc
analytical method validation and validation of hplcvenkatesh thota
 
Analytical method validation by manoj ingale(best ppts)
Analytical method validation by manoj ingale(best ppts)Analytical method validation by manoj ingale(best ppts)
Analytical method validation by manoj ingale(best ppts)Indus Biotech Pvt.Ltd.
 
Stochastic Process
Stochastic ProcessStochastic Process
Stochastic Processknksmart
 
Internal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation labInternal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation labAnkit Raiyani
 
Quality assurance part_2
Quality assurance part_2Quality assurance part_2
Quality assurance part_2ThorikulHuda2
 

Similar to The Analytical Method Transfer Process SK-Sep 2013 (20)

Analytical mehod validation explained sadasiva
Analytical mehod validation explained sadasivaAnalytical mehod validation explained sadasiva
Analytical mehod validation explained sadasiva
 
Analytical mehod validation explained sadasiva
Analytical mehod validation explained sadasivaAnalytical mehod validation explained sadasiva
Analytical mehod validation explained sadasiva
 
The late-stage analytical method lifecycle - Risk-based validation and extens...
The late-stage analytical method lifecycle - Risk-based validation and extens...The late-stage analytical method lifecycle - Risk-based validation and extens...
The late-stage analytical method lifecycle - Risk-based validation and extens...
 
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
Risk-based Analytical Method Validation and Maintenance Strategies SK-Sep13
Risk-based Analytical Method Validation and Maintenance Strategies SK-Sep13Risk-based Analytical Method Validation and Maintenance Strategies SK-Sep13
Risk-based Analytical Method Validation and Maintenance Strategies SK-Sep13
 
Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdfANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
 
validation and calibration of HPLC
validation and calibration of HPLCvalidation and calibration of HPLC
validation and calibration of HPLC
 
Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)
Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)
Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)
 
Analytical Method Validation.pptx
Analytical Method Validation.pptxAnalytical Method Validation.pptx
Analytical Method Validation.pptx
 
Validation of Analytical Procedures.pdf
Validation of Analytical Procedures.pdfValidation of Analytical Procedures.pdf
Validation of Analytical Procedures.pdf
 
Analytical Method Development
Analytical Method DevelopmentAnalytical Method Development
Analytical Method Development
 
Points to Consider in QC Method Validation and Transfer for Biological Products
Points to Consider in QC Method Validation and Transfer for Biological ProductsPoints to Consider in QC Method Validation and Transfer for Biological Products
Points to Consider in QC Method Validation and Transfer for Biological Products
 
analytical method validation and validation of hplc
analytical method validation and validation of hplcanalytical method validation and validation of hplc
analytical method validation and validation of hplc
 
Analytical method validation by manoj ingale(best ppts)
Analytical method validation by manoj ingale(best ppts)Analytical method validation by manoj ingale(best ppts)
Analytical method validation by manoj ingale(best ppts)
 
Stochastic Process
Stochastic ProcessStochastic Process
Stochastic Process
 
Internal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation labInternal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation lab
 
Quality assurance part_2
Quality assurance part_2Quality assurance part_2
Quality assurance part_2
 

The Analytical Method Transfer Process SK-Sep 2013

  • 1. 1 The Analytical Method Transfer Process PDAAMD-AMV Workshop Baltimore, MD 7-8 October 2013 Stephan O. Krause, Ph.D. Principal Scientist, Regulatory Science, Development MedImmune
  • 2. 2 The Analytical Method Transfer Process Agenda:  Introduction to Analytical Method Transfer (AMT)  Mapping out the Overall Product Development and Tech Transfer Process  Method Types and Acceptance Criteria  Analytical Platform Methods vs. New Methods  AMT Example Krause/PDA, 2012
  • 3. 3 The Analytical Method Life Cycle Krause/PDA, 2012
  • 4. CQA Development, CMC Changes, Specifications 4
  • 5. 5 The Five General AMV/AMT Classes AMV Class Description Typical Risk / Uncertainty Level (1=Low, 5=High) Suggested Prospective AMV Studies AMV Class No. Analytical Method Product / Process Sample A New New 4-5 Full Validation B New Old (Validated) 3-4(1) Full Validation Plus AMR(2) Studies C Analytical Platform Technology (not validated “as run”) New 2-3 Partial Validation D Old (Validated) New 1-2 Partial Validation or Verification E Compendial New 1-2 Verification per USP <1226> (1) If a new analytical method (forced method replacement) is needed due to supply reasons, the risk level can be generally considered higher because no other option may exist. Unforced test method replacements can be considered to be a lower risk level as more time may be available to optimize the method performance. (2) AMR = Analytical Method Replacement. A study to confirm that a new analytical method can perform equally or better than the existing one. From Krause, PDA/DHI 2007.
  • 6. 6 Risk-Based AMT Protocol Acceptance Criteria Krause/PDA, 2012
  • 7. 7 General AMT Strategy Krause/PDA, 2012  Co-validation/Co-qualification – this may be used early in the life cycle of a test method when appropriate.  Comparative study – AMT study performed concurrently by sending and receiving laboratories. Acceptance criteria determine the equivalence of the two laboratories. Historical and validation data may be used when appropriate for parts of the method transfer study. The sending laboratory typically has collected a significant amount of historical data for test method performance results in addition to test results for the samples to be tested at the receiving laboratory.  Performance Verification - The receiving laboratory may already perform the method for a similar product or for another type of sample for the same product. In this case, a formal method transfer may not be required. Any reduced prospective study considered should be properly justified.
  • 8. 8 Retrospective and Prospective Use of Data for AMV Studies from other Processes Prior to AMV – New Method
  • 9. 9 Retrospective and Prospective Use of Data for AMV Studies from other Processes Prior to AMV – Platform Method
  • 10. 10 “Fixed” vs. “Variable” AMT Execution Plan  A fixed execution plan does not integrate test method result variation and the minimum number of test replicates needed to obtain a desirable statistical confidence in the transfer results. A fixed execution matrix can be more advantageous when transferring multiple products to/from multiple locations.  A variable execution plan does consider test method result variation and may require a larger data sets, especially for test methods with relatively high result variation. For example, a variable execution plan may be advantageous when transferring bioassays with an expected high degree of test result variation. Krause/PDA, 2012
  • 11. 11 Typical “Fixed” AMT Execution Plan for Late-Stage/Commercial Products Laboratory Day Analyst Instrument Replicates Sending 1 1 1 3 Sending 1 2 2 3 Sending 2 1 1 3 Sending 2 2 2 3 Sending 3 1 1 3 Sending 3 2 2 3 Receiving 1 1 1 3 Receiving 1 2 2 3 Receiving 2 1 1 3 Receiving 2 2 2 3 Receiving 3 1 1 3 Receiving 3 2 2 3 From Krause, PDA/DHI 2007.
  • 12. 12 Analytical Method Transfer (AMT) Example  A validated analytical method for potency is to be transferred from the original QC laboratory to another QC laboratory to release drug product (DP). The analytical method generates potency (dose) results for lyophilized DP.  The vials are available in three nominal doses between 500 – 2000 IU/vial using an identical formulation. Release testing is performed using three replicate preparations from each of three vials.  Before analysis the content of a vial is reconstituted with 5.0 mL of WFI water and the potency is measured in IU/mL (100 – 400 IU/mL).  The samples and a product-specific reference standard are prepared similarly. The analytical method procedure and statistical evaluation are performed with the parallel-line concept.  The “variable” AMT execution plan is used.
  • 13. 13 AMT Study Design and Acceptance Criteria Characteristics Evaluated Accuracy/Matching: The relative difference between lab means should at 90% confidence not be less than -Θ= 10% and not more than +Θ = 10%. The 10% difference limit was set with consideration of product specification. Intermediate Precision: RSD 6 % for all sample types, with appropriate homoscedasticity throughout the potency range (from validation results). This means that any RSD from a sample of n=8 should not exceed 9.43 % Number of Replicates Nreplicates = at least 23 independent replicates The confidence interval for the lab-to-lab difference for N determinations to less than the [10%, +10%]. As above the 10% difference limit was set with consideration of product specification. Samples to test Nlevel = 3 The range of potency/dosing results is covered by: Lowest dose 500 IU/vial or 100 IU/mL Medium dose 1000 IU/vial or 200 IU/mL Highest dose 2000 IU/vial or 400 IU/mL Testing design, each sample Number of operators, n = 2 Number of days, n = 2 Number of replicates per day per operator, n = 2 N = 8 in each lab for each of n= 3 potency levels. Results are converted to “% recoveries vs. expected” to allow pooling Total NTotal = 24 individual observations will be recorded for each laboratory. N=24 individual observations are needed as N=23 is the minimum number of replicates calculated.
  • 14. 14 AMT Results from Sending and Receiving Labs Theoretical Potency Level in IU/mL Operator Day Replicate Sending lab Receiving lab Experimental Potency in IU/mL %Recovery vs. Theoretical Potency Experimental Potency in IU/mL %Recovery vs. Theoretical Potency 100 1 1 1 103 103.0 95 95.0 100 1 1 2 104 104.0 99 99.0 100 1 2 1 108 108.0 104 104.0 100 1 2 2 101 101.0 103 103.0 100 2 1 1 94 94.0 93 93.0 100 2 1 2 99 99.0 96 96.0 100 2 2 1 102 102.0 92 92.0 100 2 2 2 104 104.0 100 100.0 200 1 1 1 212 106.0 208 104.0 200 1 1 2 208 104.0 192 96.0 200 1 2 1 191 95.5 199 99.5 200 1 2 2 201 100.5 195 97.5 200 2 1 1 204 102.0 208 104.0 200 2 1 2 206 103.0 211 105.5 200 2 2 1 198 99.0 203 101.5 200 2 2 2 200 100.0 183 91.5 400 1 1 1 375 93.8 383 95.8 400 1 1 2 401 100.3 401 100.3 400 1 2 1 408 102.0 389 97.3 400 1 2 2 388 97.0 391 97.8 400 2 1 1 402 100.5 408 102.0 400 2 1 2 415 103.8 421 105.3 400 2 2 1 406 101.5 415 103.8 400 2 2 2 410 102.5 403 100.8
  • 15. 15 AMT Result Summary from Sending and Receiving Labs (1) Raw data was used unrounded. Upper and lower 90% CIs were calculated using equation below. (2) N=24 data points are not independent. Separate and Pooled Potency Levels Evaluated Sending lab Receiving lab Statistical Parameters %Recovery vs. Theoretical Potency Statistical Parameters %Recovery vs. Theoretical Potency TOST with acceptance criteria [-10%, +10%] (1) N1 24(2) N2 24 Mean1 101.1 Mean2 99.3 SD 3.5(2) SD 4.2 RSD 3.4 RSD 4.2 Pooled SD(2) 3.9 Mean1-Mean2 1.8 t-value 1.679 Upper 90% CI limit(1) 4 (3.6) Lower 90% CI limit(1) 0 (-0.1) Transfer Acceptance Conclusion Pass Θ+<      +××±−<Θ− −+− 21 2,2121 11 )( 21 nn stxx pnnα
  • 16. 16 Graphical Representation of Potency Results Per Potency Level Between Laboratories  The boxes represent the 25th – 75th percentile distribution of the results for the two laboratories. Medians (line in the box) and means (cross in the box) are approximately centered while the medians are equidistant from the box hinges, providing a visual indication for a normal data distribution(s) among data points within each laboratory set.  One potential outlier (lower open circle outside of the whiskers) is observed in the sending lab, however, this does not change the overall interpretation for the demonstration of lab-to-lab equivalence.  The variation in the test results (wider 25th – 75th percentile boxes) appears to be higher in the receiving laboratory which may be attributed to less test method execution experience. 80 90 100 110 120 Sending lab Receiving lab %Recoveryversustheoreticalpotency(in%)
  • 17. 17 Graphical Representation of the Combined Percent Recoveries Between Laboratories for all Concentration Levels
  • 18. 18 Summary  AMTs may occur at different product development stages.  AMT study execution matrix and the acceptance criteria should be developed based on risk(s).  Many thanks to: • Rashmi Rawat (Product Quality Reviewer, CDER) • Pat Cash (Sr. Director, Analytical Biochemistry, MedImmune) • Mark Schenerman (VP, Analytical Biochemistry, MedImmune) • Martin Van Trieste (SVP, Quality, Amgen) • Rich Levy (SVP, PDA) • Pierre Douette, Ph.D., Eurogentec S.A., Belgium • Michael Warncke, Ph.D., Bayer HealthCare, USA • Earl K. Zablackis, Ph.D., Sanofi Pasteur, USA Krause/PDA, 2012