SELECTING AN ANALYTICAL METHOD
Presented by :
RAMYA SAVITHRI K
UP222505
Department of Chemistry
Introduction
Accuracy
Precision
Sensitivity
Selectivity
Robustness
Ruggedness
Scale of operation
Equipment, time and cost
Making the final choice
References
Analytical chemistry is a branch of chemistry which deals with the study of the seperation,
identification, qualitative and quantitative determination of the compositions of different
substances.
An analytical method is the application of a technique to a specific analyte in a specific matrix.
The best method is chosen based on the requirements of the analysis. Some of the criterias which
are considered for selecting the method are,
1. Accuracy
2. Precision
3. Sensitivity
4. Selectivity
5. Robustness
6. Ruggedness
7. Scale of operation
8. Equipment,Time and cost
1. ACCURACY
• Accuracy is how closely the result of an experiment agrees with the “true” or “expected”
value.
• It is defined as ‘A measure of agreement between the experimental value and
theoretical value’
• Errors are of two types,
a. Absolute error [ Ea]
Ea=Xi-Xt
b. Relative error [Er]
Er= Xi-Xt x 100
Xt
If the relative error is within 1%, the value is said to be highly accurate.
If the relative error is between 1-5%, the value is moderately accurate.
If the relative errror is more than 5%, the value is considered to be of low accuracy.
2. PRECISION
• When a sample is analysed several times, the individual results are rarely the
same. Instead the results are randomly scattered
• It is defined as ‘A measure of the variabilities of results obtained in
different trials of the same experiment.’
• The closer the agreement between the individual analysis, the more precise the
results.
• Precision doesn’t imply accuracy.
In determining the concentration of K+ in serum,
The results in (a) are more precise than those in (b), but (b) may have more
accurate values.
3. SENSITIVITY
• Sensitivity is the ability of a method to demonstrate that two samples have different
amount of analyte.
• It is defined as ‘The change in signal per unit change in the amount of analyte’.
• It is equivalent to the proportionality k
If nA is the number of moles of analyte, SA is the signal obtained by analyte,
and CA is the concentration of analyte, k is given by
𝐒𝐀
𝐧𝐀
= k or SA = knA , In total analysis method.
𝐒𝐀
𝐂𝐀
= k or SA =kCA , In concentration method.
• The method with greatest k value [Sensitivity] is best able to discriminate among smaller
amounts of analyte.
4. SELECTIVITY
• An analytical method is selective, if its signal is a function of only the amount of
analyte present in the sample.
• It is defined as ‘The measure of methods of freedom from interferences’
• The signal of sample, containing interferent is given by,
SSample = SA+ SI
where, SA & SI are signals of analyte and interferent respectively
or
SSample = kAnA+ kInI (In total analysis method)
SSample = kACA+ kICI (In concentration method)
where,kA & kI are sensitivities of analyte and interferent respectively.
nA & nI are no: of moles and CA & CI are concentrations of analyte and
interferent respectively.
• The selectivity of a method is given by selectivity coefficient KA,I,
KA,I =
𝐤𝐈
𝐤𝐀
where kI and kA are sensitivities of interferent and analyte respectively.
• The selectivity coefficient greater than +1 or lesser than -1 indicates that the
method is more selective for the interferent than for the analyte.
or kI =KA,I*kA
Substituting this equation in the above equations,
SSample = kA (nA+ KA,I*nI) (In total analysis method) ............(1)
SSample = kA (CA+ KA,I*CI) (In concentration method) ............(2)
• For a method to be selective, KA,I*nI should be smaller than nA in (1) and KA,I*CI
should be smaller than CA in (2).
5. ROBUSTNESS
• For a method to be useful it must provide reliable results. Unfortunately, methods are
subject to a variety of chemical & physical interferences that contribute uncertainity
to the analysis.
• When a method is relatively free from chemical interferences it can be applied to the
determination of analytes in a wide variety of sample matrices. Such methods are
considered robust.
6. RUGGEDNESS
• Random variations in experimental conditions also introduce uncertainity.
• If a method’s sensitivity is highly dependent on experimental conditions such as
temperature, acidity or relaxation time, then slight changes in those conditions may
lead to significantly different results.
• A rugged method is relatively insensitive to changes in experimental conditions
7. SCALE OF OPERATION
• Another way of narrowing the choice of methods is to consider the scale on which the
analysis must be conducted.
• Three limitations of particular importance are
i. The amount of sample available for the analysis.
ii. The concentration of analyte in the sample.
iii. The absolute amount of analyte needed to obtain a measurable signal.
The first and second limitations define the scale of operation.
The last limitation positions a method within the scale of operation.
8. EQUIPMENT, TIME AND COST
Analytical methods can be compared in terms of their need for equipment, the time
required to complete an analysis, and the cost per sample.
I. Equipment
Methods relying on insrumentation are equipment-intensive and may require significant
operator training.
II. Time
The time needed to complete an analysis for a single sample is often fairly similar from
method to method. This is however not completely true, because, much of this time is
spent preparing the solutions and equipments are in place, the number of samples that can
be analysed per hour differs substantially from method to method. This is a significant
factor for laboratories that handle a high volume of sample.
III. Cost
The cost of analysis is determined by many factors including the cost of
necessary equipments and reagents, the cost of hiring analysts and the number
of samples that can be processed per hour.
In general methods relying on insruments cost more per sample than other
methods.
MAKING THE FINAL CHOICE
• Selecting a specific method requires a careful balance among these design
criterias.
• The most important criteria among these that should be considered during the
selection of analytical method is accuracy.
For example,
In determining the concentration of lead in drinking water, Graphite Furnace
Atomic Absorption Spectroscopy is the commonly used method because, it meets
all those criterias.
ANALYTICAL CHEMISTRY by B.K. Sharma
https://www.brainkart.com
https://chem.libretexts.org
https://www.asdlib.org
RAMYA SAVITHRI K SELECTION OF AN ANALYTICAL METHOD.pptx

More Related Content

Similar to RAMYA SAVITHRI K SELECTION OF AN ANALYTICAL METHOD.pptx

Assay-Method validation-PPT _slide
Assay-Method validation-PPT _slideAssay-Method validation-PPT _slide
Assay-Method validation-PPT _slide
Bhanu Prakash N
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
Gaurav Kr
 
Importance of instumental analysis
Importance of instumental analysisImportance of instumental analysis
Importance of instumental analysis
Dr. Nilesh Badgujar
 
International conference on harmonisation validation of analytical procedures
International conference on harmonisation validation of analytical proceduresInternational conference on harmonisation validation of analytical procedures
International conference on harmonisation validation of analytical procedures
noushin javidvand
 
Selection and calibration of analytical method & calibration methods
Selection and calibration  of analytical method & calibration methodsSelection and calibration  of analytical method & calibration methods
Selection and calibration of analytical method & calibration methods
Tapeshwar Yadav
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
ceutics1315
 

Similar to RAMYA SAVITHRI K SELECTION OF AN ANALYTICAL METHOD.pptx (20)

ICH-Q2 AMV
ICH-Q2 AMVICH-Q2 AMV
ICH-Q2 AMV
 
Assay-Method validation-PPT _slide
Assay-Method validation-PPT _slideAssay-Method validation-PPT _slide
Assay-Method validation-PPT _slide
 
Spectroscopy.ppt
Spectroscopy.pptSpectroscopy.ppt
Spectroscopy.ppt
 
B.S 4- Class 1-Introduction to analytical chemistry
B.S 4- Class 1-Introduction to analytical chemistryB.S 4- Class 1-Introduction to analytical chemistry
B.S 4- Class 1-Introduction to analytical chemistry
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
Item 2. Verification and Validation of Analytical Methods
Item 2. Verification and Validation of Analytical MethodsItem 2. Verification and Validation of Analytical Methods
Item 2. Verification and Validation of Analytical Methods
 
Chapter 1
Chapter 1Chapter 1
Chapter 1
 
analytical method validation
analytical method validationanalytical method validation
analytical method validation
 
Analytical Method Validation
Analytical Method Validation Analytical Method Validation
Analytical Method Validation
 
Importance of instumental analysis
Importance of instumental analysisImportance of instumental analysis
Importance of instumental analysis
 
International conference on harmonisation validation of analytical procedures
International conference on harmonisation validation of analytical proceduresInternational conference on harmonisation validation of analytical procedures
International conference on harmonisation validation of analytical procedures
 
Ich guidelines on validation for analytical method/equipments
Ich guidelines on validation for analytical method/equipmentsIch guidelines on validation for analytical method/equipments
Ich guidelines on validation for analytical method/equipments
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
Selection and calibration of analytical method & calibration methods
Selection and calibration  of analytical method & calibration methodsSelection and calibration  of analytical method & calibration methods
Selection and calibration of analytical method & calibration methods
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
Validation of Analytical Procedures.pdf
Validation of Analytical Procedures.pdfValidation of Analytical Procedures.pdf
Validation of Analytical Procedures.pdf
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
Analytical Method Validation basics by Dr. A. Amsavel
Analytical Method Validation  basics by Dr. A. AmsavelAnalytical Method Validation  basics by Dr. A. Amsavel
Analytical Method Validation basics by Dr. A. Amsavel
 
Method validation
Method validationMethod validation
Method validation
 

Recently uploaded

Formation of low mass protostars and their circumstellar disks
Formation of low mass protostars and their circumstellar disksFormation of low mass protostars and their circumstellar disks
Formation of low mass protostars and their circumstellar disks
Sérgio Sacani
 
Chemical Tests; flame test, positive and negative ions test Edexcel Internati...
Chemical Tests; flame test, positive and negative ions test Edexcel Internati...Chemical Tests; flame test, positive and negative ions test Edexcel Internati...
Chemical Tests; flame test, positive and negative ions test Edexcel Internati...
ssuser79fe74
 
Bacterial Identification and Classifications
Bacterial Identification and ClassificationsBacterial Identification and Classifications
Bacterial Identification and Classifications
Areesha Ahmad
 
Pests of cotton_Sucking_Pests_Dr.UPR.pdf
Pests of cotton_Sucking_Pests_Dr.UPR.pdfPests of cotton_Sucking_Pests_Dr.UPR.pdf
Pests of cotton_Sucking_Pests_Dr.UPR.pdf
PirithiRaju
 
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 bAsymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
Sérgio Sacani
 
Module for Grade 9 for Asynchronous/Distance learning
Module for Grade 9 for Asynchronous/Distance learningModule for Grade 9 for Asynchronous/Distance learning
Module for Grade 9 for Asynchronous/Distance learning
levieagacer
 

Recently uploaded (20)

Formation of low mass protostars and their circumstellar disks
Formation of low mass protostars and their circumstellar disksFormation of low mass protostars and their circumstellar disks
Formation of low mass protostars and their circumstellar disks
 
Site Acceptance Test .
Site Acceptance Test                    .Site Acceptance Test                    .
Site Acceptance Test .
 
GBSN - Microbiology (Unit 2)
GBSN - Microbiology (Unit 2)GBSN - Microbiology (Unit 2)
GBSN - Microbiology (Unit 2)
 
Chemical Tests; flame test, positive and negative ions test Edexcel Internati...
Chemical Tests; flame test, positive and negative ions test Edexcel Internati...Chemical Tests; flame test, positive and negative ions test Edexcel Internati...
Chemical Tests; flame test, positive and negative ions test Edexcel Internati...
 
Bacterial Identification and Classifications
Bacterial Identification and ClassificationsBacterial Identification and Classifications
Bacterial Identification and Classifications
 
Pests of cotton_Sucking_Pests_Dr.UPR.pdf
Pests of cotton_Sucking_Pests_Dr.UPR.pdfPests of cotton_Sucking_Pests_Dr.UPR.pdf
Pests of cotton_Sucking_Pests_Dr.UPR.pdf
 
Pulmonary drug delivery system M.pharm -2nd sem P'ceutics
Pulmonary drug delivery system M.pharm -2nd sem P'ceuticsPulmonary drug delivery system M.pharm -2nd sem P'ceutics
Pulmonary drug delivery system M.pharm -2nd sem P'ceutics
 
PSYCHOSOCIAL NEEDS. in nursing II sem pptx
PSYCHOSOCIAL NEEDS. in nursing II sem pptxPSYCHOSOCIAL NEEDS. in nursing II sem pptx
PSYCHOSOCIAL NEEDS. in nursing II sem pptx
 
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 bAsymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
 
pumpkin fruit fly, water melon fruit fly, cucumber fruit fly
pumpkin fruit fly, water melon fruit fly, cucumber fruit flypumpkin fruit fly, water melon fruit fly, cucumber fruit fly
pumpkin fruit fly, water melon fruit fly, cucumber fruit fly
 
GBSN - Biochemistry (Unit 1)
GBSN - Biochemistry (Unit 1)GBSN - Biochemistry (Unit 1)
GBSN - Biochemistry (Unit 1)
 
FAIRSpectra - Enabling the FAIRification of Spectroscopy and Spectrometry
FAIRSpectra - Enabling the FAIRification of Spectroscopy and SpectrometryFAIRSpectra - Enabling the FAIRification of Spectroscopy and Spectrometry
FAIRSpectra - Enabling the FAIRification of Spectroscopy and Spectrometry
 
Proteomics: types, protein profiling steps etc.
Proteomics: types, protein profiling steps etc.Proteomics: types, protein profiling steps etc.
Proteomics: types, protein profiling steps etc.
 
High Profile 🔝 8250077686 📞 Call Girls Service in GTB Nagar🍑
High Profile 🔝 8250077686 📞 Call Girls Service in GTB Nagar🍑High Profile 🔝 8250077686 📞 Call Girls Service in GTB Nagar🍑
High Profile 🔝 8250077686 📞 Call Girls Service in GTB Nagar🍑
 
SAMASTIPUR CALL GIRL 7857803690 LOW PRICE ESCORT SERVICE
SAMASTIPUR CALL GIRL 7857803690  LOW PRICE  ESCORT SERVICESAMASTIPUR CALL GIRL 7857803690  LOW PRICE  ESCORT SERVICE
SAMASTIPUR CALL GIRL 7857803690 LOW PRICE ESCORT SERVICE
 
9999266834 Call Girls In Noida Sector 22 (Delhi) Call Girl Service
9999266834 Call Girls In Noida Sector 22 (Delhi) Call Girl Service9999266834 Call Girls In Noida Sector 22 (Delhi) Call Girl Service
9999266834 Call Girls In Noida Sector 22 (Delhi) Call Girl Service
 
Module for Grade 9 for Asynchronous/Distance learning
Module for Grade 9 for Asynchronous/Distance learningModule for Grade 9 for Asynchronous/Distance learning
Module for Grade 9 for Asynchronous/Distance learning
 
Locating and isolating a gene, FISH, GISH, Chromosome walking and jumping, te...
Locating and isolating a gene, FISH, GISH, Chromosome walking and jumping, te...Locating and isolating a gene, FISH, GISH, Chromosome walking and jumping, te...
Locating and isolating a gene, FISH, GISH, Chromosome walking and jumping, te...
 
GBSN - Microbiology (Unit 1)
GBSN - Microbiology (Unit 1)GBSN - Microbiology (Unit 1)
GBSN - Microbiology (Unit 1)
 
Kochi ❤CALL GIRL 84099*07087 ❤CALL GIRLS IN Kochi ESCORT SERVICE❤CALL GIRL
Kochi ❤CALL GIRL 84099*07087 ❤CALL GIRLS IN Kochi ESCORT SERVICE❤CALL GIRLKochi ❤CALL GIRL 84099*07087 ❤CALL GIRLS IN Kochi ESCORT SERVICE❤CALL GIRL
Kochi ❤CALL GIRL 84099*07087 ❤CALL GIRLS IN Kochi ESCORT SERVICE❤CALL GIRL
 

RAMYA SAVITHRI K SELECTION OF AN ANALYTICAL METHOD.pptx

  • 1. JSS MAHAVIDYAPEETHA JSS SCIENCE AND TECHNOLOGY UNIVERSITY SELECTING AN ANALYTICAL METHOD Presented by : RAMYA SAVITHRI K UP222505 Department of Chemistry
  • 3. Analytical chemistry is a branch of chemistry which deals with the study of the seperation, identification, qualitative and quantitative determination of the compositions of different substances. An analytical method is the application of a technique to a specific analyte in a specific matrix. The best method is chosen based on the requirements of the analysis. Some of the criterias which are considered for selecting the method are, 1. Accuracy 2. Precision 3. Sensitivity 4. Selectivity 5. Robustness 6. Ruggedness 7. Scale of operation 8. Equipment,Time and cost
  • 4. 1. ACCURACY • Accuracy is how closely the result of an experiment agrees with the “true” or “expected” value. • It is defined as ‘A measure of agreement between the experimental value and theoretical value’ • Errors are of two types, a. Absolute error [ Ea] Ea=Xi-Xt b. Relative error [Er] Er= Xi-Xt x 100 Xt If the relative error is within 1%, the value is said to be highly accurate. If the relative error is between 1-5%, the value is moderately accurate. If the relative errror is more than 5%, the value is considered to be of low accuracy.
  • 5. 2. PRECISION • When a sample is analysed several times, the individual results are rarely the same. Instead the results are randomly scattered • It is defined as ‘A measure of the variabilities of results obtained in different trials of the same experiment.’ • The closer the agreement between the individual analysis, the more precise the results. • Precision doesn’t imply accuracy. In determining the concentration of K+ in serum, The results in (a) are more precise than those in (b), but (b) may have more accurate values.
  • 6. 3. SENSITIVITY • Sensitivity is the ability of a method to demonstrate that two samples have different amount of analyte. • It is defined as ‘The change in signal per unit change in the amount of analyte’. • It is equivalent to the proportionality k If nA is the number of moles of analyte, SA is the signal obtained by analyte, and CA is the concentration of analyte, k is given by 𝐒𝐀 𝐧𝐀 = k or SA = knA , In total analysis method. 𝐒𝐀 𝐂𝐀 = k or SA =kCA , In concentration method. • The method with greatest k value [Sensitivity] is best able to discriminate among smaller amounts of analyte.
  • 7. 4. SELECTIVITY • An analytical method is selective, if its signal is a function of only the amount of analyte present in the sample. • It is defined as ‘The measure of methods of freedom from interferences’ • The signal of sample, containing interferent is given by, SSample = SA+ SI where, SA & SI are signals of analyte and interferent respectively or SSample = kAnA+ kInI (In total analysis method) SSample = kACA+ kICI (In concentration method) where,kA & kI are sensitivities of analyte and interferent respectively. nA & nI are no: of moles and CA & CI are concentrations of analyte and interferent respectively.
  • 8. • The selectivity of a method is given by selectivity coefficient KA,I, KA,I = 𝐤𝐈 𝐤𝐀 where kI and kA are sensitivities of interferent and analyte respectively. • The selectivity coefficient greater than +1 or lesser than -1 indicates that the method is more selective for the interferent than for the analyte. or kI =KA,I*kA Substituting this equation in the above equations, SSample = kA (nA+ KA,I*nI) (In total analysis method) ............(1) SSample = kA (CA+ KA,I*CI) (In concentration method) ............(2) • For a method to be selective, KA,I*nI should be smaller than nA in (1) and KA,I*CI should be smaller than CA in (2).
  • 9. 5. ROBUSTNESS • For a method to be useful it must provide reliable results. Unfortunately, methods are subject to a variety of chemical & physical interferences that contribute uncertainity to the analysis. • When a method is relatively free from chemical interferences it can be applied to the determination of analytes in a wide variety of sample matrices. Such methods are considered robust. 6. RUGGEDNESS • Random variations in experimental conditions also introduce uncertainity. • If a method’s sensitivity is highly dependent on experimental conditions such as temperature, acidity or relaxation time, then slight changes in those conditions may lead to significantly different results. • A rugged method is relatively insensitive to changes in experimental conditions
  • 10. 7. SCALE OF OPERATION • Another way of narrowing the choice of methods is to consider the scale on which the analysis must be conducted. • Three limitations of particular importance are i. The amount of sample available for the analysis. ii. The concentration of analyte in the sample. iii. The absolute amount of analyte needed to obtain a measurable signal. The first and second limitations define the scale of operation. The last limitation positions a method within the scale of operation.
  • 11.
  • 12. 8. EQUIPMENT, TIME AND COST Analytical methods can be compared in terms of their need for equipment, the time required to complete an analysis, and the cost per sample. I. Equipment Methods relying on insrumentation are equipment-intensive and may require significant operator training. II. Time The time needed to complete an analysis for a single sample is often fairly similar from method to method. This is however not completely true, because, much of this time is spent preparing the solutions and equipments are in place, the number of samples that can be analysed per hour differs substantially from method to method. This is a significant factor for laboratories that handle a high volume of sample.
  • 13. III. Cost The cost of analysis is determined by many factors including the cost of necessary equipments and reagents, the cost of hiring analysts and the number of samples that can be processed per hour. In general methods relying on insruments cost more per sample than other methods.
  • 14. MAKING THE FINAL CHOICE • Selecting a specific method requires a careful balance among these design criterias. • The most important criteria among these that should be considered during the selection of analytical method is accuracy. For example, In determining the concentration of lead in drinking water, Graphite Furnace Atomic Absorption Spectroscopy is the commonly used method because, it meets all those criterias.
  • 15. ANALYTICAL CHEMISTRY by B.K. Sharma https://www.brainkart.com https://chem.libretexts.org https://www.asdlib.org