TECHNICAL SPECIFICATIONS: Display-->Type : LED, Parameters : SpO2, PR, Pleth bar., Information : low power ::SPO2-->Measure Range : 35-99%, Resolution : ±1%, Accuracy : 75%-99%:±2, 50%-75%:±3, 0%-50%:unspecified::Pulse Rate-->Range : 30-254 BPM, Resolution : ±1bpm, Accuracy : ±2bpm::Battery--> Type : 2 AAA Alkaline batteries or rechargeable, batteries Power Consumption : less than 30Ma, Voltage : 2.6-3.6V::Environmental-->Operating Temperature : 0'C-50'C, Storage Temperature : -10'C - 60'C, Operating Humidity : 15%RH - 95%RH, Storage Humidity : 10%RH - 95%RH, It can be used by a variety of sports enthusiasts, including mountain climbers, hikers, skiers, bikers, and others interested in measuring oxygen saturation and pulse rate, it is not for medical use. This oximeter is for anyone who is interested in knowing their oxygen saturation and heart rate. Applications would be for use during exercise or daily activities, during air travel for the passenger or pilot, and for extreme sports such as mountain climbing. This oximeter is not a medical device and does not require a prescription from your medical professional.

OPERATOR’S MANUAL
VER1.0C615
Fingertip Pulse Oximeter
General Description
Oxygen Saturation is a percentage of Oxyhemoglobin (HbO2) capacity, compounded
with oxygen, by all combinative hemoglobin (Hb) capacity in blood. In other words, it
is consistency of Oxyhemoglobin in blood. It is a very important parameter for the
Respiratory Circulation System. Many respiratory diseases can result in oxygen
saturation being lowered in human blood. Additionally, the following factors can
reduce oxygen saturation: Automatic regulation of organ dysfunction caused by
Anesthesia, Intensive Postoperative Trauma, injuries caused by some medical
examinations. That situation might result in light-headedness, asthenia, and vomiting.
Therefore, it is very important to know the oxygen saturation of a patient so that
doctors can find problems in a timely manner.
The fingertip pulse Oximeter features small size, low power consumption, convenient
operation and portability. It is only necessary for a patient to put one of his fingers into
the fingertip photoelectric sensor for diagnosis, and a display screen will show oxygen
saturation. It has been proven in clinical experiments that it also features high
precision and repeatability.
Measurement Principle
Principle of the Oximeter is as follows: A mathematical formula is established making
use of Lambert Beer Law according to Spectrum Absorption Characteristics of
Reductive hemoglobin(RHb) and Oxyhemoglobin (HbO2) in glow and near-infrared
zones. Operation principle of the instrument: Photoelectric Oxyhemoglobin Inspection
Technology is adopted in accordance with Capacity Pulse Scanning and Recording
Technology, so that two beams of different wavelength of lights (660nm glow and
940nm near infrared light) can be focused onto a human nail tip through a clamping
finger-type sensor. A measured signal obtained by a photosensitive element, will be
shown on the Oximeter’s display through process in electronic circuits and
microprocessor shown on the Oximeter’s display through electronic circuits and a
microprocessor.
Diagram of Operation Principle
1. Red and Infrared-ray Emission Tube
2. Red and Infrared-ray Receipt Tube
Product Operation Scope
Fingertip Pulse Oximeter is a portable non-invasive, spot-check, oxygen saturation of
arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patients at home, and
hospital (including clinical use in internal medicine/surgery, anesthesia, intensive care
and etc). The device is not intended for continuously monitoring.
The PULE OXIMETER requires no routine calibration or maintenance other than
replacement of batteries.
Technical Specifications
1. Display: LED
2. SpO2:
Measurement range:70~99%
Accuracy: 70%~99%, ±3%; <70% no definition
3. Pulse Rate:
Measure range:30~235 bpm
Accuracy:30~100bpm ±2bpm; 101~235bpm, ±2%
Pulse Intensity: Bar graph Indicator
4. Power Requirements:
Two AAA alkaline Batteries
Power consumption: Less than 30mA
Low power indication:
Battery Life: Two AAA 1.5V, 800mAh alkaline batteries could be continuously
operated as long as 30 hours.
5. Dimension:
Length: 58mm
Width: 32mm
Height: 37mm
Weight: 33g (without batteries)
6. Environment Requirements:
Operation Temperature: 5~40℃
Storage Temperature: -20~55℃
Ambient Humidity: ≤80%, no condensation in operation;
≤93%, no condensation in storage
7.Measurement Performance in Low Perfusion Condition: The pulse wave is
available without failure when the simulation pulse wave amplitude is at 0.6% using
the test equipment (BIO-TEK INDEX Pulse Oximeter Tester).
8.Interference Resistance Capacity against Ambient Light: Device works
normally when mixed noise produced by BIO-TEK INDEX Pulse Oximeter Tester.
Precautions for use
1 Do not use the pulse oximeter in an MRI or CT environment
2 Do not use the pulse oximeter in situations where alarms are required. The
device has no alarms.
3 Explosion hazard: Do not use the pulse oximeter in an explosive atmosphere.
4 The pulse oximeter is intended only as an adjunct in patient assessment. It
must be used in conjunction with other methods of assessing clinical signs and
symptoms.
5 Check the pulse oximeter sensor application site frequently to determine the
positioning of the sensor and circulation and skin sensitivity of the patient.
6 Do not stretch the adhesive tape while applying the pulse oximeter sensor. This
may cause inaccurate readings or skin blisters.
7 Before use, carefully read the manual.
8 The pulse oximeter has no SpO2 alarms; it is not for continuous monitoring, as
indicated by the symbol.
9 Prolonged use or the patient’s condition may require changing the sensor site
periodically. Change sensor site and check skin integrity, circulatory status, and
correct alignment at least every 4 hours.
10 Inaccurate measurements may be caused by autoclaving, ethylene oxide
sterilizing, or immersing the sensors in liquid may cause inaccurate readings.
11 Significant levels of dysfunctional hemoglobins(such as carbonxy-hemoglobin
or methemoglobin).
12 Intravascular dyes such as indocyanine green or methylene blue
13 SpO2 measurements may be adversely affected in the presence of high
ambient light. Shield the sensor area (with a surgical towel, or direct sunlight,
for example) if necessary.
14 Excessive patient movement may cause accurate readings.
15 Venous pulsations may cause accurate readings.
16 Placement of a sensor on an extremity with a blood pressure cuff, arterial
catheter, or intravascular line may cause accurate readings.
17 The patient has hypotension, severe vasoconstriction, severe anemia, or
hypothermia.
18 The patient is in cardiac arrest or in shock.
19 Fingernail polish or false fingernails may cause inaccurate SpO2 readings.
Follow local ordinances and recycling instructions regarding disposal or
recycling of the device and device components, including batteries.
Product Properties
1. Easy to use.
2. Small in volume, light in weight.
3. Lower power consumption.
4. Low voltage warning will be indicated in visual window when battery voltage is low.
5. The product will automatically be powered off when no signal is in the product for
longer than 8 seconds.
Operation Instructions
1. Install two AAA batteries into battery cassette, and then close its cover.
2. Open the clamp as shown in the picture below.
3. Insert one finger into the flexible fingertip of the Oximeter fully.
4. Press the power switch once on the front panel.
5. The patient’s finger should remain still while the Oximeter is working.
6. Read corresponding data from display screen.
When you insert a finger into the Oximeter, your nail surface must be upward.
NOTE: Please use the medical alcohol to clean the rubber touching the finger inside of
Oximeter, and clean the test sensor using alcohol before and after operation. (The
rubber inside of the Oximeter is medical rubber, which has no toxin and no harm to the
skin of human being).
Product Accessories
1. One strap
2. Two batteries
3. One user’s manual
Strap Installation
1. Thread thinner end of the strap through the loop.
2. Thread thicker end of the strap through the threaded end before pulling it tightly.
Brief Description of Front Panel
The Pulse Bar graph displays the strength of the pulse rate signal. The height of the
bar graph shows the patient’s pulse strength.
Battery Installation
1. Put the two AAA batteries into battery cassette in correct polarities.
2. Press the button down along the arrow and push the battery cover shown as
below:
Notes:
Battery polarities must be correctly installed to avoid
damage to the device.
Please insert or remove batteries in right order, or
may cause damage to the device bracket.
Please remove batteries if the Oximeter will not be
used for a long time.
Maintenance and Storage
1. Replace batteries in time when low voltage lamp is lighted.
2. Clean surface of the fingertip Oximeter before it is used in diagnosis for patients.
3. Remove batteries inside the battery cassette if the Oximeter will not be operated
for a long time.
4. It is best to preserve the product in a place where ambient temperature –20~55
℃ and humidity is <93%, no condensation.
5. It is recommended that the product should be kept in a dry environment anytime.
A wet ambient might affect its lifetime and even might damage the product.
6. Please follow the law of the local government to deal with used batteries.
Calibrating the Oximeter
1. The functional tester cannot be used to assess the accuracy of the oximeter.
2. The test method used to establish the SpO2 accuracy is clinical testing. The
oximeter used to measure the arterial haemoglobin oxygen saturation levels and
these levels are to be compared to the levels determined from arterial blood
sampling with a CO-oximeter.
3. Index 2 that made by Bioteck company is a function tester. Set Tech to 1, R curve
to 2, and then users can use this particular calibration curve to measure the
oximeter.
Declaration
EMC of this product complies with IEC60601-1-2 standard.
The materials which the user can come into contact have no toxicity and no action on
tissues, comply with ISO10993-1, ISO10993-5 and ISO10993-10.
Guidance and manufacture’s declaration – electromagnetic
emissions-for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration – electromagnetic emission
The Pulse Oximeter is intended for use in the electromagnetic environment
specified below. The customer or the user of the Pulse Oximeter should assure
that it is used in such and environment.
Emission test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1 The Pulse Oximeter uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emission
CISPR 11
Class B The Pulse Oximeter is suitable for use in all
establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
Possible Problems and resolutions
Problems Possible reason Solution
SpO2 or PR can
not be shown
normally
1.Finger is not plugged correctly
2.Patient’s Oxyhemoglobin value is too
low to be measured.
1. Retry by plugging the finger.
2. Try some more times, If you can make
sure about no problem existing in the
product. Please go to a hospital for exact
diagnosis.
SpO2 or PR is
shown unstably
1.Finger might not be plugged deeply
enough.
2.Finger is trembling or patient’s body
is in movement status.
1.Retry by plugging the finger.
2.Try not to move.
The Oximeter
can not be
powered on
1.Power of batteries might be
inadequate or not be there at all.
2.Batteries might be installed
incorrectly.
3.The Oximeter might be damaged.
1.Please replace batteries
2.Please reinstall the batteries
3.Please contact with local customer
service centre.
Indication lamps
are suddenly off
1.The product is automatically
powered off when no signal is
detected longer than 8 seconds
2. Lower power
1.Normal
2.Replace the batteries
“Error3”or
“Error4” is
displayed on
screen.
1 Low power
2 Receiving tube being shielded or
damaged together with broken
connector.
3.Mechanical malfunction for
receive-emission tube
4 Amp circuit malfunction.
1 Change new batteries.
2 Please contact with local customer
service center.
3 Please contact with local customer
service center.
4 Please contact with local customer
service center.
Symbol Definitions
Symbol Definition
Type BF applied part
Attention, consult accompanying documents.
SpO2﹪ Oxygen saturation
Pulse rate (BPM)
Low power indication
NOT for continuous monitoring
SN Serial No.
Applicable models
SM-210 SM-220
Note: The illustrations used in this manual may differ slightly from the appearance of the
actual product.
Manufactured for:IRoams Network
Add:713 W. Duarte Rd #G810
Arcadia, CA 91007 USA
ALL RIGHTS RESERVED
Issue date: 2011-2-18

Recomendados

Santamedical SM-230 OLED Finger Pulse Oximeter von
Santamedical SM-230 OLED Finger Pulse OximeterSantamedical SM-230 OLED Finger Pulse Oximeter
Santamedical SM-230 OLED Finger Pulse OximeterGurin Products, LLC
118 views2 Folien
Sm 110 pulse oximeter von
Sm 110 pulse oximeterSm 110 pulse oximeter
Sm 110 pulse oximeterGurin Products, LLC
442 views4 Folien
Wearable Bio Sensor PPT von
Wearable Bio Sensor PPT Wearable Bio Sensor PPT
Wearable Bio Sensor PPT PRITAM SHARMA
485 views27 Folien
Innovative gadgets in anesthesia and medicine von
Innovative gadgets in anesthesia and medicine Innovative gadgets in anesthesia and medicine
Innovative gadgets in anesthesia and medicine ikramdr01
930 views118 Folien
Ijsea 2nd publication von
Ijsea 2nd publicationIjsea 2nd publication
Ijsea 2nd publicationSheela Arokia Mary
761 views4 Folien
Iaetsd mobile icu using android von
Iaetsd mobile icu using androidIaetsd mobile icu using android
Iaetsd mobile icu using androidIaetsd Iaetsd
366 views5 Folien

Más contenido relacionado

Was ist angesagt?

Saline level indicator von
Saline level indicatorSaline level indicator
Saline level indicatorAravind Gurram
1.1K views4 Folien
use of pulse oximetry von
use of pulse oximetryuse of pulse oximetry
use of pulse oximetryJeya Rajathurai
20.3K views15 Folien
Report von
ReportReport
ReportAjay Poshak
2.9K views32 Folien
Bio instrumentation von
Bio instrumentationBio instrumentation
Bio instrumentationMalik Zaeem
2.8K views51 Folien
Wearable Biosensors Presentation von
Wearable Biosensors PresentationWearable Biosensors Presentation
Wearable Biosensors PresentationShubham Rokade
3.1K views12 Folien
Development of wearable_bio von
Development of wearable_bioDevelopment of wearable_bio
Development of wearable_biosaikrishnapulipati
1.5K views24 Folien

Was ist angesagt?(17)

Bio instrumentation von Malik Zaeem
Bio instrumentationBio instrumentation
Bio instrumentation
Malik Zaeem2.8K views
Wearable Biosensors Presentation von Shubham Rokade
Wearable Biosensors PresentationWearable Biosensors Presentation
Wearable Biosensors Presentation
Shubham Rokade3.1K views
Final Project Slides Buchheit Harings Hussein von Scott Buchheit
Final Project Slides Buchheit Harings HusseinFinal Project Slides Buchheit Harings Hussein
Final Project Slides Buchheit Harings Hussein
Scott Buchheit88 views
Sensors and systems by shivam gupta von Shivam Gupta
Sensors and systems   by shivam guptaSensors and systems   by shivam gupta
Sensors and systems by shivam gupta
Shivam Gupta1.1K views
A Low Cost Wearable Medical Device for Vital Signs Monitoring in Low-Resource... von IJECEIAES
A Low Cost Wearable Medical Device for Vital Signs Monitoring in Low-Resource...A Low Cost Wearable Medical Device for Vital Signs Monitoring in Low-Resource...
A Low Cost Wearable Medical Device for Vital Signs Monitoring in Low-Resource...
IJECEIAES25 views
Newer affordable gadgets in anesthesia and medicine von dr tushar chokshi
Newer affordable gadgets in anesthesia and medicineNewer affordable gadgets in anesthesia and medicine
Newer affordable gadgets in anesthesia and medicine
dr tushar chokshi261 views
Pm-120 tens unit electronic pulse massager with 6 mode and rechargeable battery von Devendra Tiwari 7838896746
Pm-120 tens unit electronic pulse massager with 6 mode and rechargeable batteryPm-120 tens unit electronic pulse massager with 6 mode and rechargeable battery
Pm-120 tens unit electronic pulse massager with 6 mode and rechargeable battery
IRJET- Intravenous Fluid Level Indicator von IRJET Journal
IRJET- Intravenous Fluid Level IndicatorIRJET- Intravenous Fluid Level Indicator
IRJET- Intravenous Fluid Level Indicator
IRJET Journal247 views

Destacado

Marieb ch13b von
Marieb ch13bMarieb ch13b
Marieb ch13bElaine Briosos
2.1K views27 Folien
The Respiratory System von
The Respiratory SystemThe Respiratory System
The Respiratory SystemDang Thanh Tuan
5.8K views73 Folien
Regulation of Respiration von
Regulation of RespirationRegulation of Respiration
Regulation of RespirationShahd Al Ali
7.4K views14 Folien
Regulation of respiration von
Regulation of respirationRegulation of respiration
Regulation of respirationRASHA
3.8K views31 Folien
Chapter 22: Respiratory System (#4) von
Chapter 22: Respiratory System (#4)Chapter 22: Respiratory System (#4)
Chapter 22: Respiratory System (#4)Caroline Tokarski
5.6K views26 Folien
Chemical control of respiration von
Chemical control of respirationChemical control of respiration
Chemical control of respirationAnwar Siddiqui
42.6K views20 Folien

Destacado(12)

Regulation of Respiration von Shahd Al Ali
Regulation of RespirationRegulation of Respiration
Regulation of Respiration
Shahd Al Ali7.4K views
Regulation of respiration von RASHA
Regulation of respirationRegulation of respiration
Regulation of respiration
RASHA 3.8K views
Chemical control of respiration von Anwar Siddiqui
Chemical control of respirationChemical control of respiration
Chemical control of respiration
Anwar Siddiqui42.6K views
Transport of oxygen and carbon dioxide in blood (1) von Lubna Abu Alrub,DDS
Transport of oxygen and carbon dioxide in blood (1)Transport of oxygen and carbon dioxide in blood (1)
Transport of oxygen and carbon dioxide in blood (1)
Lubna Abu Alrub,DDS71.6K views
Unit 3 respiratory system and gas exchange von maswazi10
Unit 3 respiratory system and gas exchangeUnit 3 respiratory system and gas exchange
Unit 3 respiratory system and gas exchange
maswazi1064.6K views
REGULATION OF RESPIRATION von Dr Nilesh Kate
REGULATION OF RESPIRATIONREGULATION OF RESPIRATION
REGULATION OF RESPIRATION
Dr Nilesh Kate65.7K views
Adventitious breath sounds von jcp1962
Adventitious breath soundsAdventitious breath sounds
Adventitious breath sounds
jcp196221.8K views
Pulmonary auscultation von VASS Yukon
Pulmonary auscultationPulmonary auscultation
Pulmonary auscultation
VASS Yukon29.7K views

Similar a Santamedical SM-220 Finger Pulse Oximeter

Santamedical SM-240 OLED Finger Pulse Oximeter von
Santamedical SM-240 OLED Finger Pulse OximeterSantamedical SM-240 OLED Finger Pulse Oximeter
Santamedical SM-240 OLED Finger Pulse OximeterGurin Products, LLC
301 views2 Folien
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe... von
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...IRJET Journal
26 views4 Folien
Oximeter von
OximeterOximeter
OximeterAdil Khan
1K views5 Folien
IRJET- Development of Portable Device for Measurement of Blood Glucose, T... von
IRJET-  	  Development of Portable Device for Measurement of Blood Glucose, T...IRJET-  	  Development of Portable Device for Measurement of Blood Glucose, T...
IRJET- Development of Portable Device for Measurement of Blood Glucose, T...IRJET Journal
60 views5 Folien
Md90 user manual von
Md90 user manualMd90 user manual
Md90 user manualMeditech Equipment Co.,Ltd
495 views22 Folien
IRJET- Patient Monitoring System von
IRJET- Patient Monitoring SystemIRJET- Patient Monitoring System
IRJET- Patient Monitoring SystemIRJET Journal
14 views4 Folien

Similar a Santamedical SM-220 Finger Pulse Oximeter(20)

IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe... von IRJET Journal
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
IRJET Journal26 views
IRJET- Development of Portable Device for Measurement of Blood Glucose, T... von IRJET Journal
IRJET-  	  Development of Portable Device for Measurement of Blood Glucose, T...IRJET-  	  Development of Portable Device for Measurement of Blood Glucose, T...
IRJET- Development of Portable Device for Measurement of Blood Glucose, T...
IRJET Journal60 views
IRJET- Patient Monitoring System von IRJET Journal
IRJET- Patient Monitoring SystemIRJET- Patient Monitoring System
IRJET- Patient Monitoring System
IRJET Journal14 views
FINGUER TIP OXIMETER von ACN
FINGUER TIP OXIMETERFINGUER TIP OXIMETER
FINGUER TIP OXIMETER
ACN836 views
Abstract1 von susinu
Abstract1Abstract1
Abstract1
susinu3K views
IRJET - A Novel Health Monitoring System for Trance Patients von IRJET Journal
IRJET -  	  A Novel Health Monitoring System for Trance PatientsIRJET -  	  A Novel Health Monitoring System for Trance Patients
IRJET - A Novel Health Monitoring System for Trance Patients
IRJET Journal19 views
Jay 5 bw von Jeroxxx
Jay 5 bwJay 5 bw
Jay 5 bw
Jeroxxx2.8K views
Pulse oximeters- Application and Maintenance von shashi sinha
Pulse oximeters- Application and MaintenancePulse oximeters- Application and Maintenance
Pulse oximeters- Application and Maintenance
shashi sinha865 views
IRJET- The Multiphysiological Parameter Monitoring System von IRJET Journal
IRJET-  	  The Multiphysiological Parameter Monitoring SystemIRJET-  	  The Multiphysiological Parameter Monitoring System
IRJET- The Multiphysiological Parameter Monitoring System
IRJET Journal22 views
Bio-Medical Smart Saline Bottle von IRJET Journal
Bio-Medical Smart Saline BottleBio-Medical Smart Saline Bottle
Bio-Medical Smart Saline Bottle
IRJET Journal13 views

Más de Gurin Products, LLC

HDD docking station von
HDD docking stationHDD docking station
HDD docking stationGurin Products, LLC
286 views9 Folien
Scientific Concept SpO2 of Pulse Oximeter von
Scientific Concept SpO2 of Pulse OximeterScientific Concept SpO2 of Pulse Oximeter
Scientific Concept SpO2 of Pulse OximeterGurin Products, LLC
807 views14 Folien
Liztek USB 3.0 Universal Docking Station, Dual Monitor, for Laptop, Ultrabook... von
Liztek USB 3.0 Universal Docking Station, Dual Monitor, for Laptop, Ultrabook...Liztek USB 3.0 Universal Docking Station, Dual Monitor, for Laptop, Ultrabook...
Liztek USB 3.0 Universal Docking Station, Dual Monitor, for Laptop, Ultrabook...Gurin Products, LLC
180 views1 Folie
Liztek PSS-100 Portable Wireless Bluetooth Speaker von
Liztek PSS-100 Portable Wireless Bluetooth SpeakerLiztek PSS-100 Portable Wireless Bluetooth Speaker
Liztek PSS-100 Portable Wireless Bluetooth SpeakerGurin Products, LLC
155 views1 Folie
Liztek HDDT2BS Dual Bay USB 3.0 Super Speed to 2.5 and 3.5 inch SATA Hard Dis... von
Liztek HDDT2BS Dual Bay USB 3.0 Super Speed to 2.5 and 3.5 inch SATA Hard Dis...Liztek HDDT2BS Dual Bay USB 3.0 Super Speed to 2.5 and 3.5 inch SATA Hard Dis...
Liztek HDDT2BS Dual Bay USB 3.0 Super Speed to 2.5 and 3.5 inch SATA Hard Dis...Gurin Products, LLC
130 views1 Folie
Liztek USB 3.0 to HDMI Video Graphics Adapter Card von
Liztek USB 3.0 to HDMI Video Graphics Adapter CardLiztek USB 3.0 to HDMI Video Graphics Adapter Card
Liztek USB 3.0 to HDMI Video Graphics Adapter CardGurin Products, LLC
252 views46 Folien

Más de Gurin Products, LLC(20)

Liztek USB 3.0 Universal Docking Station, Dual Monitor, for Laptop, Ultrabook... von Gurin Products, LLC
Liztek USB 3.0 Universal Docking Station, Dual Monitor, for Laptop, Ultrabook...Liztek USB 3.0 Universal Docking Station, Dual Monitor, for Laptop, Ultrabook...
Liztek USB 3.0 Universal Docking Station, Dual Monitor, for Laptop, Ultrabook...
Liztek PSS-100 Portable Wireless Bluetooth Speaker von Gurin Products, LLC
Liztek PSS-100 Portable Wireless Bluetooth SpeakerLiztek PSS-100 Portable Wireless Bluetooth Speaker
Liztek PSS-100 Portable Wireless Bluetooth Speaker
Liztek HDDT2BS Dual Bay USB 3.0 Super Speed to 2.5 and 3.5 inch SATA Hard Dis... von Gurin Products, LLC
Liztek HDDT2BS Dual Bay USB 3.0 Super Speed to 2.5 and 3.5 inch SATA Hard Dis...Liztek HDDT2BS Dual Bay USB 3.0 Super Speed to 2.5 and 3.5 inch SATA Hard Dis...
Liztek HDDT2BS Dual Bay USB 3.0 Super Speed to 2.5 and 3.5 inch SATA Hard Dis...
Liztek USB 3.0 to HDMI Video Graphics Adapter Card von Gurin Products, LLC
Liztek USB 3.0 to HDMI Video Graphics Adapter CardLiztek USB 3.0 to HDMI Video Graphics Adapter Card
Liztek USB 3.0 to HDMI Video Graphics Adapter Card
SantaMedical Professional Clinical Large LCD Non-Contact Infrared Thermometer von Gurin Products, LLC
SantaMedical Professional Clinical Large LCD Non-Contact Infrared Thermometer SantaMedical Professional Clinical Large LCD Non-Contact Infrared Thermometer
SantaMedical Professional Clinical Large LCD Non-Contact Infrared Thermometer
2-in-1 Professional Clinical RY210 Large LCD Non-contact Infrared Thermometer... von Gurin Products, LLC
2-in-1 Professional Clinical RY210 Large LCD Non-contact Infrared Thermometer...2-in-1 Professional Clinical RY210 Large LCD Non-contact Infrared Thermometer...
2-in-1 Professional Clinical RY210 Large LCD Non-contact Infrared Thermometer...

Último

PATIENTCOUNSELLING in.pptx von
PATIENTCOUNSELLING  in.pptxPATIENTCOUNSELLING  in.pptx
PATIENTCOUNSELLING in.pptxskShashi1
43 views16 Folien
INTRODUCTION TO PHARMACEUTICAL VALIDATION SCOPE and MERITS OF VALIDATION.pptx von
INTRODUCTION TO PHARMACEUTICAL VALIDATION SCOPE and MERITS OF VALIDATION.pptxINTRODUCTION TO PHARMACEUTICAL VALIDATION SCOPE and MERITS OF VALIDATION.pptx
INTRODUCTION TO PHARMACEUTICAL VALIDATION SCOPE and MERITS OF VALIDATION.pptxABG
117 views40 Folien
Myocardial Infarction Nursing.pptx von
Myocardial Infarction Nursing.pptxMyocardial Infarction Nursing.pptx
Myocardial Infarction Nursing.pptxAsraf Hussain
17 views73 Folien
MENSTRUAL CYCLE.pdf von
MENSTRUAL CYCLE.pdfMENSTRUAL CYCLE.pdf
MENSTRUAL CYCLE.pdfRutvikunvar Raualji (PT)
21 views24 Folien
Pulmonary Embolism for Nurses.pptx von
Pulmonary Embolism for Nurses.pptxPulmonary Embolism for Nurses.pptx
Pulmonary Embolism for Nurses.pptxAsraf Hussain
35 views31 Folien
Correct handling of laboratory Rats ppt.pptx von
Correct handling of laboratory Rats ppt.pptxCorrect handling of laboratory Rats ppt.pptx
Correct handling of laboratory Rats ppt.pptxTusharChaudhary99
16 views12 Folien

Último(20)

PATIENTCOUNSELLING in.pptx von skShashi1
PATIENTCOUNSELLING  in.pptxPATIENTCOUNSELLING  in.pptx
PATIENTCOUNSELLING in.pptx
skShashi143 views
INTRODUCTION TO PHARMACEUTICAL VALIDATION SCOPE and MERITS OF VALIDATION.pptx von ABG
INTRODUCTION TO PHARMACEUTICAL VALIDATION SCOPE and MERITS OF VALIDATION.pptxINTRODUCTION TO PHARMACEUTICAL VALIDATION SCOPE and MERITS OF VALIDATION.pptx
INTRODUCTION TO PHARMACEUTICAL VALIDATION SCOPE and MERITS OF VALIDATION.pptx
ABG117 views
Myocardial Infarction Nursing.pptx von Asraf Hussain
Myocardial Infarction Nursing.pptxMyocardial Infarction Nursing.pptx
Myocardial Infarction Nursing.pptx
Asraf Hussain17 views
Pulmonary Embolism for Nurses.pptx von Asraf Hussain
Pulmonary Embolism for Nurses.pptxPulmonary Embolism for Nurses.pptx
Pulmonary Embolism for Nurses.pptx
Asraf Hussain35 views
Complications & Solutions in Laparoscopic Hernia Surgery.pptx von Varunraju9
Complications & Solutions in Laparoscopic Hernia Surgery.pptxComplications & Solutions in Laparoscopic Hernia Surgery.pptx
Complications & Solutions in Laparoscopic Hernia Surgery.pptx
Varunraju9134 views
communication and nurse patient relationship by Tamanya Samui.pdf von TamanyaSamui1
communication and nurse patient relationship by Tamanya Samui.pdfcommunication and nurse patient relationship by Tamanya Samui.pdf
communication and nurse patient relationship by Tamanya Samui.pdf
TamanyaSamui131 views
ICH AND WHO GUIDELINES FOR VALIDATION OF EQUIPMENTS.pptx von ABG
ICH AND WHO GUIDELINES FOR VALIDATION OF EQUIPMENTS.pptxICH AND WHO GUIDELINES FOR VALIDATION OF EQUIPMENTS.pptx
ICH AND WHO GUIDELINES FOR VALIDATION OF EQUIPMENTS.pptx
ABG64 views
VarSeq 2.5.0: VSClinical AMP Workflow from the User Perspective von Golden Helix
VarSeq 2.5.0: VSClinical AMP Workflow from the User PerspectiveVarSeq 2.5.0: VSClinical AMP Workflow from the User Perspective
VarSeq 2.5.0: VSClinical AMP Workflow from the User Perspective
Golden Helix100 views
Top Ayurvedic PCD Companies in India Riding the Wave of Wellness Trends von muskansbl01
Top Ayurvedic PCD Companies in India Riding the Wave of Wellness TrendsTop Ayurvedic PCD Companies in India Riding the Wave of Wellness Trends
Top Ayurvedic PCD Companies in India Riding the Wave of Wellness Trends
muskansbl0155 views

Santamedical SM-220 Finger Pulse Oximeter

  • 1. OPERATOR’S MANUAL VER1.0C615 Fingertip Pulse Oximeter General Description Oxygen Saturation is a percentage of Oxyhemoglobin (HbO2) capacity, compounded with oxygen, by all combinative hemoglobin (Hb) capacity in blood. In other words, it is consistency of Oxyhemoglobin in blood. It is a very important parameter for the Respiratory Circulation System. Many respiratory diseases can result in oxygen saturation being lowered in human blood. Additionally, the following factors can reduce oxygen saturation: Automatic regulation of organ dysfunction caused by Anesthesia, Intensive Postoperative Trauma, injuries caused by some medical examinations. That situation might result in light-headedness, asthenia, and vomiting. Therefore, it is very important to know the oxygen saturation of a patient so that doctors can find problems in a timely manner. The fingertip pulse Oximeter features small size, low power consumption, convenient operation and portability. It is only necessary for a patient to put one of his fingers into the fingertip photoelectric sensor for diagnosis, and a display screen will show oxygen saturation. It has been proven in clinical experiments that it also features high precision and repeatability. Measurement Principle Principle of the Oximeter is as follows: A mathematical formula is established making use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive hemoglobin(RHb) and Oxyhemoglobin (HbO2) in glow and near-infrared zones. Operation principle of the instrument: Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse Scanning and Recording Technology, so that two beams of different wavelength of lights (660nm glow and 940nm near infrared light) can be focused onto a human nail tip through a clamping finger-type sensor. A measured signal obtained by a photosensitive element, will be shown on the Oximeter’s display through process in electronic circuits and microprocessor shown on the Oximeter’s display through electronic circuits and a microprocessor. Diagram of Operation Principle 1. Red and Infrared-ray Emission Tube 2. Red and Infrared-ray Receipt Tube Product Operation Scope Fingertip Pulse Oximeter is a portable non-invasive, spot-check, oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patients at home, and hospital (including clinical use in internal medicine/surgery, anesthesia, intensive care and etc). The device is not intended for continuously monitoring. The PULE OXIMETER requires no routine calibration or maintenance other than replacement of batteries. Technical Specifications 1. Display: LED 2. SpO2: Measurement range:70~99% Accuracy: 70%~99%, ±3%; <70% no definition 3. Pulse Rate: Measure range:30~235 bpm Accuracy:30~100bpm ±2bpm; 101~235bpm, ±2% Pulse Intensity: Bar graph Indicator 4. Power Requirements: Two AAA alkaline Batteries Power consumption: Less than 30mA Low power indication: Battery Life: Two AAA 1.5V, 800mAh alkaline batteries could be continuously operated as long as 30 hours. 5. Dimension: Length: 58mm Width: 32mm Height: 37mm Weight: 33g (without batteries) 6. Environment Requirements: Operation Temperature: 5~40℃ Storage Temperature: -20~55℃ Ambient Humidity: ≤80%, no condensation in operation; ≤93%, no condensation in storage 7.Measurement Performance in Low Perfusion Condition: The pulse wave is available without failure when the simulation pulse wave amplitude is at 0.6% using the test equipment (BIO-TEK INDEX Pulse Oximeter Tester). 8.Interference Resistance Capacity against Ambient Light: Device works normally when mixed noise produced by BIO-TEK INDEX Pulse Oximeter Tester. Precautions for use 1 Do not use the pulse oximeter in an MRI or CT environment 2 Do not use the pulse oximeter in situations where alarms are required. The device has no alarms. 3 Explosion hazard: Do not use the pulse oximeter in an explosive atmosphere. 4 The pulse oximeter is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods of assessing clinical signs and symptoms. 5 Check the pulse oximeter sensor application site frequently to determine the positioning of the sensor and circulation and skin sensitivity of the patient. 6 Do not stretch the adhesive tape while applying the pulse oximeter sensor. This may cause inaccurate readings or skin blisters. 7 Before use, carefully read the manual. 8 The pulse oximeter has no SpO2 alarms; it is not for continuous monitoring, as indicated by the symbol. 9 Prolonged use or the patient’s condition may require changing the sensor site periodically. Change sensor site and check skin integrity, circulatory status, and correct alignment at least every 4 hours. 10 Inaccurate measurements may be caused by autoclaving, ethylene oxide sterilizing, or immersing the sensors in liquid may cause inaccurate readings. 11 Significant levels of dysfunctional hemoglobins(such as carbonxy-hemoglobin or methemoglobin). 12 Intravascular dyes such as indocyanine green or methylene blue 13 SpO2 measurements may be adversely affected in the presence of high ambient light. Shield the sensor area (with a surgical towel, or direct sunlight, for example) if necessary. 14 Excessive patient movement may cause accurate readings. 15 Venous pulsations may cause accurate readings. 16 Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line may cause accurate readings. 17 The patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia. 18 The patient is in cardiac arrest or in shock. 19 Fingernail polish or false fingernails may cause inaccurate SpO2 readings. Follow local ordinances and recycling instructions regarding disposal or recycling of the device and device components, including batteries. Product Properties 1. Easy to use. 2. Small in volume, light in weight. 3. Lower power consumption. 4. Low voltage warning will be indicated in visual window when battery voltage is low. 5. The product will automatically be powered off when no signal is in the product for longer than 8 seconds. Operation Instructions 1. Install two AAA batteries into battery cassette, and then close its cover. 2. Open the clamp as shown in the picture below. 3. Insert one finger into the flexible fingertip of the Oximeter fully. 4. Press the power switch once on the front panel. 5. The patient’s finger should remain still while the Oximeter is working. 6. Read corresponding data from display screen. When you insert a finger into the Oximeter, your nail surface must be upward. NOTE: Please use the medical alcohol to clean the rubber touching the finger inside of Oximeter, and clean the test sensor using alcohol before and after operation. (The rubber inside of the Oximeter is medical rubber, which has no toxin and no harm to the skin of human being). Product Accessories 1. One strap 2. Two batteries 3. One user’s manual Strap Installation 1. Thread thinner end of the strap through the loop. 2. Thread thicker end of the strap through the threaded end before pulling it tightly. Brief Description of Front Panel
  • 2. The Pulse Bar graph displays the strength of the pulse rate signal. The height of the bar graph shows the patient’s pulse strength. Battery Installation 1. Put the two AAA batteries into battery cassette in correct polarities. 2. Press the button down along the arrow and push the battery cover shown as below: Notes: Battery polarities must be correctly installed to avoid damage to the device. Please insert or remove batteries in right order, or may cause damage to the device bracket. Please remove batteries if the Oximeter will not be used for a long time. Maintenance and Storage 1. Replace batteries in time when low voltage lamp is lighted. 2. Clean surface of the fingertip Oximeter before it is used in diagnosis for patients. 3. Remove batteries inside the battery cassette if the Oximeter will not be operated for a long time. 4. It is best to preserve the product in a place where ambient temperature –20~55 ℃ and humidity is <93%, no condensation. 5. It is recommended that the product should be kept in a dry environment anytime. A wet ambient might affect its lifetime and even might damage the product. 6. Please follow the law of the local government to deal with used batteries. Calibrating the Oximeter 1. The functional tester cannot be used to assess the accuracy of the oximeter. 2. The test method used to establish the SpO2 accuracy is clinical testing. The oximeter used to measure the arterial haemoglobin oxygen saturation levels and these levels are to be compared to the levels determined from arterial blood sampling with a CO-oximeter. 3. Index 2 that made by Bioteck company is a function tester. Set Tech to 1, R curve to 2, and then users can use this particular calibration curve to measure the oximeter. Declaration EMC of this product complies with IEC60601-1-2 standard. The materials which the user can come into contact have no toxicity and no action on tissues, comply with ISO10993-1, ISO10993-5 and ISO10993-10. Guidance and manufacture’s declaration – electromagnetic emissions-for all EQUIPMENT and SYSTEMS Guidance and manufacture’s declaration – electromagnetic emission The Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the user of the Pulse Oximeter should assure that it is used in such and environment. Emission test Compliance Electromagnetic environment – guidance RF emissions CISPR 11 Group 1 The Pulse Oximeter uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. RF emission CISPR 11 Class B The Pulse Oximeter is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. Possible Problems and resolutions Problems Possible reason Solution SpO2 or PR can not be shown normally 1.Finger is not plugged correctly 2.Patient’s Oxyhemoglobin value is too low to be measured. 1. Retry by plugging the finger. 2. Try some more times, If you can make sure about no problem existing in the product. Please go to a hospital for exact diagnosis. SpO2 or PR is shown unstably 1.Finger might not be plugged deeply enough. 2.Finger is trembling or patient’s body is in movement status. 1.Retry by plugging the finger. 2.Try not to move. The Oximeter can not be powered on 1.Power of batteries might be inadequate or not be there at all. 2.Batteries might be installed incorrectly. 3.The Oximeter might be damaged. 1.Please replace batteries 2.Please reinstall the batteries 3.Please contact with local customer service centre. Indication lamps are suddenly off 1.The product is automatically powered off when no signal is detected longer than 8 seconds 2. Lower power 1.Normal 2.Replace the batteries “Error3”or “Error4” is displayed on screen. 1 Low power 2 Receiving tube being shielded or damaged together with broken connector. 3.Mechanical malfunction for receive-emission tube 4 Amp circuit malfunction. 1 Change new batteries. 2 Please contact with local customer service center. 3 Please contact with local customer service center. 4 Please contact with local customer service center. Symbol Definitions Symbol Definition Type BF applied part Attention, consult accompanying documents. SpO2﹪ Oxygen saturation Pulse rate (BPM) Low power indication NOT for continuous monitoring SN Serial No. Applicable models SM-210 SM-220 Note: The illustrations used in this manual may differ slightly from the appearance of the actual product. Manufactured for:IRoams Network Add:713 W. Duarte Rd #G810 Arcadia, CA 91007 USA ALL RIGHTS RESERVED Issue date: 2011-2-18