4. Grids are used to reject scatter.
Placed between the breast and image
receptor.
Moving grid with a grid ratio of 4:1 to
5:1 are typical .
Higher grids ratio provide greater x-
ray scatter removal but also a greater
dose.
Grids frequencies ranges from 30 to
50 lines/cm for moving grids and up
to 80 lines/cm for stationary grids.
5. Creates better images by reducing x-ray
scatter .
Automatic self –adjusting tilt paddle
ensures uniform compression of the
breast tissue .
Reduces procedures time by about 10 %
and enhance patient comport.
Copper as grid strip and air as
interspace with 3.8:1 grid ratio
6.
7. Compression is achieved with a low
attenuation lexan paddle attached
to compression device.
Parallel to the receptor surface
10 to 20 newton of force is typically
used
Improves spatial resolution
,contrast resolution and reduces
patient dose.
The principle drawback of
compression ; patient discomport.
8.
9. AEC employs phototimer to measures X-ray intensity
and quality.
Kept after the image receptor ;to minimize the object
to image distance -improves spatial resolution.
Automatically calculates the exposures time that is
required to produced optimum density on the film.
When the required optimum density is produced ,the
AEC terminate the exposure and reduces the dose.
10. Two types of AEC used ionization chamber and solid
state diode.
Allows assessment of breast composition and selection
of proper target/filter combination
For thick dense breasts are imaged better with Rh/Rh ;
thin fatty breast are imaged better with Mo/Mo.
11. Four types of images receptor have been used in
mammography.
1. Direct –exposure film
2. Xeroradiography
3. Screen-film
4. Digital detector
12. Direct –exposure film ; once used for small body parts
measuring less than 10 cm. Requires 10 to 100 time
more exposure .no longer used.
Xeroradiography; this techniques gives low contrast
images.
13. • Used single phosphor screen and single emulsion flim
with cubic grain emulsion.
• The emulsion surface of the film must face the screen
to prevent the resolution due to light diffusion in the
screen.
• The use of intensifying screen to increase the speed of
imaging system resulting in a low patient dose and
increase image contrast.
• Gadolinium oxysulphide (Gd2O2S:Tb)is used as
phosphor and resolution of approx 12 lp/mm
14.
15. No film or chemical processing is used
Thallium activated cesium iodide used as phosphor.
This phosphor can be grown as parallel needle like
light pipes for improved spatial resolution(5lp/mm)
.
16. Magnification techniques are used frequently in
mammography to produced images up to twice the
normal size.
Magnification mammography requires special
equipment such as micro focus tubes, adequate
compression , patient positioning devices.
The purpose is to investigate small ,suspicious lesions
or micro calcifications.
17. Magnification of 1.5x to
2.0 x is used
Increased effective
resolution of the image
receptor by the
magnificator.
Small focal spot
size(0.1mm)used
Reduction of scatter
18. High patients dose
Long exposure times
(small focal spot ,low
MA)
Patient motion and blur
19. FDA Dose limit
3 mGy (w/grid)
Mean glandular dose
Single view
4.5cm compressed breast
Average composition