FLUOROSCOPIC IMAGING
MODERATOR- DR.RAMAKANTH M.D.R.D.
PRESENTER- DR.JAYA ADITYA
First generation fluoroscopes
• Consisted of an x-ray tube, an x-ray table, and a fluoroscopic screen.
• The fluorescent material in the screen was copper activated zinc
cadmium sulfide that emitted light in the yellow-green spectrum.
• A sheet of lead glass covered the screen, so the radiologist could stare
directly into the screen without having the x-ray beam strike his eyes.
• Fluoroscopic examinations were carried out in a dark room.
• The solution came in the form of the x-ray image intensifier.
IMAGE INTENSIFIER DESIGN
• Contains four basic elements:
1. Input phosphor and photocathode
2. Electrostatic focusing lens
3. Accelerating anode
4. Output phosphor
• After an x-ray beam passes through the patient, it enters the image
intensifier tube.
• The input screen absorbs x-ray photons and converts their energy into
light photons. The light photons strike the photocathode, causing it to
emit photoelectrons.
• As the electrons flow from the cathode to anode, they are focused by
an electrostatic lens, which guides them to the output screen.
• The electrons strike the output screen, which emits the light photons
that carry the fluoroscopic image
The image is carried in the following order
x-ray
photons
light
photons
next by
electrons
light
photons.
Input Phosphor
• The input fluorescent screen in image intensifiers is cesium iodide
(Csl).
• Csl is deposited on a thin aluminium substrate during the deposition
process the crystals of CsI grow in tiny needles perpendicular to the
substrate.
• When one of these needle-shaped crystals absorbs an x-ray photon
and produces light, the light will be transmitted to the photocathode
with little scattering.
• 4 physical characteristics of cesium iodide make it superior to the
older zinc -cadmium sulfide screens.
• The vertical orientation of the crystals
• A greater packing density
• More favorable effective atomic number
• Packing density of cesium iodide is three times greater than that of
zinc-cadmium sulfide.
Photocathode
• The photocathode is a photoemissive metal.
• When light from the fluorescent screen strikes the
photocathode,photoelectrons are emitted.
• The numbers are proportional to the brightness of the screen.
Electrostratic Focusing Lens
• The lens is made up of a series of positively charged electrodes that
are usually plated onto the inside surface of the glass envelope.
• These electrodes focus the electron beam as it flows from the
photocathode toward the output phosphor.
Point inversion
• Electron focusing inverts and reverses the image.
• This is called a point inversion because all the electrons pass through
a common focal point on their way to the output phosphor.
• Each point on the input phosphor is focused to a specific point on the
opposite side of the output phosphor.
• The input phosphor is curved to ensure that electrons emitted at the
peripheral regions of the photocathode travel the same distance as
those emitted from the central region.
• The image on the output phosphor is reduced in size.
• This is one of the principal reasons why it is brighter.
Accelerating Anode.
• Located in the neck of the image tube.
• Its function is to accelerate electrons emitted from the photocathode
toward the output screen.
Output Phosphor
• Is silver activated zinc-cadmium sulphide.
• Crystal size and layer thickness are reduced to maintain resolution in
the minified image.
• The number of light photons is increased approximately 50-fold.
• A thin layer of aluminum is plated onto the fluorescent screen to
prevent light from moving retrograde through the tube and activating
the photocathode.
• The glass tube of the image intensifier is about 2 to 4 mm thick, and is
enclosed in a lead-lined metal container.
• The lead lining protects the operator from stray radiation.
Viewing the output of an image intensifier
• Directly- Mirror optical system
• Indirectly- Closed circuit television.
Mirror optical system
• Light is reflected and focused several times.
• This can be done with only minimal loss of brightness.
•Disadvantages-
• Image is only visible in a small viewing angle.
• Freedom of movement is limited by the mirror.
• Only one observer can view the image.
Closed circuit television
• In modern systems, Viewing the output of an image intensifier is done
via a closed circuit television chain.
• We could record the film image from the TV camera signal.
• It is better to expose the film directly to the output phosphor of the
image intensifier tube.
• To maintain continuous TV viewing while exposing the film, we must
split the light from the II output into two paths at the time of film
exposure.
• During routine fluoroscopy, all the light output of the II is directed to
the TV camera.
• When film mode is selected, a semi transparent mirror is positioned
in the light beam.
Film mode is
selected
About 90% goes
to the film
camera
10% produce a
satisfactory TV
image
• In some systems, the image is coupled to the TV camera by a
fiberoptic bundle.
• Fiberoptic coupling precludes filming directly from the output
phosphor of the image intensifier tube.
• A fiber face plate is a bundle of fine optically shielded glass fibers
(several thousand per mm2) that is a few millimeters thick.
BRIGHTNESS GAIN -Two methods
• The first compares the luminance of an intensifier output screen to
that of a Pattersontype B-2 fluoroscopy screen when both are
exposed to the same quantity of radiation.
• The brightness gain is the ratio of the two illuminations:
• Intensifier luminance
• Brightness gain = Patterson B-2 luminance
• Pattersontype B-2 fluoroscopic screens, however, vary from one batch
to another, and deteriorate at an unpredictable rate with time.
• Because of this,ICRU, has recommended a second method of
evaluation.
• The conversion factor is a ratio of the luminance of the output
phosphor to the input exposure rate:
cd/m2
• Convers1on factor =mRisec
• Candelas-Output screen luminance
• The conversion factor usually equals about 1% of the brightness gain.
• The brightness gain of an image intensifier comes from two
completely unrelated sources-
• Minification gain
• Flux gain
• Brightness gain = Minification gain x Flux gain
Minification Gain
• The brightness gain from minification is produced by a reduction in
image size.
• The quantity of the gain depends on the relative areas of the input
and output screens.
• Minification gain = (d1/d2)2
Flux Gain
• For each light photon from the input screen, 50 light photons are
emitted by the output screen.
• Flux gain increases the brightness of the fluoroscopic image by a
factor of approximately 50.
one light photon from the input
screen eject one electron from the
photocathode.
The electron is accelerated to the
opposite end of the tube, gaining
enough energy to produce 50 light
photons at the output screen
IMAGING CHARACTERISTICS
• Contrast
• Lag
• Distortion
Contrast
A 1/4in. thick lead disc
is placed over the
center of the input
screen
The input phosphor is
exposed to a specified
quantity of radiation
Brightness is
measured at the
output phosphor
• Contrast is the brightness ratio of the periphery to the center of the
output screen.
• Contrast ratios range from approximately 10:1 to better than 20:1.
• Two factors tend to diminish contrast-
1) The input screen does not absorb all the photons in the x-ray beam.
• Some are transmitted through the intensifier tube, and a few are
eventually absorbed by the output screen.
• These transmitted photons contribute to the illumination of the
output phosphor produce a background of fog that reduces image
contrast.
• 2) Retrograde light flow from the output screen.
• Most retrograde light flow is blocked by a thin layer of aluminum on
the back of the screen.
Some light
photons
penetrate
through the
aluminium.
Pass back
through the
image tube
Activate the
photocathode
Cathode emits
photoelectrons
Electrons
produce "fog“
Reduce image
contrast
Lag
• Persistence of luminescence after x-ray stimulation has been
terminated.
• With older image tubes, lag times were 30 to 40 ms.
• With Csl tubes, lag times are about 1ms.
Distorsion
• The electric fields that accurately control electrons in the center of
the image are not capable of the same degree of control for
peripheral electrons.
• Peripheral electrons tend to flare out from an ideal course, do not
strike the output phosphor where they ideally should.
• The result is unequal magnification, which produces peripheral
distortion.
• The amount of distortion is always greater with large intensifiers.
Pincushion effect
Vignetting
• The center of the output screen is brighter than the periphery.
• The peripheral image is displayed over a larger area of the output
screen, and thus its brightness gain from minification is less than that
in the center.
• A fall-off in brightness at the periphery of an image is called
vignetting.
MULTIPLE-FIELD IMAGE INTENSIFIERS
• Resolve the conflicts between image size and quality.
• They can be operated in several modes, including a 4.5- in., a 6-in., or
9-in. mode.
• Field size is changed by applying a simple electronic principle:
• The higher the voltage on the electrostatic focusing lens, the more
the electron beam is focused.
• In the 9-in. mode, the electrostatic focusing voltage is decreased.
• The electrons focus to a point, or cross, close to the output phosphor,
and the final image is actually smaller than the phosphor.
• In the 6-in. mode the electrostatic focusing voltage is increased, and
the electrons focus farther away from the output phosphor.
• After the electrons cross, they diverge, so the image on the output
phosphor is larger than in the 9-in. mode.
THANK YOU

More Related Content

What's hot

Radiographic exposure and image quality
Radiographic exposure and image qualityRadiographic exposure and image quality
Radiographic exposure and image quality
Rad Tech
 

What's hot (20)

X ray beam rest
X ray beam restX ray beam rest
X ray beam rest
 
X ray films - mamita
X ray films - mamitaX ray films - mamita
X ray films - mamita
 
Attenuation
AttenuationAttenuation
Attenuation
 
image intensifier.pptx
image intensifier.pptximage intensifier.pptx
image intensifier.pptx
 
Silver recovery method converted
Silver recovery method convertedSilver recovery method converted
Silver recovery method converted
 
DAY LIGHT PROCESSING ASSIGNMENT.pptx
DAY LIGHT PROCESSING ASSIGNMENT.pptxDAY LIGHT PROCESSING ASSIGNMENT.pptx
DAY LIGHT PROCESSING ASSIGNMENT.pptx
 
Filters
FiltersFilters
Filters
 
Automatic exposure control
Automatic exposure controlAutomatic exposure control
Automatic exposure control
 
Xray beam restrictors 1436711280968
Xray beam restrictors 1436711280968Xray beam restrictors 1436711280968
Xray beam restrictors 1436711280968
 
Fluoroscopy-Rohit.pptx
Fluoroscopy-Rohit.pptxFluoroscopy-Rohit.pptx
Fluoroscopy-Rohit.pptx
 
fluoroscopy imaging
 fluoroscopy imaging  fluoroscopy imaging
fluoroscopy imaging
 
Viewing and storage of x ray film
Viewing and storage of x ray filmViewing and storage of x ray film
Viewing and storage of x ray film
 
Radiographic Image contrast & image resolution
Radiographic Image contrast & image resolutionRadiographic Image contrast & image resolution
Radiographic Image contrast & image resolution
 
Fluoroscopy
FluoroscopyFluoroscopy
Fluoroscopy
 
Radiographic exposure and image quality
Radiographic exposure and image qualityRadiographic exposure and image quality
Radiographic exposure and image quality
 
conventional fluoroscopy imaging system
conventional fluoroscopy imaging systemconventional fluoroscopy imaging system
conventional fluoroscopy imaging system
 
Sensitometry & characteristic curve
Sensitometry & characteristic curveSensitometry & characteristic curve
Sensitometry & characteristic curve
 
X ray tube
X ray tubeX ray tube
X ray tube
 
Dry Film Technology Swastik
Dry Film Technology  SwastikDry Film Technology  Swastik
Dry Film Technology Swastik
 
X ray tube
X ray tubeX ray tube
X ray tube
 

Similar to Fluoroscopic Imaging.pptx

Floroscopynewmicrosoftofficepowerpoint97 2003presentation2-131218145222-phpapp01
Floroscopynewmicrosoftofficepowerpoint97 2003presentation2-131218145222-phpapp01Floroscopynewmicrosoftofficepowerpoint97 2003presentation2-131218145222-phpapp01
Floroscopynewmicrosoftofficepowerpoint97 2003presentation2-131218145222-phpapp01
Ruta Angel
 
Flouroscopic imging
Flouroscopic imgingFlouroscopic imging
Flouroscopic imging
Sahith Reddy
 
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
mr_koky
 
flouroscopicimging-120617043323-phpapp01.pdf
flouroscopicimging-120617043323-phpapp01.pdfflouroscopicimging-120617043323-phpapp01.pdf
flouroscopicimging-120617043323-phpapp01.pdf
VishnuDutt40
 
Electron microscopy
Electron microscopyElectron microscopy
Electron microscopy
Ganga Huvin
 

Similar to Fluoroscopic Imaging.pptx (20)

fluoroscopy-3.pptx
fluoroscopy-3.pptxfluoroscopy-3.pptx
fluoroscopy-3.pptx
 
Fluoroscopy
 Fluoroscopy  Fluoroscopy
Fluoroscopy
 
Floroscopynewmicrosoftofficepowerpoint97 2003presentation2-131218145222-phpapp01
Floroscopynewmicrosoftofficepowerpoint97 2003presentation2-131218145222-phpapp01Floroscopynewmicrosoftofficepowerpoint97 2003presentation2-131218145222-phpapp01
Floroscopynewmicrosoftofficepowerpoint97 2003presentation2-131218145222-phpapp01
 
Fluroscopy
Fluroscopy Fluroscopy
Fluroscopy
 
Flouroscopic imging
Flouroscopic imgingFlouroscopic imging
Flouroscopic imging
 
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
 
flouroscopicimging-120617043323-phpapp01.pdf
flouroscopicimging-120617043323-phpapp01.pdfflouroscopicimging-120617043323-phpapp01.pdf
flouroscopicimging-120617043323-phpapp01.pdf
 
Mi ppt fi print this
Mi ppt fi print thisMi ppt fi print this
Mi ppt fi print this
 
Image quality in fluroscopy
Image quality in fluroscopyImage quality in fluroscopy
Image quality in fluroscopy
 
Various types of microscopes and microscopy
Various types of microscopes and microscopy   Various types of microscopes and microscopy
Various types of microscopes and microscopy
 
Fluroscopic imaging ppt
Fluroscopic imaging pptFluroscopic imaging ppt
Fluroscopic imaging ppt
 
Flouroscopic imaging
Flouroscopic imagingFlouroscopic imaging
Flouroscopic imaging
 
Flouroscopy
FlouroscopyFlouroscopy
Flouroscopy
 
Confocal microscope presentation pt
Confocal microscope presentation  ptConfocal microscope presentation  pt
Confocal microscope presentation pt
 
Flim screen physics
Flim screen physicsFlim screen physics
Flim screen physics
 
CR DR system.pptx
CR DR system.pptxCR DR system.pptx
CR DR system.pptx
 
Electron microscopy
Electron microscopyElectron microscopy
Electron microscopy
 
Introduction to cathode Ray Tube.pptx
Introduction   to   cathode Ray Tube.pptxIntroduction   to   cathode Ray Tube.pptx
Introduction to cathode Ray Tube.pptx
 
Molecular fluorescence spectroscophy
Molecular  fluorescence spectroscophyMolecular  fluorescence spectroscophy
Molecular fluorescence spectroscophy
 
Measurement of optical fiber refractive index
Measurement of optical fiber refractive indexMeasurement of optical fiber refractive index
Measurement of optical fiber refractive index
 

More from Srinath Chowdary (20)

k3.pptx
k3.pptxk3.pptx
k3.pptx
 
HSG AND FISTULOGRAM.pptx
HSG AND FISTULOGRAM.pptxHSG AND FISTULOGRAM.pptx
HSG AND FISTULOGRAM.pptx
 
DOC-20220623-WA0005..pptx
DOC-20220623-WA0005..pptxDOC-20220623-WA0005..pptx
DOC-20220623-WA0005..pptx
 
PACS.pptx
PACS.pptxPACS.pptx
PACS.pptx
 
5_6282657037939640532.pptx
5_6282657037939640532.pptx5_6282657037939640532.pptx
5_6282657037939640532.pptx
 
CNS3.pptx
CNS3.pptxCNS3.pptx
CNS3.pptx
 
IVP.pptx
IVP.pptxIVP.pptx
IVP.pptx
 
TRAUMATIC DIAPHRAGMATIC INJURY.pptx
TRAUMATIC DIAPHRAGMATIC INJURY.pptxTRAUMATIC DIAPHRAGMATIC INJURY.pptx
TRAUMATIC DIAPHRAGMATIC INJURY.pptx
 
PLANNING OF X-RAY, CT ROOMS AND QUALITY ASSURANCE.pptx
PLANNING OF X-RAY, CT ROOMS AND QUALITY ASSURANCE.pptxPLANNING OF X-RAY, CT ROOMS AND QUALITY ASSURANCE.pptx
PLANNING OF X-RAY, CT ROOMS AND QUALITY ASSURANCE.pptx
 
CNS2.pptx
CNS2.pptxCNS2.pptx
CNS2.pptx
 
Journal pcnsl-1.pptx
Journal pcnsl-1.pptxJournal pcnsl-1.pptx
Journal pcnsl-1.pptx
 
Journal pcnsl.pptx
Journal pcnsl.pptxJournal pcnsl.pptx
Journal pcnsl.pptx
 
ppt.pptx
ppt.pptxppt.pptx
ppt.pptx
 
RADIATION DETECTION AND MEASUREMENT ppt 2.pptx
RADIATION DETECTION AND MEASUREMENT ppt 2.pptxRADIATION DETECTION AND MEASUREMENT ppt 2.pptx
RADIATION DETECTION AND MEASUREMENT ppt 2.pptx
 
Copy-CNS 1.pptx
Copy-CNS 1.pptxCopy-CNS 1.pptx
Copy-CNS 1.pptx
 
GENERATORS
GENERATORS GENERATORS
GENERATORS
 
X RAY TUBES & PRODUCTION OF.pptx
X RAY TUBES & PRODUCTION OF.pptxX RAY TUBES & PRODUCTION OF.pptx
X RAY TUBES & PRODUCTION OF.pptx
 
5_6134079791160100714.pptx
5_6134079791160100714.pptx5_6134079791160100714.pptx
5_6134079791160100714.pptx
 
CVJ lines and angles.pdf
CVJ lines and angles.pdfCVJ lines and angles.pdf
CVJ lines and angles.pdf
 
fractures and interpretation of x-rays-1.pptx
fractures and interpretation of x-rays-1.pptxfractures and interpretation of x-rays-1.pptx
fractures and interpretation of x-rays-1.pptx
 

Recently uploaded

Abort pregnancy in research centre+966_505195917 abortion pills in Kuwait cyt...
Abort pregnancy in research centre+966_505195917 abortion pills in Kuwait cyt...Abort pregnancy in research centre+966_505195917 abortion pills in Kuwait cyt...
Abort pregnancy in research centre+966_505195917 abortion pills in Kuwait cyt...
drmarathore
 
➥🔝 7737669865 🔝▻ Muzaffarpur Call-girls in Women Seeking Men 🔝Muzaffarpur🔝 ...
➥🔝 7737669865 🔝▻ Muzaffarpur Call-girls in Women Seeking Men  🔝Muzaffarpur🔝  ...➥🔝 7737669865 🔝▻ Muzaffarpur Call-girls in Women Seeking Men  🔝Muzaffarpur🔝  ...
➥🔝 7737669865 🔝▻ Muzaffarpur Call-girls in Women Seeking Men 🔝Muzaffarpur🔝 ...
amitlee9823
 
Abortion pills in Jeddah |+966572737505 | Get Cytotec
Abortion pills in Jeddah |+966572737505 | Get CytotecAbortion pills in Jeddah |+966572737505 | Get Cytotec
Abortion pills in Jeddah |+966572737505 | Get Cytotec
Abortion pills in Riyadh +966572737505 get cytotec
 
一比一原版(Otago毕业证书)奥塔哥理工学院毕业证成绩单学位证靠谱定制
一比一原版(Otago毕业证书)奥塔哥理工学院毕业证成绩单学位证靠谱定制一比一原版(Otago毕业证书)奥塔哥理工学院毕业证成绩单学位证靠谱定制
一比一原版(Otago毕业证书)奥塔哥理工学院毕业证成绩单学位证靠谱定制
uodye
 
Kothanur Call Girls Service: 🍓 7737669865 🍓 High Profile Model Escorts | Bang...
Kothanur Call Girls Service: 🍓 7737669865 🍓 High Profile Model Escorts | Bang...Kothanur Call Girls Service: 🍓 7737669865 🍓 High Profile Model Escorts | Bang...
Kothanur Call Girls Service: 🍓 7737669865 🍓 High Profile Model Escorts | Bang...
amitlee9823
 
Call Girls In RT Nagar ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In RT Nagar ☎ 7737669865 🥵 Book Your One night StandCall Girls In RT Nagar ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In RT Nagar ☎ 7737669865 🥵 Book Your One night Stand
amitlee9823
 
Abortion Pill for sale in Riyadh ((+918761049707) Get Cytotec in Dammam
Abortion Pill for sale in Riyadh ((+918761049707) Get Cytotec in DammamAbortion Pill for sale in Riyadh ((+918761049707) Get Cytotec in Dammam
Abortion Pill for sale in Riyadh ((+918761049707) Get Cytotec in Dammam
ahmedjiabur940
 
Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...
Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...
Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...
amitlee9823
 
CHEAP Call Girls in Hauz Quazi (-DELHI )🔝 9953056974🔝(=)/CALL GIRLS SERVICE
CHEAP Call Girls in Hauz Quazi  (-DELHI )🔝 9953056974🔝(=)/CALL GIRLS SERVICECHEAP Call Girls in Hauz Quazi  (-DELHI )🔝 9953056974🔝(=)/CALL GIRLS SERVICE
CHEAP Call Girls in Hauz Quazi (-DELHI )🔝 9953056974🔝(=)/CALL GIRLS SERVICE
9953056974 Low Rate Call Girls In Saket, Delhi NCR
 
➥🔝 7737669865 🔝▻ Deoghar Call-girls in Women Seeking Men 🔝Deoghar🔝 Escorts...
➥🔝 7737669865 🔝▻ Deoghar Call-girls in Women Seeking Men  🔝Deoghar🔝   Escorts...➥🔝 7737669865 🔝▻ Deoghar Call-girls in Women Seeking Men  🔝Deoghar🔝   Escorts...
➥🔝 7737669865 🔝▻ Deoghar Call-girls in Women Seeking Men 🔝Deoghar🔝 Escorts...
amitlee9823
 
(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...
(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...
(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...
motiram463
 
怎样办理斯威本科技大学毕业证(SUT毕业证书)成绩单留信认证
怎样办理斯威本科技大学毕业证(SUT毕业证书)成绩单留信认证怎样办理斯威本科技大学毕业证(SUT毕业证书)成绩单留信认证
怎样办理斯威本科技大学毕业证(SUT毕业证书)成绩单留信认证
tufbav
 

Recently uploaded (20)

Abort pregnancy in research centre+966_505195917 abortion pills in Kuwait cyt...
Abort pregnancy in research centre+966_505195917 abortion pills in Kuwait cyt...Abort pregnancy in research centre+966_505195917 abortion pills in Kuwait cyt...
Abort pregnancy in research centre+966_505195917 abortion pills in Kuwait cyt...
 
➥🔝 7737669865 🔝▻ Muzaffarpur Call-girls in Women Seeking Men 🔝Muzaffarpur🔝 ...
➥🔝 7737669865 🔝▻ Muzaffarpur Call-girls in Women Seeking Men  🔝Muzaffarpur🔝  ...➥🔝 7737669865 🔝▻ Muzaffarpur Call-girls in Women Seeking Men  🔝Muzaffarpur🔝  ...
➥🔝 7737669865 🔝▻ Muzaffarpur Call-girls in Women Seeking Men 🔝Muzaffarpur🔝 ...
 
Top Rated Pune Call Girls Katraj ⟟ 6297143586 ⟟ Call Me For Genuine Sex Serv...
Top Rated  Pune Call Girls Katraj ⟟ 6297143586 ⟟ Call Me For Genuine Sex Serv...Top Rated  Pune Call Girls Katraj ⟟ 6297143586 ⟟ Call Me For Genuine Sex Serv...
Top Rated Pune Call Girls Katraj ⟟ 6297143586 ⟟ Call Me For Genuine Sex Serv...
 
Get Premium Pimple Saudagar Call Girls (8005736733) 24x7 Rate 15999 with A/c ...
Get Premium Pimple Saudagar Call Girls (8005736733) 24x7 Rate 15999 with A/c ...Get Premium Pimple Saudagar Call Girls (8005736733) 24x7 Rate 15999 with A/c ...
Get Premium Pimple Saudagar Call Girls (8005736733) 24x7 Rate 15999 with A/c ...
 
Abortion pills in Jeddah |+966572737505 | Get Cytotec
Abortion pills in Jeddah |+966572737505 | Get CytotecAbortion pills in Jeddah |+966572737505 | Get Cytotec
Abortion pills in Jeddah |+966572737505 | Get Cytotec
 
一比一原版(Otago毕业证书)奥塔哥理工学院毕业证成绩单学位证靠谱定制
一比一原版(Otago毕业证书)奥塔哥理工学院毕业证成绩单学位证靠谱定制一比一原版(Otago毕业证书)奥塔哥理工学院毕业证成绩单学位证靠谱定制
一比一原版(Otago毕业证书)奥塔哥理工学院毕业证成绩单学位证靠谱定制
 
Kothanur Call Girls Service: 🍓 7737669865 🍓 High Profile Model Escorts | Bang...
Kothanur Call Girls Service: 🍓 7737669865 🍓 High Profile Model Escorts | Bang...Kothanur Call Girls Service: 🍓 7737669865 🍓 High Profile Model Escorts | Bang...
Kothanur Call Girls Service: 🍓 7737669865 🍓 High Profile Model Escorts | Bang...
 
Call Girls In RT Nagar ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In RT Nagar ☎ 7737669865 🥵 Book Your One night StandCall Girls In RT Nagar ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In RT Nagar ☎ 7737669865 🥵 Book Your One night Stand
 
Abortion Pill for sale in Riyadh ((+918761049707) Get Cytotec in Dammam
Abortion Pill for sale in Riyadh ((+918761049707) Get Cytotec in DammamAbortion Pill for sale in Riyadh ((+918761049707) Get Cytotec in Dammam
Abortion Pill for sale in Riyadh ((+918761049707) Get Cytotec in Dammam
 
HLH PPT.ppt very important topic to discuss
HLH PPT.ppt very important topic to discussHLH PPT.ppt very important topic to discuss
HLH PPT.ppt very important topic to discuss
 
Shikrapur Call Girls Most Awaited Fun 6297143586 High Profiles young Beautie...
Shikrapur Call Girls Most Awaited Fun  6297143586 High Profiles young Beautie...Shikrapur Call Girls Most Awaited Fun  6297143586 High Profiles young Beautie...
Shikrapur Call Girls Most Awaited Fun 6297143586 High Profiles young Beautie...
 
Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...
Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...
Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...
 
Top Rated Pune Call Girls Shirwal ⟟ 6297143586 ⟟ Call Me For Genuine Sex Ser...
Top Rated  Pune Call Girls Shirwal ⟟ 6297143586 ⟟ Call Me For Genuine Sex Ser...Top Rated  Pune Call Girls Shirwal ⟟ 6297143586 ⟟ Call Me For Genuine Sex Ser...
Top Rated Pune Call Girls Shirwal ⟟ 6297143586 ⟟ Call Me For Genuine Sex Ser...
 
CHEAP Call Girls in Hauz Quazi (-DELHI )🔝 9953056974🔝(=)/CALL GIRLS SERVICE
CHEAP Call Girls in Hauz Quazi  (-DELHI )🔝 9953056974🔝(=)/CALL GIRLS SERVICECHEAP Call Girls in Hauz Quazi  (-DELHI )🔝 9953056974🔝(=)/CALL GIRLS SERVICE
CHEAP Call Girls in Hauz Quazi (-DELHI )🔝 9953056974🔝(=)/CALL GIRLS SERVICE
 
➥🔝 7737669865 🔝▻ Deoghar Call-girls in Women Seeking Men 🔝Deoghar🔝 Escorts...
➥🔝 7737669865 🔝▻ Deoghar Call-girls in Women Seeking Men  🔝Deoghar🔝   Escorts...➥🔝 7737669865 🔝▻ Deoghar Call-girls in Women Seeking Men  🔝Deoghar🔝   Escorts...
➥🔝 7737669865 🔝▻ Deoghar Call-girls in Women Seeking Men 🔝Deoghar🔝 Escorts...
 
Call Girls Pimple Saudagar Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Pimple Saudagar Call Me 7737669865 Budget Friendly No Advance BookingCall Girls Pimple Saudagar Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Pimple Saudagar Call Me 7737669865 Budget Friendly No Advance Booking
 
(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...
(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...
(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...
 
怎样办理斯威本科技大学毕业证(SUT毕业证书)成绩单留信认证
怎样办理斯威本科技大学毕业证(SUT毕业证书)成绩单留信认证怎样办理斯威本科技大学毕业证(SUT毕业证书)成绩单留信认证
怎样办理斯威本科技大学毕业证(SUT毕业证书)成绩单留信认证
 
Sector 18, Noida Call girls :8448380779 Model Escorts | 100% verified
Sector 18, Noida Call girls :8448380779 Model Escorts | 100% verifiedSector 18, Noida Call girls :8448380779 Model Escorts | 100% verified
Sector 18, Noida Call girls :8448380779 Model Escorts | 100% verified
 
Pooja 9892124323, Call girls Services and Mumbai Escort Service Near Hotel Th...
Pooja 9892124323, Call girls Services and Mumbai Escort Service Near Hotel Th...Pooja 9892124323, Call girls Services and Mumbai Escort Service Near Hotel Th...
Pooja 9892124323, Call girls Services and Mumbai Escort Service Near Hotel Th...
 

Fluoroscopic Imaging.pptx

  • 1. FLUOROSCOPIC IMAGING MODERATOR- DR.RAMAKANTH M.D.R.D. PRESENTER- DR.JAYA ADITYA
  • 2. First generation fluoroscopes • Consisted of an x-ray tube, an x-ray table, and a fluoroscopic screen. • The fluorescent material in the screen was copper activated zinc cadmium sulfide that emitted light in the yellow-green spectrum. • A sheet of lead glass covered the screen, so the radiologist could stare directly into the screen without having the x-ray beam strike his eyes. • Fluoroscopic examinations were carried out in a dark room. • The solution came in the form of the x-ray image intensifier.
  • 3.
  • 4. IMAGE INTENSIFIER DESIGN • Contains four basic elements: 1. Input phosphor and photocathode 2. Electrostatic focusing lens 3. Accelerating anode 4. Output phosphor
  • 5. • After an x-ray beam passes through the patient, it enters the image intensifier tube. • The input screen absorbs x-ray photons and converts their energy into light photons. The light photons strike the photocathode, causing it to emit photoelectrons. • As the electrons flow from the cathode to anode, they are focused by an electrostatic lens, which guides them to the output screen. • The electrons strike the output screen, which emits the light photons that carry the fluoroscopic image
  • 6. The image is carried in the following order x-ray photons light photons next by electrons light photons.
  • 7. Input Phosphor • The input fluorescent screen in image intensifiers is cesium iodide (Csl). • Csl is deposited on a thin aluminium substrate during the deposition process the crystals of CsI grow in tiny needles perpendicular to the substrate. • When one of these needle-shaped crystals absorbs an x-ray photon and produces light, the light will be transmitted to the photocathode with little scattering.
  • 8. • 4 physical characteristics of cesium iodide make it superior to the older zinc -cadmium sulfide screens. • The vertical orientation of the crystals • A greater packing density • More favorable effective atomic number • Packing density of cesium iodide is three times greater than that of zinc-cadmium sulfide.
  • 9.
  • 10. Photocathode • The photocathode is a photoemissive metal. • When light from the fluorescent screen strikes the photocathode,photoelectrons are emitted. • The numbers are proportional to the brightness of the screen.
  • 11. Electrostratic Focusing Lens • The lens is made up of a series of positively charged electrodes that are usually plated onto the inside surface of the glass envelope. • These electrodes focus the electron beam as it flows from the photocathode toward the output phosphor.
  • 12. Point inversion • Electron focusing inverts and reverses the image. • This is called a point inversion because all the electrons pass through a common focal point on their way to the output phosphor.
  • 13.
  • 14. • Each point on the input phosphor is focused to a specific point on the opposite side of the output phosphor. • The input phosphor is curved to ensure that electrons emitted at the peripheral regions of the photocathode travel the same distance as those emitted from the central region. • The image on the output phosphor is reduced in size. • This is one of the principal reasons why it is brighter.
  • 15. Accelerating Anode. • Located in the neck of the image tube. • Its function is to accelerate electrons emitted from the photocathode toward the output screen.
  • 16. Output Phosphor • Is silver activated zinc-cadmium sulphide. • Crystal size and layer thickness are reduced to maintain resolution in the minified image. • The number of light photons is increased approximately 50-fold.
  • 17.
  • 18. • A thin layer of aluminum is plated onto the fluorescent screen to prevent light from moving retrograde through the tube and activating the photocathode. • The glass tube of the image intensifier is about 2 to 4 mm thick, and is enclosed in a lead-lined metal container. • The lead lining protects the operator from stray radiation.
  • 19. Viewing the output of an image intensifier • Directly- Mirror optical system • Indirectly- Closed circuit television.
  • 20. Mirror optical system • Light is reflected and focused several times. • This can be done with only minimal loss of brightness. •Disadvantages- • Image is only visible in a small viewing angle. • Freedom of movement is limited by the mirror. • Only one observer can view the image.
  • 21.
  • 22. Closed circuit television • In modern systems, Viewing the output of an image intensifier is done via a closed circuit television chain. • We could record the film image from the TV camera signal. • It is better to expose the film directly to the output phosphor of the image intensifier tube.
  • 23. • To maintain continuous TV viewing while exposing the film, we must split the light from the II output into two paths at the time of film exposure. • During routine fluoroscopy, all the light output of the II is directed to the TV camera. • When film mode is selected, a semi transparent mirror is positioned in the light beam.
  • 24. Film mode is selected About 90% goes to the film camera 10% produce a satisfactory TV image
  • 25.
  • 26. • In some systems, the image is coupled to the TV camera by a fiberoptic bundle. • Fiberoptic coupling precludes filming directly from the output phosphor of the image intensifier tube. • A fiber face plate is a bundle of fine optically shielded glass fibers (several thousand per mm2) that is a few millimeters thick.
  • 27. BRIGHTNESS GAIN -Two methods • The first compares the luminance of an intensifier output screen to that of a Pattersontype B-2 fluoroscopy screen when both are exposed to the same quantity of radiation. • The brightness gain is the ratio of the two illuminations: • Intensifier luminance • Brightness gain = Patterson B-2 luminance
  • 28. • Pattersontype B-2 fluoroscopic screens, however, vary from one batch to another, and deteriorate at an unpredictable rate with time. • Because of this,ICRU, has recommended a second method of evaluation. • The conversion factor is a ratio of the luminance of the output phosphor to the input exposure rate: cd/m2 • Convers1on factor =mRisec • Candelas-Output screen luminance • The conversion factor usually equals about 1% of the brightness gain.
  • 29. • The brightness gain of an image intensifier comes from two completely unrelated sources- • Minification gain • Flux gain • Brightness gain = Minification gain x Flux gain
  • 30. Minification Gain • The brightness gain from minification is produced by a reduction in image size. • The quantity of the gain depends on the relative areas of the input and output screens. • Minification gain = (d1/d2)2
  • 31. Flux Gain • For each light photon from the input screen, 50 light photons are emitted by the output screen. • Flux gain increases the brightness of the fluoroscopic image by a factor of approximately 50.
  • 32. one light photon from the input screen eject one electron from the photocathode. The electron is accelerated to the opposite end of the tube, gaining enough energy to produce 50 light photons at the output screen
  • 34. Contrast A 1/4in. thick lead disc is placed over the center of the input screen The input phosphor is exposed to a specified quantity of radiation Brightness is measured at the output phosphor
  • 35. • Contrast is the brightness ratio of the periphery to the center of the output screen. • Contrast ratios range from approximately 10:1 to better than 20:1. • Two factors tend to diminish contrast- 1) The input screen does not absorb all the photons in the x-ray beam. • Some are transmitted through the intensifier tube, and a few are eventually absorbed by the output screen.
  • 36. • These transmitted photons contribute to the illumination of the output phosphor produce a background of fog that reduces image contrast. • 2) Retrograde light flow from the output screen. • Most retrograde light flow is blocked by a thin layer of aluminum on the back of the screen.
  • 37. Some light photons penetrate through the aluminium. Pass back through the image tube Activate the photocathode Cathode emits photoelectrons Electrons produce "fog“ Reduce image contrast
  • 38. Lag • Persistence of luminescence after x-ray stimulation has been terminated. • With older image tubes, lag times were 30 to 40 ms. • With Csl tubes, lag times are about 1ms.
  • 39. Distorsion • The electric fields that accurately control electrons in the center of the image are not capable of the same degree of control for peripheral electrons. • Peripheral electrons tend to flare out from an ideal course, do not strike the output phosphor where they ideally should. • The result is unequal magnification, which produces peripheral distortion. • The amount of distortion is always greater with large intensifiers.
  • 41. Vignetting • The center of the output screen is brighter than the periphery. • The peripheral image is displayed over a larger area of the output screen, and thus its brightness gain from minification is less than that in the center. • A fall-off in brightness at the periphery of an image is called vignetting.
  • 42. MULTIPLE-FIELD IMAGE INTENSIFIERS • Resolve the conflicts between image size and quality. • They can be operated in several modes, including a 4.5- in., a 6-in., or 9-in. mode. • Field size is changed by applying a simple electronic principle: • The higher the voltage on the electrostatic focusing lens, the more the electron beam is focused.
  • 43.
  • 44. • In the 9-in. mode, the electrostatic focusing voltage is decreased. • The electrons focus to a point, or cross, close to the output phosphor, and the final image is actually smaller than the phosphor. • In the 6-in. mode the electrostatic focusing voltage is increased, and the electrons focus farther away from the output phosphor. • After the electrons cross, they diverge, so the image on the output phosphor is larger than in the 9-in. mode.

Editor's Notes

  1. By a radiologist who had dark-adapted his eyes by wearing red goggles for 20 to 30 minutes prior to the examination.
  2. without distorting their geometric configuration
  3. Substrate by a process called vapour deposition
  4. Phosphor thicknesses have been reduced comparably from approximately 0.3 mm to 0.1mm
  5. (commonly a combination of antimony and cesium compounds).
  6. The anode has a positive potential of 25 to 35 k V relative to the photocathode,
  7. Because the electrons are greatly accelerated, they emit more light photons from the output screen than were originally present at the input screen.
  8. This can be done with only minimal loss of brightness.
  9. the resulting image would be degraded by the television chain.
  10. (often termed"partially silvered")
  11. defined as the luminous intensity so a conversion factor of 100 is about the same as brightness gain of 10,000
  12. Most x-ray image intensifiers have an input screen from 5 to 9 in. in diameter. output screen approximately 1 in. in diameter. intensifier has a gain of 81.
  13. in the same way that scattered x-ray photons produce fog and reduce contrast in a radiographic image
  14. the further .an electron is from the center of the intensifier, the more difficult it is to control.
  15. This is the cine image of a wire screen the wires curve out at the periphery; this effect is most noticeable at the corners.
  16. In summary, the center of the image intensifier screen has better resolution, a brighter image, and less geometric distortion
  17. The 9-in. mode is used when it is necessary to view large anatomic areas. When size is unimportant, the 4.5- or 6-in. mode is used because of better resultant image quality.
  18. In the 6-in. mode, the optical system "sees" only the central portion of the image, the part derived from the central 6 in. of the input phosphor